Professional Documents
Culture Documents
RLU /s(A)
RLU/s (B)
1. Gharib H, Ryan RJ, Mayberry WE, & Hockett T,
Number
Sample
(ng/ml)
Value
Mean
repeat the dose response curve. expected values established by the Manufacturer only until an "Radioimmunoassay for Triiodothyronine (T3): Affinity and
Well
I.D.
5. The addition of signal reagent initiates a kinetic reaction, in-house range can be determined by the analysts using the Specificity of Antibody for T3", J Clinical Endocrinol, 33, 509
therefore the signal reagent(s) should be added in the same method with a population indigenous to the area in which the (1971).
sequence to eliminate any time-deviation during reaction. laboratory is located. 2. Chopra IJ, Ho RS & Lam R, "An improved radio-immunoassay
A1 100468 6. Failure to remove adhering solution adequately in the of triiodothyronine in human serum", J Lab Clinical Med, 80,
Cal A 100000 0 aspiration or decantation wash step(s) may result in poor 14.0 PERFORMANCE CHARACTERISTICS
B1 99532 729 (1971).
replication and spurious results. 3. Young DS, Pestaner LC, and Gilberman U, "Effects of Drugs
C1 88644 7. Use components from the same lot. No intermixing of 14.1 Precision on Clinical Laboratory Tests", Clinical Chemistry, 21, 3660
Cal B 86989 0.5 reagents from different batches.
D1 85333 The within and between assay precision of the T3 AccuLite™ (1975).
8. Patient specimens with T3 concentrations above 7.5 ng/mL CLIA assay were determined by analyses on three different levels 4. Sterling L, Diagnosis and Treatment of Thyroid Disease,
E1 71996 may be diluted ½ with ‘0’ serum reference. The sample’s of pool control sera. The number, mean value, standard deviation Cleveland, CRC Press, 9, 51 (1975).
Cal C 72275 1.0 concentration is obtained by multiplying the result by the
F1 72553 (σ) and coefficient of variation for each of these control sera are 5. Tietz N, Textbook of Clinical Chemistry, 3rd Ed, WB Saunders,
dilution factor, 2. presented in Table 2 and Table 3. Philadelphia (1999).
G1 46658 9. Accurate and precise pipetting, as well as following the exact TABLE 2 6. Harrison, Principles of Internal Medicine, 12th Ed, McGraw Hill,
Cal D 45952 2.5 time and temperature requirements prescribed are essential.
H1 45247 Within Assay Precision (Values in ng/ml) New York (1998).
Any deviation from Monobind’s IFU may yield inaccurate
A2 26556 results. Sample N X σ C.V.
7. Surks MI, Chopra IJ, Mariash CN, “American Thyroid
Association Guidelines for use of laboratory tests in thyroid
Cal E 26887 5.0 10. All applicable national standards, regulations and laws, Low 20 0.80 0.05 6.3%
B2 27219 disorders, JAMA, 263, 1529-1532 (1990).
including, but not limited to, good laboratory procedures, Normal 20 1.56 0.09 5.8% 8. Gruhn JG, Barsano CP, Kumar Y,”The development of tests of
C2 19549 must be strictly followed to ensure compliance and proper High 20 3.06 0.10 3.3 % thyroid function”, Arch Pathol Lab Med, 111, 84-100 (1987).
Cal F 18710 7.5 device usage.
D2 17871 TABLE 3 9. Bartalena L, “Recent achievements in studies on thyroid
11. It is important to calibrate all the equipment e.g. Pipettes, Between Assay Precision (Values in ng/ml) hormone-binding proteins”, Endocrine Rev, 11, 47-64(1990).
E2 71096 Readers, Washers and/or the automated instruments used
Ctrl 1 71745 1.0 with this device, and to perform routine preventative Sample N X σ C.V. 10. Engler D, Burger AG, “The deiodination of the iodothyronines
F2 72393 and of their derivatives in man”, Endocrine Rev, 5, 151 (1984).
maintenance. Low 20 0.81 0.08 9.9%
G2 49482 12. Risk Analysis- as required by CE Mark IVD Directive Normal 20 1.57 0.16 10.2%
Ctrl 2 49732 2.2 Revision: 4 Date: 060712 DCO:0538
98/79/EC - for this and other devices, made by Monobind, High 20 2.89 0.27 9.3%
H2 49981 Cat #: 175-300
can be requested via email from Monobind@monobind.com. *As measured in several experiments in duplicate.
A3 64506
Patient 63817 1.4 12.2 Interpretation Size 96(A) 192(B) 480(D) 960(E)
B3 63127 14.2 Sensitivity
1. Measurements and interpretation of results must be The Triiodothyronine test system procedure has a sensitivity of A) 1ml set 1ml set 2ml set 2ml set x2
performed by a skilled individual or trained professional. 0.126 ng/ml. The sensitivity was ascertained by determining the
*The data presented in Example 1 and Figure 1 is for illustration 2. Laboratory results alone are only one aspect for determining variability of the 0 ng/ml serum calibrator and using the 2σ (95% B) 1 (1.5ml) 2 (1.5ml) 1 (8ml) 2 (8ml)
only and should not be used in lieu of a dose response curve patient care and should not be the sole basis for therapy, certainty) statistic to calculate the minimum dose.
Reagent (fill)
prepared with each assay. In addition, the RLU’s of the particularly if the results conflict with other determinants. C) 1 (13ml) 2 (13ml) 1(60ml) 2 (60ml)
calibrators have been normalized to 100,000 RLU/sec for the A 3. For valid test results, adequate controls and other 14.3 Accuracy
calibrator (greatest light output). This conversion minimizes parameters must be within the listed ranges and assay D) 1 plate 2 plates 5 plates 10 plates
The T3 AccuLite™ CLIA test was compared with a reference
differences caused by efficiency of the various instruments that requirements. enzyme immunoassay method. Biological specimens from E) 1 (20ml) 1 (20ml) 1 (60ml) 2 (60ml)
can be used to measure light output. 4. If test kits are altered, such as by mixing parts of different hypothyroid, euthyroid and hyperthyroid populations were used
kits, which could produce false test results, or if results are (The values ranged from 0.010ng/ml – 7.30ng/ml). The total F) 1 (7ml) 2 (7ml) 1 (30ml) 2 (30ml)
FIGURE 1 incorrectly interpreted, Monobind shall have no liability. number of such specimens was 110. The least square regression
5. If computer controlled data reduction is used to interpret the equation and the correlation coefficient were computed for the T3 G) 1 (7ml) 2 (7ml) 1 (30ml) 2 (30ml)
120000
results of the test, it is imperative that the predicted values AccuLite™ CLIA in comparison with the reference method. The
100000 for the calibrators fall within 10% of the assigned data obtained is displayed in Table 4.
concentrations. TABLE 4
80000 6. Total serum triiodothyronine concentration is dependent Lease Square
upon a multiplicity of factors: thyroid gland function and its Mean Regression Correlation
RLU
60000 Unknown #1 regulation, thyroxine binding globulin (TBG) concentration, Method (x) Analysis Coefficient
and the binding of triiodothyronine to TBG (3, 4). Thus, total
40000 This 1.35 y = -15.0 + 1.085(x) 0.972
triiodothyronine concentration alone is not sufficient to
Method
20000 assess clinical status.
Reference 1.38
7. A decrease in total triiodothyronine values is found with
0 protein-wasting diseases, certain liver diseases and
0 2 4 6 8
administration of testosterone, diphenylhydantoin or Only slight amounts of bias between this method and the
T3 Values in ng/ml
salicylates. A table of interfering drugs and conditions, which reference method are indicated by the closeness of the mean
affect total triiodothyronine values, has been compiled by the values. The least square regression equation and correlation
Journal of the American Association of Clinical Chemists3. coefficient indicates excellent method agreement.