The product consists of a cuff, hose, adhesive tape, and cross ties.
单击添加标题
Schematic diagram of cylindrical cuff structure Schematic diagram of conical cuff structure
All tourniquets are available in different wrap colors, and different adhesive tape buckle colors for identification.
Other Accessories
Hose Connectors
Sleeves
Product parameters
editoion
Tourniquet Cuff Specification
editoion
Tourniquet Cuff Specification
editoion
Tourniquet Cuff Specification
editoion
editoion
Label
2.Open the sterile sheet bag, remove the gauze sleeve, and wrap the gauze sleeve
around the biofilm.
①
3.Then take out the cuff and wrap it on the gauze sleeve, fix the adhesive tape
and fasten it with the horizontal ties on both sides of the cuff to prevent the
product from being inflated and twisted. The cuff should be loosened and
tightened appropriately. Generally, the length of the overlapping part of the
tied cuff should be 50mm to 150mm, too long or too short is not appropriate,
which will bring poor cuff stability (cuff twisting) and uneven pressure
distribution, and even produce abnormal surgical blood leakage. When fixing
② the cuff, care should be taken to adjust the direction of the trachea to
facilitate the connection of the trachea to the inflation equipment.
4.Connect the air tube to the inflatable output interface of the inflatable device such
③ as the electric air pressure hemostat. The interface of the inflatable device should
have a locking effect to ensure that it will not automatically disconnect when
working.
5.After the above work is completed, set the appropriate parameters on the inflatable equipment and
start the equipment to inflatable products. When the product is removed at the end of the work, the
air tube should be disconnected from the inflatable equipment, then loosen the straps and adhesive
tape, remove the cuff and gauze cover, and dispose of the product as medical waste.