Professional Documents
Culture Documents
Vista 120 SC
WARNING
A WARNING statement provides important
information about a potentially hazardous
situation which, if not avoided, could result in
death or serious injury.
CAUTION
A CAUTION statement provides important
information about a potentially hazardous
situation which, if not avoided, may result in minor
or moderate injury to the user or patient or in
damage to the medical device or other property.
NOTE
A NOTE provides additional information intended
to avoid inconvenience during operation.
Contents
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . 163
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . 169
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
SpO2 accessories . . . . . . . . . . . . . . . . . . . . . . . 170
NIBP accessories . . . . . . . . . . . . . . . . . . . . . . . 171
TEMP accessories . . . . . . . . . . . . . . . . . . . . . . 172
Other accessories. . . . . . . . . . . . . . . . . . . . . . . 173
WARNING
Portable RF communications equipment
(including peripherals such as antenna cables
and external antennas) should be used no
closer than 30 cm (12 inches) to any part of
Vista 120 SC, including cables specified by
the manufacturer. Otherwise, degradation of
the performance of this equipment could
occur.
Sterile accessories
CAUTION
Risk of medical device failure and of patient injury
Do not use sterile-packaged accessories if the
packaging has been opened, is damaged or if
there are other signs of non-sterility.
Single-use accessories must not be reused,
reprocessed, or resterilized.
Installing accessories
CAUTION
Risk of device failure
Install accessories to the basic device in
accordance with the instructions for use of the
basic device. Make sure that there is a safe
connection to the basic device.
WARNING
Devices connecting with monitor should be
equipotential.
WARNING WARNING
Do not rely exclusively on the auditory alarm If several items of medical equipment are
system for patient monitoring. Adjustment of interconnected, pay attention to the sum of
alarm volume to a low level or off during the leakage currents, otherwise it may cause
patient monitoring may result in a hazard to shock hazard. Consult the service personnel.
the patient. Remember that the most reliable
method of patient monitoring combines close WARNING
personal surveillance with correct operation In monitoring mode, when monitoring is in
of monitoring equipment. process, if the power supply is off and there is
no battery for standby, the monitor will be off.
WARNING The settings configured by the user can be
Accessory equipment connected to the stored, and settings not configured by user
analog and digital interfaces must be certified keep no change. That is, the last settings used
according to the respective IEC/EN standards will be recovered when the power is restored.
(e.g. IEC/EN 60950 for data processing In ward round or spot-checking mode, the
equipment and IEC/EN 60601-1 for medical patient type will be restored to adult by
equipment). Furthermore all configurations default, and the monitor is in status with no
shall comply with the valid version of the patients, and other settings processing after
standard IEC/EN 60601-1. Therefore anybody, power off are the same with that in monitoring
who connects additional equipment to the mode.
signal input or output connector to configure
a medical system, must make sure that it WARNING
complies with the requirements of the valid The device and accessories are to be
version of the system standard IEC/EN 60601- disposed of according to local regulations
1. If in doubt, consult our technical service after their useful lives. Alternatively, they can
department or the local distributor. be to the dealer or the manufacturer for
recycling or proper disposal. Batteries are
WARNING hazardous waste. Do NOT dispose them
Only patient cable and other accessories together with house-hold garbage. At the end
supplied by Dräger can be used. The of their life hand the batteries over to the
performance and electric shock protection applicable collection points for the recycling
cannot be guaranteed, and the patient may be of waste batteries. For more detailed
injured otherwise. Prior to use, check if the information about recycling of this product or
casing of a disposable accessory is intact. Do battery, contact the local Dräger sales
not use it if its casing is damaged. representative.
WARNING WARNING
When interfacing with other equipment, a test If leakage or foul odor is detected, ensure that
for leakage current must be performed by there’s no fire around.
qualified biomedical engineering personnel
before using with patients. WARNING
The packaging is to be disposed of according
to local or hospital’s regulations; otherwise, it
may cause environmental contamination.
Place the packaging at the place which is
inaccessible to children.
WARNING WARNING
This equipment is not intended for home Only items that have been specified as part of
usage. the system or specified as being compatible
with the system can be connected to the
WARNING system.
Do not service or maintain the monitor or any
accessory which is in use with the patient. WARNING
Connecting any accessory (such as external
WARNING printer) or other device (such as the
The appliance coupler or mains plug is used computer) to this monitor makes a medical
as isolation means from supply mains. system. In that case, additional safety
Position the monitor in a location where the measures should be taken during installation
operator can easily access the disconnection of the system, and the system shall provide:
device. a) Within the patient environment, a level of
safety comparable to that provided by
WARNING medical electrical equipment complying
Assembly of the monitor and modifications with IEC/EN 60601-1, and
during actual service life shall be evaluated
b) Outside the patient environment, the
based on the requirements of IEC60601-1.
level of safety appropriate for non-
medical electrical equipment complying
WARNING with other IEC or ISO safety standards.
The monitors are MR Unsafe. The monitors
are not intended for use in an MRI WARNING
environment.
The medical electrical equipment needs to be
installed and put into service according to the
WARNING EMC Information provided in this user
Only recommended batteries can be used for manual.
the monitor.
WARNING
WARNING Portable and mobile RF communications
In monitoring mode, without use of data store equipment can affect medical electrical
function, all data measured (including trend equipment; refer to the recommended
data, review data, alarm events and so on) are separation distances provided in this user
cleared either when the monitor is turned off manual.
or when the monitor is powered down in the
process of monitoring. WARNING
Using accessories other than those specified
WARNING may result in increased electromagnetic
Additional multiple socket-outlets or emission or decreased electromagnetic
extension cords can’t be connected to the immunity of the monitoring equipment.
system.
WARNING WARNING
The monitor should not be used adjacent to or When the monitor is used with HF surgical
stacked with other equipment. If adjacent or equipment, the transducer and the cables
stacked use is necessary, the user must check must be avoided from conductive connection
that normal operation is possible in the to the HF equipment. This is to protect against
necessary configuration before starting burns to the patient.
monitoring patients.
WARNING
WARNING To protect the monitor from damage during
Do not touch accessible parts of medical or defibrillation, for accurate measurement
non-medical electrical equipment in the information and to protect against noise and
patient environment and the patient other interference, use only accessories
simultaneously, such as USB connector, VGA specified by Dräger.
connector or other signal input/output
connectors. WARNING
No modification of this equipment is allowed
WARNING without authorization of the manufacturer. If
SHOCK HAZARD - Don’t connect electrical this equipment is modified, appropriate
equipment, which has not been supplied as a inspection and testing must be conducted to
part of the system, to the multiple portable ensure continued safe operation.
socket-outlet supplying the system.
WARNING
WARNING Clinical decision making based on the output of
SHOCK HAZARD - Don’t connect electrical the device is left to the discretion of the provider.
equipment, which has been supplied as a part
of the system, directly to the wall outlet when WARNING
the non-medical equipment is intended to be The monitor is equipped with wireless AP/Wi-
supplied by a multiple portable socket-outlet Fi to receive RF electromagnetic energy.
with an isolation transformer. Therefore, any other equipment complying
with CISPR radiation requirements may also
WARNING interfere with the wireless communication and
Operation of the equipment exceeding make it interrupted.
specified physiological signal or the
operational specification may cause WARNING
inaccurate results. Wireless LAN equipment contains an
intentional RF radiator that has the potential
WARNING of interfering with other medical equipment,
The equipment can provide protective means including patient implanted devices. Be sure
to prevent the patient from being burned when to perform the electromagnetic compatibility
used with HF SURGICAL EQUIPMENT. The test before installation and any time new
equipment can protect against the effects of medical equipment is added to the Wireless
the discharge of a defibrillator. Use only LAN coverage area.
Dräger-approved accessories.
WARNING CAUTION
If the protective grounding (protective earth) Do not immerse transducers in liquid. When using
system is doubtful, the monitor must be solutions, use sterile wipes to avoid pouring fluids
supplied by internal power only. directly on the transducer.
WARNING CAUTION
Portable RF communications equipment Do not use autoclave or gas to sterilize the
(including peripherals such as antenna cables monitor, recorder or any accessories.
and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the CAUTION
monitor, including cables specified by the
The device and reusable accessories may be
manufacturer. Otherwise, degradation of the
sent back to the manufacturer for recycling or
performance of this equipment could result.
proper disposal after their useful lives.
WARNING
CAUTION
The monitor is suitable for use in the presence
of electrosurgery. When the monitor is used Disposable devices are intended for single use
with HF surgical equipment, user (doctor or only. They should not be reused as performance
nurse) should be cautious about patient could degrade or contamination could occur.
safety.
CAUTION
WARNING Remove a battery whose life cycle has expired
Make sure networking function is used in a from the monitor immediately.
secure network environment.
CAUTION
WARNING To ensure patient safety, use only parts and
To avoid malicious tampering and theft of data accessories manufactured or recommended by
transmitted by the network, it is Dräger.
recommended to switch on the encryption
function. After the encryption function is CAUTION
turned on (it is set to on by default), the Before connecting the monitor to the AC power,
monitor will authenticate the accessed Vista make sure the voltage and the power frequency
120 CMS and GATEWAY devices and encrypt are consistent with the requirements indicated on
the transmitted data to ensure the security. the device label or in this user manual.
CAUTION CAUTION
Electromagnetic Interference - Ensure that the Protect the device against mechanical damage
environment in which the patient monitor is resulting from falls, impacts, and vibration.
installed is not subject to any sources of strong
electromagnetic interference, such as radio CAUTION
transmitters, mobile telephones, microwaves, etc. Do not touch the touch screen with a sharp object.
CAUTION
Keep the environment clean. Avoid vibration.
Keep it far away from corrosive medicine, dust
area, high temperature and humid environment.
CAUTION NOTE
A ventilated environment is required for monitor When there's measurement beyond range, invalid
installation. Do not block up the ventilation grille at measurement or no measurement value, it will
the back of the device. display -?-.
CAUTION NOTE
The device must be connected to the ground to The materials with which the patient or any other
avoid signal interference. person can come into contact conform with the
standard of EN ISO 10993-1: 2009.
CAUTION
NOTE
To protect eyes from damage, don't look directly
at supplementary light for long time. In normal use, the operator shall stand in front of
the monitor.
CAUTION
Poor connection might be caused by frequently
plugging and unplugging the power cord. Check
the power cord regularly and replace it in time.
NOTE
Position the device in a location where the
operator can easily see the screen and access the
operating controls.
NOTE
The monitor can only be used on one patient at a
time.
NOTE
If the monitor gets damp or liquid pours on the
monitor, contact the service personnel of Dräger.
NOTE
This monitor is not a device for treatment
purposes.
NOTE
The pictures and interfaces in this manual are for
reference only.
NOTE
Regular preventive maintenance should be carried
out every two years. The user is responsible for
any requirements specific to his country.
CAUTION
When the monitor is returned for maintenance,
disposed of, or removed from the medical
institution for other reasons, it is necessary to
ensure that all patient data are removed from the
monitor (Refer to Section Deleting data stored in
the storage device).
CAUTION
Protect the privacy for the information and the
data displayed on the screen, and for the
information and the data stored in the monitor.
NOTE
Log files generated by the monitor are used for
system troubleshooting and do not contain
protected medical data.
Application
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Intended user . . . . . . . . . . . . . . . . . . . . . . . . . 22
Intended use
The monitor is intended to be used by healthcare The monitor is intended for measuring and
professionals whenever there is a need for monitoring physiologic parameters, including
monitoring the physiological parameters of functional arterial oxygen saturation (SpO2), Pulse
patients. The monitor is intended for monitoring, Rate (PR), Non-Invasive Blood Pressure (NIBP)
analyzing, recording, and alarming of multiple and Temperature (TEMP).
physiological parameters in healthcare
The monitor is intended for use only by clinical
environments for adult, pediatric, and neonatal
professionals or under their guidance. It must only
patients.
be used by persons who have received adequate
training.
The monitor is not intended for MRI environments.
F3000 TEMP Module is not for neonates.
CAUTION
Device is for use in health care facilities only and
exclusively by persons with specific training and
experience in its use.
Intended user
Overview
Front view . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Rear view. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Side view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Bottom view . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . 31
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Front view
A Alarm When an alarm occurs, the alarm indicator will light or flash. The alarm level is
indicator/Stand color coded;
by indicator
Breathing light indicates the standby status.
B AC power When the monitor is connected to AC power, the indicator is in green.
indicator
C Battery Refer to the section Battery power indicator for details.
indicator
D Power On/Off When connected to the AC power supply or battery powered, press the key to turn
switch the monitor on. Press the key again and hold for 3 seconds to turn the monitor off.
An indicator is built in this switch. It turns on when the monitor is on and turns off
when the monitor is off.
E NIBP Press to inflate the cuff and perform NIBP measurement. Press again to stop the
measurement measurement and deflate the cuff.
F Admit/Create Admit or create new patient;
new patient
Press this button to return to the main interface.
G Switch patient In main interface, press the button to switch the patient type;
type
In other interfaces, press the button to return to the main interface.
H Menu Press the button to open the main menu when there is no menu open. Press it
again to exit.
I Rotary knob The user can turn the rotary knob clockwise or counter-clockwise to highlight the
desired item. Press the rotary knob to select the item.
J TAT-5000S- Provides Temporal Artery (TA) Temperature in about 3 seconds.
RS232
Thermometer
K F3000 TEMP Measures patient’s oral, axillary or rectal temperatures.
module
Rear view
A Portable handle
B Power supply interface
C Power cord Used to prevent the power cord from detaching. Place the latch on the power cord
safety latch and press it down firmly to ensure that it secures the power cord.
D Equipotential If the monitor is used with other devices, connect this terminal to eliminate
grounding potential ground differences between devices.
terminal
E Network Network interface: standard RJ45 interface that connects the monitor to Vista 120
interface/Nurse CMS 1.62 and above or Vista 120 Gateway.
call port
Nurse call port: it connects the monitor to the hospital’s nurse call system. Alarms
indications are alerted through the nurse call system if configured to do so.
F USB Used to connect the USB devices.
interfaces.
Side view
Left side
Right side
A Speaker
B Recorder
C NIBP port
D Dräger SpO2 port
E Nellcor SpO2 port
F TEMP communication interface
Bottom view
A Heat sink
B Battery compartment door
Abbreviations
Hb Hemoglobin
Hb-CO Carbon mono-xide hemoglobin
HR Heart rate
ICU Intensive care unit
ID Identification
IEC International Electrotechnical
Commission
IEEE Institute of Electrical and
Electronic Engineers
LCD Liquid crystal display
LED Light emitting diode
MAP Mean arterial pressure
Symbols
Menu Manufacturer
Warning
Battery indicator (Background: Yellow; Symbol &
outline: black)
Non-ionizing electromagnetic
USB port
radiation
Defibrillation-proof type BF
This way up
applied part
Recycle Non-sterile
Getting started
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Initial inspection . . . . . . . . . . . . . . . . . . . . . . . 36
Overview
NOTE
The monitor installations and settings must be
configured by the authorized hospital personnel.
Initial inspection
Before unpacking, check the packaging and ensure Open the package carefully and remove the
that there are no signs of mishandling or damage. monitor and accessories. Check that the contents
If the shipping cartons are damaged, contact the are complete and that the correct options and
local supplier for assistance. accessories have been delivered.
If the user has any question, contact the local
supplier.
To connect the AC power cable: 2 Connect the power cable provided with the
monitor. Connect the other end of the power
1 Make sure the AC power supply complies with
cable to a grounded power outlet.
the following specifications: 100 V-240 V,
50 Hz/60 Hz. NOTE
Connect the power cable to the socket specialized
for hospital use.
NOTE
Only use the power cable supplied by Dräger.
WARNING
Changing date and time will affect the storage
of trend data.
NOTE
If the system is not used for a longer period of time,
its system time may be inaccurate. In this case,
readjust the system time after powering on.
NOTE
If the system time cannot be saved and resumes
the default value after restart, contact the service
department of Dräger to replace the button cell on
main board.
Operation
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Using keys. . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Permanent keys . . . . . . . . . . . . . . . . . . . . . . . . 43
Shortcut keys . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Hardkeys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Pop-up keys . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Work modes . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Operating modes . . . . . . . . . . . . . . . . . . . . . . 45
Demo Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Standby Mode. . . . . . . . . . . . . . . . . . . . . . . . . . 45
Night Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Setting languages. . . . . . . . . . . . . . . . . . . . . . 47
Using mouse . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Overview
The user may frequently use the follow functions: Everything the user needs to operate the monitor is
accessible from its user interface. Screen elements
SpO2 monitoring (Refer to Monitoring SpO2 for
include measurement data, waveforms, screen
more information.)
keys, information fields, alarms fields and menus.
PR monitoring (Refer to Monitoring PR for more The monitor can be configured in a number of
information.) different ways. For example, the user can access
an item through the on-screen setup menu, via a
NIBP monitoring (Refer to Monitoring NIBP for
hard key, or via a shortcut key. This Instructions for
more information.)
Use describes how to access items via an on-
TEMP monitoring (Refer to Monitoring TEMP screen menu.
for more information.)
Alarm (Refer to Alarms for more information.)
A B C
D
E
H I J K L M N OP QR S
Measuring oral TEMP in ADU mode For device with F3000 TEMP
module only.
Taking Temporal Artery (TA) Temperature via TAT- For device with TAT-5000S-
5000S-RS232 Thermometer RS232 Thermometer only.
Using keys
The monitor has four different types of keys. If the In monitoring mode, the displayed keys are: Alarm
key sound is enabled, the monitor gives a normal Mute, Admission, Review, NIBP measurement,
key sound when the operation is valid. General, Standby mode, Record, Barcode
scanning, Night Mode and Score.
In ward round mode, the displayed keys are:
Permanent keys Review, NIBP AVG, NIBP measurement, General,
Standby mode, Barcode scanning and Score.
A permanent key is a graphical key that is In spot-checking mode, the displayed keys are:
permanently located at the bottom of the main Record, Review, NIBP AVG, NIBP measurement,
screen allowing fast, direct access to functions. General, Standby mode, Barcode scanning and
Score.
Menu, to display the main setup menu
Audio alarm paused/off
NIBP measurement
Hardkeys
Standby mode
A hardkey is a physical key on a monitoring device,
such as the main menu key on the front panel.
NIBP average measurement Refer to the illustration in Front view for more
information.
Night mode
Pop-up keys
Review (in ward round mode)
Pop-up keys are task-related graphical keys that
appear automatically on the screen when required.
For example, the confirmation pop-up key appears
only when the user needs to confirm a change.
Work modes
NOTE
The history data in each mode can be viewed only
in corresponding mode.
Operating modes
NOTE
In night mode, the sound of key and pulse is
muted; the alarm volume and screen brightness
are down to their minimum; the settings including
key volume, PR volume, alarm volume and screen
brightness are unavailable.
Adjusting volume
To check the monitor information, select Menu > About includes: software version, serial number,
Common Function > Device Information. device name and so on.
Monitor information includes Config, About,
Network includes: Network type, Local Net No.,
Network.
Server IP (such as Vista 120 CMS), Local Mask
Config includes: the configuration monitor No., Local Gateway No., Mac address,
supported (such as Wi-Fi, Wired, USB, Barcode, Communication protocol and so on.
etc.), the currently used configurations are marked
with √. NOTE
The contents related with network can be set in
User Maintain > Network.
Setting languages
To set Chinese input method: 2 Select Common > Chinese Input option on the
pop-up dialog box to toggle between On and
1 Select Menu > User Maintain, then type the
Off.
maintenance password into the displayed
dialog box.
The user can disable touch screen operation by of screen. To enable the touch screen operation,
selecting and holding the permanent key select and hold the menu key for three
for three seconds. A message of Screen Locked seconds again.
To enter the barcode setup menu, select Menu > If the scanner is connected for the first time, the
User Maintain. After entering the required monitor will pop up a confirmation message to ask
password, select Common > Barcode Setup. user whether the new USB device is added as
Then the user can set MRN, last name, first name scanner. Choose Yes to add as scanner, choose
and so on. No to add as USB device.
User can also check relevant scanner device NOTE
information in Scanner Management.
The start and end code should be set before using
scanner to update patient, otherwise the barcode
can’t be recognized correctly. After setting start
and end code, user should also set male code and
female code to distinguish the gender.
Using mouse
Monitoring mode
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Admitting a patient . . . . . . . . . . . . . . . . . . . . . 50
Patient category . . . . . . . . . . . . . . . . . . . . . . . . 50
Quick admit . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Admit by barcode . . . . . . . . . . . . . . . . . . . . . . 51
Event marking. . . . . . . . . . . . . . . . . . . . . . . . . 53
Overview
Admitting a patient
The monitor displays physiological data and Doctor: Enter the attending doctor for the
records the information in trends. This allows the patient.
user to monitor unadmitted patients. It is however
Gender: Male, Female and N/A.
important to admit patients properly so that the user
can identify the patient on recordings, reports, and Type: Choose the patient type, either Adult,
networked devices. Pediat, or Neonat.
During admission, the patient category setting BloodType: N/A, A, B, AB and O.
determines the algorithm that the monitor uses to
Date of Birth: Enter the patient’s date of
process and calculate measurements. These
birth.
include safety limits that are applied for specific
measurements and alarm limit ranges. Date of Admission: Enter the patient’s date
of admission.
To admit a patient:
Height: Enter the patient’s height.
1 Select Menu > Patient Management > New
Patient, or press the Admit/Create new patient Weight: Enter the patient’s weight.
Height unit: cm or inch.
hardkey on the front panel, then a Weight unit: kg or lb.
message is displayed to ask the user to confirm
to update patient. NOTE
2 Click No to cancel this operation; click Yes, the Admitting new patient will clear the history data in
Patient Info interface is displayed. the monitor associated with the patient.
WARNING
Changing the patient category may change
the relevant configuration (such as alarm
limits). Always check alarm limits to make
sure that they are appropriate for the patient.
Quick admit
If the user does not have the time or information to 3 Click No to cancel this operation; click Yes to
fully admit a patient, the user can use Quick Admit continue and the Quick Admit interface is
to quickly admit a patient and fill in the rest of the displayed.
patient information later. To quickly admit a patient:
4 Configure Type to the correct setting.
Admit by barcode
NOTE
Patient information obtained from network server
cannot be edited.
NOTE
Alarm review The monitor can store a maximum of 200 alarm
events. As soon as the alarm event storage is full,
To review the alarm event, press the Review the earliest alarm event will be replaced by the
latest one.
Event marking
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Import Patient . . . . . . . . . . . . . . . . . . . . . . . . . 56
Delete patient . . . . . . . . . . . . . . . . . . . . . . . . . 58
Overview
Ward round mode is used for parameter Select Menu > Patient Management to check the
measurement and for multiple patients’ round data patient list. If the monitor is powered off, the patient
management (maximum 1000 patients). list will be stored. The user can also create, edit,
import and delete patients in this interface.
3 Use Select Patient permanent key, and then b If the patient information via MRN hasn’t
click icon + in popup window, or been stored in the monitor before:
4 Scan patient barcode by using barcode When monitor is connected with network
server and ADT Query (in User
Maintain) is also set to on, the monitor
scanning shortcut key . The barcode is will automatically inquire for patient
based on patient MRN. information from the server via MRN.
The corresponding patient information in
1) In the interface of creating new patient, scan
the monitor will be updated after user’s
the barcode:
confirmation.
The monitor will update the patient
information from the identified MRN. In this NOTE
interface, user can complete relevant Either Bed No. or MRN should be input.
information inputting.
Import Patient
To import patient to the patient list, select Patient If USB flash disk is selected, prior to importing,
Management > Import and choose Source (USB set the import data file as following steps:
flash disk or Network).
… … … … …
Step 2: Save the file as CSV with comma delimiter If Network is selected, input the query
format. The file name should be fixed as conditions (Department, Admission Date),
‘ImportPatientList.csv’, and the encoding format of
the file should be UTF-8. and then click query button . A list
including all the patients that meet the query
NOTE
conditions is displayed. Select patients from the
In the patient list of CSV data file, first name, last list, and click Import.
name and MRN cannot be more than 63 bits for
each.
User can also directly click shortcut key
NOTE
In importing process, the monitor will import the > icon to import patient from Network.
patient information that meets the requirements,
and the wrong patient information will be During importing, if the total patients in the monitor
prompted. The wrong patient information needs to exceed upper limit (1000 patients), a confirmation
be revised and re-imported. box will be popped up. Choose OK to enter Patient
Management interface, and user should manually
Step 3: Choose patients and click Import. delete the patients till the monitor can import
patients again. Choose No to return to Import
Patient interface, and user should delete patients
to be imported, then import again.
If monitor is connected with network server and patients > click Refresh List, and then all eligible
ADT Query (in User Maintain) is also set to on, patients will be refreshed to the window of Select
Patient.
NOTE
During importing or refreshing, if MRN or bed No.
has conflicts, the monitor will replace all conflicting
patient information. The measurement status will
keep unchanged if there is a MRN conflict, and
restore to the initial unmeasured status if there is a
bed No. conflict.
Delete patient
Choose a specific patient for measurement and the Select New Round to clear all patients’
patient information will be displayed on the main measurement status and measurement time.
interface. Measurement status will all return to ‘not
measured’ status. When the user sets Scale to
Current in Review interface, Current
Click Select Patient permanent key : measurement data will be cleared.
All patients’ bed No., name, measurement time
and measurement status will be displayed. The
circle next to bed No. shows measurement
status by colour. Green means the patient has
normal measurement value; no colour means
the patient has invalid measurement value or no
measurement value.
Click one patient to choose patient for current
measurement, and the patient information will
be displayed on the main interface.
Click icon + to create new patient.
Ward round record confirmation Click the icon on the right of one ward round
record to enter Patient Measurement Data
In main interface: window. In this window, user can view the
corresponding original data (20 groups at
Generate manually: by clicking Save, the most), including patient information, parameters
measurement data in main interface will be stored value and this ward round record. User can also
as a round record in Review menu. User can Save click Record to output the original data in this
multiple ward round records for a patient. interface, or modify the ward round record and
The measurement time of ward round record is click Save to complete.
subject to the time of clicking Save. If monitor is powered off, ward round record will be
stored. 1 MB space can store 100 groups of ward
round records at least, including patient information
Ward round record review and 20 original records.
Spot-checking mode
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Admit patient. . . . . . . . . . . . . . . . . . . . . . . . . . 62
Overview
Admit patient
If monitor is connected to network server, the 3 Scan patient barcode by using barcode
patient information can be synchronized to the scanning shortcut key.
network server after admission. To admit patient,
In creating new patient interface, user can scan
1 Select Menu > Patient Management, click the barcode through scanner. The patient
New Patient to enter new patient interface. information is directly updated to this interface
according to barcode information, and user
2 Select New Patient permanent key, or press
should Confirm it.
In other interfaces, serial No. will be
Admit/create new patient hardkey automatically added by one, and the patient
(without key sound), the monitor will information is directly updated to main interface
automatically admit new patient (that is, series according to barcode information.
No. will be automatically added by one and
other setting items are empty by default).
Select Patient Management > Patient Info, or If the monitor is powered off and restarted, the bed
directly click patient information area in main No.1 patient will be automatically selected as
interface to view or modify patient information current measured patient, the patient type is adult
(serial No. is not editable). and the other information is blank.
Select Menu > Review, or directly select Review Set Scale to display spot-checking data in
shortcut key to enter review interface. specified period. The valid spot-checking data
in 12 hours before the latest valid spot-checking
Click and to turn page.
data will be displayed by default.
Networked monitoring
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Cybersecurity . . . . . . . . . . . . . . . . . . . . . . . . . 66
Wi-Fi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
e-link function . . . . . . . . . . . . . . . . . . . . . . . . . 69
Gateway communication . . . . . . . . . . . . . . . . 70
HL7 communication . . . . . . . . . . . . . . . . . . . . 71
Overview
NOTE
When selecting dynamic IP mode, check the IP
address from Vista 120 CMS.
Cybersecurity
For more security operations, select Menu > User Step 1, in CA Certificate column, select
Maintain and input user maintain password > corresponding CA certificate whose suffix
Security. In this menu: should be ‘.cer’.
Select User Password > Modify to enter Step 2, in Client Certificate column, select
Modify User Password submenu, the user can corresponding client certificate whose suffix
change the password according to the prompts. should be ‘.cer’.
For safety considerations, change the
Step 3, in Private Key Certificate column,
password periodically, and a combination of
select corresponding private key whose
words and numbers is recommended. If Old
suffix should be ‘.pem’.
Password is forgotten, contact Dräger Service.
Step 4, in Password column, input
NOTE corresponding password.
When the monitor is turned on for the first time,
Step 5, click Confirm.
modify the User Maintain password according to
the prompts. The default initial User Maintain Steps of importing certificate for HL7:
password can be found in Section Passwords.
Step 1, in CA Certificate column, select
corresponding CA certificate whose suffix
Set CMS/Gateway Encryption to Off, AES
should be ‘.cer’.
(default) or TLS when user connects the
monitor with network server (Vista 120 CMS or Step 2, in Server Certificate column, select
gateway). corresponding server certificate whose
suffix should be ‘.cer’.
Click Load Certificate to install/upgrade the
Certificate via USB flash drive. The certificate Step 3, in Private Key Certificate column,
issued by Certificate Authority (CA) is select corresponding private key whose
recommended and self-signed certificate suffix should be ‘.pem’.
should be avoided.
Step 4, in Password column, input
Steps of importing certificate for Vista 120 CMS corresponding password.
and Gateway:
Step 5, click Confirm.
NOTE NOTE
CA certificate is the root certificate of CA The storage path for certificate related to TLS
institution. enterprise-level encryption method:
For Vista 120 CMS\gateway: root directory of USB
NOTE flash drive\certs\cms\
Client certificate/Server certificate are the SSL For HL7: root directory of USB flash drive
certificate obtained from CA institution and it is \certs\hl7\
recommended to use OV-SSL certificate type.
Set Firewall to On to protect against hacker
attacking.
NOTE
The private key shall be generated by the RSA- Click Firewall Rules to check rule details.
2048 algorithm and be encrypted by using the Set Packets Limit value for traffic monitoring. If
AES-256 algorithm. The password for encryption the data traffic per minute exceeds the
shall be at least 8 characters (a combination of threshold, the monitor will trigger the alarm to
letters and numbers). remind the user.
Set Auto Login to On/Off. When it is set to On,
NOTE
the user can directly enter the main interface;
The certificate format is as follows: For CA and otherwise, the user needs to enter a password
Client certificate/Server certificate, ‘.cer’ is for identity authentication.
supported; for private key, ‘.pem’ is supported. All
the certificates should comply with X.509 format. Setting User Login Timeout allows the user to
lock the screen when the monitor has been idle
for a certain amount of time, so that others
cannot access the monitor without entering the
password. If Never is chosen, the user has no
need to login again once the authorization is
successful.
Wi-Fi
Wi-Fi modules are optional to be configured in the 4 Choose a network from the interface, in which
monitors. And the user should configure the the user can check the network’s encryption
settings on the monitor following the steps below information (Security). The user will be
before connecting the monitor to a wireless prompted to enter the password of that network
network: if a password is required. After entering the
password and setting the IPv4 address, the
1 Select Menu > User Maintain, and input the
user can click to connect the network.
password.
5 Or select to connect the hidden networks.
2 In the User Maintain menu, select Network.
After entering Network Name, Security,
3 In the Network menu, select Wi-Fi from the password and setting the IPv4 address, the
Network Type list. And click Config to open the user can click to connect the hidden
Wi-Fi Setup interface. The available networks network.
will be listed in this interface.
WARNING WARNING
Before monitoring patient, the Network Type If Wi-Fi is unavailable, restart the monitor
(wired or Wi-Fi) should be selected, it is not (refer to Section Front view) to restore Wi-Fi
allowed to be switched during monitoring. function under the precondition of ensuring
Otherwise, the Wi-Fi may be unavailable. patient’s safety.
e-link function
Gateway communication
The monitor can be connected to the Gateway, For more information about Gateway
which provides clinicians with the capability of communication, refer to Vista 120 Gateway User
viewing and collecting patient data remotely and Manual.
the data exchange of selected clinical and
administrative information between Vista 120
Network and the hospital network.
To set the monitor server IP address, select User
Maintain > Network > Server IP. Make sure the
monitor share the same server IP with the
computer in which the Vista 120 CMS software is
installed.
HL7 communication
The monitor supports HL7 protocol to upload data. Set HL7 Encryption method to Off or TLS
Select Menu > User Maintain and input user (default).
maintain password > Security. In this menu:
To set HL7 IP address of the client-side, select
Set HL7 to On/Off. When it is set to On, set User Maintain > Network > HL7 IP.
Firewall to On to protect against hacker
For more information about HL7 communication,
attacking.
refer to HL7 Communication Protocol Service
Manual.
Alarms
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Alarm category . . . . . . . . . . . . . . . . . . . . . . . . 74
Physiological alarms. . . . . . . . . . . . . . . . . . . . . 74
Technical alarms. . . . . . . . . . . . . . . . . . . . . . . . 74
Prompts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Alarm levels . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Controlling alarm . . . . . . . . . . . . . . . . . . . . . . 76
Setting parameter alarm . . . . . . . . . . . . . . . . . . 76
Audio alarm paused . . . . . . . . . . . . . . . . . . . . . 77
Audio alarm off . . . . . . . . . . . . . . . . . . . . . . . . . 78
Alarm reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Latching alarms . . . . . . . . . . . . . . . . . . . . . . . 79
Testing alarms . . . . . . . . . . . . . . . . . . . . . . . . 80
Alert setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Adjustable range of alert limits . . . . . . . . . . . . . 81
Nurse Call . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Overview
WARNING
A potential hazard can exist if different alarm
presets are used for the same or similar
equipment in different areas, e.g., an intensive
care unit or a cardiac operating room.
Alarm category
Alarm levels
In terms of severity, the device’s alarm levels can A medium level alarm warns the operator of a
be classified into three categories: high level medium priority alarm condition which requires
alarms, medium level alarms and low level alarms. prompt operator response. Failure to respond to
the cause of the alarm condition is likely to
1 High level alarms
result in reversible injury of the patient.
A high level alarm intensively warns the
3 Low level alarms
operator of a high priority alarm condition which
requires immediate operator response. Failure A low level alarm reminds the operator of a low
to respond to the cause of the alarm condition is priority alarm condition which requires
likely to result in death or irreversible injury of response. And the response time for a low
the patient. priority alarm condition can be greater than that
for a medium priority alarm condition. Failure to
2 Medium level alarms
respond to the cause of the alarm condition is
likely to result in discomfort or reversible minor
injury of the patient.
The high/medium/low-level alarms are indicated by
the system in following different ways:
Standard
The sound pressure range for standard audible The parameter area has two flash methods to
alarm signals is from 45 dB to 85 dB. prompt alarms: background flash and text flash.
User can click respective parameter area to enter
When different level alarms occur at the same time,
parameter setup menu > Alarm > Visual Effect:
alarm sound and alarm indicator prompt the highest
level alarm, alarm messages display in turn. 1 Text Flash: text flashes with frequency of 1 Hz.
2 Background Flash: background flashes with
frequency of 1 Hz.
WARNING WARNING
Do not rely exclusively on the audible alarm Ensure the volume is properly set up. When
system for patient monitoring. Adjustment of the sound pressure of audible alarm is below
alarm volume to a low level or off during or equivalent to the ambient noise, it may be
patient monitoring may result in patient difficult for the operator to distinguish the
danger. Remember that the most reliable audio alarm.
method of patient monitoring combines close
personal surveillance with correct operation
of monitoring equipment.
Controlling alarm
Upper arrow or
lower arrow to
increase or
decrease the
alarm limit
Setting value of
low alarm limit
Setting value of
high alarm limit
NOTE
If a new alarm occurs during the audio alarm off
period, the new alarm will not be sounding.
Latching alarms
In monitoring mode, user can enter Menu > User In ward round or spot-checking mode, it is acted as
Maintain > Alarm to set the alarm level of SpO2 prompt and the level is not user-selectable.
Sensor Off to High, Med. or Low. It is set to Low
by default.
WARNING
The deleted alarm events are beyond retrieve.
Use this function cautiously.
Testing alarms
When the user switches the monitor on, the monitor NOTE
will prompt two “Di” tone that means the audio in
The monitor will reboot in 3 seconds due to self-
selftest is normal. The user must check that the
test failure.
alarm indicator light is normal (it lights in red, yellow
and blue in turn during starting up). This indicates
that the visible and auditory alarm indicators are
functioning correctly. For further testing of
individual measurement alarms, perform the
measurement on a subject or use a simulator.
Adjust alarm limits and check that appropriate
alarm behavior is observed.
Alert setup
In ward round mode or spot-checking mode, the Alert information area displays prompts with
monitor provides vital signs alarms (SpO2 No blue background. (Alert information area is the
Pulse, SpO2 Desat), and if physiological physiological alarm area in monitoring mode.)
parameters of the currently monitored patient
NOTE
exceed the predefined limit, the monitor will give
alert information instead of physiological alarms. For TEMP module, TEMP alert can only be used in
predictive mode of quick TEMP and in infrared
Enter the respective parameter setup menu to set TEMP.
alert limit and alert switch.
Alert can be indicated by the system in following NOTE
different ways: Generally, monitor displays SYS alert value. When
User can select Background Flash or Text multiple values are abnormal, alert value will be
Flash in the parameter area. displayed in order of SYS, DIA and MAP.
NOTE NOTE
Prior to monitoring, make sure that the limit Setting limits to extreme values may cause the
settings are appropriate for the patient. alert system to become ineffective. It is
recommended to use the default settings.
TEMP
Patient Type High Limit Low Limit
ADU/PED/NEO +42 ºC (+107.6 ºF) +35.5 ºC (+95.9 ºF)
Nurse Call
NOTE
Before using the function of nurse call, check if it is
functioning normally.
Alarm information
Prompts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
WARNING
During monitoring, the physiological alarms
SpO2 No Pulse and SpO2 Desat are preset to
be on and cannot be turned off.
SYS High SYS measuring value is above upper alarm limit. User-selectable
SYS Low SYS measuring value is below lower alarm limit. User-selectable
DIA High DIA measuring value is above upper alarm limit. User-selectable
DIA Low DIA measuring value is below lower alarm limit. User-selectable
MAP High MAP measuring value is above upper alarm limit. User-selectable
MAP Low MAP measuring value is below lower alarm limit. User-selectable
TEMP
TEMP High Measuring value of TEMP is above upper alarm limit. User-selectable
TEMP Low Measuring value of TEMP is below lower alarm limit. User-selectable
SpO2
High
Stop using measuring
(monitoring
function of SpO2
mode)
SpO2 module failure or module, and notify
SpO2 Comm Fail Low
communication failure biomedical engineer or
(spot-checking
manufacturer’s service
or ward round
staff.
mode)
NIBP
High
Stop using measuring
(monitoring
function of NIBP module,
mode)
NIBP module failure or and notify biomedical
NIBP Comm Fail Low
communication failure. engineer or
(spot-checking or
manufacturer’s service
ward round
staff.
mode)
Check the connections
and the wrapped cuff to
see whether they are all
NIBP Leak NIBP pump, valve, cuff or prepared well.
Low
(Dräger NIBP) tube has a leakage. If failure persists, notify
biomedical engineer or
manufacturer’s service
staff.
NIBP Excessive Pressure has exceeded the Measure again, if failure
Low
Pressure specified upper safety limit. persists, stop measuring
function of NIBP module
and notify biomedical
NIBP Init Pressure The initial pressure is too engineer or
Low
High high during measuring manufacturer’s service
staff.
Notify biomedical
NIBP Aux Pressure has exceeded the
engineer or
Excessive second safety limit as High
manufacturer’s service
Pressure specified.
staff.
NIBP Cuff Type The cuff type used isn’t Confirm the patient type
Error consistent with the patient Low and change the cuff.
type.
Check whether the
Atmospheric pressure or airway is occluded or
NIBP System system pressure is pressure sensor works
Pressure abnormal. The valve is Low properly. If the problem
Abnormality occluded so that deflation is still exists, contact the
failed. service personnel.
NIBP System Contact the service
Failure NIBP is not calibrated. High
personnel.
Use other methods to
Cuff is too loose or patient
NIBP Weak Signal Low measure blood
pulse is too weak.
pressure.
TAT-5000S-RS232 Thermometer
High
Stop using measuring
(monitoring
function of TEMP
mode)
Temp Comm Fail TEMP module failure or module; notify biomedical
Low
communication failure. engineer or
(spot-checking or
Manufacturer’s service
ward round
staff.
mode)
F3000 TEMP
High
Stop using measuring
(monitoring
function of TEMP
mode)
TEMP module failure or module; notify
Temp COMM Fail Low
communication failure. biomedical engineer or
(spot-checking or
Manufacturer’s service
ward round
staff.
mode)
Others
High
(monitoring
mode)
Change the battery or
Battery Low Battery Low Low
charging.
(spot-checking or
ward round
mode)
Recorder Out Of
Recorder Out Of Paper Low Install the paper.
Paper
Prompts
Message Cause
SpO2 Search Pulse SpO2 module is analyzing the patient signal and searching for the
pulse to compute the saturation, when sensor is connected with
patient.
Message Cause
SpO2 Noisy Signal There is interference with SpO2 measurement signals due to patient
movement, ambient light, electrical interference or else. (Nellcor
SpO2)
Manual Measuring In manual measuring mode.
Continual Measuring In continuous measuring mode.
Auto Measuring In automatic measuring mode.
AVG Measuring In average measuring mode.
Measurem. Canceled Press the “Start/stop NIBP measurement” button to stop the
measurement.
Calibrating During calibrating.
Calibrat. Canceled Calibration is over.
Leak. Test Running The leakage test is in process.
Leak.Test Canceled Pneumatic test over.
Resetting NIBP module in resetting.
Please Start NIBP module is in idle status.
Done NIBP measurement is completed.
Measure time out No measuring result after the module entering Predict state for 30 s.
TEMP Is Warming Up User takes the sensor out of the bracket and TEMP is warming up
Warm-up Over The monitor displays this message after taking the sensor out of the
bracket and warm-up is over.
Predict Over After the Predict measuring is over, the data and message display
on the interface.
Incomplete parameter input, In Warning-Score System interface, parameters are not completely
unable to score input.
No WIFI module detected No Wi-Fi module is detected.
Quick Predict Over Quick prediction measurement is completed.
PlaceProbeOnMeasurePlace Probe isn’t placed on the measurement site.
SpO2 alarm limits are listed as follows (unit %): TEMP alarm limits are listed as follows:
Adjustable Range Patient Type Adjustable Range
SpO2 20 to 100 +35.5 °C (+95.9 ºF) ~ +42 °C
ADU/PED/NEO
(+107.6 ºF)
SpO2 Desat Limits are listed as follows (unit %):
Adjustable Range
SpO2 Desat Limit 20 to 99
NOTE
User can set the range through User Maintain >
Alarm > SpO2 Desat Limit. SpO2 Desat Limit
should be less than or equal to SpO2 alarm low
limit.
User Interface
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Overview
The user can select the display parameters based Exit the menu and the screen will adjust the
on the monitoring and measurement requirements. parameters automatically.
Click Menu > Parameters Setup to select the
required parameters from the popup interface >
Confirm.
NOTE
The colors won’t be affected by the default factory
Settings.
Monitoring SpO2
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Measuring SpO2 . . . . . . . . . . . . . . . . . . . . . . . 99
Measurement Procedure . . . . . . . . . . . . . . 99
Overview
WARNING WARNING
Do not use the SpO2 sensors if the packaging Neonate SpO2 sensor can only be used when
or the sensor is damaged and return them to required, no more than 20 min at a time.
the vendor.
WARNING
WARNING Use only Dräger permitted sensors and
If the SpO2 sensor does not work properly, extension cables with the oximeter. Other
reconnect the sensor or use a new one. sensors or extension cables may cause
improper monitor performance and/or minor
WARNING personal injury.
Correct and proper sensor application: if the
sensor is too loose, it might compromise the WARNING
optical alignment, and even cause the sensor High oxygen levels may predispose a
to fall off. If the sensor is too tight, (such as premature infant to retrolental fibroplasia. If
the application site is too large or becomes this is a consideration do NOT set the high
too large due to edema), excessive pressure alarm limit to 100%, which is equivalent to
and local tissue ischemia, hypoxia and lack of switching the high limit alarm off.
nutrition may occur on the application site.
Prolonged and continuous monitoring may WARNING
increase the risk of skin irritations or When serious arrhythmia is present, the SpO2
lacerations. To avoid these damages, users pulse rate may differ from ECG heart rate but
should periodically check surrounding skin of this does not indicate an inaccurate PR (SpO2)
application site according to the patient's value.
condition and pressure sensor, inspect the if
there is sign of stress-related damage in WARNING
surrounding tissue, and regularly change the
Misapplied sensor or sensor that becomes
application site. For the patients whose fluid is
partially dislodged may cause either over or
easy to transfer and/or the patients with
under reading of actual arterial oxygen
systemic or localized edema, users should
saturation.
inspect the skin and change application site
more frequently.
NOTE NOTE
Avoid placing the sensor on extremities with an Functional tester or simulator can not be used to
arterial catheter, intravascular venous infusion assess the SpO2 accuracy. However, it can be
line, or inflated NIBP cuff. When measuring SpO2 used to demonstrate that a particular monitor
on the limb with inflated NIBP cuff, turn on the reproduces a calibration curve that has been
NIBP Simul function. independently demonstrated to meet a particular
accuracy.
NOTE
When a trend toward patient deoxygenation is NOTE
indicated, analyze the blood samples with a If the surrounding temperature increases, the
laboratory co-oximeter to completely understand operator should pay attention to the site of poor
the patient’s condition. perfusion, and increase the frequency of checking
the skin and changing the measurement site to
NOTE prevent burns. If the initial skin temperature is less
than 35 °C, the temperature of all the listed
SpO2 waveform is not directly proportional to the
sensors on the skin will not exceed 41 °C during
pulse volume.
working.
NOTE
NOTE
The device is calibrated to display functional
The cumulative use time for SpO2 sensor in a
oxygen saturation.
single patient should be less than 30 days.
Measuring SpO2
1 Select the correct patient category setting Before Applying the Sensor
(adult/pediatric and neonatal), as this is used to
Be sure to understand all warnings listed in the
optimize the calculation of the SpO2 and pulse
previous section before applying any sensor to a
numerics.
patient. Also,check the sensor as follows:
2 During measurement, ensure that the
Check the sensor outside and inside. To inspect
application site:
the inside, gently open the sensor cavity and
has a pulsatile flow, ideally with a good check splits on or next to the transparent
circulation perfusion. silicone that covers the optical elements.
has not changed in its thickness, causing an Any sensor showing signs of damage or
improper fit of the sensor. alteration must not be used for further patient
monitoring; instead, dispose of it using proper
disposal procedures.
Measurement Procedure
WARNING
Inspect the application site every two to three
hours to ensure skin quality and correct
optical alignment. If the skin quality changes,
move the sensor to another site. Change the
application site at least every four hours. For
neonate, change the measuring site every
20 minutes.
Measurement limitations
The user can check the quality of the pleth wave NOTE
and the stability of the SpO2 values to assess
The SpO2 accuracy has been validated in
whether the sensor functions properly and whether
controlled human studies against arterial blood
the SpO2 readings are valid. Always use these two
sample reference measured with a CO-oximeter.
indications simultaneously to assess the validity of
SpO2 measurements are statistically distributed,
a SpO2 reading.
only about two-thirds of the measurements can be
Generally, the quality of the SpO2 pleth wave expected to fall within the specified accuracy
reflects the quality of the light signals obtained by compared to CO-oximeter measurements. The
the sensor. A wave of poor quality manifests a volunteer population in the studies are composed
decline of the signal validity. On the other hand, the of healthy men and women from age 19 to 37 (for
stability of the SpO2 values also reflects the signal Dräger SpO2 module), from 18 to 50 (for Nellcor
quality. Different from varying SpO2 readings SpO2 module), with various skin pigmentations.
caused by physiological factors, unstable SpO2 Note that the study population was healthy adults
readings are resulted from the sensor’s receiving and not in the actual intended use population.
signals with interference. The problems mentioned
above may be caused by patient movement, wrong NOTE
sensor placement or sensor malfunction. To obtain The pulse rate accuracy is obtained by comparison
valid SpO2 readings, try to limit patient movement, to the pulse rate generated with an arterial oxygen
check the placement of the sensor, measure simulator (also an electronic pulse simulator).
another site or replace the sensor.
NOTE
During monitoring, if the monitor's reading differs
significantly from the patient's physiological
condition, it indicates that the signal may be
disturbed, resulting in an inaccurate reading. In
this case, the artifact can disguise as a similar
reading, causing the monitor to fail to send an
alarm. In order to ensure reliable monitoring, it is
necessary to regularly check whether the sensor is
wearing properly and the signal quality is good.
* Only applicable to the Dräger SpO2 module. PI is indicated by a value ranging from 0 to 10. The
bigger the value is, the better the perfusion and the
PI is a numeric value indicating perfusion level. It
signal quality will be. The perfusion level and the
reflects the perfusion level at the monitoring site.
signal quality are at their maximum when the value
As the measurement of SpO2 is based on the reaches 10. When PI is below 2, it indicates the low
pulsation caused by the blood flow through the perfusion and the poor signal quality at the monitoring
vessel, PI is in relation to the strength of the pulse. site; the user needs to reposition the sensor or find a
Also, the user can use PI as a signal quality indicator better site.
for the measurement of SpO2.
The PI value will be displayed in the SpO2 parameter
area.
Setting sensitivity
The different sensitivity indicates different refresh 1 Select the SpO2 Setup menu;
frequency. High indicates the refresh frequency of
2 Select Sensitivity on the interface and select
SpO2 value is the most frequent. To change the
the desired sensitivity from the popup list.
sensitivity, follow the steps:
* Only applicable to the Nellcor SpO2 module. With the SatSeconds technique, upper and lower
* Not applicable to Vista 120 CMS. SpO2 alarm limits are set in the same way as
traditional alarm management. However, the user can
also set a SatSeconds limit that allows monitoring of
Describing SatSeconds SpO2 below the selected lower alarm limit and above
the selected upper alarm limit for a period of time
With traditional alarm management, upper and lower before an alarm is triggered.
alarm limits are set for monitoring oxygen saturation. The method of calculation is as follows:
During monitoring, as soon as an alarm limit is
violated by as little as one percentage point, an alarm The number of percentage points that the SpO2 falls
is immediately triggered. When the SpO2 level outside the alarm limit is multiplied by the number of
fluctuates near an alarm limit, the alarm is triggered seconds that the SpO2 level remains outside that
each time the limit is violated. Such frequent alarms limit. This can be stated as an equation:
can be distracting. Points × Seconds = SatSeconds
Where:
Points = SpO2 percentage points outside of the limit
Seconds = number of seconds that SpO2 remains at
that point outside of the limit
50 SatSeconds
Alarm Point
SECONDS
Alarm Response with SatSeconds
Monitoring PR
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
PR source . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Overview
PR source
Setting PR volume
Monitoring NIBP
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Measuring PR . . . . . . . . . . . . . . . . . . . . . . . . . 117
Overview
This monitor uses the oscillometric method for The blood pressure measurements determined
measuring NIBP. It can be used for adult, pediatric with this device comply with the American National
and neonatal patients. It is also intended for use Standard for Electronic or Automated
with pregnant, including pre-eclamptic patients. Sphygmomanometers (ISO 81060-2:2013) in
relation to mean error and standard deviation. In
Oscillometric devices measure the amplitude of
clinical investigation method with a reference
pressure changes in the occluding cuff as the cuff
sphygmomanometer, the fifth Korotkoff sound was
deflates from above systolic pressure. The
used to determine adult diastolic pressure, and the
amplitude suddenly increases as the pulse breaks
fourth Korotkoff sound was used to determine
through the occlusion in the artery. As the cuff
pediatric diastolic pressure.
pressure decreases further, the pulsations increase
in amplitude, reach a maximum (which The invasive blood pressure is used to determine
approximates to the mean pressure), and then the neonate pressure in clinical investigation, and
diminish. the arterial reference sites include umbilical artery,
arteria cruralis, axillary artery, brachial artery,
dorsalis pedis, and radial artery.
WARNING WARNING
Do not measure NIBP on patients with sickle- Do not apply the cuff to a limb that has an
cell disease or any condition where skin intravenous infusion or catheter in place. This
damage has occurred or is expected. could cause tissue damage around the
catheter when infusion is slowed or blocked
WARNING during cuff inflation.
Do not measure NIBP on the arm of the same
side with a mastectomy. WARNING
Do not attach the cuff to a limb being used for
WARNING IV infusions as the cuff inflation can block the
Use clinical judgment to decide whether to infusion, potentially causing harm to patient.
perform frequent blood pressure
measurements on patients with severe blood WARNING
clotting disorders because of the risk of Do not apply the cuff to a limb where
hematoma in the limb fitted with the cuff. intravascular access or therapy, or an arterio-
venous (A-V) shunt is present, otherwise, it
may result in injury to the patient.
WARNING NOTE
Ensure that the correct patient type is selected If the user spills liquid onto the equipment or
before performing measurements. Do not accessories, particularly if there is a chance that it
apply the higher adult inflation, overpressure can get inside the tubing or the measurement
limits and measurement duration for neonatal device, contact the service personnel.
patients. Not using the neonate mode on a
neonatal patient can block the blood flow, NOTE
potentially causing harm to the patient. Continuous use of the automatic measuring mode
for short intervals may lead to the discomfort of
WARNING patient. Continuous measuring and automatic
Make sure that the air tubing connecting the measuring in neonatal or pediatric mode may
blood pressure cuff and the monitor is neither result in tissue damage or ischemia to the patient.
blocked nor tangled.
NOTE
WARNING NIBP measurement can be affected by extremes
Measuring of blood pressure can temporarily of temperature, humidity and altitude.
cause malfunctioning of other medical
monitoring devices on the same limb. NOTE
NIBP measurement value should be explained by
WARNING qualified professionals.
NIBP readings can be affected by the
measurement site, the position of the patient, NOTE
exercise, or the patient's physiologic The pulse rate based on the NIBP measurement
conditions. may differ from the heart rate based on the ECG
waveform. NIBP measures the number of
WARNING peripheral pulse pulsations, and the heart rate is
Continuous cuff pressure due to connection measured by the electrical signal of the heart.
tubing kinking can block the blood flow, and When the electrical signals of the heart
may result in injury to the patient. occasionally fail to cause the peripheral blood
vessels to pulse or the patient's peripheral
WARNING perfusion is poor, the difference happens.
Verifying the calibration is only applicable for
adults, and it cannot be operated in automatic NOTE
measuring interval. Continuous measuring The cumulative use time for NIBP cuff in a single
cannot be operated in automatic measuring patient should be less than 30 days.
interval either.
NOTE
It is suggested that the user should not start NIBP
measuring when the low battery displays, or the
monitor may be turned off automatically.
Measurement limitations
Measurements are impossible with pulse rate Patients with excessive and continuous
extremes of less than 40 bpm or greater than movement such as shivering or convulsions.
240 bpm, or if the patient is on a heart-lung
Patients with rapid blood pressure changes.
machine.
Patients with severe shock or hypothermia that
The measurement may be inaccurate or impossible
reduces blood flow to the peripheries.
in the following situations:
Patients with obesity, where a thick layer of fat
A regular arterial pressure pulse is hard to
surrounding a limb dampens the oscillations
detect.
coming from the artery.
Patients with cardiac arrhythmias.
Patients on an edematous extremity.
Measurement methods
Measurement procedures
To obtain accurate measurements, the following 4 Select a measurement mode and unit (mmHg,
operating steps need to be observed: cmH2O or kPa, 1 mmHg = 0.133 kPa,
1 mmHg=1.36 cmH2O) in the NIBP Setup
1 Ensure the patient position in normal use,
menu. Refer to section Operation Prompts for
including
details.
Comfortably seated or lie flat, legs
uncrossed;
5 Press the button on the front panel to
Feet flat on the floor start a measurement.
Back and arm supported 6 Wait until the first reading is taken.
Relax as much as possible, neither talking
nor applying external pressure against the NOTE
cuff. Rest for five minutes in a quiet The width of the cuff is either approximately 40%
environment. of the limb circumference or 2/3 of the upper arm
length. The inflatable part of the cuff should be
2 Connect the air hose and switch on the monitor. long enough to encircle 80-100% of the limb. The
Apply the blood pressure cuff to the patient's wrong size of cuff can cause erroneous readings.
arm and follow the instructions below. Ensure If the cuff size is in question, use another cuff with
that the cuff is completely deflated. suitable size to avoid errors.
Apply the appropriate size cuff to the patient
(About the cuff size selection, refer to Section NOTE
NIBP accessories), and make sure that the If an NIBP measurement is suspect, repeat the
symbol "Φ" is over the artery. Ensure that measurement. If the user is still uncertain about
middle of the cuff is at the level of the right the reading, use another method to measure the
atrium of the heart and the cuff is not wrapped blood pressure.
too tightly around the limb. Excessive tightness
may cause discoloration and eventual ischemia NOTE
of the extremity.
Make sure the cuff is well connected. A leak of air
may cause measurement error.
NOTE
Select the cuff with the suitable size. An unsuitable
cuff may cause incorrect measurements.
NOTE
Avoid incursion of liquid into the cuff. If this
happens, desiccate the cuff completely.
Cuff Usage
NOTE
3 Check whether the patient type is appropriately NIBP parameter area can display real-time cuff
selected. Clicking patient information area or pressure value till displaying SYS value.
pressing Switch Patient Type hardkey can
change patient Type.
1 Manual Measurement
Access the NIBP Setup > Function menu and
key directly to start measurement. During
set the Measure Mode item to Manual. Then measurement:
Invalid value will not be used for counting
press the button on the front panel to
average. Invalid value every measurement will
start a manual measurement. During
result in invalid average;
measurement, pressing this button at any time
can stop measurement. There will have sounding of DU… after entire
measurement is finished. At that time, a final
2 Automatic Measurement
average value can be displayed. Meanwhile,
Access the NIBP Setup > Function menu and alert information of the last measurement and
set the Measure Mode item to AUTO, select technical alarms of the last measurement will be
cleared, and alert information can work
time interval as need, then press the normally. The value for each measurement and
button on the front panel. During measurement: final average value will be stored in ward round
records/spot-checking trend table which uses
If measurement is failed or monitor receives the icon * to mark the average value;
command for suspending, measurement will be
stopped. The next measurement will be In specified interval, if interval and times have
continued after interval. If alarms of NIBP been changed again, the whole average
Excessive Pressure or NIBP Aux Excessive measurement will be ended and measurement
Pressure are present, the whole measurement times will be zero;
will be stopped. Manually operation is Other operations, such as changing patient
necessary for a new auto measurement. type, entering DEMO or standby mode and so
During interval, if interval is changed, after on, can stop the measurement. Meanwhile,
ending of new interval countdown, monitor will parameter interface will be updated;
start measurement automatically. At the end of the last measurement, if user
During interval, if user manually starts the presses NIBP measurement key before monitor
measurement, the next auto measurement will gets average value with DU… sounding, a new
start after the ending of countdown which starts round of average measurement will be started
from the latest manual measurement; During with prompt NIBP AVG-1 (n) and NIBP average
interval, if user makes a continuous value of the last round will not be displayed. '-1
measurement, the whole measurement will be (n)' means the first measurement of N times.
over on the basis of continuous measurement. 5 Stopping Measurement
3 Continuous Measurement
During measurement, press the button on
the front panel at any time to stop measurement.
Resetting NIBP
Available for Dräger module only. When the pressure does not work properly and the
system fails to give a message for the problem, pick
Reset in the User Maintain > NIBP menu to
activate self-test procedure, and thus restore the
system from abnormal performance.
Calibrating NIBP
NIBP is not user-calibrated. Cuff-pressure qualified service professional. See the Service
transducers must be verified and calibrated, if Manual for details.
necessary, at least once every two years by a
Leakage test
Monitor Cylinder
To change the inflation value: If other inflation values are chosen, the
preset value by users will be adopted as the
1 Select NIBP Setup > Function > Inflation
inflation value when measuring blood
value;
pressure.
2 Choose AUTO or other inflation values from the
If AUTO is chosen, the default value will be
pull-down list.
adopted as the inflation value when
measuring blood pressure.
Measuring PR
Access NIBP Setup > Function, and set PR to On Monitor will not record PR from NIBP;
or Off.
PR from NIBP will be displayed in NIBP review
The NIBP parameter area displays PR from table but not in other review tables.
NIBP;
Monitoring TEMP
Introduction CAUTION
Always store this thermometer in a clean, dry
The Exergen Temporal Scanner, Model TAT-
place where it will not become excessively cold
5000S-RS232 Thermometer has the capability of (-4 °F/-20 °C), or hot (122 °F/50 °C) or humid
taking Temporal Artery (TA) Temperature in about (max RH 93% non-condensing, at 50 to 106 kPa).
3 seconds. The Exergen Temp scanner has the
ability to measure the patient’s Arterial temperature
in °F or °C. CAUTION
The thermometer is not shockproof. Do not drop it
Vista 120 SC monitor with Exergen Temp scanner or expose it to electrical shocks.
is indicated for use in adult, pediatric and neonatal
patient populations.
CAUTION
WARNING Do not Autoclave. Note cleaning and sterilizing
No modification of this equipment is allowed. procedures in this manual.
WARNING CAUTION
Do not use this thermometer if it is not In addition to the battery, there are no other parts
working properly, if it has been exposed to available to the user for maintenance. When the
temperature extremes, damaged, been battery power is low, the user shall replace the
subject to electrical shocks or immersed in battery according to the instructions in this
water. manual. For service, repair or adjustments, return
the thermometer to the manufacturer.
WARNING
CAUTION
Not suitable for use in the presence of
flammable anesthetic mixtures. Never drop or insert any object into any opening,
unless stated in this manual.
WARNING
CAUTION
Do not take temperature over scar tissue,
open sores or abrasions. If the thermometer is not used regularly, remove
the battery to prevent possible damage due to
chemical leakage.
CAUTION
Use this product only for its intended use as
CAUTION
described in this manual.
Disposal of used Exergen scanner must be
performed in accordance with current medical
CAUTION
practices or local regulations regarding disposal
The operating environmental temperature range of infectious, biological medical waste.
for this product is 60 to 104 °F (15.5 to 40 °C).
CAUTION
Follow the battery manufacturer’s
recommendations or the hospital policy for the
disposal of used batteries.
Sensor Lens
Metal Probe Cone
ON Button
LED Display
Battery
Compartment
Access Door
Front Rear
Using the Exergen Temp Scanner Clicking: Each fast click indicates a rise to a
higher temperature, similar to a radar detector.
To Scan: Depress the ON button. The
Slow clicking indicates that the thermometer is
thermometer will continually scan for the
still scanning, but not finding any higher
highest temperature (peak) as long as the ON
temperature.
button is depressed.
To Retain or Lock Reading: The reading will Factors That May Affect Measurement
remain on the LED display for 30 seconds after Accuracy
button is released. If measuring room
Sweating
temperature, the temperature will remain on the
LED display for only 5 seconds. When a fever resolves, the body may bring its
temperature down by sweating. The temp scanner
To Restart: Depress the ON button to restart. It
detects this reduction in temperature immediately.
is not necessary to wait until the LED display is
However, sweating also causes extra cooling of the
clear, the thermometer will immediately begin a
skin. As a result the reading given by the temp
new scan each time the ON button is
scanner may be low. The user should therefore
depressed.
either wait until the sweating has stopped (wiping of
Alternate sites when temporal artery or behind the forehead is not recommended, since the
ear are unavailable sweating immediately begins again), or use the
following method:
Femoral artery: slowly slide the scanner
across groin. 1 Scan the temperature as normal, keeping the
button depressed;
Lateral thoracic artery: slowly scan side-to-
side in the area - midway between the axilla and 2 Gently nestle the temp scanner on the neck
the nipple. directly behind the ear lobe;
3 Release the button and read the temperature.
Measuring Procedure The artery behind the ear lobe does not provide a
sufficiently accurate reading. However, this area is
Measure only the side of the head exposed to the less affected by sweating than the forehead. In
environment. Anything covering the area to be addition, during sweating, increased blood flow
measured (hair, hat, wig, bandages, etc.) would produces higher skin temperatures, equivalent to
insulate the area, resulting in falsely high readings. TA temperature, resulting in a good reflection of
body temperature.
Slide the thermometer straight across the forehead,
not down the side of the face. Midline on the Environmental Effects
forehead, the TA is about a millimeter below the The temp scanner measures the temperature of the
skin, whereas at the side of the face, the TA is much surrounding environment. For this measurement to
deeper, and measuring there would result in falsely be accurate, it needs to have become acclimated to
low readings. the temperature of the room in which it is to be
Wait about 30 seconds before measuring the same used.
person again to avoid excessive cooling of the skin.
CAUTION
An infant is frequently swaddled in blankets and
Allow the temp scanner to acclimate for at least
clothing covering the neck area. Unless visible
30 minutes before use if it is taken from a cold
sweating, one measurement at the TA area is
room into a hot room, or vice versa. Avoid holding
typically all that is required. Should the user feel the
the temp scanner by the probe as it will mistake
temperature is low, then push aside any clothing or
the temperature of hand for that of the room.
blankets covering the neck area for about 30
seconds or so, and repeat the measurement on the Movement
neck behind the ear.
If the patient is agitated, or squirms away before
completing the measurement, just keep the button
depressed and continue the measurement without
having to wait.
Alternate site:
If the temporal artery is covered, then the area
behind the ear, if exposed, can be an alternate
site.
Type of Abnormal
Possible Cause Action Taken
Temperature
Dirty Lens. Clean lens of scanner every two weeks.
Releasing the button before finished Release the button after finished
measuring. measuring.
Measuring when an ice pack or wet Remove ice pack or wet compress, wait
compress is on the forehead. 2 minutes, and re-take temperature.
Complete diaphoresis includes
diaphoresis of area behind the ear and
Abnormally low
suggests that the temperature is rapidly
temperature
Measuring a completely diaphoretic dropping. Use an alternative method of
patient. temperature measurement in these
cases until the patient is dry and the
temporal artery measurement can be
repeated.
Scan straight across forehead. The
Improperly scanning down the side
temporal artery is closest to skin in that
of the face.
area.
Confirm measurement site has not
recently been in contact with heat
Anything covering the area to be
insulators such as hats, blankets, and
Abnormally high measured would insulate and
hair. Scan the area not covered or wait
temperature prevent heat from dissipating,
about 30 seconds for the previously
resulting in false high readings.
covered area to equilibrate to the
environment.
Setting Exergen Temp scanner unit 1 Insert a pointed object (e.g., the end of
straightened paperclip) into the hole on the right
NOTE rear of the temp scanner (looking at the back of
the temp scanner).
If the monitor and Exergen Temp scanner units are
each set to a different unit (e.g., one °F and the 2 Pushing in and release the battery
other °C), the monitor will automatically switch to compartment access door.
the Temp units set in the Exergen Temp scanner, 3 Remove the battery compartment access door
once a valid temperature measurement has been from the temp scanner.
completed.
4 Remove the 9 V battery from the temp scanner
Follow this procedure to change the temp scanner battery compartment - There is no need to
Temp LED display units: disconnect the battery from the temp scanner.
5 Locate the F/C switch, and slide to desired
display temp units: °C or °F (marked on the
switch).
6 Reinstall the 9 V battery back into the temp 7 Reinstall the battery compartment access door.
scanner battery compartment.
F/C Switch
F/C Switch
Replacing Exergen Temp scanner’s 4 Remove the 9 V battery from the battery
battery compartment of the temp scanner.
5 Disconnect the battery from the battery
When the Exergen temp scanner battery fails to harness.
hold a charge it will need to be replaced or if the
Exergen temp scanner will not be used regularly, CAUTION
remove the battery to prevent possible damage due If the battery leaks, remove it carefully. Do not
to chemical leakage. allow bare skin to touch leaking fluids.
Follow this procedure to remove/replace the
battery: CAUTION
1 Insert a pointed object (e.g., the end of Properly dispose of used batteries. Do not wrap
straightened paperclip) into the hole on the right them in metal or aluminum foil. Wrap them in
rear of the temp scanner (looking at the back of paper before disposing of them, batteries may
the temp scanner). explode if overheated.
2 Pushing in and release the plastic battery 6 Connect a new 9 V battery to the battery
compartment door. harness, noting the proper polarity of the 9 V
3 Remove the plastic battery compartment door battery pins.
from the temp scanner. 7 Install a new 9 V battery into the temp scanner
battery compartment.
8 Position the battery compartment door over the 9 Gently push down on the compartment door
battery compartment by aligning the tabs with until the door is fully seated into the temp
the slots at the bottom of the battery scanner’s recess. A click should be heard when
compartment. the latch is engaged.
NOTE
Use only high quality alkaline batteries in the
Exergen temp scanner.
Introduction WARNING
Use probe covers supplied by the
The monitor with the F3000 module measures
manufacturer with this thermometer only. Use
patient temperatures by oral, axillary or rectal
of any other probe cover will result in
means.
erroneous temperature readings.
The monitor can only measure temperature of adult
and pediatric patients. If the user measure WARNING
temperature of neonate patient, the monitor will not The thermometer and probe covers are Non-
display data. sterile. Do not use on abraded tissue.
The electromagnetic compatibility of this device
has been verified by test according to the WARNING
EN60601-1-2: 2015 requirements. To limit cross contamination, use Blue
devices for Oral and Axillary temperature
WARNING
taking only.
Do not use this device near flammable
anesthetics. Not suitable for use in the
WARNING
presence of flammable anesthetic mixture
with air, oxygen or nitrous oxide. Use RED devices only for RECTAL
temperatures.
WARNING
WARNING
Do not use this thermometer without first
installing a new probe cover. Thoroughly dry all electrical contacts on both
probe and thermometer after washing, or
device may fail to function properly.
WARNING
Do not reuse the disposable probe covers.
WARNING
For re-calibration, service or integrity checks,
refer to a qualified Biomedical Technician or
return to the manufacturer.
WARNING NOTE
Do not open the F3000 module. No user- Even though this device has been designed to
serviceable parts inside. Opening of the minimize the effects of electromagnetic
module may affect calibration and voids interference, it does generate radio frequency
warranty. energy. If not used in accordance with the
instructions, the device could cause interference in
WARNING other equipment operating within its vicinity. If the
device is causing interference, the following
Disposal of used probe covers must be
actions may be taken in an attempt to correct the
performed in accordance with current medical
interference:
practices or local regulations regarding
disposal of infectious, biological medical Re-orient or re-locate the receiving device.
waste.
Increase the separation between the devices.
NOTE
To aid infection control, never switch boxes
between blue and red isolation chambers. Also,
never switch probes between blue and red
isolation chambers. Keep similar colors together.
3 Remove probe from the probe well. This 6 To change probes, remove the isolation
automatically turns on the thermometer. chamber as described previously.
4 To help remind the user to apply or remove a 7 Grasp the sides of the L-shaped connector
probe cover, a probe icon with flashing probe piece with one hand and then using other hand
cover will be displayed when the probe is pull backward on the latch holding the end of
withdrawn from the probe well and following a the L-shaped connector.
completed temperature measurement.
8 Once free of the latch, slide the L-shaped
5 Insert the probe end into a cover in the box. connector out of isolation chamber.
Push the handle firmly until the cover snaps into
9 To replace, properly align the top of the L-
place.
shaped connector to the slot on the back of the
6 Take appropriate temperature measurement isolation chamber.
(oral, axillary or rectal).
10 Then slide the connector up into the slot
7 Eject the used cover into bio-waste container by pressing firmly on the bottom of the connector
pressing top button. until it “clicks” into place.
8 Remove, discard and replace box when empty.
Measuring mode
Changing isolation chambers and
probes Predictive Mode
When MEASURE MODE is set to PREDICT, the
NOTE monitor operates in Predictive Mode to provide fast
For aiding in infection control, use only the Blue and accurate temperature measurements.
probe and Blue isolation chamber for Oral and
Axillary temperature taking. The Red probe and Quick Predictive Mode
Red isolation chamber must only be used for rectal
temperature taking. When MEASURE MODE is set to QUICK
PREDICT, the monitor operates in Quick Predictive
NOTE Mode which is an oral predictive measurement
mode intended for situations where fast
Do not attach a Red probe to a Blue isolation
temperature measurements are desired.
chamber or vice-versa.
Quick Predictive Mode allows clinicians to rapidly
1 To remove or replace any isolation identify patients with “normal” body temperatures. If
chamber/probe assembly, grasp the isolation the patient temperature is outside of the “normal”
chamber from each side as shown. range, the monitor will automatically switch into its
2 Squeeze inward releasing the snaps and slide standard predictive mode to provide a more
the isolation chamber up to pull off. accurate reading.
3 To replace, align probe well finger with opening Quick Predictive Mode is not available when in Low
in the top of the unit. Temp. Mode.
8 If three short beeps are heard, it means the unit TEMP setup for F3000 module
switched to Monitoring Mode for this
temperature only. In TEMP SETUP interface, the following settings
9 Two beeps are sounded when measurement is are available:
complete and the final temperature is displayed. MEASURE MODE: Set the measuring mode to
10 Eject the used cover into a bio-waste container PREDICT, QUICK PREDICT or MONITOR.
by pushing top button. MEASURE POS: Set the measuring position to
ORAL, AXILLARY or RECTAL.
Rectal Temperature Taking Low Temp. Mode: Activate /deactivate the Low
1 Make certain that the Red isolation Temp. Mode by setting it to On/Off.
chamber/probe unit is attached.
NOTE
2 Withdraw the probe and apply a probe cover. The QUICK PREDICT mode is for oral
Thermometer turns on automatically a beep will measurement only.
be heard when the probe completes warm-up.
3 Apply lubrication if desired. NOTE
Low Temp. Mode can be set to On only when
4 Insert the probe into the patient’s rectum. To
measure mode is PREDICT.
ensure proper tissue contact, angle the probe
slightly after insertion.
NOTE
5 Depth of insertion is recommended at 1/2” to Make sure all settings of TEMP Setup are properly
3/4” (12 mm ~ 19 mm) for adults and 1/4” to 1/2” set up every time before the user withdraws the
(6 mm ~ 13 mm) for children. probe from the probe well. If the user modifies the
6 If three short beeps are heard, it means the unit settings immediately a measurement is
switched to Monitoring Mode for this completed, the new settings will be effective for the
temperature only. next measurement.
7 Two beeps are sounded when measurement is
complete and the final temperature is displayed.
8 Eject the used cover into a bio-waste container
by pushing top button.
NOTE
Probe movement during a measurement can
affect the thermometer’s ability to measure the site
temperature and may lengthen the time required to
obtain a reading.
NOTE
If a beep is not heard 10 seconds after withdrawing
the probe from the probe well and starting
temperature measurement in Predictive Mode or
Quick Predictive Mode, check the physical
connection of the F3000 module.
Warning-Score system
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134
Overview
Warning-Score interface
Warning-Score method
Warning-Score criteria
Value
3 2 1 0 1 2 3
HR (bpm) ≤40 41~50 51~100 101~110 111~129 ≥130
SYS (mmhg) ≤70 71~80 81~100 101~199 ≥200
RESP (rpm) ≤8 9~20 21~24 25~29 ≥30
TEMP (°C) ≤35 35.1~36.0 36.1~37.9 38~38.9 ≥39
SpO2 (%) <86 86~91 92~93 ≥94
Consciousness NC A V P U
Value
3 2 1 0 1 2 3
RESP (rpm) ≤8 9~11 12~20 21~24 ≥25
SpO2 (%) ≤91 92~93 94~95 ≥96
TEMP (°C) ≤35.0 35.1~36.0 36.1~38.0 38.1~39.0 ≥39.1
SYS (mmHg) ≤90 91~100 101~110 111~219 ≥220
HR (bpm) ≤40 41~50 51~90 91~110 111~130 ≥131
Consciousness A V/P/U
Oxygen Yes No
Value
3 2 1 0 1 2 3
HR (bpm) ≤40 41~50 51~100 101~110 111~129 ≥130
SYS (mmhg) ≤70 71~80 81~100 101~199 ≥200
RESP (rpm) <9 9~14 15~20 21~29 ≥30
TEMP (°C) <35.0 35.0~38.4 ≥38.5
Consciousness A V P U
Age <50 50~70 >70
Warning-Score result
Trend table provides the monitored patient’s scores in Warning-Score interface. EWS, NEWS and
during a period of time; it includes score time, score MEWS can respectively support 1200 groups of
parameters and value, EWS/NEWS/MEWS score. trend review at least.
To check the trend table, click Trend Table button
To configure the storage device, select Menu > After configuring the appropriate storage device,
System Setup > Storage > Storage Device, and click exit. If the storage device is successfully
choose the storage medium from the pop-up list as starting data storing, the monitor will be indicated
desired.
by the symbol . If there is no enough space in
In ward round or spot-checking mode, Storage storage device, or the storage device is read-
Device is fixed as Internal Storage Device.
only/damaged, the symbol will be displayed.
NOTE
In ward round or spot-checking mode, if storage The monitor will stop storing data in the storage
space of the system is <200 M, a prompt will be device under the following circumstances:
given out. In order to store data normally, please There is no enough space in the storage device
clean the space in time. for storing data.
CAUTION
It is recommended to format the USB flash drive
to the FAT file type via PC prior to use.
In monitoring mode, to export data of one Choose the patient from the list after selecting
patient from the internal storage device to the Menu > Review > History Patient, click Export
removable device: Data on the Review menu, and then input User
Maintain password.
To export data of all patients, select Menu > In ward round or spot-checking mode, User
Review > History Patient, click Export all data, Maintain password is necessary for exporting
and then input User Maintain password. ward round records/spot-checking data in the
specified time to the removable device , and
NOTE also to the internal storage device. If no
In monitoring mode, the data in internal storage removable device is used, the data will be
device is not encrypted. Protect privacy exported to internal storage device directly.
information.
Choose Menu > Review > Export, and confirm the
operation as needed.
In ward round and spot-checking mode, the formats
of exported CSV data file names are ‘XXX- Round-
Records’ and ‘XXX-Spot-Review.csv’ respectively,
in which the XXX means export time.
NOTE
No need to restart the monitor after formatting is
successful. The internal storage device can be
identified and loaded automatically.
NOTE
If formatting is failed, try again. Restart the monitor
and retry the formatting, or contact the service
personnel of the manufacturer if formatting is failed
repeatedly.
Recording
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146
Overview
Waveform record is printed at the rate of Auto recording interval is set by the user and
12.5 mm/s, 25 mm/s or 50 mm/s; default setting the waveform displays in real time.
is 25 mm/s.
NOTE
49 mm ~ 50 mm wide printout paper.
Do not use the strip recorder with a low battery as
Record one SpO2 waveform; default setting is recorder usage will more quickly deplete the
off. battery.
User-selectable real-time recording time and
waveform.
NOTE
The Record shortcut key enables users to
manually start or stop recording.
NOTE
Proper operation
If the monitor is not configured with the recorder
When the strip recorder is operating, do not pull function, it will indicate Recorder Setup Needed
the paper. after the Record button is pressed.
Replacing paper
NOTE
Be careful when inserting papers. Avoid damaging
the thermo-sensitive print head. Do not leave the
strip recorder door open except to insert paper or
troubleshoot printing problems.
Using battery
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150
Overview
WARNING WARNING
Before using the rechargeable lithium-ion Do not unplug the battery when working is in
battery (hereinafter called battery), be sure to process.
read the user manual and safety precautions
thoroughly. WARNING
Do not heat or throw the battery into a fire.
WARNING
The service life of the battery depends on the WARNING
service frequency and time. The service life of
Do not use, leave the battery close to fire or
the battery is about three years if the battery is
other places where temperature may be above
well maintained and stored. The service life of
60 °C.
the battery may shorten if it is used
inappropriately. If the battery life is exhausted
and not replaced in time, it may cause damage WARNING
or heat to the device. Do not immerse, throw, or wet the battery in
water/seawater.
WARNING
Periodic checks on the battery performance WARNING
are required. Change the battery if necessary. Do not destroy the battery: do not pierce the
battery with a sharp object such as a needle;
WARNING do not hit with a hammer, step on or throw or
drop to cause strong shock; do not
Do not connect the positive (+) and negative (-)
disassemble or modify the battery.
terminals with metal objects, and do not put the
battery together with metal objects, which can
result in short circuits. WARNING
The recommended battery can only be used
for this monitor. Do not solder the leading wire
and the battery terminal directly.
WARNING WARNING
If liquid leaking from the battery gets into the Do not use a battery with serious scratch or
eyes, do not rub the eyes. Wash them well with deformation.
clean water and go to see a doctor
immediately. If liquid leaks of the battery WARNING
splash onto the skin or clothes, wash well with
When the monitor is running on battery power,
fresh water immediately.
do not replace the battery during monitoring
patients; or the monitor will be powered off,
WARNING which may result in patient injury.
Keep away from fire immediately when
leakage or foul odor is detected. WARNING
Do not place battery in the monitor with the (+)
WARNING and (-) in the wrong way.
Stop using the battery if abnormal heat, odor,
discoloration, deformation or abnormal WARNING
condition is detected during use, charge, or
Due to high ambient temperature, the monitor
storage. Keep it away from the monitor.
gives out Battery Error alarm. To protect the
battery from damage, take battery out for
cooling and then charge.
Battery Error
The performance of rechargeable batteries may 3 Disconnect monitor from mains power and let
deteriorate over time. Battery maintenance as the monitor run until there is no battery power
recommended here can help to slow down this left and the monitor shuts off.
process.
4 The running time of the battery reflects the
1 Disconnect the patient from the monitor and battery performance.
stop all monitoring and measurement.
If the running time is obviously less than the
2 Switch the monitor power on and charge the specified time in the specification, change the
battery for more than 6 hours continuously. battery or contact the service personnel.
To install or replace the battery, follow the 3 Insert the new battery into the battery
procedure: compartment.
4 Close battery latch.
Reprocessing
Reprocessing of patient-specific
accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 160
Categorization of accessories . . . . . . . . . . . . . 160
Reprocessing of the category "[1] SpO2
Sensors", "[2] NIBP Cuffs" . . . . . . . . . . . . . . . . 160
Storage and transport . . . . . . . . . . . . . . . . . . . . 161
Other agents and reprocessing procedures . . . 161
Safety instructions
General points
Keep the monitor, cables and accessories free of Unless otherwise specified, do not immerse any
dust and dirt. To prevent the device from damage, part of the equipment or any accessories in
follow the procedure: liquid.
Use only recommended cleaning substances Do not pour liquid onto the system.
and disinfectants listed in this manual. Others
Do not allow liquid to enter the case.
may cause damage (not covered by warranty),
reduce product lifetime or cause safety Never use abrasive material (such as steel wool
hazards. or silver polish).
Always dilute according to the manufacturer’s Inspect the monitor and reusable accessories
instructions. after they are cleaned and disinfected.
Safety information
Check the products for signs of wear and If the user spills liquid on the equipment,
replace them if necessary. battery, or accessories, or they are accidentally
immersed in liquid, contact the service
personnel or Dräger service engineer.
Information on reprocessing
Follow the national infection prevention policies Follow the infection prevention policies and
and reprocessing regulations. reprocessing regulations of the healthcare facility
(e.g., concerning the reprocessing cycles).
Classification Explanation
Classification Explanation
Non-critical Components that come into contact only with skin that is intact
Semi-critical Components that carry breathing gas or come into contact with mucous
membranes or pathologically altered skin
Critical Components that penetrate skin or mucous membranes or come into contact with
blood
Before Reprocessing
After reprocessing, there are no special – Avoid recontamination and damage during
requirements for storage and transport of the transport
product. However, the following must be observed:
All further information on storage and transport
– Store dry and free of dust included in the accompanying documents must be
observed.
Categorization of accessories
Dräger states that oxygen-releasing agents and Other surface disinfectants are used at one's own
chlorine-releasing agents may cause color change risk.
in some materials. Color change does not indicate
that the product is not functioning correctly.
After Reprocessing
Maintenance
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 164
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 165
Overview
WARNING WARNING
Failure on the part of the responsible The maintenance operations such as software
institution to implement a satisfactory upgrade of the device shall only be conducted
maintenance schedule may cause undue by Dräger-qualified service professionals. The
equipment failure and possible health history data in monitor may be deleted due to
hazards. software upgrade. Before software upgrade,
backup the data in the monitor to avoid data
WARNING loss. For data backup methods, refer to the
section Exporting data stored in the internal
If the monitor is mechanically damaged, or if it
storage device to export the data, or refer to
is not working properly, do not use it. Contact
the section Uploading data to network server
technical personnel.
to load the data to HIS system.
WARNING WARNING
If the user finds any problems with the Any serious incident that has occurred in
equipment, contact service personnel or relation to the device should be reported to
authorized supplier. the manufacturer and the competent authority
of the Member State in which the user and/or
patient is established.
Inspection
The overall check of the monitor, including the If all functions are in good conditions.
safety check, should be performed only by qualified
If the grounding resistance and leakage current
personnel every 24 months, and each time after fix
meet requirement.
up.
If any damage or abnormality is found, don’t use the
The following items should be checked:
monitor and contact local Customer Service
If the environment condition and power supply Center.
meet requirement.
Maintenance shall be carried out at least once
If the power supply cord has damage and every two years, or as specified by local
insulativity meets requirement. regulations. The following tasks are for Dräger-
qualified service professionals only. Contact an
If the device and accessories have damage.
Dräger-qualified service provider if the monitor
Specified accessories. needs a safety or performance test. Clean and
disinfect equipment to decontaminate it before
If the alarm system can work properly.
testing or maintaining it.
If the recorder can work properly and the paper
meets the requirement.
Battery performance
Repair
Disposal
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 168
Overview
CAUTION
The device and its components must be
disinfected and cleaned before disposal!
Accessories
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 170
Overview
SpO2 accessories
NIBP accessories
TEMP accessories
Other accessories
NOTE
The part description may vary depending on
context, but the part number is constant.
Product specification
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 176
Classification . . . . . . . . . . . . . . . . . . . . . . . . . 176
NIBP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 180
SpO2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183
PR. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
TEMP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 185
Wi-Fi. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187
e-link . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Interfaces. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
Nurse Call. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 188
USB Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . 188
Wired Network Interface . . . . . . . . . . . . . . . . . . 189
Overview
NOTE
The performance of the equipment with ☆ mark is
determined to be essential performance.
Classification
Compliant with Standards IEC 60601-1: 2005+A1 :2012; IEC 60601-1-2: 2014;
EN 60601-1: 2006+A1 :2013; EN 60601-1-2: 2015;
IEC 60601-2-49: 2011
Physical specifications
Function configuration
Temperature
Humidity
Altitude
Display
Display Messages
Battery specification
Number 1
Recorder
Trace 1
Data management
Data review
Data storage
In monitoring mode, a single piece of patient data maximally contains the following information:
Patient information MRN, name, date of birth, date of admission, gender, type, height,
weight, blood type, doctor, bed No., department
In ward round mode, storage data maximally contains the following information:
Ward round MRN, name, type, bed No., ward round record Up to 800 thousand sets
record and original record
SpO2 Storage starting time, SpO2 value, PR value Up to 20 sets for a single patient
NIBP SYS, DIA, MAP, PR, measurement time Up to 20 sets for a single patient
1 GB space for data storage: ≥100 thousand sets of ward round records. Up to 800 thousand sets of
ward round records are supported (one ward round record has 20 original records).
NIBP
Technique Oscillometry
times 3/5
☆ Measuring Range
Adult/Pediatric 120 s
Neonate 90 s
Pre-inflation Pressure
SunTech Module
Method Oscillometric
times 3/5
☆ Measuring Range
Adult 130 s
Pediatric 90 s
Neonate 75 s
Overpressure protection
Pre-inflation Pressure
Adult Mode 120 mmHg, 140 mmHg, 150 mmHg, 160 mmHg, 180 mmHg,
200 mmHg, 220 mmHg, 240 mmHg, 260 mmHg, 280 mmHg
Default: 160 mmHg
Pediatric Mode 80 mmHg, 100 mmHg, 120 mmHg, 140 mmHg, 150 mmHg,
160 mmHg, 180 mmHg, 200 mmHg, 220 mmHg, 250 mmHg
Default: 120 mmHg
Neonatal Mode 60 mmHg, 70 mmHg, 80 mmHg, 90 mmHg, 100 mmHg, 120 mmHg,
140 mmHg
Default: 90 mmHg
SpO2
Resolution 1%
☆ Accuracy
Sensor
PI
Resolution 1
Nellcor Module
Resolution 1%
If sensor is used for neonate as recommended, the accuracy will be larger than adult
by ±1.
NOTE
The information about wavelength range can be
especially useful to clinicians (for instance, when
photodynamic therapy is performed).
PR
TEMP
Arterial Heat Balance Range for Body 34.5°C ~ 43 °C (94 °F ~ 110 °F)
Temperature 1
Resolution 0.1 °C or °F
Battery type and life 9-volt alkaline battery, providing 15,000 readings 2
NOTE WARNING
1 Automatically applied when temperature is within The monitor may not meet performance
normal body temperature range, otherwise reads specifications if stored or used outside
surface temperature. temperature and humidity ranges. When
2 Approximate number of readings when scanning moving the monitor from a storage location,
for 5 seconds and reading the temperature display wait at least one-hour or more prior to use to
for 3 seconds before turning off the thermometer. allow the monitor to adjust to room
temperature.
Working temperature 10 °C ~ 40 °C
Resolution 0.1 °C
Typical measurement time (after insertion Oral (Quick Predictive Mode): (3~5) s (non-fever
into measurement site) temps); (8~10) s (fever temps)
Axillary: (8~12) s
Rectal: (10~14) s
Stability 0.14 °C
NOTE
The direct mode refers to monitoring mode, while
adjusted mode refers to predictive mode and quick
predictive mode.
Wi-Fi
Latency ≤1 s
Jitter ≤1 s
PER ≤10%
e-link
I/U Ratio ≤1 dB
PER ≤10%
Interfaces
Nurse Call
USB Interfaces
EMC Information
Electromagnetic emissions
NOTE
The EMISSIONS characteristics of Vista 120 SC
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class
B is normally required) Vista 120 SC might not
offer adequate protection to radio-frequency
communication services. The user might need to
take mitigation measures, such as relocating or re-
orienting the equipment.
Electromagnetic immunity
Electromagnetic immunity
following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which Vista 120 SC is used exceeds the applicable RF compliance level above, Vista 120
SC should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating Vista 120 SC.
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
c The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz,
5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to
18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to
54.0 MHz.
Test Immunity
Band a) Maximum Distance
frequency Service a) Modulation b) test level
(MHz) power (W) (m)
(MHz) (V/m)
Pulse
385 380-390 TETRA 400 modulation b) 1.8 0.3 27
18 Hz
FM c)
GMRS 460, ± 5 kHz
450 430-470 deviation 2 0.3 28
FRS 460
1 kHz sine
710
Pulse
LTE Band 13,
745 704-787 modulation b) 0.2 0.3 9
17
217 Hz
780
810 GSM
800/900,
870 Pulse
TETRA 800,
800-960 modulation b) 2 0.3 28
iDEN 820,
18 Hz
930 CDMA 850,
LTE Band 5
Bluetooth,
WLAN, Pulse
2450 2400-2570 802.11 b/g/n, modulation b) 2 0.3 28
RFID 2450, 217 Hz
LTE Band 7
5240
Pulse
WLAN
5500 5100-5800 modulation b) 0.2 0.3 9
802.11 a/n
217 Hz
5785
Test Immunity
Band a) Maximum Distance
frequency Service a) Modulation b) test level
(MHz) power (W) (m)
(MHz) (V/m)
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.
Vista 120 SC is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of Vista 120 SC can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and Vista 120 SC as recommended below, according to the maximum output power of
the communications equipment.
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Default settings
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200
Overview
Default configuration
To set default configuration, select Menu > Default To check the configuration currently used, select
Setup. On the Profile menu, users can choose a Menu > Default Setup. The one labeled with ● is
factory configuration (adult, pediatric or neonate) current configuration. If there's no labeled
based on the patient category. Also, users can configuration, it means the currently used
choose a user configuration saved in the monitor if configuration is not one of them.
it is available. For more information about user
configuration, refer to User configuration.
User configuration
Alarm Settings
Pause Time 120 s
Alarm Latch Off
PR default settings
Interval 1 minute
Interval (Average) 1 minute
Measurement times 3
(Average)
PR Switch On
Auto Recording Off
Alert Switch On
Alert High Limit (SYS) 160 120 90
Alert Low Limit (SYS) 90 70 40
Alert High Limit (MAP) 110 90 70
Alert Low Limit (MAP) 60 50 30
Alert High Limit (DIA) 90 70 60
Alert Low Limit (DIA) 50 40 20
Directive 93/42/EEC
concerning Medical Devices
Manufacturer:
26 80 025 – GA 6498.020 en
© Shanghai Dräger Medical Instrument Co., Ltd.
Edition: 2 – 2020-02
Dräger reserves the right to make modifications
to the equipment without prior notice.