CERTIFICATE OF CALIBRATION

HYTORC CARIBBEAN -TT


LP #37,13 Milton Road
Couva,Trinidad
Phone:1868-340-4113
P.O.#:
CUSTOMER Trinidad Generation Unlimited S.O.#:
TOOL MODEL/ AVANTI 3 SERIAL# 1803
DESCRIPTION HY-AVANTI 3
STATED ACCURACY: +/- 3% OF HYTORC STANDARD PRESSURE/TORQUE CONVERSION CHART
TORQUE RANGE 445 TO 2969 Lb.ft. PRESSURE RANGE: 1500 TO 10000 psi

PROCEDURE & METHOD USED


Hydraulic Torque Wrench was calibrated on "Torque Master" Automated Hydraulic Torque Wrench Calibration System in accordance with
Hydraulic Torque Tool Calibration Procedure #319 version 2.0.

CALIBRATION EQUIPMENT USED


Torque Specialities TORQUE MASTER Automated Hydraulic Calibration System.
[ X ] TSD2, s/n 6289, 0 to 20000 ft.lb Torque Cell with TSD6500-4 s/n 20811 Instrument, NIST Tracebility Reference # 500641, Calibration
Date: 4/11/22, Calibration Due Date: 4/2023, Accuracy within ±0.75% Indicated Value.
[ X ] TSD10KPT, s/n 136110, 0 to 10000 psi Pressure Transducer with TSD6500-2 s/n 20811 Instrument, NIST Traceability Reference #
500639, Calibration Date: 04/11/22, Calibration Due Date: 4/2023, Accuracy within ±0.25% Indicated Value.

Calibrations performed at 67°F, Relative Humidity 69%.

HYTORC
PRESSURE AS LEFT AS FOUND MIN MAX
STANDARD
psi Lb.ft. Lb.ft. Lb.ft. Lb.ft.
1500 445 449 420 432 458
2000 597 610 570 579 615
3000 904 927 865 877 931
4000 1210 1240 1164 1174 1246
5000 1511 1548 1455 1466 1556
6000 1811 1857 1750 1757 1865
7000 2103 2155 2030 2040 2166
8000 2393 2453 2300 2321 2465
9000 2681 2757 2587 2601 2761
10000 2969 3050 2879 2880 3058

AS FOUND/ AS LEFT TOOL HY-AVANTI 3 SERIAL # 1803 IS WITHIN STANDARDS OF ACCURACY.

CALIBRATED BY: Chester Felix APPROVED BY: Tony Febles


Certificate#: 4-20k-220825105226 Lab Manager
Calibration Interval: 1 Year CALIBRATION DATE: 24/08/2022
Calibration Due Date:24/08/23 DATE TOOL PUT IN SERVICE:____________
Uncertainty 0.6% of reading
Calibrations are in accordance with requirements of ISO/IEC 17025:2017. Calibration results were obtained using equipment capable of producing results that are traceable to NIST and
through NIST to the International System of Units (SI). Expanded uncertainties expressed at approximately 95% confidence levels using a coverage factor K=2 was accounted for in
making compliance/non compliance decision with specificiation, and are available upon request within 30 days of calibration date.Results are valid only to the above item calibrated at
the time of test. This certificate shall not be reproduced except in full without the written permission of Hytorc. Pass/Fail results are based on +/- tolerance only. Many factors may
cause an item to drift out of tolerance before the next scheduled calibration.