You are on page 1of 30

Downloaded from http://jicru.oxfordjournals.

org/ at University of Glasgow on July 9, 2016


ICRU REPORT 38
ICRU REPORT 38

Dose and Volume

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Specification for
Reporting Intracavitary
Therapy in Gynecology

Issued: 1 March 1985


First Reprinting December 15,1991

INTERNATIONAL COMMISSION ON RADIATION

UNITS AND MEASUREMENTS

7910 WOODMONT AVENUE

BETHESDA, MARYLAND 20814

U.S.A.
THE INTERNATIONAL COMMISSION
ON RADIATION UNITS AND MEASUREMENTS
INDIVIDUALS PARiflCIPATING IN THE PREPARATION OF THIS REPORT

Commission Membership During Preparation of Commission Sponsor


This Report
A. W AMBERSIE
H. O. WYCKOFF, Chairman Universite Catholique de Louvain
A. ALLISY, Vice Chairman Bruxelles, Belgium
R. S. CASWELL, Secretary
G.E.D.ADAMS
G. COWPER
P.EDHOLM
J. R. GREENING Report Committee
D.HARDER D. CHASSAGNE. Co-Chairman
A. M. KELLERER Institut Gustave Roussy
H. H. ROSSI Villejuif, France

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


W. K. SINCLAIR A. DUTREIX, Co-Chairman
J. VAN DER SCHOOT Institut Gustave Roussy
A. W AMBERSIE Villejuif, France
L. S. TAYLOR, Honorary Member and Member
P.ALMOND
Emeritus M. D. Anderson Hospital and Tumor Institute
Houston, Texas, U.S.A.
Current Commission Membership J. M. V. BURGERS
Antoni van Leeuwenhoek Ziekenhuis
H. O. WYCKOFF, Chairman Amsterdam, The Netherlands
A. ALLISY, Vice Chairman M.BuSCH
R. S. CASWELL, Secretary Strahlenklinik der Gesamthochschule
G.E.D.ADAMS Essen, Germany
G.COWPER C. A. JOSLIN
L. FEINENDEGEN Cookridge Hospital
I. ISHERWOOD Leeds, United Kingdom
A. M. KELLERER
J. R. MALLARD
H. H. ROSSI
W. K. SINCLAIR
J. VAN DER SCHOOT Consultants to the Report Committee
A. WAMBERSIE M.COHEN
L. S. TAYLOR, Honorary Chairman and Member McGill University
Emeritus Royal Victoria Hospital
Montreal, Canada
Executive Secretary T.LANDBERG
Malmo Allmiinna Sjukhus
W.R.NEY Malmo, Sweden

The Commission wishes to express its appreciation to the individuals involved in the preparation of this report
for the time and effort they devoted to this task and to express its appreciation to the organizations with which they
are affiliated.

Copyright © International Commission on Radiation Units and Measurements 1985


(For detailed information on the availability of this and other ICRU Reports, see page 20)
Preface
Scope of ICRU Activities Current Program
The International Commission on Radiation Units The Commission has divided its field of interest into
and Measurements (ICRU), since its inception in 1925, twelve technical areas and has assigned one or more
has had as its principal objective the development of members of the Commission the responsibility for
internationally acceptable recommendations re- identification of potential topics for new ICRU activities
garding: in each area. A body of consultants has been constituted
(1) Quantities and units of radiation and radioac- for each technical area to advise the Commission on the
tivity, need for ICRU recommendations relating to the tech-
(2) Procedures suitable for the measurement and nical area and on the means for meeting an identified
application of these quantities in clinical radiology and need. Each area is reviewed periodically by its sponsors

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


radiobiology, and consultants. Recommendations of such groups for
(3) Physical data needed in the application of these new reports are then reviewed by the Commission and
procedures, the use of which tends to assUre uniformity a priority assigned.
in reporting. The technical areas are:
The Commission also considers and makes similar Radiation Therapy
types of recommendations for the radiation protection Diagnostic Radiology
field. In this connection, its work is carried out in close Nuclear Medicine
cooperation with the International Commission on Radiobiology
Radiological Protection (ICRP). Radioactivity
Radiation Physics-X Rays, Gamma Rays and Electrons
Radiation Physics-Neutrons and Heavy Particles
Policy Radiation Protection
Radiation Chemistry
The ICRU endeavors to collect and evaluate the latest Critical Data
data and information pertinent to the problems of ra- Theoretical Aspects
Quantities and Units
diation measurement and dosimetry and to recommend
the most acceptable values and techniques for current The actual preparation of ICRU reports is carried
use. out by ICRU report committees. One or more Com-
The Commission's recommendations are kept under mission members serve as sponsors to each committee
continual review in order to keep abreast of the rapidly and provide close liaison with the Commission. The
expanding uses of radiation. currently active report committees are:
The ICRU feels it is the responsibility of national Absolute and Relative Dosimetry at High Doses
organizations to introduce their own detailed technical Characterization of Irradiation for Materials-Effect Studies
procedures for the development and maintenance of Chemical Dosimetry
standards. However, it urges that all countries adhere Clinical Dosimetry for Neutrons
Computer Uses in Radiotherapy
as closely as possible to the internationally recom- Definitions and Terminology for Computed Tomography
mended basic concepts of radiation quantities and Definitions of Physical Parameters to Specify Performance of
units. Imaging Instruments
The Commission feels that its responsibility lies in Determination of Absorbed Dose Distribution Around a Source
developing a system of quantities and units having the Used for Interstitial Therapy
Dose Specifications for Reporting Interstitial Therapy
widest possible range of applicability. Situations may Measurement of Dose Equivalent
arise from time to time when an expedient solution of Modulation Transfer Function for Screen-Film Systems
a current problem may seem advisable. Generally Practical Determination of Dose Equivalent Index
speaking, however, the Commission feels that action Quality Assurance in External Beam Therapy
based on expediency is inadvisable from a long-term Stopping Power
Tissue Equivalent Materials
viewpoint; it endeavors to base its decisions on the
long-range advantages to be expected.
The ICRU invites and welcomes constructive com-
ICRU Reports
ments and suggestions regarding its recommendations
and reports. These may be transmitted to the In 1962 the ICRU, in recognition of the fact that its
Chairman. triennial reports were becoming too extensive and in
Iv ... Preface

some cases too specialized to justify single-volume Council for International Organizations of Medical Sciences
publication, initiated the publication of a series of re- Food and Agriculture Organization of the United Nations
International Council of Scientific Unions
ports, each dealing with a limited range of topics. This International Electrotechnical Commission
series was initiated with the publication of six re- International Labor Office
ports: International Organization for Medical Physics
International Radiation Protection Association
ICRU Report lOa, Radiation Quantities and Units International Union of Pure and Applied Physics
ICRU Report lOb, Physical Aspects of Irradiation United Nations Educational, Scientific and Cultural Organiza-
ICRU Report lOc, Radioactivity tion
ICRU Report 10d, Clinical Dosimetry
ICRU Report lOe, Radiobiological Dosimetry The Commission has found its relationship with all
ICRU Report 10f, Methods of Evaluating Radiological Equip- of these organizations fruitful and of substantial benefit
ment and Materials
to the ICRU program. Relations with these other in-
These reports were published, as had been many of ternational bodies do not affect the basic affiliation of
the previous reports of the Commission, by the United the ICRU with the International Society of Radi-
States Government Printing Office as Handbooks of the ology.
National Bureau of Standards. Operating Funds
In 1967, the Commission determined that in the fu-

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


ture the recommendations formulated by the ICRU In the early days of its existence, the ICRU operated
would be published by the Commission itself. This re- essentially on a voluntary basis, with the travel and
port is published by the ICRU pursuant to this policy. operating costs being borne by the parent organizations
With the exception of ICRU Reports lOa, lOd, and lOe, of the participants. (Only token assistance was originally
the other reports of the "10" series have continuing available from the International Society of Radiology.)
validity and, since no subsequent reports were designed Recognizing the impracticability of continuing this
specifically to supersede them, they will remain avail- mode of operation on an indefinite basis, operating
able until the material is essentially obsolete. All future funds were sought from various sources.
reports ofthe Commission, however, will be published Financial support has been received from the fol-
under the ICRU's own auspices. Information about the lowing organizations:
availability of ICRU Reports is given on page 20. ADAC Laboratories
Agfa-Gavaert, N.V.
ICRU's Relationships With Other Organizations Atomic Energy Control Board
B.A.T. Cigaretten Fabriken GMBH
In addition to its close relationship with the Inter- Central Electricity Generating Board
national Commission on Radiological Protection, the CGR Medical Corporation
ICRU has developed relationships with other organi- Commission of the European Communities
Council for International Organizations of Medical Sciences
zations interested in the problems of radiation quan- Danish Scientific Fund
tities, units and measurements. Since 1955, the ICRU Dutch Society for Radiodiagnostics
has had an official relationship with the World Health Eastman Kodak Company
Organization (WHO) whereby the ICRU is looked to for E.l. du Pont de Nemours and Company
primary guidance in matters of radiation units and Ford Foundation
Fuji Photo Company
measurements and, in turn, the WHO assists in the General Electric Company
world-wide dissemination of the Commission's recom- Gilbert X-ray Company
mendations. In 1960, the ICRU entered into consulta- Hitachi Medic
tive status with the International Atomic Energy International Atomic Energy Agency
International Radiation Protection Association
Agency. The Commission has a formal relationship with
International Society of Radiology
the United Nations Scientific Committee on the Effects Italian Radiological Association
of Atomic Radiation (UNSCEAR), whereby ICRU ob- Japan Industries Association of Radiation Apparatus
servers are invited to attend UNSCEAR meetings. The John och Augusta Perssons stiftelse
Commission and the International Organization for Konishi-Roku Photo Company
National Cancer Institute of the U.S. Department of Health and
Standardization (ISO) informally exchange notifica-
Human Services
tions of meetings and the ICRU is formally designated N.V. Philips Gloeilampenfabrieken
for liaison with two of the ISO Technical Committees. Philips Medical Systems, Incorporated
The ICRU also corresponds and exchanges fmal reports Picker Corporation
with the following organizations: Pyne Corporation
Radiological Society of North America
Bureau International de Metrologie Legale Rockefeller Foundation
Bureau International des Poids et Mesures Shimadzu Corporation
Commission of the European Communities Siemens Aktiengesellschaft
Preface . •• V

Society of Nuclear Medicine ipating in ICRU activities, (2) travel costs involved in
Statens laegevidenskabelige Forskningsrad ICRU meetings, and (3) meeting facilities and ser-
Toshiba Corporation
United Nations
vices.
U.S. Bureau of Radiological Health of the Food and Drug Admin- In recognition of the fact that its work is made pos-
istration sible by the generous support provided by all of the or-
World Health Organization ganizations supporting its program, the Commission
Xerox Corporation expresses its deep appreciation.
In addition to the direct monetary support provid-
ed by these organizations, many organizations provide HAROLD O. WYCKOFF
indirect support for the Commission's program. This Chairman, ICRU
support is provided in many forms, including, among Bethesda, Maryland, U.S.A.
others, subsidies for (1) the time of individuals partic- 15 January 1985

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Contents
Preface .................................................... iii
1. Introduction............................................. 1
2. Definition of Terms and Concepts Currently Used in
Intracavitary Therapy ................................... 4
2.1 Treatment Techniques. . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1.1 Radium Substitutes ........................... . ..... 4
2.1.2 Simulation of Linear Sources ......................... 4
2.1.3 Dose Rates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.1.4 Afterloading Techniques ............................. 5
2.2 Absorbed-Dose Pattern and Volumes. . . . . . . . . . . . . . . . . . . . . . 5

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


2.2.1 Absorbed-Dose Pattern .............................. 5
2.2.2 Volumes ............... . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.3 Specification of Radioactive Sources ...................... 8
3. Recommendations for Reporting Absorbed Doses and
Volumes in Intracavitary Therapy. . . . . . . . . . . . . . . .. . . . . . . . . 9
3.1 Introduction ........ . ...... . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Description of the Technique ........... . ................ 9
3.2.1 The Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2.2 Simulation of Linear Sources .. . ...................... 9
3.2.3 The Applicator ..................................... 9
3.3 Recommendations for Reporting . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3.1 Total Reference Air Kerma . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3.2 Description of the Reference Volume .................. 10
3.3.3 Absorbed Dose at Reference Points .................... 11
3.3.4 Calculation of Dose Distribution ...................... 12
3.4 Definition of the 60 Gy Reference Volume in
Special Situations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.4.1 One Linear Source Only ........................... . . 14
3.4.2 Vaginal Sources Only. . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . 14
3.4.3 Rigid Applicator .................................... 14
3.4.4 Uterine Packing (In Endometrial Cancer) .............. 14
4. Time-Dose Pattern .......... . ......... . . . . . . . . . . . . . . . . . . . 15
4.1 Radiobiological Considerations ........................... 15
4.2 Recommendations for Reporting Time-Dose Pattern ........ 15
5. Conclusions and Final Recommendations .................. 16
Appendix
Specification of Radioactive Sources Used in
Intracavitary Therapy .... . ............................... 17
References.. ... . .. .. ... .. .. . .. . . .. . .. . .. .. . .. ...... .. . .. .. . 18
ICRU Reports ........................... . . . . . . . . . . . . . . . . . . . 20
Index...................................................... 23
Dose and Volume Specification
for Reporting Intracavitary
Therapy in Gynecology

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


1. Introduction

In 1978, the ICRU published Report 29, Dose Spec- system (Paterson, 1948)1. Most of the systems used
ification for Reporting External Beam Therapy with throughout the world are derived from these three basic
Photons and Electrons (ICRU, 1978). The present re- systems.
port deals with the problem of dose and volume speci- As used here, the term "system" denotes a set of rules
fication for reporting intracavitary therapy. The prin- taking into account the source strengths, geometry and
cipal subject is dose and volume specification for the method of application in order to obtain suitable dose
intracavitary treatment of cervix carcinoma. However, distributions over the volume(s) to be treated. For re-
the concepts developed in this report are designed to be porting, the system includes recommendations for
applicable to other types of intracavitary applications, specifying the application and possibly, as in the
such as uterine body, etc. Manchester system, for calculating the dose rate (or
Intracavitary therapy is often used in combination dose) at specific points.
with external beam therapy and, therefore, the same Typical applications according to the Stockholm
terminology and definitions of terms are advocated. In system and the historical Paris system are presented in
particular, it would be desirable, whenever possible, to Figures 1.1 and 1.2, respectively. In both systems, ap-
use the concepts of target volume, treatment volume plications have been reported in terms of "mg.h" (mil-
and irradiated volume, as defined for external beam ligram hours), i.e., the product of the total mass or ra-
therapy. However, due to the high dose-gradient around dium contained in the sources (in mg) and of the dura-
the sources (throughout the tumor and target volume), tion of the application (in hours).
the specification of the target absorbed dose in terms The Manchester system (Paterson and Parker, 1934),
of the dose absorbed either at one or several reference derived from the original Paris system, initiated about
point(s) within the target volume is not considered 1920, was designed to deliver a constant dose rate to
meaningful; therefore, the approach used in external defmed points near the cervix, irrespective of variation
beam therapy cannot be used. in size and shape of the uterus and vagina. In the
The problem of dose specification is still more com- Manchester system, an application was specified in
plicated when two or more target volumes are identified, terms of the "dose" in roentgens delivered at specific
some being treated either by external, or by intracavi- points such as points "A" and "B" (Figure 1.3). Points
tary therapy and some by both methods. Furthermore, A and B are still widely used throughout the world, al-
attention should be paid to the different time-dose though their exact meaning and their definition have
patterns linked to each treatment method. not always been interpreted in the same way in different
Based on clinical experience, different systems have centers and even in a given center over a period of time.
been proposed for the treatment of cervix carcinoma. In particular, some centers relate point A to anatomical
Three basic systems have been developed: the Stock- references in the patient, others to the geometry of the
holm system (Kottmeier, 1964)1, the Paris system sources.
(Lamarque and Coliez, 1951)1 and the Manchester The different methods of definition provide different
values for the calculated dose rate to point A. As a result,
1The references quoted are not the first references but those which where a prescribed dose to point A is used to calculate
seem to be the most comprehensive. the total time for an insertion, different values of time
1
2 • • . 1. Introduction

Fig. 1.1. The Stockholm system. Typical treatment of a cervix


carcinoma with a radium application (uterus normal in size and
shape). The intrauterine rod-shaped applicator is loaded with 53-88
mg radium (74 mg in the example shown). The vaginal applicator
usually consists of a flat box containing 60-80 mg radium (70 mg in
the example shown), but in special cases other forms of vaginal ap-
plicators may be used. Classically, the two applicators are not fixed
to each other, but fixed or semi-fixed combinations have been de-
veloped. The vaginal applicator is held against the cervix and lateral
fornices by careful and systematic gauze packing. Typically, 2 or 3
applications are given with 3-week intervals, each application lasting
for 27-30 hours. Modifications of the Stockholm method, using larger
amounts of radium, allow for application times of 10-18 hours at each
treatment. The figure shows a typical application and the amount of
radium in the two applicators as well as the dose-rate (Figure from
~
o Scm Walstam, 1954).

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Fig_ 1.2. The historical Paris system. Typical radium application
for a treatment of cervix carcinoma consisting of: 3 individualized
vaginal sources (one in each lateral fornix and one central in front of
the cervical os), 1 intrauterine source made of 3 radium tubes (in so-
called tandem position). Possible variations of the Paris system in-
clude: 2 vaginal sources (or only one) in case of a narrow vaginal cavity;
only 2 intrauterine radium tubes (or only one) in case of short uterus.
The active length of the sources is usually 16 mm, their linear activity
being between 6 and 10 mg per cm, and their strength between 10 and
15 mg of radium. The total activity used is one of the lowest in use for
such treatments and implies a typical duration of the application of
6-8 days. Typically, the ratio of the total activity of the vaginal sources
to the total activity of the uterine sources should be 1 (with variations
between 0.66 and 1.5). As far as the method of application is con-
cerned, the historical Paris system does not imply any fixed distance
between the vaginal sources, nor connection between vaginal and
uterine sources (Pierquin, 1964).

B B

Fig. 1.3. The Manchester system. Definition of points" A" and "B".
In the classical Manchester system, point A is dermed as a point 2 em
lateral to the central canal of the uterus and 2 em up from the mucous
mem brane of the lateral fornix, in the axis of the uterus. Point B is
defined as being in the transverse axis through points A, 5 cm from
the midline. In clinical practice, dose calculations are often made from
radiographs and point A is taken 2 em up from the flange of the in-
trauterine source and 2 em lateral from the central canal as indicated
in the figure (Meredith, 1967). In a typical application, the loading of
intrauterine applicators varied between 20 and 35 mg of radium and
between 15 and 25 mg of radium for each vaginal ovoid. The resultant
treatment time to get 8000 R at point A was 140 hours.
1. Introduction... 3

will be obtained, depending on the method used. It is (i) radium is being progressively replaced by gamma-
therefore important to give a precise description of the ray emitting substitutes (137CS, 192Ir and 60CO)2;
method used to calculate the dose rate at specific (ii) new sets of sources are available, for which the old
points. systems are not always suitable;
On the other hand, the Manchester system is based (iii) SI units are now being widely used in the field of
on a set of strict rules concerning the method of insert- radiology; and
ing the radium sources (in particular the choice of vag- (iv) computers are routinely used for calculations and
inal ovoids and their separation, and the relative complete sets of intracavitary dose distribution
strengths ofthe intrauterine and intravaginal sources). in several planes are now becoming available, from
Assuming that the Manchester definition of point A is which new information and relationships can be
used, this will be the location where there is least vari- derived.
ation in the dose rate from one source arrangement to
another one.
It is opportune now to reconsider dose and volume 2 Californium-252, a neutron emitter, is also used but will not be
specification for several reasons: discussed in this document.

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


2. Definition of Terms and Concepts Currently Used in Intracavitary Therapy

2.1 Treatment Techniques

2.1.1 Radium Substitutes used, involving different activities or different spacing


in order to modify and improve the shape of the relevant
When radium was the only available radionuclide, the
isodose surfaces (Cardis and Kjellman, 1967; Chassagne
characteristics of the sources used for intracavitary
applications were similar and their activities were of the et al., 1969; Bjornsson, 1980; Van der Laarse, 1981).
In the simple case where a linear source is simulated
same order of magnitude. Typically, the elementary
radium tubes most frequently used were 1.5 to 2.2 cm by a series of regularly spaced point sources of equal
long, with source strengths ranging from 5 to 20 mg of activity, each point source is equivalent to a given length
radium and a total Pt filtration of 1 to 2 mm (see Ap- of the linear source. The linear source simulated by the
pendix). point sources is longer than the distance between the
The replacement of radium by 137Cs, 192Ir and 60CO extreme point sources (see Figure 2.1). At distances
larger than half the spacing between the point sources,

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


may be accomplished according to two options. In the
the dose distribution around the point sources is similar
first option, the new sources (mainly 137CS) are similar
to that around a linear source of the same total activity
in size and shape and have an output similar to radium
(Dutreix and Wambersie, 1968).
sources. The same technique of application can then be
Linear sources can also be simulated by moving point
applied and the clinical experience gained with radium
sources of appropriate activity. Variations of the type
remains fully relevant. The principal advantages of the
of movement, continuous or stepwise, of the speed and
replacement of the radium concern radiation protection'
these include no contamination from leakage and les~
dwell times of the source at different positions, will
modify the shape of the isodose surfaces (Henschke et
shielding in the case of 137CS and 192Ir.
al., 1966; Joslin et al., 1969; Twiss and Bradshaw, 1970;
The second option takes advantage of the improved
Busch, 1973; Busch et al., 1977; Himmelmann et al.,
technology in the preparation of the sources, the in-
1980; Von Essen, 1980).
creased specific activities available with artificial ra-
dionuclides and the technical possibility of making
miniaturized sources. These new types of source are 2.1.3 Dose Rates
designed to allow the use of new and better techniques
of application, for instance, afterloading, and to improve In conventional brachytherapy with radium tubes,
the dose distribution by providing, for instance, greater the dose rate at the point or surface where the dose is
flexibility in the selection of the relative source activi- prescribed, lies between 0.4 and 2 grays per hour. It is
ties, source lengths or both. common practice to refer to this type of treatment as
low dose-rate brachytherapy.
In addition, the large range of source activities at
present available allows the therapist to modify com- The introduction of remote afterloading devices into
clinical medicine (see Section 2.1.4. below) has pre-
pletely the time-dose pattern of application. Irradiation
times ranging from 6 to 8 days, with radium in the Paris
!-E=lcm-!
system, could be reduced to a few minutes (O'Connell
• • ::;VI
r
et al., 1967). Thus, continuous low dose-rate irradiation
can be replaced by fractionated high dose-rate irra-
diation. It should be stressed that such drastic changes
in the time-dose pattern require a modification of the
,,--=--=='-<:':-::-~--=-">'-'=:--=--
~
I
<--
5
- - -' /
I
total dose and that the clinical experience accumulated E'l
u
over many decades with relatively low dose-rate radium
applications can no longer be applied without careful
evaluation, not only as far as normal tissue tolerance is
concerned, but also for effects on the tumor. Ii I, " ' " ,,, I
2.1.2 Simulation of Linear Sources o 2 3 4cm
In many of the modern application devices, linear Fig. 2.1. Comparison of dose distributions between "n" small
sources are simulated by a set of point sources. These sources of activity ao spaced by a distance E, and a linear source of
length I = nE and of total activity a = nao. The curves are labeled 5,
point sources may be of equal activity and equally 2, 1 for relative dose rate. The dotted lines result from a series of point
spaced, thus simulating a uniform linear activity. In sources and the full lines from a linear source. The dose is not cor-
other applications, more complex arrangements are rected for oblique filtration (Dutreix and Wambersie, 1968).

4
2.2 Absorbed-Dose Pattern and Volumes • •• 5

sented an opportunity for investigating different dose afterloading techniques as well as modifying their ir-
rates and, in particular, very high dose rates. In this radiation procedures, attention needs to be given to the
Report, the term high dose rate refers to any dose rate problems of dose specification and dose-time relation-
higher than 0.2 gray per minute (12 grays per hour), ships. As already stressed, the clinical experience ac-
although it usually refers to dose rates as high as 2 to 5 cumulated with radium techniques cannot be applied
grays per minute, i.e., treatment sessions of a few min- to new irradiation conditi{)ns without careful consid-
utes duration. eration.
Some radiotherapists are now exploring intermediate
dose rates, between 2 and 12 grays per hour and we
2.2 Absorbed-Dose Pattern and Volumes
propose to refer to such dose rates as medium dose
rates.
2.2.1 Absorbed-Dose Pattern
Such definitions of low dose rate, medium dose rate
and high dose rate are arbitrary and debatable. In this The pattern of the absorbed dose to the soft tissue in
connection, it must be stressed that the treatment du- intracavitary therapy differs from the dose pattern
ration should always be clearly reported (see Section encountered in external therapy. In external beam
4). therapy, one aim is to reduce variations in dose over the
target volume. In general, variations greater than ±10%

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


are not deemed acceptable. Outside the treatment
2.1.4 Afterloading Techniques volume the dose falls off more or less steeply. For ex-
Modifications of application techniques in intra- ample, pelvic irradiation with a 60Co "box technique"
cavitary therapy are often made for purposes of radia- (see Figure 2.2a) produces a dose distribution within the
tion protection. As already discussed, some advantage treatment volume which is rather flat; the dose falls off
is obtained by replacing radium by ')I-emitters of lower rapidly at the edges of the treatment volume and in the
energy (the shielding problems are easier and the risk lateral directions (Fields 3 and 4) remains fairly con-
of contamination is also avoided). However, the main stant at about 30% of the dose at the center of the
improvement in radiation protection derives from the treatment volume. With intracavitary therapy, the dose
introduction of afterloading techniques. is maximum adjacent to the sources and at the center
In afterloading techniques, unloaded applicators or of the treated volume, and it falls off continuously with
guides are introduced into the patient cavities, ac- distance from the sources. Consequently, the size of the
cording to a chosen plan and without exposure of the treatment volume cannot be deduced from a simple
staff. The treatment sources are inserted later, after the inspection of the isodose pattern. Therefore, the ra-
position of the applicators has been radiographically diotherapist has to indicate which dose level defines the
checked using dummy sources. The dose distribution treatment volume. In Figure 2.2c, the absorbed dose has
can also be computed so that the position of the guides been arbitrarily normalized to 100% at a distance of 2
can be corrected if necessary before the sources are in- cm from the uterine axis in order to allow a comparison
serted. of the absorbed dose patterns. Although the dose gra-
Afterloading techniques help to ensure the correct dient (dose change per unit distance) at the border of
and safe positioning of the radioactive sources in the the treatment volume is similar for either technique, as
patient, in addition to facilitating the transfer of the shown in Figure 2.2c, the dose patterns differ substan-
sources to and from a shielded storage container. Af- tially within the treatment volume as well as in healthy
terloading techniques may be manual or remotely tissues distant from the treated region.
controlled.
In remotely controlled afterloading techniques, 2.2.2 Volumes
source insertion and removal are operated from a con-
trol panel distant from the patient, thus eliminating The definitions given in ICRU Report 29 for external
exposure of the clinical staff. With some of the more beam therapy will be modified where necessary for in-
advanced devices, the sources are automatically re- tracavitary therapy situations.
tracted into the storage container when the door of the Target Volume
patient's room is opened. While either manual or me- The target volume contains those tissues that are to
chanical afterloading techniques may be used for low be irradiated to a specified absorbed dose according to
dose-rate applications, remote afterloading techniques a specified time-dose pattern. For curative treatment,
are mandatory for medium and high dose-rate appli- the target volume consists of the demonstrated tu-
cations in order to ensure good radiation protection of mor(s), if present, and any other tissue with presumed
the staff. tumor.
Afterloading is an important aspect of intracavitary For any given situation, there may be more than one
therapy. Moreover, as more centers are now adopting target volume. This is particularly the case for treat-
Ox

Field No Size Weighting


(cm x cm) factors
17 x 20 37 "/.
2 17 x 20 37 "/.
3 10 x 20 13 "/.
L. 10 x 20 13 "/.
3

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


oI , , ,
5, 10cm
, ,
2-
I
2.2a

'/
; --- .....
,
1/;-.... "
..... , ''
\ \' ,
'/,_
-_
-,'
..... ,,,
,\ \

I
~"
\'I \ \
\
\\ '\
,
"
I I , \
I I I \ \ \
I I \ \ \ \

A ...j....£_I_.l--->.
1110\\\ \ \
7050 30 20
,\\ ~
\ \\ . . . - - _
I _ __
\
10 cGy,h- 1
B
\ ,
' ......... ",'
\
,
\
\
\

~c...=..:..;:..:...-~2::.....::ccm~
..... "
\ \" \I II
\ I I I
® /,, I I
J I I I I
_/ " I I I
_",,/ I I I
____ - .... ",I' /1
/

---
--- -
..,'"
-"
/

/
/
1

, _------/
.....

-----IRIDIUM WIRES
/
-" '"

RADIUM TUBES
uterine tubes 3 x 10 mg (same reference air kerma
rates as radium tubes )
vaginal tubes 2 x 15 mg
filtration 1 mm Pt correction made for oblique
fill ration
2.2b
2.2 Absorbed-Dose Pattern and Volumes . . . 7

1//'
I
I//l
100
I~I
l;/j
I

'"
01
tU

~ 50
v
....
'"
a.
'"
III
o 3
a

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


o
-'0 -5 +5 +10 em

2.2c
Fig. 2.2. Comparison of absorbed dose patterns between external and intracavitary therapy. 2.28 shows the dose distribution for an irradiation
of the pelvis with a 6OCO teletherapy machine and four fields-the so-called "box technique" (see ICRU Report 29, Figure 3.4). The border of
the target area is indicated by the heavy dotted line. 2.2b shows a typical dose distribution in an intracavitary application (IAEA, 1972; Dutreix
et al., 1982). 2.2c shows the dose variation along a lateral axis. The full line refers to external therapy (Figure 2.2a, along the beam axis 3 and
4). The dotted line refers to intracavitary therapy (Figure 2.2b, along the AB axis). The absorbed dose has been normalized to 1000/0 at a distance
of 2 cm from the uterine axis. The shaded areas (2 cm wide) show the high dose-gradient regions beyond the border of each of the two treatment
volumes.

ment of cervix carcinoma where, depending on the ex- ume. In intracavitary therapy, it is not feasible to ex-
tent of disease, some target volumes may be treated press the value of the isodose surface enclosing the
partly or totally by an intracavitary application, while treatment volume as a percentage of the dose at any
others are treated mainly by external therapy, The point within the treatment volume. This is because of
target volume(s) must always be described, indepen- the steep dose gradient around the sources and the
dently of the dose distribution, in terms of the patient's variation encountered in the techniques used. It is
anatomy, topography and tumor volume, For the pur- necessary to plan the treatment to include the target
pose of reporting intracavitary treatment, that part of volume within the treatment volume.
the target volume which has to be irradiated by means Reference Volume
of intracavitary application should be separately de- The reference volume is defined as the volume en-
scribed, It should also be stated whether this volume will closed by the reference isodose surface. In order to fa-
receive treatment only by intracavitary application or cilitate intercomparisons between radiotherapy centers,
will receive treatment by both modalities. The physical it is necessary to agree upon a reference dose level. The
dimensions of the target volume must be given. Exam- treatment dose level defining the treatment volume may
ples illustrating target volumes in patients with carci- be equal to or different from this reference dose level.
noma of the cervix are given in Figure 2.3. For reporting intracavitary therapy, it is necessary to
Treatment Volume determine the dimensions of the reference volume.
The treatment volume is defined as the volume en- Irradiated Volume
closed by a relevant isodose surface selected by the ra- The irradiated volume is that volume, larger than the
diotherapist, and encompasses at least the target vol- treatment volume, which receives an absorbed dose
8 ... 2. Definition of Terms and Concepts Currenlly Used in Intracavlly Therapy

A. ONLY INTRACAVITARY TREATMENT B. COMBINED INTRACAVITARY AND


EXTERNAL BEAM THERAPY

Targllt lIOIumll of intra- Trutnwnt volumll of


Targllt 'w'Olumll Trcatmlznt volumll cavitary trutmcnt in tracavilary trut mllnt
(cllrvix rllgion + corpus)

\
·• .J

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


·!./ \
II
.,,. \
Fillid i\ ;,/. ._./ /
~\~::':'~~0~" -_ .. - .. -~.---- -1
Targllt volumll of Trllatmllnt volume of
utllrnal bum utllrnal bum thllrapy
thuapy

Fig. 2.3. Treatment of cervix carcinoma. Stage II.b. Examples of target volumes and treatment volumes. A-Tumor volume (hatched): 4 cm
X 3 cm X 2 em. The almost pear-shaped target volume includes the entirety of palpable and visible tumor as well as the whole uterus. The planned.
pear-shaped treatment volume encompasses the target volume. Additional methods such as combined surgery or combined external radiation
may be used. B-Tumor volume (hatched): 9 cm X 5 cm X 4 cm. The target volume for brachytherapy will only include the major parts of the
tumor volume. but will include the whole uterus. To treat the rest of the tumor volume and further subclinical disease throughout the pelvis
and the regional lymphatics. external beam therapy will be used. and brachytherapy will be used as boost therapy.

considered to be significant in relation to tissue toler- 2.3 Specification of Radioactive Sources


ance. The significant absorbed dose level can be ex-
pressed as a percentage (e.g., 50%) of the agreed dose It is recommended that radioactive sources be spec-
level (see Section 3.3.2.a). ified in terms of "reference air kerma rate". The ref-
Organs at Risk erence air kerma rate of a source is the kerma rate to air,
Organs at risk are those radiosensitive organs in or in air, at a reference distance of 1 meter, corrected for
near the target volume which would influence treatment air attenuation and scattering. For this purpose this
planning and/or the prescribed dose. For example, in quantity is expressed in J.LGy·h- 1 at one metre.
an intracavitary application for cervix carcinoma, the The total reference air kerma is the sum of the
main organs at risk are rectum, bladder, ureters and products of the reference air kerma rate and the dura-
possibly sigmoid colon. tion of the application for each source.
3. Recommendations for Reporting Absorbed Doses and Volumes in Intracavitary
Therapy
3.1 Introduction the position of the radioactive sources is limited. How-
As the absorbed dose to soft tissue from intracavitary ever, for other gynecological intracavitary situations the
applications is so highly nonuniform throughout the same philosophy can be adopted, but some of the nu-
target volume, the concepts of maximum, mean, median merical values and definitions may need to be modified
and modal target absorbed dose, as defined in ICRU according to the type of application.
Report 29 (ICRU 1978), are not relevant. The minimum 3.2 Description of the Technique
target absorbed dose is the only useful concept and is,
by definition, equal to the treatment absorbed dose It is recommended that the technique be described
level. on the basis of the guidance given below.

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


For external beam therapy, it has been recommended
that the target absorbed dose be the absorbed dose at 3.2.1 The Sources
one or more specification points which are representa- (i) radionuclide;
tive of the dose distribution throughout the target vol- (ii) reference air kerma rates;
ume. These specification points could be established (iii) shape, filtration, etc.
with respect to the target volume or to the beam axes,
or both. 3.2.2 Simulation of Linear Sources
The situation is more difficult in intracavitary ther- a) When a linear source is simulated by a set of point
apy due to the steep dose gradient in the vicinity of the sources, the activity of these point sources and their
sources, i.e., throughout the tumor or the target volume. separation(s) must be indicated.
Under these conditions, the specification of the target b) When moving sources are used to simulate a set of
absorbed dose in terms of the absorbed dose at specific different sources in fixed position, in order to pro-
point(s), in the vicinity of the sources, becomes less duce an appropriate dose distribution, the following
meaningful and a different approach is required. In- indications are required:
stead of a target-dose specification, a volume specifi- (i) type of movement (continuous or stepwise, step
cation is recommended. distance);
Specification of an intracavitary application in (ii) unidirectional or oscillating movement;
terms of the "reference volume" enclosed by the ref- (iii) range of movement or oscillation;
erence isodose surface is proposed in this Report. (iv) speed in different sections of the applicator, or
However, as the different isodose surfaces are close to dwell times of the source at different posi-
each other, the indication of the reference volume must tions.
be supplemented, for safety reasons, by the indication
of the total reference air kerma which, for radium 3.2.3 The Applicator
treatments, is proportional to the number of "mg·h". In Reference to the applicator is sufficient when a
addition, a record of absorbed dose at reference points complete description has already been published, pro-
related to organs at risk or to bony structures is rec- vided that there is no significant difference between the
ommended. The time-dose pattern shall also be clearly applicator used and the one described in the literature.
indicated (see Section 4). To avoid confusion, it is recommended that the appli-
The present recommendations do not imply a modi- cator be described, including the name of the manu-
fication of the method used for the calculation of the facturer. The description should include information
treatment duration, but they imply the calculation of on the following points:
specific quantities to be determined for reporting. (i) rigid (or not), consequently with fixed known ge-
The recommendations presented in this Report must ometry (or not) of the complete applicator;
be considered a minimum requirement for reporting. (ii) rigid uterine source with fixed curvature (or
On the other hand, the reported parameters will be not);
meaningful only to the extent that the technique of the (iii) connection between vaginal and uterine applica-
particular intracavitary application has been com- tors, i,e., fixed, loose (semi-fixed), free;
pletely described. (iv) type of vaginal sources, number and orientation of
This report deals mainly with the treatment of cervix line sources, special sources (box, ring, etc.);
carcinoma for which the anatomical region of interest (v) high atomic number shielding materials in vaginal
is similar for every patient and the possible variation in applicator (or not).
9
10 ... 3. Recommendations for Reporting Absorbed Doses and Volumes In Intracavitary Therapy

3.3 Recommendations for Reporting Plane a


Three approaches are proposed to specify intracavi-
tary application for cervix carcinoma; they complement
each other and it is recommended that they be com-
bined.

3.3.1 Total Reference Air Kerma


When radium was used exclusively, the product of the
"quantity of radioactive material" and the duration of
the treatment was given as minimal information. This
is the "concept of the mg·h". It becomes increasingly
difficult to use the product mg·h for specifying an in-
tracavitary application with other radionuclides, even
if milligrams of radium are replaced by the term "mil-
ligram-radium equivalent" which can be misleading (see

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Appendix). As noted in Section 2.3, the use of total
reference air kerma is proposed in this Report. This
quantity is unambiguous and easy to calculate. It is Plane b
proportional to the integral dose to the patient and can
also serve as a useful index for radiation protection of
personnel.
In addition, the inverse square law applied to the total
reference air kerma allows evaluation, to a reasonable
-
approximation, of the absorbed dose delivered during
treatment at distances from the sources down to 20-10
cm. When the distance of a point P from the center C
of the volume occupied by the sources is larger than 2.5
times the largest dimension of that volume, the dose rate
obtained at P from the actual distribution of sources .... a
differs by less than 4% from that obtained by assuming
Fig. 3.1. Geometry for measurement of the size of the pear -shaped
that all of the sources are located at C (Dutreix et al., 60 Gy isodose surface (broken line) in a typical treatment of cervix
1982). carcinoma using one rod-shaped uterine applicator and 2 vaginal
However, the simple calculation of the total reference applicators. Plane a is the "oblique" frontal plane that contains the
air kerma does not allow one Wderive, even roughly, the intrauterine device. The oblique frontal plane is obtained by rotation
of the frontal plane around a transverse axis, Plane b is the "oblique"
absorbed dose in the immediate vicinity of the sources sagittal plane that contains the intrauterine device. The oblique
(Le., in the tumor or target volume) and, in particular, sagittal plane is obtained by rotation of the sagittal plane around AP
it should be stressed that an accurate and simple rela- axis. The height (db) and the width (d w) of the reference volume are
tion cannot be derived between total reference air kerma measured in plane a as the maximal sizes parallel and perpendicular
and absorbed doses at specific points such as A and B to the uterine applicator, respectively. The thickness (d t ) of the ref-
erence volume is measured in plane b as the maximal size perpen-
(see Figure 1.3.).
dicular to the uterine applicator.

3.3.2 Description of the Reference Volume


The description of the reference volume, i.e., the tis-
sue volume encompassed by a reference isodose surface, dose pattern should be clearly stated (see Section
is proposed for specification in reporting. The reasons 4.2).
for this approach are described in Sections 2.2.1.,2.2.2. When intracavitary therapy is combined with ex-
and 3.1. ternal beam therapy, the isodose level to be considered
a. Dose Level is the difference between 60 Gy and the dose delivered
An absorbed dose level of 60 Gy is widely accepted at the same location by external beam therapy. For
as the appropriate reference level for classical low example, if a dose of 20 Gy were delivered to the whole
dose-rate therapy. When two or more intracavitary pelvis by external beam therapy, the isodose level to be
applications are performed, the absorbed dose to con- considered would be 60 - 20 Gy = 40 Gy. Nevertheless,
sider is that r~sulting from all applications. The time- it is recognized that the combined dose does not nec-
3.3 Recommendations for Reporting • • • 11

essarily produce the same effect as a similar dose from means of characterizing an intracavitary application
intracavitary therapy alone. and/or of specifying the target absorbed dose, particu-
For intracavitary therapy at medium or high dose larly if rigid source combinations are not used. Such
rates, the therapist has to indicate the dose level which points are not recommended.
he believes to be equivalent to 60 Gy delivered at the b. Reference Points Relatively Close to the Sources but
classical low dose rate and should clearly state this. Related to Organs at Risk
b. Reference Volume: Description of the Pear-Shaped The determination and specification of the absorbed
Volume dose to organs at risk (bladder, rectum, etc.) are ob-
When uterine source(s) are combined with vaginal viously useful with respect to normal tissue tolerance
sources, or when the uterine source is more heavily limits. However, such information will be meaningful
loaded at the lower end, the tissue volume to be de- only to the extent that it is obtained and expressed in
scribed presents a pear shape with its longest axis precise and well-codified ways.
coincident with the intrauterine source. This reference (i) Calculated Values: Reference points for the ex-
volume is defined by means of three dimensions (see pression of the absorbed dose to the bladder and the
Figure 3.1): absorbed dose to the rectum (see Figure 3.2.) have
(i) the height (d h ) is the maximum dimension along been proposed by Chassagne and Horiot (1977).

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


the intrauterine source and is measured in the The bladder reference point is obtained as fol-
oblique frontal plane containing the intrauterine lows. A Foley catheter is used. The balloon must be
source; filled with 7 cm3 of radio-opaque fluid. The catheter
(ii) the width (d w ) is the maximum dimension per- is pulled downwards to bring the balloon against the
pendicular to the intrauterine source and is mea- urethra. On the lateral radiograph, the reference
sured in the same oblique frontal plane; point is obtained on an anterio-posterior line drawn
(iii) the thickness (d t ) is the -maximum dimension through the center of the balloon. The reference
perpendicular to the intrauterine source and is point is taken on this line at the posterior surface
measured in the oblique sagittal plane containing of the balloon. On the frontal radiograph, the ref-
the intrauterine source. erence point is taken at the center of the balloon.
The point of reference for the rectal dose is ob-
The definitions of dh, d w and d t are proposed in order
tained as follows. On the lateral radiograph, an
to minimize the number of calculations. These dimen-
anteroposterior line is drawn from the lower end
sions are usually expressed in cm. The volume estimated
of the intrauterine source (or from the middle of the
from the intersections of the surface of a pear-shaped
intravaginal sources). The point is located on this
volume on two conventional planes does not necessarily
line 5 mm behind the posterior vaginal wall. The
represent the size of the true reference volume. How-
posterior vaginal wall is visualized, depending upon
ever, for most applications they do not differ from the
the technique, by means of an intravaginal mould
maximum dimensions of the reference volume by more
or by opacification of the vaginal cavity with a
than 1 or 2 mm.

3.3.3 Absorbed Dose at Reference Points


Several reference points are in current use. Some are
relatively close to the sources and related either to the balloon
7cm 3 point
sources or to organs at risk; others are relatively far from
the sources and are related to bony structures. The
following definitions apply to the case where the doses
are calculated from two perpendicular radiographs, AP
and lateral. When other methods are used, such as ste-
reographic x-ray films, oblique perpendicular radio-
graphs or transverse sections (CT scans), the calcula-
tions need to be modified.
a. Reference Points Close to the Sources and Related nlctal
to the Sources
As such points are located in a region where the dose
gradient is high, any inaccuracy in the determination
of distance results in large uncertainties in the absorbed
doses evaluated at these points. Such calculated ab- Fig. 3.2. Determination of the reference points for bladder and
sorbed doses do not, therefore, seem an appropriate rectum (see text).
12 ... 3. Recommendations for Reporting Absorbed Doses and Volumes In Intracavitary Therapy

the tolerance level. This type of measurement re-


quires special care in positioning the measuring
probe. An example is given in Figure 3.3_ (O'Connell
et al., 1967; Joslin et al., 1972).
c. Reference Points Related to Bony Structures
(I) (i) The lymphatic trapezoid is obtained as follows (see
Line taken
by probe ---.- Figure 3.4): A line is drawn from the junction of
Doso-rato 81-82 to the top of the symphysis_ Then a line is
motor
drawn from the middle of that line to the middle of
the anterior aspect of U. A trapezoid is constructed
in a plane passing through the transverse line in the
Mothod A
pelvic brim plane and the midpoint of the anterior
aspect of the body of L4. A point 6 cm lateral to the
midline at the inferior end of this figure is used to
give an estimate of the dose rate to mid-external
iliac lymph nodes (labeled R. EXT and L. EXT for

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


right and left external, respectively). At the top of
the trapezoid, points 2 cm lateral to the midline at
the level of L4 are used to estimate the dose to the
low para-aortic area (labeled R. PARA and L.
PARA). The midpoint of a line connecting these
two points is used to estimate the dose to the low
common (labeled R. COM and L. COM) iliac lymph
nodes (from Fletcher, 1980).
Doso-rat.
mlZhzr (ii) The pelvic-wall reference point (Chassagne et al.,
1977) can be visualized on an AP and a lateral ra-
diograph and related to fixed bony structures. This
Mdhod B
point is intended to be representative of the ab-
Fig. 3.3. Measurement of the rectal dose rate. Following the inser- sorbed dose at the distal part of the parametrium
tion of the source applicators, either pre-loaded (low-activity treat- and at the obturator lymph nodes (see Figure 3.5.).
ment) or manually loaded with low-activity sources identical in design On an AP radiograph, the pelvic-wall reference
to those used during high-activity after-loaded treatment, the rectal
dose is measured_
point is intersected by the following two lines: a
Method A: The measuring probe is moved relative to a rigid horizontal line tangential to the highest point of the
guide tube inserted into the rectum and held in position. The acetabulum, a vertical line tangential to the inner
point of maximum rectal dose rate is noted and the distance aspect of the acetabulum. On a lateral radiograph,
"d" in cm from the anal verge deduced. the highest points of the right and left acetabulum,
Method B: The measuring probe is moved so that the tip of
the probe is moved along the mid-line of the recto-vaginal sep-
in the cranio-caudal direction, are joined and the
tum until the point of maximum dose rate is reached_ Distance lateral projection of the pelvic-wall reference point
is taken as a direct reading on the central tube at the anal is located at the mid-distance of these points.
verge. The dose-rate and distance are recorded. Evaluation of the absorbed dose at reference points,
The major disadvantage of method" A" is that the probe tip cannot related to well-defined bony structures and lymph node
follow closely the surface of the anterior rectal wall. However, allow-
areas, is particularly useful when intracavitary therapy
ance for the distance of the probe sensor from the vagino-rectal sep-
tum needs to be taken into account. is combined with external beam therapy. It is also useful
in helping to avoid an overdose when intracavitary
therapy is to be followed by surgery.

radio-opaque gauze used for the packing. On the AP


3.3.4. Calculation of Dose Distribution
radiograph, this reference point is at the lower end
of the intrauterine source or at the middle of the The present recommendation, in particular the de-
intravaginal source(s). scription of the reference volume encompassed by the
(ii) Monitoring of the Absorbed Dose Rate to the 60-Gy isodose surface, necessitates the computation of
Rectum.- An alternative to calculating the rectal complete dose distributions in several planes.
dose is to measure the dose, or dose rate, at different The respective planes for which the dose distribunon
points along the anterior rectal wall to ensure that is to be computed will depend on the technique IUhl (I.e
no area of the rectal mucosa receives a dose above particular clinical situation. However, as a millimum
3.3 Recommendations for Reporting • . • 13

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Fig. 3.4. Determination of the lymphatic trapezoid (see text). On the left is an anteroposterior view and on the right a lateral view.

A P X - RAY FILM LATERAL X -RAY FILM


Fig. 3.5. Determination of the right (RPW) and left (LPW) pelvic wall reference points (see text).
14 . . . 3. Recommendations for Reporting Absorbed Doses and Volumes In Intracavitary Therapy

requirement, it is recommended that the dose distri- -in the case of vaginal irradiation with a central source
butions be computed in two planes: the oblique frontal from a cylindrical applicator.
plane and the oblique sagittal plane both containing the In estimating the volume (Section 3.3.2.b) in this
intrauterine source. simple case, the width is equal to the thickness.
When practicable, it is recommended that dose dis-
tributions be calculated in additional sets of planes and
that these dosimetric data be correlated with that ob- 3.4.2 Vaginal Sources Only
tained from radiographs or CT sections, in order to When only vaginal sources are present, width is the
determine the absorbed dose at any relevant anatomical largest dimension from right to left in an oblique frontal
point. plane through the main axis of the vagina.
While this additional information will be of value in Thickness is the largest dimension in a direction
assessing effects in any individual patient, it will also perpendicular to the above oblique plane. Height is
provide: measured along the vaginal axis, and is commonly
(i) the possibility of comparing the methods of spec- shorter than the other two dimensions.
ification used in different centers and of evaluating
their respective merits;
3.4.3 Rigid Applicator

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


(ii) the possibility of comparing the methods of spec-
ification used in historical series (mg·h, points "A" Provided that there is a fixed connection between
and "B") with the methods recommended in the vaginal and uterine sources, pre-calculated isodose
present Report; surfaces can be obtained for given loadings of the ap-
(iii) the possibility of deriving new clinical and radio- plicator. Therefore, pre-calculated dimensions of height,
biological data and correlations which could im- width and thickness can be given (IAEA, 1972).
prove treatment techniques and develop further
the method of specification.
3.4.4 Uterine Packing (In Endometrial Cancer)
In connection with uterine packing, the same defi-
3.4 Definition of the 60 Gy Reference Volume in
nitions of height, width and thickness given in Section
Special Situations
3.3.2 can be used. However, two facts need to be
noted:
3.4.1 One Linear Source Only
-width and thickness are usually located at the level
In some situations only one linear source is of the uterine fundus (the pear-shaped volume is re-
present: versed),
-in the case of a narrow vagina with a uterine source -height should be determined in the oblique frontal
protruding into the vaginal cavity, plane, which gives the maximum dimension.
4. Time-Dose Pattern

4.1 Radiobiological Considerations dose rate may be significant (Van Limbergen et al.,
1985). Its location and thickness will vary with the total
Radiobiological effects often depend on dose rate but
reference air kerma rate and the application time. In
to a greater or lesser extent according to the actual level
addition, the time between fractions is important in the
of the dose rate and the nature of the effect (Hall, 1978).
case of fractionated high dose-rate intracavitary
For example, it is generally observed for mammalian
therapy.
cells irradiated in vitro that for dose rates larger than
It can be concluded that any significant change in the
about 1 Gy·min- 1 survival is independent of dose rate,
source strength and the time-dose pattern should be
but becomes increasingly dependent on dose rate at
made with the greatest care. Furthermore, early tissue
lower rates. For dose rates below about 1 Gy·h- 1 survival
reactions alone should not be used to select the pre-
again becomes independent of dose rate.
scribed dose since late reactions, which are most rele-
The relationship between dose rate and tissue re-
vant, depend to a greater extent on dose rate.
sponse in vivo is complex because many factors have to

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


be taken into account such as cell distribution, repair
of potentially lethal lesions, and, for tumors, reoxyge- 4.2 Recommendations for Reporting Time-Dose
nation. As far as cell proliferation is concerned, its role Pattern
is probably a minor one in brachytherapy applica- Sufficient data on the effect of dose-rate on various
tions. tissues are not available at the present time. Therefore,
In each brachytherapy application, there is neces- no correction factors for dose rate can be recommended,
sarily both a marked dose gradient and a marked dose but the duration of the application should be stated.
rate gradient as a function of distance from the When more than one application is performed, the
sources. duration of each should be reported, as well as the time
The possible importance of changes in time factors interval(s) between them.
(dose rate) in determining the biological or clinical effect Similarly, when external beam therapy and intra-
should not be underestimated since there may be a shell cavitary therapy are combined, the time-dose schedule
of tissues surrounding the sources where differences in of the whole treatment should be reported.

15
5. Conclusions and Final Recommendations
In intracavitary therapy, due to the high dose-rate different tissues or organs are directly proportional to
gradient, the specification of the target absorbed dose the total reference air kerma.
as that dose observed at one (or several) reference Third, it is recommended that the reference volume
point(s) within the target volume or close to the sources, be described in terms of the height, width and thick-
does not appear to be meaningful. Under these condi- ness of the volume enclosed in the 60 Gy isodose surface
tions, the same approach to reporting as in external for cervix-carcinoma treatment by low dose rates. For
beam therapy cannot be recommended. higher dose rates a dose level lower than 60 Gy has to be
The following are proposed in this Report (see Table selected.
0.1). Fourth, the absorbed dose at reference points in or-
gans at risk (rectum, bladder) should be determined
First, as already pointed out in ICRU Report 29 (computed or measured) and expressed in well-codified
(ICRU, 1978), any method of specification will be ways to provide additional safety limits.
meaningful only to the extent that the treatment In addition, the absorbed dose(s) at reference point(s)

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


technique has been completely described. related to bony structures (lymphatic trapezoid and
Second, the total reference air kerma should be pelvic-wall reference points) should also be reported.
stated. For a given method of application (source ge- Finally, the time-dose pattern should be completely
ometry and relative loads), the doses delivered at the specified.

TABLE 5.1-List of data needed for reporting intracavitary therapy in gynecology


Data Needed Paragraph Page

DESCRIPTION OF THE TECHNIQUE USED 3.2 9


TOTAL REFERENCE AIR KERMA (cGy at 1 metre) 3.3.1 10
DESCRIPTION OF THE REFERENCE VOLUME 3.3.2 10
-Dose level if not 60 Gy 3.3.2.a 10
-Dimensions of the reference volume (height, width, thickness) 3.3.2.b 11
ABSORBED DOSE AT REFERENCE POINTS 3.3.3 11
Bladder reference point 3.3.3b 11
Rectal reference point 3.3.3b 11
Lymphatic trapezoid 3.3.3c 12
Pelvic wall reference point 3.3.3c 12
TIME DOSE PATTERN 4 15

16
APPENDIX
Specification of Radioactive Sources Used in Intracavitary Therapy

Only some aspects of the problem of the specification terms of its output, i.e., exposure rate at a reference
of sources will be reported here since this topic will be distance. The reference exposure rate in milliroentgens
dealt with at length in a forthcoming ICRU Report and per hour at 1 meter was used (mR·h- l ·m 2) (Wambersie
has been treated recently (Dutreix et at., 1982). et at., 1973; NCRP, 1974; Dutreix, 1974; Dutreix and
(1) When radium tubes were the only radioactive Wambersie, 1975).
sources used in intracavitary therapy, their strength was (6) Recently, the Comire Fran~ais pour la Mesure des
specified in terms of the mass of radium in mg con- Rayonnements Ionisants (CFMRI, 1983) has recom-

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


tained in the tube. mended that radioactive sources be specified in terms
(2) When artificial radionuclides became available, of "reference air kerma rate" ("debit de kerma normal
the sources were first specified in terms of their activity dans l'air"). The reference air kerma rate of a source is
in mCi. the kerma rate to air, in air, at a reference distance of
(3) Due to the influence of self-absorption and fil- one metre, corrected for air attenuation and scattering.
tration within the source and its sheath, the "contained This quantity is expressed in ~Gy·h-l at 1 m. For ex-
activity" was of little practical interest and the concept ample, a "point" source containing 1 mg of radium, 0.5
of "apparent activity" was introduced. The apparent mm Pt filtration, produces an exposure rate of 8.25 ±
activity of a source was defined as the activity of a point 0.10 R·h- l at 1 cm or 8.25 X 10- 4 R·h- l ·m 2. Taking into
source of the same radionuclide which would deliver the account the conversion factor from exposure to air
same exposure rate in air at the same distance from the kerma (0.873 cGy·R-l) and the correction for the dif-
center ofthe actual source (the distance should be large ference in filtration (0.93), a source containing 10 mg of
enough so that the actual source can be considered as radium, 1 mm Pt filtration, produces an air kerma rate
a "point source"). of 67 ~Gy·h-l at 1 m.
Instead of the term apparent activity, the expression (7) The total reference air kerma is the sum of the
"equivalent activity" has also been used (IAEA, 1967 products of the reference air kerma rate and the dura-
and ICRU, 1970), but it could lead to some confusion tion of the application for each source.
with the expression "milligram-radium equivalent". By way of example, we consider an application per-
This is because sources of different radionuclides with formed with 25 mg of radium as an intrauterine source
the same apparent activity will not give equivalent dose and 40 mg of radium as the vaginal sources. If the ra-
distributions. dium tubes are filtered by 1 mm Pt, 65 mg of radium
(4) In order to compare radium substitutes directly yield a reference air kerma rate of 435 ~Gy·h-l·m2.
with radium itself, sources were specified in "milli- Therefore, for example (for radioprotection purposes),
gram-radium equivalent". The radium equivalent mass at one metre from the patient the dose equivalent rate
(mg-Ra equivalent) of a source is the mass of radium will not exceed about 0.4 mSv·h- l .
filtered by 0.5 mm of platinum which leads to the same Assuming that the sources are left in situ for 6 days
exposure rate as that from the radioactive source of (144 hours), the total reference air kerma is equal to 435
interest at the same distance. Thus, the dose rate at a X 144 X 10-6 Gy or 6.27 cGy at 1 m. The dose delivered
given distance from a typical 10 mg radium tube filtered to the sigmoid for example (assumed to be 10 cm from
by 1 or 2 mm of platinum is less than the dose rate de- the center of the sources) is approximately 6.27 X 102
livered at the same distance by a 10 mg Ra equivalent cGy or 6.27 Gy, neglecting tissue attenuation and the
source which was standardized to 0.5 mm platinum. tissue-kerma to air-kerma ratio.
(5) Later, the strength of a source was specified in

17
References
BJORNSSON, M. (1980). "Some new treatment methods the uterine cervix by remote afterloading with cycling
with afterloading techniques" p. 102 in High Dose- sources," Am. J. Roentgenol. 96,45.
Rate After-Loading in the Treatment of Cancer of HIMMELMANN, A., KARLSTEDT, K, and RAGNHULT,
the Uterus, Bates T. D. and Berry R. J., Eds., Br. J. I. (1980). "Early experience with high dose-rate
Radiol. Special report 17. (Ralstron) treatment in Goteborg" p. 106 in High
BUSCH, M. (1973). "Three-dimensional optimization Dose-Rate Afterloading in the Treatment of Cancer
in intracavitary application of radium and other of the Uterus, Bates T. D. and Berry R. J., Eds., Br.
sealed gamma sources," p. 658 in Proceedings XIII J. Radiol. Special report 17.
International Congress of Radiology. Madrid,1973, IAEA (1967). International Atomic Energy Agency,
Vol. 2 (Excerpta Medica, Amsterdam). Physical Aspects of Brachytherapy, Reports Series
BUSCH, M., MAKOSKI, B., SCHULTZ, U., and SAUER- no. 75 (International Atomic Energy Agency,
WEIN, K (1977). "Das Essener Nachlade-Verfahren Vienna).
fur die intracavitare Strahlentherapie," Strahlen- IAEA (1972). International Atomic Energy Agency,

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


therapie 153,581. Atlas of Radiation Dose Distributions. Vol. IV,
CARDIS, R. and KJELLMAN, J. (1967). "Un nouvel ap- Brachytherapy Isodose Charts. Sealed radium
pareil pour la radiotherapie endocavitaire; Ie cervi- sources. M. Stovall, L. H. LanzI, W. S. Moos, Eds.
tron 2," Industries Atomiques, 11, 1. (International Atomic Energy Agency, Vienna).
CHASSAGNE, D., DELOUCHE, G., ROCOPLAN, J. A., ICRU (1970). International Commission on Radiation
PIERQUIN, B., and GEST, J. (1969). "Description et Units and Measurements, Specification of High
premiers essais du Curietron," J. Radiol. Electrol. 50, Activity Gamma-Ray Sources, ICRU Report 18
910. (International Commission on Radiation Units and
CHASSAGNE, D., GERBAULET, A., DUTREIX, A., and Measurements, Bethesda, Maryland).
COSSET, J. M. (1977). "Utilisation pratique de la ICRU (1978). International Commission on Radiation
dosimetrie par ordinateur en curietherapie gyneco- Units and Measurements, Dose Specification for
logique," J. Radiol. Electrol. 58, 387. Reporting External Beam Therapy with Photons
CHASSAGNE, D. and HORIOT, J. C. (1977). "Proposi- and Electrons, ICRU Report 29 (International
tions pour une definition commune des points de Commission on Radiation Units and Measurements,
reference en curietherapie gynecologique," J. Radiol. Bethesda, Maryland).
Electrol. 58, 371. JOSLIN, C. A., LIVERSAGE, W. E., and RAMSEY, N. W.
CFMRI (1983). Comite Fran~ais pour la Mesure des (1969). "High dose-rate treatment moulds by after-
Rayonnements Ionisants, Recommandations pour loading techniques," Br. J. Radiol. 42, 108.
la determination des doses absorbees en curieth- JOSLIN, C. A., SMITH, C. W., and MALLIK, A. (1972).
erapie, Rapport CFMRI no. 1 (Bureau National de "The treatment of cervix cancer using high activity
Metrologie, Paris). 6OCo sources," Br. J. Radiol. 45, 257.
DUTREIX, A. (1974). "Une nouvelle methode de speci- KO'ITMEIER, H. L. (1964). "Surgical and radiation
fication des sources radioactives en curietherapie," treatment of carcinoma of the uterine cervix," Acta
J. Radiol. Electrol. 55,781. Obst. Gynec. Scand. 43, Suppl. 2, 1.
DUTREIX, A. and WAMBERSIE, A. (1968). "Etude de LAMARQUE, P. and COLIEZ, R. (1951). "Les cancers des
la repartition des doses autour de sources ponctuelles organes genitaux de la femme" p. 2549 in Electro-
alignees. Application en curietherapie gynecolo- radiotherapie, Delherm, L., Ed. (Masson, Paris).
gique," Acta Radiol. Ther. 7, 389. MEREDITH, W. J. (1967). Radium dosage: The Man-
DUTREIX, A. and WAMBERSIE, A. (1975). "Specifica- chester System (Livingstone, Edinburgh).
tion of gamma-ray brachytherapy sources," Br. J. NCRP (1974). National Council on Radiation Protec-
Radiol. 48, 1034. tion and Measurements, Specification of Gamma-
DUTREIX, A., MARINELLO, G., and WAMBI!,;H,l:;IE, A. Ray Brachytherapy Sources, NCRP Report No. 41
(1982). Dosimetrie en Curietherapie (Masson, (National Council on Radiation Protection and
Paris). Measurements, Bethesda, Maryland).
FLETCHER, G. H. (1980). "Squamous cell carcinoma of O'CONNELL, D., JOSLIN, C. A., HOWARD, N., RAMSEY,
the uterine cervix" p. 720 in Textbook of Radio- N. W., and LIVERSAGE, W. E. (1967). "Treatment of
therapy (Lea & Febiger, Philadelphia). uterine carcinoma using the Cathetron," Br. J. Ra-
HALL, E. J. (1978), Radiobiology for the Radiologist diol. 40, 882.
(Harper & Row, Hagerstown, Maryland). PATERSON, R. (1948). The Treatment of Malignant
HENSCHKE, U. K, HILARIS, B. S., and MAHAN, G. D. Disease by Radium and X-Rays (Edward Arnold,
(1966). "Intracavitary radiation therapy of cancer of London).
18
Relerences ••. 19
PATERSON, R. and PARKER, H. M. (1934). "A dosage carcinoma," J. Eur. Radioth. (to be published in ·
system for gamma ray therapy," Br. J. Radiol. VII, 1985).
592. VON ESSEN, C. F. (1980). "Clinical application of the
PIERQUIN, B. (1964). Precis de Curietherapie, Endo- Brachytron: the San Diego technique for treatment
curietherapie et Plesiocurietherapie (Masson, of cancer of the cervix" p. 117 in High Dose-Rate
Paris). Afterloading in the Treatment of Cancer of the
TwIss, D. H. and BRADSHAw, A. L. (1970). "Automatic Uterus, Bates T. D. and Berry R. J., Eds., Br. J. Ra-
source positioning for Cathetron treatments," Br. J. diol. Special report 17.
Radiol. 43, 48. WALSTAM, R. (1954). "The dosage distribution in the
VAN DER LAARSE, R. (1981). "Computerized radiation pelvis in radium treatment of carcinoma of the cer-
treatment planning," Thesis, University of Am- vix," Acta Radiol. 42, 237.
sterdam. WAMBERSIE, A., PRIGNOT, M., and GUEULETTE, J.
VAN LIMBERGEN, E., CHASSAGNE, D., GERBAULET, (1973). "A propos du remplacement du radium par
A., and HAlE, C. (1985). "Different dose rates in Ie cesium-137 en curietherapie gynecologique," J.
preoperative endocavitary brachytherapy for cervical Radiol. Electrol. 54,261-270.

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


IeRU Reports
IeRU Reports are distributed by the IeRU Publications' office. Information
on prices and how to order may be obtained from:
IeRU Publications
7910 Woodmont Avenue, Suite 1016
Bethesda, Maryland 20814
U.S.A.
The currently available IeRU Reports are listed below.
ICRU
Report No. Title
lOb Physical Aspects of Irradiation (1964)

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


10c Radioactivity (1963)
lOf Methods of Evaluating Radiological Equipment and Ma-
terials (1963)
12 Certification of Standardized Radioactive Sources (1968)
13 Neutron Fluence, Neutron Spectra and Kerma (1969)
14 Radiation Dosimetry: X Rays and Gamma Rays with Max-
imum Photon Energies Between 0.6 and 50 MeV (1969)
15 Cameras for Image Intensifier Fluorography (1969)
16 Linear Energy Transfer (1970)
17 Radiation Dosimetry: X Rays Generated at Potentials of 5
to 150 k V (1970)
18 Specification of High Activity Gamma-Ray Sources
(1970)
20 Radiation Protection Instrumentation and Its Application
(1971)
22 Measurement of Low-Level Radioactivity (1972)
23 Measurement of Absorbed Dose in a Phantom Irradiated by
a Single Beam of X or Gamma Rays (1973)
24 Determination of Absorbed Dose in a Patient Irradiated by
Beams of X or Gamma Rays in Radiotherapy Procedures
(1976)
25 Conceptual Basis for the Determination of Dose Equivalent
(1976)
26 Neutron Dosimetry for Biology and Medicine (1977)
27 An International Neutron Dosimetry Intercomparison
(1978)
28 Basic Aspects of High Energy Particle Interactions and
Radiation Dosimetry (1978)
29 Dose Specification for Reporting External Beam Therapy
with Photons and Electrons (1978)
30 Quantitative Concepts and Dosimetry in Radiobiology
(1979)
31 Average Energy Required to Produce an Ion Pair (1979)
32 Methods of Assessment of Absorbed Dose in Clinical Use of
Radionuclides (1979)
33 Radiation Quantities and Units (1980)
34 The Dosimetry of Pulsed Radiation (1982)
35 Radiation Dosimetry: Electron Beams with Energies Be-
tween 1 and 50 MeV (1984)
36 Microdosimetry (1983)
37 Stopping Powers for Electrons and Positrons (1984)
20
ICRU Reports ..• 21
38 Dose and Volume Specification for Reporting Intracavitary
Therapy in Gynecology (1985)

Binders for ICRU Reports are available. Each binder will accommodate
from six to eight reports. The binders carry the identification, "ICRU Re-
ports," and come with label holders which permit the user to attach labels
showing the Reports contained in each binder.
The following bound sets of ICRU Reports are also available:
Volume I. ICRU Reports lOb, lOc, lOf
Volume II. ICRU Reports 12, 13, 14, 15, 16, 17, 18,20
Volume III. ICRU Reports 22, 23, 24, 25, 26
Volume IV. ICRU Reports 27, 28, 29, 30, 31, 32
Volume V. ICRU Reports 33, 34, 35, 36
(Titles of the individual Reports contained in each volume are given in the
list of Reports set out above.)

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


The following ICRU Reports were superseded by subsequent Reports and
are now out of print:
ICRU
Report No Title and Reference*
1 Discussion on International Units and Standards for X-ray
Work, Brit. J. Radio!. XXIII, 64 (1927).
2 International X-Ray Unit of Intensity, Brit. J. Radio!' (new
series) I, 363 (1928).
3 Report of Committee on Standardization of X-ray Mea-
surements, Radiology 22,289 (1934).
4 Recommendations of the International Committee for Ra-
diological Units, Radiology 23, 580 (1934).
5 Recommendations of the International Committee for Ra-
diological Units, Radiology 29, 634 (1937).
6 Recommendations of the International Commission on
Radiological Protection and of the International Com-
mission on Radiological Units, National Bureau of Stan-
dards Handbook 47 (U.S. Government Printing Office,
Washington, D.C., 1951).
7 Recommendations of the International Commission for
Radiological Units, Radiology 62,106 (1954).
8 Report of the International Commission on Radiological
Units and Measurements (ICRU) 1956, National Bureau
of Standards handbook 62 (U.S. Government Printing
Office, Washington, D.C., 1957).
9 Report of the International Commission on Radiological
Units and Measurements (ICRU) 1959, National Bureau
of Standards Handbook 78 (U.S. Government Printing
Office, Washington, D.C., 1961).
lOa Radiation Quantities and Units, National Bureau of Stan-
dards Handbook 84 (U.S. Government Printing Office,
Washington, D.C., 1962).
lOd Clinical Dosimetry, National Bureau of Standards Handbook
87 (U.S. Government Printing Office, Washington, D.C.,
1963).
lOe Radiobiological Dosimetry, National Bureau of Standards
Handbook 88 (U.S. Government Printing Office, Wash-
ington, D.C., 1963).
22 ... ICRU Reports

11 Radiation Quantities and Units (International Commission


on Radiation Units and Measurements, Washington, D.C.,
1968).
19 Radiation Quantities and Units (International Commission
on Radiation Units and Measurements, Washington, D.C.,
1971).
19S Dose Equivalent [Supplement to ICRU Report 19] (Inter-
national Commission on Radiation Units and Measure-
ments, Washington, D.C., 1973)
21 Radiation Dosimetry: Electrons with Initial Energies Be-
tween 1 and 50 Me V (International Commission on Ra-
diation Units and Measurements, Washington, D.C.,
1972).
• References given are in English. Some of the Reports were also published in other languages.

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Index
Absorbed dose at reference points, 11 Rectal reference point, 11
Absorbed-dose pattern, 5 Reference air kerma rate, 8, 17
Activity, 17 Reference dose level, 7,10
Manual afterloading, 5 Reference exposure rate, 17
Remotely controlled afterloading, 5 Reference isodose surface, 7
Apparent activity, 17 Reference points, 1-2,9, 11-12, 13
Applicator, 9 Bladder reference point, 11
Lymphatic trapezoid, 12
Bladder reference point, 11 Rectal reference point, 11
Points A and B, 1-2
Calculation of dose distribution, 12 Reference volume, 7, 9,10,11, 14
Conclusions and final recommendations, 16

Downloaded from http://jicru.oxfordjournals.org/ at University of Glasgow on July 9, 2016


Simulation of linear sources, 4, 9
Definition of terms and concepts currently used in intracavitary Special situations, 14
therapy, 4 Source specification, 17
Definition of the 60 Gy reference volume in special situations, 14 Activity, 17
Description of the technique, 9 Apparent activity, 17
Description of the volume, 10 Mass of radium, 17
Dose rate, 4-5 Milligram-radium equivalent, 17
High dose rate, 5 Reference air kerma rate, 17
Low dose rate, 4 Reference exposure rate, 17
Medium dose rate, 5 Specification, 8, 9
Duration of the treatment, 9, 15 Intracavitary application, 9
Radioactive sources, 8, 9, 17
Irradiated volume, 1,7 Target dose specification, 9
Volume specification, 9
Linear sources, 4, 9 Stockholm system, 1-2
Simulation, 4, 9 Systems, 1-2
Lymphatic trapezoid, 12 Manchester system, 1-2
Paris system, 1-2
Manchester system, 1-2 Stockholm system, 1-2
Milligram hours (mg·h), 1, 10
Target volume, 1,5
Organs at risk, 8, 9 Time-dose pattern, 9, 15
Total reference air kerma, 8, 9, 10, 17
Paris system, 1-2 Treatment techniques, 4, 5, 9
Pear shaped volume, 11 Afterloading techniques, 5
Pelvic-wall reference point, 12-13 Treatment volume, 1, 7

Radioactive sources, 4, 8, 9 Volume, 1,5,7,9, 10-11


Radiobiological considerations, 15 Irradiated volume, 7
Radium substitutes, 4 Pear shaped volume, 11
Recommendations for reporting absorbed doses and volumes in Reference volume, 7, 9, 10, 11
intracavitary therapy, 9 Target volume, 1, 5
Recommendations for reporting time-dose pattern, 15 Treatment volume, 1,7
Rectal dose rate (measurement), 12 Volume specification, 9

23

You might also like