A calibration method for mass concentration of aerosol photometer

Zhang Ailiang 1 ÿ Liu Yue1 ÿ Chen Qiyue 1 ÿ Wang Tingting 2 , Shi Chanzhen 1 , Zhang Zitong 1

(1. Shanghai Institute of Metrology and Testing Technology, Shanghai 201203; 2.

Shanghai Institute of Quality Supervision and Inspection Technology, Shanghai 200233)

Abstract: The internal mechanism and working principle of aerosol photometer are introduced, and a calibration device and calibration method for mass

concentration of aerosol photometer based on filter sampling method are proposed. The concentration stability is less than 5%, and the mass concentration indication

error of the aerosol photometer is less than ±10%. The traceability of the mass concentration of the aerosol photometer is solved, which can effectively improve the

measurement accuracy of the device. Key words: metrology; aerosol photometer; mass concentration; calibration method

Article number: 1000-1158 (2020) 06-0765-04

ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

ZHANGAi-liang1 , LIUYue1 , CHENQi -yue1 , WANGTing-ting2 , SHIChan- zhen1 , ZHANGZi -tong1

ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿ
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ


ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ±ÿÿÿÿÿÿÿ
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
ÿÿÿÿÿ
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

Therefore, it cannot meet the demand for traceability of the testing equipment. The

1 Introduction Shanghai Institute of Metrology and Testing Technology has undertaken the task of

formulating the calibration specification for the "Aerosol Photometer" issued by the

With the gradual improvement of the cleanliness requirements of production and National Environmental Chemical Metrology Technical Committee in 2018. In this

purification areas in pharmaceutical and medical and health enterprises and institutions, specification, the technical indicators and calibration methods for the mass concentration

the challenges to clean technology will also reach an unprecedented depth and breadth of aerosol photometers need to be formulated. .

[1] . At present, aerosol photometers are widely used to detect high-efficiency filters and

filtration terminals of purification products to verify whether the relevant areas and 2 Working principle and optical measurement system

equipment meet the purification requirements. Therefore, the accuracy of the aerosol

photometric count value directly affects the final test result in a clean environment. The working principle of the aerosol photometer (hereinafter referred to as the instrument)

Compared with the previous detection of ambient air cleanliness by dust particle counters, is: the instrument extracts the air samples at the upstream and downstream ends of the filter

aerosol photometers can more intuitively and accurately reflect the leak points of the system under test at a constant flow rate, the sampling unit delivers the air sample to the

filtration system under inspection. At present, this detection method has been widely used measurement unit, and the measurement unit measures the aerosol in the air sample. The mass

by national standards and industry standards, as a widely used detection equipment, but there is no domestic is measured, and the data processing unit analyzes the measurement results and calcul
concentration

Received date: 2019-02-28; Revised date: 2019-08-27


Fund Project: Scientific Research Project of Shanghai Municipal Bureau of Quality and Technical Supervision (2018-21); Scientific Research Project of Shanghai Institute of Metrology and Testing Technology (100RJ1811)
Machine Translated by Google
ÿÿÿ Journal of Metrology June 2020

The ratio of the two calculates the leakage rate of the system under test, and finally the measurement result is output The retention efficiency of quasiparticles is not less than

by the display unit. The schematic diagram of the instrument detection principle is shown in Figure 1. 99.99%. 3.2 Polyalphaolefins

Polyalphaolefin (PAO) is a kind of synthetic base oil, which is a kind of synthetic

base oil and is widely used. PAO-4 (chemical code 68649-12-7) aerosol refers to the

generation of poly-alpha-olefin oil (PAO) with a kinematic viscosity of 4m2/s, which is

generated by aerosol photometer fogging and mixing device. Polydisperse Aerosol. 3.3

Sampling and weighing system (1) Sampling device: the flow range is 0~50L/min, and the

stability of 4h should be less than 5%; (2) Flow standard device: The range is 0~50L/min,

the maximum

Figure 1 Schematic diagram of detection principle

ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

The light scattering chamber of the measurement unit is a complex optoelectronic


Allowable error ±1.0%; cumulative sampling volume can be displayed;
unit that contains a conical light sensor attached to the tip, and a pair of collimating lenses
(3) Balance: Accuracy level ?, the actual division value is less than 1 × 10-6g; (4)
adjusts the incoming light from the light source and focuses it into the sampling conical
Aerosol fogging and mixing device: It can generate PAO-4 aerosol with controllable
cavity. A small hole forms a conical dark area around the photomultiplier tube, preventing

light from reaching the photomultiplier tube directly. A condenser lens facing away from concentration and uniform mixing. The interior is equipped with multiple III-A

the light source directs the scattered light to the photomultiplier tube through a conical dark brushed gold nozzles, which generate polydisperse aerosols from PAO-4 oil through the

area. Its optical principle is: based on the Mie scattering theory of light, the particle size is Venturi effect. Using the filter membrane sampling method to measure the concentration

close to or larger than the wavelength of the incident light; the scattered light intensity is of the device's fog outlet, the stability of the aerosol mass concentration within 20min

asymmetric in all directions, and the forward scattering along the incident direction is the
should be less than 5% [4] . Figure 3 shows the structure of the ÿ-A wire drawing gold
strongest, and the larger the particle is. The larger the value, the stronger the forward
nozzle [5] .
scattering [2] . The scattered light intensity is proportional to the particle number

concentration. If the relative density of the particles is known, the ratio of the particle

number to the mass is a constant, so the scattered light intensity is proportional to the

particle mass concentration. The purpose of indirect measurement of aerosol concentration

can be achieved by measuring the scattered light intensity caused by aerosol under a

specific light path [3] . The schematic diagram of the optical measurement system of the

instrument is shown in Figure 2.

Fig.2 Diagram of optical


measurement system

3 Mass concentration calibration equipment


Fig.3 Structure diagram of type III-A wire drawing gold nozzle

ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
3.1 Filter

Membrane The filter membrane is circular with a diameter of (37 ± 0.25) mm; the
4. Calibration method of mass concentration indication error
pore size is less than 2 ÿm; the thickness is 0.2 to 0.25 ÿm; at a clean air flow rate of 0.45

m/s, the pressure It should be less than 3kPa; the mass increase value after exposure to
4.1 Preparation before calibration (1)
35% RH air for 24h should not exceed 10ÿg; glass fiber filter membrane, quartz filter

The visual inspection of the filter membrane should ensure that the edges are flat, the thickness is uniform,
membrane and other inorganic filter membranes can be selected.
Machine Translated by Google
Volume 41 Issue 6 Zhang Ailiang et al.: A Calibration Method for Mass Concentration of Aerosol Photometer ÿÿÿ

No burrs, no pollution, no pinholes or any damage.

(2) Before sampling, the filter membrane should be balanced and weighed 5 Verification of calibration results

in a constant temperature and humidity environment, that is, 15-30 °C, (50±5)% RH,
5.1 Stability verification of the mist outlet of the aerosol misting and mixing device
and the filter membrane should be placed for at least 24 hours; The wet equipment
Since the stability of the mass concentration of the misting port of the aerosol
remains the same; after the filter is equilibrated, the filter is weighed using an
misting and mixing device is directly related to the calibration result of the mass
analytical balance.
concentration of the aerosol photometer, it is necessary to verify the stability of the
4.2 Mass concentration calibration
aerosol concentration of the misting port [7] . By adjusting the relevant parameters of

Set the internal reference value of the instrument to be calibrated to 100 mg/m3 . Perform the aerosol fogging and mixing device, a stable aerosol can be generated.

the calibration according to the calibration system shown in Figure 4. Select an aerosol photometer with a repeatability better than 1.5% and connect it to

the fogging port of the aerosol fogging and mixing device. After the instrument is

stable, read the initial value of the aerosol photometer and start timing. The

concentration value was read once every 5 min, for a total of 5 times. Take the

maximum and minimum values of the five readings, and calculate the 20-min stability

of the aerosol fogging and mixing device according to formula (4).

Figure 4 Schematic diagram of the instrument concentration indication error calibration system

ÿÿÿÿÿÿÿÿÿ ×
ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿ ÿÿÿ ÿÿÿ
r
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
In the formula: ÿÿ is the stability of the aerosol fogging mixing device; ÿmax is the maximum value

Use the aerosol fogging and mixing device to generate PAO-4 aerosol with a mass concentration
of the five concentration values, mg/m3 ; ÿmin is the minimum value of the five concentration

of 1 to 100 mg/m3 . Connect the fogging ports I and II of the aerosol photometer fogging and mixing
values, mg/m3 ; ÿ is the maximum value of the five concentration values. Average, mg/m3 .
device to the filter membrane respectively through hoses. The sampling port of the sampling device and
Three kinds of aerosols with different mass concentrations were generated
the downstream sampling port of the instrument to be calibrated shall conduct equal flow sampling at the
by using the aerosol fogging and mixing device.
same time, and the measured value Cs1 of the filter membrane sampling device shall be calculated
ÿÿÿ
And record the concentration measurement value Cm1 of the instrument being calibrated .
according to formula (1).

ÿÿÿÿÿÿÿÿ ÿÿÿ
V

Where: Cs is the measured value of the filter membrane sampling device, mg/m3 ; w1 is the mass of the

filter membrane before sampling, mg; w2 is the mass of the filter membrane after sampling, mg; V is the

sampling volume, m3 . The sampling port of the exchange filter sampling device is connected to the

downstream sampling port of the instrument to be calibrated, and the misting port of the aerosol

mist mixing device is connected. At the same time, equal flow sampling is carried out, the measured value

Cs2 of the membrane sampling device is calculated and the concentration measurement of the instrument
Fig.5 Stability curve of misting port I
under calibration is recorded. value Cm2 . Calculate the ith single concentration indication error ÿi of the
ÿÿÿÿÿ ÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿ
instrument according to formula (2), mg/m3 . (Cm1 + Cm2 ) - (Cs1 + Cs2 ) ÿi=

ÿÿÿ
ÿ

In the formula: Cs1 is the calculated value of the filter membrane sampling device connected to the misting

port I, mg/m3 ; Cs2 is the calculated value of the filter membrane sampling device connected to the misting

port II, mg/m3 ; Cm1 is the instrument to be calibrated connected to the misting port. The measured value

of ÿ, mg/ m3 ; Cm2 is the measured value of the instrument to be calibrated connected to the fogging port

ÿ, mg/m3 . Repeat the above operation 4 times, and calculate the mass concentration indication

error ÿ of the calibrated instrument according to formula (3), mg/m3 .


ÿ

ÿ ÿÿ Fig.6 Fogvent II stability curve


ÿÿÿ
dÿ ÿÿÿ
ÿ
Machine Translated by Google
ÿÿÿ Journal of Metrology June 2020

Table 1 Stability of the mass concentration of the mist nozzle at different concentrations The stability of the fog concentration of the fog mixing device is better than 5%, and the

ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ The error of the mass concentration indication value of the glue photometer is less than ±10%. thus effective

ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
Solve the problem of traceability of mass concentration of aerosol photometer

Aerosol fog concentration Mist mouth I Mist mouth II It ensures the accuracy of mass concentration measurement of aerosol photometer and can improve the

ÿÿÿ·ÿÿÿ stability/(%) stability/(%) Final test results for clean environments.

ÿÿ ÿÿÿ ÿÿÿ
[ references]
ÿÿ ÿÿÿ ÿÿÿ

[1] Huang Tinglei, Li Min, Chen Yazhu. Development of Optical Particle Counting Device
ÿÿÿ ÿÿÿ ÿÿÿ
[J]. Chinese Journal of Metrology, 2006, 27(1): 88-90.
From the data in Table 1, it can be seen that there are two aerosol mist mixing devices.
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
The stability of the concentration of the fogging port within 20min is less than 5%, which is in line with the technical requirements.
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
technical requirements. Therefore, the misting port of the aerosol misting and mixing device is unstable ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

The systematic error introduced by the property is under control, the device can be used for gas [2] Wang Shimin, Zhu Zhen, Ye Mao, et al. Mie theory in light scattering particle size measurement

Sol photometer mass concentration calibration work. On two improved numerical calculation methods [J]. Journal of Metrology, 1999,

ÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
5.2 Verification of filter sampling calibration results

ÿÿÿÿÿ ÿÿ ÿÿÿ ÿÿ ÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿ


Use 3 types of mainstream aerosol photometers on the market
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
The above calibration method performs aerosol mass concentration calibration. Calibration result
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
See Table 2.
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

Table 2 Calibration results of aerosol mass concentration [3] Wang Zhengqi, Hu Guangyong, Zhang Yuchen. Aerosol Photometer R&D Analysis

ÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ [J]. Medical Equipment, 2014, 27(8): 18-19.

ÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ


filter sampling Tested instrument single quality relatively
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
Concentration Display quality Concentration indication Indication
model
average value Concentration average error error ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

ÿÿÿ·ÿÿÿ ÿÿÿ·ÿÿÿ ÿÿÿ·ÿÿÿ ÿÿÿÿ [4] Zhang Ailiang, Gu Lingling, Gao Zheng, et al. Aerosol for leak detection of high efficiency filter

Research on the calibration method of glue photometer [J]. Cleaning and Air Conditioning Technology, 2016,
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ ÿÿÿ
ÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿ
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ ÿÿÿ
ÿÿÿÿÿÿÿÿ ÿÿÿÿ ÿÿÿÿ ÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ on
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿ ÿÿÿ
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ

ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ


ÿÿ
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ ÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
ÿÿÿÿÿ
ÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿÿ ÿÿÿ
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ ÿÿÿ
ÿÿÿÿÿÿÿÿÿÿ ÿÿÿ ÿÿÿÿÿÿ ÿÿÿÿÿÿÿÿÿ ÿÿÿÿ ÿÿ ÿÿÿÿÿÿÿ
ÿÿÿÿÿ ÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
ÿÿÿÿÿÿ
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ ÿÿÿÿ [6] HJ618-2011, Determination of ambient air PM10 and PM2.5 weight
ÿÿÿÿÿÿ
ÿÿÿÿÿ ÿÿÿÿÿ ÿÿÿÿ Law [S].
ÿÿÿÿÿ
[7] Zhang Wenge, Liu Wei, Xu Xiao. National level for PM2.5 monitor detection

The development of reference material [J]. Chinese Journal of Metrology, 2019, 40(1): 159-
From the data in Table 2, it can be found that the aerosol hair mist is used for mixing.
ÿÿÿÿ
The device and filter sampling method were used to calibrate the aerosol photometer and were calibrated
ÿÿhangÿ ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿationÿÿÿÿÿ
The relative indication error of the mass concentration of the aerosol photometer is less than
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
±10%. The calibration results meet the technical requirements of these three types of instruments
ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ
Therefore, this method can effectively solve the problem of traceability of such instruments.

need.

First author: Zhang Ailiang (1982-), male, Anhui

6 Conclusion From Hefei, Shanghai Institute of Metrology and Testing Technology

Senior engineer, mainly engaged in environmental measuring instruments

In this paper, an aerosol photometer mass concentration parameter is given. Metrology calibration technology and standardization research.

Calibration device and calibration method, and experimentally verified aerosol emission ÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿÿ