You are on page 1of 19

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/334497706

Experience in commissioning The Halcyon linac

Article  in  Medical Physics · July 2019


DOI: 10.1002/mp.13723

CITATIONS READS

18 2,742

17 authors, including:

Yuting Li Song Gao


University of Texas MD Anderson Cancer Center University of Texas MD Anderson Cancer Center
20 PUBLICATIONS   220 CITATIONS    73 PUBLICATIONS   759 CITATIONS   

SEE PROFILE SEE PROFILE

Laurence E Court Lei Dong


University of Texas MD Anderson Cancer Center University of Pennsylvania
405 PUBLICATIONS   6,096 CITATIONS    484 PUBLICATIONS   14,134 CITATIONS   

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

“Archimedean TEI ATHENS EPEAEK II” with title: «Automatic Diagnosis System of Thyroid Nodules Cancer "Department of Medical Instruments Technology T.E.I. Athens.
View project

Ultra-high-dose rate (FLASH) proton therapy View project

All content following this page was uploaded by Lei Dong on 13 January 2020.

The user has requested enhancement of the downloaded file.


Experience in commissioning the halcyon linac
Tucker Netherton, Yuting Li, Song Gao, Ann Klopp, Peter Balter, and Laurence E Court
Department of Radiation Physics, University of Texas MD Anderson Cancer Center, Houston, TX 77030, USA
Ryan Scheuermann, Chris Kennedy, Lei Dong, James Metz, and Dimitris Mihailidisa)
Perelman Center for Advanced Medicine, University of Pennsylvania, Philadelphia, PA 19104, USA
Clifton Ling, Mu Young Lee, Magdalena Constantin, Stephen Thompson, Juha Kauppinen,
and Pekka Uusitalo
Varian Medical Systems Inc, Palo Alto, CA 94304, USA

(Received 15 March 2019; revised 10 June 2019; accepted for publication 1 July 2019;
published 27 August 2019)
Purpose: This manuscript describes the experience of two institutions in commissioning the new
HalcyonTM platform. Its purpose is to: (a) validate the pre-defined beam data, (b) compare relevant
commissioning data acquired independently by two separate institutions, and (c) report on any signif-
icant differences in commissioning between the Halcyon linear accelerator and other medical linear
accelerators.
Methods: Extensive beam measurements, testing of mechanical and imaging systems, including the
multi-leaf collimator (MLC), were performed at the two institutions independently. The results were
compared with published recommendations as well. When changes in standard practice were necessi-
tated by the design of the new system, the efficacy of such changes was evaluated as compared to
published approaches (guidelines or vendor documentation).
Results: Given the proper choice of detectors, good agreement was found between the respective
experimental data and the treatment planning system calculations, and between independent measure-
ments by the two institutions. MLC testing, MV imaging, and mechanical system showed unique
characteristics that are different from the traditional C-arm linacs. Although the same methodologies
and physics equipment can generally be used for commissioning the Halcyon, some adaptation of
previous practices and development of new methods were also necessary.
Conclusions: We have shown that the vendor pre-loaded data agree well with the independent mea-
sured ones during the commission process. This verifies that a data validation instead of a full-data
commissioning process may be a more efficient approach for the Halcyon. Measurement results
could be used as a reference for future Halcyon users. © 2019 American Association of Physicists in
Medicine [https://doi.org/10.1002/mp.13723]

Key words: beam data verification, commissioning, Halcyon, TPS verification.

1. INTRODUCTION version (Halcyon Version 1.1) and repeated commissioning


measurements. The linac at MDACC was a non-clinical,
This study describes the combined experiences of two institu- research only machine. University of Pennsylvania success-
tions in commissioning a recently released radiotherapy treat- fully treated the first patient on September 14, 2017 using the
ment device which has some significant differences production version and has treated over 250 patients to date.
compared with conventional C-arm Linacs. Prior to its clini- The goal of this work is to: (a) validate the pre-defined beam
cal release, the vendor (Varian Medical Systems, Palo Alto, data, (b) compare relevant commissioning data acquired from
CA) contracted and worked with two academic institutions, two separate institutions, and (c) report on any significant dif-
the University of Texas MD Anderson Cancer Center ferences in commissioning between the Halcyon linear accel-
(MDACC) and the Hospital of University of Pennsylvania erator and other medical linear accelerators.
(UPENN), for independent clinical validation of a pre-release
version of the new treatment machine and pre-release version
2. BACKGROUND
of the Eclipse treatment planning software. The research was
conducted independently without any communication The main features of this new treatment device are sum-
between the two institutions until two months prior to the marized in Table I. It has an enclosed, ring-mounted gantry
public announcement of the new system. Then, all commis- and a 6 MV flattening filter free (FFF) beam. This beam is
sioning data gathered by the two institutions were compared. used for both treatment and imaging for Halcyon version 1.0.
The new linac was named as HalcyonTM Version 1.0 (Varian Future versions of Halcyon (Halcyon 2) will include a kV
Medical Systems, Palo Alto, CA). Later, UPENN replaced imaging system. Because the gantry is enclosed, it is allowed
the pre-release version of Halcyon system with the production to rotate significantly faster than on conventional C-arm

4304 Med. Phys. 46 (10), October 2019 0094-2405/2019/46(10)/4304/10 © 2019 American Association of Physicists in Medicine 4304
4305 Netherton et al.: Commissioning the Halcyon 4305

TABLE I. Features of the Halcyon treatment device (with Eclipse treatment Association of Physicists in Medicine (AAPM) and IAEA
planning system). documents2–11. A large portion of the test results (common to
both Halcyon and other linacs) are available via an electronic
Halcyon Feature Description
supplement. The characteristics of the clinical (1.1) and pre-
Beam model Pre-defined clinical (1.0) units are very similar, with the only exception
Ring mounted gantry 15 s rotation time being that the MLC was redesigned for the clinical (1.1) unit
Enclosed gantry bore 100 cm diameter (tongue-and-groove vs no tongue-and-grove). Furthermore,
Treatment and imaging beam 6 MV FFF because the physical design of this treatment device is differ-
Nominal dose rate 800 MU/min ent from previous Varian linacs, we have included some of
Dual layer, stacked MLC design Dynamic MLC, 5 cm/s leaf speed our specific experience in the setup and use of QA devices
Maximum field size 28 9 28 cm2 for this specific unit.
Varian IGRT Couch Coplanar with mounted camera
Alignment at isocenter IGRT only, no light field, no ODI
3. MATERIALS AND METHODS
3.A. Independent beam data validation
treatment devices – with a maximum rotation speed of 360°
in 15 s. The Halcyon 1.0 linac only permits the delivery of The performance and stability of the Varian TrueBeam is
dynamic multi-leaf collimator (MLC) plans in clinical mode well documented across multiple institutions.12,13 The cre-
(field-in-field, IMRT, or RapidArcâ, or irregular surface ation of a RBD set is useful as (a) a robust method of ensur-
compensator), and has a dual-layered, stacked MLC that pro- ing general consistency across machines of the same design
jects to a maximum field size of 28 cm2 9 28 cm2 at isocen- and (b) a reliable second check for institutions that perform
ter. Each leaf is 1 cm wide with 58 upper and 56 lower MLCs their own modeling. It has yet to be shown how the perfor-
for a total of 114 leaves. The top layer is displaced by 0.5 cm mance of the Halcyon varies from institution to institution
laterally relative to the lower layer1 to reduce linter-leaf leak- while using its own unique, and vendor-supplied RBD data.
age. The leaves have a maximum speed of 5.0 cm/s, twice In the present study, beam data (beam profiles, percent-depth
that of current Varian MLCs. There are no moving collimat- dose, output factors, and off-axis ratios) were measured and
ing jaws because the dual-stack MLC design is designed to compared. In addition, the measured results were compared
provide adequate shielding and allows for “per-leaf tracking” to the vendor-supplied beam measurements.
as opposed to the jaw tracking feature in other C-arm linacs.
In regard to the typical workflow on the Halcyon, the
3.A.1. Percent depth dose measurements
patient (or physics QA equipment) is first positioned using
lasers outside of the Halcyon gantry housing. The couch is The percent depth dose (PDD) of 6 MV FFF photon
then translated into the bore by a specific distance, moving beams for different field sizes using CC13 ionization cham-
the patient to the treatment position. MV images are then bers (IBA) and diodes (SunNuclear) were measured in the
acquired to verify that the patient is accurately positioned for Blue2 water tank (IBA). Because the diameter of the treat-
treatment. As there is neither light field, optical-distance-indi- ment bore is 100 cm, some other scanning water tanks may
cator, nor mechanical distance measurement device, accurate be too large to use for measurements on the Halcyon. In addi-
patient set-up relies largely on MV imaging. In-line of the tion, the water tank cannot be aligned with lasers, light field,
radiation source, on the opposite side of the gantry is an elec- or mechanical measuring device at isocenter. The user must
tronic portal imaging device (EPID) positioned in front of a first align the water tank to lasers at an external setup posi-
beam stopper. In the versions that we tested (v1.0 and 1.1), all tion, then shift the tank to isocenter for subsequent alignment
treatments were image-guided, using either MV orthogonal using image guidance. When the desired source to water sur-
projections or MV CBCT. For each image method, there is a face distance (SSD) is 100 cm, the air-to-water interface can
choice of “low-dose” or “high-quality” mode. Since, the dose be easily visualized by taking images with the gantry at either
from this imaging beam is included in the treatment plan, the 90 or 270 degrees in service mode. The 100 cm SSD is
choice of the imaging technique must be made before treat- achieved by adjusting the table height until the air-water inter-
ment planning commences. face is aligned with the crosshair overlay of the isocenter
Varian provides a representative beam data set (RBD), location in the service mode images. However, when the
consisting of percentage depth doses (PDDs), profiles and desired SSD is not 100 cm, measuring the appropriate table
output factors. This RBD set is used by Varian to create the shifts from images is complicated due to beam divergence
pre-configured Halcyon optimization and dose calculation and the blurring of the air-to-water interface (Note: The Var-
models (PO & AAA) in Eclipse 15.1.1. treatment planning ian user manual thoroughly describes this setup procedure for
system (TPS)1. all users, we recommend that this resource be used). To
This study describes the pre-clinical testing of the Halcyon achieve an SSD of 90 cm, the gantry is positioned at 84.3
1.0 (two units, one at each institution), and the subsequent degrees, and the table height is adjusted until the distance
testing of a clinical version of the device at UPENN. The test- between the isocenter projection in the image and the sharp
ing is exhaustive and largely based on relevant American air-to-water interface is 10.05 cm. This distance is measured

Medical Physics, 46 (10), October 2019


4306 Netherton et al.: Commissioning the Halcyon 4306

using the measuring tools available in service mode and range of field sizes: 2 9 2, 4 9 4, 6 9 6, 8 9 8, 10 9 10,
relies on a trigonometric relationship indicated in the user 20 9 20, and 28 9 28 cm2. Measurements were carried
manual. Further alignment of the ion chamber’s measurement out at five different depths: 1.3 cm (dmax), 5 cm, 10 cm,
volume to machine isocenter was achieved using the Blue2 20 cm, and 30 cm. The beam profiles were also measured
Phantom central axis measurement utility, and verified with at five different depths along the diagonal direction of the
an anterior-posterior EPID image using ~ 25–50 MU. The largest field size 28 9 28 cm2 at SSD of 90 cm. The
OmniPro –Accept software (IBA) was used to analyze the choices of detector for beam profiles were CC13 for FS
resulting scanning data. Percent depth dose curves for field >6 9 6, diode for FS <=6 9 6 (MDACC), and CC04 for
sizes: 2 9 2, 4 9 4, 6 9 6, 8 9 8, 10 9 10, 20 9 20, and all field sizes (UPENN). Representative beam data set was
28 9 28 cm2 were compared on a point-by-point basis scanned with a CC13 for FS >4 9 4 and diode for FS<=
against the RBD from each institution for depths beyond 4 9 4. For the measured beam profiles, the data were first
dmax. The measurement devices used were CC13 and diodes smoothed then corrected for central axis discrepancies and
(<6 9 cm2) at MDACC, CC04 at UPENN, and CC13 and made symmetrical. The beam profiles were then renormal-
diode (<8 9 8 cm2) at Varian for the RBD, respectively. ized to 100% by the values at central axis. A point-by-point
Comparison against TPS PDDs were completed in section G comparison between institutional and RBD was used to
(beam model verification). evaluate beam profiles. High dose regions, penumbra, and
low dose tails were individually analyzed using point-by-
point and distance to agreement at every field size for
3.A.2. Dose in the buildup region
depths 1.3 cm and 10.0 cm (cross-plane and in-plane). To
For both institutions, dose in the buildup region was mea- define each region of the profile, inflection points were
sured in a solid water phantom using parallel-plate chamber found by plotting the first derivatives of the beam profiles
(Table II). The source-to-chamber distance was 100 cm and for each field size (Fig. 1). The full-width-half-maximum of
dmax was searched for different field sizes. Various thick- each first derivative curve was used to define penumbra,
nesses of the solid water blocks were used for depth dose low-dose tail, and high dose regions. Since the 80–20%
measurements. Measurements at different depths were nor- definition of penumbra is only directly applicable to flat-
malized to the measured dose at dmax. Since the polarity tened profiles, these regions must be explicitly defined in
effect of the Advanced Markus parallel-plate chamber is very order for MPPG5a guidelines to be applied. These regions
large in the buildup region (MDACC), measurements were were analytically determined for the RBD and for the TPS
done in both +300 V and 300 V, and the results averaged. data and were used in sections 4.A and 4.D.
Flatness and symmetry were calculated from in-plane (ra-
dial) and cross-plane (transverse) scans for 2 9 2 cm2 to
3.A.3. Beam profiles
28 9 28 cm2 field sizes at an SSD of 90 cm and different
Beam profiles were measured at different depths along depths in water. Although flatness is recommended by
in-plane and cross-plane directions at SSD of 90 cm for a AAPM for flattened beams, flatness is not an appropriate

TABLE II. Institutional measurement device Table.

Type Model Institution Tests Performed

Cylindrical ion chambers IBA CC04 UPENN All PDD, all beam profiles
IBA CC04 MDACC ≤6x6 cm2 output factors, phantom imaging dose
IBA CC13 MDACC ≥6x6 cm2 output factors, ≥4x4 cm2 PDD and beam profiles
IBA CC13 UPENN TG 119 testing
IBA CC13 Varian ≥6 9 6 cm2 PDD and beam profiles, all FS output factors
Exradin A12 UPENN TG51
PTW TN30013 MDACC TG51, monthly output standard, DLG and transmission
Diodes Sun Nuclear EDGE MDACC ≤4 9 4 cm2 PDD and ≤6 9 6 beam profiles
Sun Nuclear EDGE Varian <6 9 6 cm2 PDD
Array Detectors Sun Nuclear QailyQA3 MDACC Daily output standard
Sun Nuclear IC Profiler MDACC Output vs gantry, flattness and symmetry
Sun Nuclear Map Check 2 UPENN IMRT patient specific QA
Sun Nuclear Arc Check MDACC, UPENN RapidArc patient specific QA
Portal Dosimetry MDACC, UPENN IMRT and RapidArc patient specific QA
Plane Parallel PTW 23343 Markus UPENN Surface dose
PTW Advanced Markus MDACC Surface dose
PPC40 UPENN DLG and transmission

DLG, Dosimetric leaf gap; MDACC, MD Anderson cancer center; UPENN, university of pennsylvania.

Medical Physics, 46 (10), October 2019


4307 Netherton et al.: Commissioning the Halcyon 4307

FIG. 1. Methodology to determine dose regions for a FFF beam. Field size = 10 9 10 cm2, depth = 10 cm. Red curve = TPS, Black curve = RBD using a
CC13 ionization chamber. Blue region = low-dose tails, White region = penumbra, Red region = high-dose region. Here, regions are based on TPS (red). FFF,
flattening filter free; RBD, representative beam data set; TPS, treatment planning system. [Color figure can be viewed at wileyonlinelibrary.com]

metric for this FFF beam. Instead, off axis ratio (including etc.), it presents challenges for testing by restricting access
diagonal normalized flatness) is the appropriate metric to to machine components and preventing direct measure-
characterize the open field beam quality. ment. Nearly all information regarding mechanical perfor-
mance relative to treatment isocenter must be derived from
irradiation of film, EPID imaging, or other phantom-based
3.A.4. Output factor measurements
radiation measurement.
To verify Halcyon RBD set, output factors were measured An additional consideration when evaluating a QA pro-
with the ionization chamber set at a depth of 5.0 cm and SSD gram for Halcyon is the implementation of the machine
at 95 cm in the Blue2 water phantom. The SSD was set using performance check procedure (MPC),14 which must be exe-
the procedure in the Varian “Instructions for Use” manual. cuted daily prior to treatment, as part of the Halcyon
The measured results for different field sizes were normalized workflow. The MPC utilizes a drum phantom with imbed-
to that of a 10 9 10 cm2 field to yield the output factors. A ded fiducials mounted to a specific position on the treat-
CC04 ion chamber was used for field sizes less than ment couch. The MPC automatically sets various machine
6 9 6 cm2, CC13 ion chamber was used for field sizes parameters and acquires 33 radiation images with the
greater than 6 9 6 cm2 (MDACC). The RBD used a CC13 EPID. Based on these measurements, the MPC software
ion chamber for all field sizes. performs analysis and verifies that the machine’s dosimet-
ric and mechanical performance are within their respective
tolerances. Another version of the MPC has been shown
3.B. Mechanical system validation against
previously to provide excellent results and long-term stabil-
specifications
ity for QA of TrueBeam.14 We have also carried out mea-
Due to the similarities in the method of treatment deliv- surements, in which errors were introduced intentionally,
ery, Halcyon mechanical testing requirements are generally to test the ability of the MPC in detecting the errors.15
consistent with C-arm linacs.3 While Halcyon’s bore The methods described below provide an independent
geometry eliminates the need to test physical indicators of check of the necessary QA procedures to ensure accurate
the treatment field (optical distance indicator, light field, treatment delivery.

Medical Physics, 46 (10), October 2019


4308 Netherton et al.: Commissioning the Halcyon 4308

available with the Halcyon (including different isocenter posi-


3.B.1. Lasers (virtual-iso) versus machine treatment
tion and field size combinations). The phantom was then posi-
isocenter
tioned at the Halcyon and the imaging doses measured for the
All lasers - lateral (left and right), and vertical (top) - were same conditions using a cross-calibrated CC04 ion chamber.
checked against the virtual treatment isocenter. The Winston- The experimentally measured doses were compared with
Lutz (WL) phantom was used for these tests. The tests were those calculated by the Eclipse. For more details on our
done as follows: (a) Align the center of the ball with the exhaustive studies into the imaging dose with the Halcyon,
lasers, (b) move couch into the treatment isocenter position readers are referred to Li et al.17 and Mihailidis et al.18.
by applying the machine’s automated shift from virtual to
treatment isocenter, and record the actual couch coordinates.
3.D. Beam model verification, end-to-end testing,
(c) Align the center of WL ball to the treatment isocenter
and treatment plan validation
(digital crosshair) by adjusting the couch position using
orthogonal images—final couch coordinates were compared As part of the overall dosimetric performance validation of
to that obtained from the “load” operation. the Halcyon, various aspects of the MPPG5a testing were
performed to ensure that the pre-defined beam model calcula-
tions matched commissioning measurements. The equipment
3.B.2. Gantry and collimator angle indicators
used for validation was taken from the recommendations of
As the Halcyon’s gantry is not directly accessible to its user, Tables I and II of MPPG5a.6 Four categories of tests included
(no light field), new methods need to be devised to evaluate in report are TPS model comparison, basic photon beam vali-
the accuracy of the gantry angle indicator. Two methods were dation, heterogeneous photon beam validation, and RA/
evaluated: (a) gantry spoke shot referenced to bubble level IMRT validation.
indicator, and (b) observation of fiducial movement in portal For TPS model comparisons the tests performed were: (a)
images at various couch vertical positions. In the first test, a dose in test plan vs clinical calibration condition, and (b) dose dis-
horizontal line was marked on a radiochromic film (EBT3) tribution calculated in planning system vs. commissioning data
with the aid of a bubble level indicator, prior to star shot irradi- (PDD, off-axis output factors). In a situation where the user can-
ation at various gantry angles. The film was digitized, and the not obtain a set of measured beam data to compare with the TPS,
angle of the spokes relative to the bubble reference line were the RBD can be requested from the vendor. For PDD and beam
evaluated using the measuring tools in the ImageJ software.16 profile data comparisons against the TPS, MPPG5a recom-
The measured gantry angles were then compared to the corre- mended distance to agreement for penumbra (3 mm), high dose
sponding values of gantry positions indicators. region point dose agreement (2.0%), and low-dose profile tails
In the second test, where the gantry angle is evaluated based point dose agreement (3%) criteria were used. Dose region speci-
on fiducial image projections under couch motion, the portal fications for these flattening-filter-free beam profiles were calcu-
imager, a WL pointer phantom, variable couch positions, and lated as mentioned in section B (independent beam validation).
in-house software were used. This test’s goal is to address the The following basic beam validation tests were performed:
accuracy of the angle given by the digital gantry angle indica- (a) PDD and profile comparison of small square MLC shaped
tor. Since this test was found to be sensitive to setup error, the fields, (b) PDD and profile comparison of asymmetric MLC
experiment was repeated with intentional table setup perturba- shaped fields, (c) point dose and profile comparison of large
tions of 1 mm in the direction sensitive to each measurement. field with extensive MLC blocking, (d) point dose and profile
Cardinal gantry angle was calculated using the portal images comparison of half-beam block at minimally anticipated SSD,
of a WL pointer as a function of table height. and (e) point dose comparison of oblique incident beam.
Collimator angle indicator accuracy was assessed by ana- Exported doses from the Eclipse TPS used a 0.25 cm grid size.
lyzing EPID images of long narrow fields defined by MLCs The following tests were performed for heterogeneity cor-
acquired with the collimator digitally set at 0° and 90°. An rection validation: (a) Point dose comparison (with AAA for
in-house (MDACC) machine QA software was used for ana- treatment dose) using an anthropomorphic phantom (CIRS)
lyzing the images. and (b) point dose comparison (with modified AAA used for
imaging dose) using an anthropomorphic phantom (CIRS).
Dosimetric leaf gap (DLG) measurements were performed
3.C. Imaging systems validation and performance
for the Halcyon MLC system prior to any dynamic delivery
Evaluation of the imaging performance of the Halcyon testing and was performed by using the procedure and dicom
closely followed standard practice (as defined in MPPG2a7) files provided by the manufacturer. Finally, for IMRT/RA dose
to establish baseline values, as expected for any medical lin- validation the tests performed were: (a) AAPM TG-119 testing,
ear accelerator. However, the imaging dose evaluation (b) clinical tests (end-to-end), and (c) external review using
revealed some interesting features for Halcyon and its testing IMRT and RA (IROC end-to-end). Output factor and PDD ver-
is described below. ification for very small MLC field sizes were also performed
A CT of an anthropomorphic thorax phantom (CIRS) was in the above sections, since all fields on the Halcyon are
imported into the Eclipse TPS and the imaging doses calcu- defined by the MLC. Point measurements described in AAPM
lated for 11 different points for each of the imaging techniques TG 1198 were performed using an IBA CC13 cylindrical ion

Medical Physics, 46 (10), October 2019


4309 Netherton et al.: Commissioning the Halcyon 4309

chamber in a solid water phantom. All plans were generated For a range of depths (1–15 mm) and square field sizes
using both IMRT and RA delivery techniques. Measured dose (6 9 6, 10 9 10, and 20 9 20 cm2), ratios of surface doses
values were compared against the average dose calculated in between institutions were at worst 1.007.
Eclipse within the chamber collecting volume.
For the external review process, we created plans and irra-
4.A.3. Beam profiles
diated the anthropomorphic head (RA plan) and thorax phan-
tom (IMRT plan) from IROC-Houston, following the RBD profile data is presented here for all field sizes mea-
standard guidelines. Patient-specific QA measurements were sured (2 9 2 cm2 to 28 9 28 cm2) at depths 1.3 cm and
also performed at UPENN for all plans using either the Sun 10.0 cm. The vendor used crossline data to model the TPS.
Nuclear Corporation MapCHECKâ 2 (IMRT), ArcCHECKâ Data presented from each institution had crossline and in-line
(RA) diode arrays, or portal dosimetry. Distance to agreement measurements averaged to compare with RBD. The dose
(DTA) analysis was performed on all measurements using regions (high-dose, penumbra, and low-dose) were based on
two methods: 1) 3%/3mm, global normalization, 10% thresh- the location of the inflection points and width of the penum-
old, and 2) 2%/2 mm, local normalization, 10% threshold. bra from the RBD set. The maximum point difference
between the RBD and measured data in the high dose region
is as follows for various field sizes at each institution for
4. RESULTS depth = 1.3 cm. These results are also featured above in
Table III. All high-dose region doses matched RBD high-dose
4.A. Independent Beam Data validation region doses within 1.6% for both institutions. All maximum
point dose differences were within the penumbra region. In
4.A.1. Percent depth dose measurements
the RBD Data set for depth = 1.3 cm, the penumbra ranged
For all field sizes measured (2 9 2 cm2 to 28928 cm2), from 0.3 cm to 0.7 cm for field sizes 2 9 2 to 28 9 28 cm2.
PDD curves from both institutions matched the vendor pro- The measured penumbra regions matched RBD penumbra
vided RBD within 1.0% (step size = 1 mm). For field sizes regions within 2 mm distance to agreement for all field sizes
greater than 2 9 2 cm2, this maximum difference was 0.6%. at depth = 1.3 cm. Low dose tails were within 0.9% (maxi-
The maximum point difference between the RBD and mea- mum point difference at all points) of RBD low dose tails for
sured data is as follows for various field sizes (cm2) at each both institutions, except for one point (5.3%, UPENN) for a
institution and is featured in Table III: (for MDACC and point on the edge of the penumbra (2 9 2 cm2).
UPENN). The maximum point difference between the RBD and
measured data in the high dose region is as follows for vari-
ous field sizes (cm2) at each institution for depth = 10.0 cm.
4.A.2. Buildup region dose
These results are also featured above in Table III. All highest
Dose in the buildup region was measured in a solid block maximum point dose differences were within the penumbra
phantom using a parallel-plate chamber (both institutions). region. In the RBD set for depth = 10.0 cm, the penumbra
The source-to-chamber distance was 100 cm and dmax was ranged from 0.4 cm to 0.8 cm for field sizes 2 9 2 to
searched for different field sizes. Various thicknesses of the 28 9 28 cm2. The measured penumbra regions matched
blocks were used for surface depth dose measurements. Mea- RBD penumbra regions within 2 mm distance to agreement
surements at different depths were normalized to the mea- for all field sizes at depth = 10.0 cm. Low dose tails were
sured dose at dmax. within 1.7% (maximum point difference at all points) of
RBD low dose tails for both institutions, except for one point
(4.7%, UPENN) for a point on the edge of the penumbra
TABLE III. Maximum percent difference, measured data vs RBD. (2 9 2 cm2).

Beam profile Beam profile


High dose Region High dose Region 4.B. Output factor measurements
PDD d = 1.3 cm d = 10.0 cm
For all symmetric and asymmetric output factors mea-
FS [cm2] MDACC UPENN MDACC UPENN MDACC UPENN sured, the maximum ratios between the measured data and
the RBD were 1.008 and 1.009 for each institution (MDACC,
2 –0.4 –1 –1.3 –1.6 –1.3 –2.1
UPENN), respectively, and for field sizes >2 9 2 cm2.
4 –0.4 –0.3 –0.5 –0.3 –0.6 –0.5
6 –0.3 0.4 –0.7 –0.1 –0.8 –0.1
8 –0.3 –0.6 0.3 –0.2 –0.3 –0.3
4.C. Mechanical system validation
10 –0.4 –0.4 0.4 –0.3 0.5 –0.1
20 0.3 0.3 0.7 –0.4 0.7 N/A 4.C.1. Lasers (virtual-iso) versus machine treatment
28 –0.3 0.5 1.0 0.4 0.7 N/A isocenter

MDACC, MD Anderson cancer center; PDD, percentage depth dose; RBD, repre- Each of the three lasers indicating the virtual isocenter
sentative beam data set; UPENN, university of pennsylvania. was tested independently. All virtual isocenter laser axes were

Medical Physics, 46 (10), October 2019


4310 Netherton et al.: Commissioning the Halcyon 4310

found to indicate the treatment isocenter position to within


4.E. Beam model verification, end-to-end testing,
1 mm accuracy.
and treatment plan validation
For beam model validation, the MPPG5a recommended
4.C.2. Gantry and collimator angle indicators
limit for measured dose distributions matching the TPS is 2%
Comparison of the measured gantry angles from the spoke within the high dose region, 3 mm distance to agreement
shot on the EBT3 film, referenced to marked bubble level within the penumbra region, and 3% in the low-dose tails
line, to the digital readout are within 0.2°. region. Detector choices as a function of institution and FS
Gantry angle indicator accuracy was also tested by are specified in Table II. For all field sizes measured
observing fiducial movement in portal images under verti- (2 9 2 cm2 to 28 9 28 cm2), PDD curves from both institu-
cal couch motion. On average, gantry angles were veri- tions matched the exported TPS data within a maximum
fied to be within 0.19° of expected gantry angles. When point difference of 1.0% (step size = 1 mm). For field sizes
1 mm setup error was introduced, gantry angles were cal- greater than 2x2 cm2, this maximum point difference was
culated to be within 0.49° of expected angles, on aver- 0.7%. The maximum point difference between the TPS data
age, showing that, even in the presence of a small setup and measured PDD data (Table III) is as follows for various
error, this test is reasonable for checking that the gantry field sizes (cm2) at each institution.
angles are within specifications. For the collimator angle TPS profile data is presented here for all field sizes mea-
indicator accuracy testing, the results show that the mea- sured (2 9 2 cm2 to 28 9 28 cm2) at depths 1.3 cm and
sured collimator angles are 0.13° (different from digital 10.0 cm. Data presented from each institution had crossline
setting by 0.13°) and 89.86° (different from digital setting and in-line measurements averaged to compare with TPS data.
by 0.14°). The dose regions (high-dose, penumbra, and low-dose) were
based on the location of the inflection points and width of the
penumbra from the TPS data set. The maximum point differ-
4.D. 4.3. Imaging system validation and
ence between the TPS data and measured data (Table IV) in
performance
the high dose region is as follows for various field sizes (cm2)
For high-dose MV-MV mode, the mean extra-target at each institution for depth = 1.3 cm. All high-dose region
doses to the heart, left lung, right lung and spine were doses matched TPS high-dose region doses within 3.3% for
1.18, 1.64, 0.80 and 1.11 cGy per image set, respectively. both institutions on a point-by-point basis. All highest maxi-
The mean in-target doses were 3.36, 3.72, 2.61 and mum point dose differences were within the penumbra region.
2.69 cGy per image set, respectively. For high-dose MV- In the TPS data set for depth = 1.3 cm, the penumbra ranged
CBCT mode, the mean imaging doses to the heart, left from 0.3 cm to 0.5 cm for field sizes 2 9 2 to 28 9 28 and
lung, right lung and spine were 8.45, 7.16, 7.19 and was tighter than the penumbra from the RBD and institutional
6.51 cGy per image, respectively. The accuracy of TPS measured data. The measured penumbra regions matched
calculated imaging dose compared to measured dose, on TPS penumbra regions within 1 mm distance to agreement
the anthropomorphic phantom gave average differences of for all field sizes at depth = 1.3 cm. Low dose tail point dif-
0.05, 0.35, 0.35, 0.09 and 0.01 cGy for the heart, ference from the TPS (Table IV) is as follows for both institu-
left lung, right lung, spine, and breast, respectively. For tions for depth = 1.3 cm.
more details on our exhaustive study into the imaging The maximum point difference between the TPS data and
dose with the Halcyon, readers are referred to Li et al.17 measured data (Table IV) in the high dose region is as follows
and Mihailidis et al.18 for various field sizes (cm2) at each institution for

TABLE IV. Maximum percent difference, measured data vs TPS.

Beam profile Beam profile Beam profile Beam profile


High dose region High dose region Low dose tails Low dose tails
PDD d = 1.3 cm d = 10.0 cm d = 1.3 cm d = 10.0 cm
2
FS [cm ] MDACC UPENN MDACC UPENN MDACC UPENN MDACC UPENN MDACC UPENN

2 0.3 –1.0 –1.3 –0.1 –1.1 –3.3 0.0 0.4 –2.7 6.8
4 –0.3 0.4 0.4 –0.3 –1.4 –4.9 –2.9 1.0 –3.8 3.8
6 –0.4 0.3 –1.0 –2.5 –1.0 –4.7 –4.3 4.8 –2.6 4.0
8 0.6 –0.7 –0.8 –0.3 –5.8 –4.2 4.1 –0.5 8.9 3.2
10 0.5 0.4 –2.6 –1.4 –4.1 –4.2 –6.0 –1.8 –9.0 –2.4
20 0.6 0.4 1.1 0.7 –3.3 N/A 3.1 0.5 –2.4 N/A
28 0.6 0.7 –3.3 –1.6 –2.9 N/A 5.4 2.2 6.8 N/A

MDACC, MD Anderson cancer center; PDD, percentage depth dose; TPS, treatment planning system; UPENN, university of pennsylvania.

Medical Physics, 46 (10), October 2019


4311 Netherton et al.: Commissioning the Halcyon 4311

depth = 10.0 cm. When a diode was used, point-dose agree- TABLE VII. Results of the IROC independent evaluation of the delivered dose
ment was within 1.4% in the high dose-region for field sizes profiles (head and neck phantom).
2 9 2 to 6 9 6 cm2 (MDACC). All high-dose region doses
Film plane Gamma indexa Criteria Acceptable
matched TPS high-dose region doses within 5.8% for both
institutions on a point-by-point basis for larger field sizes Axial 99% 85% Yes
using ion chambers (CC13 and CC04). All highest maximum Sagittal 99% 85% Yes
point dose differences were within the penumbra region. In a
Percentage of points meeting Gamma index criteria of 7% and 4 mm.
the TPS data set for depth = 10.0 cm, the penumbra was
0.3 cm for all field sizes. The measured penumbra regions
matched TPS penumbra regions within 1 mm distance to TABLE VIII. Results of the IROC independent evaluation of the delivered
agreement for all field sizes at depth = 10.0 cm. Low dose point doses (thoracic phantom, IMRT plan).
tail point difference from the TPS (Table III) is as follows for
both institutions for depth = 10.0 cm. Location IROC-H vs. Inst. Criteria Acceptable
The results for DLG and MLC transmission for the distal PTV_TLD_sup 0.97 0.92–1.05 Yes
leaf bank are shown in Table V for both institutions along PTV_TLD_inf 0.97 0.92–1.05 Yes
with the values in the pre-configured treatment planning sys-
tem.
The doses delivered to the IROC head and neck phantom
TABLE IX. Results of the IROC independent evaluation of the delivered dose
using a RA plan, as assessed by IROC-Houston, are shown in profiles (thoracic phantom).
Tables VI and VII. The doses delivered to the IROC thoracic
phantom using an IMRT plan, as assessed by IROC-Houston, Film Plane Gamma indexa Criteria Acceptable
are shown in Tables VII and VIII. The phantom irradiation
Axial 98% 80% Yes
results listed in the table above do meet the criteria estab-
Coronal 99% 80% Yes
lished by IROC Houston in collaboration with the coopera-
Sagittal 99% 80% Yes
tive study groups.
Average over 3 planes 99% 85% Yes
Regarding TG-119, confidence limits for cumulative dose
in high and low dose regions were within 3% for IMRT and a
Percentage of points meeting Gamma index criteria of 7% and 5 mm.
RA (Table IX). A negative mean dose (lower measured dose
than calculated) was found for points in the high dose region
for both IMRT and RA (Table X). TABLE X. TG 119 Results.
Patient specific QA results averaged 100% and 99.6% of
IMRT RA
points passing 3%, 3 mm gamma analysis with global nor-
malization for IMRT and RA, respectively. When analyzed High dose Low dose High dose Low dose
using the more stringent 2%, 2 mm, gamma with local nor-
Mean –0.87% 0.08% –0.60% –0.08%
malization, the results averaged 95.9% and 94.4% of points
Standard Deviation 1.08% 0.57% 0.99% 0.82%
Confidence Limit 3.00% 1.19% 2.53% 1.69%
TABLE V. Dynamic leaf gap and MLC transmission measurements.
Mean, standard deviation, and confidence limit of percent difference between cal-
UPENN UPENN culated and measured ion chamber doses relative to prescription dose for IMRT
MDACC Prototype clinical Eclipse and RA TG 119 plans in high and low dose regions.

DLG (mm) 0.08 0.13 0.13 0.10


passing for IMRT and RA, respectively. For RA delivery, the
Transmission (%) 0.5 0.4 0.4 0.5
average passing rate was 99.3 % (96.6–100%) for 3%/3mm
DLG, Dosimetric leaf gap; MDACC, MD Anderson cancer center; UPENN, uni-
global normalization analysis, and 93.1% (83.3–98.9%) for
versity of pennsylvania. 2%/2 mm local normalization. For static gantry IMRT deliv-
ery, the average passing rate was 98.0% (91.3–100%) for 3%/
TABLE VI. Results of the IROC independent evaluation delivered point doses 3 mm, and 77.7% (63.9–92.2%) for 2%/2 mm.
(head and neck phantom, RA plan).

Location IROC-H vs Inst. Criteria Acceptable 5. DISCUSSION

Primary PTV sup/ant 0.99 0.93–1.07 Yes We have described the combined experience of two insti-
Primary PTV inf/ant 0.98 0.93–1.07 Yes tutions in commissioning the Halcyon linac. The Halcyon
Primary PTV sup/post 0.98 0.93–1.07 Yes presents some new challenges in commissioning, mostly
Primary PTV inf/post 1.00 0.93–1.07 Yes related to its new geometry, and the absence of a light field
Secondary PTV sup 0.99 0.93–1.07 Yes and a mechanical distance measuring device. However, new
Secondary PTV inf 0.98 0.93–1.07 Yes methods were developed to use imaging to position the QA
equipment. Once developed, these became very

Medical Physics, 46 (10), October 2019


4312 Netherton et al.: Commissioning the Halcyon 4312

straightforward and easy to use. Other than this, we were able MLC motions. Thus, exhaustive end-to-end tests and initial
to use the same equipment to commission the Halcyon as for characterization of a large range of patient-specific quality
other new treatment devices at our institutions. assurance tests may be more reflective of beam model verifi-
Although this report is on the commissioning experience cation. Other work that also details clinically acceptable
of two institutions, each institution initially implemented its agreement of validation measurements with the pre-defined
commissioning process independently. It was only after each TPS is described by Roover et al and employs end-end tests,
institution had completed their work that the results were IMRT/VMAT validation, and external dosimetric audits on
shared and compared. In most cases, both groups used simi- the Halcyon system.19 We did not examine the minimum data
lar methods and had similar experiences. requirements for spot-checking this system, and should be the
The Eclipse system is pre-loaded with a Halcyon beam focus of future work. Since the completion of this work, ongo-
model, which reduces the need to collect extensive beam ing quality assurance has not indicated the need to change any
data. Instead of collecting data to create and test the beam of the pre-determined beam data in the clinical system.
model in the treatment planning system, the user collects the There are some differences between the two pre-clinical
relevant beam data to check what is already in the Eclipse systems and the clinical systems which are now commercially
treatment planning system. Because this was our first experi- available, or will become available in the future. One differ-
ence with this linac, both institutions collected extensive ence is in the design of the MLCs. The pre-clinical systems
beam data—although we expect that this could be signifi- have MLCs which did not have a tongue-and-groove design,
cantly reduced for most institutions. We experienced no but the clinical systems do have such a design – meaning the
major challenges or disagreements between the measured systems that are currently commercially available have a ton-
data and the data provided by the vendor (RBD and Eclipse gue-and-groove design that is similar to those on other linacs
TPS calculated data). A minor difference we noticed, which from the same vendor. Transmission measurements from the
was exceeding MPPG5a recommendations, was that maxi- pre-clinical machine were 0.5% while clinical systems were
mum point differences in the high dose region (d = 10 cm) 0.4% (Table V). While we did not determine whether this dif-
between measured profiles and exported TPS profiles, were ference was statistically significant or due to the absence of
greater than 2% for multiple field sizes (Table IV). We tongue-and-groove from the pre-clinical system, transmission
believe this to be attributed to the use of ion chambers vs measurements on the clinical Halcyon 1.1 system were within
diodes, as well as to the fact that a misalignment at depth by the range specified by Lim et al in their publication describ-
1 mm can result in a large point percent difference when ing the characterization of the Halcyon MLC system.20 Other
close to gradient regions (in this case, 5% per mm). When features of the Halcyon can also be expected to change over
dose regions are analytically defined and plotted, measure- time. One key example is the availability of kilo-voltage
ments can then be analyzed according to the limits defined in imaging. Our test systems, in common with some systems
MPPG5a. In addition, the effect of TPS modeling on the currently available, only offer megavoltage imaging. The Hal-
RBD beam profiles can be visually observed. Penumbra, in cyon is designed, however, to facilitate the addition of kV
the case of the profiles exhibited in Fig. 1, is effectively imaging with the imaging beamline perpendicular to the
reduced by half (8–3 mm) –perhaps attributed to the removal treatment beamline (the same as with most other medical
of volume dependency from the ionization chamber. In terms linacs) – at the time of writing, this has just become commer-
of the difference between detector choice (MDACC vs cially available and has been characterized by Cai et al.21
UPENN) for scanning measurements, using a CC13 on a Kilovoltage imaging may allow a means to facilitate equip-
10 9 10 cm2 field (depth = 10 cm) vs a CC04 yielded a ment setup during commissioning – although we found that
0.2 cm difference in penumbra—small compared to the the MV imaging was sufficient, and may be preferable as it
dimensions of the field. These differences in penumbra were uses the treatment beam line. Results from other commis-
small for large field sizes (>8 9 8 cm2), with the larger sioning tests are provided in the supplemental material (Data
chamber having a larger penumbra due to volume averaging. S1 & Appendix S1).
Likewise, a 0.2 cm difference between MDACC (EDGE Another example of changes that can be expected in future
diode) and UPENN (CC04) penumbra was observed for pro- versions of the Halcyon is the capability to treat multi-isocen-
files at 2 9 2 cm2. This difference in penumbra is significant ter plans. This is needed to allow treatment of larger treat-
due to the fact that it is 10% of the field size. Thus, larger ment volumes. When that capability is available, it will be
chambers would not be appropriate to compare dose profiles necessary to include that testing in the commissioning pro-
calculated in a TPS to those taken with scanner measure- cess, including the end-to-end tests.
ments. For small field sizes, diodes more closely matched rel-
ative TPS beam profiles than did a CC04 chamber.
6. CONCLUSIONS
End-to-end tests were all successful, and the results would
have been acceptable for the systems to be released to the We have reported our experiences and results for commis-
clinic. It is noted that two of the three systems tested in our sioning the Halcyon unit. Although the user needs to get com-
work were pre-clinical systems. It is important to note that fortable with the new geometry of this system, we found that
while MPPG5a recommends many tests related to static, open we could use the same equipment to commission the Halcyon
beam characterization, the Halcyon only treats with dynamic as is used for traditional (C-arm linacs in our clinics). Because

Medical Physics, 46 (10), October 2019


4313 Netherton et al.: Commissioning the Halcyon 4313

penumbra is not defined as is in flattened beams, we recom- 14. Clivio A, Vanetti E, Rose S, et al. Evaluation of the machine perfor-
mend analytically calculating beam regions based on inflec- mance check application for TrueBeam linac. Radiat Oncol.
2015;10:97.
tion points. Good agreement was found between our 15. Li Y, Netherton T, Nitsch PL, et al. Independent validation of machine
respective experimental data and the treatment planning sys- performance check for the Halcyon and TrueBeam linacs for daily qual-
tem calculations for the correct use of detector. Thus, we have ity assurance. J Appl Clin Med Phys. 2018;19:375–382.
16. Low DA, Li Z, Drzymala RE. Minimization of target positioning error in
shown that the vendor pre-loaded data agree well with the
accelerator-based radiosurgery. Med Phys. 1995;22:443–448.
independent measured ones during the commission process. 17. Li Y, Netherton T, Nitsch PL, et al. Normal tissue doses from MV
This verifies that a data validation instead of a full-data com- image-guided radiation therapy (IGRT) using orthogonal MV and MV-
missioning process may be a more efficient approach. This CBCT. J Appl Clin Med Phys. 2018;19:52–57.
18. Mihailidis D, Ling CC, Brady L, et al.Evaluation of MV imaging dose
work should be useful for new users of the Halcyon device. for the first clinical Halcyon system. ESTRO; 2017; Vienna, Austria. p.
PO-0992.
19. De Roover R, et al. Validation and IMRT/VMAT delivery quality of a
CONFLICTS OF INTEREST preconfigured fast-rotating O-ring linac system. Med Phys.
2019;46:328–339.
This work was, in part, funded by Varian Medical Sys- 20. Lim TY, Dragojevic I, Hoffman D, Flores-Martinez E, Kim GY. Charac-
tems. terization of the Halcyon TM multileaf collimator system. J Appl Clin
Med Phys. 2019;20:106–114.
21. Cai B, et al. Characterization of a prototype rapid kilovoltage x-ray
a)
Author to whom correspondence should be addressed. Electronic mail: image guidance system designed for a ring shape radiation therapy unit.
Mihailidis@pennmedicne.upenn.edu. Med. Phys. 2019;46:1355–1370.

REFERENCES SUPPORTING INFORMATION

1. Thompson S, Constantin M, Kokkonen P. Eclipse Photon Beam Models Additional supporting information may be found online in
for Halcyon. Varian Medical Systems: Technical Perspectives; 2017. the Supporting Information section at the end of the article.
2. McEwen M, DeWerd L, Ibbott G, et al. Addendum to the AAPM's TG-
51 protocol for clinical reference dosimetry of high-energy photon Fig. A1. Las Vegas Phantom images.
beams. Med Phys. 2014;41:041501. Table A1. TG51 measurement for Halcyon.
3. Klein EE, Hanley J, Bayouth J, et al. Task group 142 report: quality
Table A2. Radiation field size verification for different MLC
assurance of medical acceleratorsa). Med Phys. 2009;36:4197–4212.
4. Almond PR, Biggs PJ, Coursey BM, et al. AAPM's TG-51 protocol for settings.
clinical reference dosimetry of high-energy photon and electron beams. Table A3. Isocenter Verification with Winston-Lutz (radius,
Med Phys. 1999;26:1847–1870. mm).
5. Smith K, Balter P, Duhon J, et al. Medical physics practice guideline
8.a.: linear accelerator performance tests. J Appl Clin Med Phys.
Table A4. Normalized dose measurements for varying gantry
2017;18:23–39. speed, dose rate and MLC speed.
6. Smilowitz JB, Das IJ, Feygelman V, et al. Medical physics practice Table A5. Distance from MV imaging to radiation isocenter
guideline 5.a.: commissioning and QA of treatment planning dose calcu- as calculated using the Logos XRV-124 system (Average of
lations - megavoltage photon and electron beams. J Appl Clin Med Phys.
2015;16:14–34.
four trials shown).
7. Fontenot JD, Alkhatib H, Garrett JA, et al. Medical physics practice Table A6. Basic MV imager performance measurements
guideline 2.a: commissioning and quality assurance of X-ray-based (MCACC and UPENN).
image-guided radiotherapy systems. J Appl Clin Med Phys. Table A7. Contrast and spatial resolution measurements (us-
2014;15:4528.
8. Ezzell GA, Burmeister JW, Dogan N, et al. IMRT commissioning: multi- ing CATPHAN).
ple institution planning and dosimetry comparisons, a report from Table A8. MVCBCT geometric accuracy measurements
AAPM task group 119. Med Phys. 2009;36:5359–5373. (CATPHAN).
9. AGENCY IAE. Commissioning and quality assurance of computerized Table A9. CT number accuracy measurements using the
planning systems for radiation treatment of cancer. Vienna: International
Atomic Energy Agency; 2004. CATPHAN (MD Anderson) and QUART phantom
10. Das IJ, Cheng CW, Watts RJ, et al. Accelerator beam data commission- (UPENN).
ing equipment and procedures: report of the TG-106 of the therapy phy- Table A10. CT image uniformity measurements using the
sics committee of the AAPM. Med Phys. 2008;35:4186–4215.
CATPHAN (MD Anderson) and QUART phantom
11. Bissonnette JP, Balter PA, Dong L, et al. Quality assurance for image-
guided radiation therapy utilizing CT-based technologies: a report of the (UPENN).
AAPM TG-179. Med Phys. 2012;39:1946–1963. Data S1. Data supplement for PDD, beam profile, output fac-
12. Chang Z, Wu Q, Adamson J, et al. Commissioning and dosimetric char- tors, and surface dose (VARIAN, UPENN, MDACC).
acteristics of TrueBeam system: composite data of three TrueBeam
machines. Med Phys. 2012;39:6981–7018.
Appendix S1. Supplemental text describing shielding
13. Glide-Hurst C, Bellon M, Foster R, et al. Commissioning of the Varian requirements, output calibration, mechanical tests, radiation
TrueBeam linear accelerator: a multi-institutional study. Med Phys. tests, MLC and gantry tests, and imaging system tests
2013;40:031719. (UPENN, MDACC).

Medical Physics, 46 (10), October 2019



 

1  APPENDIX

2  A. Shielding requirements.

3  The Halcyon linac has a beam-stop, which should allow for flexibility of installation in an

4  existing vault. Radiation surveys and other measurements should be the same as with other

5  linacs, following applicable State, Federal, and other regulations. A radiation survey was

6  conducted to validate the shielding adequacy. With a lateral beam at maximum field size

7  irradiating a phantom, a maximum exposure rate of ~9 R/hr was measured at 2 m from isocenter

8  in both lateral and longitudinal directions (for a 28x28 cm2 field on a phantom, with the gantry

9  rotated to the contralateral side of the measurement point); the lateral position being downstream

10  of the beam direction. A maximum leakage (all MLC closed) exposure rate of ~2 R/hr was

11  measured at 2 m from the isocenter directly behind the bore and laterally to the gantry with the

12  source proximal to the point of measurement.

13   
14  B. Output Calibration

15  The Halcyon system allows MU calibration to be performed in three alternative geometries. At

16  our institutions we followed the TG51 protocol1 (including the addendum for FFF beams2) to

17  calibrate the Halcyon to 1.00cGy/MU to water, measured in a water phantom, at the depth of

18  maximum dose, with SSD = 100 cm. We used the IBA Blue2 Phantom and Sun Nuclear 1D water

19  phantoms for this experiment. First, the water tank was aligned to the external setup lasers on the

20  treatment table and a water-proof farmer chamber (Exradin A12 / PTW TN30013) was inserted

21  for the output measurements. The water tank was aligned and SSD set as described above. Percent

22  depth dose measurements were performed to determine the beam quality correction factor kQ with

23  and without the lead foil as described.2,1



 

24  For UPENN Halcyon 1.0, the measurements (Table A1) showed that PDD10=63% and

25  63.9% without and with the lead foil, respectively. Similarly, MDACC Halcyon 1.0 corresponding

26  values differed less than a percent for PDD(10) with and without lead. For Halcyon 1.0, equation

27  (1) and Table I in Ref. 2 were used to calculate kQ for the A12 and TN30013 chambers without

28  lead (0.9974 and 0.9958, respectively) and with lead (0.9962 and 0.9951). This small difference

29  in KQ (with and without lead) is also reported by Lloyd et al3. The correction for change in radial

30  beam profile over the chamber volume (Prp) was 0.25%, calculated by averaging the profile scan

31  results over the chamber length. The output calibration was measured at 10 cm depth and corrected

32  to dmax using the clinical PDD10 for a 10x10 cm2 reference field at 100 SSD, as per TG-51. An

33  adjustment of about 1% was made relative to the output established at installation. For safety

34  considerations, the Halcyon is restricted to a limit of 5% adjustment relative to the factory setting.

35  Varian service must be contacted in that limit needs to be exceeded.

36  C. Mechanical tests

37  Coincidence of radiation isocenter during collimator or gantry rotation

38  Evaluation of the radiation isocenter was performed using the “star-shot” method with EBT3 films.

39  For the gantry rotation star shot, the EBT3 film was positioned vertically in the plane of the

40  isocenter. For the collimation star shot the film was horizontal at source-to-film distance 100cm.

41  An in-house MLC QA software was used for analysis. The tolerance was: ±2 mm (non-

42  SRS/SBRT).4 For collimator rotation, the radiation isocenter for collimator rotation was found

43  within 0.3 mm diameter circle (both institutions). For gantry rotation, the radiation isocenter for

44  gantry rotations was found within 0.8 mm diameter circle (MDACC) and 0.4 mm diameter circle

45  (UPENN).

 

46  MPC performs isocenter size checks and combines collimator rotations and gantry

47  rotations. We have assessed the ability of the MPC to correctly identify intentionally inserted errors

48  for the Halcyon unit – this data is reported elsewhere5. The average isocenter size (radius, the size

49  of treatment isocenter is defined as the maximum distance of the beam’s central axis from the

50  idealized isocenter) of 30 measurements is 0.79 mm which is twice that found with film. A possible

51  explanation for this disagreement could be due to the sparsity of gantry angles chosen with film.

52  Though the angles chosen were in 30o increments, discrete angles with the largest

53  deviations from isocenter could be missed (see maximum deviations depicted in Section describing

54  Isocenter Verification using WL test). In addition, star shots do not fully capture gantry wobble in

55  the in/out direction. It is suggested that sparse, film-based star shots alone should not be used for

56  this reason. A WL-type test is recommended as described below in the “Isocenter Verification”

57  section.

58 

59  Table travel

60  The measurement of the physical treatment table travel was performed by using a ruler versus

61  the machine lasers of 10.0 cm movements in each direction. The treatment table was moved by

62  digital setting in the system. The actual travel distances were measured using the ruler. The

63  results for the difference of the distance measured by using ruler and digital setting for each

64  movement revealed agreement better than 1 mm for all travel axes.

65 

66  Digital distance indicators.



 

67  Known and measured distances (using digital tools at the linac console) were compared based on

68  images of a polystyrene board with graticules to assess scaling. The results showed that the

69  differences between measured and expected points 10 cm apart are less than 0.5 mm.

70 

71  D. Radiation tests

72  Radiation field size verification

73  The graticule board was placed perpendicular to the beam central axis at SSD=100 cm, collimator

74  at 0o. Metal BBs were placed on the graticules at known positions. Images of the board with BBs

75  for different field size settings were acquired. The distance from field edge to the center of the BB

76  on images was measured for symmetric fields: 1010 cm2, 2020 cm2 and 2626 cm2.

77  For asymmetric fields, MLCs were digitally set for different field size openings, and

78  images were acquired. The distance from the center crosshair to the field edges was measured

79  manually using digital measurement tools in the service mode. The difference between digital

80  settings and the measured distances were also calculated.

81  We found the agreements were within 1.3 mm for all fields, as seen in Table A2. Small

82  uncertainty in the values on the order of 0.5 mm is expected since all BB’s were hand-placed.

83  Using a graticule board with machined high-contrast inserts would be desirable for more accurate

84  measurements.  

85  Isocenter Verification

86  The WL pointer phantom was aligned to the isocenter. Then, 36 images at different gantry angles

87  (from 0o to 350o, 10o increment) and collimator at 0o were acquired using MV imaging system.

88  The Sun Nuclear Machine software was used to analyze the images. The method used to compute

89  3D target position error is based on the study by Low, et al.6. Y-axis is the gantry rotation axis, Z-

 

90  axis coincides with the central axis of the radiation beam, pointing into the collimator, and X is

91  assigned to form a right-hand coordinate system. The origin is defined as the center of each

92  imaging field. The horizontal, vertical and total 2D offsets versus gantry angle were determined.

93  Although the WL test result were within the 2 mm tolerance, agreement is better at the cardinal

94  angles than at intermediate angles, especially for X-axis. In general, a sparse, film-based star-shot

95  test alone is not recommended, but if further verification of the WL test are desired, performing a

96  star-shot at the angles with maximum offset is suggested. Isocenter verification the results are

97  given in Table A3.

98 

99  E. MLC and Gantry tests

100  Picket fence test

101  A picket fence pattern was delivered to the EPID using dynamic QA fields to assess MLC

102  alignment and picket position accuracy. The picket fence pattern images were analyzed using in-

103  house software to evaluate the picket spacing and slope relative to the imager coordinates

104  (MDACC). Both proximal and distal banks of leaves were used for “2-bank” tests, while only the

105  distal bank of leaves were used for the “lower-bank” tests. Static tests were performed at the

106  cardinal gantry angles, and a dynamic test was performed under RA delivery. The static MLC

107  picket fence testing results for different gantry angles revealed a maximum picket spacing

108  difference of 0.7 mm, and a maximum slope of 0.17 degrees. The RA MLC picket fence test

109  resulted in a maximum picket spacing difference of 0.4 mm, and a maximum slope of 0.21 degrees.

110 

111  MLC speed, gantry speed and dose rate testing



 

112  Patterns of equivalent dose delivered under varying levels of dose rate, gantry speed, and MLC

113  speed were captured by the portal imager, and normalized to an open-field image.7 Three test files

114  were provided by Varian for delivery of dose rate levels from 128-800 MU/min, gantry speeds

115  from 3.7-23 deg/s, and MLC speeds from 1-5 cm/s respectively. The images were analyzed to

116  verify that the normalized dose in each delivered pattern was within the 2% specification. The

117  acquired EPID images were analyzed to determine the normalized dose for each delivered pattern,

118  and the results are presented below for dose rate, gantry speed, and MLC speed (Table A4). The

119  tolerances for these tests are ±2%.

120   

121   

122  F. Imaging system tests

123  Coincidence of radiation isocenter and MV imager isocenter

124  Coincidence of the radiation and imaging isocenters are checked as part of the MPC process.

125  We independently evaluated this using the Logos XRV-124 device (Logos Systems

126  International, Scotts Valley CA). The XRV-124 is a conical scintillator that converts an incident

127  beam into two visible spots in the beam direction that can be used to calculate the location of the

128  beam vector in the device’s native coordinate system. A CT-based treatment plan was created in

129  the Eclipse treatment planning system that consisted of eight 2x2 cm square fields at equally-

130  spaced gantry angles over the full 360° gantry rotation. The clinical IGRT process was used to

131  align the device to the MV imaging isocenter prior to delivery of the test treatment plan. The

132  isocenter location was calculated in the XRV software by finding the average location of the

133  mid-points of the shortest line segment between every set of 2 beams in the test treatment plan

134  described above. Radiation isocenter size and distance to the MV imaging isocenter were also

135  calculated. Four trials of IGRT alignment were performed, with the average of calculated values

 

136  reported. The isocenter size for all four trials with the XRV-124 system was found to be 0.7 mm.

137  The average distance of the isocenter location to the MV imaging isocenter was 0.65 mm. The

138  principal components of the distance between MV and radiation isocenters is shown in Table A5.

139   

140  MV imager performance

141  The Las Vegas (Portal Vision) phantom was imaged using both high-quality and low-dose

142  techniques to give QA baselines (suggested by MPPG2a8). MV images taken with Halcyon 1.0 are

143  shown in Figure A1, and all visible holes were recorded in Table A6 (Varian’s specification: Big

144  right dot in 5th row is visible). Our results showed that image quality meet Varian’s specification

145  at both institutions.

146 

147  MVCBCT performance

148  Image quality tests for the MVCBCT mode were performed with the Catphan phantom, and with

149  the QUART DVT_VN phantom that ships with the clinical system. Scans were performed with

150  a 20 cm field length for the high-quality and low-dose techniques. These tests included low- and

151  high-contrast resolution, geometric accuracy, CT number accuracy, and CT uniformity.

152  Results of the image quality tests from MDACC are shown in Table A7, Table A8, Table A9,

153  and Table A10. The results from UPENN results were similar.

154 

155  REFERENCES 

156  1. Almond PR, Biggs PJ, Coursey BM, et al. AAPM's TG-51 protocol for clinical reference dosimetry
157  of high-energy photon and electron beams. Med Phys 1999;26:1847-70.
158  2. McEwen M, DeWerd L, Ibbott G, et al. Addendum to the AAPM's TG-51 protocol for clinical
159  reference dosimetry of high-energy photon beams. Med Phys 2014;41:041501.
160  3. Lloyd, S. A. M. et al. TG-51 reference dosimetry for the HalcyonTM: A clinical experience. J. Appl.
161  Clin. Med. Phys. 19, 98–102 (2018).

 

162  4. Klein EE, Hanley J, Bayouth J, et al. Task Group 142 report: Quality assurance of medical
163  acceleratorsa). Medical Physics 2009;36:4197-212.
164  5. Li Y, Netherton T, Nitsch PL, et al. Independent Validation of Machine Performance Check for the
165  Halcyon and TrueBeam Linacs for Daily Quality Assurance. J Appl Clin Med Phys 2018;19:375-382.
166  6. Low DA, Li Z, Drzymala RE. Minimization of target positioning error in accelerator-based
167  radiosurgery. Med Phys 1995;22:443-8.
168  7. Ling CC, Zhang P, Archambault Y, Bocanek J, Tang G, Losasso T. Commissioning and quality
169  assurance of RapidArc radiotherapy delivery system. Int J Radiat Oncol Biol Phys 2008;72:575-81.
170  8. Fontenot JD, Alkhatib H, Garrett JA, et al. AAPM Medical Physics Practice Guideline 2.a:
171  Commissioning and quality assurance of X-ray-based image-guided radiotherapy systems. J Appl
172  Clin Med Phys 2014;15:4528.
173 
174 

175 

176 

177 

178 

View publication stats

You might also like