116 About calibration requests

About calibration requests


The system can request calibration automatically, or you
can request calibration manually.

The cause of the calibration request is displayed on the


Calibration > Status window.

Cause Description
Changeover You load a new reagent pack with no valid lot calibration on the system
Daily The calibration is defined as required daily
Failed The calibration fails one or more quality criteria
Manual A manual request for calibration by the operator
QC Violation The QC results are outside the confidence limits
Calibration requests

Related topics
• About calibration (115)
• Validating QC results (152)

Roche Diagnostics
cobas e 411 analyzer · Software Version 03-02 · Operator's Manual · Version 3.3.1