Professional Documents
Culture Documents
XK3190 YHT7 Manual
XK3190 YHT7 Manual
Weighing Indicator
User Manual
V1.01
Dear users:
-1-
K7+E
7 Operating environment
Power supply AC220V; 50, 60Hz (-2%~+2%)
Operating temperature 0℃~40℃
Storage and transport temperature -25℃~ 55℃
Relative humidity ≤85%RH
Fuse 500mA
8 Weight Approx. 2 kg
-2-
K7+E
K7+E
RS232
Power switch AC scoreboard Load cell
-3-
K7+E
II. Load cell Connection
1. The Load cell is connected through 9-pin plug socket (hole). Figure 2-3 shows the meaning of each
pin.
2. Please use 4-core shielded cable.
▲!The connections of Loadcell and indicator must be reliable, and the shielded cable of
Loadcell must be reliably grounded. Connections wires shall not be plugged and pulled when
the indicator is in a Power-up State in order to prevent static electricity damaging to the
indicator or load cell.
▲!Since both Load cell and indicator are static-sensitive device, anti-static measures must be
practically taken in the use, and welding or other strong-electric operations on weighing
platform are strictly prohibited. In the thunderstorm season, reliable lightning protection
measures must be taken to prevent lightning damaging the senor and instrument and to ensure
the operator safety and the safe operation of weighing equipment and related equipment.
-4-
K7+E
1. Interface parameter
① Signal: RS232C
① Baud rate: /1200/2400/4800/9600
2. Communication method
Method 1 P5=2 continuous transmission
Transmitted data is weight.( gross weight, net weight or tare weight, decided by parameter P4 ).
Gross weight format is : ww000.000kg or ww000.000lb
Net weight format is: wn000.000kg or wn000.000lb
Tare weight format is: wt000.000kg or wt000.000lb
Note: Above decimal point position is determined by decimal point position setting of the indicator .
-5-
K7+E
Note: = data format , ASCII code.
<Weight data (including decimal point)>: 6 digits (including decimal point) of weight data with
symbol, ASCII code.
The lower digits are in front, and the higher digits and symbol digits are in the rear. Negative symbol
bit is "-", and positive symbol is "0".
For example, if the weight showing on the indicator is -500.00kg, the serial output data will be "=
00.005-".
If the weight showing on the indicator is 500.00kg, the serial output data will be "= 00.0050".
Large screen
See scheme 2-4 for wiring details. Scoreboard signal is current loop signal of 20mA constant current and
is output in a serial way through binary code, with a baud rate of 600. Each frame has 11 data bits, including
one start bit (0), eight data bits (LSB in front), one flag bit, and one stop bit (1).
In the calibration and parameter setting status, some keys will perform the following functions:
1.ZERO key performs the "plus 1" function. After the ZERO key is pressed, the number displayed in the
last bit will be "plus 1", automatic zero will be made after it is added to 9.
2.TARE key performs the "shift" function. After the TARE key is pressed, the number shown in the last bit
will move to the right for one bit, and will move to the highest bit automatically after press tare key
when it is in the last bit at the right .
3."# " key performs “input” function, after the # key is pressed the date will be input to indicator which
was set.
4.CLEAR key performs "exit" function, after Clear key is pressed the indicator will exit the calibration state
or setting state.
5."*" key performs" switching parameter" function, after the * key is pressed the indicator will come into
the next parameter setting state .
-6-
K7+E
When the displayed weight in weighing status is positive and the STABLE light is on, press the
TARE key to deduct the indicated weight as tare weight . In this case, the indicator will show a net weight
of "0" and the NET WEIGHT light is on. Press the TARE key again when gross weight is 0, the indicator
will clear the tare weight value.
【Note】Tare is not workable when gross weight is negative number.
1. Inner code checking:
Press[#] and [*] key together in the normal weighing status , the indicator will display the inner
code.
2. HOLD, average value:
You can choose this function by function setting P2. You can start or stop this function by [#]key
after your choose this function.
Choose 2 for P2 parameter, press [#] key, average value light is on, now the indicator can estimate
automatically ,and after stabilization the indicator will calculate and display average value, and will
display the average value after lighting, and also will cancel display average value after current value
back to zero, now can weigh the next average value ;
Choose 3 for P2 parameter, press [#] key in normal weighing status and peak value light is on, and
indicator now in the auto peak value working status, and it can keep the maximum weighing data, and
will cancel the data hold after the weight back to zero.
In above operation steps, we can cancel the average or hold function by press[#]key anytime;and
will start this function after press[#] key every time when startup .
3. Manual accumulation of measured value:
Press the[*] key when the measured value is larger or equal to 20 divisions and the data is stable in
the normal weighing state, the indicator will perform "manual accumulation" function. In this case, the
indicator will display the total accumulation data (in two steps): [total = ] (indicating that the content
shown below is the amount of accumulation data) will show the accumulation data [******] in about 1
second. The times of accumulation will be then indicated (in two steps): [n = ] (indicating that the times
is shown below) will show the times of accumulation [ ***] in about 1 second. The
ACCUMULATION indicating light is then on.
Note: the maximum times of accumulation is 9999 (when accumulation result is ensured to be
≤999999); The accumulation results will be held before they are cleared and the data will not be
lost after power-off. When the [*] key is pressed, if the net weight is less than 20 divisions, only the
accumulation value will be displayed, and the accumulation of weight and times will not be made.
After lead sealing is broken, please insert short connecting ring to connect pin J7 on the indicator main
board, and allows the indicator to be calibrated. (circuit board is setting as default that allows
calibration, new indicator can omit this step),then operate as following steps(after calibration, lead
sealing again):
(1) Press【#】key during startup initialization, the indicator will come into calibration state.
1、 Division setting:
Display 【d X 】 press【tare】key to choose 1、2、5、10、20、50,press【#】key for confirmation,
and will enter into next parameter setting, press【tare】 key for automatic
-8-
K7+E
step-by -step cycle display.
Display 【d X 】
Display 【d 1 】
Display 【d 2 】
Display 【d 5 】
Display 【d 10 】
Display 【d 20 】
Display 【d 50 】
Display 【d 1 】
For example, press【#】key when it shows【d 5】,now the division is setting to 5, and will come into
decimal point setting state automatically.
2、 decimal point setting :
Display 【P X】 press【tare】key to choose0、1、2、3,press【#】for confirmation, and will come
into next parameter setting automatically. press【tare】key for automatic step-by-step cycle display.
Display 【P 0】
Display 【P 0.0】
Display 【P 0.00】
Display 【P 0.000】
Display 【P 0】
For example, press【#】key when it shows【P 0.000】,now the decimal point setting is 0.000,and will
come into full capacity setting state automatically.
3、 Full capacity setting:
Display 【FULL 】 press【tare】key to come into the number input state.
Display 【0 0 0 0 0 0】 press【tare】key,indicate symbol▼will move to right one step by step to
which input position you want, and press【zero】to add value to adjust the
number you need, press【tare】to adjust all the numbers you need until input
the full capacity you need, and press【#】for confirmation and indicator will
come into next parameter setting.
For example, when it displays 【0 2 5 0 0 0】 press【#】key for confirmation and will come into zero point
setting calibration state.
4、 Zero calibration:
Display 【nOLOAD】 Insure there’s no loads on the platform and wait until the stable indicate
symbol▼display, then press【#】key,zero calibration finished, and indicator
come into full capacity calibration state.
5、 Full capacity calibration:
Display 【AdLOAD】 Put weights on the platform and press【tare】key to come into input state
after stabilization.
Display 【0 0 0 0 0 0 】 Press【tare】key,the indicate symbol▼move to right and to the position which
can select number input, press【zero】key to add number and until input the
number you need and then press【tare】key , the indicate symbol▼move to
right and to the position which you can select number input ,press【zero】
key to add number until get the number as same as the weight value of the
weight. press【#】key for confirmation and finish the full capacity
calibration state.
Display 【 End】
6、 Press【clear】key and back to weighing state and new parameters take effects. At this time you can also
keep pressing【#】key to set other parameters.
(2)Press【#】key during startup initialization, the indicator will come into calibration state.
-9-
K7+E
Fast Zero calibration:
Press【Function】key in any time before display 【nOLOAD】,the indicator will save the parameter
of division, decimal point, full capacity,and come into the zero calibration state directly. Press【zero】
key when the stabilization symbol ▼ appear, it displays【 End】,it means that the indicator save the full
capacity calibration parameter before and press【clear】key, the indicator save the parameters and back to
weighing state.
- 11 -
K7+E
[Err 3] Zero position is too high or there is heavy goods on platform when startup.
[Err 7] The calibration short-circuit ring did not connected
[Err 8] Loadcell signal line connect in reverse,please connect Loadcell line in right way .
[------] Out of display range , display value should between -99999~999999
[A oL] Out of the maximum times of accumulation or weigh of accumulation,
now the accumulation is not workable , please do the operation after clearance.
2.It is not suitable to place the indicator in the dusty and vibration environment and also avoid
using in the moist environment.
3.Signal source and weighing indicator should be reliably connected, and system should be well
grounding. It should be far away with strong electric field, strong magnetic field. Load cell and
indicator should be far away with strong corrosive, inflammable, explosive object.
▲ Do not use under inflammable gas or inflammable steam; do not use under pressure
container canning system.
▲ Lightning frequent areas, it must be installed reliable lightning arrester, to ensure
operators safety and prevent damage of instruments and relevant equipment due to
lightning.
▲ Signal source and indicator are static sensitive equipments, when using must
earnestly adopt anti-static measure. It is prohibited in measuring device for welding
operation or other strong electric field operation; in thunderstorms season, we must
implement the reliable lightning-protection measures to prevent signal source and the
indicator damage caused by lightning strike, ensure operators safety of weighing
equipment and related equipment safety operation.
4. Never use strong solvents (e.g., benzene, nitro class oil) to clean the housing.
5. Liquid or electricity conducting particles should not be poured into the indicator, in case the
indicator damage and electric shock.
6. You should cut off power supply of indicator and relevant device before you pull-in and out
the connecting cable of indicator and external device.
You should cut off power supply before pull-in and out connecting cable of signal source.
7. Company advice for customers: start to use our indicator after test. The company is only
responsible for the quality of indicator, the biggest compensation is not more than twice of
- 12 -
K7+E
indicator value, the company is not responsible failure of the whole system.
8. Output interfaces of indicator must be strictly in accordance to the user's manual, you should
not alter any connection. If there is failure when using the indicator, you should
immediately unplug it, and send to manufacturers for repair. Ordinary non-professional
manufacturers should not repair it to avoid bigger damage.
9. Since invoice date, the indicator has a one-year free repair period. If any non-artificially
failure happens under regular using conditions within this period, the user can send the indicator
with its guarantee card (of the correct number) back to our service station or supplier for repair.
The indicator shouldn’t be open without authorization, otherwise free guarantee will be cancelled
10. Battery
① The battery will be charged after power cord is connected to AC 220V power supply. So
please remove the battery if it is not used frequently.
In order to avoid internal over-heat and battery over-charging, the current will be limited. If
you feel that charging is too slow, please buy a special charger for external charging. Please note
that the wire ends connected to battery shall not be connected inversely (red +, black -), or the
indicator may burn out.
Be sure to fully charge the battery before using the built-in battery for the first time!
② When the indicator is battery-powered, it will automatically switch to AC supply mode
once the AC power supply is connected; in this case, the AC indicating lamp will light on. There
are battery level indications at the lower left corner of indicator screen. The indicator will
automatically cut off the DC power supply after the last indicating lamp goes out. Please charge
the battery immediately in this situation. The current battery voltage can be displayed when the
indicator is started. Please pay attention to this data from time to time.
③ Please charge the battery for about 10-16 hours before its first use so as to avoid a too low voltage
caused by self-discharging of battery which may be mistakenly taken as failure.
④ The battery shall be charged for about 10-16 hours at a time during the normal operation. If the
indicator is not to be used for long, the battery shall be charged for 10-16 hours every two months in order to
extend its service life.
⑤ Four grade battery voltage display, and battery voltage are :6.19V、5.99V、5.77V、5.55V。The
indicator will turn off when the last light is off.
The built-in battery of indicator is a consumable part that is not within the range of "three
guarantees".
- 13 -
K7+E
- 14 -