Tourniquet Cuff

Main structure and composition

The product consists of a cuff, hose, adhesive tape, and cross ties.

1、Adhesive tape (barbed button) 2、Hose


3、Cuff tape 4、Adhesive tape (hair buckle)
5、Cross tie

单击添加标题

Schematic diagram of cylindrical cuff structure Schematic diagram of conical cuff structure

All tourniquets are available in different wrap colors, and different adhesive tape buckle colors for identification.
Other Accessories

Hose Connectors

Sleeves
Product parameters

Conditions of use and Main material: TPU


frequency of use: non-
autoclavable, >25 times.

Maximum use pressure.


Maximum storage life: Suitable for 650mmHg
greater than three years hemostat device with 4
hours continuous use time

Production Environment & Inspection


Tourniquet Cuff Specification

editoion
Tourniquet Cuff Specification

editoion
Tourniquet Cuff Specification

editoion
Tourniquet Cuff Specification

editoion

Suitable for 8" tourniquets cuff

Suitable for 12"-24" tourniquets cuff

Suitable for 26"-38" tourniquets cuff

Suitable for over 38" tourniquets cuff


Tourniquet Cuff Specification

editoion

1.Packaging:Disposable tourniquet cuff in a sterile pouch,reusable tourniquet cuff in a paper-plastic bag


2.Sterilization: radiation sterilization
3.Not made with natural rubber latex
4.Shelf life:Greater than three years
5.Meets ISO13485 standard
6.Manufacturing cerified ISO13485,FDA listing
Packaging
1、The package of disposable tourniquet cuff
Long white box, each containing 10 tourniquets, outer box size 56 * 44 * 44cm, a total of 40 in one box

2. The package of reusable tourniquet cuff


small white box, with an outer box size of 66 * 54 * 44cm, and a total of 48 in one box

Label

Tourniquet label display Packaging label display


Method of use
1.Wrap the biofilm around the upper end of the patient's surgical site (the static end is the upper
end of the limb) before use, and the biofilm will be procured and used by the clinical use unit
itself.

2.Open the sterile sheet bag, remove the gauze sleeve, and wrap the gauze sleeve
around the biofilm.


3.Then take out the cuff and wrap it on the gauze sleeve, fix the adhesive tape
and fasten it with the horizontal ties on both sides of the cuff to prevent the
product from being inflated and twisted. The cuff should be loosened and
tightened appropriately. Generally, the length of the overlapping part of the
tied cuff should be 50mm to 150mm, too long or too short is not appropriate,
which will bring poor cuff stability (cuff twisting) and uneven pressure
distribution, and even produce abnormal surgical blood leakage. When fixing
② the cuff, care should be taken to adjust the direction of the trachea to
facilitate the connection of the trachea to the inflation equipment.

4.Connect the air tube to the inflatable output interface of the inflatable device such

③ as the electric air pressure hemostat. The interface of the inflatable device should
have a locking effect to ensure that it will not automatically disconnect when
working.

5.After the above work is completed, set the appropriate parameters on the inflatable equipment and
start the equipment to inflatable products. When the product is removed at the end of the work, the
air tube should be disconnected from the inflatable equipment, then loosen the straps and adhesive
tape, remove the cuff and gauze cover, and dispose of the product as medical waste.