You are on page 1of 6

TS3-1 M to mA metering PCB TS3-8 COM

TS3-2 XL Com. Of R2 LF TS3-9 ST. By Hand Sw


TS3-3 A1 TS1-1 TS3-10 X-Ray
TS3-4 XS Com. Of R1 SF TS3-11 0 of Tube Rotor
TS3-5 R.L. TS3-12 I of Tube Rotor
TS3-6 R.L./ Bucky Com. TS3-13 II of Tube Rotor
TS3-7 Bucky TS3-14 ---TS3-17
TS3-15 LBD thru’ fuse
TS3-16 LBD 12V
KI
A2 230 V A1
Booster Tr.
B1 2

3
C1

30 ohms
D2 Booster Resistance

ATTS2 K1 To VC Comn
SwP com1
K2

K2
400 V P2

SwP com2
P1

D2 T2
ON Sw
ATTS5 X_Ray Contr

Com of ON &OFF SW

OFF Sw

K3
230 V

Ad contact
L2 L1 N
Minor KVP Com
X-Ray
Contr

P1

Phase2
Phase 1
Neutral

9415080083
300 mA Auto Transformer 2 Phase input 01.04.10
Main Wdg. with 12 and 14 swg 110x50 mm Bobbin Volt/ Turn 1.47V
AT no. Turns Total Turns Volt Connected to
1 0 0 -205.8 Phase 1- L1
2 17 17 -180.8 -------------------------------------------- T1
3 17 34 -155.8
4 90 124 -23.5
5 16 140 0 Volts A1,TS3-1 T2 TS1-1
6 16 156 23.5
7 21 177 54.4 TS1-2
8 21 198 85.3 RK1 T3
9 11 209 101.43 RK2 T4
10 11 220 117.6 RK3 T5
11 11 231 133.8 RK4 T6
12 11 242 150 RK5 T7
13 12 254 167.6 RK6 T8 TS1-3
14 12 266 185.2 RK7 VC15 T9
15 3 269 189.6 VC14 T10
16 3 272 194 VC13 T11
17 3 275 198.45 RK8 VC12 T12
18 3 278 202.9 VC11 T13
19 3 281 207.3 VC10 T14
20 3 284 211.7 RK9 VC9 T15
21 3 287 216.1 VC8 T16
22 3 290 220.5 VC7 T17
23 3 293 224.9 RK10 VC6 T18
24 3 296 229.3 VC5 T19 TS1-4
25 3 299 233.7 VC4 T20
26 3 302 238.14 Rk11 VC3 T21
27 3 305 242.55 VC2 T22
28 3 308 247 VC1 T23
29 3 311 251.4 RK12 T24 TS1-5
AUXILLARY-1 with 12 or 13 swg
AT no. Turns Total Turns Volt Connected to
30 0 0 0 mk1
31 1 1 1.47 mk2
32 1 2 2.95 mk3
33 1 3 4.4 mk4
34 1 4 5.9 mk5
35 1 5 7.4 mk6
36 1 6 8.8 mk7
37 1 7 10.22 mk8
38 1 8 11.7 mk9
39 1 9 13.3 mk10
40 1 10 14.7 mk11
41 1 11 16.2 mk12
AUXILLARY-2 22 swg LBD Connections
44 0 0 13.2 TS3-16
43 9 9 0 TS3-15 thru, fuse
AUXILLARY -3 24 swgTimer
46 0 0 0 TS1-6
45 13 13 19.9 TS1-7
AUXILLARY-4 24 swg mAS
48 0 0 0 TS1-8
47 7 7 9.8 TS1-9
AUXILLARY -5 24 swg
49 0 0 -8.8 TS1-10
50 6 6 0 TS1-11 5V dc supply for kvp
51 6 6 8.8 TS1-12
TIMER PCB PIN DATA

Pin no. 1 - 0 V DC unregulated TS2-1 Pin no. 10 K3 AB TS2-10


Pin no. 2 - 19V AC Thru’ fuse F-3 TS1-7 Pin no. 11 -02 sec. Contact, K3 B TS2-11
Pin no. 3 - 0V AC TS1-6 Pin no. 12 ----------- Open
Pin no. 4 - K4 B, Fuse F-4in TS2-4 Pin no. 13 K9 RE Cl-1 TS2-12
Pin no. 5 - K3 A, 0V DC regulated TS2-5 Pin no. 14 K10 RE Cl-1 ----- ------TS2-13
Pin no. 6 - K4 A TS2-6 Pin no. 15 G com. of Tech. Sel. TS2-14
Pin no. 7 - K4 AB TS2-7 Pin no. 16 K9 RE Cl-1 ----------TS2-12
Pin no. 8 - K12 RE Coil Cl-1 TS2-8 Pin no. 17 H1 of Tech. Sel , kVp PCB pin no. 6
Pin no. 9 -Com. of timer RT Chain TS2-9 Pin no. 18 K11 Cl-1

TERMINAL STRIP DATA


TS1-1 -0 V AC TS1-7 -19.9 V AC- Timer Supply

TS1-2 -53 V AC TS1-8 -0 V AC, -MAS PCB


TS1-3 -166 V AC- Space Charge primary TS1-9 -9 V AC, -MAS PCB

TS1-4 - 230 V AC Volt meter TS1-10 - kVp PCB


TS1-5 -250 V AC- To X-Ray Cont. CL1 TS1-11 -kVp PCB 8-0-8 for KVP
TS1-6 -0 V AC - Timer Supply TS1-12 -kVp PCB Supply

TS2-1 -Pin no. 1 of Timer PCB TS2-10 -K3 AB, Pin no. 10 of Timer PCB
TS2-2 - TS2-11 -Pin 11, 02 sec. Contact Of Timer Sw.,
TS2-3 - TS2-12 -K11B Timer pin no. 13, 16
TS2-4 - Fuse F-4 out, K4 B TS2-13 -K10 Cl 1 Timer pin no. 14
TS2-5 - K3A, 0V DC reg. Pin 5 of Timer PCB TS2-14 -G com of tech Sel.
TS2-6 -Pin no. 6 of Timer PCB, K4 A TS2-15 -0 V ac TS1-1 Timer Pin no. 15 Fuse F1
TS2-7 -Pin 7 of Timer PCB, K4 AB TS2-16 -230 V ac TS1-4 --- Fuse F2
TS2-8 - K12 Cl-1, Pin 8 of Timer PCB
TS2-9 -Com. Of Timer Sw. RT Chain

TS3-1 M to mA metering PCB TS3-8 COM


TS3-2 XL Com. Of R2 LF TS3-9 ST. By Hand Sw
TS3-3 A1 TS1-1 TS3-10 X-Ray
TS3-4 XS Com. Of R1 SF TS3-11 0 of Tube Rotor
TS3-5 R.L. TS3-12 I of Tube Rotor
TS3-6 R.L./ Bucky Com. TS3-13 II of Tube Rotor
TS3-7 Bucky TS3-14 ---TS3-17
TS3-15 LBD thru’ fuse
TS3-16 LBD 12V

300 mA Space Charge Transformer Bobbin 15 no. 1”

Primary 30 swg 900 Turns


1 0 0 TS1-5
2 900 900 Com. Of Minor kVp Sw.
2ndary 26 swg.
1 0 0
2 36 36
3 9 45 50 mA 13
4 9 54 100 mA 16
5 12 66 200 mA 19
6 12 78 300 mA 22
7 12 90
8 12 102
9 12 114

Relay Pin Data

K3 Exposure Start 24V DC

CL1 TS2-4 CL2 K5-B

AB TS2-10 CD TS2-16 EF
A TS2-5 C E
B TS2-11 D K10 AB F

K4 Rotor Start Current sense 6V DC

CL1 ( to ac current thru’ I of rotor sensing Bridge rectifier)CL2

AB TS2-7 CD P2 of Space Charge EF -ve of X- Ray LED


A TS2-6 C E 7 mA of mA Meter
B TS2-4 D Com. Of Minor kVp Sw F 350 mA of mA Meter

K5 Rotor Run Current sense 6V DC

CL1 ( to ac current thru’ O of rotor sensing Bridge rectifier) CL2

AB TS3-10 CD TS2-16 EF
A C E
B K3 CL2 D ‘Com.’ of Bucky Sw. F

K8 St. By /Rotor start Relay 24V DC

CL1 K9AB CL2 TS3-9

AB TS2-16 CD EF
A C Pin no. 8 of kVp PCB E
B Fuse F5 D Pin no. 9 of kVp PCB F filament circuit

K9 O/L Relay 24V DC

CL1 K11 AB CL2 TS2-1

A C -ve o f O/L LED E


B TS2-4 D F
AB K8 CL1 CD TS2-1 EF

K10 Timer Relay 24V DC

CL1 TS2-13 CL2 TS2-1

A C E
B X- Ray Contr. CL2 D F
AB K3 D CD EF

K11 100KVP O/L relay 24V DC

CL1 Timer PCB Pin no.18 CL2 TS2-1

A C E
B TS2-12 D F
AB K9 CL1 CD EF

K12 Rotor Change over 24V DC


CL1 TS2-8 CL2 TS2-1

AB TS3-11 CD Bridge of K4 EF K12 C &5 ohm 10W res


A TS2-15 C K12 EF &5 ohm 10W res E TS3-12 & 13 ohm 5W res
B To 5 ohms &K5 bridge D TS3-12 & 13 ohm 5W res
F To 13 ohm 5W res