Professional Documents
Culture Documents
Gizeaddis L.
Jimma University
Jimma Institute of Technology
Biomedical Engineering Department
Introduction
In conventional radiography:
In medical tomography:
For objects that can be identified in both images, the two films
provide location information.
2D (parallel slices)
1D set of
projections
of that slice
Computer system
Data acquisition unit
Electronics
of the
Computer
Detectors system
Operators
Image
workplace
evaluation
X-ray tube
Image
processor
generator
Moveable
patient bed
CT- Scanner Generations
Pencil Beam: inefficient use of the x-ray source , excellent scatter rejection
Fan Beam: linear detectors, scattered radiation accounts around 5%
Open Beam: used in projection radiography, highest detection of scatter
CT- Scanner Generations
Ring artifacts
resulting from
defective detector
element in 3rd
generation single-
slice scanner.
CT- Scanner Generations
The equation is true only if the gain terms cancel each other out,
If there is electronic drift in one or both of the detectors, the gain
changes between detectors.
3rd vs. 4th generation
5th generation: stationary/stationary
I t I 0 e t
ln(I 0 / I t ) t
Reconstruction Algorithms
Simple backprojection
Filtered backprojection
Image Reconstruction: ART
Image Reconstruction: ART
Image Reconstruction: Simple backprojection
Starts with an empty image matrix, and the value from each ray
in all views is added to each pixel in a line through the image
corresponding to the rays path
y y
object x x
X-ray
tube
Data Acquisition field Reconstruction field
Blur
Simple Backprojection
Backprojection: linear
Backprojection: two projections linear
Backprojection: multiple projections
Filtered backprojection
Projection Data, P
x y
Filtered Backprojection
Convolution kernel
Filter
Filtered Backprojection
Unfiltered Backprojection
Filtered Backprojection
Projection and Radon transform
Where p(r) is the projection of the function (x,y) along the angle
Projection and Radon transform
Thus, CT values are relatively stable for a single organs and are to a
large extent independent of the X-ray tube spectrum.
Hounsfield Scale
Hounsfield Scale
A narrow window
produces a very
high contrast
image,
4 3
STOP
Table & Patient
Indexed To The
Next Scanning Position Tube & Detectors
Slow Down & Stop
Section thickness Vs index
Contiguous spiral
Pitch = 1, that is 10mm/10mm
Extended spiral
Pitch = 2, that is 20mm/10mm
Overlapping spiral
, that is 5mm/10mm
Pitch
SMALL PITCH
LARGE PITCH
Pitch
examination
Beam collimation
As a spectrum of x-rays
passes layers of tissue, the
lower-energy photons in the
x-ray spectrum are
attenuated to a greater
degree than the higher-
energy components of the
spectrum.