You are on page 1of 6

SYMPOSIUM: NEONATOLOGY

A clinical and lungs, which do not participate in gas exchange. After birth, when
the lungs are filled with air, pulmonary vascular resistance falls,

echocardiographic allowing blood to circulate through the lungs. This removes the
physiological need for the DA, which closes through a number of

approach to evaluation of mechanisms. The DA usually begins shunting from left-to-right,


then closes functionally in the first two to three days, but can

patent ductus arteriosus take longer than three weeks to close anatomically.
In some babies, particularly preterm babies, this process of

in preterm infants closure often does not occur, resulting in a patent ductus arte-
riosus (PDA). The persistence of a PDA occurs as a result of a
number of factors, often related to gestational age. This includes
Kunal Babla immaturity of the musculature within the ductal wall, persistence
Sandeep Shetty of high levels of circulating prostaglandins, high sensitivity of the
preterm duct to vasodilators such as prostaglandins and nitric
Anay Kulkarni oxide, thrombocytopenia, impaired platelet function and adrenal
insufficiency. In addition, conditions such as hypoxia and sepsis
can further promote persistence of PDA.
Abstract The prevalence of PDA falls as gestational age increases, with
The ductus arteriosus provides a physiological right-to-left shunt in up to 56% of infants born under 28 weeks’ gestation affected.
fetal life, before closing early in the postnatal period. In extremely pre- Other authors have reported rates as high as 90% in infants born
term infants, it commonly does not close and remains as a patent duc- at under 24 weeks’ gestation.
tus arteriosus, and can have significant effects on the haemodynamic While it is outside the scope of this paper to discuss the ad-
status of the baby. There is no clear definition for haemodynamically vantages and disadvantages of medical or surgical treatment, we
significant patent ductus arteriosus in preterm infants and debate per- will provide a summary of the strengths and limitations of clin-
sists regarding the benefits of medical or surgical intervention to close ical, echocardiographic and biochemical methods used to define
patent ductus arteriosus. Echocardiography remains the current gold haemodynamically significant PDA (hsPDA), and a concise
standard diagnostic approach. There are several echocardiographic summary of how to apply them in clinical practice.
parameters that can be used to assess patent ductus arteriosus in pre-
term infants, each with their own strengths and limitations. Owing to
Pathophysiology associated with a haemodynamically
these limitations, there is no single echocardiographic parameter
significant PDA
that allows us to define a haemodynamically significant patent ductus
arteriosus. We therefore recommend an approach that encompasses The left-to-right shunt associated with hsPDA results in increased
clinical assessment, and echocardiographic assessment of duct char- pulmonary circulation and therefore, an increased return of blood
acteristics, pulmonary overcirculation and systemic hypoperfusion. to the left side of the heart. This increased pulmonary circulation
Combining these assessments and interpreting them in the context can result in breathlessness, hypoxia and radiographic features of
of the baby’s cardiac anatomy and clinical status will allow clinicians pulmonary plethora. The increased return of blood to the left side
to make a more objective and informed judgement regarding the hae- of the heart can result in left-sided volume overload and dilata-
modynamic significance of a patent ductus arteriosus in preterm tion of the left atrium and ventricle. Concurrently, the left-to-right
infants. shunt can give features of ‘ductal steal’ resulting in low systemic
diastolic pressures and wide pulse pressures.
Keywords echocardiography; heart disease; patent ductus arterio-
It is postulated that hsPDA is associated with a number of adverse
sus; PDA; preterm
outcomes in preterm infants. These include intraventricular hae-
morrhage, bronchopulmonary dysplasia, necrotising enterocolitis
and mortality. There is therefore considerable interest in treatment
Introduction
methods, and defining which babies require treatment for hsPDA.
In fetal life, the ductus arteriosus (DA) serves to allow blood to
pass from the pulmonary artery to aorta and bypass the fluid-filled Bedside assessment
Clinical signs are generally less reliable than echocardiographic
findings. Our recommendation would be that these should not be
Kunal Babla BSc (Hons) MBBS MSc MRCPCH MAcadMEd, Senior Clinical relied upon for diagnostic purposes. Echocardiography remains
Fellow in Paediatric Cardiology and ST8 Trainee in Neonatal the gold standard for diagnosing the presence of a PDA. How-
Medicine, Royal Brompton Hospital, London, UK. Conflicts of ever, given that not all practitioners will readily have access to
interest: none declared. echocardiography, we will discuss the strengths and limitations
Sandeep Shetty MBBS DCH DNB (Paediatrics) FRCPCH (UK) MDres of key clinical findings as part of this review, and have sum-
Consultant Neonatologist, St George’s University Hospital, London, marised them in Table 1.
UK. Conflicts of interest: none declared. PDA characteristically presents with tachypnoea, tachycardia,
Anay Kulkarni MBBS DCH DNB (Paediatrics) FRCPCH (UK) Consultant bounding pulses, a wide pulse pressure and a systolic murmur.
Neonatologist, St George’s University Hospital, London, UK. Given the likely clinical status and coexisting problems (e.g.,
Conflicts of interest: none declared. sepsis, respiratory distress syndrome) of a preterm infant,

PAEDIATRICS AND CHILD HEALTH 30:4 129 Ó 2020 Published by Elsevier Ltd.
SYMPOSIUM: NEONATOLOGY

A summary of the sensitivity and specificity of a range of clinical and echocardiographic parameters, including commonly
used cut off values for defining hsPDA. Note that sensitivity and specificity figures may vary between studies and timing
of echocardiography.
Parameter Timing after birth Cut-off value Sensitivity Specificity
for hsPDA (%) (%)

Clinical/ Murmur Daily from day 1 to day 7 21e79 86e100


radiographic Pulses Daily from day 1 to day 7 31e71 61e86
parameters Pulses 3e7 days 43 74
Increased cardiothoracic ratio 3e7 days 8 94
Duct and shunt PDA Diameter <31 hours 1.5 mm 81 85
characteristics PDA Diameter 12e48 hours 1.5 mm 94 73
PDA:Left Pulmonary 4 days 0.5 78 80
Artery diameter
PDA:Left Pulmonary 3 days 0.5 92 55
Artery diameter
PDA Diameter:weight 12e48 hours >1.5 mm/kg 94 73
Pulsatile flow pattern 24 hours 91 59
Pulsatile flow pattern Daily for 7 days 93 100
Pulsatile flow pattern 3 days 67 78
Growing pattern Daily for 7 days 64 81
Vmax:Vmin First 48 hours >1.9 88 90
1/2 SDT 7 days <90 ms 63 100
Pulmonary LA:Ao <31 hours 1.5 29 91
overcirculation E/A ratio 3 days >1 10 6
Left ventricular output <31 hours 300 ml/kg/min 26 92
Systemic Descending aorta diastolic flow <31 hours Absent/ 68 85
hypoperfusion retrograde

Table 1

tachypnoea and tachycardia are non-specific findings in the haemodynamically insignificant PDA, and in turn, hsPDA.
context of PDA. Heart rate was found to be higher in infants with However, the measured blood pressures have all been within the
hsPDA but still within the normal range, and therefore unlikely expected normal range. Diastolic blood pressure has been found
to be of significant clinical value in determining the presence of to correlate slightly better with the presence of a hsPDA, most
a PDA. likely representing the ductal steal phenomenon as described
above. The classically described wide pulse pressure in the
Murmurs presence of hsPDA does not have strong diagnostic value. Studies
The presence of a murmur was found to be of greater sensitivity have shown that the pulse pressure in the first week of life in
by day seven in comparison to day three (79% vs 31%). The extreme preterm infants does not differ significantly between
specificity of the sign remained relatively high at 86e94%. The those with hsPDA and those without.
results suggest that the presence of a murmur is most likely to be
due to PDA, however the absence of one does not exclude the Chest radiograph
presence of a PDA. However, it should be considered that mur- Chest x-rays will be helpful in the detection of pulmonary plethora
murs could be produced by other congenital defects, most and to help exclude other causes of tachypnoea and the need for
commonly a patent foramen ovale or ventricular septal defect. respiratory support, such as pneumothorax or consolidation. A
cardiothoracic ratio better than 60% has been shown to be highly
Pulse volume specific for the presence of hsPDA (94%) but has very low
Increased pulse volumes do not have sufficient clinical accuracy sensitivity (8%). Chest x-rays are therefore unlikely to be helpful
to be relied on as an isolated marker for hsPDA. They have been in the diagnosis of PDA, but will help investigate other pathology.
found to have sensitivities of 43e71% with a specificity of 61
e86% in two studies. When combined with the presence of a Echocardiographic evaluation of hsPDA
murmur the positive predictive value of increased pulse volume
for hsPDA was found to be 77%. As discussed above, echocardiography remains the gold standard
investigation for diagnosis and evaluation of hsPDA in preterm
Blood pressure infants. Prior to evaluating hsPDA, normal cardiac anatomy
Systolic and mean blood pressure have been found to be higher should be demonstrated. There are three main components to
in infants with a closed duct in comparison to those with the echocardiographic evaluation of hsPDA: duct and shunt

PAEDIATRICS AND CHILD HEALTH 30:4 130 Ó 2020 Published by Elsevier Ltd.
SYMPOSIUM: NEONATOLOGY

characteristics, pulmonary overcirculation and systemic hypo- overestimation of PDA diameter. Studies have shown significant
perfusion. Using these three components, we will examine the inter- and intra-observer variability when measuring PDA
strengths and limitations of each individual parameter to evalu- diameter both in 2D and on colour Doppler imaging.
ation of hsPDA. These are summarised in Table 1. PDA diameter can be indexed against LPA diameter (cut-off
value 0.5) or bodyweight (1.4mm/kg), offering an element of
Duct and shunt characteristics standardisation for patient size, but carries the same inherent
Within the characteristics of the duct, we include three main limitations as described above with regards to PDA diameter.
features: the patency, the size and the direction of the shunt
through the duct. Assessment of flow direction and character: as described
above, simple inspection of the colour Doppler over the PDA can
Patency of the duct: the PDA can be visualised on echocardi- give an impression of the direction of the shunt. However, given
ography through a number of different echo windows. Current the high heart rate in preterm infants, right-to-left shunting in
consensus is that the preferred window to obtain a clear 2D diastole can be easily missed using this approach. Continuous
image is through a left-sided high parasternal “ductal view”. This wave (CW) Doppler through the PDA is a more objective
classically allows the PDA to be visualised leaving the descend- approach to defining the direction and characteristics of the
ing aorta and entering the main pulmonary artery adjacent to the shunt.
left pulmonary artery, as illustrated in Figure 1. Doppler flow patterns through the PDA been characterised as
Using colour Doppler can help to show flow through the non-restrictive (left-to-right) restrictive (left-to-right), bidirec-
duct and therefore demonstrate its patency. Most commonly, tional, and right-to-left. These are illustrated in Figure 2. Non-
the PDA will show as red colour flow (in contrast to the blue restrictive shunts can be objectively defined as having a high
flow of the pulmonary arteries), representing a left-to-right systolic to diastolic velocity gradient (Vmax:Vmin 2.0),
shunt. A right-to-left shunt will show as blue flow through whereas restrictive shunts have a low gradient (Vmax:Vmin
the PDA. This can be easily overlooked and care should be <2.0). A pulsatile, non-restrictive left-to-right Doppler pattern is
taken to ensure that a right-to-left shunt is not missed, as this most sensitive (93.5%) and specific (100%) for predicting the
could be representative of a duct-dependent cardiac lesion or development of hsPDA, whereas a growing pattern (bidirectional
pulmonary hypertension. with right-to-left shunt 30% of the cycle) has less sensitivity
(64.5%) and specificity (81%).
Ductal diameter: ductal diameter is one of the most commonly In addition, measuring the half-systolic decay time (1/2SDT)
used parameters to define an hsPDA. The most commonly used can be derived by measuring the time taken to halve peak sys-
absolute value to define hsPDA is >1.5mm. PDA diameter should tolic velocity across a PDA on CW Doppler. A 1/2SDT (on day
be measured in end-systole at the pulmonary end, or the nar- seven of life) of <90ms showed 62.5% sensitivity and 100%
rowest part of the PDA. We recommend measuring duct diameter specificity for hsPDA.
in 2D imaging as the use of colour Doppler can result in

Figure 1 Echocardiographic images of a patent ductus arteriosus (PDA) demonstrating a left to right shunt. The PDA inserts into the pulmonary
artery (PA) at point A, emerging from the aorta (Ao) at point B. Blue flow signifies normal forward flow through the pulmonary artery. Red flow
signifies flow from the aorta to the pulmonary artery via the PDA.

PAEDIATRICS AND CHILD HEALTH 30:4 131 Ó 2020 Published by Elsevier Ltd.
SYMPOSIUM: NEONATOLOGY

Figure 2 Doppler flow profiles for ductal flow measured from a left-sided high parasternal view.

Pulmonary overcirculation Left ventricular end diastolic diameter (LVEDD): LVEDD can
In the presence of a left-to-right shunt via a PDA, left ventricular be measured in the parasternal long axis view, using M-mode
output represents the total pulmonary venous return, and with the cursor placed through the tips of the mitral valve leaf-
therefore parameters measuring left ventricular output, volume lets, and then standardised by calculating z-scores. LVEDD can
and pressure loading can be used as surrogates for the degree of be helpful for reviewing left ventricular volume loading over
shunting at a ductal level. One should consider that the presence time.
of a shunt via a patent foramen ovale (PFO) can impact mea- Pulmonary vein d wave velocity: using a high parasternal
surements of volume and pressure in the left atrium and view with colour Doppler, the pulmonary veins can be visualised
ventricle. entering the left atrium. Using pulsed-wave Doppler on the lower
pulmonary veins can provide a flow profile (systolic, diastolic
Left ventricular output: left ventricular output can be derived by and atrial contraction components) of pulmonary venous return
measuring the velocity-time integral (stroke distance) from a to the left side of the heart. A diastolic velocity of 0.5 m/s has
pulsed-wave Doppler of flow at the aortic valve and measuring been associated with the presence of hsPDA. This approach has
the size of the aortic annulus. The measurements are applied to the advantage that it is not affected by shunting at the atrial level.
the equation: There is however, currently limited evidence for the use of this
parameter for defining hsPDA.
Blood flow (ml/kg/min) ¼ (pr2 (cm2)  VTI (cm)  HR (bpm))/ Left pulmonary artery diastolic velocity: the presence of and
body weight (kg) exclusively left-to-right shunt through an hsPDA is likely to cause
where r is the radius of the aortic annulus, to calculate blood flow. forward flow in diastole in the left pulmonary artery. This can be
A calculated output of greater than 300e400 ml/kg/min has been quantified using pulsed-wave Doppler in high parasternal short
associated with the presence of hsPDA. This approach has been axis view and measuring the mean and end-diastolic velocity. A
shown to have errors of up to 28% in comparison to gold suggested end-diastolic flow velocity cut-off value for a large
standard measurement techniques, as small errors in recording shunt is 0.5 m/s or higher. Though there is some evidence for its
Doppler profiles, measurement of VTI and aortic annulus radius use in the literature, its accuracy has been calibrated against
can lead to clinically significant errors in the final value. LVO/SVC flow ratio, which carries its own limitations, as
Calculating a ratio of left ventricular output to superior vena described earlier.
cava flow has not been found to provide additional benefit to the
assessment of hsPDA. Left side pressure loading: left-sided pressure loading can be
investigated by interrogating the flow of blood from the left
Left heart volume loading: Left atrium to aortic root ratio atrium to left ventricle, by taking an apical 4-chamber view, and
(LA:Ao): left atrium to aortic root ratio is a popular parameter, measuring a pulsed-wave Doppler just below the mitral valve
and has been shown to be higher in babies with hsPDA. It can be annulus. The recorded Doppler profile will delineate the flow of
measured in the parasternal long axis view, using M-mode, with blood into the left ventricle during diastole due to passive filling
the cursor placed perpendicular to the aorta at the level of the of the ventricle (E wave) and active filling caused by atrial
aortic valve. The most commonly applied cut-off value for contraction (A wave) (Figure 3).
determining an hsPDA is 1.5. The sensitivity and specificity of Mitral valve E/A ratio: mitral valve E/A ratio gives an
LA:Ao is variable across the literature at 29e89% and 29e91% objective measure of left ventricular filling by early passive filling
respectively, with relatively poor repeatability between versus active filling. Preterm infants have relatively impaired
observers. diastolic ventricular relaxation, meaning the E/A ratio is usually

PAEDIATRICS AND CHILD HEALTH 30:4 132 Ó 2020 Published by Elsevier Ltd.
SYMPOSIUM: NEONATOLOGY

Figure 3 Doppler flow profiles of mitral inflow in diastole. In this


example, A wave is larger than E wave (E/A ratio <1). IVRT: Isovolumic
relaxation time.

<1 (E wave smaller than A wave). In the presence of the sig-


nificant left-to-right shunt through an hsPDA, atrial pressure will
increase, meaning that the proportion of early diastolic filling will
be increased in comparison to normal. This may lead to the E/A
ratio becoming >1 (E wave larger than A wave), representing an
hsPDA. Care should be taken when interpreting this in the
presence of a left-to-right shunt through a PFO, which may act to
offload pressure in the left atrium, and therefore give a falsely
low E/A ratio.
Isovolumic relaxation time (IVRT): following closure of the
aortic valve, there is a short period of time before the mitral
valve opens, during which the left ventricle relaxes. This is the
IVRT. In the presence of a significant atrial pressure load, atrial
pressure will exceed to that in the left ventricle, and force the
mitral valve open sooner, thereby reducing the IVRT. This can
be measured using the same Doppler profile as that used to Figure 4 Doppler profiles of flow in abdominal aorta or coeliac axis.
measure the E/A ratio. An IVRT of less than 30e40ms has been
associated with hsPDA. Care should be taken to ensure that the
Doppler profile is as clear as possible to ensure accuracy in Biochemical assessment of hsPDA
measuring the IVRT. In addition to echocardiographic assessments, blood markers
have also been used to assess the significance of the PDA. Brain
Systemic hypoperfusion
Natriuretic Peptide (BNP) has been used as a marker of left atrial
In addition to echocardiographic assessment for systemic hypo-
stretch. High BNP levels have been shown to be associated with
perfusion, one should consider assessing end-organ perfusion.
hsPDA, and have been shown to fall in the presence of the
This includes assessing urine output, GI tract function, and
closing or closed PDA.
monitoring lactate levels.
Although not a direct assessment for hsPDA, one should
Echocardiographic assessment can be performed by check-
consider performing basic renal function, full blood count and
ing flow in the abdominal aorta, coeliac axis or in the middle
liver function tests when considering medical treatment such as
cerebral artery. Normally, there is forward flow in diastole in
ibuprofen or paracetamol. Thrombocytopenia may reduce the
post-ductal arteries. In the presence of hsPDA, diastolic flow
chance of successful closure of the duct. One should also ensure
may be absent or reversed, as illustrated in Figure 4. Recording
that renal function is adequate before and during a course of non-
a pulsed-wave Doppler profile in any of these arteries will give
steroidal anti-inflammatory therapy, and that liver function is
an objective assessment of flow direction. Care should be taken
monitored if using paracetamol.
to ensure that the angle of insonation is as low as possible
(ideally less than 20 ) to minimise errors in the Doppler profile Combined echocardiographic parameters
recorded. One should ensure normal aortic valve function, as it As demonstrated in Table 1 and in the text above, each indi-
is also possible to see reversed aortic flow in the presence of vidual parameter for the assessment of hsPDA has its own
severe aortic regurgitation. Interpretation of Doppler profile in strengths and limitations. We would therefore recommend the
the middle cerebral artery may be more difficult in the presence use of several parameters in combination in order to build a
of a significant intraventricular haemorrhage in a preterm clearer picture of the haemodynamic effect of the PDA. A scoring
infant. system has suggested in the literature, and includes clinical

PAEDIATRICS AND CHILD HEALTH 30:4 133 Ó 2020 Published by Elsevier Ltd.
SYMPOSIUM: NEONATOLOGY

criteria as well as echocardiographic measurements to determine Jain A, Shah PS. Diagnosis, evaluation, and management of patent
the magnitude of hsPDA. ductus arteriosus in preterm neonates. JAMA Pediatrics 2015; 169:
863e72. https://doi.org/10.1001/jamapediatrics.2015.0987.
Conclusion Laere DV, Overmeire BV, Gupta S. Application of NPE in the assess-
ment of a patent ductus arteriosus, 2018. https://doi.org/10.1038/
When performing echocardiographic assessment of PDA, it is
s41390-018-0077-x.
important to ensure that normal anatomy has been demonstrated
Mcnamara PJ, Sehgal A. Towards rational management of the patent
and the measurements are interpreted with caution in the context
ductus arteriosus: the need for disease staging. Arch Dis Child
of additional shunts and cardiac lesions.
Fetal Neonatal 2007; 92: 424e7.
There is ongoing debate about the benefits of treating hsPDA,
Schwarz CE, Preusche A, Baden W, Poets CF, Franz AR. ‘Repeat-
and indeed how to define hsPDA. We have discussed a number
ability of echocardiographic parameters to evaluate the hemody-
of clinical, echocardiographic and biochemical parameters for the
namic relevance of patent ductus arteriosus in preterm infants: a
assessing hsPDA in preterm infants. There is no parameter that
prospective observational study’, BMC Pediatrics. BMC Pediatr
singularly allows us to clearly define the significance of a PDA.
2016; 16: 1e5. https://doi.org/10.1186/s12887-016-0552-7.
We therefore recommend the use of a combination of parameters
Shepherd JL, Noori S. What is a hemodynamically significant PDA in
to assess the duct characteristics, pulmonary overcirculation and
preterm infants? Congenit Heart Dis 2019; 14: 21e6. https://doi.
systemic hypoperfusion. A org/10.1111/chd.12727.
Skelton R, Evans N, Smythe J. A blinded comparison of clinical and
FURTHER READING echocardiographic evaluation of the preterm infant for patent
Alagarsamy S, Chhabra M, Gudavalli M, Nadroo AM, Sutija VG, ductus arteriosus. J Paediatr Child Health 1994; 30: 406e11.
Yugrakh D. Comparison of clinical criteria with echocardiographic https://doi.org/10.1111/j.1440-1754.1994.tb00689.x.
findings in diagnosing PDA in preterm infants. J Perinat Med 2005; Urquhart D, Nicholl R. Detection of patent ductus arteriosus in the
33: 161e4. https://doi.org/10.1515/JPM.2005.030. preterm neonate. Arch Dis Child 2003; 88: 85e6.
Benitz WE. Patent ductus arteriosus in preterm infants. Pediatrics
2016; 137. https://doi.org/10.1542/peds.2015-3730.
Bin-Nun A, Kasirer Y, Mimouni F, Schorrs I, Fink D, Hammerman C.
Practice points
Wide pulse pressure is not associated with patent ductus arterio-
sus in the first week of life. Am J Perinatol, 2019; 16e9. https://doi.
C Clinical assessment alone is insufficient to determine the
org/10.1055/s-0038-1677475.
presence and haemodynamic significance of a PDA
Chakkarapani AA, Gupta S. Preterm patent ductus arteriosus: what the
C Echocardiography should be used to assess duct and shunt
research tells us. Paediatrics and Child Health (United Kingdom),
characteristics, evidence of pulmonary overcirculation and
2019; https://doi.org/10.1016/j.paed.2019.03.002.
systemic hypoperfusion.
Fink D, El-Khuffash A, McNamara PJ, Nitzan I, Hammerman C. Tale of
C Clinicians should not depend on one parameter alone to
two patent ductus arteriosus severity scores: similarities and dif-
define hsPDA.
ferences. Am J Perinatol 2018; 35: 55e8. https://doi.org/10.1055/
C Normal cardiac anatomy should be demonstrated before
s-0037-1605576.
interpreting findings with the published cut-off values for
Groves AM, Singh Y, Dempsey E. Introduction to neonatologist-
hsPDA
performed echocardiography. Pediatr Res 2018; 84. https://doi.
C Echocardiographic findings may be significantly affected by
org/10.1038/s41390-018-0076-y.
coexisting shunts at atrial or ventricular level.
Hamrick SE, Hansmann G. Patent ductus arteriosus of the preterm
infant. Pediatrics 2010; 125: 1020e30. https://doi.org/10.1542/
peds.2009-3506.

PAEDIATRICS AND CHILD HEALTH 30:4 134 Ó 2020 Published by Elsevier Ltd.

You might also like