Professional Documents
Culture Documents
Reagent (fill)
often cause these interactions and have been known to be
problems for all kinds of immunoassays (Boscato, LM, Stuart, and elevated concentrations were assayed. The total number of C) 1 plate 2 plates
3.000 such specimens was 241. The least square regression equation
MC. “Heterophilic antibodies: a problem for all immunoassays” D) 1 (20ml) 1 (20ml)
Clin. Chem. 1988: 3427-33). For diagnostic purposes, the and the correlation coefficient were computed for the PSA E) 1 (7ml) 2 (7ml)
2.500
results from this assay should be used in combination with AccuBind® ELISA test method in comparison with the reference
F) 1 (7ml) 2 (7ml)
clinical examination, patient history and all other clinical method. The data obtained is displayed in Table 4.
Absorbance(s)
2.000
G) 1 (8ml) 2 (8ml)
findings.
1.500 TABLE 4
4. For valid test results, adequate controls and other parameters
must be within the listed ranges and assay requirements. Least Square Correlation
1.000 Method Mean Regression Analysis Coefficient
Patient 5. If test kits are altered, such as by mixing parts of different kits,
which could produce false test results, or if results are This Method (X) 5.62 y = -0.0598+0.98(X) 0.987
0.500
incorrectly interpreted, Monobind shall have no liability. Reference (Y) 5.57
0.000 6. If computer controlled data reduction is used to interpret the
0 20 40 60 80 100 120 Only slight amounts of bias between the PSA AccuBind® ELISA
results of the test, it is imperative that the predicted values for
test system and the reference method are indicated by the
the calibrators fall within 10% of the assigned concentrations.
PSA Values in ng/ml closeness of the mean values. The least square regression
7. PSA is elevated in benign prostate hypertrophy (BPH).
Clinically, an elevated PSA value alone is not of diagnostic equation and correlation coefficient indicates excellent method
11.0 Q.C. PARAMETERS value as a specific test for cancer and should only be used agreement.
in conjunction with other clinical manifestations (observations) 14.4 Specificity:
In order for the assay results to be considered valid the and diagnostic procedures (prostate biopsy). Free PSA
following criteria should be met: No interference was detected with the performance of PSA
determinations may be helpful in regard to the discrimination of AccuBind® ELISA test system upon addition of massive amounts
1. The absorbance (OD) of calibrator F should be > 1.3. BPH and prostate cancer conditions.5 of the following substances to a human serum pool.
2. Four out of six quality control pools should be within the 8. Due to the variation in the calibration used in PSA/ fPSA test
established ranges. kits and differences in epitopic recognition of different Substance Concentration
antibodies, it is always suggested that the patient sample
Acetylsalicylic Acid 100 µg/ml
12.0 RISK ANALYSIS should be tested with PSA/ fPSA tests made by the same
Ascorbic Acid 100 µg/ml
manufacturer. (Monobind Inc. offers a fPSA ELISA test that
Caffeine 100 µg/ml
The MSDS and Risk Analysis Form for this product are available should be used for consistency reasons, when needed.)
CEA 10 µg/ml
on request from Monobind Inc.
AFP 10 µg/ml
13.0 PERFORMANCE CHARACTERISTICS CA-125 10,000 U/ml
12.1 Assay Performance
1. It is important that the time of reaction in each well is held hCG 1000 IU/ml
Healthy males are expected to have values below 4 ng/ml.4 hLH 10 IU/ml
constant to achieve reproducible results.
2. Pipetting of samples should not extend beyond ten (10) hTSH 100 mIU/ml
TABLE I
minutes to avoid assay drift. hPRL 100 µg/ml
Expected Values for PSA AccuBind® ELISA Test System
3. Highly lipemic, hemolyzed or grossly contaminated Healthy Males <4 ng/ml
specimen(s) should not be used. 15.0 REFERENCES
4. If more than one (1) plate is used, it is recommended to repeat It is important to keep in mind that establishment of a range of
the dose response curve. values, which can be expected to be found by a given method for 1. Christensson A, Laurell CB, Lilja H, Eur J Biochem, 194, 755-
5. The addition of substrate solution initiates a kinetic reaction, a population of "normal"-persons, is dependent upon a multiplicity 63 (1990).
which is terminated by the addition of the stop solution. of factors: the specificity of the method, the population tested and 2. Watt KW, et al., Proc Nat Acad Sci USA, 83, 3166-70 (1986).
Therefore, the substrate and stop solution should be added in the precision of the method in the hands of the analyst. For these 3. Chen Z, Prestiglacomo A, Stamey T, Clin Chem, 41, 1273-82
the same sequence to eliminate any time-deviation during reasons, each laboratory should depend upon the range of (1995).
reaction. expected values established by the Manufacturer only until an 4. Wild D, The Immunoassay Handbook, Stockton Press, 452,
6. Plate readers measure vertically. Do not touch the bottom of in-house range can be determined by the analysts using the (1994).
the wells. method with a population indigenous to the area in which the 5. Junker R, Brandt B, Zechel C, Assmann G, Clin Chem, 43,
7. Failure to remove adhering solution adequately in the laboratory is located. 1588-94 (1997).
aspiration or decantation wash step(s) may result in poor 6. Prestigiacomo AF, Stamey TA, “Physiological variations of
replication and spurious results. serum prostate antigen in the (4-10 ng/ml) range in male
14.0 PERFORMANCE CHARACTERISTICS
8. Use components from the same lot. No intermixing of reagents volunteers’, J Urol, 155, 1977-80 (1996).
from different batches. 7. Stamey TA, McNeal JE, Yemoto CM, Sigal BM, Johnstone IM,
14.1 Precision
9. Patient specimens with PSA concentrations above 100 ng/ml “Biological determinants of cancer progression in men with
The within and between assay precisions of the PSA AccuBind®
may be diluted (for example 1/10 or higher) with normal female prostate cancer” , JAMA 281, 1395-1400 (1999).
ELISA test system were determined by analyses on three different
serum (PSA = 0 ng/ml) and re-assayed. The sample’s 8. Chen Z, Prestigiacomo A, Stamey T, “Purification and
levels of control sera. The number, mean value, standard
concentration is obtained by multiplying the result by the characterization of Prostate Specific Antigen (PSA)
deviation and coefficient of variation for each of these control sera
dilution factor (10). are presented in Table 2 and Table 3. Complexed to α 1 - Anticymotrypsin: Potential reference
10. Accurate and precise pipetting, as well as following the exact Material for International Standardization of PSA
time and temperature requirements prescribed are essential. TABLE 2 Immunoassays”, Clin Chem, 41/9, 1273-1282 (1995).
Any deviation from Monobind’s IFU may yield inaccurate Within Assay Precision (Values in ng/ml) 9. Horton GL, Bahnson RR, Datt M, Cfhan KM, Catalona WJ and
Landenson JH, “Differences in values obtained with two
results. Sample N X σ C.V.
11. All applicable national standards, regulations and laws, assays of Prostate Specific Antigen”, J Urol, 139, 762-72
Level 1 20 1.06 0.06 5.2% (1988).
including, but not limited to, good laboratory procedures, must
Level 2 20 3.56 0.18 5.1%
be strictly followed to ensure compliance and proper device 10. Stenman UH, Leinonen J, Alfthan H, Rannikko S, Tuhkanen K
Level 3 20 23.07 0.88 3.8% and Alfthan O,”A complex between prostate specific antigen
usage.
12. It is important to calibrate all the equipment e.g. Pipettes, TABLE 3 and α1-anticymotrypsin is the major form of prostate specific
Readers, Washers and/or the automated instruments used Between Assay Precision* (Values in ng/ml) antigen in serum of patients with prostate cancer: assay of
with this device, and to perform routine preventative complex improves clinical sensitivity for cancer”, Cancer Res,
maintenance.
Sample N X σ C.V.
51, 222-26 (1991).
13. Risk Analysis - as required by CE Mark IVD Directive 98/79/EC Level 1 20 0.98 0.08 8.5%
- for this and other devices, made by Monobind, can be Level 2 20 3.35 0.19 5.7% Revision: 6 Date: 2022-MAY-01 DCO: 1557
requested via email from Monobind@monobind.com. Level 3 20 23.17 0.95 4.1% MP2125 Product Code: 2125-300
*As measured in ten experiments in duplicate.
12.2 Interpretation
1. Measurements and interpretation of results must be 14.2 Sensitivity
performed by a skilled individual or trained professional.