You are on page 1of 1

PL-TEC-EFR-015 CALIBRATION CERTIFICATE CORRECTION FORM

LABORATORY Room 21.5 3.5


DATE OF UNCERTAINITY ADJUSTED RANGE ADJUSTED RANGE
No INSTRUMENT NAME SERIAL NUMBER DUE DATE CURRENT USE SET POINT RANGE FACTOR CORRECTION RANGE FINAL RANGE
CALIBRATION (MU) (CORRECTION) (UNCERTAINITY) Humidity 50 20
1 Lec Medical Refregrator 5.02002E+16 24/10/2022 24/10/2023 Refrigrator 5 3 -0.06 2.27 2 _ 8 2.06 _ 8.06 4.33 _ 5.79 4.33 _ 5.79 Refrigrator 5 3
2 Biobase yc010025200114 24/10/2022 24/10/2023 Refrigrator 5 3 0.48 0.72 2 _ 8 1.52 _ 7.52 2.24 _ 6.8 2.24 _ 6.8 Refrigrator (Blood Bank) 3.5 2.5
3 Panasonic 15090388 24/10/2022 24/10/2003 Refrigrator 5 3 -0.94 1.54 2 _ 8 2.94 _ 8.94 4.48 _ 7.4 4.48 _ 7.4 Freezer (-18.0°C to -25.0°C) -21.5 3.5
4 Thermoscientific 1.53617E+14 24/10/2022 24/10/2023 Freezer (-18.0°C to -25.0°C) -21.5 3.5 1.07 1.85 -25 _ -18 -26.07 _ -19.07 -24.22 _ -20.92 -24.22 _ -20.92 Deep Freezer (-20.0°C to -40.0°C) -30 10
5 MAS 01 Room 21.5 3.5 0.39 0.5 18 _ 25 17.61 _ 24.61 18.11 _ 24.11 18.11 _ 24.11 Deep Freezer (-50.0°C to -90.0°C) -70 20
6 MAS 01 Humidity 50 20 -7.56 2.1 30 _ 70 37.56 _ 77.56 39.66 _ 75.46 39.66 _ 75.46 Incubator 36 1
7 - - - _ - - _ - - _ - - _ - Microbiology Incubator 34 1
8 MAS 02 Room 21.5 3.5 0.39 0.5 18 _ 25 17.61 _ 24.61 18.11 _ 24.11 18.11 _ 24.11 Hot Air Oven 60 0
9 MAS 02 Humidity 50 20 -7.06 2.1 30 _ 70 37.06 _ 77.06 39.16 _ 74.96 39.16 _ 74.96 Platelet Shaker Incubator 22 2
10 MAS 03 Room 21.5 3.5 0.19 0.5 18 _ 25 17.81 _ 24.81 18.31 _ 24.31 18.31 _ 24.31 Water Bath 37 1
11 MAS 03 Humidity 50 20 -5.56 2.1 30 _ 70 35.56 _ 75.56 37.66 _ 73.46 37.66 _ 73.46
12 MAS 04 Room 21.5 3.5 -0.31 0.5 18 _ 25 18.31 _ 25.31 18.81 _ 24.81 18.81 _ 24.81
13 MAS 04 Humidity 50 20 -9.46 2.1 30 _ 70 39.46 _ 79.46 41.56 _ 77.36 41.56 _ 77.36
14 MAS 05 Room 21.5 3.5 -0.21 0.5 18 _ 25 18.21 _ 25.21 18.71 _ 24.71 18.71 _ 24.71
15 MAS 05 Humidity 50 20 -6.26 2.1 30 _ 70 36.26 _ 76.26 38.36 _ 74.16 38.36 _ 74.16
16 MAS 06 Room 21.5 3.5 -0.61 0.5 18 _ 25 18.61 _ 25.61 19.11 _ 25.11 19.11 _ 25.11
17 MAS 06 Humidity 50 20 -3.86 2.1 30 _ 70 33.86 _ 73.86 35.96 _ 71.76 35.96 _ 71.76
18 Alphatechnics 1 Room 21.5 3.5 -0.1 0.13 18 _ 25 18.1 _ 25.1 18.23 _ 24.97 18.23 _ 24.97
19 Alphatechnics 2 Refrigrator 5 3 -0.1 0.13 2 _ 8 2.1 _ 8.1 2.23 _ 7.97 2.23 _ 7.97
20 _ - - - _ - - _ - - _ - - _ -

Guide Line
Apply the "FINAL RANGE" in the "Temperature Monitoring Form".

Revision No: 4 1 Revision Date: 28/12/2022