You are on page 1of 28

T a p e r e d H i p S y s t e m

T a p e r e d H i p S y s t e m
Robert B. Bourne, M.D.,
F.R.C.S.(C)
London, Ontario, Canada

Professor Ernesto DeSantis


Rome, Italy

Wayne M. Goldstein, M.D.


Chicago, Illinois

Gianni L. Maistrelli, M.D.,


F.R.C.S.(C)
SURGICAL TECHNIQUE Toronto, Ontario, Canada
COMPLETED IN CONJUNCTION WITH
John W. McCutchen, M.D.
Orlando, Florida

Cecil H. Rorabeck, M.D.


London, Ontario, Canada

James P. Waddell, M.D.


Toronto, Ontario, Canada

Nota Bene: The technique descrip-

tion herein is made available to the

healthcare professional to illustrate

the authors’ suggested treatment for the

uncomplicated procedure. In the final

analysis, the preferred treatment is that

which addresses the needs of the patient.


INTRODUCTION
The Synergy Tapered Hip System capitalizes on the All Synergy femoral stems are made from high-
excellent clinical results of proximal to distal strength, forged titanium and feature a 3° proximal to
tapered stem designs. The Synergy system features a distal taper. Parallel and longitudinally arranged flutes
variety of stem designs that provide different meth- on the stems provide good axial and rotational stability.
ods of stem fixation and that also address different
patient demand types. All of the stems in the The 12/14 taper of the Synergy stems accepts 22, 26,
Synergy system are implanted with one simple set of 28, and 32 mm femoral heads in a variety of neck
surgical instruments. lengths. The circulotrapezoidal neck of the Synergy
stems is smaller in the anterior/posterior direction
This surgical technique focuses only on the cementless than most competitive stems and is polished. The
stems of the Synergy Tapered Hip System. smaller A /P neck dimension allows for improved
prosthesis range of motion. If impingement of the
Synergy femoral stems are available in two femoral stem and acetabular component occurs, the
circumferential surface finishes; porous-coated, polished neck will reduce the amount of debris that is
and HA-coated. The porous-coated design pro- generated. The distal bullet tip of the Synergy stems
motes bone ingrowth and three-dimensional has also been polished in order to minimize distal
interlocking between the implant and the bone. endosteal irritation and reduce the risk of thigh pain.
The HA-coated stem design features a high shear
strength, hydroxylapatite coating that helps pro- The Synergy instrumentation is packaged in two
mote secure fixation. Both the porous-coated and sterilization trays. The instruments are arranged in the
the HA-coated stems are available in standard and sterilization trays in the same order that they are
high offset versions. required in surgery. This makes for a simple, straight-
forward surgical technique that is highly reproducible.

4
Standard Stem

High Offset Stem

5
STEM SPECIFICATIONS

SPECIFICATIONS NECK HEIGHT MM


Conical When Femoral Head Component Selected Is:
Neck Cross Stem A-P* M-L* Cat. No. Size –3 +0 +4 +8 +12 +16
Size Angle Section Length Width Width 8** 25 27 29 32 35 37
8** 131° 8 mm 120 mm 12 mm 27 mm
91-1390 9 26 28 30 33 35 38
9 131° 9 mm 135 mm 13 mm 28 mm
91-1390 10 26 28 31 33 36 39
10 131° 10 mm 140 mm 14 mm 29 mm
11 131° 11 mm 145 mm 15 mm 30 mm 91-1458 11 27 29 32 34 37 39
12 131° 12 mm 150 mm 16 mm 31 mm 91-1390 12 28 30 32 35 37 40
13 131° 13 mm 155 mm 17 mm 32 mm 91-1457 13 28 30 33 35 38 41
14 131° 14 mm 160 mm 18 mm 33 mm 91-1390 14 29 31 34 36 39 41
15 131° 15 mm 165 mm 19 mm 34 mm 91-0314 15 30 32 33 37 39 42
16 131° 16 mm 170 mm 21 mm 35 mm 91-1390 16 30 32 35 37 40 43
17 131° 17 mm 175 mm 22 mm 36 mm 17 31 33 35 38 41 43
18 131° 18 mm 180 mm 23 mm 38 mm 91-1390 18 32 33 36 39 41 44
19** 131° 20 mm 185 mm 24 mm 39 mm 19** 32 34 37 39 42 44
20** 131° 21 mm 190 mm 25 mm 40 mm 20** 33 35 37 40 43 45

NECK OFFSET MM NECK LENGTH MM


When Femoral Head Component Selected Is: When Femoral Head Component Selected Is:
Cat. No. Size –3 +0 +4 +8 +12 +16 Cat. No. Size –3 +0 +4 +8 +12 +16
8** 31 34 37 40 43 46 8** 26 29 33 37 41 45
91-1390 9 32 34 37 40 43 46 91-1390 9 27 30 34 38 42 46
9 HO 38 40 43 46 49 52 9 HO 31 34 38 42 46 50
91-1390 10 33 35 38 41 44 47 91-1390 10 28 30 34 38 42 46
10 HO 39 41 44 47 50 53 10 HO 32 35 39 43 47 51
91-1458 11 34 36 39 42 45 48 91-1458 11 28 31 35 39 43 47
11 HO 40 42 45 48 51 54 11 HO 33 36 40 44 48 52
91-1390 12 34 37 40 43 46 49 91-1390 12 29 32 36 40 44 48
12 HO 40 43 46 49 52 55 12 HO 33 36 40 44 48 52
91-1457 13 35 37 40 43 46 49 91-1457 13 30 32 36 40 44 48
13 HO 41 43 46 49 52 55 13 HO 34 37 41 45 49 53
91-1390 14 36 38 41 44 47 50 91-1390 14 30 33 37 41 45 49
14 HO 44 46 49 52 55 58 14 HO 36 39 43 47 51 55
91-0314 15 37 39 42 45 48 51 91-0314 15 31 34 38 42 46 50
15 HO 45 47 50 53 56 59 15 HO 37 40 44 48 52 56
91-1390 16 37 40 43 46 49 52 91-1390 16 31 34 38 42 46 50
16 HO 45 48 51 54 57 60 16 HO 37 40 44 48 52 56
17 38 40 43 46 49 52 17 32 35 39 43 47 51
17 HO 46 48 51 54 57 60 17 HO 38 41 45 49 53 57
91-1390 18 39 41 44 47 50 53 91-1390 18 32 35 39 43 47 51
18 HO 47 49 52 55 58 61 18 HO 39 42 46 50 54 58
19** 40 42 45 48 51 54 19** 33 36 40 44 48 52
20** 40 43 46 49 52 55 20** 34 37 41 45 49 53

For use with Smith & Nephew 12/14 taper femoral heads only.

**This measurement is for the HA stems. The porous-coated stems have 0.5 mm additional thickness.
**These stem sizes are only available in the porous coated version and by special request.

6 ™
STEM SPECIFICATIONS

0
10
Neck Length
20 Standard Offset
High Offset
30

–3 and +16 femoral heads


available in 28 mm and
32 mm only.
*Denotes skirted heads.

NOTE: For illustration purposes


only. Surgical templates are avail-
able by contacting your Smith
& Nephew Representative or
Customer Service.

NOT ACTUAL SIZE


7
INDICATIONS AND PREOPERATIVE PLANNING

INDICATIONS
Both cemented and cementless femoral components
have demonstrated long-term durability in the
treatment of severe arthritic conditions of the hip.
An algorithm has been developed that helps define Figure 1. Postoperative
which implant matches each patient’s needs. The lateral radiograph of a
algorithm takes into consideration patient age, bone porous-coated Synergy
stem showing three-
stock, disease type, and implant cost. The algorithm point fixation.
is shown below in Table 1.
Table 1. Algorithm for Implant Selection.
Cementless Cemented
Age Under 75 Over 75
Medullary Canal Funnel Shaped Cylindrical
Disease Osteoarthritis Inflammatory it contacts the medial cortex of bone on both femurs.
Cost Type High-Demand Low-Demand If the patient’s legs are of equal length, the line that
has been drawn will contact both femurs at the same
PREOPERATIVE PLANNING level. If the patient’s legs are of unequal length, the
The goal of preoperative planning is to determine lines will contact the femurs at different levels along
the correct stem size, level of the femoral neck cut, the femur. Select a reference point along the femur,
and proper head and stem offset combination. such as the bottom of the lesser trochanter. The dis-
tance between the line that has been drawn and the
Preoperative templating requires at least an antero-
reference point on both femurs is measured. The
posterior radiograph of the pelvis and a lateral
radiograph of the affected hip. If the opposite hip difference in these measurements indicates the
is unaffected by disease, it can often provide accu- patient’s leg length discrepancy. This technique
rate sizing information for the femoral stem. The is shown in Figure 2.
Synergy stems gain immediate, rigid fixation
through three-point contact with the femur. This
is best appreciated by viewing a lateral postoperative
radiograph as shown in Figure 1.
As can be seen in Figure 1, the stem has direct con-
tact with hard cortical bone at three points: proxi-
mally at the posterior aspect of the femur, anteriorly
in the midsection of the stem, and posteriorly above
the polished distal tip of the stem.
To determine if a patient has a leg length discrepancy,
the anteroposterior radiograph should be used. Draw
a line tangential to both of the ischia or both of the Figure 2. Anteroposterior radiograph demonstrating leg
obturator foramens. This line should extend out until length inequality.


8
INDICATIONS AND PREOPERATIVE PLANNING

Note: Using this method of templating for leg length


discrepancy assumes the patient has a normal, sym-
metrical pelvis and has neutral limb positioning.
Intraoperatively, leg length restoration can be veri- SYNERGY™ STANDARD AND HIGH OFF
SET SIZE 11
fied by measuring the distance between a pin in the 42 mm 0
MM

iliac wing and a mark on the greater trochanter 36 mm 10

+16* 20
before hip dislocation. This measurement should +12*
+8 30
29 mm +4
be recorded. It is compared later in surgery to +0 40

0 –3 50
a measurement using the same reference points 10
60
20
after the implant trials are in place. 30
Neck Length
Standard Offset: 31 mm
High Offset: 36 mm
70

Porous 80
Coating
For use with Smith & Nephew
Another method of leg measurement consists of the 11
Level

HA
12/14 femoral heads only. 90

Coating -3 and +16 femoral heads 100


surgeon placing the foot of the affected limb on top Level available in 28 mm and
32 mm only. 110

of the unaffected limb. The relative position of one 11 mm *Denotes skirted heads
120

knee to another is then analyzed. This second 130

SIZE 140

method is slightly less accurate than the previously 145 mm


11 150

mentioned technique; however, both methods pro- 160

™ 170

vide a reasonable degree of accuracy in restoring 180

limb length equality. 120% MAGNIFICATION


190

200

When determining which size Synergy stem to use, 210

Figure 3. Anteroposterior radiograph demonstrating proper


©1998 Smith & Nephew, Inc.,
Memphis, TN, U.S.A. 220

the anteroposterior and the lateral radiographs templating of a femur.


Synergy is a trademark of
Smith & Nephew, Inc.
01/98 7138-0370

should be templated. Using the anteroposterior


radiograph, place the femoral templates over the
proximal femur of both the affected and unaffected
hips. The junction of the lateral femoral neck and
greater trochanter serves as a good reference point
for placement of the X-ray templates. Place a mark
at this junction and in the center of the femoral
Figure 4.
head. Align the lateral shoulder of the prosthesis Anteroposterior
with the mark at the junction. Find the appropriate radiograph of a
stem that fits and fills the proximal femur and properly implanted
porous-coated
whose neck length matches the center of the femoral Synergy stem.
head. This is demonstrated in Figure 3.
It is important to check that the stem fits properly
into the femur on the lateral radiograph. As stated
earlier, it is the lateral radiograph that shows best
where three-point fixation will occur.
A properly implanted porous-coated Synergy stem
that provides both normal leg length and offset
is shown in Figure 4.


9
SHORT TECHNIQUE
1. FEMORAL OSTEOTOMY 2. PREPARE ACETABULUM

If acetabular reconstruc-
tion is required, prepare
acetabulum using the
surgical technique for
the intended acetabular
component.

3. CANAL PREPARATION 4. FEMORAL REAMING

5. BROACH ASSEMBLY/DISASSEMBLY 6. FEMORAL BROACHING


10
SHORT TECHNIQUE
7. CALCAR PREPARATION 8. TRIALING
(OPTIONAL)

9. STEM/IMPACTOR ASSEMBLY 10. STEM INSERTION

11. FINAL TRIAL REDUCTION 12. FEMORAL HEAD ASSEMBLY


11
SURGICAL TECHNIQUE

1. Femoral
Osteotomy
Use the osteotomy guide
to determine level of
femoral neck resection.
Osteotomize the femoral
neck (Figure 1).

Figure 1

2. Prepare
Acetabulum
If acetabular reconstruction
is required, prepare the
acetabulum using the
surgical technique for
the intended acetabular
component.


12
SURGICAL TECHNIQUE

3. Femoral Canal
Preparation
Open the medullary canal
with the box osteotome
(Figure 2). Sound the
femoral canal using the Figure 2
canal finder and modular
T-handle (Figure 3).

NOTE: It is important to
stay lateral with both the
box osteotome and canal
finder. This helps ensure
that femoral reaming and
broaching will be in align-
ment with the femoral axis.

Figure 3


13
SURGICAL TECHNIQUE

4. Femoral
Reaming
Begin reaming the femoral
canal with the smallest
Figure 4
tapered reamer. Continue
enlarging the canal with
successive reamers. Stop
reaming when the templated
size reamer mark reaches the
medial calcar (Figure 4).

NOTE: It is important to stay


lateral with the femoral ream-
ers to help ensure alignment
with the femoral axis.


14
SURGICAL TECHNIQUE

5. Broach Assembly/
Disassembly
Assemble the broach to the broach
handle by placing the broach post
in the clamp. Use thumb to lock the
clamp onto the broach. A modular
anteversion handle can be assem- Figure 5
bled to the broach handle to pro-
vide version control (Figure 5).
Disassemble the broach from the
broach handle by placing two
fingers (index and middle) into
the rectangular slot. Apply pressure
to the release bar by squeezing
two fingers toward thumb resting
on the medial side of the broach
handle frame (Figure 6).

Figure 6


15
SURGICAL TECHNIQUE

6. Femoral
Broaching
Start the broaching procedure
along the axis of the femur
with a broach at least two
sizes smaller than the last
reamer used. Successively
broach up to the templated
size. Stop broaching when Figure 7
the top of the templated size
broach coincides with the
neck resection (Figure 7).
NOTE: It is important to
stay lateral with the femoral
broaches. This helps ensure
that the final implant will be
aligned with the femoral axis.

Check for stability by retro-


verting the broach in the
canal with the anteversion
handle attached. Watch for
movement between the
broach and bone. If there
is little or no movement, the
templated size is appropriate
and final implantation may
proceed. If there is gross
movement, ream one size
larger and broach to this
new size. Check for stability
again and repeat procedure
until stability is acquired.


16
SURGICAL TECHNIQUE

7. Calcar
Preparation (Optional)
Figure 8

With the final broach fully seated,


remove the broach handle and
ream the calcar with the appropri-
ate calcar reamer, if desired
(Figure 8).

8. Trial Reduction
Place the standard or high
offset trial neck (as determined by
templating) onto the broach post.
Select the trial femoral head of Figure 9
desired diameter and neck length
and reduce the hip to assess
stability (Figure 9 and Table 1).
If trialing for the universal Bipolar
or Unipolar, trial according to the
appropriate technique for the
selected device.

Reduce the hip and evaluate


range of motion.

FEMORAL HEAD AND NECK LENGTH OPTIONS


Trial
Color 22 mm 26 mm 28 mm 32 mm
Green — — –3 –3
Yellow +0 +0 +0 +0
Red +4 +4 +4 +4
White +8 +8 +8 +8
Blue +12* +12* +12* +12*
Black — — +16* +16*
Table 1. Femoral head and neck length options.
*Denotes skirted heads.


17
SURGICAL TECHNIQUE

9. Range of Motion
1. Soft tissue tension – some shuck is
normal when applying a longitudinal
Figure 10
distraction force to the hip. Shuck
should not be excessive, and the hip
should not dislocate (Figure 10).
2. Anterior stability – place the leg in
full adduction, full extension and hyper-
extension, while exerting an external
rotation force. If the hip cannot be fully
extended, it may be too tight. If it dislocates Figure 11
easily, it is too loose and impingement
must be addressed or component
malposition exists (Figure 11).
3. Posterior stability – place the leg in neutral
adduction and 90° flexion. Gradually rotate
internally. If it dislocates with minimal inter-
nal rotation, it is too loose and impingement
must be addressed or component malposition Figure 12
exists (Figure 12).
4. Sleep position – place the leg in the “sleep
position” with the operated leg semiflexed,
adducted and internally rotated over the
other leg. Apply axial force to try to dislo-
cate. This position represents a dangerous
unstable position that may be adopted by Figure 13
a patient sleeping on their nonoperated
side (Figure 13).


18
SURGICAL TECHNIQUE

10. Stem/Impactor
Assembly
Place the stem inserter pommel through
Figure 14
the stem inserter frame and stand upright
so that the threaded tip is pointed up
(Figure 14). Screw the implant onto
the threaded tip as far as possible. Figure 15

Flip the assembly over so that the stem


tip is now pointing down (Figure 15).
Engage the frame tines into the slots
adjacent to the threaded hole on the
stem. Screw the pommel until assembly
is secure (Figure 16).

Figure 16

11. Stem Insertion


Insert the selected femoral stem
into the canal. Apply hand pressure
and rotate the stem into the correct
position. Use gentle mallet blows
to seat the stem to the position of
the neck resection (Figure 17).
Check stem stability.
If the implant has stopped moving Figure 17
with gentle mallet blows and is not
completely seated, remove the stem
and repeat the same size reaming
and broaching steps.
CAUTION: Do not use excessive
force to seat the stem.


19
SURGICAL TECHNIQUE

12. Final Trial


Reduction
Figure 18

A final trial reduction may be


performed at this time using
trial femoral heads (Figure 18).

13. Femoral
Assembly
Head
Clean and dry the neck taper
with a clean, sterile cloth.
Place the prosthetic femoral
head on the neck taper and
firmly impact with the femoral Figure 19
head impactor and a mallet
several times (Figure 19).


20
SYNERGY FEMORAL STEM & HEAD COMPONENTS

C A T A L O G
Porous-Coated Stems HA-Coated Stems
Titanium 6Al-4V Titanium 6Al-4V
Standard High Offset Standard High Offset
Size Length Cat. No. Cat. No. Size Length Cat. No. Cat. No.

I N F O R M A T I O N
8* — 7130-6608 — 9 135 mm 7130-6709 7130-6409
9 135 mm 7130-6619 7130-6109 10 140 mm 7130-6710 7130-6410
10 140 mm 7130-6610 7130-6110 11 145 mm 7130-6711 7130-6411
11 145 mm 7130-6611 7130-6111 12 150 mm 7130-6712 7130-6412
12 150 mm 7130-6612 7130-6112 13 155 mm 7130-6713 7130-6413
13 155 mm 7130-6613 7130-6113 14 160 mm 7130-6714 7130-6414
14 160 mm 7130-6614 7130-6114 15 165 mm 7130-6715 7130-6415
15 165 mm 7130-6615 7130-6115 16 170 mm 7130-6716 7130-6416
16 170 mm 7130-6616 7130-6116 17 175 mm 7130-6717 7130-6417
17 175 mm 7130-6617 7130-6117 18 180 mm 7130-6718 7130-6418
18 180 mm 7130-6618 7130-6118
19* 185 mm 7192-6107 —
20* 190 mm 7192-6108 —

*Available by special request

Cemented Synergy Stems Conquest Fx™ Stems


Forged CoCr CoCr
Standard High Offset
Size Length Cat. No. Cat. No. Size Length Cat. No.
9 110 mm 7131-6009 7131-6209 9 125 mm 7131-6509
10 115 mm 7131-6010 7131-6210 10 130 mm 7131-6510
11 120 mm 7131-6011 7131-6211 11 135 mm 7131-6511
12 125 mm 7131-6012 7131-6212 12 140 mm 7131-6512
13 130 mm 7131-6013 7131-6213 13 145 mm 7131-6513
14 135 mm 7131-6014 7131-6214 14 150 mm 7131-6514
15 140 mm 7131-6015 7131-6215 15 150 mm 7131-6515
16 140 mm 7131-6016 7131-6216 16 150 mm 7131-6516
17 140 mm 7131-6017 7131-6217 17 150 mm 7131-6517
18 150 mm 7131-6518

21
SYNERGY FEMORAL STEM & HEAD COMPONENTS
I N F O R M A T I O N

Zirconia 12/14 Taper Femoral Heads


Neck Length 22 mm 26 mm 28 mm
+0 — 7132-0026 7132-0028
+4 7132-0422 7132-0426 7132-0428
+8 7132-0822 7132-0826 7132-0828
C A T A L O G

CoCr 12/14 Taper Femoral Heads


Cobalt Chromium – ASTM F 799
Neck Length 22 mm 26 mm 28 mm 32 mm
–3 — — 7130-2803 7130-3203
+0 7130-2200 7130-2600 7130-2800 7130-3200
+4 7130-2204 7130-2604 7130-2804 7130-3204
+8 7130-2208 7130-2608 7130-2808 7130-3208
+12 7130-2212 7130-2612 7130-2812 7130-3212
+16 — — 7130-2816 7130-3216

22
SYNERGY INSTRUMENTATION

C A T A L O G
I N F O R M A T I O N
Femoral Instrumentation Tray No.1
Cat. No. 7136-6201

Tapered Reamer
Cat. No. Size
Osteotomy Guide 7136-6209 8-9-10
Cat. No. 7136-4000 7136-6211 11-12
7136-6213 13-14
7136-6215 15-16
7136-6217 17-18

Box Osteotome
Cat. No. Size Broach Handle
7136-4002 Small (2 Per Set)
7136-4003 Large Cat. No. 7136-4007

Anteversion
Canal Finder Handle
Cat. No. 7136-4001 Cat. No. 7136-4012

T-Handle
Cat. No. 7136-4006

23
SYNERGY INSTRUMENTATION
I N F O R M A T I O N

Femoral Instrumentation Tray No. 2


Cat. No. 7136-6203

Trial Neck, Standard


Cat. No. Size
Tapered Broach
C A T A L O G

7136-6408 8-13
Cat. No. Size 7136-6414 14-18
7136-6308 8
7136-6309 9
7136-6310 10
7136-6311 11
7136-6312 12
7136-6313 13 Trial Neck, High Offset
7136-6314 14 Cat. No. Size
7136-6315 15 7136-6508 8-13
7136-6316 16 7136-6514 14-18
7136-6317 17
7136-6318 18

Calcar Reamer
Stem Inserter Cat. No. Size
Frame 7136-4004 Small
Cat. No. 7136-4008 7136-4005 Large

Cemented Stem Inserter Femoral Head Impactor


Cat. No. 7136-4014 Cat. No. 7136-4009

Stem Inserter Pommel Slap Hammer Weight


Cat. No. 7136-4011 Cat. No. 7136-4010

Trial 12/14 Taper Femoral Heads


Neck Color 22 mm 26 mm 32 mm
Length Code Optional Optional 28 mm Optional
–3 Green — — 7135-2803 7135-3203
+0 Yellow 7135-2200 7135-2600 7135-2800 7135-3200
+4 Red 7135-2204 7135-2604 7135-2804 7135-3204
+8 White 7135-2208 7135-2608 7135-2808 7135-3208
+12 Blue 7135-2212 7135-2612 7135-2812 7135-3212
+16 Black — — 7135-2816 7135-3216

24
OPTIONAL SYNERGY INSTRUMENTATION

C A T A L O G
I N F O R M A T I O N
Collared Trial Neck Sterilization Tray
Cat. No. 7136-6204

Collared Trial Neck, Collared Trial Neck,


Standard High Offset
Cat. No. Size Cat. No. Size
7136-6908 8-9 7136-6918 8-9
7136-6910 10-11 7136-6920 10-11
7136-6912 12-13 7136-6922 12-13
7136-6914 14-15 7136-6924 14-15
7136-6916 16-17 7136-6926 16-17

Fully Toothed Broach


Cat. No. Size Cat. No. Size
7136-6708 8 7136-6714 14
7136-6709 9 7136-6715 15
7136-6710 10 7136-6716 16
7136-6711 11 7136-6717 17
7136-6712 12 7136-6718 18
7136-6713 13

25
IMPORTANT MEDICAL INFORMATION
Warnings and Precautions
Total Hip System
IMPORTANT NOTE Taper Sleeves b. insufficient quantity or quality of bone support, e.g., osteoporo-
Total hip replacement arthroplasty has become a successful procedure sis, or metabolic disorders which may impair bone formation,
A taper sleeve is required to be impacted on the small taper femoral
for relieving pain and restoring motion in patients who are disabled and osteomalacia; and
component prior to impacting a femoral head size 26, 28, or 32 mm. A
from hip arthropathy. The goals of total hip replacement are to
taper sleeve is required to attach a unipolar head. Unipolar taper c. infections or other conditions which lead to increased bone
decrease pain, increase function, and increase mobility.
sleeves are available in small, large, and 12/14 tapers. Never place resorption.
MATERIALS
more than one taper sleeve on a femoral component.
The Total Hip System is manufactured from materials as outlined 2. Mental or neurological conditions which tend to impair the patient's
below. The component material is provided on the outside carton label. Femoral Heads ability or willingness to restrict activities.
Cobalt chromium (22, 26, 28, and 32 mm) and ceramic (22, 26, 28, and 3. Physical conditions or activities which tend to place extreme loads
Component Material Material Standards
32 mm) heads are available in multiple neck lengths for proper on implants, e.g., Charcot joints, muscle deficiencies, multiple joint
Femoral Components Ti-6Al-4V ASTM F 136 and ISO 5832/3 or anatomic and musculature fit. Heads are highly polished for reduced disabilities, etc.
or ASTM F 1472 and ISO 5832/3 friction and wear. The following zirconia ceramic heads are available
4. Skeletal immaturity.
Co-Cr-Mo or for use only with small (0.404) and large (.564) taper femoral compo-
nents. 5. The zirconia ceramic head is contraindicated for use with any other
ASTM F 799 and ISO 5832/12 product than an UHMW polyethylene cup or a metal backed
Zirconia Head
or UHMW polyethylene cup.
Ceramic Diameter Neck Length
ASTM F 75 and ISO 5832/4 42-7815 32 mm Standard 0 mm Contraindications may be relative or absolute and must be carefully
Acetabular shells Ti-6Al-4V ASTM F 1472 and ISO 5832/3 42-7816 32 mm Long 4 mm weighted against the patient's entire evaluation and the prognosis for
42-7817 32 mm X-Long 8 mm possible alternative procedures such as non-operative treatment,
Proximal pads
42-7818 28 mm Standard 0 mm arthrodesis, femoral osteotomy, pelvic osteotomy, resection arthro-
Taper sleeves 42-7819 28 mm Long 4 mm plasty, hemiarthroplasty and others.
Distal sleeves 42-7820 28 mm X-Long 8 mm Conditions presenting increased risk of failure include: osteoporo-
Fixation screws and pegs Note: 32 mm heads with a -3 mm neck length are not available for use sis, metabolic disorders which may impair bone formation, and
Hole covers with the small taper stems. osteomalacia.
Acetabular components UHMWPE ASTM F 648 In addition to the components listed above, the following components Possible Adverse Effects
Acetabular liners are available for use only with small (0.404) taper femoral components 1. Wear of the polyethylene articulating surfaces of acetabular com-
Femoral centralizers PMMA Not applicable Zirconia Head ponents has been reported following total hip replacement. Higher
Ceramic Diameter Neck Length rates of wear may be initiated by the presence of particles of
Acetabular spacer pods
7132-0002 22 mm Long 4 mm cement, metal, or other debris which can develop during or as a
X-ray marking wire Co-Cr-Mo ASTM F 90 and ISO 5832/5 7132-0006 22 mm X-Long 8 mm result of the surgical procedure and cause abrasion of the articu-
Acetabular Reconstruction CP Titanium ASTM F 67 and ISO 5832/2 Note: 22 mm Zirconia Ceramic Heads used with small (0.404) taper lating surfaces. Higher rates of wear may shorten the useful life of
Ring femoral components are not available in the USA. the prosthesis, and lead to early revision surgery to replace the
Acetabular worn prosthetic components.
Reinforcement Ring The following zirconia ceramic heads are available for use only with 2. With all joint replacements, asymptomatic, localized, progressive
12/14 taper femoral components: bone resorption (osteolysis) may occur around the prosthetic com-
Femoral Heads Co-Cr-Mo ASTM F 799 and ISO 5832/12
Zirconia Head ponents as a consequence of foreign-body reaction to particulate
Zirconia ISO 13356
Ceramic Diameter Neck Length wear debris. Particles are generated by interaction between com-
Ceramic ponents, as well as between the components and bone, primarily
7132-0028 28 mm Standard 0 mm
7132-0428 28 mm Long 4 mm through wear mechanisms of adhesion, abrasion, and fatigue.
Porous titanium components and porous Co-Cr-Mo components are 7132-0828 28 mm X-Long 8 mm Secondarily, particles may also be generated by third-body parti-
coated with commercially pure (C.P.) titanium beads (ASTM F 67 and 7132-0026 26 mm Standard 0 mm cles lodged in the polyethylene articular surface. Osteolysis can
ISO 5832/2) and Co-Cr-Mo beads (ASTM F 75), respectively. 7132-0426 26 mm Long 4 mm lead to future complications necessitating the removal or replace-
Hydroxylapatite coatings include HA (ASTM F 1185) that is applied 7132-0826 26 mm X-Long 8 mm ment of prosthetic components.
either on a grit blasted or porous surface. NOTE: HA coated porous 7132-0422 22 mm Long 4 mm 3. Loosening, bending, cracking, or fracture of implant components
implants are not available in the USA. 7132-0822 22 mm X-Long 8 mm may result from failure to observe the Warnings and Precautions
Zirconia ceramic femoral heads are yttria stabilized zirconia ceramic. Acetabular Components below. Fracture of the implant can occur as a result of trauma,
Acetabular components can be one piece all polyethylene or two- strenuous activity, improper alignment, or duration of service.
Some of the alloys needed to produce orthopedic implants contain
some metallic components that may be carcinogenic in tissue cultures piece components consisting of a titanium shell and a polyethylene 4. Dislocations, subluxation, decreased range of motion, or lengthen-
or intact organism under very unique circumstances. Questions have liner. Please see Warnings and Precautions for specific information ing or shortening of the femur caused by improper neck selection,
been raised in the scientific literature as to whether or not these alloys on screws, pegs and hole covers use. Acetabular reinforcement and positioning, looseness of acetabular or femoral components, extra-
may be carcinogenic in implant recipients. Studies conducted to eval- reconstruction rings are used with an all polyethylene acetabular neous bone, penetration of the femoral prosthesis through the shaft
uate this issue have not identified conclusive evidence of such phe- component. of the femur, fracture of the acetabulum, intrapelvic protrusion of
nomenon, in spite of the millions of implants in use. Femoral components and femoral heads are designed for use with any acetabular component, femoral impingement, periarticular calcifi-
Smith & Nephew polyethylene acetabular component or polyethylene- cation, and/or excessive reaming.
DESCRIPTION OF SYSTEM
lined, metal-backed acetabular component having an appropriately- 5. Fracture of the pelvis or femur: post-operative pelvic fractures are
The Total Hip System consists of femoral components, proximal sized inside diameter. usually stress fractures. Femoral fractures are often caused by
pads, taper sleeves, distal sleeves, acetabular components, fixation
INDICATIONS, CONTRAINDICATIONS, AND ADVERSE EFFECTS defects in the femoral cortex due to misdirected reaming, etc.
screws and pegs, hole covers, centralizers, and femoral heads.
Intraoperative fractures are usually associated with old congenital
Components may be grit blasted, porous coated, hydroxylapatite Hip components are indicated for individuals undergoing primary and deformity, improper stem selection, improper broaching, and/or
(HA) coated, or HA porous coated. All implantable devices are revision surgery where other treatments or devices have failed in reha- severe osteoporosis.
designed for single use only. bilitating hips damaged as a result of trauma or noninflammatory
degenerative joint disease (NIDJD) or any of its composite diagnoses 6. Infection, both acute post-operative wound infection and late deep
Femoral Components
of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital wound sepsis.
Femoral components are available in a variety of sizes. Porous coated epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. 7. Neuropathies; femoral, sciatic, peroneal nerve, and lateral femoral
components are coated for biological ingrowth. Proximally and distally
Hip components are also indicated for inflammatory degenerative joint cutaneous neuropathies have been reported. Temporary or perma-
modular femoral components accept proximal pads and distal sleeves,
disease including rheumatoid arthritis, arthritis secondary to a variety nent nerve damage resulting in pain or numbness of the affected limb.
respectively. Non-porous femoral components can feature PMMA cen-
tralizers that help produce a uniform thickness of cement in a concen- of diseases and anomalies, and congenital dysplasia; old, remote 8. Wound hematoma, thromboembolic disease including venous
tric manner. osteomyelitis with an extended drainage-free period, in which case, the thrombosis, pulmonary embolus, or myocardial infarction.
patient should be warned of an above normal danger of infection post-
Femoral components are available with a small, large (14/16), or operatively; treatments of nonunion, femoral neck fracture and 9. Myositis ossificans, especially in males with hypertrophic arthritis,
12/14 global taper (gage diameters 0.404, 0.564, and 0.500 inches, trochanteric fractures of the proximal femur with heads involvement limited pre-operative range of motion and/or previous myositis.
respectively). that are unmanageable using other techniques; endoprosthesis, Also, periarticular calcification with or without impediment to joint
femoral osteotomy, or Girdlestone resection; fracture-dislocation of the mobility can cause decreased range of motion.
Small taper femoral components mate and lock directly with a 22 mm
metal or ceramic head. The small taper also mates with a taper hip; and correction of deformity. 10. Trochanteric nonunion usually associated with early weight bearing
sleeve which, in turn, mates with either metal or ceramic heads (26, Acetabular reinforcement and reconstruction rings are intended to be and/or improper fixation of the trochanter, when a transtrochanteric
28, or 32 mm), bipolar or unipolar components. used in primary and revision surgeries where the acetabulum has the surgical approach is used.
Large taper femoral components mate and lock with either metal heads deficiencies of the acetabular roof, anterior or posterior pillar, medial 11. Although rare, metal sensitivity reactions and/or allergic reactions
(26, 28, or 32 mm), ceramic heads (22, 28 or 32 mm), bipolar or unipo- wall deficiency, and / or protrusion as a result of the indications listed to foreign materials have been reported in patients following joint
lar components. previously. replacement.
Femoral components with a 12/14 taper mate and lock with either Some of the diagnoses listed above and below may also increase the 12. Damage to blood vessels.
metal heads (22, 26, 28, or 32 mm), ceramic heads (26 or 28 mm), chance of complications and reduce the chance of a satisfactory result.
13. Traumatic arthrosis of the knee from intraoperative positioning of
bipolar or unipolar components. Contraindications the extremity.
Small, large, and 12/14 taper femoral component tapers are 1. Conditions that would eliminate or tend to eliminate adequate 14. Delayed wound healing.
machined to mate and lock with ceramic heads, thus preventing rota- implant support or prevent the use of an appropriately-sized
tion of the ceramic head on the stem, the latter would cause wear of 15. Aggravated problems of the affected limb or contralateral extremity
implant, e.g.:
the stem taper. caused by leg length discrepancy, excess femoral medialization, or
a. blood supply limitations; muscle deficiency.

26
16. Failure of the porous coating/ substrate interface or hydroxylapatite 7. Modular heads and femoral components should be from the same and patient care, are extremely important. Gradual weight bearing
coating/ porous coating bonding may result in bead separation manufacturer to prevent mismatch of tapers. is begun after surgery in ordinary total hip arthroplasty. However,
delamination. with trochanter osteotomy or certain complex cases, weight-bear-
8. Clean and dry stem taper prior to impacting the femoral head or
17. Stem migration or subsidence has occurred in conjunction with ing status should be individualized with the non or partial weight-
taper sleeve. The modular femoral head component must be firmly
compaction grafting procedures usually resulting from insufficient bearing period extended.
seated on the femoral component to prevent disassociation.
graft material or improper cement techniques. Varus stem align- 2. Patients should be warned against unassisted activity, particularly
9. Take care, when positioning and drilling screw and peg holes, to
ment may also be responsible. use of toilet facilities and other activities requiring excessive motion
avoid penetration of the inner cortex of the pelvis, penetration of the
WARNINGS AND PRECAUTIONS of the hip.
sciatic notch, or damage to vital neurovascular structures.
The patient should be warned of surgical risks, and made aware of Perforation of the pelvis with screws that are too long can rupture 3. Use extreme care in patient handling. Support should be provided
possible adverse effects. The patient should be warned that the device blood vessels causing the patient to hemorrage. Do not place a to the operative leg when moving the patient. While placing the
does not replace normal healthy bone, that the implant can break or screw in the center hole of the acetabular prosthesis. patient on bedpans, changing dressings, and clothing, and similar
become damaged as a result of strenuous activity or trauma, and that activities, precautions should be taken to avoid placing excessive
Placement of drills and screws in the anterior or medial portions of
it has a finite expected service life and may need to be replaced in the load on the operative part of the body.
the prosthesis is associated with a high risk of potentially fatal vas-
future. Do not mix components from different manufacturers. Additional cular injury. 4. Postoperative therapy should be structured to regain muscle
Warnings and Precautions may be included in component literature. strength around the hip and a gradual increase of activities.
Bone screws must be completely seated in the holes of the shell to
Preoperative allow proper locking for the acetabular component liner. If the 5. Periodic x-rays are recommended for close comparison with imme-
1. Use extreme care in handling and storage of implant components. tapered pegs need to be removed from the shell after impaction of diate postoperative conditions to detect long-term evidence of
Cutting, bending, or scratching the surface of components can sig- the pegs, do not reuse the pegs or the peg shell holes. Use new changes in position, loosening, bending and/or cracking of compo-
nificantly reduce the strength, fatigue resistance, and/or wear char- pegs and different shell holes, or a new shell if necessary. nents or bone loss. With evidence of these conditions, patients
acteristics of the implant system. These, in turn, may induce inter- should be closely observed, the possibilities of further deterioration
10. USE ONLY REFLECTION® TITANIUM BONE SCREWS, UNIVER-
nal stresses that are not obvious to the eye and may lead to frac- SAL CANCELLOUS BONE SCREWS, TAPERED PEGS, AND evaluated, and the benefits of early revision considered.
ture of the component. Implants and instruments should be pro- HOLE COVERS with the Reflection Acetabular Component and 6. Prophylactic antibiotics should be recommended to the patient sim-
tected from corrosive environments such as salt air during storage. USE ONLY OPTI-FIX® TITANIUM BONE SCREWS AND UNIVER- ilar to those suggested by the American Heart Association for con-
Do not allow the porous surfaces to come in contact with cloth or SAL CANCELLOUS BONE SCREWS with the Opti-Fix Acetabular ditions or situations that may result in bacteremia.
other fiber-releasing materials. Component. The Reflection Interfit and the Reflection For Screws
PACKAGING AND LABELING
Only (FSO) shells accept Universal Cancellous, Reflection screws,
2. Allergies and other reactions to device materials, although infre-
and tapered screw-hole covers, not pegs. Tapered pegs can only Components should only be accepted if received by the hospital or sur-
quent, should be considered, tested for (if appropriate), and ruled
be used with Reflection V Shells. The threaded center hole in geon with the factory packaging and labeling intact.
out preoperatively.
Reflection Shells only accepts the threaded hole cover, not screws STERILIZATION/RESTERILIZATION
3. Fixation and expected longevity of components expected to be left or pegs. The InterFit threaded hole cover is only for use with
in place at revision surgery should be thoroughly assessed. Reflection Interfit. The Reflection threaded hole cover can be used Most implants are supplied sterile and have been packaged in protec-
with both Reflection and InterFit shells. Refer to product literature tive trays. The method of sterilization is noted on the package label. All
4. Surgical technique information is available upon request. The sur-
for proper adjunctive fixation and hole cover usage. radiation sterilized components have been exposed to a minimum of
geon should be familiar with the technique.
25 kilo Grays of gamma radiation. If not specifically labeled sterile, the
5. Intraoperative fracture or breaking of instruments can occur. 11. Prior to seating modular components, surgical debris including tis- implants and instruments are supplied non-sterile and must be steril-
Instruments which have experienced extensive use or excessive sue must be cleaned from the surfaces. Debris, including bone ized prior to use. Inspect packages for punctures or other damage prior
force are susceptible to fracture. Instruments should be examined cement, may inhibit the component locking mechanism. If the shell to surgery.
for wear, or damage, prior to surgery. is to be cemented in place, remove extraneous cement with a plas-
tic sculps tool to ensure proper locking of the liner. During liner Metal Components
6. Do not cold water quench ceramic components and never sterilize insertion, make sure soft tissue does not interfere with the Nonporous metal components may be initially sterilized or resterilized,
ceramic heads while fixed on the stem taper. (See sterilization sec- shell/liner interface. Chilling the liner reduces the impaction force if necessary, by steam autoclaving in appropriate protective wrapping,
tion, below.) required to seat the liner. Modular components must be assembled after removal of all original packaging and labeling. Protect the
7. Select components such that the Zirconia ceramic head always securely to prevent disassociation. Debris inhibits the proper fit and devices, particularly mating surfaces, from contact with metal or other
articulates with a UHMW polyethylene cup or a metal backed locking of modular components which may lead to early failure of hard objects which could damage the product. The following process
UHMW polyethylene cup. Zirconia ceramic should never articulate the procedure. Failure to properly seat the acetabular liner into the parameters are recommended for these devices:
against metal because severe wear of the metal will occur. shell can lead to disassociation of the liner from the shell.
• Prevacuum Cycle: 4 pulses (Maximum = 26.0 psig [2.8 bars] &
8. Select only Smith & Nephew femoral components that indicate 12. Avoid repeated assembly and disassembly of the modular compo- Minimum = 10.0 inHg [339 millibars]) with a minimum dwell time of 4
their use with ceramic heads. This is important because the taper nents which could compromise the critical locking action of the minutes at 270°F to 275°F (132°C to 135°C), followed by a 1 minute
on the stem is machined to tightly mate and lock with the ceramic locking mechanism. purge and at least 15 minutes of vacuum drying at 10 inHg (339 mil-
head thus preventing rotation of the ceramic head on the stem. 13. Care is to be taken to assure complete support of all parts of the libars) minimum.
Also, an improperly dimensioned taper could result in fracture of device embedded in bone cement to prevent stress concentration • Gravity Cycle: 270°F to 275°F (132°C to 135°C) with a minimum
the ceramic head. which may lead to failure of the procedure. During curing of the dwell time at temperature of 15 minutes, followed by a 1 minute
9. The zirconia ceramic head is composed of a new ceramic material cement, care should be taken to prevent movement of the implant purge and at least 15 minutes of vacuum drying at 10 inHg (339 mil-
with limited clinical history. Although mechanical testing demon- components. libars) minimum.
strates that when used with a polyethylene acetabular component, 14. If components are to be left in place at revision surgery, they should Smith & Nephew does not recommend the use of low temperature
the yttria stabilized zirconia ball produces a relatively low amount first be thoroughly checked for signs of looseness, etc. and gravity cycles or flash sterilization on implants.
of particulates, the total amount of particulate remains undeter- replaced if necessary. The head/neck component should be
mined. Because of the limited clinical and preclinical experience, changed only when clinically necessary. Do not resterilize femoral prostheses with ceramic heads seated on
the biological effect of these particulates can not be predicted. the stem.
15. Once removed from the patient, implants previously implanted
Intraoperative should never be reused, since internal stresses which are not visi- If porous coated implants are inadvertently contaminated, return the
ble may lead to early bending or fracture of these components. unsoiled prosthesis to Smith & Nephew for resterilization. DO NOT
1. The general principles of patient selection and sound surgical judg-
RESTERILIZE porous coated implants. The porous coating requires
ment apply. The correct selection of the implant is extremely impor- 16. With the congenitally dislocated hip, special care should be taken special cleaning procedures.
tant. The appropriate type and size should be selected for patients to prevent sciatic nerve palsy. Also, note that the femoral canal is
with consideration of anatomical and biomechanical factors such often very small and straight and may require an extra-small Plastic Components
as patient age and activity levels, weight, bone and muscle condi- straight femoral prosthesis; however, a regular-sized prosthesis Plastic components may be resterilized by ethylene oxide gas. The fol-
tions, any prior surgery and anticipated future surgeries, etc. should be used when possible. Note that the true acetabulum is lowing parameters are recommended as starting points for cycle vali-
Generally, the largest cross-section component which will allow rudimentary and shallow. A false acetabulum should not ordinarily dation by the health care facility:
adequate bone support to be maintained is preferred. Failure to be utilized as a cup placement site for anatomical and biomechan-
use the optimum-sized component may result in loosening, bend- ical reasons.
ing, cracking, or fracture of the component and/or bone. Sterilant Temp. Humidity Maximum Concen- Exposure
17. With rheumatoid arthritis, especially for those patients on steroids, Pressure tration Time
2. Correct selection of the neck length and cup, and stem position- bone may be extremely osteoporotic. Care should be taken to pre- 10% EtO 130˚F 40-60% 28 PSIA 550-650 120
ing, are important. Muscle looseness and/or malpositioning of vent excessive penetration of the acetabular floor or fracture of the 90% HCFC (55˚C) (1930 millibar) mg/L minutes
components may result in loosening, subluxation, dislocation, medial acetabular wall, femur, or greater trochanter.
and/or fracture of components. Increased neck length and varus 10% EtO 100˚F 40-60% 28 PSIA 550-650 6
positioning will increase stresses which must be borne by the 18. Revision procedures for previous arthroplasty, Girdlestone, etc., 90% HCFC (38˚C) (1930 millibar) mg/L hours
stem. The component should be firmly seated with the compo- are technically demanding and difficult to exercise. Common 100% EtO 131˚F 30-60% 10 PSIA 736 30
nent insertion instruments. errors include misplacement of the incision, inadequate exposure (55˚C) (689 millibar) mg/L minutes
or mobilization of the femur, inadequate removal of ectopic bone,
3. Care should be taken not to scratch, bend (with the exception of or improper positioning of components. Postoperative instability
the Reconstruction Rings) or cut metal components during surgery Suggested initial starting point for aeration validation is 12 hours at
as well as excessive blood loss can result. In summary, increased 122˚F (50˚C) with power aeration. Consult aerator manufacturer for
for the reasons stated in Number One of the "Preoperative" section operative time, blood loss, increased incidence of pulmonary
of "Warnings and Precautions." more specific instructions.
embolus and wound hematoma can be expected with revision
4. A +12 mm or +16 mm femoral head should not be used with procedures. Ceramic Components
any small taper stems. 19. Prior to closure, the surgical site should be thoroughly cleaned of Do not resterilize ceramic femoral heads.
5. Distal sleeves should not be used to bridge cortical defects cement, bone chips, ectopic bone, etc. Ectopic bone and/or bone INFORMATION
that lie within 25 mm of the tip of the base stem. spurs may lead to dislocation or painful or restricted motion.
Range of motion should be thoroughly checked for early contact For further information, please contact Customer Service at (800) 238-
6. Matrix small taper stem sizes 8S–10L must have a minimum neck or instability. 7538 for calls within the continental USA and (901) 396-2121 for all
length of +8 mm when used with a bipolar component; and small international calls.
taper stem sizes 12S–16L must have a minimum neck length of Postoperative
Caution: Federal Law (USA) restricts this device to sale by or on
+4 mm when used with a bipolar component. 1. Postoperative directions and warnings to patients by physicians, the order of a physician.

3433198 Rev. 0 10/98

27
Synergy is a trademark of Smith & Nephew, Inc. ©1999 Smith & Nephew, Inc. 8/99 7138-0613

You might also like