You are on page 1of 9

Analytical

Methods
View Article Online
TECHNICAL NOTE View Journal

Direct estimation of ethanol as a negative peak


Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

from alcoholic beverages and fermentation broths


Cite this: DOI: 10.1039/c6ay01075j
by reversed phase-HPLC†
M. Nisha,a Mukund Shankar,b Nagasathiya Krishnan,c Lilly M. Saleena,a Mathur Rajeshb
and M. Vairamani*d

A relatively simple and straightforward technique utilizing reversed phase liquid chromatography with
a PDA-UV detector was used for the detection and quantification of ethanol. Samples of alcoholic
beverages and fermentation broths were injected without any prior chemical pre-treatment or structural
modification. The C18 column was equilibrated with water as a mobile phase having a high UV absorbing
compound, acetone, added in an optimized percentage. Ethanol when injected resulted in a negative
Received 12th April 2016
Accepted 24th April 2016
peak within 4 minutes of runtime, and the absolute area/height of the peak was used for quantification.
The method was validated for linearity with the detection ranging from 0.1–100% v/v. Intra-day and
DOI: 10.1039/c6ay01075j
inter-day precision experimental results were within acceptable limits. The accuracy and repeatability of
www.rsc.org/methods this method can reduce the dependency of using specific columns and detectors for the analysis of ethanol.

respectively.1 Use of methyl ethyl ketone as an UAC for ethanol


Introduction detection in alcoholic beverages is reported.7 It highlights the
Indirect ultraviolet (UV) determination of analytes of biological feasibility of using a C18 column in HPLC with methanol : water in
samples, solvents and impurities in pharmaceutical drugs a ratio of 30 : 70 as the mobile phase with methyl ethyl ketone
using Reversed Phase (RP)-HPLC is a relatively older tech- added in small quantities. The linear range of detection was
nique.1–4 The technique relies on detecting compounds with low limited to a maximum of 40% ethanol. Other polar UACs such as
UV absorbing properties and the detection is facilitated by acetone and methyl isopropyl ketone were also suggested. However
adding a minimal quantity of a high UV absorbing compound acetone's full potential as an UAC has not been reported yet.
(UAC) to the mobile phase which equilibrates itself throughout The present work explores the possibility of using acetone as
the column and sets a high UV absorbing background.4 When an UAC in detecting ethanol in two types of samples. The
a low UV absorbing sample is injected, a positive or a negative mobile phase used in the study is water with acetone added in
peak at the retention time (RT) corresponding to the sample an optimized percentage. Water in general has a very low UV
emerges. An UAC peak generally referred to as the system peak cut-off but acetone has a broader UV absorbance range of 190-
also appears with opposite polarity to that of the sample due to 330 nm. Ethanol absorbs UV only between 190 and 210 nm.8,9
disturbance in the UAC equilibration on the column.5 The Negative peaks in a chromatogram are generally observed when
polarities of the sample peak vary with respect to the system the UV absorbance drops below the calibrated mobile phase
peak based on its relative charge and relative RT in comparison baseline.10 Thus in the present study, UV for detection is
to the eluting UAC.6 deliberately chosen with a wavelength above the cut-off of
UACs such as 1-phenthyl-2-picolinium and nicotinamide are ethanol and in the absorbing range of acetone. When acetone in
used for the analysis of carboxylic acids and aliphatic alcohols, water as the mobile phase is equilibrated on the column, a high
UV absorbing background is set. Ethanol absorbs less UV
a
compared to the high UV absorbing background which will
Department of Bioinformatics, School of Bioengineering, SRM University,
Kattankulathur, Tamil Nadu, India
reverse its peak polarity to negative. No additional UAC or
b
Department of Chemical Engineering, School of Bioengineering, SRM University, buffers have been used in ethanol quantication because that
Kattankulathur, Tamil Nadu, India will lead to a possible increase in the number of retained system
c
Department of Biotechnology, School of Bioengineering, SRM University, peaks and thereby reduce the sensitivity of the detection.11
Kattankulathur, Tamil Nadu, India
d
School of Bioengineering, SRM University, Kattankulathur-603203, Kancheepuram
District, Tamil Nadu, India. E-mail: dean@biotech@ktr.srmuniv.ac.in; Tel: +91–44– Experimental
27417812
† Electronic supplementary information (ESI) available: Standard methanol This article does not contain any studies with human partici-
quantication and analysis. See DOI: 10.1039/c6ay01075j pants or animals performed by any of the authors.

This journal is © The Royal Society of Chemistry 2016 Anal. Methods


View Article Online

Analytical Methods Technical Note

Chemicals Peak conrmation


Acetone and ethanol of HPLC grade and ltered Milli-Q water Since the study involved only three chemical entities, in every
for dilutions were used throughout the study. run, the peak patterns detected will represent acetone, water or
ethanol singly or in combination. The method is acceptable if
there are no interfering peaks in the retention time of the
Instrumentation
analyte peak. Pure water, ethanol and acetone samples were
HPLC. All the preliminary feasibility and parameter opti- injected individually into the HPLC system in separate runs
mizations were performed with an Agilent HPLC 1260 (mobile phase: 4% acetone in water) and their retention times
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

INFINITY equipped with an Agilent 1260ALS autosampler. were conrmed. For further validation of acetone and ethanol,
Detection of the samples was carried out through a 13 mL ow the peaks emerging at the RT 3.14  0.13 min and 3.49  0.17
cell with a photodiode array detector (Agilent 1260DADVL). min were collected and the samples were subjected to GC-MS.
The detector peak width and the response time were set as >0.1 Water was not conrmed because the mass spectrometer was
min (2.0 s response time) (2.5 Hz). A reversed phase Zorbax scanned only from 40 to 100 amu.
Eclipse Plus C18 column with an ID, length and pore size of 4.6
mm, 250 mm and 5 mm respectively was used. All the outputs
were processed through Agilent Chemstation-Open lab so- Method validation
ware. Peak areas in arbitrary units and peak heights in milli Data analysis. All statistical analyses were independently
absorbance units (mAU) were considered for quantitative performed for both the methods (methods 1 and 2) using
analyses. Microso Excel 2010 and Minitab Release-14.
Parameters standardized in the Agilent LC system were then Linearity. Linearity between the concentration and its cor-
subsequently applied to a Shimadzu UFLC system for all responding response (peak area) values arising from triplicate
precision, accuracy, sensitivity, standard calibration and test injections for each ethanol concentration was analysed through
sample quantication experiments. The temperature controlled the least-square method. The analysis of variance (ANOVA) test
autosampler (SIL 20AC HT) was maintained at 25  C and was performed to establish the lack of t and statistical
separation was carried out in the same aforementioned C18 signicance of the regression equation.12–14
column. The data were analysed with Lab Solution soware Precision, accuracy and recovery. The closeness between the
version 5.31 (CBM20A). The total runtime was 6 min with peaks estimated value and its corresponding actual observed mean
detected with a UV detector (SPD20A). values was calculated. Precision with repeatability (5 consecu-
Gas chromatography-mass spectrometry. Analysis was per- tive hours) and interday variations (5 consecutive days) was
formed in an Agilent 7890B GC equipped with an autosampler analysed by calculating the relative standard deviations (%
and a HP-5MS capillary column with an ID, length and lm of RSDs). Accuracy and precision were checked in intermittent
250 mm, 30 m and 0.25 mm respectively. Helium was used as the concentrations in the linear range (three injections per
carrier gas with 1 mL min1 column ow rate. 1 mL of the concentration) for the instrument. The recovery percentage for
collected sample was injected at a split ratio of 1 : 10 and the the mean observed values to the known standards was subse-
inlet temperature was maintained at 250  C with the septum quently calculated.13,15,16
purged at 3 mL min1. The column temperature was held at 40 Limits of detection and quantication. The minimum

C for 5 min and ramped further to 100  C at 10  C min1. The detectable range was calculated to be 3.3v/a and the minimum
mass spectrometer (Agilent 5977A MSD) was scanned between quantiable limit was calculated to be 10v/a where ‘v’ is the
40 and 100 amu at a rate of 1.562 u s1 [N ¼ 2] with a solvent standard deviation and ‘a’ is the slope of the calibration plot as
delay of 0.5 min. The total runtime was 5 min. per ICH guidelines.13,15,17,18

Method optimization Test sample analysis


The PDA detector was scanned across wavelengths ranging from Alcoholic beverages. The need for sample preparation may
190 to 400 nm with 235 nm used for all concerned experiments. purely depend on the nature of the analyte.
Water was used as a mobile phase with acetone added in an For commercial alcoholic beverages such as whiskey and
optimized percentage. Acetone percentages were varied from 1 vodka no sample preparation is suggested. In method 1, 1 mL of
to 10% v/v from which 4% (v/v) acetone in water was used in all clear (vodka) and coloured (whiskey) alcoholic beverages were
the experiments. The standard ow rate was set at 1 mL min1. directly injected, quantied and compared with the claimed
Standard ethanol solutions were prepared in water (0.1–100% v/v) values as per the label contents.
for all analyses and validation experiments. Injection volumes Bioethanol from fermentative broth. Clostridium thermo-
were varied from 1 to 20 mL and optimized for different ethanol cellum DSM 1313 was cultivated in standard DSM media with
concentration ranges based on the sensitivity. 1% cellulose and anaerobically fermented for 120 hours at 58
Method 1. An injection volume of 10 mL for 0.1–5% v/v  1  C.19 The nal hour broth samples were analysed for bio-
ethanol concentrations was considered. ethanol. Fermentative broths are known to contain UV
Method 2. An injection volume of 1 mL for >5% v/v ethanol absorbing compounds such as proteins. To avoid any such
concentrations was considered. interference, 50 mL of centrifuged samples were distilled

Anal. Methods This journal is © The Royal Society of Chemistry 2016


View Article Online

Technical Note Analytical Methods

until a constant weight of 10 g was collected. 10 mL of the the peak occurrence, peak separation, and steadiness of the
distillate was immediately subjected to analysis in method 2. baseline, 235 nm was selected from the UV scan range. The
Standards were also prepared as per the above procedure. The selected UV wavelength lies within the UV absorbing range of
bioethanol from the fermented broth sample was also ana- acetone and above the cut-off range of ethanol.8,9 Fig. 4 shows
lysed with a GC-FID (Chemito GC 7610 system, 10% SE30 the formation of a negative peak for the analyte (ethanol)
column)20 to compare the values produced from the current during the detection. The quenching effect is due to drop in
method. the mAU value in comparison to the higher UV absorbing
mobile phase (4% v/v acetone). An ideal chemical component
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

Results and discussion such as water having low UV absorbing properties, high
polarity and miscibility with solvents such as acetone and
Peak conrmation
ethanol is a suitable candidate for a mobile phase.21 Its inter-
Under the optimized method conditions, a negative peak cor- action with the hydrophobic C18 column is at the minimum
responding to ethanol was observed at a retention time (RT) of and provides a background with least noise. It also allows
3.49  0.17 min as conrmed by GC-MS. A co-eluting peak lower polar compounds (acetone and ethanol) to partially
always preceded this negative analyte peak. One of the co- interact with the column components and allows separation
eluting peaks corresponded to a negative peak of water at 3.13  and detection.22 The optimization of acetone as an added UAC
0.02 min. This was ascertained by injecting pure water into the was performed for better negative peak emergence, peak
optimized aqueous mobile phase containing 4% acetone separation with constant RT and a steady baseline. Lower
(Fig. 1). The other co-eluting peak corresponded to a positive acetone concentrations (1–3%) gave poorly resolved broadened
peak of acetone at a RT of 3.14  0.13 min as conrmed by GC- peaks. Higher acetone concentrations (5–10%) depicted
MS and is the system peak. However, it is noteworthy that all the a trend of decreased sample RTs within 2 min of runtime. No
three peaks never emerged together for any of the ethanol sample peaks were detected for acetone concentrations >10%.
concentrations. Two kinds of peak pairs were always noted The mobile phase was thus optimized at 4% (v/v) acetone in
based on the ethanol concentration (Fig. 2 and 3). water. Reports suggest that the sensitivity of indirect photo-
(i) A negative peak pair (3.13  0.02 min for water, 3.49  metric estimation depends entirely on the added UAC, which
0.17 min for ethanol) was observed for <3.0% (v/v) ethanol in turn relies on charge, ion pairing mechanisms, pH, pro-
concentrations. tolysis and other buffering agents in the mobile phase.2,6,11
(ii) A pair of positive and negative peaks (3.14  0.13 min for However, the present work essentially detects the negative
acetone, 3.49  0.17 min for ethanol respectively) was observed peak of ethanol directly thus enabling linearity from 0.1 to
for >3.5% (v/v) ethanol. 100% (v/v). During optimization, injection volumes (>5 mL) for
The ethanol peak was observed for all concentrations (0.1– ethanol concentrations above 25% resulted in either
100% v/v) which inferred the non-dependence of ethanol on the a deformed system peak or merged system and sample peaks.
system peak. Whereas, at lower injection volumes (<5 mL) samples with
ethanol concentrations below 0.5% remained undetected. To
Method optimization maintain the linearity two injection volumes were thus
The developed method is an attempt to utilize a polar solvent considered 10 mL for 0.1–5% (method 1) and 1 mL for
to elute a polar sample in a reversed phase column. Based on concentrations above 5% (v/v) (method 2).

Fig. 1 Relative absorbance of water is low compared to that of the 4% acetone mobile phase. A single negative peak detected when pure water
was injected.

This journal is © The Royal Society of Chemistry 2016 Anal. Methods


View Article Online

Analytical Methods Technical Note


Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

Fig. 2 Chromatograms of standard ethanol (0.1–5% v/v). The injection volume is 10 mL.

Fig. 3 Chromatograms of standard ethanol (10–100% v/v).The injection volume is 1 mL.

Anal. Methods This journal is © The Royal Society of Chemistry 2016


View Article Online

Technical Note Analytical Methods

column, which may migrate through the column pores as an


isolated pocket resulting in an early emergence.23 Since the
standard ethanol samples were prepared in an aqueous solution,
the presence of this negative peak at 3.13  0.02 min is very
evident. The height and area of the water peak are inversely
related to the increasing concentrations of ethanol. However, this
tendency is observed only in <3.0% (v/v) ethanol.
Acetone peak. The injection of pure acetone into the aqueous
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

mobile phase containing 4% acetone resulted in a single posi-


tive peak at 3.14  0.13 min (Fig. 5). The positive peak indicates
a higher absorbance of absolute acetone compared to that of 4%
acetone as the mobile phase (Fig. 1). The area of the system

Table 1 Data analysis of method 1a

Method 1
Fig. 4 The UV absorbance (A235) value of 4% acetone in water solution
was considered as a reference and values of absolute acetone, ethanol Injection volume 10 mL
and water were read. Absolute acetone had the highest absorbance Linear range (% v/v) 0.1–5
whereas ethanol and water gave negative absorbance relative to the Replicates 3
4% acetone solution. It infers that any ethanol solution will have lower Regression equationb Y ¼ 29 843X  177
absorbance relative to the 4% acetone and vice versa for any acetone Correlation coefficient R2 0.996
solution above 4%. This justifies the negative peaks for ethanol and
water and positive peaks for acetone. Source of variation

Residual
Regression error Lack of t Pure error
Peak occurrences
Degrees of 1 19 5 14
Water peak. It is noteworthy that a negative peak at 3.13  freedom
0.02 min appears when pure water was injected into an optimized Sum of squares 2.01  1010 7.76  107 2.77  107 4.98  107
aqueous mobile phase (Fig. 1). A similar observation was re- Mean sum of 2.01  1010 4.08  106 5.55  106 3.56  106
ported by Jan Sbrek et al. (2005), where they injected pure water squares
F value 4917.55 1.56
into a RP-HPLC system containing an aqueous mobile phase with
acetate buffer, copper acetate and alkyl sulfonates.6 The negative p 0.235
polarity of the water peak is due to the lower UV absorbance of F critical 2.31 (a ¼ 0.1)
water than that of the 4% acetone mobile phase (Fig. 4). There is No evidence of lack of t
no corresponding system peak for water since the aqueous a
ANOVA results for linearity assessment with weighted regressions. b Y:
solvent is not adsorbed onto the column. The high polarity of negative peak area, X: ethanol concentration (% v/v).
water resists itself from interacting with the hydrophobic

Fig. 5 Relative absorbance of absolute acetone is high compared to that of the 4% acetone mobile phase. A single positive peak detected when
absolute acetone was injected.

This journal is © The Royal Society of Chemistry 2016 Anal. Methods


View Article Online

Analytical Methods Technical Note

peak (acetone) was also proportional to the increasing ethanol propose that ethanol does an equimolar replacement of acetone
concentrations and was detected only at >3.5% v/v concentra- from the column. This justies a proportional change in the
tions of ethanol. Since ethanol and acetone are polar solvents, system peak intensity and also its prior emergence to the
the incoming ethanol molecules may replace the bound acetone negative ethanol peak. The variation observed in the system
molecules in the column. This may be due to various contrib- peak intensity was due to the concentration of the desorbed
uting factors such as individual hydrophobicity and dipole– UAC at the ow cell.
dipole interactions among them.24 As mentioned by Arvidsson Ethanol peak. The negative peak at 3.49  0.17 min that
et al.11 any disturbance in the equilibrium will cause a change in appeared for 100% ethanol injection was conrmed to be due to
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

the concentration of the retained mobile phase components ethanol by GC-MS. The late emergence of the negative ethanol
(acetone) resulting in their migration as separate zones peak can be attributed to a new equilibrium state which is
throughout the column. Since acetone and ethanol both have achieved soon aer disturbance of the previous equilibrium by
polar and non-polar ends with similar polarity indices, we desorption of the mobile phase component (acetone) by
ethanol. The process of achieving this state may occur in several
different ways,6 out of which the process of solvation and
Table 2 Data analysis of method 2a desorption of the analyte molecule by the incoming mobile
phase component is relevant here. The as-obtained peak
Method 2 polarities are in accordance with the published report where the
Injection volume 1 mL sample peak is negative when it emerges aer the system peak
Linear range (% v/v) 10–100 and when the sample is uncharged or has the same charge as
Replicates 3 the UAC. However, the ion pairing mechanism as mentioned in
Regression equationb Y ¼ 32 516X + 19 629 the article11 may not be relevant to our study, since the analyte
Correlation coefficient R2 1
and the UAC both essentially are polar solvents.
Source of variation
Method validation
Residual
Regression error Lack of t Pure error Both the UV response (negative mAU) and the peak area were
proportional to the increasing ethanol concentrations (Fig. 2
Degrees of 1 13 3 10 and 3). The regression coefficients of the linear equation were
freedom
determined and are mentioned in Tables 1 and 2. Based on the
Sum of squares 1.69  1013 7.32  109 1.65  109 5.67  108
Mean sum of 1.69  1013 5.63  108 5.51  108 5.67  108 residual plot of method 1 weighted regression was considered.
squares The residual plots for method 2 showed a homoscedastic spread
F value 30 006 0.97 and no weighted analysis was performed. A linear relationship
between the analyte concentration and its corresponding
p 0.444
response for both the methods was established with no
F critical 2.73 (a ¼ 0.1)
No evidence of lack of t evidence of lack of t with p $ 0.1 from the analysis of variance
a b
(Tables 1 and 2).
ANOVA results for linearity assessment. Y: negative peak area, X:
The limit of detection was calculated to be 0.062% and the
ethanol concentration (% v/v).
limit of quantication was 0.19%. However, 0.1% (v/v) was
observed to lie within the linear quantiable range with no

Table 3 Validation results for methods 1 and 2. Calibration curve for method 1 with weighted linear regression

Inter-day variationsa Intra-day variationsb


Ethanol
Standard concentration Mean Standard
Mean (area a.u) deviation % RSD % accuracy % recovery (% v/v) (area a.u) deviation % RSD % accuracy % recovery

Method 1
2981.00 140.68 4.72 96.30 103.84 0.10 2964.80 184.06 6.21 95.28 104.96
15 133.40 190.85 1.26 97.46 102.61 0.50 15 083.52 103.68 0.69 97.76 102.29
30 495.00 331.31 1.09 96.19 103.96 1.00 30 255.40 692.21 2.29 96.09 104.06
152 239.40 1462.39 0.96 97.90 102.14 5.00 154 254.60 2008.61 1.30 96.51 103.61

Method 2
345 668.42 3559.12 1.03 99.88 99.88 10.00 344 385.22 2137.40 0.62 99.73 100.27
823 402.60 3277.72 0.40 98.84 98.85 25.00 823 199.72 5757.19 0.70 98.86 98.88
1 669 659.60 10 621.72 0.64 99.09 100.92 50.00 1 660 322.40 11 271.35 0.68 98.53 100.46
3 236 016.80 22 298.87 0.69 99.29 99.30 100.00 3 248 955.20 26 180.64 0.81 99.25 99.25
a
Mean of 5 intermittent injections (each triplicate) on a daily basis. b Mean of 5 intermittent injections (each triplicate) on an hourly basis.

Anal. Methods This journal is © The Royal Society of Chemistry 2016


View Article Online

Technical Note Analytical Methods

evidence of lack of t. The usage of lower measurement viz. For precision analysis the use of a non-thermostated auto-
0.1% may purely depend on an agreement between the labo- sampler introduced errors in the analysis (data not shown).
ratory, end user's data and the intended purpose.13 Since the study involved volatile solvents such as acetone and
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

Fig. 6 Chromatograms of test samples (A) whiskey, (B) vodka, and (C) bioethanol fermentation broth.

This journal is © The Royal Society of Chemistry 2016 Anal. Methods


View Article Online

Analytical Methods Technical Note

ethanol the use of a temperature controlled autosampler is approach is simple, user-friendly and does not require any
crucial. Repeatability was thus achieved with a temperature chemical modication or the use of any secondary UACs or
controlled autosampler; with the % RSDs of intra-day and inter- costly reagents. A HPLC instrument with a thermostated auto-
day precision experiments falling within the acceptable limits. sampler is suggested for accurate and precise results. The
The experiment also gives information on the errors which method is applicable for the detection of ethanol in fermenta-
seem to decrease proportionately with increasing ethanol tive samples and alcoholic beverages and its avenue can be
concentrations. % accuracy and % recovery mentioned in Table widened by analysing its feasibility in medical and pharma-
3 also take into account the systematic errors. ceutical samples. The feasibility of methanol quantication was
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

also studied with one of the representative peaks linear across


Test sample analysis all methanol concentrations. The scope of this study can be
extended to higher alcohols as well.
Injection of commercial alcoholic beverages, both clear and
coloured, gave clean chromatograms which are benecial for
easy detection and quantication. The percentage of ethanol
content detected in the two commercial alcoholic beverages
Conflict of interest
whiskey and vodka was 43.03  0.25% (v/v) and 43.01  0.10% The authors declare that they have no conict of interest.
(v/v) respectively. The labelled values for both the samples are
42.8% (v/v) of ethanol. Samples derived from the anaerobic
fermentation broth of Clostridium thermocellum were checked Acknowledgements
for the presence of bioethanol at the end of a 120 hour batch
and was found to be 0.114  0.01% (v/v). The result was Support by the Department of Science and Technology (DST)
comparable to the mean value as obtained with the GC-FID under Women Scientist Scheme A (SR/WOS-A/LS-196/2011(G)),
(0.117% v/v). The results are also in agreement with the pub- New Delhi for author N. M. is gratefully acknowledged. We also
lished data.19 When the fermentative broth was centrifuged, extend our acknowledgement to SRM University for providing
ltered and injected directly, the presence of certain early instruments and facilities.
eluting positive peaks was observed but did not interfere with
the concerned peaks. However, this will purely depend on the
nature of the fermentative medium composition. It is suggested References
to perform a simple distillation as mentioned in the “Experi-
1 A. Weston and P. R. Brown, High performance liquid
mental” section to avoid the interference of all the non-essential
chromatography and capillary electrophoresis: principles and
positive peaks (Fig. 6).
practices [Internet], Academic Press, 1997, [cited 2015 Sep
8], p. 280, available from, https://books.google.com/books?
Feasibility of methanol quantication and analysis
id¼CQW9rPV1UfwC&pgis¼1.
The developed method also could be extended to the analysis of 2 X. Shen and S. A. Tomellini, Indirect photometric and
methanol in ethanol. The presence of methanol in spurious uorometric detection in high-performance liquid
non-commercial liquors is a global concern causing serious chromatography: a tutorial review, Crit. Rev. Anal. Chem.,
health issues in consumers.25 Trials for the detection of meth- 2007, 37, 107–126.
anol separately and in combination with ethanol have been 3 Francesc A. Esteve-Turillas, Wolfgang Bicker,
attempted. Standard methanol in water (0.1–100% v/v) was Michael Lämmerhofer, Thomas Keller and
injected under the same conditions followed for the ethanol Wolfgang Lindner, Determination of ethyl sulfate –
quantication. A positive peak at 2.77  0.01 min varied linearly a marker for recent ethanol consumption – in human
for all methanol concentrations singly or in combination with urine by CE with indirect UV detection, Electrophoresis,
ethanol. Marginal shis in the RTs were observed even for the 2006, 27, 4763–4771.
characteristic peaks of ethanol (ESI†). Further investigations 4 P. Herné, M. Renson and J. Crommen, Indirect UV detection
into peak trends and their behaviours and method validation of non-ionic substances in liquid chromatography by
will be done in future. addition of a UV-absorbing uncharged compound to the
mobile phase, Chromatographia, 1984, 19(1), 274–279.
Conclusion 5 L. L. M. Glavina and F. F. Cantwell, Origin of indirect
detection in the liquid chromatography of a neutral sample
Direct identication of a negative peak linear to ethanol with an ionic probe using an ODS bonded phase and
concentration is studied and conrmed in the present work. aqueous mobile phase, Anal. Chem., 1993, 65, 268–276.
Even though linearity in the ethanol quantication was 6 S. Jan, P. Coufal, B. Zuzana and T. Eva, System peaks and
observed from 0.1–100% (v/v), the use of two injection volumes their positive and negative aspects in chromatographic
for two sets of concentration ranges (0.1–5% and >5–100%) is techniques, J. Sep. Sci., 2005, 28, 1263–1270.
suggested to avoid peak splitting and peak merging at higher 7 J. M. Betz and J. G. Nikelly, Determination of ethanol in
ethanol concentrations. This method proves to be an alternative alcoholic beverages by liquid chromatography using the UV
to other ethanol detection techniques in chromatography. This detector, J. Chromatogr. Sci., 1987, 25, 391–394.

Anal. Methods This journal is © The Royal Society of Chemistry 2016


View Article Online

Technical Note Analytical Methods

8 UV cutoffs for solvent, wavelength cutoffs for common simultaneous determination of cisplatin and the complex
mobile phases [Internet], [cited 2015 Sep 8], available from: cis,cis,trans-diamminedichlorodihydroxoplatinum(IV) in
http://www.molecularinfo.com/MTM/G/G3/G3-1/G3-1-11.html. a aqueous solutions, Quı́mica Nova, 2011, 34(8), 1450–1454.
9 The Royal Society of Chemistry, Modern chemical 18 Corporation S., Introduction to HPLC: LC world talk
techniques [Internet], [cited 2015 Dec 3], available from, special issue volume 1 [Internet], [cited 2015 Sep 10],
http://media.rsc.org/Modern chemical techniques/MCT4 available from: http://www.shimadzu.de/sites/default/les/
UV and visible spec.pdf. LC_World_Talk_Special_Issue_Volume1.pdf.
10 Agilent Technologies Inc, Tips and tricks of HPLC system 19 P. J. Weimer and J. G. Zeikus, Fermentation of Cellulose and
Published on 27 April 2016. Downloaded by UNIVERSITY OF NEBRASKA on 03/06/2016 15:44:33.

troubleshooting [Internet], [cited 2015 Dec 3], available Cellobiose by Clostridium thermocellum in the Absence and
from, https://www.agilent.com/cs/library/slidepresentation/ Presesnce of Methanobacterium thermoautotrophicum,
Public/Tips_and_Tricks_HPLC_Troubleshooting.pdf. Appl. Environ. Microbiol., 1977, 33(2), 289–297.
11 E. Arvidsson, J. Crommen, G. Schill and D. Westerlund, 20 H. Pontes, P. G. Pinho De, S. Casal, H. Carmo, A. Santos,
Indirect detection in reversed-phase liquid T. Magalhães, et al., GC Determination of Acetone,
chromatography response deviations in ion-pairing Acetaldehyde, Ethanol, and Methanol in Biological
systems, J. Chromatogr., 1989, 461, 429–441. Matrices and Cell Culture, J. Chromatogr. Sci., 2009, 47,
12 J. Vial and A. Jardy, Taking into account both preparation 272–278.
and injection in high-performance liquid chromatography 21 R. E. Schirmer, Modern Methods of Pharmaceutical Analysis
linearity studies, J. Chromatogr. Sci., 2000, 38, 189–194. [Internet], CRC Press, 2nd edn, 1990 [cited 2015 Dec 3], p.
13 Michael Thompson, Stephen L. R. Ellison and Roger Wood, 464, available from, https://books.google.com/books?
Guidelines for Single- Laboratory Validation of Methods of id¼7NDrUh2HgxIC&pgis¼1.
Analysis (IUPAC Technical Report), Pure Appl. Chem., 2002, 22 M. W. Dong, Modern HPLC for practicing scientists [Internet],
74, 835–855. John Wiley & Sons, 2006 [cited 2015 Dec 3], p. 320,
14 A. Antonio, F. Zielinski, C. M. Braga, I. M. Demiate, available from, https://books.google.com/books?
F. L. Beltrame, A. Nogueira, et al., Development and id¼RYOHFneTy-UC&pgis¼1.
optimization of a HPLC-RI method for the determination 23 S. C. Moldoveanu and V. David, Essentials in Modern HPLC
of major sugars in apple juice and evaluation of the effect Separations [Internet], Newnes, 2012 [cited 2015 Dec 3], p.
of the ripening stage, Food Sci. Technol., 2014, 34(1), 38–43. 550, available from: https://books.google.com/books?
15 L. Huber, Validation of analytical methods [Internet], Agilent id¼Dj2diIVudQwC&pgis¼1.
Technologies, [cited 2016 Mar 25], available from, https:// 24 P. Jandera, Stationary and mobile phases in hydrophilic
www.agilent.com/cs/library/primers/Public/5990- interaction chromatography: a review, Anal. Chim. Acta,
5140EN.pdf. 2011, 692(1–2), 1–25.
16 E. Staes, E. Rozet, Bernard Učakar, P. Hubert and V. Préat, 25 D. W. Lachenmeier, K. Schoeberl, F. Kanteres, T. Kuballa,
Validation of a method for the quantitation of ghrelin and E. M. Sohnius and J. Rehm, Is contaminated unrecorded
unacylated ghrelin by HPLC, J. Pharm. Biomed. Anal., 2010, alcohol a health problem in the European Union? A review
51, 633–639. of existing and methodological outline for future studies,
17 Y. Ramos, C. Hernández, L. A. Fernandez, M. B. Y. E. Veliz Addiction, 2011, 106(suppl. 1), 20–30.
and R. Small, Optimization of a HPLC procedure for

This journal is © The Royal Society of Chemistry 2016 Anal. Methods

You might also like