Professional Documents
Culture Documents
strip comprises a sample pad, a glass fiber, a 2. Reading card: Insert the magcard into the analyzer and click
“read card” on the machine. (only read one time for the same
nitrocellulose membrane (the detection zone
Strip item of the same batch) Then select the calibration curve for the
is coated with a 25-OHVD-BSA conjugate,
corresponding sample type.
and the quality control zone is coated with a
3. Sampling. Use a pipettor to add 20ul sample to 80ul buffer A(add
chicken IgY), absorbent paper, PVC board.
to a tube suitable for Diagreat’s incubator hole) , and then add
Buffer A Main component is phosphate buffered
Beijing Diagreat Biotechnologies Co., Ltd
30ul buffer B to former compound for full mixing. Connect QC product evaluation effectiveness: test results of QC products
power supply for Diagreat’s incubator. (Temperatures has been in test kit boxes are all within QC range.
set by Diagreat. Please do not adjust the temperature). The above Inter bottle difference of QC products: CV≤15%.
compound was put into the incubator tube location for 15 NOTES
minutes incubation. This product is for in vitro diagnostic use only.
4. Add samples: Take out the strip from the bag and use a pipette to The operation steps cannot be omitted or simplified.
take 80 ul of the step 3 sample and add it to the sample well of The reagent contains preservatives. It is strictly forbidden to
the strip. inhale or touch the skin. Once touched, immediately flush the
5. Test: After sample added, please insert the strip into Diagreat touched area with water.
incubator for 10 mins incubation. Please strictly control the time This product should be stored at 2 °C - 8 °C. It should be used as
for 10 mins. After incubation, please insert the strip into soon as possible after opening, to avoid being placed in the air
Diagreat analyzer to read the result.(For D10 analyzer, click for too long to result in moisture. Expired products cannot be
“quick test”). used.
6. The dry fluorescence immunoassay analyzer measures and Disinfect or treat all samples and reagents with potential
analyzes the optical signal to quantify the concentration of the pollution according to local regulations.
substance to be tested. INDEX OF SYMBOLS
REFERENCE INTERVAL
This product selected 200 healthy human serum in Beijing, using
95% confidence interval, and combined with reference to determine
the reference interval.
The sample reference interval is: 30-80 ng/mL. The above reference
interval is for reference only. Each laboratory should establish an For technical assistance, please contact:
expected normal reference interval through experiments. Diagreat’s Technical Services
INTERPRETATION OF TEST RESULT Tel: (8610) 69502385
25-OHVD is normal at 30-80 ng/mL, <30 ng/mL is not E-mail: info@diagreat.com
conducive to bone health, >150ng/mL may cause poisoning.
LIMITATIONS OF THE TEST Beijing Diagreat Biotechnologies Co., Ltd
This kit is limited to in vitro diagnostics. Floor 1-3, Building 19, No. 2 Yard, Huanke Middle Road, Tongzhou
Diagnosis and treatment cannot rely solely on this test result. District, Beijing, China.
Please consider both clinical history and other laboratory test Tel: (8610) 69502385
results. Fax: (8610) 69507615
PRODUCT PERFORMANCE www.diagreat.com
Appearance: The appearance of the reagent should be neat and
tidy, and the text symbol should be clearly marked. There is no
leakage. The test strip is intact. Freeze-dried products are
colorless or pale yellow after redissolution.
Analytical Sensitivity: 1ng/mL. Revision No: 01
Linear range: 3-100ng/mL, and the correlation coefficient Date of last revision: 2021/6/2
R≥0.990.
Repeatability (Precision): Repeated testing (20 ± 4) ng / mL and
(80 ± 16) ng / mL samples, the result coefficient of variation
CV≤15%.
Inter batch difference: Testing (20 ±4) ng / mL and (80 ±16) ng /
mL samples, the relative difference R≤15%.
Accuracy: The recovery rate is 85% to 115%.
Beijing Diagreat Biotechnologies Co., Ltd