You are on page 1of 67

Nuclear Medicine and Molecular

Imaging: the Requisites 5th Edition


Janis P. O'Malley
Visit to download the full and correct content document:
https://ebookmass.com/product/nuclear-medicine-and-molecular-imaging-the-requisit
es-5th-edition-janis-p-omalley/
THE REQUISITES

NUCLEAR MEDICINE
AND MOLECULAR IMAGING
5th EDITION

JANIS P. O’MALLEY, MD, FACR Series Editor


Professor of Radiology
University of Alabama at Birmingham JAMES H. THRALL, MD
Division of Molecular Imaging & Therapeutics Radiologist-in-Chief Emeritus
Birmingham, Alabama Massachusetts General Hospital
Distinguished Juan M. Taveras Professor of
Radiology
HARVEY A. ZIESSMAN, MD Harvard Medical School
Professor of Radiology Boston, Massachusetts
Division of Nuclear Medicine and Molecular Imaging
The Johns Hopkins University
Baltimore, Maryland
Elsevier
1600 John F. Kennedy Blvd.
Ste 1800
Philadelphia, PA 19103-2899

NUCLEAR MEDICINE AND MOLECULAR IMAGING:


THE REQUISITES, 5th EDITION ISBN: 978-0-323-530378
Copyright © 2021, Elsevier Inc. All rights reserved.
2014, 2006, 2001, 1995 by Mosby, Inc., an affiliate of Elsevier Inc.

No part of this publication may be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying, recording, or any information storage and retrieval system, without
permission in writing from the Publisher. Details on how to seek permission, further information about the
Publisher’s permissions policies, and our arrangements with organizations such as the Copyright Clearance
Center and the Copyright Licensing Agency can be found at our website: www.elsevier.com/permissions.

This book and the individual contributions contained in it are protected under copyright by the Publisher
(other than as may be noted herein).

Notices

Knowledge and best practice in this field are constantly changing. As new research and experience broad-
en our understanding, changes in research methods, professional practices, or medical treatment may
become necessary.
Practitioners and researchers must always rely on their own experience and knowledge in evaluating
and using any information, methods, compounds, or experiments described herein. In using such infor-
mation or methods, they should be mindful of their own safety and the safety of others, including parties
for whom they have a professional responsibility.
With respect to any drug or pharmaceutical products identified, readers are advised to check the most
current information provided (i) on procedures featured or (ii) by the manufacturer of each product to
be administered, to verify the recommended dose or formula, the method and duration of administra-
tion, and contraindications. It is the responsibility of practitioners, relying on their own experience and
knowledge of their patients, to make diagnoses, to determine dosages and the best treatment for each
individual patient, and to take all appropriate safety precautions.
To the fullest extent of the law, neither the Publisher nor the authors, contributors, or editors as-
sume any liability for any injury and/or damage to persons or property as a matter of products liability,
negligence, or otherwise or from any use or operation of any methods, products, instructions, or ideas
contained in the material herein.

Library of Congress Control Number: 2019941296

Senior Content Strategist: Kayla Wolfe


Content Development Specialist: Ann R. Anderson
Publishing Services Manager: Shereen Jameel
Project Manager: Radhika Sivalingam
Designer: Brian Salisbury

Printed in the People’s Republic of China

Last digit is the print number: 9 8 7 6 5 4  


3 2 1
CONTRIBUTOR

Frederic H. Fahey, DSc


Director of Nuclear Medicine/PET Physics, Department of Radiology,
Boston Children’s Hospital, Boston; Professor, Department of
Radiology, Harvard Medical School, Boston, Massachusetts
Chapter 1: Radioactivity and Radionuclides;
Chapter 2: Radiation Detection and Ancillary Instrumentation;
Chapter 3: Single-Photon Emission Tomography, Positron Emission
Computed Tomography, and Hybrid Imaging

iii
F OR EWOR D

Nuclear Medicine and Molecular Imaging: The Requisites is a new Readers of Nuclear Medicine and Molecular Imaging: The Requi-
title for this well-received text now appearing its fifth edition. sites will feel this vitality almost palpably as they work their way
The change in title reflects the remarkable advances in trac- through the book. In particular, PET and PET/CT have become
er-based medical imaging that have taken place over the last two cornerstones in cancer diagnosis and management with PET/
decades. These advances have clearly extended the diagnostic MRI becoming more important in both cancer diagnosis and
utility and value of nuclear medicine and molecular imaging in neurological studies.
clinical patient care. The Requisites in Radiology titles have now become old
Predictably much of the fifth edition of Nuclear Medicine and friends to generations of radiologists. The original intent of the
Molecular Imaging: The Requisites is focused on new molecular series was to provide the resident or fellow with a text that might
imaging methods and the latest advances in their clinical appli- be reasonably read within several days at the beginning of each
cation including positron emission tomography (PET), SPECT/ subspecialty rotation and perhaps reread several times during
CT, PET/CT and PET/MRI hybrid imaging. The intense interest subsequent rotations or during board preparation. The series is
in hybrid imaging provides clear recognition of the increased not intended to be exhaustive but rather to provide the basic
value now placed on functional and molecular information in conceptual, factual, and interpretive material required for clin-
disease diagnosis. ical practice. After more than 30 years of experience with the
Although retitled, Nuclear Medicine and Molecular Imaging: series, it is now clear that the books are also sought out by prac-
The Requisites continues to follow the format of the first four ticing imaging specialists for the efficiency of their presenta-
editions. The basic science chapters are designed to present tion format and the quality of their material. With more people
important principles of physics, instrumentation, and nuclear reaching the point of requiring re-certification, the Requisites
pharmacy in the context of how they help shape clinical prac- books are again proving helpful.
tice. The physics content of the fifth edition has been expanded The first four editions of Nuclear Medicine and Molecular
and integrated to reflect current technology. Topics on regula- Imaging: The Requisites were well received in the radiology
tory issues, radiation safety and quality control have been added and nuclear medicine community. For the retitled fifth edi-
as well as material on the “non-interpretive” aspects of nuclear tion, Dr. Janis M. O’Malley and Dr. Harvey A. Ziessman have
medicine and molecular imaging practice. again done a terrific job in putting together this substantially
The clinical chapters continue to follow a logical progression updated edition. Congratulations to them. I expect that this
from basic principles of tracer distribution and localization to fifth edition will be deemed to be even more outstanding than
practical clinical applications. Knowledge of how radiopharma- its predecessors.
ceuticals localize temporally and spatially in normal and dis- We hope that Nuclear Medicine and Molecular Imaging: The
eased tissues is the best deductive tool available for analyzing Requisites will serve residents in radiology as a concise and
images. The best use of new tracers such as Ga-68 DOTA, the useful introduction to the subject and will also serve as a very
F-18 amyloid agents and F-18 PMSA agents requires this kind manageable text for review by fellows and practicing nuclear
of knowledge of the underlying mechanisms of disease and medicine specialists and radiologists.
therewith tracer localization. James H. Thrall, MD
Adding new tracers and new SPECT, PET, PET/CT, and Radiologist-in-Chief Emeritus, Massachusetts General Hospital
PET/MRI applications to the nuclear medicine armamentar- Distinguished Juan M. Taveras Professor of Radiology, Harvard
ium has injected new, unprecedented, vitality into the specialty. Medical School, Boston, Massachusetts

iv
P R E FA C E T O T H E 5 T H E D I T I O N
OF THE REQUISITES, NUCLEAR
MEDICINE

This is the 5th edition of Nuclear Medicine: The Requisites. How- agents highlighted. The popular chapter, “Pearls, Pitfalls, and
ever, we are now titled Nuclear Medicine and Molecular Imaging: Frequently Asked Questions,” again provides an excellent con-
The Requisites. We are deeply indebted to Dr. James H. Thrall for cluding summary. Protocols and key facts are again organized
developing the concept of the Requisite series and his involve- in many boxes and tables for easy identification.
ment as a coauthor of Nuclear Medicine: The Requisites for the Over the years, it has been an honor to help guide and train
first four editions, in 1995, 2001, 2006, and 2014. Since the last the most incredibly gifted physicians and introduce them to
edition, there have been many exciting changes in the field, par- the incredible field of nuclear medicine. Our students have
ticularly concerning new PET agents and therapy techniques. also taught us a lot along the way, and their feedback has been
The 5th edition builds on the success of the prior editions, pro- essential as we continue to strive to improve as educators. In
viding a concise, easy to read review that is suitable not only for addition, hearing from colleagues around the globe how the text
radiology and nuclear medicine residents and fellows preparing has helped them or their trainees has been another wonderful
for their service rotations or board exams but also serves a use- way to continue developing new material. Hopefully, our expe-
ful tool for those in practice at all levels of expertise, particularly riences are reflected here and provide a foundation for another
when targeting knowledge gaps during maintenance of board successful text.
certification reviews.
All chapters have been significantly updated and contain
numerous stunning new images. The first section of the book
ACKNOWLEDGEMENTS
is again devoted to technical matters: basic principles and con- We would like to thank those who have contributed to the
cepts of radiation production, instrumentation and detection, preparation of this book. Selected images were provided by:
radiopharmaceuticals and quality control, radiation safety, and Suzanne Lapi, PhD; Kirk Fry, MD, PhD; Jonathon McConathy,
regulatory matters. New topics have been introduced including MD, PhD; Steven P Rowe, MD, Bital Savir Baruch MD, Corina
PET/MR, and important facts the Authorized Users of radio- M Millo, MD; Khun Visith Keu, MD; Lauren L Radford, PhD;
pharmaceuticals are highlighted. The second section of the Mark Muzi, PhD; Les Foto; Farrokh Dehdashti, MD; and Hong-
book is focused on clinical imaging and therapy, emphasizing gang Liu. Hong-gang Liu also produced some of graphics in
physiological mechanisms and pharmacokinetics. Because animated images. Suzy Lapi and Jon McConathy helped with
of the rapid progress in oncology, particularly in the areas of editing the chapters on brain imaging and Molecular Imaging.
prostate cancer and neuroendocrine tumor imaging and treat- We would also like to thank our spouses and family for all
ment, these section have been extensively updated with import- their support during our work on each of these editions.
ant details on the use of newly approved imaging and therapy Janis O’Malley thanks her mom, Lanis Petrik.

v
PART 1 Basic Principles 10 Gastrointestinal System, 220
11 Urinary Tract. 256
1 Radioactivity and Radionuclides, 1
12 Oncology: F-18 Fluorodeoxyglucose Positron
2 Radiation Detection & Ancillary Instrumentation, 14
Emission Tomography, 288
3 Single-Photon Emission Computed Tomography, Positron
13 Oncology-Beyond Fluorodeoxyglucose, 339
Emission Tomography, and Hybrid Imaging, 28
14 Central Nervous System, 364
4 Radiopharmaceuticals, 42
15 Inflammation and Infection, 409
5 Molecular Imaging, 64
16 Cardiovascular System, 441
17 Pearls, Pitfalls, and Frequently Asked Questions, 480
PART 2 Clinical Studies
6 The Skeletal System, 75
Appendix 1: Dosimet ry, 493
7 The Pulmonary System, 125
Appendix 2: The Periodic Table of the Elements, 495
Index, 496
8 Endocrine System, 152
9 Hepatic, Biliary, and Splenic Scintigraphy, 180

vi
PART 1 Basic Principles

1
Radioactivity and Radionuclides

In nuclear medicine, radiopharmaceuticals given to the patient Elements are organized in the periodic table of the elements
emit the radiation used to create images or perform therapy. In (see Appendix 2). All atoms of the same element have the same
order to understand how these agents perform and what safety number of protons. The proton number is also referred to as the
considerations are involved in their use, it is necessary to be famil- atomic number or Z. Thus, all carbon atoms have 6 protons, all oxy-
iar with some basic aspects of the physics behind radioactive decay. gen atoms have 8 protons, and all iodine atoms have 53 protons—
This chapter discusses radioactive molecules, different types of that is, they have Z numbers of 6, 8, and 53, respectively. Atoms of
radioactive decay, and how these emissions interact with matter. a particular element can, however, have a varying number of neu-
trons (referred to as the neutron number, N). For example, in addi-
ATOMIC STRUCTURE OF MATTER tion to their 8 protons, some oxygen atoms have 8 neutrons, and
others have 7 or 10 neutrons. The total number of nucleons (Z plus
Electronic Structure of the Nucleus N) is known as the atomic mass or atomic number, A. Therefore, in
All matter is made up of atoms, which in turn are made up of the oxygen example, A would be 15, 16, and 18 for atoms that have
protons, electrons, and neutrons. Positively charged protons 8 protons plus 7, 8, or 10 neutrons, respectively.
and uncharged neutrons have a similar mass and are known as Unlike an element, which is characterized only by its number
nucleons because they are located in the nucleus. Although of protons (Z), a nuclide is a nuclear entity characterized by a
much less massive, electrons orbiting the nucleus possess an certain nuclear composition of protons and neutrons as well as
opposite negative charge equal in magnitude to that of the pro- a certain energy level. Shorthand notation has been agreed on to
tons (Table 1.1). Some properties of atomic particles are listed, describe the makeup of specific nuclides:
along with important constant values, in Table 1.2.
Atomic mass Element
The attraction of the opposite charges keeping the electron in
orbit around the nucleus is known as an electrostatic force (or A
ZX N
coulombic force; the coulomb is the unit for electric charge). On
the other hand, there is also a repulsive, electrostatic force in the
nucleus from the like-charged protons pushing apart. The Proton number Neutron number
nucleus is held together by the attractive strong nuclear force
each nucleon exerts on the other nucleons. Although more To illustrate this, consider the element iodine, which has 53
powerful than electrical forces, these strong forces act only over protons (Z = 53). If one particular nuclide of the element iodine
extremely short distances. The actual atomic mass is less than has 78 neutrons (N = 78), the atomic mass (A) of 53 + 78 equals
the sum of the masses of all its nucleons. This difference in mass, 131. It can be written as:
or mass deficit, is manifest in the nuclear binding energy hold-
A
ing the nucleus together (as related by the equation E = mc2).
131
I
53 78
TABLE 1.1 Properties of Atomic Particles Z
Mass N
Particle Charge (amu or u)a Mass (MeV)b Mass (kg)
Because the atomic number can be inferred by the element’s
Proton +1 1.0073 938.21 1.673 × 10–27
symbol, and N = A – Z, this can be shortened:
Neutron 0 1.0087 939.51 1.675 × 10–27
Electron –1 0.000549 0.511 9.11 × 10–31 131 131 131
I I I
53 78 78
aOne amu = 1.661 × 10–27 kg or 1/12 atomic mass carbon (1 nucleon
from carbon-12 atom). This can also be written as I-131 or iodine-131. The term iso-
bEnergy as related by E = mc2.
tope describes nuclides of the same element, that is, nuclides
1
2 PART 1 Basic Principles

TABLE 1.2 Summary of Physical Constants BOX 1.1 Important Terms Related
Unit of charge 1 amp·sec to Atomic Matter
Coulomb (C) 6.24 × 1018 electrons Nucleon—components of the atomic nucleus: protons and neutrons
Electron volt (eV) 1.602 × 10–19 J Atomic number Number of protons, or Z
Charge of 1 electron –1.6 × 10–19 C Neutron number Number of neutrons, or N
Charge of 1 proton + 1.6 × 10–19 C Atomic mass Sum of the nucleons—protons and neutrons (Z + N)—or
Planck’s constant (h) 6.63 × 10–34 m2·kg/s atomic number or A
Elements Atoms with the same number of protons (Z)
Avogadro’s number 6.02 × 1023 molecules/g·mole
Nuclides Nuclear entity comprised of a particular number of protons
Calorie (cal) 4.2 Joules (Z) and neutrons (N) as well as energy state of the
Speed of light in a vacuum 3.0 × 108 m/sec nucleus
Angstrom (Å) 10–10 m Radionuclides Unstable nuclides: isotopes emitting radiation attempting
to reach stability
Isotopes Atoms with the same number of protons: P for proton,
with the same number of protons (Z) but potentially differing P for isotope.
atomic numbers. Radioisotopes are isotopes that undergo radio- Isotones Atoms with the same number of neutrons: N for isotone
active decay. For example, some common isotopes of iodine are Isobars Atoms with the same atomic number A: A for isobar
as follows: Isomer Nuclide with same Z and N (so same A) but a different
energy level
131
I I
125
I
124 123
I
53 78 , 53 72, 53 71, 53 70
M for metastable; M for isomer.
In medicine, different isotopes have varying properties, such
as the types of radiation they emit and how long they remain
radioactive, which can determine their usefulness. For example,
the beta and high-energy gamma emitter 131I (I-131) is used for
treating thyroid cancer and performing thyroid uptake mea-
surements; 125I (I-125), a low-energy gamma and x-ray emitter,
is used in biological assays and prostate cancer brachytherapy;
124I (I-124), a positron emitter, can image thyroid cancer with a

positron emission tomography (PET) scanner; and 123I, a mod-


erate-energy gamma emitter, is very commonly used to image
benign thyroid diseases and thyroid cancers as well as to calcu- N M L K Nucleus
Z = 19
late thyroid activity (radioactive iodine uptake).
In addition to isotopes, other special terms include isotones,
nuclides with the same number of neutrons
(e.g.,148 O, 137 N, 126 C where N = 6); and isobars, those with simi-
lar atomic mass numbers (e.g., 14O, 14N, 14C). Nuclides that
have the same Z and N numbers (and, therefore, A) but differ in
their energy states are called isomers. A well-known example of
an isomer in nuclear medicine is technetium-99 (Tc-99) and its
metastable state technetium-99m (Tc-99m). Several key terms
Fig. 1.1 Bohr model diagram of the potassium atom. Potassium has an
to know concerning atomic structure are listed in Box 1.1. atomic (Z) number of 19; that is, it has 19 protons in the nucleus and 19
orbital electrons.
Structure of the Orbital Electrons
Our understanding of the atom has evolved, but it is still useful ELECTROMAGNETIC RADIATION
to picture the classic Bohr atom (Fig. 1.1) with electrons arrang-
ing themselves into discrete orbital shells (Table 1.3). The inner- Electromagnetic (EM) radiations, such as visible light, have
most shell is referred to as the K shell, and subsequent shells are long been known to have a duality to their nature: behaving in
referred to as L, M, N, O, and beyond. Each shell holds only a set some situations as a wave and in others as a particle, or photon.
maximum number of electrons, given by 2n2, where n is the The EM spectrum (Fig. 1.3) varies in wavelength and frequency,
shell number). Based on this, for example, the K shell (n = 1) from low-energy radio waves up to high-energy x-rays and
contains 2 electrons, and the L shell (n = 2) has 8. gamma (γ) rays as used in medical imaging and therapy.
Because electrons are bound by the electrical forces, energy The unit of energy typically used in atomic and nuclear phys-
is required to remove an electron from an atom. This orbital ics is the electron volt (eV), which is the amount of energy an
binding energy (BE) is characteristic for each particular atom, electron garners when crossing an electronic potential difference
depending on its Z number, as well as which shell is involved of 1 volt. One eV is equivalent to 1.6 × 10–9 joules. EM radiations
(i.e., it is harder to remove an inner-shell electron than an out- travel at the speed of light (c) with the known relationship:
er-shell electron; Fig. 1.2). c = vλ
Chapter 1 Radioactivity and Radionuclides 3

TABLE 1.3 Terms Used to Describe where ν is frequency, λ is the wavelength, and c = 3 × 108 m/s.
Electrons The photon energy (E) is related to the frequency of the EM
wave by
Term Comment
Electron Basic elementary particle E = hv
Orbital electron Electron in one of the shells or orbits in an atom where h is Planck’s constant (6.626 × 10–34 J/s)
Electron in the outermost shell of an atom; responsible
Valence electron for chemical characteristics and reactivity Relating these equations, v = c so E = hc , thus:
Electron ejected from an atomic orbit by energy
Auger electron released during an electron transition 12.4
E (keV) = with the λ measured in angstroms Å .
Electron ejected from an atomic orbit because of Å
internal conversion phenomenon as energy is given
Conversion electron off by an unstable nucleus Visible light has energy slightly less than 1 eV, whereas x-rays
Electron ejected from an atomic orbit as a conse- and gamma rays have energies in the range of several thousand
quence of an interaction with a photon (photoelectric eV (or keV) to tens of millions eV (MeV).
interaction) and complete absorption of the photon’s
X-rays and gamma-ray photons do not differ in their energy
Photoelectron energy
levels but in their origin. X-rays are generated from interactions
Electron ejected from orbit after absorbing a portion of
outside the nucleus, whereas gamma rays are generated by tran-
Compton electron a photon’s energy during Compton scatter
sitions within the nucleus. Once created, nothing distinguishes
an x-ray from a gamma ray (e.g., A 100-keV x-ray is absolutely
identical and indistinguishable from a 100-keV gamma ray).
Potential Binding
Energy Energy Production of X-Rays
orbital shell
X-rays are produced in two ways: (1) as a result of the transition
Highest Lowest of atomic electrons from one orbit to another and (2) from the
N
deacceleration of passing charged particles as they interact with
other charged particles, usually as a result of columbic electrical
M
interactions.
L
Characteristic X-Rays
In the first instance, excited electrons may be removed from their
K
atomic orbit or elevated to a higher-energy orbit. An electron from
an outer orbit can drop down to fill the vacancy, and the excess
energy, the difference in the binding energy of the shells, can be
nucleus emitted as an x-ray photon, a fluorescent x-ray (Fig. 1.4A). This is
also known as a characteristic x-ray because it is specific to not only
each element but also to the orbital shell from which it originated.
Lowest Highest Consider the case of fluorescent or characteristic x-rays from elec-
Fig. 1.2 Orbital binding energy. tronic transitions within an iodine atom with the following binding

Photon energy 106 103 100 101 106 1010


electron Volt (eV) 1MeV 1keV 1eV

Increasing Wavelength

0.0001 nm 0.01 nm 10 nm 1000 nm 0.01 cm 1 cm 1m 100 m

Ultra-
Gamma rays X-rays Infrared Radio waves
violet
Radar TV FM AM

Visible light
Fig. 1.3 Electromagnetic energy spectrum. Photon energies (eV) and wavelengths of x-rays and gamma
ultraviolet, visible light, infrared, and radio waves.
4 PART 1 Basic Principles

Characteristic
Orbital Orbital energy transfer
X-ray
electron electron
vacancy e- vacancy e-

e- e-
M L K n e- M L K n Auger
electron

A B
Fig. 1.4 Interactions may result in (A, left) emission of characteristic (fluorescent) x-rays or (B, right) Auger
electrons.

energies: K = 35, L = 5, and M = 1 keV. Thus, the energy of the flu- 160
orescent x-rays resulting from the transition of electrons from the L Neutron rich
to the K shell (referred to as Kα fluorescent x-rays) is 30 keV 140 (Beta minus decay) Proton rich
(35 − 5 keV) and that from the transition from the M to the K shell
(referred to as Kβ x-rays) is 34 keV (35 − 1 keV). (Electron
capture
120 and
Auger Electrons positron
decay)
There is an alternative outcome to characteristic x-ray emission, 100
where the transition can cause an outer-shell electron to be ejected.


ility
This electron is called an Auger (pronounced oh-zhey) electron
Neutron number N

tab
80
(see Fig. 1.4B). The kinetic energy (KE) of the resultant Auger elec-

f “s
tron is determined by the binding energy of the orbits involved:

eo

N
Lin
60


KEAuger = BEInner − BEOuter − BEAuger

Z
Using the example of the iodine atom binding energies again for
40
the transitions shown in Fig. 1.4B, the calculation would then be:
KEAuger = BEK shell − BEL shell − BEM shell
20
= 35keV − 5keV − 1keV = 29keV
The probability of an Auger electron being emitted is greater
in lower Z elements and from outer shells where the binding 0
energy is lower. X-ray fluorescence, on the other hand, is the 20 40 60 80 100
more likely outcome when binding energy is higher, such as in Atomic number Z
higher Z elements and from inner-shell electrons. Fig. 1.5 Graph of neutrons (N) versus protons (P) for various nuclides. For
elements with a low atomic number, the two are roughly equal (Z = N).
Particle Deceleration and Bremsstrahlung X-Rays With increasing atomic number, the relative number of neutrons increases.
Stable nuclear species tend to occur along the “curve of stability.”
X-rays can also be produced as a charged particle deaccelerates
as it passes an atom. In nuclear medicine, this commonly
involves electrons or beta particles passing through soft tissue.
RADIOACTIVITY AND RADIOACTIVE DECAY
In this case, the negatively charged particle is slowed as it inter- The protons and neutrons can only exist in the nucleus in a
acts with the positively charged nucleus it is passing, causing it limited number of combinations. The remaining unstable
to slow. The energy it loses is emitted as radiation referred to as atoms may transform to a lower-energy stable state through
bremsstrahlung (from the German for “braking”) radiation. The radioactive decay (or disintegration), with the excess energy
magnitude of the bremsstrahlung production increases linearly resulting in either particulate emissions or electromagnetic
with the kinetic energy of the incident electron and the Z num- radiation. The initial nuclide, or radionuclide, is known as the
ber of the target material. Thus, bremsstrahlung x-ray produc- parent, and the resultant one after radioactive decay is known
tion is more likely to occur at higher energies and with high-Z as the daughter. Although the daughter nucleus created from a
targets. As a result, radiographic systems can generate x-rays by radioactive decay has a lower energy than the parent nucleus,
directing an energetic electron beam into a tungsten (Z = 74) it may not be stable, and thus subsequent radioactive decays
target. On the other hand, the intensity of the bremsstrahlung may result.
radiation is relatively low when beta particles pass through soft Fig. 1.5 shows a plot of the stable nuclides as a function of
tissue. the Z number on the x-axis and the N number on the y-axis. At
Chapter 1 Radioactivity and Radionuclides 5

B C
Fig. 1.6 “Proton-rich” radionuclides that decay by positron emission can be made in a cyclotron or particle
accelerator. (A) Varying in size and appearance, cyclotrons may be self-shielded or housed in a thick cement
vault (as shown) to reduce radiation exposure. Beam lines (arrow) extending from the central unit direct high-
speed charged particles to bombard desired targets. (B) The bottom of the cyclotron contains the accelerating
electrodes (short arrows). (C) Electromagnetic fields created by a large magnet (arrowhead) in the upper por-
tion of the cyclotron constrain the particles to circular orbits. (Photos courtesy of Anthony F. Zagar, University
of Alabama at Birmingham.)

low Z numbers, stable elements tend to have equal numbers of Proton-rich radionuclides can be created by bombarding a cer-
protons and neutrons (e.g., carbon-12, nitrogen-14, and oxy- tain target material with high-energy protons that can overcome
gen-16) and lie along or near the Z = N line. However, as the nuclear forces. Typically, a particle accelerator such as a cyclotron is
nucleus becomes larger, the repulsive force of the nuclear pro- used, increasing kinetic energy by accelerating charged particles to
tons grows, and more neutrons are necessary in the stable high speeds in a spiral path using alternating high-frequency volt-
nucleus to provide additional attractive nuclear force. Other age and electromagnetic fields (Fig. 1.6). Conversely, in artificial
factors also contribute to the stability and instability of the production of neutron-rich radionuclides, one typically must use a
nucleus. For example, nuclides with even numbers of protons nuclear reactor to bombard a target with a neutron flux (Fig. 1.7).
and neutrons tend to be more stable than those with odd Z and
N configurations. Modes of Radioactive Decay
Unstable nuclides fall to either the right or the left of the A decay scheme is a way to illustrate the transition from parent
curve of stability, with those to the right considered proton rich to daughter nuclides. In a decay scheme, higher energy levels
and those to the left neutron rich. As unstable radionuclides are toward the top of the figure, and higher Z numbers are to the
decay to entities that are closer to the curve of stability, pro- right of the figure. Transitions that lead to a reduction in energy
ton-rich radionuclides tend to decay in a manner that will are represented by an arrow pointing down. If it also results in a
reduce the Z number and increase the N number, and neu- daughter nuclide with a change in the Z number, the arrow will
tron-rich radionuclides tend to decay in a way that decreases the point to the left with a decrease in the Z number and to the right
N number and increases the Z number. if Z is increased.
6 PART 1 Basic Principles

Alpha Decay
An unstable heavy atom may decay to a nuclide closer to the
curve of stability by emitting an alpha particle (α) consisting of
2 protons and 2 neutrons (essentially an ionized helium atom):
A A −4 A −4 4
Z
X→ Z−2Y + α = Z−2Yjo 9 + 2 He

The daughter nucleus may not be stable, and thus the emis-
sion of an alpha particle often will lead to the emission of a
series of radiations until the nucleus is stable. The decay scheme
for the decay of radium-226 (Ra-226) to radon-222 (Rn-222) is
shown in Fig. 1.8.

Beta-Minus Decay
Neutron-rich radionuclides tend to stabilize by decreasing
the number of neutrons through a radioactive-decay process
referred to as beta-minus (β–), also known as negatron or
beta decay. Factors such as weak forces between nucleons
transfer energy, transforming a neutron into a proton (N – 1
and Z + 1). This is an isobaric transition with no change to
Fig. 1.7 Fission and neutron capture radionuclide production in a nuclear
the atomic mass (A). An example of the beta-minus decay
reactor. Samples can be lowered into the reactor as shown, with water scheme for I-131 is shown in Fig. 1.9. Excess energy is emit-
acting as shielding against neutrons. The blue glow is caused by the ted from the nucleus as an antineutrino and a negative beta
emission of electrons from the radioactive products; when charged par- particle (or negatron). This process can be written as
ticles move faster than the speed of light in a medium such as water, follows:
the emitted radiation is called Cherenkov radiation. (Courtesy of the Uni-
A A
versity of Missouri Research Reactor Center.) X
Z N
= Z+1YN−1 + β − + −υ antineutrino

226
Ra (1600 yr)
88

448 keV


2
186 keV


1

222
Rn (3.8 days)
86
Fig. 1.8 Alpha decay. The emission of an alpha particle (2 protons and 2 neutrons) results in the atomic
number (Z) decreasing by 2 and the atomic mass (Z + N) decreasing by 4. Decay of radium-226 to the daugh-
ter Rn-222 shows the arrow pointed down, indicating a decrease in energy, and to the left because of the
decrease in Z.
Chapter 1 Radioactivity and Radionuclides 7

131 18
I (8 days)
53 9 F (110 min)


364 keV


 (81%)

18
8O (stable)
131 Fig. 1.10 Positron (β+) decay results in a loss of 1 proton (Z – 1) in pro-
Xe (stable)
54 ton-rich radionuclides. Because 1 neutron is gained, the atomic mass of
Fig. 1.9 Beta minus (β –) decay scheme for iodine-131 to the daughter the daughter is unchanged, another example of isobaric transition. F-18
Xe-133. β– decay (negatron emission) results in the daughter with one decay by positron emission results in the daughter product, O-18. The
more proton in the nucleus (Z + 1), so the arrow points to the right. arrow points down and to the left, indicating the decrease in Z.
Because a neutron is lost (N – 1), this is an isobaric transition with the
atomic mass unchanged.
201
81Tl (73 hr)
The antineutrino is very difficult to measure because it has
virtually no mass or charge associated with it, only energy. The
negative beta particle is indistinguishable from an electron
with the same mass and electric charge, differing only in that EC
the beta particle is emitted from the nucleus and the electron
orbits the nucleus. In addition to the Mo-99 used to make
Tc-99m, several β–-emitting radionuclides play an important
167 keV
role in nuclear medicine for therapy applications: I-131, phos-
 (3%)
phorus-32 (P-32), yttrium-90 (Y-90), and lutetium-177 32 keV
 (10%)
(Lu-177).
201
Beta-Plus (Positron) Decay 80 Hg (stable)
Unstable proton-rich radionuclides can reduce Z and increase Fig. 1.11 Electron capture is an alternate transition that can occur to
N numbers through either beta-plus decay or electron capture. In reduce the proton number and does not require that an energy thresh-
beta-plus decay, the parent nucleus emits a positively charged old be met. Tl-201 decays by electron capture, with the daughter (Hg-
beta particle, a positron (β+). The resulting daughter nucleus has 201) containing one fewer proton (Z – 1) than the parent.
one fewer proton and one more neutron than the parent, an iso-
baric transition: Electron Capture. An alternative to beta-plus decay for
A A proton-rich radionuclides is electron capture (EC). In this
X = Y
Z N Z−1 N+1
+ β + + υ neutrino process, an inner-shell, orbital electron is absorbed into the
average β + kinetic energy: Eβ + ≈ Emax /3 nucleus, leading to the reduction of Z and increase of N by 1.
The positron has the same mass as a beta-minus particle or However, no energy threshold exists for EC to occur. In cases in
electron, with a charge of the same magnitude but the opposite. which the transition energy is less than the 1022-keV threshold,
In fact, the positron is the antiparticle of the electron; if they are EC is the only possible decay process, but either process is
brought into close contact, they will be annihilated and trans- possible when the energy is greater than 1022 keV. For F-18,
formed into two 511-keV photons, traveling at 180 degrees in positron decay occurs 97% of the time, and EC occurs 3% of the
opposite directions. This annihilation process is the basis of time. The capture of an orbital electron leads to an inner-shell
PET imaging. The 511-keV value derives from the energy equiv- vacancy, which in turn leads to the emission of fluorescent
alence of the mass of the beta particle, similar to the rest mass of x-rays or Auger electrons. Radionuclides that decay through EC
an electron (using E = mc2 as previously discussed). exclusively include thallium-201 (Tl-201; Fig. 1.11), gallium-67
For positron decay to occur, the transition energy must be in (Ga-67), and indium-111 (In-111). These are all produced in a
excess of a 1022-keV threshold (twice 511 keV) to overcome the cyclotron.
production of the positron and addition of an orbital electron to Isomeric Transition. In some cases, an excited radionuclide
maintain electric neutrality. These radionuclides are typically decays from one energy level to another while retaining the
produced using a cyclotron. Some positron-emitting radionu- same Z and N numbers. This is referred to as an isomeric
clides of interest include fluorine-18 (F-18; Fig. 1.10), nitro- transition because the nuclide decays from one isomer (energy
gen-13 (N-13), carbon-11 (C-11), gallium-68 (Ga-68), and level) to another. This transition may result in the emission of a
rubidium-82 (Rb-82). gamma ray, the energy of which is determined by the energy
8 PART 1 Basic Principles

99m reciprocal of Tm, the fraction of the radioactive atoms that decay
43 Tc (6.01 hr) per unit time, is referred to as the decay constant, λ:
1
λ=
Tm
Thus, the number of atoms (dN) that decay in a short time
 140.5 keV (89%) interval (dt) is given by:
dN = λdt
Integrating this equation over time leads to:
N = N0e−λt
where N0 is the initial number of radioactive atoms, and N is the
99 number remaining after some time, t.
43 Tc This equation describes exponential decay in which a certain
Fig. 1.12 Isomeric transitions involve a change in the energy state of a fraction of the material is lost in a set period. This fraction is
radionuclide, such as Tc-99m to Tc-99.
referred to as the decay fraction, DF:

99 DF = e−λt
42Mo (2.8 days or 66 hr)
Thus, the number of radioactive atoms remaining, N, is also
given by:

N = N0 × DF
142.7 keV
Also, Nd is the number of atoms that have decayed in time, t,
 140.5 keV
and can be calculated with:
 Nd = N0 × (1 − DF)
0.0
The time necessary for half of the material to decay is defined
99 5 as the half-life (T1/2). The half-life is related to the mean life and
43Tc (2.1  10 yr)
the decay constant by the following equations:
Fig. 1.13 Decay scheme of Mo-99. Beta-minus emission to Tc-99m,
followed by isomeric transition to Tc-99. 0.693
T1/2 = ln (2) Tm =
difference of the initial and eventual energy levels. In some λ
cases, an alternate process called internal conversion can occur, Alternatively, one can determine the decay constant from the
resulting in the emission of an orbital electron, a conversion half-life by:
electron. The kinetic energy is calculated as the difference in the 0.693
two energy levels minus the electron’s binding energy. λ=
T1/2
Perhaps the most important isomeric transition for
nuclear medicine involves technetium. The term metastable One can also express the radioactive decay equation using
(i.e., almost stable) is used if the daughter nucleus remains the half-life:
in the excited state for a considerable amount of time (>1 − 0.693t
microsecond, which is long by nuclear standards). Mo-99 N = N0e T1/2

decays to an excited, or metastable, Tc-99m that in turn


If a sample contains 10,000 radioactive atoms at a partic-
transitions to Tc-99 (Figs. 1.12 and 1.13). Tc-99m has a
ular point in time, one half-life later, there will be 5000
6-hour half-life. Tc-99m is so commonly used because of its
atoms; another half-life later, there will be 2500 atoms; and
reasonable half-life, as well as its gamma-ray energy (140
so on. This process of a certain fraction of the material
keV) and lack of beta- or alpha-particle emissions. Another
decaying in a certain time is representative of exponential
example of isomeric transition is seen in the decay scheme of
decay (Fig. 1.14A). When graphed using a log scale on the
I-131 (see Fig. 1.9). Xenon-131 (Xe-131), formed from the
y-axis (semilog plot), the result is a straight line with the
beta-minus decay of I-131, is in an excited state and imme-
negative slope equal in magnitude to the decay constant
diately decays by isomeric transition with the emission of a
(Fig. 1.14B).
364-keV gamma ray.
The amount of activity (A) is the number of nuclear transfor-
Radioactive Decay Calculations mations—decays or disintegrations—per unit time. The activity
is characterized by the number of radioactive atoms in the sam-
Atoms in a sample containing a certain number (N) of radioac-
ple, N, divided by the mean time to radioactive decay, Tm:
tive atoms will not all decay at the same time but with a mean
time (Tm) that is characteristic of a particular radionuclide. The A = N/Tm
Chapter 1 Radioactivity and Radionuclides 9

Radioactivity (fraction remaining) 1 1.0

Radioactivity (fraction remaining)


0.9
0.8
0.7
0.6
0.5 0.1
0.4
0.3
0.2
0.1
0 0.01
0 6 12 18 24 30 36 42 0 6 12 18 24 30 36 42
A Time (hr) B Time (hr)
Fig. 1.14 Decay plot for Tc-99m. (A) Standard graph showing the exponential loss. (B) Semilog graph.

Activity is thus the product of the decay constant and the


number of radioactive atoms: BOX 1.2 Conversion of International
System and Conventional Units of
A = λN Radioactivity
Conversely, if the amount of activity of a particular radionu- Mathematical
clide is known, the number of radioactive atoms can be c (centi-) 10–2
calculated: m (milli-) 10–3
k (kilo-) 103
N = A/λ μ (micro-) 10–6
Because the activity is directly related to the number of M (mega-) 106
n (nano-) 10–9
radioactive atoms, all of the equations for radioactive decay
G (giga-) 109
apply to activity, as well as the number of atoms:
p (pico-) 10–12
A = A0e−λt T (tera-) 1012

And Conventional Unit


1 curie (Ci) = 3.7 × 1010 disintegrations per second (dps)
A = A0e−(0.693t)/(T 1/2)
SI Unit
The units associated with activity are the becquerel (1 Bq = 1 1 becquerel (Bq) = 1 dps
disintegration per second) and the curie (1 Ci = 3.7 × 1010 disin- Curies → Becquerels
tegrations per second; Box 1.2). Their relationship is as follows: 1 Ci = 3.7 × 1010 dps = 37 GBq
1 mCi = 37 MBq, and 1 MBq = 27 μCi 1 mCi = 3.7 × 107 dps = 37 MBq
1μCi = 3.7 × 104 dps = 37 KBq
Example 1. The radiopharmacy is preparing a dose of an
I-123–labeled agent (13-hour half-life) for the clinic. If 10 mCi Becquerels → Curies
is to be administered at 1 pm, how much activity should be 1 Bq = 1 dps = 2.7 × 10–11 Ci = 27 pCi
1 MBq = 106 dps = 2.7 × 10–5 Ci = 0.027 mCi
placed in the syringe at 7 am?
1 GBq = 109 dps = 27 mCi
A = A0e−(0.693t)/(T 1/2) Probability of radiative losses directly proportional to Z of target and energy
incident particle.
Thus
(−0.693)(6 hr)/(13 hr)
10mCi = A0e

(−6.693)(6 hr)/(13 hr)


A = A0e−(0.693t)/(T 1/2)
A0 = 10mCi/e = 10mCi/0.726 = 13.8mCi

Example 2. The staff at the nuclear medicine clinic is testing Thus


their equipment with a cobalt-57 source (270-day half-life) that
–0.693 × 243 days
was calibrated to contain 200 MBq on January 1 of this year.
A = 200e 270 days = 200 MBq × 0.536 = 107 MBq
How much activity remains on September 1 (243 days)?
10 PART 1 Basic Principles

INTERACTIONS BETWEEN RADIATION AND The β– kinetic energy is variable because it shares energy with
the antineutrino produced during the decay event. The maxi-
MATTER mum kinetic energy of the beta particle (Emax) is defined by the
Charged-Particle Interactions With Matter difference in the energy levels of the parent and daughter
A charged particle may transfer energy in different ways. First, nuclide. However, it is the average kinetic energy that is used
it can be attracted and slowed by the opposite charge of the when calculating the impact of the β– on cells and tissues, esti-
nucleus or orbiting electrons in target material atoms. The mated as 1/3 of Emax (Eβ ≈ Emax /3), similar to the calculation
resulting kinetic energy loss is released as radiation (radiative previously described for positrons.
losses). Bremsstrahlung radiation occurring with a β– emitter is
one example of this type of interaction. The energy of radiative Photon Interactions in Matter
losses is directly proportional to the Z number of the target as High-energy photons (gamma rays, x-rays, bremsstrahlung
well as to the incident particle’s energy. radiation, and annihilation radiation) can also transfer energy
Charged particles can also directly transfer energy to the to the electrons, nuclei, or atoms as a whole that they encoun-
atom’s orbital electrons, resulting in electron excitations and ter. Unlike charged particles, which directly create ionized
ionizations. While excited, electrons can temporarily move to a atoms, the high-energy photons act indirectly, transferring
shell farther from the nucleus. As de-excitation occurs, trans- their energy to charged particles, specifically electrons, which
ferred energy leads to the emission of Auger electrons or elec- in turn create most of the excitations and ionizations that occur
tromagnetic radiation. This radiation can have a wide range of in the matter. Thus they are considered secondary ionizing
energies, including visible or ultraviolet radiation for outer-shell radiations.
transitions and fluorescent x-rays for the inner-shell At low energy levels (a few keV), photons are scattered in a
transitions. manner that does not deposit energy, referred to as Rayleigh
When the energy from charged-particle interactions pro- scattering. Photons with energy in excess of several MeV can
duces ionized electrons and atoms in tissues, the majority of the result in pair production of a negatron and a positron (effectively
ionized electrons are low energy. However, some interactions a negative and positive electron). However, in energy ranges
result in high-energy electrons, referred to as delta rays, which most common in nuclear medicine (from several tens of keV to
in turn can also cause excitation and ionization. In the energies approximately 1 MeV), the two most prominent modes of pho-
of practical interest in nuclear medicine, nearly all of the energy ton interaction are the photoelectric effect and Compton scatter-
from a charged-particle interaction (greater than 99%) is ing. Factors involved in the various types of interaction are
expended in excitation and ionization (or collisional losses) outlined in Table 1.4.
compared with radiative losses.
The rate at which a material causes a charged particle to lose Photoelectric Effect
energy (per unit length of the matter) is referred to as its stop- The photoelectric effect (Fig. 1.15) occurs when a photon trans-
ping power. A related quantity is linear energy transfer (LET), fers all of its energy to an orbital electron, causing it to be ejected
which is the amount of energy deposited locally (i.e., not lost to from the atom and creating an electron-shell vacancy. The
energetic electrons, delta rays, or radiative loss) per unit length. kinetic energy of the liberated photoelectron equals the incident
The stopping power and LET values depend on the type of radi- photon’s energy minus the binding energy of the electron’s ini-
ation, its energy, and the density of the material through which tial orbital shell. As the shell vacancy is filled by electrons from
it travels. Radiation with a higher LET value has been shown to outer shells, fluorescent x-rays and Auger electrons are also
cause more damage to cells. Alpha particles have a higher LET emitted. Paradoxically, the probability of photoelectric interac-
than beta particles or electrons. tions is highest for tightly bound orbiting electrons (i.e., those in
Although densely ionizing, alpha particles deposit their the inner shells of high Z elements). These electrons are most
energy over a very short distance, a small fraction of a millime- likely the ones in the innermost orbital shell where the binding
ter in soft tissues. Also, although they are easily stopped by the energy is just under the photon’s energy. In addition, the chance
skin, these high-energy particles cause substantial cell death of this interaction dramatically decreases as incident photon
when internalized, making them both extremely dangerous if energy increases. The probability of the photoelectric interac-
accidentally ingested as well as highly effective in therapeutic tion (or PPE) is given by:
applications (e.g., Ra-223 in prostate cancer).
3 3
Comparatively, β– particles travel for much longer distances, PPE α Z /E
ranging from several millimeters to several centimeters depend-
ing on their initial energy. They can be stopped by material such where Z = atomic number, and E = incident photon kinetic
as a thin sheet of aluminum or a few millimeters of soft tissue. energy.
Chapter 1 Radioactivity and Radionuclides 11

TABLE 1.4 Photon Interactions in Matter


Resulting
Effect of Target Incident Photon Particle Secondary Photon
Interaction Occurrence Material Z E Range (E0) Target Emissions Emissions
Compton Predominant in Nearly independent of Z Mid E range (≈26 keV–30 Outer-shell e– Recoil e– Scattered photon
soft tissues at Depends on e– density MeV) Degrades image
diagnostic E (therefore on target
range density)
Hydrous > anhydrous
material
Photoelectric Predominates in High-Z materials Low-E photon Innermost-shell Photoelectrons Characteristic x-ray
effect shielding and Z3/E3 e– possible Auger e–
detector crys-
tals/PMT
e– cascade produces
Characteristic x-ray ↑ with ↑ Z-detector mate- ↑ when weakly bound valence (e.g.,
rial, shielding PMT photocathode materials)
Auger e– ↑ in soft tissues e– binding E less a
factor in tissue
Pair production Not typically 1.02-MeV minimum but Usually nucleus, β+ and β– Annihilation photons (from
seen in ener- actually sometimes orbital (or e+ and e–) β+)
gies used in >>1.02 MeV (not present at e– Two 5110 keV at 180
medicine diagnostic E ranges) degrees
E, Energy; Z, atomic number (number of protons); PMT, photomultiplier tube; e–, electron; β+, positron (or a positive electron); β –, negatron (same
as negative electron or beta-minus particle).

In soft tissues (low Z), the photoelectric effect is much less occurring. Higher-energy incident photons lose less energy to
common than Compton scatter. It is, however, more prevalent the electron, and deflection is less significant (i.e., the scatter
in the high-Z materials used for shielding (e.g., lead) or for pho- angle is narrower), such that both the scattered photon and elec-
ton detection (sodium iodide crystals). Photoelectric effect tron tend to travel in a more forward direction.
interactions can also occur in the gamma camera’s photomulti- Whereas the photoelectric effect is important at lower
plier tubes, which contain high-Z materials, such as cesium. energies and more likely involves inner-shell electrons in
high-Z materials, Compton scatter predominates in soft tis-
Compton Scatter sues in the moderate-energy ranges of gamma-ray and x-ray
The incident photon does not disappear in Compton scatter. photons in nuclear medicine imaging and tends to involve
Rather, it transfers a portion of its energy to an orbital electron outer-shell electrons. Because the energy of the incident pho-
(a Compton electron), which is then ejected from the atom. The ton is much greater than the shell’s binding energy, the colli-
photon is deflected or scattered, at an angle θ from its original sion occurs as if involving a free electron. Compton
path (Fig. 1.16). The electron and scattered photon may go on to interactions tend to depend on electron density but are rela-
ionize or excite other atoms. tively independent of the Z number or incident photon
The sum of the kinetic energies of the scattered photon and energy. Because electron density is fairly consistent among
the Compton electron will equal the initial photon’s energy. the atoms in soft tissues, the probability increases with
With lower-energy incident photons, more energy is transferred increasing material density rather than its Z number. Electron
to the electron, and there is greater backscatter of the resulting density is higher when hydrogen atoms are present, so tissues
photon (i.e., the angle between the incident and scattered pho- with high water content are more affected than anhydrous
tons tends to be wider), with even 180 degrees of backscatter tissues.
12 PART 1 Basic Principles

Photoelectron

Nucleus

Nucleus

Vacancy

Incident photon
A B

Characteristic
x-ray

Vacancy
Nucleus
Electron
transition
between
shells

C
Fig. 1.15 Photoelectric absorption. (A) An incident photon interacts with an orbital electron. (B) The electron
is ejected from its shell, creating a vacancy. The electron is either ejected from the atom or moved to a shell
further from the nucleus. (C) The orbital vacancy is filled by the transition of an electron from a more distant
shell. Consequently, a characteristic x-ray is given off.
Chapter 1 Radioactivity and Radionuclides 13

SUGGESTED READING
Chandra R, Rahmin A. Nuclear Medicine Physics: The Basics. 8th ed.
Philadelphia: Williams & Wilkins; 2012.
Cherry SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine. 4th
ed. Philadelphia: WB Saunders; 2012.
Compton
Eckerman KF, Endo A. MIRD: Radionuclide Data and Decay Schemes.
Nucleus
electron 2nd ed. Reston, VA: Society of Nuclear Medicine; 2008.
Loevinger R, Budinger TF, Watson EE. MIRD Primer for Absorbed
Dose Calculations. Reston, VA: Society of Nuclear Medicine; 1988.
Powsner RA, Powsner ER. Essentials of Nuclear Medicine Physics. 3rd
ed. West Sussex, UK: Wiley-Blackwell; 2013.

Angle Saha GP. Physics and Radiobiology of Nuclear Medicine. 4th ed. New
of scatter Compton- York: Springer; 2013.
scattered
photon

Incident
photon
Fig. 1.16 Compton scatter. An incident photon interacts with an outer
or loosely bound electron, giving up a portion of its energy to the elec-
tron and undergoing a change in direction at a lower kinetic energy level.
2
Radiation Detection & Ancillary
Instrumentation
Janis M. O’Malley, Harvey Ziessman, Frederic Fahey

The passage of radiation, such as x-rays and gamma rays, through of the detector. For photons, the intrinsic efficiency, DI, is given
a given material leads to ionizations and excitations that can be to first order by:
used to quantify the amount of energy deposited. This property
DI = (1 − e ‐ μx )
allows measurement of the level of intensity of a radiation beam
or small amounts of radionuclides, including from within the where μ is the linear attenuation coefficient for the material of
patient. The appropriate choice of detection approach depends interest at the incident photon energy, and x is the thickness of
on the purpose. In some cases, the efficient detection of min- the detector. Thus the intrinsic efficiency can be improved by
ute amounts of the radionuclide is essential, whereas in other using a thicker detector or choosing a photon energy and detec-
cases the accurate determination of the energy or location of the tor material that optimizes the value of μ.
radiation deposited is most important. A variety of approaches The extrinsic efficiency is the fraction of photons or particles
to radiation detection are used, including those that allow for emitted from the source that strike the detector. It depends on
in vivo imaging of radiopharmaceuticals. the size and shape of the detector and the distance of the source
from the detector. If the detector is a considerable distance from
the source (i.e., a distance that is >5 times the size of the detec-
RADIATION DETECTION tor), the extrinsic efficiency, DE, is given by:
Consider the model of a basic radiation detector, as shown DE = A/ 4πd 2
in Fig. 2.1. The detector acts as a transducer that converts
radiation energy to electronic charge. Applying a voltage where A is the area of the detector, and d is the distance from the
across the detector yields a measurable electronic current. source to the detector. This equation defines the inverse square
Radiation detectors typically operate in either of two modes, law. For example, if the source-to-detector distance is doubled,
current mode or pulse mode. Detectors that operate in cur- the intensity of the radiation beam is reduced by a factor of 4.
rent mode measure the average current generated within the The total detection efficiency is the product of the intrinsic and
detector over some characteristic integration time. This aver- extrinsic efficiencies:
age current is typically proportional to the exposure rate to DT = DI × DE
which the detector is subjected or the amount of radioactivity
within the range of the detector. In pulse mode, each individ- In pulse mode, the pulse height is proportional to the energy
ual detection is processed with respect to the peak current (or deposited within the detector. However, the uncertainty in the
pulse height) for that event. This pulse height is proportional energy estimation, referred to as the energy resolution, depends
to the energy deposited in the detection event. The histogram on the type of detector used and the energy of the incident radi-
of pulse heights is referred to as the pulse-height spectrum. It ation. For a photon radiation source of a particular energy, the
is also referred to as the energy spectrum because it plots a feature associated with that energy is referred to as the photo-
histogram of the energy deposited within the detector. peak, as shown in Fig. 2.2. The width of the photopeak, as char-
Certain properties of radiation detectors characterize their oper- acterized by the full width at half of its maximum (FWHM)
ation. Some are applicable to all detectors, whereas others are used value normalized by the photon energy represented as a per-
for detectors that operate in pulse mode. These characterizations centage, is used as a measure of the energy resolution of the
are not only useful for describing the operation but can also give detector.
insight into the benefits and limitations of the particular detector. When the detector is subjected to a radiation beam of low
The detection efficiency depends on several factors, includ- intensity, the count rate is proportional to the beam intensity.
ing the intrinsic and extrinsic efficiency of the detector. The However, the amount of time it takes for the detector to process
intrinsic efficiency is defined as the fraction of the incident radi- an event limits the maximum possible count rate. Two models
ation particles that interact with the detector. It depends on the describe the count rate limitations: nonparalyzable and paralyz-
type and energy of the radiation and the material and thickness able. In the nonparalyzable model, each event takes a certain

14
Chapter 2 Radiation Detection & Ancillary Instrumentation 15

Voltage source
+ –
+ Anode Geiger-Müller
Incident region

Amplitude of output pulse


ionizing Air or
radiation e– other
e–
e–
Current
+ e– e– gas I measuring
+ + + + device
Ionization
chamber region
– Cathode
Fig. 2.1 Block diagram of basic detector. The radiation detector basi- Proportional
cally acts as a transducer, converting radiation energy deposited into an counter region
electrical signal. In general, a voltage has to be supplied to collect the
signal, and a current or voltage measuring device is used to measure
the signal. In some instances, the average current over a characteris-
tic integration time is measured, which is referred to as current mode.
In other cases, the voltage pulse of each detection event is analyzed, Applied voltage
referred to as pulse mode. (From Cherry, Sorenson JA, Phelps ME. Fig. 2.3 Amplitude of gas detector output signal as a function of applied
Physics in Nuclear Medicine. 3rd ed. Philadelphia: WB Saunders; 2003.) voltage. This graph shows the relationship between the magnitude of
the output signal from the gas detector (related to the amount of ion-
ized charge collected) as a function of the voltage applied across the
detector. There is no signal with no voltage applied. As the voltage is
increased, the detector signal starts to increase until the saturation volt-
age is reached, the start of the plateau defining the ionization chamber
region, where all of the initially liberated charge is collected. Further
increasing the voltage leads to the proportional counter region, at which
the liberated electrons attain sufficient energy to lead to further ion-
ization within the gas. Finally, the Geiger-Müller region is reached, at
which each detection yields a terminal event of similar magnitude (i.e.,
a “click”). (From Cherry, Sorenson JA, Phelps ME. Physics in Nuclear
Medicine. 3rd ed. Philadelphia: WB Saunders; 2003.)

A gas radiation detector is filled with a volume of gas that


acts as the sensitive material of the detector. In some cases it
is air, and in others it is an inert gas such as argon or xenon,
depending on the particular detector. Electrodes are located at
Fig. 2.2 Spectrum for technetium-99m (Tc-99m) in air. The energy res- either end of the sensitive volume. The detector circuit also con-
olution is characterized by the width of the photopeak (the full width tains a variable voltage supply and a current detector. As radi-
at half maximum [FWHM]) normalized by the photon energy. For the ation passes through the sensitive volume, it causes ionization
particular detector system illustrated, the FWHM is 18 keV. The energy
in the gas. If a voltage is applied across the volume, the result-
resolution of the detector system for Tc-99m is 13% (100 × 18/140).
ing ions (electrons and positive ions) will start to drift, causing
a measurable current in the circuit. The current will last until
amount of time to process, referred to as the dead time, which all of the charge that was liberated in the event is collected at
defines the maximum count rate at which the detector will satu- the electrodes. The resulting current entity is referred to as a
rate. For example, if the dead time is 4 μs, the count rate will sat- pulse and is associated with a particular detection event. If only
urate at 250,000 counts per second. With the paralyzable model, the average current is measured, this device operates in current
the detector count rate not only saturates but can “paralyze”— mode. If the individual events are analyzed, the device is oper-
that is, lose counts at very high count rates. Gamma cameras, for ating in pulse mode.
example, are paralyzable systems. Fig. 2.3 shows the relationship between the charge collected
The three basic types of radiation detectors used in nuclear in the gas detector and the voltage applied across the gas vol-
medicine are gas detectors, scintillators, and semiconductors. ume. With no voltage, no electric field exists within the volume
These three operate on different principles and are typically to cause the ions liberated in a detection event to drift, and thus
used for different purposes. no current is present and no charge is collected. As the voltage is
Gas detectors are used every day in nuclear medicine for increased, the ions start to drift, and a current results. However,
assaying the amount of radiopharmaceutical to be administered the electric field may not be sufficient to keep the electrons and
and to survey packages and work areas for contamination. How- positive ions from recombining, and thus not all of the originally
ever, because of the low density of gas detectors, even when the liberated ions are collected. This portion of Fig. 2.3 is referred
gas is under pressure, the sensitivity of gas detectors is not high to as the recombination region. As voltage is increased, the level
enough to be used for clinical counting and imaging applications. is reached at which the strength of the electric field is sufficient
16 PART 1 Basic Principles

for the collection of all of the liberated ions (no recombination). efficiency because they affect the linear attenuation coefficient
This level is referred to as the saturation voltage, and the result- of the scintillation material. The amount of emitted light affects
ing plateau in Fig. 2.3 is the ionization chamber region. When both energy and, in the gamma camera, spatial resolution. Res-
operating in this region, the amount of charge collected is pro- olution is determined by the statistical variation of the collected
portional to the amount of ionization caused in the detector and light photons, which depends on the number of emitted pho-
thereby to the energy deposited within the detector. Ionization tons. Finally, the response time affects the temporal resolution
detectors or chambers typically operate in current mode and of the scintillator. The most common scintillation crystalline
are the detectors of choice for determining the radiation beam material used in nuclear medicine is thallium-doped sodium
intensity level at a particular location. They can directly mea- iodide (NaI) with lutetium oxyorthsilicate (LSO) or lutetium
sure this intensity level in either exposure in roentgens (R) or air yttrium oxyorthosilicate (LYSO) most commonly used in posi-
kerma in rad. Dose calibrators and the ionization meters used to tron emission tomography (PET).
monitor the output of an x-ray device or the exposure level from Once the light is emitted in a scintillation detector, it must be
a patient who has received a radiopharmaceutical are examples collected and converted to an electrical signal. The most com-
of ionization (or ion) chambers used in nuclear medicine. monly used device for this purpose is the photomultiplier tube
If the voltage is increased further, the drifting electrons (PMT). Light photons from the scintillator enter through the
within the device can attain sufficient energy to cause further photomultiplier entrance window and strike the photocathode,
ionizations, leading to a cascade event. This can cause substan- a certain fraction of which (approximately 20%) will lead to the
tially more ionization than with an ionization chamber. The total emission of photoelectrons moving toward the first dynode. For
ionization is proportional to the amount of ionization initially each electron reaching the first dynode, approximately a million
liberated; therefore these devices are referred to as proportional electrons will eventually reach the anode of the photomultiplier
counters or chambers. Proportional counters, which usually tube. Thus the photomultiplier tube provides high gain and
operate in pulse mode, are not typically used in nuclear med- low noise amplification at a reasonable cost. Other solid-state
icine. If the voltage is increased further, the drifting electrons light-detection approaches are now being introduced into
attain the ability to cause a level of excitations and ionizations nuclear medicine devices. In avalanche photodiodes (APDs),
within the gas. The excitations can lead to the emission of ultra- the impinging light photons lead to the liberation of electrons
violet radiation, which also can generate ionizations and fur- that are then drifted in the photodiode, yielding an electron ava-
ther excitations. This leads to a terminal event in which the level lanche. The gain of the APD is not as high as with the PMT (sev-
of ionization starts to shield the initial event, and the level of eral hundred compared with about a million), but the detection
ionization finally stops. This is referred to as the Geiger-Müller efficiency is substantially higher (approximately 80%). A second
process. In the Geiger-Müller device, every event leads to the solid-state approach is the silicon photomultiplier tube (SiPMT).
same magnitude of response, irrespective of the energy or the This device consists of hundreds of very small APD channels
type of incident radiation. Thus the Geiger-Müller meter does that operate like small Geiger-Müller detectors—that is, each
not directly measure exposure, although it can be calibrated in detection is a terminal event. The signal from the SiPMT is the
a selected energy range to milliroentgens per hour (mR/hr). number of channels that respond to a particular detection event
However, the estimate of exposure rate in other energy ranges in the scintillator. SiPMTs have moderate detection efficiency
may not be accurate. The Geiger-Müller survey meter is excel- (approximately 50%) and operate at low voltages. One further
lent at detecting small levels of radioactive contamination and advantage of APDs and SiPMTs compared with PMTs is that
thus is often used to survey radiopharmaceutical packages that they can operate within a magnetic field. Thus the development
are delivered and work areas within the nuclear medicine clinic of positron emission tomography/magnetic resonance (PET/
at the end of the day. MR) scanners has involved the use of either APDs or SiPMTs.
Solid-state technology is used to detect the light from a scin-
tillation detector and also can be used to directly detect gamma
SCINTILLATION DETECTORS rays. The detection of radiation within a semiconductor detector
Some crystalline materials emit a large number of light pho- leads to a large number of electrons liberated, resulting in high
tons upon the absorption of ionizing radiation. This process is energy resolution. The energy resolution of the lithium-drifted
referred to as scintillation, and these materials are referred to germanium (GeLi) semiconductor detector has approximately
as scintillators. As radiation interacts within the scintillator, a 1% energy resolution compared with the 10% energy resolution
large number of excitations and ionizations occur. On de-exci- associated with a sodium iodide scintillation detector. However,
tation, the number of light photons emitted is directly propor- thermal energy can lead to a measurable current in some semi-
tional to the amount of energy deposited within the scintillator. conductor detectors such as GeLi, even in the absence of radia-
In some cases, a small impurity may be added to the crystal to tion, and thus these semiconductor detectors must be operated
enhance the emission of light and minimize the absorption of at cryogenic temperatures. On the other hand, semiconductor
light within the crystal. Several essential properties of scintil- detectors such as cadmium telluride (CdTe) or cadmium zinc
lating materials can be characterized, including density, effec- telluride (CZT) can operate at room temperature. CdTe and
tive Z number (number of atomic protons per atom), amount CZT do not have the excellent energy resolution of GeLi, but
of light emitted per unit energy, and response time. The density at approximately 5%, it is still significantly better than that of
and effective Z number are determining factors in the detection sodium iodide.
Chapter 2 Radiation Detection & Ancillary Instrumentation 17

contamination. It is routinely used to determine whether there


is contamination on packages of radiopharmaceutical that are
delivered to the clinic and to test working surfaces and the
hands and feet of workers for contamination. GM meters often
are equipped with a test source of cesium-137, with a very small
amount of radioactivity, that is affixed to the side of the meter.
On calibration, the probe is placed against the source, and the
resulting exposure rate is recorded. The probe is tested daily
using the source to ensure that the meter’s reading is the same
at the time calibration. The GM meter should be calibrated on
an annual basis.
The ionization chamber meter (ion chamber) operates in
current mode and assesses the amount of ionization within an
internal volume of gas (often air) and thus can directly measure
exposure or air kerma rate. The ion chamber is used to evalu-
Fig. 2.4 Energy spectrum for technetium-99m (Tc-99m) in air for a ate the exposure rate at various locations within the clinic. For
gamma scintillation camera with the collimator in place. Note the iodine example, it could be used to measure the exposure rate in an
escape peak at approximately 112 keV. The 180-degree backscatter
uncontrolled area adjacent to the radiopharmaceutical hot lab-
peak at 90 keV merges with the characteristic x-ray peaks for lead (Pb).
The Compton edge is at 50 keV. oratory. The ion chamber is also used to evaluate the exposure
rate at a distance from a patient who has received radionuclide
therapy (e.g., iodine-131 for thyroid cancer) to determine that
The pulse height spectrum corresponding to the detection the patient can be released without exposing the general public
of the 140-keV gamma rays from technetium-99m (Tc-99m) is to unacceptable radiation levels. The ion chamber also should
illustrated in Fig. 2.4. The photopeak corresponds to events where be annually calibrated.
the entire energy of the incident photon is absorbed within the The dose calibrator is an ionization chamber used to assay
detector. These are the events of primary interest in most count- the amount of activity in vials and syringes. This includes the
ing experiments, and thus the good events are within an energy assay of individual doses before administration to patients, as
acceptance window about the photopeak. Other events corre- required by regulation. The dose calibrator operates over a very
spond to photons scattered within the detector material and wide range of activities, from tens of microcuries to a curie
depositing energy, which can range from very low energy from (hundreds of Bq to tens of GBq). The device is also equipped
a very-small-angle scatter to a maximum 180-degree scatter (in with variable settings for each radionuclide to be measured,
the spectrum referred to as the Compton edge). Events below the with typically about 10 buttons for ready selection of the radio-
Compton edge correspond to these scattered events. In some nuclides commonly used in the clinic. In addition, buttons are
cases, photons can undergo multiple scatters and possibly result available for user-defined radionuclide selection. Others can be
in events between the Compton edge and the photopeak. Pho- selected by entering the appropriate code for that radionuclide
tons scattered within the patient and then detected may also into the system.
result in events in this energy region. Finally, the pulse-height The dose calibrator is used to assay the activity administered
spectrum will be blurred depending on the energy resolution of to the patient, and thus a comprehensive quality control pro-
the detector. Thus in Fig. 2.4, the photopeak has approximately gram is necessary. Regulations specify dose calibrator quality
a 10% spread because of the energy resolution associated with control program must meet the manufacturer’s recommenda-
NaI, rather than the narrow spike that might be expected from tions or national standards. Typically, the program comprises
the emission of a monoenergetic gamma ray. four basic quality control tests: geometry, accuracy, linearity,
and constancy.
The geometry protocol tests that the dose calibrator provides
ANCILLARY NUCLEAR MEDICINE EQUIPMENT the same reading for the same amount of activity irrespective of
Besides the imaging equipment in the nuclear medicine clinic, the volume or orientation of the sample. A reading of a certain
other additional ancillary equipment may be necessary from amount of activity in a 0.5-mL volume is obtained. The volume
either a medical or regulatory point of view or to otherwise is then increased by augmenting the sample with amounts of
enhance the operation of the clinic. This equipment will be nonradioactive water or saline and taking additional readings.
reviewed, including the quality control required for proper The subsequent readings should not vary from the original
operation. readings by more than 10%. The geometry test is performed
As previously discussed, the two basic radiation meters com- during acceptance testing and after a major repair or move of
monly used in the nuclear medicine clinic are the Geiger-Müller the equipment to another location.
(GM) meter and the ionization chamber. Both are gas detec- For accuracy, calibrated sources (typically cobalt-57 and
tors, although they operate differently. With the GM meter, all 137Cs) are assayed; the resultant reading cannot vary by more

detections lead to a terminal event of the same magnitude—a than 10% from the calibrated activity decay corrected to the
“click.” The device is excellent for detecting small amounts of day of the test. The accuracy test should be performed during
18 PART 1 Basic Principles

acceptance testing, annually thereafter, and after a major repair of standards of known activity concentration (kilobecquerel per
or move. milliliter), the patient’s GFR can be estimated.
The linearity protocol tests that the dose calibrator operates The thyroid probe consists of an NaI crystal on a stand with
appropriately over the wide activity range to which it is applied. the associated counting electronics. The patient is administered
The device is tested from 10 μCi (370 kBq) to a level higher a small amount of radioactive iodine. The probe is placed at a
than that routinely used in the clinic and perhaps as high as 1 certain distance from the thyroid, and a count is obtained. In
Ci (37 GBq). The activity readings are varied by starting with addition, a count is acquired of a known standard at the same
a sample of the radioactivity of Tc-99m at the highest value to distance. The thyroid uptake of iodine can be estimated from
be tested (e.g., tens of gigabequerels). The activity readings are these measurements.
then varied by either allowing the source to radioactively decay The quality control program for both the well counter and
over several days or using a set of lead shields of varying thick- the thyroid probe includes the energy calibration, the energy
nesses until a reading close to 370 kBq is obtained. Each reading resolution, the sensitivity, and the chi-square test. For the energy
should not vary by more than 10% from the line drawn through calibration, the energy window is set for the calibration source
the calculated activity values. The linearity test should be per- of a particular radionuclide—for example, the 662-keV peak of
formed during acceptance testing, quarterly thereafter, and after Cs-137. The amplifier gain is varied until the maximum count is
a major repair or move. found that corresponds with the alignment of the window with
The constancy protocol tests the reproducibility of the read- the 662-keV energy peak. In addition, the counts in a series of
ings compared with a decay-corrected estimate for a reference narrow energy windows across the peak can be measured to
reading obtained from the dose calibrator on a particular day. estimate the energy resolution. A standard window can be set,
Today’s constancy reading cannot vary from the decay-cor- and the counts of a known calibration source can be counted
rected reference reading by more than 10%. The constancy test and normalized by the number of nuclear transformations to
varies from accuracy in that it evaluates the precision of the estimate the sensitivity in counts per transformation (or counts
readings from day to day rather than accuracy. The constancy per second per becquerel). Finally, the chi-square test evaluates
test should be performed on every day that the device is used to the operation of the counter by comparing the uncertainty of
assay a dose to be administered to a patient. the count to that expected from the Poisson distribution.
There are two nonimaging scintillator devices, the well
counter, and the thyroid probe that are routinely used in the
nuclear medicine clinic. The well counter is used for both radi-
THE PATIENT AS A RADIOACTIVE SOURCE
ation protection and clinical protocols. The thyroid probe can In nuclear medicine, the patient is administered a radiophar-
provide clinical studies with a fraction of the equipment costs maceutical that distributes according to a specific physiological
and space requirements of the use of nuclear imaging equip- or functional pathway. The patient is then imaged using exter-
ment. However, these devices also require comprehensive qual- nal radiation detectors to determine the in vivo distribution
ity control programs. and dynamics of the radiopharmaceutical through which the
The well counter consists of an NaI crystal with a hole in it, patient’s physiology can be inferred, providing this essential
allowing for test tubes, and other samples can be placed within information to the patient’s doctor to aid in diagnosis, prog-
the device for counting. The samples to be placed in the counter nosis, staging, and treatment. The equipment used to acquire
is practically surrounded by the detector, with a geometrical these data will be described in the sections ahead. Single-pho-
efficiency in excess of 90%. Thus the well counter can measure ton emission computed tomography (SPECT) and PET are
very small amounts of radioactivity, on the order of a kilobec- described in the next chapter. However, before examining how
querel. The well counter should not be confused with the dose the instrumentation operates, it is instructive to understand the
calibrator, which is a gas-filled ionization chamber that can nature of the signal itself—that is, the radiation being emitted
measure activities up to 37 GBq. It is used to test packages of from within the patient.
radiopharmaceuticals to ensure that no radioactivity has been The radiopharmaceutical is administered to the patient most
spilled on the outside of the package or leaked from the inside. commonly by intravenous injection but also in some cases
The device also can be used to measure removable activity from through other injection routes, such as intraarterial, intraper-
working surfaces where radioactivity has been handled or from itoneal, or subdermal. In other cases, the radiopharmaceutical
sealed sources such as calibration sources to ensure that the may be introduced through the gastrointestinal tract or through
radioactivity is not leaking out. the breathing of a radioactive gas or aerosol. After administra-
The well counter can also be used for the assay of biologi- tion, the path and rate of uptake depend on the particular radio-
cal samples for radioactivity for a variety of clinical evaluations. pharmaceutical, the route of administration, and the patient’s
For example, after the administration of Tc-99m diethylenetri- individual physiology. However, the characteristics and param-
amepentaacetic acid (DTPA), blood samples can be counted at eters associated with the radiopharmaceutical in vivo distribu-
several time points (e.g., at 1, 2, and 3 hours) to estimate the tion and dynamics are of considerable clinical importance. In
patient’s glomerular filtration rate (GFR). The amount of radio- some cases, the enhanced uptake of the radiopharmaceutical in
activity in a 0.2-mL blood sample will be very small, and thus the certain tissues (e.g., the uptake of fluorodeoxyglucose [FDG] in
well counter is the appropriate instrument for these measure- tumors) may be of most clinical importance, whereas in other
ments. By making these measurements and the measurements cases it may be the lack of uptake (e.g., the absence of Tc-99m
Chapter 2 Radiation Detection & Ancillary Instrumentation 19

sestamibi in infarcted myocardium). In the first case, this would


be referred to as a hot-spot imaging task, and in the latter would
be a cold-spot task. In other situations, it may be the rate of
uptake (wash in) or clearance (wash out) that may be considered
the essential characteristic of the study. In a Tc-99m mercapto-
acetyltriglycine (MAG3) renal study, fast wash in may indicate a
well-perfused kidney, and delayed clearance may indicate renal
obstruction. In the Tc-99m DTPA counting protocol described
previously, a slow clearance of the radiopharmaceutical from
the blood would indicate a reduced GFR. In some cases, the
ability to discern uptake in a particular structure that is adjacent
to other nonspecific uptake may require the ability to spatially
resolve the two structures, whereas other tasks may not require
such specific resolution. The choice of instrumentation, acqui-
sition protocol, and data-processing approach fundamentally
depend on the clinical task at hand. Fig. 2.5 Energy spectrum from a gamma camera with the techne-
To characterize the rate, location, and magnitude of radio- tium-99m (Tc-99m) activity in the patient. Note the loss of definition
of the lower edge of the Tc-99m photopeak. This spectrum illustrates
pharmaceutical uptake within the patient, the emitted radia-
the difficulty of discriminating Compton-scattered photons within the
tion must be detected, in most cases, by detectors external to patient using pulse-height analysis.
the patient’s body. Some instruments are specially designed
for internal use—for example, interoperative radiopharma-
ceutical imaging—but in most the cases, the imaging device only allowing photons to be counted within a narrow energy
is located outside the body while detecting radiation inter- window about the photopeak energy) will lead to the elimination
nally. This requirement limits the useful emitted radiations of a significant number of scattered photons from the nuclear
for nuclear medicine imaging to energetic photons—that medicine image. In contrast to the case of a point source, a more
is, gamma rays and x-rays. The amount of overlying tissue challenging clinical case with regard to scatter may be the imag-
between the internally distributed radiopharmaceutical and ing of a cold-spot feature, such as an infarction in a myocardial
the radiation detector may vary from several centimeters to perfusion scan or a renal scar in a Tc-99m DMSA scan. In these
as much as 20 to 30 cm. Alpha and beta particles will not be cases, scattered photons in the neighboring tissue may be dis-
of use in most cases because their ranges in tissue are lim- placed into the cold spot, leading to a loss in image contrast and
ited to a few millimeters, and thus they will not exit the body an inability to properly discern the extent of the feature. It must
and cannot be measured by external radiation detectors. Even also be kept in mind that in a true clinical case, the distribution
x-rays and gamma rays must have energies in excess of 50 of the radiopharmaceutical is unknown, and background levels
keV to penetrate 10 cm of tissue. On the other hand, once the in other tissues may compromise the situation. The pulse-height
radiation exits the patient, it is best that the radiation not be spectrum from a patient is shown in Fig. 2.5.
so energetic as to be difficult to detect with reasonable-size
detectors. Thus the radiation types optimal for most nuclear
medicine imaging applications are x-rays and gamma rays in
GAMMA CAMERAS
the 50- to 600-keV energy range, depending on the equip- In the earliest days of nuclear medicine, counting devices simi-
ment and collimation being used. lar to the thyroid probe described in the previous section were
Consider a situation in which a radiopharmaceutical labeled used to evaluate the amount of activity in a particular tissue.
with Tc-99m leads to a point source at some depth within the For example, probes could be used to evaluate the iodine uptake
patient’s body. The 140-keV gamma rays will be emitted isotrop- of the thyroid gland. However, it was not long before clinicians
ically from the point source. Therefore it would be advantageous realized that it would be helpful to not only know the total
to place the radiation detector close to the source or to place sev- uptake of the radiopharmaceutical within the tissue of inter-
eral detectors around the source to collect as many of the emitted est but also to be able to discern the spatial distribution of the
photons as possible. In fact, acquiring data from several angles uptake within the tissue. In the early 1950s, Benedict Cassan
may allow the source to be better localized. Those emitted pho- attached a focused collimator to an NaI crystal and a mecha-
tons that exit the body without interaction and are subsequently nism for acquiring the counts from the patient at multiple loca-
detected will yield the highest quality spatial information. tions in a raster fashion and plotting the spatial distribution of
Conversely, those photons that scatter within the patient com- the counts. This device, the rectilinear scanner, provided nuclear
promise spatial information. Photons that undergo very-small- medicine images of physiological function. As a result the term
angle scatter will perhaps not be of much consequence, but those scan, as in a thyroid or bone scan, has remained in the nuclear
that undergo scatter at larger angles will not be of much use. Not- medicine lexicon. However, these scans took a long time to
ing that the Compton-scattered photons have less energy than acquire and did not allow for the acquisition of time-sequence
the incident photons, and that small-angle scatter leads to less or dynamic studies. Still, the rectilinear scanner continued to be
energy loss than large-angle scatter, energy discrimination (i.e., used in nuclear medicine clinics through the late 1970s.
20 PART 1 Basic Principles

Collimator
Crystal

Positioning
pulses

Position computer
X-ray or Gamma
gamma ray (X)
camera
photon oscilloscope
(Y) and
computer
memory

PMTs Logic pulses

Addition Pulse
circuit height
analyzer

(z)
“Z” pulse
Fig. 2.6 Schematic of gamma scintillation camera. The diagram shows a photon reaching the NaI crystal
through the collimator and undergoing photoelectric absorption. The photomultiplier tubes (PMTs) are opti-
cally coupled to the NaI crystal. The electrical outputs from the respective PMTs are further processed through
positioning circuitry to calculate (x, y)-coordinates and through additional circuitry to calculate the deposited
energy of the pulse. The energy signal passes through the pulse height analyzer. If the event is accepted, it is
recorded spatially in the location determined by the (x,y)-positioning pulses.

In the mid-1950s, Hal Anger developed his first prototype The detection material of the gamma camera is typically a sin-
of the gamma camera, which allowed a section of the body to gle, thin large-area NaI scintillation crystal. Some smaller cameras
be imaged without a raster scan, opening the door for the pos- rely on a 2D matrix of smaller crystals, but most rely on a single
sibility of both dynamic and physiologically gated studies. Fur- large crystal. In the most common gamma camera designs, the NaI
ther developments of the technology took place over the next 10 crystal is about 30 cm × 50 cm in area and 9.5 mm thick. Some
years, and the first commercial gamma camera was introduced cameras designed for imaging only photons with energies below
in the mid-1960s. With further advances that have improved and 150 keV may have thinner crystals. Others used more commonly
stabilized the operation of the instrument, along with the addi- for higher-energy photons may be thicker, but the 9.5-mm thick-
tion of tomographic capability, the gamma camera remains the ness provides a reasonable compromise because it detects more
most commonly used imaging device in the nuclear medicine than 85% of the photons with energies of 140 keV or lower and
clinic. stops about 28% of the 364-keV gamma rays emitted by I-131. NaI
A block diagram of the gamma camera is shown in Fig. 2.6. is hygroscopic and thus damaged by water. It is hermetically sealed
Gamma rays emitted from within the patient pass through the and has a transparent light guide on the side adjacent to the PMT
holes of an absorptive collimator to reach the NaI crystal. On array and aluminum on the side closer to the collimator. The NaI
interaction of the gamma ray with the NaI scintillating crystal, crystal is the most fragile component of the gamma camera, being
thousands of light photons are emitted, a portion of which are susceptible to both physical and thermal shock. When the collima-
collected by an array of PMTs. By taking weighted sums of the tor is not in place, the bare NaI crystal must be treated with extreme
PMT signals within the associated computer, the two-dimen- care. In addition, the environment in the room must be controlled
sional (2D) x- and y-location and the total energy of the detec- so that the air temperature is maintained at a reasonable level (18-
tion event deposited are estimated. If the energy deposited is 24°C) and is not subject to wide variations over a short period.
within a prespecified energy window (e.g., within 10% of the The PMT array consists of about 60 to 100 photomultiplier
photopeak energy), the event is accepted, and the location of tubes that are each about 5 cm in diameter. The PMTs are usually
the event recorded. In this manner, the gamma camera image hexagonal and arranged in a hexagonal close-packed array to col-
is constructed on an event-by-event basis, and a single nuclear lect as many light photons as possible. Although PMTs are used
medicine image may consist of hundreds of thousands of such in practically all gamma cameras, some small camera designs are
events. Each component of the gamma camera will be described. using avalanche photodiodes to collect the scintillation light. The
Chapter 2 Radiation Detection & Ancillary Instrumentation 21

signal from each PMT is input into the gamma camera host com-
puter. First, the sum of all of the PMT signals is used to estimate
the energy deposited in the detection event. In addition, each
PMT has a weight associated with its position in both the x- and
y-direction. For example, the PMTs on the left side of the camera
may have a low weight, and those on the right side of the cam-
era would have a higher weight. For a particular detection event,
if the weighted sum of the signal is low, the event would be on
the left side, and if it were high, it would be on the right side of
the camera. However, the weighted sum as described is depen-
dent not only on the position of the event but also on the total
amount of light collected, which is directly proportional to the
energy deposited. Therefore the sum must be normalized by
the energy estimate. This approach to determining the position
of the detection event is often referred as Anger logic, in honor
of the developer of the gamma camera, Hal Anger. This leads
to an estimate of the detection event location to within 3 to 4
mm, which is referred to as the intrinsic spatial resolution of the
camera.
However, distortions can occur in images with respect to both
the energy and position estimates. Detection events directly
over PMTs lead to the collection of slightly more light than the
events between PMTs and therefore to a slightly higher pulse
height. Energy calibration notes the shift in the pulse height
spectrum as a function of position. Subsequently, an opposite
shift is applied on an event-by-event basis, leading to improved
energy resolution and greater energy stability. In addition, there
is an inherent nonlinearity, with events being bunched over
PMTs and spread out between PMTs. Analogous to energy cal-
ibration, linearity calibration determines the spatial shift from
linearity as a function of position across the entire field of view.
Again, these shifts in both energy and position are applied on
an event-by-event basis, providing an image that is free of lin-
ear distortion. A very-high-count uniformity calibration map
Fig. 2.7 Pinhole collimator. The image is inverted. The image is mag-
is acquired that characterizes the remaining nonuniformities nified if the distance from the aperture to the object is smaller than
inherent in the gamma camera acquisition process. These uni- the distance from the aperture to the gamma camera crystal. Spatial
formity calibration maps are used to generate uniformity cor- resolution improves and sensitivity decreases as the aperture diameter
rections that are applied during each acquisition. decreases. The sensitivity also decreases with the source-to-aperture
distance, according to the inverse square law. In general, the pinhole
collimator provides the best spatial resolution and the lowest sensitivity
COLLIMATORS of any collimator used in nuclear medicine.

Although the NaI crystal, PMTs, and electronics can estimate the
location of a detection event to within 3 to 4 mm, the direction- The simplest is the pinhole collimator (Fig. 2.7). It consists of
ality of the event is not known. Gamma rays from a point source a single, small hole or aperture located a set distance (typically
could be detected anywhere across the field of view, and the counts on the order of 20 cm) from the surface of the NaI crystal. Pho-
detected at a particular location in the NaI crystal could have also tons from one end of the source that pass through the aperture
originated from practically anywhere within the patient. Thus col- will be detected on the opposite side of the detector. In addi-
limation is required to determine the directionality of the detected tion, objects that are closer to the aperture will be magnified
event. Because gamma rays cannot be easily focused, absorptive compared with those farther away. If b is the distance from the
collimation must be used—that is, all photons not heading in the aperture to the object and f is the distance from the aperture to
desired direction will be absorbed by the collimator, and those the detector, the amount of magnification, M, is given by:
heading in the correct direction will be allowed to pass. There- M = f/b
fore absorptive collimation is inherently very insensitive because
practically all of the emitted photons will be absorbed and only a Magnification can be of significant value when imaging
very few will be accepted. In general, only 0.01% (i.e., 1 in 10,000) small objects using a camera with a large field of view. Magnifi-
of the photons emitted from the patient will be accepted by the cation will minimize the effect of the intrinsic spatial resolution
collimator and incorporated into the image. of the camera and thus enhance overall system resolution. The
22 PART 1 Basic Principles

of the collimators commonly used in nuclear medicine. It is often


used when imaging small organs (e.g., the thyroid gland) with a
gamma camera with a large field of view or in special cases when
a very-high-resolution spot view image is required, such as when
trying to discern which bone in the foot may be demonstrating
increased radiopharmaceutical uptake on a bone scan.
The multihole collimator provides substantially better geo-
metrical sensitivity compared with the pinhole collimator
because the object is viewed through many small holes rather
than through a single hole. The most commonly used multi-
hole collimators consist of a very large number of parallel holes
with absorptive septa between the holes to restrict the emit-
ted gamma rays from traversing from one hole to a neighbor-
ing hole. The holes are typically hexagonal and arranged in a
hexagonal, close-packed array (Fig. 2.8). A typical low-energy,
Fig. 2.8 Multihole, parallel-hole collimator. The collimator shown at the parallel-hole collimator may have hole diameters and lengths
top has longer holes designed to provide higher resolution. However,
it would also have a lower sensitivity than the collimator shown at the
of about 1 and 20 mm, respectively, and septal thicknesses
bottom. The septal thickness and thus energy rating are the same for between holes of about 0.1 mm. No magnification occurs with a
both collimators. parallel-hole collimator. The collimator spatial resolution
depends on the diameter (d) and the length (a) of the collimator
holes and the distance from the source to collimator (b):
collimator spatial resolution of the pinhole, RPH, is determined RP = (d / a) (a + b)
by the diameter or the aperture, d (typically 4-6 mm) and the dis-
tances from both the object and the NaI crystal to the aperture. A parallel-hole collimator with either small or long holes will
provide the best spatial resolution (Fig. 2.8, top). Similar to the
RPH = d × (f + b) /f
pinhole collimator, the spatial resolution of the parallel-hole
It must be kept in mind that spatial resolution is typically char- collimator is best at the surface of the collimator and degrades
acterized by the size of an imaged point source, and thus a large with distance from the collimator. The geometrical sensitivity
value corresponds to poor resolution, and a very small value indi- of the parallel-hole collimator also depends on the thickness of
cates excellent spatial resolution. A system with 1-mm spatial the septa between the holes (t) in addition to the hole diameter
resolution will lead to an image with greater acuity than one with and length:
5-mm spatial resolution. For this reason, the term high resolution 2 2
can be ambiguous because it may be unclear whether this system GP ≈ 1 / 16 (d / a) d / (d + t)
has very high resolution or a high R value (poor resolution). Based The geometrical sensitivity will be the highest for a col-
on the earlier formula, better spatial resolution is attained using limator with the thinnest interhole septa. On the other hand,
a smaller aperture (small d value) with the source as close to the the septa must be thick enough to minimize septal penetration
pinhole aperture as possible. In fact, all gamma camera collimators when a photon enters one hole, traverses the septa, and enters
provide the best spatial resolution very close to the collimator, and the neighboring hole. The septa are typically designed to be as
spatial resolution will degrade as the object is moved farther from thin as possible while limiting the amount of septal penetration
the collimator. The geometrical sensitivity of the pinhole collima- to less than 5% of photons striking the septa. As a result, colli-
tor, GPH, depends on the area of the pinhole (πd2) compared with mators designed for higher energy photons (over 200 keV) will
the squared distance of the source from the pinhole (b2): require thicker septa than those designed for lower energies.
2 Converse to collimator spatial resolution, the best geometrical
GPH ≈ 1/16 (d / b)
sensitivity is attained with a collimator with either large or short
Thus the geometrical collimator sensitivity is highest with a large holes. Thus again, a trade-off exists between spatial resolution
aperture diameter and drops off as the inverse square of the dis- and geometrical sensitivity. Because the geometrical sensitivity
tance from the source to the aperture—that is, it follows the inverse is proportional to (d/a)2 and the spatial resolution is propor-
square law. A larger aperture diameter leads to better geometri- tional to (d/a), the geometrical sensitivity of the collimator is
cal sensitivity but poorer spatial resolution, whereas the converse roughly proportional to the square of the spatial resolution:
is true for a smaller aperture diameter. As is true in some other
GP ∝ R2P
instances in nuclear medicine imaging, a trade-off occurs between
sensitivity and spatial resolution such that improvement in one Finally, it is notable that the geometrical sensitivity of a paral-
area may cause degradation of another. The choice of whether to lel-hole collimator does not depend on the distance between the
use high sensitivity or better resolution may depend on the clinical source and the collimator. The sensitivity is the same at the surface
imaging task at hand, but often a compromise will lead to a rea- as it is at a distance removed from the surface. This fact may be
sonable value for both parameters. The pinhole collimator typically counterintuitive because it might be expected that the sensitivity
provides the best spatial resolution and the lowest sensitivity of all would drop off with distance as it does with the pinhole collimator.
Chapter 2 Radiation Detection & Ancillary Instrumentation 23

The extrinsic or system spatial resolution (RE) depends on


both the intrinsic and collimator geometrical spatial resolution
(RI and RC, respectively). To first order, the relationship between
these is given by:

RE = R2I = R2C

Based on this equation, the larger of the two values, the intrinsic
or the collimator resolution, will dominate the system resolu-
tion. Except at distances very close to the collimator face, the
collimator spatial resolution is substantially greater than the
intrinsic resolution, and thus the collimator spatial resolution
is, in general, the more important factor. In cases involving
magnification, the intrinsic spatial resolution, RI, is modified
by magnification, thus minimizing the effect of intrinsic spatial
resolution on system spatial resolution:

( )2
RI
RE = = R2C
M

The system spatial resolution and the collimator geometrical


sensitivity vary as a function of the distance from the radioac-
tive source to the collimator (Fig. 2.10). The system spatial reso-
lution of all of the collimators degrades with increasing distance
from the collimator (see Fig. 2.10). The pinhole provides the
best spatial resolution, followed by the converging collimator
and two types of parallel-hole collimators, the high-resolution
and general-purpose collimators. On the other hand, the pin-
Fig. 2.9 Converging-hole collimator. With this collimator, objects are hole collimator has the poorest geometrical sensitivity, which
magnified, which tends to minimize the blurring effects of the intrinsic varies as the square of the distance (see Fig. 2.10). For the two
spatial resolution and thus provide higher system spatial resolution. In parallel-hole collimators, the sensitivity does not vary with dis-
addition, the sensitivity increases with distance as the collimator’s focal
distance is approached, and thus it provides both improved spatial reso-
tance, and the sensitivity of the converging collimator improves
lution and higher sensitivity, but with a decreased field of view. with distance.
The standard gamma camera can be used for various stud-
ies; however, some other nuclear medicine imaging devices are
In fact, the sensitivity of a single hole of the collimator does go designed for very specific clinical applications. These often use
down with distance, but the degrading spatial resolution leads to novel approaches to either gamma ray or light detection. In some
the irradiation of more holes, and these two facts cancel each other. instances, they use a semiconductor detector such as CZT. In other
The converging multihole collimator provides both enhanced cases, they may use avalanche photodiode or silicon PMTs for light
spatial resolution and improved sensitivity. With the converging detection in conjunction with a scintillator. The most notable clin-
collimator (Fig. 2.9), the direction of the holes is focused at a ical planar imaging application for these types of devices is breast
point some distance from the collimator surface. The distance imaging. The compact size allows the device to stay close to the
from the collimator to the focal point is typically on the order breast, resulting in high spatial resolution. In addition, the camera
of 50 cm and thus far beyond the boundaries of the patient. The can be designed with limited dead space between the edge of the
focusing provides magnification similar to that with the pinhole field of view and the patient, allowing imaging close to the chest
collimator. As a result, the spatial resolution is typically slightly wall. All of these characteristics result in improved imaging with
better than that with a parallel collimator but not as good as this device relative to the standard gamma camera.
that with a pinhole collimator. In addition, the geometrical
sensitivity of the converging collimator improves as the source
approaches the focal point, and thus the sensitivity improves at
QUALITY CONTROL
distances farther from the collimator. On the other hand, the To ensure proper operation of any medical device, including the
field of view is slightly reduced at greater distances because of gamma camera, it is essential that a comprehensive quality con-
the increased magnification. The converging collimator is used trol program be applied. This involves acceptance testing of the
in applications similar to those with the pinhole collimator— device before its initial use and a program of routine tests and
that is, for imaging smaller objects using a camera with a large evaluations applied on a regular basis. It is essential that the per-
field of view and to achieve a magnified image with slightly formance be evaluated regularly to ensure that the images ade-
improved spatial resolution. quately demonstrate the in vivo distribution of the administered
24 PART 1 Basic Principles

12 25

Extrinsic sensitivity (rel. units)

Ext spatial resolution (mm)


10 20
8
15
6
10
4

2 5

1 6 11 1 6 11
Distance from collimator (cm) Distance from collimator (cm)
Fig. 2.10 System sensitivity (left) and spatial resolution (right) as a function of the source-to-collimator dis-
tance. The system spatial resolution of all collimators degrades (increases in value) with distance. The pinhole
collimator (blue) provides the best spatial resolution (lowest value) but the lowest sensitivity, which varies
according to the inverse square law. The converging collimator (green) provides very good spatial resolution
with a sensitivity that increases with distance as the source approaches the focal distance of the collimator.
The high-resolution parallel-hole collimator (pink) has good resolution and reasonable sensitivity. The gener-
al-purpose, parallel-hole collimator (red) has poorer spatial resolution than the high-resolution collimator but
with a 50% increase in sensitivity. It is noted that the sensitivity of the two parallel-hole collimators does not
vary with distance.

is the central 50% of the area of the UFOV. The U.S. National
Electronic Manufacturers Association, a trade association of
electronic manufacturers, has defined parameters for gamma
camera manufacturers to use to characterize the performance
of their equipment. Although some of these parameters may be
difficult to assess in the clinic, they still provide the basis for
many of the quantitative measures used in the gamma camera
quality control program.
The uniformity (or flood) test evaluates the consistency of the
response of the gamma camera to a uniform flux of radiation (Fig.
2.11). It should not be confused with the high-count uniformity
calibration. The uniformity test can be applied either intrinsically
or extrinsically. For the intrinsic test, the collimator is removed and
exposed to the radiation from a point source of small activity
(about .05 mCi (2 MBq) at a distance far enough to ensure uniform
Fig. 2.11 Uniformity (flood) phantom image. irradiation of the camera’s field of view (at least 2 m). Extrinsic
flood images are acquired with a large-area uniform source con-
radiopharmaceutical and that any quantitation performed with taining approximately 3 mCi to 10 mCi (111 to 370 MBq). This
the camera yields values that are as accurate and precise as possible. may consist of a thin water-filled source into which the radionu-
Gamma camera quality control involves tests that are either clide of choice (e.g., Tc-99m) is injected and thoroughly mixed.
quantitative or qualitative. For the quantitative tests, vari- More commonly, a solid, sealed, large-area source of Co-57 (122-
ous parameters are used to measure the characteristics of the keV gamma ray, 270-day half-life) is used. For routine testing, 5 to
gamma camera system. Some of these parameters are evaluated 20 million counts are acquired and the images evaluated qualita-
intrinsically (i.e., without a collimator, to characterize the optics tively for any notable nonuniformities. The daily flood should be
and electronics of the system) and other parameters extrinsi- acquired before administering the radiopharmaceutical to the first
cally to include the collimator. If extrinsic tests are performed patient to ensure that the camera is working properly. Extrinsic
frequently (e.g., daily), they should be performed using the col- floods for all collimators used with the camera may be acquired on
limators most commonly used in the clinic. However, it may an annual basis.
be best to perform the tests with all of the collimators used in Gamma camera spatial resolution can be evaluated either
the clinic at least annually. Certain parameters may be evalu- intrinsically or extrinsically, qualitatively or quantitatively. In
ated in different parts of the gamma camera’s field of view. The general, it may be evaluated quantitatively only during accep-
useful field of view (UFOV) is the portion of the field of view the tance testing and perhaps during annual testing using very small
manufacturer has designated to be the proper extent for clinical point or line sources. The spatial resolution is characterized by the
imaging. Although the UFOV typically covers more than 95% width of the image of the small source. Typical values for intrinsic
of the total field of view, it may not extend to the very edge of the spatial resolution range from 3 to 4 mm. Extrinsic values depend
NaI crystal or collimator face. The central field of view (CFOV) on the particular collimator being evaluated and the distance
Chapter 2 Radiation Detection & Ancillary Instrumentation 25

at which the test was performed, but for collimators commonly bar phantom image can also be used to qualitatively test spatial
used in the clinic, the extrinsic spatial resolution at 10 cm ranges linearity by evaluating the straightness of the bars in the image.
from about 8 to 12 mm. In the clinic, a qualitative assessment of Other performance parameters or characteristics that can be
extrinsic spatial resolution, typically a four-quadrant bar phan- tested include the sensitivity, energy resolution, count rate per-
tom (Fig. 2.12), is more commonly performed. Each quadrant formance, and multiwindow registration. The sensitivity is most
of this phantom comprises alternating lead and spacing equal to commonly evaluated extrinsically using a small-area source (e.g.,
the width of the bars of varying sizes (e.g., 2.0, 2.5, 3 0, and 3.5 10 × 10 cm) of known activity (typically approximately
mm). For extrinsic spatial resolution at the collimator surface, the 1.1 -3.2 mCi (40 - 120 MBq)` of Tc-99m) placed on the collima-
phantom is placed on the collimator with the large-area unifor- tor being evaluated, counted for 1 minute and reported as the
mity source on top of it. The user reviews the resultant image and counts per minute per unit activity. For parallel-hole collimators,
determines how many of the quadrants of the phantom can be the distance of the source to the collimator is inconsequential
discerned as separate bars. In general, the bars that can be dis- because the sensitivity does not vary with distance. For the pin-
cerned should be approximately 60% of the quantitative spatial hole or focusing collimators, a standard distance such as 10 cm
resolution value. Thus, if the intrinsic spatial resolution is 3.5 mm, should be used. The sensitivity value will obviously depend on
it should be possible to discern 2-mm bars of the four-quadrant the collimator being evaluated, ranging from 5.0 to 8.5 cpm/kBq
bar phantom. The extrinsic spatial resolution at the surface of the for typical high-resolution and general-purpose collimators. For
most commonly used collimator in the clinic is qualitatively eval- the energy resolution, a pulse height (or energy) spectrum is
uated routinely on either a weekly or monthly basis. The num- acquired of a known radionuclide, typically Tc-99m, and the
ber of quadrants that can be discerned should be compared with width of the photopeak is determined in a manner similar to that
those determined during acceptance testing. The four-quadrant used for spatial resolution normalized by the gamma-ray energy.
A typical gamma camera will have an energy resolution of 9% to
11% at 140 keV. As discussed previously, radiation detectors take
a certain amount of time to process each event and, if events are
registered too quickly, some may be lost as a result of dead time
or count rate losses. The count rate performance can be evalu-
ated by using two sources of reasonably high activity to calculate
the dead time value in microseconds or by varying the exposure
rate to which the camera is exposed and recording the observed
count rate. For modern cameras, the maximum observable count
rate is typically between 200,000 and 400,000 counts per second
(cps). Finally, the multiwindow registration can be characterized.
As previously discussed, the gamma camera position estimate
obtained using Anger logic must be normalized by the energy
deposited so that the position estimate does not vary as a func-
tion of photon energy. For this test, point sources of gallium-67,
which emits photons of three different energies (90, 190, and 300
keV), are placed in several locations within the gamma camera
Fig. 2.12 Four-quadrant bar phantom image for spatial resolution. field of view and the image location of each of the points sources

TABLE 2.1 Gamma Camera Quality Control Summary


Parameter Comment
Daily
Uniformity Flood field; intrinsic (without collimator) or extrinsic (with collimator)
Window setting Confirm energy window setting relative to photopeak for each radionuclide used with each patient

Weekly or Monthly
Spatial resolution Requires a “resolution” phantom such as the four-quadrant bar
Linearity check Qualitative assessment of bar pattern linearity

Annually
System uniformity High count flood with each collimator
Multiwindow registration For cameras with the capability of imaging multiple energy windows simultaneously
Count rate performance Vary counts using decay or absorber method
Energy resolution Easiest in cameras with built-in multichannel analyzers
System sensitivity Count rate performance per unit of activity for each collimator
26 PART 1 Basic Principles

are evaluated to make sure that they do not vary depending on appropriate quality control program for a specific gamma camera
which photopeak was imaged. depends on manufacturer recommendations and the clinical use
These parameters should be evaluated and compared with and stability of performance for that particular camera.
manufacturer specifications during acceptance testing. It is highly The x- and y-location of each accepted detection event is digi-
recommended that these tests be performed by a qualified nuclear tally stored within the acquisition host computer associated with
medicine physicist. After acceptance testing, quality control tests the gamma camera. These data can be captured in two ways—
will be run at various frequencies (daily, weekly, monthly, quar- matrix and list mode. In matrix mode, a specific matrix size (64 ×
terly, or annually), and the evaluation may be qualitative rather 64, 128 × 128, 256 × 256, and so on) is predetermined depending
than quantitative in these cases. Table 2.1 summarizes the rec- on the assumed spatial resolution of the imaging task. For tasks
ommended frequency for each of the described tests. These rec- that involve higher spatial resolution, a larger matrix would be
ommendations are for the typical gamma camera, and the most required. The chosen matrix size is mapped to the field of view,
and each estimated (x,y)-location is assigned to a particular pic-
ture element or pixel within the image matrix. The value of that
pixel is then incremented by 1. In this manner, a 2D histogram
of the event locations is generated, and at the end of the acquisi-
tion, the value in a particular pixel is the total number of events
assigned to that pixel during the data acquisition process. An
example with a 6 × 6 matrix is shown in Fig. 2.13. To display
the image, the number of counts in a particular pixel is assigned
a color or gray value according to a certain color scale lookup
table on the computer monitor. An example might be that the
pixel with the most counts is assigned the color white, pixels with
no counts are assigned black, and all other pixels are assigned a
A shade of gray. Alternatively, the colors of the rainbow could be
used, with violet indicating zero counts and red indicating the
highest count. In addition, many other color tables could be used.
In many cases, more than one image is acquired during the
imaging procedure. In some cases, a time-sequence of image
frames, also known as a dynamic study, may be acquired. For
example, a frame may be acquired every minute for 20 minutes. A
multiphase study may be acquired, in which ten 30-second frames
are followed by five 60-second frames, followed by five 120-second
frames. In other instances, the data acquisition may be associated
with a physiological gating signal such as the electrocardiogram
(ECG) or a respiratory gate. In the cardiac example, counts from
different parts of the heart cycle could be placed in different
B frames, resulting in frames from the end of diastole to the end of
systole and back again. In matrix mode, these multiframe acquisi-
tions would be obtained by establishing the desired number of the
frames in the computer a priori. During the acquisition process,
the appropriate pixel for each event would need to be determined,
in addition to the appropriate frame within the heart cycle.
In list mode, the (x,y)-location of each event is stored using
the highest level of digitization possible as a stream. In addition,
timing and physiological gating marks may be stored periodi-
cally. For example, a timing mark may be stored every millisec-
ond, as can the time of the R peak in the ECG (Fig. 2.14). After
the acquisition is complete, the user can then select the desired
matrix size and the temporal or physiological framing rate a
posteriori. Based on these criteria, a postacquisition program
C is run to format the data as defined. The user could then decide
Fig. 2.13 Digital image. Consider a nuclear medicine image acquired to reformat the data to a different set of parameters. In this way,
in matrix mode using a 6 × 6 matrix. (A) The calculated positions based list mode acquisition is very flexible because it does not require
on Anger logic a number of events. (B) A 6 × 6 matrix superimposed
the user to define the acquisition matrix and framing a priori.
onto these events demonstrates into which of the pixels of the matrix
each event would fit. (C) The number of events (dots) in each pixel is On the other hand, it typically requires more computer storage
recorded and assigned a particular shade of gray or another color to and running the formatting program to view the data. For these
create the digital matrix. reasons, matrix mode is most commonly used.
Chapter 2 Radiation Detection & Ancillary Instrumentation 27

Fig. 2.14 List-mode data acquisition. In list mode, the (x,y)-position of each detection event is determined at
the highest available resolution and stored in sequence. In addition, time markers and physiological signals
(such as the timing of the R wave from the electrocardiogram trigger) are periodically stored. Once the acqui-
sition is completed, the desired matrix size, time sequence, and physiological gate framing can be selected,
and a formatting program is run to provide the acquired data for viewing and analysis.

Each pixel in the planar nuclear medicine image can be con- which the patient’s physiology or function can be inferred to
sidered its own detector, and thus the total counts in a pixel are further define the patient’s medical picture. A rigorous quality
governed by Poisson statistics similar to a well counter or a thy- control program must be maintained for all equipment used in
roid probe. Therefore the standard deviation of the pixel counts the nuclear medicine clinic to ensure the integrity of the data
is simply estimated by the square root of the pixel counts. In obtained from the patient. The quality control program for
addition, the sum of Poisson distributed values is also a Poisson the gamma camera includes acceptance testing and tests that
distributed value. Therefore, if a region of interest is defined on need to be performed on a routine basis. The nuclear medicine
a planar nuclear medicine image and the pixel values within that image acquired with the gamma camera provides a snapshot
region are added, the result is also Poisson distributed. Nuclear of the patient’s in vivo radiopharmaceutical distribution from
medicine studies are often quantified by defining regions of a certain view and at a particular point in time. These images
interest (ROIs) over features of interest and subsequently com- can also be acquired as a dynamic (time-sequence) study or
paring the counts. In some cases, the counts from different in conjunction with a physiological gate such as the ECG.
views of the patient can be combined to provide more accurate ROIs can be drawn about specific features to provide regional
quantitation. For example, taking the geometrical mean (square quantitation or TACs of dynamic processes. Nuclear medicine
root of the product of the counts) of similar regions from oppo- instrumentation continues to evolve, including the develop-
site, conjugate views such as the anterior and posterior views, ment of devices designed for a specific clinical task, such as
can provide an estimate that, to first order, does not depend breast imaging. It is expected that this development will con-
on the depth of the activity within the body. Theoretically, this tinue in the years ahead.
approach works for point sources, but it also has been shown
to work reasonably well for extended sources. The counts are
determined from the ROIs drawn about each lung, right and
SUGGESTED READING
left, on images acquired from both the anterior and posterior Chandra R, Rahmim A. Nuclear Medicine Physics: The Basics. 8th ed
views of the patient. The geometrical mean of the counts in each Philadelphia: Williams & Wilkins. 2018.
lung is calculated, and the differential function of each lung is Cherry SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine. 4th
estimated by dividing the counts for that lung by the sum of ed. Philadelphia: WB Saunders. 2012.
counts for both lungs. International Atomic Energy Association. Nuclear Medicine Physics: A
Handbook for Teachers and Students. Vienna, Austria: Internation-
In a dynamic study, the region of interest counts in each
al Atomic Energy Agency. 2014.
frame can be plotted as a function of time. The resulting plot International Atomic Energy Association. Planning a Clinical PET
is referred to as the time–activity curve (TAC). In the case of Centre. Vienna, Austria: International Atomic Energy Agency.
relatively short-lived radionuclides, each value along the plot 2010.
should be decay corrected to the beginning of the acquisition or International Atomic Energy Association. Quality Assurance for PET
the time of radiopharmaceutical administration. In the example and PET-CT Systems. Vienna, Austria: International Atomic Ener-
of a Tc-99m MAG3 renal study, a TAC can be used to evaluate gy Agency. 2019.
both renal perfusion and clearance of the agent. International Atomic Energy Association. Quality Assurance of SPECT
Radiation detection and counting form the cornerstone Systems. Vienna, Austria: International Atomic Energy Agency. 2009.
of nuclear medicine. Detectors of all types—gas detectors, International Atomic Energy Association. Quality Control Atlas for
scintillators, and semiconductors—are used every day in the Scintillation Camera Systems. Vienna, Austria: International Atom-
ic Energy Agency. 2019.
nuclear medicine clinic. Some are used for ancillary purposes
National Electrical Manufacturers Association. Performance Mea-
that support the clinic, such as those used in the context of surements of Positron Emission Tomographs. Rosslyn, VA: National
radiation protection. Others are used specifically to acquire Electrical Manufacturers Association. 2018.
biological data for a particular clinical purpose. Most nota- Powsner RA, Powsner ER. Essentials of Nuclear Medicine Physics. 3rd
bly, the gamma camera is used to obtain images of the in vivo ed. Malden, MA: Blackwell Science. 2012.
distribution of the administered radiopharmaceutical from
3
Single-Photon Emission Computed
Tomography, Positron Emission Tomography,
and Hybrid Imaging

DATA ACQUISITION OF EMISSION reviews the current approaches to the acquisition and recon-
struction of SPECT and PET, including the use of hybrid imag-
TOMOGRAPHYN ing such as PET/CT, PET/MR and SPECT/CT, and the quality
Conventional or planar radionuclide imaging suffers a major lim- control necessary to ensure high-quality clinical results.
itation in the loss of object contrast as a result of background All tomographic modalities used in diagnostic imaging,
radioactivity. In the planar image, radioactivity underlying and including SPECT, PET, CT, and magnetic resonance imaging
overlying the object of interest is superimposed on that coming (MRI), acquire raw data in the form of projection data at a vari-
from the object. The fundamental goal of tomographic imaging ety of angles about the patient. Although SPECT and PET use
systems is a more accurate portrayal of the three-dimensional different approaches to acquiring these data, the nature of the
(3D) distribution of radioactivity in the patient, with improved data is essentially the same. Image reconstruction involves the
image contrast and definition of image detail. This is analogous to processing of these data to generate a series of cross-sectional
the way computed tomography (CT) provides better soft tissue images through the object of interest.
contrast than planar radiography. The Greek tomo means “to cut”; The geometries associated with the acquisition of SPECT
tomography may be thought of as a means of “cutting” the body and PET are illustrated in Fig. 3.1. In the simple SPECT example
into discrete image planes. Tomographic techniques have been using a parallel-hole collimator, the data acquired at a particular
developed for both single-photon and positron imaging, referred location in the gamma camera crystal originated from a line
to as single-photon emission computed tomography (SPECT) passing through that point perpendicular to the surface of the
and positron emission tomography (PET), respectively. sodium iodide (NaI) crystal face and is referred to in the figure
Restricted or limited-angle tomography keeps the plane of as the line of origin (see Fig. 3.1, left). Thus the data at this point
interest in focus while blurring the out-of-plane data in much can be seen to represent the sum of counts that originated along
the same way as conventional x-ray tomography. Various this line, or ray, referred to as a ray sum. These ray-sum values
restricted-angle systems have been investigated, including across the patient are referred to as the projection data for this
multi-pinhole collimator systems, pseudo-random, coded- cross-sectional slice at this particular viewing angle. For PET,
aperture collimator systems, and various rotating slant-hole the ray sum represents the data collected along a particular line
collimator systems. Although clinical use has been limited, of response (LOR) connecting a pair of detectors involved in a
resurgent interest has been shown for specific imaging applica- coincidence detection event (see Fig. 3.1, right).
tions, including those designed for cardiac and breast imaging. For a SPECT acquisition, the projection image acquired at
Tomographic approaches that acquire data over 180 or 360 each angle consists of the stack of projections for all slices within
degrees provide a more complete reconstruction of the object the camera field of view at that angle. Fig. 3.2, on the right, shows
and therefore are more widely used. Rotating gamma camera projections from a SPECT brain scan at five different viewing
SPECT systems offer the ability to perform true transaxial angles. For a particular slice (see Fig. 3.2, dashed white line), a
tomography. PET uses a method called annihilation coincidence row of the projection data for each angle can be stacked such that
detection to acquire data over 360 degrees without the use of the displacement along the projection is on the x-axis and the
absorptive collimation. The most important characteristic of viewing angle is on the y-axis (see Fig. 3.2, right). This plot is
these approaches is that only data arising in the image plane of referred to as the sinogram, because the resulting plot of a point
interest are used in the reconstruction of the tomographic source resembles a sine-wave plot turned on its side. A more
image. This is an important characteristic leading to improved complicated object such as a brain scan can be perceived as many
image contrast compared with methods using restricted-angle such sine waves overlaid on top of each other for each point
tomography. As will be discussed, the reconstruction of these within the object. The sinogram, represents the full set of projec-
data has historically been done with filtered back-projection. tion data necessary to reconstruct a particular single slice. A sep-
However, iterative techniques such as ordered subsets expecta- arate slice is made in the sinogram for each cross-sectional slice
tion maximization (OSEM) are increasingly used. This chapter through the object. The set of projection views and the set of

28
Chapter 3 Single-Photon Emission Computed Tomography 29

detectors that may be involved in an annihilation coincidence


detection event. These data thus represent the ray sum along
this LOR. The data associated with a particular LOR are charac-
terized in the sinogram by their distance from the center of the
gantry (on the x-axis) and their angle of orientation (on the
y-axis). In this manner, PET data acquisition directly into sino-
grams may be more straightforward than into projection views.
In a PET detection event, the two detectors involved in the coin-
cidence event are identified, and the LOR is recorded. The loca-
tion in the sinogram corresponding to that particular LOR is
localized, and its data are incremented. After the collection of
Fig. 3.1 Single-photon emission computed tomography (SPECT) and many such events, the projection data are represented by a set of
positron emission tomography (PET) acquisition geometries. For SPECT sinograms for each PET slice. However, these data also can be
(left), the gamma camera rotates about the patient, acquiring a projec- displayed as projection views similar to those acquired in
tion image at each angle. Each projection image represents the projec- SPECT studies. This simple example illustrates the acquisition
tions of many slices acquired at that angle. For PET (right), the patient
is located within a ring of detectors. A positron annihilation event leads
of PET data in 2D mode, in which each cross-sectional slice
to two photons emitted in opposite directions. When two events are basically is acquired separately. Most current PET scanners
detected within a small timing acceptance window (5-12 ns), they are acquire data only in 3D mode, in which LORs cut across the
considered to be from the same event and are assumed to have origi- parallel cross-section slices. The corresponding projection data
nated along the line of response that connects the two detectors. will include oblique views or sinograms through the object.
With time-of-flight PET (discussed later in this chapter), it is
necessary to record not only the LOR but also the time differ-
ence between the two detections involved in the annihilation
coincidence detection event, which will also be incorporated
into the reconstruction of these data.
Tomographic data can be acquired in a dynamic or gated
approach. For example, a PET study can be acquired as a time
sequence of scans that might be simple or multiphase (e.g., ten
5-second frames, four 30-second frames, and five 60-second
frames). In addition, the tomographic study can be acquired in
association with a physiological gate such as the electrocardio-
gram (ECG) or a respiratory signal. For example, myocardial
perfusion SPECT is acquired in conjunction with the ECG. In
dynamic or gated tomographic acquisitions, a full set of projec-
tion data acquired at each time or gate point is to be recon-
structed separately.

Tomographic Reconstruction
Images, like time signals, can be considered as either a spatial
variation of the signal or a sum of signals of varying frequencies.
It is intuitive to consider images as a spatial variation in the sig-
nal because some part of the image will be bright and other parts
will be dark. In nuclear medicine, the bright and dark areas may
correspond to regions of high and low radiopharmaceutical
uptake, respectively. Conversely, it is not intuitive to consider an
Fig. 3.2 Single-photon emission computed tomography (SPECT) projec-
tion images and sinograms. Right, projection images of a SPECT brain
image to comprise signals of varying frequency, although this is
scan at five different viewing angles. For a particular slice (indicated by the in fact the case. On the other hand, we do naturally perceive
dashed white line), the projection data can be stacked to form the sino- audio signals in terms of frequencies. A choral performance
gram (left). (From Henkin RE. Nuclear Medicine. St. Louis: Mosby; 2006.) comprises sopranos, altos, tenors, and basses, and the combina-
tion of these voices hopefully leads to a very pleasurable experi-
sinograms are alternative means of displaying the projection ence. On the other hand, we cannot perceive a presentation of
data associated with a tomographic acquisition. Each projection the audio signal as a temporal variation of the signal and intui-
view displays the projection data across all slices with a separate tively identify it as music. The music is fully described by either
image for each angle, whereas the sinogram displays the projec- representation, and there may be cases in which either the tem-
tion data across all angles with a separate sinogram for each slice. poral (i.e., real) or the frequency representation is the best
The geometry of PET acquisition (see Fig. 3.1, right) involves approach for considering the audio data. The same is true for
the data acquired along a particular LOR connecting two image data, except the variations are in space rather than time.
30 PART 1 Basic Principles

Backprojections

Image
profiles
Intersection of of point
backprojected source
rays at location
of point source

A B
Fig. 3.3 Simple back-projection. (A) The counts in each position along the projection are back-projected across
the reconstruction matrix because the algorithm has no knowledge as to the origin of the event. This process
is referred to as simple back-projection. (B) Simple back-projection leads to streak artifacts that render all but
the simplest objects discernable.

Image data may be best represented in either spatial (real) or caused by uneven sampling of frequency space during the
frequency space. The mathematician Joseph Fourier noted in back-projection process, where low frequencies are sampled at a
1807 that any arbitrary signal can be generated by adding a large much higher rate than higher frequencies. To compensate for
number of sine and cosine signals of varying frequencies and this, a filter, called the ramp filter, is applied during the recon-
amplitudes. The plot of amplitude as a function of frequency is struction that increases linearly with frequency (Fig. 3.4).
referred to as the Fourier transform, and it defines the compo- Applying back-projection in conjunction with such filtration is
nents of the image at each frequency. The low frequencies pro- referred to as filtered back-projection. With a very large number
vide the overall shape of the object, whereas the high frequencies of accurate, noiseless projections, filtered back-projection will
help define the sharp edges and fine detail within the image. yield an excellent, almost perfect reconstruction.
Audio signals can be manipulated by emphasizing certain fre- However, with true clinical data, the projections are noisy,
quencies (low or high); the same is true for images. Image noise and thus the ramp filter will tend to accentuate the high-frequency
is typically present in all frequencies; if the low frequencies are noise in the data. Therefore a windowing filter is applied, in
emphasized, the image may be less noisy but blurry, whereas addition to the ramp filter, to smoothly bring the filter back to
emphasizing the high frequencies will accentuate both the edges zero at frequencies above the pertinent content in the study.
of the objects and the noise. Such image manipulation is referred Commonly used windowing functions include the Hamming
to as filtering because it allows certain spatial frequencies to be and Butterworth filters (see Fig. 3.4). With these filters, a cutoff
realized while removing others. frequency is defined, which is the point at which they return to
Since the initial development of CT 40 years ago, filtered zero, with no higher frequencies being incorporated into the
back-projection has been the most common approach to recon- reconstructed image. Noting that low frequencies yield the
structing medical tomographic data, including SPECT, PET, overall shape and high frequencies yield the sharp edges and
and CT, although iterative techniques were introduced into the fine detail, the appearance of the resultant reconstructed image
clinic for use with PET more than a decade ago. However, fil- can be altered by varying the cutoff frequency. Selecting a cutoff
tered back-projection is still used in SPECT and remains the frequency that is too low will yield a blurry reconstruction (Fig.
most common method for CT. In back-projection, it is assumed 3.5, A, far left), and one that is too high will yield a noisy recon-
that all of the data detected at a particular point along the pro- struction (see Fig. 3.5, C, second from the right). However, an
jection originated from somewhere along the line emanating appropriate choice for cutoff frequency will provide an image
from this point. For SPECT using parallel-hole collimation, this that is a fair compromise between noise and detail (see Fig. 3.5,
would be the line of origin passing through the detection point B, second from left). With an appropriate choice of cutoff fre-
and perpendicular to the NaI crystal surface. For PET, events quency, filtered back-projection is a simple, fast, and robust
would be assumed to have come from the LOR connecting the approach to image reconstruction.
two detectors involved in the annihilation coincidence detec- Iterative reconstruction provides an alternative to filtered
tion event. In general, back-projection makes no assumptions of back-projection that tends to be less noisy, tends to have fewer
where along the line the event occurred, and thus the counts are streak artifacts, and often allows for the incorporation of certain
spread evenly along the line. In other words, the counts are physical factors associated with the data acquisition into the
back-projected along the line of origin or LOR. All of the counts reconstruction process, leading to a more accurate result. In
from every location along every projection are back-projected iterative reconstruction, an initial guess as to the 3D object that
across the reconstructed image (Fig. 3.3, A). The result is could have led to the set of acquired projections is estimated. In
referred to as simple back-projection; it has substantial streak addition, a model of the imaging process is assumed that may
artifacts that, in all but the simplest objects, render the recon- incorporate assumptions regarding photon attenuation and
structed image indiscernible (see Fig. 3.3, B). These streaks are Compton scatter. It may also include other assumptions
Chapter 3 Single-Photon Emission Computed Tomography 31

1.0 regarding the data-acquisition process, such as estimates of the


device’s spatial resolution that vary with position within the
Ramp field of view; for example, the variation of collimator spatial res-
olution as a function of the distance between the object and the
Amplitude

0.5
collimator can be incorporated into the reconstruction process.
Based on this model and the current estimate of the object, a
Cut-off new set of projections is simulated that is then compared with
frequency the real, acquired set. Variations between the two sets, parame-
terized by either the ratio or difference between pixel values, are
then back-projected and added to the current estimate of the
0.25 0.5 object to generate a new estimate (Fig. 3.6). These steps are
Frequency
(cycles/pixel)
repeated, or iterated, until an acceptable version of the object is
reached. The goodness of the current estimate is typically based
1.0 on statistical criteria such as the maximum likelihood. In other
words, the process generates an estimate of the object that has
Hamming
the highest statistical likelihood to have led to the set of acquired
projection data. A commonly used approach for the reconstruc-
Amplitude

0.5 tion of SPECT and PET data is the maximum-likelihood expec-


tation maximization (MLEM) algorithm.
Iterative reconstruction often leads to a more accurate recon-
struction of the data than that obtained through filtered
back-projection. However, a large number of iterations, perhaps
as many as 50, may be required to generate an acceptable esti-
0.25 0.5 mation, and each iteration may take about the same time as a
Frequency
(cycles/pixel)
single filtered back-projection; thus the iterative approach may
1.0
take 50 times longer to reconstruct. One approach to reducing
the number of iterations is to organize the projection data into a
series of ordered subsets of evenly spaced projections and
Butterworth
update the current estimate of the object after each subset rather
Amplitude

than after the complete set of projections. If the data are orga-
0.5 nized into 15 subsets, in general, the data can be reconstructed
about 15 times faster while generating a result of similar image
quality. A similar result can be produced with 15 ordered sub-
sets and 3 iterations as would be obtained with 45 iterations
Cut-off frequency using the complete set. The most common approach that uses
ordered subsets in the clinic is referred to as OSEM. Fig. 3.5, D
0.25 0.5
Frequency (far right) shows an OSEM reconstruction compared with a fil-
(cycles/pixel) tered back-projection of the same object. The use of faster algo-
Fig. 3.4 Ramp, Hamming, and Butterworth filters. The ramp filter is rithms such as OSEM and the development of faster computers
a “high-pass” filter designed to reduce background activity and the have allowed iterative reconstruction of SPECT and PET data in
star artifact. Hamming and Butterworth filters are “low-pass” filters
5 minutes or less, which is considered acceptable for clinical
designed to reduce high-frequency noise.

A B C D
Fig. 3.5 Effect of different filtration on reconstruction. (A) Single-photon emission computed tomography
(SPECT) study reconstructed with a cutoff frequency that is too smooth. The image is very blurry. (B) SPECT
study reconstructed with an appropriate cutoff frequency, with a moderate noise level and sharpness. (C)
SPECT study reconstructed with a cutoff frequency that is too sharp. The level of detail is good, but an
excessive amount of image noise is present. (D) SPECT study acquired with iterative reconstruction ordered
subsets expectation maximization (OSEM).
32 PART 1 Basic Principles

Current correct for attenuation. A similar analytic method has been


estimate Real
Simulated projections
developed for PET imaging, primarily of the brain.
projections The major limitation of the analytic approach occurs when
multiple types of tissue, each with a different attenuation coeffi-
cient, are in the field of view. This can be particularly problem-
atic for cardiac imaging, in which the soft tissues of the heart are
Modify surrounded by the air-containing lungs and the bony structures
estimate of the thorax. To correct for nonuniform attenuation, a trans-
Compare mission scanning approach is incorporated into the attenuation
correction. In essence, a CT scan of the thorax is obtained using
an x-ray tube. Older SPECT and PET systems also have used
Back-project
radionuclide sources for this purpose. The technique is similar
Difference
to the use of CT, except radioactive sources incorporated into
Fig. 3.6 Iterative reconstruction process. A set of simulated projections
the scanner are used rather than an x-ray tube. The data are
is generated from an initial guess of the object. This is compared with much noisier and require segmentation into the different tissue
the real projection data, and the difference is back-projected and added types before the attenuation map can be created. Manufacturers
to the initial guess. This process is iterated until the differences between are moving away from the radioactive source methodology.
the simulated and real projections are within an acceptable level. A hybrid SPECT-CT or PET-CT scanner is used to acquire a
CT over the same axial range as the SPECT or PET scan. The CT
work. With the development of even faster computers, iterative scan is acquired with a tube voltage of 80 to 120 kVp, leading to
reconstruction may be routinely applied to the larger data sets an effective energy of about 40 to 60 keV. The range of the
associated with CT in the near future. tube-current time product (milliamperes) is variable, depend-
ing on whether the CT scan is acquired for diagnostic purposes,
Attenuation Correction for anatomical correlation, or for attenuation correction. Thus
A special problem of both SPECT and PET imaging is the scans could be acquired with as little as 4 mA and as high as 400
attenuation of emissions in tissue. Photons emitted from deeper mA. A lookup table is used to convert the Hounsfield units in
within the object are more likely to be absorbed in the overlying the reconstructed CT scan to attenuation coefficients for the
tissue than those emitted from the periphery. Therefore the sig- desired photon energy. The resulting attenuation map can then
nals from these tissues are attenuated. To obtain an image where be applied as a postreconstruction correction or incorporated in
the signal is not depth dependent, an attenuation correction the reconstruction process.
must be performed to compensate for this effect. Good evidence
indicates that studies that have not traditionally been attenua- Display of Emission Tomographic Data
tion corrected, such as myocardial perfusion imaging, benefit A particular advantage of gamma camera rotational SPECT is
from proper attenuation correction. Two fundamentally differ- that a volume of image data is collected simultaneously. PET data
ent approaches are used for attenuation correction: analytic may be acquired in several steps, but the resultant reconstructed
methods and those that incorporate transmission data into the data are also a volume. The pixel size for SPECT is the same in
process. Both are designed to create an image attenuation cor- the three axes; for PET, the axial sampling might be slightly dif-
rection matrix, in which the value of each pixel represents the ferent from that in the transverse plane. However, in either case,
correction factor that should be applied to the acquired data. once the transaxial tomographic volume is reconstructed, it eas-
Some approaches are applied during reconstruction, whereas ily can be resorted into other orthogonal planes. Thus the sagittal
others are applied after reconstruction to the resultant images. and coronal images can be directly generated from the recon-
For portions of the body consisting almost entirely of soft structed volume represented by the set of transaxial slices.
tissue, an assumption of near-uniform attenuation can be made, The data can be reformatted into planes oblique to the origi-
and an analytic or mathematical approach such as the Chang nal transverse planes. This is particularly useful in cardiac imag-
algorithm can be used. The Chang algorithm is a postrecon- ing, in which the long axis of the heart does not coincide with
struction approach. After the object is reconstructed, an outline any of the three major axes of the reconstructed data. It is desir-
of the body part is defined on the computer for each tomo- able to reorient the data to obtain images that are perpendicular
graphic slice. From this outline, the depth, and therefore the and parallel to the long axis of the left ventricle, which can be
appropriate correction factor, for each pixel location inside the readily accomplished from the original volume data set. The
outline can be computed. A correction matrix is generated, and computer operator defines the geometry of the long axis of the
a multiplicative correction is applied on a pixel-by-pixel basis. heart, and the data are reformatted to create cardiac long-axis
The linear attenuation coefficient for technetium-99m (Tc-99m) and short-axis planes oblique to the transaxial slices (Fig. 3.7).
in soft tissue is 0.15/cm. This applies only to “good” geometry— The optimum angulation is highly variable across patients.
that is, a point source with no scatter. Thus a value for Tc-99m of Another useful strategy is to view tomographic data as a
approximately 0.12/cm is often used to compensate for scatter. sequence of planar images from different viewing angles in
At a depth of 7 cm in a liver SPECT study, almost 60% of the closed-loop cine. In the early days of SPECT imaging, this was
corresponding activity is attenuated. The observed count value accomplished by viewing the closed-loop cine of the raw projec-
would have to be multiplied by a factor of 2.5 (0.4 × 2.5 = 1) to tion data. This is still done in many cardiac imaging software
Chapter 3 Single-Photon Emission Computed Tomography 33

Fig. 3.7 Cardiac single-photon emission computed tomography (SPECT) images reformat data into multiple
planes. The top two rows are short-axis views obtained perpendicular to the long axis of the left ventricle. The
middle two rows are horizontal long-axis images, and the bottom two rows are vertical long-axis images. The
patient has a large fixed perfusion defect involving the inferior wall of the left ventricle. The ability to reformat
the data allows for more precise and accurate localization of abnormalities.

packages for quality control. However, these data tend to be The most common device used for SPECT is the rotating
noisy, making it difficult to view small variations in intensity. gamma camera, which consists of one or more gamma camera
Currently, a common approach is to reproject the transaxial heads mounted onto a special rotating gantry. Nearly all
images to generate a series of planar images that have the benefit gamma cameras marketed today incorporate SPECT capabil-
of greatly reduced noise. The reprojection method often used is ity. Early systems used a single gamma camera head, whereas
the maximum-intensity projection scan (MIPS), created by modern systems more commonly have two detector heads.
reprojecting the hottest point along each particular ray for any Dual-head systems that allow flexibility in configuration
given projection. These MIPS images emphasize areas of between the heads are very popular. For body imaging, the
increased accumulation of radioactivity while providing an heads are typically arrayed parallel to each other; for cardiac
overall impression of the area of increased radioactivity in rela- imaging, they are often placed at right angles (Fig. 3.8). Some
tion to the normal structures in individual tomographic slices. cameras are permanently configured in the 90-degree position
In some cases the MIPS images are distance weighted to make for dedicated cardiac imaging. Multiple heads are desirable
activity that is farther from the viewer appear less intense, because they allow more data to be collected in a given period.
thereby enhancing the 3D effect. Rotational SPECT is photon poor compared with x-ray CT,
and thus SPECT imaging protocols commonly take 10 to 30
Single Photon Emission Computed Tomography minutes for the acquisition of a data set. Therefore it is desir-
SPECT allows true 3D image acquisition, reconstruction, and able to obtain as many counts as possible while completing the
display of the radiopharmaceuticals routinely used in conven- imaging within a reasonable time to limit the effects of patient
tional nuclear medicine. Over the past 30 years, SPECT has motion and to minimize pharmacokinetic changes during the
developed, particularly in the field of nuclear cardiology, to the imaging time. Rotational SPECT has highlighted the need to
point at which SPECT has become the standard imaging method. improve every aspect of gamma camera system performance.
In SPECT, a series of projection images is acquired about the Flood field nonuniformities are translated as major artifacts in
patient. In most cases, these projection images are acquired by tomographic images because they distort the data obtained
rotating the imaging device about the object, but in other cases from each view or projection. Desirable planar characteristics
they may be acquired by viewing the object with multiple devices of a camera to be used for SPECT are an intrinsic spatial reso-
or through multiple pinhole apertures. These projection data are lution of 3.5 mm (as estimated by the full width at half maxi-
then reconstructed as described in the previous section, leading mum [FWHM]), linearity distortion of 1 mm or less, and
to the generation of a series of slices through the object. corrected integral uniformity within 3%. All contemporary
34 PART 1 Basic Principles

Two camera heads Two camera


180 degrees apart heads
(parallel) 90 degrees apart
(perpendicular)

Gantry

Fig. 3.8 Two configurations for a dual-detector single-photon emission computed tomography (SPECT)
­system.

rotational SPECT systems have online energy and uniformity


BOX 3.1 Image-Acquisition Issues
correction, as described in Chapter 2.
for Single-Photon Emission Computed
Recently, dedicated SPECT systems have been developed for
Tomography
cardiac imaging only. These cameras may use Anger logic for
event positioning; however, they are distinctly different in that Collimator selection
they are not large, single-crystal detectors as are found in the Orbit
Matrix size
traditional gamma camera, and many use solid-state detectors
Angular increment: number of views
of cadmium zinc telluride (CZT) rather than NaI scintillating 180- vs. 360-degree rotation
material. These detectors often use a pixelated design with Time per view
detector elements of approximately 2 × 2 mm. Because of their Total examination time
multicrystal design, the scintillation-based systems often use
either position-sensitive photomultiplier tubes or photodiodes
for light detection. The systems that use CZT have higher intrin- general-purpose collimators, even with fewer counts. The
sic efficiency and enhanced energy resolution (6% at 140 keV use of multihead SPECT systems allows the operator to gain
compared with 9%-11% compared with NaI). This allows for back some of the counts lost when using high-resolution col-
the reduction of Compton scatter in the images and may also limators by longer acquisition at each step or projection
enhance the ability to perform dual-isotope acquisitions (e.g., angle.
Tc-99m and iodine-123). Finally, the detectors in these systems In addition to the parallel-hole collimators routinely used for
have physical design characteristics that improve sensitivity. For planar and SPECT imaging, there are special focused collimator
instance, multiple detectors or pinhole apertures may be view- options specifically designed for SPECT imaging of the brain
ing the heart simultaneously. These improvements in sensitivity and the heart. These typically are a type of converging collima-
can be used to shorten the acquisition time or lower the quan- tor that permits more of the camera crystal to be used for radia-
tity of injected radioactivity and thereby lower the patient’s radi- tion detection. These collimators cause magnification of the
ation dose. Each system has different design characteristics, object and an increase in sensitivity proportional to the level of
acquisition procedures, and quality control methods. Although magnification. Thus given a parallel-hole collimator and a
these devices are promising, their use remains quite limited; focused collimator with the same spatial resolution, the focused
therefore the rest of this section focuses on the rotating collimator will have an improvement in sensitivity compared
camera. with the parallel-hole collimator. The use of focused collimators
Box 3.1 summarizes factors that must be considered in per- results in a geometrical distortion that must be accounted for in
forming SPECT with a rotating gamma camera. In addition to the reconstruction.
the calibrations described earlier and standard gamma camera The orbit selected (circular or noncircular) depends on the
quality control, careful attention to each of these factors will organ of interest (Fig. 3.9). Almost all systems today offer both
result in the high-quality SPECT images. circular and noncircular orbits. The ideal orbit keeps the detec-
Although collimator selection is generally limited to those tor as close to the object of interest as possible during the acqui-
supplied by the manufacturer, the specific choice depends on sition because the best resolution is at the face of the collimator
the clinical imaging task at hand. For a given septal thickness for parallel-hole collimators. For imaging the trunk of the body,
and hole diameter, collimators with longer channels provide most cameras use a noncircular orbit for this reason. Both cir-
better resolution but at a cost of lower sensitivity. However, cular and noncircular orbits may be used for imaging the brain
even though SPECT is relatively photon poor, collimator depending on whether the operator is able to position the detec-
selection should favor high resolution over high sensitivity tors to clear the shoulders. When using special focused collima-
when possible because high-resolution collimators provide tors, the orbit is often determined automatically by the system
improved image quality compared with high-sensitivity or that keeps the organ of interest in the focused area.
Another random document with
no related content on Scribd:
possession of a Negro, called John Conny. Ships
constantly stop here to water, as the most convenient
Place for taking in any large Quantity, and pay, each
Ship, an Ounce of Gold for the Privilege.
W. Thence, and anchored the 15th at Dixcove, an 14.
English Factory. This, Succonda, Anamaboo, and
others, tho’ called Factories, are the Residence only of
two or three People from the principal one at Cape
Corso, who have Commission over and above their
Salary, for what Trade they transact.
W. Dixcove, and anchored next Day before Cape Corso 16.
Castle, our African Company’s chief Fort; the
Residence of their Governor, stiled Director General;
two Merchants, a Secretary, Chaplain, Surgeon,
Factors, Writers, Artificers, and a Company of
Soldiers; with Buildings and Conveniencies inside, for
themselves or Slaves.
A. Anamaboo (just below it) a noted Place of stopping, June
for all our Windward trading Ships, to compleat their 26.
Slave Cargoes.
A. and W. Montford; the 30th, Barkee, and then Shallo. 28.
Through the whole from Sierraleon, it may be
observed, that Wood, Candles, or any other Ships
Necessaries are hard to get; the former, not from a
Want in the Country (it being over-run) but an
impassable Beach, where there is no navigable River,
and the Diffidence of the Negroes, where it might be
best supplied; and the other, because Merchant-Ships
do not expect a Trade of that sort, and therefore
unprovided.
A. Whydah. The whole Coast runs in a strait Line July 4.
(without Gulphs or Bays) is thick set with Trees, a
Tendency of the Sea with the Wind, and every where a
very rough and turbulent Beach.
W. Whydah, and arrived the 28th at the Island of 20.
Princes, belonging to the Portuguese. In our
Approach, saw every Day abundance of Whale,
Thresher, and Petrel.——Cleaned our Ships, heaving
down by one another, but became exceeding sickly by
the Fatigue, each burying three and four Men a Day,
for six Weeks together.
W Princes, both having purchased their Anchors with Sep.
difficulty. 20.

A St. Thomas’s, another Portuguese Island (the 28.


principal of three) on this Coast, abounding with fresh
Provisions, especially Hogs and Fowls, exchanged like
other Places of Poverty, at very easy Rates.
W Hence, and stretched with our Starboard Tacks to Oct. 5.
the Westward, designing to reach as far to Windward
as possible, that if any Pyrates should be on the
Coast, we might have them under our Lee. The 20th
we fell in with Cape Apollonia, went from thence the
23d, and anchored at Axim.
W Axim, and came to Cape 3 Points, where neglecting 24.
to pay John Conny his Duties for Water, he panyarr’d
some of our Men, till satisfied.
Left Cape 3 Points, and arrived next Day in Cape Corso 30.
Road again; it being every where confirmed to us in this
Return down, that the Pyrates in August last (the time we were at
Princes) had committed great Ravages upon the Merchant Ships.
W. Cape Corso Road, leaving the Weymouth, (now too Nov.
disabled to weigh her Anchor,) and plying to 10.
Windward, fell in with Succonda the 15th, repeating
our Visits in a Month’s Cruise, to Dixcove, cquedah,
Cape 3 Points, Axim, Cape Apollonia, Assinee,
Bassam, Jaque a Jaques, &c. our Purpose in it being
to secure Trade, air a sickly Ship, be in the way of
Intelligence, and impress Men from the Merchant-
Ships. Many prevented this indeed, by escaping to us
themselves from ill Treatment (they said) bad or short
Diet; but then, as more again on the same Pretence
took on with the Pyrates, it shews Caprice and
Humour to be the principal Point that determines
Seamen to this or that Service.
A De Elmina, the Dutch African Company’s principal 1721/2
Fort, of great Trade, there being seldom less than 5
Jan. 6.
or 6 Sail of Dutch Ships in the Road, often more.
A. Cape Corso Road, and left it the 10th in pursuit of the 7.
Pyrates; the Governor here, having received two or three
Expresses, that they had chased and taken a Ship nigh Axim, a
Place we had just come from.
A Apong to Leeward, not following too fast lest we over- 11.
shot them, but after certain Intelligence that the
Rogues had passed this Road (off at Sea) we
followed.
A Accra, a considerable trading Place, (for Salt 12.
particularly) and where the Dane, the Dutch, and
English, have a Castle.
A Whydah, and learned that the Pyrates had plundered Jan 15.
and ransomed 11 Sail of Ships, and left the Place two
Days before, on the Report of our following them.
W. Thence, and followed the Pursuit, coming before the 19.
Isle of Princes the 29th, and found the Portuguese
Strangers to the News.
A. The Mouth of the River Gabone, a snug Harbor we Feb. 1.
thought, for their Reception, the Navigation being
difficult; but finding by our Boats we had missed them,
left it the 3d, and continued our Search to Cape
Lopez.
Made the Cape, and soon after discovered the three
Pyrate Ships at Anchor in that Bay. One of them upon
the Heel, righted at sight of us, slipped her Cable and
chased, bending some of her Sails as she came out,
by which we judged the Rashness of our Enemy, who
fell a Prize to us before Night.
Recovered the Cape again, and found the Prize’s 10.
Consorts (according to expectation) very easy in the
Bay, and stayed so long that we doubted whether they
would stir for us; but at length, as their Eyes cleared in
our nearer Advance, all mad and frightned, they cut
their Cable, set their Sails, up went the black Flag,
and down their Courage; they continued a running
Fight, while only our chace Guns could play upon
them, and struck presently when our Broadside
reached, without the least Damage done to us.
A. Cape Lopez Bay, seizing there the third Pyrate Ship, Feb.
that had been deserted for a better Escape or 12.
Defence in the other.
W. Thence, having wooded and watered, bound with 18.
our Prizes and Prisoners to Cape Corso; the General,
and chief Merchants there, being in the Commission,
(brought out of England with us) for the Tryal of them.
Stopped at Princes, from the 21st to the 24th.
A. Cape Corso Road; the Pyrates in this Passage were Mar.
very troublesome to us, from a Project or two they 15.
had formed for their Deliverance, and hoped by the
Weakness of our Ship’s Company, would have
succeeded.
W. Cape Corso, the General’s Daughter of the Coast 1722
taking a Passage with us to England, a fair, flaxen- May 1.
hair’d, young Lady, tho’ born of a Mulatto.
I shall here observe at leaving the Center, that in respect
to Trade, Guinea needs only this threefold Division,
viz. the Gold, the Ivory, and the Slave-Coast; all to
Windward of this, might be called the one, and all to
Leeward the other; not because either of these Parts
of Trade would be entirely wanted in such respective
Division, but each abounds more under that
Denomination.
A Whydah, and left it the 5th, arriving at Cape Lopez May 3.
the 26th, where both Ships wooded, watered, and
purchased Wax for making Candles, now exceeding
scarce; and is the most convenient Place for Ships of
War, at leaving the Country.
W. Cape Lopez, and after a few Days at Sea, by foggy June 5.
Weather lost Company with our Consort the Swallow.
Made Cape Augustine in Brasil, a Portuguese Colony, July 1.
and anchored the 4th in Pernambuca Road, the next
great Port of Trade in this Province, to Bahia.
W Brasil, having found the Trade-Winds blow home, 12.
and increased in their Strength to this Continent,
bringing a dangerous Swell into the Road.
A Barbados, took in a Supply of Rum and Provisions, August
and left it the 9th. 3.

A Port-Royal in Jamaica, where we found the Swallow 23.


had arrived, a Week before.
A Hurricane[39] that drove the Prize ashore, blew away Aug.
all our Masts, with other Damages that detained us 28.
here 6 Months to repair.
W Port-Royal, and anchored at the Kays. 1722/3
W. The Kays, bound for England. Jan. 1.
A Donna Maria Bay in Hispaniola (the Windward Feb. 7.
Passage) to water, &c. 19.
W. Thence, and arrived at Spithead, April 8th, whence 22.
we were ordered to Woolwich, and paid off May 11th,
1723.

F I N I S .
E R R ATA .

P. line
32 19 for to r. too.
67 22 for he r. they.
72 27 After r. Aft.
75 24 and will be paid not only &c.
115 4 Ch. x.
115 5 Ezion
125 16 r. some other Parts.
FOOTNOTES:
[1] Put metaphorically for a Ship’s Cockpit; and answers to the
dark Cellaring of a House.
[2] At the Changes of the Moon appears a Pillar of Fire in the
North, which darting its Rays every way, moves from Place to
Place, enlightning not only Greenland, but Iceland and Norway;
and sometimes further, till the returning Sun obscures it. (Harris,
p. 635. Vol. II.)
[3] Finis Terræ, the Westermost Extremity of Europe, and formerly
thought a Ne plus ultra.
[4]
Sold 2 half-worn Suits for a Pipe of Wine.
3 Second-hand Wigs — — Ditto.
Loaf-Sugar sells 20d. per lib.
Cheshire-Cheese, 8.
Bisket, 2.
Beef per piece 10.
Bought Citron at 15d.
Lemons per C. 20.
[5] Vide Harris’s Voyages.
[6] Ten-pounders are like Mullets, but full of small Bones, like
Herring-bones.
[7] Old-wives; a scaly, flat Fish, half as thick as long, called so
from some Resemblance the Face is fancied to have, with that of
a Nun’s.
[8] Cavalloes; a bright, silver-colour’d Fish, with a prickly Ridge on
each side, half its length.
[9] Barricudoes; a well-tasted Fish, one Foot and an half long, not
wholesome if the Roof of the Mouth be black.
[10] Sucking-Fish; something like the Dog-Fish; underneath he
has an oval Flat, of three Inches and an half over, granulated like
a Nutmeg-grater; with this he sticks so fast, as difficultly to be torn
from the Deck. He often infests the Shirk, sticks fast, and sucks
his Nourishment from him.
[11] Cat-Fish, so called from four slender Fibres like Whiskers,
sprouting from the under part of his Mouth.
[12] Lollas, are Places cleared of Wood, but barren; the
Habitations only of Bug a bugs, the Species of an Ant; build not
above a Foot and half high; are whitish, smaller than the common
sort, sting, and devour Cloaths.
[13] Lugars; open, clear Places, sowed with Rice, &c.
[14] Plantanes and Bananoes are a very common Fruit, shaped
like Cucumbers, but slender and longer; peeled of their Coat, they
are roasted and eat as Bread, fried, or eaten raw. The latter is the
juicier, and of a preferable Taste. The Plant bearing them grows
as high as a Cherry-tree, with a Leaf three Yards long, and one
over; an admirable Detergent in foul, sanious Ulcers, stripped of
the inner Skin, and applied as you do Housleek in Corns.
[15] The Pine-Apple is their Prince of Fruits; does not grow so
high, but about the Bigness of a Pæony; a beautiful green and
yellow; firm and juicy as a Melon; eaten with Wine and Sugar.
Some of strong Fancy, imagine all sorts of Fruit to be tasted in it;
to me, it always left a stinging abstergent Flavour.
[16] Lime-trees, about as big as our Apple, arise by several Roots,
and have an oval Leaf; the Fruit smaller, but of sharper Scent and
Flavour than Lemons. In the Woods also are many Sevil-Orange
Trees, the Fruit largest and best tasted of any I ever met.
[17] Papais, the Size of a moderate Melon, green as that, and full
of Seeds, which thrown out, and the outside pared, is used with
Meat, buttered and salted. They grow 20 or 30 Foot high. Bosman
says, Male and Female (the Alcoran, that all Fruits grow so, p.
213.) the Male blossoming, but bearing no Fruit.
[18] Rice is sown in swampy Grounds; grows the height of our
Wheat, and from the top of the Stems shoot very slender Stalks,
bearing the Rice grained one above another to a vast Increase; a
Peck yielding above 40 Bushels: Yet such is their Idleness, there
is often a Deficiency supplied from Sherbro, &c.
[19] The Civet is about as large as a Ram Cat, comes from about
Sherbro; it’s Head like a Foxes. The Male only affords this, at the
rate of 3 or 4 Grains a day, gathered with a Quill out of a little Cod
or Hole, near the Intestin. rectum.
[20] General Phips at Cape Corso, was so nettled at this (he
receiving but 19 for 21) that it took his Stomach off Victuals two or
three Days.
For as in Fight the Gun or Drum
Will make the Warriour’s Stomach come;
So eke in Play; if two miss Fire,
The Stomach palls with wax’ning Ire.
[21] The Word Fetish is used in a double Signification among the
Negroes: It is applied to Dress and Ornament, and to something
reverenced as a Deity (a Lake, a Stone, a Tree, &c.) both so far
agree, as to be regarded as a Charm. That by a Peculiarity, and
this by some inherent Essence, can attract Good, or divert Evil.
Here they sometimes hide the Fetish in secret parts of the Woods;
on urgent Occasions make a sort of Appeal to them, separating
some the Friday, some the Saturday, and keep within doors the
whole day, in a Moaning, or what you may call a Devotion to it.
[22] Salaries 80l. per Ann.
[23] Boiled by the Negroes to the bigness of half-penny Rolls, and
an Accy purchases nine a day of them for a Month. The English
bake it.
A lean Goat you may get by chance for five Accys; a Muscovy
Duck, a Parrot, or couple of Chickens, for one.
[24] Miscell. Curiosa. Vol. iii. has a Journal of the Weather at
Cape Corso for 12 Months, from Mr. Hillier, who says, that was a
Year of the most Rain that could be remembred.
[25] Tittwees, like a large Wolf or Mastive, very fierce, and rob
their Towns in the Night, of what Kid or Poultry they find.
Tigers, not so adventurous, but are seen by them sometimes:
There are two now in the Castle.
Serpents. I have heard the Gentlemen of the Factory say, they
have been seen here 30 foot long, able to swallow a Child whole;
(Bosman says, a Man, or a full-grown Deer.)
Deer. Those whose Feet are tipped, and used as Tobacco-
Stoppers, are the bigness of a large Cat. The General had one in
his Kitchen, the Feet as thick as the middle Finger; whence I
judge, those very slender ones we see, are the Abortives of this
Animal.
[26] These sort of Tryals have much the same View with the
Water of Jealousy among the Jews, or Ordeal with our Saxon
Ancestors, that is, a Tryal by Fire or Water: The former was
proving their Innocency by walking on hot Plough-Shears un-hurt:
The latter was used hot or cold. They run their Arm into it scalding
hot; or the Priest gave an Imprecation to a Draught of Holy-Water.
The Person swore to his Innocence, and being tied Hands and
Feet, was thrown into a River or Pond; if he sunk, he was
adjudged innocent, if he floated, guilty: And these ways continued
till K. Hen. III.
Another way with the Saxons, was single Combat; if a Woman,
she appointed her Champion.
Another, since we are upon Tryal, was by two Ounces of Bread
and Cheese taken after the Communion, the Priest thus
imprecating; May it stick in your Throat, turn pale, your Limbs
convulsed, &c. if guilty; but if innocent, may you swallow it easily,
&c.
Rapin.
[27] Hæmac is a Brasil word, and signifies a Net slung to rest in;
made there from the Rind of a Tree.
[28] Milton. B. 10, & 11.
[29] A Negrish Name.
[30] See the Appendix to the Navy-Surgeon, in which are Physical
Observations on the Moisture and Density of the Air.
[31] There is a square Fort on the Larboard Point of the Bay, and
Anchorings about a League from it.
[32] Some pretend to have found what they call a material
Thunder-bolt; such a one is said to have fell on the Turkish
Mosque at Adrianople A. D. 1693; and such are shewn in the
Museums of Princes. At Copenhagen they have a large piece of
metallick Substance, said to be Thunder-bolt.
[33] A Word used by our Sailors, for the Grout is made of it.
[34] Moquissin is a name given to any thing they think has an
incomprehensible Virtue. V. Geographic. Atlas.
[35] The Portuguese, who trade hither from Erasil, chuse their
Cargoes all Boys and Girls, if they can, as more ductile for
Conversion; there being Fathers appointed to instruct them in
their Creed, and to baptize them, on their arrival; but then they are
Papists.
[36] Made of a peculiar Earth from Germany, and bear (those that
are good) the most intense heat.
[37]
There’s but the twinkling of a Star,
Between a Man of Peace and War.
Hud.
[38] At this Place I would observe, in relation to heaving the Lead,
that there is a Nisus in Bodies of Water from below upwards,
which makes ’em to sink neither so fast, nor so direct, at any
considerable Depth, as near the Surface; all at 200 Fathom or
less, being bottomless; i. e. unfathomable.
This Nisus, or resisting Motion to the Descent of Bodies, is not
only perceptible in the Lead, but more sensibly declares itself,
first, in that black or green Skim, seen sometimes on the Surface
(even smelling) after long Calms, the Product of some intestine
Motion.
2. That Divers, or any floating Bodies, emerge with greater Force
than they sunk.
3. Mr. Boyle’s 20th Experiment observes, that a glass Bubble let
open into the Receiver, on the Exsuction of the Air, the Water in it
manifestly rises a greater Height; consequently the Expansion
and Rarefaction of the Air by the Heat of the Sun, makes room for
this Spring in the Water, to exert itself; and therefore the Tides
themselves would more difficultly yield to the distant Attractions of
the Sun and Moon (I should think) without adding to that Theory
this conjoined Force, or natural Propensity of the Sea, to swell
before.
In respect to sinking the Lead, also may be added, a greater
Coldness, and a greater Saltness of the Sea, in proportion to the
Depths; (both which are very probable,) and will create a greater
Buoyancy, or Resistance to sinking, as will likewise the drawing
out a greater Quantity of Line, (less apt to demerge.) So that
although falling Bodies in Air, have their Velocities encreased, the
nearer they approach the Earth, yet contrarily in Water, it
diminishes with the Descent.
[39] Depend much on the preceding Season, (hot and dry
Weather) apt to raise greater Plenty of elastick Vapours on the
Terra firma, and will explode themselves now here, now there, as
the greater Rarefaction of Air (more towards one Island than
another) may invite.
B O O K S lately Printed for, and Sold by Cæsar Ward
and Richard Chandler, at the Ship, between the
Temple-Gates in Fleet-Street, and at their Shop in
Scarborough.
1. A Voyage to Guinea, Brasil, and the West-Indies, in His Majesty’s Ships the
Swallow and Weymouth. Describing the several Islands and Settlements, Madeira,
the Canaries, Cape de Verd, Sierraleon, Sesthos, Cape Apollonia, Cabo Corso,
and others on the Guinea Coast; Barbadoes, Jamaica, &c. in the West Indies. The
Colour, Diet, Languages, Habits, Manners, Customs and Religions of the
respective Natives, and Inhabitants. With Remarks on the Gold, Ivory, and Slave-
Trade; and on the Winds Tides and Currents of the several Coasts. By John
Atkins, Surgeon in the Royal Navy.
Illi Robur & Æs triplex
Circa Pectus erat qui fragilem truci
Commisit Pelago Ratem
Primus ——.
Horat.
2. An Enquiry into the Contents, Virtues, and Uses of cold and hot mineral
Springs, particularly the Waters of Scarborough; in a Letter to Robert Robinson of
Lincoln’s-Inn, Esq; Recorder of that Corporation. Price 1s.
3. The Christian Turk, or, the instructive and entertaining Adventures of Prince
Jakaya, Son to Mahomet III. Emperor of the Turks: Who lost his Succession to the
Ottoman Empire, by being privately bred up in the Christian Religion. The whole
intermix’d with several delightful Histories of the most illustrious Personages, and
containing the greatest Variety of Events ever yet published. In two Volumes
12mo. Price 4s. The Second Edition. Translated from the Original French.
4. The Life of Osman the Great, Emperor of the Turks, who first attempted the
Conquest of Poland, and the uniting that Kingdom to the Dominions of the
Ottoman Porte. Compiled from the Memoirs of three successive Ambassadors
from the King of France, to the Grand Signior, by Madam de Gomez; Author of La
Belle Assemblée. The whole comprizing the greatest Variety of extraordinary
Incidents and surprizing Revolutions that ever happen’d in so short a space of
Time, even in that fluctuating Empire. Faithfully Translated from the French, by
John Williams, Esq; in two neat Pocket Volumes. Price 4s.
5. The Vocal Miscellany; a Collection of above 800 celebrated Songs, many of
which were never before Printed; with the Names of the Tunes prefixed to each
Song, in two Pocket Volumes. Price Six Shillings.
N. B. Either Volume may be had alone, Price Three Shillings.
6. The Platonick Lovers: Consisting of original Letters in Prose and Verse, that
pass’d between an English Lady and an English Gentleman in France, (under the
borrow’d Names of Clio and Strephon;) who took an Affection to each other, by
reading accidentally their occasional Compositions. With a Critical Essay,
containing some Remarks upon the Nature of Epistolary and Elegiac Poetry; and
on the most beautiful Passages in these Epistles. By the ingenious Mr. John
Porter. The Third Edition corrected. In a neat Pocket Volume. Price 2s.
“But oh! encroaching Mortal as thou art,
Let still thy Spirit have the greatest Part;
You may admire me all the Ways you can;
Give me the Lover, but keep back the Man.”
Clio to Strephon, p. 79.
7. Poems on several Occasions, viz. The Progress of Musick. A Hymn to Sleep.
Mira and Colin, a Song. The Bee. The lost Muse. The Girdle. To Mira; with the
Miscellaneous Works of Mr. Pope. An Ode to Lycidas. The Candle. CORVUS a
very common Case. The Flea. To Fulvia singing. The Constant Shepherd. A
Supportable Misfortune. The Gift. Mira’s Picture. Cupid’s Reply. The Advice to
Mira. To Lycidas in the Country. A Paraphrase on some of the Odes of Anacreon,
being an Essay towards the Translation of that Poet. On Happiness. To Apollo,
between the Author and his Friend, in the Manner of Horace’s first Satire, second
Book. An Ode on his Majesty’s Birth-Day. The Plague of Wealth; or, the Poet’s
Diary: Occasion’d by the Author’s receiving 50l. as a Premium for the foregoing
Ode. By Matthew Pilkington, A. M. Revised and Corrected by the Rev. Dr. Swift.
Octavo. Price 3s. 6d.
8. Letters: Containing an Account of what seemed most remarkable in
travelling through Switzerland, Italy, and some Parts of Germany, &c. in the Years
1685 and 1686. Written by Dr. Gilbert Burnet, late Lord Bishop of Sarum. 8vo.
Price 4s. 6d.
9. C. Julius Cæsar’s Commentaries of his Wars in Gaul, and Civil War with
Pompey. To which is added, a Supplement to his Commentary of his Wars in Gaul;
as also Commentaries of the Alexandrian, African, and Spanish Wars; with the
Author’s Life. Adorned with Sculptures from the Designs of Palladio. Made English
from the original Latin, by Colonel Martin Bladen, one of the Lords Commissioners
of Trade and Plantations. 8vo. Price 6s.
10. The Adventures of Proteus, or Love in several Shapes, &c. Being a most
diverting Sett of Novels, never before published. By Mrs. Herbert. The Second
Edition. 8vo. Price 4s. 6d.
11. The Frauds of Romish Monks and Priests, set forth in eight Letters, lately
written by a Gentleman in his Journey to Italy, and published for the Benefit of the
Publick. The Fifth Edition in 2 Vol. 12mo. Pr. 5s.
12. Glossarium Antiquitatum Britannicarum, five Syllabus Etymologicus
Antiquitatum veteris Britanniæ atque Iberniæ, temporibus Romanorum. Auctore
Willielmo Baxter, Cornavio, Scholæ Merciarium Præfecto. Accedunt Viri Cl. D.
Edvardi Lhuidii, Cimeliarchæ Ashmol. Oxon. de Fluviorum, Montium, Urbium, &c.
in Britanniâ Nominibus, Adversaria posthuma. Editio Secunda. 8vo. Price 6s.
13. The English Pleader: Being a select Collection of various Precedents of
Declarations on Actions brought in the Courts of King’s-Bench and Common-Pleas
at Westminster, in Case, Debt, Covenant, Trespass, Trespass and Assault,
Ejectment, Replevin, Prohibition, &c. Taken from the Rolls in the Treasury of the
said Courts, and Forms settled by Counsel and Special Pleaders, since the
Commencement of the Act of Parliament for the Laws being in the English
Language. To which are added, The Forms of Pleas and Issues, both General and
Special, with Replications thereto: As also Judgments in both Courts on the
several Actions: And likewise Forms and Precedents of Recoveries and Concords
of Fines; with the Method of suffering and passing the same. By an Eminent
Attorney of Lincoln’s-Inn. 8vo. Price 5s.
14. The English Clerk’s Instructor in the Practice of the Courts of King’s-Bench
and Common-Pleas, viz. in filling up and suing out Writs of the first Process; in
drawing Declarations, Pleas, and Demurrers; making up Issues and Paper Books;
entering Judgments; suing out Writs of Execution; Writs of Error, Originals,
Outlawries, &c. and in passing of Fines; and several other Matters necessary for
the Practice of an Attorney. To which are added, several Choice Precedents of
Declarations never before printed. By Daniel Reading, Attorney at Law. In 2 Vol.
8vo. Pr. 10s.
15. A Guide to Justices: Being modern English Precedents, for the Direction of
Justices of the Peace, in making out Warrants, Mittimus’s, Recognizances,
Supersedeas’s, Affidavits, Informations, Inquisitions, Summons, Precepts,
Certificates for the Poor, and Warrants for their Removal, Bonds, &c. Necessary
for all Justices of the Peace, Deputy-Lieutenants, Commissioners of Sewers, &c.
to assist them in the Execution of their several Offices. Approved by, and published
at the Request of Sir John Dutton, Sir Robert Cann, Sir William Codrington, Barts
Robert Kendal, William Blathwaite, Giles Earle, John Stephens, John Temple,
Thomas Cooke, John Small, Esqrs; and others his Majesty; Justices of the Peace
for the County of Gloucester. By Joseph Higgs Gent. who has been Clerk to the
Commission of the Peace, from the Reign of King Charles II. to the present time.
8vo. Price 3s. 6d.
16. The Justice’s Case-Law: Or, A concise Abridgment of all the Cases of Crown
Law relating to Justices of Peace and their Business and Proceedings; with
References to the Reports, and other Law-Books. Digested in a new Alphabetical
Method, very necessary for Justices of the Peace, as a Direction how to proceed
in their Offices. By a Gentleman, who hath many Years filled the Chair of the
Quarter-Sessions. 8vo. Price 4s.
17. Lex Mercatoria: Or, The Merchant’s Companion. Containing all the Laws and
Statutes relating to Merchandize, wherein our Trade with foreign Nations, and
amongst ourselves, is concisely treated of, under the Heads of Merchants and
Owners of Ships, Masters, Mariners, Pilots, &c. with Precedents of Instruments
and Writings made use of in all Cases relating to Trade. The Second Edition,
corrected and enlarged. 8vo. Price 5s.
18. The Laws of England concerning the Game: Of Hunting, Hawking, Fishing,
and Fowling, &c. and of Forests, Chases, Parks, &c. And also concerning Setting-
Dogs, Greyhounds, Lurchers, Nets, Tunnels, Low bells, Guns, and all Manner of
Engines and Instruments mentioned in the several Statutes to destroy the Game:
Shewing who are qualified by Law to keep and use them; the Punishments of
those who keep them, not being qualified. The whole being a summary Collection
of all the Statute-Law concerning the Game; and of all the Cases, Resolutions, and
Judgments in the several Courts of Record at Westminster relating thereunto.
Together with two Precedents of Declarations; the one in an Action of Trespass
against an idle and dissolute Person for Hunting, &c. in which the Plaintiff (if he
recover) is to have full Costs; and the other in an Action of Debt upon the Statute
of 8 Geo. I. for a pecuniary Penalty, forfeited by that Law, in which the Plaintiff (if
he recover) shall have double Costs. Continued down to the present Time. By
William Nelson, of the Middle-Temple, Esq; The Second Edition. 12mo. Price 2s.
6d.
19. An exact and compleat Collection of all the Laws relating to Elections of
Members to serve in the House of Commons, from Magna Charta, down to the
Year 1734. Digested under the Heads of the Electors, the Elected, and the
Returning Officers. To which is added, An Alphabetical Index, referring to the
several Sections; whereby may be seen, at one View, the Care of the Legislature
to preserve the Freedom of Elections. Price 2s. bound.
20. The Navy Surgeon: Or, A Practical System of Surgery. Illustrated with
Observations on such remarkable Cases as have occurred to the Author’s Practice
in the Royal Navy. To which is added, A Treatise on the Venereal Disease, the
Causes, Symptoms, and Method of Cure by Mercury. An Enquiry into the Origin of
that Distemper; in which the Dispute between Dr. Dover, and Dr. Turner,
concerning Crude Mercury, is fully considered; with Useful Remarks thereon. Also
an Appendix, containing Physical Observations on the Heat, Moisture, and Density
of the Air on the Coast of Guinea; the Colour of the Natives; the Sicknesses which
they and the Europeans trading thither are subject to; with a Method of Cure.
Address’d to the Surgeons of the Navy Royal. By John Atkins, Surgeon in the
Royal Navy. 12mo. Pr. 3s.
21. Prosodia Chirurgica: Or, A Memoria Technica, calculated for the Use of old
Practitioners, as well as young Students in Surgery. Being a Lexicon, wherein all
the Terms of Art are accounted for, their most received Sense given, and an exact
Definition of them from the best Greek Authors: Also their Pronunciation, as to
Quantity, determined by proper Marks over each Syllable. The Second Edition.
Price 2s.
22. The Gardener’s Universal Calendar: Containing an Account of the several
Monthly Operations in the Kitchen-Garden, Flower-Garden, and Parterre,
throughout the Year; and also Experimental Directions for performing all Manner of
Works in Gardening, whether relating to Sowing, Planting, Pruning, Herbs,
Flowers, Shrubs, Trees, Ever-greens, &c. with the Products of each Month, in a
Method wholly new, taking in the whole Business of Gardening and Husbandry.
12mo. Price 2s.
23. The Compleat Fruit and Flower-Gardener: Containing Directions for raising
the most curious Flowers, and cultivating the choicest Fruits. By J. Cowell,
Gardener at Hoxton. To which are added, Two new and curious Essays concerning
the best Method of Pruning Fruit-Trees, sewing the great Disadvantage of the
Knife. II. A Discourse concerning the Improvement of the Potatoe, and the various
Ways of preparing and dressing it for the Table. By R. Bradley, F. R. S. The Third
Edition. Adorned with curious Figures. 8vo. Price 3s. 6d. bound.
24. A Practical Grammar of the English Tongue: Or, A Rational and easy
Introduction to Speaking and Writing English correctly and properly. Peculiarly
adapted to the Nature and Genius of the Language, and free from the hard and
unnecessary Terms of the Latin Rudiments. The whole treated of in expressive
Terms and familiar Style, and in the most natural and instructive Method, viz. That
of Question and Answer. Designed for the Use of Schools: And though calculated
for such as require only an English Education, may yet be a useful Foundation to
those who are designed for higher Studies. By William Loughton, Schoolmaster at
Kensington. Price 1s. 6d. or 15s. a Dozen, for the Use of Schools.
The ingenious Author of the Republick of Letters, for September 1734, has this
Character of Mr. Loughton’s Grammar. “The Author (says he) in his Preface, gives
his Reasons for compiling A Practical English Grammar, which, it must be own’d,
he has done in such a Manner, as to render it the best of the Kind: His Rules are
plain and easy, concise and clear; he has laid aside the old Terms, and made use
of such as are more expressive of the several Parts of Speech; his way of
Question and Answer renders it intelligible to young Capacities.—I heartily wish it
more generally introduc’d into our English Schools, since it would give all those
whose Education is confin’d to their Mother Tongue, an adequate Notion of it;
teach the Fair Sex to write more correctly; and expedite the Studies of such who
are design’d for the Learned Languages, by furnishing them with a proper Idea of
the several Parts of Grammar before their Entrance upon the Latin Rudiments.”
25. A General Dictionary, Historical and Critical, in which a new and accurate
Translation of that of the Celebrated Mr. Bayle, with the Corrections and
Observations printed in the late Edition at Paris, will be included, and interspers’d
with several thousand Lives, never before publish’d. The Whole containing the
History of the most Illustrious Persons of all Ages and Nations; particularly those of
Great-Britain and Ireland, distinguished by their Rank, Actions, Learning, and other
Accomplishments. With Reflections on those Passages of Mr. Bayle, which seem
to favour Scepticism and the Manichee System. By the Rev. Mr. Bernard, the Rev.
Mr. Birch, F. R. S. Mr. John Lockman, and other Hands; and the Articles relating to
Oriental History by Mr. George Sale, Editor of the late Edition of the Alcoran in 4to.
N. B. This Work is publish’d Monthly at 3s. per Number, containing 20 Sheets
each. Two Volumes are already finish’d, and the Remainer will be publishied in the
same Manner, till the Whole is completed.
Proposals for the said Work, with a Specimen annex’d, may be had of C. Ward
and R. Chandler, at the Ship between the Temple-Gates in Fleet-Street; and at
their Shop at Scarborough: At both which Places Subscriptions are taken in.
TRANSCRIBER’S NOTE
The numerical dates in the Sidenotes (on pages 255 to 265) have all been
italicized for consistency eg Mar. 10..
Except for those changes noted below, all misspellings in the text, and
inconsistent or archaic usage, have been retained.
Pg 35: ‘some hunred Leagues’ replaced by ‘some hundred Leagues’.
Pg 48: the footnote anchor [16] was missing, and has been added to ‘[16]Limes,
Oranges’.
Pg 68: ‘weighed Auchor’ replaced by ‘weighed Anchor’.
Pg 97: ‘our Rigth of’ replaced by ‘our Right of’.
Pg 97: ‘The Heigth of’ replaced by ‘The Height of’.
Pg 104: ‘ars sold’ replaced by ‘are sold’.
Pg 132: ‘a Change full as bad’ replaced by ‘a Charge full as bad’.
Pg 193: ‘joined also in in a’ replaced by ‘joined also in a’.
Pg 199: ‘ridiculous, scaramouch’ replaced by ‘ridiculous, and scaramouch’
(catchword ‘and’ was missing).
Pg 202: ‘we re- our’ replaced by ‘we repeated our’.
Pg 207: ‘off the 1sland’ replaced by ‘off the Island’.
Pg 209: ‘Food or Necessarieis’ replaced by ‘Food or Necessaries’.
Pg 243: ‘Bitts of 7d½’ replaced by ‘Bitts of 7½d’.
Pg 261: ‘1721/2’ inserted before ‘Jan 6.’ as a new Sidenote.
Pg 263: ‘1722’ inserted before ‘May 1.’ as a new Sidenote.
Pg 265: ‘1722/3’ inserted before ‘Jan 1.’ as a new Sidenote.
Catalog: ‘by furnish- them’ replaced by ‘by furnishing them’.

You might also like