5mVp~±5mVp
Measurement accuracy Maximum mean difference: ±5 mmHg HR measuring range 15bpm~350bpm
Maximum standard deviation: 8 mmHg HR measuring accuracy ±1% or ±2bpm, whichever is greater
Measuring range HR alarm delay time ≤10s
Systolic pressure Adult 40mmHg~275mmHg Sensitivity selection ×1/4, ×1/2, ×1, ×2
Pediatric 40mmHg~200mmHg Sweeping speed 25mm/s
Neonate 40mmHg~135mmHg
ECG noise level ≤30µVp-p
Diastolic pressure Adult 10mmHg~210mmHg
ECG input loop current ≤0.1µA
Pediatric 10mmHg~150mmHg
Neonate 10mmHg~95mmHg Differential input impedance ≥10MΩ
Upto 12000 groups of NIBP, 3000 groups SpO2 and event record
Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available
ECG cable NIBP cuff Skin temp probe Infrared ear probe SpO2 sensor
PC-900PLUS
PC-900PLUS
Upto 12000 groups of NIBP, 3000 groups SpO2 and event record
Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available
ECG cable NIBP cuff Skin temp probe Infrared ear probe SpO2 sensor
Technique Oscillometric method Input dynamic range ±(0.5mVp~5mVp)
Pressure measuring range 0 mmHg~300mmHg Differential input impedance ≥10MΩ
Typical measurement time <30 seconds (adult cuff) Frequency response 1~20Hz (Operation)
SYS measuring range Adult: 40mmHg~275mmHg 0.67~40Hz (Monitoring)
Pediatric: 40mmHg~200mmHg 0.05~150Hz (Diagnostic)
Neonate: 40mmHg~135mmHg CMRR ≥90dB (Diagnostic)
MAP measuring range Adult: 20mmHg~230mmHg ≥105dB (Monitoring & Operation)
Pediatric: 20mmHg~165mmHg
Neonate: 20mmHg~110mmHg Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 an d Auto
Sweeping speed 6.25, 12.5, 25, 50mm/s, tolerance: ±10%
DIA measuring range Adult: 10mmHg~210mmHg
Pediatric: 10mmHg~150mmHg HR display range 15bpm~350bpm
Neonate: 10mmHg~95mmHg HR display accuracy ±1% or ±2bpm, whichever is greater
Measuring accuracy Maximum mean difference: ±5 mmHg
Maximum Standard deviation: 8 mmHg
Measuring mode Manual, Auto, STAT
Thermometer
Multi-parameter Patient Monitor
NIBP ECG
Technique Oscillometric Input dynamic range ±0.5mVp~±5mVp
Typical measuring time <30 seconds (typical adult cuff) HR measuring range 15bpm~350bpm
Initial cuff inflation pressure Adult <175mmHg HR measuring accuracy ±1% or ±2bpm, whichever is greater
Pediatric <135mmHg HR alarm delay time ≤10s
Neonate <65mmHg
Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 and Auto
Overpressure protection limit Adult 300mmHg Sweeping speed 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
Pediatric 240mmHg
Neonate 150mmHg ECG noise level ≤30µVp-p
ECG input loop current ≤0.1µA
Measuring range
Differential input impedance ≥10Mohm
Systolic pressure Adult 40mmHg~275mmHg
Common-mode rejection ratio ≥105dB (Monitoring mode)
Pediatric 40mmHg~200mmHg
(CMRR) ≥89dB (Diagnostic mode)
Neonate 40mmHg~135mmHg
Diastolic pressure Adult 10mmHg~210mmHg Time constant ≥0.3s (Monitoring mode)
Pediatric 10mmHg~150mmHg ≥3.2s (Diagnostic mode)
Neonate 10mmHg~95mmHg
Mean arterial pressure Adult 20mmHg~230mmHg
Pediatric 20mmHg~165mmHg
RESP
Neonate 20mmHg~110mmHg RR measuring range 0rpm~120rpm
RR measuring accuracy ±5% or ±2rpm, whichever is greater
Measurement accuracy Maximum mean difference: ±5 mmHg
Maximum standard deviation: 8 mmHg
Measurement mode Manual, Auto, STAT
Others
Automatic measuring intervals 1~480min Power supply 100~240Vac, 50/60Hz
Built-in battery 4400mAh Lithium battery
TEMP Display 12.1 inch (Resolution 800*600)
Measuring range 21.0°C~50.0°C Alarming mode Audible-visual alarm
Measuring accuracy ±0.2°C for range from 25.0°C~45.0°C Networking port Ethernet port
FDA510(k)
Shenzhen Creative Industry Co., Ltd.
Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
12.1” high resolution TFT display with LED backlight Up to 9 waveforms simulteneously display
Arrhythmia analysis and S-T segment measurement 72-hour ECG waveform data storage and recall
Dual IBP with EtCO2 Dual IBP with cardiac output
Protection against defibrillator discharge 2000-hour data trends with graphic and tabular view
Adult/Pediatric/Neonate measurement modes 2000 groups event, ARR and SpO2 storage
Visual and audible alarms; Networking capability Built-in lithium battery; Touch screen optional
NIBP
TEMP
Technique: Oscillometric method
Typical measurement time: <30 seconds (adult cuff) Measuring range: 21.0~50.0℃
NIBP
SYS: 40~275mmHg (Adult) Measuring accuracy: ±0.2℃from 25~45℃
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate)
DIA: 10~210mmHg (Adult) Other Specifications
NIBP measuring range: 10~150mmHg (Pediatric)
Power supply: AC 100V-240V, 50/60Hz, 60VA
10~95mmHg (Neonate)
Built-in lithium battery: 11.1V/4400mAh
MAP: 20~230mmHg (Adult)
NIBP measuring range: 20~165mmHg (Pediatric) Display: 10.4 inch TFT display
20~110mmHg (Neonate) Alarming method: 3 levels audible-visible alarm
NIBP measuring accuracy: Mean difference: ±5mmHg Networking: Ethernet
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode
Auto measuring intervals: 1-480min Standard configuration
ECG, Respiration, SpO2, PR, NIBP, Temperature
SpO2
Options
Technique: Dual-wavelength optical method
Measuring range: 0%~100% Touch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG
Measuring accuracy: Arms is not greater than 2% for Cerebral State Monitoring, Central Monitor Station
SpO2 range 70~100%.
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Low perfusion performance: As low as 0.3%.
Patient Monitor
Accessoy box for standard configuration
cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation
Software upgrade
SpO2 sensor NIBP cuff ECG cable Temperature probe HL7 protocol connect
to hospital system
Bed to bed view via
central monitor station
Technical Specifications
ECG CO2
K12
Patient Monitor
Input dynamic range: ±(0.5mVp~5mVp) Technique: Infrared optical method
Differential input impedance: ≥10MΩ Sampling mode: Sidestream or Mainstream
Bandwidth: 0.05~150Hz (Diagnostic) Measuring range: 0~150mmHg
0.5~40Hz (Monitoring) Measuring accuracy: 0~40mmHg ±2mmHg
1~20Hz (Operation) 41~70mmHg ±5% of reading
CMRR: ≥90dB (Diagnostic) 71~100mmHg ±8% of reading
≥105dB (Monitoring & Operation) 101~150mmHg ±10% of reading
Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto Flow rate: 50ml/min ±10 ml/min (Sidestream)
Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
HR measuring range: 15~350bpm Cerebral State Monitoring (CSM)
HR accuracy: ±1% or ±2bpm, whichever is greate EEG sensitivity: ±400µV
Pacemaker pulse detection and rejection function Noise level: <2µVp-p, <0.4µV rms (1~250Hz)
CMRR: >140dB
Input impedance: >50Mohm
RESP CSI and update: 0-100. filter: 6-42Hz, 1 sec. update
Measuring range: 0~120rpm EMG%: 0-100 (logarithmic)
Measuring accuracy: ±5% or ±2 rpm, whichever is greater filter: 75-85 Hz, 1 sec. update.
BS%: 0-100. filter: 2-42 Hz, 1 sec. update
TEMP
IBP
Measuring range: 21.0~50.0℃
Technique: Strain gauge transducer
Measuring accuracy: ±0.2℃from 25~45℃
Input sensitivity: 5µV/V/mmHg
Measuring range: -50~300mmHg
NIBP
Measuring accuracy: ±2% or ±4mmHg, whichever is greater
Technique: Oscillometric method Measuring positions: ART, RAP, PA, LAP, CVP
Typical measurement time: <30 seconds (adult cuff) ICP, AUXP1, AUXP2
SYS: 40~275mmHg (Adult) Calibration: zero calibrating
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate) Cardiac Output (C.O.)
DIA: 10~210mmHg (Adult) Blood temperature
NIBP measuring range: 10~150mmHg (Pediatric) measuring: range: 23-43℃, accuracy: ±0.5℃
10~95mmHg (Neonate) Injecta temperature
MAP: 20~230mmHg (Adult) measuring: range: 0-20℃, accuracy: ±0.5℃
NIBP measuring range: 20~165mmHg (Pediatric) Measuring range: 0.2~20 L/min
20~110mmHg (Neonate) Measuring accuracy: ±0.2 L/min or ±10%, whichever is greater
NIBP measuring accuracy: Mean difference: ±5mmHg
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode Other Specifications
Auto measuring intervals: 1-480min Power supply: AC 100V-240V, 50/60Hz, 60VA
Built-in lithium battery: 11.1V/4400mAh
Display: 12.1 inch TFT display
SpO2 Alarming method: 3 levels audible-visible alarm
Technique: Dual-wavelength optical method Networking: Ethernet
Measuring range: 0%~100%
Measuring accuracy: Arms is not greater than 2% for Standard configuration
SpO2 range 70~100%. ECG, Respiration, SpO2, PR, NIBP, Temperature
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Options
Low perfusion performance: As low as 0.3%.
2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG
Cardiac Output, Cerebral State Monitoring, CMS, Touch Screen
Patient Monitor
Accessoy box for standard configuration
cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation
Software upgrade
TEMP
IBP
Measuring range: 21.0~50.0℃
Technique: Strain gauge transducer
Measuring accuracy: ±0.2℃from 25~45℃
Input sensitivity: 5µV/V/mmHg
Measuring range: -50~300mmHg
NIBP
Measuring accuracy: ±2% or ±4mmHg, whichever is greater
Technique: Oscillometric method Measuring positions: ART, RAP, PA, LAP, CVP
Typical measurement time: <30 seconds (adult cuff) ICP, AUXP1, AUXP2
SYS: 40~275mmHg (Adult) Calibration: zero calibrating
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate) Cardiac Output (C.O.)
DIA: 10~210mmHg (Adult) Blood temperature
NIBP measuring range: 10~150mmHg (Pediatric) measuring: range: 23-43℃, accuracy: ±0.5℃
10~95mmHg (Neonate) Injecta temperature
MAP: 20~230mmHg (Adult) measuring: range: 0-20℃, accuracy: ±0.5℃
NIBP measuring range: 20~165mmHg (Pediatric) Measuring range: 0.2~20 L/min
20~110mmHg (Neonate) Measuring accuracy: ±0.2 L/min or ±10%, whichever is greater
NIBP measuring accuracy: Mean difference: ±5mmHg
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode Other Specifications
Auto measuring intervals: 1-480min Power supply: AC 100V-240V, 50/60Hz, 60VA
Built-in lithium battery: 11.1V/4400mAh
Display: 15 inch TFT display
SpO2 Alarming method: 3 levels audible-visible alarm
Technique: Dual-wavelength optical method Networking: Ethernet
Measuring range: 0%~100%
Measuring accuracy: Arms is not greater than 2% for Standard configuration
SpO2 range 70~100%. ECG, Respiration, SpO2, PR, NIBP, Temperature
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Options
Low perfusion performance: As low as 0.3%.
Touch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG,
Cardiac Output, Cerebral State Monitoring, Central Monitor Station,
Multi-Gas Monitoring
Patient Monitor
Accessoy box for standard configuration
cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation
Software upgrade
SpO2 sensor NIBP cuff ECG cable Temperature probe HL7 protocol connect
to hospital system
Bed to bed view via
central monitor station
ECG SpO2
Input dynamic range ±0.5mVp~±5mVp SpO2 measuring range 0%~100%
SpO2 measuring accuracy 2% for range from 70% to 100%
HR measuring range 15bpm~350bpm
Low perfusion performance As low as 0.4%
HR measuring accuracy ±1% or ±2bpm, whichever is greater
PR measuring range 0bpm~250bpm
Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 and Auto
PR measuring accuracy ±2bpm or ±2%, whichever is greater
Sweeping speed 6.25, 12.5, 25, 50mm/s
NIBP
Technique Oscillometric Dual IBP(Optional)
Typical measuring time <30 seconds (typical adult cuff) Measuring range -30mmHg ~ 300mmHg
Measuring accuracy ±4mmHg or ±2%, whichever is greater
Measuring range
Systolic pressure Adult 40mmHg~275mmHg
Pediatric 40mmHg~200mmHg Cardiac Output(Opional)
Neonate 40mmHg~135mmHg Measuring range 0.2 ~ 20l/min
Diastolic pressure Adult 10mmHg~210mmHg Measuring accuracy 0.2l/min or 10%, whichever is greater
Pediatric 10mmHg~150mmHg
Blood Temperature(BT) 23°C ~ 45°C
Neonate 10mmHg~95mmHg
Injector Temperature(IT) 0°C ~ 20°C
Mean arterial pressure Adult 20mmHg~230mmHg
Pediatric 20mmHg~165mmHg
Neonate 20mmHg~110mmHg
Measurement accuracy Maximum mean difference: ±5 mmHg Cerebral State Monitoring(Optional)
Maximum standard deviation: 8 mmHg Measuring range
Measurement mode Manual, Auto, STAT Cerebral State Index(CSI) 0~100
Automatic measuring intervals 1~480min
Burst Suppression Ratio(BS%) 0~100%
Measuring accuracy ±5% or ±2rpm, whichever is greater Alarming mode Audible-visual alarm
Networking port Ethernet port
Arrhythmia analysis and S-T segment measurement 72-hour ECG waveform data storage and recall
Dual IBP with EtCO2 Dual IBP with cardiac output
Protection against defibrillator discharge 2000-hour data trends with graphic and tabular view
Adult/Pediatric/Neonate measurement modes 2000 groups event, ARR and SpO2 storage
Visual and audible alarm; Networking capability Built-in lithium battery; Touch screen optional
Shenzhen Creative Industry Co., Ltd. Early warning score | Synchronous12-lead ECG
Add: Floor 5, BLD 9, Balwangxin High-Tech Industrial Park, Songbai Road, Xili Street Tel: + 86 755 2643 3514
Nanshan District, 518110 Shenzhen, China
Fax: + 86 755 2643 0930
https://en.lepumedical.com E-mail: info@creative-sz.com
HD
Minimalist Design
The AI VIEW can be connected to a central monitoring station (CMS) via wired and * 28 types of arrhythmia analysis to monitor patients' heart rate and give accu-
wireless networks and enables bedside observation, as well as direct connection to rate results;
a third-party electronic medical record (EMR) via the HL7 output.
OxyCRG IBP&CO2
Multiple mode options
* EWS (early warning score) is a physiologic
scoring system- RESP, HR, NIBP, level-of-consciousness, For the needs of different scenarios, AI VIEW provides a variety of monitoring
TEMP, etc used by medical service to. modes, which greatly improves the experience for doctors and patients.
AIView VX
manufacturing and marketing high-tech medical devices and equipment. Today, Lepu Medical has
grown into a global leading medical technology company in the fields of cardiovascular interventions, Tablet Patient Monitor
structural heart diseases, cardiac rhythm management, anesthesia and critical care, in vitro diagnostics
The Revolution of Minimalist
and general surgery. Our vision is to develop Lepu Medical into one of the largest supplier on the
platforms of medical device, medicine, healthcare products, healthcare service and new medical mode.
Early Warning Score EWS Intuitive UI Design Portable and unique split design of main
unit and base.
NEWS2, MEWS, CART International Access to most functions in 2 steps
early warning score. or fewer. Innovative NIBP pump valve design.
Ultra-thin main unit with concealed
handle, only 2.2 cm thick.
Independent base design with ergonomic
Thoughtful Design of Built-in Camera Scanner handle, built-in 3 channel thermal printer
Screen Lock
Block accidental touches and easy cleaning.
Quick input patient information. and accessory box.
Ergonomic & Concise Design
Multi-parameter measurement.
3/5/12 leads ECG, Dual TEMP, SPO2, NIBP, RESP, PR & ETCO2.
Advanced SPO2 anti-interference.
Low perfusion SPO2 algorithm with visual alarm.
Multi-Lead ECG algorithm with ST & QT analysis.
Detect and Analyze up to 28 types of arrhythmia event.
Also provide advanced configuration options, such as EtCO2 provide Respironic and
Chinese brand with mainstream and sidestream.
PI 0.6
EWS Early Warning Score
PI 0.2 AIView VX monitors provide a point-of-care EWS tools to assist clinicians or caregivers to
recognize the early signs of clinical deterioration in order to initiate early intervention.
Compliant with National Early Warning Score 2 (NEWS2), Modified Early Warning
Score (MEWS) and Cardiac Arrest Risk Triage Score (CART).
Data Storage Initiative display of scoring value and radar-gram on screen to remind clinicians anytime.
WiFi
AIView VX
CMS EMR
ECG Half-screen View ALL ECG Trace View OxyCRG View
HL7