7E-A Portable Phlegm Suction Unit

Service Manu al

P l e a s e r e a d t h e User , s m a n u a l c a r e f u l l y b e f o r e u s i n g t h e
product!
Ref. No.: YY-JS.YD-2042
Contents of Service m a n u al

Overview of safety………………………………………………2-4

Characteristic of product…………………………………………5

Handling…………………………………………..……………5-6

Debugging…..………………………………………………7-9

Use and maintenance…………………………………………10-12

Other precautions……………………………………………13-14

1
Ref. No.: YY-JS.YD-2042
Overview of safety
1. Warning:
This product was precisely manufactured, assembled and grounded. Do not try to uninstall or repair
the equipment. All the repair shall be performed by authorized repair center and by qualified personnel.
2. Key safety measures
When use this electric product, child shall be especially taken care, and the following safety
protection measures shall be complied with:
◆ Danger - Reduce the risk of electric shock
1) Shut down the power on the end of each use;
2) Do not fetch the equipment falling in water, shut down the power firstly;
3) Do not place or store the equipment in the location easily affected by water drop or other liquid.
4) Do not contact with equipment in humidity;
5) Do not uninstall the equipment. Consult the qualified service personnel for service.
6) Periodic electric safety checks
◆ Warning - Reduce the risk of burn, electric shock, fire and body hurt.
1) When the power is on, leaving of personnel is not allowed;
2) When the children or personal is close to the equipment, they should be monitored.
3) In this manual, the instructions were described. Do not use the accessories beyond the
recommendation of manufacturer; otherwise the performance of equipment will be lowered.
4) In cases of the following circumstance, do not use the equipment, return it to service center for check
or repair: damage of power line or plug, abnormal running of equipment, falling and destroy of the
equipment, equipment falling in water, etc.
5) Power line shall keep away from heat or heating surface.
6) Do not block the air intake of the product; soft cloth and articles such as floss shall be avoided on
intake.
7) Electric and magnetic environment guidance in use

2
Ref. No.: YY-JS.YD-2042

Guidance and declaration of manufacturer -Electromagnetic emission


The 7E-A Portable Phlegm Suction Unit is intended for use in an environment specified below. The
customer of the user of 7E-A Portable Phlegm Suction Unit should assure that the unit is userd in such
an environment.
Emission test Compliance Electromagnetic environment- regulations

RF emissions Group 1 The 7E-A Portable Phlegm Suction Unit uses RF energy
CISPR 11 solely for its internal function. Therefore, its RF emission is very
low and is not likely to cause any interference in nearby electronic
equipment.
RF emissions Class B The 7E-A Portable Phlegm Suction Unit is suitable for use in
CISPR 11 all establishments, including domestic and those directly
connected to the public low-voltage power supply network that
supplies building used for domestic purposes.
Emission of Class A
harmonics IEC
61000-3-2

Voltage fluctuations / Complies


flicker emissions IEC
61000-3-3

3
Ref. No.: YY-JS.YD-2042

Guidance and manufacture’s declaration-electromagneti c immunity

The 7E-A Portable Phlegm Suction Unit is intended for use in the electromagnetic
environment specified below.The customer of the user of the 7E-A Portable Phlegm
Suction Unit should as surethat it is used in such and environment.

Electrom agnetic
Immunity test IEC60601 test level Compliance level
environment-guidance
Floorsshould be wood,concrete
Electrostaic orceramic tile. If floors are
±6KV contact
discharge(ESD) covered with synthetic
IEC 61000-4-2 ±8KV air material.the relative humidity
should be at least 30%
Electrostaic fast ±2KV for power supply Mains power quality should be
transient/burst lines that of a typical commercial
IEC 61000-4-4 ±1KV for input/out lines or hospital environment.
±1KV lins to lins Mains power quality should be
Surge
±2KV lins to earth that of a typical commercial or
IEC 61000-4-5
hospital environment.
<5%UT
(>95%dip in UT)
for 0.5 cycle
Mains power quality should
40%UT be that of a typical commercial
Voltage dips, short
(60% dip in UT) or hospital environment.if the
interruptions and
voltage variations on for 5 cycle user of the(7E-A)requires
continued operation during
power supply input
lines 70%UT power mains,it is
(30% dip in UT) recommended that the(7E-A)be
GB/T 17626.11(IEC
for 25 cycle powered from an
61000-4-11) uninterruptible power supply
<5%UT or a battery。
(>95%dip in UT)
for 5 s

Power frequency magnetic


Power frequency fields should be at levels
(50/60Hz)magnetic characteristic of a typical
3 A/m 3 A/m
field location in a typical
(IEC 61000-4-8) commercial or hospital
environment.

NOTE:UT is the a.c.mains voltage prior to applicayion of the test level。

4
Ref. No.: YY-JS.YD-2042
Characteristics of the product
Sketch map of 7E-A:
Fluid storage bottle

Filter
Machine box

Handle

Negative pressure adjustment


Power cable orientation indicator

Negative adjustment valve

Negative pressuregauge Power indicator lamp Power switch

Sketch map of 7E-A whole machine

Fuse
F1.5AL250V,Φ5×20

Base

Sketch map of 7E-A whole machine (underside)

HANDLING
1. Overview
Type 7E Portable Phlegm Suction Unit is a new generation non-lubricant negative pressure pump
suction apparatus developed and manufactured from similar products. It is applicable for purulence, blood,
sputum and viscosity fluid suction in operation of medical unit and other situations applicable (it is not
applicable for abortion suction).
2. Structure characters and operational principle
► Uses non-lubricant pump, can prevent oil pollution of environment.

► Low noise

5
Ref. No.: YY-JS.YD-2042

► All-plastic panel, with good water-erosion-proofing performance.


► Fluid storage bottle is designed as embedded structure, cubic shape negative pressure
gauge, all-plastic shell design.
► This equipment will not produce positive pressure in running, which ensured the
reliability and safety in use.
► The negative pressure adjustment system can adjust the pressure steplessly according to requirement.
► The volume is small, weight is light, taking is convenient, especially suitable for cruise
diagnostic need of various circumstances.
► Operational principle: Negative
vacuometer 负压 调 节 阀 adjustment valve
真空表

vent
排 气口 空气 过滤器
filter

真空泵 至To
吸 suction
引 管 tube

吸气口
inspiration 溢流阀
消 muffle
音器 Overfall 贮液瓶
valve Fluid storage bottle

Vacuum pump
3. Main technique indicators
1) Power: AC230V±10%(253V~207V), 50Hz±1 Hz;
2) Input power: ≤90VA
3) Minimum negative pressure: ≥0.075MPa;
4) Negative pressure adjustment scope: 0.02MPa~Minimum negative pressure
5) Evacuation rate: ≥15L/min;
6) Fuse tube: F1.5AL250V,φ5×20
7) Fluid storage bottle: 1000mL, 1 piece
8) Noise: ≤60 dB(A)
9) Weight: 4.4kg
10) Size: 280×196×285 (mm)
► This equipment is not applicable for use in situation with flammable or explosive gas.
► Regulation for operation: short-term running
► Electric safety requirements: belongs to Class II Type B equipment.
► This equipment belongs to “High vacuum/low flow velocity” electric suction device.
4. Normal operation conditions

Environment temperature: 5℃~35℃ Relative humidity: 30%~80%


Atmospheric pressure scope: 86kPa~106kPa
Note: when the storage temperature is lower than 5℃, the equipment shall be laid in normal operation temperature
environment for at least 4 hours.

6
Ref. No.: YY-JS.YD-2042
Debugging
1. Unload from carton
The user shall check the appearance, accessories types, amount and consistentce with instruction
manual of product before install and debut the product. If there is missing, please contact the
supplier and manufacture in time.

2. Connect the tube (refer to tube connection sketch map, not connect the suction tube
temporarilly)

Note: before installation, paste a few distilled 2 5


water on bottle inputting part of bottle cover,
which is benefit to enhance pressing and
4 To suction
至吸痰管
air-proofing.
3 tube

3, Connect to power supply 2


Connect the power plug to power supply, power indicator 1
lamp lighting means power supply is connected.
1.缓冲瓶
1. Buffering 2.吸引软管 3.空气过滤器
bottle 2.suction soft tube
Note: Power cutting device of this product is power 3. air filter 5.吸引软导管
4.blue mark
4.蓝色标记
plug, power receptacle shall be reliably grounded.
5. suction soft catheter

4. Verify the tube


► Screw on the negative pressure adjustable valve by clockwise, block the intake orifice with finger or

rubber head of burette, or fold and pinch the soft suction tube.
► Turn on the suction apparatus, run the equipment, there shall be allophone. The index finger on

vacuum meter will ascend to minimum negative pressure value quickly; unloose the intake orifice, the
index finger will go back to less than 0.02MPa, which shows the tube is correctly connected.
► Connect the suction tube, when connect F6 suction tube, negative pressure value
of negative pressure system shall be less than 0.06MPa; connect F8, negative pressure
value shall be less than 0.04MPa; connect F12 suction tube, negative pressure value
shall be less than 0.03MPa. The above-mentioned circumstances show the suction
apparatus is normal, suction tube is smooth.
Note: IF the suction tube is not smooth, dredge according to following method: fold
the suction soft catheter into V shape (there shall be no fluid in fluid storage
bottle), make the negative pressure value achieve maximum value, recover the
suction soft catheter to original shape quickly, repeat several times to make
the suction tube smooth.

7
Ref. No.: YY-JS.YD-2042
5. Adjust the negative pressure
► Block the intake orifice, turn on the switch of suction apparatus, adjust the negative pressure
adjustable valve, index value of pressure meter shall be in scope of .02MPa~limit negative pressure
value.
► In clinical use, the negative pressure adjustable valve is used for control the necessary

negative pressure value in suction.


► The negative pressure will ascend when the index finger of negative pressure adjustable valve

rotate by clockwise.
► The negative pressure shall be lowered to less than 0.02 MPa before shutting down

the equipment.

6. Check and test the flooding apparatus


► Open the bottle piston, clear the valve orifice, press the rubber valve patch on floater to flat, the
valve shall not with deficiency such as warp, mangle, and the connection with floater shall be good.
The floater shall freely move in floater rack and without blocking.
► Raise the bottle piston by manual to make the floater contact with water surface vertically, then

descend the bottle cover slowly, the floater shall float.


► Cover the bottle cover closely, there is suction soft tube on intake orifice, screw on the adjustable

valve, run the suction apparatus. Drop the suction soft tube in water bucket or simulate the actual use,
suck the fluid into fluid storage bottle with flooding apparatus. The ascend of fluid level will take the
float ascend of floater, until the close of valve orifice, suction will stop automatically. Finally the fluid
level will change along with the suction method.
►Loose the adjustable valve, shut down the suction apparatus, open the bottle piston,
empty the fluid storage bottle, when screw on the piston again, the float shall be locating
on the bottom of floater rack, and the valve orifice shall be on “ON” status.
The flooding apparatus conforming to above-mentioned requirements can run normally and can
be used for clinical application.

Note 1: When shutting off the flooding apparatus, if the fluid level ascend continually, there are 2
potential causes:
(1) caused by residual negative pressure in fluid storage bottle.
(2) valve orifice not completely closed.
For the first case, when the suction soft tube leave the suction fluid and drop in it again, the fluid
level of fluid storage bottle will not ascend; for the second case, the fluid level will ascend still, this
shall be observed carefully, if the fluid storage bottle is near full, the suction soft tube shall be taken
out from suction fluid immediately, shut down the suction apparatus, stop to suck, investigate the
cause of deficiency of valve orifice.
Note 2: After the valve orifice closed by floater, the floater perhaps still sucked on valve orifice
This is caused by negative pressure in tube. Loose or shut off the suction apparatus (loose the
negative pressure in tube). The floater will get away from valve orifice under gravity (pulling off the
floater by manual is forbidden to prevent the separation of rubber valve patch with floater).
3. After shut off the equipment, release the negative pressure before open the bottle piston.
4. Using of suction apparatus in case of the flooding apparatus and guide tube uninstalled is forbidden.

7. Shutting down
Shut down the suction apparatus firstly, pull off the plug from power receptacle to cut off the

8
Ref. No.: YY-JS.YD-2042
power.

8. The symbols and meaning involved in this equipment on safety requirement


Symbol Meaning Symbol Meaning
Caution, consult ACCOMPANYING
~ Alternating current
DOCUMENTS

Protective grounding IPX0 Degrees of protection:0

Class II equipment TYPE B equipment

9
Ref. No.: YY-JS.YD-2042
Use and maintenance
1. Use and maintenance
►Before use, install and debug the procedure to check the suction apparatus to ensure the performance,
connect the disinfected suction soft tube and abdominal cavity suction tube.
Note: For the use of equipment-attached suction tube, please consult instructions in package
before use.
► In use, the negative pressure adjustable valve uses negative pressure necessary, note the fluid level in
fluid storage bottle. if the level of former fluid storage bottle ascend to specified capacity (still
applicable if the integral equipment leans for 10 degrees), shall stop to suck, waiting for empty before
use the backup fluid storage bottle again. The backup fluid storage bottle can prevent the fluid entering
the pump as buffer bottle.
Note: If the flooding apparatus shut off, and the fluid level ascend still, the countermeasures see
“Check and Test the Flooding Apparatus”.
► Emergency measure in use:
(1) If there is pum and phlegm, which glues and blocks the suction tube, turn to loose
the negative pressure adjustment valve quickly, release the negative pressure,
change the suction tube before suct the pum again.
(2) If suction tube is hardly to remove after suction of pum or human body tissue,
turn to loose the negative pressure adjustment valve according to
above-mentioned method.
Note 1: Before suction, fold the suction soft catheter into V shape, turn on the machine
to allow the negative to necessary scope, insert the suction tube into pum
blocking part of sufferer, then recover the suction soft catheter to original
shape to perform suction, the effect will be better.
Note 2: Specification and size of suction tube is selected by medicinal personnel
according to clinical need.
Note 3: The use of suction apparatus shall be under instruction of medicinal personnel,
strictly follow the operation procedure in scope, any question please contact
the supplier or manufacturer.

Turning on and operation time do not over 30 minutes


Suction amount do not over maximum fluid level warning mark
In case of insufficient negative pressure, please turn on the bottle cover and connector head

2. Change of air filter


The air filter suck the bubble or dust, which will cause the cooler of filter membrane change from
gray to dark, and will cause suction force of intake orifice decrease even disappear, while the
negative pressure on pressure meter will ascend continually up to 0.04MPa or more, at that time the
air filter shall be changed in time with air filter manufactured by our company.

10
Ref. No.: YY-JS.YD-2042
Note 1: Shutting down of flooding device, blocking of tube in use will also cause reducing even
disappearance of suction force and increasing of negative pressure value. Refer to
“Stroubleshooting of usual failures”
Note 2: The air filter shall be changed frequently and destroyed centralizedly.

3. Change of fuse tube


The fuse tube is installed on back of base, if necessary to change, shut down the power first. Loose
the index finger with cross screw or “-” shape screw by counterclockwise, then change.

4. Maintenance after use


► Before shut down the equipment, sucking little clean water into suction apparatus recommended to
clean the interior wall of tube.
► After shut down the equipment, empty the fluid storage bottle, clear the dirty on bottle and piston with

soft brush or cloth, then flush with clean water and disinfect. (including flooding valve, valve patch and
floater shall not separate)
► Fluid storage bottle, piston and each tube can be disinfected with Kangweida disinfection piece

(0.5g/piece) with the preparation proportion of 1: 200 concentration, soak for 1 hour.
Fluid storage bottle and the cover can be sterilized 30 times with disinfectant. If the air tightness can not
reach 0.075MPA, replace the cover seals.

Note: Fluid storage bottle is plastic product, do not crash with acute article in washing and use, and
prevent falling.
► The surface of equipment can be cleared with disinfected warm cloth. The fluid shall be prevented to

enter the gap of equipment.


► The equipment not in use shall be located in dry and clear area and shall be operated periodically

(usually once per 6 months).


Note: Check the flooding apparatus, guide tube and other tubes before the next use and the orifice of guide
tube facing on interior wall of fluid storage bottle shall be ensured.

11
Ref. No.: YY-JS.YD-2042

5. Troubleshooting of usual failures

S/N Failure Analysis of Causes Correction Method Remark


1 Minimum negative 1) Leakage of bottle 1) wash the dirty 1) Interior
pressure less than 90kPa orifice orifice of bottle, cover components check
the stopper closely or and repair shall be
2) Leakage of conduit change the stopper performed by
connection professional
2) Stuff up the stopper personnel
3) Loosing of again to each joint
adjustable valve 2) breaking of suction
3) screw on the soft tube shall be
adjustable valve change
2 Minimum negative 1) Turn off of flooding 1) Turn the adjustable 1) empty the air
pressure higher than apparatus valve by storage bottle in
40kPa, but the suction counter-clockwise time
force of conduit orifice 2) Jam of conduit after shutting down
obviously decrease or the equipment, release 2) The terminal
3) Jam of air filter indicated with blue
desapear the negative pressure
in conduit then screw point is the orifice of
on. admission

2) Dredge, wash or
change the soft tube
3) Change with air
filter of our company
3 Indicator light of power 1) Loose of receptacle 1) Repair or change the Performed by
voltage normality not receptacle professional repair
bright 2) Fuse tube fused personnel (refers
2) change the fuse tube electric cricuit
3) Damage of
indicator light 3) change the indicator principle chart)
light
4 Fuse tube fused 1) Over high voltage 1) check the circuit to
exclude the failure
2) Failure of internal
lines 2) adjust or change the
relay
3) Failure of relay
3) check pump body
4) Block of pump and electric machinery
body, increase of
current
Note: I f there i s failu re o f pu mp, the unloading and repa ir shal l be operated b y
special ty personn el , please contac t supplier or manufacturer if necessary (Please shut
down the power before checking the circuitry or unloading the case).
Note: Please check appearance condition of the power cable insulation skin,elasticity of the plug and
receptacle, on-off condition of the power switch, fastening condition of the electric components before the
next use. Make sure the periodic electric safety. If any questions, please contact the supplier or
manufacturer

12
Ref. No.: YY-JS.YD-2042
Other precautions
1. Limit conditions in transportation and storage
1. Temperature of environment: -40℃~+55℃ 2. Relative humidity: 10%~93%
3. Atmospheric pressure: 70kPa~106kPa
Note: The suction apparatus shall be stored in house without erosive gas and with good aeration.
The violent shake in transportation shall be avoided.
2. The electric principle chart (Electric repair shall be conducted by specialty
personnel)
FU1 s

~220 V HL M

c

3. Post-sales service
► If there is non-man-made quality issue in one week after distribution, it is our company to response
for returning, change and reapair. In normal use and normal maintenance circumstances, if the
Electric Suction Apparatus can not be normally used, if there is quality issue in one and half a year
since deliveyr (or used for 1 year), the company is responsible for repair for free or change for free;if
there is quality issue after one and half a year, user can contact the post-sales service department,
office or agents for repair for free based on invoice and repair guarantee card;If it can not be
normally used after half a year, the company can provide accessories and repair with reasonable
charge.
► The following circumstances are not included in repair guarantee scope: ①
Whole machine damage or distortion caused by crash; ② water or rain entering
the whole machine; ③ Water, blood, pum or glue fluid entering pump and caused
pump not normally work by user; ④ Articles easily to damage or easily to consume:
air filter, suction tube, fuse.
► If necessary, the power supply circuit and materials necessary for repair can be provided by our
company. If there is question on electric circuit repair, please contact with the manufacturer.
4. List of accessories
 suction soft catheter ( Φ7×Φ12, length 2m) 1 pc
Danyang Wanbeng Medical Apparatus Co.,Ltd.
 suction tube ( F8, F12) each 1 copy
Suzhou Weikang Medical Apparatus Co.,Ltd.
 Fuse tube ( F1.5AL250V,Φ5×20) 2 pcs
Chanzhou China-Lay Electric CO.,LTD
 Air filter 2 pcs
Yangzhou bailifeng Electric Appliances CO.,LTD
Instruction Manual , Repair Guarantee Card , Conformity Certificate each 1 copy
Danyang dianxing presswork CO.,LTD

13
Ref. No.: YY-JS.YD-2042
5. Treatment of waste
The treatment of waste shall be conforming to law and regulations.
The change caused by improvement of equipment and electric circuit or appearance
will not be informed additionally.

14