Technique Oscillometric Input dynamic range ±0.

5mVp~±5mVp
Measurement accuracy Maximum mean difference: ±5 mmHg HR measuring range 15bpm~350bpm
Maximum standard deviation: 8 mmHg HR measuring accuracy ±1% or ±2bpm, whichever is greater
Measuring range HR alarm delay time ≤10s
Systolic pressure Adult 40mmHg~275mmHg Sensitivity selection ×1/4, ×1/2, ×1, ×2
Pediatric 40mmHg~200mmHg Sweeping speed 25mm/s
Neonate 40mmHg~135mmHg
ECG noise level ≤30µVp-p
Diastolic pressure Adult 10mmHg~210mmHg
ECG input loop current ≤0.1µA
Pediatric 10mmHg~150mmHg
Neonate 10mmHg~95mmHg Differential input impedance ≥10MΩ

Mean arterial pressure Adult 20mmHg~230mmHg Common-mode rejection ratio ≥105dB


Pediatric 20mmHg~165mmHg
Neonate 20mmHg~110mmHg

Measurement mode Manual, Auto, STAT


Overpressure protection limit Adult 300mmHg Measuring range 21.0℃~50.0℃
Pediatric 240mmHg
Neonate 150mmHg Measuring accuracy ±0.2℃ for range 25℃~45℃

Transducer Dual-wave length LED Power supply 100~240Vac, 50/60Hz

SpO2 measuring range 35%~100% Built-in battery 11.1V/4400mAh Li-ion battery

SpO 2 measuring accuracy 3% for range from 70% to 100%


Low perfusion performance As low as 0.4%

PC-900SN: SpO2, PR, NIBP monitoring


PC-900SNT: SpO2, PR, NIBP and TEMP monitoring
PC-900SNTE: SpO2, PR, NIBP, TEMP and ECG monitoring
Measuring range 30bpm~240bpm
Measuring accuracy ±2bpm or ±2%, whichever is greater

Nellcor SpO2, Infrared Ear temp probe, Printer, Clamp, Trolley

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
3.5 inch high resolution LCD display of SpO2/ECG waveform

Table top design with user friendly operation

Upto 12000 groups of NIBP, 3000 groups SpO2 and event record

Historic data records can be reviewed in waveform, list or trend graph

Data export and software upgrading can be done by USB cable

Flexible configurations among SpO2, NIBP, TEMP and ECG

Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available

ECG cable NIBP cuff Skin temp probe Infrared ear probe SpO2 sensor
PC-900PLUS
PC-900PLUS

Technique Oscillometric Input dynamic range ±0.5mVp~±5mVp


Measurement accuracy Maximum mean difference: ±5 mmHg HR measuring range 15bpm~350bpm
Maximum standard deviation: 8 mmHg HR measuring accuracy ±1% or ±2bpm, whichever is greater
Measuring range HR alarm delay time ≤10s
Systolic pressure Adult 40mmHg~275mmHg Sensitivity selection ×1/4, ×1/2, ×1, ×2
Pediatric 40mmHg~200mmHg Sweeping speed 25mm/s
Neonate 40mmHg~135mmHg
ECG noise level ≤30µVp-p
Diastolic pressure Adult 10mmHg~210mmHg
ECG input loop current ≤0.1µA
Pediatric 10mmHg~150mmHg
Neonate 10mmHg~95mmHg Differential input impedance ≥10MΩ

Mean arterial pressure Adult 20mmHg~230mmHg Common-mode rejection ratio ≥105dB


Pediatric 20mmHg~165mmHg
Neonate 20mmHg~110mmHg

Measurement mode Manual, Auto, STAT


Measuring range 21.0℃~50.0℃
Overpressure protection limit Adult 300mmHg
Pediatric 240mmHg Measuring accuracy ±0.2℃ for range 25℃~45℃
Neonate 150mmHg

Power supply 100~240Vac, 50/60Hz


Transducer Dual-wave length LED Built-in battery 11.1V/4400mAh Li-ion battery
SpO2 measuring range 35%~100% Display 7 inch touch screen

SpO 2 measuring accuracy 3% for range from 70% to 100%


Low perfusion performance As low as 0.4%

PC-900SN: SpO2, PR, NIBP monitoring


PC-900SNT: SpO2, PR, NIBP and TEMP monitoring
PC-900SNTE: SpO2, PR, NIBP, TEMP and ECG monitoring
Measuring range 30bpm~240bpm
Measuring accuracy ±2bpm or ±2%, whichever is greater

Nellcor SpO2, Infrared Ear temp probe, Printer, Clamp, Trolley

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
PLUS

7 inch touch screen of SpO2/ECG waveform and numeric display

Table top design with user friendly operation

Upto 12000 groups of NIBP, 3000 groups SpO2 and event record

Historic data records can be reviewed in waveform, list or trend graph

Data export and software upgrading can be done by USB cable

Flexible configurations among SpO2, NIBP, TEMP and ECG

Nellcor compatible SpO2, SunTech NIBP and Infrared ear probe available

ECG cable NIBP cuff Skin temp probe Infrared ear probe SpO2 sensor
Technique Oscillometric method Input dynamic range ±(0.5mVp~5mVp)
Pressure measuring range 0 mmHg~300mmHg Differential input impedance ≥10MΩ
Typical measurement time <30 seconds (adult cuff) Frequency response 1~20Hz (Operation)
SYS measuring range Adult: 40mmHg~275mmHg 0.67~40Hz (Monitoring)
Pediatric: 40mmHg~200mmHg 0.05~150Hz (Diagnostic)
Neonate: 40mmHg~135mmHg CMRR ≥90dB (Diagnostic)
MAP measuring range Adult: 20mmHg~230mmHg ≥105dB (Monitoring & Operation)
Pediatric: 20mmHg~165mmHg
Neonate: 20mmHg~110mmHg Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 an d Auto
Sweeping speed 6.25, 12.5, 25, 50mm/s, tolerance: ±10%
DIA measuring range Adult: 10mmHg~210mmHg
Pediatric: 10mmHg~150mmHg HR display range 15bpm~350bpm
Neonate: 10mmHg~95mmHg HR display accuracy ±1% or ±2bpm, whichever is greater
Measuring accuracy Maximum mean difference: ±5 mmHg
Maximum Standard deviation: 8 mmHg
Measuring mode Manual, Auto, STAT

Measuring range 0~120rpm


Measuring accuracy ±5% or ±2 rpm, whichever is greater

Measuring range 21℃~50℃ Power supply AC 100V~240V, 50/60Hz


Measuring accuracy ±0.2℃ for range from 25℃~45℃ Lithium battery 14.4V/2200mAh
Probe type Skin/Rectal probe Display 7 inch TFT display, 800x480 resolution
Alarming 3 levels audible-visible alarm
Networking Ethernet

Transducer Dual-wavelength LED


SpO2 measuring range 0%~100%
SpO2 measuring accuracy 2% for range from 70%~100%
SpO2, PR, NIBP, ECG, RESP, TEMP
Low perfusion performance As low as 0.3%
PR measuring range 0bpm~250bpm
PR measuring accuracy ±2bpm or ±2%, whichever is greater.
Central monitor system, Trolley, Wall mount, Clamp, Carry bag

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
7” high resolution(800x480) color display 12000 groups NIBP list and 2000 groups arrhythmia event
Portable design with optional carry bag 60-hour ECG waveform and 1000-hour trend data recall
Arrhythmia analysis and ST segment analysis Pitch tone selection and brighteness adjustment
Protection against defibrillator interference Pacemaker pulse detection
Data export via USB cable with software Central monitoring system optional

Central monitor station Carry Bag Data Management Software Trolley

Thermometer
Multi-parameter Patient Monitor

NIBP ECG
Technique Oscillometric Input dynamic range ±0.5mVp~±5mVp
Typical measuring time <30 seconds (typical adult cuff) HR measuring range 15bpm~350bpm
Initial cuff inflation pressure Adult <175mmHg HR measuring accuracy ±1% or ±2bpm, whichever is greater
Pediatric <135mmHg HR alarm delay time ≤10s
Neonate <65mmHg
Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 and Auto
Overpressure protection limit Adult 300mmHg Sweeping speed 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
Pediatric 240mmHg
Neonate 150mmHg ECG noise level ≤30µVp-p
ECG input loop current ≤0.1µA
Measuring range
Differential input impedance ≥10Mohm
Systolic pressure Adult 40mmHg~275mmHg
Common-mode rejection ratio ≥105dB (Monitoring mode)
Pediatric 40mmHg~200mmHg
(CMRR) ≥89dB (Diagnostic mode)
Neonate 40mmHg~135mmHg
Diastolic pressure Adult 10mmHg~210mmHg Time constant ≥0.3s (Monitoring mode)
Pediatric 10mmHg~150mmHg ≥3.2s (Diagnostic mode)
Neonate 10mmHg~95mmHg
Mean arterial pressure Adult 20mmHg~230mmHg
Pediatric 20mmHg~165mmHg
RESP
Neonate 20mmHg~110mmHg RR measuring range 0rpm~120rpm
RR measuring accuracy ±5% or ±2rpm, whichever is greater
Measurement accuracy Maximum mean difference: ±5 mmHg
Maximum standard deviation: 8 mmHg
Measurement mode Manual, Auto, STAT
Others
Automatic measuring intervals 1~480min Power supply 100~240Vac, 50/60Hz
Built-in battery 4400mAh Lithium battery
TEMP Display 12.1 inch (Resolution 800*600)
Measuring range 21.0°C~50.0°C Alarming mode Audible-visual alarm
Measuring accuracy ±0.2°C for range from 25.0°C~45.0°C Networking port Ethernet port

SpO2 Standard Configuration


Transducer Dual-wave length LED ECG, RESP, SpO2, NIBP, TEMP, PR
SpO2 measuring range 0%~100%
SpO2 measuring accuracy 2% for range from 70% to 100% Option
Low perfusion performance As low as 0.4% 2-IBP, EtCO 2 , Nellcor SpO 2 , SunTech NIBP, Cardiac Output, Built-in printer,
PR measuring range 0bpm~250bpm Cerebral State Monitoring, Central monitoring system, Touch screen
PR measuring accuracy ±2bpm or ±2%, whichever is greater

FDA510(k)
Shenzhen Creative Industry Co., Ltd.
Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
12.1” high resolution TFT display with LED backlight Up to 9 waveforms simulteneously display

Arrhythmia analysis and S-T segment measurement 72-hour ECG waveform data storage and recall
Dual IBP with EtCO2 Dual IBP with cardiac output
Protection against defibrillator discharge 2000-hour data trends with graphic and tabular view
Adult/Pediatric/Neonate measurement modes 2000 groups event, ARR and SpO2 storage
Visual and audible alarms; Networking capability Built-in lithium battery; Touch screen optional

Cerebral state monitoring 9 waveform display

Skin temperature NIBP cuff ECG leadwire SpO 2 sensor


Technical Specifications
ECG CO2
K10
Patient Monitor
Input dynamic range: ±(0.5mVp~5mVp) Technique: Infrared optical method
Differential input impedance: ≥10MΩ Sampling mode: Sidestream or Mainstream
Bandwidth: 0.05~150Hz (Diagnostic) Measuring range: 0~150mmHg
0.5~40Hz (Monitoring) Measuring accuracy: 0~40mmHg ±2mmHg
1~20Hz (Operation) 41~70mmHg ±5% of reading
CMRR: ≥90dB (Diagnostic) 71~100mmHg ±8% of reading
≥105dB (Monitoring & Operation) 101~150mmHg ±10% of reading
Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto Flow rate: 50ml/min ±10 ml/min (Sidestream)
Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
HR measuring range: 15~350bpm
HR accuracy: ±1% or ±2bpm, whichever is greate IBP
Pacemaker pulse detection and rejection function
Technique: Strain gauge transducer
Input sensitivity: 5µV/V/mmHg
Measuring range: -50~300mmHg
RESP
Measuring accuracy: ±2% or ±4mmHg, whichever is greater
Measuring range: 0~120rpm Measuring positions: ART, RAP, PA, LAP, CVP
Measuring accuracy: ±5% or ±2 rpm, whichever is greater ICP, AUXP1, AUXP2
Calibration: zero calibrating

NIBP
TEMP
Technique: Oscillometric method
Typical measurement time: <30 seconds (adult cuff) Measuring range: 21.0~50.0℃
NIBP
SYS: 40~275mmHg (Adult) Measuring accuracy: ±0.2℃from 25~45℃
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate)
DIA: 10~210mmHg (Adult) Other Specifications
NIBP measuring range: 10~150mmHg (Pediatric)
Power supply: AC 100V-240V, 50/60Hz, 60VA
10~95mmHg (Neonate)
Built-in lithium battery: 11.1V/4400mAh
MAP: 20~230mmHg (Adult)
NIBP measuring range: 20~165mmHg (Pediatric) Display: 10.4 inch TFT display
20~110mmHg (Neonate) Alarming method: 3 levels audible-visible alarm
NIBP measuring accuracy: Mean difference: ±5mmHg Networking: Ethernet
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode
Auto measuring intervals: 1-480min Standard configuration
ECG, Respiration, SpO2, PR, NIBP, Temperature

SpO2
Options
Technique: Dual-wavelength optical method
Measuring range: 0%~100% Touch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG
Measuring accuracy: Arms is not greater than 2% for Cerebral State Monitoring, Central Monitor Station
SpO2 range 70~100%.
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Low perfusion performance: As low as 0.3%.

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
0123
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
K10 10.4” display with LED backlight
9-waveform on screen
360-degree visible indicator
with 3-level alarm

Patient Monitor
Accessoy box for standard configuration

Li-ion battery up to 4 hours Integral 3-channel


continuous monitoring thermal recorder

Parameter case for optional parameters

Features Comprehensive calculations for clinical application

10.4” high resolution display 9 traces on-screen waveforms


Touch screen optional and maximal up to 13
* Hemodynamics calculation
* Respiration calculation

User customized NIBP measuring


* Oxygenation calculation

cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation

Versatile clinical calculations for HL7 protocol, Bed to bed view


application convenience and 12-lead ECG available

Software upgrade

SpO2 sensor NIBP cuff ECG cable Temperature probe HL7 protocol connect
to hospital system
Bed to bed view via
central monitor station
Technical Specifications
ECG CO2
K12
Patient Monitor
Input dynamic range: ±(0.5mVp~5mVp) Technique: Infrared optical method
Differential input impedance: ≥10MΩ Sampling mode: Sidestream or Mainstream
Bandwidth: 0.05~150Hz (Diagnostic) Measuring range: 0~150mmHg
0.5~40Hz (Monitoring) Measuring accuracy: 0~40mmHg ±2mmHg
1~20Hz (Operation) 41~70mmHg ±5% of reading
CMRR: ≥90dB (Diagnostic) 71~100mmHg ±8% of reading
≥105dB (Monitoring & Operation) 101~150mmHg ±10% of reading
Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto Flow rate: 50ml/min ±10 ml/min (Sidestream)
Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
HR measuring range: 15~350bpm Cerebral State Monitoring (CSM)
HR accuracy: ±1% or ±2bpm, whichever is greate EEG sensitivity: ±400µV
Pacemaker pulse detection and rejection function Noise level: <2µVp-p, <0.4µV rms (1~250Hz)
CMRR: >140dB
Input impedance: >50Mohm
RESP CSI and update: 0-100. filter: 6-42Hz, 1 sec. update
Measuring range: 0~120rpm EMG%: 0-100 (logarithmic)
Measuring accuracy: ±5% or ±2 rpm, whichever is greater filter: 75-85 Hz, 1 sec. update.
BS%: 0-100. filter: 2-42 Hz, 1 sec. update

TEMP
IBP
Measuring range: 21.0~50.0℃
Technique: Strain gauge transducer
Measuring accuracy: ±0.2℃from 25~45℃
Input sensitivity: 5µV/V/mmHg
Measuring range: -50~300mmHg
NIBP
Measuring accuracy: ±2% or ±4mmHg, whichever is greater
Technique: Oscillometric method Measuring positions: ART, RAP, PA, LAP, CVP
Typical measurement time: <30 seconds (adult cuff) ICP, AUXP1, AUXP2
SYS: 40~275mmHg (Adult) Calibration: zero calibrating
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate) Cardiac Output (C.O.)
DIA: 10~210mmHg (Adult) Blood temperature
NIBP measuring range: 10~150mmHg (Pediatric) measuring: range: 23-43℃, accuracy: ±0.5℃
10~95mmHg (Neonate) Injecta temperature
MAP: 20~230mmHg (Adult) measuring: range: 0-20℃, accuracy: ±0.5℃
NIBP measuring range: 20~165mmHg (Pediatric) Measuring range: 0.2~20 L/min
20~110mmHg (Neonate) Measuring accuracy: ±0.2 L/min or ±10%, whichever is greater
NIBP measuring accuracy: Mean difference: ±5mmHg
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode Other Specifications
Auto measuring intervals: 1-480min Power supply: AC 100V-240V, 50/60Hz, 60VA
Built-in lithium battery: 11.1V/4400mAh
Display: 12.1 inch TFT display
SpO2 Alarming method: 3 levels audible-visible alarm
Technique: Dual-wavelength optical method Networking: Ethernet
Measuring range: 0%~100%
Measuring accuracy: Arms is not greater than 2% for Standard configuration
SpO2 range 70~100%. ECG, Respiration, SpO2, PR, NIBP, Temperature
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Options
Low perfusion performance: As low as 0.3%.
2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG
Cardiac Output, Cerebral State Monitoring, CMS, Touch Screen

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
0123
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
K12 12.1” display with LED backlight
9-waveform on screen
360-degree visible indicator
with 3-level alarm

Patient Monitor
Accessoy box for standard configuration

Li-ion battery up to 4 hours Integral 3-channel


continuous monitoring thermal recorder

Parameter case for optional parameters

Features Comprehensive calculations for clinical application

12.1” high resolution display 9 traces on-screen waveforms


Touch screen optional and maximal up to 13
* Hemodynamics calculation
* Respiration calculation

User customized NIBP measuring


* Oxygenation calculation

cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation

Versatile clinical calculations for HL7 protocol, Bed to bed view


application convenience and 12-lead ECG available

Software upgrade

SpO2 sensor NIBP cuff ECG cable emperature


T probe HL7 protocol connect
to hospital system
Bed to bed view via
central monitor station
Technical Specifications
ECG CO2
K15
Patient Monitor
Input dynamic range: ±(0.5mVp~5mVp) Technique: Infrared optical method
Differential input impedance: ≥10MΩ Sampling mode: Sidestream or Mainstream
Bandwidth: 0.05~150Hz (Diagnostic) Measuring range: 0~150mmHg
0.5~40Hz (Monitoring) Measuring accuracy: 0~40mmHg ±2mmHg
1~20Hz (Operation) 41~70mmHg ±5% of reading
CMRR: ≥90dB (Diagnostic) 71~100mmHg ±8% of reading
≥105dB (Monitoring & Operation) 101~150mmHg ±10% of reading
Sensitivity selection: ×1/4, ×1/2, ×1, ×2, ×4 and Auto Flow rate: 50ml/min ±10 ml/min (Sidestream)
Sweeping speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
HR measuring range: 15~350bpm Cerebral State Monitoring (CSM)
HR accuracy: ±1% or ±2bpm, whichever is greate EEG sensitivity: ±400µV
Pacemaker pulse detection and rejection function Noise level: <2µVp-p, <0.4µV rms (1~250Hz)
CMR R: >140dB
Input impedance: >50Mohm
RESP CSI and update: 0-100. filter: 6-42Hz, 1 sec. update
Measuring range: 0~120rpm EMG%: 0-100 (logarithmic)
Measuring accuracy: ±5% or ±2 rpm, whichever is greater filter: 75-85 Hz, 1 sec. update.
BS%: 0-100. filter: 2-42 Hz, 1 sec. update

TEMP
IBP
Measuring range: 21.0~50.0℃
Technique: Strain gauge transducer
Measuring accuracy: ±0.2℃from 25~45℃
Input sensitivity: 5µV/V/mmHg
Measuring range: -50~300mmHg
NIBP
Measuring accuracy: ±2% or ±4mmHg, whichever is greater
Technique: Oscillometric method Measuring positions: ART, RAP, PA, LAP, CVP
Typical measurement time: <30 seconds (adult cuff) ICP, AUXP1, AUXP2
SYS: 40~275mmHg (Adult) Calibration: zero calibrating
NIBP measuring range: 40~200mmHg (Pediatric)
40~135mmHg (Neonate) Cardiac Output (C.O.)
DIA: 10~210mmHg (Adult) Blood temperature
NIBP measuring range: 10~150mmHg (Pediatric) measuring: range: 23-43℃, accuracy: ±0.5℃
10~95mmHg (Neonate) Injecta temperature
MAP: 20~230mmHg (Adult) measuring: range: 0-20℃, accuracy: ±0.5℃
NIBP measuring range: 20~165mmHg (Pediatric) Measuring range: 0.2~20 L/min
20~110mmHg (Neonate) Measuring accuracy: ±0.2 L/min or ±10%, whichever is greater
NIBP measuring accuracy: Mean difference: ±5mmHg
Standard deviation: 8mmHg
NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode Other Specifications
Auto measuring intervals: 1-480min Power supply: AC 100V-240V, 50/60Hz, 60VA
Built-in lithium battery: 11.1V/4400mAh
Display: 15 inch TFT display
SpO2 Alarming method: 3 levels audible-visible alarm
Technique: Dual-wavelength optical method Networking: Ethernet
Measuring range: 0%~100%
Measuring accuracy: Arms is not greater than 2% for Standard configuration
SpO2 range 70~100%. ECG, Respiration, SpO2, PR, NIBP, Temperature
PR measuring range: 30~250bpm
PR measuring accuracy: ±2bpm or ±2%, whichever is greater
Options
Low perfusion performance: As low as 0.3%.
Touch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG,
Cardiac Output, Cerebral State Monitoring, Central Monitor Station,
Multi-Gas Monitoring

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
0123
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
K15 15” display with LED backlight
9-waveform on screen
360-degree visible indicator
with 3-level alarm

Patient Monitor
Accessoy box for standard configuration

Li-ion battery up to 4 hours Integral 3-channel


continuous monitoring thermal recorder

Parameter case for optional parameters

Features Comprehensive calculations for clinical application

15” high resolution display 9 traces on-screen waveforms


Touch screen optional and maximal up to 13
* Hemodynamics calculation
* Respiration calculation

User customized NIBP measuring


* Oxygenation calculation

cycles up to 5-phase
Data export and software upgrade * Drug concentration calculation
* Renal function calculation

Versatile clinical calculations for HL7 protocol, Bed to bed view


application convenience and 12-lead ECG available

Software upgrade

SpO2 sensor NIBP cuff ECG cable Temperature probe HL7 protocol connect
to hospital system
Bed to bed view via
central monitor station
ECG SpO2
Input dynamic range ±0.5mVp~±5mVp SpO2 measuring range 0%~100%
SpO2 measuring accuracy 2% for range from 70% to 100%
HR measuring range 15bpm~350bpm
Low perfusion performance As low as 0.4%
HR measuring accuracy ±1% or ±2bpm, whichever is greater
PR measuring range 0bpm~250bpm
Sensitivity selection ×1/4, ×1/2, ×1, ×2, ×4 and Auto
PR measuring accuracy ±2bpm or ±2%, whichever is greater
Sweeping speed 6.25, 12.5, 25, 50mm/s

Common-mode rejection ratio ≥105dB (Monitoring mode)


EtCO2(Optional)
(CMRR) ≥89dB (Diagnostic mode)
Measuring range 0mmHg ~ 150mmHg
Measuring accuracy ±2mmHg for range from 0~40mmHg

NIBP
Technique Oscillometric Dual IBP(Optional)
Typical measuring time <30 seconds (typical adult cuff) Measuring range -30mmHg ~ 300mmHg
Measuring accuracy ±4mmHg or ±2%, whichever is greater
Measuring range
Systolic pressure Adult 40mmHg~275mmHg
Pediatric 40mmHg~200mmHg Cardiac Output(Opional)
Neonate 40mmHg~135mmHg Measuring range 0.2 ~ 20l/min
Diastolic pressure Adult 10mmHg~210mmHg Measuring accuracy 0.2l/min or 10%, whichever is greater
Pediatric 10mmHg~150mmHg
Blood Temperature(BT) 23°C ~ 45°C
Neonate 10mmHg~95mmHg
Injector Temperature(IT) 0°C ~ 20°C
Mean arterial pressure Adult 20mmHg~230mmHg
Pediatric 20mmHg~165mmHg
Neonate 20mmHg~110mmHg
Measurement accuracy Maximum mean difference: ±5 mmHg Cerebral State Monitoring(Optional)
Maximum standard deviation: 8 mmHg Measuring range
Measurement mode Manual, Auto, STAT Cerebral State Index(CSI) 0~100
Automatic measuring intervals 1~480min
Burst Suppression Ratio(BS%) 0~100%

Forehead EMG Ratio(EMG%) 0~100%


Temperature Measuring accuracy: ±2 or ±10%, whichever is greater
Measuring range 21.0°C~50.0°C
Measuring accuracy ±0.2°C for range from 25.0°C~45.0°C
Others
Power supply 100~240Vac, 50/60Hz
Respiration Rate Built-in battery 4400mAh Lithium battery

Measuring range 0rpm~120rpm Display 12.1 inch TFT touch screen

Measuring accuracy ±5% or ±2rpm, whichever is greater Alarming mode Audible-visual alarm
Networking port Ethernet port

Shenzhen Creative Industry Co., Ltd.


Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street
Nanshan District, 518110 Shenzhen, PEOPLE’S REPUBLIC OF CHINA
Tel: +86 755 2643 3514 Fax: +86 755 2643 0390
www.creative-sz.com info@creative-sz.com
12.1” high resolution TFT display with LED backlight Up to 9 waveforms simulteneously display

Arrhythmia analysis and S-T segment measurement 72-hour ECG waveform data storage and recall
Dual IBP with EtCO2 Dual IBP with cardiac output
Protection against defibrillator discharge 2000-hour data trends with graphic and tabular view
Adult/Pediatric/Neonate measurement modes 2000 groups event, ARR and SpO2 storage

Visual and audible alarm; Networking capability Built-in lithium battery; Touch screen optional

Cerebral state monitoring 9 waveform display

SpO2 sensor NIBP cuff ECG leadwire Skin temperature


Lepu Medical Technology ( Beijing ) Co., Ltd was established in 1999. It is specialized in developing,
manufacturing and marketing high-tech medical devices and equipment. Today, Lepu Medical has
grown into a global leading medical technology company in the fields of cardiovascular interventions,
structural heart diseases, cardiac rhythm management, anesthesia and critical care, in vitro diagnostics
and general surgery. Our vision is to develop Lepu Medical into one of the largest supplier on the
platforms of medical device, medicine, healthcare products, healthcare service and new medical mode.

AIView V10/V12 Patient Monitor

Shenzhen Creative Industry Co., Ltd. Early warning score | Synchronous12-lead ECG
Add: Floor 5, BLD 9, Balwangxin High-Tech Industrial Park, Songbai Road, Xili Street Tel: + 86 755 2643 3514
Nanshan District, 518110 Shenzhen, China
Fax: + 86 755 2643 0930
https://en.lepumedical.com E-mail: info@creative-sz.com

Science and technology cherish life


The new generation of Lepu AI intelligent multi-parameter monitor, with hardware
ergonomic design and software flat interactive design, will meet the different
Flat design
needs in different departments of hospital. It changes the working mode of the
medical staffs and brings them unprecedented use experience. Clear and concise menu, no difficulty for doctors
and nurses to learn, any routine operation can
be completed in 2 steps at most, significantly
improving the efficiency of use.

Synchronous 12-lead ECG


Support 3.5.6.10 multiple electrode connections, up to 12 leads of ECG can be displayed
on the same screen, which is equivalent to equipping a professional ECG machine.

HD

GFF High resolution Fanless design Anti electric- IPX2


touch screen 1920*1080 shock CF type waterproof

Minimalist Design

* Capacitive touch screen, easy and convenient to operate;


* Ultra-high resolution provides clearer picture quality and better visual effects;
* Tilted screen ergonomically designed for better bedside viewing;
* Dual alarm lights facilitate long-distance identification of alarm conditions;
* Fanless design reduces noise to provide a better monitoring environment.
Excellent IT workflow Powerful clinical features

The AI VIEW can be connected to a central monitoring station (CMS) via wired and * 28 types of arrhythmia analysis to monitor patients' heart rate and give accu-
wireless networks and enables bedside observation, as well as direct connection to rate results;
a third-party electronic medical record (EMR) via the HL7 output.

* ST&QT analysis, HRV analysis, HRV analysis time ≥ 5min;

* With the unique anti-motion algorithm blood oxygen technology MoveOXY,


which ensures that accurate measurements in the process of patient movement
and in the case of low temperature or weak peripheral circulation;
WiFi
LAN
AIView V10/V12 CMS EMR
* Optional with 2-IBP,ETCO2,Nellcor SpO2, Suntech NIBP,to meet a greater range
of clinical needs.
HL7

ST&QT analysis HRV analysis

OxyCRG IBP&CO2
Multiple mode options
* EWS (early warning score) is a physiologic
scoring system- RESP, HR, NIBP, level-of-consciousness, For the needs of different scenarios, AI VIEW provides a variety of monitoring
TEMP, etc used by medical service to. modes, which greatly improves the experience for doctors and patients.

Quickly determine the degree of illness of a patient;


Predict the development trend of diseases;
Night Mode Privacy Mode
Assist medical staff to understand the overall situation
of patients from the vast and complex monitoring data;
Prompt nursing staff to request a medical review and
take early action.
Intubation Mode Monitoring Paused Mode

Portable and easy to move


* Handle integrated design, easy to transfer and store accessories.

* A variety of mounting solutions to meet the needs of


different scenarios in the hospital.

Wall mounted Trolley


Lepu Medical Technology ( Beijing ) Co., Ltd was established in 1999. It is specialized in developing,

AIView VX
manufacturing and marketing high-tech medical devices and equipment. Today, Lepu Medical has
grown into a global leading medical technology company in the fields of cardiovascular interventions, Tablet Patient Monitor
structural heart diseases, cardiac rhythm management, anesthesia and critical care, in vitro diagnostics
The Revolution of Minimalist
and general surgery. Our vision is to develop Lepu Medical into one of the largest supplier on the
platforms of medical device, medicine, healthcare products, healthcare service and new medical mode.

Lepu Medical Technology (Beijing) Co., Ltd


Add: Building 7, No. 37 Chaoqian Rd., Changping District,Beijing,China 102200
Tel: + 86 755 2643 3514 Fax: + 86 755 2643 0930 E-mail: info@creative-sz.com
https://en.lepumedical.com

Science and technology cherish life


AIView VX has a minimalist and unique design, advanced technology and innovative
patented design to streamline caregivers’ workflow and improve work efficiency.
Various combinations can meet the needs of different clinical scenarios, such as
general ward, emergency room, recovery unit etc.

Unibody Accessory design with Patent


Multifunctional adapter equipped with 3/5 ECG leads, SPO2 sensor and 2 TEMP sensor
in one.

Ultra-thin Tablet Design Ultra-light and Tablet


Innovative NIBP pump valve design, Design
≤2.2 cm thickness. Main unit ≤1.35KG (Include Battery).

Unibody Accessory Design High Resolution Screen

Ergonomic & Concise Design


HD
3/5/12 leads ECG leads, SPO2 sensor and 1280*800 pixel, 10.1’’ Touch capacitive
2 TEMP sensor are all-in-one design. display.

Unibody accessory design.

Early Warning Score EWS Intuitive UI Design Portable and unique split design of main
unit and base.
NEWS2, MEWS, CART International Access to most functions in 2 steps
early warning score. or fewer. Innovative NIBP pump valve design.
Ultra-thin main unit with concealed
handle, only 2.2 cm thick.
Independent base design with ergonomic
Thoughtful Design of Built-in Camera Scanner handle, built-in 3 channel thermal printer
Screen Lock
Block accidental touches and easy cleaning.
Quick input patient information. and accessory box.
Ergonomic & Concise Design
Multi-parameter measurement.
3/5/12 leads ECG, Dual TEMP, SPO2, NIBP, RESP, PR & ETCO2.
Advanced SPO2 anti-interference.
Low perfusion SPO2 algorithm with visual alarm.
Multi-Lead ECG algorithm with ST & QT analysis.
Detect and Analyze up to 28 types of arrhythmia event.

Also provide advanced configuration options, such as EtCO2 provide Respironic and
Chinese brand with mainstream and sidestream.

SpO2 low perfusion

PI 0.6
EWS Early Warning Score
PI 0.2 AIView VX monitors provide a point-of-care EWS tools to assist clinicians or caregivers to
recognize the early signs of clinical deterioration in order to initiate early intervention.

Compliant with National Early Warning Score 2 (NEWS2), Modified Early Warning
Score (MEWS) and Cardiac Arrest Risk Triage Score (CART).
Data Storage Initiative display of scoring value and radar-gram on screen to remind clinicians anytime.

1000 240 140 2000


cases hours hours sets

Event Trends and Table Full disclosure waveform NIBP measurement


of all parameters
Ergonomic & Concise Design Excellent and Valuable IT Solutions
AIView VX monitors can be connected to a central monitoring station (CMS) via wire-
less networks and enable bedside observation, as well as directly connect to a third
party Electronic Medical Record (EMR) by HL7 output.
Caregivers could set parameter, alarming threshold, patient ID, data management
and alarm event on CMS directly to help to streamline the workflow and improve
efficiency.
General View Big Numerical View Mini Trend View

WiFi
AIView VX
CMS EMR
ECG Half-screen View ALL ECG Trace View OxyCRG View

HL7

Flexible Mounting and Carrying Solutions


Split design of main unit and base are suitable for various clinical scenarios.

- Concealed handle and Accessory box

- The wide range of mounting solution