Professional Documents
Culture Documents
LABORATORY PROCEDURES
by
Jim Sequeira
498 Water Treatment
TABLE OF CONTENTS
Chapter 11. LABORATORY PBOCEDURES
Page
OBJECTIVES 500
woBDS..............._.. .. ... ................ 501
LESSON 1
'l
0
1.2'l lmportance of Representative Sampling... ........528
1 1.21 1 Source Water Sampting ...........
'l'1.212 ln-Plant Sampting......................
'11.213 Distribution System Sampting
... ........529
11.22 Types of Samples ...
........530
1 1.220 Grab Samptes........................... ........530
1 1.221 Composite Samptes.................. ........530
11.23 SamplingDevices...
.......530
1 1.24 Sampling Techniques ..............................
.......531
'1
1,240 Surface Samp|ing.....................
53'l
1 1.24't Depth Samptin9........................
531
1 1.242 Water Tap Samptang ...................
533
i j.243 First_Draw Samples....................
533
11.25 Sampting Containers and preservation of Samples ..
533
1 1.26 Reporting................
533
11.27 Additional Reading .
533
DISCUSSION AND REVIEW QUESTIONS
s33
LESSON 3
4. Colilorm Bacteria
544
DISCUSSION AND REVIEW OUESTIONS
565
LESSON 5
5. Hardness................._........
566
6. Jar Test.......-.....................
567
7. pH ...-,....:..,.......................
570
8. Temperature............_......_.
57'l
9. Turbidity.............................
572
ARITHMETIC ASSIGNMENT
574
AODITIONAL READING........
574
574
574
575
500 Water Treatment
OBJECTIVES
Chapter 11. TABOBATORY PFOCEDURES
-{
_r _Ar
6/)'
Lab Procedures 501
WORDS
Chapter 11. LABOHATORY PROCEDURES
ELANK BLANK
A.bottle containin-q only dilutaon water or dislilled water; the sample being tested is nol added. Tests are frequenfly
run on a SAM-
PLE and a BLANK and lhe ditferences are compared. The procedure h6lps to eliminata or reduce test resjlt errors that
could be
caused when the dilulion waler or disttlled waler used is contaminaled.
BUFFER
BUFFEF
A solution or liquid whose chemical makeup neutralizes acids or bases without a great change in pH
BUFFEB CAPACITY
BUFFER CAPACITY
A measure of the capacity of a solution or I iquid to neutralize acids or bases. This is a measure of the capacity of water lor oflering a
resislance to changes in pH
COMPOUND COMPOUND
A pure substance composed of two or more elements whose composition is constant. For example, table salt (sodium chloride,
NaCI) is a compound,
ELEMENT ELEMENT
A substance which cannot be separated inlo its constituent parts and still retain its chemical identity. For example, sodium (Na) is an
element.
GHAVIMETRIC GRAVIMETRIC
A means of measuring unknown concentrations of water quality indicators in a sampla by WEIGHING a precipitats or residue of tho
Sample,
Lab procedures 503
INDICATOB (CHEMICAL)
A substance that gives a visible change,
usually of color; at a desired point in a chemical
reaction, n"""r",,f:'::;:; ::ltfln:l
INOFGANIC
INORGANIC
ffi:T[!il:i"i:""""11; i3,iii,:l;ii,fll[ffhil,',fl:i;Tl"1%;,a5llii:. rnorsanic substances are o, mrnerarorisin,
whereas
INORGANIC WASTE
,i wasre mareriar such as sand, sarr, iron,
carcium, and orher minerar materiars
**'es are chemicar su6"t""""" or.i"".ii
wiici.:re^.Tt s]islrry gtfg*, ,ffijff]:""ffiJn.t
;:I;:?,Ufl""" ";;:il;;;$rg""ic wasres are chemicar subsrances or an animar
or
,It or MOLAR
A molar solulion consists ol one oram molecular it or MOLAFT
werght ol a compound dissolved in enough
water Io make one liter o, solution. A
ej[ I"f:,1?::iJflil' ;,'!p;i;;;f
- -- - '- --'r lf,lr,t "r " "#i,","i, iJii"l Il, ,," ,i"b*[i*iig-h.r ;liruric acio (H.so.)
sorurron. would consist ",",p,", in enouin JitiiiJo'*aieiio
or 98 grams ol H2so. dissorveo
mare one rite, or
MENISCUS (meh-NIS-cuss)
The curved surface ot a column of liquid (warer, MENISCUS
oil, mercury) in a small tube,, when the liquid
with water)' rhe curve rorms a varrev wets the sides ol ths container (as
1i'v6jn rn" .ij6. #'l"J*o ,"'r"rrvi, ir,," ;#;;" ,
""riiri.gi-he ro-p o;'in" iiqiJii reuert"".rlr,,
a meniscus forms in a'measurrng oevice, "i. hi, or upward
l'Jfl:"Jj* or rn" sampte is determined by the bonom
of rhe
WATER
MEFCUFY
(READ (READ
BOTTOM) TOP)
:i. : i
MOLE
MOLECULAR WEIGHT
OHGANIC OHGANTC
Substances that comelrom animal or plant sources. Organic substances always contain carbon, (lnorganic mat€rials are chemical
substances of mineral origin.) Also see INORGANIC.
ORGANISM ORGANISM
Any form ol animal or plant life. Also see BACTEBIA.
REAGENT (re-A-gent)
A pureschemical substance that is used to make new producls REAGENT
or is used in chemical tests to measure, detect,
or examine other sub_
REDUCTTON (re-DUCK-shun)
Beduction is the addilion of hvdroocn BEDUCTION
removal ot oxygen, or the addition of
electrons to an element or compound. under
fi:IHTJJ:ri:'."','J""13fl$"^?iJifr'
*rr"
."#p""-,"i. ,'""iJiJ.ioio ooo,-p,ooucins hydrosen ana'robic
surride (H,s) and orher com-
BEPRESENTATIVE SAMPLE
STANDARD METHODS
STANDARD MErHoDs FoR rHE EXAMtNAfloN oF wArER
AND wASrF!!.111,
can Public Hearth Association A.",i"an wate, wo.is ,i..'".,r'ili ie*wol, zt.";t Eoition. A joint ffi:"T::,Tr:'ff"":
{APHA),
which outlines the acceoled
laboratorir.proceo*". ,."0 io iiiri", eiii,ii,ilii neo"ration gerl
"roii,,"
"rurvi"fi;iil,jl ii1,l_*1,er and wastewaler. Avairabre ,rom American
[1;:H3']:,i:,..?6ii;;3?i;i";:j?iffiffiru*#1,ffi;xJ,","1't'oaozgs oi;";;ii":'ilid;;;['t"ofr.emoers,$rgaso:
STANDABD SOLUTION
A solurion in which the exact concenrration STANDAHD soLUTloN
of a chemicar or compound is known.
STANDAHDIZE
Tocompare with a standard. STANDABDIZE
I
f(l) ln wet chemislrv, to find out the exacl
slrength ol a solution by comparing it with
a slandard of known strength. This information
[ ,^.
rs used to adjust rhe strength by addirg ,"i" ;tJ#.J,""J r"ti'Jrlostrnc" oi.soru"a.
i
' iffi?"'?,iLJ:1fi'fiL?tli:ff#J:fl0,f"','""15,"",g1.,n,s arrows vou ro ad,usr the instrument so thar it
reads accurarery, or
I
sTERtLtzATtoN (STARE_uh_tuh_zAy_shun)
I
TEFllLlzATloN
[]he removal o. destruclion of all microord:nismc in^r,,.ri^^ ^^.!^^^-:^ -,- ,.
' "'-'oorganisms' including pathogenic
Imr, olsrrrrercrior.'r and oiher bacteria, vegetative forms and
spores. compare
I
i SUPERNATANT (sue.per-NAy-tenl)
SUPEBNATANT
luquid removed kom setfled sludoe. Suocrnar:.r commonlv .^,^-^
re,ers .^ .L - ,:, !..
Im,it rr". .rrr""" ^^--^hr, to the liquid between the studge on rhe bottom and th6 scum on
I
"ir"or.in"o";:;;i;j;:f"'n"t"t
isuBFAcrANT (sir-FAc-renl)
for surrace-acrive agent. The active suRFAcTANT
lAbbreviation agent in detergents that possesses a high
creaning ability.
InrRArE
(TlE-trate)
:To flrFArE a sample' ..
a chemical solution oJ known strength
is added drop.by drop untir a certain
TTTBATE
coror change, precrpitate, or pH
#iill"}'lli,i;ffii: iiXffi]:"'"i:H:f:lJX;fl";til*fJ"ffi liul5in! *"'"nu.i;"i;;#,;t iffi;ji i,,:;ements (0 , -i o
506 Water Treatment
VOLUMETRIC VOLUMETRIC
A measurement based on the volume ot some factor. Volumehic titration is a means of measuring unknown concentrations of wat6r
quality indicators in a sample BY DETERMINING THE VOLUME ot litrant or liquid reagent neede-d to complete particular reactions.
i
Lab Procedures 507
CHAPTERll. LABORATORYPROCEDURES
(Lessonlof5Lessons)
I1.O BASIC WATER LABORATOBY PROCEOUBES l\rany limes in the water laboratory we use smaller amounts
than a meler, a lite( or a gram. To express these smaller
11.00 lmportance of Laboraiory procedures amounts, prelixes are added lo the names of the base metric
Water trealment processes cannot be controlled etfeclively unit. There are many pretixes in use; however, we commonly
unless the operatorhas some means to check and evaluate th; use two or three prelixes more than any others in the labora_
quality ol water being treated and produced. Laboratory qualily tory
control tests provide the necessary information to monitor th6
lreatment processes and to ensure a sale and pleasant-tasting Pretix Abbreviation Meonlng
drinking waterfor allwho use it. By relating laboratory resultst;
lreatment operations, the water lreatment or supply syslem c6nti- c 1l'100 oli ot 0.01 times
operalor can first select the most etfective operational proce-
dures, then delermine lhe efficiency of the treatmenl proc- milli- m 1/1,000 oI; or 0.001 times
esses, and idenlify potential problems before they atfect fin- micro- p 1/1,000,000 of; or 0.000001 tim€s
ished walerquality. Forthese reasons, a clear understanding ol
laboratory procedures is a must lor every watenarorks operaior.
Onecentimeter (cm)is 1/100 (on€ hundredth) oIa meter, one
milliliter (mL) is 'll1,000 (ons thousandth) of a titer, and tikewis€,
one microgram (pgm) is 1/1,000,000 (one mi ionth) of a gram.
NOTICE EXAMPLES:
fhe aollection ol a bad san?le or a baa bborat ory (1) Convert 3 grams into milligrams.
rcault i5 abou't aE uselul as no .esullo. To prefinl 1 milligram = 1 mg = 1/1,000 grams
bad results requirest (1) aongtant maintenancc anl
therefore, 1 gram = 1,000 milligrams
calibration ol labontory cquipment, and (2) use ot
correat lab procedures, A!'c, reoultg of lab te1t5 arc (3 gramsx1,000 mq/gram) = 3,OOO mg
of no value to anyone unlesa thcy arc used, (2) Convert 750 milliliters (mL) to lit€rs.
mL = l/1,000 tirer
1
Type ot English Metric Metric Larger amounts than a mete( liter, or gram can be ex-
Measuremenl System Name Abbreviation pressed using such prefixes as kilo- meaning '|,OOO. A kilo-
gram is 1 ,000 grams-
Length inch meter m The Celsius (or centigrade) temperature scale is used in the
foot
yatd
water laboratory rather than the more familiar Fahrenheit
scale.
Temperalure Fahrenheit Celsius
Volume quart
Fahrenhelr (.F) Celsius ("C)
liter L
gallon Freezing point of water 32 O
To convert Celsius to Fahrenheit, the following tormula can Amrnonium Hydroxide NHlOH
be used: Calcium Carbonate CaC03
Temperalure,'F = 9/5("C) + 32.F Chlorotorm cHct3
Copper Sulfate CUSOa
EXAMPLE: Convert 35'C to'F
Ferric Chloride FeC13
Temperature, "F = 9/5("C) i 32'F
Nitric Acid HNO3
= 9/5(35'C) + 32'F Phenylarsine Oxide C6HrAsO
= 63+32 Potassium lodide KI
= 95"F Sodium Bicarbonate NaHC03
'11.02 Chemical Names and Formulas Sodium Hydroxide NaOH
Sulfuric Acid HrSOl
ln the laboratory, chemical symbols are used as shorthand
for the names of the elements. The names and symbols for
some of these elements are listed below. ' 14.3 HrO. Alum in the dryform based on 17./. Alroi.
I
t
5
/_
P/PEIS. Pipets are used to deliver accurate volumes and
range in size from 0..1 mL to 1OO mL.
I !
I
t
_----.'*_.],_Fri=
Pipet
OUESTIONS (pie-PET),
Write your answers in a notebook and then compare your Volumetric
answers with those on page S7S.
11.0A Why are laboratory quatity control tests important?
11.08 What does the prefix milti- mean? ,'.."'.'5,. 0
1:j
o o
Beaker
-l-
7
510 Water Treatment
Burel
(byur-RET)
Kjeldahl Flask Flask.
(KELL-doll) Distilling
r0r
iri
Automatic
Buret
Bottle, Bottle,
Reag6nt BOO
Funnel
Flask,
Boaling
. Flat Bottom
A Buchner lunnel is used to separate solids from a mrxture
It is used with a filter tlask and a yacuum.
,R
()
Funnel.
Flask,
Boiling
Round Bottom
p F:\
':! Buchner
with
Short Neck Perforated Plate
Flask,
Filtsring
Lab Procedures 511
Separaiory funnels are used to separate one chemical OTHER LABWARE AND EQUIPMENT.
mixture. from another. The separated chemical usually is dis-
solved in one or two layers of liquid.
l-)
Separatory Funnel
Condenser
K
r(l::--h,
1,.)->:=!::==r'{
''--ru-rrl!-Y
Dish, Petri
rilmt
Test Tube Culture Tube
without Lip
Desiccator
(DESS-uh-KAY-to0
\l
Color Comparison
Tubes, Nessler
Thermometer, Dial I
512 Waler Treatment
'1
I k*.q
yt*,aa {
i
,l
s:{61
Oven, Mechanical Convection -.-N
Clamp, Dish,
Safety Tongs
*!-. ,
Hot Plate
Clamp, Flask,
Safety Tongs
.+
1
Clamp, Test Tube
,g
..!
a.
I
o
Muffle Furnace. Electric Clamp Holder
qr
Lab Procedures 5.13
7..
!
I
-
I
.t
Clamp, Utitity
Fume Hood
Clamp
Tripod, Concenlric
Ring
Burner. Bunsen
at lD
-tD
v
Triangle,
Fused Portable pH Meter
(Cou.t€sy ol HACH Coinpany)
-l
5 t4 Water Treatment
Crucible (CREW-suh-bull),
Porcelain
Crucible,
Gooch
Porcelain
Pipet Bulb
Dish,
Evaporating
4{ l/
+-s
H fdrion
P" pafl B
5 1
f
E Laboratory Turbidimeter
(ratio mode--€N/OFF-is keypad sel€ctable)
Test Paper, pH 1-11 (Coirriosy ot HACH C{mp.rry)
Lab Procedures 515
i5
1t
( t
Autoclave
lncubator
lPemisio. ot Alue M Elettu)
\
-
ir-
I
DireclReading Colorimeter
(free chlorine residual)
(coudesy ol HAcr Compar'y)
Portable Spectrophotometer
(Coutusy or HACH Comoaiy)
516 Water Treatment
aaa
t
I
a o I
I
I
:\- .!'.-
>
-
0 0 6.
w.ash! =
E
95.5560 am.
Balance, Analytical
(Pemisslon ol M6nl6D
Lab Procedures 517
a E
I
I
I
c
I
t
o
..=i:--
,.8t---
Anperomettic Titrator
11.1'l Use of Laboratory Glassware When a buret is lilled with liquid, lhe surtace ol the liquid is
curved. This curve ol the surface is called the meniscus (melF
BURETS
NIS-cuss). Dependang on the liquid, the curve lorms a valley,
A buret is used lo give accurate measurements of liquid vol- as with wate( ortorms a hill, as with mercury Sjnce most soiu-
umes. The stopcock conlrols the amount of liquid that willflow tions used in the laboratory are water based, always read h6
from the buret. A glass stopcock must be lubricated (slopcock bottom of the meniscus with your eye at the same Isv6l (FtgurB
grease) and should not be used with alkaline solutions. A 1 1 .1 ). lf you have the meniscus at eye level, the closest marks
Teflon stopcock never needs to be lubricated. that go all lhe way around the buret will appear as strsjgrt
lines, not circles.
GRADUATED CYLINOERS
The graduated cylinderor "graduate'is one ol th6 most olten
used pieces ot laboratory equipment, This cylind€r is mad€
either of glass or of plastic and ranges in sizes from 1O mL to4
liters. The graduate is used to measure volumes of liquid with
an accuracy less than burets but greater than beakers or flasks.
Graduated cylinders should never be heated in an open flame
because they will break.
Stopcock
Buret
7- b t_
Lt
'-ot!
t-:
?b?Eg LINE
oF vtrroN 4 : aeu!6cu,h
<9ge9' I
6
Fig. 1 1.1 How to read meniscus
Lab Procedures 519
Volumetric Pipet
(=JErl,_B l
Graduated or Measuring Pipet
Serological Pipet
11.1C Why should graduated cylinders never be heated in Anolher procedure is to place approximately the desired
an open flame? weight in the weighing djsh. Weigh this amount exac y, Then
add a proportionate amount of distilled water.
'11.12 Chemical Solutions
EXAMPLE 1
Many laboratory procedures do nol give the concentrations
of standard solutions in grams/liter or milligrams/liter. lnstead, The directions for preparing a standard reagent indicate that
lhe concentrations are usually given as NORMALITY (N),1 you should weigh out 7.6992 grams and dilute to ona liter. You
which is lhe standard designation lor solution skengths in weigh out 7.5371 grams. How much waler should be added to
chemistry. produce the desired concentration or normality ol ihe standard
reagent?
EXAMPLES:
0.025 N H,SO. means a 0.025 normal solution of sulfuric
acid
2 A/NaOH means that the normality of a sodium hy-
droxide solution is 2
The larger the number in front of the M the more concen-
kated the solution. For example, 1 A/ NaOH solution is more
concenlraled than a 0.2 N NaOH solution.
Another method of specitying the concenlration of solutions
uses lhe 'a + b system." This means that "a'volumes of con-
centrated reagent are diluted with "b" volumes of distilled Known Unknown
water to form lhe required solution.
= 7.6992 gm
Desired Weight, gm Water, mL
EXAMPLES: AclualWeight, gm = 7.5371 sm
1+1 HCI means 1 volume of concentrated HCI is diluted The chemical should be dilut€d to how many milliliters?
with 1 volume of distilled water
(Actual Weight, gm)(l,000 mL)
'l + 5 H"SO. means 1 volume of concentraled sulfuric acid Ditule to mL _
Desired Weight, gm
is diluted with 5 volumes of distilled water
(7.5371 gm)(1,000 mL)
When the exact concentration of a prepared chemical solu-
tion is known, it is referrdd to as a "standard solution." Many 7.6992 gm
times, standard solutions can be ordered already prepared
from chemical supply companies. Once a standard has been = 979 mL
prepared, it can then be used lo standardize other laboratory
The 7.5371 grams oI chemical should bB diluted to 979 mL.
solutions. To standardize a solution means to determine and
adiust its concentration accurately, thereby making it a stand-
ard solution. "Standardization" is the process oI using one so-
1l.13 Titrations
lution of known concentration to determine the concentration A titration involves the measured addition of a standardized
of another solution. This action often involves a procedure solution, which is usually in a buret, to another solution in a
called a "litralionl llask or beaker. The solution in ihe buret is rsferred to as the
' N or Nomal. A normal solution conlarns one gram equivalent weight ol reactant (compound) per liler ol solution. The equivalont weight ol an
acid is that weight which conlains one gram atom of ionizable hydrogen or ils chemlcal equivalent. For sxample, th€ equivalent wgight ol sullurlc
acid (H,SO4) is 49 (98 divided by 2 because there are two replaceabl€ hydrogen ions). A one N solution ol sulluric acid would consist of 49 grams
ol HrSOa dissolved in enough water lo make one liler of solulion.
T Lab Procedures 521
1. Flecord volume
!
ol sample.
I
,/E
is no standard taboratory
lorm. Most operators usu-
.,There
ally develop their own data sheets tor recordtng test results
and olher important data. These data sheets shiould be pre-
<_
pared in a manner that makes it easy for you 3. Add titrant
to record results, until
revrew lhem. and recover these results when jl is necessary. end point is
Each trealment planl wi have diflerent needs tor coffeitinlj reached.
and recording dala and may require several different dala
o-r
worksheets. Figures I 'l .3 and 1 1.4 illustrale two typical labo-
mlory worksheets.
I
522 Water Treatment
SAMPLE SOUFICE
DATE SAI,4PLED: BY
o
F
tr
-,i:
411
)t!!1.",/,
l,/./ W i, L;i
)
tVl 't;;!
t
E o
] - (.) -E ! !
E
i io
E
(.) o
,l.i
,i/) 'ril r{i 'it :-Sl :a-i N'
F ifi r7, *: N .,I lS ::-i 11,:
z F (E
i/,t
:))'!
tfi 'f'lt -ir\ rr L"/, sq)
)t1 tiY \ :15\ S to
ad i
a
=- ltl r<
rt (.)
s \ o
zZ)o
<F o.6
a
o
o-
O( rr
uJ
o
=i ?.)-: L'-l
'!;i w ilt frt #
,/n ',1:+
',i/r
rt t!:, 'iv) l l/ ?i,
I
FE
o
9
I '
z
!F
aoo
)
524 Water Treatment
1. Corrosive lvlaterials the eyeball and irritate the eye, the National lnstitute of 0c-
cupational Safety and Health (NIOSH) recommends an eye
ACIDS injury hazard evaluation be conducted in accordance with
the safety guidelines listed in Current lntelligence Bulletin
a. Examples: Hydrochloric or muriatic (HCl), hydrofluoric (ClB) 59 2005-1 39 (www.cdc.gov/niosh/docs/2005-1
(HF), glacial acetic (CH3COOH), nitric (NHO3), and sut- 394.
furic (H,SOJ.
EASES (Causlics)
a. Examples: Sodium hydroxide (caustic soda or lye,
NaOH), quicklime (CaO). hydrated time (Ca(OH)r), and
alkaline iodine-azide solution (us6d in dissolved oxygen
test).
b. Bases are exlremely corrosive to skin, clolhing, and
leather. Caustics can quickly and permanenty cloud
vision if not immediately flushed out of eyes. Determine
location of safety showers and eye wash stalion before
starting to work wilh dangerous chemicals.
c. Commercially available spill cleanup materials should
be kept on hand for use in the event of an accidental
spill. A jug of ordinary vinegar can be kept on hand to
neutralize bases and it will nol harm your skin.
MISCELLANEOUS CHEMICALS
a. Examples; Alum, chlorine, fenic salts (ferric chloride),
and other strong oxidants.
2. Toxic Materials
Examples:
a. Solids: Cyanide compounds, chromium, orthotolidine,
cadmium, mercury, and other heavy metals. DO NOT PIPET HAZARDOUS
b. Liquids: Chloroform and other organic solvenls. LIOUIDS BY MOUTH.
c. Gases: Chlorine, ammonia, sulfur dioxide, and chlorine
dioxide.
3. Explosive or Flammable Materials 3. Never pipet hazardous materials by your mouth.
Store heavy items on or as near to the floor as possible. 1 1.1 622 PBOPER LABOFATORY TECHNIOUES
y1UTlLE LIQIJIOS' that may escape
as a gas. such as
€lher, musl be kepl away from heat sources, sunlight, and . Faulty technique is on€ of th6 chief caus€s of accidents and,
eleckic switches. Store acids and bases in separate-slorage because it involves the human element, is one of lhe most dif_
cabinets designed for acid and base storage. ficult to correct.
Cap and_secure cytinders of gas in slorage to prevent rolling Because ol their nature and prevalence in the laboralory
orlipping._They shoutd atso be placed awat from any possibl; acids and other corrosive materials conslitute a series of hai_
sources of heat or open llames. ards ranging from poisoning. burning, and gassing through ex_
plosion. Always llush the oulsides ot acid'Oottle-s with "water
belore opening them. Do noi lay the stopper down on the
tl counlertop where a person mjght lay a hand or rest an arm on
iil iii it. Keep all acids tighfly stoppered when not in use and make
lrl sure no spilled acid remains on the floor, table, or bot e atter
lli use. To avoid splashing of acid, do not pour water into acid;
Always pour acid into water.
jelly for ease in ioining connections. Never use oil or grease. When working with chlorine and other toxic substancsq
Wear gloves when making such connections, and hold the always wear a selr-contained breathing apparatus. lt possibls,
tubing as close lo the end being inserted as possible to pre- lry to clear the almosphere with adequate ven lalon bsfom
vent bending or breaking. Also, never try to torce rubber tubing entry
or stoppers from glassware, Cut off the rubber or materials.
WASTE DISPOSAL A good sarety program requires con-
A first-aid kit must be available in the laboratory stant care in disposal of laboralory waste. Corrosive materi
als should never be poured down an ordinary sink or drain.
BURNS. All glassware and porcelain look cold after the red
These substances can corrode away the draln pipe or kap.
from heating has dasappeared. The red is gone in seconds but
Corrosive acids should be neutralized and poured down
the glass is hot enouoh to burn for several minutes. Afler heat-
corrosion-resislant sinks and sew€rs using large quantities ol
ing a piece ol glass, put it out of the way until cold.
water to dilute and flush the acid.
Spattering from acids, caustic materials, and strong oxidiz-
To protect mainlenance personnel, use separato cov€red
ing solutions should be washed otl immediately with large
quantities of water. Every worker in the water laboratory containers to dispose of broken glass.
should have access to a sink and an emergency deluge show-
er. Keep vinegar and soda handy to neutralize acids and )
bases (caustic materials).
Many safeguards against burns are available. Gloves, o
safety tongs, aprons, and emergency deluge showers are but
a few examples. Never decide it is too much trouble to pul on a
pair ol gloves or use a pair of tongs to handle a dish or flask
that has been healed-
0
t
}P
I
t\
-l :{ #,1
r: Ir {
n h
) iiirl : U,I lrll iirtr lr II
electrical equipment; and D, combustible metals. Fire extin- Calilornia Water Pollution Conlrol Association's OPERATORS'
guishers are also classified as A, B, C, or D lo correspond with LABORATORY MANUAL. Some ol the ideas and materlal also
the class of lire each will extinguish. came from lhe FISHEB SAFETY MANUAL.
Class A lires: ordinary combustibles such as wood, paper,
cloth, rubber, many plastics, dried grass, hay, and stubble. 1 1.164 Actdltlonal Beacllng
Use foam, water, soda-acid, carbon dioxide gas, or almost '1. FISHER SAFETY CATALOG. Obtain trom Fisher Scientitic
any lype of exlinguisher. Company, Safety Division,4500 Turnberry Drive, Suite A
Class B lires: flammable and combuslible liquids such as (Cuslomer Service), Hanover park, IL 60133 or phone
gasoline, oil, grease, tar, oil-based paint, lacquer, and sol- (800) 7i 2-6733.
venls, and also flammable gases. Use foam, carbon dioxide, 2. GENERAL INDUSTRY, OSHA. SAFEW AND HEALTH
or dry chemical extinguishers. STANDAqDS (CFFI, Title 29, Labor pt. 1900-1910 (most
Class C fires: energized electrical equipment such as slart- recent edition)). Obtain from the US Gov€rnment Printing
ers. breakers, and motors. Use carbon dioxide or dry chemical Otfice, PO box 979050, Si. Louis, MO 63't97-9OOO. Order
extinguishers to smother the fire; both types are nonconduc- No. 869-062-001 09-6. Price, $61.00.
lors ol electricity. 3. See SAFEry page 1-38, STANOARD METHODS, ZIst
Class D fires: Edition,2005.
combustible metals such asmagnesium,
sodium, zinc, and potassium. Operators rarely encounler this
type oi fire. Use a Class D extinguisher or use fine dry soda OUESTIONS
ash, sand, or graphite to smother the fire. Consult with your
local lire department about the best melhods to use for spe- Write your ansrwers in a notebook and then compare your
cilic hazards that exisl at you. facility. answers with lhose on page 575.
Multipurpose e)dinguishers are also available, such as a 1'1.1G List live laboratory hazards.
Class BC carbon dioxide extinguisher that can be used to 1
'1.1H Why should you not work alone in a laboratoM
smother Class B and Class C fires. A multipurpose ABC
carbon dioxide extinguisher will handle most laboratory fire sit- 11.11 True or False? You may add acid to water, but nevsr
uations. (When using carbon dioxide extinguishers, remem- water to acid.
ber that the carbon dioxide can displace oxygen-take appro-
priate precautions.)
11.'1J How would you dispose of a corosive acid?
At the end of each lesson in this c hapter, you will lind some 3. How are pipets emptied or drained?
dscussion and review questions. The purpose of these ques-
is to indicate to you how well you understand the material 4. How would you litrat€ a test solution?
lhe lesson, Write the answers to these questions in your
5. List as many of the seven basic rules for working in a labo-
mlebook
ratory as you can remember.
, How do the operators of water lreatment plants use th€ re- 6. Why should work with certain chemicals be conducted
sults lrom laboratory tests? under a ventilated laboratory tume hood?
Why must chemicals be properly labeled? 7. Why should waler nol be poured on certain types of firss?
528 Water Treatment
' Reprcsentative Sampl6. A sample portion ot malerial or waler lhat is as nearly identical in contenl and consistency as possible to that in the
larger body ol male.ialor water being sampled.
Lab Procedures 529
Calculate the flushing time for lhe service line in minutes Hov!,evei since operators themselves are often the sampler-
analyzer, continuous analysis would leave little lime for any-
Flushing (Pipe Volume, gal)(2)
Time, min - Flow, gal/min
thing but sampling and testing. Except for tests that cannot
wait due to rapid physical, chemical, or biological changss ol
(1.8s3 galx2) the sample (such as tests for dissolved oxygen, pH, and tem-
perature) a lair compromise may be reached by taking sam-
0.5 GPM ples throughoul the day at hourly or two.hour intervals. Each
sample should be r€frigorated immediaiely after it is col.
= 7.3 min Iected. At the end ol 24 hours, a portion of each sample is
mixed with the other samples. The size of the portion is in
direct proportion io the llow when the sample was collecled
or = 7 min + (0.3 min)(60 sec/min) and the total size of sample needed for testing. For example, il
hourly samples were collecled when the llow was 't.2 MGD.
use a 12-mL portion sample, and when the flow was '1.5 i,,lGD,
= 7 min and 18 sec use a 15-mL porlion sample. The resulting mixture of portions
of samples is called a COMPOSITE S/AMPLE tn no case,
QUESTIONS however, should a composite sample be collected ror bacterio-
logical examination.
Write your answers in a notebook and then compare your
answers with those on page 575. When the samples are taken, they can either be s€t aside to
be combined lat6r or combined as they are collected. ln both
1 1.2A What are frequently the grealest sources of errors in cases, they should be stored at a lemperature ol4"C untilthey
laboratory tests? are analyzed.
11.28 Why must a representative sample be collected?
11.2C How are sampling points sel€cted in a distribution
system?
coRo
. XNOT
RING r*
sroPPER
-/?
BING -------}.
GLASS SAMPLE
BOITLE
WIRE CAGE
5-POUND
r--z?777_ LEAD WEIGHT
Fig. 1 1.5 Sectional view ol homemade depth sampler Fig. 11.6 Distribution system sampling station
11.24 Sampling Techniques from above. Be sure to label the container when sampling is
completed.
11.240 Surtace Sampling
A surface sample is obtained by grasping the sample con- 11.241 Depth Sampling
lainer at the base with one hand and plunging the boftle mouth Several additional pieces of equipment are needed for col-
down into the water to avoid introducing ant materiat floating lection of depth samples from basins, tanks, lakes, and r6s€r-
on the surlace. Position the mouth of lhe bot e into lhe current voirs. These depth samplers require lowering the sample
and away from the hand of lhe collector (see Figure I i.7). lf device and conlainer to the desired water depth, then opening,
lhe waler is not flowing, then an artificial current aan be creat- filling, and closing the container, and returning the device t,o
ed by moving the bottle horizontally in the drrection il is point- the surface. Although deplh measurements are best made
ed and away lrom the sampler. Tip the botUe stighlty upward to with a pre-marked steel cable, the sample depths can be de-
allow arr lo exit so the bottte can fi . Tigh y sto;per and tabel termined by pre-measuring and marking a nylon rope al inter-
lhe botlle. vals with a non-smearing ink, paint, or fingernail polish. One of
Place the bot e in a weighted frame that holds the botfle se- the most common commercial devices is called a Kemmerer
curely when sampling from a walkway or other structure above sampler (see Figure 11.8). This type ot depth sampler con-
a body ol water. Flemove the stopper or lid and lower the sists ol a cylindrical tube that contains a rubber slopper or
device lo the water surlace. A nylon rope, which does not valv€ al each end. The device is lowered into the water in the
absorb water and wtll not rot. is recommended. Face the bot e open position and the waler sample is trapped in the cylinder
moulh upstream by swinging lhe sampling device downstream when the valves are closed by the dropped messenger.
and then allow it to drop into the water, without slack in the Figures 1 1 .5 and 1 1.9 show typicat depth samplers. These
rcp€, Pull lhe sample device rapidly upstream and out of the samplers are low€red to the desired depth. A jerk on the cord
water simulating (imitating) the scooping motion of the hand. will remove the stopper and allow the bot e in the depth sam-
sampling described previously. Take care not to dislodge dirt pler to fill. Good samples can be co ectod in depths ol wat6r
0I olher debris that might fall into the open sample container up to 40 feet (12 m).
532 Water Treatment
I
{
I
i
I
c
()
c
o () I
() ()
o
oe
,l
Fig. 11.8 Kemmerer depth samplet. (A) nyton line, (B) nes-
senger, (C) catch set so that the sampler is open, (D) top
rubbet valve, (E) connecting rod between the valves, (F)
lube body, (G) boftom rubber valve, (H) knot at the boftom ot
the suspension line, and (l) rubber tubing attached to the
sp ri ngloaded ch eck valve.
(Source: US EPA 'MicdDlogical Method. For Monhoiin! Fig. 11.9 Depth sanpler
the ENircn@.ni O.cember 1 978) (Pe.frlsson ol HACH Company)
Lab Procedures 533
11.242 Water Tap Sampling regulatory agency as required by the Safe Drinking Water
Acl.
To collect samples from waler main conneclions, lirst ,lush
lhe service line lor a brief period of time. Samples should not 11.27 Additional Reading
be laken from drinking lounlains, resirooms, or taps that have
aerators. Aerators can change water quality indicators such as 1. See page 1-27, COLLECTION AND PRESERVATION OF
pH and dissolved oxygen, and can harbor bacteda under SAMPLES, STAN DARD M ETHODS, 21st Edition.
some conditions. Do not sample from taps surrounded by ex-
cessive foliage (leaves, flowers) or from taps that are dirty, cor- 2. WATER OUAuW SAMPLING GUIDELINES, Second Edi-
roded, or are leaking. Never collect a sample from a hose or tion, January 2005. Obtaln from American Water Works
any other attachment fastened to a faucet. Care must be taken Association (AWWA), CA-NV Section, 10574 Acacia
to be sure that the sample collector does not come in contact Street, Suite D6, Bancho Cucamonga, CA 91730. Price to
with the laucet members, $32.00; nonmembers, $37.00; price includes
cost of shipping and handling.
11.243 First-Draw Samples
The Lead and Copper Flule calls for lirst-draw or first-flush
QUESTIONS
samples. These are waler samples taken at the custome/s Write your answers in a notsbook and then compare your
tap after the water stands motionless in the plumbing pipes for answers with those on pages 575 and 576.
at leasl six hours. This usually means laking a sample early in
ihe day belore water is used in the kitchen or bathroom. 11.2D What are the two general types of samples collected
by water lreatment personnel?
11.25 Sampling Containers and Preservalion ol Samples 11.2e Lisl three water quality indicators that are usually
The shorter the time that elapses between the actual collec- measured with a grab sample.
lion of the sample and the analysis, the more reliable your re- 11.2F How would you collect a depth sample trom a lake?
sults will be. Samples should be preserved il they are not
going to be analyzed immediately due to remoteness of the 11.2G Samples should not b€ collected ,rom water taps
laboratory or workload. Preservation of some types of sam- under whal conditions?
ples is essential to prevent deterioration of lhe sample. Some
water quality indicators, such as residual chlorine and temper-
11.2H What information should be recorded when a sample
is collected?
alure, require immediate analysis, while others can be pre-
served and lransporled to the laboratory. A summary of ac-
ceplable sample containers, preservative, and maximum time
between sampling and analysis is shown on Table '1 1.2.
Whatever type ol container you use, clearly identify the
fud of L e+e ottZ of 5 l.omow
oyL
sample location, dale and time of collection, name of collector,
and any other pertinent inlormation. L gOgNOeV
'11.26 Reporting wocE?uQE4
The waler system owner (waler utility agency) is respon-
sible for reporting lab resulls at regular frequencies to the Please answer the discussion and review questions next.
TABLE 11.2 RECOMMENDATION FOB SAMPLING AND PRESEBVATION OF SAMPLES ACCOFDING TO MEASUHET/IENT'
PHYSICAL PROPEBTIES
Color 500 P,G Cool,4"C 48 hours
Conductance 500 PG Cool, 4'C 28 days
Hardnessd 100 EG HNO3 to pH <2, 6 months
H,SO4 to pH <2
Odor 200 G only Cool,4"C 6 hours
pHd NG Det. on site lmmediately
Flesidue, Filterable 100 qG Cool, 4"C 7 days
Temperature 1,000 EG Det. on site lrnmediately
Turbidity '100 PG Cool,4'C 48 hours
METALS (Fe, Mn)
Dissolved or Suspended 200 EG Filter on site, 6 months
HNO3 to pH <2
Tolal 100 PG Filter on site. 6 months l
HNO3 to pH <2
I
INORGANICS, NONMETALLICS
Acidity 100 EG Cool, 4'C 14 days
Alkalinity 200 EG Cool, 4'C 14 days
Bromide 100 RG None Req. 28 days
Chloride 50 P,G None Fleq. 28 days
Chlorine, Total Residual 500 BG Det. on site lmmediately
Cyanide, Total and Amenable 500 qG Cool, 4'C, 14 days
to Chlorination NaOH to pH >'12,
0.6 gm ascorbic acid€
Fluoride 300 D None Req. 28 days
lodide 100 EG Cool, 4'C 24 hours
Nitrogen
Ammonia 500 P,G Cool, 4"C, 28 days
H2SO. to pH <2
Kjeldahl and Organic 500 P,G Cool, 4"C, 28 days
HrSO. to pH <2
Nilrale-Nitrite 204 qG Cool, 4"C, 28 days
H,SO{ to pH <2
Nitrate 100 RG Cool, 4'C 48 hours
Nitrite 100 P,G Cool, 4'C 48 hours
Dissolved Oxygen
Probe 300 G with top Det. on site lmmediately
Winkler 300 G with top Fix on sile, store 8 hours
Phosphorus in dark
Orthophosphate 50 P,G Filter on site, 48 hours
Cool. 4'C
Elemental 50 Cool, 4"C 48 hours
Total 50 BG Cool,4'C, 28 days
H,SO4 to pH <2
Silica 50 P Cool, 4'C 28 days
Sulfate 100 P,G Cool, 4"C 28 days
Sulfide 100 EG Cool,4'C, 7 days
add zinc acetate plus
HrSO4 to pH >9
Sullite 50 PG Det. on site lmmediately
"Required Containers, Preservation lechniques, and Holding Times.' CODE OF FEDERAL REGUL,4IONS, Protection of the Environment, 40, parts
136'149, 2007. Ihis publicat;on is available Irom the US Government Printing Oflice, Superintendent of Documenls, pO Box 979050, St. Louis, MO
63197-9000. Order No. 869-062-00160-6. Price, $61.00.
Polyethylene (P) or Glass (G). For melals, potyethylene with a polypropylene cap (no liner) is preferred.
Holding limes lisled above are recommended tor properly preserved samples based on currentli available data- lt is recognized lhat for some
sample types, exlension ol lhese limes may be possible while lor other lypes, lhese limes may be loo tong. Where shippind regulalions prevent
the use of lhe proper preservalion lechnique or the holding time is exceeded, such as the caie of a 24-hour composiid, tn-e fiial reportio oata
for these samples should indicate the specific variance.
Hardness and pH are usually considered chemical properties of water rath6r than physical properties.
Use ascorbic acid only i, residual chlorine is present.
SPECIAL NO|E:whenever you collecl a sample lor a bacteriological test (coliforms), be sure to use a sterile plastic or glass bot €. ll the sample
conlains any chlorine residual, sutticieft sodium thiosulfaie should be added to neutralize all ol th€ chlorine residual. Usually, two d.ops (b.l
mL) of ten percenl sodium lhiosullate lor every 100 mL o, sample is sufficient, unless you are disinfecting mains or storage tanks.
Lab Procedures 53S
There are ,five alkalinity condiljons posstble in a water 1. Take a clean beaker and add 1OO ml ol samDle (or olher
.
ple.(l) bicarbonate atone, (2) bicarbonale and carbonale.sam_t3)
sample volume that will give a titration voluma ol less
caroonate atone, (4) carbonate and hydroxide, and (5) than 50 mL of acid titrant)
hvdroxl
oe arone. Ihese five conditions may be distinguisifrei 2. Place of pH meter into beaker
and containing
quantities determined from the resutrs ot -electrodes
aciO fitritons OyiiiJ sample.
method given below.
3. Stir sample slowly (with a magnetic stirrer if possible).
B. What ls Tested
4. Check pH of sampte. ll pH is 9.3 or betow, then there is
Sample Common Range, mg/L no phenotphthalein alkaljnity present and you can go to
Step 6.
Raw and Trealed Surface Water 20-300 5. ll the pH is grealer than 9.3, hlrate very carelullv to a oH
We Water 80-500 of 8.3 wilh 0.02 ,V H,SO4. Hecord th; arorni ot i6iJ
used to this point.
C. Apparatus Required
6. Continue lo titrate to pH 4.5 with O.O2 N H,SO..6 Flecord
pH meter graduated cylinder (tOO mL) rhe totat amount of acid used lrom starting
reference electrode ioinilo finish.
buret (25 mL)
glass electrode buret support
7. Calculate total and phenotphthalein (if presenl) atkatini-
magnetic stirrer ties.
beaker (250 mL)
magnetic slir-bar wash bottle E Example
analytical balance desiccator
Results rrom atkalinity titrations on a ,inished water
D Reagents were as follows:
sample
,
536 Water Treatment
(Alkalinity)
"' f-".
pHIETEi pH lilETEn
oao olo
,
IEE
1. Add 100 mL ol 2. Place electrodes' of pH 3. Tikate down to pH 8.3 (if
sample. meler in beaker necessary), with
0.02 N H2SO4.
--,>
9H EA
oao '.E
4. Continue lo titrate
to pH 4.5.
. Some pH meters
have a single .combination,, electrode.
Lab Procedures 537
(Chlorine Ftesidual)
G. Calculations
2 99! !1gu 310-i.1 METH1DS FOR CHEMTCAL
Phenolphthalein Alkalinily, mg,/L as CaCO3 SIS OF WATER AND rrylASIES, March - ANAL4-
" 1919.-
_AxNx5O,OOO
ml of Sample QUESTIONS
Write your answers in a nolebook and then
(0.5 mL) x (0.02 N x 50,000 compare your
) ansiwers wilh those on page 576.
x 100 mL
1 1.3A What chemicals used in water treatmenl
will cause
changes in atkatinity?
= 5 mg/L
Total Alkalinity, mg/L as CaCO3 1 1.38 Why is tt important to know the alkatinity ol
a water
sampl6?
_ BxNx50,OO0
mL of Sample 2. Chlorine Bestdual
(6.8 ml ) x (0.02 N ) x 50,000 A. Discussion
100 mL Chlorine is not only an excellent disinfectant
, but also serves
l"-
j"::t_ rlh .iron, manganese, protei,
= 68 mg/L ano many taste_ and odor-producing compounds "rOitu""e.-.-JiiO",
fi H. lnterpretation of Besults
the test results and the informalion given
prove the quatity of treated water. t"
conrrot microorganrsms rhat mioht
and flocculalion. keeps lilter me;ja tree
.t"rti"
to helo im_
iooir.'.,, in[rir" n![J'ii
*iir, Jo-"jri5ioi
,.,From
diff€rent types
below, the of sf ir" gro*t"h;, irj
ol alkalinjty contained in a water sample helps bleach out undesirable color
d€lermined.
can be
There
are two general types o, residual chlorine produced
, in
Alkalinity, mg/L as CaCOJ chlorinated water. They ari: (1) free residual
chlorine en.t r2l
combtned residual chlorine. Free resiOuat ctrtorine
Titration r#rs'-tJ
Result Bicarbonate Carbonate :l:XX, l.p, llq*hrorus. acid fsoclt. a,j r,e iiipociiJ,iil
Hydroxide ,. uomDrnect residual chtorine genera[y
:"Jj-,_y_y, rdfers to the
cnronne-ammonia- compounds of .orocnfor"ri.r,ne (NHrcDl
P=0 T 0 drchtoramine (NHctJ. and trichtoramine
P is less
0 tllcr, ol' nitrijeiiii_
chloride). Both types ol residuals act as dGinfectanis.
f but
lhan !,4 r -2P 2P 0
differ in their capacity to produce a gerr+ree
,raiei s;pi[
P = ,/zT 0 during the same contact time.
2P 0
P is greater
]t,o, a the
I
to positive aspects o, chtonnatron,
lhan th T 0 2r -2P 2P -T _]::9
may be some adverse effects. potentially carcinoge;ic
there
P=T 0 0 T organjc compounds such as chtorotorm ;;;r:
;nO ortreiir_ifr,fs
where P = phenolphthalejn alkalinity be formed.during lhe chlorination pro"""". io,ini.i.! m-av
| = total atkatinity "ii
:9y."T9 "fi"gr:. the operator shouta le ramiriar wirii tir-e-co#
centrattons
Example:The example in ,,G, above gave ol free and combined residual cntorine proOuced
the following resulls
phenolphlhalein atkatinity S mO/L ::^11"a191
Iremety
suo.pty.rg owing chtorinarion. eorh resiOuais
rmportanl jn producing potable ii! e-xl
= a water ttrat is n.,t
sale to dnnk but is also free of objectronaOte
tastes ""i,,
tolal alkalinity = 68 mg/L oOor"l
Since. the phenolphthalein atkalinity (5
B. What ls Tested "nO
mg/t) is tess lhan
one half of lhe toral atkatinity (6S
mg/ti fro; r;"i;;, i;;; Common Flanoe
liere is bicarbonate and carbbnate
atiatinlty in fne watei. Residual Chlorine, mo/L
Source Free -
--Tle.bicarbonate alkalinity in this case is egual to T _ 2p or Total
68 mg/L - (2 x 5 mg/L) = SS mg/L asCaCO" Chlorinated Raw Surface
Tte carbon_ate atkatinity is equat to
mg4 as CaC03.
2p or 2 x S mg/L = 10 Water (prechtorination) 0.3 to 3 0.5 to s
Chlorinated Finished
L Precautions Surface Water
l (Post-Chtorination)
l l. The sample should be analyzed
O.2 lo 1 0.3 to 'l.5
as soon as possible, at Well Water 0.2 to 'l
leasl within a few hours afteicollectron. 0.2 lo 1
l
2. Thesampte should nol be agilated.
C. [4ethods
warmed, fjltered, di_
luted, concentrated, or allerei in any Thereare-eightm€thods listedlormeasuring
way. residualchlorine
nlhe 21sl Edrtron ot
J. Reference
STANDARD METHODi.
mos-l_practicat and appropriate procedure
il;;t"; i;;
;i
in any particuiai
1. rn-starce generally depends upon
See page 2-27, STANDARD METHOD,S, 2.lst Editjon. the characteristics;ithe water
oerng examined. The AMpEROMETB\CT titrarion
mertroO iii
I Arnperomettic (am"p Un.o+,lEt-ricr1.
,| A method ot measurement that records
etectric current . generaled, rather than recording volt-
ol measulng concentrar'tions ot cenaln sutslinceJ-* flowino or
Amperomekic ti lralion is a means .
age.
---"'Y
538 Water Treatment
(Chlorine Residual)
standard of comparison for delerminlng free or combined DPD Colorimstric Msthod (Field Comparator Kit)
chlorine residual. This method is relatively free ol interferences
but does require greater operator skill to obtain good results. ln Use reagents suppliod by kit manufacturer.
addition, Ihe titration instrument is expensive.
DPD Titrimetric Method
The DPD$ methods are simpler to perform than ampero- 1. 1 + 3 H2SO.. Carefully add 10 mL concentrated sulfurjc
metric titration but are subject to interferences due to manga- acid to 30 mL distilled waler. Cool.
nese. Field comparator kits are available from several sup-
pliers such as Orbeco-Hellige, Wallace & Tiernan, and HACH. 2. Phosphate buffer solution.
3. DPD indicator solutaon.
4. Standard ferrous ammonium sulfate (FAS) titrant, O.OO2B2
0
OPD (pronounce as separate leliers). A method of mea€uring the chlorine residual in waler. Th€ residual may be determined
by either titrating
or comparing a developed color wilh color standards. DpD sta;ds for N,N-diethyl-p-phenytene_diamine.
^s Some regulalory
agencies require the use ol the oPD Tilrimetric Metnoo f cntorine iesioual lesting is used in ptace ol some ot th€ colilorm tosts.
Lab Procedures 539
(Chlorine Besidual)
[*,,*,* )!,)t
"n,.,,r,
1. Place 5 mL each of buffer reagent and DPD indicator in a
mL ol sample
Reading A, mL
Fieading B, mL
Beading C, mL -,
00 mL
.4 mL
.6 ml
.7 mL
250-mL flask and mix.
Calculalion
2. Add 100 ml of sample and mix.
READING CHLORINERESIDUAL
3. Tilrate rapidly with standard FAS titrant until red cotor is
discharged (disappears).
A = mg/L kee residual chlorine
B-A = mg/l monochloramine
4. Record amount of FAS used (Fleading A). lf combined re- C-B = mg/L dichloramine
sidual chlorine fractions are desired, continue to next C-A = mg/L combined available chlorine
t slep. C = mg/L lotal residual chlorine
FOR COMBINED RESIDUAL CHLORINE Example: The concenlrations ol the ditterent types of re-
sidual chlorine presenl can be calculated trom the
Monochloramine
inlormalion given in (H).
5. Add one very small cryslal ol Kl and mix. ll monochlor- READING CHLOBINE BESIDUAL
amine is present, the red color will reappear.
1.4mL = 1 .4 mg/L free residual chlorine
6. Continue titrating carefully until red color again disap- B-A = 1.6-1.4 = 0.2 mg/L monochloramine
pears.
C-B = 1.1 mg/L dichloramine
7. Becord reading of FAS in lhe buret used to this point (this C-A = 2.7 - 1.4 = 1.3 mg/L combined available chlorine
includes amount used above). This is Fleading B. 2.7 mL = 2.7 mgy'L total residual chlorine
t oichloramine J. lnterpretalion of Besults
8. Add several crystals Kl (about 1 gm) and mix until dis- 1. Any chlorine taste and odor that may result from chlorina-
solved. tion would generally be from the dichloramine or nitrogen
trichlorid6 nuisance residuals.
9. Lel stand 2 minutes.
2. Free residual chlorination produces the best results when
10. Continue titrating until red color again disappears. Record the lree r€sidual mak6s up more than 80 percent ol the
amount of FAS used to this point. This includes amounts total residual. Howeyer, this will not be the case if am-
used in two previous titrations for lree and monochlor- monia is added to the water being treated to torm chlor-
amine. This is Reading C. amines in ordsr to prevent the formation o, THMs.
iIOIE Manufacturers of laboratory equipment are continu- K. Reference
ally developing faster and more accurate ways to
measure chlorine residual. lf you have new equip- See page 4-56, STANDARD METHODS, 21st Edition.
ment, Iollow the manufacturer's procedures.
G. Precaulions
(Chlorine Besidual)
ML FAS T_
FBEE RESIDUAL
CHLORINE
ML FAS
COMBINED RESIDUAL
CHLORINE
l\,lonochloramine
\ 1
4. Add one crystal Kl
A
5. Titrate with FAS until red
color disappears.
Dichloramine
ML FAS
I 0
(Chlorine Demand)
3. Chlorine Demand stock solution, add 25 mL gl the doslng solution. Note:
A. Discussion ln measuring the volume of the dosing solution, notice
that '1 mL of the 0.025 N thiosullate titrant is equal to
The chlorine demand of water is the difference between the 0.9 mg chlorine.
amount of chlorine app|ed (or dosed) and lhe amount of tree,
combined. or total residual chlorine remaining at the end of a d. Tilrate with standardized 0.025 N thiosulfate titrant
specifjc contact period. The chlorine demand varies with the until the yellow iodine color almost disappears.
amount of chlorine applied, tength of contacl time. pH, and e. Add 1 to 2 mL starch indicator solution.
lemperature. Also, lhe presence of organics and reducing
agents in water will inf,uence the chlorine demand. The chlo: f. Continue to titrate until blue color djsappears.
rine demand test should be conducted with chlorine gas or
wilh granular hypochlorite, depending upon which tor;t you (mL Thiosulfale used) v Nx 35'45
mg/L cl as cl'/ml
' -
usuallv use lor chlorination mL ot Dosing Solution
The chlorine demand test can be used to determine the where fV = normality ol Thiosulfate Titrant
best chlorine dosage to achieve specific chlorinalion objec- 3. Acetic acid, concentrated (glacial).
tives. The measurement of chlorine demand is pertormed by
treating a series of water samples in question witir tnown but 4. Potassium iodide, Kl, crystals.
varying amounts of chlorine or hypochlorite. After the desired
contact time, calculation of residual chlorine in the samples
5. Slandard sodium thiosulfate 0.025 N.
will demonstrale which dosage satisfied the requiremenG of 6. Chlorine demand-free water. prepare chlorine demand_
the chlorine demand in lerms of the residual desiied. free waler lrom good-quality distilled or deionized water
by adding sutficient chtorine to give 5 mg/L free chtorjne
B. What Is Tested residual. Atter standing 2 days. this soluiion should con_
tain al least 2 mg/L lree residua, chlorjne. If not, discard
Common Range
Chlorine Demand, mg/L and obtain better quality water. Bemove remaining lree
chlorine by placing the sotution jn the direct sunlight:After
SurfaceWater O.S to 5 several hours, measure total chlorine residual, Do not use
Well Water 0.1 to 1.3 until last trace of chlorine has been removed.
C. Apparatus 7. Starch indicator
ln addition to the apparatus describecl under one of the E. Procedure
melhods for chlorine residual, the following jtems are required:
Flasks, Erlenmeyer
1. lvleasure a 500.m1sample into each of five to ten 1,OOO-
(1,OOO mL) mL flasks or botfles.
Pipets (5 and 10 ml) 2. To the first ftask, add an amount ol chlorine that leaves no
Graduated cylinder (5OO mL) residual at the end of the contact time. Mix.
Flask, volumetric ('1,OOO mL) 3. Add increasing amounts of chlorine to successive por_
Flask, Erlenmeyer (250 mL) tions of the sample and mix.
Buret (25 mL) 4. Measure residual chlorine after the specific contact time.
Flecord resulls.
0. Reagents
addition to the reagents described under the method you
5. O-n graph paper, plot the residual chlorine (or the amount
.ln of chlorine consumed) versus chlorine dosage.
will use to determine chlorine residual, the following items;re
also needed: E Precautions
1, Stock chlorine solution. Obtain a suitable solution from a 1. Dose sample porlions at time intervals that will leave you
chlorinator solution line or by purchasing a botfle ol enough time for chlorine residual testing at predei€r_
household bteach (.Clorox, or simitar produit). Store in a mined contact times.
dark, cool place to.malntain chemical strenoth. House_
hold bleach products usua y contain approximately live 2. Conduct test over the desired contact time. lf test objec_
percent available chtorine, which is about 5O,OOO mg7l. tive is to dupiicate your plant contact time, lhen mitch
plant detention time as closely as possible.
2. Chlorine dosing solution. lf using household bleach, care-
tully pipet about 1O mL of the bte;ch into a .t ,oOO-mL votu_ 3. Keep samples in the dark, protected from sunlight.
melric flask. Fill to the mark with chlorine demand-free 4. Keep temperature as constant as possible.
waler and slandardize. lf usinq a stock chlorine solution
obtained from a chlorinator solution line, simply standard- 5. Sterilize all glassware if test has a bacteriologic objec-
ize this solulion direc y. tive.
Slandardization: G. Example
a, Place 2 mL acetic acid and 20 mL chlorine demand- A raw water sample was collected trom a river to delermine
free water in a 250-mL flask. chlorine demand.
b. Add about 1 gm Kl crystals. Contact Time = 30 minutes (plant Detention Time)
c. Measure into the tlask a suitable volume of chlorine PH = 76
dosing solulion. lf using household bleach as your Temperature = 15"C
542 Waler Treatment
(Chlorine Demand)
OUTLINE OF PBOCEDURE FOB CHLOBINE DEMAND
m
1. Measure 500 mL water sample into each container.
0
Lab Procedures 543
(Chlorine Demand)
Flesults from the chlorino demand test we.e as follows: Figure 11.1'l is a plot of the chtorine demand curve. Note
ChtorineAdded, Total Residual that the far right end of the curve is flat, which indicates that
Flask No. mglL Chlorine after 30 min, mg/L the chlorine d€mand has b€en satisfied. The curv6 reveals
that lhe chlorine demand will increase as chlorine is added l;
1 0.5 0.36
water until you have gone past the breakpoint.
2 1.O 0.82
3 1.5 1.14
cHLOiTNE DE[|AriD, rc/a
4 2.O 0.60
c 2.s o.75
6 3.0 1.25
H. Calculation
Calculate the chlorine demand by using the formula:
Chlorine Oemand,
_ Chlorine Added, Total Residual Chlorino,
mglL-mg/L-mg/L
Added. Totat Residual Chtorine
Chtorine
Frask No. mglL Chtorine. mg/L = Demand, mo/L
a
11.3F How do€s lhe operator use the results Irom lhe
chlo_
rine demand test?
- 11.3G Calcxlate the chlorine demand lor a raw water
sam_
ple if chtorine is added at Z.O mg/t anO
lhe-free
i0r residual chlorine after 30 minutes w;s 0.4 ngL. --
fudrtLoaaothof bl.o*0rc
olL
c}ltoRDtElooEo, nc,r u\goeffioce
Fig. 11.10 Briakpoinl chlorination curve
wocEPucth
Please answer the discussion and review questions
next.
CHAPTERll. LABORATOHYPROCEDURES
(Lesson4ol5Lessons)
to Facullative (F
AcK'ul-TAY'tive). Facullalive bacleriacan use eilherdissolved molecularorygen oroxygen obtained lrom lood materials such as
sulrale or nitrale ions. ln olher words, facullative bacteria can Jive under aerobic or anaerobia aonditions.-
Lab Procedures 545
(Coliform)
The basic procedure involves filtering a known volume of water specifically designed reagsnt formulation that is specific only
through a membrane lilter of optimum pore size for full colilorm to coliform bacteria. lt provides a specific indicator for the
bacteria retention. As the water passes through the micro- target microbes, lotalcoliforms, and E coli As the medias nu-
scopic pores, bacleria are entrapped on the upper surlace of trients are metabolized, yellow color and fluorescence are re-
the filter. The membrane lilter is lhen placed in conlact wilh leased, confirming the presence of total colilorms and E coll
either a paper pad saturated with liquid medium or directly over respectively.
an agar (gelatin-like) medium to provide proper nutrients tor
bacterial groMh. Following incubation under prescribed condi- 6. Colisure Method
tions ol time, lemperalure, and humidity, the cultures are exam- The Colisure test is a presence/absence test for colilorm
ined lor colilorm colonies, which are counted directly and re- bacteria. A sample of waler is edded to a dehydrated mBdium
corded as a density ol coliforms per 100 mL of water sample. and, atter 28 to 48 hours, the medium is examined for the
There are certain important limitations to membrane filter presence ol colitorms. (Th€ Colisure test is available from
methods. Some types of samples cannot be filtered because IDEXX Laboratories, '1 IDEXX Driva, Westbrook, ME 04092.)
ol excessive turllidity, high non-colilorm bacterial densities, or
heavy metal (bactericidal) compounds. ln addition, coliforms
C. Whal ls Tested
conlained in chlorinated supplies sometimes do not give char- Common Range
acteristic reactions on the media and hence special proce- Total Colitorms per 100 mL
dures must then be used.
Surlace Waters 50 to 1,000,000
3. Membrane Filter N,lethod-Escherichia cor(EPA N4ethod)
Treated Water Supplies 0
This method combines information from a 1985 EPA publi-
cation (Test methods tor Escherichia coll and Enterococci in Well Water 0to50
Water by the Membrane Filter Procedure, EPA 600-4-85-076)
and a subsequent fiIarch 2000 manual (lmproved Enumera-
D. Materials Required
lion Methods lor the Recreational Water Quality lndicators: 1. Sampling Bottles
Enterococci and Escherichia cofi, ePN821lR-971004). This
method is a membrane filter (MF) procedure for identifying Botlles of glass or other mat€rial that are watertight, resist-
and counting E. coli in walet. The method incorporates the
ant to the solvent action ol water, and capable of being steri.
lized may be used for bacteriologic sampling. Plastic bottles
use of a selective and differential medium, mTEC agar, fol-
made of nontoxic materjals are also satisfactory and eliminate
lowed by incubation with urea substrate medium.
the possibility of breakage during transport. The bottles
The E coli test is used as a measure ol recrealional water should hold a sutficient volume of sample tor all tests, permit
quality. Epidemiological studies have led to the development proper washing, and maintain the samples uncontaminated
ol criteria that can be used to promulgate recreational waler until examinations are complete.
slandards based on established relationships between heallh
ellecls and water quality. The signilicance ol finding E coli in
Before sterilization by autoclave, add 0.1 mL 10 percent
recreational water samples is the direct relationship between
sodium thiosulfate per 4-ounce bottle (120 ml). This will neu-
lhe density ol E coll and the risk of gastrointestinal illness as-
tralize a sample containing about 15 mg/L residual chlorine. lf
sociated with swimming in the water. This tesl can be applied
the residual chlorine is not nautralized, it would continue to be
toxic to the colilorm organisms remaining in the sample and
lo kesh, estuarine, and marine waters. Since a wide range of
give false results.
sample volumes or dilutions can be analyzed by the MF tech-
nique, a wide range ol E, coll levels in water can be idenlified When filling botlles with sample, do not llush out the sodium
and counted. thiosulfate or contaminate the bottl€ or sampl6. Fill boilles ap-
proximately three-quarters full and start the test in the labora-
1. Presence/Absence (P/A) Method
tory within six hours. ll the samples cannol be processed
The presence/absence (P/A) method for total coliform is a wilhin one hour, they should be held below 10'C for not longer
simple modificalion of the multiple tube (or MPN) method. lhan six hours.
Eolh lactose broth and lauryl tryptose broth are the essential
irgredients of the P/A media. ln addition, bromocresol purple 2. Glassware
isadded as a pH indicator All glassware must be thoroughly cleansed using a suitablg
The procedure involves the addition of '100 mL of sample to detergent and hot waler (160'F or 71'C), rinsed with hot water
50 mL ol sterile P/A broth in a 250-mL milk dilution bottle. This ('180"F or 82'C) to remove alllraces of residual delergent, and
iroculaled media is then incubated at 35"C and inspecled finally rinsed with distilled or deionized water.
allet 24 and 48 hours for an acid reaction. A distinct yellow
mlor forms in the media when acid conditions exist lollowing
3. Water
klose lermentation. ll gas is also produced, gently shaking Only distilled water or demineralized water that has been
he bottle will result in a toaming reaction. Any amount of gas tested and found free from traces of dissolved metals and bac-
oracid (yellow color) constitutes a positive presumptive test re- tericidal and inhibitory compounds may be used lor prepara-
qjidng confirmation in the same manner as the IVIPN method. tion ot culture media.
t ColilertrM Method 4, Butfered" Dilution Water
Colilert, a registered trademark of IDEXX Laboratories, Prepare a stock solution by dissolving 34 grams ol KHZPOT
flovides simultaneous deteclion, specific identification, and in 500ml distilled water, adjusting the pH lo 7.2 with 1 N
oflfirmation ol total coliforms and E coli in water. Colilert is a NaOH and dilute to one liter. Prepare dilution wat6r by adding
&r,ter A solution or liquid whose chemical makeup n€ukalizes acids or bases without a great change in pH
545 Water Treatment
(Coliform)
1.25 mL of the stock solution and 5.0 mL magnesium sulfate media stor€d in the refrigerator should be used within
(50 qrams MgSO". z H,O dissolved in one liter of waler) to 1 96 hours.
liter distilled water. This solution can be dispersed into various
size dilution blanks or used as a sterile rinse lor the membrane
The rosolic acid solution should be stored in the
lilter test. dark and discarded atter two weeks or sooner il its
color changes from dark red to muddy brown.
5. Media Preparation
M E D IA_P R E S E NC qABS E N C E ( P/A) M ETHO D
Careful media preparalion is necessary lor meaningful bac-
teriological lesting. Attention must be given to the quality, mix- a. P/A broth
ing, and sterilization ofthe ingredients. The purpose ot this care Prepare this media lollowing the instructions on the
is to ensure thal il the bacteria being tested for are indeed media bottle label.
present in the sample, every opportunity is presented lor the
development and ultimale identification. Bacteriological identi- b. Brilliant green bile broth
lication is often done by noting changes in the medium; conse- Same as for MPN method.
quently, the composition of lhe medium must be standardized.
Much otthe tedium of media preparation can be avoided by pur- c. EC broth
chase of dehydrated media (Difco, BBL, or equivalent) from
local scientilic supply houses. The operator is advised to make Same as for MPN method.
use of these products, and, if only a limited amount of lesting is M ED I A--COLI LERT TM
M ETHOO
to be done, consider using tubed, prepared media.
a. Available lrom: IDEXX Laboratori€s
MEDIA_MPN (TOTAL COLIFOBM) 'l IDEXX Drive
Westbrook, ME 04092
a. Lauryl tryptose broth
(800) 321-0207
For the presumptive coliform test, dissolve the rec-
ommended amount ol lhe dehydraled media in distilled 6. Media Storage
water. Dispense solution into fermentalion tubes con- Culture media should be prepared in balches of such size
taining an inverted glass vial (see illustration of the tube that lhe entire batch willbe used in less than one we€k.
with vial on page 549). For 1o-ml water portions from
samples, double-strength media is required while ali 7. Autoclaving
other inoculations require single strenglh. Direclions
for preparation are given on the media bottle label. Steam auloclaves are used for the sterilization of the liquid
media and associated apparatus. They sterilize (kill all organ-
b. Brilliant green bile (BGB) broth isms) at a relalively low temperature of 121"C within ',5 min-
utes using moist heat.
For the confirmed colilorm test, dissolve 40 grams of
the dehydrated media in one liter of distilled water. Dis- Components of the media, particularly sugars such as lac-
pense and sterilize as with lauryl tryptose broth. tose, may decompose at higher temperatures or longer heat-
ing times. For this reason, adherenc6 to time and temperatu16
c. EC broth schedules is vital. The maximum elapsed tim6 for exposure ot
Prepare this media following the instructions on the the media to any heat (kom the time lhe autoclave door is
bottle. closed to unloading) is 45 minutes. Preheating the autoclavo
can reduce total heating time.
MEDIA-MEMBBANE FILTER M ETHOD (TOTAL
coLtFoRM) Autoclaves operale in a manner similar to the familiar kitch-
en pressure cooker:
a. M-Endo broth
a. Water is heated in a boiler to produce steam.
Prepare this media by dissolving 48 grams of the de-
hydrated product in one liter oI distilled water that con- b. The steam is vented to drive out air.
tains 20 mL of ethyl alcoholl'? per liter. Heat solution to c. The steam vent is closed when the air is gone.
boiling only--do not autoclave. Promptly remove solu-
tion lrom heat and cool. Prepared media should be d. Continued heat raises the pressure io 15 lbs/sq in
stored in a refrigerator and used within 96 hours. (103.4 kPa or 1.05 kg/sq cm); (at this pressure, pure
steam has a temperature of 12'1'C at sea level only).
b. LES Endo agar
e. The heat and pressure are maintained for 1 5 minutes
Prepare this media, used lor the two-step proce-
dure, following lhe instructions on the boltle. f. The heat is turned off.
c. l\4-FC media g. The steam vent is opened slowly to vent steam until at-
mospheric pressure is reached. (Fast venting will
Flehydrate in dislilled water containing 10 mL ol 1
cause the liquids to boil and over,low lubes.)
percent rosolic acid in 0.2 N NaOH. Heat just to boiling
then cool to below 45"C. Do not autoclave. Prepared h. Slerile material is removed to cool.
''? ln some states, the ethyl alcohol used in bacteriological m6dia pr€paralion cannot be lhe speclally denatured alcohol sold by supply houses to
people withoul an alcohol permit. One way to obtain suitable ethanol undor lhese circumslances is lo buy a brand ol gthanol sold lor human con.
sumption.
Lab Procedures 547
(Colilorm)
ln autoclaving termentation tubes, a vacuum is formed in way.ol a coniaminated pipet. Clean, sterile pipets must be
the inner iubes. As the tubes cool, the inner lubes are filled used lor each separate sampls.
with sterile medium. Capture ol gas in this inner tube from the
culture ol bacleria is the evidence of fermentation and is re_ FOB UNTREATED WATER SAMPLES
corded as a positive test.
a. Shake the samplo bot e vigorously 20 times before re-
E. Procedure for Testing Total Coliform Bacteria moving sample volumes.
irulliple Tube Fermentation Method b. Pipet 10 ml of sample direc y into each ot the lirst livs
tubes. Each tube must contain 10 mL lauryl tryptoss
1. Genoral Discussion broth (doubl€ skength).
The tesl for colirorm bacteria is used to measure the suit- NOIE You must realiz€ that the sample voluma
ability of a water for human use. The test is not only uselul in applied to the first fiv€ tubes will depend upon the typs
determining the bacterial quality ol a linished water. but il can of water being tested. The sample volume apptiei to
be used by lhe operator in the treatment plant as a guide to each tube can vary from 1O mL for high-quality waters
achieving a desired degree ot treatment. to as low as 0.00001 mL (applied as 1 mL of a dilut€d
Colilorm bacteria are detected in water by placing portions sample) for highly polluted raw water samples.
of a sample of the water in lauryl tryptose broth. L;urvl trvo_ NOIE When delivoring the sample into the cullure
lose broth is a standard bacteriological media containing lac_ medium, deliver sample portions of i mL or less down
tose (milk) sugar in tryptose brolh. The colitorm bacterii are into the cullure tube nearth6 surface ofthe medium. Do
those that can grow in this media at 35"C temperature and are not deliver smallsample volumes at the top of the tube
able lo ferment and produce gas from the lactose within 48 and allowthem to run down inside the lube:too much ol
hours. Thus, to detect these bacteria, the operator need onlv the sample willfail to reach the culture medium.
inspecl the lermentalion lubes lor gas. A schematic of lhe coli:
Iorm test procedure is shown in Figure .11.12. A/OTE; Use 1o-ml pipets lor 1O-ml sample porlions,
and l-ml pipets for portions ol .t ml or iess. Handle
Dinkjng water coliform samples must be jOO mL in volume, sterile pipet only near the mouthpiece, and protect the
so the fivelube multiple tube fermentation method is not ap- delivery end from enernal contamination.
propriate. The lest uses len tubes with a 1O-mL sample in
each tube. c. Pipet 1 mL ol water sample into each of the next five
lauryl tryptose broth (single strength) tubes.
2. Materials Needed
d. Pipet 1/10 mL (0.1 mL) ol water sample into each ol
FOP UNTFIEATED WATEB SAMPLES the next.five lauryl tryptose broth (single strength)
a. Fitleen sterite tubes containing .lO tubes. This makes a 0.1 (1 to tO) dilution.
mL ol lauryl tryp-
tose broth are needed for each sample. Use fiv; iubes At this point, you have t5 tubes inoculated and can
lor each dilution. place these three sels of lubes in the incubator; how-
b. Dilution tubes or btanks containing 99 ever, your sample specimen may show gas production
ml of sterile in all fermentation tubes. This means your simple was
bulfered distilled water.
not diluted enough and you have no usable resulls.
c. A quantity of 'l-ml and 1O-mL serotogical pipets. The
1-mL pipets should be gradualed in 0.1_mL incre- e. To make a 1/100 (0.01) ditution, add 1 mL ol well_
mixed water sampl€ to 99 mL of sterile butfered dilu-
ments.
tion water. Mix thoroughly by shaking. Add 1 mL lrom
d. lncubator set at 35" j 0.5.C. this bottle direc y into each of five more lauryt tryptos€
broth tubes.
e. Thermometer verified to be accurate by comparison
with a National lnstitute of Standards and Technology f. To make a 1/1.000 (0.001) dilution, add 0.1 mL from
(NIST) certified thermomeler the 1/100 dilution bonle direc y into each of live more
lauryl tryptose broth tub€s.
FOR DRINKING WATER SAMPLES
I 3.
with a National lnstjtute of Standards and Technology
(NIST) certified thermomeler, or equivalent. ''
Technique for lnoculation ol Sample (Figures 11.13 and
11 14) The first time a sample is analyzed, 25 tubes of tauryl tryp-
.
tose brolh shoutd be prepared. Once you find out wtrit dill.
All inoculations and dilutions ol waler samples must be ac- tions give usable results for determining the MpN index. you
flilale and made so that no contaminants from the air, equip- will only _need to prepare t5 tubes to analyze subsequint
ment, clothes, or fingers reach lhe sample, either direcfly dr 6y
samples lrom the same source.
-.
548 Water Trealment
(Colilorm)
1. Presumptive Test
No gas
no colilorm present
discard tubes
2. Conlirmed Test
FECAL COLIFORM
1. Presumptive Test
(same as for TOTAL COLIFOHM)
2. Fecal Coliform Test
Fig. 11.12 Schematic outline ol tesl procedure for totalcolilorm and fecal coliform-qultiple tube lermenlation method
lrqr
Lab Procedures 549
(Coliform)
q!{19! 4 lncubation (Totat Coliform)
I0mLIOEACHTIjBE
a. 24-Hour Lauryl Tryptose (LT) Broth presumptive Test
-l NO DILL]TION
Place all inoculated LT broth tubes in 35" t O.S"C incu-
0 0 0 0 0 bator. After 24 t 2 hours have elapsed, examine each
tube lor gas rormation in inverled vial (inner tube). Mark
plus (+) on report form such as shown in Figure 11..1S for
all tubes that show presence of gas. Mark mlnus (-) for all
tubes showing no gas formation. lmmediately perform
confirmation test on atl positive (+) tubes (See Faiagraph
0 0 0 0 0 c. below.'24-Hour Brilliant green bite (BGB) Confirma-
lion Test"). The negative (-) tubes musl be reincubated
for an additional24 hours.
b. 48-Hour Lauryl Tryptose Broth presumptive Test
0 0 0 0 0 Record both positive and negative iubes at the end of
48 t 3 hours. lmmediately perlorm total coliform contir-
mation test on all new positive tubes. For drinking water
samples, all positive tubes should also be tested for fecal
coliforms fE. colr.
0 0 0 0 0 c. 24-Hour Brilliant green bile (BGB) Confirmation Test
0l mt TO EACH TUB€
,Confirm all presumptive tubes that show gas at 24 or
DILUTION
$
Fig. 11-13 Colitorm bacteria test-raw water
FOR UNTREAIED WAIER
OATE NO 50uRca
28 Rivar
Raw
88 o
k:l 21
l0 0.1
24
l
---oT-l
ITTr.I,l; TII] rr, rrrl 14
g !ll!
6^\ 0 0 0 toltol i 66 tol
NO DILUTION 4E 48 ITt-F
$
Fig. 11.14 Colilorn bacteria test-4rinking water
.OR ORIt]KING WATER
City
70
Hall
FOR DBINKING WATEB SAMPLES
a. Shake the sample bottle vigorously 20 times before re- PNESUMPTIVI CONFIRMTO MPN/r00 mt
moving a sample volume. t0
b. Pipet 10 mL ol sample direc y into each ol ten tubes 24 24 I T o
containing '10 mL of doubte-strength lauryl tryptose
48 48 I !lll I
broth. Fig. 11.15 Becorded cotilorm test rcsults
550 Water Treatment
(Coliform)
Always sterilize inoculation loops and needles in ,lame production tor each tube. Gas in any quantity is a
immediately before transter ol culture; do not lay loop positive test.
down or touch it to any nonsterile object belore making
the lransfer. After slerilization in a flame, allow sufficient
(7) As soon as results are recorded, discard all
time lor cooling, in the ai( to prevent lhe heat of the loop tubes. This is a 24-hour tesl for EC broth inocula-
tions and not a 48-hour test.
from killing lhe bacterial cells being transferred. Sterile
wood applicator sticks also are used to transfer cullures, (8) Transfer any additional 48-hour gas-positive
especially in the lield where a flame is not available for tubes lrom the presumptive tesl to corrgspond-
sterilization. lf using hardwood applicators, sterilize by ingly labeled tubes ol EC broth. lncubato tor 24 t
autoclaving belore use and discard after each transrer. 2 hours at 44.5" ! O.2"C and record results on
Aiter 24 hours have elapsed, inspect each of the BGB data sheel.
tubes for gas formation. Those with any amount of gas (9) Codily results using the same procedure as lor
are considered positive and are so recorded on the data total colilorms and determine MPN of fecal coli-
sheet. Negative BGB tubes are reincubated for an addi' forms per,I00 mLof sample (Figure'11.15).
lional 24 hours.
ing them. Transler one loop-full of culture lrom Read MPN as 50 per 100 mL from Table 1't.3
each gas-positive culture in presumptive test to Fleport results as 500/100 mL
the correspondingly labeled tube ol EC brolh. we added one zero to 50 because we started with the 1
lrom the water bath, shake genlly, and record gas Flead MPN as 1.1 perl00mLfromTabl6 11.4
Lab Procedures 55'l
(Coliform)
,I
TABLE 1.3 MPN INDEX FOB VARIOUS COMBINATIONS OF POSITIVE AND NEGATIVE BESULTS
lN A PLANTING SERIES OF F|VE 1O-mL, F|VE .t-ml, AND
FIVE 0.1-mL pORTIONS OF SAMpLE
1 4
0 1
5
0 0 4
0 2 1 6
0 3 0 6
1 0 0 2
l 0 1 4
1
,l
0 2 o
0 8
1 1 0 4
1 1 1 6
1 1 2 8
2 0 6
2
2
'I
I
'10
2
3 n 8
3 1 t0
1 4 0 11
2 0 0 4
2 0 1 7
2 0 2 o
2 0 3 2
2 1 0 7
1
,l
1 9
2 2 12
2 2 0
2 2 ,,
12
2 2 2 14
2 3 0 12
2 3 1 14
2 4 0 15
3 0 0 8
0 1 11
0 2 13
3 1 0 11
3 1
,l
14
3 1 2
3 1
20
3 2 0 14
3 2 1 17
3 2 20
552 Waler Treatment
(Coliform)
TABLE 11.3 MPN INDEX FOR VARIOUS COMBINATIONS OF POSITIVE AND NEGATIVE RESULTS
IN A PLANTING SEBIES OF FIVE 1O-mL, FIVE 1.m1, AND
FIVE 0.1-mL PORTIONS OF SAMPLE (conrnued)
3 3 0 17
3 3 1
3 4 0
3 4 1 24
3 0 25
4 0 0 '13
4 0
,1
17
4 0 2 21
4 0 3 25
4 1 0 17
,1
4 1 21
4 1 2 26
4 2 0 22
4 2 '1
26
4 2 2
4 3 0 27
4 1
4 3 2 39
4 4 0 34
4 4 1 40
4 5 0 41
,|
4 5 48
5 0 0
,1
5 0 30
5 0 2 40
5 0 3 60
5 0 4 80
5 0 30
,l
5 50
5 60
5 3 80
5 2 o 50
2 1 70
5 2 2 90
5 2 3 130
5 2 4 150
5 2 180
5 3 0 80
'1
3 1 10
5 3 2 140
3 3 170
5 3 4 210
5 3 250
5 4 0 130
4 1 170
5 4 2 220
5 4 280
f-.
Lab Procedures 553
(Colilorm)
TABLE 11.3 MPN INDEX FOR VARIOUS COMBINATIONS OF POSITIVE AND NEGATIVE HESULTS
IN A PLANTING SERIES OF FIVE 1O.mT, FIVE l.mL, AND
F|VE 0.1-mL pOBT|ONS OF SAMpLE (continued)
66BB
1 1.1
0 2 2.2
3.6
4 5.1
5 6.9
6
0.00'l mL O OUTOF 5
7 12.O
0 0 0 0 B 16.1
I 23.0
Fig. 11.16 Besults ol colilorm test-untreated water 1 0 >23.0
554 Water Treatment
(Coliform)
When you wish to summarize with a single MPN value the 3. Determine the mL of sample in all tubes.
resulls from a series of samples, the arithmetic mean, the me-
dian, or the geometric mean may be used (see following sec- Number mL of Sampte
tions for formulas and examples).
Sample Size, ml Tubes
ol in Att Tubes
10 ml 5 (10 mrxs)
1.0 ml 5 (1.0 mL)(5)
FOBMULAS 0.1 mL 5 (0.1 mLXs) = 0.5 ml
The median is the middle value in a set or group of data. = 10.8 MPN Coliforms/'l00 mL
There are just as many values larger than the median as there
are smaller than the median. To determine the median, the
dala should be written in ascending (increasing) or descend-
ing (decreasing) order and lhe middle value identified.
= 11 MPN Coliforms/'|00 mL
EXAMPLE 5
Median = Middle Value ol a Group of Data
Flesults from IVIPN tests during one week were as follows:
The median and geometric mean are used when a sample
contains a few very large values. This frequen y happens DAySMTWTHFS
when measuring the MPN of raw water. lf X is a measurement i\.4PN/100 mL 2 8 14 6 10 26 4
(MPN) and n is the numberof measurements, then
Estimate the ('1) mean, (2) median, and (3) geometric mean
Geometric Mean .................................. (X")],, of the data in MPN/100 mL.
= [(X,X&XX.)
1. Calculate the mean.
This calculation can be easily performed on many electronic
calculators using the power tunction, y'. sum of All MPNs
t4ean. MPN/'1oo ml -
Number of MPNS
EXAMPLE 4 2+8+ 14+6+ 10+26+4
_
The results from an MPN test using five ,ermentaiion tubes 7
for each dilution on a sample of water were as follows; 70
Sample Size, mL 10 1.0 0.'1 7
(Coliform)
4. Summary k. One moist incubator at 35"C; auxiliary incubalor dish
with cover.
1. Mean = 10 MPN/100 mL
2. l,4edian 8 MPN/100 mL
l. Enrichment media-lauryl tryptose broth (for pre-
= enrichment technique).
3. Geometric Mean = 7.5 N4PN/100 mL m. A binocular, widejield, dissecling microscope is rec-
As you can see trom the summary the geometric mean ommended for counting. The light source should be a
more nearly describes most of the MPNS. This is the cool white, f luorescent lamp.
reason why the geometric mean is sometimes used to de-
scribe the results of MPN tests when there are a lew very 3 Selection of Sample Size
large values that would distort the mean.
The size of the sample ot ALIQUATI1 will be governed
8. Belerence by the expected bacterial densiry An ideal quantity will
result in the groMh oI 20 to 80 coliform colonies, but not
See page 9-48, STANDARD \IETHODS, 21st Edition. more than 200 bacterial colonies of all types. The table
below lists suggested sample volumes for MF lotal coli-
Membrane Filter Method form testing.
1. General Discussion Ouantities Filtered (mL)
ln addition to the fermenlation tube test lor coliform 100 '10 1 0.1 0.0'1
bacteria, anolher test is used for these same bacteria in
Well Water x
water analysis. This test uses a cellulose ester filler,
Drinking Water x
called a membrane filter, the pore size ol which can be
Lakes x x x
manufactured to close tolerances. Not only can the pore
size be made to selectively trap bacteria from water fil-
Fiivers x x xx
tered through lhe membrane, but nutrients can be dif- When less than 20 mL of sample is to be filtered, a
lused (lrom an enriched pad) through the membrane to small amount ot sterile dilulion water should be added to
grow these bacteria into colonies. These colonies are rec- the funnel before filtration. This increase in water volume
ognizable as coliform because the nutrients include fuch- aids in unlform dispersion of the sample over the mem-
sin dye, which peculiarly colors the colony. Knowing the brane filter.
number of colonies and the volume ot water liltered, the
operator can then compare the water tesled with water
quality standards.
4. Preparation of Petri Dish for Membrane Filter
a. Sterilize forceps by dipping in alcohol and passing
A two-step pre-enrichment technique for chlorinated quickly lhrough Bunsen burner flame to burn off the al-
samples is included at the end of this section. Chlorinated cohol. An alcohol burner may be used also.
bacteria that are still living have had their enzyme sys-
lems damaged and require a 2-hour enrichmenl media b. Place sterile absorbent pad into sterile petri dish.
before contact with the selective M-Endo media.
c. Add '1.8 to 2.0 mL lvl-Endo medium to absorbent pad
2. Materials Needed using a sterile pipet. Flemove excess media.
a. One sterile membrane filter having a 0.45p pore size. 5. Procedure for Filtration of Unchlorinated Samples
b. One sterile 47-mm pelridish wjth lid. All liltrations and dilutions of water samples must be ac-
c. One sterile tunnel and support stand. curate; no contaminants from the air, equipment, clothes,
or fingers should reach the specimen either directly or by
d. Two sterile pads. way oI the contaminated pipet.
e. One receiving flask (side-arm, 1,000 mL). a. Secure tubing trom pump and bypass to receiving
flask. Place palm of hand on flask opening and start
f. Vacuum pump, trap, suction or vacuum gauge, con- pump. Adjusl suction to % atmosphere with hose
nection sections of plastic tubing, glass T-hose clamp
clamp on pressure bypass. Turn pump switch to OFE
to adjust pressure bypass.
g. Forceps (round-tipped tweezers), alcohol, Bunsen b. Set sterile filter support stand and funnel on receiving
flask. Loosen wrapper. Rotate funnel counterclockwise
burner, grease pencil.
to disengage pin. Secure wrapper.
h. Sterile, buffered, distilled water tor rinsing
c. Place petri dish on bench vrith lid up. Write identifica-
i. M-Endo media. tion on lid with grease pencil.
Slerile pipets-two s-ml graduated pipets and one 1- d. Unwrap sterile pad container. Light Bunsen burner.
mL pipetlor sample or one 10-mL pipet lor larger sam-
ple. Quantity of 1-mL pipets if dilution of sample is nec-
e. Unwrap membrane filter container.
essary. Also, quantity of dilution water blanks if dilu. f. Sterilizo forc€ps by dipping in alcohol and passing
lion of sample is necessary. quickly through Bunsen burnsr to burn off th6 alcohol,
rr ,4rqoot (AL-li'kwot). Represenlalive portion of a sample. Ollen an equally divided porlion of a sample
J t
556 Water Treatment
(Coliform)
g. Center membrane filter on lilter stand with forceps SPECIAL NOTE:
after lifting funnel. l\.4embrane filter with printed grid
should show grid uppermost (Fig. I, next paqe). lnexperienced persons often have great difiiculty with
h. Replace funnel and lock against pin (Fig. ll). connected colonies, with mirror reflections ol fluorescent
tubes (which are confused with metallic sheen), and with
i. Shake sample or diluted sample. Measure proper ali- water condensate and particulate matter, which are occa-
quot with sterile pipet and add to funnel. sionally mistaken for colonies. Thus, there is a lendency
j. Add a small amount o, the sterile dilution water to fun- for inexperienced persons to make errors on the high sida
nel. This will help check for leakage and also aid in dis- in MF counts. Al least live apparent coliform colonies
persing smallvolumes (Fig. lll). should be transterred to lauryl tryptose broth tubes for
k. Now start vacuum pump. verification as coliform organisms.
'a Aseptic (a-SEP-tick). Free lrom the tiving germs ol disease, lermentation, or putrefaclion. Sterile,
Lab Procedures 557
(Colilorm)
OUTLINE OF PFIOCEOUFE FOFI INOCULATION OF MEMBHANE FILTER
Fig. I Fig. ll
1. Cenler membrane lilter on 2. Place funnel
filter holder. Handle mem- onto filter
brane only on outer 3/r6 holder
inch with forceps sterilized
betore use in ethyl or methyl
alcohol and passed lightly
through a llame.
I
(r,
5. lncubate in
inverted
4. Flemove membrane filter lrom
position for
filter holder with sterile
22 t 2 houts.
forceps. Place membrane
3. Pour or pipet sample on pad. Cover with petri 6. Counl coliform-
aliquot into funnel. top. appearing colonies
Avoid spattering. Alter on membrane.
suction as applied, rinse
lour times with sterile,
bullered, dislilled water.
558 Water Treatment
(Coliform)
9. Alternate Nlembrane Filter Procedure for FecalColi{orm EXAMPLE:
The following method is a way to test for fecal coliform di- A total of 2 colonies grew after tiltering a 100-mL
reclly without first running the tolal coliform test as a separate sample.
procedure. This membrane filter procedure for recal coliform
uses an enriched lactose medium (M-FC broth) that depends Fecal Coliforms Fecal Colonios Counted x'100 mL
on an incubation temperature of 44.5't 0.2"C for its selectiv- per 100 mL Sample Volume Filtered, mL x 100 mt
ity. Since the temperature is critical, incubation takes place in (2 Colonies)(100 mL)
a water bath using watertight plastic bags. = rco ,rxroo -L)
a. Materials Required (0.02x100)
(1) M-FC media. 100 mL
<.\
j. Water bath maintained at 44.5'r 0.2'C
k. Absorbent pads.
(Colilorm)
9- x 50-mm culture dish to a 4- to s-mm depth (approx- with urea substrale medium and allow to sit at room
imately 4 to 6 mL) and allow to sotidify. The tinal pH temperature lor 15 to 20 minutes.
should be 7.310.2. Store in a refrigerator
k. After incubation on the urea substrate at room lemper-
m. Urea substrate medium ature, count and record the number ol yellow, yellow-
green, or yellow-brown colonies on the membrane fil-
Preparation: Add dry ingredients to 100 mL of re-
agent-grade distilled water in a flask. Stir to dissolve ters, ideally containing 20 to 80 colonies.
and adjust to pH 3 lo 4 with 1 N HCl. The substrate so- 4. Calculalion of Besults
lution should be a straw-yellow color at this pH.
Select the membrane filter with an acceptable number of
n. Nutrient agar (Dirco 0001, BD431r472) yellow, yellow-green, or yellow-brown colonies (20 to 80) on
Preparation: Add 23 grams ol nutrient agar to 1 L of the urea substrate and calculate the number ol E coli per 1OO
. reagent-grade distilled water and mix well. Heat to
mL according to the lollowing general formula:
boiling to dissolve the agar completely. Dispense in _ No. E coli Colonies Counted x 100 mt
screw-cap tubes and autoclave al 121.C ('15lbs pres- E. coli llOO mL
Volume in mL of Sample Filtered x 100 mL
sure) fo|l5 minutes. Remove the lubes and slant. The
finalpH should be 6.8 ! 0.2. 5. Reporting Besults
3. Procedure Fleporl the results as E coliper'100 mL of sample.
a. Prepare the mTEC agar and urea subsirate medium 6. Example
as directed in (l) and (m), respectively.
A total ol 40 E coll coloni€s were count€d atter filtering a
b. Mark the petri dish and report form wiih the sample 50-mL sample.
identification and voiume.
E. coti llOO mL _ No. E 6oli Colonies Counted x .100 m L
c. Place a sterile membrane filter on the tilter base, grid Sample Volume Filtered, mL x 100 mL
side up, and attach the funnel to the base; the mem-
40 E. coli 100 mL
brane filter is now held between the funnel and the
(50 mL)(100 mL)
base.
0.8 E. colilmL 100 mL
d. Shake the sample bottle vigorously about 25 times to '1
00 m L
distribute lhe bacteria uniformly and measure the de-
sired volume of sample or dilution inlo the funnel.
= 80 E. coli /10O mL
e. For ambient surlace waters and wastewaters, select 7. References
sample volumes based on previous knowledge of pol-
lution level to produce 20 to 80 E coll colonies on the EPA Test Method: Escherichia coli (E. coli) in Water by Mem-
membranes. Sample volumes of 'l io 100 mL are nor- brane Filtration Using l\rembrane-Thermotolerant Escherichia
mally tested at hall-log intervals, lor exampte, 100, 30, colAgar (mTEC), Method 1 '103.1, 2002.
10,3 mL.
See pages 9-48 and 9-56, STANDARD METHODS,2lst
t l. Smaller sample sizes or sample dilutions can be used Edition.
I lo minimize the interference of turbidity or for high bac-
lerial densities. Multiple volumes of the same sample Presence/Absence Method
or sample dilutions may be liltered and the results may 1. General Discussion
be combined.
The presence/absence (P/A) test for the coliform group
g. Filter the sample and rinse the sides of the funnel at in drinking wat6r is a simple modilication of the multiple
least twice with 20 to 30 mL ol sterile buffered rinse tube procedure described above. One large test portion
water. Turn off the vacuum and remove the funnel lrom (100 mL) in a single culture bottle is used to obtain quati-
I the filter base.
tative information on the presence or absence ol coli-
h. Use sterile lorceps to aseptically remove the mem- forms. The media used js a mixture of laclose and lauryl
brane filter from lhe filter base and roll it onto the tryptose broths with bromocresol purple added to indicate
mTEC agar to avoid the formation of bubbles between pH. Following incubalion, gas or acid (yellow color oI
Ihe membrane and lhe agar surface. Fleseat the mem- media) is produced if colilorms are present. The p/A test
brane if bubbles occur. Close the dish, invert, and incu- procedure is outlined in Figure 11.18.
bate ai 35' 10.5" lor 2 hours. 2. Materials Needed
i. Alter a 2-hour incubation at 35. t 0.5"C. transler the
a. P/A broth. Prepare according to the instructions on the
plate to a Whirl-Pakd bag, seal the bag, place the bag
bottle.
with lhe plate inverted in a test-tube rack, and pul the
rack in a 44.5" I 0.2"C water bath for 22 to 24 hours. b. 250-mL screw-cap. milk dilution bottle with 50 mL ster-
ile triple-strength P/A broth.
j. After 22 to 24 hours, removo th6 plate from the water
bath. Place an absorbenl pad in a new petri dish or the c. Autoclave for sterilization of m€dia and glassware.
lid of the same petri dish and salurate the pad with
urea substrate medium. Aseptically hansfer the mem- d. lncubator set at 35" t 0.5"C.
brane lrom mTEC agar to the absorbent pad saturated e. Sterile, '100-mL graduated cylinder.
560 Water Treatmenl
(Coliform)
OUTLINE OF PBOCEDURE FOR TOTAL COLIFORM
1. Presumplive Test
2. Conlirmed Tesl
FECAL COLIFORM
Fig . 1 1 . 1 8 Schematic outline ol test procedure for total colitorm and lecal colifonn-presence/absence melhod
T Lab Procedures 56,
lnoc u latio n
I00 mL
O No gas P/A pres umptive
but yellow color pos
present ltive
NO CAs
Fecal E NO GAS
Date No
Total Colif orm Fecal
t Sou rce Presumptive Confirmed Coliform
4/8/o8 29 jty Halt
C
+
I
I
i
i
562 Water Treatment
(Coliform)
ColilertrM Method
1- General Discussion '10 mL
Colilert provides a media that contains specific indica- lillline
tor nutrients for total colilorm and E. coli. As these nutri-
ents are metabolized, yellow color and fluorescence are
released confirming the presence of total coliform and E.
coli respectively. Non-coliform bacteria are chemically
suppressed and cannot metabolize the indicator nulri-
ents. Consequently, they do not inlerlere wilh the identifi-
cation ot the target microbes. Total colilorms and E. coli \ t)
are specifically and simultaneously detected and identi-
fied in 24 hours or less.
Although the Colilert method can yield presence/
absence results within 24 hours and is less cumbersome
to perlorm than the m€mbrane filtration (NlF) method, op-
erators should be aware of the limitations ol these tests in e. Cap ths tubes tightly.
evalualing samples lor regulatory purposes. The results f. Mix vigorously to dissolve the reagent by repeated in-
lor the Colilert and MF tests are not always comparable, version of carrier.
which may be due to:
. /VOIE; High calcium sall concenlrations in cerlain walers
lnterferences in the sample lhat may suppress or may cause precipilat€. This will not atfect test results.
mask bacterial groMh
. Greater sensitivity of lhe Colilert media
. Added stress to organisms related to filtering
. The fact that ditlerent media may obtain betler groMh
for some bacteria 4
Both test methods are approved by the US Environ-
mental Proteclion Agency for reporting under lhe Safe
Drinking Water Program. When these tests produce con-
flicting bacteriological results, however, the safest course
of action is to increase monitoring and treatmenl efforts
until the results for both tests are negative.
2 l\,4aterials Needed
(Cotiform)
c. Samples are negative for total colitorms if no color is Head MPN as 2.2 per 1OO ml for totat cotirorm and <.1..1 for E
observed at 24 hours. Should a sample be so ligh y colifrom Table 1 1.S.
yellow alter 24 hours' incubalion that vou cannot d;fin-
itively read it relative to the posilive;omparalor tube,
you may incubate it up to an additional 4 hours. lf the
lnoculation Results
sample is coliform positive, the color will intensify. ll it
does not intensify, consider the sample negative. 10 mL
2 ol 10 - yellow
d. To ,ind theconcentration of total coliforms or E. colipet 0 of 10 - fluorescence
100 ml, compare the number of posjtjve reaclion
tubes per sample set (.10 tubes) to the standard MpN
Yellow Yellow
(Mosl Probable Number) chart shown on Table 1 1.5 or no fluorescence no fluorescence
in STANDARD METHODS.
DATE
MPN/]00 mL
NO SOURCE 24 HRS TOTAL COLIFOHM E. COU
1701 ----+----+---- Yellow
4t16t0B x29 Main St 2.2 <l.1
Fluorescence
(Coliform)
.11,3L
QUESTIONS Estimate the most probable number (MPN) ol coli-
lorm group bacteria for a raw water sample from th€
Write your answers in a notebook and then compare your following test results:
answers with those on page 576.
11.3H Why are drinking waters not tested for specific patho-
ml ol sample 1 0.1 0.01 0.001
gens (such as Cryptosporidium ot Giadia)2 dilutions 0 -1 -2 -g
readings (+ tubes) 5 5 3 't
11.31 To perform coliform lests, how would you decide how
'l
many samples to test? 1.3M How is the number of colilorms estimated by the
membrane filter method?
11.U Why should sodium thiosulfale be added to coljform
1 1.3N What are the incubation conditions for chlorinated
sample bottles?
samples when using the membrane filter method
11.3K Steam autoclaves sterilize (kill atl organisms) at (enrichment technique)?
a
pressure of _psi and temperature ol_"C 11.30 Why is the lotal coliform tesl not an ad€quate meas-
during a time period (at sea level). ure of lhe extent of human wastes in wateaT
-minute
CHAPTERll. LABORATORYPROCEDURES
(Lesson5ol5Lessons)
5. Hardness Funnel
Hot plate
A. Discussion Flask, volumetric (1,000 mL)
Hardness is caused principally by lhe calcium and magne-
sium ions commonly present in water. Hardness may also be
D, Fleagents
caused by iron, manganese, aluminum, strontium, and zinc if NOIe Standardized solutions are available already pre-
present in significant amounts. Because only calcium and pared from laboratory chemical supply compa-
magnesium are present in significant concentrations in most nles,
waters. hardness can be defined as the total concentration of
calcium and magnesium ions expressed as the calcium 1. Bufler solulion
carbonate (CaCO3) equivalent. 2. Standard EDTA or CDTA titrant. EDTA is disodium
There are two types or classifications oI water hardness: ethylene-diaminetetraacetats dihydrate, also called (ethyl-
carbonate and noncarbonate. Carbonate hardness is due to enedinitrilo)-tetraacetic acid disodium salt. CDTA is diso-
calcium/magnesium bicarbonate or carbonale. Hardness that dium-CDTA (1,2 cyclohexanediaminetetraacetic acid).
is due to calcium/magnesium sulfate, chloride, or nitrate is 3. lndicator solulion.
called noncarbonaie hardness.
Hard water can cause incrustations (scale) when the water
evaporales, or when heated in household hot water heaters
and piping. Hardness-producing substances in water also OUTLINE OF PROCEDURE
combine with soap lo lorm insoluble precipitates. The
common method ol minimizing lhese and other problems due mr EOT
to hardness is water supply softening. This procedure is dis-
f 1
cussed in Volume ll, Chapter 14, "Softeningl'
B. What ls Tested
Common Range
6
mg/L as CaCOr
G. (Jar Test)
Calculation
3. Beakers (1,OOO mL)
Hardness, mg/t as CaCO3 _ A x 1,000 4. Pipers (10 mL)
mL ot Sample
5. Flask, volumetric (l,OOO mL)
_ 10.5 mL x 1,OOO
6. Balance, analytical
50 mL o, Sampte
D. Beag€nts
= 2_tO mg/L '1. Stock Coagulant Solulion
H. Precautions
a. Dry alum, Alr(SO,)3 . 14.3 HrO.15 Dissolve
1. Some melal ions interfere with this procedure by arum (r / percent) jn 600 ml distilled water
1O.O om drv
causing con-taine;
lading or indistinct end points. ln tnese in a 1,000-mL votumelric ftask. Filt to miri.
reagent should be used. you mav Iitrate
cases, an inhibitor
wiff, tion contains 1O,OOO mgr'l or iO mg/mt.
ilil;;:
et'
ard CDTA solution or EDTA sotution. "iit "i"ni-
2. The titration should be completed withjn
o
!E:,0, :!f, .Ar,(SoJ"
.
4e.6 H,o.,5 rhe operator
tve minules lo :11r,,",_y:r]ry,rh"
strength of rhe atum with a hydrom-
minimize CaC03 precipitation. erer. Lrqutd atum is usually shipped as 8.0
to 8.5 oer_
3. A sample volume should be selected that requires cent Al2Oj and conlains a6out b.g6 pound";;;li;-
less minum surate (48.5 percent orvl pii
than 15 mL of EDTA titrant to be used.
gravity 1.325). This converts ro leb,OO6l"rri,i flplJiii"
4. For titralions ol samples containing fore, add,'l5_6 mL tiquid atum to a f
mg/t:fi;;;:
low
hardness concen_ .OOO-mf"vof umeirt
(anons (tess Ihan 150 mg/L,
as CaCO.) a larger sampte lasx and ,i to mark. This solution contains IO,OOO
votume should be used- mglLot t0 mg/mL.
L Reference c. Table I 1.6 indicates the slrengths of stock solulions lor
varlOUS ClOSageS.
See page 2-37, STANDARD METHOD.S, 2jst Edition.
TABLE 11.5 DRY CHEMICAL CONCENTRATIONS
USED
QUESTIONS FOR JAR TESTING
'
Write your answers in a notebook
- ers with and then compare your 1 ml
Added to
ans those on page 576. Approx. Do3age GramVLlter 1 Litor Stock Sotution
Required, mg/Lb lo Preparec Sample Eq!.ts
11.3P What^are the principal hardness_causing Conc., mg/t (%)
ions in
l-10 mg/t
water? l gmtL 1 mllL 1,000 mg/a (0.17c)
10-50 mg/t 10
11.3Q What problems are caused by hardness EnlL 10 mgtL 10,000 mg/a (1.0?,
50 500 mg/L 1o0 g.t/L 1o0 mg/L
in waler? 100,000 mgla fi0.0%)
(Jar Test)
6. Beduce the stining speed for the ne)t 30 minules to 20 There are a number of tests lhat can be performed to im-
RPM.16 prove the jar test and the interpretation of lhe results. These
tests include:
OUTLINE OF PROCEDUFE 1. Alkalinity (before and afler)
2. pH (before and atter)
\I 3. Turbidity of SUPEFNATANT" (before and after)
4. Filtered turbidity ol supernatant
See Figure 1 1.22 for a helplul jar test data sheet.
NO 'NO 2 lro. s; NO,1 NO.5 NO6
After estimating the optimum coagulant dosage, run another
jar test with the optimum coagulant dosage constant, bui vary
the pH. These results will give you the optimum pH.
1. Add 1,000 ml lo each of 6 beakers Alkalinity must be monitored very carelully belore and atter
the jar test. Alkalinity must always be at least half of the coagu-
lant dose. For example, if the opiimum coagulant dose is 50
mg/l. the total alkalinity must be at least 25 mg/|. lI the natural
alkalinity is less than 25 mg/l, adjust lhe total alkalinity up to
25 mg/L by adding lime. For additional inlormalion on how to
calculate the amount ol lime needed lo increase the alkaliniry
0 see page 172, Section 5.243, "Arithmetic for Solids-Contact
Clarilication;'and page 173, Example 6.
2. Add increasing dosages of coagulant The following series ol alum dosages were added:
etl a,9
a
5
6
1.8
2.O
18
20
3. Stir for appropriate time period. Evaluate floc quality H. lnterpretation of Besults
Results of the above jar testing were recorded as follows:
7. Observe and evaluate each beaker as to lhat specific
dosage's floc quality. Record results. Beaker No. Alum Dose, mg/L Floc Ouality
r5
Use stirring speeds and times lhat are similar to aclual condilions in your water trealmenl planl.
t7
Supemalant \sue-perNAY.lent). Liquid removed lrom settled sludge. Supernatant commonly relgrs lo thB liquid bolwesn the sludge on the
bottom and lhe scum on lhe water surlace ol a basin or conlainer.
Lab Procedures 569
=o
-aE
=<
=
za.
il4sr
z Az2
= =H?
=
o -*rL
O
J;Y
=.; iii
> -.'!
9onI
>OJ
iio
o
(,
Ntht st
Nlv! 0e
4 Ntht 9z q)
o e
F N ll/'l 0Z
I o
z NIIEt Is
F 9 1
F
o a NIW OI .q
l- G
.9
F o NI S q
IC
F.'
lst
Hd
\
oflsntov
lsoo lr-
orv -cvo3
SSOO
cvo 3
A.tlO t€ a nl
o3all rB
>".3k
r >,.3k
tr
F
>",3a
: >"8e
I ZIS
3td htvs
cc
c(6
1>
'l
(pH)
Write your answers in a notebook and lhen compare your Common Range
answers with those on page 576. Surface water 6.5 to 8.5
11.3R What is the purpose of the jar test? Finished waier 7.5 to 9.0
Well water 6.5 to 8.0
11.3S What stirring speeds are used during the jar tests lor
optimum alum dosage?
C. Apparatus
1. pH meter
2. Glass electrode
3. Fleferenceelectrode
4. Magnelic stirrer
5. Magnetic stir-bar
D. Reagents
1. Butfer tablets for various pH value solutions (available
from laboratory chemical supply houses).
2. Distalled water.
7. pH
E. Procedure
A. Discussion
Due to the ditference between the various makes and
The pH ol a water indicates the intensity of its acidic or models ol pH meters commercially available, specific in-
basic strength. The pH scale runs irom 0 to 14. Water having structions cannot be provided for the correct operatjon of
a pH of 7 is at the midpoint of the scale and is considered all instruments. ln each case, follow the manufactureds in-
neutral. Such a water is neither acidic nor basic. A pH of structions for preparing the electrodes and operating the
grealer than 7 indicates basic waler The stronger the basic instrument.
intensity, the greater the pH. The opposite is true of lhe acidi-
ly. The stronger the intensity of the acidity, the lower will be 2. Standardize the instrument (pH meteo against a butfer
solution with a pH close to that of the sample.
the pH.
pH SCALE 3. Rinse electrodes thoroughly with distilled water after re-
moval trom butler solution.
O+ INCREASINGACID --_-7 - _- . INCREASING BASE -}14 4. Place electrodes in sample and measure pH.
5. Bemove electrodes lrom sample, rinse thoroughly with
, Neulral. distilled water.
Mathematically, pH is the logarithm (base '10) of the recipro- 6. lmmerse electrode ends in beaker of pH 7 buffer storage
cal ot the hydrogen ion activity, or the negative logarithm of lhe solution,
hydrogen ion activity.
1
7. Turn meter to standby (or OFF).
pH = Log *-
lH'l
= -Los [H*]
F. Precautions
'1,
For example, if a water has a pH of then the hydrogen ion 1. To avoid laulty instrument calibration, prepare fresh buffer
activity [H.] = 10 1 = 0.1. lf the pH is 7, then [H.] = 10i = solutions as needed, and ai least once per week (from
0.0000001. A change in the pH of one unit is caused by commercially available buffertablets).
changing the hydrogen ion level by a lactor ol 10 (10 times).
2. pH meter, buffer solution, and samples should all be near
ln a solution, bolh hydrogen ions [H.] and the hydroxyl ions the same temperature because temperature variations
IOH-I are always present. At a pH of 7. the activity of both hy- will give somewhat erroneous results. Allow a few minutes
drogen and hydroxyl ions equals 10-7 MOLES'8 per liter. for the probes to adjust to the butfers before calibrating a
When the pH is less than 7, the activity of hydrogen ions is pH meter to ensure accurate pH readings.
greater than lhe hydroxyl ions.
3. Watch for erratic results arising from electrodes, faulty
pH plays an important role in the water treaiment processes connections, or fouling of electrodes with aily precipi
such as disinfection, coagulation, soflening, and corrosion tated matter. Films may be removed from electrodes by
control. The pH test also indicates changes in raw and fin- placing isopropanol on a tissue or a Q-tip and cleaning
ished water quality. the probe.
(Temperature)
4. The temperature compensator on the pH meter adjusts gas concentrations. and water slability with respect lo calcium
the meler lor changes in electrode response with te;per- carbonate, rn addition lo its acceptability by consumors tor
alure. However, the pH ol water also changes with t;m_ drinking.
peralure and the pH meter cannot compensate for this
change. B. What ls Tested
5. We. recommend standardizing lhe pH meter using a pH Common Flange, oC
buller sotution close to the pH ol the sample. nowevei it Baw and Treated Surface Water
you use another butfer solution with a ditferent pH to de- 5to25
termine the calibration of the pH meler for a range ol pH Well Water 10 to 20
values and the pH meter does not give the pH of ihe
second butfer, follow the manufacturer,s directions to C. Apparatus
adiu_st the, pH meter. This procedure is calted adjusting
the "slope' of the pH meter. 1. One National lnstitute of Standards and Technology
(NIST) thermometer lor calibration ot the other thermori-
6. lt.you are measuring pH in colored samples or samples eters.
with high solids content, or if you are taking measure_
ments that need to be reported to the US EpA;you should 2. One Fahrenheit, mercuryJilled, '1" subdivided thermom-
use a pH electrode and meter instead ol a cblorimetric eter.
method or test papers. pH meters are capable of provid- 3. O-ne Celsius (Iormerly called Centigrade), mercuryJilled,
t
ing 0.1 pH accuracy in most applications. ln contrast, 'l' subdivided thermometer.
colorimetric lests provide : 0.1 pH accuracy onty in a lim-
iled range. pH papers provide even less accuracy. 4. One m6talcas€ to fit each lhermometer.
G. lnterpretation ol Flesutts NOIEr There are three types of thermom€ters and two
scales.
The pH ol water has a very imporlant tnfluence on lhe effec-
lr!€ness of chlorine disinfeclion. Chlorination is a chemical re- SCALES
action in which chlorine is an oxidizing agent. Chlorine is a more
efleclive oxidant at lower pH values. Simply slated. a chlorine
1. Fahrenheit, marked "F.
Esidual.ol 02 mg,/t at a pH of 7 is
'10.
iust as etfective as I mg/L at 2. Celsius, marked "C (formerly Centigrade).
a pH ol Therefore. live times as much chlorine is requiied to
do the same disinfecting job at pH 10 as it does at pH 7. THERMOMETER STYLES
The finished water of some treatment laciljties is adjusted 1. Total imm€rsion. This type of thermometer must be totally
or caustic soda to a slighfly basic pH tor the purpose
with.lime immersed when read. Readings with this type of thermom. I
dmrnrmrztng corrosion in the distribution system. eter will chang€ most rapidly when removed from the liquid
to be recorded.
H. Rererence
2. Parlial immersion. This type thermotneter will have a solid
See page 4-90, STANDARD METHODS. 21st Edirion. line (water-level indicator) around the stem below the point
where the scale starts.
OUEST]ONS 3. Dial. This type has a dial that can be easity read while the
i Wrile your answers in a notebook and then compare your thermometer is still immersed. Oial thermometer readjngs
0,$vers with thoso on page 576. should be checked (calibrated) against the NIST thermo;-
eter. Some dialthermometers can be recalibrated (adjusted)
il,3l What does the pH of a water indicate? to read at a set temperature against ihe NIST thermometei
ll.3u What precautions should be exercised when using a D. Beagents
i pH meter?
None required.
(, (l
i
E, Procedure I
t, Temperature
4. When measuring th€ temperature of well water samples,
allow the waler lo continuously overllow a small container
Discussion (a polystyrene cotfee cup is ideal). place th8 thermometer
in the cup. After ther€ has b6€n no change in th6 tempsr-
Temperature is one ol the most frequen y taken tests in the ature reading for ono minuts, record th6 tsmp€rature. ih6
industry Accurate water temperature readi ngs are impor- t€mperature ot waler samples collected from a distrlbu.
not only for historical purposes but also b ecauss of its in-
tion system mainly depends on th6 soiltemp6rature at the
on chemical reaction rates, biological growth, dissolved depth of the water main.
572 Water Treatment
(Turbidity)
F. Precautions suspended matter such as silt, linely divided organic and inor-
ganic matter, and microscopic organisms such as algae.
To avoid breaking or damaging a glass thermometer, store it
in a shielded metal case. Check your thermometer's accuracy The accepted method used to measure turbidity is called
against the NIST-certilied thermometer by measuring lhe tem- the nephelometric method. The nephelometric turbidimeter or
peralure ol a sample \,vith bolh thermometers simultaneously. nephelomeler (Figure 1'1.23) is designed to measure particle-
Some of the poorer quality thermometers are substantially in- reflected light at an angle of 90 degrees to the source beam.
accurate (off as much as 6'F or 3"C). The greater the intensity of scattered light, the higher lhe tur.
bidity.
G. Example
To measure the temperature of treated water, a sample was
obtained in a gallon bottle, the thermometer immediately im- NEPHELOMETEF
mersed, and a temperature ol 15"C recorded after the reading
METER
(-
became constanl.
c
H. Calculation
Normally, we measure and record temperatures using a ther- PHOTOCELL
momelerwith the proper scale. We could, however, measure a TURBIOITY PARTICLES
temperature in "C and convert to'F. The following formulas are
LAMP
I! SCATTER LIGHT
used lo convert temperalures from one scale to lhe other.
LENS
I 7II
1. Il we measure in 'F and wanl 'C,
"C=11"P-.r"t,
9
SAMPLE CELL
2. lf we measure in "C and want "F,
{vEBTTCAL VrEW}
"r=9t"o*sz"r
5
Fig. 11.23 Nephelometer
(Pemission ol I^IACH Compar,y)
3. Sample Calculation
The measured treated water temperature was 15'C. Whal fhe TURBIDITY UNlf (IU) r'g is an empirical quantity lhat is
is the lemperature in "F? based on the amount ol light that is scattered by particles ol a
polymer reference standard called formazin, which produces
"F = 9/s("C) +32"F particles that scafter light in a reproducible manner. Formazin.
= 9/5(15"C) + 32'F the primary turbidity standard, is an aqueous (waler-based)
suspension of an insoluble polymer formed by the condensa-
tion reaction between hydrazine sulfate and hexamethylene-
tetramine.
= 59"F
l. Fleference Secondary turbidity standards are suspensions of various
materials formulated to match the primary formazin solutions.
See page 2-61, STANDARD METHODS, 21st Edition. These secondary standards are generally used because of
their convenience and the instability ol dilute lormazin primary
QUESTIONS slandard solutions. Examples of these secondary standards
include standards that are supplied by the turbidimoier manu-
Write your answers in a notebook and then compare your facturer with the instrumenl. Periodic checks ol these second-
answers with those on page 577. ary standards against the primary lormazin slandard are a
must and will provide assurance of measurement accuracy.
1 1.3V Why are temperature readings important?
Th€ turbidity measurement is one of the most important
11.3W Why should the lhermometer remain immersed in
tests the plant operator performs. The Safe Drinking Water Act
the liquid while'being read?
stipulates specific monitoring requiremenls lor turbidity. Tur
11.3X Why should thermometers be calibrated against an bidity of treated surface water musl be measured every four
accurate NIST-certified thermometer? hours or conlinuously and the results reported to the appropri-
ate authority. Turbidity testing is the most critical tool in recog-
9. Turbidity nizing changes in raw water quality, detecling problems in co-
agulalion and sedimentation, and troubleshooting filtration
A. Discussion problems.
The term turbidity is simply an expression of the physical The maximum contaminant level (MCL) tor turbidity is one
cloudiness of water. Turbidity is caused by the presence ol TU with five TU allowed under certain circumstances. Water
1'1
Turbidity Unils (TU). Turbidity units are a m€asure ol th6 cloudin6s6 oI wat€r. ll measursd by a neph€lomskic (delloctod lighl) lnslrum€ntal pro-
cedure, turbidity units are expressed in nephelomelric turbidity units (NTU) or simply TU. Those turbidity units obtained by visual molhods ar6 6x.
pressed in Jackson Turbidity Unils (JTU) which are a measure of th6 cloudiness ol water; they are used to indicats the clarity ol water. Thers is no
real connection beh/veen NTUS and JTUS. The Jackson lurbadimeler is a visual method and lhe nephelom8ler is an inslrumental method based on
dellected lighl.
Lab Procedures 523
(Turbidity)
lreatment plant operators should strive to produce
a finjshed the mark and mix. The lurbidity of this suspension is
water with a turbidity of 0.1 TU or less.
considered 400 NTU exacfly.
. For.additional intormation
lurbidity requirements
on monitoring requirements and d. Prepare solutions and suspensions monthly.
of the Surface Wa'ier teatment Buie
ll[Il.^=, rrllel. chaprer zz, ,,ortnitng witei
Erons. Seclion 22.221, and the poster includeO wrtn tiis
hd; 3. Slandard lurbidjty suspensions: Dilute .tO.OO
mL stock
rurototty suspension to 100 mL with turbidityjree water.
manual. Prepare weekly. The turbidity of rnis suspeirsion ts
Oe-
lined as 40 NTU-
B. What ls Tested
4. Oilute_ turbidity standards: Dilute portions of the
standard
Common Range, NTU turbidity suspension wilh turbidity-free *ut"l' u. *lrirrO.
Untreated Surlace Waler '1 Prepare weekly.
to 300
Filtered Water 0.03 to 0.50 E. Procedure
Well Water 0.05 to 1.0+ 1. Turbidimeter calibration: The manulaclurer.s ooeratino in-
structions should be followed. Measure your sianOarO"so-
C. Apparatus lutions on the turbidimeter covering the iang" ot int"i;.
lf the inslrument is already calibral;d in sta;dard turbiditu
1. Turbidimeter: To minimize ditferences in turbidity
meas- units, this procedure will check the accuracy of the
I urements, rigorous specilications lor turbidimeiers cali'-
are bration scales. At least one standard shouid be;u;;n
necessary The turbidimeter should have the lollowing
im- each. instrument range lo be used. Some in"tirr"ni"
portant characteristics: permit adjustments of sensitivity so that
scate vatueJwi
a. The turbidimeter should consist ol a nephelomeler correspond to turbidities. Fleliance on an initrJmeni
with a light source illuminating the sample.'and manufaclurer's scattering standards lor cafiOrafin!
one oi tfie in_
more photoelectric detectors to indicale the intensity not an acceptabte practice unless rhe-y are in
scaflered light at a 90. angle wilh a readout device.
ol :tM91j
acceplabty" close agreement with prepared standaids.
lf a
precalibrat€d scale is not supplied, tiren
b. The light source should be an intense tungsten curves should be prepared for each usable r.ange
fila- ""liOr"tion
oi ih"
ment lamp. rnslrumenl-
c. The tolal drslance lraveled by the lighl through the 2. Turbidities less than 40 units: Shake the sample
to thor-
sampte should be less than about 5 ceitimelers: oughly disperse the suspended sotids. Wait ,ntit ai,
OrU-
d. The instrument should have several measurement ores otsappea( lhen pour the sample into the turbidimeter
ranges. The instrument should be able to measure ruoe. Head the turbidity direc y lrom the instrument scale
lrom 0 to 1OO turbidity units, with sufficient or trom lhe appropriate calibration curve.
sensitivitu
{on the lowest scale) to detect dif,erences ol 0.O2 or 3. Turbidities exceeding 40 uniis: Dilute the sample with
one
tess in. filtered walers having turbidilies ol less or more volumes of turbidity{ree water until ihe lurbiditv
tha;
one unit rals below 40 units. The turbidity of the orioinal samole il
2. Sample lubes. These are usually provided with the rnstru- then compuled from lhe turbidity ot tne dituted sampte
ment. and the dilution tactor. I
F. Example i
0 Reagents
sampte was coltected and the lurbidity was
l. Turbidityjree water: pass dis lled water through a ,^.tj?'?.oi,
round to be greater than 40 units when first checked. 30 mL of
mem-
brane tilteriaving a pore size no grealer than this sampte was then diluted with 60 mL of turbiOity-11g;;fter.
012 microns
llrlElte lroT taboralory suppty houses). Discard the
lirsl200 mL coltected. lf lillration does not reduce
This diluted sampte showed a turbidity of 25 units.
turbiditv
olthe disti ed water. use unfittered disti ed wa;; - - - ' NTU Found in Diluted Sample, A = 25 NTU
2. Slock formazin turbidity suspension:e mL of Dilution Water Used, B = 60 mL
a. Solution I-Dissotve 1.OOO gm hydrazine sullate
in dis-
mL ol Sample Volume Taken lor Dilution. C = 30 mL
tr ed water and dilule to I OO mL in a volumetnc ,lask. G. Calcutation
b. Solution.ll--Drssotve 10.00 gm hexamelhylenetetra-
mrne rn distiled water and dilute to 100 mL in Nephelometric Turbiditv Units (NTU) (A,*?
a volu- =
metric Ilask.
c. ln a 100-mL volumetric llask. add (using _ (2s NTU)X(60 mL + so mL)
S_mL volumet-
rc prpels) 5.0 mL Sotution I and 5.0 m7 of Sotutron 30 mL
ll.
(Turbidity)
H. lnterpretalion ot Flesults Section A.139, "Laboratory Procedurss." Check all ol the
arithmetic in this section using an electronic pock6t calcula'
Report turbidity results as lollows tor. You should be able to get the sams answers.
Turbidity Reading Record to Nearest
I1.5 ADOITIONALREADING
0.0 lo 1.0 0.05
1. NEw YORK MANIJAL, Chaplet 4,' 'water Chemistry,'
1 to 10 0.1
Chapter 5,' "Microbiology,' and Chapter 21,' "Laboratory
10 to 40 1
Examinations ol Waler."
40 to 100 5
2. TExAs MAN|)AL, Chapler 6,* 'Water Chemistry,' and
100 to '1 ,000 10 Chapter 12,"Laboratory Examinations.'
>1,000 100
' Depends on odition.
(Lesson5of5Lessons)
Write the answers to these questions in your notebook. The 2'1. what precautions should be exercised when taking tem-
question numbering continues from Lesson 4. perature measuremgnls?
19. What precautions should be considered when perform- 22. Why should turbidity be measured in a sampl€ as soon as
ing lh€ hardness determination on a water sample? possibl€?
20. How could you estimate the most etfective dose ol alum
or a polymer in a water treatment process?
.T
Lab Procedures 575
SUGGESTED ANSWERS
Chapter 11. LABORATORY PROCEDURES
sizes lrom 10 mL to accurate for test results to have any significant meaning. With-
1,000 mL volumes used out a representative sample, the test results will not
in titralions reflect actual water conditions.
'1
1.3C The benefats of chlorinating water include:
11.30 The total coliform test is not an adequate measure of
the extent o, human wasles in water because it
1. Disinlection measures both fecal colilorms (E col, and soil coli-
2. lmproved quality of water due to chlorine reacting lorms. Additional tesls such as the MF-mTEC
with iron, manganese, protein substances, sul- method are needed to specifically identity and count
fide, and many taste- and odor-producing com- lecalcolilorms.
pounds
3. Control of microorganisms thal might interfere ANSWERS TO OUESTIONS IN LESSON 5
wilh coagulation and llocculalion
4. Keeps filler media free oi slime groMhs Answers lo queslions on page 567.
5. Helps bleach out undesirable color 1 1.3P The principal hardness-causing ions in water are caL
11.3D A potential adverse effect from chlorination is the cium and magnesium.
possibility of the formation of carcinogenic chloror- '11.3O Problems caused by hardness in water include: (1)
ganic compounds such as chloroform and other incrustations when water evaporates or scale when
THMs. healed. and (2) formation ol precipitates when com-
Answers lo questions on page 543. bined with soap.
'1
1.3E Conditions that can cause variations in the chlorine Answers to questions on pag€ 570.
demand of water include: (1) the amount of chlorine 11.38 The jar test is us€d to: (1) dolermine lh€ correct
applied, (2) length of contacl time, (3) pH, (4) lemper- coagulant dosage, and (2) evaluate new coagulants
ature, (5) organics, and (6) reducing agenls. or polymers.
11.3F The operator uses lhe chlorine demand test to deter-
11.3S Speeds used during the lar lest are as follows:
mine the best chlorine dosage to achieve specific
chlorination objectives. 1. 80 RPN4 for the first minute
1
'1.3G chror,ne 2. 20 RPN,i lor the nexl 30 minutes
Demand. = Chlorne Added. mg/L - Free Residual Chlodne, mg/L 3. Stop stirring (0 RPM) for the next 30 minules
lr,gJL
= 20 nglL- O-a .'..glL These speeds and times should be adjusted if nec-
essary to be similar to actual conditions in the wat€r
= 1.6 mgy'L treatment plant.
ANSWERS TO QUESTIONS IN LESSON 4 Answers to questions on page 571.
Answers to questions on page 565 'l 1.3T The pH o, a h/ater indicales lhe intensity ol its basic
or acidic strength.
11.3H Drinking waters are not tested for specific pathogens
because the tests are very time-consuming and 11.3U Precautions to be exercised wh€n using a pH meler
require special techniques and equipment. include:
'1
1.31 The number ol samples required for colilorm tests is 1 .
Prepare tresh butfer solution weekly for calibration
purposes.
generally based on the population served by the
water system. 2. Have pH meter, samples, and butfer solutions all
near the same temperature,
11.3J Sodium thiosulfate should be added to sample bot- 3. Watch lor erratic results arising from olectrodes.
tles for coliform tests to neulralize any residual chlo- faulty connections, or fouling of electrodos with
rine in the sample. lf residual chlorine is not interrering matier.
neutralized. it will continue to be toxic to the coliform 4. The pH of water changes with temperature.
organisms remaining in lhe sample and give false 5. lf testing waters with a rangs of pH values, the
(negalive) results. slope of th€ pH meter may require adjusting.
t
Lab Procedures 577
t
Answers to questions on page 572 Answers to questions on page 574.
11.3V Temperature readings are important because tem- 11.3Y Turbidity in waler can be caused by the presence ol
peralure influences chemical reaction rates, biologi-
cal groMh, dissolved gas concentrations, and water
suspended matter such as silt, finely divided organic
and inorganic matter, and microscoptc organisms
I
stability with respecl to calcium carbonate. Also, con- such as algae,
SUmers are sensitive to the temperature of the water
they drink. 11 .32 Turbidity is measured by the nephelometric method.
E
11.3W The thermometer should remain immersed in the
liquid while being read for accurate results. When
11.3X
removed from the liquid, the reading will change.
All thermometers should be calibrated against an
Qt E
accurale National lnstitule of Standards and Technol
ogy (NIST) lhermomeler because some poorer qual-
ity thermomelers are substantially inaccurate (otf as
much as 6"F or 3'C). I
I
E
E
T
!
E
Ir
FI
EI
EI
Ei
Bi
Ir
E
E
I
t
!
i
di
I
I
lr
I
Ijli
ll
I
i
I