You are on page 1of 2

Excellent performance, high reliability, and ease of use BC-700 & BC-720

Auto Hematology Analyzer with ESR

BC-700 & BC-720


Key Specifications Physical Specifications Auto Hematology Analyzer with ESR

Above and Beyond


Principles Dimensions
WBC (IMG/Neu/Mon/Lym/Eos/Bas), NRBC/RET*,PLT-H/PLT-O*/IPF: 500D x 325W x 450H mm
labXpert Applicable to different sample types
SF Cube ^ Cell Analysis Technology
Coming as standard configuration Peripheral blood/Capillary blood
^S: Scatter; F: Fluorescence; Cube: 3D analysis Weight
Same software as BC-6 series Pre-diluted blood/Body fluid
≤35Kg

RBC, PLT Voltage


RFID Floating screen Focusing Flow-DC Impedance Method 100V-240V~ (±10%)
Encryption key management Switching between different analysis
modes with a single touch Frequency
HGB
Colorimetric method 50Hz/60Hz (±1Hz)

2 ESR Power input


1
Photometric method 300VA

Number of measuring parameters (whole blood): 109 External output


Number of reportable parameters: 41 LAN x 1 , USB x 4 (Specifications: DC 5V; 500mA;
3 4
WBC Bas# Bas% Neu# Neu% Eos# Eos% Lym# Lym% Mon# USB 2.0 x 3; USB 3.0 x 1)
Mon% IMG# IMG% RET%* RET#* RHE* IRF* LFR* MFR* HFR*
RBC HGB MCV MCH MCHC RDW-CV RDW-SD HCT NRBC# Normal Operating Environment
NRBC% PLT PLT-I PLT-H PLT-O* MPV PDW PCT P-LCR P-LCC
IPF ESR Ambient temperature:
Number of research parameters: 68* 10℃ ~ 35℃
An all-in-one solution that goes above
Number of measuring parameters (body fluid): 18 Relative humidity:
and beyond your expectations Number of reportable parameters: 7 30% ~ 85%
WBC-BF TC-BF# MN# MN% PMN# PMN% RBC-BF DESIGN
AWARD
Number of research parameters: 11 Atmospheric pressure: 2021
70.0kPa ~ 106.0kPa^
Sample volume ^Note : Required altitude for normal operation:
Superior functionality CD (whole blood): 23ul -400m ~ +3000m
Excellent performance
CD+ESR (whole blood): 160ul
Integrated CBC + ESR detection
Reliable results for Predilute: 20ul
abnormal and interfered samples
provides a comprehensive
tool for inflammation detection
Performance
Optical PLT-H ensures reliable Data storage capacity
PLT results even with interference Parameter Linearity Range Precision Carryover
Up to 150,000 results including numeric and graphical WBC 0-500×109 /L ≤2.5% (≥4.51×109 /L) ≤ 1.0%
information *
RBC 0-8.60×1012/L ≤1.5% (≥3.5×1012/L) ≤ 1.0%
HGB 0-260g/L ≤1.0% (110-180g/L) ≤ 1.0%
Throughput
HCT 0-75% ≤1.5% (30%-50%) ≤ 1.0%
CD 80t/h CDR 45t/h CD+ESR 40t/h
≤ 1.5(SD) (≤20×109/L)^
PLT* 0-5000×109 /L ≤ 1.0%
Analysis Mode ≤ 2.5% (≥100×109/L)^

Sample Type Analysis Mode ≤15% (RBC ≥ 3.00×1012/L


RET* 0-0.8×1012/L ≤ 1.0%
RET%: 1.00% ~ 4.00%)
Integrated CBC + ESR analyzer CBC, CBC + DIFF, CBC + DIFF + RET*, CD + ESR,
SF Cube 3D fluorescent platform Optical PLT-H in every CD test Whole blood ESR ≤1.8(SD)(0~20mm/h) ≤ 1.0%
IMG/NRBC reportable parameters
CDR + ESR*, CD/WBC-3X, CDR/PLT-5X*,
Integrated RET* & body fluid analyzer
Accurate measurement of low WBC/PLT count and other modes
^ Note: Applicable only to CDR/PLT-O 5x and CR/PLT-O 5x models
Anti-aging capability:
24h (room temperature) Predilute CBC, CBC + DIFF, CDR*, and other modes
48h (refrigerated transportation) Items marked with an asterisk (*) apply only to BC-720
CBC + DIFF

Interactive excellence
www.mindray.com
Refined details for the optimal iHelp videos
user experience Floating screen P/N:ENG-BC-700 & BC-720 ESR-210285X6P-20220115
Ease of maintenance ©2022 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved.
BC-700 & BC-720 Auto Hematology Analyzer with ESR BC-700 & BC-720 Auto Hematology Analyzer with ESR
Above your expectations Beyond your expectations
SF Cube fluorescent technology allows reliable counting and differentiation of Limitations of traditional PLT counting
abnormal samples
In the traditional impedance method, PLTs are subject to interferences that may lead to falsely high or falsely low results
(as shown in the figure). Once an error report is generated, it will directly affect the judgment and decision-making of
clinicians. The results reported at the clinical decision level are related to patient safety. Therefore, accurate PLT results are
critical in clinical practice.
Interference of RBC fragments
Fragment
PLT
DIFF channel RET channel
More refined and reliable cell Fragment area

Falsely high
differentiation Normal

MON RET PLT


PLT Result
3D fluorescent analysis technology allows
HFC
IMG
RBC Impedance method: 89
LYM 0 10 20 30 fL Reference value: 26
reliable differentiation of immature and other NEU

EOS
abnormal cells, such as immature
granulocytes (IMGs), reticulocytes (RETs*), and GHOST BASO PLT-O IPF
Interference of large PLTs

0 10 20 30 fL
immature platelet fraction (IPF). PLT
Large PLTs area
Falsely low

Large PLTs PLT Result


Impedance method: 66
0 10 20 30 fL Reference value: 98

More reliable measurements for


low-value samples
WBC6-Part Differentiation

Optical PLT-H in every CD test


PLT-O Measurement

Conventional Conventional
The BC-700 & BC-720 3D fluorescence analysis
platform is designed with multiple counting
In order to solve the above problem, we have developed a brand new parameter PLT-H. CD + ESR in one test provide reliable
It combines small PLTs from the conventional impedance method and large PLTs from the optical method. The solution
WBC-3x and PLT-O 5x analysis modes to help
ensure higher reliability for low-value WBC
can resist the interferences in conventional PLT detection without requiring extra reagents. ESR results with greater ease
and PLT samples. In addition, the PLT Schematic diagram of PLT-H
The BC-700 series integrates an automatic ESR module in a hematology analyzer. It can also generate both CBC & ESR
de-aggregation function can reduce the
results in one test within 1.5 min. In addition, it saves the costs that would otherwise be incurred for the purchase,
cumbersome review work. Normal sample Interfered sample maintenance, consumables, and storage space of a separate ESR analyzer. Compared with the traditional Westergren
WBC 3x PLT-O 5x method, this method performs better in quality traceability, repeatability, speed, safety, and level of automation.

Impedance channel
Small PLTs

Small PLTs

Volume
PLT-H Accurate Cost-effective Automatic
More comprehensive alarm messages Small PLTs are not interfered
for abnormalities Small PLTs (< 10 fL) are free from interferences in impedance channel • Great correlation with the Westergren • The integrated instrument is capable • Report CBC + ESR results together
method of both CBC and ESR detection; within 1.5 min;
PLT-H • Same QC and calibrator as in the • Takes up the space of only one analyzer. • The measurement results are
The analyzer provides a detailed list of over 40 BC-6000 series protected against the influence of
prompt messages, including WBC message, RBC Scattergram of DIFF Channel Scattergram of DIFF Ghost • Combined examination helps to avoid subjective factors;
message, and PLT message. This allows laboratory Volume
the interferences of dehydration, • Automation can reduce the biosafety
FS
Large PLTs Large PLTs polycythemia vera and anemia on hazards that may otherwise be
technicians to intuitively and quickly identify Optical PLT-H in every CD test ESR results; introduced by a manual method.
DIFF channel

abnormal samples and proceed further with the Large PLTs


Volume

samples in a timely manner. This in turn helps to Fragments


Large PLTs are
from lysed
avoid missed diagnosis of blood disease and false Ghost
RBCs +
Small PLTs
not interfered in
DIFF channel
reports. RETs

Magnified view of ghost

You might also like