You are on page 1of 3

User instructions

Control panel

Door button
Press here to open door. The door can only
be opened if the power is on.

Instrument
Washer Disinfector

The display shows the current


®
status and error messages

Status indicators

Ready press start READY


to start another
wash
PRE WASH

Pre-wash
WASH

Wash
RINSE

Rinse
DISINFECT

Disinfection
COMPLETE

Washing cycle ended


Remove items
CAREFUL - Items
! may be hot !

!
Start button
Press once to start the program or to pause an
ongoing program.

In case of error, acknowledge receipt of error


message on this button.

KEN IWD1, IWD2 Section/page: 5.1


Subject to alterations UK: 28.03.2007 ref.: M-IWD1_IWD2-UK
User instructions

Program start IWD 1 and IWD 2.


Fill the basket. For the optimum washing result, the basket should not be
overfilled.
Large items, such as trays and bowls, should not be placed flat in the
basket, since they would then keep the water from hitting other items.
Scissors and similar instruments should be opened 90°. Place small items
in a closed wire basket, so they will not run down the drain and damage
the pump.
Do not overfill the basket; place items so as to allow water jets to hit all
items from the top and the bottom.

IWD 2 is also able to wash and disinfect up to 6 hollow cavity


instruments. The flexible Quick-Click System is used for instruments
from 1 to 18 mm. Notice the colours red for small, blue for medium and
yellow and green for large diameters on the different connections. Plug off
connections that are not in use.
IWD2 cannot wash hollow cavity instruments that are all clogged; it must
thus be checked that water through-flow is possible, before the instrument
is placed in the basket.

Push the basket into the washing chamber and check that it touches the
coupling at the back of the washing chamber.
Close the door; then push the start button to activate a washing program.
! The door is locked throughout the program sequence. Do not try to
force the door open!

When the program is done, the door opens - approx. 30mm , thereby
allowing vapours to escape and the items to dry (the heat from the items
will make the water evaporate).
Note! Be careful if you remove the items just after the program has
! ended. If the program includes disinfection, the items will be hot!

Operating disturbances In case of error in the program sequence, the program will stop and the
display will show the error.
Press the start button to acknowledge receipt of the error message. Drain
the water out of the Washer; the door can now be opened.
The items are not disinfected correct the error and restart the
Washer!

! Regarding error types and remedial action, cf. the Instruction Manual.

KEN IWD1, IWD2 Section/page: 5.2


Subject to alterations UK: 28.03.2007 ref.: M-IWD1_IWD2-UK
User instructions

Washing programs
IWD 1 and IWD 2 offer up to 6 different programs. Programs 1 to 4 are the
wash program options. Programs 5 and 6 are test and demonstration
programs that may not be used as wash programs. Programs 3 and 4 comply
with EN15883 recommendations regarding disinfection.

Program 1 Program 2 Program 3 Program 4


Enzyme soap Alkaline soap Enzyme soap Alkaline soap

Pre-wash duration 3 minutes 3 minutes 3 minutes 3 minutes


Wash duration 5 minutes 5 minutes 5 minutes 5 minutes
Wash temperature 43-45°C 53°C 43-45°C 53°C
Rinse 1 and 2 duration 2 minutes 3 x 1 minute 2 minutes 3 x 1 minute
Disinfection duration 2 minutes 2 minutes 2 minutes 10 minutes
Disinfection temperature 80-85°C 80-85°C 90°C 80-85°C

Program change Use the following procedure for program change: Load Instruments
1. Display must show message shown in this frame. Press START
Open door by pressing door button.

2. With door open, push door button and keep it down.


Set Clock & Prog
3. Use the start button to scroll through time and date PROG 1
settings, until the following display comes up.

4. Keep the start button down to scroll through the programs 1 to 6.


Release the start button when the desired program is shown in the
display.

5. Press the door button to save the change and leave the menu.
When a program has been selected, this program will be used until the
procedure is repeated and another choice made.
We recommend using program 3 as the default program.

KEN IWD1, IWD2 Section/page: 5.3


Subject to alterations UK: 28.03.2007 ref.: M-IWD1_IWD2-UK