Flowmeter Calibration Certificate Certificate No. HTM-065-10

Customer : Flowhire Ltd Date: : 16/01/2024 0.1930


Meter : TSI Flowmaster 250DL Job Ref. : 240116B 0.1352
Size : 65mm -0.2849
Serial No : 3579 Cal. Fluid : Water -0.4442
Flow Range : 500 to 2000 L/min Fluid Temp. : 13.3 to 15.8 degC 0.0000
Accuracy : +/- 2.0% Calibration Ref. : FQM001
Tag No. : HTM-065-10

Final (5 Points / 3 Repeats Lin)


Run Nr. Flow Rate Meter under test Reference meter Deviation Flow %
L/min Lts Lts Error

1a 2000 6144 6134.30 9.70 0.16 0.1930


1b 2000 6194 6183.43 10.57 0.17
1c 2000 6087 6071.83 15.17 0.25 -0.2849
2a 1500 4448 4441.13 6.87 0.15
2b 1500 4529 4524.50 4.50 0.10
2c 1500 4547 4540.13 6.87 0.15
3a 1000 2955 2961.53 -6.53 -0.22
3b 1000 2948 2958.57 -10.57 -0.36 -0.2849
3c 1000 3001 3009.33 -8.33 -0.28
4a 500 1496 1503.33 -7.33 -0.49
4b 500 1511 1517.83 -6.83 -0.45
4c 500 1589 1595.30 -6.30 -0.39

2.000

1.000 % Error

0.000
4 3 2 1

-1.000

-2.000
Calibration Traceable to National Standards
Average Run Number

Calibration Traceable to UKAS Standards

Sales Director Calibration Engineer