Professional Documents
Culture Documents
CXDI GC-H
CXDI GC-S
CXDI GC-T
For CXDI-1 System
Operation Manual
1. Introduction .........................................................................................1
2. Settings ...............................................................................................2
3. Communication ...................................................................................9
1. Introduction .......................................................................................13
2. Settings .............................................................................................14
3. Communication .................................................................................21
Pages 1 through 12 are for the users of the CXDI system software version 3.1.
See pages 13 through 24 if your CXDI system software is version 4.0 or 4.1.
1. Introduction
This software enables communication between the CXDI and the X-ray generator.
Thus, by setting the exposure parameters for the X-ray generator from the CXDI, the
procedure of exposure can be simplified.
However, it is essential to always confirm that the exposure parameters displayed on
the exposure screen and the exposure parameters displayed on the X-ray generator are
the same before performing the exposure.
This software is set to send either anatomical code or exposure parameters during the
installation of the software by the service personnel.
Refer to section 2.1 when sending anatomical code, or section 2.2 when sending
exposure parameters to the X-ray generator.
–1–
2. Settings (For CXDI Software Version 3.1)
2. Settings
NOTE: If you do not set any anatomical code for the exposure mode, it will be re-
garded as “0” having been set.
From the exposure screen, choose system > EDIT BODY PARTS > PARA EDIT.
Enter the anatomical code to be sent to the X-ray generator in less than 256
characters using the displayed keyboard. If you want to set an identical anatomi-
cal code for multiple exposure mode keys, put a serial number after the ana-
tomical code with a hyphen between it and the code, as follows:
–2–
2. Settings (For CXDI Software Version 3.1)
Or choose CANCEL.
When <OK> is chosen, a confirmation message will appear. Choose <OK> to
confirm the change or choose CANCEL.
* Some X-ray generators need the size of the patient to be entered when anatomical
code is used.
From the exposure screen, choose system > SETUP MENU > CUSTOMIZE.
–3–
2. Settings (For CXDI Software Version 3.1)
–4–
2. Settings (For CXDI Software Version 3.1)
From the exposure screen, choose system > SETUP MENU > EDIT BODY
PARTS > PARA EDIT.
Error will occur unless “kV”, “mA” and “msec” are set.
–5–
2. Settings (For CXDI Software Version 3.1)
Phototimer key
Focal spot
Touch the key and select the size of the focal spot.
Body size
Touch the key and select the size of the patient.
Phototimer
Each time the key is touched, the number and brightness of phototimers change.
The bright phototimer is the one that will be used.
Density
Touch the key of the required density.
–6–
2. Settings (For CXDI Software Version 3.1)
6 Touch EXIT
–7–
2. Settings (For CXDI Software Version 3.1)
–8–
3. Communication (For CXDI Software Version 3.1)
3. Communication
–9–
3. Communication (For CXDI Software Version 3.1)
– 10 –
4. When Error Occurs (For CXDI Software Version 3.1)
If any of the following error messages appears, take the countermeasures depending
on the displayed message.
If error message other than indicated below appears, please contact a Canon represen-
tative or distributor, and let us know in detail about the message that is displayed.
– 11 –
4. When Error Occurs (For CXDI Software Version 3.1)
(2) If the CXDI cannot communicate with the X-ray generator after restarting the
CXDI with the power of the X-ray generator turned ON, turn the power of the X-
ray generator OFF and then ON again.
– 12 –
1. Introduction (For CXDI System Software Versions 4.0 and 4.1)
1. Introduction
This software enables communication between the CXDI and the X-ray generator.
Thus, by setting the exposure parameters for the X-ray generator from the CXDI, the
procedure of exposure can be simplified.
However, it is essential to always confirm that the exposure parameters displayed on
the exposure screen and the exposure parameters displayed on the X-ray generator are
the same before performing the exposure.
This software is set to send either anatomical code or exposure parameters during the
installation of the software by the service personnel.
Refer to section 2.1 when sending anatomical code, or section 2.2 when sending
exposure parameters to the X-ray generator.
– 13 –
2. Settings (For CXDI Software Version 4.0 and 4.1)
2. Settings
NOTE: If you do not set any anatomical code for the exposure mode, it will be re-
garded as “0” having been set.
From the exposure screen, choose System > EDIT EXPOSURE MODE.
Enter the anatomical code to be sent to the X-ray generator in less than 256
characters using the displayed keyboard. If you want to set an identical anatomi-
cal code for multiple exposure mode buttons, put a serial number after the
anatomical code with a hyphen between it and the code, as follows:
Anatomical code - Serial number
– 14 –
2. Settings (For CXDI Software Versions 4.0 and 4.1)
For example, in order to set an anatomical code of “12” for three exposure
mode buttons, enter it as follows:
12-01, 12-02, 12-03
It does not matter how many digits there are in the serial number as long as the
total number of digits in the anatomical code and the serial number comes to
256 or less.
Or choose CANCEL.
When OK is chosen, a confirmation message will appear. Choose OK to confirm
the change or choose CANCEL.
* Some X-ray generators need the size of the patient to be entered when anatomical
code is used.
From the exposure screen, choose System > SETUP MENU > SYS. SETUP.
– 15 –
2. Settings (For CXDI Software Version 4.0 and 4.1)
– 16 –
2. Settings (For CXDI Software Versions 4.0 and 4.1)
From the exposure screen, choose System > EDIT EXPOSURE MODE.
Error will occur unless “kV”, “mA” and “msec” are set.
– 17 –
2. Settings (For CXDI Software Version 4.0 and 4.1)
5 Touch GENERATOR
Phototimer key
Focal spot
Touch the key and select the size of the focal spot.
Body size
Touch the key and select the size of the patient.
Phototimer
Each time the key is touched, the number and brightness of phototimers change.
The bright phototimer is the one that will be used.
Density
Touch the key of the required density.
6 Touch EXIT
– 18 –
2. Settings (For CXDI Software Versions 4.0 and 4.1)
– 19 –
2. Settings (For CXDI Software Version 4.0 and 4.1)
– 20 –
(For CXDI System Software Versions 4.0 and 4.1)
3. Communication
– 21 –
3. Communication (For CXDI Software Versions 4.0 and 4.1)
– 22 –
(For CXDI System Software Versions 4.0 and 4.1)
If any of the following error messages appears, take the countermeasures depending
on the displayed message.
If error message other than indicated below appears, please contact a Canon represen-
tative or distributor, and let us know in detail about the message that is displayed.
– 23 –
4. When Error Occurs (For CXDI Software Version 4.0 and 4.1)
(2) If the CXDI cannot communicate with the X-ray generator after restarting the
CXDI with the power of the X-ray generator turned ON, turn the power of the X-
ray generator OFF and then ON again.
– 24 –
Blank Page
CANON INC.
20-2 Kiyohara-Kogyodanchi, Utsunomiyashi, Tochigiken, Japan
Telephone: (81)-28-667-5711
CANON EUROPA N. V.
Bovenkerkerweg 59-61, P. O. Box 2262, 1180 EG Amstelveen, The Netherlands
Telephone: (31)-20-545-8926