Professional Documents
Culture Documents
1 CPC BINCONET
APU FLUIMAC
5 GDS HONEYWELL
6 FTS DANFOSS
7 FMU ABB
9 T-Strainer WOOSUNG
ARA-15iXTS-MPPC
Copyright
This publication contains information that is protected by copyright. No part of it may be reproduced in any form or by
any means or used to make any transformation/adaptation without the prior written permission from the copyright
holders.
Trademark
Product names or trademarks (VGA, DPMS, VESA, DVI, PS/2, PC AT, IBM, PIVOT, MS-DOS, WINDOWS,
MICROSOFT, INTEL) are for identification purpose only and are the properties of the respective owners.
ARA-15iXTS-MPPC Panel PC
1. Introduction
- ARA-15iXTS-MPPC is all-in-one Panel PC that combines 15” LCD panel and Main Board 3rd
Generation Intel® CoreTM processors.
- The resolution is 1024 x 768 and can use touch screen and PS/2 mouse at the same time.
2. Specifications
2.1 General Specifications
ARA-15iXTS-MPPC Panel PC
ARA-15iXTS-MPPC Panel PC
3. Introduction
4. External Components
5. Dimensions
Picture 1. Panel PC Dimensions
7/10
User’s Manual
ARA-15iXTS-MPPC Panel PC
7. Packing List
If anything that listed below is absence, please contact retailer your retailer or BINCONET MARIN
E SOLUTIONS Support Help Desk.
Panel PC
Accessories
Panel PC Manual
Software Drivers Install CD-ROM
OS CD ( Windows 7 Pro English Version)
Touch Screen Driver Install CD-ROM
9. Safety Precaution
-The safety recommendations outlined in this section are to be read, understood and followed before operating the
product.
- Keep this information in a safe place for future reference. Failure to comply with any of the following safety
procedures could result in serious hazard.
8/10
User’s Manual
ARA-15iXTS-MPPC Panel PC
9/10
User’s Manual
ARA-15iXTS-MPPC Panel PC
If you strongly wipe down the surface of the system with a sponge and a piece of cloth, in case of
plastic equipment,
The coating may be fallen away.
When cleaning the system case or the surface of the TFT LCD, wipe with a slightly moistened, soft
fabric.
10/10
User’s Manual
ANU Mering pump
(IWAKI)
( )Country codes
IWAKI
IWAKI CO.,LTD. 6-6 Kanda-Sudacho 2-chome Chiyoda-ku Tokyo 101-8558 Japan
Australia IWAKI Pumps Australia Pty. Ltd. TEL : (61)2 9899 2411 FAX : 2 9899 2421 Italy
TEL:(81)3 3254 2935 FAX:3 3252 8892(http://www.iwakipumps.jp)
IWAKI Europe GmbH, Italy Branch TEL : (39)0444 371115 FAX : 0444 335350
Electromagnetic Metering Pump
Austria IWAKI EUROPE GmbH TEL : (49)2154 92540 FAX : 2154 9254 48 Korea IWAKI Korea Co.,Ltd. TEL : (82)2 2630 4800 FAX : 2 2630 4801
Belgium IWAKI Belgium n.v. TEL : (32)1367 0200 FAX : 1367 2030 Malaysia IWAKIm Sdn. Bhd. TEL : (60)3 7803 8807 FAX : 3 7803 4800
China IWAKI Pumps (Shanghai) Co., Ltd. TEL : (86)21 6272 7502 FAX : 21 6272 6929 Norway IWAKI Norge AS TEL : (47)23 38 49 00 FAX : 23 38 49 01
China
China
China
IWAKI Pumps (Guangdong) Co., Ltd. TEL : (86)750 3866228
GFTZ IWAKI Engineering & Trading (Guangzhou) TEL : (86)20 8435 0603
GFTZ IWAKI Engineering & Trading (Beijing) TEL : (86)10 6442 7713
FAX : 750 3866278
FAX : 20 8435 9181
FAX : 10 6442 7712
Singapore
Spain
Sweden
IWAKI Singapore Pte. Ltd.
IWAKI Europe GmbH, Spain Branch
IWAKI Sverige AB
TEL : (65)6316 2028
TEL : (34)93 37 70 198
TEL : (46)8 511 72900
FAX : 6316 3221
FAX : 93 47 40 991
FAX : 8 511 72922
EH-E Type
Denmark IWAKI Nordic A/S TEL : (45)48 24 2345 FAX : 48 24 2346 Switzerland IP Service SA TEL : (41)26 674 9300 FAX : 26 674 9302
Finland IWAKI Suomi Oy TEL : (358)9 2745810 FAX : 9 2742715 Taiwan IWAKI Pumps Taiwan Co., Ltd. TEL : (886)2 8227 6900 FAX : 2 8227 6818
France
Germany
Holland
IWAKI France S.A.
IWAKI EUROPE GmbH
TEL : (33)1 69 63 33 70
TEL : (49)2154 9254 0
IWAKI Europe GmbH, Netherlands Branch TEL : (31)547 293 160
FAX : 1 64 49 92 73
FAX : 2154 9254 48
FAX : 547 292 332
Taiwan
Thailand
U.K.
IWAKI Pumps Taiwan (Hsin-chu) Co., Ltd.
IWAKI (Thailand) Co.,Ltd.
IWAKI Pumps (UK) LTD.
TEL : (886)3 573 5797
TEL : (66)2 322 2471
TEL : (44)1743 231363
FAX : (886)3 573 5798
FAX : 2 322 2477
FAX : 1743 366507
Instruction Manual
Hong Kong IWAKI Pumps Co., Ltd. TEL : (852)2 607 1168 FAX : 2 607 1000 U.S.A. IWAKI AMERICA Inc. TEL : (1)508 429 1440 FAX : 508 429 1386
Indonesia IWAKI Singapore (Indonesia Branch) TEL : (62)21 690 6606 FAX : 21 690 6612 Vietnam IWAKI pumps Vietnam Co.,Ltd. TEL : (84)613 933456 FAX : 613 933399
Read this manual before use of product
T656-2 '13/09
Thank you for selecting the electromagnetic metering pump EH-E series. This instruction
manual deals with “Safety Section” “Product outline” “Installation Section” “Operation
Section” and “Maintenance Section”.
Please read through this manual carefully to ensure the optimum performance, safety and
service of the EH-E series.
Contents
IMPORTANT INSTRUCTION .......................................................... 1
Safety Instruction .............. .............................................................. 2
Outline ........................................................................................... 5
1. Unpacking ............................................................... 6
2. Operating Principle ................................................. 6
3. Identification Code .................................................. 7
4. Features ... .............................................................. 9
5. Specification ......................................................... 10
6. Operating Function 11
7. Display and Keys .................................................. 13
Maintenance .................................................................................. 44
1. Troubleshooting .................................................... 45
2. Maintenance and Inspection ................................. 46
3. Disassembly and Assembly .................................. 49
4. Optional Accessories ............................................ 53
5. Exploded Views and Dimension Drawing .............. 53
Important instructions
For the Safe and
Correct Handling of the Pump
● "Safety Instruction" section deals with important details about handling of the product. Before
use, read this section carefully for the prevention of personal injury or property damage.
● Observe the instructions accompanied with "WARNING" or "CAUTION" in this manual. These
instructions are very important for protecting users from dangerous situations.
Types of Symbols
Indicates that “Warning” or “Caution” must be exercised. Inside this triangle, a con-
crete and practical image provided as a warning or caution message is depicted.
Indicates a prohibited action or procedure. Inside or near this circle, a concrete and
practical image of the activity to be avoided is depicted.
Export Restrictions
Technical information contained in this instruction manual might be treated as controlled
technology in your countries, due to agreements in international regime for export control.
Please be reminded that export license/permission could be required when this manual is
provided, due to export control regulations of your country.
-1-
Safety instructions
WARNING
● Turn off the power supply
Working without disconnecting the power supply cause an electrical shock. Before
engaging upon any working procedures involving the pump, make sure to turn the
power supply switch off and to stop the pump and other related devices.
Electrical Shock
● Terminate operation
When you detect or become aware of a dangerous sign or abnormal condition during
operation, terminate the operation immediately and start from the beginning again.
● No remodeling
Never remodel a pump. Otherwise, a serious accident may result. Iwaki will not be
responsible for any accident or damage of any kind which is caused by the user
remodeling the pump without first obtaining permission or instructions from lwaki. No Remodeling
● Wear protectors
If you touch or come in contact with any type of hazardous chemical liquid, including
but not limited to chemicals, you may experience a serious injury. Wear protective Wear protective
gear (protective mask, gloves, etc.) during the pump operation. gear
-2-
Safety instructions
CAUTION
● Qualified operators only
The pump operator and pump operation supervisor must not allow any operators who
have little or no knowledge of the pump to run and operate the pump. Pump operators
must have a sound knowledge of the pump and its operation. Prohibited
● Ventilate
Poisoning may result during an operation which involves, toxic or odorous liquid.
Ventilate the operating site sufficiently. Caution
● Spill-out accident
Protective measures should be taken against any accidental spill-out or leakage of
the operating liquid as a result of unexpected damage on the pump or the related pip-
Caution
ing.
● Damaged pump
Never operate a damaged pump. A damaged pump may cause leakage or electrical
shock. Prohibited
● Frequent stop and start of pump should be done by using STOP function (ON and
OFF of STOP terminal). If you can not use STOP function and are forced to operate
pump by turning OFF and ON of power source, ON and OFF of power source should
be limited to six times an hour.
-3-
Safety instructions
CAUTION
● Arrange grounding
Do not operate the pump without connecting the grounding wire. Otherwise, an elec-
trical shock may result. Make sure the grounding wire is connected with grounding
Grounding
terminal.
-4-
Outline
1. Unpacking ........................................... 6
4. Features ............................................. 9
5. Specification ..................................... 10
-5-
Outline
1. Unpacking
2. Operating Principle
The EH Series electronic metering pump consists
of a pump unit, a drive unit, and a control unit. The
Drive unit drive unit is an electromagnetic solenoid. When the
Stroke length Valve solenoid coil is energized by the control unit the
adjusting dial armature shaft moves forward due to the magnetic
force of the solenoid. The shaft is attached to a
PTFE faced diaphragm which is part of the pump
Pump
unit. The diaphragm is forced into the pump head
Control unit
head cavity decreasing volume and increasing pressure
which forces liquid in the pump head out though
the discharge check valves. When the solenoid
coil is de-energized, a spring returns the armature
to its starting position. This action pulls the dia-
phragm out of the head cavity increasing volume
and decreasing pressure. Atmospheric pressure
Plunger Spring Diaphragm then pushes liquid from the supply tank through
the suction check valves to refill the pump head.
-6-
Outline
3. Identification Code
• Pump identification
EH - E 31 PC - 23U P E 8 -
(1) (2) (3) (4) (5) (6) (7) (8) (9)
-7-
Outline
(6) Power code
P with a plug
No symbol without a plug
EHC - 100 P E - * *
(1) (2) (3) (4) (5)
(1) Controller name EH
10 0 100VAC 90-126VAC
11 U 110/115VAC 90-126VAC
(2) Voltage codes 20 J 200VAC 180-220VAC 50/60Hz
20 E 220/230/240VAC 198-264VAC
23 U 230VAC 207-253VAC
P: with a plug
(3) Lead wire
No code: without a plug
(4) Controller E
0-99
(5) Special configuration HV: High viscosity type
-8-
Outline
4. Features
Driving unit Stroke length adjusting knob
●Electromagnet and spring force makes diaphragm ●To adjust discharge capacity per stroke in the
reciprocate responding to the command of control range of 20 to 100%.
unit.
● Control unit
Pump operation is done electronically by control unit in which microcomputer is built-in.
● Easy operation
Pump operation and flow rate control is done by keys on control unit, which enables simple control of flow
rate.
-9-
Outline
5. Specification
• Pump specification
Standard types
Output capacity Output per stroke Maximum pressure Stroke frequency Permissible stroke
Model (ml/min) (ml/stroke) (MPa) (spm) length % (mm)
31 340 0.19-0.94 1
36 520 0.29-1.44 0.7 (0.6) 20-100
0-360
46 750 0.42-2.08 0.4 (0.3-1.5mm)
56 1250 0.69-3.47 0.2
- 10 -
Outline
6. Operating Function
• Manual operation
Stroke rate can be set from 0 to 360 spm with keys ▲ and ▼, and pump start and stop can be done with
keys START/STOP. Both setting can be done while the pump stops or operates.
Operation chart
ON ON ON
START/STOP key
Operation
360 spm
0 A1 A2 20 mA
SET point 1 SET point 2
Current Current
- 11 -
Outline
• Digital input (pulse multiply) operation
Pump makes strokes from 1 to 999 responding to external pulse signal. Stroke rate is the spm set for man-
ual operation. The pulses which came while operation are put in memory up to 255 pulses. (It is possible to
make the pulses not to be put in memory.)
Ext. pulse
Pump stroke
1 2 3 4 5 1 2 3 4 5
External pulse
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5
Pump stroke
- 12 -
Outline
7. Display and Keys
• Controller display and panel
NOTE: Two keys pushing means to push simultaneously two keys of EXT key and UP or DOWN key.
- 13 -
Outline
• Basic display
1
2
1
2
1
2
- 14 -
Outline
• Alarm display
Display for excess spm at EXT operation (Pulse multiply operation). (OVER
1 lights.)
2
In multiply operation, visible if next pulse comes when the pump is making
the strokes. While visible, max. 255 pulses can be put in storage. You can
set not to put them in storage.
WAIT EXT STOP OVER SET
Display for excess spm at EXT operation (Pulse dividing operation). (OVER
1 lights.)
2
In dividing operation, visible if signal which exceeds max. preset number
of strokes comes. While visible, the pump runs at fixed max. number of
strokes. You can set so that excessive input pulses can be put in storage up
WAIT EXT STOP OVER SET
to 255 pulses.
- 15 -
Installation
2. Installation ........................................ 19
3. Tubing ............................................... 20
- 16 -
Installation
1. Notes on Operation
Operators and maintenance service staff must read the instruction manual thoroughly before using the prod-
ucts. Do not operate the pump system unless all of the contents in the manual are completely understood.
CAUTION
- 17 -
Installation
3. Install the pump at the place convenient for the
maintenance/inspection works in the future.
Securely fix the pump so that the pump can not
vibrate horizontally.
CAUTION
Liquid splash on driving unit and control
unit may cause failure or accident. Pay
attention the liquid not to be splashed.
- 18 -
Installation
2. Installation
CAUTION
When you feel danger or abnormality
during installation or mounting works,
stop works immediately to check its
reason.
1) Install pump at place of ambient temperature of
40 deg. C or below and humidity of 40% RH or
below (No dew condensation should be in the
control unit.) where is easy and convenient for
maintenance and inspection works.
Cover the pump when installing it out of doors for
protection against direct sunlight and wind & rain.
- 19 -
Installation
3. Tubing 1. Tubing for types VC, V6, PC, VM, FC
Use PVC braided hose or so which corresponds
Hose to connection bore of pump.
Hose is fixed by hose stopper. Put fitting nut and
hose stopper on hose and insert hose end into
fitting spacer and then push the hose stopper to
the bottom of fitting spacer and tighten fitting nut.
Fitting nut
Make sure that the R-side of the hose stopper is
facing to the fitting spacer.
Hose stopper NOTE: Especially for type FC, make sure to face
the R-side of the stopper to the fitting
spacer. Otherwise, hose may be dam-
aged.
Insufficient or wrong mounted hose stopper may
cause liquid leakage, failed priming or damaged
Face R-side to fitting spacer. hose.
Fitting nut is made of plastics. Pay attention not
Fitting spacer to tighten it excessively. Excessive tightening
may break nut.
CAUTION
Fitting nut · Use specified size of tube. Otherwise
tube may come off.
Remove · Insert the fitting spacer into tube as deep
check valve
as possible. Otherwise liquid may leaks
or tube may come off.
· Pay attention not to lose O ring (gasket
for FC). If O ring (gasket for FC) is not
mounted, it may cause liquid leakage or
Return
to tank failed self-priming.
2. Tubing for SH type
Connections are Rc1/4 or NPT1/4 (EH-E31,
36SH) and Rc3/8 or NPT3/8 (EH-E46, 56SH).
Apply seal tape to connected pipe and securely
tighten it so that liquid can not leak nor air be
sucked in.
Check valve 3. Tubing for bleeding
When you do bleeding, return the discharge
side tube to the suction side tank. Remove
Three way joint the check valve if it is mounted. Refer to item
Air vent "Bleeding" of "Operation" section.
valve Discharge hose
CAUTION
Pump
The EH-E pump is not equipped with built-
in air-vent valve. Mount the air-vent valve
referring to the illustration on left.
Built-in type air-vent valve is available as
an option for types VC, V6 and PC.
- 20 -
Installation
4. Mounting check valve
In case the pump is installed as mentioned
below, make sure to install the check valve
to avoid over-feeding.
Suction side Discharge side 3) In case the discharge side liquid level is high-
Fig. A er than the suction side one but the height
difference between the two levels is 5 meters
or less. (Fig. C)
Fig. B
5m
Fig. C
- 21 -
Installation
Precautions when mounting check valve
1) Install the check valve at the end of discharge
tubing. It should be separated the pump by 1
meter or more.
- 22 -
Installation
4. Electrical Wiring
1. Precautions on wiring
CAUTION
● Only qualified operator/service staff should be in charge of the related electrical
arrangement and control of the power source. Failure to observe this instruction may
result in injury to person or damage to assets.
● Frequent stop and start of pump should be done by using STOP function (ON and OFF
of STOP terminal). If you can not use STOP function and are forced to operate pump
by turning OFF and ON of power source, ON and OFF of power source should be lim-
ited to six times an hour.
● Surge voltage
The electronic circuit of the control unit may be
affected by excessively high surge voltage. So,
Surge absorbing element
do not operate the pump near high-power electri-
cal equipment of 220V or above that generates
high surge voltage.
P Under unavoidable circumstances, take either of
the following measures.
(1) Use a surge absorbing element (such as a
varistor with surge resistance of 2000A or
P more) at the pump power supply connection.
Noise-cut transformer (2) Use a noise-cut transformer.
- 23 -
Installation
2. Wiring
Power cord and external signal cord are connected according to the procedure as bellow.
CAUTION
· Never do the wiring when pump is operating. Otherwise you will get an electric shock or
the pump will be failed due to the short-circuit.
· Wait one minute or more after the power is switched off to start the wiring works
because the internal circuit will be electrified just after it is switched off.
· Use the cord of which outer diameter is 7, 8 mm. If smaller cord is used, enough tight-
ness can not be obtained
· Wiring of external and stop signals
Do not combine cables for EXT and STOP signals with power cable. Do not use coaxial
cable (5-wires cable or so) for wiring power cable, EXT and STOP signals cable all
together. Otherwise noise may be generated in EXT and STOP cables because of
induction effect from power cord, which may result in pump failure or failed operation.
1 2 4 5 6 7 8
Pulse Input
Control unit
Stop Input
Power Ground
+ - Analog Input
Terminal box gasket + - Sensor Power
DC 12 V
Terminal box 10 mA max.
Cord nut
Connector
Cord gasket
Cord
Fig. 5
- 24 -
Installation
1) Remove four screws with screw driver to take off
terminal box.
CAUTION
Check if connector is securely mounted.
Wrongly or insufficiently mounted connec-
tor may cause failed operation.
CAUTION
If terminal box is mounted by other screws
than attached ones, it may cause failed
sealing.
CAUTION
Use the cord of outer diameter of 7.8 mm.
Smaller diameter cord may cause wrong
sealing resulting in failed pump.
- 25 -
Installation
● Wiring for analog input operation
Analog input operation means the pump operation
by input current signal of 0 to 20 mA to change the
stroke rate in proportion to 0 to 360 spm.
For the wiring, connect the wires to the terminals
and and fix them by screws. Tightening
+ - torque is 0.4 N•m or less. is plus and is
minus. (Inside resistance : 250 Ohm)
CAUTION
Pay attention to polarity. Wrong connection
leads to unintentional operation. If signals
are entered with wrong connection for a
long time, the pump may break down.
+ - ● Wiring for stop function
Stop function means the method to provisionally
stop the pump operation by external potential
free contact signal or open collector signal.
For the wiring, connect the wires to and of the
connector.
ON • When used with open collector
Pump Pay attention to the polarity. is plus and is minus.
(Max. chargeable voltage is 5 V, current 1.1 mA)
CLOSE
External OPEN • When used with mechanical contact such as relay
contact or so. Use the electronic purpose one of which the
minimum applicable load of 1 mA or below.
CAUTION
Frequent stop and start of pump should
be done by using STOP function (ON and
OFF of STOP terminal). If you can not use
STOP function and are forced to operate
pump by turning OFF and ON of power
source, ON and OFF of power source
should be limited to six times an hour.
● Wiring for pulse input function
Pulse input function means the pulse multiply or divid-
+ - ing operation by external potential free contact signal
or open collector input.
For the wiring, connect the wires to and with
screw driver. Tightening torque is approx. 0.4 N•m.
• When used with open collector
Pay attention to the polarity. is plus and is minus.
(Max. chargeable voltage is 5 V, current 1.1 mA)
When mechanical contact such as relay or so is
used, use the one designed for electronic pur-
pose of which the minimum applicable load of 1
mA or below.
- 26 -
Operation
2. Operation .................................................. 32
- 27 -
Operation
1. Preparation for Operation
CAUTION
• Do not operate the pump with discharge-side valve completely closed.
Operating the pump with discharge-side valve fully closed may lead to liquid leakage or
pipe rupture.
• Do not run pump dry.
Dry operation of the pump over a long time (longer than 30 minutes) causes the pump
to overheat and the pump unit (pump head, valve guide etc.) to become deformed or the
pump head attachment to become loose, which may result in liquid leakage trouble.
• Keep the pump head firmly assembled.
If mounting bolts on the pump head are loosened, liquid leakage may result.
* Fasten the 4 or 6 hex. socket bolts tightly before starting the initial pump operation.
(The bolts may be loosened during storage or transportation of the pump, depending
upon the condition.)
* Fastening torque: 2.55 N·m.
Tighten all the bolts fully by applying an equal amount of torque in a diagonal order
among the bolts.
• Frequent stop and start of pump should be done by using STOP function (ON and OFF
of STOP terminal). If you can not use STOP function and are forced to operate pump by
turning OFF and ON of power source, ON and OFF of power source should be limited
to six times an hour.
1-1. Bleeding
Bleeding is a process to eliminate the air which remains in the suction tube or pump head.
Make sure to carry the bleeding prior to the initial operation of the pump or when the liquid in the tank was
replaced or the pump rested for a long time.
Starting and stopping pump operation
1) Switch on the power and ON lamp (green)
START/STOP
1
lights and goes to WAIT mode after the ver-
2 sion number was displayed. (When the pump is
EXT
switched on for the first time.)
WAIT EXT STOP OVER SET
The word WAIT is lit during WAIT mode.
EH-E ON
ON lamp
START/STOP key
EH-E ON
- 28 -
Operation
Fitting nut Bleeding for types VC, V6, PC and FC
1) Return the discharge side hose to the tank and
start the pump. Remove the check valve if it is
Remove mounted.
check valve
2) Run the pump for 10 minutes for bleeding.
To tank
4) Close the adjusting screw by turning it clockwise.
To discharge
line
- 29 -
Operation
1-2. Adjustment of Discharge Capacity
Adjustment of the discharge capacity can be done by adjusting the stroke length and by adjusting the stroke
rate but basically it is done by adjusting the stroke rate. Stroke length adjustment is an auxiliary way when
the stroke rate adjustment is not enough.
Pay attention to the following for the correct adjustment.
1) For the gaseous liquid such as sodium hypochlorite and hydrazine solution, set the stroke length at 100 %
or near. Discharge flow may be reduced if the stroke length is short.
2) When the back pressure is high at discharge side, set the stroke length at 100 % or near to adjust dis-
charge capacity by stroke rate.
In case the reaction is greatly influenced by discharge capacity per pump shot in the application of neu-
tralization or titration, adjust discharge capacity by stroke speed with shortened stroke length to minimize
discharge capacity per shot.
50 180spm
25
0 2025 50 75 100
Stroke length is adjusted. %
75
Disch. capacity
50
25
0
180 360
Stroke rate spm
- 30 -
Operation
3. Adjustment of stroke length
(1) Power on the pump to start and adjust the discharge capacity by turning the stroke length adjustment
knob.
(2) Figure below shows the relation between the stroke length and discharge capacity.
Stroke length can be adjusted from 0 to 100 % but actually set the length between 50 and 100 %.
75
Disch. capacity
50
25
0 25 50 75 100
Stroke length %
CAUTION
Do not turn the stroke length adjusting knob while pump stops.
● Full operation
Start pump for full operation after bleeding and discharge adjustment are finished.
Operation is done by manually or by any function of control unit.
● Manual operation
Starting pump : Turn on power to light ON lamp (green) to come to WAIT mode.
Push START/STOP key (for the first time to turn on power) for WAIT lamp to go out and
start pump.
Stop pump : Push START/STOP key to light WAIT lamp to stop pump. Every time to push START/STOP
key, pump starts and stops.
● EXT operation
Refer to next section of "Operation".
● When the pump is stopped for a long period (more than 1 month)
When the pump is stopped for a long period, clean the wet-end of pump and tubing by operating the pump
with clean water for 30 minutes before stopping the pump.
When the pump is started after stopping, if it does not suck up and discharge the liquid, clean the valve set
and remove foreign matters adhered. Do bleeding and adjust the discharge capacity.
- 31 -
Pulse memory switching over
▲ ▼ ▲ ▼
Program Ver display
1
EXT
1 1
EXT OP.
V3.0E 2 EXT OP. mode setting / -- OF 2 X -- ON 2
WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
EXT
Pulse dividing memory OFF Pulse multiplying memory OFF
▲
Pulse dividing op. 1 1 1
EXT OP. 360 2 dIG 2 ANA 2
1 WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
/ 1 START START ▼
2 STOP STOP Digital mode setting display Analog mode setting display
WAIT EXT STOP OVER SET START
MANUAL OP. STOP
- 32 -
at digital mode set at analog mode set
/ 2 2 / 1 2 A04.0 2 A04.1 2
WAIT EXT STOP OVER SET
MANUAL OP. spm change WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
▼ ▼
EXT
EXT EXT ▲
Set value change
1 1
P 0 2 P 1 2
▲ WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
1 1 ▼
X 2 X 1 EXT
2 2 EXT
WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
▼
Pulse multiply op. mode ▼
1 1
A20.0 2 A19.9 2
WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
▲
EXT
▼
1 1
P360 2 P359 2
WAIT EXT STOP OVER SET WAIT EXT STOP OVER SET
▲
Operation
Notes to Overview Operating Scheme
1) ------> means automatic movement. After the program Ver is displayed, it automatically moves to the sta-
tus at which the power was off last time. (When the pump is powered for the first time, it comes to WAIT
mode.)
2) For the pump start by manual operation, push START/STOP key at WAIT mode. To stop the pump, push
START/STOP key again.
3) EXT operation starts by pushing EXT key at WAIT mode and stops by pushing START/STOP key.
4) For the switching over between analog and digital operation, push EXT and ▲ keys simultaneously.
Parameter is changed by ▲ and ▼ keys and the set value is fixed by START/STOP key.
5) To move to pulse memory switching mode, push EXT key while dIG is selected at analog/digital switching
mode. (If ANA is selected, you can not come to this mode.)
Figures are changed by ▲ and ▼ keys and switching between multiply and dividing is done by EXT key.
Push START/STOP key to return to WAIT mode.
6) To go to EXT setting mode, push EXT and ▲ key simultaneously at waiting mode.
Parameter is changed by pushing ▲ and ▼ keys when you are in analog operation, and when you are in
digital operation, by pushing ▲ and ▼ keys to change parameter and push EXT key to toggle between
dividing and multiplying operation.
- 33 -
Operation
2-2. Setting and Operation of Controller
■ Manual operation
(1) Power ON
START/STOP
1 When the power is on, a green lamp lights up and the word-
2
ing "V3.0E" appears, then the stroke rate for manual opera-
EXT
WAIT EXT STOP OVER SET
tion is displayed and comes to WAIT mode. (In case the
pump is powered on for the first time.) If it does not come to
EH-E ON
WAIT mode, push once START/STOP key to come to WAIT
mode.
EH-E ON
Blink
EH-E ON
- 34 -
Operation
■ Automatic operation
1. Analog signal operation
(1) Power ON
When the power is on, a green lamp lights up and the
START/STOP
1
2 wording "V3.0E" appears, then the stroke rate for manual
EXT
operation is displayed and comes to WAIT mode. (In
WAIT EXT STOP OVER SET
case the pump is powered on for the first time.) If it does
not come to WAIT mode, push once START/STOP key to
EH-E ON
come to WAIT mode.
EH-E ON
EH-E
EH-E ON
EH-E
- 35 -
Operation
(6) Push EXT key to confirm the value at SET point 1 and
START/STOP
1
move to the setting of stroke rate for the current value of
2 SET point 1. The words PO is displayed and SET and 1
EXT
light.
WAIT EXT STOP OVER SET
EH-E ON
(7) Setting of stroke rate for the current value at SET point 1
START/STOP
1 Set the stroke rate for the current value at SET point 1 by
2
▲ and ▼ keys. Setting range is 0 - 360 spm. The pump
EXT
is set at 0 spm when it is shipped from the factory.
WAIT EXT STOP OVER SET
EH-E
(8) Push EXT key to confirm the set current value at SET
START/STOP
point 1 and move to the setting of the input current value
1
2 at SET point 2.
EXT
Display shows A20.0 and SET and 2 light.
WAIT EXT STOP OVER SET
EH-E ON
(10) Push EXT key to confirm the set current value at point 2
START/STOP
1
and move to the setting of stroke rate vs current value of
2 SET point 2. Display shows P360 and SET and 2 light.
EXT
WAIT EXT STOP OVER SET
EH-E ON
(11) Setting of stroke rate for the current value of SET point 2
START/STOP
1 Push ▲ and ▼ keys to set the stroke rate for the current
2 value at SET point 2. Do not set the value which is set for
EXT
SET point 1.
WAIT EXT STOP OVER SET
Set range is 0 - 360 spm.
EH-E
The pump is set at 360 spm when it is shipped.
- 36 -
Operation
(12) Push START/STOP key to confirm the set value and
move to WAIT mode.
START/STOP
1
2
EXT
WAIT EXT STOP OVER SET
EH-E ON
Alarm indication
At the analog input operation, if the current exceeding
START/STOP
360 spm comes, the word OVER lights. While the OVER
1
2 is lighting, the pump operates at 360 spm.
EXT
WAIT EXT STOP OVER SET
EH-E ON
EH-E ON
- 37 -
Operation
(2) Move to EXT operation mode.
Push ▲ key and EXT key simultaneously.
START/STOP
1
2 The word dIG is displayed and SET lights.
EXT
(If the word ANA is displayed, push ▼ key to change to
WAIT EXT STOP OVER SET
dIG. Every time you push ▼ key, the words ANA and dIG
switch over.) When the pump is shipped from the factory,
EH-E ON
it is set at dIG.
EH-E ON
(5) Push EXT key to confirm ON, OFF of pulse memory and
START/STOP
1
return to EXT operation mode.
2 The display shows dIG and SET lights.
EXT
WAIT EXT STOP OVER SET
EH-E ON
EH-E ON
- 38 -
Operation
(7) Switching of operation mode
START/STOP
Push EXT key and ▼ key simultaneously and / 1
1
2 (Dividing operation mode) is displayed and SET lights. (If
EXT X 1 is displayed, go to next (8).
WAIT EXT STOP OVER SET
EH-E
EH-E ON
EH-E ON
- 39 -
Operation
(11) Push EXT key to operate the pump.
START/STOP
Pump starts to operate, WAIT goes out and ON lamp
1
2 blinks when a pulse comes.
EXT The pump automatically stops after it operated for the
WAIT EXT STOP OVER SET stroke number which is set at above item (8). During the
operation, the display shows pre-set stroke number and
EH-E ON
EXT lights. Push START/STOP key to stop the pump
Blink and go to WAIT mode. To start the pump again, push
EXT key.
Alarm indication
If next pulse comes during the multiplying operation,
START/STOP
OVER lights. If the pulse memory is set to ON, max. 255
1
2 times pulse is put in memory and the pump continues
EXT
operation.
WAIT EXT STOP OVER SET
EH-E ON
- 40 -
Operation
(3) Move to pulse memory (The function to memorize the
START/STOP
pulses (max. 255 pulses) which come while the pump
1
2 does the multiply operation.)
EXT
Push EXT key and / -- OF or / -- ON is displayed and
WAIT EXT STOP OVER SET SET lights.
EH-E ON
EH-E ON
EH-E ON
EH-E ON
- 41 -
Operation
(7) Switching of operation mode
START/STOP
Push EXT key and ▼ key simultaneously, and / 1 (Pulse
1
2 dividing operation mode) is displayed and SET lights.
EXT
WAIT EXT STOP OVER SET
EH-E
EH-E ON
EH-E ON
- 42 -
Operation
(11) Push EXT key to start the pump. WAIT disappears and
START/STOP
ON lamp blinks.
1
2 Dividing ratio is displayed and EXT lights.
EXT Stop the pump, push START/STOP key to come to WAIT
WAIT EXT STOP OVER SET mode.
When the pump is started next time, push EXT key.
EH-E
Alarm indication
If pulses which exceeds upper stroke rate come in divid-
START/STOP
ing operation, OVER lights.
1
2 If the pulse memory is set at ON, max. 255 inputs are put
EXT in memory and the pump continue to run for this pulses.
WAIT EXT STOP OVER SET
EH-E
- 43 -
Maintenance
1. Troubleshooting ................................ 45
- 44 -
Maintenance
1. Troubleshooting
WARNING
● Wear protector
You may be injured if you touch chemical liquid. When you work on
pump, be sure to wear protector such as protective mask, safety gloves Wear protective
or so. gear
- 45 -
Maintenance
2. Maintenance and Inspection
2-1. Daily inspection
Pay attention to following items during pump operation and if you find any abnormality, stop pump immedi-
ately to take countermeasures referring to Troubleshooting on page 45. When the time comes to replace the
wear parts, replace it by new one.
No. Check item Description Check method
If pump discharges liquid ● Check if liquid is discharged. ● Flow meter or visual.
normally. ● Check if suction or discharge pressure is ● Refer to figures on
1 normal. nameplate.
● Check if liquid characteristics are not ● Visual or auditory
changed, crystallized nor adhered.
Abnormal noise or vibration. ● Pump may generate abnormal sound or ● Visual or auditory
2
vibration when it runs abnormally.
If no liquid leaks through pip- ● Tighten the parts where leakage happens. ● Visual inspection
3 ing or any pump parts or if ● Air is sucked in if you see many bubbles in
no air is not sucked in pump. discharged liquid.
Check pump head tightening bolts periodically and tighten them when necessary.
14
14
11 14
31
13 14 11
Valve set 12 2 set
11 13
11 13 12
13 12
17
12 17
17
Approx.
8000 hrs
Diaphragm 7 1
*Refer to
"Exploded
O ring 30 * views" for quan-
tity required.
- 46 -
Maintenance
FC type
Parts Q'ty Time to replace Remarks
EH-E31 • E36 EH-E46 • E56
14
11
14
13
12 11
Valve set 14 2 set
13
11 12
13 14
12 17
14
17
Approx
8000 hrs
Diaphragm 7 1
*Refer to
"Exploded
Gasket *
30 views" for quan-
tity required.
SH type
Parts Q'ty Time to replace Remarks
EH-E31 • E36 EH-E46 • E56
14
11
14
13
14 11
Valve set 12 2 set
14 13
11 14
12
13
31 Approx
14 8000 hrs
12
31
Diaphragm 7 1
- 47 -
Maintenance
HP6 type
Parts Q'ty Time to replace Remarks
EH-E36
14
11
13
12
14
Approx
8000 hrs
Diaphragm 7 1
*Refer to
"Exploded
Gasket 17 *
30 views" for quan-
tity required.
- 48 -
Maintenance
3. Disassembly and Assembly
WARNING
• Inspection and maintenance
Inspection, maintenance, disassembly and assembly of pump must
be done within the procedure and extent described in this instruction
manual. Never disassemble the parts beyond the extent described in
this manual (Do not disassemble electromagnetic driving part and elec-
tronic circuit board). The product is not guaranteed if it is disassembled
or modified beyond the extent described in this manual. IWAKI is not
responsible to the damage and accident caused by the disassembly and
modification beyond the extent.
• Wear protector
You may be injured if you touch chemical liquid directly or chemical liq-
uid splashes on you. Wear protective cloth, mask or safety gloves or so Wear protective
● How to depressurize
EH-E pump has no air vent valve except SH type. Install an air vent valve at discharge side to open it to
release pressure. (Refer to item "Tubing" on page 20.
SH type has an air vent valve and can release pressure by following procedure.
(Take the same procedure to release air for types VC, V6 and PC which equip air vent valve as option.)
1. Stop pump.
2. Turn an air vent adjusting screw to left by two turns to fully open air vent port.
3. Confirm that liquid came out from air vent port and pressure was released.
NOTE : When the liquid does not come out, it is possible the pressure is not released. In this case continue
to run pump until you confirm the liquid came out from air vent port and the pressure was released.
- 49 -
Maintenance
3-2. Replacement of valve set
<Disassembly>
Fitting nut
1. Loosen fitting nut to remove hose paying
attention to liquid coming out.
Connection port
2. Loosen connection port by wrench to remove
it and take valve set out of pump head.
<Assembly>
Assemble valve set in reverse order to disassembly
paying attention to the following.
● Arrange and insert valve set in correct order and
direction.
● Wrongly ordered and directed valve set causes
failed pumping such as leakage and reduced flow
rate or so.
● Do not forget to put O ring.
- 50 -
Maintenance
Exploded views of valve set
VC, V6, PC, VM types FC type SH type
17
14 31 · 36 31 · 36 31 · 36
14
Same for 11 Same for Same for
11
disch. and 13
disch. and disch. and
13 suc. 12 suc. suc.
14
12
11
11 13
12
13
12
14 14
31
31 46 · 56
14
Same for
Disch.
14 11 disch. and
Disch.
10 suc.
11
11
13 14
17
12
17 46
46
17
17
14 14
11
Suc.
11
Suc.
13
13 12
14 HP6 type
12 31
31 14
14
14
11
Same for
52 disch. and
13 suc.
17 56 56 12
14 Same for
14 Same for 14
disch. and disch. and
11
suc. suc.
13
11
12
14
13
17
12
- 51 -
Maintenance
3-3. Replacement of diaphragm
<Disassembly>
1. Set stroke adjusting knob at zero.
2. Remove four or six hex. socket head bolts with
hex. wrench and take pump head off.
3. Hold periphery of diaphragm and turn it coun-
ter clockwise to remove it from plunger pin.
(Diaphragm is screwed in plunger pin.)
WARNING
● Do not lose diaphragm spacer. Zero
to several diaphragm spacers are put
between retainer and plunger pin for the
purpose of adjusting the position of dia-
phragm. Number of spacer depends on
each pump and there is the pump which
has no spacer.
<Assembly>
Assemble pump in the reverse order to disassem-
bly paying attention to following.
1. Confirm that a stroke length adjusting knob is set
at 0%.
2. Put a retainer and diaphragm spacer on the
threaded part of new diaphragm and screw it in
plunger pin. Put the recessed side of retainer on
stepped part of diaphragm. Pay attention for the
retainer not to drop.
3. Start pump and adjust stroke length at 50% and
then stop pump and turn off power.
4. Mount pump head on pump body with hex. sock-
et head bolts. (Tightening torque : 2.55 N•m)
- 52 -
Maintenance
4. Optional Accessories
Specification of check valve and back pressure valve
Set pressure Wet-end Applicable Pump wet-end
Model MPa material pump model code
CA-2VC VC
PVC
CA-2VE V6
0.17±0.04 EH-E31, E36
CA-2V
GFRPP PC
CA-2E
CA-3VH VC
PVC
CA-3VEH 0.17±0.04 EH-E46 V6
CA-3VH GFRPP PC
CA-3VCL VC VM
PVC
CA-3VEL 0.05±0.04 EH-E56 V6
CA-3VL GFRPP PC
CS-1S 0.2±0.03 SUS316 EH-E31, E36 SH
CS-2S 0.2±0.03 SUS316 EH-E46 SH
CS-2SL 0.1±0.02 SUS316 EH-E56 SH
0.2±0.02 EH-E31, E36 FC
BVC-1TV PVDF
0.1±0.02 EH-E46, E56 FC
Pump body
(Backup retainer :
for E56 only)
Valve set
Base gasket
Pump head
Pump base
Retainer
Diaphragm
Gasket
Cord nut
Cord gasket
Terminal box
Connector gasket
Controller body
- 53 -
Maintenance
Exploded view of pump head (VC, V6, PC, VM, FC types)
4 4 4
28 28
28
29
29 30
30 29
30
3
3
14 3
14
11 31
14 14
13 18
12 11 11 10
14 9
13 18 13
11 9 12 7
12
13 18 7 14
12 9 14
7 17 17
14
17
1 1 1
17
20 21 14 17
19 17
11 21 14 14
13
20 20 21
19 11 19
12 11
14 13
13
11 12 12
13 14 14
12 31
14 14
3
3 3
30
30 30 29
29 29
28
28 28
4 4 4
- 54 -
Maintenance
Exploded view of pump head (SH type)
EH-E31 · 36 EH-E46 · 56
- 55 -
Maintenance
Exploded view of pump head (HP6 types)
EH-E36
- 56 -
Maintenance
■ Outline dimension (VC, V6, PC, VM, FC)
● EH-E31 (243)
(198) 16.5
OUT
(246)
163
(78)
IN
17
154 (28) 146
114 25.5
7 7
132
● EH-E36 (243)
(198) 16.5
OUT
(246)
(79)
IN
17
114 25.5
7
132
- 57 -
Maintenance
● EH-E46
(247)
(199) 19
OUT
(255)
163
(70)
IN
17
154 (29) 146
114 25.5
7 7
132
● EH-E56 (259)
(209) 21.5
OUT
(266)
163
(59)
IN
17
114 25.5
7 7
132
- 58 -
Maintenance
■ Outline dimension (SH)
● EH-E31SH (249)
(197) (23)
Rc1/4"
OUT
AIR
ODØ4
(300)
(256)
163
Rc1/4"
(97)
IN
17
114 25.5
7 7
132
(249)
OUT
AIR
(303)
(259)
163
(94)
IN
17
114 25.5
7
132
- 59 -
Maintenance
● EH-E46SH (254)
(198) (27)
OUT
AIR
(328)
(276)
163
(92)
IN
114 25.5
7 7
132
● EH-E56SH (265)
(208) (28)
OUT
AIR
(331)
(279)
163
(79)
IN
17
114 25.5
7 7
132
- 60 -
Maintenance
● EH-E36HP6 (245)
(196) (20)
OUT
(255)
163
IN
(75)
17
154 (26) 146
114 25.5
7 7
132
- 61 -
( )Country codes
IWAKI
Electromagnetic Metering Pump
IWAKI CO.,LTD. 6-6 Kanda-Sudacho 2-chome Chiyoda-ku Tokyo 101-8558 Japan
TEL:(81)3 3254 2935 FAX:3 3252 8892(http://www.iwakipumps.jp)
Australia IWAKI Pumps Australia Pty. Ltd. TEL : (61)2 9899 2411 FAX : 2 9899 2421 Italy IWAKI Europe GmbH, Italy Branch TEL : (39)0444 371115 FAX : 0444 335350
Austria IWAKI EUROPE GmbH TEL : (49)2154 92540 FAX : 2154 9254 48 Korea IWAKI Korea Co.,Ltd. TEL : (82)2 2630 4800 FAX : 2 2630 4801
Belgium IWAKI Belgium n.v. TEL : (32)1367 0200 FAX : 1367 2030 Malaysia IWAKIm Sdn. Bhd. TEL : (60)3 7803 8807 FAX : 3 7803 4800
EH-E Type
China IWAKI Pumps (Shanghai) Co., Ltd. TEL : (86)21 6272 7502 FAX : 21 6272 6929 Norway IWAKI Norge AS TEL : (47)23 38 49 00 FAX : 23 38 49 01
China IWAKI Pumps (Guangdong) Co., Ltd. TEL : (86)750 3866228 FAX : 750 3866278 Singapore IWAKI Singapore Pte. Ltd. TEL : (65)6316 2028 FAX : 6316 3221
China GFTZ IWAKI Engineering & Trading (Guangzhou) TEL : (86)20 8435 0603 FAX : 20 8435 9181 Spain IWAKI Europe GmbH, Spain Branch TEL : (34)93 37 70 198 FAX : 93 47 40 991
China GFTZ IWAKI Engineering & Trading (Beijing) TEL : (86)10 6442 7713 FAX : 10 6442 7712 Sweden IWAKI Sverige AB TEL : (46)8 511 72900 FAX : 8 511 72922
Denmark IWAKI Nordic A/S TEL : (45)48 24 2345 FAX : 48 24 2346 Switzerland IP Service SA TEL : (41)26 674 9300 FAX : 26 674 9302
Instruction Manual
Finland IWAKI Suomi Oy TEL : (358)9 2745810 FAX : 9 2742715 Taiwan IWAKI Pumps Taiwan Co., Ltd. TEL : (886)2 8227 6900 FAX : 2 8227 6818
France IWAKI France S.A. TEL : (33)1 69 63 33 70 FAX : 1 64 49 92 73 Taiwan IWAKI Pumps Taiwan (Hsin-chu) Co., Ltd. TEL : (886)3 573 5797 FAX : (886)3 573 5798
Germany IWAKI EUROPE GmbH TEL : (49)2154 9254 0 FAX : 2154 9254 48 Thailand IWAKI (Thailand) Co.,Ltd. TEL : (66)2 322 2471 FAX : 2 322 2477
Holland IWAKI Europe GmbH, Netherlands Branch TEL : (31)547 293 160 FAX : 547 292 332 U.K. IWAKI Pumps (UK) LTD. TEL : (44)1743 231363 FAX : 1743 366507
Hong Kong IWAKI Pumps Co., Ltd. TEL : (852)2 607 1168 FAX : 2 607 1000 U.S.A. IWAKI AMERICA Inc. TEL : (1)508 429 1440 FAX : 508 429 1386
Indonesia IWAKI Singapore (Indonesia Branch) TEL : (62)21 690 6606 FAX : 21 690 6612 Vietnam IWAKI pumps Vietnam Co.,Ltd. TEL : (84)613 933456 FAX : 613 933399
Read this manual before use of product
T656-2 '13/09
Date of preparation Document No. Version
2020-10-20 TCSH-002 Rev.3
Prepared by Reviewed by Approved by QC
Lim, Heonjin Lee, Gwangho Kwon, Kyungahn
Subject Document type
DPD TRO Sensor Manual Technical document
Manual
Techcross
Confidential
Document
i
i 페이지 1/4/2021
Date of preparation Document No. Version
2020-10-20 TCSH-002 Rev.3
Prepared by Reviewed by Approved by QC
Lim, Heonjin Lee, Gwangho Kwon, Kyungahn
Subject Document type
DPD TRO Sensor Manual Technical document
Table of contents
1. Overview ....................................................................................................................... 4
1.1. Revision History ............................................................................................................ 4
2. Product specification ..................................................................................................... 4
3. Safety ............................................................................................................................ 5
3.1. Safety information labelling ........................................................................................... 5
4. Product component ....................................................................................................... 6
5. Measuring principle ....................................................................................................... 7
5.1. Analytical method .......................................................................................................... 7
5.2. Measurement procedure ............................................................................................... 7
6. Installation ..................................................................................................................... 8
6.1. Installation of the product .............................................................................................. 8
6.2. Installation of pipes ....................................................................................................... 8
6.3. Electrical connections ................................................................................................... 9
6.3.1. Power connection ............................................................................................. 10
6.3.2. Communication (RS-485) linkage .................................................................... 10
6.3.3. Analog output (4~20mA) connection ................................................................ 10
6.3.4. DI and DO connection .......................................................................................11
6.4. Installation of reagents ................................................................................................ 12
7. Product operation ........................................................................................................ 14
7.1. Operation .................................................................................................................... 14
7.2. Keypad and Display information ................................................................................. 16
7.3. Calibration ................................................................................................................... 16
7.4. Setting ......................................................................................................................... 18
7.4.1. Menu configuration ........................................................................................... 18
7.4.2. Communication settings ................................................................................... 19
7.4.2.1. Modbus address setting ................................................................................. 19
Confidential
Document
ii
ii 페이지 1/4/2021
Date of preparation Document No. Version
2020-10-20 TCSH-002 Rev.3
Prepared by Reviewed by Approved by QC
Lim, Heonjin Lee, Gwangho Kwon, Kyungahn
Subject Document type
DPD TRO Sensor Manual Technical document
Confidential
Document
iii
1. Overview
This manual explains DPD TRO Sensor that measures TRO(total residual oxidant) content in the water. The manual’s
contents are correct by thorough inspection or repair. However, they can be a little bit different from the product if
it is improved.
2. Product specification
Table 1. Product specification
range
16 Humidity 0~95% Relative Humidity (Non-Condensing)
17 Altitude ≤ 2000 m
3. Safety
This manual explains the basic guidelines that must generally be implemented in the product including the
installation, operation, and maintenance. Please make sure to read all the contents of the manual before unpacking,
installing, or operating the product. Please make sure to pay attention to all kinds of danger and warnings. Please
be informed that noncompliance may result in user or equipment damage.
4. Product component
Product component can largely be divided into 6 parts as shown in the following Figure 1.
1) Sample water inlet section: This section plays a role of supplying sample water to measurement cells according
to measurement procedures.
2) Reagent supply section: This section plays a role of supplying reagents to measurement cells by compressing
pleated reagent packs.
3) Measurement section: This section consists of LED and a photo sensor, and a peristaltic pump and plays a role
of mixing sample water with reagents in measurement cells and analyzing TRO concentration.
4) Control section : This section is responsible for overall operation and control of this equipment and plays a role
of providing user interface.
5) Sample water outlet section : This plays a role of discharging sample water in the measurement cells.
6) Air vent : This vent prevent sample water from flowing backward being connected to the measurement cells and
makes them smoothly discharged.
Control
section
Air vent
Reagent
supply section
Sample water
inlet section
Measurement
section
Sample water
Figure 1. Block diagram of the product
drain port
5. Measuring principle
5.1. Analytical method
DPD TRO Sensor (Hereafter, referred to as ‘DTS’) is equipment that measures TRO in the water. It uses the DPD
colorimetric method that does DPD(N,N-diethyl-p-phenylenediamine sulfate), the indicator, and buffer solutions.
When the suitable amount of an indicator and a buffer solution are mixed with a sample water, they are changed
to red in proportion to concentration of TRO in the water. It measures and digitizes the concentration by using
LED(Sources of light) and inform users of them. TRO concentrations are calculated by differences in absorbance.
1) Flushing : When measurement begins, open the inlet and outlet valves and flush it.
2) Measure the zero (0) point: Put the suitable quantity of the sample water in a measurement cell by closing the
discharge valve and adjusting the inlet valve and measure the zero point.
3) Discharge: Discharge the sample water measured to be the zero point and prepare the following procedure.
4) Measure the inflow rate of the sample water : Check the flow rate of the sample water that are now flowed into
equipment by controlling the inlet valve.
5) Put a reagent in equipment : Put the suitable amount of the sample water in the measurement cell by controlling
the inlet value according to the quantity that they are flowed into equipment. Put the reagent in equipment in the
middle of this process so that the it can react to the sample water.
6) Measure TRO : Analyze absorbance of the sample water developed by the reagent.
7) Discharge: When measurement ends, discharge the sample water from the measurement cell and wait for the
next measurement command.
6. Installation
6.1. Installation of the product
This product was designed to be installed by attaching it to the wall surface. If there is no wall surface, you can use
it by installing it on a good place for adhesion. But, it is recommended to install it in the closest place to the
position of sample water sampling.
When equipment is installed, it is recommended to fix it with M6(1/4”) screws by using mounting holes. Please see
the following Figure 2.
When sample water are discharged, they are discharged being mixed with reagents. It is the best method
to ask local authorities about how to dispose of waste even though used chemicals are in low concentration.
Connection
of the air vent
(10mm Tubing connection)
Electrical wiring must be worked by specialists according to the instructions. Please connect equipment
according to electricity rules of countries, states, and regions.
The DO function shows the measurement and abnormal state of this equipment. When the outputs are enabled,
the voltage of 24VDC is output. Connect the cable to the No.11(+ Output of 24VDC – DO1) and No.12(+ Output
of 24VDC - DO2)에 of the terminal block in order. For cable specifications, it is recommended to use 20AWG. Please
see the following Figure 5.
1) Connect the pipes of the reagent pack : Put the reagents dissolved in the reagent pack and connect it to a
Pharmed tube and Reducer Barb fittings.
- If the pharmed tube connected to the reagent pack is not inserted into the peristaltic pump, the reagents can
be leaked. Please be careful.
- Please close the lid tightly to prevent reagent from leaking.
Barb fitting
Barb fitting
1/16”
3/16”
2) Lift the lever of Peristaltic pump to prepare for inserting the pharmed tube.
3) Insert the tube into the peristaltic pump as shown in Figure 8. and lower the lever to secure the pharmed tube.
4) Secure the block to the top of the pump so that the pharmed tube is fully seated. (secure with screw)
5) Install the reagent pack: Lift up the compressor of the reagent pack.
And insert the Silicon tube(1/8”) into Barb fittings of the reagent pack which were in advance prepared.
6) Fix the reagent pack: Insert the reagent pack into the compressor by lifting it upside down.
If the pharmed tube connected to the reagent pack is not inserted into the peristaltic pump, the reagents
can be leaked. Please be careful.
7) Inject reagents: Execute the automatic injection of reagents by pressing the top menu button. For more
information, please see the contents of 7.3.4.2.
7. Product operation
7.1. Operation
Check if all the pipes and electrical wiring were completely installed before operating the product. When
everything’s all set, prepare for measurement by powering equipment. When you power equipment, the software
version is shown and is changed to the screen to prepare measurement in the product and. Please see Figure 12.
You can start measurement after that. Measurement is done according to the normal procedure. The following
procedure explains the general measurement procedure.
1) Start measurement : You can start measurement with two methods in this product. When measurement starts,
measurement is arranged during a cycle and all the process is automatically done.
1-1) Supply DI signal: The voltage of 24VDC is referred to a start signal through the DI terminal block. When
the start signal is referred, the product immediately starts measurement.
1-2) RS-485 Communication : It’s the method to give a start signal by using RS-485 communication. When the
start signal is transmitted to communication like DI input, measurement immediately begins.
2) Bypass & Flushing : When measurement begins, Bypass is arranged as sample water are discharged from the
outlet. And at the same time, the measurement cells are flushed. The bypass time can be set by 5 second in the
Menu by users.
3) Measure the 0 point: Put a sample water in the measurement cell and start measuring the zero point by
turning on LED while it is not mixed with a reagent after stopping equipment. The zero point is adjusted every
cycle. And when measurement ends, the sample water is discharged from the outlet.
4) Analyze TRO: Put the flow rate of the sample water that is now flowed into equipment in the measurement cell
by dividing it into proper amounts. Operate equipment by putting liquid forms of the buffer solution and the
indicator simultaneously so that they can well be mixed with the sample water. When the proper amount is put in
the measurement cell, measure and analyze current TRO.
5) Discharge the sample water: When the analysis of TRO of the sample water is done, it is digitized through the
display window and is shown to users. And the sample water that the analysis is finished is discharged. After that,
it waits for next measurement.
1 2 3 4
Figure 13. Keypad and Display
7.3. Calibration
When the measured values of the instrument need to be calibrated, can adjust the values according to the following
procedures.
1) For the samples of sample water that you try to measure, measure TRO by using other equipment(Use equipment
that has been calibrated) that becomes the criteria (Pocket photometers, iodometry, etc.).
2) Press the mode key after measurement of the product is at least once done. Values cannot be adjusted while
TRO never be once measured.
3) When it is changed to the ADJUST screen, go to it by pressing the set key. At this moment, the screen to ask
your password appears. This means that you cannot adjust values except the permitted users. The first setting
password is ‘1111’.
4) When you go to the screen to enter a password, the positional number farthest to the right flickers. Can adjust
the numbers with Up and Down keys and move the set key to the left positional number. Enter 1111 in the screen
with the same method and press the set key for more than 2 seconds. If you enter the wrong number, the display
of entering the right password is again shown with an expression, ‘WRONG’.
5) When you get to enter the right password, you can go to the screen to be able to adjust measured values of
TRO. The different point from the measurement screen is that ‘A’ to display ‘Adjust’ is shown in the very left. The
measured value of TRO that was lastly measured is shown. And the measured value can be adjusted by Up and
Down keys.
6) When you press the set key for more than 2 seconds after finishing the adjustment of the measured values, the
7.4. Setting
7.4.1. Menu configuration
This product provides various setting menus to do convenience of measurement and each menu’s movement and
setting can be executed through the keypad under the display window. General menu configuration is shown in
the following figure 14. And the keypad information necessary for moving the menu is shown. For more information,
please see the contents after 7.3.2.
2) Go to the screen to be able to adjust Modbus address by pressing the set key in the M.ADDR. screen.
3) When the number displayed in the screen flickers by the set key, you can adjust the Up and Down keys.
4) Designate the set value by pressing the set key after finishing the adjustment.
2) Go to the screen to be able to change the baud rate by pressing the set key in the B.RATE screen.
3) When the number displayed in the screen flickers by pressing the set key you can adjust the Up and Down key.
4) Save the set value by pressing the set key for more than 2 seconds after the adjustment.
2) Go to the screen to be able to adjust Bypass Time by pressing the set key in the B.TIME screen.
3) When the number displayed in the screen by pressing the set key flickers, you can adjust the numbers with the
Up and Down keys.
4) Save the set value by pressing the set key for more than 2 seconds after finishing the adjustment.
This Menu must be executed after installing or changing new reagent packs.
1) Go to the INJECT screen by pressing the Down key once in the B.TIME screen.
2) Go to the FILL screen by pressing the set key in the INJECT screen.
3) When you press the set key for more than 2 seconds in the FILL screen, reagents are automatically injected. And
after finishing them, the OK screen is displayed and the injection of the reagents ends.
2) Go to the 4mA screen by pressing the set key in the 4-20mA screen. At this moment, when pressing the Down
key, go to the 20mA screen.
3) Go to the screen to be able to adjust 4~20mA output to match measured TRO by pressing the set key in the
4) When the number displayed in the screen by pressing the set key flickers, you can adjust the number with the
Up and Down keys. Save the set value by pressing the set key for more than 2 seconds after the adjustment.
2) Go to the execution screen by pressing the set key in the F.RESET screen. Factory reset is executed when pressing
the set key for 2 seconds in the YES screen. When factory reset is executed, the set values in the product are
initialized as the state that all the things are released.
8. Trouble shooting
8.1. Alarm
DTS has its own function to monitor and judge errors. If you the problems are solved, the alarms are automatically
released. For the Alarm lists and how to solve problems, please see the following table.
9. Maintenance
9.1. Periodic maintenance
Table 4. List of periodic maintenance
Tubing Replacement every year (recommend) Please refer to 6.4 2), 3), 4)
Alarm switch
for using
Figure 15. Switch to monitor reagents
reagents.
2) Abnormal condition of reagent : reagent replacement must be performed in the following cases
- Exceeds the recommended replacement cycle of two months.
- If the color of the reagent is excessively discolored.
- When measured TRO concentration is significantly lower than portable TRO sensor.
- If the ambient temperature where the DTS is installed is too high and more than one month after the reagent
change.
Cover of the
measurement cell
Brush for
washing
2) Abnormal condition of measurement cell : Cell cleaning must be performed in the following cases.
- If the cell is contaminated by foreign matter when visually checked.
- There is no problem with the reagents, but the measured TRO concentration is significantly lower than the other
sensor.
- If an alarm occurs. ( please see 8.1 “E.SENSE” )
Clean the
Figure 17. Strainer cleaning filter.
2) Abnormal condition of strainer : Strainer cleaning must be performed in the following cases.
- If the filter of strainer is contaminated by foreign substance when visually checked.
- If the strainer is clogged by foreign objects.
USER MANUAL
MANUALE D’USO
PHOENIX
PHOENIX ATEX
PHOENIX FOOD
PHOENIX FOOD ATEX
INDEX indice
4 IDENTIFICATION CODE sigla di identificazione
TRANSPORT trasporto
STORAGE stoccaggio
INSTALLATION installazione
START UP avviamento
USE utilizzo
STOP arresto
44 MAINTENANCE manutenzione
RECOMMENDATIONS raccomandazioni
DISASSEMBLY smontaggio
INSPECTION ispezione
DISPOSAL smaltimento
possibili cause
2 Rev.02
DICHIARAZIONE DI CONFORMITA’
DECLARATION OF CONFORMITY
MARCATURA ATEX
MARKING ATEX II 3/3 GD c IIB T135ºC
MARCAR ATEX
PHOENIX
Questo prodotto è conforme alle seguenti direttive CE/EX e relativi standard armonizzati:
This product complies with the following European Community Directives CE/EX and relating harmonized standards:
2006/42/CE Direttiva Macchine / Machinery Directive / Maschinenrichtlinie / Directive Machines / Directiva Máquinas
2014/34/UE: Direttiva ATEX, concernente il ravvicinamento delle legislazioni degli Stati Membri relative agli apparecchi e sistemi di protezione destinati a essere utilizzati in
atmosfera potenzialmente esplosiva.
UNI EN ISO 12100:2010 – Sicurezza del macchinario. Concetti fondamentali, principi generali di progettazione. Parte 1: terminologia di base, metodologia.
UNI EN ISO 12100:2010 – Safety of the machinery. Fundamental notions, design general principles. Part 1: Basic terminology, methods.
UNI EN ISO 12100:2010 – Sicurezza del macchinario. Concetti fondamentali, principi generali di progettazione. Parte 2: principi tecnici.
UNI EN ISO 12100:2010 – Safety of the machinery. Fundamental notions, design general principles. Part 2: Technical principles.
UNI EN ISO 3746: 2011 – Acustica. Determinazione dei livelli di potenza sonora delle sorgenti di rumore mediante misurazione della pressione sonora. Metodo di controllo con una superficie
avvolgente su un piano riflettente.
UNI EN ISO 3746: 2011 – Sound. Determination of sound power levels for noise sources by measuring the sound pressure. Monitoring method with an enveloping surface on a reflecting plate.
UNI EN ISO 11200: 2014 – Acustica. Rumore emesso dalle macchine e dalle apparecchiature. Linee guida per l’uso della norme di base per la determinazione dei livelli
di pressione sonora al posto di lavoro e in altre specifiche posizioni.
UNI EN ISO 11200: 2014 – Sound. Noise done by the machines and the equipments. Guidelines for using the basic norms for determining the sound pressure levels in
the working place and in other specific positions.
UNI EN ISO 4414: 2012 - Sicurezza del macchinario. Requisiti di sicurezza relativi a sistemi e loro componenti per trasmissioni oleoidrauliche e pneumatiche. Pneumatica.
UNI EN ISO 4414: 2012 - Safety of the machinery. Safety requirements concerning the systems and their components for hydraulic and pneumatic transmissions. Pneumatic.
UNI CEI EN ISO 80079-36:2016 - Apparecchi non elettrici destinati ad essere utilizzati in atmosfere potenzialmente esplosive. Parte 1: metodo e requisiti di base.
UNI CEI EN ISO 80079-36:2016 - Non-electrical equipments to be used in potentially explosive environments. Part 1: Method and basic requirements.
UNI CEI EN ISO 80079-37:2016 - Apparecchi non elettrici per atmosfere potenzialmente esplosive. Parte 5: protezione per sicurezza costruttiva “c”.
UNI CEI EN ISO 80079-37:2016 – Non-electrical equipments for potentially explosive environments. Part 5: protection for building safety “c”.
LA SEGUENTE CONFORMITA’ E’ RIFERITA AL PROTOTIPO DELLA PHOENIX 100 MATRICOLA NR. P0001 DEL 16.01.2012.
THIS COMPLIANCE REFERS TO PHOENIX 100 PROTOTYPE, SERIAL NUMBER P0001 OF 16.01.2012.
ESTENSIONI: la presente dichiarazione si estende anche ai modelli PHOENIX 7, PHOENIX 18, PHOENIX 30, PHOENIX 50, PHOENIX 65, PHOENIX 100, PHOENIX 101, PHOENIX 160,PHOENIX
250, PHOENIX 400,PHOENIX 500, PHOENIX 700, IN PLASTICA E METALLO.
EXTENSION: this declarations is also valid for the following versions PHOENIX 7, PHOENIX 18, PHOENIX 30, PHOENIX 50, PHOENIX 65, PHOENIX 100, PHOENIX 101, PHOENIX 160,PHOENIX
250, PHOENIX 400,PHOENIX 500, PHOENIX 700, MADE OF METAL OR PLASTIC.
ATTENZIONE: data l’innumerevole varietà di prodotti e composizioni chimiche, l’utilizzatore è ritenuto il maggior conoscitore delle reazioni e compatibilità con i materiali
costruttivi della pompa. Pertanto, prima dell’impiego, eseguire con perizia tutte le verifiche e prove necessarie al fine di evitare situazioni pericolose anche se remote che
non possono essere conosciute ed imputabili al costruttore. Per ogni controversia il Foro Competente è quello di Varese.
WARNING: since there exists an endless variety of products and chemical compositions, the user is presumed to have the best knowledge of their reaction and
compatibility with the materials used to build the pump. Therefore, before using the pump, all the necessary checks and tests must be performed with great care to avoid
even the slightest risk, an event that the manufacturer cannot foresee and of which he cannot be held responsible. Any controversy lies within competence of the Court of
Varese.
DATA 12/10/2017
3 Rev.02
IDENTIFICATION CODE sigla di identificazione
PUMP PUMP AIR FLUID BALL VERSION
BALLS O-RING CONNECTION
MODEL BODY DIAPHRAGM DIAPHRAGM SEATS ATEX ZONE1
PF/SP18 S - AISI 316 H - HYTREL T - PTFE T - PTFE S - SS AISI 316 T - PTFE 1 - BSP THREATED X
PF: S-SS AISI 316 2 - FLANGED
PF/SP30 ELECTROPOLISCHED 3 - CLAMP
PF/SP50 SP:
MECCANICHALLYLLY
PF/SP100
POLISCHED
PF/SP160
PF/SP400
PF/SP500
PF/SP700
PF/SP1000
Each pump is supplied with the serial and model abbreviation and the serial number on the rating
plate, applied onto the support side. Check these data upon receiving the goods. Any discrepancy
between the order and the delivery must be communicated immediately. In order to be able to
trace data and information, the abbreviation, model and serial number of the pump must be
quoted in all correspondence.
Ogni pompa spedita è contrassegnata dalla sigla della serie, dal modello e dal numero di serie impressi sulla
targhetta di identificazione. Si prega di confermare i dati impressi quanto prima dal ricevimento della merce. Ogni
discrepanza tra l’ordine e le informazioni impresse sulla targhetta deve essere immediatamente comunicata. Al fine
di consentire la perfetta rintracciabilità dei dati e delle informazioni, ogni corrispondenza deve riportare la sigla, il
modello e il numero di serie della pompa in oggetto.
MODEL
modello
YEAR OF MANUFACTURE
Anno di fabbricazione
SERIAL No
Numero di serie
ATEX CERTIFICATION certificazione Atex
PHOENIX ATEX - PHOENIX FOOD ATEX - CERT. ZONE 2 II 2/2 GD c IIB T135°C
4 Rev.02
GENERAL NOTES
“PHOENIX” series pumps are air-operated, double-diaphragm positive-displacement pumps, designed and manufactured
for pumping fluids that are chemically compatible with the constructive materials of the pump. The characteristics of the
fluid (pressure, temperature, chemical reactivity, specific weight, viscosity, vapor pressure) and of the environment must
be compatible with the pump characteristics and are defined in the ordering phase. Fluimac is not responsible for the
pumped liquid. The customer must ensure that there is compatibility between the pumped liquids and pump materials.
“PHOENIX” series pumps are self-priming; at the start-up the pipes can be empty.
The declared dry negative suction is referred to intake of water at a temperature of 20°C/ 68°F. The priming time and
the diaphragm’s life depend on.
- the suction circuit (total length and diameter)
- specific weight of the pumped fluid
- viscosity of the pumped fluid
- negative suction: max 5.000 cps (at 18°C / 64,4°F)
- below head suction: max 50.000 cps (at 18°C / 64,4°F) “PHOENIX” series pumps may be used dry
SS 0÷40°C / 14÷104°F
SS 95°C / 203°F 80°C / 176°F
The pump may be operated at a maximum pressure equal to 1.5 times the head value with closed delivery.
The value of the vapor pressure of the pumped fluid must be greater (of at least 3 mwc - meters of water column) than
the difference between the total absolute head value (pressure on suction level subtracted of the suction height) and
the leakages of the suction section.
The pumped fluid may contain particles suspended in different concentrations in accordance with the type of valve
assembled:
MAX DIM.
mm. 10
2,5 3 3,5 3,5 7,5 8 8,5
37 Rev.02
NOTE GENERALI
Le pompe serie "PHOENIX" sono pompe pneumatiche volumetriche a doppia membrana, progettate e costruite per il
pompaggio di liquidi compatibili chimicamente con i materiali costruttivi della pompa. Le caratteristiche del liquido
(pressione, temperatura, reattività chimica, peso specifico, viscosità, tensione di vapore) e dell’ambiente devono essere
compatibili con le caratteristiche della pompa e sono definite in sede d’ordine. La Fluimac s.r.l non è responsabile dei
liquidi pompati. Il cliente deve assicurarsi che ci sia compatibilità tra i liquidi pompati e i materiali della pompa.
Le pompe serie “PHOENIX” sono autoadescanti, all’avviamento le tubazioni possono essere vuote
La capacità di aspirazione negativa a secco dichiarata è riferita al pescaggio di acqua a temperatura di 20°C/
68°F.
Le pompe serie “PHOENIX” non possono essere usate per creare il vuoto
Assicurarsi che le caratteristiche fisico-chimiche del liquido siano state attentamente valutate.
La temperatura massima riferita ad acqua in funzionamento in continuo dipende dalla versione dei materiali
(riportata in targhetta) e dall’ ambiente in cui verrà installata la pompa:
VERSIONE MAX TEMP. MAX TEMP. L’intervallo di temperatura ambiente è in funzione della versione dei
ATEX ZONE 2 ATEX ZONE 1
materiali (riportata in targhetta):
Il valore della tensione di vapore del liquido pompato deve essere superiore (di almeno 3 m c.a.) alla differenza tra il
battente totale assoluto (pressione sul livello in aspirazione sottratta dell’ altezza di aspirazione) e le perdite nel tratto di
aspirazione
Il liquido pompato può contenere solidi sospesi in misura diversa in relazione al tipo di valvola montata:
MAX DIM.
2,5 3 3,5 3,5 7,5 8 8,5 10
mm.
38 Rev.02
OPERATING PRINCIPLE principio di funzionamento
The pneumatic distribution system sends compressed air behind one of the two diaphragms (A), which
pushes the fluid towards the delivery circuit. Simultaneously, the opposite diaphragm (B) is in the intake
phase since it is dragged by the shaft that connects it to the other diaphragm (A) under pressure; air
present behind it is discharged into the environment through the flow rate regulator present on the
pump , while a pressure drop is created in the fluid chamber which sucks the fluid from the suction
circuit. When the diaphragm (A), under pressure, reaches the stroke limit, the distributor switches the
two inputs to the chamber on the diaphragms air side, putting diaphragm (B) under pressure and
diaphragm(A) in discharge. When the pump reaches its original starting point, each diaphragm has
carried out one air discharge stroke and one fluid delivery stroke. This sequence of movements makes
up a complete pumping cycle.
Il sistema di distribuzione pneumatico invia l’aria compressa dietro una delle due membrane(A), la quale spinge il fluido
verso il circuito di mandata. Contemporaneamente la membrana opposta(B) si trova in fase di aspirazione essendo
trascinata dall’albero che la collega all’altra membrana(A) sotto pressione; l’aria presente dietro di essa viene scaricata in
atmosfera attraverso il regolatore di prestazioni presente sulla pompa mentre nella camera del fluido si crea un
abbassamento di pressione che risucchia il fluido dal circuito di aspirazione. Quando la membrana(A), sotto pressione,
raggiunge il limite della corsa il distributore commuta i due ingressi alle camera lato aria delle membrane, mandando in
pressione la membrana(B) ed in scarico la membrana(A) Nel momento in cui la pompa raggiunge il suo punto di
partenza originale, ogni membrana ha compiuto una corsa di scarico aria e una di mandata liquido. Questa sequenza di
movimenti costituisce un ciclo di pompaggio completo.
WARNING: the pneumatic supply of “PHOENIX” series pumps must be carried out with oil-free, filtered, dry and
unlubricated air Avoid pressure drops by using pipes and adjusting and controlling elements having characteristics suitable
for the pump In case of installation in atex zone, the compressor must suck air from outside the area classified as atex or
use inert gas.
ATTENZIONE: l’alimentazione pneumatica delle pompe serie “PHOENIX” deve essere eseguita con aria disoleata, filtrata, essiccata e
non lubrificata Evitare cadute di pressioni impiegando tubi ed elementi di regolazione e controllo con caratteristiche adeguate a
quelle della pompa In caso di installazione in zona ATEX il compressore deve aspirare aria esternamente alla zona classificata ATEX
ovvero utilizzare gas inerte
39 Rev.02
INSTALLATION AND USE INSTRUCTIONS
TRANSPORT
• cover the hydraulic connections
• lift the hydraulic plastic parts without mechanical stress
• for transport on irregular roads, cushion the bumps with suitable support plane
• blows and impacts may damage parts that are important for the machine operation and safety
STORAGE
• In case you need to get away for a period of time the pumps before installation, store in original boxes. The boxes
should be stored off the ground, in a closed, clean and dry.
• in the event that the packaging has not received any is intact, it will be necessary to free the pump from it-Checking
the integrity and restore a new packaging
• The storage place should be closed environment with a temperature not lower than -5 ° C, not more than 40 ° C
and with a moisture content not exceeding the value of 80%; any packaging must not be subjected to shocks, vibrations
and loads above you
INSTALLATION
• it is essential for the pump self-priming operation that the hydraulic system is leak-proof
• clean the system before connecting the pump
• the pump must not contain foreign bodies and all the seals on the hydraulic connections must be removed
• before starting the pump check the tightness of the screws of the pump bodies and manifolds
• the pump positioning is horizontal, the fluid delivery manifold must always be positioned in the upper part (see arrows on the pump casing)
• fastening may be on the floor or on the ceiling
• position the pump the closest possible to the point of collection
STOCCAGGIO
• nel caso sia necessario staccare per un periodo di tempo le pompe prima dell’installazione, conservarle nelle casse
originali. Le casse devono essere stoccate sollevate dal suolo, in un ambiente chiuso, pulito ed asciutto.
• nel caso che al ricevimento l’eventuale imballo non risulti integro, occorrerà liberare la pompa dall’imballo, verificar-
ne l’integrità e ripristinare un imballo nuovo
• il luogo di stoccaggio dev’essere un ambiente chiuso con temperatura non inferiore ai –5°C, non superiore ai 40°C
e con un tasso di umidità che non superi il valore di 80%; l’eventuale imballo non dev’essere sottoposto ad urti,
vibrazioni e carichi sovrastanti
INSTALLAZIONE
• è fondamentale per l’operazione di autoadescamento della pompa che l’impianto idraulico sia a perfetta tenuta
• pulire l’impianto prima di collegare la pompa
• la pompa non deve contenere corpi estranei e devono essere rimossi i sigilli sulle connessioni idrauliche
• prima di avviare la pompa verificare il serraggio delle viti dei corpi pompa e dei collettori
• il posizionamento della pompa è in orizzontale; il collettore di mandata del fluido deve essere posizionato sempre
nella parte superiore (vedi frecce su corpi pompa)
• il fissaggio può essere a pavimento o a soffitto
• posizionare la pompa più vicino possibile al punto di prelievo
40 Rev.02
use the plant solutions indicated in the following diagram:
1. YES: use flexible pipes reinforced with rigid spiral to connect the hydraulic circuit of the pump. Rigid piping may
cause strong vibrations and manifolds breaking. Do not use pipes with nominal diameter smaller than the
diameter of the pump connections. For negative installations and/or viscous fluids use pipes with greater
diameter related to the nominal diameter of the pump.
2. YES: pulse damper
3. YES: gate valve for delivery adjustment
4. YES: intake for gauge or protection pressure switch
5. YES: pipe anchoring
6. YES: shut-off valve
7. NO: air pockets; the circuit must be linear and short
8. YES: discharge duct around the base
9. YES: wide and rigid filtering separator in case of open tanks
10. YES: wide and rigid filtering separator in case of open tanks
11. Make it as short as possible the length of the horizontal S no vent for the air
12. Slope of the pipe to the pump
13. n the flow velocity of the fluid max. 3.5 m / s
42 Rev.02
START UP
• check the correct execution of what indicated in the INSTALLATION paragraph
• check that the intake and delivery pipes of the hydraulic circuit are correctly connected
• open the intake and delivery valves of the pump hydraulic circuit
• open the 3-way valve on the air circuit
• set the operation point requested for the pump: properly adjust the air pressure and delivery that supplies the pump.
With pressure values under 2 bar the pump may stall, with pressure values above 7 bar it is possible that breakdowns and/or
yields may
occur with consequent spillage of the pumped fluid
• for pumps with split manifold the two pumped fluids must have the same viscosity value, very different viscosity values may lead to
stall
problems and/or diaphragms breaking
• do not operate at the limits of the operation curves: the maximum head or maximum delivery (total absence of leaks and intake
height in
the delivery circuit)
• check that there are no anomalous vibrations or noise due to the too elastic support structure, unsuitable fastening or cavitation
• after 2 hours of operation stop the pump correctly and check the tightening of all the bolts on the pump
AVVIAMENTO
• verificare la corretta esecuzione di quanto indicato in INSTALLAZIONE
• verificare che i tubi di aspirazione e mandata del circuito idraulico siano correttamente collegati
• aprire le valvole in aspirazione e mandata del circuito idraulico della pompa
• aprire la valvola a 3 vie sul circuito aria
• Impostare il punto di funzionamento per il quale la pompa è stata richiesta: regolare opportunamente la pressione e la portata dell’
aria che
alimenta la pompa. Con pressioni inferiori ai 2 bar la pompa può andare installo, con pressioni superiori ai 7 bar si possono
verificare
rotture e/o cedimenti con conseguente fuoriuscita del fluido pompato
• Nel caso di pompe con collettore sdoppiato i due fluidi pompati devono avere viscosità dello stesso ordine di grandezza;
viscosità sensibilmente differenti possono portare a problemi di stallo e/o rottura delle membrane
• non operare agli estremi della curva di funzionamento: massima prevalenza o massima portata (totale assenza di perdite e di
altezza di
sollevamento nel circuito di mandata)
• constatare l’assenza di anomali vibrazioni o rumorosità dovute a struttura portante troppo elastica, inadeguato fissaggio o
cavitazione
• dopo 2 ore di funzionamento arrestare correttamente la pompa e verificare il serraggio di tutti i bulloni della pompa
USE
• do not operate valves or shunts during the pump operation
• Risk of harmful water hammers in case of incorrect or sudden operations (valves must be operated only by trained personnel)
• empty and wash accurately inside the pump in case different fluids must be pumped
• insulate or empty the pump if the fluid crystallization temperature is equal to or below the ambient temperature
• stop the pump if the fluid temperature exceeds the maximum allowed temperature indicated in the GENERAL NOTES;
if the exceeding temperature is about 20% it is necessary to inspect the status of the internal parts
• stop the pump and close the valves in case of leaks
• Wash with water only if chemical compatibility allows it ; alternatively use the suitable solvent that does not generate hazardous
exothermic
reactions
• consult the fluid supplier to decide the most suitable fire-prevention method
• Empty the pump in case of long periods of disuse (particularly with fluids which are particularly tending to crystallize)
• check that there is no gas in the delivering fluid, if there is stop the pump
UTILIZZO
• non azionare valvole o derivazioni durante il funzionamento della pompa
• rischi di dannosi colpi d’ariete in caso di errate o improvvise manovre (azionamento valvole solo da personale ad- destrato)
• svuotare e lavare accuratamente l’interno pompa in caso di pompaggio di liquidi diversi
• isolare o svuotare la pompa se la temperatura di cristallizzazione del liquido risulta uguale o inferiore alla tempera- tura ambiente
• fermare la pompa se la temperatura del liquido supera la temperatura massima consentita indicata nelle NOTE GENERALI;
se l’aumento è dell’ordine del 20 % è necessario ispezionare lo stato delle parti interne
• arrestare la pompa e chiudere le valvole in caso di perdite
• lavare con acqua solo se la compatibilità chimica lo consente; in alternativa utilizzare il solvente appropriato che non generi
reazioni
esotermiche pericolose
• contattare il fornitore del liquido per stabilire il più appropriato metodo anti-incendio
• svuotare la pompa in caso di lunghi periodi di non utilizzo (in particolare con liquidi con forte tendenza a cristallizzare)
• controllare che nel fluido in mandata non sia presente gas, in tal caso arrestare la pompa
STOP
To stop the pump, operate exclusively on the air supply closing the 3-way valve, discharging in this way residual pressure from the
pneumatic system of the pump.
WARNING never stop the pump by totally closing the suction and/or delivery valves of the hydraulic circuit
WARNING is prohibited to stop the pump by means of the complete closure of the valves in the suction and / or discharge of the
hydraulic circuit
ARRESTO
Per arrestare la pompa agire esclusivamente sull’ alimentazione dell’aria chiudendo la valvola a 3 vie, scaricando così la pressione
residua dell’ impianto pneumatico della pompa
ATTENZIONE è vietato arrestare la pompa mediante la chiusura completa delle valvole in aspirazione e/o mandata del circuito
idraulico.
43 Rev.02
MAINTENANCE
- all the operation must be carried out by qualified personnel
- do not carry out maintenance and/or repairs with the air circuit under pressure
- carry out periodic inspections (2 ÷ 30 days in accordance with the fluid pumped) to check the filtering elements cleaning
- carry out periodic inspections (3 ÷ 5 months in accordance with the fluid pumped and with the environment conditions)
to ensure the correct operation of the system start/stop units
- the presence of fluid under the pump casing may indicate failures to the pump
- damaged parts must be replaced with complete original parts and not with repaired parts
- the replacement of damaged parts must be carried out in a clean and dry place
RECOMMENDATIONS
WARNING: before performing any maintenance or repair work on the pump, disconnect the pump from the air supply line.
Disconnect the hydraulic connections and discharge the product that is being pumped
- all the operation must be carried out by qualified personnel
- use gloves, goggles and acid-resistant clothing when disconnecting from the system and washing the pump
- wash the pump before carrying out maintenance operations
- do not disperse the washing waste into the environment
DISASSEMBLY
- Bolts are the type with right thread
- Clean all the pump external surfaces using a damp cloth Ball seats removal (for all models)
- Remove the delivery and intake manifolds removing the tightening bolts
- Pull off the seats, the balls and the related cages
- Check the condition of the gasket
Diaphragms removal
PHOENIX 07-18-30-50-65-100-101-160-250-500-700-1000
- Remove the delivery and intake manifolds removing the tightening bolts
- Remove the deposits from the internal surfaces
- Remove the two pump casings
- Remove the plates that lock the diaphragms
- Remove the diaphragms and the related backing plates
- If it would be necessary to disassemble the shaft, remove one of the two diaphragms on air side and then pull off the shaft
INSPECTION
Check the absence of:
• excessive abrasion of the thermoplastic parts
• clots and/or agglomerates due to the pumped fluid
• deformations and/or surface lesions of the diaphragms
• deformations and/or breakages on the valve seats
Replace the parts: broken, cracked, deformed.
Reopen all the clogged ducts and eliminate any chemical agglomerates.
Clean all the surfaces before reassembly, particularly the OR gaskets seats (risk of leaks for dripping).
44 Rev.02
MANUTENZIONE
- tutti gli interventi devono essere presidiati da personale qualificato
- non effettuare manutenzione e/o riparazioni con il circuito aria in pressione
- effettuare ispezioni periodiche (2 ÷ 30 giorni in funzione del liquido convogliato) di pulizia sugli elementi filtranti
- Effettuare ispezioni periodiche (3 ÷ 5 mesi in funzione del liquido convogliato e delle condizioni ambientali) sulla funzionalità
degli organi di consenso/arresto dell’impianto; garantirne l’efficienza
- la presenza di liquido sotto il corpo pompa può essere indizio di guasti alla pompa
- le parti danneggiate devono essere sostituite con parti originali integre e non riparate
- la sostituzione di parti danneggiate deve essere effettuata in ambiente pulito e asciutto
RACCOMANDAZIONI
ATTENZIONE: prima di intervenire sulla pompa per qualsiasi intervento di manutenzione o riparazione, scollegare dalla
pompa la linea aria di alimentazione. Staccare le connessioni idrauliche e scaricare il prodotto che si sta pompando
- tutti gli interventi devono essere presidiati da personale qualificato
- utilizzare guanti, occhiali ed indumenti antiacido in fase di scollegamento dall’impianto e lavando la pompa
- lavare la pompa prima di effettuare operazioni di manutenzione
- non disperdere nell’ambiente lo scarico di lavaggio dei corpi
SMONTAGGIO
- La bulloneria è del tipo a filetto destro
- Provvedere alla pulizia esterna di tutte le superfici della pompa con un panno umido
ISPEZIONE
Verificare l’assenza di:
• eccessiva abrasione delle parti in materiale termoplastico
• grumi e/o agglomerati dovuti al liquido pompato
• deformazioni e/o lesioni superficiali delle membrane
• deformazioni e/o rotture sulle sedi valvola
Sostituire le parti: rotte, fessurate, deformate.
Riaprire tutti i condotti occlusi ed eliminare gli eventuali agglomerati chimici.
Pulire tutte le superfici prima del rimontaggio, in particolare le sedi di guarnizioni OR (rischio di perdite per gocciola- mento).
45 Rev.02
SAFETY RISKS
WARNING! CHEMICAL RISK. Pumps are intended for operation with different types of fluids and chemical solutions. Follow the specific internal
instructions for decontamination during the inspection or maintenance operations.
WARNING! ELECTRICAL RISK. The pump must always be earthed independently to other members connected to it. Safety requirements and
explosion risk prevention are not fulfilled if the pump is not earthed or is incorrectly earthed.
WARNING: the diaphragms (into contact with the product and external) are components extremely subject to wear. Their duration is
strongly affected by the conditions of employments and by chemical and physical stresses. By tests carried out on thousands of pumps
installed with head value from 0° to 18°C, the ordinary life exceeds
one hundred million cycles. For safety reasons, in environments with explosion risk it is necessary to disassemble and check the
diaphragms every five million cycles and to replace them every twenty mil- lion cycles.
WARNING! In the case of diaphragms total breaking, the fluid may enter in the pneumatic circuit, da- mage it and come out from the
discharge port. Therefore it is necessary to convey the air discharge in
a piping up to a safe area.
WARNING! In situations where the user foresees the possibility of exceeding the temperature limits indicated in this manual, it is
necessary to install a protection device on the equipment that prevents to achieve the maximum operating temperature allowed. If
exceeded, respect to the maximum marking temperature is not guaranteed.
REMEMBER! Safety risks to persons are mainly caused by improper use or accidental damage.
These risks may be of hand injury for operators working on the open pump, or caused by the nature of the fluids that are conveyed by this type of
pump. Therefore it is extremely important to diligently carry out all
the instructions contained in this manual in order to eliminate the causes of accidents that may lead to the pump failure and to the subsequent
outcome of fluid hazardous to persons and to the environment.
For installation and use in a potentially explosive environment, comply with these general precautions:
• ascertain that the pump is full and if possible, that the level is above it by 0.5 m
• ascertain that the fluid treated does not contain or cannot contain large solids or solids of a dangerous shape
• ensure that the intake or delivery ports are not obstructed nor limited to avoid cavitation or pneumatic motor strain;
• also ascertain that the connection piping is strong enough and cannot be deformed by the pump weight or by the intake. Also check that
the pump is not burdened by the weight of the piping.
• if the pump is to stay in disuse for a long period of time, clean it carefully by running a non-flammable liquid detergent through it that is
compatible with the pump’s construction materials;
• if the pump was turned off for a long period of time, circulate clean water it in for some minutes to avoid incrustations
• before starting, after long periods of disuse, clean the internal and external surfaces with a damp cloth, check the grounding ATEX ZONE ;
• always protect the pump against possible collisions caused by moving objects or by various blunt materials that may damage it or react with
its materials;
• protect the pump’s surrounding ambient from splashes caused by accidental pump failure;
WARNING: the air supply pressure must never be over 7 bar or below 2 bar.
WARNING: when using the pump with aggressive or toxic liquids or with liquids that may represent a health hazard you must install suitable
protection on the pump to contain, collect and signal any spills: DANGER OF POLLUTION, CON- TAMINATION, INJURIES AND/OR DEATH.
WARNING: the pump must not be used with fluids that are not compatible with its construction materials or in a place containing incompatible
fluids.
CAUTION: installing the pumps without on-off valves on the intake and delivery sides to intercept the product in case of spillage is forbidden: danger
of uncontrolled product spillage.
CAUTION: installing the pumps without on-off, three-way or check valves on the air supply piping to prevent the pumped liquid from entering the
pneumatic circuit if the diaphragms are broken is forbidden: danger of fluid entering the compressed air circuit and being discharged into the
environment.
WARNING: Should the user think that the temperature limits set forth in this manual may be exceeded during service, a protective device must be
installed on the system to prevent the maximum allowed process temperature from being reached. If exceeded, respect of the maximum temperature
marked cannot be guaranteed.
WARNING: The pumps must always be grounded irrespective of any organ to which they are connected. Lack of grounding or incorrect grounding
will cancel the requirements for safety and protection against the risk of
WARNING: the use of pumps made with non-conductive material, which become charged with static, and without suitable grounding for flammable
liquids is forbidden: RISK OF EXPLOSIONS DUE TO STATIC CHARGE.
CAUTION: Aggressive, toxic or dangerous liquids may cause serious injuries or damage to health, therefore it is forbidden to return a pump
containing such Products to the manufacturer or to a service center. You must empty the internal circuits from the product first and wash and treat
it.
CAUTION: Pumps containing aluminium parts or components coming into contact with the product cannot be used to pump III-trichloroethane,
methylene chloride or solvents based on other halogenated hydrocarbons: DANGER OF AN EXPLOSION CAUSED BY A CHEMICAL REACTION.
CAUTION: The pumps PHOENIX cannot pump Acetylene, Hydrogen, Carbon disulfide
CAUTION: The components of the pneumatic exchanger, including the shaft are made from materials that are not specifically resistant to chemical
products. If the diaphragm should break, replace these elements completely if they have come into contact with the product.
CAUTION: The air-driven motor of the PHOENIX pumps is self-lubricating and will not require any greasing. Therefore avoid using lubricated and
non-dried air.
WARNING: ascertain that during service no anomalous noise appears. In that case, stop the pump immediately. WARNING: ascertain that the fluid
at the delivery side does not contain gas. Otherwise stop the pump immediately.
WARNING: the diaphragms (in contact with the product or the external ones) are highly subject to wear. Their duration is strongly affected by the
conditions of use and by chemical and physical stress. Fields tests carried out on thousands of pumps with a head value from 0° to 18° C have shown
that normal service life exceeds one hundred million cycles. However, in places at risk of explosion, the diaphragm must be disassembled and checked
every 5 million cycles and replaced every 20 million cycles.
WARNING: Periodic controls must be made to ensure that there is no powder and/or deposits on the external and inter- nal surfaces of the pump
and, if necessary, they must be cleaned with a damp cloth.
WARNING: removal of the silencer and the air supply fitting must be done when free from powder. Before restarting the pump, ensure that no
powder has entered the pneumatic distributor.
WARNING : Protect always the site and the persons from accidental failure by installing a protection guard to old and collect any
product leakage. Danger of serious injuries and damage to health and/or objects.
46 Rev.02
PRESCRIZIONI PER LA SICUREZZA
ATTENZIONE! RISCHIO DI NATURA CHIMICA. Le pompe sono destinate al funzionamento con diverse tipologie di liquidi e soluzioni chimiche.
Seguire le specifiche istruzioni interne per la decontaminazione durante le operazioni di ispezione o manutenzione.
ATTENZIONE! RISCHIO DI NATURA ELETTRICO. La pompa deve essere sempre messa a terra indipendentemente da altro organo ad essa
collegato. La mancanza di messa a terra o non corretta messa a terra fa decadere i requisiti di sicurezza e tutela dal pericolo di esplosione.
ATTENZIONE: le membrane (a contatto con il prodotto ed esterne) sono componenti altamente soggetti ad usura. La loro durata è
fortemente influenzata dalle condizioni di impiego e dalle sollecitazioni chimi- che e fisiche. Da test eseguiti su migliaia di pompe installate
con prevalenza pari a 0°a 18°C, la durata normale supera i cento milioni di cicli. Per motivi di sicurezza, negli ambienti con pericolo di
esplosione bisogna eseguire lo smontaggio e la verifica della membrana ogni cinque milioni di cicli e la loro sostituzione ogni venti milioni di
cicli.
ATTENZIONE! Nel caso di rottura totale delle membrane, il fluido può entrare nel circuito pneumatico, danneggiarlo ed uscire dallo
scarico. Pertanto bisogna convogliare lo scarico dell’aria in una tubazione fino ad una zona sicura.
ATTENZIONE! laddove l’utilizzatore preveda il rischio di superamento dei limiti di temperatura previsti dal presente manuale, è necessario
installare sull’impianto un dispositivo di protezione che impedisca il raggiungimento della temperatura massima ammessa di processo. In
caso di superamento non è garantito il rispetto della massima temperatura di marcatura.
RICORDA! I rischi per la sicurezza delle persone nascono principalmente per uso improprio o in caso di danni accidentali. Tali rischi
possono essere di ferimento alle mani per chi opera sulla pompa aperta, oppure sono dovuti alla natura dei liquidi che vengono convogliati
da questo genere di pompe. E’ quindi di estrema importanza eseguire diligentemente tutte le istruzioni contenute in questo manuale al fine
di eliminare le cause di incidenti che possano portare alla rottura della pompa ed alla conseguente fuoriuscita di liquido pericoloso per le
persone e per l’ambiente.
Per l’installazione e l’impiego in ambiente potenzialmente esplosivo e non, rispettare le seguenti precauzioni generali:
- controllare che la pompa sia piena e il livello sia, possibilmente, al disopra di essa di 0,5m
- controllare che nel fluido trattato non vi siano o vi possano essere parti solide di dimensioni elevate o forma dannosa non ci siano
restrizioni nell’ingresso o nell’uscita della pompa per evitare fenomeni rispettivamente di cavitazione e sforzo del motore pneumatico
- controllare che le tubazioni di collegamento siano sufficientemente resistenti e che non possano deformarsi sotto il peso della pompa e
dell’aspirazione, né che la pompa subisca il peso delle tubazioni
- se la pompa deve rimanere inattiva per lunghi periodi, pulirla accuratamente facendo circolare un fluido detergente non infiammabile
compatibile con i materiali della pompa.
- se la pompa è stata spenta per lunghi periodi è opportuno far circolare acqua pulita per alcuni minuti per evitare il rischio di incrostazioni
- prima dell’avviamento, dopo lunghi periodi di sosta, effettuare la pulizia delle superfici interne ed esterne con un panno umido, controllare
la messa a terra per ZONE ATEX .
- proteggere sempre la pompa da possibili urti provocati accidentalmente da mezzi in movimento o materiali vari contundenti che possono
danneggiarla e/o reagire al contatto.
- proteggere sempre l’ambiente circostante la pompa da spruzzi provenienti da guasti accidentali alla pompa
ATTENZIONE: l’alimentazione dell’aria non deve mai risultare superiore a 7bar o inferiore a 2 bar
ATTENZIONE: Per tutti i fluidi pompati, soprattutto fluidi aggressivi, tossici o pericolosi per la salute bisogna installare sulla pompa un’adeguata
protezione per il contenimento e la raccolta e segnalazione del prodotto in caso di fuoriuscita: PERICOLO DI INQUINAMENTO,
CONTAMINAZIONE, LESIONI E/O MORTE.
ATTENZIONE: è vietato l’uso della pompa con fluidi non compatibili con i materiali dei componenti o in ambiente con presenza di fluidi non
compatibili.
ATTENZIONE: è vietata l’installazione della pompa in assenza di valvole per l’intercettazione del prodotto sull’aspirazione e sulla mandata per
eseguire il Sezionamento in caso di perdita: pericolo di fuoriuscita incontrollata del prodotto.
ATTENZIONE: è vietata l’installazione della pompa in assenza di valvola di intercettazione, valvola a 3 vie e valvola di non ritorno sul condotto di
alimentazione dell’aria per impedire che il fluido pompato entri nel circuito pneumatico in caso di rottura delle membrane: pericolo di immissione del
fluido nel circuito dell’aria compressa e scarico in ambiente
ATTENZIONE: laddove l’utilizzatore preveda il rischio di superamento dei limiti di temperatura previsti dal presente manuale, è necessario installare
sull’impianto un dispositivo di protezione che impedisca il raggiungimento della temperatura massima ammessa di processo. In caso di superamento
non è garantito il rispetto della massima temperatura di marcatura.
ATTENZIONE: la pompa deve essere sempre messa a terra indipendentemente da altro organo ad essa collegato. La mancanza di messa a terra o
non corretta messa a terra fa decadere i requisiti di sicurezza e tutela dal pericolo di esplosione.
ATTENZIONE: è vietato l’impiego per liquidi infiammabili della pompa in materiale non conduttivo, che si carica statica- mente e senza un’adeguata
messa a terra: PERICOLO DI ESPLOSIONI A CAUSA DI CARICHE STATICHE.
ATTENZIONE: fluidi aggressivi, tossici o pericolosi possono causare gravi lesioni fisiche e/o alla salute pertanto è vieta- to restituire al produttore o
ad un centro di servizio una pompa che contenga prodotti di tale specie: Svuotare e lavare il circuito interno del prodotto e provvedere al lavaggio e
trattamento prima di rispedire la pompa.
ATTENZIONE: i modelli di pompe che contengono componenti o parti in alluminio a contatto con il prodotto non possono essere impiegate per il
pompaggio di III-tricloroetano, il cloro metilene o solventi a base di altri idrocarburi alogenati: PERICOLO DI ESPLOSIONE PER REAZIONE
CHIMICA.
ATTENZIONE: le pompe della serie PHOENIX non possono pompare acetilene, idrogeno e solfuro di carbonio
ATTENZIONE: i componenti dello scambiatore pneumatico, albero compreso, sono costruiti con materiali non specificamente resistenti ai prodotti
chimici. In caso di rottura delle membrane, se vengono a contatto con il fluido, provvedere alla loro completa sostituzione.
ATTENZIONE: il motore pneumatico delle pompe PHOENIX è autolubrificante e non necessita di ulteriore lubrificante;
pertanto evitare l’impiego di aria lubrificata e non essiccata.
ATTENZIONE: verificare che durante il funzionamento non si verifichi rumorosità anomala. In tal caso bloccare immediatamente la pompa.
ATTENZIONE: controllare che nel fluido in uscita non sia presente gas, in tal caso bloccare immediatamente la pompa.
ATTENZIONE: le membrane (a contatto con il prodotto ed esterne) sono componenti altamente soggetti ad usura. La loro durata è fortemente
influenzata dalle condizioni di impiego e dalle sollecitazioni chimiche e fisiche. Da test eseguiti su migliaia di pompe installate con prevalenza pari a 0°a
18°C, la durata normale supera i cento milioni di cicli. Per motivi di sicurezza, negli ambienti con pericolo di esplosione bisogna eseguire lo
smontaggio e la verifica della membrana ogni cinque milioni di cicli e la loro sostituzione ogni venti milioni di cicli.
ATTENZIONE: Bisogna verificare periodicamente l’assenza di polveri e/o depositi dalle superfici esterne ed interne della pompa e, se necessario,
effettuare la pulizia con un panno umido.
ATTENZIONE: lo smontaggio del silenziatore e del raccordo di alimentazione dell’aria deve essere effettuato in assenza di polvere. Prima di riavviare
la pompa assicurarsi che non sia entrata polvere all’interno del distributore pneumatico.
ATTENZIONE: Proteggere sempre l’ambiente e le persone installando una protezione, in caso di
rottura accidentale della pompa, per il contenimento e la raccolta in caso di fuoriuscita del prodotto.
Pericolo di gravi danni fisici, alla salute e/o cose
Per la sostituzione di parti usurate impiegare unicamente pezzi di ricambio originali.
L’inosservanza di quanto sopra può far insorgere pericoli per l’operatore, i tecnici, le persone, la pompa e/o l’ambiente
non imputabili al costruttore.
47 Rev.02
However five general elements are important:
A- all the operations must be carried out by skilled personnel or monitored by qualified personal as appropriate
B- implement personal protection works (when the pump is installed in places involving more than occasional visits) against splashes of fluorescent fluid for
accidental breakage and conveying works
(always) of possible fluid leakages towards collection tanks
C- wear acid-resistant clothing and protection whenever operating on the pump
D- make sure that the Intake and delivery valves are correctly closed during the disassembly
E- make sure that there is no supply to the pneumatic circuit during the disassembly
It should be noted that it is very important to realize systems with pipes well arranged, identifiable, suitably equipped with shut-off valves, with comfortable
compartments and passages for operators who must inspect their status (since the pressure developed by the pump may promote failures to the system if it is
of defective construction or worn).
DISPOSAL
For type of material: separate plastic parts from metal parts and dispose of by authorized companies.
SMALTIMENTO
Per tipologia di materiale: separare le parti in plastica dalle parti in metallo e smaltire tramite imprese autorizzate.
48 Rev.02
TROUBLESHOOTING AND POSSIBLE CAUSES
DEFECT CAUSE SUGGESTION
Circuit without air Check the circuit (valves, connections, regulators, etc.)
Insufficient air pressure Adjust the air pressure
Insufficient air flow rate Check that the pipes and fittings have suitable passages
Damaged control valve Replace
Damaged pneumatic distributor Replace
Open some valves, or remove the pipes and check if the
Pump delivery or intake closed pump starts
The pump does not
1 Damaged discharge cover Replace
start
Check if air comes out of the product delivery pipe, if yes
Broken diaphragm replace the diaphragm.
Disassemble the manifolds and clean the seats or replace the
The balls do not close. balls and the seats.
Excessive intake height. Reduce the intake height.
The pump works
butit does not Install pipes with greater size especially for intake and decre-
2 Too viscous fluid. ase the pumps cycles.
pump.
Clogged intake. Check and clean.
Excessively viscous fluid. No remedy.
The pump works Clogged delivery pipe. Check and clean.
3 with slow cycles Clogged intake. Check and clean.
Internal pneumatic exchanger clogged
or defective Replace the pneumatic exchanger.
Worn shaft. Replace the pneumatic exchanger.
Ice on the discharge. Dehumidify and filter air.
The pump works Check all the air control fittings, especially the quick cou-
4 Air volume is lacking. plings.
irregularly.
Internal exchanger dirty Replace.
Intake clogs during operation. Replace the intake pipe.
Dirty air, full of condensate or oil. Check the air line.
Check the pressure with a gauge installed on thepump and
with running pump. If pressurein that point is too low related
to the mains pressure, check all the air connections, especial-
ly the quick couplings Check that all the air control devices
Insufficient air volume or pressure. have a sufficient flow rate. WARNING: In 90% of the cases
5 The pump stalls stall conditions depends on the quick couplings.
Defective distributor. Replace it.
The stop procedure was not respected. Respect the stop procedure.
The product intake pipe is not correctly
connected. Check.
Clogged pipes. Check and clean.
Install pipes with greater size especially for intake and decre-
Too viscous fluid. ase the pumps cycles.
Disassemble the manifolds and clean the seats or replace the
The balls do not close. balls and the seats.
The pump does not Check pressure with a gauge installed on the pump and with
deliver the flow rate running pump. If pressure in that point is too low related to
6 indicated on the the mains pressure, check all the air connections, especially
table. the quick couplings. Check that all the air control devices have
Insufficient air volume. a sufficient flow rate. WARNING: In 90% of the cases stall
conditions depends on the quick couplings.
49 Rev.02
ANOMALIE DI FUNZIONAMENTO E POSSIBILI CAUSE
DIFETTO CAUSA SUGGERIMENTO
Circuito senza aria Controllare il circuito(rubinetti, allacciamenti, regolatori, ecc.)
Insufficiente pressione dell’aria Regolare la pressione dell’ aria
Insufficiente portata dell’aria Controllare che tubi ed accessori abbiano passaggi adeguati
Valvola di comando danneggiata Sostituire
Distributore pneumatico danneggiato Sostituire
Mandata o aspirazione della pompa Aprire eventuali valvole, oppure rimuovere le tubazioni e
chiuse verificare se la pompa parte
La pompa non si
1 Coperchio scarico danneggiato Sostituire
avvia
Controllare se esce aria dal tubo di mandata prodotto; nel
Membrana rotta caso sostituire la membrana.
Smontare i collettori e pulire le sedi o sostituire le sfere e le
Le sfere non chiudono. sedi.
Altezza di aspirazione troppo elevata. Ridurre l’altezza di aspirazione.
La pompa funziona Installare tubi maggiorati specie in aspirazione e diminuire i
2 Fluido troppo viscoso. cicli della pompa.
ma non pompa.
Aspirazione otturata. Controllare e pulire.
Fluido troppo viscoso. Nessun rimedio.
La pompa funziona Tubo di mandata otturato. Controllare e pulire.
3 con cicli lenti Aspirazione otturata. Controllare e pulire.
Scambiatore pneumatico interno
usurato o difettoso Sostituire lo scambiatore pneumatico.
Albero usurato. Sostituire lo scambiatore pneumatico.
Ghiaccio sullo scarico. Deumidificare e filtrare l’aria.
La pompa funziona Controllare tutti gli accessori di controllo aria, in particolar
4 Manca volume d’aria. modo gli innesti rapidi.
in modo irregolare.
Scambiatore interno sporco Sostituire.
L’aspirazione si ottura durante il
funzionamento.. Sostituire il tubo di aspirazione.
Aria sporca, piena di condensa o di olio.
Verificare la linea dell’aria.
Controllare la pressione con un manometro installato sulla-
pompa ed a pompa in funzione. Se la pressione in quel punto è
troppo bassa rispetto alla pressione di rete, controllare tutti
gli attacchi dell’aria, specialmente quelli ad innesto rapido.
Volume o pressione d’aria insufficiente. Controllare che tutti i dispositivi di controllo aria abbiano
La pompa va in portata sufficiente. ATTENZIONE: Nel 90% i casi di stallo
5 stallo dipendono dagli innesti rapidi.
Distributore difettoso. Sostituirlo.
Non è stata rispettata la procedura di
arresto Rispettare la procedura di arresto.
Il tubo di aspirazione prodotto è mal
collegato. Controllare.
Tubazioni otturate. Controllare e pulire.
Installare tubi maggiorati specie in aspirazione e diminuire
Fluido troppo viscoso. I cicli della pompa.
Smontare i collettori e pulire le sedi o sostituire le sfere e le
Le sfere non chiudono. sedi..
Controllare la pressione con un manometro installato sulla
pompa ed a pompa in funzione. Se la pressione in quel punto è
La pompa non eroga troppo bassa rispetto alla pressione di rete, controllare tutti
6 la portata di tabella. gli attacchi dell’aria, specialmente quelli ad innesto rapido.
Volume d’aria insufficiente. Controllare che tutti i dispositivi di controllo aria abbiano
portata sufficiente. ATTENZIONE: Nel 90% i casi di stallo
dipendono dagli innesti rapidi
50 Rev.02
51 Rev.02
Production head and legal office:
● The INSTRUCTION MANUAL must be delivered to the pump-user , who takes diligent note of it,
keeps the file for subsequent reference. Possible modifications
do not imply updating of the existing manuals
● MANUALE D’USO deve essere consegnato all’utilizzatore della pompa, il quale deve prenderne attenta
visione, e conservarlo per successive consultazioni. Eventuali modifiche non comportano
l’aggiornamento dei manuali preesistenti.
52 Rev.02
TSU Ex-CLX
(HF scientific)
MANUAL
Chlorine and Total Residual Oxidant Monitor
For Use in Hazardous Environments
And Extended Reagent Life
Model CLX-Ex2 28034
HF scientific
3170 Old Metro Parkway
Ft. Myers, FL 33916
Phone: 239-337-2116
Toll Free: 888-203-7248
Fax: 239-332-7643
Email:hf.info@wattswater.com
Website: www.hfscientific.com
Catalog No 28920
Rev 2 (8/2019)
DECLARATION OF CONFORMITY
Ex pxb IIC T4 Gb 0°C ≤ Ta ≤ 55°C
IECEx Certificate #IECEx LC 14.0006
To the following standards:
Manufacturer’s Address: 3170 Old Metro Parkway, Fort Myers, Florida 33916-7597
Importer’s Name:
Importer’s Address:
I, the undersigned, hereby declare that the equipment specified above conforms to the above Directive and
Standard
Specifications ....................................................................................................1
1.0 Overview ...........................................................................................................2
1.1 Unpacking and Inspection of the Instrument and Accessories ...................2
1.2 Instrument Labels.........................................................................................3
1.3 The Display .................................................................................................3
1.4 The Touch Pad ............................................................................................4
1.5 Orientation ..................................................................................................5
Instruction Manual 1
Remove the instrument from the shipping crate. Carefully inspect all items to ensure that no
visible damage has occurred during shipment. If the items received do not match the order,
please immediately contact the local distributor or the HF scientific Customer Service
Department.
Informational Front of
Enclosure
From the diagram below the major components can be located. Note that two reagents are
required the buffer and the indicator. The reagents are discussed in the Routine Maintenance
section 11.0.
The measurement chamber consists of a sample inlet, a purge drain, and an overflow. The
reagent is added from the check valves integrated into the lower portion. A green LED
provides the 515 nm source lamp, a red LED is used for sample level and flow measurement.
A single detector is located 180º from the green LED. A replaceable glass cuvette separates
the LEDs from the detector and maintains the measurement path length. Sample water flows
in the inlet at the bottom, through the measurement cuvette and out through an overflow drain.
This flow is used to both fill the cuvette and flush the system.
The reagents are dispensed from two replaceable bottles. One bottle has a buffer to control
the pH; the second has an indicator that contains the DPD, which produces color when
chlorine is present in the sample. The degree of color is dependant on the amount of chlorine
in the sample water.
The measurement chamber is open to view operations. A white LED backlights the chamber
for a clearer view. The white LED will flash to attract attention in the case of a warning or
failure. Most warnings and failures are also displayed on the screen. To prevent interference,
the white LED is turned off during measurements.
During normal operation the CLX-Ex2 will run through a timed cycle. A simplified cycle
will consist of the following sequences:
• Flushing – continuous sample flow
• Purging – PURGE valve opens
• Zeroing – no flow with cuvette full
• Adding Reagents – one pulse of the reagent pump
• Mixing with sample – sample flow pulses in
• Reading resulting sample – no flow with cuvette full
• Purging – PURGE valve opens to remove reacted sample
The cycle above is simplified and does not describe all the actions and testing that occurs. The
CPU continuously diagnoses the entire system for correct operation and sample water flow.
If an error occurs, a message is posted to the message queue on the LCD screen.
The reagent is added by a single pulse of the reagent solenoid. When the reagents require
replacement, the PRIME button is pushed to bring new reagents into the system. During
PRIME the reagent solenoid is pulsed several times to draw fluid from the two reagent bottles
and fill the tubes with new reagent. A complete PRIME takes less than a minute.
The SERVICE button empties the cuvette, stops the flow of sample water, and clears any
errors. This provides a convenient way to replace reagents and the measurement cuvette. If
more extensive servicing is performed, all power to the CLX-Ex2 should be removed. If the
CLX-Ex2 is to be turned off, it is recommended that the instrument be placed in SERVICE
mode before removing power. This ensures that the cuvette is emptied and the flow is off.
Start-up
At start-up with the enclosure door open the Air Sentinel will display a RED PRESSURE
lamp on the visual indicator indicates that the system is not pressurized and no power is sent
to the instrument.
Closing the door with the air pressure applied, will cause the Air Sentinel to display a GREEN
PRESSURE lamp and a RED TIMER lamp. During this time the enclosure will undergo an
automatic dilution purge period of 5 minutes where at least 5 volumes of free space inside the
enclosure will be exchanged with clean purge air. No power is applied to the rest of the
instrument during this time.
After the enclosure air has gone through this purge period the Air Sentinel will display all
GREEN lamps. At this time the instrument is deemed safe and power is applied to rest of the
instrument.
WARNING: Do not apply power to the Air Sentinel controller unless the area has been
properly tested and is known not to contain explosive materials.
Bypass Key
WARNING: Use of the bypass key disables the safety features and safe operation of this
instrument. It is imperative that this operation be performed only after the area has
been tested and known to be safe.
Assuming the area has been tested and is known not to contain explosive materials; a bypass
service key is provided and can be used. In this operating mode power will be retained to the
rest of the instrument even though the door is open. It is imitative that this mode is only used
after area is known to be safe. This operation disables the safe operation of this instrument.
Upon turning the key, the indicator will display the RED BYPASS & TIMER lamps. This is
an indication of the maintenance mode. Power is retained to the instrument. The key cannot
be removed in this mode of operation.
After maintenance has been performed the door should be closed and latched. Air pressure
will once again pressurize the enclosure. The maintenance key can now be turned and
removed and the key cover replaced. The purge controller will display GREEN PRESSURE
& BYPASS lamps indicating a safe operation and power is kept applied to the instrument.
The TIMER lamp will remain RED for 5 minutes to indicate a bypass operation had taken
place.
Dry contact are provided as feedback as to the operation of the instrument (operating or
standby). For more information on this feature refer to sections 4.4.4 Remote Standby
Connections and 6.2 Routine Measurement with Remote Standby.
The buffer reagent is required to control the sample pH to bring the sample to neutral.
The reagents are provided already hydrated and require little operator preparation.
The indicator reagents are packaged a plastic box. Each 125 mL indicator will require the
addition and mixing of the DPD powder contained in the small brown bottle. This activates
the indicator reagent. The indicator reagents should be changed out on a 90 day basis even if
they are not fully used.
The 500 ml buffer does not require any preparation and is good for one year.
After one year all the reagents will have been consumed and replacement reagent set will be
required. This can be ordered through agent Catalog No.09991. This set will include a new
plastic box with four sets of Indicator reagents plus one 500 mL bottle of Buffer.
For proper operation power must be applied by the Air Sentinel controller to the cooling
chamber temperature controller. Clean dry compressed air must also be provided at pressures
between 5.5 and 7 bar (80 to 101.5 PSI).
The cooling chamber temperature controller governs the solenoid which controls air flow to
the vortex cooling unit. The air flow is turned on or off as required to maintain the indicating
reagent temperature at 20-25˚C. This extends the life of the reagent up to 90 days.
To provide this extended reagent life, power to the cooling power supply and air pressure
must be supplied full time.
Be sure that the instrument be mounted as close as possible to the sampling point to ensure a
quick response time (within 2-3 meters (6-10 ft) of the sampling point).
The water drain requires 12 mm OD stainless steel tubing. This fitting is also a compression
fitting so the installer need only place the supplied the nut and feral over tubing and tighten.
Properly installed and sealed connections are required to ensure the water tightness and ratings
of the instrument.
The drain must be open to the Atmosphere for proper operation.
The air connection requires 6 mm stainless steel tubing. A compression type fitting is supplied
on the base of the main enclosure that simply requires the installer to slip the supplied nut and
feral over the tubing and then tighten the nut.
4.4.1 RS-485
The RS-485 half-duplex (2-wire) digital interface operates with differential levels that
CLX-Ex2 (8/2019) Page 11
REV 2
are not susceptible to electrical interferences. This is why cable lengths up to 3000 ft
can be implemented. The last device on each bus may require terminating with a 120
ohm resistor to eliminate signal reflection on the line. Do not run RS-485 cables in the
same conduit as power. Set-up of the RS-485 is covered in 8.3 Configuring the RS–
485 Port
4.4.3 4-20 mA
The 4-20 mA output is driven by a 15 VDC power
source and can drive recorder loads up to 600 ohms.
Transformer isolation is provided on the CLX-Ex2.
Do not run 4-20 mA cables in the same conduit as power. Operation of this output is
covered in section 8.1 Setting the 4-20 mA.
To prevent damage to the instrument, and for general safety ensure that power is
disconnected to the CLX-Ex2 prior to making any connections. Polarities of the
connections are labeled beside this termination on the PC board.
The recommended cable is 22 AWG shielded twisted pair. To prevent ground loops,
connect the shield at either the CLX-Ex2 or at its destination, but not both. The grey
terminal block is removable to assist in making connections.
The 4-20mA is factory calibrated. An adjustment is available on the 4-20mA in
sections 8.12 and 8.13. In addition to making adjustments, these menus output
continuous 4 mA or 20 mA and can be used as a signal test. Remember that the
configuration mode will timeout after 15 minutes.
At any time the standby status can be check at Modbus Coil Address 00005 or Modbus
Input Address 10005. False (0) indicates standby and True (1) indicates the
measurement cycle is taking place. Control is only available from the Coil Address
00005.
WARNING: Do not apply power to the Air Sentinel controller unless the area has been
properly tested and is known not to contain explosive materials.
5.1 Start Up
When power is applied to the CLX-Ex2 for the first time, ensure that the air supply is
connected as stated in section 4.3 and the cabinet is closed and locked. Once power is
applied, the Air Sentinel will go through a purge period of 5 minutes where at least five
volumes of the enclosure are pushed through the CLX-Ex2 and out the Air Sentinel vent.
During the purge period the TIMER lamp Air Sentinel will be RED indicating the purge is
in process and no power is applied to the rest of the CLX-Ex2.
At the complete of the purge, the TIMER lamp will go GREEN and power will be applied
to the rest of the system. As long as the proper air supply conditions are met the CLX-Ex2
will continue to operate.
Using the table above, you can quickly determine the operating condition of the instrument.
Under normal operation all GREEN lamps should appear.
Use of the bypass key disables the safety features and safe operation of this instrument.
It is imperative that this operation be performed only after the area has been tested and
known to be safe.
When shipped from the factory the bypass key is shipped on the strap securing the key
cap. Once the commissioning of the system has been done these should be moved to a
known secure location. It is not advised that they be left on the strap.
To perform some maintenance operations on the CLX-Ex2 it is necessary to have the entire
instrument powered up while the enclosure door is open. Under these conditions normally the
Air Sentinel controller would determine that the enclosure is not pressurized and power will
be removed to the measurement section of the CLX-Ex2. To allow for this situation, a special
maintenance bypass key is supplied.
Located on the right side of the Air Sentinel controller is a knurled cap. Removing this cap
reveals the key slot. Insert the key, push in slightly and turn it clockwise to engage the bypass
operation. The TIMER & BYPASS lamps will turn RED to indicate the bypass mode.
Assuming the area around the CLX-Ex2 has been tested and known to be safe, the enclosure
door can now be safely opened and power will remain on to the entire system. The key cannot
be removed while operating in the maintenance bypass mode.
Do not operate the CLX-Ex2 for long periods of time in the bypass mode. This is
intended only for maintenance operations.
To return the instrument to normal, safe operation, close the enclosure door. Ensure that the
air supply is connected turn the key counter-clockwise and remove. Replace the key cover. If
the pressure in the cabinet has built up to a normal acceptable level, the Air Sentinel will show
all GREEN lamps and the instrument will continue to run.
The CLX-Ex2 Online Chlorine Monitor allows for the measurement of the chlorine of
process water on-line. The chlorine value of the process water is usually reported in
milligrams per Liter (mg/L), these units are equivalent to Parts Per Million (PPM).
The CLX-Ex2 has two available measurement ranges, 0.00 – 10.00 or 0.00 – 15.00 mg/L
(PPM). Although the unit will display above the range, these readings will not be within the
stated accuracy. As the reagents degrade due to aging, readings above the set range may
decrease in value. .
6.1 Routine Measurement
First, ensure that all plumbing and electrical connections are complete before continuing.
The following steps describe how to measure the value of chlorine of a sample using this
instrument:
1. Apply power to the instrument and allow the unit to warm up. The instrument may
have to re-calibrate the water level sense detector. This WCAL or water calibration
will take few minutes to perform.
2. When a continuous process stream is flowing through the instrument, the instrument
will display the measured chlorine level of the sample by displaying it on the LCD
screen. In addition, the equivalent signal is provided on the analog (4-20 mA) output,
or the digital (RS-485) output, depending on the options selected.
The screen depicted indicates that the system has just been
started or just entered AUTO mode from Service mode
and no readings have been taken yet.
Please note that calibrations will not be allowed until a reading is posted.
The security code (333) must be entered to gain access to CAL or CONFIG menus. Notice
that the first number in the code is flashing. The flashing indicates that this is the number to
be changed. Use the or arrows to select the first of
the three numbers in the code and then press the button
to accept the first number of the code. Now enter the
second number in the code. Proceed as with the first
number followed by . Then repeat the process for the
third number in the access code, and finish with the
button.
If the valid access code has been selected, the instrument will be directed to the calibration
mode. If the wrong access code is selected, the instrument will return to the AUTO mode.
Refer to section 8.4 Enabling the Security Access for more information.
The white LED is also used to draw attention to a problem as described in section 10.1
CLX-Ex2 Fault Detection. In these instances the white LED blinks at a constant rate
dependant on the severity of the problem, but is still turned off as described above. Please
note that any fault is always posted to message queue on the lower portion of the LCD.
1. On the CLX-Ex2, press the MODE/EXIT button once. The screen is shown
below.
4. The screen will show the current reading on the CLX-Ex2. Since there is no
chlorine, the only reading may be a slight offset due to the absorbance of the
reagents. There should be no pink color developed.
5. Press to perform a Zero calibration. When the calibration has completed the
instrument will return to AUTO measurement mode automatically.
There is a limit of ±0.20 mg/L total adjustment available. A ZERO Cal. greater than this
will cause a CAL warning and no calibration will have occurred. Enter SERVICE mode
to clear this error.
CLX-Ex2 (8/2019) Page 19
REV 2
7.3 Restore Factory Settings
If the CLX-Ex2 displays a CAL error or the calibration was incorrectly performed, it may be
preferred to restore the factory calibration. All factory defaults settings including the factory
calibration can be reset by holding down the button and then pressing and releasing the
button then releasing the button.
Please note that these operating screens can be seen except with the cabinet door closed
through the door window or with the instrument in maintenance bypass mode. See
section 5.3 for more information.
If changes to the configuration have to be performed, you will need to enter the maintenance
bypass mode see section 5.3 for more information.
The instrument has been designed to provide the ability to customize the instrument according
to needs at any time during normal operation. This mode has been split into sub-menus to
facilitate instrument configuration. This section describes how to use each of the sub-menus
to configure the instrument. While in the configuration mode, the instrument has a time-out
feature that automatically returns the system operation to the AUTO mode after a fifteen (15)
minute period of no button pushes.
Enter the CONFIG mode of the instrument by pressing the MODE/EXIT button until the
arrow beside CONFIG is illuminated, then press the ↵ button to scroll through the sub-
menus.
To exit the CONFIG mode, press the MODE/EXIT button any changes that were made
will be saved.
8.4.1 Alarm 1
Alarm 1 Function: The ALM1 is displayed and the display indicates the current
function of alarm 1 (HI, LO, OFF, or Error). Use the or buttons to cycle
through and select the desired function. Press the button to accept the selection.
If the alarm was turned OFF, a prompt will appear to set up alarm 2.
Alarm 1 Set Point: This prompt is used to select the set point for this alarm; this is
indicated by “S/P” shown on the lower row of the display. Select the desired alarm
level by using the and buttons. Once the desired set point has been set, press
the button to accept it.
The Remote Standby limits the relays to a single normally open connection.
Although the alarm 2 menu may display, this alarm can be used.
Due to the cyclic nature of the CLX, relay chatter is not an issue. There is no need
for alarm delays or hysteresis.
If extended settings are set to OFF, pressing the button will save all settings and the
CLX-Ex2 will automatically return to the normal AUTO mode of the instrument.
The factory setting is averaging of 2. Select the desired AVG using the and buttons and
press the button to accept it.
In addition to averaging, the CLX-Ex2 has a software filter that limits the change between
consecutive readings to 20%. For example, the reading following a reading of 1.00 ppm
could not be displayed as higher than 1.20 ppm or lower than 0.80 ppm. This filter also
helps smooth out large changes and eliminate reading
spikes. After 3 consecutive readings, it is assumed that
the large change in the readings is "real", and the filter
will be disabled. For example, the change between
reading 1 & 2, and readings 2 & 3 will be limited to
20%, but the change between readings 3 & 4 will not be
limited.
Note: This software filter is completely disabled when the averaging (AVG) is set to 1.
The indicator reagent has a defined life after being mixed regardless of the cycle time
setting. If the reagent is kept in the powered cooling chamber the life is 90 days.
This feature is generally not used in ballast water applications and should be set to OFF.
The use of this option may result in some slight loss of accuracy. The factory setting for this
option is OFF.
Make selections using the and buttons then press the ↵ button to exit to AUTO mode
and save all configuration settings. If the 4-20 mA in section 7.1 is turned ON, there are two
additional menus that will
appear before returning to AUTO mode.
This setting will be slightly different on each instrument as each CLX-Ex2 will be factory
set to 4.00mA. Press the ↵ button when adjustments are complete to save this setting and
move on to the 20mA adjustment.
When complete with the 20mA adjustment, press the ↵ button to exit to AUTO mode and
save all configuration settings.
Please note that these operating screens can be seen through the door window with the cabinet
door closed or in maintenance bypass mode See section 5.3 for more information on this.
The CLX-Ex2 performs continuous diagnostic monitoring. In the CLX-Ex2, there are 4
severity levels of fault detection. Level 4, 3 & 2 will allow normal operation, but warn of the
problem. Level 1 is an instrument failure and the instrument will not operate. Any faults are
displayed in a queue form in the bottom row of the LCD.
A level 4 fault is simply a screen indication that one of the alarm levels has been activated.
This fault level will not affect the 4-20 mA and will only affect the alarm activated. The
sample back light blinks at a rate of once every 4 seconds.
A level 3 fault indicates a failure or a problem that usually can be corrected by the operator.
Refer to the chart below. If any of these errors occur, the instrument will still display
readings and probably will operate correctly. These faults will self-clear when the problem
is corrected. If any of these faults occur, they may affect the 4-20mA and any alarm
dependent on the detection setting (Error). See sections 8.2 & 8.4.1 for error settings. The
sample back light blinks at a rate of once every 2.5 seconds to indicate these faults.
A level 2 fault indicates a severe problem that will usually require technical assistance from
HF scientific customer service (see section 10.6). The queued display will show POST. If this
fault occurs it will affect the 4-20mA and any alarm set for fault detection (Error). The sample
back light blinks at rate of once every 1 second.
HF scientific
3170 Old Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Toll Free (888) 203-7248
Fax: (239) 332-7643
Email:hf.info@wattswater.com
www.hfscientific.com
Specific condition of use: Parts of the enclosure are non-conducting and exceed the maximum
permissible resistance according to the IEC 60079-0. Therefore, to avoid electrostatic charge
build-up, it must not be rubbed or cleaned with solvents or a dry cloth when installed/used
within a potentially explosive atmosphere.
The CLX-Ex2 is shipped with one CLX-Ex2 Tubing/Cuvette kit, HF part #09950EX. The kit
consists of the following:
Qty Part
1 Spare Buffer Cap Assembly
1 Spare Indicator Cap Assembly
8 Pump Tubes
1 Spare Cuvette
Two replacement sets of reagent cap/tubing assemblies are supplied with the CLX-Ex2; one
for the buffer and one for the indicator.
Additional kits can be ordered from your local HF scientific distributor or representative. It is
recommended to keep one kit on hand at all times.
Generally, both pump tubes should be replaced annually. Spare cap assemblies for both the
buffer and the indicator are supplied and can be replaced when needed.
Every 90 days
1. The indicator reagent required for operating this instrument must be changed at least
every 90 days. The buffer reagent can be used up to one year.
2. The internal strainer should be checked and cleaned if necessary.
3. The glass cuvette should be inspected. Check for excessive debris on the inside surface
of the glass. It is suggested to keep a spare cuvette to replace when required. The old
cuvette may be cleaned, if possible, for future replacement.
*The 1 year buffer and 90 day indicator reagent life is primarily used on ballast water
applications. The assumption is based on the factory default of a 60 second cycle and up to
60 hours total operation per month.
Pump Tubing Replacement
The black pump tubes may need replacement more often due to the fact that they are subject
to wear. The inline check valves should not need to be replaced and should be saved. Please
note that the check valves are directional as shown below.
Steps:
1 Flush the system as described above to reduce personal contact with the reagents.
2 Press SERVICE to stop the flow of sample water and drain the cuvette.
3 Remove and retain the thumb screw on top of the pump; pull the pump hammer and
spring up and out of the way. There is no need to completely remove the hammer and
spring.
4 Working on one reagent side at a time. Replace the black tubing between optics inlet and
the check valve OUTLET. Discard the old tubing.
5 Ensure the check valve is placed into its seat.
6 Repeat steps 4 & 5 for the other reagent.
7 Replace the hammer and spring back into place and secure with the thumb screw.
8 Check the drawing on the following page to ensure correct installation.
9 Return to operation as described.
Steps:
1 Flush the system as described above to reduce personal contact with the reagents.
2 Press SERVICE to stop the flow of sample water and drain the cuvette.
3 Replace the buffer cap (blue or black cap) assembly.
4 Repeat step 3 above, for the indicator side (white cap).
5 Replace reagents, prime and return to service.
T-Strainer Cleaning
The T-strainer is integral to the instrument and must be checked
occasionally. When necessary it must be removed and cleaned. The
strainer screen may require replacement after a period of time.
Steps:
1. Press SERVICE to stop the flow of sample water and drain the
cuvette.
2. Ensure the source water is turned off.
3. The T-strainer is clamped to the intake regulator. The removal
requires the use of a flat blade screwdriver as shown in the photo.
Once the two clamps are opened the T-strainer can be removed.
4. Disconnect the top of the T-strainer and regulator from the tubing
and clear of the enclosure.
5. Remove and clean the screen and then reassemble.
6. Be sure to tighten the bowl of the strainer.
7. Be certain the clamps are tightened fully and the tubing
connections are complete. You may require pliers to close
the clamps.
8. Turn source water back on.
9. Return to normal operation.
10. Check for any leaks.
Tubes may darken due to contact with the reagent. This condition does not affect the
performance of these parts.
Once mixed the indicator has an expected life of 90 days if kept in the powered cooling
chamber. Write the mixing date on the reagent bottle labels in the area provided.
Dispose of expired reagents correctly.
To replace the reagents, press the SERVICE button; this will empty the cuvette and stop
any flow of water. Remove the cap on both bottles replace with the cap supplied with the
CLX-Ex2. The buffer is installed on the left and the indicator is installed in the cooling
chamber. The suction tube for both reagents will reach the bottom of the bottles.
To complete the replacement procedure, press the PRIME button and then the button.
This will draw enough of each reagent to completely prime the tubes and replace any old
solution. The system will automatically return to normal operation after it has primed.
Use caution when changing the reagents as they are corrosive. These reagents may
stain clothing. After changing the reagents, operators should wash their hands.
To order any accessory or replacement part, please contact the HF scientific Customer Service
Department. If for any reason technical assistance is needed regarding this instrument, please
do not hesitate to contact the HF scientific Technical Services Department.
HF scientific
3170 Old Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Toll Free: (888) 203-7248
Fax: (239) 332-7643
Email:hf.info@wattswater.com
www.hfscientific.com
HF scientific, inc.
3170 Old Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Toll free: 888-203-7248
Email: hf.info@wattswater.com
Website: www.hfscientific.com
USER MANUAL
MANUALE D’USO
PHOENIX
PHOENIX ATEX
PHOENIX FOOD
PHOENIX FOOD ATEX
INDEX indice
4 IDENTIFICATION CODE sigla di identificazione
TRANSPORT trasporto
STORAGE stoccaggio
INSTALLATION installazione
START UP avviamento
USE utilizzo
STOP arresto
44 MAINTENANCE manutenzione
RECOMMENDATIONS raccomandazioni
DISASSEMBLY smontaggio
INSPECTION ispezione
DISPOSAL smaltimento
possibili cause
2 Rev.02
DICHIARAZIONE DI CONFORMITA’
DECLARATION OF CONFORMITY
MARCATURA ATEX
MARKING ATEX II 3/3 GD c IIB T135ºC
MARCAR ATEX
PHOENIX
Questo prodotto è conforme alle seguenti direttive CE/EX e relativi standard armonizzati:
This product complies with the following European Community Directives CE/EX and relating harmonized standards:
2006/42/CE Direttiva Macchine / Machinery Directive / Maschinenrichtlinie / Directive Machines / Directiva Máquinas
2014/34/UE: Direttiva ATEX, concernente il ravvicinamento delle legislazioni degli Stati Membri relative agli apparecchi e sistemi di protezione destinati a essere utilizzati in
atmosfera potenzialmente esplosiva.
UNI EN ISO 12100:2010 – Sicurezza del macchinario. Concetti fondamentali, principi generali di progettazione. Parte 1: terminologia di base, metodologia.
UNI EN ISO 12100:2010 – Safety of the machinery. Fundamental notions, design general principles. Part 1: Basic terminology, methods.
UNI EN ISO 12100:2010 – Sicurezza del macchinario. Concetti fondamentali, principi generali di progettazione. Parte 2: principi tecnici.
UNI EN ISO 12100:2010 – Safety of the machinery. Fundamental notions, design general principles. Part 2: Technical principles.
UNI EN ISO 3746: 2011 – Acustica. Determinazione dei livelli di potenza sonora delle sorgenti di rumore mediante misurazione della pressione sonora. Metodo di controllo con una superficie
avvolgente su un piano riflettente.
UNI EN ISO 3746: 2011 – Sound. Determination of sound power levels for noise sources by measuring the sound pressure. Monitoring method with an enveloping surface on a reflecting plate.
UNI EN ISO 11200: 2014 – Acustica. Rumore emesso dalle macchine e dalle apparecchiature. Linee guida per l’uso della norme di base per la determinazione dei livelli
di pressione sonora al posto di lavoro e in altre specifiche posizioni.
UNI EN ISO 11200: 2014 – Sound. Noise done by the machines and the equipments. Guidelines for using the basic norms for determining the sound pressure levels in
the working place and in other specific positions.
UNI EN ISO 4414: 2012 - Sicurezza del macchinario. Requisiti di sicurezza relativi a sistemi e loro componenti per trasmissioni oleoidrauliche e pneumatiche. Pneumatica.
UNI EN ISO 4414: 2012 - Safety of the machinery. Safety requirements concerning the systems and their components for hydraulic and pneumatic transmissions. Pneumatic.
UNI CEI EN ISO 80079-36:2016 - Apparecchi non elettrici destinati ad essere utilizzati in atmosfere potenzialmente esplosive. Parte 1: metodo e requisiti di base.
UNI CEI EN ISO 80079-36:2016 - Non-electrical equipments to be used in potentially explosive environments. Part 1: Method and basic requirements.
UNI CEI EN ISO 80079-37:2016 - Apparecchi non elettrici per atmosfere potenzialmente esplosive. Parte 5: protezione per sicurezza costruttiva “c”.
UNI CEI EN ISO 80079-37:2016 – Non-electrical equipments for potentially explosive environments. Part 5: protection for building safety “c”.
LA SEGUENTE CONFORMITA’ E’ RIFERITA AL PROTOTIPO DELLA PHOENIX 100 MATRICOLA NR. P0001 DEL 16.01.2012.
THIS COMPLIANCE REFERS TO PHOENIX 100 PROTOTYPE, SERIAL NUMBER P0001 OF 16.01.2012.
ESTENSIONI: la presente dichiarazione si estende anche ai modelli PHOENIX 7, PHOENIX 18, PHOENIX 30, PHOENIX 50, PHOENIX 65, PHOENIX 100, PHOENIX 101, PHOENIX 160,PHOENIX
250, PHOENIX 400,PHOENIX 500, PHOENIX 700, IN PLASTICA E METALLO.
EXTENSION: this declarations is also valid for the following versions PHOENIX 7, PHOENIX 18, PHOENIX 30, PHOENIX 50, PHOENIX 65, PHOENIX 100, PHOENIX 101, PHOENIX 160,PHOENIX
250, PHOENIX 400,PHOENIX 500, PHOENIX 700, MADE OF METAL OR PLASTIC.
ATTENZIONE: data l’innumerevole varietà di prodotti e composizioni chimiche, l’utilizzatore è ritenuto il maggior conoscitore delle reazioni e compatibilità con i materiali
costruttivi della pompa. Pertanto, prima dell’impiego, eseguire con perizia tutte le verifiche e prove necessarie al fine di evitare situazioni pericolose anche se remote che
non possono essere conosciute ed imputabili al costruttore. Per ogni controversia il Foro Competente è quello di Varese.
WARNING: since there exists an endless variety of products and chemical compositions, the user is presumed to have the best knowledge of their reaction and
compatibility with the materials used to build the pump. Therefore, before using the pump, all the necessary checks and tests must be performed with great care to avoid
even the slightest risk, an event that the manufacturer cannot foresee and of which he cannot be held responsible. Any controversy lies within competence of the Court of
Varese.
DATA 12/10/2017
3 Rev.02
IDENTIFICATION CODE sigla di identificazione
PUMP PUMP AIR FLUID BALL VERSION
BALLS O-RING CONNECTION
MODEL BODY DIAPHRAGM DIAPHRAGM SEATS ATEX ZONE1
PF/SP18 S - AISI 316 H - HYTREL T - PTFE T - PTFE S - SS AISI 316 T - PTFE 1 - BSP THREATED X
PF: S-SS AISI 316 2 - FLANGED
PF/SP30 ELECTROPOLISCHED 3 - CLAMP
PF/SP50 SP:
MECCANICHALLYLLY
PF/SP100
POLISCHED
PF/SP160
PF/SP400
PF/SP500
PF/SP700
PF/SP1000
Each pump is supplied with the serial and model abbreviation and the serial number on the rating
plate, applied onto the support side. Check these data upon receiving the goods. Any discrepancy
between the order and the delivery must be communicated immediately. In order to be able to
trace data and information, the abbreviation, model and serial number of the pump must be
quoted in all correspondence.
Ogni pompa spedita è contrassegnata dalla sigla della serie, dal modello e dal numero di serie impressi sulla
targhetta di identificazione. Si prega di confermare i dati impressi quanto prima dal ricevimento della merce. Ogni
discrepanza tra l’ordine e le informazioni impresse sulla targhetta deve essere immediatamente comunicata. Al fine
di consentire la perfetta rintracciabilità dei dati e delle informazioni, ogni corrispondenza deve riportare la sigla, il
modello e il numero di serie della pompa in oggetto.
MODEL
modello
YEAR OF MANUFACTURE
Anno di fabbricazione
SERIAL No
Numero di serie
ATEX CERTIFICATION certificazione Atex
PHOENIX ATEX - PHOENIX FOOD ATEX - CERT. ZONE 2 II 2/2 GD c IIB T135°C
4 Rev.02
GENERAL NOTES
“PHOENIX” series pumps are air-operated, double-diaphragm positive-displacement pumps, designed and manufactured
for pumping fluids that are chemically compatible with the constructive materials of the pump. The characteristics of the
fluid (pressure, temperature, chemical reactivity, specific weight, viscosity, vapor pressure) and of the environment must
be compatible with the pump characteristics and are defined in the ordering phase. Fluimac is not responsible for the
pumped liquid. The customer must ensure that there is compatibility between the pumped liquids and pump materials.
“PHOENIX” series pumps are self-priming; at the start-up the pipes can be empty.
The declared dry negative suction is referred to intake of water at a temperature of 20°C/ 68°F. The priming time and
the diaphragm’s life depend on.
- the suction circuit (total length and diameter)
- specific weight of the pumped fluid
- viscosity of the pumped fluid
- negative suction: max 5.000 cps (at 18°C / 64,4°F)
- below head suction: max 50.000 cps (at 18°C / 64,4°F) “PHOENIX” series pumps may be used dry
SS 0÷40°C / 14÷104°F
SS 95°C / 203°F 80°C / 176°F
The pump may be operated at a maximum pressure equal to 1.5 times the head value with closed delivery.
The value of the vapor pressure of the pumped fluid must be greater (of at least 3 mwc - meters of water column) than
the difference between the total absolute head value (pressure on suction level subtracted of the suction height) and
the leakages of the suction section.
The pumped fluid may contain particles suspended in different concentrations in accordance with the type of valve
assembled:
MAX DIM.
mm. 10
2,5 3 3,5 3,5 7,5 8 8,5
37 Rev.02
NOTE GENERALI
Le pompe serie "PHOENIX" sono pompe pneumatiche volumetriche a doppia membrana, progettate e costruite per il
pompaggio di liquidi compatibili chimicamente con i materiali costruttivi della pompa. Le caratteristiche del liquido
(pressione, temperatura, reattività chimica, peso specifico, viscosità, tensione di vapore) e dell’ambiente devono essere
compatibili con le caratteristiche della pompa e sono definite in sede d’ordine. La Fluimac s.r.l non è responsabile dei
liquidi pompati. Il cliente deve assicurarsi che ci sia compatibilità tra i liquidi pompati e i materiali della pompa.
Le pompe serie “PHOENIX” sono autoadescanti, all’avviamento le tubazioni possono essere vuote
La capacità di aspirazione negativa a secco dichiarata è riferita al pescaggio di acqua a temperatura di 20°C/
68°F.
Le pompe serie “PHOENIX” non possono essere usate per creare il vuoto
Assicurarsi che le caratteristiche fisico-chimiche del liquido siano state attentamente valutate.
La temperatura massima riferita ad acqua in funzionamento in continuo dipende dalla versione dei materiali
(riportata in targhetta) e dall’ ambiente in cui verrà installata la pompa:
VERSIONE MAX TEMP. MAX TEMP. L’intervallo di temperatura ambiente è in funzione della versione dei
ATEX ZONE 2 ATEX ZONE 1
materiali (riportata in targhetta):
Il valore della tensione di vapore del liquido pompato deve essere superiore (di almeno 3 m c.a.) alla differenza tra il
battente totale assoluto (pressione sul livello in aspirazione sottratta dell’ altezza di aspirazione) e le perdite nel tratto di
aspirazione
Il liquido pompato può contenere solidi sospesi in misura diversa in relazione al tipo di valvola montata:
MAX DIM.
2,5 3 3,5 3,5 7,5 8 8,5 10
mm.
38 Rev.02
OPERATING PRINCIPLE principio di funzionamento
The pneumatic distribution system sends compressed air behind one of the two diaphragms (A), which
pushes the fluid towards the delivery circuit. Simultaneously, the opposite diaphragm (B) is in the intake
phase since it is dragged by the shaft that connects it to the other diaphragm (A) under pressure; air
present behind it is discharged into the environment through the flow rate regulator present on the
pump , while a pressure drop is created in the fluid chamber which sucks the fluid from the suction
circuit. When the diaphragm (A), under pressure, reaches the stroke limit, the distributor switches the
two inputs to the chamber on the diaphragms air side, putting diaphragm (B) under pressure and
diaphragm(A) in discharge. When the pump reaches its original starting point, each diaphragm has
carried out one air discharge stroke and one fluid delivery stroke. This sequence of movements makes
up a complete pumping cycle.
Il sistema di distribuzione pneumatico invia l’aria compressa dietro una delle due membrane(A), la quale spinge il fluido
verso il circuito di mandata. Contemporaneamente la membrana opposta(B) si trova in fase di aspirazione essendo
trascinata dall’albero che la collega all’altra membrana(A) sotto pressione; l’aria presente dietro di essa viene scaricata in
atmosfera attraverso il regolatore di prestazioni presente sulla pompa mentre nella camera del fluido si crea un
abbassamento di pressione che risucchia il fluido dal circuito di aspirazione. Quando la membrana(A), sotto pressione,
raggiunge il limite della corsa il distributore commuta i due ingressi alle camera lato aria delle membrane, mandando in
pressione la membrana(B) ed in scarico la membrana(A) Nel momento in cui la pompa raggiunge il suo punto di
partenza originale, ogni membrana ha compiuto una corsa di scarico aria e una di mandata liquido. Questa sequenza di
movimenti costituisce un ciclo di pompaggio completo.
WARNING: the pneumatic supply of “PHOENIX” series pumps must be carried out with oil-free, filtered, dry and
unlubricated air Avoid pressure drops by using pipes and adjusting and controlling elements having characteristics suitable
for the pump In case of installation in atex zone, the compressor must suck air from outside the area classified as atex or
use inert gas.
ATTENZIONE: l’alimentazione pneumatica delle pompe serie “PHOENIX” deve essere eseguita con aria disoleata, filtrata, essiccata e
non lubrificata Evitare cadute di pressioni impiegando tubi ed elementi di regolazione e controllo con caratteristiche adeguate a
quelle della pompa In caso di installazione in zona ATEX il compressore deve aspirare aria esternamente alla zona classificata ATEX
ovvero utilizzare gas inerte
39 Rev.02
INSTALLATION AND USE INSTRUCTIONS
TRANSPORT
• cover the hydraulic connections
• lift the hydraulic plastic parts without mechanical stress
• for transport on irregular roads, cushion the bumps with suitable support plane
• blows and impacts may damage parts that are important for the machine operation and safety
STORAGE
• In case you need to get away for a period of time the pumps before installation, store in original boxes. The boxes
should be stored off the ground, in a closed, clean and dry.
• in the event that the packaging has not received any is intact, it will be necessary to free the pump from it-Checking
the integrity and restore a new packaging
• The storage place should be closed environment with a temperature not lower than -5 ° C, not more than 40 ° C
and with a moisture content not exceeding the value of 80%; any packaging must not be subjected to shocks, vibrations
and loads above you
INSTALLATION
• it is essential for the pump self-priming operation that the hydraulic system is leak-proof
• clean the system before connecting the pump
• the pump must not contain foreign bodies and all the seals on the hydraulic connections must be removed
• before starting the pump check the tightness of the screws of the pump bodies and manifolds
• the pump positioning is horizontal, the fluid delivery manifold must always be positioned in the upper part (see arrows on the pump casing)
• fastening may be on the floor or on the ceiling
• position the pump the closest possible to the point of collection
STOCCAGGIO
• nel caso sia necessario staccare per un periodo di tempo le pompe prima dell’installazione, conservarle nelle casse
originali. Le casse devono essere stoccate sollevate dal suolo, in un ambiente chiuso, pulito ed asciutto.
• nel caso che al ricevimento l’eventuale imballo non risulti integro, occorrerà liberare la pompa dall’imballo, verificar-
ne l’integrità e ripristinare un imballo nuovo
• il luogo di stoccaggio dev’essere un ambiente chiuso con temperatura non inferiore ai –5°C, non superiore ai 40°C
e con un tasso di umidità che non superi il valore di 80%; l’eventuale imballo non dev’essere sottoposto ad urti,
vibrazioni e carichi sovrastanti
INSTALLAZIONE
• è fondamentale per l’operazione di autoadescamento della pompa che l’impianto idraulico sia a perfetta tenuta
• pulire l’impianto prima di collegare la pompa
• la pompa non deve contenere corpi estranei e devono essere rimossi i sigilli sulle connessioni idrauliche
• prima di avviare la pompa verificare il serraggio delle viti dei corpi pompa e dei collettori
• il posizionamento della pompa è in orizzontale; il collettore di mandata del fluido deve essere posizionato sempre
nella parte superiore (vedi frecce su corpi pompa)
• il fissaggio può essere a pavimento o a soffitto
• posizionare la pompa più vicino possibile al punto di prelievo
40 Rev.02
use the plant solutions indicated in the following diagram:
1. YES: use flexible pipes reinforced with rigid spiral to connect the hydraulic circuit of the pump. Rigid piping may
cause strong vibrations and manifolds breaking. Do not use pipes with nominal diameter smaller than the
diameter of the pump connections. For negative installations and/or viscous fluids use pipes with greater
diameter related to the nominal diameter of the pump.
2. YES: pulse damper
3. YES: gate valve for delivery adjustment
4. YES: intake for gauge or protection pressure switch
5. YES: pipe anchoring
6. YES: shut-off valve
7. NO: air pockets; the circuit must be linear and short
8. YES: discharge duct around the base
9. YES: wide and rigid filtering separator in case of open tanks
10. YES: wide and rigid filtering separator in case of open tanks
11. Make it as short as possible the length of the horizontal S no vent for the air
12. Slope of the pipe to the pump
13. n the flow velocity of the fluid max. 3.5 m / s
42 Rev.02
START UP
• check the correct execution of what indicated in the INSTALLATION paragraph
• check that the intake and delivery pipes of the hydraulic circuit are correctly connected
• open the intake and delivery valves of the pump hydraulic circuit
• open the 3-way valve on the air circuit
• set the operation point requested for the pump: properly adjust the air pressure and delivery that supplies the pump.
With pressure values under 2 bar the pump may stall, with pressure values above 7 bar it is possible that breakdowns and/or
yields may
occur with consequent spillage of the pumped fluid
• for pumps with split manifold the two pumped fluids must have the same viscosity value, very different viscosity values may lead to
stall
problems and/or diaphragms breaking
• do not operate at the limits of the operation curves: the maximum head or maximum delivery (total absence of leaks and intake
height in
the delivery circuit)
• check that there are no anomalous vibrations or noise due to the too elastic support structure, unsuitable fastening or cavitation
• after 2 hours of operation stop the pump correctly and check the tightening of all the bolts on the pump
AVVIAMENTO
• verificare la corretta esecuzione di quanto indicato in INSTALLAZIONE
• verificare che i tubi di aspirazione e mandata del circuito idraulico siano correttamente collegati
• aprire le valvole in aspirazione e mandata del circuito idraulico della pompa
• aprire la valvola a 3 vie sul circuito aria
• Impostare il punto di funzionamento per il quale la pompa è stata richiesta: regolare opportunamente la pressione e la portata dell’
aria che
alimenta la pompa. Con pressioni inferiori ai 2 bar la pompa può andare installo, con pressioni superiori ai 7 bar si possono
verificare
rotture e/o cedimenti con conseguente fuoriuscita del fluido pompato
• Nel caso di pompe con collettore sdoppiato i due fluidi pompati devono avere viscosità dello stesso ordine di grandezza;
viscosità sensibilmente differenti possono portare a problemi di stallo e/o rottura delle membrane
• non operare agli estremi della curva di funzionamento: massima prevalenza o massima portata (totale assenza di perdite e di
altezza di
sollevamento nel circuito di mandata)
• constatare l’assenza di anomali vibrazioni o rumorosità dovute a struttura portante troppo elastica, inadeguato fissaggio o
cavitazione
• dopo 2 ore di funzionamento arrestare correttamente la pompa e verificare il serraggio di tutti i bulloni della pompa
USE
• do not operate valves or shunts during the pump operation
• Risk of harmful water hammers in case of incorrect or sudden operations (valves must be operated only by trained personnel)
• empty and wash accurately inside the pump in case different fluids must be pumped
• insulate or empty the pump if the fluid crystallization temperature is equal to or below the ambient temperature
• stop the pump if the fluid temperature exceeds the maximum allowed temperature indicated in the GENERAL NOTES;
if the exceeding temperature is about 20% it is necessary to inspect the status of the internal parts
• stop the pump and close the valves in case of leaks
• Wash with water only if chemical compatibility allows it ; alternatively use the suitable solvent that does not generate hazardous
exothermic
reactions
• consult the fluid supplier to decide the most suitable fire-prevention method
• Empty the pump in case of long periods of disuse (particularly with fluids which are particularly tending to crystallize)
• check that there is no gas in the delivering fluid, if there is stop the pump
UTILIZZO
• non azionare valvole o derivazioni durante il funzionamento della pompa
• rischi di dannosi colpi d’ariete in caso di errate o improvvise manovre (azionamento valvole solo da personale ad- destrato)
• svuotare e lavare accuratamente l’interno pompa in caso di pompaggio di liquidi diversi
• isolare o svuotare la pompa se la temperatura di cristallizzazione del liquido risulta uguale o inferiore alla tempera- tura ambiente
• fermare la pompa se la temperatura del liquido supera la temperatura massima consentita indicata nelle NOTE GENERALI;
se l’aumento è dell’ordine del 20 % è necessario ispezionare lo stato delle parti interne
• arrestare la pompa e chiudere le valvole in caso di perdite
• lavare con acqua solo se la compatibilità chimica lo consente; in alternativa utilizzare il solvente appropriato che non generi
reazioni
esotermiche pericolose
• contattare il fornitore del liquido per stabilire il più appropriato metodo anti-incendio
• svuotare la pompa in caso di lunghi periodi di non utilizzo (in particolare con liquidi con forte tendenza a cristallizzare)
• controllare che nel fluido in mandata non sia presente gas, in tal caso arrestare la pompa
STOP
To stop the pump, operate exclusively on the air supply closing the 3-way valve, discharging in this way residual pressure from the
pneumatic system of the pump.
WARNING never stop the pump by totally closing the suction and/or delivery valves of the hydraulic circuit
WARNING is prohibited to stop the pump by means of the complete closure of the valves in the suction and / or discharge of the
hydraulic circuit
ARRESTO
Per arrestare la pompa agire esclusivamente sull’ alimentazione dell’aria chiudendo la valvola a 3 vie, scaricando così la pressione
residua dell’ impianto pneumatico della pompa
ATTENZIONE è vietato arrestare la pompa mediante la chiusura completa delle valvole in aspirazione e/o mandata del circuito
idraulico.
43 Rev.02
MAINTENANCE
- all the operation must be carried out by qualified personnel
- do not carry out maintenance and/or repairs with the air circuit under pressure
- carry out periodic inspections (2 ÷ 30 days in accordance with the fluid pumped) to check the filtering elements cleaning
- carry out periodic inspections (3 ÷ 5 months in accordance with the fluid pumped and with the environment conditions)
to ensure the correct operation of the system start/stop units
- the presence of fluid under the pump casing may indicate failures to the pump
- damaged parts must be replaced with complete original parts and not with repaired parts
- the replacement of damaged parts must be carried out in a clean and dry place
RECOMMENDATIONS
WARNING: before performing any maintenance or repair work on the pump, disconnect the pump from the air supply line.
Disconnect the hydraulic connections and discharge the product that is being pumped
- all the operation must be carried out by qualified personnel
- use gloves, goggles and acid-resistant clothing when disconnecting from the system and washing the pump
- wash the pump before carrying out maintenance operations
- do not disperse the washing waste into the environment
DISASSEMBLY
- Bolts are the type with right thread
- Clean all the pump external surfaces using a damp cloth Ball seats removal (for all models)
- Remove the delivery and intake manifolds removing the tightening bolts
- Pull off the seats, the balls and the related cages
- Check the condition of the gasket
Diaphragms removal
PHOENIX 07-18-30-50-65-100-101-160-250-500-700-1000
- Remove the delivery and intake manifolds removing the tightening bolts
- Remove the deposits from the internal surfaces
- Remove the two pump casings
- Remove the plates that lock the diaphragms
- Remove the diaphragms and the related backing plates
- If it would be necessary to disassemble the shaft, remove one of the two diaphragms on air side and then pull off the shaft
INSPECTION
Check the absence of:
• excessive abrasion of the thermoplastic parts
• clots and/or agglomerates due to the pumped fluid
• deformations and/or surface lesions of the diaphragms
• deformations and/or breakages on the valve seats
Replace the parts: broken, cracked, deformed.
Reopen all the clogged ducts and eliminate any chemical agglomerates.
Clean all the surfaces before reassembly, particularly the OR gaskets seats (risk of leaks for dripping).
44 Rev.02
MANUTENZIONE
- tutti gli interventi devono essere presidiati da personale qualificato
- non effettuare manutenzione e/o riparazioni con il circuito aria in pressione
- effettuare ispezioni periodiche (2 ÷ 30 giorni in funzione del liquido convogliato) di pulizia sugli elementi filtranti
- Effettuare ispezioni periodiche (3 ÷ 5 mesi in funzione del liquido convogliato e delle condizioni ambientali) sulla funzionalità
degli organi di consenso/arresto dell’impianto; garantirne l’efficienza
- la presenza di liquido sotto il corpo pompa può essere indizio di guasti alla pompa
- le parti danneggiate devono essere sostituite con parti originali integre e non riparate
- la sostituzione di parti danneggiate deve essere effettuata in ambiente pulito e asciutto
RACCOMANDAZIONI
ATTENZIONE: prima di intervenire sulla pompa per qualsiasi intervento di manutenzione o riparazione, scollegare dalla
pompa la linea aria di alimentazione. Staccare le connessioni idrauliche e scaricare il prodotto che si sta pompando
- tutti gli interventi devono essere presidiati da personale qualificato
- utilizzare guanti, occhiali ed indumenti antiacido in fase di scollegamento dall’impianto e lavando la pompa
- lavare la pompa prima di effettuare operazioni di manutenzione
- non disperdere nell’ambiente lo scarico di lavaggio dei corpi
SMONTAGGIO
- La bulloneria è del tipo a filetto destro
- Provvedere alla pulizia esterna di tutte le superfici della pompa con un panno umido
ISPEZIONE
Verificare l’assenza di:
• eccessiva abrasione delle parti in materiale termoplastico
• grumi e/o agglomerati dovuti al liquido pompato
• deformazioni e/o lesioni superficiali delle membrane
• deformazioni e/o rotture sulle sedi valvola
Sostituire le parti: rotte, fessurate, deformate.
Riaprire tutti i condotti occlusi ed eliminare gli eventuali agglomerati chimici.
Pulire tutte le superfici prima del rimontaggio, in particolare le sedi di guarnizioni OR (rischio di perdite per gocciola- mento).
45 Rev.02
SAFETY RISKS
WARNING! CHEMICAL RISK. Pumps are intended for operation with different types of fluids and chemical solutions. Follow the specific internal
instructions for decontamination during the inspection or maintenance operations.
WARNING! ELECTRICAL RISK. The pump must always be earthed independently to other members connected to it. Safety requirements and
explosion risk prevention are not fulfilled if the pump is not earthed or is incorrectly earthed.
WARNING: the diaphragms (into contact with the product and external) are components extremely subject to wear. Their duration is
strongly affected by the conditions of employments and by chemical and physical stresses. By tests carried out on thousands of pumps
installed with head value from 0° to 18°C, the ordinary life exceeds
one hundred million cycles. For safety reasons, in environments with explosion risk it is necessary to disassemble and check the
diaphragms every five million cycles and to replace them every twenty mil- lion cycles.
WARNING! In the case of diaphragms total breaking, the fluid may enter in the pneumatic circuit, da- mage it and come out from the
discharge port. Therefore it is necessary to convey the air discharge in
a piping up to a safe area.
WARNING! In situations where the user foresees the possibility of exceeding the temperature limits indicated in this manual, it is
necessary to install a protection device on the equipment that prevents to achieve the maximum operating temperature allowed. If
exceeded, respect to the maximum marking temperature is not guaranteed.
REMEMBER! Safety risks to persons are mainly caused by improper use or accidental damage.
These risks may be of hand injury for operators working on the open pump, or caused by the nature of the fluids that are conveyed by this type of
pump. Therefore it is extremely important to diligently carry out all
the instructions contained in this manual in order to eliminate the causes of accidents that may lead to the pump failure and to the subsequent
outcome of fluid hazardous to persons and to the environment.
For installation and use in a potentially explosive environment, comply with these general precautions:
• ascertain that the pump is full and if possible, that the level is above it by 0.5 m
• ascertain that the fluid treated does not contain or cannot contain large solids or solids of a dangerous shape
• ensure that the intake or delivery ports are not obstructed nor limited to avoid cavitation or pneumatic motor strain;
• also ascertain that the connection piping is strong enough and cannot be deformed by the pump weight or by the intake. Also check that
the pump is not burdened by the weight of the piping.
• if the pump is to stay in disuse for a long period of time, clean it carefully by running a non-flammable liquid detergent through it that is
compatible with the pump’s construction materials;
• if the pump was turned off for a long period of time, circulate clean water it in for some minutes to avoid incrustations
• before starting, after long periods of disuse, clean the internal and external surfaces with a damp cloth, check the grounding ATEX ZONE ;
• always protect the pump against possible collisions caused by moving objects or by various blunt materials that may damage it or react with
its materials;
• protect the pump’s surrounding ambient from splashes caused by accidental pump failure;
WARNING: the air supply pressure must never be over 7 bar or below 2 bar.
WARNING: when using the pump with aggressive or toxic liquids or with liquids that may represent a health hazard you must install suitable
protection on the pump to contain, collect and signal any spills: DANGER OF POLLUTION, CON- TAMINATION, INJURIES AND/OR DEATH.
WARNING: the pump must not be used with fluids that are not compatible with its construction materials or in a place containing incompatible
fluids.
CAUTION: installing the pumps without on-off valves on the intake and delivery sides to intercept the product in case of spillage is forbidden: danger
of uncontrolled product spillage.
CAUTION: installing the pumps without on-off, three-way or check valves on the air supply piping to prevent the pumped liquid from entering the
pneumatic circuit if the diaphragms are broken is forbidden: danger of fluid entering the compressed air circuit and being discharged into the
environment.
WARNING: Should the user think that the temperature limits set forth in this manual may be exceeded during service, a protective device must be
installed on the system to prevent the maximum allowed process temperature from being reached. If exceeded, respect of the maximum temperature
marked cannot be guaranteed.
WARNING: The pumps must always be grounded irrespective of any organ to which they are connected. Lack of grounding or incorrect grounding
will cancel the requirements for safety and protection against the risk of
WARNING: the use of pumps made with non-conductive material, which become charged with static, and without suitable grounding for flammable
liquids is forbidden: RISK OF EXPLOSIONS DUE TO STATIC CHARGE.
CAUTION: Aggressive, toxic or dangerous liquids may cause serious injuries or damage to health, therefore it is forbidden to return a pump
containing such Products to the manufacturer or to a service center. You must empty the internal circuits from the product first and wash and treat
it.
CAUTION: Pumps containing aluminium parts or components coming into contact with the product cannot be used to pump III-trichloroethane,
methylene chloride or solvents based on other halogenated hydrocarbons: DANGER OF AN EXPLOSION CAUSED BY A CHEMICAL REACTION.
CAUTION: The pumps PHOENIX cannot pump Acetylene, Hydrogen, Carbon disulfide
CAUTION: The components of the pneumatic exchanger, including the shaft are made from materials that are not specifically resistant to chemical
products. If the diaphragm should break, replace these elements completely if they have come into contact with the product.
CAUTION: The air-driven motor of the PHOENIX pumps is self-lubricating and will not require any greasing. Therefore avoid using lubricated and
non-dried air.
WARNING: ascertain that during service no anomalous noise appears. In that case, stop the pump immediately. WARNING: ascertain that the fluid
at the delivery side does not contain gas. Otherwise stop the pump immediately.
WARNING: the diaphragms (in contact with the product or the external ones) are highly subject to wear. Their duration is strongly affected by the
conditions of use and by chemical and physical stress. Fields tests carried out on thousands of pumps with a head value from 0° to 18° C have shown
that normal service life exceeds one hundred million cycles. However, in places at risk of explosion, the diaphragm must be disassembled and checked
every 5 million cycles and replaced every 20 million cycles.
WARNING: Periodic controls must be made to ensure that there is no powder and/or deposits on the external and inter- nal surfaces of the pump
and, if necessary, they must be cleaned with a damp cloth.
WARNING: removal of the silencer and the air supply fitting must be done when free from powder. Before restarting the pump, ensure that no
powder has entered the pneumatic distributor.
WARNING : Protect always the site and the persons from accidental failure by installing a protection guard to old and collect any
product leakage. Danger of serious injuries and damage to health and/or objects.
46 Rev.02
PRESCRIZIONI PER LA SICUREZZA
ATTENZIONE! RISCHIO DI NATURA CHIMICA. Le pompe sono destinate al funzionamento con diverse tipologie di liquidi e soluzioni chimiche.
Seguire le specifiche istruzioni interne per la decontaminazione durante le operazioni di ispezione o manutenzione.
ATTENZIONE! RISCHIO DI NATURA ELETTRICO. La pompa deve essere sempre messa a terra indipendentemente da altro organo ad essa
collegato. La mancanza di messa a terra o non corretta messa a terra fa decadere i requisiti di sicurezza e tutela dal pericolo di esplosione.
ATTENZIONE: le membrane (a contatto con il prodotto ed esterne) sono componenti altamente soggetti ad usura. La loro durata è
fortemente influenzata dalle condizioni di impiego e dalle sollecitazioni chimi- che e fisiche. Da test eseguiti su migliaia di pompe installate
con prevalenza pari a 0°a 18°C, la durata normale supera i cento milioni di cicli. Per motivi di sicurezza, negli ambienti con pericolo di
esplosione bisogna eseguire lo smontaggio e la verifica della membrana ogni cinque milioni di cicli e la loro sostituzione ogni venti milioni di
cicli.
ATTENZIONE! Nel caso di rottura totale delle membrane, il fluido può entrare nel circuito pneumatico, danneggiarlo ed uscire dallo
scarico. Pertanto bisogna convogliare lo scarico dell’aria in una tubazione fino ad una zona sicura.
ATTENZIONE! laddove l’utilizzatore preveda il rischio di superamento dei limiti di temperatura previsti dal presente manuale, è necessario
installare sull’impianto un dispositivo di protezione che impedisca il raggiungimento della temperatura massima ammessa di processo. In
caso di superamento non è garantito il rispetto della massima temperatura di marcatura.
RICORDA! I rischi per la sicurezza delle persone nascono principalmente per uso improprio o in caso di danni accidentali. Tali rischi
possono essere di ferimento alle mani per chi opera sulla pompa aperta, oppure sono dovuti alla natura dei liquidi che vengono convogliati
da questo genere di pompe. E’ quindi di estrema importanza eseguire diligentemente tutte le istruzioni contenute in questo manuale al fine
di eliminare le cause di incidenti che possano portare alla rottura della pompa ed alla conseguente fuoriuscita di liquido pericoloso per le
persone e per l’ambiente.
Per l’installazione e l’impiego in ambiente potenzialmente esplosivo e non, rispettare le seguenti precauzioni generali:
- controllare che la pompa sia piena e il livello sia, possibilmente, al disopra di essa di 0,5m
- controllare che nel fluido trattato non vi siano o vi possano essere parti solide di dimensioni elevate o forma dannosa non ci siano
restrizioni nell’ingresso o nell’uscita della pompa per evitare fenomeni rispettivamente di cavitazione e sforzo del motore pneumatico
- controllare che le tubazioni di collegamento siano sufficientemente resistenti e che non possano deformarsi sotto il peso della pompa e
dell’aspirazione, né che la pompa subisca il peso delle tubazioni
- se la pompa deve rimanere inattiva per lunghi periodi, pulirla accuratamente facendo circolare un fluido detergente non infiammabile
compatibile con i materiali della pompa.
- se la pompa è stata spenta per lunghi periodi è opportuno far circolare acqua pulita per alcuni minuti per evitare il rischio di incrostazioni
- prima dell’avviamento, dopo lunghi periodi di sosta, effettuare la pulizia delle superfici interne ed esterne con un panno umido, controllare
la messa a terra per ZONE ATEX .
- proteggere sempre la pompa da possibili urti provocati accidentalmente da mezzi in movimento o materiali vari contundenti che possono
danneggiarla e/o reagire al contatto.
- proteggere sempre l’ambiente circostante la pompa da spruzzi provenienti da guasti accidentali alla pompa
ATTENZIONE: l’alimentazione dell’aria non deve mai risultare superiore a 7bar o inferiore a 2 bar
ATTENZIONE: Per tutti i fluidi pompati, soprattutto fluidi aggressivi, tossici o pericolosi per la salute bisogna installare sulla pompa un’adeguata
protezione per il contenimento e la raccolta e segnalazione del prodotto in caso di fuoriuscita: PERICOLO DI INQUINAMENTO,
CONTAMINAZIONE, LESIONI E/O MORTE.
ATTENZIONE: è vietato l’uso della pompa con fluidi non compatibili con i materiali dei componenti o in ambiente con presenza di fluidi non
compatibili.
ATTENZIONE: è vietata l’installazione della pompa in assenza di valvole per l’intercettazione del prodotto sull’aspirazione e sulla mandata per
eseguire il Sezionamento in caso di perdita: pericolo di fuoriuscita incontrollata del prodotto.
ATTENZIONE: è vietata l’installazione della pompa in assenza di valvola di intercettazione, valvola a 3 vie e valvola di non ritorno sul condotto di
alimentazione dell’aria per impedire che il fluido pompato entri nel circuito pneumatico in caso di rottura delle membrane: pericolo di immissione del
fluido nel circuito dell’aria compressa e scarico in ambiente
ATTENZIONE: laddove l’utilizzatore preveda il rischio di superamento dei limiti di temperatura previsti dal presente manuale, è necessario installare
sull’impianto un dispositivo di protezione che impedisca il raggiungimento della temperatura massima ammessa di processo. In caso di superamento
non è garantito il rispetto della massima temperatura di marcatura.
ATTENZIONE: la pompa deve essere sempre messa a terra indipendentemente da altro organo ad essa collegato. La mancanza di messa a terra o
non corretta messa a terra fa decadere i requisiti di sicurezza e tutela dal pericolo di esplosione.
ATTENZIONE: è vietato l’impiego per liquidi infiammabili della pompa in materiale non conduttivo, che si carica statica- mente e senza un’adeguata
messa a terra: PERICOLO DI ESPLOSIONI A CAUSA DI CARICHE STATICHE.
ATTENZIONE: fluidi aggressivi, tossici o pericolosi possono causare gravi lesioni fisiche e/o alla salute pertanto è vieta- to restituire al produttore o
ad un centro di servizio una pompa che contenga prodotti di tale specie: Svuotare e lavare il circuito interno del prodotto e provvedere al lavaggio e
trattamento prima di rispedire la pompa.
ATTENZIONE: i modelli di pompe che contengono componenti o parti in alluminio a contatto con il prodotto non possono essere impiegate per il
pompaggio di III-tricloroetano, il cloro metilene o solventi a base di altri idrocarburi alogenati: PERICOLO DI ESPLOSIONE PER REAZIONE
CHIMICA.
ATTENZIONE: le pompe della serie PHOENIX non possono pompare acetilene, idrogeno e solfuro di carbonio
ATTENZIONE: i componenti dello scambiatore pneumatico, albero compreso, sono costruiti con materiali non specificamente resistenti ai prodotti
chimici. In caso di rottura delle membrane, se vengono a contatto con il fluido, provvedere alla loro completa sostituzione.
ATTENZIONE: il motore pneumatico delle pompe PHOENIX è autolubrificante e non necessita di ulteriore lubrificante;
pertanto evitare l’impiego di aria lubrificata e non essiccata.
ATTENZIONE: verificare che durante il funzionamento non si verifichi rumorosità anomala. In tal caso bloccare immediatamente la pompa.
ATTENZIONE: controllare che nel fluido in uscita non sia presente gas, in tal caso bloccare immediatamente la pompa.
ATTENZIONE: le membrane (a contatto con il prodotto ed esterne) sono componenti altamente soggetti ad usura. La loro durata è fortemente
influenzata dalle condizioni di impiego e dalle sollecitazioni chimiche e fisiche. Da test eseguiti su migliaia di pompe installate con prevalenza pari a 0°a
18°C, la durata normale supera i cento milioni di cicli. Per motivi di sicurezza, negli ambienti con pericolo di esplosione bisogna eseguire lo
smontaggio e la verifica della membrana ogni cinque milioni di cicli e la loro sostituzione ogni venti milioni di cicli.
ATTENZIONE: Bisogna verificare periodicamente l’assenza di polveri e/o depositi dalle superfici esterne ed interne della pompa e, se necessario,
effettuare la pulizia con un panno umido.
ATTENZIONE: lo smontaggio del silenziatore e del raccordo di alimentazione dell’aria deve essere effettuato in assenza di polvere. Prima di riavviare
la pompa assicurarsi che non sia entrata polvere all’interno del distributore pneumatico.
ATTENZIONE: Proteggere sempre l’ambiente e le persone installando una protezione, in caso di
rottura accidentale della pompa, per il contenimento e la raccolta in caso di fuoriuscita del prodotto.
Pericolo di gravi danni fisici, alla salute e/o cose
Per la sostituzione di parti usurate impiegare unicamente pezzi di ricambio originali.
L’inosservanza di quanto sopra può far insorgere pericoli per l’operatore, i tecnici, le persone, la pompa e/o l’ambiente
non imputabili al costruttore.
47 Rev.02
However five general elements are important:
A- all the operations must be carried out by skilled personnel or monitored by qualified personal as appropriate
B- implement personal protection works (when the pump is installed in places involving more than occasional visits) against splashes of fluorescent fluid for
accidental breakage and conveying works
(always) of possible fluid leakages towards collection tanks
C- wear acid-resistant clothing and protection whenever operating on the pump
D- make sure that the Intake and delivery valves are correctly closed during the disassembly
E- make sure that there is no supply to the pneumatic circuit during the disassembly
It should be noted that it is very important to realize systems with pipes well arranged, identifiable, suitably equipped with shut-off valves, with comfortable
compartments and passages for operators who must inspect their status (since the pressure developed by the pump may promote failures to the system if it is
of defective construction or worn).
DISPOSAL
For type of material: separate plastic parts from metal parts and dispose of by authorized companies.
SMALTIMENTO
Per tipologia di materiale: separare le parti in plastica dalle parti in metallo e smaltire tramite imprese autorizzate.
48 Rev.02
TROUBLESHOOTING AND POSSIBLE CAUSES
DEFECT CAUSE SUGGESTION
Circuit without air Check the circuit (valves, connections, regulators, etc.)
Insufficient air pressure Adjust the air pressure
Insufficient air flow rate Check that the pipes and fittings have suitable passages
Damaged control valve Replace
Damaged pneumatic distributor Replace
Open some valves, or remove the pipes and check if the
Pump delivery or intake closed pump starts
The pump does not
1 Damaged discharge cover Replace
start
Check if air comes out of the product delivery pipe, if yes
Broken diaphragm replace the diaphragm.
Disassemble the manifolds and clean the seats or replace the
The balls do not close. balls and the seats.
Excessive intake height. Reduce the intake height.
The pump works
butit does not Install pipes with greater size especially for intake and decre-
2 Too viscous fluid. ase the pumps cycles.
pump.
Clogged intake. Check and clean.
Excessively viscous fluid. No remedy.
The pump works Clogged delivery pipe. Check and clean.
3 with slow cycles Clogged intake. Check and clean.
Internal pneumatic exchanger clogged
or defective Replace the pneumatic exchanger.
Worn shaft. Replace the pneumatic exchanger.
Ice on the discharge. Dehumidify and filter air.
The pump works Check all the air control fittings, especially the quick cou-
4 Air volume is lacking. plings.
irregularly.
Internal exchanger dirty Replace.
Intake clogs during operation. Replace the intake pipe.
Dirty air, full of condensate or oil. Check the air line.
Check the pressure with a gauge installed on thepump and
with running pump. If pressurein that point is too low related
to the mains pressure, check all the air connections, especial-
ly the quick couplings Check that all the air control devices
Insufficient air volume or pressure. have a sufficient flow rate. WARNING: In 90% of the cases
5 The pump stalls stall conditions depends on the quick couplings.
Defective distributor. Replace it.
The stop procedure was not respected. Respect the stop procedure.
The product intake pipe is not correctly
connected. Check.
Clogged pipes. Check and clean.
Install pipes with greater size especially for intake and decre-
Too viscous fluid. ase the pumps cycles.
Disassemble the manifolds and clean the seats or replace the
The balls do not close. balls and the seats.
The pump does not Check pressure with a gauge installed on the pump and with
deliver the flow rate running pump. If pressure in that point is too low related to
6 indicated on the the mains pressure, check all the air connections, especially
table. the quick couplings. Check that all the air control devices have
Insufficient air volume. a sufficient flow rate. WARNING: In 90% of the cases stall
conditions depends on the quick couplings.
49 Rev.02
ANOMALIE DI FUNZIONAMENTO E POSSIBILI CAUSE
DIFETTO CAUSA SUGGERIMENTO
Circuito senza aria Controllare il circuito(rubinetti, allacciamenti, regolatori, ecc.)
Insufficiente pressione dell’aria Regolare la pressione dell’ aria
Insufficiente portata dell’aria Controllare che tubi ed accessori abbiano passaggi adeguati
Valvola di comando danneggiata Sostituire
Distributore pneumatico danneggiato Sostituire
Mandata o aspirazione della pompa Aprire eventuali valvole, oppure rimuovere le tubazioni e
chiuse verificare se la pompa parte
La pompa non si
1 Coperchio scarico danneggiato Sostituire
avvia
Controllare se esce aria dal tubo di mandata prodotto; nel
Membrana rotta caso sostituire la membrana.
Smontare i collettori e pulire le sedi o sostituire le sfere e le
Le sfere non chiudono. sedi.
Altezza di aspirazione troppo elevata. Ridurre l’altezza di aspirazione.
La pompa funziona Installare tubi maggiorati specie in aspirazione e diminuire i
2 Fluido troppo viscoso. cicli della pompa.
ma non pompa.
Aspirazione otturata. Controllare e pulire.
Fluido troppo viscoso. Nessun rimedio.
La pompa funziona Tubo di mandata otturato. Controllare e pulire.
3 con cicli lenti Aspirazione otturata. Controllare e pulire.
Scambiatore pneumatico interno
usurato o difettoso Sostituire lo scambiatore pneumatico.
Albero usurato. Sostituire lo scambiatore pneumatico.
Ghiaccio sullo scarico. Deumidificare e filtrare l’aria.
La pompa funziona Controllare tutti gli accessori di controllo aria, in particolar
4 Manca volume d’aria. modo gli innesti rapidi.
in modo irregolare.
Scambiatore interno sporco Sostituire.
L’aspirazione si ottura durante il
funzionamento.. Sostituire il tubo di aspirazione.
Aria sporca, piena di condensa o di olio.
Verificare la linea dell’aria.
Controllare la pressione con un manometro installato sulla-
pompa ed a pompa in funzione. Se la pressione in quel punto è
troppo bassa rispetto alla pressione di rete, controllare tutti
gli attacchi dell’aria, specialmente quelli ad innesto rapido.
Volume o pressione d’aria insufficiente. Controllare che tutti i dispositivi di controllo aria abbiano
La pompa va in portata sufficiente. ATTENZIONE: Nel 90% i casi di stallo
5 stallo dipendono dagli innesti rapidi.
Distributore difettoso. Sostituirlo.
Non è stata rispettata la procedura di
arresto Rispettare la procedura di arresto.
Il tubo di aspirazione prodotto è mal
collegato. Controllare.
Tubazioni otturate. Controllare e pulire.
Installare tubi maggiorati specie in aspirazione e diminuire
Fluido troppo viscoso. I cicli della pompa.
Smontare i collettori e pulire le sedi o sostituire le sfere e le
Le sfere non chiudono. sedi..
Controllare la pressione con un manometro installato sulla
pompa ed a pompa in funzione. Se la pressione in quel punto è
La pompa non eroga troppo bassa rispetto alla pressione di rete, controllare tutti
6 la portata di tabella. gli attacchi dell’aria, specialmente quelli ad innesto rapido.
Volume d’aria insufficiente. Controllare che tutti i dispositivi di controllo aria abbiano
portata sufficiente. ATTENZIONE: Nel 90% i casi di stallo
dipendono dagli innesti rapidi
50 Rev.02
51 Rev.02
Production head and legal office:
● The INSTRUCTION MANUAL must be delivered to the pump-user , who takes diligent note of it,
keeps the file for subsequent reference. Possible modifications
do not imply updating of the existing manuals
● MANUALE D’USO deve essere consegnato all’utilizzatore della pompa, il quale deve prenderne attenta
visione, e conservarlo per successive consultazioni. Eventuali modifiche non comportano
l’aggiornamento dei manuali preesistenti.
52 Rev.02
TSU Sol. valve Assy
(PAKER)
COILS & HOUSING
60
51.4
38
27
26.5 104
15
37
96.5
135
Subject to Change
aerospace
climate control
electromechanical
filtration
fluid and gas handling
hydraulics
pneumatic
process control
sealing & shielding
Solenoid Valves
27 Page 29 Page
Features
● General Purpose Solenoid Valve
● 2-Way Direct lift
● Operated on Zero MOPD
● Air, Water, Inert gas, Oil, Corrosive fluid and Steam applicapable.
● Applied Seal : FKM, Ruby, Teflon and Factory consult need for NBR, EPDM
Application
● Oil burner and Boiler
● Welding machines
● Water treatment
● Fume analysers
● Gas analysers
● Distribution of light oils
● measurement and control instruments
● Lubrication system
Specifications
● Body - Brass : 121K/F, 122K
StSt 303 :121V
● Seal - FKM, Ruby, Teflon and Factory consult need for NBR, EPDM
● Sleeve - StSt 303, 7000
● Coil - Refer the Electrical part table on the last page
● Body thread type : NPT, PT available with factory consult
EX)
VESSEL HOUSING NUT & WASHER COIL VOLTAGE REMARK
121KBG2LV00 - 4270 - 481000C2 STD / ALL
131VVG2LV00 - H - 1F - 4K KOR ONLY
- I - 1K - C2 ALL
Features
● General Purpose Solenoid Valve
● 3-Way Direct lift
● Operated on Zero MOPD
● Air, Water, Inert gas, Oil, Corrosive fluid and Steam applicapable.
● Applied Seal : FKM, Ruby and Factory consult need for NBR, EPDM
Application
● Air Compressores
● Diesel oil burners
● Water treatment plants
● Vending machine
● Packing machine
Specifications
● Body - Brass : 131K/F, 133K
303 StSt : 131V
● Seal - FKM, Ruby, PTFE and NBR with factory consult
● Sleeve - StSt 303,7000
● Coil - Refer the Electrical part table on the last page
● Body thread type : NPT, PT available with factory consult
Universal
Brass body with S coil StSt body with Scoil Brass body with H coil StSt body with H coil Plastic body with S coil
Features Specifications
● Normally closed ● Body - Brass, StSt 303
● Body : Brass, StSt 303, Plastic ● Seal - FKM and NBR with factory consult
● Coil : S or H ● Sleeve - StSt 303, 7000
● Operated on Zero MOPD ● Coil - Refer the Electrical part table on the last page
● Long life cycle ( I Million )
● Air, Water Inert gas, Oil, Corrosive fluid applicapable
● Manifold confiqurations
● Applied Seal : FKM, NBR, Silicon
● Standard FKM seal
● A miniature, High Performance
● UL & CE components recognized
● UL/CE approved for 'S' coil
● Body thread type : NPT, PF available with factory consult
St. steel 303 Body Assembly Brass Body Assembly Plastic Body Assembly
Features Specifications
● Normally Closed ● Body -Brass, StSt 303 Response Time
● Body : Brass, StSt 303 ● Seal -FKM and NBR with factory ● 25 to 40 milliseconds
● Coil : S or H consult
● Operated on Zero MOPD ● Sleeve -StSt 303, 7000 Duty Cycle
● Air, Water Inert gas, Oil, Corrosive ● Coil -Refer the Electrical part table ● Continuous Duty
fluid applicapable on the last page
Mounting
● Applied Seal : FKM, NBR
● For direct pipe mounting
● A miniature, High Performance or 2-M 5 * 0.8 mm screws
● UL/CE approved for 'S' coil
● Body thread type : NPT, PF available
with factory consult
Normally Closed
Sleeve Body Maximum
VESSEL Body Operating
COIL Pipe Orifice Orifice Seal Fluid
MODEL Kv Pressure( Kgf/cm2 )
& HOUSING Size Size Size disc Temp.
NUMBER Factor
(mm) (mm) DC AC
I-1S-XX
WV131S222GV 1.3 1.6 1.6 6 6 FKM 75
H-WS-XX
I-1S-XX
WV131S222JV PT 1/4 1.9 1.6 2.0 4.5 4.5 FKM 75
H-WS-XX
I-1S-XX
WV131S222LV 2.8 1.6 2.5 3 3 FKM 75
H-WS-XX
* Connection Thread type : Available NPT, PT. Need consultation
* Pipe size : 1/8" available, need consulation
* StSt 303 body available with factory counsult
(Unit : mm)
Brass Body
2/2Way Magnalift
VALVE SECTION
Seal
Dimension
Modle
A B C D E F G H
221G 3/8” 75 93 37.5 15 34 27 10 -
221G 1/2” 75 93 37.5 15 34 27 15 -
221G 3/4” 80 95.5 40 17.5 34 32 20 53
221G 1” 85 102.5 42.5 22.5 46 41 25 53
Dimensions
Blank Page
Features
● Normally closed
● Body : Brass
● 7000 sleeve assembled
● Coil : All of 7000 sleeve coil available
● Operated on nearly Zero MOPD
● Air, Water, Inert gas, Oil and Steam applicapable
● Applied Seal : FKM, NBR, EPDM
● Body thread type : NPT, PF available with factory consult
Specification
Major material features Respons time
● Body -Brass ● Open / Close time : 80 ~ 100 ms
● Seal -FKM,EPDM ,NBR
Ambient temperature
● Sleeve - StSt 303 ,7000
● 0 ~ 50˚C
Rated Voltage & Power Consumption
Fluid temperature
● Please refer to table #1
on the back side ● 60˚C Water & Air (NBR)
● 100˚C Oil (FKM)
● 140˚C Steam (EPDM)
Dimensions
NKV Series
(Unit : mm)
Dimension
Series(Fitting size)
A B C
NKV 3/8" 72 100 85
NKV 1/2" 72 100 85
NKV 3/4" 84 108 90
NKV 1" 95 118 95
How to Order
NKV 15 N-H-1F-C1
❶ ❷ ❸ ❹ ❺
❶ Valve series ❹E
nclosure & Coil
❷ Port size H-1F 1/2" PF Conduit
10 PT3/8" 8 Watt, "F" Class
15 PT1/2" H-1H 1/2" PF Conduit
20 PT 3/4" 14 Watt, "H" Class
25 PT1" ❺ Rated voltage code
❸ Seal material See Table #1
N NBR
F FKM
E EPDM
Features
● Normally closed
● Body : StSt 316
● 7000 sleeve assembled
● Coil : All of 7000 sleeve coil available
● Operated on nearly Zero MOPD
● Oil, gase, Steam and corrosive fluid applicapable
● Applied Seal : FKM, EPDM and NBR with factory consult
● Body thread type : NPT, PT available with factory consult
Application
● Chiller (Semiconductor)
● Food and fluid Press process
● Oil and gase applied equipment
● Printing machine
● Testing equipment
Normally Cloased
FLOW ADMISSIBLE MAX.
REFERENCE NO. PORT DIFF. PRESSURE FLUID
(L/MIN) Power
SIZE ORIFICE SEAL
( L / min ) (BAR) TEMP.(ºC) (W)
(PF) (mm)
VESSEL N0. COIL Liquids Max
HOUSING G Thread Min Water Steam
Lucifer Gobal Korea KV DC AC DC AC
I - 1K - 10 85 14 10
2995 481865 - - 10 85 9 8
4270 481000 - 0 - 10 85 8 8
221S10F 15 40 FKM
4270 486265 - 7 10 85 14 14
H - 1F - 10 85 8 8
H - 1H 3/8" 7 10 85 14 14
2995 481865 - - 10 85 - 9 8
4270 481000 - - 10 85 - 8 8
221S10E 4270 486265 - 15 40 0 4 4 - 140 EPDM 14 14
H - 1F - 10 85 8 8
H - 1H 4 4 - 140 14 14
I - 1K - 10 85 14 10
2995 481865 - - 10 85 9 8
4270 481000 - - 10 85 8 8
221S15F 15 50 0 FKM
4270 486265 - 7 10 85 14 14
H - 1F - 10 85 8 8
H - 1H 1/2" 7 10 85 14 14
2995 481865 - - 10 85 9 8
4270 481000 - - 10 85 8 8
221S15E 4270 486265 - 15 50 0 4 4 - 140 EPDM 14 14
H - 1F - 10 85 8 8
H - 1H 4 4 - 140 14 14
Dimension
unit: mm
Model
Size
221S10F 221S15F 221S20F 221S25F
A 72 72 72 83
B 100 100 117 117
Application
● Water treatment plants
● Industrial washing plants
● Irrigation Systems
● HVAC
● Cooling System
● Fire-fighting system
Fluid Temp.
• Min. : -10℃ Class F Coil or Class H Coil
• Max. : Prefer Table
Diaphragm
Manual Operator
Valve Port
Dimensions
Model
A B C D E F G H
E321G38 110 137.5 55 33 60.5 60 32 70
E321G39 140 144 75 33 67 60 40 99
E321G40 150 158.5 80 41.5 73 75 50 99
Dimensions(mm)
Application
● Fire system
● Air dryer
● Water treatment system
Application
● Oil Burner
● Blow former
● Compressor
● Industrial machine
Normally Cloased
REFERENCE NO. PORT FLOW ADMISSIBLE MAX. Power
SIZE (L/MIN) DIFF. PRESSURE FLUID Consumption
ORIFICE (BAR) TEMP.(˚C ) (W)
Valve Reference COIL (PF) ( L / min ) SEAL
HOUSING (mm)
no G Liquids Max
Min Gas Liquid Oil DC AC
Gobal Korea (Inch) KV DC AC
I - 1K 30 40 - 100 100 14 10
2995 481865 - 25 40 - 100 100 9 8
4270 481000 - 30 40 - 100 100 8 8
E321H21 8 36 0.3 FKM
4270 486265 40 40 - 100 100 14 14
H - 1F 30 40 - 100 100 8 8
1/4"
H - 1H 40 40 - 100 100 14 14
I - 1K 30 40 100 75 100 14 10
2995 481865 - 25 40 100 75 100 9 8
E321H11 8 36 0.3 NBR
4270 481000 - 30 40 100 75 100 8 8
H - 1F 30 40 100 75 100 8 8
I - 1K 30 40 - 100 100 14 10
2995 481865 - 25 40 - 100 100 9 8
4270 481000 - 30 40 - 100 120 8 8
E321H23 11 50 0.3 FKM
4270 486265 40 40 - 100 140 14 14
H - 1F 30 40 - 100 120 8 8
3/8"
H - 1H 40 40 - 100 140 14 14
I - 1K 30 40 100 75 100 14 10
2995 481865 - 25 40 100 75 100 9 8
E321H13 11 50 0.3 NBR
4270 481000 - 30 40 100 75 100 8 8
H - 1F 30 40 100 75 100 8 8
I - 1K 30 40 - 100 100 14 10
2995 481865 - 25 40 - 100 100 9 8
4270 481000 - 30 40 - 100 120 8 8
E321H25 14.5 60 0.3 FKM
4270 486265 40 40 - 100 140 14 14
H - 1F 30 40 - 100 120 8 8
1/2"
H - 1H 40 40 - 100 140 14 14
I - 1K 30 40 100 75 100 14 10
2995 481865 - 25 40 100 75 100 9 8
E321H15 15 60 0.3 NBR
4270 481000 - 30 40 100 75 100 8 8
H - 1F 30 40 100 75 100 8 8
Applications
● Ironing machines
● Tanning plants
● A Sterilizer
● Food and fluid steam Press process
● A dyeing machine
● Hospital equipment
● Tobacco plants
Normally Closed
ADMISSIBLE MAX.
PORT SIZE FLOW Power
REFERENCE NO. DIFF. PRESSURE FLUID
(PF) ORIFICE ( L / min ) (W)
(BAR) TEMP.(ºC) SEAL
G (mm) Liquids
COIL Max
VESSEL N0. HOUSING (INCH) kv Min Steam AC DC
Gobal Korea DC AC
4270 486265 - - 8 180 14 -
708N2 1/2 15 62 0.3
H - 1H-XX - 8 180 14 -
4270 486265 - - 8 180 14 -
708N3 3/4 19 79 0.3 TEFLON
H - 1H-XX - 8 180 14 -
4270 486265 - - 8 180 14 -
708N4 1 28 195 0.3
H - 1H-XX - 8 180 14 -
③ Piston Guide ⑦
④ Piston Seal ⑥
⑤ Spring Piston ⑧ ④
⑥ Piston Spring(SUS)
②
(708N2, 708N3) (708N4)
⑤ ①
⑦ Plunger Assembly
③ Dimension
Model
⑧ Float Buttom Seal OUT IN H P L F G S D W
708N2 129 115 68 36 - - - -
⑨ Coil (High Temp., Class H) 708N3 134.5 118 70 36 - - - -
708N4 - 100.5 98 54 41 59.5 93 115
24/50
24/60
24/50~60
42/50, 48/60
48/50
100/50, 110/60
110/50
110/50, 115/50
110/60
100/50, 115/60
110/60, 115/60
Consumption(Watt)
WV Coil
Housing & Coil Part 7000
Series Insulation Certification USED ON
No. Sleeve
Sleeve Class
DC AC
4270-481000 - 8 8 F NA A2 B2 S7 A4 OA 6J
OK
4270-486265 - 14 14 H NA A2 A5 OP
GP All
2995-481865 OK - 9 8 F CE A2 B2 A5
I-1K OK - 14 10 F UL, CE 8P
H-1F - 8 8 F NA A2 B2 OA 6J
GP All
OK
(KOR Markt)
H-1H - 14 14 H NA A2 A5 OP
115/60
200/50, 220/60
220/50
220/60
220/50, 230/50
200/50, 230/60
220/50, 240/60
220/60, 240/60
230/50
230/60
380/50
380/50, 440/60
12/DC
24/DC
48/DC
110/DC
220/DC
Housing & Coil
Housing type
Part No.
Application and line arrangement of water separator 2)Prevention for solid obstruction entering into the water separator
1.General applications and arrangement If large amount of solid obstruction (sand, rust) is intruded into
Please refer to the applications and arrangement the air flowing into the water separator, the obstruction may stack
on the bowl and cause the operating failure of the auto drain and
which are used normally as Figure 1.
the drain cock, and air leakage. However, the intrusion of the
Cleaning obstruction into the water separator can be prevented by setting the
(Air gun) air filter of the main line filter to the inlet side of the water
Air filter separator.
Regulator
MAINTENANCE
General
industrial Although element replacement is basically unnecessary, extremely dirty air
valve and might cause clogging due to deteriorated oil or rust. Replacement is
Compressor sylinder necessary in that case. (When pressure drop reach 0.1MPa)
Water Air filter Lubricator
separator Regulator Element (gasket, o-ring accessory) model number: AMG-EL***
※ *** is AMG size symbol. (Ex: AMG-EL150)
Painting Hexagon socket head bolt
O-ring(New) O-ring(Used)
Figure1.Application and circuit
Long inlet side piping helps the release of the heat. SMC CORPORATION AMX−OM−F005−C
-3- -4-
BA01149C/07/EN/08.20
71491465
Products Solutions Services
2020-07-31
Operating Instructions
Smartec CLD18
Conductivity measuring system
Smartec CLD18 Table of contents
Table of contents
1 About this document . . . . . . . . . . . 4 9 Diagnostics and
1.1 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 troubleshooting . . . . . . . . . . . . . . . 34
1.2 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . 4 9.1 General troubleshooting . . . . . . . . . . . . . 34
1.3 Symbols on device . . . . . . . . . . . . . . . . . . . . 5 9.2 Troubleshooting instructions . . . . . . . . . 34
9.3 Queued diagnostic messages . . . . . . . . . . 35
2 Basic safety instructions . . . . . . . 5
2.1 Requirements for personnel . . . . . . . . . . . 5 10 Maintenance . . . . . . . . . . . . . . . . . . 38
2.2 Designated use . . . . . . . . . . . . . . . . . . . . . . 5 10.1 Maintenance tasks . . . . . . . . . . . . . . . . . . 38
2.3 Workplace safety . . . . . . . . . . . . . . . . . . . . 5
2.4 Operational safety . . . . . . . . . . . . . . . . . . . 7
2.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . 7
11 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . 39
11.1 General notes . . . . . . . . . . . . . . . . . . . . . . 39
2.6 IT security . . . . . . . . . . . . . . . . . . . . . . . . . . 7
11.2 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
11.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
3 Product description . . . . . . . . . . . . . 8
3.1 Product design . . . . . . . . . . . . . . . . . . . . . . . 8
12 Accessories . . . . . . . . . . . . . . . . . . . . 40
12.1 Calibration solutions . . . . . . . . . . . . . . . . 40
4 Incoming acceptance and
product identification . . . . . . . . . . 9 13 Technical data . . . . . . . . . . . . . . . . . 40
4.1 Incoming acceptance . . . . . . . . . . . . . . . . . 9 13.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.2 Product identification . . . . . . . . . . . . . . . . . 9 13.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
4.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . 10 13.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . 41
4.4 Certificates and approvals . . . . . . . . . . . . 11 13.4 Performance characteristics . . . . . . . . . . 42
13.5 Environment . . . . . . . . . . . . . . . . . . . . . . . 42
5 Installation . . . . . . . . . . . . . . . . . . . . 11 13.6 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.1 Installation conditions . . . . . . . . . . . . . . . 11 13.7 Mechanical construction . . . . . . . . . . . . . 44
5.2 Mounting the compact device . . . . . . . . . 17
5.3 Post-installation check . . . . . . . . . . . . . . . 17 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6 Electrical connection . . . . . . . . . . 17
6.1 Connecting the transmitter . . . . . . . . . . . 17
6.2 Ensuring the degree of protection . . . . . 21
6.3 Post-connection check . . . . . . . . . . . . . . . 21
7 Operation options . . . . . . . . . . . . . 22
7.1 Overview of operating options . . . . . . . . 23
7.2 Structure and function of the operating
menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8 Commissioning . . . . . . . . . . . . . . . . 25
8.1 Switching on the measuring device . . . . 25
8.2 Display settings (Display menu) . . . . . . . 25
8.3 Configuring the measuring device . . . . . 26
8.4 Advanced settings . . . . . . . . . . . . . . . . . . 26
8.5 Calibration (Calibration menu) . . . . . . . . 31
Endress+Hauser 3
About this document Smartec CLD18
Symbol Meaning
Permitted or recommended
Reference to page
Reference to graphic
Result of a step
4 Endress+Hauser
Smartec CLD18 Basic safety instructions
Symbol Meaning
NOTICE
Applications outside specifications!
Incorrect measurements, malfunctions and even measuring point failure could result
‣ Use the product only in accordance with the specifications.
‣ Pay attention to the technical data on the nameplate.
Endress+Hauser 5
Basic safety instructions Smartec CLD18
Electromagnetic compatibility
• The product has been tested for electromagnetic compatibility in accordance with the
applicable international standards for industrial applications.
• The electromagnetic compatibility indicated applies only to a product that has been
connected in accordance with these Operating Instructions.
6 Endress+Hauser
Smartec CLD18 Basic safety instructions
During operation:
‣ If faults cannot be rectified:
products must be taken out of service and protected against unintentional operation.
2.6 IT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any
inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
Endress+Hauser 7
Product description Smartec CLD18
3 Product description
3.1 Product design
5
4
3
2
A0019184
1 Elements
1 Temperature sensor
2 Process connection
3 Leakage bore (offset by 90° in relation to the flow direction)
4 Removable housing cover
5 Window for display
6 Cable glands (M16)
7 Flow opening of sensor
8 Endress+Hauser
Smartec CLD18 Incoming acceptance and product identification
Technical data→ 40
Endress+Hauser 9
Incoming acceptance and product identification Smartec CLD18
Product page
www.endress.com/CLD18
Manufacturer's address
Endress+Hauser Conducta GmbH+Co. KG
Dieselstraße 24
D-70839 Gerlingen
10 Endress+Hauser
Smartec CLD18 Installation
4.4.2 Hygiene
FDA
All materials in contact with the product are FDA-listed materials (apart from the PVC process
connections).
EHEDG
Certified cleanability according to EHEDG Type EL Class I.
When using the sensor in hygienic applications, please note that the cleanability of the
sensor also depends on the way the sensor is installed. To install the sensor in a pipe, use
the appropriate and EHEDG-certified flow vessels for the particular process connection.
3-A
Certified according to 3-A Standard 74- ("3-A Sanitary Standards for Sensor and Sensor
Fittings and Connections Used on Milk and Milk Products Equipment").
EC Regulation No. 1935/2004
The sensor meets the requirements of EC Regulation No. 1935/2004 on materials and articles
intended to come into contact with food.
5 Installation
5.1 Installation conditions
5.1.1 Installation instructions
Hygienic requirements
‣ Easily cleanable installation of equipment according to the criteria of the EHEDG must be
free of dead legs.
‣ If a dead leg is unavoidable, it shall be kept as short as possible. Under no circumstances
shall the length of a dead leg L exceed the pipe’s inner diameter D minus the equipment’s
enveloping diameter d. The condition L ≤ D – d applies.
‣ Furthermore, the dead leg must be self-draining, so neither product nor process fluids are
retained therein.
Endress+Hauser 11
Installation Smartec CLD18
‣ Within tank installations, the cleaning device must be located so that it directly flushes the
dead leg.
‣ For further reference, see the recommendations concerning hygienic seals and installations
in EHEDG Doc. 10 and the Position Paper: “Easy cleanable Pipe couplings and Process
connections”.
For 3-A-compliant installation, please observe the following:
‣ After the device has been mounted, hygienic integrity must be guaranteed.
‣ The leakage hole must be located at the lowest point on the device.
‣ 3-A-compliant process connections must be used.
Orientations
The sensor must be completely immersed in the medium. Avoid air bubbles in the area of the
sensor.
> 1 (3.3)
A0037970
If the flow direction changes (after pipe bends), turbulence in the medium can result.
‣ Install the sensor at a distance of at least 1 m (3.3 ft) downstream from a pipe bend.
12 Endress+Hauser
Smartec CLD18 Installation
The product should flow along the hole of the sensor (see the arrows on the housing). The
symmetrical measuring channel allows flow in both directions.
a
increases for electrically insulating pipes (f > 1) and decreases
for electrically conductive pipes (f < 1). A0037972
1.05
2
1.00
1
0.95
0 10 20 30 40 a [mm]
Endress+Hauser 13
Installation Smartec CLD18
28 (1.10) 85 (3.35)
78 (3.07)
127 (5.00)
A 23 A
(0.91)
69 (2.72)
(2.10)
(0.24)
53.3
6
112 (4.41)
28 (1.10) 85 (3.35)
(2.89)
123 (4.84)
73.5
5.2
(0.20)
B B
(2.10)
53.3
(2.72)
69
20.5
(0.81)
114.5 (4.51)
12.3 (0.49)
136.5 (5.37)
123 (4.84)
32.8 (1.29)
17.1 (0.67)
C D
101.7 (4.00)
86 (3.39)
A0018942
14 Endress+Hauser
Smartec CLD18 Installation
51 (2.01)
Ø 48.3
(1.90)
Ø 44.3
(1.74)
A0019302
Endress+Hauser 15
Installation Smartec CLD18
(1.22)
31
(1.50)
(1.61)
Ø 41
Ø 38
A0022166
(2.26)
57.5
Ø 40 (1.58)
Ø 50 (1.97)
A0024073
8 Installation in DN 40 pipe with 2¼" PVC coupling nut process connection. Dimensions: mm (in)
16 Endress+Hauser
Smartec CLD18 Electrical connection
1. Mount the compact device directly on a pipe nozzle or tank nozzle via the process
connection.
2. For the 1½" threaded connection, use a Teflon tape to seal the connection and an
adjustable pin wrench (DIN 1810, flat face, size 45 to 50 mm (1.77 to 1.97 in)) to
tighten it.
3. When installing, align the compact device in such a way that the medium flows through
the flow opening of the sensor in the direction of medium flow. Use the arrow on the
nameplate to help you align the device.
4. Tighten the flange.
6 Electrical connection
LWARNING
Device is live!
Incorrect connection may result in injury or death!
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
Endress+Hauser 17
Electrical connection Smartec CLD18
mA
IOut1 MRS
Lf
mA
IOut2 L
X
Alarm Out
A0033106
9 Electrical connection
18 Endress+Hauser
Smartec CLD18 Electrical connection
L+
Out+
MRS+
IOut2+
+
IOut1+
L-
L-
Out-
MRS-
IOut2-
IOut1-
L
2
A0029684
10 Terminal assignment
NOTICE
Removing the electronics box will destroy the sensor connection!
‣ The electronics box must not be removed under any circumstances.
‣ Do not open the cover on the electronics box.
The recommended cable cross-section for the connecting cables is 0.5 mm2. The
maximum cable cross-section is 1.0 mm2.
Endress+Hauser 19
Electrical connection Smartec CLD18
4 3
1 2
A0033108
5
4
6
7 3
8
1 2
A0033109
20 Endress+Hauser
Smartec CLD18 Electrical connection
Are the transmitter and cables free from damage on the outside? Visual inspection
Are the signal cables correctly connected as per the wiring diagram?
Are the PE distributor blocks grounded (if present)? Grounding is carried out at the point
of installation.
Endress+Hauser 21
Operation options Smartec CLD18
7 Operation options
1
3
Main µS/cm
- ESC -
_
+ E
4
A0018963
22 Endress+Hauser
Smartec CLD18 Operation options
Always quit menu items / submenus at the end of the menu via "x Back".
Symbols in the editing mode:
Accept entry.
A0020597 If this symbol is selected, the entry is applied at the position specified by the user, and you quit editing mode.
Reject entry.
If this symbol is selected, the entry is rejected and you quit editing mode. The previously set text remains.
A0020598
Delete backwards.
A0020600
If this symbol is selected, the character to the left of the cursor position is deleted.
Delete all.
if this symbol is selected, the entire entry is deleted.
A0020601
Endress+Hauser 23
Operation options Smartec CLD18
Display Settings for the device display: contrast, brightness, time for alternating measured values on the display
* The air set and the correct cell constant have already been configured at the factory for the
Smartec CLD18. A sensor calibration is not necessary during commissioning.
24 Endress+Hauser
Smartec CLD18 Commissioning
8 Commissioning
8.1 Switching on the measuring device
1. Familiarize yourself with the operation of the transmitter before it is first switched on.
After power-up, the device performs a self-test and then goes to the measuring
mode.
2. If you are commissioning the device for the first time, Setup program the as described in
the following sections of the Operating Instructions.
Endress+Hauser 25
Commissioning Smartec CLD18
Out1 0/4 mA 0 to 2000000 μS/cm ‣ Enter the measured value at which the min. current value (0/4 mA) is
0 μS/cm present at the transmitter output.
Out1 20 mA 0 to 2000000 μS/cm ‣ Enter the measured value at which the max. current value (20 mA) is
0 μS/cm present at the transmitter output.
Out2 0/4 mA -50 to 250 °C ‣ Enter the measured value at which the min. current value (0/4 mA) is
0.0 °C present at the transmitter output.
Out2 20 mA -50 to 250 °C ‣ Enter the measured value at which the max. current value (20 mA) is
100.0 °C present at the transmitter output.
Damping main 0 ... 60 s Damping value for the conductivity measured value
0s
Extended setup
Advanced settings→ 26
Manual hold Off, On Function for freezing the current and alarm outputs
26 Endress+Hauser
Smartec CLD18 Commissioning
Hold release 0 to 600 s Prolongs the device hold when the hold condition no longer
0s applies
Alpha coeff. 1.0 to 20.0 %/K Coefficient for linear temperature compensation
2.1 %/K
Process check The process check checks the measuring signal for stagnation. An
alarm is triggered if the measuring signal does not change over a
specific period (several measured values).
Out1 0/4 mA 0 to 2000000 μS/cm ‣ Enter the measured value at which the min. current value
0 μS/cm (0/4 mA) is present at the transmitter output.
Out1 20 mA 0 to 2000000 μS/cm ‣ Enter the measured value at which the max. current value
0 μS/cm (20 mA) is present at the transmitter output.
Out2 0/4 mA -50 to 250 °C ‣ Enter the measured value at which the min. current value
0.0 °C (0/4 mA) is present at the transmitter output.
Out2 20 mA -50 to 250 °C ‣ Enter the measured value at which the max. current value
100.0 °C (20 mA) is present at the transmitter output.
Endress+Hauser 27
Commissioning Smartec CLD18
MRS
Setting for measuring range switching → 31
Out1 0/4 mA 0 to 2000000 μS/cm ‣ Enter the measured value at which the min. current value
0 μS/cm (0/4 mA) is present at the transmitter output.
Out1 20 mA 0 to 2000000 μS/cm ‣ Enter the measured value at which the max. current value
0 μS/cm (20 mA) is present at the transmitter output.
Out2 0/4 mA -50 to 250 °C ‣ Enter the measured value at which the min. current value
0.0 °C (0/4 mA) is present at the transmitter output.
Out2 20 mA -50 to 250 °C ‣ Enter the measured value at which the max. current value
100.0 °C (20 mA) is present at the transmitter output.
Please confirm No
No, Yes
28 Endress+Hauser
Smartec CLD18 Commissioning
1.05
2
1.00
1
0.95
0 10 20 30 40 a [mm]
14 Relationship between the installation factor (f) and the distance from wall (a)
1 Electrically conductive pipe wall
2 Electrically insulating pipe wall
Endress+Hauser 29
Commissioning Smartec CLD18
Where
k(T) = conductivity at process temperature T
k(T0) = conductivity at process temperature T0
The temperature coefficient depends on both the chemical composition of the solution and on
the temperature, and is between 1 and 5 % per °C. The electrical conductivity of the majority
of diluted saline solutions and natural waters changes in a close-to-linear fashion.
Typical values for the temperature coefficient α:
30 Endress+Hauser
Smartec CLD18 Commissioning
General
The calibration of a conductivity measuring system is always performed in such a way that the
suitable calibration solutions determine or verify the exact cell constant. This process is
described in the standards EN 7888 and ASTM D 1125, for example, and the method for
producing a number of calibration solutions is explained.
Endress+Hauser 31
Commissioning Smartec CLD18
32 Endress+Hauser
Smartec CLD18 Commissioning
Endress+Hauser 33
Diagnostics and troubleshooting Smartec CLD18
No measured value displayed No power supply connected Check the device's power supply.
Simulation
Analog out 1 Off Outputs a corresponding value at the "Analog out 1"
0 mA, 3.6 mA, 4 mA, 10 mA, 12 mA, output.
20 mA, 21 mA
Analog out 2 Off Outputs a corresponding value at the "Analog out 2"
0 mA, 3.6 mA, 4 mA, 10 mA, 12 mA, output.
20 mA, 21 mA
Reset device
34 Endress+Hauser
Smartec CLD18 Diagnostics and troubleshooting
Endress+Hauser 35
Diagnostics and troubleshooting Smartec CLD18
36 Endress+Hauser
Smartec CLD18 Diagnostics and troubleshooting
Endress+Hauser 37
Maintenance Smartec CLD18
10 Maintenance
LWARNING
Risk of injury if medium escapes!
‣ Before each maintenance task, ensure that the process pipe is unpressurized, empty and
rinsed.
The electronics box does not contain any parts that the user must maintain.
• The cover on the electronics box may be opened only by the Endress+Hauser Service
Department.
• The electronics box may only be removed by the Endress+Hauser Service Department.
‣ When performing any work on the device, bear in mind any potential impact this may have
on the process control system or on the process itself.
NOTICE
Prohibited cleaning agents!
Damage to the housing surface or housing seal
‣ Never use concentrated mineral acids or alkaline solutions for cleaning.
‣ Never use organic cleaners such as benzyl alcohol, methanol, methylene chloride, xylene or
concentrated glycerol cleaner.
‣ Never use high-pressure steam for cleaning.
38 Endress+Hauser
Smartec CLD18 Repair
11 Repair
The O-ring is defective if medium escapes from the leakage hole.
11.2 Return
The product must be returned if repairs or a factory calibration are required, or if the wrong
product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
11.3 Disposal
Endress+Hauser 39
Accessories Smartec CLD18
12 Accessories
The following are the most important accessories available at the time this documentation
was issued.
‣ For accessories not listed here, please contact your Service or Sales Center.
13 Technical data
13.1 Input
13.1.1 Measured variable
Conductivity
Temperature
40 Endress+Hauser
Smartec CLD18 Technical data
Voltage range 0 V to 30 V
Voltage High Min. 12 V
Voltage Low max. 9.0 V
Current consumption at 24 V 30 mA
Undefined voltage range 9.0 to 12 V
13.2 Output
13.2.1 Output signal
13.2.2 Load
Max. 500 Ω
13.2.3 Characteristic
Linear
Endress+Hauser 41
Technical data Smartec CLD18
13.4.3 Repeatability
13.5 Environment
13.5.1 Ambient temperature range
42 Endress+Hauser
Smartec CLD18 Technical data
13.5.3 Humidity
≤ 100 %, condensating
13.5.10 Altitude
<2000 m (6500 ft)
13.6 Process
13.6.1 Process temperature
Stainless steel process connection:
–10 to 110 °C (14 to 230 °F)
Max.130 °C (266 °F) up to 60 minutes
p (abs.)
[psi] [bar]
188.5 13
1
130.5 9
2
87 6
14.5 1
50 60 110 130 T [°C]
15 Pressure-temperature ratings
1 Stainless steel process connection
2 PVC process connection
A Process temperature increased briefly (max. 60 minutes)
44 Endress+Hauser
Smartec CLD18 Technical data
13.7.2 Weight
13.7.3 Materials
Endress+Hauser 45
Technical data Smartec CLD18
LQP
7 (0.28)
Ø 83 (3.27)
TSK
Ø 84 (3.31)
(0.67)
17.1
(1.29)
32.7
86 (3.39)
101.7 (4.00)
(1.29)
Ø 39.7
32.8
(1.56)
A0018955
46 Endress+Hauser
Smartec CLD18 Technical data
Endress+Hauser 47
Index Smartec CLD18
Index
A Menus . . . . . . . . . . . . . . . . . . . . . . . . 24
Accessories . . . . . . . . . . . . . . . . . . . . . 40 MRS . . . . . . . . . . . . . . . . . . . . . . . . . 31
Advanced Setup . . . . . . . . . . . . . . . . . . 26
Air set . . . . . . . . . . . . . . . . . . . . . . . . 33
N
Application examples . . . . . . . . . . . . . . . 15 Nameplate . . . . . . . . . . . . . . . . . . . . . . 9
C O
Calibration . . . . . . . . . . . . . . . . . . . . . 31 Operating keys . . . . . . . . . . . . . . . . . . . 23
Cell constant . . . . . . . . . . . . . . . . . . . . 31 Operation . . . . . . . . . . . . . . . . . . . . . . 22
Certificates and approvals . . . . . . . . . . . . . 11 Operational safety . . . . . . . . . . . . . . . . . . 7
Cleaning the housing . . . . . . . . . . . . . . . 38 Orientations . . . . . . . . . . . . . . . . . . . . . 12
Commissioning . . . . . . . . . . . . . . . . . . . 25
P
D Post-connection check . . . . . . . . . . . . . . . 21
Declaration of Conformity . . . . . . . . . . . . . 11 Post-installation check . . . . . . . . . . . . . . 17
Designated use . . . . . . . . . . . . . . . . . . . . 5 Power up . . . . . . . . . . . . . . . . . . . . . . 25
Device configuration . . . . . . . . . . . . . . . . 26 Product description . . . . . . . . . . . . . . . . . 8
Device diagnostics . . . . . . . . . . . . . . . . . 34 Product identification . . . . . . . . . . . . . . . . 9
Diagnostic messages . . . . . . . . . . . . . . . . 35 Product page . . . . . . . . . . . . . . . . . . . . 10
Diagnostics . . . . . . . . . . . . . . . . . . . . . 34 Product safety . . . . . . . . . . . . . . . . . . . . 7
Display settings . . . . . . . . . . . . . . . . . . . 25
R
Disposal . . . . . . . . . . . . . . . . . . . . . . . 39
Repair . . . . . . . . . . . . . . . . . . . . . . . . 39
E Residual coupling . . . . . . . . . . . . . . . . . . 33
Electrical connection . . . . . . . . . . . . . . . . 17 Return . . . . . . . . . . . . . . . . . . . . . . . . 39
Ensuring the degree of protection . . . . . . . . 21
S
I Safety instructions . . . . . . . . . . . . . . . . . . 5
Incoming acceptance . . . . . . . . . . . . . . . . 9 Scope of delivery . . . . . . . . . . . . . . . . . . 10
Installation . . . . . . . . . . . . . . . . . . . 11, 17 Symbols . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation conditions . . . . . . . . . . . . . . . 11
T
Installation examples . . . . . . . . . . . . . . . 15
Technical data . . . . . . . . . . . . . . . . . . . 40
Installation factor . . . . . . . . . . . . . . . . . 29
Temperature compensation . . . . . . . . . . . . 30
Interpreting the order code . . . . . . . . . . . . 10
Troubleshooting . . . . . . . . . . . . . . . . . . 34
IT security measures . . . . . . . . . . . . . . . . . 7
Troubleshooting instructions . . . . . . . . . . . 34
M W
Maintenance . . . . . . . . . . . . . . . . . . . . 38
Warnings . . . . . . . . . . . . . . . . . . . . . . . 4
Manufacturer's address . . . . . . . . . . . . . . 10
Wiring . . . . . . . . . . . . . . . . . . . . . . . . 17
Measuring range switch . . . . . . . . . . . . . . 31
Workplace safety . . . . . . . . . . . . . . . . . . 5
Menu . . . . . . . . . . . . . . . . . . . . . . . . . 26
Calibration . . . . . . . . . . . . . . . . . . 31
Diagnostics . . . . . . . . . . . . . . . . . . 34
Display . . . . . . . . . . . . . . . . . . . . . 25
Setup . . . . . . . . . . . . . . . . . . . . . . 26
48 Endress+Hauser
*71491465*
71491465
www.addresses.endress.com
GDS
(HONEYWELL)
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
1 Safety
Ensure that this Operating Manual is read and understood BEFORE installing / operating /
maintaining the equipment. Pay particular attention to Warnings and Cautions. All document
Warnings are listed here and repeated where appropriate at the start of the relevant chapter(s) of
this Operating Manual. Cautions appear in the sections/sub-sections of the document where they
apply.
WARNINGS
Sensepoint XCD is designed for installation and use in Zone 1 or 2 hazardous areas in
many countries including Europe and for Class 1 Division 1 or 2 area applications in North
America.
Installation must be in accordance with the recognized standards of the appropriate
authority in the country concerned.
Access to the interior of the detector, when carrying out any work, must only be conducted
by trained personnel.
Before carrying out any work ensure local regulations and site procedures are followed.
Appropriate standards must be followed to maintain the overall certification of the detector.
For installations where conduit is used, and the sensor is mounted directly to the
Sensepoint XCD, there must be a “Seal Fitting” installed for each of the cable gland entries
within 18 inches of the Sensepoint XCD. The total distance of the location of these Seal
Fittings is 18 inches. (e.g. If all 3 gland entries are to be used, 3 Seal Fittings should be
located, each within 6 inches of the wiring entrance).
If using an anti-seize compound, the threads should be thinly coated with an approved
silicone free compound e.g. petroleum jelly
To reduce the risk of ignition of hazardous atmosphere, de-classify the area or disconnect
the equipment from the supply circuit before opening the detector enclosure. Keep
assembly tightly closed during operation.
Never attempt to open a junction box/enclosure or replace/refit the sensor in potentially
hazardous atmospheres while power is still applied to the transmitter.
The detector must be earthed/grounded for electrical safety and to limit the effects of radio
frequency interference. Earth/ground points are provided inside and outside the unit. The
internal grounding shall be used as the primary equipment ground. The external terminal is
only a supplemental bonding connection where local authorities permit or require such a
connection.
Ensure that all screens/instrument earth/clean earth wiring is earthed/grounded at a single
point (either at the controller or detector - BUT NOT BOTH) to prevent false readings or
alarms that may occur due to potential earth/ground loops.
Take care when handling sensors as they may contain corrosive solutions.
Do not tamper with or in any way disassemble the sensor.
Do not expose to temperatures outside the recommended ranges.
Do not expose sensors under storage conditions to organic solvents or flammable liquids.
At the end of their working life, replacement electrochemical sensors for oxygen and toxic
gas must be disposed of in an environmentally safe manner. Disposal should be according
to local waste management requirements and environmental legislation.
Alternatively, old replaceable sensors may be securely packaged and returned to Honeywell
Analytics clearly marked for environmental disposal.
Electrochemical sensors should NOT be incinerated as this action may cause the cell to
emit toxic fumes.
Refer to the local or national regulations relative to the installation at the site. For Europe
see EN60079-29-2, EN60079-14 and EN61241-14.
Only assessed for ATEX for ignition hazards.
This equipment is designed and constructed as to prevent ignition sources arising, even in
the event of frequent disturbances or equipment operating faults.
The sensor head must be fitted with the supplied weather protection, and mounted so that
the sinter is pointing downward to provide ingress protection IPX6.
The weather protection is a potential electrostatic charging hazard. The manufacturer’s
instructions should be observed.
2
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
2 Information
This manual is for use with the Sensepoint XCD range transmitters only.
The Start-up/Surge/In rush current is dependant on the type of power supply used. The
typical start-up current for Sensepoint XCD is less than 800mA. Measure the start-up
current using the specific power supply before installation to ensure suitability for your
application.
Honeywell Analytics can take no responsibility for installation and/or use of its equipment if
not done so in accordance with the appropriate issue and/or amendment of the Operating
Manual.
The reader of this Operating Manual should ensure that it is appropriate in all details
for the exact equipment to be installed and/or operated. If in doubt, contact Honeywell
Analytics for advice.
The following types of notices are used throughout this Operating Manual:
WARNING
Identifies a hazardous or unsafe practice which could result in severe injury or
death to personnel.
Caution: Identifies a hazardous or unsafe practice which could result in minor injury
to personnel, or product or property damage.
Every effort has been made to ensure the accuracy of this document, however, Honeywell
Analytics can assume no responsibility for any errors or omissions in this document or
their consequences.
Honeywell Analytics would greatly appreciate being informed of any errors or omissions
that may be found in the content of this document.
Honeywell Analytics reserve the right to change or revise the information supplied
in this document without notice and without obligation to notify any person or
organization of such revision or change. If information is required that does not
appear in this document, contact the local distributor/agent or Honeywell Analytics.
3
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
3 Table of contents
1 Safety 2
2 Information 3
3 Table of contents 4
4 Introduction 6
4.1 Transmitter 7
4.2 Flammable, Toxic Oxygen Gas sensors 7
4.3 Accessories 8
5 Installation 10
5.1 Mounting and location 11
5.2 Mounting the transmitter 11
5.3 Installing the sensor 13
6 Electrical connections 14
6.1 Transmitter Wiring 15
6.2 Terminal connections 16
6.3 Power 17
6.4 Cabling 18
6.5 Cable and Earth/Ground regimes 18
6.6 Ground Terminal Wiring 19
7 Default configuration 20
8 Normal Operation 21
8.1 Display Screen 21
8.2 System Status 22
8.3 Magnetic Wand Activation 23
8.4 Operation Mode Structure 23
9 First time switch on (Commissioning) 24
10 Response Check and Calibration 26
10.1 Zeroing and span calibration 26
10.2 Zeroing and span calibration of Hydrogen Sulfide sensors 29
11 General Maintenance 30
11.1 Operational Life 30
12 Servicing 31
12.1 Sensor replacement 31
12.2 Replacing Modules within the Transmitter 33
12.3 Faults and Warnings 34
13 Menu’s and Advanced Configuration 35
13.1 Abort Function 35
13.2 Configuration Mode 35
13.2.1 Configuration mode operation table 38
13.3 Sensor / Gas Selection 40
13.3.1 Sensor Selection 40
13.3.2 Gas Selection 40
13.4 Review Mode 42
4
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
14 General specification 45
15 Ordering information 46
16 Warranty statement 48
17 Installation Drawing 49
17.1 Mechanical Installation Drawing 49
17.2 Electronic Connection Drawing 50
17.3 Duct Mounting Drawing 51
17.4 Collecting Cone Drawing 52
17.5 Mounting Bolt Assy Drawing 53
17.6 Mounting Bracket Drawing 54
18 Certification 55
18.1 China GB Ex and PA 55
18.2 Korea KTL 58
18.3 European ATEX 59
18.4 International IEC 61
18.5 ATEX Name Plate 63
18.6 UL Name Plate 64
18.7 Sensor Cartridges Label 65
19 Cross Interference and Cross Calibration 66
19.1 Cross Interference Table for Toxic and Oxygen 66
19.2 Cross Calibration Flammable Gas Detector 67
5
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
4 Introduction
The Sensepoint XCD (“Exceed”) comprises a gas detector transmitter and a choice
of sensors for detecting flammable gas, toxic gas and oxygen. The construction of
Sensepoint XCD allows it to be used in hazardous area locations; it may also be used in
other areas not classified as hazardous.
The transmitter features a display and three programmable relays for controlling external
equipment e.g. alarms, sirens, valves or switches. The transmitter provides an industry
standard 3-wire, 4-20mA source or sink output for connection to a dedicated gas detection
control system or PLC.
Calibration and maintenance is carried out using a Magnetic Wand, this allows a single
user to undertake routine maintenance without needing to access internal components.
Sensepoint XCD is suitable for use in Zone 1 or 2 hazardous areas (International) or Class
I Division 1 or 2 area applications (North America).
Internal Earth
Point
External Earth
Terminal Module Point (x2)
Locking Seal
Grub
screw Sensor Socket
Enclosure
Sensor Retainer
6
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
4.1 Transmitter
The transmitter enclosure has three threaded entries. The two cable/conduit entries either
side of the upper part of the transmitter housing are for connecting the power source,
signal output and relay contacts to associated signalling equipment. The bottom entry
allows direct connection of the sensor socket. There is a mounting plate incorporated into
the transmitter housing allowing for various mounting configuration options.
A local LCD provides gas type, concentration, alarm and operating status. The display
provides numerical, bar graph and icon information.
Diagnostic information may also be displayed when the transmitter is interrogated using
a magnet. The transmitter cover has a glass window which allows use of the Magnetic
Wand to activate the three user interface magnetic switches that are located on the front
of the display module. The magnet also enables a non intrusive, one-man calibration and
configuration facility for the Sensepoint XCD.
Measuring
Units
Magnetic
DOWN Magnetic
Switch UP Switch
7
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
4.3 Accessories
A range of accessories are available to allow use of Sensepoint XCD in a wide variety of
applications. These including mounting brackets, maintenance tools, weather housings,
flow housing, collecting cone, duct mounting kit and sensor junction boxes.
The Pipe Mounting Bracket (P/N: SPXCDMTMB) The Gassing Cap (P/N: S3KCAL) is an optional
is an optional accessory and may be used to accessory that may be temporarily fitted to the
allow the Sensepoint XCD to be installed onto a sensor in place of the Weather Protection for Gas
dedicated mounting post or existing structure at Response checking during commissioning and
the desired location. routine maintenance.
8
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
The Collecting Cone (P/N: SPXCDCC) is an The Duct Mounting Kit (P/N: SPXCDDMK)
optional accessory that may be fitted to the is an optional accessory that is designed for
Weather Protection. The collecting cone is applications to allow the monitoring for the
designed for use with lighter than air. presence of gas inside a Heating, Ventilation or
Air Conditioning (HVAC) duct is required.
9
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
5 Installation
WARNINGS
Sensepoint XCD is designed for installation and use in Zone 1 or 2 hazardous areas in
many countries including Europe and for Class 1 Division 1 or 2 area applications in North
America.
Installation must be in accordance with the recognized standards of the appropriate
authority in the country concerned.
Access to the interior of the detector, when carrying out any work, must only be conducted
by trained personnel.
Before carrying out any work ensure local regulations and site procedures are followed.
Appropriate standards must be followed to maintain the overall certification of the detector.
Care should be taken when removing and refitting the Sensepoint XCD plug-in Sensor to
the Sensor Socket so that damage to the connection pins can be avoided.
For installations where conduit is used, and the sensor is mounted directly to the
Sensepoint XCD, there must be a “Seal Fitting” installed for each of the cable gland entries
within 18 inches of the Sensepoint XCD. For installations where the sensor will be mounted
remotely from Sensepoint XCD, an additional “Seal Fittings” will be required: One at each of
the conduit entries for the power/signal/relay contact outputs and one at the sensor wiring
entrance. The total distance of the location of these Seal Fittings is 18 inches. (e.g. If all 3
gland entries are to be used, 3 Seal Fittings should be located, each within 6 inches of the
wiring entrance).
If using an anti-seize compound, the threads should be thinly coated with an approved
silicone free compound e.g. petroleum jelly
To reduce the risk of ignition of hazardous atmosphere, de-classify the area or disconnect
the equipment from the supply circuit before opening the detector enclosure. Keep
assembly tightly closed during operation.
Never attempt to open a junction box/enclosure or replace/refit the sensor in potentially
hazardous atmospheres while power is still applied to the transmitter.
The detector must be earthed/grounded for electrical safety and to limit the effects of
radio frequency interference. Earth/ground points are provided inside and outside the unit.
Ensure that all screens/instrument earth/clean earth wiring is earthed/grounded at a single
point (either at the controller or detector - BUT NOT BOTH) to prevent false alarms due to
earth/ground loops.
Take care when handling sensors as they may contain corrosive solutions.
Do not tamper with or in any way disassemble the sensor.
Do not expose to temperatures outside the recommended ranges.
Do not expose sensors under storage conditions to organic solvents or flammable liquids.
At the end of their working life, replacement electrochemical sensors for oxygen and toxic
gas must be disposed of in an environmentally safe manner. Disposal should be according
to local waste management requirements and environmental legislation.
Alternatively, old replaceable sensors may be securely packaged and returned to Honeywell
Analytics clearly marked for environmental disposal.
Electrochemical sensors should NOT be incinerated as this action may cause the cell to
emit toxic fumes.
Refer to the local or national regulations relative to the installation at the site. For Europe
see EN60079-29-2, EN60079-14 and EN61241-14.
Only assessed for ATEX for ignition hazards.
This equipment is designed and constructed as to prevent ignition sources arising, even in
the event of frequent disturbances or equipment operating faults.
The sensor head must be fitted with the supplied weather protection, and mounted so that
the sinter is pointing downward to provide ingress protection IPX6.
The weather protection is a potential electrostatic charging hazard. The manufacturer’s
instructions should be observed.
10
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Caution: The location of gas detectors should be made in accordance with any relevant
local and national legislation, standards or codes of practice. Always replace sensors
with a sensor of the same type.
Gas detectors should be mounted where a potential hazard of gas is most likely to be
present. The following points should be noted when locating gas sensors.
• When locating detectors consider the possible damage caused by natural events e.g.
rain or flooding.
• Consider ease of access to the gas detector for functional testing and servicing.
• Consider how escaping gas may behave due to natural or forced air currents.
Note: The placement of gas detectors should be determined following the advice of
experts having specialist knowledge of gas dispersion, experts having knowledge of the
process plant system and equipment involved, safety and engineering personnel. The
agreement reached on the location of detectors should be recorded.
The Sensepoint XCD transmitter has an integral mounting plate consisting of four
mounting holes on the transmitter body. The transmitter may be fixed directly to a surface
mounting, or to a horizontal or vertical pipe/structure, 40.0-80.0mm (1.6 to 3.1 inches) in
diameter/cross section. The Pipe Mounting Bracket accessory (optional accessory) may
be used for this purpose.
11
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
1. Fit the four spring washers, then the plain washers to the M8 x 80mm SS316 bolts.
2. Pass the four bolts through the four mounting holes of the transmitter housing.
3. Note: If the XCD sunshade is to be used then fit the sunshade and the two locking
bolts and washers to the M8 x 80 bolts.
4. Place the transmitter housing against the mounting position and fit the two “U”
channel mounting bars the other side of the Pipe/Structure.
5. Secure bolts to the threaded holes of the “U” channel mounting bars.
6. Tighten the four bolts securely, (but do not over-tighten) until the transmitter housing
cannot be moved by hand on its mounting position.
12
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
The XCD gas sensor cartridge plugs into the XCD Sensor Socket.
Sensor Socket
Weather Protection
Sensor Retainer
Locking Grub Screw
WARNINGS
Care should be taken when removing and refitting the Sensepoint XCD plug-in Sensor to
the Sensor Socket so that damage to the connection pins can be avoided.
If using an anti-seize compound, the threads should be thinly coated with an approved
silicone free compound e.g. petroleum jelly
Take care when handling old sensors as they may contain corrosive solutions.
Take care when removing or replacing the Sensor Retainer as there may be sharp edges
present on the adjoining threads.
The equipment is designed and constructed as to prevent ignition sources arising, even in
the event of frequent disturbances or equipment operating faults.
Only assessed for ATEX for ignition hazards.
The sensor head must be fitted with the supplied weather protection, and mounted so that
the sinter is pointing downward to provide ingress protection IPX6.
The weather protection is a potential electrostatic charging hazard. The manufacturer’s
instructions should be observed.
2. Use a 1.5mm Allen key (hexagonal wrench-included) to loosen the Locking Grub
Screw on the Sensor Retainer.
4. Remove the XCD Sensor Cartridge from its packaging and plug it into the XCD
Sensor Socket, taking care to align the sensor pins with the connector.
5. Refit the Sensor Retainer and tighten the Locking Grub Screw.
13
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
6 Electrical connections
WARNINGS
Sensepoint XCD is designed for installation and use in Zone 1 or 2 hazardous areas in many
countries including Europe and Class 1 Division 1 or 2 area applications in North America.
Installation must be in accordance with the recognized standards of the appropriate
authority in the country concerned.
Access to the interior of the detector, when carrying out any work, must only be conducted
by trained personnel.
Before carrying out any work ensure local regulations and site procedures are followed.
Appropriate standards must be followed to maintain the overall certification of the detector.
Care should be taken when removing and refitting the Sensepoint XCD Sensor Cartridge to
the Sensor Socket so that damage to the connection pins can be avoided.
For installations where conduit is used, and the sensor is mounted directly to the
Sensepoint XCD, there must be a “Seal Fitting” installed for each of the cable gland entries
within 18 inches of the Sensepoint XCD. The total distance of the location of these Seal
Fittings is 18 inches. (e.g. If all 3 gland entries are to be used, 3 Seal Fittings should be
located, each within 6 inches of the wiring entrance).
If using an anti-seize compound, the threads should be thinly coated with an approved
silicone free compound e.g. petroleum jelly
To reduce the risk of ignition of hazardous atmosphere, de-classify the area or disconnect
the equipment from the supply circuit before opening the detector enclosure. Keep
assembly tightly closed during operation.
The detector must be earthed/grounded for electrical safety and to limit the effects of
radio frequency interference. Earth/ground points are provided inside and outside the unit.
Ensure that all screens/instrument earth/clean earth wiring is earthed/grounded at a single
point (either at the controller or detector - BUT NOT BOTH) to prevent false alarms due to
earth/ground loops.
Take care when handling sensors as they may contain corrosive solutions.
Do not tamper with or in any way disassemble the sensor.
Do not expose to temperatures outside the recommended ranges.
Do not expose sensors under storage conditions to organic solvents or flammable liquids.
At the end of their working life, replacement electrochemical sensors for oxygen and toxic
gas must be disposed of in an environmentally safe manner. Disposal should be according
to local waste management requirements and environmental legislation.
Alternatively, old replaceable sensors may be securely packaged and returned to Honeywell
Analytics clearly marked for environmental disposal.
Electrochemical sensors should NOT be incinerated as this action may cause the cell to
emit toxic fumes.
Refer to the local or national regulations relative to the installation at the site. For Europe
see EN60079-29-2, EN60079-14 and EN61241-14.
14
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Caution: All electrical connections should be made in accordance with any relevant
local or national legislation, standards or codes of practice.
The Sensepoint XCD transmitter may be wired in either Current SOURCE or Current
SINK configuration. These two options are offered to allow greater flexibility in the type of
control system that it can be used with. SOURCE/SINK is selectable via the switch located
on the back side of the display module; accessible by removing the display module during
installation / commissioning (see section 9).
Controller Detector
+VE +VE
1 1
Signal Signal
2 3
RL
-VE -VE
3 2
Controller Detector
+VE +VE
1 1
RL
Signal Signal
2 3
-VE -VE
3 2
15
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Note: Ensure that none of the wires in the terminal area cause an obstruction when
refitting the Display Module. Ensure that the socket on the Display Module is fully engaged
in the Display Module Connector on the Terminal Module.
Ground/Earth
Point
Terminal Block
Terminal Block
Power & Signal
7 for Relays
1
2 8
3 9
4 10
11 Display Module
5 Alignment Pin
Display Module 6 12
13 Locator
Connector
14
15 Note: Terminal
Sensor Blocks are
Connector plug/socket type
and may be
removed to ease
wiring.
16
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
6.3 Power
The Sensepoint XCD transmitter requires a power supply from the controller of between
16Vdc and 32Vdc. Ensure that a minimum supply of 16Vdc is measured at the sensor,
taking into account the voltage drop due to cable resistance.
Example;
The controller is supplying a nominal 24Vdc (V controller), the detector minimum allowable
voltage is 16Vdc (V detector min), therefore the maximum allowable voltage drop between the
controller and detector is 8Vdc; this means a voltage drop of 4V in each core (+ve core
and -ve core).
+20V
(wrt Controller 0V) +24V
4-20mA Signal
16V (min) XCD Controller
+4V 0V
(wrt Controller 0V) Diagram 8: Power Connection
Power consumption of the detector is 5.0W. The current required to drive the detector at
the minimum voltage is (I = P / V), 5.0 / 16 = 312.5mA (I detector).
So, the maximum field cable loop resistance (R loop) = 8 / 0.31 = 26 Ohms, or 13 Ohms
per core, (allowing for component variations, losses, etc.).
The following tables show the maximum cable distances between the controller and
transmitter assuming a voltage drop of 4V in each core and for different cable parameters.
The tables are examples only and actual cable parameters and source power supply
voltage for the application should be used to calculate the maximum cable distance
allowed at the installation site.
6.4 Cabling
The use of industrial grade, suitably armoured field cable is recommended.
For example, screened 3 cores (plus screen 90% coverage), suitably mechanically
protected copper cable with a suitable M20 explosion-proof gland, or ¾” NPT steel
conduit, with 0.5 to 2.5 mm2 (20 to 13 AWG) conductors. Ensure the cable gland is
installed correctly and fully tightened. All unused cable/conduit entries must be sealed with
a suitable certified sealing plug (one plug is supplied).
Effective Earth/Ground bonding is important to ensure good EMC and RFI immunity.
The following diagrams show examples of how to earth/ground bond the cable at
enclosures. The same principles apply to conduit installations. These bonding techniques
provide good RFI/EMC performance. Earth/ground loops must be avoided to prevent the
risk of false signal variation.
Controller
Clean / Instrument
Earth Screen
Cable to Transmitter
Transmitter Controller
Power & Signal
Earth Screen
18
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
The Earth Screen of the field cable should be “tied to Earth” or connected to Ground at
one point only. It is common practise to adopt a STAR EARTH connection regime where
all instrumentation Screens are connected at one common point.
The Screen at the other end of the cable should be “parked” or terminated into a blank
terminal.
The following diagrams show how to install the wire into the ground screw of the
enclosure.
19
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
7 Default configuration
The Sensepoint XCD transmitter is supplied with the following default configuration.
Gas Name Default Range Lower Alarm Lower Alarm Higher Alarm Higher Alarm
Type Type
Oxygen 25.0%VOL 19.5%Vol Falling 23.5%Vol Rising
Hydrogen Sulfide 50.0ppm 10.0ppm Rising 20.0ppm Rising
Carbon Monoxide 300ppm 100ppm Rising 200ppm Rising
Hydrogen 1,000ppm 200ppm Rising 400ppm Rising
Flammable IR 100%LEL 20%LEL Rising 40%LEL Rising
Flammable CAT 100%LEL 20%LEL Rising 40%LEL Rising
Carbon Dioxide IR 2.00%Vol 0.40%Vol Rising 0.80%Vol Rising
For details of how to change the configuration of the Sensepoint XCD please refer
to section 13.
20
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
8 Normal Operation
Sensepoint XCD is supplied configured and ready for use according to the “Default
Settings” table shown above. However these setting may be tailored to a specific
application requirement using the Sensepoint XCD configuration menu system.
Access to the Sensepoint XCD transmitter’s configuration menus system is via the
Magnetic Activation Tool.
During normal operation the instrument display shows a steady GREEN backlight.
During Low and High Gas Alarm it displays a flashing RED backlight
During Fault condition the instrument display shows a flashing YELLOW backlight.
The screen is visible through the window of the transmitter’s cover. The display shows the
gas concentration (both graphically and numerically), range, units, alarm/
fault status, etc.
Note: The detector display may become sluggish in sub-zero temperatures and possibly
unclear at temperatures below -40 °C, but the detector continues its gas monitoring
function. The display is not damaged and recovers when the temperature increases.
Full Scale
Bar Graph
(Proportional Test Pass Icon
to gas reading)
Gas Type
Measuring
Unit
Gas Reading
21
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
System Status
Relay Back
Status Display Current Output
A1 A2 Fault Light
F-XX
Fault:
fault number
Circuit or sensor 0-1.0 mA Yellow,
with fault icon
error flashing
blinking
System Fault N/A 0-0.15mA
Note: In the event of processor failure the watchdog will automatically reset the system for
recovery.
W-XX
Green
warning number Dependent on
Warning Steady
with fault icon system status
blinking
0.0
Green
Normal Gas 4-20 mA
Steady
concentration
Gas
concentration. Red,
Alarm 1 4-20 mA
1st alarm icon flashing
blinking
Gas
concentration. Red,
1
Alarm 2 4-20 mA
2nd alarm icon flashing
blinking
Full scale icon
Red,
2
Over-range and reading 22mA
flashing
blinking
Inhibit icon
dependent 2 or 4mA
on Menu depending on
command. If configuration. Green
Inhibit
any relay is set Note: Steady
to inhibit relay, 2 or 17.4mA for
then inhibit relay Oxygen version
will be activated.
Note:
1. For Oxygen, A1 relay will not be activated because AL1 is configured to activate above
23.5 %Vol (in Rising Alarm configuration).
2. For Oxygen, A2 relay will not be activated because AL2 is configured to activate below
19.5 %Vol (in Falling Alarm configuration).
22
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
1. Monitoring mode, is the normal operating status while XCD measures and displays
gas concentration. The fault/warning status is periodically checked, relay contacts
are activated according to the configuration.
3. Review mode, allows the user to view the current configuration settings.
Start - Up
Further details of the information available and configuration options for the Sensepoint
XCD can be found in Section 13. of this manual.
23
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
WARNING
The following procedure requires the Transmitter Cover to be removed while carrying out
supply voltage checks. Therefore the appropriate permits to work should be sought in
preparation.
Prior to carrying out any HOT WORK ensure local and site procedures are followed.
Ensure that the associated control panel output actuation is inhibited so as to prevent false
alarms.
Caution: The following procedure should be followed carefully and only performed by
suitably trained personnel
Note: Calibration is mandatory before the detector can be used for gas monitoring. Refer
to Section 10.1 Calibration for the proper procedure.
1. Remove the transmitter housing cover and detach the display unit by lifting the semi-
circular handle and pulling the assembly directly away from the termination module
(without twisting it)
2. Configure the detector's analogue output signal configuration for Current SINK or
Current SOURCE operation using the selector switch located on the back of the
display module. The default setting is Current SOURCE.
3. Check that all electrical connections are terminated correctly as per section 6.
4. Switch On the external power supply to the transmitter at the safe area gas detection
controller (or PLC).
5. Using a Digital Multi Meter (DMM), check the Supply Voltage at the terminals 1 (24V)
and 2 (0V), this should be a minimum supply voltage of 16Vdc (Maximum supply
voltage is 32V DC)
6. Switch Off the external power to the detector.
7. Refit the Display Module and Cover.
Note: Ensure that none of the wires in the terminal area cause an obstruction when
refitting the Display Module. Ensure that the socket on the Display Module is fully engaged
in the Display Module Connector on the Terminal Module.
10. A start up sequence will then be displayed, similar to the one shown in Diagram 15.
24
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Note:
For a full description of each screen shown in Diagram 15., please refer to Section 13.3
“Review Mode” of this Manual.
25
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
1. A simple Response Check often referred to as a “BUMP TEST” is a test using calibration
gas applied to the sensor via the nozzle of the Weather Protection or using the Sensepoint
XCD Gassing Cap.
If a BUMP TEST is done via the Weather Protection nozzle it may be necessary in windy
conditions to increase the flow rate of the test gas by a further 1 LPM, OR, to shelter the
weather protection from the wind.
2. A full gas calibration of the sensor as described in the following section, using ONLY the
Sensepoint XCD Gassing Cap (Part No.: S3KCAL).
Caution: Before initial calibration allow the detector to stabilize for 30 minutes after applying
power.
When in zeroing and span calibration mode the current output from the detector is inhibited
(default 2mA) to avoid false alarms.
For Flammable gas calibration use a calibration gas concentration of between 25%LEL and
75%LEL to ensure that the required accuracy can be attained.
For Hydrogen Sulfide type sensors refer to section 10.1.2 before conducting zero and span
calibration.
To calibrate the detector, use an appropriate span gas cylinder, constant flow regulator
AND the Sensepoint XCD Gassing Cap (see section 4.7). The flow rates used for
calibration gas are as follows:
Gas Type Flow rate (L / Min)
Air or N2 for Zero 0.5 to 1.0
Flammable CAT 1 to 1.5
O2 0.5 to 1.0
H2S 0.5 to 1.0
CO 0.5 to 1.0
H2 0.5 to 1.0
Flammable IR 0.4 to 0.6
CO2 IR 0.4 to 0.6
A compressed air cylinder (20.9%Vol oxygen) should be used to perform the zero
calibration if the area where the detector is located contains any residual amount of the
target gas. If no residual gas is present then the background air can be used to perform
the zero calibration. Contact your Honeywell Analytics representative for details of suitable
calibration kits.
Note: the Oxygen sensor does not require a zeroing procedure. Background air (20.9%Vol
oxygen) can be used to span the oxygen sensor in place of a compressed air cylinder
(20.9%Vol oxygen). For oxygen sensors only do parts 1-4, 12, 13 (if compressed air
cylinder is used), 14-17 and 22 of the procedure below.
26
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
(ZERO CALIBRATION)
1. If the ambient air is NOT considered reliable to use to set the ZERO, then remove
the weather protection and fit the Gassing Cap accessory (see Section 4.3) onto the
sensor and apply a clean source of zero gas or compressed air.
2. To access the calibration menu, hold the end of the magnet over the switch located
at the top center of the detector display () for at least 3 seconds and then remove.
3. The display will indicate the first configuration mode menu ‘SEt CAL’.
4. Put the magnet over the ‘’ switch again and move to enter the Calibration menu.
5. The display will show the current gas reading, and the ‘ ’ icon flashes.
6. When the zero gas reading is stable use ‘’ to confirm zero calibration.
7. If successful the display shows ‘ZEro PASS’ (if not successful, the display shows ‘ZEro
FAIL’ and returns to configuration mode).
10. If span calibration is required use ‘’ proceed to the next step. If span calibration
is not required, use ‘’ to select ‘No’ and ‘’ to return to configuration mode.
27
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
(SPAN CALIBRATION)
11. The display shows the current calibration span gas concentration while flashing the
‘ ’ icon. Use ‘’ to change the calibration span gas concentration, and ‘’ when
required span calibration level is set.
12. The display will show the current gas reading, and the ‘ ’ icon flashes.
14. Apply the span gas to the sensor using the Sensepoint XCD Gassing Cap (see
section 4.7 for description). The live gas reading is displayed. When the reading is
stable, use ‘’ to confirm span calibration.
15. If the sensor has been replaced the following display may be shown.
16. Use ‘’ to select ‘YES’ if the sensor has been replaced or ‘No’ if it has not been
replaced.
17. If the span calibration is successful the instrument will briefly display ‘SPAn PASS’ (if
fails ‘SPAN FAIL’ displayed and returns to configuration mode).
Note: the calibration due warning counter is reset after a successful calibration. See
section 12.3 for further details of setting a calibration due warning.
18. The display alternates between “Purg gAS” and the gas reading to indicate that the
unit is expecting the span gas to be removed from the sensor.
28
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
19. Promptly switch off the calibration span gas and remove the Sensepoint XCD
Gassing Cap from the sensor to allow the gas to disperse.
20. When the reading falls below 50% of the calibration gas level the display indicates a
countdown (up to 180 seconds dependant on gas type).
22. The instrument returns to the ‘Set CAL’ menu. Activate the ‘’ or ‘’ switch to select
another menu or select ‘QuIT’ to return to normal monitoring mode.
Note: Remember to always replace the Weather Protection and other accessories.
When calibrating Hydrogen Sulphide cartridges the following should be taken into account
while following the procedure in 10.1:
1. To zero the sensor, use a compressed air cylinder of 20.9%Vol oxygen (not Nitrogen).
Do not use background air.
2. Apply the gas to the sensor for three minutes before using ‘’ to confirm the zero
calibration.
4. Apply the span gas for two minutes before using ‘’ to confirm the span calibration.
29
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
11 General Maintenance
WARNINGS
Access to the interior of the transmitter, when carrying out any work, must only be
conducted by trained personnel.
Care should be taken when removing and refitting the Sensepoint XCD plug-in Sensor
Cartridge to the Sensor Socket so that damage to the connection pins can be avoided.
Before carrying out any work ensure local regulations and site procedures are followed.
Appropriate standards must be followed to maintain the overall certification of the sensor
and transmitter.
To reduce the risk of ignition of hazardous atmosphere, de-classify the area or disconnect
the equipment from the supply circuit before opening the transmitter enclosure. Keep
assembly tightly closed during operation.
Take care when handling sensors as they may contain corrosive solutions.
Do not tamper with or in any way disassemble the sensor.
Do not expose to temperatures outside the recommended ranges.
Do not expose sensors under storage conditions to organic solvents or flammable liquids.
At the end of their working life, replacement electrochemical sensors for oxygen and toxic
gas must be disposed of in an environmentally safe manner. Disposal should be according
to local waste management requirements and environmental legislation.
Alternatively, old replaceable sensors may be securely packaged and returned to Honeywell
Analytics clearly marked for environmental disposal.
Electrochemical sensors should NOT be incinerated as this action may cause the cell to
emit toxic fumes.
Honeywell Analytics recommends that gas detectors are tested and re-calibrated on a six-
monthly basis, or according to site practice. For remote Flammable or Toxic sensors are
used with the XCD Transmitter, please refer to the individual sensor manuals for specific
recommended calibration periods.
The pellistors used in the Catalytic flammable gas sensor can suffer from a loss of
sensitivity when in the presence of poisons or inhibitors, e.g. silicones, sulphides, chlorine,
lead or halogenated hydrocarbons. The pellistors are poison resistant to maximize the
operational life of the Catalytic flammable sensor. A typical operating life, subject to the
presence of poisons/inhibitors is 36 months.
The NDIR (infrared) flammable gas sensor is not affected by the above-mentioned poisons
and therefore has a longer life span. A typical operating life is 5 years.
Typical life of a toxic gas sensor is dependant on the application, frequency and amount of
gas exposure. Under normal conditions (3 monthly visual inspection and 6 monthly test/re-
calibration), the XCD Oxygen and other toxic sensors have an expected life equal to or
greater than 24 months.
Caution: Oxygen deficient atmospheres (less than 6%V/V) may result in inaccuracy of
reading and performance.
12 Servicing
WARNINGS
Take care when handling sensors as they may contain corrosive solutions. Do not tamper
or in any way dis-assemble the sensor. Do not expose to temperatures outside the
recommended range. Do not expose sensor to organic solvents or flammable liquids.
Care should be taken when removing and refitting the Sensepoint XCD plug-in Sensor
Cartridge to the Sensor Socket so that damage to the connection pins can be avoided.
At the end of their working life, replacement electrochemical sensors for oxygen and toxic
gas must be disposed of in an environmentally safe manner. Disposal should be according
to local waste management requirements and environmental legislation.
Alternatively, old replaceable sensors may be securely packaged and returned to Honeywell
Analytics clearly marked for environmental disposal.
Electrochemical sensors should NOT be incinerated as this action may cause the cell to
emit toxic fumes.
Caution: The following procedure should be followed carefully and only performed by
suitably trained personnel.
A fault condition will be signalled by the detector if the sensor is removed with the unit
under power.
The Flammable Catalytic and Toxic ECC sensor cartridge that are used with the
Sensepoint XCD Sensor Socket have no serviceable parts. When they have reached the
end of their operational life, simply replace the cells.
31
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Sensor Socket
Sensor Cartridge
Weather Protection
Sensor Retainer
To replace the plug-in sensor of a Sensepoint XCD Sensor Socket use the following
procedure:
2. Remove the Weather Protection or other accessories from the sensor socket
thread.
3. Loosen the Locking Grub Screw and unscrew the Sensor Retainer
4. Carefully pull the old Sensor from the sensor socket without twisting.
6. Calibrate sensor
WARNING
Ensure that the same Gas Type and Range of Sensor is fitted in place of the old Sensor.
Care should be taken when removing and refitting the Sensepoint XCD plug-in Sensor
Cartridge to the Sensor Socket so that damage to the connection pins can be avoided.
The sensor head must be fitted with the supplied weather protection, and mounted so that
the sinter is pointing downward to provide ingress protection IPX6.
32
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Two replaceable module assemblies are located within the transmitter housing. The
Display Module and the Terminal Module.
The Display Module is simply removed by unplugging it from the Terminal Module (this
procedure is done during normal installation).
33
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
34
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
In Review Mode or Configuration Mode the user can escape one step back from the
current position using the Abort Function. To do this the user must activate the Enter
switch for more than 3 seconds with the Magnetic Wand. Switching between each pair of
modes or between menus and sub menus are shown in the following table.
From To Example
Configuration Mode Configuration Activate Enter switch for more than 3 seconds
sub menu Mode main menu while in a sub menu
The table below shows the functions available via the configuration menu that can be
displayed on the transmitter and accessed using the Magnetic Wand.
The instrument will show the main Menu when the “Enter” switch is activated with the
Magnetic Wand and held for at least 3 seconds.
The Menu is password protected to prevent any unauthorized changes. The password is
initially disabled and the default password is ‘0000’. If the default password is changed
to other than ‘0000’, then the password is enabled automatically and requested when
entering Configuration Mode.
With the Menu showing, the following functions can be performed: calibration, bump test,
sensor selection and configuration of parameters such as measuring range, calibration
gas level, calibration interval, inhibit current, inhibit timeout, alarm setting, relay setting,
password change, location setting, temperature unit reading, force analogue output and
alarm function checks.
While in Configuration mode, the output current of the transmitter is inhibited to prevent
false alarms.
Names, displays and descriptions for each menu item in Configuration Mode are shown in
the following table.
35
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Enable/disable password
Set Password Set password
Default – no password (Select ‘0000’)
36
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Monitoring
Mode
Menu Switch
( > 3 Seconds)
Quit/ESC command
Auto quit after no activity of timeout
No
Access permitted?
Abort
Configuration Mode
Set Range Relay Operation
(Inhibited)
Configure Set Location
Inhibit
Bump Test
* Note : “Select Sensor” and “Select Gas” menus are only available for flammable/IR sensors.
Diagram 14: Configuration Mode
37
13.2.1 Configuration mode operation table
Configuration mode allows the user to perform calibration and configure parameters such as full scale range, calibration gas level,
calibration interval, inhibit current & timeout, alarm settings, relay settings, set a password, etc. To activate Configuration mode hold
the magnet over the ENTER switch for at least 3 seconds and then remove. Configuration mode can be password protected to prevent
unauthorized personnel from changing parameters. Initially the password is set to ‘0000’ meaning it is disabled. While in Configuration
mode, the output current of the detector is inhibited to prevent false alarms.
Use the table below to help navigate the menus and make configuration changes. The menus are shown in the left hand column. Use
to select the required menu and ‘’ to enter. Follow the information and instructions in the table from left to right from the required
menu.
OK OK OK OK
GAS NAME, CURRENT CONCENTRATION AND
‘SPAN’ DISPLAYED AND ‘YES’ FLASHING ‘ ’ ICON DISPLAYED. APPLY SPAN GAS
Sensepoint XCD Technical Manual
38
‘SEnS’ DISPLAYED WITH SENSOR TYPE ‘CAL YES’ DISPLAYED TO ASK
SEL2 DISPLAYED. USE TO SELECT IF YOU NOW WANT TO PERFORM
CALIBRATION. USE ‘ ’ TO PROCEED
SENS DIFFERENT SESNOR. OR USE TO SELECT ‘No’ AND
RETURN TO MENU MODE.
IF GAS TYPE IS CHANGED,
‘9AS’ DISPLAYED WITH GAS TYPE ‘CAL YES’ DISPLAYED TO ASK IF
SEL YOU NOW WANT TO PERFORM
DISPLAYED. USE TO SELECT CALIBRATION. USE ‘ ’ TO
9AS2 DIFFERENT GAS. PROCEED OR USE TO SELECT
‘No’ AND RETURN TO MENU MODE.
IF RANGE IS CHANGED, ‘CAL’ IF RANGE IS CHANGED, ‘ALrm’
BAR GRAPH INDICATING CURRENT RANGE, DISPLAYED AND ‘YES’ FLASHING TO DISPLAYED AND ‘YES’ FLASHING TO
SEt ‘rAn9’ DISPLAYED & CURRENT RANGE ASK IF YOU NOW WANT TO ASK IF YOU NOW WANT TO MODIFY
UNIT RETURNS TO MENU MODE
rAn9 FLASHES. USE TO SELECT DIFFERENT PERFORM CALIBRATION. USE ‘ ’ TO ALARM SETTINGS. USE ‘ ’ TO
RANGE PROCEED OR USE TO SELECT PROCEED OR USE TO SELECT
‘No’. ‘No’ AND RETURN TO MENU MODE.
‘tImE’ DISPLAYED WITH ‘ ’ ICON.
‘Inhb’ DISPLAYED WITH ‘ ’ ICON
FLASHING CURRENT INHIBIT
ConF FLASHING. CURRENT INHIBIT mA VALUE
TIMEOUT PERIOD (MINUTES)
FLASHES. USE TO SELECT NEW UNIT RETURNS TO MENU MODE
Inhb FLASHES. USE TO SET NEW
VALUE (2 OR 4mA TOXICS, OR 2 OR 17.4mA
TIMEOUT. (IF SET TO 0 OUTPUT IS
OXYGEN VERSION).
PERMANENTLY INHIBITED)
‘PASS’ DISPLAYED WITH FIRST DIGIT
SEt PLACE ICON OF THE PASSCODE FLASHING.
USE TO SELECT 1ST DIGIT OF CURRENT UNIT RETURNS TO MENU MODE
PASS PASSWORD. USE TO MOVE TO NEXT
DIGIT AND SET REST OF PASSCODE.
SPXCDHMANEN Issue 2
Forc ‘Forc’ DISPLAYED WITH ‘ ’ ICON. DEFAULT UNIT TRANSMITS THE CURRENT
FORCE CURRENT ‘4.00’ FLASHES. USE IF YOU WANT TO EXIT FROM THIS
Curr TO CHANGE TO REQUIRED mA LEVEL. MENU, USE ABORT FUNCTION
39
OPr FLASHES. USE TO CHANGE TO
CHANGE TO REQUIRED RELAY-ON DELAY. REQUIRED LATCH OPTION.
REQUIRED RELAY-OFF DELAY.
‘Loc’ DISPLAYED WITH FIRST 4
CHARACTERS OF THE LOCATION STRING. ‘Loc’ DISPLAYED NEW LOCATION
USE TO CHANGE THE 1ST CHARACTER STRING. THE STRING MOVES
Set OF CURRENT LOCATION STRING. USE RIGHT-TO-LEFT TO SHOW WHOLE
Loc TO MOVE TO NEXT CHARACTER AND SET CHARACTERS TWICE. THEN UNIT
REST OF STRING. MAXIMUM 12 AUTOMATICALLY RETURNS TO
CHARACTERS CAN BE SET. MENU MODE.
tESt ‘Forc’ DISPLAYED and ’AL1’ FLASHING TO GAS NAME, CURRENT GAS
SELECT TEST-REQURED ALARM. USE ‘ ’ CONCENTRATION DISPLAYED AND UNIT RETURNS TO MENU MODE
ALrm TO FORCE ALARM1. ‘ ’ ICON AND AL ICON FLASHE.
OK OK OK OK
1
Refer to section 10.1 for detailed zero and span calibration procedures. A detector fitted with an oxygen sensor will skip the zeroing
procedure. Re-calibrate the detector if left un-powered for periods in excess of 24 hours.
2
This menu is only available and visible when the unit is fitted with flammable/IR sensor.
SPXCDHMANEN Issue 2
3
The calibration due warning counter is automatically reset after a successful calibration.
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
NOTE:
This Configuration option is not available for XCD units with EC sensors
Selecting ‘’ will move the operator to the Select Sensor screen. The first screen displays
the currently configured sensor. To select a new mV sensor, use ‘’ to move through
the list, then use ‘’ to make the selection or discard the selection and return to menu
mode by using abort function. If type of sensor is changed, calibration prompt will appear
to ask if you want to perform calibration. If the sensor type is not changed, it means that
XCD has detected the sensor type automatically and a user does not need to configure
the sensor type manually.
Use “SEL gAS” to set the target gas for sensors capable of detecting multiple gases. The
available gases:
40
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Gas selection is dependent on the type of sensor attached to the XCD. If Ir-1/Ir-3 sensor
is attached, then a user can only select CO2/mEt gas respectively. But when Cb-1 sensor
is attached to the XCD, a user can select the gas from Str1 (1*) to Str8 (8*). For more
information on star rating, please refer to section 19.2.
The current configuration of the XCD is displayed and by using the ‘’ switch to enter “SEL
gAS” menu. To select a flammable gas, use ‘’ to move through the list, then use ‘’ to
make the selection or abort function to discard the selection and return to menu mode.
If type of gas is changed, calibration prompt will appear to ask if you want to perform
calibration.
...
41
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
The instrument will enter Review mode when the “Enter” switch is activated with the
Magnetic Wand and held for around one second.
Names, displays and descriptions for each review item in Review Mode are shown in the
following table.
42
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Note:
*Power voltage and internal transmitter temperature may be different from actual value
due to measuring accuracy and internal heating components.
43
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Menu Switch
(1s to 3s)
Review Monitoring
Mode Abort Mode
44
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
14 General specification
Sensepoint XCD Transmitter
Use 3-wire, 4-20mA, gas detector transmitter for use with directly installed flammable and toxic gas
sensors. For the protection of personnel and plant from flammable and toxic gas hazards.
Electrical
Input Voltage Range: 16 to 32Vdc (24Vdc nominal)
Max Power Consumption: Max 5 Watts. at 24Vdc (see section 2 regarding maximum in
rush current)
Current output 4-20mA (Source or Sink)
≥0.0<1.0 mA Fault (refer to table 5 section 12.3 for further details).
4.0 mA to 20.0 mA Normal gas measurement
2.0 mA or 4.0 mA (17.4 mA) Inhibit (during configuration/user settings)
22.0 mA Maximum over range
Terminals 15 x screw terminals suitable for wire diameter 0.5mm 2 to
2
2.5mm (20AWG to 13AWG).
Relays 3 x 5A@250VAC. Selectable normally open or normally closed
(switch) and energized/de-energized (programmable).
Communication RS485, Modbus RTU (Pending)
Construction
Material Epoxy painted aluminium alloy or 316 Stainless Steel
Weight Aluminium alloy: 1.7kg, 316 Stainless Steel: 3.7kg
Mounting Pole or wall mounting
Entries 2 x M20 (for ATEX/IECEx/AP Approval) or 2x3/4NPT (for UL Approval)
Detectable Gases & Performance (See notes below)
Response
User Selectable Full Default User Selectable Default Cal Operating Temperature* Default alarm points
Gas Steps Time (T90) Accuracy
Scale Range Range Cal Gas Range Point
secs Min Max A1 A2
Electrochemical Sensors
O xyg e n 2 5 .0 % V /V o n ly 2 5 .0 % V /V n /a 2 0 .9 % V /V (F ixe d ) 2 0 .9 % V /V <30 < + /-0 .5 % V o l. -2 0 °C / -4 °F 5 5 °C / 1 3 1 °F 1 9 .5 V /V▼ 2 3 .5 V /V▲
H yd ro g e n S u lfid e 1 0 .0 to 1 0 0 .0 p p m 5 0 .0 p p m 1 .0p p m 2 5 .0p p m <50 < + /-1 p p m -2 0 °C / -4 °F 5 5 °C / 1 3 1 °F 1 0 .0p p m▲ 2 0 .0p p m▲
C a rb o n M o n o xid e 1 0 0 to 1 ,0 0 0 p p m 300ppm 1 0 0p p m 100ppm <30 < + /-6 p p m -2 0 °C / -4 °F 5 5 °C / 1 3 1 °F 100 p p m▲ 20 0 p p m▲
H yd ro g e n 1 ,0 0 0 p p m o n ly 1 ,0 0 0 p p m n /a 500ppm <65 < + /-2 5 p p m -2 0 °C / -4 °F 5 5 °C / 1 3 1 °F 2 0 0 p p m▲ 4 0 0 p p m▲
Catalytic Bead Sensors 3 0 to 7 0 % o f
F la m m a b le 1 to 8 * 2 0 to 1 0 0 % L E L 100% LEL 1 0 % L E L se le cte d fu ll sca le 50% LEL <25 < + /-1 .5 % L E L -2 0 °C / -4 °F 5 5 °C / 1 3 1 °F 2 0 % L E L▲ 4 0 % L E L▲
Infrared Sensors ra n g e
Certification
China GB Ex d IIC T4 GB3836.1&2 -2000, PA, (CCCF – Pending)
Korea KTL Ex d IIC T6 (-40ºC to +65ºC)
US UL - Class I, Division 1, Groups B, C and D, Class I, Division 2, Groups B, C & D, Class II,
Division, Groups E, F & G, Class II, Division 2, Groups F & G. -40ºC to+65ºC
European ATEX Ex II 2 GD Ex d IIC Gb T6(Ta -40ºC to +65ºC) Ex tb IIIC T85ºC Db IP66
International IEC Ex II 2 GD Ex d IIC Gb T6(Ta -40ºC to +65ºC) Ex tb IIIC T85ºC Db IP66
CE EN50270:2006 EN6100-6-4:2007
Environmental
IP Rating IP66 in accordance with EN60529:1992
Operating -40ºC to +65ºC/ -40ºF to +149ºF, (IR: -20ºC to +50ºC/ -4ºF to +122ºF).
Note: The detector display may become illegible at temperatures below -40°C, but the detector continues its gas monitoring
Temperature function. The display is not damaged and recovers when the temperature rises back above -20 °C.
Operating
Continuous 20-90%RH (non condensing), Intermittent 10-99%RH (non condensing)
Humidity
Operating
90-110kPa
Pressure
Storage
-25°C to +65°C (-13°F to +149°F)
Conditions
45
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
15 Ordering information
Sensepoint XCD transmitter and sensor KIT (ATEX/IECEx/AP*, ADC12 and M20 Entry)
SPXCDAAMFX ATEX/IECEx/AP* approved Methane CAT 0-100%LEL (20 to 100%LEL, 10%LEL) with ADC12, M20 Entry
SPXCDAAMRX ATEX/IECEx/AP* approved Methane IR 0-100%LEL (20 to 100%LEL, 10%LEL) with ADC12, M20 Entry
SPXCDAAMHX ATEX/IECEx/AP* approved Hydrogen Sulfide 0-50ppm (10.0 to 100.0ppm, 1ppm) with ADC12, M20 Entry
SPXCDAAMCX ATEX/IECEx/AP* approved Carbon Monoxide 0-300ppm (100 to 1000ppm, 100ppm) with ADC12, M20 Entry
SPXCDAAMB1 ATEX/IECEx/AP* approved Carbon Dioxide IR 0-2.00%VOL with ADC12, M20 Entry
Sensepoint XCD transmitter and sensor KIT (ATEX/IECEx/AP*, LM25 and M20 Entry)
SPXCDALMFX ATEX/IECEx/AP* approved Methane CAT 0-100%LEL (20 to 100%LEL, 10%LEL) with LM25, M20 Entry
SPXCDALMRX ATEX/IECEx/AP* approved Methane IR 0-100%LEL (20 to 100%LEL, 10%LEL) with LM25, M20 Entry
SPXCDALMHX ATEX/IECEx/AP* approved Hydrogen Sulfide 0-50ppm (10.0 to 100.0ppm, 1ppm) with LM25, M20 Entry
SPXCDALMCX ATEX/IECEx/AP* approved Carbon Monoxide 0-300ppm (100 to 1000ppm, 100ppm) with LM25, M20 Entry
SPXCDALMB1 ATEX/IECEx/AP* approved Carbon Dioxide IR 0-2.00%VOL with LM25, M20 Entry
Sensepoint XCD transmitter and sensor KIT (ATEX/IECEx/AP*, 316SS and M20 Entry)
SPXCDASMFX ATEX/IECEx/AP* approved Methane CAT 0-100%LEL (20 to 100%LEL, 10%LEL) with 316SS, M20 Entry
SPXCDASMRX ATEX/IECEx/AP* approved Methane IR 0-100%LEL (20 to 100%LEL, 10%LEL) with 316SS, M20 Entry
SPXCDASMHX ATEX/IECEx/AP* approved Hydrogen Sulfide 0-50ppm (10.0 to 100.0ppm, 1ppm) with 316SS, M20 Entry
SPXCDASMCX ATEX/IECEx/AP* approved Carbon Monoxide 0-300ppm (100 to 1000ppm, 100ppm) with 316SS, M20 Entry
SPXCDASMB1 ATEX/IECEx/AP* approved Carbon Dioxide IR 0-2.00%VOL with 316SS, M20 Entry
Sensepoint XCD transmitter and sensor KIT (UL, ADC12 and 3/4”NPT Entry)
SPXCDUANFX UL approved Methane CAT 0-100%LEL (20 to 100%LEL, 10%LEL) with ADC12, 3/4"NPT Entry
SPXCDUANRX UL approved Methane IR 0-100%LEL (20 to 100%LEL, 10%LEL) with ADC12, 3/4"NPT Entry
SPXCDUANHX UL approved Hydrogen Sulfide 0-50ppm (10.0 to 100.0ppm, 1ppm) with ADC12, 3/4"NPT Entry
SPXCDUANCX UL approved Carbon Monoxide 0-300ppm (100 to 1000ppm, 100ppm) with ADC12, 3/4"NPT Entry
Sensepoint XCD transmitter and sensor KIT (UL, 316SS and 3/4”NPT Entry)
SPXCDUSNFX UL approved Methane CAT 0-100%LEL (20 to 100%LEL, 10%LEL) with 316SS, 3/4"NPT Entry
SPXCDUSNRX UL approved Methane IR 0-100%LEL (20 to 100%LEL, 10%LEL) with 316SS, 3/4"NPT Entry
46
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
SPXCDUSNHX UL approved Hydrogen Sulfide 0-50ppm (10.0 to 100.0ppm, 1ppm) with 316SS, 3/4"NPT Entry
SPXCDUSNCX UL approved Carbon Monoxide 0-300ppm (100 to 1000ppm, 100ppm) with 316SS, 3/4"NPT Entry
Accessories
SPXCDCC Collecting cone for use with lighter than air gases
SPXCDMTBR Pipe Mounting bracket (inc. fixings and hex wrench set)
Spares
SPXCDMAG Magnet
47
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
16 Warranty statement
All products are designed and manufactured to the latest internationally recognized
standards by Honeywell Analytics under a Quality Management system that is certified to
ISO 9001. As such Honeywell Analytics warrants its products against defective parts and
workmanship and will repair or (at its option) replace any instruments which are or may
become defective under proper use within 12 months from date of commissioning by an
approved Honeywell Analytics representative
or 18 months from date of shipment from Honeywell Analytics, whichever is the sooner.
This warranty does not cover disposable batteries or damage caused by accident, abuse,
abnormal operating conditions or poisoning of sensor.
This warranty covers instrument and parts sold to the Buyer only by authorized distributors,
dealers and representatives as appointed by Honeywell Analytics. The warranties set out
in this clause are not pro rata, i.e. the initial warranty period is not extended by virtue of
any works carried out there under.
In no event will Honeywell Analytics be liable for any incidental damages, consequential
damages, special damages, punitive damages, statutory damages, indirect damages,
loss of profits, loss of revenues, or loss of use, even if informed of the possibility of such
damages. Honeywell Analytic's liability for any claims arising out of or related to this
product will in no case exceed the order value. To the extent permitted by applicable
law, these limitations and exclusions will apply regardless of whether liability arises from
breach of contract, warranty, tort (including but not limited to negligence), by operation of
law, or otherwise.
48
#
& &
! *(!!+!'&
*!1 / / -9EG=75@(=D9 CE=MCBG5@(=D9%CHBG98
2 3 PN
.! + #; %CHBG98
2P N3
+)&*%!++) ?; 2 3
3
%',&+!&)#+ ?;
%+)!$
+)&*%!++) $% **
*&*'))+)! **
2
2 3
!! 1)',*)+!!+!'&
,$ *
2 3
+/ 2 3
!! 2 3
K8!!6++5 OGCO
17 Installation Drawing
!K*
#+$
K8!!+ N -C@GF(A5K. 2 3
+5A6 O+CO .5@@%CHBG98
17.1 Mechanical Installation Drawing
(
!!! !+%$!*+
49
/+)&*%!++)
' +'/! $+)&*%!++)
+.'% CE &(+$&+)0
2 3
/*&*'))+)!
*+!&$***+$*&*')
' +'/! $%%$
()+&,%)(&*'&*+0(
2 3
.+ )()'' ',*!& %C@G
% CE &(+($, 2 3 Q '$
2 3
(!(%',&+!&)#+
2 3
PN
2P N 3
' '
+)&*%!++)'&&+!'&*
Sensepoint XCD Technical Manual
N A
(CJ9E F=;B5@
50
=;=G5@7CAA +
=;=G5@7CAA +
&'+* =;=G5@7CAA +
)9@5L )$0 &
))+'!&*+),+!'&%&,$')+ &!$+!$*
)9@5L )$0 '%
/'()+!&-'$+ N -C@GF.%5K
)9@5L )$0 &'
N A$''()*!*+&+' C<AF )9@5L )$0 &
('.)*,(($0$''()*!*+&+' C<A )9@5L )$0 '%
&'+.')+ ('!&+** ',$'&&+-!*)& )9@5L )$0 &'
& ')'&,!+*'*+'-'!)',&$''(* )9@5L )$0&
%,$+!($)+ ('!&+*)&'+$$'. )9@5L )$0'%
('!&+*'&+!&,!+0'*)&%0 !-,*!& )9@5L )$0&'
(()'()!+$&* $')'&,!+!&+)&$&/+)&$
)+ ('!&+*!&+ +)&*%!++))()'-!
/+)&*%!++))+!!+!'&
#+$K8!!++5A6 OGCO
) &
C & &+*
C
C
C a
' " '0.1' , - $ '# # 2
Q ( $%%
' $%% $-. &! $%% (&
-) $%% &
Q ( !" % *+ %)!
#/
' # $ %
Q ( !"! !"! !"! !"! !"! !"! ) %
SPXCDHMANEN Issue 2
#
& &
! !+%$!*+
,+%',&+!&($+ H7G%CHBG=B; C@9
Q 2 3
17.3 Duct Mounting Drawing
Sensepoint XCD Technical Manual
2 3
Q 2 3
51
Q 2 3
2 3
2 3
Q 2 3
%CHBG=B; C@9
2 3
O
Q
Sensepoint XCD Technical Manual
Q
P
&'+*
9B9E5@8E5:G5B;@9 OHB@9FFCG<9EJ=F9FG5G98
Q
!BG9EB5@)58== %5KHB@9FFCG<9EJ=F9FG5G98
52
CADCB9BG-C@HA9 7A4
+<9DCF=G=CBC::998DC=BGFFD@=G@=B9F5B8
Q 9>97GCEFGC695;E9986L (69:CE9GCC@
DEC8H7G=CB7CAA9B79F%5KD9EA=FF56@9:@5F<
(+*!$!'&CE*+))$*
&+*AHFGBCG69HF988HE=B;A5BH:57GHE9
+<9+GC69HF985FG<9A5FG9E
=B:CEA5G=CB:CEDEC8H7=B;G<=F7CADCB9BG
=A9BF=CBFCBG<=F8E5J=B;5E9:CEE9:9E9B79CB@L
@@FHE:579FGC69:E99:ECAF=B?A5E?F
A57<=B=B;A5E?FF7E5G7<9F5B8CG<9E6@9A=F<9F
) & &CI
((*) $
C &
C
C
C a
' " &"&
Q (
'
'
Q ( !" ( !
' # $ %
Q ( "#$#"% "#$#"% )
SPXCDHMANEN Issue 2
#
& &
17.5 Mounting Bolt Assy Drawing
P
53
Q
Q
) & '7G
**
C &
C
C
C a
' " ,& %!& "%*& "%&
Q (
' - & ( ')+ '() '()
Q ( !" )( *)' *)' !
' # $ %
Q ( "#$#" "#$#" "#$#" "#$#" "#$#" )
SPXCDHMANEN Issue 2
#
"# & &
5#
5
+
'#
30#
6
0
5
Sensepoint XCD Technical Manual
4
5
54
05#
%
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
18 Certification
18.1 China GB Ex and PA
China GB Ex (Chinese Version):
55
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
56
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
China PA Certification:
57
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
58
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
59
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
60
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
61
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
62
THIS DOCUMENT HAS BEEN GENERATED USDING
3001EA041 SHT. ADOBE ILLUSTRATOR AND MUST ONLY BE UPDATED BY A.I.
II 2 GD
Sensepoint XCD Technical Manual
63
certifiying body
2. Serial no. format
S/N: SIIPPPYYWWNNN
Sequence number
Week of MFG
Year of MFG
Part Code
Issue State
SUPPLIER CODE
3. Etching Depth : 0.1 mm
P/N : ______________________________
64
WARNING :To reduce the risk of Hazardous Atmos-
pheres, Disconnect the equipment from supply circuit
before opening. Keep tightly closed when in opera
-tion. Conduit runs must have a seal fitting within
18 inches of the enclosure.
SCALE 2/1
BEN 16 MAY 08
NONE
5 4 3 2 1 B A Sensepoint XCD
27/JAN/09 24/JAN/09 15/JAN/09 06/JAN/09 09/DEC/08 23/JUL/08 16/MAY/08 UL Name Plate
CHANGE REMOVE CHANGE CHANGE CHANGE CHANGE NEW
SENTENCE “CLASS2” SENTENCE SENTENCE REVISION RATING DRAWING
Gas Type
II 2 GD
Honeywell Analytics Inc. Ex d IIC T6(Ta -40 to +65 ) Gb
0344 Lincolnshire, IL, USA 60069 Ex tb IIIC T85 Db IP66 A21
Baseefa08ATEX0316X
Sensepoint XCD IECEx BAS 08.0104X
Disposable Sensor P/N :
WARNING : DO NOT SEPARATE
Rev. S/N : SIIPPPYYWWNNNN WHEN ENERGISED
18.7 Sensor Cartridges Label
Sensepoint XCD Technical Manual
Revision Number
65
N
Gas
Gas Type Applied Concentration Unit Reading Unit
type
%vol (change O2
O2 Carbon Dioxide 5 %vol 0.1
reading) per %vol CO2
Ammonia 50 ppm 0 ppm H2S
Carbon Monoxide 100 ppm <2 ppm H2S
Carbon Dioxide 5000 ppm 0 ppm H2S
Chlorine 0.5 ppm 0 ppm H2S
Ethylene 100 ppm 0 ppm H2S
H 2S
Hydrogen 100 ppm 0 ppm H2S
Hydrogen Sulfide 10 ppm 10 ppm H2S
Nitrogen Monoxide 25 ppm 0 ppm H2S
Nitrogen Dioxide 3 ppm 0 ppm H2S
Sulfur Dioxide 2 ppm 0 ppm H2S
Acetone 1000 ppm 0 ppm CO
Acetylene 40 ppm 80 ppm CO
Ammonia 100 ppm 0 ppm CO
Carbon Monoxide 100 ppm 100 ppm CO
Chlorine 2 ppm 0 ppm CO
Ethanol 2000 ppm 3 ppm CO
CO Ethylene 100 ppm 85 ppm CO
Hydrogen 100 ppm 20 ppm CO
Hydrogen 25 ppm 0 ppm CO
Iso-Propanol 200 ppm 0 ppm CO
Nitrogen Monoxide 50 ppm 8 ppm CO
Nitrogen Dioxide 800 ppm 20 ppm CO
Sulfur Dioxide 50 ppm 0.5 ppm CO
Carbon Monoxide 300 ppm ≤60 ppm H2
Hydrogen Sulfide 15 ppm <3 ppm H2
Sulfur Dioxide 5 ppm 0 ppm H2
Nitrogen Monoxide 35 ppm »10 ppm H2
H2 Nitrogen Dioxide 5 ppm 0 ppm H2
Chlorine 1 ppm 0 ppm H2
Hydrogen Cyanide 10 ppm »3 ppm H2
Hydrogen Chloride 5 ppm 0 ppm H2
Ethylene 100 ppm »80 ppm H2
66
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
However, it is not always practical to obtain every detectable type of hydrocarbon gas in
a calibration-ready, certified and verifiable form. Therefore, it is possible to carry out a
“cross calibration” using another hydrocarbon gas/air mixture.
When the Sensepoint XCD Combustible LEL sensor is to be calibrated with a gas which is
different to the gas or vapour to be detected, the following cross calibration procedure may
be followed:
Caution: Where the user calibrates any sensor using a different gas, responsibility for
identifying and recording calibration rests with the user. Refer to the local regulations where
appropriate.
Notes:
1. Table 11 lists a selection of hydrocarbon compounds and states a reference figure or
“Star Rating” according to the reaction they produce in relation to other hydrocarbons.
2. An eight star (8*) gas produces the highest output, while a one star (1*) gas produces
the lowest output.
67
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
25 Styrene 2*
26 Tetra hydrafuran 4*
27 Toluene 3*
28 Triethylamine 3*
29 Xylene 2*
Note: These settings must only be used with a calibration gas concentration of 50% LEL.
Table 12. Calibration span setting
(3) If a sensor is to be used to detect a gas other than that for which it was calibrated and
there is no intention to use an equivalent calibration gas to re-calibrate the sensor, then
the required correction factor may be obtained from Table 13.
The reading shown on the gas detector controller or transmitter display should be
multiplied by this number in order to obtain a more accurate gas concentration result.
68
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 2
Notes:
1. Since catalytic sensors require oxygen for correct operation, a mixture of gas in air
should always be used for calibration purposes.
2. Assuming average performance of the sensor, the sensitivity information in Tables 11 To
13 is normally accurate to + or - 20%.
Working Example:
If the target gas to be detected is 0-100%LEL Ethylene and the only calibration gas
available to re-calibrate the sensor is Methane (at 50% LEL concentration), the procedure
is as follows:
(1) Look up the star rating for each gas in Table 11:
(2) Then, look up the span settings for a 50% LEL calibration gas in Table 12 by selecting
the row of figures next to the 6* in the “calibration gas” column. Select the figure in the 5*
column of the “gas to be detected” section. The figure is 62.
(3) This means that during re-calibration, the span gas setting on the gas detector
transmitter or controller should be set to 62% LEL to give an accurate measuring scale for
0-100%LEL Ethylene, when using 50% LEL Methane as the calibration gas.
Please contact your local Honeywell Analytics sales or service distributor, or regional office
should further clarification or additional information be required.
69
D aejeon T el : 042)824-9872 Fax : 042)824-9873
E -mail: sdd@sd-d.co.kr
Web site: http//www.sd-d.co.kr
FTS
(DANFOSS)
Data sheet
Performance
Indicative response times
Water 0.2 m/s Air 1 m/s
t0.5 t0.9 t0.5 t0.9
ø8 mm
10 s 35 s 95 s 310 s
Accuracy < ± 0.5% FS (typ.)
< ± 1.0% FS (max.)
Max. load protection tube 100 bar
Electrical specifications
Environmental conditions
Media temperature (max. 120 °C without extension length) -50 – 200 °C
Temperature on electronics 1
)
-40 – 85 °C
Transport temperature range -50 – 85 °C
EMC – Emmision EN 61000-6-3
EMC – Immunity EN 61000-6-2
Sinusoidal 15.9 mm-pp, 5 Hz – 25 Hz –
Vibration stability 4 g, 25 Hz – 2 kHz IEC 60068-2-6
Random 7.5 gms, 5 Hz – 1 kHz IEC 600868-2-34, IEC 60068-2-36
Shock 500 g/1 ms IEC 60068-2-27
Shock resistance
Free fall IEC 60068-2-32
Enclosure (depending on electrical connections) See page 4
1
) Temperature of the electronics depends on the media temperature, extension length, ambient temperature and air velocity.
Mechanical characteristics
Wetted parts W.no. 1.4571 (AISI 316 Ti)
Materials:
Enclosure W.no. 1.4404 (AISI 316 L)
Measuring insert fixed
Net weight
0.1 – 0.15 kg
(Depending on design)
Ordering standard
Sensor
Type MBT 5560
Insertion length
0050 mm 0050 Electrical connection
0080 mm 0080 1 Plug, EN 175301-803, Pg 9
0100 mm 0100 4 Plug, AMP Econoseal, J series, Male, excl. female plug
0120 mm 0120 5 Screened cable, 2 m
0150 mm 0150 6 Plug, IEC 947-5-2, M12 × 1, male, excl. female plug
0200 mm 0200 A Flying leads
0250 mm 0250 9 Other
Preferred versions
Enclosure
IP65 IP67 IP67 IP67 IP67
Materials
Glass filled Glass filled Glass filled Glass filled
PUR
polyamid, PA 6.6 polyamid, PA 6.6 polyamid, PA 6.6 polyamid, PA 6.6
Ø30.5
E
Ø8
8
Ø
E = Extension length = 33 mm
L = Insertion length
H = 9 mm
ProcessMaster
Valid as of software version Further information
FEP300, FEP500
HygienicMaster
• 01/01/00 with HART Additional documentation on ProcessMaster
FEP300, FEP500/HygienicMaster FEH300, FEH500
FEH300, FEH500
• 01/00/00 with PROFIBUS PA or is available for download free of charge at
FOUNDATION fieldbus www.abb.com/flow.
ABB Automation Products GmbH ABB Inc. ABB Engineering (Shanghai) Ltd.
Measurement & Analytics Measurement & Analytics Measurement & Analytics
Dransfelder Str. 2 125 E. County Line Road No. 4528, Kangxin Highway,
37079 Göttingen Warminster Pudong New District
Deutschland PA 18974 Shanghai, 201319,
Tel: 0800 1114411 USA P.R. China
Fax: 0800 1114422 Tel: +1 215 674 6000 Tel: +86(0) 21 6105 6666
Email: vertrieb.messtechnikprodukte@ Fax: +1 215 674 7183 Fax: +86(0) 21 6105 6677
de.abb.com Email: china.instrumentation@cn.abb.com
1 Safety
You must read these instructions carefully prior to installing and commissioning the device.
These instructions are an important part of the product and must be kept for future reference.
These instructions are intended as an overview and do not contain detailed information on all
designs for this product or every possible aspect of installation, operation and maintenance.
For additional information or if specific problems occur that are not discussed in these
instructions, contact the manufacturer.
The content of these instructions is neither part of any previous or existing agreement, promise
or legal relationship nor is it intended to change the same.
This product is built based on state-of-the-art technology and is operationally safe. It has been
tested and left the factory in perfect working order from a safety perspective. The information in
the manual must be observed and followed in order to maintain this state throughout the period
of operation.
Modifications and repairs to the product may only be performed if expressly permitted by these
instructions.
Only by observing all of the safety instructions and all safety/warning symbols in these
instructions can optimum protection of both personnel and the environment, as well as safe and
fault-free operation of the device, be ensured.
Information and symbols directly on the product must be observed. They may not be removed
and must be fully legible at all times.
IMPORTANT (NOTE)
• An additional document with Ex safety instructions is available for measuring systems that
are used in explosion hazardous areas.
• Ex safety information is an integral part of this manual. As a result, it is crucial that the
installation guidelines and connection values it lists are also observed.
Installation, commissioning, and maintenance of the product may only be performed by trained
specialist personnel who have been authorized by the plant operator to do so. The specialist
personnel must have read and understood the manual and comply with its instructions.
Prior to using corrosive and abrasive measurement media, the operator must check the level of
resistance of all parts coming into contact with the wetted parts. ABB Automation Products
GmbH will gladly support you in selecting the materials, but cannot accept any liability in doing
so.
The operators must strictly observe the applicable national regulations with regards to
installation, function tests, repairs, and maintenance of electrical products.
Using the device in a manner that does not fall within the scope of its intended use, disregarding
this instruction, using underqualified personnel, or making unauthorized alterations releases the
manufacturer from liability for any resulting damage. This renders the manufacturer's warranty
null and void.
IMPORTANT (NOTE)
This symbol indicates operator tips, particularly useful information, or important information
about the product or its further uses. It does not indicate a dangerous or damaging situation.
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
3
300
13
4
14
5
15
6
7
16
8
17
9
18
10
19
11
12 G00629
IMPORTANT (NOTE)
Meters with 3A approval are labeled with an additional plate.
IMPORTANT (NOTE)
Meters with 3A approval are labeled with an additional plate.
5
9
6
G00632-01
• Depending on the device, the center of gravity may not be in the center of the equipment.
• The protection plates or protective caps installed on the process connections of devices lined
with PTFE / PFA must not be removed until just before installation; to prevent possible
leakage, make sure that the liner on the flange is not cut or damaged.
During operation with hot fluids, contact with the surface may result in burns.
Aggressive fluids may result in corrosion or abrasion of the parts that come into contact with the
medium. As a result, pressurized fluids may escape prematurely.
Wear to the flange gasket or process connection gaskets (e.g., aseptic threaded pipe
connections, Tri-Clamp, etc.) may enable a pressurized medium to escape.
When using internal flat gaskets, these can become embrittled through CIP/SIP processes.
If pressure shocks exceeding the device's permissible nominal pressure occur continuously
during operation, this can have a detrimental effect on the device's service life.
The device is designed for use exclusively within the stated values on the name plate and within
the technical limit values specified in the data sheets.
The following technical limit values must be observed:
• The permissible operating pressure (PS) in the permissible temperature (TS) may not
exceed the pressure-temperature ratings.
• The maximum operating temperature may not be exceeded.
• The permitted operating temperature may not be exceeded.
• The housing protection system must be observed.
• The flowmeter sensor may not be operated in the vicinity of powerful electromagnetic fields,
e.g., motors, pumps, transformers, etc. A minimum spacing of approx. 1 m (3.28 ft) should
be maintained. For installation on or to steel parts (e.g., steel brackets), a minimum spacing
of approx. 100 mm (3.94 inch) should be maintained (based on IEC801-2 and IECTC77B).
Use the original packaging or suitably secure shipping containers if you need to return the
device for repair or recalibration purposes. Fill out the return form (see the Appendix) and
include this with the device.
According to EC guidelines for hazardous materials, the owner of hazardous waste is
responsible for its disposal or must observe the following regulations for shipping purposes:
All devices delivered to ABB Automation Products GmbH must be free from any hazardous
materials (acids, alkalis, solvents, etc.).
Rinse out and neutralize hazardous materials from all hollow spaces such as between meter
tube and housing. For flowmeters larger than DN 400, the service screw (for draining
condensate fluid) at the lower point of the housing must be opened to dispose of hazardous
substances and to neutralize the coil and electrode chamber. These activities must be
confirmed in writing using the return form.
Please
Please contact Customer Center Service acc. to page 2 for nearest service location.
ABB Automation Products GmbH operates an integrated management system, consisting of:
• Quality management system to ISO 9001
• Environmental management system to ISO 14001
• Occupational health and safety management system to BS OHSAS 18001 and
• Data and information protection management system
1.16 Disposal
This product is manufactured from materials that can be reused by specialist recycling
companies.
1.16.1 Information on WEEE Directive 2012/19/EU (Waste Electrical and Electronic Equipment)
This product is not subject to WEEE Directive 2012/19/EU or relevant national laws
(e.g., ElektroG in Germany).
The product must be disposed of at a specialist recycling facility. Do not use municipal garbage
collection points. According to the WEEE Directive 2012/19/EU, only products used in private
applications may be disposed of at municipal garbage facilities. Proper disposal prevents
negative effects on people and the environment, and supports the reuse of valuable raw
materials.
If it is not possible to dispose of old equipment properly, ABB Service can accept and dispose of
returns for a fee.
G00005
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
ProcessMaster HygienicMaster
FEP300 FEP500 FEH300 FEH500
Measuring accuracy X - X -
0.4 % (optional 0.2 %) of rate
Measuring accuracy - X - X
0,3 % (optional 0.2 %) of rate
Batch functions - X - X
Presetting counter, overrun correction,
external start/stop, batch end contact
Other software functions X X X X
Mass units, editable totalizers,
Two measuring ranges - X - X
Graphic display X X X X
Line recorder function
Diagnostic functions - X - X
Detection of gas bubbles or deposits on
electrodes, conductivity monitoring,
temperature monitoring, finger print, trend
Partially filled X X - -
Recognition through partial filling
electrode (TFE)
Hardware options - X - -
• Ceramic carbide lining
• Wolfram carbide electrodes
• Double layer electrodes
Hardware options - - - X
DN 1 ... 2
Startup functions - X - X
Grounding check
Fieldbus X X X X
PROFIBUS PA, FOUNDATION fieldbus
Verifications / Diagnostic tool X X X X
ScanMaster
For devices with an integral mount design, the transmitter and the sensor form a single
mechanical unit.
The transmitter is available in two housing designs:
• Single-compartment housing:
On the single-compartment housing, the electronics area and the connection area in the
transmitter are not separated from each other.
• Dual-compartment housing:
On the dual-compartment housing, the electronics area and the connection area in the
transmitter are separated from each other.
ProcessMaster
The ProcessMaster sensor is available in two designs, which are distinguished by the design
level.
FEP311 (without explosion protection) FEP315 Zone 2, Div. 2 FEP315
Zone 1, Div. 1
1), 3) 2), 3) 2), 4) 2), 5) 1), 3) 2), 3) 2), 4) 2), 3) 2), 4)
HygienicMaster
FEH311 (without explosion FEH315 Zone 2, Div. 2 FEH315 Zone 1, Div. 1
protection)
1) 2) 1) 2) 2)
For devices with a remote mount design, the transmitter and sensor are mounted in separate locations. The
electrical connection between the transmitter and the sensor is provided by a signal cable.
When the minimum conductivity of the measuring medium is 5 µS/cm, a maximum signal cable length of 50 m
(164 ft) is possible without fitting an additional preamplifier to the sensor. With a pre-amplifier, the maximum
permissible signal cable length is 200 m (656 ft).
The transmitter is available in two housing designs:
• Single-compartment housing:
On the single-compartment housing, the electronics area and the connection area in the transmitter are not
separated from each other.
• Dual-compartment housing:
On the dual-compartment housing, the electronics area and the connection area in the transmitter are separated
from each other.
ProcessMaster
The ProcessMaster sensor is available in two designs, which are distinguished by the design level.
Sensor
FEP321 / FEP521 FEP325 / FEP525 FEP325 / FEP525
(without explosion Zone 2, Div. 2 Zone 1, Div. 1
protection)
1) 2) 1) 1)
Transmitter
FET321 / FET521 FET325 / FET525 FET321 / FET521 FET325 / FET525 FET325 / FET525 FET321 / FET521
(without explosion (Zone 2, Div. 2) (without explosion (Zone 1, Div. 1) (Zone 2, Div. 2) (without explosion
protection) protection) protection)
3) 4) 3) 4) 3) 4) 4) 3) 4) 3) 4)
HygienicMaster
Sensor
FEH321 / FEH521 (without FEH325 / FEH525 (Zone 2, Div. 2)
explosion protection)
Transmitter
FET321 / FET521 (without FET325 / FET525 (Zone 2, Div. 2) FET321 / FET521 (without explosion
explosion protection) protection)
3) 4) 3) 4) 3) 4)
3.1 Inspection
Check the devices for possible damage that may have occurred during transport. Damages in
transit must be recorded on the transport documents. All claims for damages must be claimed
without delay against the shipper and before the installation.
For transport of flanged units smaller than DN 450 use a lifting strap. Wrap the straps around
both process connections when lifting the device. Avoid chains since these may damage the
housing.
Flanged units may not be lifted by the terminal box or at the middle of the housing. Use only the
eye bolts on the device to lift and install it in the pipeline.
4 Mounting
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
4.1 General information on installation
For devices with a wafer-type design, ABB offers an installation set as an accessory that
comprises threaded rods, nuts, washers and centering sleeves for installation.
1 2 3 3 G01348
The meter tube can be installed at any location in the piping while taking the installation
conditions into account.
1. Remove protective plates, if present, to the right and left of the meter tube. To prevent
possible leakage, make sure that the liner on the flange is not cut or damaged.
2. Position the meter tube coplanar and centered between the piping.
3. Install gaskets between the surfaces; see chapter „Torque information“ on page 27.
IMPORTANT (NOTE)
For achieve the best results, ensure the gaskets fit concentrically with the meter tube
4. Use the appropriate screws for the holes in accordance with chapter „Torque information“ on
page 27.
5. Slightly grease the threaded nuts.
6. Tighten the nuts in a crosswise manner as shown in the figure. Observe the tightening
torques in accordance with chapter „Selecting gaskets“ on page 25!
First tighten the nuts to approx. 50 % of the maximum torque, then to 80 %, and finally a
third time to the maximum torque. Do not exceed the max. torque.
1 5
1 3
8 3
4 7
4 2
6 2 G00034
Fig. 11
IMPORTANT (NOTE)
The specified torques are valid only for greased threads and piping that is not subject to tensile stress.
IMPORTANT (NOTE)
If concentric reducers are installed on the device, it must be mounted in a vertical position.
G01319
Fig. 12
1 Bracket 2 Leakage hole
G00657-01 G00039-01
Fig. 13 Fig. 16
max. 45° 3°
G00038
1
G00041
Fig. 17
Fig. 14
4.4.6 Free inlet or outlet
• Do not install the flowmeter at the highest point or in the draining-
4.4.3 In- and outlet pipe sections off side of the pipeline, flowmeter runs empty, air bubbles can
form (1).
The metering principle is independent of the flow profile as long as • Provide for a siphon fluid intake for free inlets or outlets so that
standing eddies do not extend into the metering section, such as may the pipeline is always full (2).
occur after double elbows (1), in the event of tangential inflow, or
where half-open gate valves are located upstream of the flowmeter
sensor. 2
In such cases, measures must be put in place to normalize the flow
profile.
1
• Do not install fittings, manifolds, valves, etc., directly in front of the
flowmeter sensor (1).
• Butterfly valves must be installed so that the valve plate does not G00040
1 2
Fig. 15 G00042
Fig. 19
4.4.8 Installation in the vicinity of pumps 4.4.12 Installation in pipelines with larger nominal
diameters
For flowmeter primaries which are to be installed in the vicinity of
pumps or other vibration generating equipment, the utilization of Determine the resulting pressure loss when using reduction
mechanical snubbers is advantageous. pieces (1):
1. Calculate the diameter ratio d/D.
2. Determine the flow velocity based on the flow range nomograph
(Fig. 24).
3. Read the pressure drop on the Y-axis in Fig. 24.
G00561
Fig. 20
G00654
Fig. 21
1 Insulation
xx m (xx ft)
G01349
Fig. 24
Fig. 22
Change from one to two columns
Depending on the mounting position, the LCD display or transmitter housing can be rotated to
enable horizontal readings.
G00659-01
1. Loosen screws (3) and rotate housing 90° to the left or right.
2. Retighten screws (3).
4.6 Ground
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
IMPORTANT (NOTE)
If the flowmeter sensor is installed in plastic or earthenware pipelines, or in pipelines with an
insulating lining, transient current may flow through the grounding electrode in special cases.
In the long term, this may destroy the sensor, since the ground electrode will in turn degrade
electrochemically. In these special cases, the connection to the ground must be performed
using grounding plates. Install a grounding plate upstream and downstream of the device in
this case.
Use a copper wire (at least 2.5 mm² (14 AWG)) to establish the ground connection between the
sensor (1), the pipeline flanges and an appropriate grounding point.
IMPORTANT (NOTE)
• Earthing is illustrated using the example of the dual-compartment transmitter housing; in
the case of transmitters with single-compartment housing, earthing is to be performed as
shown
1. Solder the threaded nuts M6 (1) to the pipeline and connect the ground as shown in the
illustration.
2. Use a copper wire (at least 2.5 mm² (14 AWG)) to establish the ground connection between
the sensor (2) and an appropriate grounding point.
IMPORTANT (NOTE)
• Earthing is illustrated using the example of the dual-compartment transmitter housing; in
the case of transmitters with single-compartment housing, earthing is to be performed as
shown
For plastic pipes or pipes with insulating lining, the earthing for the measuring medium is
provided by the grounding plate as shown in the figure below or via grounding electrodes that
must be installed in the device (option). If grounding electrodes are used, the grounding plate is
not necessary.
1. Install the flowmeter sensor with grounding plate (3) in the pipeline.
2. Connect the terminal lug (2) for the grounding plate (3) and ground connection (1) on the
flowmeter sensor with the grounding strap.
3. Use a copper wire (min. 2.5 mm² (14 AWG)) to link the earthing terminal (1) to a suitable
earthing point.
IMPORTANT (NOTE)
• Earthing is illustrated using the example of the dual-compartment transmitter housing; in
the case of transmitters with single-compartment housing, earthing is to be performed as
shown
Ground the stainless steel model as shown in the figure. The measuring fluid is grounded via
the adapter (1) and an additional ground is not required.
Fig. 29
For devices with a meter size between DN 10 … 250, grounding plates made of conductive
PTFE are available. These are installed in a similar way to conventional grounding plates.
For devices with extended diagnostic functions different grounding conditions may be valid.
For further information read and observe chapter 9 “Extended diagnostic functions”.
Install grounding plates on each side of the flowmeter sensor. Insulate the grounding plates
from the pipe flanges and connect them to the flowmeter sensor and to functional ground.
Insulate the screw bolts for the flange connections when mounting. The insulation plates and
the insulation pipe are not included in the delivery. They must be provided onsite by the
customer.
The CCP potential must be diverted through a connecting line "A" away from the insulated
flowmeter sensor.
Fig. 31: Flowmeter sensor with grounding plate and functional ground
1 Insulated pipe A Connecting line for CCP potential
2 Functional ground ≥ 4 mm2 Cu, not included in the delivery,
3 Grounding plate to be provided onsite
B Insulated screw bolts without grounding
plates
4.6.9.2 Mixed system pipeline with CCP and functional ground potentials
This mixed system has an insulated pipeline with CCP potential and an uninsulated bar metal
pipe (L = 1/4 x flowmeter sensor size) with functional ground potential upstream and
downstream of the flowmeter sensor.
The Fig. 32 shows the preferred installation for cathodic corrosion protection systems.
5 Electrical connections
Fig. 33
IMPORTANT (NOTE)
Use wire end sleeves.
• Wire end sleeves 0.75 mm2 (AWG 19), for shielding (1S, 2S)
• Wire end sleeves 0.5 mm2 (AWG 20), for all other wires
The shields may not touch (signal short circuit).
IMPORTANT (NOTE)
Use wire end sleeves.
• Wire end sleeves 0.75 mm2 (AWG 19), for shielding (1S, 2S)
• Wire end sleeves 0.5 mm2 (AWG 20), for all other wires
The shields may not touch (signal short circuit).
"X"
L2
1
2
3 8 (0.31)
4
5 25 (0.98)
6
7
8
9
X 10 11 12 13
G01030-01
Transmitter side
Fig. 36: Flowmeter sensor side, dimensions in mm (inch)
1 Measurement potential 3, green L2 = 70 (2.76) 8 Magnet coil, M2, red L2 = 90 (3.54)
2 Signal line E1, violet L2 = 60 (2.36) 9 Magnet coil, M1, brown L2 = 90 (3.54)
3 Shield 1S L2 = 60 (2.36) 10Foil shield (D1, D2)
4 Shield 2S L2 = 60 (2.36) 11Foil shield continuity wire (D1, D2)
5 Signal line, E2, blue L2 = 60 (2.36) 12Ground wire, steel
6 Data line, D2, yellow L2 = 70 (2.76) 13SE clamp
7 Data line, D1, orange L2 = 70 (2.76)
M1 M2 D1 D2 SE 3 2S E2 E1 1S
D173D027U01
M1 M2 D1 D2 SE 3 2S E2 E1 1S
D173D031U01
G01323
IMPORTANT (NOTE)
• Use wire end sleeves.
- Wire end sleeves 0.75 mm2 (AWG 19), for shielding (1S, 2S)
- Wire end sleeves 0.5 mm2 (AWG 20), for all other wires
• The shields may not touch (signal short circuit).
en
Prepare the cable end on the transmitter side as shown in Fig. 37.
• Twist the wire mesh shield of the cable and connect to the ground terminal.
• Connect the ground wire of the cable to the SE clamp of the terminal strip.
• Connect all other wires as shown in Fig. 37.
• Connect the cable ground wire together with the foil shield continuity wire from D1, D2 to the
SE clamp of the terminal strip.
• When using the flowmeter sensor in systems with cathodic corrosion protection (CCP),
connect the cable ground wire together with the foil shield continuity wire from D1, D2 to the
SE clamp of the terminal strip.
• Connect all other wires as shown in Fig. 37.
Connections can only be made with the power supply switched off.
The device must be earthed according to instructions. The sensor is connected to the
transmitter via the signal / magnetic coil cable (part no. D173D027U01 or D173D031U01).
1S E1 E2 2S 3 D2 D1 M2 M1 1S E1 E2 2S 3 D2 D1 M2 M1
1S E1 E2 2S 3 D2 D1 M2 M1
5
1 1
6
2 2
D173D027U01 D173D031U01
G00671-02
Fig. 38
1 Earth wire. 5 Foil shield D1, D2 (D173D031U01 only).
2 Earth clamp. 6 Continuity wire of the foil shield (D1, D2)
4 Braided shield (D173D027U01 only). (D173D031U01 only).
IMPORTANT (NOTE)
The cable with the part number D173D027U01 can be used for all device designs.
The cable with the part number D173D031U01 can be used for all device designs.
• Sensor without explosion protection from nominal diameter DN 15 (models FEP321,
FEH321, FEP521, FEH521)
• Sensor for use in Zone 2, Div. 2 from nominal diameter DN 15 (models FEP325, FEH325,
FEP525, FEH525)
IMPORTANT (NOTE)
Use wire end sleeves.
• Wire end ferrules 0.75 mm2 (19 AWG), for shielding (1S, 2S)
• Wire end ferrules 0.5 mm2 (20 AWG), for all other wires
The shielding may not touch (signal short circuit).
G01312
An installation set for sealing the cable conduit is available via order number
3KXF081300L0001.
For sensors with IP rating IP 68, the maximum flooding height is 5 m (16.4 ft). The supplied
cable (part no. D173D027U01 or D173D031U01) fulfills all submersion requirements.
G00171
Fig. 40
1 Max. flooding height 5 m (16.4 ft)
The sensor is type-tested in accordance with EN 60529. Test conditions: 14 days at a flooding
height of 5 m (16.4 ft).
5.4.3.1 Connection
1. Use the supplied cable to connect the sensor and the transmitter.
2. Connect the cable in the terminal box of the sensor.
3. Route the cable from the terminal box to above the maximum flooding height of 5 m (16.4 ft).
4. Tighten the cable gland.
5. Carefully seal the terminal box. Make sure the gasket for the cover is seated properly.
IMPORTANT (NOTE)
As an option, the sensor can be ordered with the signal cable already connected to the sensor
and the terminal box already potted.
Preparation
• Complete the installation before potting in order to avoid moisture penetration. Before
starting, check all the connections for correct fitting and stability
• Do not overfill the terminal box. Keep the potting compound away from the O-ring and the
gasket / groove (see fig. Fig. 41)
• Prevent the two-component potting compound from penetrating the cable conduit if an NPT
1/2" installation is used
Procedure
1. Cut open the protective enclosure of the two-component potting compound (see packing).
2. Remove the connection clamp of the potting compound.
3. Knead both components thoroughly until a good mix is reached.
4. Cut open the bag at a corner. Perform work activity within 30 minutes.
5. Carefully fill the terminal box with the two-component potting compound until the connection
cable is covered.
6. Wait a few hours before closing the cover in order to allow the compound to dry, and to
release any possible gas.
7. Ensure that the packaging material and the drying bag are disposed of in an environmentally
sound manner.
1 2 3
5
G00676
Fig. 41
1 Packing bag 4 Max. filling level
2 Connection clamp 5 Drying bag
3 Two-component potting compound
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
The line voltage and power consumption are indicated on the name plate for the transmitter.
A circuit breaker with a maximum rated current of 16 A must be installed in the power supply
line of the transmitter.
The wire cross-sectional area of the power supply cable and the circuit breaker used must
comply with VDE 0100 and must be dimensioned in accordance with the current consumption of
the flowmeter measuring system. The cables must comply with IEC 227 and/or IEC 245.
The circuit breaker should be located near the transmitter and marked as being associated with
the device.
The power supply is connected to terminal L (phase), N (neutral), or 1+, 2-, and PE, as stated
on the name plate.
Connect the transmitter and sensor to functional earth.
IMPORTANT (NOTE)
• Observe the power supply limit values in accordance with the information on the name
plate and from chapter „Power supply“ on page 163
• Observe the voltage drop for large cable lengths and small conductor cross-sections. The
voltage at the terminals of the device may not fall below the minimum value required in
accordance with the information on the name plate and from chapter „Power supply“ on
page 163,
• Complete the electrical connection in accordance with the connection diagram
The terminals for the power supply can be found under the terminal cover (1).
G00678
Fig. 42
1Terminal cover
G01320
Fig. 43
1 Terminals (power supply)
The outer shielding of the signal and magnet coil cable is attached to the busbar via the clip (4)
(from the accessory bag in the connection area) (dual-compartment transmitter housing only).
In the case of the single-compartment transmitter housing, the outer shielding of the signal and
magnet coil cable is connected to the corresponding terminal for the signal and magnet coil
cable.
The shielding for the signal wires functions as a driven shield to transmit the measurement
signal.
The cable is attached to the flowmeter sensor and transmitter according to the connection
diagram.
1 2
6 2
5 4 3 7 3 G00677-01
IMPORTANT (NOTE)
The power supply for the optional pre-amplifier is provided via terminals 1S and 2S.
The transmitter automatically detects the sensor and switches to the required supply voltage
on terminals 1S and 2S.
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
Important (Note)
For detailed information about earthing the transmitter and the sensor, please refer to chapter
4.6 "Ground" on page 35!
A HART
24 V
+ - + - + - + -
31 32 51 52 81 82 41 42 M1 M2 D1 D2 3 2S E2 E1 1S
B M1 M2 D1 D2 3 2S E2 E1 1S
G01350
A = transmitter, B = sensor
Fig. 45
Change from one to two columns
RB[Ω]
G00475-01 450
400
The current / HART output can be operated in "active" or "passive" mode. 350
A Active: 4 ... 20 mA, HART protocol (standard), load: 250 Ω ≤ R ≤ 650 Ω 300
B Passive: 4 ... 20 mA, HART protocol (standard), load: 250 Ω ≤ R ≤ 650 Ω 250
Supply voltage for the current output: minimum 11 V, maximum 30 V. 200
17 18 19 20 21 22 23 24 25 26 27 28 29 30
U2 [V] G00592
For Ex zone 1 / Div. 1 the maximum load is 300 Ω. Max. permissible load (RB) as a function of the source
voltage (U2)
Fig. 46: (I = internal, E = external)
RB [Ω]
Imax = 220 mA
* RB UCE
750
ICE 650
G00476-04 550
The output can be configured as an "active (A)" or "passive (B)" output (in the case 450
350
of the transmitter with the dual-compartment housing, the output is configured 250
using the software; in the case of the transmitter with the single-compartment 150
50
housing, it is configured by means of jumpers on the transmitter backplane). 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
Configuration as "active" output: U2 [V] G00593
82 0V
G00477-01
Fig. 49: (I = internal, E = external)
PA- FF-
98 98
R R
PA+ FF+
C C
97 97
G00248-01
PROFIBUS PA (PA+ / PA-) FOUNDATION fieldbus (FF+ / FF-)
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the event of U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the event of
an error / FDE) an error / FDE)
Bus connection with integrated protection against polarity reversal. Bus connection with integrated protection against polarity reversal
The bus address can be set via the DIP switches in the device (with
dual-compartment transmitter housing only), the transmitter display or
the fieldbus.
The resistance R and condenser C form the bus termination. They must be installed when the device is connected to the end of the entire bus
cable. R = 100 Ω; C = 1 µF
51 51
52 52
-
V +
24V+
24V
42 42
41 41
G00791
Fig. 52: (I = internal, E = external)
5.8.3 PROFIBUS PA - Connection via M12 plug
Only in non-hazardous areas
Pin assignment
(Front view showing pin insert and pins)
PIN 1 = PA+
PIN 2 = nc
PIN 3 = PA-
PIN 4 = shield
4 3
1 2
G01003-01
Fig. 53
IMPORTANT (NOTE)
For additional information on configuring the current output, see chapter 7.2 „Configuring the
current output“
100
H2-Bus A 1 F
G00111
PROFIBUS PA ID no 0x3430
Fig. 54
Alternative 0x9700 or 0x9740
standard ID no
Configuration Directly on the device
Configuration Directly on the device
Software DAT200 Asset Vision Basic
Software DAT200 Asset Vision Basic (+
(+ HART-DTM)
PROFIBUS PA-DTM)
Transmission FSK modulation on current output
Transmission signal Acc. to IEC 61158-2
4 ... 20 mA acc. to Bell 202 standard
Cable Shielded, twisted cable (acc. to IEC
Max. signal amplitude 1.2 mAss
61158-2, types A or B are preferred)
Current output load Min. 250 Ω, max. = 560 Ω
Cable AWG 24 twisted
Max. cable length 1500 m 6.2.1 Bus topology
Baud rate 1,200 baud
• Tree and/or line structure
Display Log. 1: 1,200 Hz
• Bus termination: passive at both ends of the main bus line (RC
Log. 0: 2,200 Hz element R = 100 Ω, C = 1 µF)
Interoperability test ITK 5.20 The files can be downloaded from www.abb.com/flow.
campaign no. The files required for operation can also be downloaded from
Manufacturer ID 0x000320 http://www.fieldbus.org.
Device ID 0x0124
Configuration • Directly on the device
• Via services integrated in the
system
• National configurator
Transmission signal Acc. to IEC 61158-2
7 Commissioning
IMPORTANT (NOTE)
An additional document with Ex safety instructions is available for measuring systems that are
used in explosion hazardous areas. As a result, it is crucial that the specifications and data it
lists are also observed.
For devices without explosion protection or for operation in Zone 2 / Div. 2 the following
is valid:
The signal can be configured in "active" or "passive" mode. The current setting is contained in
the order confirmation.
„A“ 1 „B“
2 3
3
G00679-02
Fig. 57
A Integral mount design 4 Backplane (in the transmitter housing)
B Remote mount design 5 Jumper (BR901) for active / passive current output
1 Housing cover 6 Jumper (BR903) for integral / remote mount design
2 Transmitter plug-in module 7 Jumper (BR902) for hardware write protection
3 Fixing screws
IMPORTANT (NOTE)
The backplane is mounted in the transmitter housing (not the transmitter plug-in module).
7 8
„A“ 1 „B“
5
2 3
4
3
G01321
For units with PROFIBUS PA, the bus address must be checked or configured prior to start-up.
If no bus address information was supplied by the customer, the unit was shipped with its BUS
address set to “126”.
The address must be set during start-up to a number within the valid range (0 … 125).
IMPORTANT (NOTE)
The address selected may only appear once in the segment.
The PROFIBUS PA interface of the device conforms with Profile 3.01 (fieldbus standard
PROFIBUS, EN 50170, alias DIN 19245 [PRO91]).
The transmitter transmission signal is designed according to IEC 61158-2.
IMPORTANT (NOTE)
The manufacturer-specific PROFIBUS PA ID no. is: 0x3430.
The unit can also be operated with the PROFIBUS standard ID nos. 0x9700 or 0x9740.
1 2 3 4 5 6 7 8
G01033
Address setting
Switches 1, 5, 7 = ON means: 1+16+64 = 81 Bus address 81
Switch 1 2 3 4 5 6 7 8
Status Device address Address mode
Off 0 0 0 0 0 0 0 Bus
On 1 2 4 8 16 32 64 Local
7 8
„A“ 1 „B“
5
2 3
4
3
G01321
Fig. 60:
A Integral mount design 5 Jumpers (BR905, BR906) for communication
B Remote mount design 6 Jumper (BR901) for active / passive current output
1 Housing cover 7 Jumper (BR904) for active / passive pulse output
2 Transmitter plug-in module 8 Jumper (BR903) for integral / remote mount design
3 Fixing screws 9 Jumper (BR902) for hardware write protection
4 Backplane (in the transmitter housing)
IMPORTANT (NOTE)
Devices are interchanged using the parameter "ID number selector", which can only be
modified on an acyclical basis.
The manufacturer-specific GSD file "ABB_3430.gsd" is available to download from the ABB
homepage http://www.abb.com/flow.
The standard "PA1397xx.gsd" GSD files are available for download from the Profibus
International homepage: http://www.profibus.com
For devices with a FOUNDATION Fieldbus, the settings of the DIP switch must be checked
prior to start-up.
The DIP switches on the unit must be set correctly as follows:
• DIP switch 1 must be OFF.
• DIP switch 2 must be OFF.
Otherwise, the hardware write protection and the process control system prevent the unit from
recording information.
When integrating the unit in a process control system, a DD file (device description) and a CFF
file (common file format) are required. The DD file contains the device description. The CFF file
is required for segment engineering. Engineering can be performed online or offline.
The DD and CFF files are available to download from the ABB homepage
http://www.abb.com/flow.
The FOUNDATION Fieldbus interface for the device is compliant with the standards FF-890/891
and FF-902/90. The transmission signal of the transmitter is designed in accordance with IEC
61158-2.
The device is registered with the FOUNDATION fieldbus.
Registration for the FOUNDATION fieldbus is recorded under Manufacturer ID 0x000320 and
Device ID 0x0124.
1 1
On
1 2 3 4 5 6 7 8
G01033
DIP switch 2:
Hardware write protection for write access via bus (locks all blocks).
DIP switch 1 2
Status Simulation Mode Write Protect
Off Disabled Disabled
On Enabled Enabled
7 8
„A“ 1 „B“
5
2 3
4
3
G01321
Fig. 62:
A Integral mount design 5 Jumpers (BR905, BR906) for communication
B Remote mount design 6 Jumper (BR901) for active / passive current output
1 Housing cover 7 Jumper (BR904) for active / passive pulse output
2 Transmitter plug-in module 8 Jumper (BR903) for integral / remote mount design
3 Fixing screws 9 Jumper (BR902) for hardware write protection
4 Backplane (in the transmitter housing)
The bus address is automatically allocated at the FF via LAS (link active scheduler). For
address detection, a unique number is used (DEVICE_ID). This number is a combination of
manufacturer ID, device ID and device serial number.
The behavior when switching on the unit corresponds to Draft DIN IEC/65C/155/CDV of June
1996.
The mean current consumption of the device is 10 mA.
The voltage on the bus line must lie in the range of 9 … 32 V DC.
IMPORTANT (NOTE)
The upper limit of the current is electronically limited. In the event of an error, the FDE (= Fault
Disconnection Electronic) function integrated in the device ensures that the current
consumption cannot exceed a maximum of 13 mA.
1. Switch on the power supply. After switching on the power supply, the following messages
appear in succession on the LCD display:
Transmitter Sensor
3. The flowmeter is ready for operation and will operate with factory settings or settings
requested by the customer. To change the factory settings, refer to chapter 8
"Parameterization".
1) The SensorMemory is a data memory integrated in the flowmeter sensor.
IMPORTANT (NOTE)
System data must only be loaded during initial start-up. If the power supply is later switched
off, the transmitter automatically loads all data the next time the power supply is switched on
again.
A selection as described below (1-3) is not required.
IMPORTANT (NOTE)
When commissioning the device, make sure that the transmitter is assigned to the sensor
correctly. It is not possible to operate a flowmeter sensor of the 300 series with a transmitter of
the 500 series.
If the transmitter is operated with a flowmeter sensor of another series, the following message
appears on the transmitter display:
Sensor
Enter password
********** 7. Use to confirm the password. A password is not
RSTUVWXYZ012345 available as factory default; you can continue without
entering a password.
Next OK
Exit Select
Easy Setup
Language 10.Use to call up the edit mode.
German 11. Use or to select the desired language.
12. Confirm the selection with .
Next Edit
Easy Setup
Q (Flowrate) Unit 13.Use to call up the edit mode.
l/s 14.Use or to select the desired unit.
Next Edit
15. Confirm the selection with .
Easy Setup
Qmax 16.Use to call up the edit mode.
25.000 l/s 17. Use or to set the desired upper range value.
Next Edit
18.Use to confirm your setting.
Easy Setup
Totalizer/Pulse Unit 19.Use to call up the edit mode.
m³ 20.Use or to select the desired unit.
Next Edit
21. Confirm the selection with .
Easy Setup
Operation 22.Use to call up the edit mode.
Pulse Mode 23. Use or to select the required operating mode.
• "Pulse Mode": In pulse mode, pulses per unit are
Next Edit output. The relevant settings are provided in the next
menu
• "Fullscale Frequency": In frequency mode, a frequency
proportional to the flow rate is output. The maximum
frequency can be configured according to the flow
measuring range
The factory default for the operating mode is "Pulse Mode".
24. Confirm the selection with .
Easy Setup
Pulses per Unit 25.Use to call up the edit mode.
10.000 / m³ 26.Use or to set the desired value.
Next Edit
27.Use to confirm your setting.
Easy Setup
Pulse Width 28.Use to call up the edit mode.
30.00 ms 29. Use or to set the required pulse width.
Next Edit
30.Use to confirm your setting.
Easy Setup
Damping 31.Use to call up the edit mode.
30.00 ms 32.Use or to set the desired damping.
Next Edit
33.Use to confirm your setting.
Easy Setup
Iout at Alarm 34.Use to call up the edit mode.
High Alarm 35.Use or to select the alarm mode.
Next Edit
36. Confirm the selection with .
Easy Setup
Low Alarm Value 37.Use to call up the edit mode.
3.5000 mA 38. Use or to set the required current for Low
Alarm.
Next Edit
39. Confirm the selection with .
Easy Setup
High Alarm Value 40.Use to call up the edit mode.
21.800 mA 41. Use or to set the required current for High
Alarm.
Next Edit
42. Confirm the selection with .
Easy Setup
System Zero 43.Use to start automatic balancing of the zero point
for the system.
███████
Next OK
IMPORTANT (NOTE)
Prior to starting the zero point balancing, make sure
that:
• There is no flow through the sensor (close all
valves, shut-off devices etc.)
• The sensor is completely filled with the medium to
be measured
Easy Setup Enter the signal cable length between the transmitter and
Cable length the sensor. For devices with an integral mount design
5.0000 m 0.01 m must be entered.
Menu Once all parameter have been set, the main menu appears
Easy Setup again. The most important parameters are now set.
47. Use to switch to the process display.
Exit Select
IMPORTANT (NOTE)
• For additional information regarding operation of the LCD display, refer to chapter "8.1
„Operation".
• For detailed descriptions of all menus and parameters, see chapter 8.4 „Description of
parameters“.
The LCD display is provided with capacitive control buttons. These enable you to control the
device through the glass of the closed cover.
IMPORTANT (NOTE)
The transmitter automatically calibrates the capacitive control buttons on a regular basis. If the
cover should be opened during operation, the buttons' sensitivity is at first increased. As a
result, operating errors may occur. The button sensitivity will return to normal during the next
automatic calibration.
Information regarding factory settings for further parameters (unless customer has
requested a specific parameterization)
Possible parameter settings Factory setting
Qmax Depending on the size (see QmaxDN (see table)
table)
Sensor TAG Alphanumeric, max. None
20 characters
Sensor Location TAG Alphanumeric, max. None
20 characters
Q (Flowrate) Unit l/s; l/min; l/h; ml/s; ml/min; l/min
m3/s; m3/min; m3/h; m3/d;
hl/h; g/s; g/min; g/h; kg/s;
kg/min; kg/h; kg/d; t/min; t/h;
t/d
Totalizer/Pulse Unit m3; l; ml; hl; g; kg; t l
Pulses per Unit 1
Pulse Width 0,1 ... 2,000 ms 100 ms
Damping ( 1 Tau) 0,02 ... 60 sec. 1
DO1 Alarm Config Pulse F/Pulse R, Pulse F, Pulse F/Pulse R
General Alarm, Min. Flowrate
Alarm, Max. Flowrate Alarm,
Empty Pipe, TFE
Only available for
FEP500 / FEH500 are:
Gas Bubble, Conductivity,
Coating, Sensor Temp
DO1 Action Active, Passive Passive
DO2 Alarm Config F/R Signal, Pulse R, General F/R Signal
Alarm, Min. Flowrate Alarm,
Max. Flowrate Alarm, Empty
Pipe, TFE
Only available for
FEP500 / FEH500 are:
Gas Bubble, Conductivity,
Coating, Sensor Temp
Nominal diameter Min. flow range end value QmaxDN Max. flow range end value
DN " 0.02 x QmaxDN (≈ 0.2 m/s) 0 … ≈ 10 m/s 2 x QmaxDN (≈ 20 m/s)
1 1/25 0,012 l/min (0,0032 US gal/min) 0,6 l/min (0,16 US gal/min) 1,2 l/min (0,32 US gal/min)
1,5 1/16 0,024 l/min (0,0063 US gal/min) 1,2 l/min (0,32 US gal/min) 2,4 l/min (0,63 US gal/min)
2 1/12 0,04 l/min (0,0106 US gal/min) 2 l/min (0,53 US gal/min) 4 l/min (1,06 US gal/min)
3 1/10 0.08 l/min (0.02 US gal/min) 4 l/min (1.06 US gal/min) 8 l/min (2.11 US gal/min)
4 5/32 0.16 l/min (0.04 US gal/min) 8 l/min (2.11 US gal/min) 16 l/min (4.23 US gal/min)
6 1/4 0.4 l/min (0.11 US gal/min) 20 l/min (5.28 US gal/min) 40 l/min (10.57 US gal/min)
8 5/16 0.6 l/min (0.16 US gal/min) 30 l/min (7.93 US gal/min) 60 l/min (15.85 US gal/min)
10 3/8 0.9 l/min (0.24 US gal/min) 45 l/min (11.9 US gal/min) 90 l/min (23.78 US gal/min)
15 1/2 2 l/min (0.53 US gal/min) 100 l/min (26.4 US gal/min) 200 l/min (52.8 US gal/min)
20 3/4 3 l/min (0.79 US gal/min) 150 l/min (39.6 US gal/min) 300 l/min (79.3 US gal/min)
25 1 4 l/min (1.06 US gal/min) 200 l/min (52.8 US gal/min) 400 l/min (106 US gal/min)
32 1 1/4 8 l/min (2.11 US gal/min) 400 l/min (106 US gal/min) 800 l/min (211 US gal/min)
40 1 1/2 12 l/min (3.17 US gal/min) 600 l/min (159 US gal/min) 1200 l/min (317 US gal/min)
50 2 1.2 m3/h (5.28 US gal/min) 60 m3/h (264 US gal/min) 120 m3/h (528 US gal/min)
65 2 1/2 2.4 m3/h (10.57 US gal/min) 120 m3/h (528 US gal/min) 240 m3/h (1057 US gal/min)
80 3 3.6 m3/h (15.9 US gal/min) 180 m3/h (793 US gal/min) 360 m3/h (1585 US gal/min)
100 4 4.8 m3/h (21.1 US gal/min) 240 m3/h (1057 US gal/min) 480 m3/h (2113 US gal/min)
125 5 8.4 m3/h (37 US gal/min) 420 m3/h (1849 US gal/min) 840 m3/h (3698 US gal/min)
150 6 12 m3/h (52.8 US gal/min) 600 m3/h (2642 US gal/min) 1200 m3/h (5283 US gal/min)
200 8 21.6 m3/h (95.1 US gal/min) 1080 m3/h (4755 US gal/min) 2160 m3/h (9510 US gal/min)
250 10 36 m3/h (159 US gal/min) 1800 m3/h (7925 US gal/min) 3600 m3/h (15850 US gal/min)
300 12 48 m3/h (211 US gal/min) 2400 m3/h (10567 US gal/min) 4800 m3/h (21134 US gal/min)
350 14 66 m3/h (291 US gal/min) 3300 m3/h (14529 US gal/min) 6600 m3/h (29059 US gal/min)
400 16 90 m3/h (396 US gal/min) 4500 m3/h (19813 US gal/min) 9000 m3/h (39626 US gal/min)
450 18 120 m3/h (528 US gal/min) 6000 m3/h (26417 US gal/min) 12000 m3/h (52834 US gal/min)
500 20 132 m3/h (581 US gal/min) 6600 m3/h (29059 US gal/min) 13200 m3/h (58117 US gal/min)
600 24 192 m3/h (845 US gal/min) 9600 m3/h (42268 US gal/min) 19200 m3/h (84535 US gal/min)
700 28 264 m3/h (1162 US gal/min) 13200 m3/h (58118 US gal/min) 26400 m3/h (116236 US gal/min)
760 30 312 m3/h (1374 US gal/min) 15600 m3/h (68685 US gal/min) 31200 m3/h (137369 US gal/min)
800 32 360 m3/h (1585 US gal/min) 18000 m3/h (79252 US gal/min) 36000 m3/h (158503 US gal/min)
900 36 480 m3/h (2113 US gal/min) 24000 m3/h (105669 US gal/min) 48000 m3/h (211337 US gal/min)
1000 40 540 m3/h (2378 US gal/min) 27000 m3/h (118877 US gal/min) 54000 m3/h (237754 US gal/min)
1050 42 616 m3/h (2712 US gal/min) 30800 m3/h (135608 US gal/min) 61600 m3/h (271217 US gal/min)
1100 44 660 m3/h (3038 US gal/min) 33000 m3/h (151899 US gal/min) 66000 m3/h (290589 US gal/min)
1200 48 840 m3/h (3698 US gal/min) 42000 m3/h (184920 US gal/min) 84000 m3/h (369841 US gal/min)
1400 54 1080 m3/h (4755 US gal/min) 54000 m3/h (237755 US gal/min) 108000 m3/h (475510 US gal/min)
1500 60 1260 m3/h (5548 US gal/min) 63000 m3/h (277381 US gal/min) 126000 m3/h (554761 US gal/min)
1600 66 1440 m3/h (6340 US gal/min) 72000 m3/h (317006 US gal/min) 144000 m3/h (634013 US gal/min)
1800 72 1800 m3/h (7925 US gal/min) 90000 m3/h (396258 US gal/min) 180000 m3/h (792516 US gal/min)
2000 80 2280 m3/h (10039 US gal/min) 114000 m3/h (501927 US gal/min) 228000 m3/h (1003853 US gal/min)
8 Parameterization
8.1 Operation
The LCD display is provided with capacitive control buttons. These enable you to control the
device through the glass of the closed cover.
IMPORTANT (NOTE)
The transmitter automatically calibrates the capacitive control buttons on a regular basis. If the
cover should be opened during operation, the buttons' sensitivity is at first increased. As a
result, operating errors may occur. The button sensitivity will return to normal during the next
automatic calibration.
2 -----------------Menu---------------- 3
4
1
5 Exit Select 5
G00822-01
You can use the or control buttons to browse through the menu or select a number or
character within a parameter value.
Different functions can be assigned to the and control buttons. The function that is
currently assigned to them (5) is shown on the display.
Meaning
Meaning
Process display
Note
For a detailed description of the individual parameters and menus on the configuration level
refer to the sections 8.3 "Overview of parameters on the configuration level" and 8.4
"Description of parameters".
3 4 3 G00756
The process display appears on the LC display when the device is switched on. It shows
information about the device and current process values.
The way in which the current process values (2) are shown can be adjusted on the configuration
level.
8.2.1.1 Description of symbols
Symbol Description
Call up information level.
When Autoscroll mode is enabled, a symbol appears here and the operator
pages are automatically displayed one after the other.
Call up configuration level.
On the information level, the operator menu can be used to display diagnostic information and
choose which operator pages to display.
Process display
1. Use to switch to the information level.
Menu Description
Process display
Electronics
The error messages are divided into four groups in accordance with the NAMUR classification
scheme:
Symbol Description
Error / Failure
Functional check
Out of specification
Maintenance required
Additionally, the error messages are divided into the following areas:
Area Description
Electronics Error / alarm of the electronics.
Sensor Error / alarm of the flowmeter sensor.
Status Alarm due to the current device status.
Operation Error / alarm due to the current operating conditions.
Electronics
Back Select
Electronics The first line shows the area in which the error has
F123,321 occurred.
Brief description
The second line shows the unique error number.
Information
Back Exit The next lines show a brief description of the error and
information on how to remedy it.
Note
For a detailed description of the errors and information on their remedy refer to Chapter 10
"Error messages".
The device parameters can be displayed and changed on the configuration level.
Process display
1. Use to switch to the information level.
Next OK
IMPORTANT (NOTE)
There are four access levels: For the "Standard" and "Advanced" levels you can define
passwords. There are no factory default passwords.
• On the „Read Only“ level all entries are disabled. Parameters are read only and cannot be
modified.
• On the "Standard" level you can edit all parameters described in Chapter 8.4 "Description
of parameters" except for those written in italics.
• On the "Advanced" level all parameters can be modified.
• The Service menu is reserved to the customer service.
Once you have logged on to the corresponding access level, you can edit or reset the
password. Resetting to the "No password defined" state is done by selecting "" as the
password.
„A“ 1 „B“
„A“ 1 „B“
G01322
-----Parameter name----
1 2 .34 56 [unit] 3. Use to select the decimal position to be changed.
4. Use or to set the required value.
-----Parameter name----
ABC.......................... 3. Use to select the position to be changed.
4. Use or to select the required character.
ABCDEFGHIJKLMNOPQ
Next OK 5. Use to select the next position.
6. If necessary, select and set other decimal positions
using the same procedure as described in steps 3 and 4.
7. Use to confirm your setting.
This concludes the procedure for changing a parameter
value.
1 By pressing (Next) repeatedly you can move the cursor to the right. Once the cursor
reaches the end position, "Cancel" is displayed in the lower right.
2 With you can terminate editing and exit the menu item. With you can return to the
start.
IMPORTANT (NOTE)
The LCD display automatically returns to the process display three minutes after the last
button has been actuated.
Rate ADC
Analog Reset Noise Reset Max
Noise-Reset ON
Sensor QmaxDN
Qmax
Qmax2
Range Mode
Sensor Location TAG
Sensor TAG
Display 4 Language
Contrast
Operator Pages Operator Page 1 Display Mode
1st Line
Autoscroll Operator Page 2 2nd Line
Flowrate Format 3st Line
Volume Format Operator Page 3 Bargraph
Date/Time Format
Display Test
Operator Page 4
Input / Output 5 Digital Out Mode DO1 / DO2 Function Pulse F/Pulse R,
Pulse F/Logic,
Pulse VR/Logic,
Logic/Logic
FF
Show FF Address
AI1-Q Flowrate
INT1-Q Flowrate
AI2-Internal Tot Fwd
AI3-Internal Tot Rev
AI4-Diagnostics
AO-Density Adjust
DI-Alarm Info
DO-Cyclic Control
Sensor Measurements
Sensor Measurements
One shot Sensor Mea
Coil Current
Coil Resistor
Coil Voltage
Coil R. Max Alarm
Coil R. Min Alarm
Signal-Cable length
Sensor Temp Value
Adj. Sensor T. Value
Sensor T. Min Alarm
Sensor T. Max Alarm
Cond. Detection
Cond. Detection
One shot Cond.
Cond. Value
Adj. Cond. Value
Cond. Min Alarm
Cond. Max Alarm
Elec. Imp. E1-GND
Elec. Imp. E2-GND
Elec. Imp Min Alarm
Elec. Imp Max Alarm
Diagnosis Values
SNR Value
Slope Value
Slope Variation
Reference
Signal Ratio
Signal max
Signal min
NV Resets/sec
Amplification Int.
Startup F.P.
Startup F.P.
Sensor Coil R
Elec. Coat. QE1
Elec. Coat. QE2
Elec. Imp. E1-GND
Elec. Imp. E2-GND
Start Tx Startup F.P
Tx check CMR
Tx check 5 m/s
Tx check 10 m/s
On demand F.P.
Start On demand F.P.
Sensor Coil R
Elec. Coat. QE1
Elec. Coat. QE2
Elec. Imp. E1-GND
Elec. Imp. E2-GND
Start Tx Startup F.P
Tx check CMR
Tx check 5 m/s
Tx check 10 m/s
Trend Conductivity
Electrode Coat. QE1
Electrode Coat. QE2
Logtime Interval
Trend-Logging
Batching Damping
Preset Batch Tot.
Lag Corr. Mode
Lag Corr. Quantity
Batch Tot.
Total Batch Counter
Reset Batch Counter
Start/Stop Batching
IMPORTANT (NOTE)
This menu is used only to display the device parameters. The parameters are displayed
independently of the configured access level, but cannot be changed.
Device Info
Sensor Select the "Sensor" submenu.
Acquisition Select the "Acquisition" submenu.
Analog Range Select the "Analog Range" submenu.
Transmitter Select the "Transmitter" submenu.
Device Setup
....Access Control Select the "....Access Control" submenu.
Sensor Location TAG Alphanumeric, max. Enter the measuring point tag of the sensor (the
20 characters measuring point tag is shown in the upper left-hand
corner of the process display).
Sensor TAG Alphanumeric, max. Enter the TAG number of the sensor.
20 characters
Damping 0,02 ... 60 s Set the damping (the value relates to 1 T (Tau).
The value relates to a step flow rate change.
It affects the instantaneous value in the display and at
the current output.
Default setting: 1 second
Density 0,01 ... 5,0 g/cm³ If the flow is measured and indicated in the units g/s,
g/min, g/h, kg/s, kg/min, kg/h, kg/d, t/min, t/h, t/d, lb/s,
lb/min, lb/h and lb/d, a fixed density must be taken into
account for the calculations.
To convert the flow rate to mass flow units, a density
value from 0.01 to 5.0 g/cm³ can be entered.
....Low Flow Cut Off Select the "....Low Flow Cut Off" submenu.
TX Location TAG Alphanumeric, max. Enter the measuring point tag for the transmitter.
20 characters
TX TAG Alphanumeric, max. Enter the TAG number for the transmitter.
20 characters
....Operating Mode Select the "....Operating Mode" submenu.
Display
Language Deutsch, English, Français, Select the menu language.
Español, Italiano, Dansk,
Svenska, Polski, Russki,
Zhongweng, Turkce
Contrast 0 ... 100 % Contrast setting for the LCD display
Operator Pages Select the "Operator Pages" submenu.
IMPORTANT (NOTE)
Up to four user-specific operator pages (layouts)
can be configured for the process display. If
several operator pages have been configured,
these can be scrolled manually. In the factory
setting only Operator Page 1 is enabled.
Autoscroll On / Off If Multiplex mode is enabled, you can also activate the
"Autoscroll" function on the information level.
In this function, operator pages appear on the LCD
window in ten-second intervals. Manual scrolling
through pre-configured operator pages as described
above is no longer necessary.
When Autoscroll mode is enabled, the icon is
displayed on the lower left.
Default setting: Off
Flowrate Format x, x.x, x.xx, x.xxx, x.xxxx Select the decimal places for the flow indicator.
Default setting: x.xx
Volume Format x, x.x, x.xx, x.xxx, x.xxxx Select the decimal places for the flow totalizer.
Factory setting: x.xx
Date/Time Format DD-MM-YYYY, MM-DD- Set the display format for the date and time. Factory
YYYY, YYYY-MM-DD setting: YYYY-MM-DD
Display Test Start the test of the LCD display with "OK".
Input/Output
....Digital Out Mode Select the "....Digital Out Mode" submenu.
Digital Input Setup No Function, Totalizer Select the operating mode for the digital output. There
Reset(All), Flowrate to Zero, are four operating modes available:
System Zero Adjust, Totalizer • Totalizer reset for all totalizers (forward, reverse
Stop(All), Dual Range, and difference totalizer)
Start/Stop Batching 1) • External switch-off
• External zero point balancing
• External totalizer stop for all totalizers (forward,
reverse and difference totalizer)
• Switchover between measuring ranges 1 and 2
(Qmax and Qmax2)
• Start / stop of the fill function (batch) 1).
Default setting: external switch-off
IMPORTANT (NOTE)
If the fill operation is stopped before the set fill
quantity is reached, the fill totalizer is set to zero.
When the fill function is restarted, the interrupted
fill operation is not continued.
Current Select the "Current" submenu.
Fullscale Frequency 0 ... 5000 Hz Set the frequency for the upper range value in
"Fullscale Frequency" operating mode.
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
Input/Output / Current
Iout at Alarm High Alarm, Select the status of the current output in error
condition.
Low Alarm
The output "low" or "high" current is set in the
subsequent menu.
Default setting: "High".
Low Alarm Value 3,5 ... 3,6 mA Select the current for Low Alarm.
Default setting: 3.5 mA.
High Alarm Value 21 ... 23 mA Select the current for High Alarm.
Factory setting: 21.8 mA.
Iout at EP Alarm Off, Q=0%, Select the status of the current output for an empty
meter tube.
High Alarm,
• Off:
Low Alarm
Error is not output at the current output.
• Q = 0 %:
The current output assumes the value for "No
flow".
• High Alarm:
The current output assumes the value for "High
Alarm".
• Low Alarm:
The current output assumes the value for "Low
Alarm".
Default setting: Off.
Iout at Flow >103% Off, High Alarm, Low Alarm Select the status of the current output for overshoot of
the upper range value.
• Off:
Error is not output at the current output.
• High Alarm:
The current output assumes the value for "High
Alarm".
• Low AlarmThe current output assumes the value
for "Low Alarm".
Default setting: Off.
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
Process Alarm
Clear Alarm History - Allows you to clear the alarm list.
Group Masking Select the "Group Masking" submenu.
Individual Masking Select the "Individual Masking" submenu.
Alarm Simulation Off, ... A variety of alarm messages and output conditions
can be simulated.
For further information read and observe chapter
“Alarm simulation”.
Communication
HART Select the "HART" submenu.
Cyclic Data Out Select the "Cyclic Data Out" submenu.
Service Port Select the "Service Port" submenu.
PROFIBUS Select the "PROFIBUS" submenu.
The menu is displayed only for devices with
PROFIBUS PA.
FIELDBUS Foundation Select the "FF" submenu.
The menu is displayed only for devices with
FOUNDATION Fieldbus.
Communication / HART
Device Address 0 ... 15 Select the HART device address.
The HART protocol has provisions for creating a bus
with up to 15 devices (1 ... 15).
IMPORTANT (NOTE)
If an address greater than 0 is set, the device
operates in multidrop mode. The current output is
fixed at 4 mA. Apart from that, the current output
is only used for HART communication.
Default setting: 0
HART TAG 8 characters, uppercase only, Enter a HART TAG number as unique identifier for the
no special characters. device.
HART Descriptor 16 characters, uppercase Enter a HART descriptor.
only, no special characters.
HART Message Display only. Display of the alphanumeric TAG number.
HART Manf. ID Display only. Display of the HART manufacturer ID.
ABB = 26
HART Device ID Display only. Display of the HART device ID.
FEX300 / FEX500 = 30
Last HART Command Display only. Display of the most recently sent HART command.
Communication / PROFIBUS (continued) The menu is displayed only for devices with
PROFIBUS PA.
AI1-Q Flowrate Display only Current flow in the selected unit from the Transducer
Block Flow, including status.
TOT1-Q Flowrate Display only Current totalizer status in the selected unit from the
Transducer Block Flow, including status.
TOT2-Q Flowrate Display only Current totalizer status in the selected unit from the
Transducer Block Flow, including status.
AI2-Internal Tot Fwd Display only Current totalizer status of the forward totalizer in the
selected unit from the Transducer Block Flow,
including status.
AI3-Internal Tot Rev Display only Current totalizer status of the reverse totalizer in the
selected unit from the Transducer Block Flow,
including status.
AI4-Diagnostics Display only Current output value, including status. The channel
can be selected using the "AI4 Channel" parameter.
This function block delivers active values for FEX500
only. For this purpose, the sensor measurement or the
conductivity measurement must be switched on. For
FEX300 this function block delivers "0".
AI4-Channel Sensor Temp, Selection of the channel output by AI4. The
Conductivity PV_SCALE and OUT_SCALE structure is not
adapted. This channel is active for FEX500 only.
AO-Density Adjust Display only Current density output value from the Transducer
Block Flow, including status.
DI-Alarm Info Display only Current output value, including status. The channel
can be selected using the "DI Channel" parameter.
DI-Channel Maintenance, Out of Spec, Select the channel output by "DI Alarm Info".
Function Check, Failure
DO-Cyclic Control Display only Current function, including status. The function can be
selected using the "DO Channel" parameter.
DO-Channel Off, Totalizer Reset(All), Select the function for "DO Cyclic Control".
Flowrate to Zero, System
Zero Adjust, Totalizer
Stop(All), Dual Range,
Start/Stop Batching
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
Diagnostics
Diagnosis Control Select the "Diagnosis Control" submenu.
Diagnosis Values Select the "Diagnosis Values" submenu.
Fingerprints Select the "Fingerprints" submenu.
Trend Select the "Trend" submenu.
Flowrate Alarm Select the "Flowrate Alarm" submenu.
Simulation Mode Off, Flow Velocity, Q [units], Q Manual simulation of measured values.
[%],Iout, Freq on DO1, Freq The output values correspond to the simulated
on DO2, Logic DO1, Logic flowrate entered.
DO2, HART Freq, Digital in The "Configuration" information is displayed in the
lower line of the display.
Restore the Simulation mode to "Off" once completed.
The values in the "Value range" column can be
simulated.
Output Readings Select the "Output Readings" submenu.
Diagnostics / Fingerprints 1)
Factory F.P. Select the "Factory F.P." submenu.
Startup F.P. Select the "Startup F.P." submenu.
On demand F.P. Select the "On demand F.P." submenu.
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
The “fingerprint” database integrated in the transmitter allows you to compare the values at the
time of factory calibration or commissioning with the currently recorded values.
As a result, changes of the measuring system can be detected early, and the appropriate
measures can be taken.
Diagnostics / Trend 1)
Conductivity The measured values are shown as a line diagram.
Electrode Coat. QE1 When the "Trend-Logging" function is enabled, the
Electrode Coat. QE2 measured values are stored at the defined interval
(cycle time). The last 12 measured values are stored
and indicated as a line diagram. The oldest data
record is overwritten.
Logtime Interval 1 ... 45,000 min. Interval for measured value logging.
Trend-Logging ON / OFF Activate the "Trend-Logging" function.
When the "Trend-Logging" function is enabled, the
measured values are stored at the defined interval
(cycle time).
The data records can be read out by the
"ScanMaster" verification software and analyzed as
trends.
Totalizer
....Reset Vol. Totals Select the "....Reset Vol. Totals" submenu.
Totalizer / ....Batching 1)
Damping ON / OFF Switches the damping on / off.
Default setting: ON
IMPORTANT (NOTE)
In order to achieve a shorter response time for
the fill function, the damping must be switched
off.
The fill time should be > 3 seconds.
Preset Batch Tot. - Setting of the fill quantity. When the defined fill
quantity is reached, the configured digital output is
activated.
Lag Corr. Mode Automatic / Manual Select the overrun correction.
Closing the fill valve takes some time and as a
consequence more liquid is added, even though the
fill quantity is reached and the contact for closing the
valve is actuated.
With "Automatic overrun correction" the set fill quantity
is corrected with the overrun quantity.
Lag Corr. Quantity -100.000 ... 100.000 Manual entry of the overrun quantity.
In the "Process Alarm / Alarm Simulation" menu a variety of alarms can be simulated.
Parameter Description
Process Alarm
... / Alarm Simulation
Off Alarm Simulation switched off.
0-Sim.CurrentOut Simulate current output
1-Sim.Logic on DO1 Switch on/off contact output (terminals 51, 52)
2-Sim.Pulse on DO1 Simulate pulse output (terminal 51/52)
3-Sim.Logic on DO2 Switch on/off contact output (terminals 41, 42)
4-Sim.Pulse on DO2 Simulate pulse output (terminal 51/52)
5-Min Alarm Flowrate Simulate flowrate min. alarm
6-Max Alarm Flowrate Simulate flowrate max. alarm
7-Flowrate >103% Simulate flowrate > 103 % as alarm
8-Flow Simulation Run flowrate simulation
9-Calbration Mode Run transmitter alarm on simulator
10-Flowrate to Zero Simulate external output switch-off
11-Totalizer Stop Simulate external totalizer stop
12-Tot.Display<1600h Simulate display value <1600 h for Qmax.
13-Totalizer Reset Simulate external totalizer reset
14-Err.Sensor-Comms Simulate distorted communication to SensorMemory.
15-HART Address <> 0 Simulate HART Multiplex mode
16-FRAM-Com Fail Simulate FRAM error in the transmitter
17-No Sensor Simulate error "No communication to SensorMemory"
18-Sim.Digital Input Simulate digital input "ON/OFF"
19-ADC saturated Simulate "AD converter override" error
20-Error Coil circ Simulate error in coil loop
21-Coil Resistor Simulate "Coil resistance out of limits" error
22-Driver Err Uref=0 Simulate "Reference voltage = 0" error
23-El.Noise too High Simulate "Noise signal too high" error
24-DC to High Simulate "DC too high, several NV resets" error
25-Empty Pipe Simulate "Empty pipe" error
27-NV Corrupt Simulate "NV Corrupt" error
29-Electrode Imp. Simulate "Electrode impedance out of limits" error
30-Hold Last Value Simulate "Hold last good value" error
32-Digi-Pot Error Simulate "Digital potentiometer" error
33-TFE Simulate "Partial filling alarm" error
34-CurrentOut Error Simulate "Loop current output interrupted" error
35-Not Calibrated Simulate "Not calibrated" error
36-SensorIncompatib. Simulate "Calibration mode incompatible" error
37-ROM Error Simulate ROM error in the transmitter
38-RAM Error Simulate RAM error in the transmitter
39-Sim. HART Freq. Simulate a HART frequency
40-SIL Simulate "Self check alarm" error
41-Conductivity Simulate "Conductivity alarm" error
42-Elec.Coated Simulate "Electrode deposits" error
43-Gas bubble Simulate "Gas bubbles" error
44-Pulse Cut Off Simulate "Pulse output" error
46-Sensor Temp Simulate "Sensor temperature alarm" error
2
1
Q
DI O C
DO
4
M 0 t
tv
tn
5
G01046
The integrated fill function (batch) of the ProcessMaster FEP500 and HygienicMaster FEH500
models allows you to control filling processes with a filling time >3 seconds.
During this process, the fill quantity is given via an adjustable totalizer.
The fill operation is started via the digital input (DI) or via the fieldbus.
The valve is triggered via one of the digital outputs (DO) and closed again once the preset fill
quantity is reached.
The transmitter measures the overrun quantity (tn) and uses this value to calculate the overrun
quantity correction value.
Additionally, the low flow cut-off can be activated if required.
8.6.1 Configuration
The fill function is configured via menus using the LCD indicator.
1. If the fill operation is to be controlled via the digital input (DI), the following settings must be
applied in the "Input/Output" menu.
Input/Output
Digital Input Setup Start/Stop Batching Select the operating mode for the digital
1) output.
• Start / stop of the fill function (batch)
1).
Totalizer / ....Batching 1)
Damping Off Switches the damping on / off.
Default setting: ON
IMPORTANT (NOTE)
In order to achieve a shorter response time for the fill function,
the damping must be switched off.
The fill time should be > 3 seconds.
Preset Batch Tot. - Setting of the fill quantity. When the
defined fill quantity is reached, the
configured digital output is activated.
Lag Corr. Mode Automatic / Manual Select the overrun correction.
Closing the fill valve takes some time and
as a consequence more liquid is added,
even though the fill quantity is reached
and the contact for closing the valve is
actuated.
With "Automatic overrun correction" the
set fill quantity is corrected with the
overrun quantity.
Totalizer / ....Batching 1)
Lag Corr. Quantity -100.000 ... 100.000 Manual entry of the overrun quantity.
Batch Tot. Read only Once a fill operation has been started,
the quantity already filled is shown here.
The totalizer restarts at zero for each fill
operation initiated and then counts up to
the set fill quantity.
Total Batch Counter Read only Total number of fill operations.
4. To display all the data relevant to the fill operation in the process display, one of the operator
pages in the "Display" menu must be configured accordingly.
Software D200S062U01
Software Version Type of changes Operating instruction
00.01.01 Original software OI / FEP300 / FEH300, rev. A
00.01.02 Function extension, incorporated OI / FEP300 / FEH300, rev. A
new HART commands
00.02.00 Optimized measured value OI / FEP300 / FEH300, rev. B
processing
00.02.01 Optimized measured value OI / FEP300 / FEH300, rev. B
processing
00.02.04 Optimized boot sequence OI / FEP300 / FEH300, rev. B
Software D200S069U01
01.01.02 Optimized access to the service OI / FEX300 / FEX500, rev. C
menu. Implemented TFE
functionality. Added diagnostic
functions and batch mode
(for 500 series, only)
01.01.04 Optimized sensitivity of the keys OI / FEX300 / FEX500, rev. D
on the display
01.01.06 Optimized view on the display OI / FEX300 / FEX500, rev. D
01.02.00 Totalizer preset for ProcessMaster OI / FEX300 / FEX500, rev. E
300 implemented.Error with
Swedish menu navigation
corrected.
01.02.01 TFE functionality optimized for OI / FEX300 / FEX500, rev. F
integral mount design.
01.03.01 Software optimized for ScanMaster OI / FEX300 / FEX500, rev. F
verification tool.
IMPORTANT (NOTE)
• The extended diagnostic functions are available for ProcessMaster 500 and
HygienicMaster 500 only.
• The "Partial Filling Detector" function is not available for HygienicMaster 500.
• When using the extended diagnostic functions the external flowmeter sensor must not be
provided with a preamplifier.
• To facilitate initial start-up, the extended diagnostic functions are deactivated (factory
default).
• To use the extended diagnostic functions, a "start-up fingerprint" must be created during
start-up of the flowmeter.
• Each diagnostic function (e.g. Gas Bubble Detector or Electrode Deposit Detector) can be
individually activated. Once activated, the diagnostic function must be calibrated according
to the conditions on site and the limit values must be set.
Optionally, a measuring electrode (TFE electrode) is available for detecting a partially filled
flowmeter sensor. The alarm for partial filling is output via the programmable digital output.
Conditions for using the function:
• Nominal diameter from DN 50 (2“) with sensor design level B
• Max. signal cable length for version with external transmitter 200 m (656 ft).
• Conductivity of the measuring medium: 20 μS/cm ... 20,000 μS/cm
• The function is only available for ProcessMaster 300 / 500 without explosion protection or
with explosion protection for Zone 2 / Div. 2.
Additional installation conditions:
• The flowmeter sensor must be installed horizontally with the terminal box pointing upward.
Gas bubbles in the fluid are detected by using an adjustable maximum limit value. When this
limit value is exceeded, an alarm is tripped via the programmable digital output, depending on
the configuration.
Conditions for using the function:
• This function is available in the nominal diameter range 1) of DN 10 … 300 (3/8 “ ... 12 “).
• The signal cable length of the remote transmitter must not exceed a maximum value of 50 m
(164 ft) .
• For this function, the conductivity of the measuring medium must be in the range
20 μS/cm ... 20,000 μS/cm.
Additional installation conditions:
• The flowmeter sensor can be installed either horizontally or vertically. Vertical installation is
preferred.
1) The specified nominal diameter range is valid for ProcessMaster, only. The nominal diameter range valid for
HygienicMaster is DN 10 … 100 (3/8 “ ... 4 “).
This function provides the opportunity to detect coatings on the measuring electrodes by using
an adjustable maximum limit value.
When the set limit value is exceeded, an alarm is tripped via the programmable digital output,
depending on the configuration.
Conditions for using the function:
• This function is available in the nominal diameter range 2) of DN 10 … 300 (3/8 “ ... 12 “).
• The signal cable length of the remote transmitter must not exceed a maximum value of 50 m
(164 ft) .
• For this function, the conductivity of the measuring medium must be in the range
20 μS/cm ... 20,000 μS/cm.
Additional installation conditions:
• When using plastic tubes, install a grounding plate at the front and back of the device.
The conductivity of the fluid is monitored by using an adjustable minimum / maximum limit
value.
When the value falls below or exceeds the set limit value, an alarm is tripped via the
programmable digital output, depending on the configuration.
Conditions for using the function:
• This function is available in the nominal diameter range 1) of DN 10 … 300 (3/8 “ ... 12 “).
• The signal cable length of the remote transmitter must not exceed a maximum value of 50 m
(164 ft) .
• For this function, the conductivity of the measuring medium must be in the range
20 μS/cm ... 20,000 μS/cm.
Additional installation conditions:
• When using plastic tubes, install a grounding plate at the front and back of the device.
• There must not be any deposits on the measuring electrodes.
1) The specified nominal diameter range is valid for ProcessMaster, only. The nominal diameter range valid for
HygienicMaster is DN 10 … 100 (3/8 “ ... 4 “).
The impedance between the electrode and ground is monitored by using a minimum / maximum
limit value. This enables the transmitter to detect an electrode fine short or leakage.
When the value falls below or exceeds the set limit value, an alarm is tripped via the
programmable digital output, depending on the configuration.
Conditions for using the function:
• This function is available in the nominal diameter range 1) of DN 10 … 300 (3/8 “ ... 12 “).
• The signal cable length of the remote transmitter must not exceed a maximum value of 50 m
(164 ft) .
• For this function, the conductivity of the measuring medium must be in the range
20 μS/cm ... 20,000 μS/cm.
Additional installation conditions:
• When using plastic tubes, install a grounding plate at the front and back of the device.
• There must not be any deposits on the measuring electrodes.
• The measuring tube must always be completely full, and the fluid must feature only minor
conductivity variations.
This function includes the monitoring of the sensor temperature and the monitoring of the
resistance of the flowmeter sensor's coils.
1) The specified nominal diameter range is valid for ProcessMaster, only. The nominal diameter range valid for
HygienicMaster is DN 10 … 100 (3/8 “ ... 4 “).
9.1.7 Trend
The device has an internal memory where the measured value for the electrode deposits and
the conductivity are cyclically stored as a data set with an adjustable time (1 min ... 45000 min).
A maximum of 12 data sets is stored. When the thirteenth record is stored, the oldest data set is
overwritten automatically.
The data sets can be read out or analyzed as a trend using the external diagnostic tool
(ScanMaster).
9.1.8 Fingerprint
The “fingerprint” database integrated in the transmitter allows you to compare the values at the
time of factory calibration or commissioning with the currently recorded values.
This function allows you to check the electrical grounding of the device.
While the check is in progress, no flow measurement can take place.
Conditions for using the function:
• The measuring tube must be completely full.
• No flow must occur in the flowmeter sensor.
Additional installation conditions:
• The flowmeter sensor must not be provided with a preamplifier.
....Grounding Check Once the earthing check has been started, the frequency
Grounding Check 46,75 range up to 250 Hz is measured. The four most intensive
48,65 frequencies of the spectrum are shown at the right of the
50,08 display.
52,00 The corresponding amplitudes and the power spectrum
Back OK over the frequency range can be called up for display using
the following parameters.
....Grounding Check 7. Use or to select the "Power Spectrum" entry.
Power Spectrum 8. Use to display the parameter.
335
Back OK
The measured values indicate possible disturbances to the earthing line of the device at the
time of the test.
No or minor disturbance:
• When the power spectrum is below 1000
• When the four measured amplitude values are above 10
In the "Diagnostics / Diagnosis Control / ..." menu, limit values for the diagnostic values can be
specified.
In order to simplify their setting, recommendations for the individual limit values are shown here.
The values indicated are only intended as a rough guide and may need to be adapted in line
with on-site conditions.
The coil resistance depends on the measuring medium temperature Tmedium and the ambient
temperature.
Parameter
Tmedium R coil min alarm R coil max alarm
-40 °C Factory for Fingerprint Factory for Fingerprint
(-40 °F) (coil resistance) x 0.71 (coil resistance) x 0.79
-20 °C Factory for Fingerprint Factory for Fingerprint
(-4 °F) (coil resistance) x 0.81 (coil resistance) x 0.89
0 °C Factory for Fingerprint Factory for Fingerprint
(32 °F) (coil resistance) x 0.9 (coil resistance) x 1.0
20 °C Factory for Fingerprint Factory for Fingerprint
(68 °F) (coil resistance) x 0.95 (coil resistance) x 1.05
60 °C Factory for Fingerprint Factory for Fingerprint
(140 °F) (coil resistance) x 1.19 (coil resistance) x 1.31
90 °C Factory for Fingerprint Factory for Fingerprint
(194 °F) (coil resistance) x 1.28 (coil resistance) x 1.42
130 °C Factory for Fingerprint Factory for Fingerprint
(266 °F) (coil resistance) x 1.43 (coil resistance) x 1.58
180 °C Factory for Fingerprint Factory for Fingerprint
(356 °F) (coil resistance) x 1.62 (coil resistance) x 1.79
Electrode deposit monitoring is switched off (factory default). Monitoring can be switched on in
the "Diagnostics / Diagnosis Control / Coating Detector" menu.
IMPORTANT (NOTE)
The deposit value QE is the mean value of Startup Fingerprint QE1 and QE2. The value is
determined using the following formula:
QE = (Startup Fingerprint QE1 + Startup Fingerprint QE2) / 2
Electrode impedance monitoring is switched off (factory default). Monitoring can be switched on
in the "Diagnostics / Diagnosis Control / Cond. Detection" menu.
The limit values for parameters "Elec.Imp.Min Alarm" and "Elec.Imp.Max Alarm" depend on
the measuring media conductivity and must be determined on site.
Recommended settings
• Elec.Imp.Min Alarm = 0.2 x average impedance value
• Elec.Imp.Max Alarm = 3.0 x average impedance value
IMPORTANT (NOTE)
The average impedance value is the value of Startup Fingerprint "Elec. Imp. E1-GND" and
"Elec. Imp. E2-GND". The value is determined using the following formula:
Average impedance value = (Startup Fingerprint "Elec. Imp. E1-GND" + Startup Fingerprint
"Elec. Imp. E2-GND") / 2
10 Error messages
Electronic
unit
Back Select
Electronic unit The first line shows the area in which the error has
F123.321 occurred.
Brief description
The second line shows the unique error number.
Information
Back Exit The next lines show a brief description of the error and
information on how to remedy it.
10.2.1 Errors
10.2.4 Maintenance
Error no. / Text on the LCD Current Digital output Pulse Display Error
Range display output behavior output maskable?
behavior behavior
F254.038 RAM Error in Transmitter Iout at Alarm General Alarm 0 Hz 0% No
Contact
Electronics ABB Service
unit
F253.037 ROM Error in Transmitter Iout at Alarm General Alarm 0 Hz 0% No
Contact
Electronics ABB Service
unit
F252.017 No Sensor Memory Iout at Alarm General Alarm 0 Hz 0% No
Check wiring
Sensor Check switch SW3
F251.040 Self Check Alarm Iout at Alarm General Alarm 0 Hz 0% No
Electronics
unit
F250.016 Tx. memory fault Iout at Alarm General Alarm 0 Hz 0% No
detected
Electronics Contact
unit ABB Service
F248.036 Incompatible snsTx+ Iout at Alarm General Alarm 0 Hz 0% No
snr are not the same
Sensor series
F246.032 Defect digital Iout at Alarm General Alarm 0 Hz 0% No
potentiometer
Electronics Transmitter Hardware
unit fault
ABB Service
F245.047 Stack NV Corrupt Iout at Alarm General Alarm 0 Hz 0% No
Contact
Electronics ABB Service
unit
F244.031 Internal supply voltage Iout at Alarm General Alarm 0 Hz 0% No
error
Electronics Contact
unit ABB Service
F236.024 DC to High Iout at Alarm General Alarm 0 Hz 0% No
Lot of NV-Resets
Operation Refer to instr. Manual
F232.022 Driver Error Iout at Alarm General Alarm 0 Hz 0% No
Uref = 0
Electronics Check wiring for
unit open circuit
Check fuse
F228.020 Error in Iout at Alarm General Alarm 0 Hz 0% No
Coil circuit
Electronics Check wiring for
unit short circuit
F226.019 AD Converter Iout at Alarm General Alarm 0 Hz 0% No
saturated
Electronics Check empty pipe
unit or Galv. Voltage
Error no. / Text on the LCD Current Digital output Pulse Display Error
Range display output behavior output maskable?
behavior behavior
An alarm is Current value No response Current value Current No
C190.045 simulated value
Configuration Switch off
alarm simulation
Tx Simulator/ Current value Current value Current value Current Mask group
C186.009 Calibrator mode value
Configuration Switch off
Calibrator Mode
Hold last good Current value No response Current value Current Mask group
C185.030 known value value
Operation Switch OFF
Noise Reduction
ABB Service
The Flowrate 4 mA (0 % flow) No response 0 Hz 0% Mask group
C184.010 is set to zero
Configuration Check digital in
terminals 81,82
Flowrate Current Value No response, Current value Current Mask group
C182.008 Simulation or High Alarm Min/Max or General value
Configuration Switch off (flow > 105 %) Alarm
Simulation Mode
Error no. Text on the LCD Current Digital output Pulse Display Error
range display output behavior output maskable?
behavior behavior
Simulation Current value No response Current value Current Mask group
C154.018 Digital In value
Configuration Switch off
Simulation Mode
Coil resistor Current value No response Current value Current Mask group
C149.021 out of limits value
Sensor Check wiring
Contact
ABB Service
Empty Pipe Programmed Programmed alarm 0 Hz 0% Mask single
S148.025 Check Pipe alarm alarm
Operation
Gas Bubble Alarm No response No response No response No Mask group
S149.021
response
Operation
S146.043 Partially Current value Programmed alarm Current value Current Mask group
filled pipe(TFE) value
Operation Check Pipe
Or adjust
Detector
S144.033 Electrode Programmed Programmed alarm Current value Current Mask group
Coating Alarm alarm value
Operation
S143.042 Conductivity Alarm Current value Programmed alarm Current value Current Mask group
value
Operation
S142.041 Sensor and or Housing Current value Programmed alarm Current value Current Mask group
Temperature to high value
Operation
S141.046 Flowrate >103% Current value Programmed alarm Current value Current Mask group
Check Flowrate value
Operation Check Range
Setting
Max Alarm Programmed Collective Alarm Current value Current Mask single
S140.007 Flowrate alarm value alarm
Operation
Min Alarm Current value Programmed alarm Current value Current Mask single
S136.006 Flowrate value alarm
Operation
Coil resistor Current value Programmed alarm Current value Current Mask single
S132.05 out of limits value alarm
Operation Check wiring
Contact
ABB Service
Electr.Impedance Current value No response Current value Current Mask group
S124.029 too high value
Operation Coating?
Conductivity?
Empty Pipe?
Short-ciruit E1 Current value No response Current value Current Mask group
S122.026 E2 with shield. value
Operation
Error no. Text on the LCD Current Digital output Pulse Display Error
range display output behavior output maskable?
behavior behavior
S120.023 Electrode Noise too high Current value No response Current value Current Mask group
Switch on value
Operation Noise Reduction
Sensor setup Current value Current value Current value Current Group
S110.035 Cal-Status value mask
Sensor Set Cal-Status
to calibrated
Pulse output is Current value No response Maximum Current Group
S108.044 cutted off possible value value mask
Operation Check pulse out
configuration
M099.027 NV Corrupt Current value No response Current value Current Group
value mask
Elektronics
unit
Current Low Alarm No response Current value Current Single Alarm
M94.034 out fault value mask
Electronics Comms. to MSP
unit Check wiring!
20mA passive?
Check BR901!
Errors Current value No response Current value Current Group
M90.014 Sensor Comms value mask
Sensor Bad EMC
environment
Check wiring
Display value is Current value No response Current value Current Group
M80.012 <1600h at Qmax value mask
Operation Change eng.
Unitfor Totalizer
-System Startup- Once the device has been switched on, the sensor
NO SENSOR DETECTED calibration data and the transmitter settings are loaded
Device Resetting from the SensorMemory into the transmitter.
If it is not possible to establish a communication with the
Offline SensorMemory1), the shown message appears on the LCD
display.
The device will restart after the progress bar is complete until either the communication with the
SensorMemory1) is re-established successfully or the process is canceled by selecting "Offline".
In Offline mode the device can be operated or parameterized, but no measurement is
performed.
In Offline mode the error message "F252.017" is set.
IMPORTANT (NOTE)
When commissioning the device, make sure that the transmitter is assigned to the sensor
correctly. It is not possible to operate a flowmeter sensor of the 300 series with a transmitter of
the 500 series.
If the transmitter is operated with a flowmeter sensor of another series, the following message
appears on the transmitter display:
Sensor
11 Maintenance
Repair and maintenance activities may only be performed by authorized customer service
personnel.
When replacing or repairing individual components, original spare parts must be used.
NOTICE - Potential damage to parts!
The electronic components of the printed circuit board can be damaged by static electricity
(observe ESD guidelines).
Make sure that the static electricity in your body is discharged before touching electronic
components.
The flowmeter sensor is largely maintenance-free. The following items should be checked
annually:
• Ambient conditions (air circulation, humidity),
• Seal integrity of the process connections,
• Cable entry points and cover screws,
• Operational reliability of the power supply feed, the lightning protection, and the station
ground.
The flowmeter sensor electrodes must be cleaned when the flowrate information on the
transmitter changes when recording the identical flowrate volume. If the display shows a higher
flowrate, the contamination is insulating. If a lower flowrate is displayed, the contamination
results in a short-circuit.
For repairs to the lining, electrodes or magnet coil, the flowmeter must be returned to the head
office in Göttingen.
IMPORTANT (NOTE)
When sending the flowmeter sensor for repair to the head office of ABB Automation Products
GmbH, complete the return form in the appendix and include with device.
When cleaning the exterior of meters, make sure that the cleaning agent used does not corrode
the housing surface and the gaskets.
11.2 Gaskets
Some device designs are shipped with special gaskets. These gaskets must be used and
installed properly to prevent leakage and ensure 3A conformity.
For all other device designs, use commercially available gaskets made from a compatible
material for the measuring medium and prevailing temperature (rubber, PDFE, lt, EPDM, silicon,
Viton, etc.) or use 3A-compliant gasket material for HygienicMaster devices.
IMPORTANT (NOTE)
A wafer type sensor is installed without gaskets directly in the pipeline.
IMPORTANT (NOTE)
When replacing the transmitter or sensor, ensure that they are assigned correctly. It is not
possible to operate a sensor from the 300 series with a transmitter from the 500 series.
The series (e.g. ProcessMaster 300 or ProcessMaster 500) is shown on the name plate of the
transmitter or sensor.
11.3.1 Transmitter
4 3
1 2 3
3 G00689
G00689-02
1 2
G00698-03
Fig. 69
1. Switch on the power supply. After switching on the power supply, the following messages
appear in succession on the LCD display:
Transmitter Sensor
3. The flowmeter is ready for operation and will operate with factory settings or settings
requested by the customer. To change the factory settings, refer to chapter 8
"Parameterization".
1) The SensorMemory is a data memory integrated in the flowmeter sensor.
IMPORTANT (NOTE)
System data must only be loaded during initial start-up. If the power supply is later switched
off, the transmitter automatically loads all data the next time the power supply is switched on
again.
A selection as described below (1-3) is not required.
1 2
1
G00690-02
Fig. 70
No. Name of part Order number
1 Fuse (1.0 A) for power supply, suitable for all devices D151B003U05
2 Fuse (0.25 A) for the coil circuit in the field housing, suitable for D151B003U02
all devices
1 2 3
G00878-01
Fig. 71
No. Name of part Order number
1 Housing cover for transmitter with single-compartment housing MJFA9915
with integral mount design
2 Universal backplane for transmitter with dual-compartment D685A1156U01
housing
3 Front housing cover for transmitter with dual-compartment D612A197U01
housing with integral mount design (standard, Ex Zone 2 / Div. 2)
Front housing cover for transmitter with dual-compartment D612A197U02
housing with integral mount design (Ex Zone 1 / Div. 1)
1 2
4
3
G00879-01
Fig. 72
No. Name of part Order number
1 Contact board assy for dual-compartment transmitter housing D682A016U01
2 Universal backplane for dual-compartment transmitter housing D685A1156U01
3 Housing cover for transmitter with single-compartment housing MJBX9905
with remote mount design
4 Backplane for transmitter with single-compartment housing with 3KXF002058U0100
remote mount design
1 2
G00876-01
Fig. 73
No. Name of part Order number
1 Front housing cover for transmitter with dual-compartment D612A197U01
housing with remote mount design (standard, Ex Zone 2 / Div. 2)
Front housing cover for transmitter with dual-compartment D612A197U02
housing with remote mount design (Ex Zone 1 / Div. 1)
2 Universal backplane for dual-compartment transmitter housing D685A1156U01
G00880
Fig. 74
No. Name of part Order number
for FEH model for FEP model
1 Connection board (without preamplifier) D685A1090U01 D685A1090U01
Connection board (with preamplifier) D685A1089U01 D685A1089U01
G00877
Fig. 75
No. Name of part Order number
1 O-ring D101A034U06
2 Cable gland for Zone 1 / Div. 1, plastic, black, M20 x 1.5 D150A004U15
13 Performance specifications
Change from one to two columns
A
Fig. 76
1) In the case of hazardous area design for ambient temperature of -40 °C (40 °F)
2) Cable Gland with M20x1,5 thread or NPT thread with plug to be specified through
modelcode
Change from one to two columns
IMPORTANT (NOTE)
When using the device in potentially explosive areas, the additional temperature specifications in the section titled "Ex
relevant specifications" on the data sheet or in the separate Ex safety instructions (SM / FEX300 / FEX500 / ATEX /
IECEX) or (SM / FEX300 / FEX500 / FM / CSA) must be observed.
14.5.6 Overview sensor design level "C" ASME flange, stainless steel, up to DN 400 (16") (CL150/300) up
to DN 1000 (40") (CL150)
100 1450.4
CL600
90 1305.3
80 1160.3
Electrode design:
60 870.2
Pressure rating
50 725.2
Hard rubber
CL300
Sensor Size
40 580.2
Standard
30 435.1
PTFE 20
CL150
290.1
10 145.0
0
DN 25 (1") X X — X X -30 -10 10 30 50 70 90 110 130 150 170 [°C]
-22 14 50 86 122 158 194 230 266 302 338 [°F]
DN 32 (1 1/4") X X — X X TS G00591
DN 40 (1 1/2") X X X X X Fig. 78
DN 50 (2") X X X X X
DIN flange, steel, up to DN 600 (24")
DN 65 (2 1/2") DIN PN 10, X X X X X
PS PS
DN 80 (3") DIN PN 16, X X X X X [bar] [PSI]
DN 400 (16") X X X X X 40 PN 40
580.2
DN 500 (20") X X X X X 20
PN 25
290.1
PN 16
DN 600 (24") X X X X X 10 145.0
PN 10
0
-30 -10 10 30 50 70 90 110 130 150 170 [°C]
-22 14 50 122 122 158 194 230 266 302 338 [°F]
14.6 Material load TS G00588
Fig. 79
Limits for the permissible fluid temperature (TS) and permissible
pressure (PS) are calculated on the basis of the lining and flange
material used in the device (refer to the name plate on the device). ASME flange, steel, up to DN 400 (16") (CL150/300); up to
DN 1000 (40") (CL150)
PS
[bar]
14.6.1 Flowmeter sensor design level “B” 110
100
90 1305.3 50
CL300
80 1160.3 40
70 1015.3 30
PN 63
60 870.2 20
CL150
10
50 725.2
0
40 580.2 -30 -10 10 30 50 70 90 110 130 150 170 [°C]
PN 40
-22 14 50 86 122 158 194 230 266 302 338 [°F]
30 435.1 TS G00590
20
PN 25
290.1
Fig. 80
PN 16
10
PN 10
145.0 JIS 10K-B2210 flange
0 Nominal Material PN TS PS
10 30 70 110 130 150 170 [°C]
-30
-22
-10
14 50 86
50
122 158
90
194 230 266 302 338 [°F] diameter
TS G00589
32 ... 400 Stainless steel 10 -25 ... 180 °C 10 bar
Fig. 77 (1 1/4 ... 16") (-13 ... 356 °F) (145 psi)
32 ... 400 Steel 10 -10 ... 180 °C 10 bar
(1 1/4 ... 16") (14 ... 356 °F) (145 psi)
DIN flange, stainless steel, DN 700 (28") up to DN 1000 (40") ASME flange, stainless steel, DN 25 ... 400 (1 ... 24“)
PS PS
[bar] [psi] PS PS
[bar] [psi]
17 246.5 CL2500
350 5076
16 DN 700 PN 16 232.0
Fig. 81 TS G01337
Fig. 83
0 0
-10 0 10 20 30 40 50 60 70 80 90 100 110 120 130 °C
14 32 50 68 86 104 122 140 158 176 194 212 230 248 266 °F
TS
G01336
Fig. 86
G01340
G01341
The process connections are made of one of the materials listed below:
1) 1.4301 (AISI 304), 1.4307, 1.4404 (AISI 316L) 1.4435 (AISI 316L), 1.4541 (AISI 321)
1.4571 (AISI 316Ti), ASTM A182 F304, ASTM A182 F304L, ASTM A182 F316L, ASTM
A182 F321, ASTM A182 F316TI, ASTM A182 F316, 0Cr18Ni9, 0Cr18Ni10,
0Cr17Ni13Mo2, 0Cr27Ni12Mo3, 1Cr18Ni9Ti, 0Cr18Ni12Mo2Ti
2) 1.0038, 1.0460, 1.0570, 1.0432, ASTM A105, Q255A, 20#, 16Mn
Change from one to two columns
Preamplifier
Maximum signal cable length between flowmeter sensor and
transmitter:
a) Without preamplifier:
• Max. 50 m (164 ft) for conductivity ≥ 5 µS/cm
A preamplifier is required for cables > 50 m (164 ft).
b) With preamplifier
• Max. 200 m (656 ft) for conductivity ≥ 5 µS/cm
FEH321
Flange -20 °C (-4 °F) 60 °C (140 °F) -25 °C (-13 °F) 180 °C (356 °F)
FEH325
1) The following is valid for the low temperature design (option): -40°C (-40°F).
2) For CIP/SIP cleaning, higher temperatures are permitted for limited time periods; refer to the table „Max. permissible cleaning temperature“ on page 171.
40 580.0
PN 40
35 507.5
Fig. 89
30 435.0
25
PN 25
362.5
JIS 10K-B2210 wafer type design
20 290.0 Nominal Material PN TS PS [bar]
15
PN 16
217.5 diameter
10 145.0 DN 32 ... 100 1.4404 10 -25 ... 130 °C 10
PN 10
5 72.5 (1 1/4 ... 4") 1.4435 (-13 ... 266 °F) (145 psi)
0 0 1.4301
-30 -10 10 30 50 70 90 110 130 150 170 190 [°C]
TS
-22 14 50 86 122 158 194 230 266 302 338 374 [°F]
G00215
Fig. 87
16 Appendix
Customer details:
Company:
Address:
Contact person: Telephone:
Fax: Email:
Device details:
Type: Serial no.:
Reason for the return/description of the defect:
Was this device used in conjunction with substances which pose a threat or risk to health?
Yes No
If yes, which type of contamination (please place an X next to the applicable items)?
Biological Corrosive/irritating Combustible (slightly/extremely
combustible)
Toxic Explosive Other Toxic substances
Radioactive
We hereby state that the devices/components shipped have been cleaned and are free from any dangerous or
poisonous substances.
IMPORTANT (NOTE)
All documentation, declarations of conformity and certificates are available in ABB's download
area.
www.abb.com/flow
IMPORTANT (NOTE)
This is a class A device (industrial sector). This device can cause radio interferences in
residential areas. In this case, the operator may be required to take appropriate measures to
remedy the fault.
abb.com/flow
—
We reserve the right to make technical changes or modify the contents of this document
without prior notice. With regard to purchase orders, the agreed particulars shall prevail.
ABB does not accept any responsibility whatsoever for potential errors or possible lack of
information in this document.
We reserve all rights in this document and in the subject matter and illustrations contained
therein. Any reproduction, disclosure to third parties or utilization of its contents – in whole or
in parts – is forbidden without prior written consent of ABB.
© 2019 ABB
All rights reserved 3KXF231300R4201
Ex-FMU
(ABB)
SM/FEX300/FEX500/ATEX/IECEX-X1
ProcessMaster / HygienicMaster
FEX300, FEX500
IMPORTANT (NOTICE)
This symbol indicates operator tips, particularly useful information, or important information
about the product or its further uses. It does not indicate a dangerous or damaging situation.
Devices suitable for use in potentially explosive atmospheres feature the corresponding Ex
mark on their name plates.
The design intended for use in Ex Zones 2, 21, and 22 is identified by the letter "M" in the
model number, while the design intended for use in Zones 1 and 21 is identified by the letter
"L".
Example:
FEP315-100A1S1D2B0A1A0M1A1C1, FEP325-100A1S1D2B0A1A1M1A0Y1,
FET325-1A0M1A1C1
Important
For further information about the instruments' explosion protection approval please refer to the
Ex test certificates (available on the product CD or under www.abb.com/flow).
G00886 G00883
ATEX ATEX
Certificate: FM08ATEX0080 Certificate: FM08ATEX0080
DN3-300: II 2G Ex d e ia ma IIC T6 … T2 II 2G Ex d e ia ma IIC T6 … T2
>DN300 : II 2G Ex d e ia IIC T6 … T2
II 2 D Ex tD iaD A21 IP6X T70°C … Tmedium II 2 D Ex tD iaD A21 IP6X T70°C … Tmedium
IEC IEC
Certificate: FME08.0004 Certificate: FME08.0004
DN3-300: Ex d e ia ma IIC T6 … T2 Gb Ex d e ia ma IIC T6 … T2 Gb
>DN300 : Ex d e ia IIC T6 … T2 Gb
Ex tD iaD A21 IP6X T70°C … Tmedium Ex tD iaD A21 IP6X T70°C … Tmedium
ATEX ATEX
Certificate: FM08ATEX0038, FM08ATEX0080 Certificate: FM08ATEX0038, FM08ATEX0080
II 3 G Ex nA nC IIC T4 … T3 II 3 G Ex nA nC IIC T4 … T3
II 2 D Ex tD A21 IP6X T70 °C … Tmedium II 2 D Ex tD A21 IP6X T70 °C … Tmedium
IEC IEC
Certificate: FME08.0004 Certificate: FME08.0004
Ex nA nC IIC T4 … T3 Ex nA nC IIC T4 … T3
Ex tD A21 IP6X T70 °C … Tmedium Ex tD A21 IP6X T70 °C … Tmedium
Important
For further information about the instruments' explosion protection approval please refer to the
Ex test certificates (available on the product CD or under www.abb.com/flow).
If the FEP325 flowmeter sensor is combined with the FET321 or FET325 transmitter (Zone 2),
the maximum permissible signal cable length is 50 m (164 ft) with a minimum conductivity of
5 µS/cm, without a pre-amplifier. The same is valid for the FEP525 flowmeter sensor with
FET521 or FET525 (Zone 2) transmitter.
If the FEP325 flowmeter sensor is combined with the FET325 transmitter (Zone 1), 10 m
(32.8 ft) signal cables are permanently connected to the transmitter. The same is valid for the
FEP525 flowmeter sensor with FET525 (Zone 1) transmitter.
Flowmeter sensor
ProcessMaster 300 / ProcessMaster 500
FEP325.....L.... / FEP525.....L....
In Ex area, Zone 1, 21, 22
G00862
ATEX
Certificate: FM08ATEX0080
DN3-300: II 2G Ex e ia ma IIC T6 … T2
>DN300: II 2G Ex e ia IIC T6 … T2
II 2 D Ex tD iaD A21 IP6X T85 °C … Tmedium
IEC
Certificate: FME08.0004
DN3-300: Ex e ia ma IIC T6 … T2 Gb
>DN300 : Ex e ia IIC T6 … T2 Gb
Ex tD A21 IP6X T85 °C … Tmedium Gb
Transmitter
In Ex area, Zone 1, 21, 22 In Ex area, Zone 2, 21, 22 Outside the Ex area
FET325.....L.... / FET525.....L.... FET325.....M.... / FET525.....M.... FET321 / FET521
G00863
ATEX ATEX -
Certificate: FM08ATEX0038,
Certificate: FM08ATEX0080 No Ex approval!
FM08ATEX0080
II 2 (2) G Ex d e [ia] IIC T6 II 3 G Ex nA IIC T4
II 2 (2) D Ex tD [iaD] A21 IP6X T70°C II 2 D Ex tD A21 IP6X T70°C
IEC IEC
Certificate: FME08.0004 Certificate: FME08.0004
Ex d e [ia Gb] IIC T6 Gb Ex nA IIC T4
Ex tD [iaD] A21 IP6X T70°C Ex tD A21 IP6X T70°C
Without a pre-amplifier, the maximum permissible signal cable length is 50 m (164 ft) with a
minimum conductivity of 5 µS/cm.
With a pre-amplifier, the maximum permissible signal cable length is 200 m (656 ft).
Flowmeter sensor
ProcessMaster 300 / ProcessMaster 500 HygienicMaster 300 / HygienicMaster 500
FEP325.....M.... / FEP525.....M.... FEH325.....M.... / FEH525.....M....
In Ex area, Zone 2, 21, 22 In Ex area, Zone 2, 21, 22
ATEX ATEX
Certificate: FM08ATEX0038, FM08ATEX0080 Certificate: FM08ATEX0038, FM08ATEX0080
II 3 G Ex nA IIC T6 … T3 II 3 G Ex nA IIC T6 … T3
II 2 D Ex tD A21 IP6X T85 °C … Tmedium II 2 D Ex tD A21 IP6X T85 °C … Tmedium
IEC IEC
Certificate: FME08.0004 Certificate: FME08.0004
Ex nA IIC T6 … T3 Ex nA IIC T6 … T3
Ex tD A21 IP6X T85 °C … Tmedium Ex tD A21 IP6X T85 °C … Tmedium
Transmitter
(The flowmeter sensors illustrated above can be combined with the transmitters described
below)
In Ex area, Zone 2, 21, 22 Outside the Ex area
FET325.....M.... / FET525.....M.... FET321 / FET521
ATEX -
Certificate: FM08ATEX0038, FM08ATEX0080 No Ex approval!
II 3 G Ex nA IIC T4
II 2 D Ex tD A21 IP6X T70°C
IEC
Certificate: FME08.0004
Ex nA IIC T4
Ex tD A21 IP6X T70°C
FEH315 or FEH515
HygienicMaster 500
Important
All documentation, declarations of conformity, and certificates are available in ABB's download
area.
www.abb.com/flow
2
12
3
13
4 PN 16
14
5
15
6
16
7
17
8
18
9
19
10
20
11
21
G00629
Important
Meters with 3A approval are labeled with an additional plate.
1
FEP325 11
2
12
3
4
5
6
13
7
14
8
9
15
10
16
17
18
G00630
Important
Meters with 3A approval are labeled with an additional plate.
4 9
G00864
4 9
6
10
7
G00632-01
Release the cover safety device by screwing in the Allen screw (1).
Before closing the housing make sure that the gasket (O-ring) for the housing cover is seated
correctly.
After closing the housing, lock the housing cover by unscrewing the Allen screw (1) to prevent
opening the cover accidentally.
1 1
G00758
Fig. 5
1 2 3
G00759
Fig. 6
Black cable fittings are installed by default when the device is supplied. If signal outputs are
connected to intrinsically safe circuits, we recommend that you replace the black caps on the
corresponding cable fittings with the blue ones supplied.
Important
In order to provide for the required temperature resistance, devices in low-temperature design
(optional, ambient temperature down to -40 °C (-40 °F) are delivered with a cable gland made
from metal.
When operating these devices on intrinsically safe current circuits, remove the metal cable
glands.
G00654
Fig. 7
1 Insulation
3
2
3 G00761
Fig. 8
1. Loosen the Allen screws (3) on the front and back sides, but do not remove entirely.
2. Loosen screws (2) and rotate the transmitter housing (1) by 90° to the left or right.
3. Retighten screws (2) and Allen screws (3).
Risk of explosion!
When the screws for the transmitter housing are loosened, the explosion protection is
suspended.
Tighten all screws (2, 3) for the transmitter housing prior to commissioning.
A temperature of 70 °C (158 °F) at the cable entry is assumed for the Ex calculations.
Therefore, the cables used for the supply power and the signal inputs and outputs must have a
minimum specification of 70 °C (158 °F).
4.1.1 Models FEP315 / FEP515, FEH315 / FEH515, FEP325 / FEP525 and FET325 / FET525 in Zone 1 with
HART protocol
FEP315 FEH315
FEP515 FEH515
A
24 V
6
1+ 2- + - + - + - + -
L N 31 32 51 52 81 82 41 42 M1 M2 D1 D2 3 2S E2 E1 1S SE
6
6
{
{
{
{
{
1 PE 2 3 4 5 FEP325 FET325
FEP525 FET525
7 8 9 10 11 12 13
6
M1 M2 D1 D2 3 E2 E1 SE
B
6 G00866-01
Fig. 9
A Transmitter 4 Digital input: (terminal 81/82)
B Flowmeter sensor Only available in conjunction with "passive" current output.
1 Supply power: Function can be configured locally using software: External output
See name plate switch-off, external totalizer reset, external totalizer stop, other
2 Current output (terminals 31/32) Data for the optocoupler: 16 V ≤ U ≤ 30 V, Ri = 2 kΩ
Dependent upon the device design, an "active" or a "passive" 5 Digital output DO2 (terminal 41/42)
output will be available. The output is always a "passive" output (optocoupler).
For devices designed for use in Ex Zone 1, the current output Data for the optocoupler: Umax = 30 V, Imax = 220 mA
cannot be reconfigured locally. Function can be configured locally as "Pulse Output" or "Digital
• Active: 4 ... 20 mA, HART protocol (standard), load: 250 Ω ≤ R ≤ Output" using software. Factory setting is "Digital Output", flow
300 Ω direction signaling.
• Passive: 4 ... 20 mA, HART protocol (standard), 6 Equipotential bonding
load: 250 Ω ≤ R ≤ 650 Ω, 7 Brown
Supply voltage for the current output: minimum 11 V, maximum 8 Red
30 V at terminals 31/32. 9 Orange
3 Digital output DO1 (terminal 51/52) 10 Yellow
The output is always a "passive" output (optocoupler). 11 Green
• Data for the optocoupler: Umax = 30 V, Imax = 220 mA, 12 Blue
Function can be configured locally as "Pulse Output" or "Digital 13 Violet
Output" using software. Factory setting is "Pulse Output".
• Configuration as pulse output. Maximum pulse frequency:
5,250 Hz, pulse width: 0.1 … 2,000 ms. The pulse factor and
pulse width are interdependent and are calculated dynamically.
• Configuration as contact output. Function: System alarm, empty
pipe alarm, max./min. alarm, flow direction signaling, other
All inputs and outputs are electrically isolated from each other and from the supply power.
The electrical specifications given are operating values.
A
FF+ FF- 6
1+ 2- PA+ PA- + -
L N 97 98 41 42 M1 M2 D1 D2 3 2S E2 E1 1S SE
6
6
{
{
{
{
{
1 PE 2 3 4 5 FEP325 FET325
FEP525 FET525
7 8 9 10 11 12 13
6
M1 M2 D1 D2 3 E2 E1 SE
B
6 G01019
Fig. 10
A Transmitter 5 Digital output DO2 (terminal 41/42)
B Flowmeter sensor The output is always a "passive" output (optocoupler).
1 Supply power: Data for the optocoupler: Umax = 30 V, Imax = 220 mA
See name plate Function can be configured locally as "Pulse Output" or "Digital
2 Digital communication (terminal 97 / 98) Output" using software. Factory setting is "Digital Output", flow
• PROFIBUS PA in acc. with IEC 61158-2 (PA+ / PA-) direction signaling.
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the 6 Equipotential bonding
event of an error / FDE) 7 Brown
Bus connection with integrated protection against polarity 8 Red
reversal 9 Orange
The bus address can be set via the DIP switches in the device, 10 Yellow
the transmitter display or the fieldbus. 11 Green
or 12 Blue
• FOUNDATION Fieldbus in acc. with IEC 61158-2 (FF+ / FF-) 13 Violet
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the
event of an error / FDE)
Bus connection with integrated protection against polarity
reversal
3 Not assigned
4 Not assigned
All inputs and outputs are electrically isolated from each other and from the supply power.
The electrical specifications given are operating values.
For devices with PROFIBUS PA or FOUNDATION Fieldbus the bus termination must conform to the FISCO model or the explosion protection
regulations, respectively.
6a
A
24 V
1+ 2- + - + - + - + -
L N 31 32 51 52 81 82 41 42 M1 M2 D1 D2 3 2S E2 E1 1S SE
6 / 6a
{
{
{
{
{
1 PE 2 3 4 5
7 8 9 10 11 12 13
M1 M2 D1 D2 3 E2 E1 SE
B
6 G00867
Fig. 11
A Transmitter 4 Digital input: (terminal 81/82)
B Flowmeter sensor Function can be configured locally using software: External output
1 Supply power: switch-off, external totalizer reset, external totalizer stop, other
See name plate Data for the optocoupler: 16 V ≤ U ≤ 30 V, Ri = 2 kΩ
2 Current output (terminals 31/32) 5 Digital output DO2 (terminal 41/42)
The current output can be configured locally as an "active" or The output is always a "passive" output (optocoupler).
"passive" output. Data for the optocoupler: Umax = 30 V, Imax = 220 mA,
• Active: 4 ... 20 mA, HART protocol (standard), fmax ≤ 5250 Hz,
load: 250 Ω ≤ R ≤ 650 Ω Function can be configured locally as "Pulse Output" or "Digital
• Passive: 4 ... 20 mA, HART protocol (standard), Output" using software. Factory setting is "Digital Output", flow
load: 250 Ω ≤ R ≤ 650 Ω, direction signaling.
Supply voltage for the current output: minimum 11 V, maximum 6 Equipotential bonding
30 V at terminals 31/32. 6a Functional ground (only with flowmeter sensor FET321 outside
3 Digital output DO1 (terminal 51/52) the hazardous area)
The digital output can be configured locally as an "active" or 7 Brown
"passive" output. 8 red
• Active: U = 19 ... 21 V. Imax = 220 mA, fmax ≤ 5250 Hz 9 Orange
• Passive: Umax = 30 V, Imax = 220 mA, fmax ≤ 5250 Hz 10 yellow
Function can be configured locally as "Pulse Output" or "Digital 11 Green
Output" using software. Factory setting is "Pulse Output". 12 blue
• Configuration as pulse output. Maximum pulse frequency: 13 Violet
5,250 Hz, pulse width: 0.1 … 2,000 ms. The pulse factor and
pulse width are interdependent and are calculated dynamically.
• Configuration as contact output. Function: System alarm, empty
pipe alarm, max./min. alarm, flow direction signaling, other
All inputs and outputs are electrically isolated from each other and from the supply power.
The electrical specifications given are operating values.
6a
A
FF+ FF-
1+ 2- PA+ PA- + -
L N 97 98 41 42 M1 M2 D1 D2 3 2S E2 E1 1S SE
6 / 6a
{
{
{
{
{
1 PE 2 3 4 5
7 8 9 10 11 12 13
M1 M2 D1 D2 3 E2 E1 SE
B
6 G01006
Fig. 12
A Transmitter 5 Digital output DO2 (terminal 41/42)
B Flowmeter sensor The output is always a "passive" output (optocoupler).
1 Supply power: Data for the optocoupler: Umax = 30 V, Imax = 220 mA
See name plate Function can be configured locally as "Pulse Output" or "Digital
2 Digital communication (terminal 97 / 98) Output" using software. Factory setting is "Digital Output", flow
• PROFIBUS PA in acc. with IEC 61158-2 (PA+ / PA-) direction signaling.
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the 6 Equipotential bonding
event of an error / FDE) 7 Brown
Bus connection with integrated protection against polarity 8 Red
reversal 9 Orange
The bus address can be set via the DIP switches in the device, 10 Yellow
the transmitter display or the fieldbus. 11 Green
or 12 Blue
• FOUNDATION Fieldbus in acc. with IEC 61158-2 (FF+ / FF-) 13 Violet
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the
event of an error / FDE)
Bus connection with integrated protection against polarity
reversal
3 Not assigned
4 Not assigned
All inputs and outputs are electrically isolated from each other and from the supply power.
The electrical specifications given are operating values.
For devices with PROFIBUS PA or FOUNDATION Fieldbus the bus termination must conform to the FISCO model or the explosion protection
regulations, respectively.
All inputs and outputs are electrically isolated from each other and from the supply power.
Special connection conditions:
The output circuits are designed in such a way that they can be connected to both intrinsically-safe and non-intrinsically-safe circuits. It is not
permitted to combine intrinsically safe and non-intrinsically safe circuits. On intrinsically safe circuits, equipotential bonding must be in place
along the entire length of the cable used for the current outputs.
The rated voltage of the non-intrinsically safe circuits is UM = 60 V.
Provided that rated voltage UM = 60 V is not exceeded if connections are established to non-intrinsically safe external circuits, intrinsic safety is
still guaranteed.
If the installation is changed from type of protection "e" to "i" or vice versa, the device must be checked in accordance with the instructions
contained in section 6.4 "Changing the type of protection".
Variant 1 Intrinsically safe fieldbus connection in acc. with FISCO, intrinsically safe connection of the digital output
1) Intrinsically safe single-channel or multi-channel barriers (supply isolators) with resistance characteristic must be used.
Variant 2 Intrinsically safe fieldbus connection (not in acc. with FISCO!), intrinsically safe connection of the digital output
1) Intrinsically safe single-channel or multi-channel barriers (supply isolators) with resistance characteristic must be used.
Variant 3 Fieldbus connection in acc. with FNICO (Zone 2, Div. 2), connection of digital output (Zone 2, Div. 2)
1) Single-channel or multi-channel barriers (supply isolators) with resistance characteristic must be used.
All inputs and outputs are electrically isolated from each other and from the supply power.
Special connection conditions:
The output circuits are designed in such a way that they can be connected to both intrinsically-safe and non-intrinsically-safe circuits. It is not
permitted to combine intrinsically safe and non-intrinsically safe circuits. On intrinsically safe circuits, equipotential bonding must be in place
along the entire length of the cable used for the signal outputs.
The rated voltage of the non-intrinsically safe circuits is UM = 60 V.
Provided that rated voltage UM = 60 V is not exceeded if connections are established to non-intrinsically safe external circuits, intrinsic safety is
still given.
If the installation is changed from type of protection "e" to "i" or vice versa, the device must be checked in accordance with the instructions
contained in section 6.4 "Changing the type of protection".
Important
The maximum permissible fluid temperature depends on the lining and flange material, and is limited by the operating
values in Table 1 and the explosion protection specifications in Tables 2 ... n.
class
Not thermally Thermally Not thermally Thermally Not thermally Thermally
insulated insulated insulated insulated insulated insulated
Gas & Gas & Gas & Gas & Gas & Gas &
Gas Gas Gas Gas Gas Gas
dust dust dust dust dust dust
NT 130 °C 90 °C 30 °C 80 °C 40 °C
T1
HT 180 °C 120 °C 20 °C 120 °C 20 °C
NT 130 °C 90 °C 30 °C 80 °C 40 °C
T2
HT 180 °C 120 °C 20 °C 120 °C 20 °C
DN 3 ... DN 100
NT 130 °C 90 °C 30 °C 80 °C 40 °C
T3
HT 180 °C 120 °C 20 °C 120 °C 20 °C
NT 120 °C 90 °C 30 °C 80 °C 40 °C
T4
HT 120 °C 120 °C 20 °C 120 °C 20 °C
NT 85 °C 70 °C 30 °C 80 °C 40 °C
T5
HT 85 °C 85 °C 20 °C 85 °C 20 °C
NT 70 °C 70 °C 30 °C 70 °C 40 °C
T6
HT 70 °C 70 °C 20 °CC 70 °C 20 °C
NT 130 °C 90 °C 30 °C 80 °C 40 °C
T1
HT 180 °C 120 °C 20 °C 120 °C 20 °C
NT 130 °C 90 °C 30 °C 80 °C 40 °C
T2
DN 125 ... DN 2000
Important
The standard version includes explosion protection for gases and dust.
• If the installation location for the device is classified as a potentially explosive area for gases and dust, the
temperature data in the "Gas & dust" columns in the table must be taken into consideration.
• If the installation location for the device is classified as a potentially explosive area for gases only, the temperature
data in the "Gas" columns in the table must be taken into consideration.
class
Not thermally Thermally Not thermally Thermally Not thermally Thermally
insulated insulated insulated insulated insulated insulated
Gas & Gas & Gas & Gas & Gas & Gas &
Gas Gas Gas Gas Gas Gas
dust dust dust dust dust dust
NT 130 °C 110 °C 110 °C 110 °C 110 °C
T1
HT 180 °C 160 °C 150 °C 160 °C 150 °C
NT 130 °C 110 °C 110 °C 110 °C 110 °C
T2
HT 180 °C 160 °C 150 °C 160 °C 150 °C
DN 3 ... DN 100
Important
The standard version includes explosion protection for gases and dust.
• If the installation location for the device is classified as a potentially explosive area for gases and dust, the
temperature data in the "Gas & dust" columns in the table must be taken into consideration.
• If the installation location for the device is classified as a potentially explosive area for gases only, the temperature
data in the "Gas" columns in the table must be taken into consideration.
class
Not thermally Thermally Not thermally Thermally Not thermally Thermally
insulated insulated insulated insulated insulated insulated
Gas & Gas & Gas & Gas & Gas & Gas &
Gas Gas Gas Gas Gas Gas
dust dust dust dust dust dust
NT 130 °C 110 °C 20 °C 80 °C 40 °C
T1
HT 180 °C 120 °C 20 °C 120 °C 20 °C
NT 130 °C 110 °C 20 °C 80 °C 40 °C
T2
HT 180 °C 120 °C 20 °C 120 °C 20 °C
DN 3 ... DN 100
NT 130 °C 110 °C 20 °C 80 °C 40 °C
T3
HT 180 °C 120 °C 20 °C 120 °C 20 °C
NT 120 °C 110 °C 20 °C 80 °C 40 °C
T4
HT 120 °C 120 °C 20 °C 120 °C 20 °C
NT 85 °C 85 °C 20 °C 80 °C 40 °C
T5
HT 85 °C 85 °C 20 °C 85 °C 20 °C
NT 70 °C 70 °C 20 °C 70 °C 40 °C
T6
HT 70 °C 70 °C 20 °C 70 °C 20 °C
Important
The standard version includes explosion protection for gases and dust.
• If the installation location for the device is classified as a potentially explosive area for gases and dust, the
temperature data in the "Gas & dust" columns in the table must be taken into consideration.
• If the installation location for the device is classified as a potentially explosive area for gases only, the temperature
data in the "Gas" columns in the table must be taken into consideration.
For devices with remote mount design for use in FM / cFM Div. 1 or FM / cFM Div. 2 the signal
cable between the flowmeter sensor and the transmitter must have a minimum length of 5 m
(16.4 ft).
6a
FEH315 / FEH515 FEP325 / FEP525 FET325 / FET525
FEH325 / FEH525
6
6
6
A
24 V
1+ 2- + - + - + - + -
L N 31 32 51 52 81 82 41 42 M1 M2 D1 D2 3 2S E2 E1 1S SE
6 / 6a
{
{
{
{
{
1 PE 2 3 4 5
7 8 9 10 11 12 13
M1 M2 D1 D2 3 E2 E1 SE
B
6 G00873-01
Fig. 13
A Transmitter 4 Digital input: (terminal 81/82)
B Flowmeter sensor Function can be configured locally using software: External output
1 Supply power: switch-off, external totalizer reset, external totalizer stop, other
See name plate Data for the optocoupler: 16 V ≤ U ≤ 30 V, Ri = 2 kΩ
2 Current output (terminals 31/32) 5 Digital output DO2 (terminal 41/42)
The current output can be configured locally as an "active" or The output is always a "passive" output (optocoupler).
"passive" output. Data for the optocoupler: Umax = 30 V, Imax = 220 mA, fmax ≤ 5250
• Active: 4 ... 20 mA, HART protocol (standard), load: 250 Ω ≤ R ≤ Hz
650 Ω Function can be configured locally as "Pulse Output" or "Digital
• Passive: 4 ... 20 mA, HART protocol (standard), Output" using software. Factory setting is "Digital Output", flow
load: 250 Ω ≤ R ≤ 650 Ω direction signaling.
Supply voltage for the current output: minimum 11 V, maximum 6 Equipotential bonding
30 V at terminals 31/32. 6a Functional ground (only with transmitter FET321 outside the
3 Digital output DO1 (terminal 51/52) hazardous area)
The digital output can be configured locally as an "active" or 7 Brown
"passive" output. 8 red
• Active: U = 19 ... 21 V. Imax = 220 mA, fmax ≤ 5250 Hz 9 Orange
• Passive: Umax = 30 V, Imax = 220 mA, fmax ≤ 5250 Hz 10 yellow
Function can be configured locally as "Pulse Output" or "Digital 11 Green
Output" using software. Factory setting is "Pulse Output". 12 blue
• Configuration as pulse output. Maximum pulse frequency: 13 Violet
5,250 Hz, pulse width: 0.1 … 2,000 ms. The pulse factor and
pulse width are interdependent and are calculated dynamically.
• Configuration as contact output. Function: System alarm, empty
pipe alarm, max./min. alarm, flow direction signaling, other
All inputs and outputs are electrically isolated from each other and from the supply power.
The electrical specifications given are operating values.
6a
FEH315 / FEH515 FEP325 / FEP525 FET325 / FET525
FEH325/ FEH525
6
6
6
A
FF+ FF-
1+ 2- PA+ PA- + -
L N 97 98 41 42 M1 M2 D1 D2 3 2S E2 E1 1S SE
6 / 6a
{
{
{
{
{
1 PE 2 3 4 5
7 8 9 10 11 12 13
M1 M2 D1 D2 3 2S E2 E1 1S SE
B
6 G01020
Fig. 14
A Transmitter 5 Digital output DO2 (terminal 41/42)
B Flowmeter sensor The output is always a "passive" output (optocoupler).
1 Supply power: Data for the optocoupler: Umax = 30 V, Imax = 220 mA,
See name plate fmax ≤ 5250 Hz,
2 Digital communication (terminal 97 / 98) Function can be configured locally as "Pulse Output" or "Digital
• PROFIBUS PA in acc. with IEC 61158-2 (PA+ / PA-) Output" using software. Factory setting is "Digital Output", flow
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the direction signaling.
event of an error / FDE) 6 Equipotential bonding
Bus connection with integrated protection against polarity 6a Functional ground (only with flowmeter sensor FET321 outside
reversal the hazardous area)
The bus address can be set via the DIP switches in the device, 7 Brown
the transmitter display or the fieldbus. 8 red
or 9 Orange
• FOUNDATION Fieldbus in acc. with IEC 61158-2 (FF+ / FF-) 10 yellow
U = 9 ... 32 v, I = 10 mA (normal operation), I = 13 mA (in the 11 Green
event of an error / FDE) 12 blue
Bus connection with integrated protection against polarity 13 Violet
reversal
3 Not assigned
4 Not assigned
All inputs and outputs are electrically isolated from each other and from the supply power.
The electrical specifications given are operating values.
For devices with PROFIBUS PA or FOUNDATION Fieldbus in Zone 2 / Div 2 the bus termination must conform to the FNICO model or the
explosion protection regulations, respectively.
All inputs and outputs are electrically isolated from each other and from the supply power.
1) Single-channel or multi-channel barriers (supply isolators) with resistance characteristic must be used.
Important
The maximum permissible fluid temperature depends on the lining and flange material, and is limited by the operating
values in Table 1 and the explosion protection specifications in Tables 2 ... n.
class
T1 110 °C 3)
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 40 °C 180 °C 40 °C
100 °C 2)
NT 130 °C 130 °C --- --- 130 °C --- --- 80 °C 40 °C --- ---
T2 110 °C 3)
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 40 °C 180 °C 40 °C
100 °C 2)
NT 130 °C 130 °C --- --- 130 °C --- --- 80 °C 40 °C --- ---
T3 110 °C 3)
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 40 °C 180 °C 40 °C
100 °C 2)
NT 130 °C 130 °C --- --- 130 °C --- --- 80 °C 40 °C --- ---
T4 110 °C 3)
HT 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 40 °C 130 °C 40 °C
Important
The standard version includes explosion protection for gases and dust.
• If the installation location for the device is classified as a potentially explosive area for gases and dust, the
temperature data in the "Gas & dust" columns in the table must be taken into consideration.
• If the installation location for the device is classified as a potentially explosive area for gases only, the temperature
data in the "Gas" columns in the table must be taken into consideration.
class
110 °C 2)
NT 130 °C 130 °C --- --- 130 °C 130 °C --- --- 110 °C --- ---
T2 120 °C 3)
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C
110 °C 2)
NT 130 °C 130 °C --- --- 130 °C 130 °C --- --- 110 °C --- ---
T3 120 °C 3)
HT 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C 180 °C
110 °C 2)
NT 130 °C 130 °C --- --- 130 °C 130 °C --- --- 110 °C --- ---
T4 120 °C 3)
HT 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C 130 °C
NT 95 °C 95 °C --- --- 95 °C 95 °C --- --- 95 °C 95 °C --- ---
T5
HT 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C 95 °C
NT 80 °C 80 °C --- --- 80 °C 80 °C --- --- 80 °C 80 °C --- ---
T6
HT 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C 80 °C
Important
The standard version includes explosion protection for gases and dust.
• If the installation location for the device is classified as a potentially explosive area for gases and dust, the
temperature data in the "Gas & dust" columns in the table must be taken into consideration.
• If the installation location for the device is classified as a potentially explosive area for gases only, the temperature
data in the "Gas" columns in the table must be taken into consideration.
• Any unused connections must be sealed in accordance with IEC 60079 prior to
commissioning using the plugs supplied.
Important
Commissioning and operation must be performed in accordance with ATEX 137 or BetrSichV
(EN60079-14). Only properly trained personnel are authorized to carry out commissioning in
Ex areas.
6.2 Notes on combining the FEP325 flowmeter sensor with the FET325 transmitter or the FEP525
flowmeter sensor with the FET525 transmitter
G00875
Fig. 15
When combining the FEP325 flowmeter sensor with the FET325 transmitter, you must ensure
that the flowmeter sensor is assigned to the transmitter correctly.
The same is valid for combining the FEP525 flowmeter sensor with the FET525 transmitter.
The flowmeter sensor is identified by means of the suffix (X01, X02, etc.) that forms part of the
order number on the name plate.
The associated transmitter is identified by means of the suffix (Y01, Y02, etc.) that forms part of
the order number on the name plate.
Important
The outputs' type of protection remains unaffected by this. The devices connected to these
outputs must conform to the applicable regulations for explosion protection.
4
2 3
1 2 1 2 1 2
3 G00874-01
31 32 41 42 51 52 81 82
A B C D
G00884
G00885
• Visual inspection
Zone 2: • Special precautions/checks are not necessary.
Non-sparking (nA)
2 Zone 1: Zone 1: • Visual inspection (no damage to the threads of
Intrinsically safe circuits Ex d, non-intrinsically the cover or the cable fittings)
safe circuits
Zone 2: • Special precautions/checks are not necessary.
Non-sparking (nA)
3 Zone 2: Zone 1: • Perform the checks described under No. 1.
Non-sparking (nA) Intrinsically safe circuits
Zone 1: • Visual inspection (no damage to the threads of
Ex d, non-intrinsically the cover or the cable fittings)
safe circuits
Risk of explosion!
When the housing cover is open, the explosion protection is suspended.
Before opening the housing switch off the power to all connection lines and wait at least 20
minutes.
Important
Users must also observe the information in the operating and commissioning instruction for
the device.
7.2.1 Transmitter
1 2 3
4 3
G00689
G00689-01
Fig. 17
Replace the transmitter plug-in as follows:
1. Switch off the supply power and wait at least 20 minutes before the next step.
2. Release the cover safety device (4) and housing cover (1).
3. Loosen screws (3) and pull out transmitter plug-in (2).
4. Replace transmitter plug-in and retighten screws (3).
5. Close the housing cover (1) and lock the cover by unscrewing the screw (4).
1 2
G00698
G00698-01
Fig. 18
Replace the flowmeter sensor as follows:
1. Switch off the supply power and wait at least 20 minutes before the next step.
2. Open the cover safety device (4) and housing cover (1).
3. Disconnect the signal cable (if necessary, remove the sealing compound).
4. Install the new sensor according to the installation instructions.
5. Complete the electrical connection according to the connection diagram.
6. Close the housing cover (1) and lock the cover by unscrewing the screw (3).
1 2
1
G00690-01
Fig. 19
No. Name of part Order number
1 Fuse (1.0 A) for power supply, suitable for all devices D151B003U05
2 Fuse (0.25 A) for the coil circuit in the field housing, suitable for D151B003U02
all devices
G00878
Fig. 20
No. Name of part Order number
1 Cable gland without explosion protection, plastic, gray, M20 x 1.5 D150A008U15
Cable gland for Zone 2 / Div. 2, plastic, gray, M20 x 1.5 D150A008U23
Seal plug for Zone 2 / Div. 2, plastic, M20 x 1.5 D150Z007U09
Cable gland for Zone 1 / Div. 1, plastic, black, M20 x 1.5 D150A004U15
Cable gland for Zone 1 / Div. 1, plastic, blue, M20 x 1.5 D150A008U13
2 O-ring 118 x 3.7 (covered) D101A034U05
1+2
1+2
4
G00879
Fig. 21
No. Name of part Order number
1 Screw D004G108AU01
2 Spring washer D085D020AU20
3 Contact board assy D682A016U01
4 Cable gland without explosion protection, plastic, gray, M20 x 1.5 D150A008U15
Cable gland for Zone 2 / Div. 2, plastic, gray, M20 x 1.5 D150A008U23
Seal plug for Zone 2 / Div. 2, plastic, M20 x 1.5 D150Z007U09
5 Cover for supply power connection (not shown) D355H303U01
G00876
Fig. 22
No. Name of part Order number
1 Cable gland without explosion protection, plastic, gray, M20 x 1.5 D150A008U15
Cable gland for Zone 2 / Div. 2, plastic, gray, M20 x 1.5 D150A008U23
Seal plug for Zone 2 / Div. 2, plastic, M20 x 1.5 D150Z007U09
Cable gland for Zone 1 / Div. 1, plastic, black, M20 x 1.5 D150A004U15
Cable gland for Zone 1 / Div. 1, plastic, blue, M20 x 1.5 D150A008U13
2 O-ring 118 x 3.7 (covered) D101A034U05
1 2
3
G00880
Fig. 23
No. Name of part Order number
for FEH model for FEP model
1 O-ring D101A034U06 D101A034U06
2 Connection board (without preamplifier) D685A1090U01 D685A1090U01
Connection board (with preamplifier) D685A1089U01 D685A1089U01
3 Cable gland for Zone 2 / Div. 2, plastic, gray, D150A008U23 D150A008U23
M20 x 1.5
G00877
Fig. 24
No. Name of part Order number
1 O-ring D101A034U06
2 Cable gland for Zone 1 / Div. 1, plastic, black, M20 x 1.5 D150A004U15
The version of the meter in your possession meets the requirements of the
CE mark
following European directives:
- EMC directive 2004/108/EC
- Low voltage directive 2006/95/EC
- Pressure equipment directive (PED) 97/23/EC
- ATEX directive 94/9/EC
Explosion Identification for intended use in potentially explosive atmospheres according
Protection to:
- IEC standards
- FM Approvals (US)
- NEPSI (China)
- GOST
Important
All documentation, declarations of conformity, and certificates are available in ABB's download
area.
www.abb.com/flow
FLOW
FLOW
FLOW FLOW
FLOW
FLOW
5 TV5 TV6 TH9
HORIZONTAL TYPE
8
FLOW FLOW FLOW FLOW FLOW FLOW
7 ◈Operation◈
■ Prior to the system operation, open the outlet of the strainer
and a valve, which is close by, and then gradually open the i-
4 nlet of the strainer and a valve, which is close by, after opera-
ting the system. At this time, it is important to open a valve
gradually to prevent the damage of the filter(8).
■ If a pressure difference between the strainer front and near is
6 increased, the filter(8) should be cleaned because there is a f-
oreign substance.
5
◈Maintenance and Repair◈
■ The water leak inspection should be executed periodically an-
d at a proper time, the filter(8) should be cleaned and repla-
ced.
■ When the filter(8) is blocked by a foreign substance , it will
ITEM ITEM be the reason of a potential damage to the connected equip-
DESCRIPTION DESCRIPTION ment by occurring the pressure difference between the strai-
NO. NO.
ner front and near therefore when the pressure between the
BODY INLET front and the near is increased, the filter should be cleaned.
1 6 STUD BOLT
FLANGE
BODY OUTLET
2 7 COVER ◈Filter cleaning and Replacement◈
FLANGE
■ Block a flow of a fluid by closing a front and a rear valve of
3 BODY 8 FILTER the strainer. (If a valve is not installed, block a inflow of a fluid.)
■ Gradually open the drain plug(4) and completely discharge a
4 DRAIN PLUG 9 COVER GASKET pressure and fluid which is inside. At this time, if the operat-
ing oil contains a fluid except of a water or a operation tem-
5 NUT perature is over 120℉(48.9℃), a fluid should be discharged
to the safety space which is away from a operator. A operator
◈Installation◈ should wear a safety equipment.
■ Check the gasket to be inserted between the flange of the pi- ■ Release the nut(5) and then dismantle the cover(7).
pe line and the flange of the strainer extremity. If the strainer ■ Remove a foreign substance after taking out the filter(8).
and the pipe line are connected by welding, rust and other u- ■ Confirm the filter(8) whether it is damaged or not. If it is da-
nknown material have to be removed. maged, replace it. Change a cover gasket(9).
■ When the strainer is installed, the flow direction of the fluid ■ Remove a foreign substance which is inside of the strainer.
and the flow direction(Arrow) which is marked on the product ■ Assemble in reverse order of a dismantlement. When the filt-
should be in line and should be installed to be able to chan- er(8) is assembled, assemble with considering a inflow directi-
ge and clean the filter(8). At this time, horizontality and vert- on of a fluid and the shape of the filter such as the picture
icality of the strainer, which is connected with the pipe line, which is on the left side.
should be checked.
21
WOOSUNG FLOWTEC Co.
Head office / Factory
1461-1, Songjung-dong, Gangseo-gu, Busan, Korea
Installation, Operation &
Tel : +82-51-831-1533 Fax : +82-51-831-1534 Maintenance Manual
Branch Office(Seoul)
2F, Soojung B/D, #225-2 Seokchon-dong, Songpa-gu, Seoul, Korea Tee Strainers
Tel : +82-2-412-6822 Fax : +82-2-412-6823 Angle Flow Type (M/N:TA1/2)
Web : www.woosungflowtec.com
FLOW
FLOW
1 FLOW FLOW FLOW FLOW
FLOW
FLOW
TV1 TV2 TV3 TV4
5
HORIZONTAL TYPE
FLOW
FLOW FLOW FLOW FLOW
FLOW
8
TH1 TH2 TH3 TH4
FLOW
9
FLOW
7
TH5 TH6 TH7 TH8
◈Operation◈
4 ■ Prior to the system operation, open the outlet of the strainer
and a valve, which is close by, and then gradually open the i-
nlet of the strainer and a valve, which is close by, after opera-
ting the system. At this time, it is important to open a valve
6 gradually to prevent the damage of the filter(8).
■ If a pressure difference between the strainer front and near is
increased, the filter(8) should be cleaned because there is a f-
oreign substance.
5
20
Mixing pump
(DESMI)
OPERATION AND MAINTENANCE INSTRUCTIONS
Tel.: +45 96 32 81 11
Fax: +45 98 17 54 99
E-mail: desmi@desmi.com
Internet: www.desmi.com
These operation and maintenance instructions apply to the DESMI ESL Monobloc pump.
The pump is a single-stage vertical "in-line" centrifugal pump (i.e. horizontal inlet and outlet on
the same line) equipped with stainless steel shaft, mechanical shaft seal, and closed impeller.
The impeller is mounted on the motor shaft at a shaft extension. The pumps are equipped with
a mechanical seal shaft and have holes in the bracket for inspection of leaks.
1.1 DELIVERY
- Check on delivery that the shipment is complete and undamaged.
- Defects and damages, if any, to be reported to the carrier and the supplier
immediately in order that a claim can be advanced.
2. TECHNICAL DATA
The pumps are manufactured in various material combinations which appear from the type
number on the name plate. See below.
ESLXXX-YYY-MR-Z
The pumps can be delivered in other material combinations according to agreement with the
supplier.
Any use of the pump is to be evaluated on the basis of the materials used in the pump. In case
of doubt, contact the supplier.
Pumps in material combinations A and C are primarily used for fresh water.
Pumps in material combination D are primarily used for seawater.
If the pumps are designed for special purposes the following is to be indicated:
Pump No. :
Pump type :
Application :
Comment :
The capacity of the pump appears from the name plate on the pump. If the pump has been
delivered without motor, the pump capacity is to be indicated on the plate when mounting the
motor.
The permissible loads on the flanges appear from the following table. The values apply to
standard pumps in bronze (Rg5) and cast iron
(GG20). As to pumps in SG iron (GGG40), NiAlBz
or stainless steel the values are to be increased by
a factor 1.5.
Horizontal pipeline 25 250 320 250 480 300 150 260 420
at right angles to the 40 400 500 400 750 400 200 300 550
shaft 50 500 600 550 1000 450 250 350 600
Suction and pressure 65 650 840 750 1340 510 310 380 700
flanges above level 80 800 950 850 1500 550 350 400 750
of installation 100 1000 1250 1150 2000 650 400 500 900
In connection with the permissible loads on the flanges the following is to be observed:
2 2
F
calculated M calculated 2
F M
Max. permissibl e Max. permissibl e
3. INSTALLATION
3.1 MOUNTING/FASTENING
Insert the pump in the pipeline in the same way as a valve. The pump can be inserted in both
horizontal and vertical pipes, but not in a horizontal pipe if the motor is placed below horizontal
level. The max. loads on the flanges appear from the technical description.
When installing the pump check that it is earthed to avoid an electrical potential in the pump.
At installations pumping hot or very cold liquids, the operator must be aware that it is
dangerous to touch the pump surface and, consequently, he must take the
necessary safety measures.
3.2 WIRING
Wiring to be carried out by authorised skilled workmen according to the rules and
regulations in force.
4. TRANSPORT/STORAGE
The weights of the pumps in A and D combination (without motor) are stated in the following
table, and the pumps are to be lifted as shown below.
5. DISMANTLING
Before dismantling the pump make sure that it has stopped. Empty the pump of
liquid before it is dismantled from the piping system. If the pump has been
pumping dangerous liquids you are to be aware of this and take the necessary
safety measures.
If the pump has been pumping hot liquids, take great care that it is drained before it is
removed from the piping system.
Before dismantling 1. Remove nut (15) and 2. Remove screw (20) and
stud (14); washer (23);
Before dismantling 1. Remove the flush pipe (29) only 2. Remove nut (15) and stud
forESL40/50-180N, remove venting (14);
plug (5), muff (9) and sleeve (21) for
ESL80/100-180N and ESL65-180M
3. Remove screw (20, 24 in 4. Remove the impeller (2, 5 in 5. Remove shaft seal (22, 522 in
ESL40/50-180N) and washer ESL40/50-180N) and the key (19) ESL40/50-180N) and rear cover
(23, 25 in ESL40/50-180N); (4, 7 in ESL40/50-180N);
5.3 INSPECTION
When the pump has been dismantled, check the following parts for wear and damage:
If it is necessary to use tools for assembling, then protect the sliding surface of the seat to
prevent it from being scratched or cut. Lubricate the inner surface of the slide ring rubber
bellows with soapy water and push it over the shaft. The use of a conical fitting bush as
shown on the assembly drawing is recommended to avoid that the rubber bellows is cut.
Push the slide ring over the shaft with the hand. If the rubber bellows is tight, use a fitting
tool and take care that the slide ring is not damaged. If the carbon ring is not fixed, it is
important to check that it is fitted correctly, i.e. the chamfered/lapped side is to face the seat.
The carbon ring can be held by a little grease.
When using soapy water on the shaft, the bellows will settle and seat in abt. 15 minutes, and
until then tightness should not be expected. After start, check by viewing the leak hole that
there are no leaks.
71/80 M6 10Nm
90/100/112/132/160 M8 24Nm
6.7 SHAFT
When the pump has been assembled, check that the shaft rotates freely. In case the shaft
has been dismantled in the combination, tap the shaft towards the shaft end of the electric
motor by means of a plastic hammer, and fasten the pointed screws (first the middle screw)
according to the below table. Check that the wobble, measured as close to the shaft end as
possible, is within the limits indicated in the table.
71 M6 10Nm 70um
80 M6 10Nm 70um
90 M8 24Nm 70um
Pumps which are not in operation during frost periods are to be drained to avoid frost
damage. Remove the plug at the bottom to empty the pump. Alternatively, it is possible to
use anti-freeze liquids in normal constructions.
8. START-UP
A centrifugal pump will not function until it has been filled with liquid between the foot
valve and somewhat above the impeller of the pump.
The liquid also serves as coolant for the shaft seal. In order to protect the shaft seal
the pump must not run dry.
ATTENTION
For safety reasons the pump is only allowed to operate against closed discharge valve for a
short time (max. 5 minutes and at a max. temperature of 80°C for standard pumps).
Otherwise there is a risk of damage to the pump and, at worst, of a steam explosion. If the
pump is not monitored, the installation of a safety device is recommended.
8.1 START-UP
Before starting the pump check that:
Start the pump for a moment to check the direction of rotation. If the direction is correct (i.e.
in the direction of the arrow) the pump may be started.
In order to protect the shaft seal the pump must never run dry.
If there is a positive suction lift, open the valve on the suction side of the pump and evacuate
the air through the vent plug (5, 25 in rear cover in ESL40/50-180N). If there is no positive
suction lift, the plant must be equipped with a foot valve or a non-return valve on the suction
side.
Prime the pump while the vent plug (5, 25 in rear cover in ESL40/50-180N) is kept open.
When the pump has been filled with water, and the air has been evacuated, the system
can be started. Check the direction of rotation. If the direction is incorrect interchange 2 of
the power cords. The direction of rotation is indicated by an arrow. Make sure the bearings
are running correctly and that they do not become warm.
A considerably smaller delivery head than expected will increase the quantity of liquid
delivered, causing increased power consumption and perhaps cavitation in pump and piping.
In the pump the impeller may show signs of heavy erosion caused by cavitation (corrosion)
which may at times render an impeller unfit for use in a very short time. Not unusually do
similar erosions occur in pipe bends and valves elsewhere in the piping system.
Therefore, after start-up, it is necessary to check either the quantity of liquid delivered or the
power consumption of the pump e.g. by measuring the current intensity of the connected
motor. Together with a reading of the differential pressure the quantity of water delivered can
be determined against the characteristics of the pump.
Should the pump not function as intended, please proceed according to the fault-finding list.
Bear in mind, though, that the pump was carefully checked and tested at the factory and that
the majority of faults stem from the piping system
The pump uses too 1. Counter-pressure too low Insert orifice plate or check
much power 2. The liquid is heavier valve/Contact DESMI
than water Contact DESMI
3. Foreign body in pump
Dismantle the pump, remove the
4. Electric motor is cause
running on 2 phases Check fuses, cable connection,
and cable
The pump makes 1. Cavitation in pump Suction lift too high/ Suction line
noise wrongly dimensioned/Liquid
temperature too high
- Before inspection of a pump without guard check that the pump cannot be started
unintentionally.
- The repairman must be familiar with the type of liquid which has been pumped as well
as the safety measures he is to take when handling the liquid.
11. REPAIRS
The following parts are exposed to wear: Shaft seal (22,522 in ESL40/50-180N), impeller
(2)/(7)/(5), wear rings (16)and (17) (not including in ESL25-180 and ESL40/50-180N)and the
ball bearings of the motor.
When motor and pump have been separated, the wear ring (17) in the pump casing (1) can
be pulled out.
In order to replace the upper wear ring (16) it is necessary first to dismantle the impeller. Press
the new seal rings into place with an even push all over the face of the ring. Normal diameter
difference between seal ring and impeller 0.3-0.4 mm.
When the impeller has been dismantled, remove the key (19) and then the stop ring (21) (Not
applicable for ESL 25-180 and ESL40/50/80/100-180N, ESL65-180M) which is locked with a
pointed screw. The sealing element itself can now be pulled off the shaft.
Remove the seat ring from the top piece by pressing with a pair of screwdrivers or the like
under the collar off the seat. If the seat cannot be removed, dismantle the top piece from the
motor to permit pressing from the back.
Grease the new seat ring and the inserted 0-ring well with glycerine – not oil – and press it
into place in the casing. Check that the seat is fitted correctly by knocking lightly with a piece
of wood. The seat ring is to be treated very carefully to prevent the sliding surface from being
scratched.
It is necessary to use a conical bush as shown below to prevent the rubber bellows from being
damaged during mounting.
Grease the back of the carbon ring with glycerine to ensure that it stays in its correct place
during the mounting. Grease the shaft, the conical bush and the rubber bellows amply with
glycerine.
Press the shaft seal on to the shaft until it resists. Press the driving ring only.
Finally place the stop ring (21) (Not applicable for ESL 25-180 and
When the pump has been mounted, ventilate the pump casing carefully before putting the
pump into operation.
When the pump has been running for 1-2 min., ventilate it again, after which it can start normal
operation.
Do not forget to check the direction of rotation if there is a risk that the power cords to motor
have been exchanged.
The ball bearing in the front bearing of the motor can be replaced without disassembling the
pump.
If the ball bearing in the rear bearing of the motor is to be replaced, it is necessary to
disassemble the pump and dismantle the pump shaft (as described in paragraph 5.1):
Dismantle the top piece from the motor. Remove the pointed screws (11, 311 in ESL40/50-
180N) in the pump shaft (6,206 in ESL40/50-180N). The pump shaft can now be pulled off the
motor shaft.
When assembling the pump again, check the motor shaft for damages in the form of marks,
burrs, or upsetting of the shaft end.
Damages, if any, are to be remedied by means of a file and fine emery cloth. This work has to
be done very carefully, especially on new motors, the shaft ends of which are often damaged
during transport. If this is not done carefully, the throw will be bigger than permitted and/or the
pump shaft will scratch the motor shaft when fitted.
Normally, the pump shaft is to go relatively smoothly on the motor shaft (manually or by means
of light strokes with a plastic hammer).
DESMI Pumping Technology A/S 17
www.desmi.com
Tagholm 1
9400 Nørresundby – Denmark
Tel: +45 96 32 81 11 Fax: +45 98 17 54 99
E-mail: desmi@desmi.com
Tighten the pointed screws (the one in the middle first) in accordance with the torque indicated
in the table and finally check that the throw stays within the indicated limits by means of an
indicator clock.
The following working pressures (pressure in piping incl. the pressure increase caused by the
pump) and number of revolutions are allowed in standard pumps.
Max. Max.
working working pressure[bar]
Pump Max.RPM
pressure[bar] Bronze SG-iron
ESL25-180 /Cast
16iron 30 3600
ESL40-180N 16 30 3600
ESL40-180 16 30 3600
ESL50-180N 16 30 3600
ESL50-180 16 30 3600
ESL65-180M 10 16 3600
ESL65-180 10 16 3600
ESL80-180N 10 16 3600
ESL80-180 10 20 3600
ESL80-180B 10 20 3600
ESL80-180E 10 16 3600
ESL100-180N 10 16 3600
ESL100-180 10 16 3600
The max. working pressure for NiAlBz and stainless steel pumps is 1.5 times max. working
pressure for bronze (RG5).
The above-mentioned max. working pressure is a design value – delivered pumps are
pressure tested according to actual application requirements and actual flange standards.
For instance the above-mentioned max. working pressure is NOT valid for pumps approved
by a classification society. Pumps approved by classification societies have been pressure
tested according to the requirements of these societies, i.e. a test pressure of 1.5 x the
permissible working pressure. The test pressure is stated in the test certificate and stamped
into the discharge flange of the pump.
DESMI Pumping Technology A/S 18
www.desmi.com
Tagholm 1
9400 Nørresundby – Denmark
Tel: +45 96 32 81 11 Fax: +45 98 17 54 99
E-mail: desmi@desmi.com
13. EU DECLARATION OF CONFORMITY
DESMI Pumping Technology A/S, hereby declare that our pumps of the ESL Monobloc type
are manufactured in conformity with the following essential safety and health requirements in
the COUNCIL DIRECTIVE 2006/42/EC on machines, Annex 1.
Pumps delivered by us connected with prime movers are CE-marked and comply with the
above requirements.
Pumps delivered by us without prime movers (as partly completed machinery) must only be
used when the prime mover and the connection between prime mover and pump comply with
the above requirements.
No damage materials are used in DESMI pumps – please refer to DESMI Green Passport (can be sent on
request – contact a DESMI sales office) – i.e. common recycling companies can handle the disposal at
end-of-life. Alternatively the pump and motor can be returned to DESMI at end-of-life for safe
recycling.
1 Pump casing
2 Impeller
02 Plug
3 Gasket
4 Motor bracket
5 Vent plug
6 Shaft
10 Set screw
11 Pointed screw
14 Stud
15 Nut
19 Key
20 Nut
Mechanical
22
shaft seal
23 Washer
25 Cover
26 Plug
27 Washer
28 Guard
96 Manometer
97 Nipple
98 Sleeve
100 Base plate
200 Screw
1 Pump casing
2 Plug
3 O-ring
4 Casing cover
5 Vent plug
6 Shaft
7 Impeller
8 Shaft flange
10 Set screw
11 Pointed screw
14 Stud
15 Nut
16 Wear ring
18 Plug
19 Key
20 Screw
21 Stop ring
Mechanical
22
shaft seal
23 Washer
24 Pointed screw
51 Base plate
52 Screw
96 Manometer
97 Nipple
98 Sleeve
1 Pump casing
2 Plug
3 O-ring
4 Casing cover
5 Vent plug
6 Shaft
7 Impeller
8 Shaft flange
10 Set screw
11 Pointed screw
14 Stud
15 Nut
16 Wear ring
17 Wear ring
18 Plug
19 Key
20 Screw
21 Stop ring
Mechanical
22
shaft seal
23 Washer
24 Pointed screw
51 Base plate
52 Screw
96 Manometer
97 Nipple
98 Sleeve
1 Pump casing
2 Plug
3 O-ring
4 Casing cover
5 Vent plug
6 Shaft
7 Impeller
8 Shaft flange
10 Set screw
11 Pointed screw
14 Stud
15 Nut
16 Wear ring
17 Wear ring
18 Plug
19 Key
20 Screw
21 Stop ring
Mechanical
22
shaft seal
23 Washer
24 Pointed screw
96 Manometer
97 Nipple
98 Sleeve
100 Base plate
200 Screw
1 Pump casing
2 Plug
3 O-ring
4 Casing cover
5 Vent plug
6 Shaft
7 Impeller
8 Shaft flange
10 Set screw
11 Pointed screw
14 Stud
15 Nut
16 Wear ring
17 Wear ring
18 Plug
19 Key
20 Screw
21 Stop ring
Mechanical
22
shaft seal
23 Washer
24 Pointed screw
96 Manometer
97 Nipple
98 Sleeve
100 Base plate
200 Screw
1 Pump casing
2 Plug
3 O-ring
4 Casing cover
5 Vent plug
6 Shaft
7 Impeller
8 Shaft flange
10 Set screw
11 Pointed screw
14 Stud
15 Nut
16 Wear ring
17 Wear ring
18 Plug
19 Key
20 Screw
21 Stop ring
Mechanical
22
shaft seal
23 Washer
24 Pointed screw
96 Manometer
97 Nipple
98 Sleeve
100 Base plate
200 Screw
1 Pump casing
5 Impeller
7 Rear cover
14 Stud
15 Nut
16 Washer
17 Set screw
18 Plug
19 Key
23 Washer
24 Screw
25 Plug
26 Nipple
28 Nipple
29 Pipe
51 Base plate
52 Screw
96 Manometer
97 Nipple
98 Sleeve
104 Motor bracket
108 Motor coupling
206 Shaft W. Flange
311 Pointed screw
522 Shaft seal
703 O-ring
810 Set screw
811 BH-Screw
812 Guard
1 Pump casing
2 Impeller
3 O-ring
4 Rear cover
5 Venting plug
6 Shaft W. Flange
7 Motor bracket
8 Motor coupling
9 Muff
10 Set screw
11 Pointed screw
14 Stud
15 Nut
16 Wear ring
17 Wear ring
18 Plug
19 Key
20 Set screw
21 Sleeve
22 Shaft seal
23 Washer
24 Washer
25 Set screw
96 Manometer
97 Nipple
98 Sleeve
100 Base plate
200 Screw
201 Inter. flange
202 Set screw
811 BH-Screw
812 Guard
D DA DB DC H HA HB HF I J K KA KB L LA LB N x M
Type TN Flange Max. Max. Min. Max. Max.
mm mm mm mm mm mm mm mm mm mm mm mm mm mm mm mm mm
1"
1 ESL25- 180 10 25 115 68 85 754 644 110 60 240 275 220 110 110 300 150 150 4xø14
25 mm
1½"
2 ESL40-180N 10 40 150 90 110 878 778 100 60 285 320 244 116 128 360 180 180 4xø18
40 mm
1½"
3 ESL40- 180 10 40 150 90 110 960 840 120 60 285 320 253 118 135 350 175 175 4xø18
40 mm
4 2"
ESL50-180N 10 50 165 102 125 878 780 98 60 285 320 260 123 137 400 200 200 4xø18
50 mm
2"
5 ESL50- 180 10 50 165 102 125 935 830 105 60 285 320 267 145 122 400 200 200 4xø18
50 mm
2½"
6 ESL65- 180 10 65 185 122 145 878 828 155 100 285 320 295 137 158 440 220 220 4xø18
65 mm
7 2½"
ESL65- 180M 10 65 185 122 145 878 828 155 100 285 320 295 137 158 440 220 220 4xø18
65 mm
3"
8 ESL80- 180 10 80 200 138 160 966 830 136 120 285 320 320 150 170 480 240 240 8xø18
80 mm
9 3"
ESL80- 180N 10 80 200 138 160 990 830 160 120 285 320 310 140 170 480 240 240 8xø18
80 mm
10 3"
ESL80- 180B 10 80 200 138 160 966 830 136 120 285 320 320 150 170 480 240 240 8xø18
80 mm
11 3"
ESL80- 180E 10 80 200 138 160 990 830 160 120 285 320 310 140 170 480 240 240 8xø18
80 mm
12 4”
ESL100- 180 10 100mm 100 220 158 180 1003 833 170 140 285 320 350 160 190 540 270 270 8xø18
13 4”
ESL100- 180N 10 100mm 100 220 158 180 1003 833 170 140 285 320 350 160 190 540 270 270 8xø18
Table with thread sizes of manometer, drain and vent plug hole (if any) on the next page.
Table of Contents
Abbreviation and Denomination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Warning Levels Used in Manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Handling and Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Service Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Mounting Recommendations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Bleeder Valve Positions on LPU. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Use with other Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
More Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
1
Reference Manual Damcos LPU: Table of Contents
RM 8000-100-01 Ed 01 July 2020
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .155
2
Damcos LPU: Abbreviation and Denomination Reference Manual
July 2020 RM 8000-100-01 Ed 01
DI Digital In
DO Digital Out
FC Fail Close
FO Fail Open
HP Hand Pump
3
Reference Manual Damcos LPU: Abbreviation and Denomination
RM 8000-100-01 Ed 01 July 2020
4
Damcos LPU: Handling Reference Manual
July 2020 RM 8000-100-01 Ed 01
Handling
A description of issues concerning
5
Reference Manual Damcos LPU: Handling
RM 8000-100-01 Ed 01 July 2020
6
Damcos LPU: Warning Levels Used in Manuals Reference Manual
July 2020 RM 8000-100-01 Ed 01
Warning! Indicates a potentially hazardous situation which, if not avoided, could result in death or seri-
ous injury.
Caution! Indicates a potentially hazardous situation which, if not avoided, may result in property damage.
Warnings
All personnel that handles Damcos LPU shall take part of and understand the following information prior to han-
dling the system. See warnings on next page.
7
Reference Manual Damcos LPU: Safety
RM 8000-100-01 Ed 01 July 2020
Safety
8
Damcos LPU: Handling and Storage Reference Manual
July 2020 RM 8000-100-01 Ed 01
9
Reference Manual Damcos LPU: Commissioning
RM 8000-100-01 Ed 01 July 2020
Commissioning
Qualified commissioning is crucial for the service life and functional safety of the hydraulic system.
Note! LPU-Ex has it’s own User Manual placed inside the electrical housing.
Please read this manual carefully before installing LPU Ex.
Preparation Instruction
Before first time starting up the system, LPU, actuator and possible hydraulic piping must be fully de-aired.
All electrical cables to LPU must to be connected.
Note! If the LPU is connected with hydraulic piping, please note the following:
This will influence the required tank size and the system performance, due to compressibility of the oil in
piping and pressure drop in piping.
Several other parameters influence the system; Temperature, Flow, Valve torque, Actuator type.
It may be required to use larger diameter piping for some LPU, or adjust the valve speed to be quite low.
In case you have trouble with the LPU system performance, please consult Emerson.
Before installation check pipe length from LPU to actuator. Check below if maximum pipe length and LPU tank
capacity match. The values below are for piping ID 8 mm. for larger pipe diameter, max pipe length must be
reduced accordingly to keep the same pipe volume
KF+KF KF+KFR
Tank size (ml) KC 65 KC 125 KC 250 KC 325 KC 400 KF 65
R 125 250
10
Damcos LPU: Commissioning Reference Manual
July 2020 RM 8000-100-01 Ed 01
Commissioning Procedure
1. First check the power supply to the substation that controls the LPU. Voltage must correspond to LPU.
(Please be aware, that switching power to an LPU-Power will make the LPU run for 7 seconds. If on/off relays
has been used, all LPU-Power will run 7 seconds or more simultaneously after switching power to substation).
Check cable entry in cable glands and that cables are fastened in the glands. Minimum bending radius for
cable must be OK.
2. If refill is required, refill with a hydraulic oil, as specified for the system if required. The LPU is normally pre-
filled with Shell Tellus S2V32 (earlier Tellus T 32), for deck mounting Shell Tellus S4VX (earlier Tellus Arctic
32).
See “Refill of Oil in LPU” on page 89.
3. Open valve.
If the valve has not started opening within 10 seconds, open bypass valve at actuator for 5 seconds and close
firmly again. If the valve still doesn’t start opening within 10 seconds, start trouble shooting.
4. At fully open position, the pressure will rise to 150 bar and the LPU will run at that pressure for an individual
extra time, depending partly on the valve operating time. If it doesn’t stop, check position indicator sig-
nal, pressure switch signal and electrical connection. For bulkhead mounted units check oil level again. If oil
level is low the system is not properly de-aired.
5. Check position indication while the valve is fully open.
6. Close the valve again. If LPU becomes noisy (more than by pressure rise) during opening while the actuator
stops moving before full opened position, check oil level again.
7. Check position indication while the valve is fully closed. If position indication has to be adjusted, adjust in
closed position and check in open position. If continuous position indication has to be adjusted at circuit
board (Power Control), adjust in closed position first, then in open position.
8. Open / close the valve several times checking opening/closing speed.
9. Check for external oil leakage.
10.After 1 week or after operating valve several times, check for external oil leakage.
11
Reference Manual Damcos LPU: Commissioning
RM 8000-100-01 Ed 01 July 2020
• Oil level is controlled by measuring the distance from cap to piston with closed or fully opened actuator, see
picture above.
Oil level must be 30-32 mm for 250 tank and 27-29 mm for 1000 tank.
• If oil refill is required, close the valve, connect pressure connection from hand pump to B-quick connection
for LPU.
Note! See “Measuring Oil Level in LPU Level Tank” on page 81for complete instruction on how to measure
the oil level.
12
Damcos LPU: Commissioning Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! Fail-Open actuators must be set in CLOSED position, and the solenoid valve kept activated while
leaving the mode with blue LED, by activating magnetic input LOCAL once. LED turns steady white.
Solenoid valve stays activated.
4. Activate the ADJUST mode and See Adjust Mode in User Manual.
the position indicator is adjusted
in closed valve position.
Note! ADJUST may not be needed when a PCB has been replaced without changing or adjusting the posi-
tion indicator.
13
Reference Manual Damcos LPU: Service Area
RM 8000-100-01 Ed 01 July 2020
Service Area
There should be enough space around the actuators to ensure easily access and service.
To be able to perform emergency operation by emergency key or hand pump there must also be enough space
around the actuator enabling this operation.
All Measurements in mm
14
Damcos LPU: Service Area Reference Manual
July 2020 RM 8000-100-01 Ed 01
All Measurements in mm
All Measurements in mm
15
Reference Manual Damcos LPU: Service Area
RM 8000-100-01 Ed 01 July 2020
16
Damcos LPU: Mounting Recommendations Reference Manual
July 2020 RM 8000-100-01 Ed 01
Mounting Recommendations
The LPU is especially developed for mounting directly at Damcos valve actuators types BRC 125-16000, BRCF
125-16000, KC 65-400, KF 65-250 and KFR 125-250. Mounting at actuator types BRC 32000, KC 600 and BRCF
HT is possible, but not recommended due to runtime capabilities.
Before direct mounting at small pipes the strength of pipes should be considered. Alternatively the LPU could be
bulkhead mounted instead of mounted directly on actuator.
LPU can be placed at almost all environment onboard a ship, but attention must be paid to hazardous areas and
submersion. The LPU can be short time flooded (3 bar for 24 hours or 1,5 bar for 72 hours), but permanent sub-
mersion must be avoided.
For LPU placed on deck or in very corrosive environments a special coating must be applied to LPU and actuator.
Environmental temperature for LPU is normally -5°C to +70°C and with special oil -25°C to +70°C. Oil viscosity
must be considered depending on environmental temperatures.
17
Reference Manual Damcos LPU: Bleeder Valve Positions on LPU
RM 8000-100-01 Ed 01 July 2020
Valve bleed positions on the LPU. The lower pictures shows the bleed valve position on the hydraulic block.
LPU Version 3 are equipped with an automatic bleeder valve function. The automatic bleeder function enables
the LPU to be installed in all directions without disabling the unit. This will ease the installation and reduce the
number of opening points and hence minimize the risk of introducing dirt into the system.
Due to this new tank bleed design the LPU bleed valve is placed at fixed location and is not to be relocated
depending on actuator and LPU orientation.
Exceptions
For horizontal placed BRCF actuator the bleed valve at top of BRCF should be placed at top position, see picture
above.
18
Damcos LPU: Bleeder Valve Positions on LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
If you have an LPU of older design (version 1 or 2), then it is extremely important that bleeder valve position is cor-
rectly chosen when mounting LPU with actuator on pipe. Other positions are plugged when not used! This is nor-
mally done at installation.
Note! LPU version 3 has automatic bleeder function and the position of the bleeder valve will be the same
regardless to position of the LPU.
For KC/KF/KFR bleed valve on top of actuator has no connection to LPU tank and must always be pres-
ent!
19
Reference Manual Damcos LPU: Use with other Equipment
RM 8000-100-01 Ed 01 July 2020
Caution!
When using the portable hand pump with a double acting LPU the pipe pressure must be
released before remote operation.
For a single acting LPU there is no restriction.
Usage of Portable Hand Pump where LPU is Hydraulically Connected via Pipes
There are cases where the portable hand pump must be connected remotely to the LPU via pipes (block). This
implicates that the access to the LPU is difficult and that there will be no local feedback. See pictures below.
In these cases pipe connectors including stop and quick connectors are recommended to be added to the LPU.
DPI inside
20
Damcos LPU: Use with other Equipment Reference Manual
July 2020 RM 8000-100-01 Ed 01
Bulkhead Mounted
If valve and actuator are placed in a way so mounting LPU directly on actuator is not possible (due to narrow space
for LPU or actuator placed in hazardous or submerged area etc.) LPU can be bulkhead mounted on a B-block, and
hydraulically connected to actuator with pipes.
If the position indicator is located at the actuator (DPI or NAMUR type) with electrical cable to LPU. A cable is then
connected between position indicator and LPU.
If a Volumetric Position Indicator is used, a BB-VPI block is used, serving both as B-block and VPI.
By default the BB-VPI has internal wire connection to LPU, so no external cables are added.
When using VPI with LPU, pipe length should be no more than 30 meters each, since no compensation blocks can
be used.
Mounting on Actuator
LPU can be placed directly at Damcos actuators type BRC, BRCF, KC, KF and KFR and Bettis actuators type BHH,
BHHF, BL, BLF and BLFR. Other actuators can be operated using B-block or special mounting block placed at actu-
ator or bulkhead, but please be aware of flow, pressure and maximum runtimes.
21
Reference Manual Damcos LPU: Use with other Equipment
RM 8000-100-01 Ed 01 July 2020
22
Damcos LPU: General Description Reference Manual
July 2020 RM 8000-100-01 Ed 01
General Description
A short description
the system
21
Reference Manual Damcos LPU: General Description
RM 8000-100-01 Ed 01 July 2020
22
Damcos LPU: Description Reference Manual
July 2020 RM 8000-100-01 Ed 01
Description
The LPU - (Local Power Unit) - is an integrated electro-hydraulic system for remote control of valves and actuators
and are especially developed for bulkhead mounting or direct on valve actuators, primarily onboard ships. The
LPU consists of a hydraulic pump which is driven by an asynchronous capacitor motor and several valve functions.
The LPU offers installation in both safe and hazardous area in two basic versions:
• LPU-S (fail safe) for single acting actuator
• LPU-D (fail set) for double acting actuator
The LPU offer is further expanded with six different electrical standard versions all equipped with different levels of
functions and possibilities:
– LPU Basic
– LPU Plus
– LPU Elite
– LPU P-NET
– LPU Power
– LPU Ex
The LPU program offers installation both on the actuator and for separate installation on the bulkhead in safe and
hazardous area and can control both single and double acting actuators.
23
Reference Manual Damcos LPU: More Information
RM 8000-100-01 Ed 01 July 2020
More Information
Installation, Service, Maintenance, Operation and Technical Data can be found in the following documentation:
Data Sheets
• Damcos Local Power Unit - General – SD 1507-2EXX
• LPU-D - Hydraulic Data for Double-Acting Operated Actuator – SD 1502-2EXX
• LPU-S - Hydraulic Data for Single-Acting Spring Operated Actuator – SD 1501-2EXX
• Power Controlled LPU – SD 1503-2EXX
• P-NET Controlled LPU System – SD 1506-2EXX
• LPU Basic, Plus and Elite Controlled System – SD 1507-2EXX
• LPU-S-Ex-d-IIC – SD 1510-2EXX
• LPU-D-Ex-d-IIC – SD 1511-2EXX
Manuals
• Reference Manual LPU – RM 8000-100-01
• User Manual LPU – UM 8000-100-02
• User Manual Aperio – UM 9000-30-001
Contact Information
Emerson Automation Solutions
Damcos A/S
Aaderupvej 41
DK-4700 Naestved
T +45 5578 7200
F +45 5578 7272
www.Emerson.com/marine
24
Damcos LPU: Electrical Product Connection Reference Manual
July 2020 RM 8000-100-01 Ed 01
wiring diagrams
and power
25
Reference Manual Damcos LPU: Electrical Product Connection
RM 8000-100-01 Ed 01 July 2020
26
Damcos LPU: Cable Connection Cover Reference Manual
July 2020 RM 8000-100-01 Ed 01
LPU Basic, Plus and Elite P-NET controlled LPU Power controlled LPU
controlled system
Examples of Cable Connection Cover for different systems.To the left the numbered connections as explained in the dia-
gram below.
Terminal 1 2 3 4 5 6 7 8 9 10 11 12
Basic 24 0 DI DI DO DO Close
VDC VDC Open Close Open
Terminal X1 X2 X3 X4 X5 X6 X7
Basic PE/Ground N(AC) 230 VAC
Power PE/Ground
27
Reference Manual Damcos LPU: Cable Connection Cover
RM 8000-100-01 Ed 01 July 2020
Terminal 1 2 3 4 5 6 7 8 9 10 11 12 J1
Basic Y3 Y2 Y1 Y1
Plus Y3 Y2 Y1 Y1
Elite Y3 Y2 Y1 Y1
Power Y1 Y1
P-NET Y3 Y2 Y1 Y4 Y5 Y6 Y7 Y1
230 V Power:
(chain/loop)
28
Damcos LPU: Cable Connection Cover Reference Manual
July 2020 RM 8000-100-01 Ed 01
Connection Layout
The connection layout for Basic, Plus and Elite are the same for both double and single acting valve.
29
Reference Manual Damcos LPU: Cable Connection Cover
RM 8000-100-01 Ed 01 July 2020
30
Damcos LPU: Encapsulation Top Cover - Internal Wiring Reference Manual
July 2020 RM 8000-100-01 Ed 01
1 2 3 4 5 6 7 8 9 10 11 12
31
Reference Manual Damcos LPU: PCB, Printed Circuit Board - Wiring
RM 8000-100-01 Ed 01 July 2020
32
Damcos LPU: PCB, Printed Circuit Board - Wiring Reference Manual
July 2020 RM 8000-100-01 Ed 01
33
Reference Manual Damcos LPU: PCB, Printed Circuit Board - Wiring
RM 8000-100-01 Ed 01 July 2020
34
Damcos LPU: Cable Glands Reference Manual
July 2020 RM 8000-100-01 Ed 01
Cable Glands
Cable Glands sizes and numbers are decided at order.
There are room for two M20x1.5 and two M25x1.5 on the Cable Connection Cover.
Not used cable entrances will be plugged!
IP Class requirement IP68. For more technical data about cable glands and cables please read the data sheets for
the specific system (see “Data Sheets” on page 24).
35
Reference Manual Damcos LPU: Cable Glands
RM 8000-100-01 Ed 01 July 2020
36
Damcos LPU: General Service Instructions for LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
37
Reference Manual Damcos LPU: General Service Instructions for LPU
RM 8000-100-01 Ed 01 July 2020
38
Damcos LPU: General Service Instructions for LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
General Considerations
Important notes on LPU equipment and systems:
• use original spare parts and tools
• use lifting facilities and transport equipment - see “Handling and Storage” on page 9
• use handling instructions
• electrical cables are only to be connected by qualified personnel
39
Reference Manual Damcos LPU: General Service Instructions for LPU
RM 8000-100-01 Ed 01 July 2020
Maintenance
As LPU is designed for limited running capabilities and has a closed oil reservoir regular maintenance work is not
required. However, if running in major temperature variations the recommendation is to change the oil every 5th
year. Furthermore, if the LPU is used with a BRCF it is recommended to change the oil in case of a breakdown in
the spring housing. In that case the suction filter should be changed as well. LPU Basic, Plus and Elite will raise a
maintenance flag after 7000 operations as a reminder.
A hydraulic oil, as specified for the system (32 cSt40°C, Viscosity Index (VI) > 140, for temperature below -5°C VI >
300) has to be used. The LPU is normally pre-filled with Shell Tellus S2V32 and for deck mounting (on request)
Shell Tellus S4VX. If other oil brand is to be used it must be compatible with above mentioned Shell oil to ensure
proper function of LPU as there normally will be a rest of the old oil in small parts of LPU.
Suction filter is designed to last for a long time, so regular exchange is not required, but at breakdown the oil
could be contaminated and for this reason the suction filter should be exchanged as well.
Note! If green oil is specified, please order the LPUs with Panolin Atlantis. Exchanging mineral oil to green oil
is troublesome, since the LPU cannot be properly emptied without dismantling the LPU.
40
Damcos LPU: General Service Instructions for LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
41
Reference Manual Damcos LPU: General Service Instructions for LPU
RM 8000-100-01 Ed 01 July 2020
42
Damcos LPU: Detailed Product Overview Reference Manual
July 2020 RM 8000-100-01 Ed 01
43
Reference Manual Damcos LPU: Detailed Product Overview
RM 8000-100-01 Ed 01 July 2020
44
Damcos LPU: Overview LPU-D Reference Manual
February 2017 RM 8000-100-01 Ed 01
Overview LPU-D
Cable Glands Cable Connection Cover
Encapsulation
Top Part Sight Glass
for LED 1000 ml Tank
w switch
Solenoid PCB
Encapsulation Control
Base Part Block D 250 ml Tank
Capacitor w switch
Motor Small
or Large
250 ml Tank
The LPU-D is designed for the control of double-acting actuators requiring hydraulic pressure for operation in
both opening and closing the valve. The LPU-D can be mounted either directly on actuator or remotely on bulk-
head.
In the modular design of the LPU the Base Hydraulic Block acts as the rack and the Control Block as the heart of
the hydraulic function. For LPU-D these blocks are named Base Hydraulic Block D and Control Block D.
The LPU standard versions comes with several electrical alternatives with different levels of possibilities and func-
tions, such as LPU Basic, Plus, Elite and P-NET.
For further information about the different versions, please see separate PDS.
45
Reference Manual Damcos LPU: Overview LPU-D
RM 8000-100-01 Ed 01 February 2017
A and B
Pump
Quick Connection for
Open/Close
emergency operation
Speed adjustment
B to T
Tank filling valve
DPI
Plug for adjusting
Position Indicator and
easy exchange of DPI
Control Block D
SV Solenoid Valve
46
Damcos LPU: Overview LPU-D Reference Manual
February 2017 RM 8000-100-01 Ed 01
47
Reference Manual Damcos LPU: Hydraulic Functions
RM 8000-100-01 Ed 01 February 2017
Hydraulic Functions
Solution
When activating motor and solenoid valve the actuator/valve will open. When valve is fully opened motor will be
stopped and a few seconds later solenoid valve will be de-activated. When activating motor without activating
solenoid valve the valve will close. When valve is fully closed motor will be stopped.
When motor is de-activated the valve will be kept in position by means of hydraulic lock.
When emergency opening or closing by hand pump the oil will be pumped from hand pump to actuator opening
port and return oil from actuator will be led back to hand pump tank so no oil is added to LPU system during
emergency operation. When hand pump is not activated any longer the position will be kept by means of hydrau-
lic lock.
When closing valve after emergency opening this can be done in 2 ways: Turn portable hand pump direction valve
to close and pump to close or by remotely closing the valve.
Small BRC actuators can alternatively be emergency operated by means of handle to turn shaft directly after
opening of by-pass valve.
Benefits
• Remote operation by activating/deactivating motor and solenoid valve
• Emergency operating by use of firmly mounted hand pump or portable hand pump
• Remote control after hand pump operation will automatically be engaged
• No risk of oil emptying due to wrong handling of emergency operation
• Possibility of pressure switch to ensure “keep-closed function”
• Release valves to avoid major pressure increase due to heating of oil
48
Damcos LPU: Hydraulic Functions Reference Manual
February 2017 RM 8000-100-01 Ed 01
Background
It will be of advantage to adjust position feedback without opening the electrical compartment.
Solution
Activating the ADJUST function the internal LED will help adjusting position indicator by use of a combination of
LED light and blinking. Position indicator can thus be mechanically adjusted to the right level.
Benefits
Easy adjustment of position indicator without opening cover of LPU.
49
Reference Manual Damcos LPU: Hydraulic Functions
RM 8000-100-01 Ed 01 February 2017
Background
The LPU is developed with several safety functions and equipment to secure a smooth and safe use at all times,
following regulations and handling instructions.
A firmly mounted hand pump is used for emergency operation.
A portable hand pump is used for emergency operation or filling of oil into LPU and Actuator.
Solution
All LPUs are provided with quick connections for connection of a portable hand pump for emergency operation of
the valve.
Benefits
• Quick connections on front of the LPU for the portable hand pump
• The firmly mounted HP-LPU can be used with all double-acting Damcos actuators such as BRC and KC actua-
tors
• The firmly mounted hand pump offers fast emergency operation without need of any other remote facilities
• Safe operating of hand pump without risk of draining oil out of the unit
• Portable hand pump can be used for filling of oil into LPU and Actuator
• LPU Elite offers a possibility to set LPU into Local mode prior to manual operation - thus issuing a warning to
the remote system, and blocking the remote control.
50
Damcos LPU: Operating LPU-D 3 Reference Manual
July 2020 RM 8000-100-01 Ed 01
Operating LPU-D 3
1. Tank for hydraulic oil
2. Suction filter
3. Electrical motor
4. Pump safety valve
5. Solenoid valve
6. Double pilot operated check valve, opening
side
7. Tank filling valve
8. Actuator relief valve, opening side
9. In-line filter
10.Quick connection for hand pump opening
with filter
11.Double-acting actuator
12.Quick connection for hand pump closing
with filter
13.In-line filter
14.Actuator relief valve, closing side
15.Double pilot operated check valve, closing
side
16.Pressure switch (D2 or D3)
17.Variable displacement pump
To Open Valve
When the motor and solenoid valve are activated the oil is sucked from tank through the suction filter to the
pump and pumped through the solenoid valve and the pilot operated check valve (6) to the actuator port B. This
causes the actuator to open the valve. The oil from actuator port A flows back through the pilot operated check
valve (15) (which is opened by the pressure in the B-line) and returns through the solenoid valve to the tank.
When the valve is fully open, the pressure rises to 150 bar, which causes the pump safety valve to open so that the
oil flows back to tank. The motor and the solenoid valve are then de-energized.
The actuator is now hydraulically locked in position by the pilot operated check valves.
Close Valve
Closing the valve follows exact the same procedure, except that the solenoid valve is not activated which causes
the ports A and B to be reversed.
When the motor is running the direction of oil- flow is solely determined by the activation of the solenoid valve.
51
Reference Manual Damcos LPU: Operating LPU-D 3
RM 8000-100-01 Ed 01 July 2020
52
Damcos LPU: Emergency Operation LPU-D Version 3 Reference Manual
July 2020 RM 8000-100-01 Ed 01
53
Reference Manual Damcos LPU: LPU-D, Options for Double Acting Actuator
RM 8000-100-01 Ed 01 July 2020
All LPU-D are equipped with quick connections for Emergency operation by Portable hand pump.
All Plus and Elite units can, if magnets are ordered, be locally operated by magnets whenever power to
the unit is intact 230 VAC and 24 VDC.
There is no option to have a separate hand pump bulkhead mounted (HP should always be mounted
either with Actuator or with LPU).
54
Damcos LPU: LPU-D, Options for Double Acting Actuator Reference Manual
July 2020 RM 8000-100-01 Ed 01
Bulkhead mounted LPU with HP and VPI feedback (left) and external (DPI) feedback at actuator (right).
55
Reference Manual Damcos LPU: Emergency Operations with Hand Pump
RM 8000-100-01 Ed 01 July 2020
Emergency Opening
1. Connect hand pump P-line to B port on LPU-D.
2. Connect hand pump T-line to A port on LPU-D.
3. Pump to open.
4. Disconnect hand pumphand pump.
Emergency Closing
1. Connect hand pump P-line to port A on LPU-D.
2. Connect hand pump T-line to port B on LPU-D.
3. Pump to close.
4. Disconnect hand pump.
56
Damcos LPU: Emergency Operations with Firmly Mounted Hand PumpReference Manual
July 2020 RM 8000-100-01 Ed 01
Note! LPU Elite offers a possibility to set LPU into Local mode prior to manual operation – thus issuing a warn-
ing to the remote system, and blocking the remote control.
57
Reference ManualDamcos LPU: Emergency Operations with Firmly Mounted Hand Pump
RM 8000-100-01 Ed 01 July 2020
58
Damcos LPU: Overview LPU-S Reference Manual
July 2020 RM 8000-100-01 Ed 01
Overview LPU-S
Encapsulation
Top Part 1000 ml Tank
Sight Glass with switch
for LED
Bracket
Solenoid for PCB
Capacitor
Motor Small
or Large 250 ml Tank
Base Hydraulic
Block S
Pump Cover External
Feedback
DPI
The LPU-S is designed to control spring closing or spring opening actuators by mounting either directly on actua-
tor or remotely on bulkhead. Oil pressure is used for opening the valve by pressing the springs together. Closing
valve by means of mechanical springs.
In the modular design of the LPU the Base Hydraulic Block acts as the rack and the Control Block as the heart of
the hydraulic function. For LPU-S these blocks are named Base Hydraulic Block S and Control Block S.
The LPU standard versions comes with several electrical alternatives with different levels of possibilities and func-
tions, such as LPU Basic, Plus, Elite and P-NET. For further information about the different versions, please see sep-
arate PDS.
Note! The LPU-S has a built-in quick closing function and therefore does not need an external QC-Block.
For fast closing of LPU-S, see section about Quick Closing Function.
61
Reference Manual Damcos LPU: Overview LPU-S
RM 8000-100-01 Ed 01 July 2020
B to T
Tank filling or
Pump spring release
Opening
Speed adjustment
DPI
Plug for adjusting
Position Indicator and
easy exchange of DPI
Control Block S
SV Solenoid Valve
62
Damcos LPU: Overview LPU-S Reference Manual
July 2020 RM 8000-100-01 Ed 01
63
Reference Manual Damcos LPU: Hydraulic Functions
RM 8000-100-01 Ed 01 July 2020
Hydraulic Functions
Solution
The built-in quick closing function is used to block flow back to the LPU tank during closing and to control oil flow
directly back to actuators A-port.
Benefits
• Better control of oil flow within the LPU-S due to the blocking function of oil back to the actuator during clos-
ing
• Faster and safer closing due to a larger closing oil flow feasibility and hence speed and safety is increased
64
Damcos LPU: Hydraulic Functions Reference Manual
July 2020 RM 8000-100-01 Ed 01
Solution
The control block is equipped with functions to ensure remote control and emergency handling. Contains quick
closing function to allow faster closing of valves.
When activating motor and solenoid valve the actuator/valve will open. When valve is fully opened motor will be
stopped, but solenoid valve is still activated to keep valve in open position. When de-activating solenoid valve the
valve will close due to internal springs in actuator.
When emergency operating by hand pump the passage to solenoid valve is blocked and opening of actuator by
means of hand pump will then be possible. Return oil from actuator will be led back to hand pump, so no oil is
added to LPU system during emergency operation.
When closing valve after emergency opening this can be done in 3 ways: Turn portable hand pump direction valve
to close or open by-pass valve at actuator or LPU or start remote opening for few seconds and then remotely close
the valve.
Contains quick closing function to allow faster closing of valves.
Benefits
• Remote operation by activating/deactivating motor and solenoid valve.
• Emergency opening by use of firmly mounted hand pump or portable hand pump
• Valve closing after emergency opening by use of hand pump, by-pass valve or remote operation
• No risk of oil emptying due to incorrect handling of emergency operation
• Possibility of pressure switch to ensure “keep-open function” or “keep-close function”
• Release valve to avoid major pressure increase due to heating of oil
65
Reference Manual Damcos LPU: Hydraulic Functions
RM 8000-100-01 Ed 01 July 2020
Background
It will be of advantage to adjust position feedback without opening the electrical compartment.
Solution
Activating the ADJUST function the internal LED will help adjusting position indicator by use of a combination of
LED light and blinking. Position indicator can thus be adjusted to the right level.
Benefits
Easy adjustment of position indicator without opening cover of LPU. Simply by means of a magnet and an 8mm
Allen key, you can mechanically adjust the position indicator - On/OFF or Analogue. No electrical connections to
be made, no exposure to water or dirt.
66
Damcos LPU: Hydraulic Functions Reference Manual
July 2020 RM 8000-100-01 Ed 01
Background
The LPU is developed with several safety functions and equipment to secure a smooth and safe use at all times,
following regulations and handling instructions.
A firmly mounted hand pump is used for emergency operation.
A portable hand pump is used for emergency operation or filling of oil into LPU and Actuator.
Solution
All LPUs are provided with quick connections for connection of a portable hand pump for emergency operation of
the valve.
Benefits
• Quick connections on front of the LPU for the portable hand pump
• The firmly mounted HP-LPU can be used with all single-acting Damcos actuators such as BRCF and KF actua-
tors
• The firmly mounted hand pump offers fast emergency operation without need of any other remote facilities
• Safe operating of portable hand pump without draining the LPU for oil
• Portable hand pump can be used for filling oil into LPU and Actuator
67
Reference Manual Damcos LPU: Hydraulic Functions
RM 8000-100-01 Ed 01 July 2020
68
Damcos LPU: Operating LPU-S Reference Manual
July 2020 RM 8000-100-01 Ed 01
Operating LPU-S
1. Tank for hydraulic oil
2. Suction filter
3. Electrical motor
4. Pump safety valve
5. Non-return valve
6. Pressure switch (S2 or S3)
7. Shuttle valve
8. In-line filter
9. Quick connection for hand pump,
opening with filter
10.Fail safe actuator, single-acting
11.By-pass / tank filling valve
12.Quick connection for hand pump suction
13.Non-return valve
14.Actuator relief valve
15.Solenoid valve
16.Throttle valve / closing speed adjustment
17.Pilot operated non return valve
18.Variable displacement pump
To open Valve
To move the valve towards open, the motor (3) is activated.
The oil is led from tank through the pump and through the non-return valve (5), directly to the actuator B port. To
prevent the oil from flowing back to tank, the solenoid valve (15) must be energized. When the valve is fully open,
the pressure rises to 150 bar which causes the pump safety valve (4) to open and the oil flows back to tank. The
motor is de-energized.
The actuator is now hydraulically locked in position by the solenoid valve.
67
Reference Manual Damcos LPU: Operating LPU-S
RM 8000-100-01 Ed 01 July 2020
Close Valve
To move the valve towards closed, the solenoid valve is de-energized. The springs then move the actuator, press-
ing the oil back from the actuator B port, through the throttle valve (16) and the solenoid valve (15) to the LPU
tank.
68
Damcos LPU: Emergency Operation LPU-S Version 3 Reference Manual
July 2020 RM 8000-100-01 Ed 01
69
Reference Manual Damcos LPU: LPU-S, Options for Single Acting Actuators
RM 8000-100-01 Ed 01 July 2020
All spring loaded actuators can be emergency operated to set the valve in safe position by opening the
actuator bypass valve manually, releasing the spring package.
All Plus and Elite units can, if magnets are ordered, be locally operated by magnets whenever power to
the unit is intact 230 VAC and 24 VDC.
There is no option to have a separate hand pump bulkhead mounted (HP should always be mounted
either with Actuator or with LPU).
Direct mounted LPU (BA) permanently mounted at actuator KF for emergency operation with DPI feedback. See
above to the right.
70
Damcos LPU: Emergency Operation LPU-S Version 3 Reference Manual
July 2020 RM 8000-100-01 Ed 01
Bulkhead mounted LPU with HP and VPI feedback (left) and external (DPI) feedback at actuator (right).
71
Reference Manual Damcos LPU: Emergency Operations with Hand Pump
RM 8000-100-01 Ed 01 July 2020
Emergency Opening
1. Connect hand pump P-line to B port on LPU-S.
2. Connect hand pump T-line to A port on LPU-S.
3. Pump to open.
4. Disconnect hand pump.
72
Damcos LPU: Emergency Operation LPU-S Version 3 Reference Manual
July 2020 RM 8000-100-01 Ed 01
73
Reference ManualDamcos LPU: Emergency Operations with Firmly Mounted Hand Pump
RM 8000-100-01 Ed 01 July 2020
74
Damcos LPU: Actuator Speed Adjustment on LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
LPU-D
Tools
• 4 mm Allen Key
75
Reference Manual Damcos LPU: Actuator Speed Adjustment on LPU
RM 8000-100-01 Ed 01 July 2020
LPU-S
Tools
• 4 mm Allen Key
76
Damcos LPU: Actuator Speed Adjustment on LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Tools
• 4 mm Allen Key
• 11 mm Wrench
77
Reference Manual Damcos LPU: Actuator Speed Adjustment on LPU
RM 8000-100-01 Ed 01 July 2020
Note! It is recommended to run a LEARN or TIME procedure after speed adjustment of LPU. These procedures
are found in the User Manual.
78
Damcos LPU: Pump Pressure Adjustment for LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Tools
• 5 mm Allen Key with ball head
The key most be long enough to reach
down to hexagon in Base Hydraulic block,
see picture next page.
• Pressure Gauge with adapter.
79
Reference Manual Damcos LPU: Pump Pressure Adjustment for LPU
RM 8000-100-01 Ed 01 July 2020
Caution! Pump pressure must be in the range of 90-150 bar. Adjustment beyond this range is possible for
testing purposes, but is not allowed for working actuators.
80
Damcos LPU: Measuring Oil Level in LPU Level Tank Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! Just measuring the oil level in standard tank is not possible. A portable hand pump must be used, and
the tank is re-filled according to the description in “Refill of Oil in LPU” on page 89.
81
Reference Manual Damcos LPU: Measuring Oil Level in LPU Level Tank
RM 8000-100-01 Ed 01 July 2020
Measure level of oil in LPU tank with switch using portable hand pump
82
Damcos LPU: Measuring Oil Level in LPU Level Tank Reference Manual
July 2020 RM 8000-100-01 Ed 01
83
Reference Manual Damcos LPU: Measuring Oil Level in LPU Level Tank
RM 8000-100-01 Ed 01 July 2020
84
Damcos LPU: Emptying Oil from LPU Reference Manual
March 2019 RM 8000-100-01 Ed 01
Tools
• Hand pump
To be connected to B port
• Allen key - 4 mm
To open valve B to T
• Wrench
To remove the Bleeder valve (level tank)
• Level meter, such as a caliper
To measure oil level in the LPU tank
85
Reference Manual Damcos LPU: Emptying Oil from LPU
RM 8000-100-01 Ed 01 July 2020
86
Damcos LPU: Emptying Oil from LPU Reference Manual
March 2019 RM 8000-100-01 Ed 01
87
Reference Manual Damcos LPU: Emptying Oil from LPU
RM 8000-100-01 Ed 01 July 2020
88
Damcos LPU: Refill of Oil in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Caution! The cleanliness class of the oil must be of NAS 9 or better. Make sure that no dirt or foreign particles
can enter the system when refilling oil into the system. Keep caps on quick connections mounted in
order to keep out dirt from the oil in the system.
Tools
• Hand pump
To be connected to B port
• Allen key - 4 mm
To open valve B to T
• Wrench
To remove the Bleeder valve (level tank)
• Level meter, such as a caliper
To measure oil level in the LPU tank
89
Reference Manual Damcos LPU: Refill of Oil in LPU
RM 8000-100-01 Ed 01 July 2020
Refill oil in LPU with 250 ml tank (standard) using portable hand pump
90
Damcos LPU: Refill of Oil in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Refill oil in LPU with tank with switch using portable hand pump
91
Reference Manual Damcos LPU: Refill of Oil in LPU
RM 8000-100-01 Ed 01 July 2020
92
Damcos LPU: Replace Pump in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Warning! Disconnect the Power from the LPU before continue replacing the pump!
Note! This procedure requires the LPU tank to be emptied from oil.
Tools
• Hand pump
To be connected to B port
• Allen key
To open valve B to T
• Wrenches
To remove screws and Bleeder valve (level tank)
• Level meter
To measure oil level in the LPU tank
• Blunt tool
To break the cover on the spare pump
95
Reference Manual Damcos LPU: Replace Pump in LPU
RM 8000-100-01 Ed 01 July 2020
96
Damcos LPU: Replace Pump in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
7b
8. Remove stator.
97
Reference Manual Damcos LPU: Replace Pump in LPU
RM 8000-100-01 Ed 01 July 2020
98
Damcos LPU: Replace Pump in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
99
Reference Manual Damcos LPU: Replace Pump in LPU
RM 8000-100-01 Ed 01 July 2020
100
Damcos LPU: Replace Pump in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
101
Reference Manual Damcos LPU: Replace Pump in LPU
RM 8000-100-01 Ed 01 July 2020
102
Damcos LPU: Replace Hydraulic Block in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! This procedure requires the LPU tank to be emptied from oil.
Tools
• Hand pump
To be connected to B port
• Allen keys
To open valve B to T and for open and close LPU cover
101
Reference Manual Damcos LPU: Replace Hydraulic Block in LPU
RM 8000-100-01 Ed 01 July 2020
Warning! Disconnect the Power from the LPU before continue disassemble the unit!
102
Damcos LPU: Replace Hydraulic Block in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
LPU-D LPU-S
103
Reference Manual Damcos LPU: Replace Hydraulic Block in LPU
RM 8000-100-01 Ed 01 July 2020
104
Damcos LPU: Replace Hydraulic Block in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
105
Reference Manual Damcos LPU: Replace Hydraulic Block in LPU
RM 8000-100-01 Ed 01 July 2020
106
Damcos LPU: Replace DPI in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Caution! When mounting DPI, please leave the DPI adjusting screw protruding 1-2 mm from the block.
If the adjusting screw is tightened too far, the DPI will be crushed!
Tools
• 8 mm Allen keys with ball head (Pin wrench and 3 mm Allen Key for painted LPU)
To remove the DPI plug
• Magnets
To run the LPU locally and initiate Adjust and Calibrate procedures as required
Can be ordered from Damcos
107
Reference Manual Damcos LPU: Replace DPI in LPU
RM 8000-100-01 Ed 01 July 2020
108
Damcos LPU: Replace DPI in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
109
Reference Manual Damcos LPU: Replace DPI in LPU
RM 8000-100-01 Ed 01 July 2020
110
Damcos LPU: Replace DPI in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
DPI-C
111
Reference Manual Damcos LPU: Replace DPI in LPU
RM 8000-100-01 Ed 01 July 2020
112
Damcos LPU: Replace DPI in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
113
Reference Manual Damcos LPU: Replace DPI in LPU
RM 8000-100-01 Ed 01 July 2020
114
Damcos LPU: Replace Motor in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! This procedure requires the LPU tank to be emptied from oil.
Tools
• Hand pump
To be connected to B port
• Allen keys
To open valve B to T and for open and close LPU cover
Caution! Please be aware that if other ID number of motor is used, capacitor might be required to be
exchanged as well!
To replace capacitor please see “Replace Motor Capacitor in LPU” on page 123.
117
Reference Manual Damcos LPU: Replace Motor in LPU
RM 8000-100-01 Ed 01 July 2020
Warning! Disconnect the Power from the LPU before continue disassemble the unit!
118
Damcos LPU: Replace Motor in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
119
Reference Manual Damcos LPU: Replace Motor in LPU
RM 8000-100-01 Ed 01 July 2020
120
Damcos LPU: Replace Motor in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
121
Reference Manual Damcos LPU: Replace Motor in LPU
RM 8000-100-01 Ed 01 July 2020
122
Damcos LPU: Replace Motor Capacitor in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
Tools
• Allen keys
To open valve B to T and for open and close LPU cover
Caution! Please be aware that different types of capacitors for LPU exist, so check that new capacitor has
same ID number or description as the one to be replaced.
123
Reference Manual Damcos LPU: Replace Motor Capacitor in LPU
RM 8000-100-01 Ed 01 July 2020
Warning! Disconnect the Power from the LPU before continue disassemble the unit!
124
Damcos LPU: Replace Motor Capacitor in LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
125
Reference Manual Damcos LPU: Replace Motor Capacitor in LPU
RM 8000-100-01 Ed 01 July 2020
126
Damcos LPU: Replace Solenoid Valve Coil Reference Manual
July 2020 RM 8000-100-01 Ed 01
Tools
• Hand pump
To be connected to B port
• Allen keys
To open valve B to T and for open and close LPU cover
125
Reference Manual Damcos LPU: Replace Solenoid Valve Coil
RM 8000-100-01 Ed 01 July 2020
Warning! Disconnect the Power from the LPU before continue disassemble the unit!
126
Damcos LPU: Replace Solenoid Valve Coil Reference Manual
July 2020 RM 8000-100-01 Ed 01
127
Reference Manual Damcos LPU: Replace Solenoid Valve Coil
RM 8000-100-01 Ed 01 July 2020
128
Damcos LPU: Replace Printed Circuit Board (PCB) Reference Manual
July 2020 RM 8000-100-01 Ed 01
Tools
• Allen keys
To open LPU cover
131
Reference Manual Damcos LPU: Replace Printed Circuit Board (PCB)
RM 8000-100-01 Ed 01 July 2020
Replace PCB
Warning! Disconnect the Power from the LPU before continue disassemble the unit!
132
Damcos LPU: Replace Printed Circuit Board (PCB) Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! It is recommended to run ADJUST, LEARN or TIME procedure after installation of the new PCB card.
These procedures are found in the User Manual.
133
Reference Manual Damcos LPU: Replace Printed Circuit Board (PCB)
RM 8000-100-01 Ed 01 July 2020
134
Damcos LPU: Replace 250 ml Tank Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! This procedure requires the LPU tank to be emptied from oil.
Tools
• Allen keys
To open tank and tank parts.
135
Reference Manual Damcos LPU: Replace 250 ml Tank
RM 8000-100-01 Ed 01 July 2020
136
Damcos LPU: Replace 250 ml Tank Reference Manual
July 2020 RM 8000-100-01 Ed 01
137
Reference Manual Damcos LPU: Replace 250 ml Tank
RM 8000-100-01 Ed 01 July 2020
138
Damcos LPU: Replace 250 ml Tank Reference Manual
July 2020 RM 8000-100-01 Ed 01
139
Reference Manual Damcos LPU: Replace 250 ml Tank
RM 8000-100-01 Ed 01 July 2020
140
Damcos LPU: Replace 250 ml Level Tank (with Switch) Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! This procedure requires the LPU tank to be emptied from oil.
Tools
• Allen keys
To open tank and tank parts.
139
Reference Manual Damcos LPU: Replace 250 ml Level Tank (with Switch)
RM 8000-100-01 Ed 01 July 2020
140
Damcos LPU: Replace 1000 ml Level Tank (with Switch) Reference Manual
July 2020 RM 8000-100-01 Ed 01
Note! This procedure requires the LPU tank to be emptied from oil.
Tools
• Allen keys
To open tank and tank parts.
141
Reference Manual Damcos LPU: Replace 1000 ml Level Tank (with Switch)
RM 8000-100-01 Ed 01 July 2020
142
Damcos LPU: Troubleshooting Guide LPU Reference Manual
July 2020 RM 8000-100-01 Ed 01
143
Reference Manual Damcos LPU:
RM 8000-100-01 Ed 01 July 2020
144
Damcos LPU: Trouble Shooting Reference Manual
July 2020 RM 8000-100-01 Ed 01
Trouble Shooting
Valve/Actuator/LPU Bypass valve at actuator, Check all bypass valves and make
don't move after block or LPU open. sure they are closed!
receiving remote
command. AUX in is activated. Make sure that the valve can be
operated, then turn OFF AUX IN.
LPU is in LOCAL mode. Make sure that the valve may be See Setting and Con-
operated, then turn OFF LOCAL trolling LPU in LOCAL
mode. mode in User Man-
ual.
No electrical power or too Find reason for no power and See “Commissioning
low voltage (AC or DC). connect power to LPU. Procedure” on
page 11.
No hydraulic oil. Check oil level in tank and possi- See “Measuring Oil
bly refill. Level in LPU Level
Look for leakage. Tank” on page 81.
Solenoid valve defect, not Check power to solenoid valve, See “Replace Sole-
activated or not de-acti- possibly change coil or complete noid Valve Coil” on
vated. valve. page 125.
See “Change Sole-
noid Valve”.
145
Reference Manual Damcos LPU: Trouble Shooting
RM 8000-100-01 Ed 01 July 2020
Valve/Actuator/LPU Bypass valve at actuator, Check all bypass valves and make
don't move after block or LPU open. sure they are closed!
receiving local com-
mand (magnet) AUX in is activated. Make sure that the valve is opera-
ble, then turn OFF AUX IN.
No electrical power or too Find reason for no power and See “Commissioning
low voltage (AC or DC). connect power to LPU. Procedure” on
page 11.
No hydraulic oil. Check oil level in tank and possi- See “Refill of Oil in
bly refill. Look for leakage. LPU” on page 89.
Solenoid valve defect, not Check power to solenoid valve, See “Replace Sole-
activated or not de-acti- possibly change coil or complete noid Valve Coil” on
vated. valve. page 125.
146
Damcos LPU: Trouble Shooting Reference Manual
July 2020 RM 8000-100-01 Ed 01
Double acting Quick connector not firmly Check that quick connectors are See “Emergency
Actuator/LPU Emer- mounted. clean and fasten cap. Operation LPU-D Ver-
gency operation is sion 3” on page 53.
not working
Bypass valve at actuator, Check all bypass valves and make
block or LPU open. sure they are closed.
Single acting Actu- Quick connector not firmly Check that quick connectors are See “Emergency
ator/LPU Emer- mounted. clean and fasten cap. Operation LPU-S Ver-
gency operation is sion 3” on page 69.
not working
Bypass valve at actuator, Check all bypass valves and make
block or LPU open. sure they are closed.
147
Reference Manual Damcos LPU: Trouble Shooting
RM 8000-100-01 Ed 01 July 2020
Oil leaking from From BRCF spring cham- Check if piping corresponds to See “Commission-
bleed valve ber: wrong pipe dimen- dimensioning rules. ing” on page 10.
sioning.
Leakage from bleed valve Release the pressure from tank, See “Refill of Oil in
in main block: Oil filling has then fill at lower pressure. LPU” on page 89.
taken place with too high
pressure.
Bleed valve defective. For level tank change bleed See “Replace 250 ml
valve, for suction tank change Tank” on page 135.
tank.
Wrong setup of pipe Check rules for pipe mounted See “Commission-
mounted hand pump. hand pump. ing” on page 10.
Hydraulic system over- Release a little oil from system See “Bleeder Valve
filled. and check again. The purpose of Positions on LPU” on
bleed valve is to release excess oil page 18.
to avoid too high internal pres-
sure in case of heating, decom-
pression etc.
Need for repeated Leakage from BRCF spring Check if piping corresponds to See “Commission-
oil filling chamber: wrong pipe dimensioning rules. ing” on page 10.
dimensioning.
Wrong setup of pipe Check rules for pipe mounted See “Commission-
mounted hand pump. hand pump. ing” on page 10.
Bleed valve defective. For level tank change bleed See “Replace 250 ml
valve, for suction tank change Tank” on page 135.
tank.
148
Damcos LPU: Trouble Shooting Reference Manual
July 2020 RM 8000-100-01 Ed 01
Noise from pump Natural noise due to pres- Please be aware that sudden
surization of hydraulic oil. change of oil pressure that
always happen in a hydraulic
pump will make noise.
Compare with other LPU.
Low oil level. Check oil level and refill. See “Measuring Oil
Level in LPU Level
Tank” on page 81.
Pump piston sticking. Clean area for pump and replace See “Replace Pump
pump. in LPU” on page 95.
Ball bearing defective. Clean area for pump and See “Replace Pump
replaced bearing. Check if pump in LPU” on page 95.
should be replaced too due to
contamination of parts.
149
Reference Manual Damcos LPU: Trouble Shooting
RM 8000-100-01 Ed 01 July 2020
Actuator/LPU is Wrong speed adjustment. Adjust opening and closing See “Actuator Speed
moving too slow speed (see MaxTimeError as Adjustment on LPU”
well). on page 75.
Too low oil level. Check oil level and refill tank. See “Measuring Oil
Level in LPU Level
Tank” on page 81.
150
Damcos LPU: Trouble Shooting Reference Manual
July 2020 RM 8000-100-01 Ed 01
Motor overload error Motor Duty/cycle is too high Move valve less often when
to be continued continu- in regular use.
ously or motor has run too
long compared to the nomi- Check oil level. See “Measuring Oil
nal runtime. Level in LPU Level
Tank” on page 81.
Position error Setpoint and actual position Check LPU functions, adjust See ADJUST proce-
differ more than accepted position indicator. dure in User Manual
and possibly calibrate
analog position indi-
cator.
Max time Error Operation time (opening or Check LPU functions. See TIME procedure
closing) > max time limit. in User Manual.
See “Actuator Speed
Adjustment on LPU”
on page 75.
151
Reference Manual Damcos LPU: Trouble Shooting
RM 8000-100-01 Ed 01 July 2020
Motor high current Too high motor or pump Reestablish correct power
error torque, defective motor or supply, check voltage while
capacitor, too high AC Volt- LPU is running.
age.
Check power supply and See “Commissioning
LPU functions. Procedure” on
page 11.
Oil level low alarm by Low oil level in the LPU tank! Refill oil tank. See “Refill of Oil in
level tank indicator. LPU” on page 89.
Low DC voltage 24 VDC supply voltage below Reestablish correct power See Local Operation
limit. supply, check voltage at and Feedback in User
LPU. Manual.
Pressure error Pressure not achieved or has Check LPU system for: See Local Operation
fallen -and cannot be rebuilt – tightness and Feedback in User
by AutoCorrection. – bypass valves closed Manual.
– oil level
– de-airing
– pressure switch function
– sufficient ExtraTime
Temperature low LPU temperature below low Consider changing to Arctic See “Maintenance”
limit inside encapsulation. oil. Avoiding repeated alarm on page 40.
requires changed factory
setting of minimum tem-
perature
Maintenance recom- Maintenance interval Call for Emerson service See “Reset Mainte-
mended exceeded (operation count). engineer to check all LPU nance Flag” in User
and reset the indicator. Manual.
Min time error Operation time < minimum Check oil level, possibly See “Measuring Oil
time limit. refill and de-air LPU. Level in LPU Level
Tank” on page 81.
152
Damcos LPU: Trouble Shooting Reference Manual
July 2020 RM 8000-100-01 Ed 01
Position indicator Position indicator electrical Adjust the position indica- See ADJUST proce-
error warning. (analog signal level tor. dure in User Manual.
is out of range or OPEN and
CLOSED switches are both Subsequently calibrate ana- See Calibrating Ana-
closed). log position indicator. logue Position Indica-
tors in User Manual.
Pressure switch error Pressure switch does not go Check the pressure switch See “Replace Hydrau-
OFF at low pressure, or the connections and the resis- lic Block in LPU” on
switch cannot be detected by tance of the switch (2K2 page 101.
means of the internal resis- when OPEN).
tor.
Analog input error Analog 4-20 mA output out Check the control output
of range or bad connection. signal through LPU <20V.
Analog output error 4-20 mA current loop error. Check the output current.
Oil level sensor error Oil level switch signal level Check connections.
not in range (disconnected).
Replace the level tank. See “Replace 250 ml
Tank” on page 135.
Sensor error Climate chip communication Disconnect DC power for See “Replace Printed
(RH and T) error. one second. Circuit Board (PCB)”
If the error persists on page 131.
exchange the PCB.
153
Reference Manual Damcos LPU: Trouble Shooting
RM 8000-100-01 Ed 01 July 2020
154
Damcos LPU: Index Reference Manual
February 2017 RM 8000-100-01 Ed 01
Index
A HP-LPU ................................................................ 20
Actuator speed adjustment ................................. 75 Hydraulic Block replacement in LPU ........... 101, 107
Hydraulic Functions
LPU-D ............................................................ 48
B
LPU-S............................................................. 62
Base Hydraulic Block D ........................................ 46
Base Hydraulic Block S ......................................... 60
Bettis actuators ................................................... 20 I
Bleeder valve positions ........................................ 18 Internal Connections – Terminal to Terminal ....... 28
Bulkhead mounting............................................. 20
L
C LPU and Optional Equipment ............................... 41
Cable Connection Cover ...................................... 27 LPU Setup for Not Configured LPU (type0) ........... 13
Cable Glands ....................................................... 35 LPU-D .................................................................. 45
Check of oil level....................................... 82, 87, 91 LPU-S ................................................................... 59
Commissioning ................................................... 10
Contact Information............................................ 24 M
Control Block D ................................................... 46 Maintenance........................................................ 40
Control Block Function Motor Capacitor replacement in LPU ................. 121
LPU-D ........................................................... 48 Motor replacement in LPU ................................. 115
LPU-S ............................................................ 63 Mounting on actuator.......................................... 20
Control Block S .................................................... 60 Mounting Recommendations .............................. 17
Customer Cable Connection................................ 27
N
D NAMUR................................................................ 20
Damcos actuators ............................................... 20
Description.......................................................... 23
O
DPI ...................................................................... 20
Oil and refilling instruction................................... 89
Oil in LPU ............................................................. 89
E Oil level check.......................................... 82, 87, 91
Emergency Operation of LPU-D ........................... 53 Operating LPU-D.................................................. 51
Emergency Operation of LPU-S............................ 69 Operating LPU-S .................................................. 67
Empty LPU of oil .................................................. 85 Overview
Encapsulation Top Cover - Internal Wiring ........... 31 LPU-S............................................................. 59
Overview of LPU-D ............................................... 45
F
Feedback Adjustment Function P
LPU-D ........................................................... 49 PCB replacement ............................................... 129
LPU-S ............................................................ 64 Printed Circuit Board - Wiring............................... 32
Pump pressure adjustment for LPU ...................... 79
G Pump replacement in LPU .................................... 93
General Considerations ....................................... 39
Q
H Quick Closing Function LPU-S .............................. 62
Handling instruction.............................................. 9
Hand pump ......................................................... 20
155
Reference Manual Damcos LPU: Index
RM 8000-100-01 Ed 01 February 2017
R
Recommended refill oil........................................ 89
Refill of Oil ........................................................... 89
Refill of Oil in LPU................................................. 89
Replace 1000 ml Level Tank (Pressure Tank) ...... 141
Replace 250 ml Level Tank (Pressure Tank) ........ 139
Replace 250 ml Tank (Pressureless) ................... 133
Replace Hydraulic Block in LPU .................. 101, 107
Replace Motor Capacitor in LPU ......................... 121
Replace Motorin LPU ......................................... 115
Replace Printed Circuit Board ............................ 129
Replace Pump in LPU ........................................... 93
Replace Solenoid Valve Coil ............................... 125
S
Safe hand pump Function
LPU-D............................................................ 50
LPU-S ............................................................ 65
Service Area......................................................... 14
Solenoid Valve Coil replacement ....................... 125
Speed adjustment on LPU.................................... 75
Storage instruction................................................ 9
T
Tank Level Switch
LPU-D............................................................ 47
LPU-S ............................................................ 61
Tank replacement.............................. 133, 139, 141
Transport instruction............................................. 9
Troubleshooting Guide...................................... 143
W
Warning Levels ...................................................... 7
Warning messages ................................................ 7
156
Emerson Automation Solutions © 2020 Emerson. All rights reserved.
Damcos A/S
The Emerson logo is a trademark and service mark of Emerson Electric Co. Damcos and the
Aaderupvej 41 Damcos logotype are trademarks of Damcos A/S, Rosemount and the Rosemount logotype
DK-4700 Naestved are registered trademarks of Rosemount Inc. Damcos A/S is a member of the Emerson Process
T +45 5578 7200 Management family of companies. All other marks are the property of their respective owners.
F +45 5578 7272
www.Emerson.com/marine The contents of this publication are presented for information purposes only, and while effort
has been made to ensure their accuracy, they are not to be construed as warranties or guaran-
tees, expressed or implied, regarding the products or services described herein or their use or
applicability. Standard Terms and Conditions of Sale can be issued by contacting Damcos A/S.
We reserve the right to modify or improve the designs and specifications of our products at
any time without notice.
Damcos A/S accepts no responsibility for any errors that may appear in this publication.
※ Actuators that can be opened by emergency operation actuator : BRC125 - BRC500
3. Open crossover valve at the actuator and turn the key counter clockwise to open valve an clockwise
to close valve.
Manual
Techcross
Confidential Document
i
i 페이지 11/30/2020
Date of preparation Document No. Version
2020-10-14 TC31-20-PTS-001 Rev.2
Prepared by Reviewed by Approved by QC
Park, Jihun Lee, Gwangho Kwon, Kyungahn
Subject Document type
Pocket TRO Sensor Manual Technical document
Table of contents
1. Overview ....................................................................................................................... 3
1.1. Revision History ............................................................................................................ 3
2. Product specification .................................................................................................... 3
2.1. Specification of Reagent .............................................................................................. 4
2.2. Drawing ......................................................................................................................... 4
3. Safety ............................................................................................................................ 4
3.1. Safety information labelling ......................................................................................... 5
4. Product component ...................................................................................................... 5
4.1. Keypad .......................................................................................................................... 6
4.2. Display ........................................................................................................................... 7
5. Measuring principle ...................................................................................................... 7
5.1. Analytical method ......................................................................................................... 7
5.2. Measurement method .................................................................................................. 7
5.3. Self-test ......................................................................................................................... 8
5.4. Calibration ..................................................................................................................... 8
6. Maintenance ................................................................................................................. 8
6.1. Clean the instrument .................................................................................................... 8
6.2. Clean the measurement cell ........................................................................................ 8
7. Troubleshooting ............................................................................................................ 9
Confidential Document
ii
ii 페이지 11/30/2020
Date of preparation Document No. Version
2020-10-14 TC31-20-PTS-001 Rev.2
1. Overview
This manual explains Pocket TRO Sensor (Hereafter, referred to as ‘PTS’) that measures TRO(total residual oxidant)
content in the water. The manual’s contents are correct by thorough inspection or repair. However, they can be a
little bit different from the product if it is improved.
2. Product specification
Table 1. Product specification
10 Operation temperature 0 ~ 55 ℃
2.2. Drawing
3. Safety
This manual explains the basic guidelines that must generally be implemented in the product including the
installation, operation, and maintenance. Please make sure to read all the contents of the manual before unpacking,
installing, or operating the product. Please make sure to pay attention to all kinds of danger and warnings. Please
be informed that noncompliance may result in user or equipment damage.
4. Product component
Product components consist as below.
1) Storage case
2) Main body
4.1. Keypad
1
3
Key Contents
Set zero value before a measurement.
1. Zero key Push and hold for 2 seconds to enter or exit menu mode.
In menu mode and selected menu, scrolls down.
2. Power key Set the power to on and off.
Start a sample measurement.
3. Measurement key In menu mode, selects menu option shown.
In selected menu, scrolls up or selects option.
4.2. Display
2 3 4
5. Measuring principle
5.1. Analytical method
PTS is equipment that measures TRO in the water. It uses the DPD colorimetric method that does DPD(N,N-diethyl-
p-phenylenediamine sulfate) reagent. When a pack of reagent is mixed with a sample water, they are changed to
red in proportion to concentration of TRO in the water. It measures and digitizes the concentration by using
LED(Sources of light) and inform users of them. TRO concentrations are calculated by differences in absorbance.
5.3. Self-test
Check the LED, Photosensor status before calibration.
Note: If the result is Need calibration, do 5.4. Calibration. If the result is Good status, just use it as it is.
5.4. Calibration
If you want to calibrate, do 5.3. Self-test first.
6. Maintenance
6.1. Clean the instrument
Clean the exterior of the instrument with moist and soft cloth and then wipe the instrument dry as necessary.
Note: Take care to avoid scratches on the interior and exterior surface of the measurement cell.
7. Troubleshooting
Error Description Solution
There is ambient light in the cell holder. Identify the cap fully
E-00 Ambient light error
installed into the cell holder.
The zero measurement is too low. Check the degree of
E-01 Zero-measurement error
turbidity. If it is not clear, fill clean water again.
E-02 No Zero error Measure a blank solution to set the zero.
Identify if there is blockage in the cell. If present, clean the
E-03 Blockage error
cell holder. If there is not any blockage factor, test again.
The reading is less than minimum range of the instrument.
E-04 Low range error
Check sample water and measurement cell and test again.
The reading is more than maximum range of the instrument.
E-05 High range error
Check sample water and measurement cell and test again.
E-06 Low battery power The voltage of battery is too low. Replace the batteries.