assay include high affinity and specificity antibodies (enzyme and 1. Wash Buffer the horizontal axis of the graph (the duplicates of the unknown
One (1) vial contains a strong acid (0.5M H2SO4) Store at
immobilized), with different and distinct epitope recognition, in Dilute contents of wash solution to 1000ml with distilled or may be averaged as indicated). In the following example, the
2-8°C.
excess, and native antigen. In this procedure, the immobilization deionized water in a suitable storage container. Diluted buffer average absorbance (0.721) intersects the dose response
I. Product Instructions
takes place during the assay at the surface of a microplate well can be stored at room temperature (2-30°C) for up to 60 days. curve at (58.4U/ml) CA 15-3 concentration (See Figure 1).
through the interaction of streptavidin coated on the well and Note 1: Do not use reagents beyond the kit expiration date.
exogenously added biotinylated monoclonal anti-CA15-3 antibody.
EXAMPLE 1 including, but not limited to, good laboratory procedures, must TABLE 3 Revision: 4 Date: 2019-Jul-16 DCO: 1353
Sample Well Mean Value be strictly followed to ensure compliance and proper device Between Assay Precision* (Values in U/ml) MP5625 Product Code: 5625-300
Abs (A)
I.D. Number Abs (B) (U/ml) usage. Sample N X σ C.V.
A1 0.044 12. It is important to calibrate all the equipment e.g. Pipettes, Level 1 10 22.2 2.0 9.1% Size 96(A) 192(B)
Cal A 0.043 0 Readers, Washers and/or the automated instruments used Level 2 10 58.5 3.85 6.6%
B1 0.042 A) 1ml set 1ml set
with this device, and to perform routine preventative Level 3 10 104.6 9.33 8.9%
C1 0.204 B) 1 (12ml) 2 (12ml)
Cal B 0.198 10 maintenance.
Reagent (fill)
D1 0.191 *As measured in ten experiments in duplicate.
13. Risk Analysis- as required by CE Mark IVD Directive 98/79/EC - C) 1 (12ml) 2 (12ml)
E1 0.560 for this and other devices, made by Monobind, can be 14.2 Sensitivity D) 1 plate 2 plates
Cal C 0.543 40
F1 0.525 requested via email from Monobind@monobind.com. The CA 15-3 procedure has a analytical sensitivity of 0.2 U/ml at E) 1 (20ml) 1 (20ml)
G1 1.103 three (3) SD from the zero calibrator. The functional sensitivity F) 1 (50ml) 2 (50ml)
Cal D 1.064 100
H1 1024 12.2 Interpretation (20% CV) was found to be 1.25U/ml. G) 1 (12ml) 2 (12ml)
A2 1.784 1. Measurement and interpretation of results must be
Cal E 1.777 200 performed by a skilled individual or trained professional. 14.3 Accuracy H) 1 (8ml) 2 (8ml)
B2 1.770
2. Laboratory results alone are only one aspect for determining The CA 15-3 AccuBind® ELISA test system was compared with a
C2 2.431
Cal F 2.438 400 patient care and should not be the sole basis for therapy, reference method. Biological specimens from low, normal, and
D2 2.445 elevated concentrations were assayed. The total number of such
particularly if the results conflict with other determinants.
A3 0.737 3. The reagents for the test system have been formulated to specimens was 43. The least square regression equation and the
Patient 0.721 58.4
B3 0.705 eliminate maximal interference; however, potential interaction correlation coefficient were computed for the CA 15-3 in
between rare serum specimens and test reagents can cause comparison with the reference method. The data obtained is
erroneous results. Heterophilic antibodies often cause these displayed in Table 4.
Figure 1 interactions and have been known to be problems for all kinds
of immunoassays (Boscato LM, Stuart MC. ‘Heterophilic TABLE 4
3 Method Mean Least Square Correlation
antibodies: a problem for all immunoassays’ Clin. Chem.
2.5 1988:3427-33). For diagnostic purposes, the results from this Regression Analysis Coefficient
assay should be in combination with clinical examination, Monobind (y) 180.2 y=-0.219+1.008(x) 0.99
2 patient history and all other clinical findings. Reference (x) 178.6
Absorbance