You are on page 1of 227

BRF 90°/15°

RF TILTING TABLE

TECHNICAL MANUAL
Emission date: 28-11-06
Revision date: 30-05-19 [Rev.03] (R 5.0)

Technical Manual code: DBQ117-EN


BRF 90°/ 15°

THIS DOCUMENT WAS ORIGINALLY EDITED IN ITALIAN LANGUAGE

Page 2 of 227
Emission date: 28-11-06
Revision date: 30-05-19 [Rev.03] (R 5.0)

Technical Manual code: DBQ117-EN


BRF 90°/ 15°
WARNINGS

Given that tilting table BRF is a diagnostic medical device, its use is only considered
“appropriate” if performed by specialised medical or paramedic staff.
All users of BRF unit described herein are strongly recommended to read carefully all instruction
in this manual and to respect recommendations reported about safety before attempting any
operation of installation, service or use of these devices.
Installation, service and maintenance of all devices which are part of the system is reserved to
technical personnel, qualified and trained, with adequate experience and professional
background.
Even if the unit BRF unit has been designed and manufactured to ensure utmost safety during
use, installation and examination, a lack of respect of the indication herein reported can result in
a potential risk for user, patient and things.

The system configuration must not be changed during the lifetime of the device.

IONIZING RADIATION PROTECTION

An excess of X-Ray dose is a well known risk for health, and an improper use of Radio diagnostic
devices can result as well in a possible risk for health of patient and user.
Users of X-ray units, like those described in this manual, and technicians in charge of installation
and maintenance, are mandatory required to be informed on risk connected to exposure to an
excess of direct and diffracted X-rays and to keep safe themselves as well as patient and all
people eventually exposed to the same risk. A proper use of all standard means of protection is
needed in this purpose.
Specifically is mandatory that users and all personnel eventually required in the radiological
room will be informed on the International regulations for X-Ray protection and respect all
indication mentioned there.

SPECIFIC WARNINGS

All mechanical components and mobile parts which are included in the standard featuring of this
unit shall be used carefully and precisely, in respect of what described herein and according to
the indication of the present manual. A lack of respect of these indications may result in a
collision with patient, users, other operator or object present in the X-Ray room.
It is forbidden for all non authorized personnel to access to the internal part of the devices,
especially removing protective cores and carters: only skilled and qualified personnel can
remove these carters and attempt intervention to powered parts or other components for
ordinary and non ordinary maintenance.
It’s requested utmost attention and care while using mobile phones, sources of
electro-magnetic radiation and other potential sources of interferences to electrical
and electronic components. Given the medical device BRF is not in need of any specific
shielding to ensure safety, use of such devices like those here mentioned should be
prohibited in the immediate proximity of the medical device and, in every case, in the
X-Ray room.

Intellectual property rights

The content of the present manual, technical, planning and design, is an intellectual property of
BMI BIOMEDICAL INTERNATIONAL SRL.
Readers are required to contact directly BMI BIOMEDICAL INTERNATIONAL SRL and mention all
violation to this right, may it be in form of unauthorized copy, improper diffusion or other.

Page 3 of 227
Technical Manual code: DBQ117-EN

INDEX BRF 90°/ 15°


INDEX
WARNINGS_____________________________________________ (Page: 3)

I - DESCRIPTION OF THE MEDICAL DEVICE BLADE (Page: 8)

I.1 – UNIT BLOCK DESCRIPTION


I.2 – DEFINITIONS

II - COMBINATION AND RELATED CLINICAL APPLICATION (Page: 15)

III – TECHNICAL AND FUNCTIONAL FEATURE, DIMENSIONS ENCUMBRANCES AND


WEIGHTS (Page: 18)

IV – NORMS AND DECLARATION OF CONFORMITY (Page: 25)

V – SAFETY (Page: 27)

V.1 – DECLARATION OF RESPONSIBILITY


V.2 – PROTECTION CLASSES
V.3 – PROTECTION AGAINST COLLISION
V.4 – PROTECTION AGAINST IONIZING RADIATIONS
V.5 – EMC COMPATIBILITY
V.6 – CABLES AND WIRES

VI – UNPACKING (Page: 38)

VI.1 – UNPACKING PROCEDURE: SINGLE PACKAGE


VI.2 – UNPACKING PROCEDURE: DISMANTLE PACKAGE
VI.2.1 – DISMANTLE PACKAGE: BASE
VI.2.2 – DISMANTLE PACKAGE: COLUMN AND SFD

VII – INSTALLATION (Page: 45)

VII.1 - PRELIMINARY CHECK


VII.2 - ENVIRONMENTAL CONDITIONS
VII.3 - RADIOLOGICAL TABLE BLADE INSTALLATION
VII.4 – BASE INSTALLATION PROCEDURE
VII.5 – ATTACHMENT OF THE BASE TO THE FLOOR
VII.6 – COLUMN INSTALLATION PROCEDURE (ONLY FOR UNITS SHIPPED
DISASSEMBLED)
VII.7 – SPOT FILM DEVICE INSTALLATION PROCEDURE (ONLY FOR UNITS SHIPPED
DISASSEMBLED)
VII.8 – ELECTRICAL CONNECTION PROCEDURE (ONLY FOR UNITS SHIPPED
DISASSEMBLED)
VII.9 – COVER INSTALLATION PROCEDURE
VII.10 – REMOVE THE COUNTERWEIGHT BLOCK

VIII – INSTALLATION AND USE OF ACCESSORIES (Page: 73)

VIII.1 – FOOT SUPPORT


VIII.2 - LEADED X-RAY PROTECTIVE SCREENS
VIII.3 - PAIR OF HAND-GRIPS
VIII.4 - SHOULDER SUPPORT
VIII.5 – PRESSING BAND
VIII.6 – LATERAL CASSETTE SUPPORT
VIII.7 – PAIR OF LEG SUPPORTS

Page 4 of 227
Technical Manual code: DBQ117-EN

INDEX BRF 90°/ 15°


VIII.8 – FLUOROSCOPY PEDAL
VIII.9 - PATIENT SCREEN

IX – INSTALLATION AND CENTRING X-RAY COMPONENTS (Page: 85)

IX.1 – FITTING THE IMAGE INTENSIFIER


IX.2 – FITTING THE X-RAY TUBE ASSEMBLY AND THE ATTACHMENT FLANGE OF THE
COLLIMATOR
IX.3 – COLLIMATOR INSTALLATION PROCEDURE
IX.4 – CONNECTION AND POSITIONING OF HIGH VOLTAGE CABLES
IX.5 – CHECKING OF PERPENDICULARITY BETWEEN THE X-RAY BEAM AND THE
COLLIMATOR
IX.6 – VERIFICATION OF CENTRING OF THE COLLIMATOR AND OF THE FORMAT OF THE
IMAGING FIELD
IX.7 - VERIFICATION OF COMPRESSOR CORRECT WORKING

X – SYSTEM COMPONENTS (Page: 99)

X.1 - SPOT FILM DEVICE


X.2 – TABLE

XI - COMPONENTS POSITION (Page: 103)

XI.1 – BRAKES POSITION


XI.2 – MICROSWITCHES POSITION
XI.3 – MOTORS POSITION
XI.4 – POTENTIOMETERS POSITION
XI.5 – SUMMARY OF COMPONENTS POSITION
XI.6 – SFD COMPONENTS POSITION
XI.7 – COLUMN COMPONENTS POSITION

XII – FUSES (Page: 110)

XIII – INTERFACE SIGNALS (Page: 111)

XIII.1 - X-RAY HT GENERATOR


XIII.2 - IMAGE INTENSIFIER / TV CHAIN / COLLIMATOR
XIII.3 – VARIOUS
XIII.4 - POTTER (SUPPORT CONNECTIONS)

XIV – INVERTERS (Page: 112)

XIV.1 - INVERTERS DESCRIPTION


XIV.2 - INVERTER CONNECTION SIGNALS
XIV.3 - INVERTERS SETTING
XIV.4 – MOVING MOTORS BY INVERTER

XV – HOW TO PROGRAM ELECTRICAL BOARDS (Page: 118)

XV.1 - ELECTRICAL BOARDS POSITION


XV.2 – ACCESS TO TABLE CONTROL BOARD (PCB101)
XV.3 – ACCESS TO BASE KEYBOARD (PCB102)
XV.4 – ACCESS TO GENERATOR INTERFACE BOARD (PCB103)
XV.5 – ACCESS TO COLUMN CONTROL BOARD (PCB104)
XV.6 – ACCESS TO SFD KEYBOARD (PCB1)
XV.7 – ACCESS TO SFD CONTROL BOARD (PCB2)

Page 5 of 227
Technical Manual code: DBQ117-EN

INDEX BRF 90°/ 15°


XV.8 - ELECTRICAL BOARDS FIRMWARE DESCRIPTION
XV.9 - FIRMWARE INSTALLATION PROCEDURE

XVI – SETUP AND CHECK PROCEDURES (Page: 133)

XVI.1 – ACCESS TO SETUP MODE


XVI.2 – MENU AND SUBMENU LIST
XVI.3 – MENU AND SUBMENU NAVIGATION BUTTONS LIST
XVI.4 – SDF SETUP
XVI.4.1 – MOTOR CASSETTE setup menu
XVI.4.2 – MOTOR SHUTTER setup menu
XVI.4.3 – LANGUAGE setup menu
XVI.4.4 – EJECT CASSETTE setup menu
XVI.4.5 – IMAGE INTENSIFIER setup menu
XVI.4.6 – DEFAULT PARAMETERS setup menu
XVI.4.7 – EXIT to SFD setup men
XVI.5 – TABLE SETUP
XVI.5.1 – TILTING setup menu
XVI.5.2 – LONGITUDINAL PANEL setup menu
XVI.5.3 – TRANSVERSAL PANEL setup menu
XVI.5.4 – LONGITUDINAL SFD setup menu
XVI.5.5 – DEFAULT PARAMETERS setup menu
XVI.5.6 – EXIT to table setup menu
XVI.6 – COLUMN SETUP
XVI.6.1 – COLUMN UP/DOWN setup menu
XVI.6.2 – COLUMN POSITION SFD Z setup menu: SFD in/out
XVI.6.3 – COLUMN SPEED DOWN menu
XVI.6.4 – EXIT to column setup menu
XVI.7 – COLLIMATOR SETUP
XVI.7.1 – COLLIMATOR OPEN AND CLOSE setup
XVI.7.2 – COLLIMATOR POSITION TUBE RX setup
XVI.7.3 – COLLIMATOR EXPOSITION setup
XVI.7.4 – COLLIMATOR 300x300 MIN setup
XVI.7.5 – COLLIMATOR 300X300 MAX setup
XVI.7.6 – EXIT to collimator setup menu

XVII – CLEANING OPERATIONS AND ORDINARY MAINTENANCE (Page: 182)

XVIII – ORDINARY AND EXTRAORDINARY MAINTENANCE (Page: 183)

XIX – WARRANTY (Page: 185)

XX – LABELLING AND IDENTIFICATION OF PRODUCTS (Page: 187)

XXI – DISPOSAL (Page: 188)

XXII – SYMBOLS (Page: 189)

XXIII - REFERENCES AND MANUFACTURER ADDRESS (Page: 190)

XIV – CE MARK (Page: 191)

Page 6 of 227
Technical Manual code: DBQ117-EN

INDEX BRF 90°/ 15°


ANNEX A – BLOCK DIAGRAM (Page: 192)

ANNEX B – ELECTRICAL BOARDS (Page: 203)

B.1 - SFD KEYBOARD (PCB1)


– L AYOUT
– CONNECTIONS
B.2 - SFD CONTROL BOARD (PCB2)
– L AYOUT
– CONNECTIONS
B.3 - SFD-CASSETTE X & Y FORMAT BOARD (PCB3)
– L AYOUT
B.4 - SFD-JOYSTICK BOARD (PCB4)
– L AYOUT
– CONNECTIONS
B.5 - BASE CONTROL BOARD (PCB101)
– L AYOUT
– CONNECTIONS
– EMERGENCY SIGNALS
B.6 - BASE KEYBOARD (PCB102)
– L AYOUT
B.7 - INTERFACE BOARD (PCB103)
– L AYOUT
– LED DESCRIPTION
– CONNECTIONS
– ELECTRICAL SCHEMES
B.8 - COLUMN & SFD CONTROL BOARD (PCB104)
– L AYOUT
– CONNECTIONS

ANNEX C – INSTALLATION REPORT (Page: 225)

Page 7 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


I – Description of the medical device BRF 90°/15°

BRF is an X-ray unit that is combined with other devices to compose a general
radiological system for clinical diagnosis on patient. Typical application is intended
with patient in prone or orthostatic position.
The radiological table BRF is intended with use exclusively in a hospital
environment, only operated by technical skilled personnel.

Description:

BRF is a table for radio-fluorography examinations, with a tilting table moving


from the vertical position (+90°) to the Trendelenburg position (-15°), fitted with
an automatic Spot Film Device (SFD) positioned above the examination table. The
SFD had a receiver for an image intensifier (I.I.) with a nominal input field up to
30 cm diameter and a relative television system.
The automatic SDF is connected mechanically to the X-ray tube fitted below the
examination table, X-Ray tube had an automatic collimator which remains
permanently centred with the field filmed by the SFD itself and with the input
screen of the I.I. in all examination positions.
For radiographic filming, the tilting table could be supplied with:
• an additional potter-buck placed below the tabletop
• a second X-ray source placed to a suspended support (fixed to the ceiling).

BRF unit is equipped with a solid and strong frame, realized with extensive use of
iron sheet that not compromising the total weight of the unit.
The state of the art technology has been used for the electronic and
electromechanical design to ensure reliability, maximum performance and total
operating safety.

BRF 90/15 tilting table is easy to use, with mechanical, electrical & electronic
components needed to build a complete, analogical and/or digital radio-
fluorography system with a specific configuration.

Standard accessories for all BRF models:


• foot support
• two hand grips
• shoulder support
• leaded X-ray protections curtains

Optional accessories (only for “2 tubes version” – mod. AR4000/AR4002):


• pressing band
• lateral cassette support

Page 8 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


• leg supports (2x)
• fluoroscopy pedal
• patient screen

Pic. 1-1 shows the main components of BRF tilting table.

Support column spot film


device
Examination
Table (table top)
Main control Spot film
panel end
Spot film Foot end
device
Window aperture
Spot film device Potter
bucky
Compressor

Head end

Secondary
X-ray generator Emergency control panel
collimator stop

Pic. 1-1
Main components of the BRF tilting table

Page 9 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


Clinical applications field of the tilting table 90/15 BRF regards general contrast
graphic examinations and specifically for digestive system exams with analogical
and\or digital acquisition techniques (depending on the system configuration).
“2 X-ray tubes versions” of BRF unit is a universal system, suitable for use in
general radiographic examinations (chest and skeleton) with the patient in an
orthostatic, inclined or horizontal position.

I.1 - Unit blocks description

View from above


Spot Film Device
HEAD / FOOT

COLUMN
SPOT FILM
DEVICE Longitudinal (y)

Panel
SFD

IN
IN

CASSETTE
SFD OUT

Panel
OUT
Trasversal (x)
Panel Panel
(HEAD) (FOOT)

Page 10 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


Frontal view

SFD Head/Foot
Seriografo HEAD / FOOT

Device UP
Spot Film

COLUMN
SFD
Spot Film
Device

Tilt (UP)

Tilt (DOWN)
COLLIMATOR

BASE

Page 11 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


Lateral view

SFD IN / OUT

SFD UP
COLUMN

SFD

SFD DOWN
Panel IN Panel OUT

COLLIMATOR

Collimator
IN/OUT

BASE

Page 12 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


I.2 - Definitions

In this document, are used the following definitions:

- SFD: Spot Film Device, device able to load a X-Ray cassette in different
shape and perform the appropriate divisions.

- TABLE: Tilting device with Table Top.

- Operator: X-Ray technician using the Tilting Table

- Trasversal cassette side: the cassette side parallel to the SFD short
side (also defined as “Cassette X-axis”).

- Longitudinal cassette side: the cassette side parallel to the SFD long
side, on this side slides the carriage tray (also defined as “Cassette Y-
axis”).

- Cassette Movement: defined as “IN / OUT”.


1. IN: cassette moves from the loading side to the parking side.
2. OUT: cassette moves from the parking side to the loading side.

- Shutters Movement: defined as “OPEN / CLOSE / OVERLAP”.


1. OPEN: shutters move from the closing position to the
zero-position.
2. CLOSE: shutters move from zero-position to the closing position (
parallel division)
3. OVERLAP: shutters move from the full-closing position to the
overlap position to perform cross-divisions

- Longitudinal PANEL movement: defined as “HEAD / FOOT”:


1. HEAD: panel moves from low to high position.
2. FOOT: panel moves from high to low position.
All movements are activated by joystick and base keyboard.

- Trasversal PANEL movement: defined as “IN / OUT”.


1. IN: panel moves from operator side to the SFD-column.
2. OUT: panel moves from the SFD-column to operator side.
All movements are activated by joystick and base keyboard.

- Tilting movement: defined as “UP / DOWN”.


1. UP: defined as the table movement to the vertical position
(+90°).
2. DOWN: defined as the table movement to Trendelenburg position
(-15°).
All movements are activated by joystick and base keyboard.

Page 13 of 227
Technical Manual code: DBQ117-EN

I - DESCRIPTION OF MEDICAL DEVICE BRF 90°/ 15°


- SFD longitudinal movement: defined as “HEAD / FOOT”.
1. HEAD: the entire SFD moves to patient head side.
2. FOOT: the entire SFD moves to patient foot side.
All movements are activated by joystick.

- SFD vertical movement: defined as “UP / DOWN”.


1. UP: the entire SDF moves away from patient to ceiling.
2. DOWN: the entire SDF moves to floor and approaching patient
All movements are activated by joystick.

- SFD trasversal movement: defined as “IN / OUT”.


1. IN: SFD moves from operator side to SFD column.
2. OUT: SFD moves from SFD column to operator side.
All movements are activated manually by operator.

- COLLIMATOR movement: defined as “IN / OUT”.


1. IN: collimator moves from operator side to SFD column.
2. OUT: collimator moves from SFD column to operator side.
All movements are activated following the IN/OUT SFD position (tracking
function).

Page 14 of 227
Technical Manual code: DBQ117-EN

II - COMBINATION AND RELATED


CLINICAL APPLICATION
BRF 90°/ 15°
II – Combination and related clinical application

For functional purposes and applications, tilting table BRF must be incorporated
into a complete analogical and/or digital radio-fluorography system and
therefore must provide fitting for an X-ray generator tube assembly, an image
(intensifier with relative TV system and connection to an high voltage
generator) and to a digital image acquisition system where necessary.

In the “two-tubes” versions, the tilting table must also be combined with a
support hanging from the ceiling with relative X-ray generator and X-ray
collimator, where necessary, associated with additional imaging equipment.

In particular, for what regards the X-ray tube assembly (fitted below the table)
and the image intensifier (fitted on top of the automatic SFD), all those
produced by leading manufacturers with application features appropriate for the
foreseen tests can be used, in observance of the above specifications and of the
following mechanical compatibility (see Chapter “III – Technical and
functional feature, dimensions encumbrances and weights”).

The BRF unit can be combined with devices that may or may not be
manufactured by BMI BIOMEDICAL INTERNATIONAL itself. In case of
combination with a device not manufactured by BMI BIOMEDICAL
INTERNATIONAL, it is mandatory to contact BMI BIOMEDICAL INTERNATIONAL
itself to check compatibility with BRF. In any case, all devices used in
combination with BRF must be CE marked and conform to international
rules and laws for safety.

Warning
In the eventual combination of BMI BIOMEDICAL INTERNATIONAL SRL devices
with other manufacturer devices (main elements and accessories), please check
compatibility through these pages: where no mention of clear compatibility is
declared in this manual the user is required to check the conformity to the
mentioned norms and regulations and the eventual presence of non predictable
risks due to lack of compatibility between the devices.
It is mandatory to contact BMI BIOMEDICAL INTERNATIONAL SRL in this
connection, as well as the manufacturer of each device.

Here follows the list of devices for which BMI Biomedical International has
already checked and approved compatibility:

• All BMI BIOMEDICAL INTERNATIONAL Products.

• Collimator:
It is a standard supply of the unit, any different collimators are not
supported.

Page 15 of 227
Technical Manual code: DBQ117-EN

II - COMBINATION AND RELATED


CLINICAL APPLICATION
BRF 90°/ 15°
• Tube – Housing Complex:

All devices with followings features:

o Anode angle equal or above 15°


o Maximum housing diameter: 185 mm
o housing horn angles, in relation to exit axis : 90°
o hole-spacing ray window: 65 mm

65

Anode side view

Compatibility of any other devices require manufacturer approval.

• HT Generators:

Any R/F HV generator, subject to the manufacturer's authorization.

• Cassette holder Potter Bucky Type:

Cassette Holder/
Removable Vibrating
Description Supported Detector
Grid Grid
Removable FPD

BPB Potter Analogue films - FPD cassette


NO YES YES
Bucky size wi fi e CR Cassette

DGT850 YES NO NO FPD Rayence Mod. 1717SGC/SCC

DGT960 NO NO NO FPD Rayence Mod. 1717SGC/SCC


Analogue films - FPD cassette
DGT550 NO NO YES
size wi fi e CR Cassette
Analogue films - FPD cassette
DGT500 YES NO YES
size wi fi e CR Cassette
DGT900 YES NO NO FPD TOSHIBA FDX4343R-C

DGT960 NO NO NO FPD Rayence Mod. 1717SGC/SCC

Page 16 of 227
Technical Manual code: DBQ117-EN

II - COMBINATION AND RELATED


CLINICAL APPLICATION
BRF 90°/ 15°
• Ionization Chamber:
o Manufacturer: Claymount
Model “SSMC 601”;

Compatibility of any other devices are subjected to the manufacturer's


authorization.

• Image intensifier:
o Maximum dimensions 12”

Compatibility of any image intensifier are subjected to the manufacturer's


authorization.

• Antiscattering grid:
o Maximum dimensions 380 x 380 mm
o Focal distance cm 90

Page 17 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
III – Technical and functional feature, dimensions
encumbrances and weights

The Radiological table BRF is commercialized in 4 different configurations,


identified by 4 different codes:

CODE DESCRIPTION

Tilting table for radio-fluorography exams (2 X-ray tubes version, one of them on
AR4000 a suspended support): supplied with automatic SFD and image intensifier (I.I.),
table top which can be moved longitudinally and transversally and with an
additional potter-Bucky below the examination table.
Tilting table for radio-fluorography exams (1 X-ray tubes version – below the
AR4001 table): supplied with automatic SFD and image intensifier (I.I.), table top which
can be moved longitudinally and transversally
Tilting table for radio-fluorography exams (2 X-ray tubes version, one of them on
a suspended support): supplied with automatic SFD and image intensifier (I.I.),
AR4002 table top which can be moved only transversally and with an additional potter-
Bucky below the examination table.
Tilting table for radio-fluorography exams (1 X-ray version – below the table):
AR4003 supplied with automatic SFD and image intensifier (I.I.), table top which can be
moved only transversally
Tilting table +90°/-15°
Standard set of examination accessories

Page 18 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
TECHNICAL FEATURES

TABLE

- Height of table top when horizontal 880 mm

- Max patient weight 180 kg

- Structure tabletop POLIRAY laminate

- Absorption tabletop 0.6 mm Al eq. (100 kV, SEV 2.7


mmAl )

- Surface tabletop Flat

- Measurements tabletop 2100x770 mm

- Tilting motorized, from +90° to –15°

- Tilting speed continuously or variable

- Longitudinal movement motorized, 600 mm (head) / 400 mm


(foot)

(Models AR4000 and AR4001)

- Speed of long. movement continuously variable up to 35 mm/s

- Transversal movement motorized ± 100 mm

AUTOMATIC SPOT FILM DEVICE

- Technology microprocessors (16 bit)

- Window aperture structure PVC

- Absorption 0.5 mm Al eq. (100 kV, HVL 2.7 mm


Al)

- Compatible image intensifier max. input field: 16” – three fields

- Cassette loading from front side with self-centring


system

- Cassette format from 18x24 to 35x35 cm in both

Page 19 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
directions

- SFD divisions 4 transverse and 2 longitudinal with


possibility of cross divisions of 4-6–8.
Indication of possible divisions
depending on cassette format
identified by the system

- Radioscopy-radiography time depends on cassette format and spot


film division

- Rapid sequence available for all the spot film divisions

- Longitudinal movement motorized, 900 mm, at all inclinations


of the table, from vertical to
Trendelenburg

- Speed of long. movement 80 mm/s

- Transversal movement manual, 725 mm, blocked by ELM


brake

- Compression movement motorized, 3 different speeds


selectable, counterbalanced and
blocked by ELM brakes

- Compression range 300mm

- Distance window aperture-table maximum 500 mm, minimum 200


top surface mm

- Compressor pyramidal trunk shape, can be moved


onto the field with motorized
movement

- Compression force two available values: 3 kg and 5 kg

- SFD parking lateral right, manually controlled


movement of 725 mm, blocked by
ELM brakes

- Standard equipment grid oscillating, 36 l/cm, r 12, f.d. 90 cm

- Distance window aperture - film 56 mm

- Distance window aperture - I.I. 110 mm

Page 20 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
- Camera sensitive to AEC Optional

X-RAY TUBE - COLLIMATOR

- Movements motorized and synchronised with


longitudinal and transversal
movements of the SFD

- Distance focus-film from 690 mm to 990 mm (depends on


X-ray tube assembly type)

- Distance focus-I.I. variable from 744 mm to 1044 mm

- X-ray collimator motorized shutters

− Operative mode automatic, depending on the format


of the cassette and division of the
SFD, with compensation of the FFD,
and manual

− shape of radiated fields rectangular and square

− number of leaves 4 pairs (2 dedicated to extra-focal


radiation)

− material leaves Pb+Fe

− Filtration 0.3 mm Al eq.

POTTER BUCKY (Models AR4000 and AR4002)

- Movement manual, 1150mm at all inclinations

- Movement brake ELM

- Distance film–patient support Depends on potter bucky model

- Standard fitted grid Depends on potter bucky model

- Cassette format Depends on potter bucky model

Page 21 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
OVERALL MEASUREMENTS AND WEIGHT
(worst case)

- Height (with table vertical) 2800 mm

- Length (with table horizontal) 3100 mm

- Width 2000 mm

- Weight 1100 kg

(without intensifier or X-ray tube assembly unit)

ELECTRICAL POWER SUPPLY

- Voltage 230 Vac

- Frequency 50/60 Hz

- Power 2 kVA

Page 22 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
X-RAY COMPONENTS FEATURES

X-ray tube assembly

65

Anode side view

- max diameter tube : 185 mm


- housing horn angles, in relation to exit axis : 90°
- outlet terminals high voltage cables at 45°
- hole-spacing ray window : 65 mm

Image intensifier

- max. nominal input field: 30 cm


- max. weight: 50 kg

The warning stated below applies in all cases.

High voltage cables

- Due to the tube it could be necessary to use cables with 45° pipe in order
not to limit the run of the tube

Page 23 of 227
Technical Manual code: DBQ117-EN

III - TECHNICAL AND FUNCTIONAL


FEATURE
BRF 90°/ 15°
The below illustrations show the total measures and movements of the tilting
table 90/15 BRF, in the worst case of models available.

ALWAYS ADD TO 952mm THE


IMAGE INTENSIFIER HEIGHT

AGGIUNGERE SEMPRE ALLA


QUOTA DI 952mm L’ALTEZZA
DELL’INTENSIFICATORE DI
BRILLANZA

Page 24 of 227
Technical Manual code: DBQ117-EN

IV - NORMS AND DECLARATION OF


CONFORMITY
BRF 90°/ 15°
IV – Norms and declaration of conformity

BRF X-ray devices are designed and manufactured in conformity to the Dlg
46/97, that receipt the EU Directive 93/42/CEE, and further modification and
integration such as those provided by the Dlg 25 January 2010, n. 37, that
receipt the EU Directive 2007/47/CE. BRF devices have been classified as
Medical Devices in Class IIb.

Product classification according to Annex IX (Classification Criteria)

The medical device “BRF”, subject of this manual has been classified as active
medical device in Class IIb, according to rule 10, of 3rd paragraph that refer to
additional rules applicable to active medical devices, in annex IX of the Dlg
46/97, that receipt the EU Directive 93/42/CEE, and further modification and
integration such as those provided by the Dlg 25 January 2010, n. 37, that
receipt the EU Directive 2007/47/CE.
Hence the last comma of rule 10 states what follows “devices intended to emit
ionizing radiation and designed for diagnosis, radiotherapy or interventional
radiology, including those devices that controls them or directly influence their
performances are included in class IIb” . The device subject to this document,
has a direct influence on the result of radio diagnostic applications: as a matter
of fact, as it includes proper patient positioning means and acting as direct
influence in the imaging process, it falls into given definition.

Classification Rationale

Moreover the Spot film device or the digital detector has a dialogue with the
generator, influencing so the x-ray emission, which is not happening if not
approved.

The product has a life period of 10 years, shelf life depends on the use made out
of the product and on the level of maintenance and consumption.

NORM REFERENCES

Tilting table 90/15 BRF has been designed and manufactured in conformity with
the directive EC 93/42EEC, it is certified as a Medical Device class II b and as
such bears the symbol of conformity CE0051, complete with the code number of
the notified body appointed to issue such certification.
BRF tilting table complies, furthermore, with the Standards:

Page 25 of 227
Technical Manual code: DBQ117-EN

IV - NORMS AND DECLARATION OF


CONFORMITY
BRF 90°/ 15°
EN 60601-1:2007/A1:2012 Medical electrical equipment. Part 1: General requirements for basic
safety and essential performance
EN 60601-1-2:2015 Medical electrical equipment. Part 1: General requirements for safety 2
– Collateral Standard: Electromagnetic compatibility – Requirements
and tests.
EN 60601-1-3:2010 Medical electrical equipment. Part 1: General requirements for basic
safety and essential performance – Collateral Standard: Radiation
protection in diagnostic X-ray equipment.
EN 60601-1-6/A1:2016 Medical electrical equipment. Part 1: General requirements for basic
safety and essential performance – Collateral standard: Usability
EN 62366-1:2015 Medical devices. Application of usability engineering to medical devices
EN 60601-2-54:2009/A1:2015 Particular requirements for the basic safety and essential performance
of X-ray equipment for radiography and radioscopy
EN ISO 14971:2012 Medical devices – Application of risk management to medical devices
EN 62304:2006/A1:2015 Medical device software – Software life-cycle process
ISO 15223-1: 2012 Medical devices -- Symbols to be used with medical device labels,
labelling and information to be supplied -- Part 1: General
requirements
ISO 10993-1:2018 Biological evaluation of medical devices Evaluation and testing within a
risk management process.

Rule application
The X-Ray table BRF is CE Marked in conformity to what stated in annex II of
MDD 93/42, as described in art.11 and respond to all essential requirements in
annex I.

The Declaration of Conformity is available under request, at the archive of the


manufacturer.

Essential performances
The devices is designed and manufactured in such a way as to:
• Keeping of electrical and mechanical integrity in expected life
• Safety movements of the tilting table, meaning it cannot harm
accidentally the patient.
• Ability to maintain safety even in case of first failure (electric /
mechanical)
• Keeping electromagnetic immunity
• Suspension systems security

Page 26 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


V – Safety

The BRF units must be used only by qualified and duly trained personnel, in
respect of what specified in this manual about safety.

In all cases when the BRF units are combined and made working with devices
from other manufacturer, the compatibility of those devices must be specified in
this manual and the user is required to take care that no risk can be generated
to patients and users.

In these situations it is mandatory to take contact with the manufacturer and


with the manufacturers of all devices in the room or with qualified experts for
the needed controls of compatibility.

The X-Ray units BRF requires ordinary, periodic operation of cleaning and
maintenance, as specified in the appropriate section of this manual.

Moreover it is absolutely forbidden to attempt any kind of examination or


investigations that may result in a spread of liquids of any kind on the tabletop
of the BRF Table, in order to avoid damages to the electronic components.

Tilting table BRF must be used by qualified personnel in observance of the


specifications reported in this manual for what regards safety.
The necessary precautions must therefore be taken to prevent use of the
appliance by non-authorised and/or non-qualified personnel.

Before commencing any operation, the personnel appointed to clinical use of the
appliance must be informed about the safety devices and means of protection
described in this manual and fitted to the appliance, and on their correct use.
The technicians responsible of installation, ordinary and extraordinary
maintenance must be previously authorised by the manufacturer.
Any work on electrical components, cables, protective casings of the control
panels and live areas must be performed only by authorised technicians.

Tilting table BRF is not designed to operate in conditions where there is a


potential risk of explosion, such as in the presence of mixtures of anaesthetic
gases or other gases (oxygen, nitrogen monoxide…) or of inflammable agents,
such as those used to clean or disinfect skin or parts of the appliance itself.

In the case that tilting table BRF is used with appliances or equipment produced
by other manufacturers the compatibility of which has not been specified in this
manual, the operator is responsible for checking that no risks are created for
the patient and/or operator.
In these cases it is best to contact the manufacturers of the appliances and/or
accessories, or experts, to verify the situation as required.

BRF tilting table requires cleaning, disinfecting, ordinary and extraordinary


maintenance as specified below in this manual.

Page 27 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°

The operator must make sure that the system has been disconnected from
power supply before commencing any cleaning or disinfecting of the table.

V.1 – Declaration of responsibility

The manufacturer declines all responsibility for damages occurred to things or


human beings as well as for risky situation caused directly or indirectly by:
o Improper use of the system or use that differs in any way to what
described in this manual,
o Installation of the units performed in a different way than what stated
in this manual
o Maintenance and service operations are not performed in respect of
what stated in this manual

It is here reminded that manufacturer is responsible of the safety of its products


only if Installation, service and repair operations are performed by technically
skilled personnel authorized by manufacturer with a suitable training course.

The manufacturer declines any responsibility for damage to people and/or


things, or for dangerous situations caused directly or indirectly by:

- use of the unit by non-authorised and/or non-qualified personnel


- improper use of the unit not conforming to the instructions of this manual for
what concerns operating procedures and safety precautions
- technical intervention by non specialised and/or non authorised personnel
- installation and/or ordinary maintenance operations not carried out in
observance of the instructions contained in this manual
- mechanical and/or electrical modifications made to the appliance, including
during repair, not previously authorised in written form by the manufacturer

V.2 – Protection Classes

BRF units are so classified:

- Protection type against electric contacts …………………. Class 1


- Degrees of protection against electric contacts………. Type B

Protection against electrical discharges has been achieved through earth


connection on all parts and metal covers off the unit.

The X-Ray systems cannot operate at all, in every situation where a potential
risk of explosion is detected, as in presence of anaesthetic gases or flammable

Page 28 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


agents, such as for example those used for cleaning or disinfection of the skin or
of the parts of the units.
V.3 – Protection against collision

The BRF unit is not provided of safety devices that may impede an eventual
collision with any object and person present in the radio diagnostic room.

The movements of tilting table BRF have been designed so as to prevent


collision of any part of the appliance with the floor or ceiling, as well as any
collision between the various parts of the appliance.

By mean of this it is requested the utmost attention from the operator during
the use.

Moreover, during installation phase, it is necessary to provide an adequate area


around the x-ray unit, which is keep free form any obstacles to the movement
of the unit in every working modality. For more details, refers to Chapter “III –
Technical and functional feature, dimensions encumbrances and
weights”.

V.4 – Protection against ionizing radiations

It is full responsibility of the installation staff to ensure that the installation site
has adequate protection for the ionizing radiation. The user has mandate to
check that this obligation is fulfilled properly.

It is a full responsibility of the company or person that handle the installation to


verify that the rooms intended for the installation are adequately shielded
against ionizing radiation. The final user is mandatory committed to control the
shielding before attempting any use of the device.

It is full responsibility of the user to evaluate the effects which may result by an
exposure o x-ray radiation, direct and diffracted, and provide in order tot have
the exams conducted in respect of the safety radioprotection norms, especially
for what is up for:

- Careful limitation of the surface exposed to the X-Ray beam, in respect of


the anatomic area of interest

- Us of leaded protective shields, on patient and user, for any possible


application where this is not impeding the exam

- Use of stationary and mobile barriers and shields, which shall be given in
the installation.

Page 29 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


- Use of adequate leaded anti X clothes such as gloves an collars any time
the user is necessary aside of the patient during the examination. Please
note to always have care of being out of the primary field.

- Check the perfect closure of the doors for access and exit to the X-Ray
room. Always have care to leave in the x-ray room only the people strictly
necessary for the good result of the examination

- Check correct use and positioning of the personal dose measurer of the
personnel, according g to the indication of the radioprotection specialist.

It is responsibility of the client to install the necessary X-ray room shielding


where the radio-fluorography unit is located and to install the protective barriers
for the operators.
It is duty of the operator to assess the injurious effects which might be
produced by exposure to direct and diffuse X-rays and to make sure that
examinations are conducted in strict observance of safety regulations
concerning radioprotection, with especial regard to:
- effective limitation of the area struck by the primary X-ray to the relevant
anatomic area
- use of leaded rubber aprons whenever possible to protect anatomical parts
on the edges of the irradiated area
- use of fixed and mobile X-ray protection provided with the system and
individual appliance
- use of appropriate protective clothing against X-rays (e.g. leaded aprons)
whenever the examination requires the presence of the operator with the
patient, in any case positioning him/herself so as to keep out of the primary
X-ray beam
- perfect sealing of the access doors to the radio-diagnostic area before any
exposure, and expulsion of any personnel whose presence is not strictly
necessary to perform the examination
- correct positioning of the personal exposure meter as instructed by the
radioprotection chief.

Tilting table BRF has been constructed with adequate lead shielding of the parts
where the presence of leaking or diffused radiation might be present.
For adequate protection from X-rays during radiographic and radioscopic
exposure, the operator is advised to stay in the areas shown in Pic. 5.1 and 5.2,
and to use the leaded screens supplied with the tilting table, which have to be
attached to the sides of the SFD.
When the tabletop is in a horizontal position, or inclined from –15° to +45°, it is
recommended for the operator to remain, for safety’s sake, in zone “A” and to

Page 30 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


bring the leaded screens on the longitudinal side of the SFD along the side of
the table, as shown in Pic. 5-1
When the tabletop is inclined by more than +45°, it is recommended for the
operator to remain in zone “B” and to position the leaded screens on the
transversal side of the SFD towards the foot of the table in Pic. V-2.
Table 1 below shows the values measured in the areas occupied as above.
Note that the values measured conform to the limits established by the
provisions of IEC EN 60601–1–3 quoted earlier.

Position of table Area occupied CM from ground mGy without screens mGy with screens
20 0 0
40 0.162 0
60 0.513 0.061
80 0.503 0.035
0° A 100 1.129 0.052
120 0.764 0.043
140 0.721 0.026
160 0.156 0.087
180 0.017 0
20 0.318 0
40 0.320 0
60 0.045 0
80 0.044 0
90° B 100 0.094 0
120 0.031 0
140 0.072 0
160 0.046 0
180 0.025 0

Table 1 - Readings in the recommended areas occupied by the operator during use of the BRF
tilting table

Zone X-ray
“A” screens
60x60cm side

Pic. 5-1 - Recommended occupation area where operator are protected against X-rays, by
leaded screens, when table top are in horizontal position or inclined between –15° and +45°

Page 31 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


Area
“B”
60x60cm

X-ray screen
position
side

Pic. 5-2 - Area for occupation by the operator, recommended for the purposes of protecting
against X-rays, and position of the leaded screens, should the table top of the tilting table be in
an inclined position of > +45°

V.5 – EMC Compatibility

This electronic equipment complies with EMC 2014/30/EU Directive, defining the
maximum emission levels allowed for the electronic equipment and the required
immunity against electromagnetic field.
This, however, does not exclude that signals coming from transmitters, like
mobile phones or similar mobile radio transmitters, even if complying with EMC
standards, can affect the correct operation of the electro medical equipment if
used nearby and with high transmitting power. It is therefore necessary to avoid
using these radio transmitters near electronically controlled medical systems, to
exclude any interference.
Electronic unit complying with EMC standards has been designed to avoid any
risk of malfunctioning caused by electromagnetic interferences, when working in
standard conditions. However, radio signals coming from high-frequency
transmitters at a high transmitting power, if used near the electronic
equipment, may cause electromagnetic incompatibility. Any transmission with a
mobile radio transmitter must be avoided.
Mobile phones must be off in the area near the unit. These rules must be
complied with when the unit is operating.
This electronic equipment is suitable for professional healthcare facility
environment.
WARNING: USE OF THIS EQUIPMENT ADJACENT TO OR STACKED WITH OTHER
EQUIPMENT SHOULD BE AVOIDED BECAUSE IT COULD RESULT IN IMPROPER
OPERATION. IF SUCH USE IS NECESSARY, THIS EQUIPMENT AND THE OTHER

Page 32 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


EQUIPMENT SHOULD BE OBSERVED TO VERIFY THAT THEY ARE OPERATING NORMALLY.

WARNING: ALL CABLES AND ACCESSORIES THAT ARE REPLACEABLE MUST BE


REQUESTED TO THE MANUFACTURER.

WARNING: USE OF ACCESSORIES, TRANSDUCERS AND CABLES OTHER THAN THOSE


SPECIFIED OR PROVIDED BY THE MANUFACTURER OF THIS EQUIPMENT COULD RESULT
IN INCREASED ELECTROMAGNETIC EMISSIONS OR DECREASED ELECTROMAGNETIC
IMMUNITY OF THIS EQUIPMENT AND RESULT IN IMPROPER OPERATION.

WARNING: PORTABLE RF COMMUNICATIONS EQUIPMENT (INCLUDING


PERIPHERALS SUCH AS ANTENNA CABLES AND EXTERNAL ANTENNAS) SHOULD BE
USED NO CLOSER THAN 30 CM (12 INCHES) TO ANY PART OF THE DEVICE,
INCLUDING CABLES SPECIFIED BY THE MANUFACTURER. OTHERWISE, DEGRADATION
OF THE PERFORMANCE OF THIS EQUIPMENT COULD RESULT.

WARNING: NOT TAKING EMC MEASURES INTO ACCOUNT ON THE WIRING MAY
RESULT IN INCREASED EMISSIONS OR DECREASED IMMUNITY. EN IEC 60601-
1-2:2015 MUST BE FOLLOWED FOR BEING COMPLAINT WITH EMC GUIDELINES.
WARNING: THE EMISSIONS CHARACTERISTICS OF THIS EQUIPMENT MAKE IT
SUITABLE FOR USE IN INDUSTRIAL AREAS AND HOSPITALS. IF IT IS USED IN A
RESIDENTIAL ENVIRONMENT, THIS EQUIPMENT MIGHT NOT OFFER ADEQUATE
PROTECTION TO RADIO-FREQUENCY COMMUNICATION SERVICES. THE USER MIGHT
NEED TO TAKE MITIGATION MEASURES, SUCH AS RELOCATING OR RE-ORIENTING THE
EQUIPMENT.

Page 33 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


Precautions to be taken to prevent adverse events to the PATIENT and the
OPERATOR due to Electromagnetic Disturbances are listed in the column
“Electromagnetic environment –guidance” in the tables below.
Manufacturer's declaration - Electromagnetic emissions
The Device is suitable for use in the specified electromagnetic environment. The
purchaser or user of the Device should assure that it is used in an electromagnetic
environment as described below:
Emissions test Compliance Electromagnetic environment - guidance
This Device uses RF energy only for its
RF emissions internal function. Therefore, the RF emission
Group 1
CISPR 11 is very low and not likely to cause any
interference in nearby electronic equipment.
RF emissions This Device is suitable for use in all
Class A
CISPR 11 establishments other than domestic and
Harmonic emissions Not those directly connected to the low voltage
IEC 61000-3-2 applicable power supply network which supplies
Voltage fluctuations / buildings used for domestic purposes.
Not
flicker emissions
applicable
IEC 61000-3-3

Page 34 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


Manufacturer's declaration - Electromagnetic immunity
The Device is suitable for use in the specified electromagnetic environment. The purchaser or
user of the Device should assure that it is used in an electromagnetic environment as described
below:
Immunity IEC 60601-1-2 Level of Electromagnetic environment -
tests test level compliance guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrostatic ± 8 kV contact
In case of high discharges at the
discharge ± 8 kV contact (except for
joystick, very rare in consideration
(ESD) ± 2/4/8/15 kV air joystick ± 2 kV)
of the use of the device and the
IEC 61000-4-2 ± 2/4/8/15 kV air
movement of the operators, the
display could be unreadable and
the device shall be restarted in
order to restore its operation.
Portable and mobile RF
Radiated
communications equipment should
electromagneti 3 V/m IEC 60601-1-2
be used no closer to any part of the
c field 80 MHz to 2.7 GHz Test level
Device, including cables.
IEC 61000-4-3
Minimum distance 30 cm
± 2 kV for power
Electrical fast supply lines Mains power quality should be that
IEC 60601-1-2
transient/burst ± 1 kV for of a typical commercial or hospital
Test level
IEC 61000-4-4 input/output lines > environment.
3m
0.5/1 kV differential
Mains power quality should be that
Surge mode IEC 60601-1-2
of a typical commercial or hospital
IEC 61000-4-5 0.5/1/2 kV common Test level
environment.
mode
Conducted Portable and mobile RF
disturbances 3 V 150 kHz to 80 communications equipment should
IEC 60601-1-2
induced by RF MHz be used no closer to any part of the
Test level
fields 6V ISM frequencies Device, including cables.
IEC 61000-4-6 Minimum distance 30 cm
Voltage dips. Mains power quality should be that
10 ms – 0% a 0°,
short of a typical commercial or hospital
45°, 90°, 135°,
interruptions environment. If the user of the
180°. 225°, 270°,
and voltage IEC 60601-1-2 Device requires continued
315°
variations on Test level operation during power mains
20 ms – 0% a 0°
power supply interruptions, it is recommended
500 ms – 70% a 0°
input lines that the Device be powered from
5 s – 0%
IEC 61000-4- an uninterruptible power supply or

Page 35 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°


11 a battery.
Power Power frequency magnetic fields
frequency should be at levels characteristic of
IEC 60601-1-2
(50/60 Hz) 30 A/m a typical location in a typical
Test level
magnetic field commercial or hospital
IEC 61000-4-8 environment.

Guidance and manufacture's declaration


Compliance Test Levels for ENCLOSURE PORT IMMUNITY to RF wireless
communications equipment
Test IMMUNITY
Max Distanc
Frequenc Banda) TEST
Service a) Modulation b) Power e
y (MHz) LEVEL
(W) (m)
(MHz) (V/m)
Pulse
380 -
385 TETRA 400 modulationb) 1,8 0,3 27
390
18 Hz
FM c)
430 - GMRS 460, ±5 kHz
450 2 0,3 28
470 FRS 460 deviation
1 kHz sine
710 Pulse
704 - LTE Band 13,
745 modulation b) 0,2 0,3 9
787 17
780 217 Hz
810 GMS 800/900
TETRA 800, Pulse
800 -
870 IDEN 820, modulation b) 2 0,3 28
960
CDMA 850, 18 Hz
930 LTE Band 5
GSM 1800;
1720 CDMA 1900;
GSM 1900; Pulse
1700 -
1845 DECT; modulation b) 2 0,3 28
1990
LTE Band 1, 3, 217 Hz
1970 4, 25;
UMTS
Bluetooth;
WLAN; Pulse
2400 -
2450 802.11 b/g/n; modulation b) 2 0,3 28
2570
RFID 2450; 217 Hz
LTE Band 7
5240 Pulse
5100 -
WLAN 802.11 a/n modulation b) 0,2 0,3 9
5500 5800
217 Hz
5785
a) For some service, only the uplink frequencies are included
b) The carrier shall be modulated using a 50% duty cycle square wave signal
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used
because while it does not represent actual modulation, it would be worst case.

Page 36 of 227
Technical Manual code: DBQ117-EN

V - SAFETY BRF 90°/ 15°

V.6 – Cables and wires

WARNING! Changing factory cables (following list) with others non compatibles
or with different features may affect the electromagnetic safety of the device

Cable Description Length Note

3G 2,5 General power supply 15m -

WARNING! To avoid electric shock this EM device MUST be connected to a


power source properly earthed

WARNING! Changing the non detachable POWER SUPPLY CORD, non


detachable is only possible by service authorized staff. For a proper procedure
of substitution and attachment of power supply cord, please take direct contact
with the manufacturer.

SUGGESTED EXTERNAL SAFETY MEANS

three-polar differential switch type B, 30 mA 16A

LINE IMPENDANCE

Mail line impedance <0,1Ω

Page 37 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


VI – Unpacking

BRF unit could be packed singly or dismantled. If BRF unit is entire it is singly
packed and protected in a wooden cage or in a wooden crate. The dimensions
and the unpacking procedure are the same for packing in a wooden cage or in a
wooden crate. If BRF unit is dismantled it is packed in two different wooden
cages or in two different wooden crates. The dimensions and the unpacking
procedure are the same for packing in a wooden cage or in a wooden crate.

BRF unit single package:

Dimensions:
• 2420x1700xh2020mm
Weight:
• Cage Kg.215
• Crate Kg.240

BRF unit dismantled package:

Package1
Dimensions:
• 2350x1400xh1180mm
Weight:
• Cage Kg.153
• Crate Kg.188

Package2
Dimensions:
• 2350x1400xh1000mm
Weight:
• Cage Kg.149
• Crate Kg.184

To move this type of packing it is mandatory to use proper gear (for example a
pallet truck).

Page 38 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


The image below describes the wooden cage used for the single package which
contains the single complete unit

The image below describes the wooden cages used for the dismantled unit
package.
Package 1 (lower one in the image) contains the base.
Package 2 (upper one in the image) contains column, spot film device,
collimator and accessories.

Page 39 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°

Page 40 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


VI.1 – Unpacking procedure: single package

1. Remove the top cover:


a. Unscrew all the screws in the top
b. Lift and remove the top cover

2. Remove the front cover and the ceiling of the cage wooden, remove also
the wooden boards that block the carriage. Remove the “carriage kit” (red
circle)

Page 41 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


3. Pull the BRF unit rotating the wheel with the lever.

4. Put the slider in the front side and fix them with the given screws. Align
them this the guide already fixed on the bottom of the cage.

Page 42 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


5. Put out from the cage the BRF unit. Beware to not make this operation too
fast!!!

6. Bring the BRF in the installation site.

Page 43 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


VI.2 – Unpacking procedure: dismantle package

VI.2.1 – DISMANTLE PACKAGE: BASE

The unpacking procedure of the case (or crabe) contains the base is the
same of the unpacking procedure of the single package

VI.2.2 – DISMANTLE PACKAGE: COLUMN AND SFD

1. Open up the column cage and remove the column from the package.

2. Remove the column from the cage and put it in vertical position on a
elevating carriage.

NOTE: FOR THIS OPERATION 3 MEN IS NEEDED

Page 44 of 227
Technical Manual code: DBQ117-EN

VI - UNPACKING BRF 90°/ 15°


3. Move the column near the base (see Chapter “VII – Installation”).
4. Remove the SFD from the package.

NOTE: KEEP AN EYE TO THE COMPRESSOR ON THE LOWER PART OF TH SFD

5. mount the SFD as explane in the chapter “VII – Installation”.

Page 45 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII – Installation

Installation of the medical device BRF (every code and models) must be
performed only by qualified personnel, trained by BMI Biomedical
International during the mandatory training course.

VII.1 - Preliminary check

Verify that the room of destination of the X-Ray units has dimensions and
structures wide enough to host the devices, considered the encumbrances
already described in this manual and the need for installation of an X-ray
generator and its control console. As well, verify that the room allows hosting all
accessories and other objects eventually needed for examination.

It is important to check that the immediate surroundings of the X-ray device is


allowing enough free space for the operator movement and to avoid collision
with objects and people. (please also ref to Pic. 8-1)

Moreover verify all component and devices not manufactured by BMI


BIOMEDICAL INTERNATIONAL, needed for the use of the X-ray system, (such as
the X-ray generator or the X-ray tube) and other accessories eventually
designed to be hosted in the room in association with the X-Ray system. All
these objects must respond to the following parameters:

- CE mark of conformity, where required;


- Perfect compatibility (mechanical, electrical and functional for all
component of the system;
- Position in the room which helps as much as possible the devices access
by operators and patient, and made as ease as possible examination
performing

For what concerns the elements which are not part of the functional
configuration of the system, which have anyway a use during some examination
(e.g. service stool) and some furniture elements (such as wardrobes or
desktops): for all these elements take care that their presence will not result as
an obstacle for device movement and ordinary work of the system.

Room features:

The room designed for installation of the system must respond to the following
features:
- Conformity and certification in line of international and local laws for room
designed to host an X-Ray source
- Electric plant in line of international and local laws, and according to all
devices designed to work in the base of the same power plant. For BMI
BIOMEDICAL INTERNATIONAL devices, please ref to the present manual,

Page 46 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


for other complementary or alternative devices, please ref to the proper
manuals and documents.
- Walls, floors and ceilings adequate to insert of fixing parts, and strong
enough to handle the weight of the different structures installed.

Safe Zones:

A B

850 850

Pic. 8-1
Installation of the BRF tilting table: “safe zones” A – B – C – D.

The drawing of Pic. 8-1 indicates the “safe zones” for the tilting table BRF model
AR4001 where the operator and other authorised personnel can stand protected
by X-Ray emission and avoiding risk of collision with the moving parts of the
device.

The drawing above also shows the minimum distances from the walls of the
room where the machine is being installed and defines the minimum distances
areas required for the tilting table to be operational, including therefore the
“safe zones” as above, distances to be borne in mind even more wherever the
tilting table should form part of a radio-fluorographic system including other
tables or complementary equipment.
The distances shown in the drawing in Pic. 8-1 must obviously be suitably
adapted wherever the tilting table (models AR4000 and AR4002) is used with a
lateral column support or suspended column (or any other suitable device),
depending on the overall size and the movements of the said columns and the
examinations foreseen (such as radiography of the thorax).

Page 47 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


It should also be noted that BRF unit can be shipped fully assembled in a single
crate: should this be the case the door providing access to the room must be at
least 1600mm wide.

Electromagnetism:

The BRF tilting table has been designed and tested so as to achieve the highest
standards of electromagnetic protection and compatibility.
However, it is recommended that the unit be installed and used in areas free of
influence from those devices which might cause interference and negatively
affect the performance of the system: sources of radio emission, UPS and other
devices

VII.2 - Environmental conditions

All devices described in the present manual can be used in the following
environmental conditions:

- Temperature : from + 10 °C (≈ 50 °F) a + 40 °C (≈ 104 °F)


- Maximum relative humidity: 80 %
- Maximum Altitude ( sea level ) : 3500 m

VII.3 - Radiological table BRF installation

For the installation of the BRF tilting table the following assembly kit is available
(Code BLD999):

- N°2 Bottles of powder paint (1 RAL9016BF and 1 RAL5014) for potential


fixings (mix a very little quantity of colour powder with paint thinner till
obtain a homogeneous solution)
- N° 9 – Plates 10/10 (for counterweight)
- N° 6 – Plates 20/10 (for counterweight)
- N° 9 – Plates 50/10 (for counterweight)
- N° 1 – CHI014 (special wrench Ø14)
- N° 1 – CHI017 (special wrench Ø17)
- N° 1 – CHI20 (special “blended” wrench Ø20)
- N° 1 – CHI120 (special “straight” wrench Ø20)
- N° 1 – User and maintenance manuals

Standard equipment also includes, save specific request by the customer:

- N°1 Power supply cable (without plug) length 15 m

Page 48 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


The Technical department responsible for installation of the table and relative
maintenance is supplied with:

- A sight to verify that the axis of the X-ray is perpendicular to the SFD
- A test phantom to verify centring of the collimator

When the BRF tilting table is shipped fully assembled, it’s packed in a single
cage or crate containing the fully mounted table, relative power supply cables,
assembly kit, instruction and maintenance manual and any accessories.
Crate: External Size Cm 242x170xh202, Weight: Kg 1420

When the BRF tilting table is shipped disassembled, it’s packed in 2 cage or
crate. The first cage or crate contain the base table and the relative power
supply cable. The second cage or crate contain the SFD, column, collimator,
fluoroscopy pedal (optional), assembly kit, instruction and maintenance manual
and any accessories.

Cage1: External Size Cm 237x140xh118, Weight: Kg 850


Cage2: External Size Cm 237x140xh100, Weight: Kg 520

Crate1: External Size Cm 237x140xh118, Weight: Kg 844


Crate2: External Size Cm 237x140xh100, Weight: Kg 517

Page 49 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII.4 – Base installation procedure

NOTE: pictures display only cages, procedure to unpack crated items is identical

1. Remove the front and top cover of the wooden cage, remove also the
wooden boards that block the carriage. Remove the “carriage kit” (red
circle)

2. Lift the BRF unit rotating the wheel with the lever.

Page 50 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


3. Position the slider in the front side and fix them with the given screws.
Align them this the guide already fixed on the bottom of the cage.

4. Put out from the cage the BRF unit. NOTE: please be careful to perform
this operation slowly and carefully.

5. Bring the BRF in the installation site.

6. Lower a bit the BRF rotating the wheel\levers (point 2) with the previous
lever and mark the floor next to the holes of fixation in the table base.
Please ref to the following picture:

Page 51 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


7. Lower to the floor, the BRF rotating the wheel\levers (ref point 2).

8. Remove with a wrench (size 17) the fixing screws in the picture. Remove
the bars inside the lateral support in both side.

9. Remove the bars in both sides unscrewing the screws in the red circles.

Page 52 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


Remove the lateral bars by sliding them out from one side and the other

10. Fix the BRF to the floor with the proper screws provided in the “carriage
kit”. (more information in the following section)

Page 53 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII.5 – Attachment of the base to the floor

The base of the BRF tilting table is provided with 5 attachment holes with a
diameter of 18 mm and its bearing surface is 1 m2.
The floor which the machine is placed on must be level and smooth so as to
enable full contact in all parts of the bearing surface of the said base.

Pic. 8-2 specifies the measurements to respect in attaching the base to the floor
and especially the minimum distances from the end wall of the installation
room.
Plugs for M12 screws with minimum tensile strength of 800 kg must be
used for such attachment (see “carriage kit”).

Ingombro del tavolo


table encumbance

Pic. 8-2
Detail of attachment of the base of the tilting table to the floor and relative
distances

Page 54 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


NOTE:

All the operations in the following sections must be performed according to


relative paragraphs and specifications of all other devices used with BRF (e.g
image intensifier, collimator, X-ray tube assembly)

Page 55 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII.6 – Column installation procedure (only for units shipped
disassembled)

1. Remove the 4 screws (M6) on the transversal movement of the tabletop

2. Move the tabletop to the frontside to access to rear cover.

3. Remove the rear cover unscrewing the 4 screws in the picture (one of
them is under the tabletop, move it to discover the screw as per point 2):

Page 56 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


4. Remove the supports in the following picture

5. Open up the column cage and remove the column from the package.

Page 57 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°

6. Remove the column from the cage and put it in vertical position on a
elevating trolley.

NOTE: FOR THIS OPERATION 3 MEN ARE NEEDED

7. Position the column near the base.

8. Remove the lateral cover of the column.

Page 58 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°

9. Remove the bearing in the on the lateral side of the colimn (4 screws M8).

10. Elevate the column (using elevating trolley mentioned in point 6) till the
bearing in the yellow circle will be higher than the linear guideway in table
base.

11. Push the column aside the base and lower the column till the bearings
touch the linear guideway in table base.

Page 59 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°

12. Mount the column with the tube support. As show in the following picture.
Please note: these screws are holding the weight of all the column and
spot film device, it is very important that these are kept tight and strong.

Page 60 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


13. Remount the bearing mentioned in point 9 and fix the column to the
column chain.

14. Remove the back side cover of the column for the cables connection.

Page 61 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII.7 – Spot film device installation procedure (only for units shipped
disassembled)

1. Remove the SFD from the package.

NOTE: PLEASE PAY ATTENTION TO THE COMPRESSOR ON THE LOWER PART


OF THE SFD

Page 62 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°

2. Remove the upper cover of the SFD .

3. Verify the presence of the “bearing protection” inside the bearings

Page 63 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


4. Remove the bumper in the backside of the SFD.

5. Pull up the SFD and slide it in the sfd support from the front side

Page 64 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


6. Pay attention to remove the ball-bearings protection during the insertion
of the SFD. NOTE: DO NOT REMOVE THE BALL BEARINGS PROTECTION
BEFORE THIS PHASE!

Page 65 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


7. Set and adjust the SFD potentiometer very carefully according to the
following procedure: the SFD run is very long and it must be positioned
with great attention:
• rotate the potentiometer in the sense of the green arrow to reach
the end of the run
• When the end run is reached rotate the potentiometer a little bit in
the opposite way (yellow arrow) to avoid it to work near the end run

This operation allow to make use of the entire run of the potentiometer without
cause the broken of the potentiometer

8. Block the SFD with the bumper removed previously (Point 4).

Page 66 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


9. Verify the movement of the SFD.

Page 67 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII.8 – Electrical connection procedure (only for units shipped
disassembled)

1. Cut the plastic bend from the cable support.

2. Block the support of the chain to the column base with 2 M4 screws.

3. Connect the cable from the base with the cable from the column

Page 68 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


4. Have the CAN-BUS and the GROUND (green-yellow) cables passing
through the the hole show in the picture.

5. Insert the cable in the cable-support:

Page 69 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


6. Connect the CAN-BUS cable to the column-board. Pass the GROUND cable
under the board and connect it to the free connector.

7. Connect the tubolar cable support to the back side of the SFD:

8. Connect the GROUND cables to the equipotential bar (1), the CN1
connector (2)and the CAN-BUS cable (3).

1 2 3

Page 70 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


VII.9 – Cover installation procedure

1. Remount the front cover with 5 “M5” screws

2. Remount the lateral cover of the column and the top cover.

Page 71 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°

VII.10 – Remove the counterweight block

NOTE: remove the counterweight block only after the complete


assembly of the BRF table.

1. Put the BRF at 0° (horizontal position)

2. Remove the cover on the bottom of the column.

3. Remove the nut that block the counterweight bar, after removal the block
will slide freely.

4. Reinstall the cover.

Page 72 of 227
Technical Manual code: DBQ117-EN

VII - INSTALLATION BRF 90°/ 15°


Procedure to add counterweights

1. Tilt the table to 89°.

2. Remove the M12 nut on the bar.

3. Insert the new counterweights and push it far deep to compat them to the
others.

4. Remount the nut (M12) and tight very hard.

5. Check the balance of the SFD.

Page 73 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
VIII – Installation and use of accessories

The following accessories are installed on all models of the BRF titling table:
• foot support
• protective leaded curtains
• handgrips (2x)
• shoulder support

Is also possible have in optional the following accessories:


• pressing band
• lateral cassette support
• leg supports (2x)
• fluoroscopy pedal
• patient screen

VIII.1 – Foot support

It is a fundamental and an essential accessory of the tilting table.


Assembling foot support must precede any clinical use of BRF unit and it is
practically permanent, remove it only during periodic cleaning operations or for
specific examinations which require images of the patient in a strictly horizontal
position.

1 3

1
3

Foot support and attachment details

To fit the foot support at the “foot end” of the table, proceed as follows:
1. Tilt the table and bring the table top to horizontal position.
2. Firmly grip the handles (1) and position the guides (2) over table top profiles
at the “foot end” of the table.
3. Keeping the 2 buttons (3) pressed and insert the guides of the foot support
into the profiles of the table top: push all the way back and release the
buttons to ensure it is blocked into position.

Page 74 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°

To dismantle, proceed as follows:


1. Move the table top in a horizontal position.
2. Grip the handles (1) as above.
3. Press the buttons (3) and keeping them pressed.
4. Pull the guides of the foot support out of the profiles of the table top.

When the installation is complete, the foot support is automatically identified by


the anti-collision system which prevents possible collisions with other parts of
the machine during movement.

VIII.2 - Leaded X-ray protective screens

According to paragraph “V.4 – Protection against ionizing radiations”, the leaded


screens must be attached to the sides of SFD, whenever radiographic or
radioscopic exposure is performed.

Page 75 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
The following picture show the fixation mechanism

Is it possible to use leaded screen in the following positions

Page 76 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
VIII.3 - Pair of hand-grips

The hand-grips protect the patients’ hands preventing them from being struck
by objects during table sliding movement. Handgrips avoid that patient hands
hang loose from the sides of the table, give also a sense of security to the
patient and help them to stay still during the examination.
In the Trendelenburg position, they act as a support.

Use of these accessories is heartily recommended.


BRF standard supplies include two hand-grips.

To fix handgrips to the table top metal profiles, hooking in and then blocking
them rotating relative knob.

blocking
bloccaggio

Relative assembly detail into the longitudinal metal profiles of the table top; the
handgrip can slide along the profile of the table and be blocked into the desired
position by rotating the specially designed knob

It is best to fit the handgrips once the patient has been positioned on the table
top, after adjust the position according to the height of the patient by sliding
them along the profiles of the examination table. Be sure that they are properly
blocked in place before starting the examination.

Page 77 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
VIII.4 - Shoulder support

The shoulder support is used to hold up the patient during examinations


requiring the Trendelenburg inclination.

2
2

head 1

Pic. 8-1
Shoulder support supplied with the BRF and relative assembly detail

To fit the shoulder support, proceed as follows:

1. Positioning the patient with the table top in the horizontal position (0°)
2. Put the shoulder support near the “head end” of the table and insert the
guides (1) as shown in Pic.8-1.
3. Slide the shoulder support along the table top profiles to ensure the right
position in relation to the patient height.
4. Block the shoulder support in place using the knobs (2).

VIII.5 – Pressing Band

The compression belt, usable when it is necessary to perform an X-Ray exam to


elastic anatomical parts (abdomen), this belt reduces the quantity of widespread
radiation with a great enhancement of image quality.

The extremity of the pressing band accessory and the gear which holds the
band itself are designed for a fixation on the aluminium profiles of all BMI
BIOMEDICAL INTERNATIONAL tabletops as displayed in Pic.8-2.

Page 78 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
3

1
.

Pic.8-2

Details of the mounting parts of pressing band and gear on tabletops


aluminium profiles.
1- Hook of fixation to pressing band extremity; 2- Grip to operate on the gear:
A clockwise rotation will revise the pressure of the band; 3- Unlocking device of
the pressing band.

Install the pressing band on the tabletop profile form any side (head or feet)
and position it in the desired area of examination; ensure that is fixed in a
proper way.

Installation procedure:
1. Unlock the band by mean of device nr 3 in Pic. 8-2 in order to act on
desired area of examination; fix the extremity of the band to the opposite
profile with hook nr 1 in Pic. 8-2.
2. Attach to extremity of the band to the other side of the tabletop by mean
of the hook (detail 1 in Pic. 8-2).
3. Act on the knob (detail 2 ion Pic. 8-2) to tighten the band.
4. When got the desired pressure on the band, this is automatically blocked
by detail 3.
5. Push detail 3 to release the band

It is recommended to warn the patient before pressure with the band, and its
effect and purposes.
VIII.6 – Lateral cassette support

This support accepts cassettes from size 24x30 cm to 35x35 cm; it is helpful
when the examination on progress requires latero-lateral projections of certain
anatomical parts. The use of this device made easier to perform this
examination without force the patient to uncomfortable positions

Page 79 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°

(B) (A)
Fig.8-3

Lateral cassette support: (A)- detail of mounting part for cassette support on
the aluminium profile of tabletop with blocking gear (1)
(B)- Lateral view of cassette support.

In Pic.8-3 (A) and (B) some mounting part detail of the lateral cassette support
are displayed to show the installation on the aluminium support, by mean of the
proper guides.

Installation procedure:
1. Unlock the blocking knob counter clockwise
2. Insert the cassette holder along the aluminium profile
3. Adjust the position in reason of imaging area
4. stabilize the position by locking the blocking knob clockwise

The specific configuration of this support allows the use of X-ray cassettes with
size 24x30 cm to 35x35 cm in both directions.
For a more practical use and a better imaging result it is suggested to employ
cassette incorporating the X-ray grid.

VIII.7 – Pair of leg supports

The leg supports are used mainly in examinations of the urological and genital
organs.
They must be fitted to the metal profiles of the table top by hooking and then
blocking them as shown in detail in Pic. 8-4.

Page 80 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°

Pic. 8-4
Leg support supplied with the BRF tilting table (two leg supports as standard
fittings) and relative detail of fitting to the longitudinal metal profiles of the
table top; the leg support can move along the section of the table and is
blocked in position by rotating the knob 1.
The leg support can also be adjusted in height to adapt to the size of the
patient, by loosening the knobs 2 and 3, and then blocking them again once the
desired height has been reached.

With the patient lying on the tabletop in a horizontal position, hook the leg
supports to the profiles, sliding them along to the relative part of the body.
Make sure that they are firmly attached to the profiles before positioning the
patient’s legs, and adjust the height if necessary.

VIII.8 – Fluoroscopy pedal

The fluoroscopy pedal is simple to install and allow the user to have a hand free
during examination

Page 81 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
The fluoroscopy pedal has a cable with two wires that must be connected on the
interface board of the unit (PCB103); the wire of the cable must be connected in
the pin 7 and 10 of the connector M5 as show hereunder.

Pin
Pin

VIII.9 - Patient Screen

The kit of the new Patient Screen Support Contain:

• L-shaped Support
• 2x M4x12 screws
• 2x M4 nuts

1. Join the Patient Screen with the support: make attention the two part are
aligned.

Page 82 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
2. With the two part joined, make two holes in correspondence of the blue
arrows.

3. Use the Screws and the nuts to fasten the Patient Screen and the support

Page 83 of 227
Technical Manual code: DBQ117-EN

VIII - INSTALLATION AND USE OF


ACCESSORIES
BRF 90°/ 15°
4. Place the patient screen on the BRF’s SFD as shown in the following
picture.

Page 84 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
IX – Installation and centring x-ray components

ATTENTION: during every start up system and after every reset of the boards,
the first 40 second will be needed by the system to enable the acoustic and
visual signal of safety (micro-switches, emergency button, etc.). Nevertheless,
every safety feature is active and the system is perfectly safe even in these 30
seconds.

IX.1 – Fitting the image intensifier

Remove the protective casing of the SFD.

A) B)

Mounting the image intensifier on the SFD

Fix the attachment plate to the image intensifier (A); then attach the mentioned
plate to the SFD (B).
Put the casing back on.

Page 85 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
IX.2 – Fitting the x-ray tube assembly and the attachment flange of
the collimator

Order of operations:

- Power the unit at 220 Vac using the cable provided, connected to the panel
fitted to the base of the table

- Tilt the table towards the vertical position so to have access the frame “B”
supporting the X-ray tube assembly and the attachment flange of the
collimator

Removal of the frame “B” supporting the X-ray tube assembly and the
attachment flange of the collimator, placed below the tabletop of the tilting
table

- Remove the frame “B” by unscrewing the four security bolts

Page 86 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
- Fix the X-ray tube assembly to the tube support, and to the sealing flange of
the collimator as per picture above; the same four screws hold the tube
assembly and the collimator flange.

Detail of assembly of the X-ray tube and the attachment flange of the
collimator to the support frame “B”

Assembly of the frame “B”, with attached X ray tube unit and collimator flange,
below the examination table

- Put the frame “B” back on, with attached X-ray complex and collimator
flange, below the examination table and attach it using the four security bolts
removed before; do not tighten these bolts as in next steps these bolts will
be used to adjust the collimator centring.

Page 87 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
IX.3 – Collimator installation procedure

1. Remove the back side cover of the collimator.

2. Put the collimator in the collimator tray under the tabletop.

3. Connect the cable and fix the collimator as described in the collimator
manual.

4. Verify the correct working of the BRF


Page 88 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
IX.4 – Connection and positioning of high voltage cables

The connection of the high voltage cables to the X-ray tube assembly must be
done according to the procedure shown in the follows picture.

A B

C D

High voltage cables connections

A. Cables wiring starts from tube; Cables must have the right length to permit
the correct movement of the tube support. Cables get into the given
“slide” and are fixed on by the given angled metal support.
B. Cables goes round behind column support bearing and are fixed with the
given metal support (it is necessary to remove and fix again the bearing
support: be careful to the washers used as spacers)
C. Cables pass under the column and exit through the apposite hole at the
lateral end of the column
D. Cables are fixed by the apposite angle metal support and exit toward floor
at the down end of the column
For this operation the table may be kept in the vertical position, or inclined
slightly in order to proceed in the most comfortable position.

Page 89 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
IX.5 – checking of perpendicularity between the x-ray beam and the
collimator

Sequence of operations:

- Put the table back in horizontal position

- With the table in a horizontal position, move the table top completely to the
“head end” (left with operator in front of unit).
The SFD too must be positioned longitudinally at the end of its
run at the “head end “ of the table (left with operator in front of unit).

b
c

b
c

SW105

A) B)

Access to the X-ray tube assembly under the table and to the attachment
flange of the collimator: first stage.

- Turn off the power supply using the switch (base of the table or power supply
panel)

- Unscrew the four “b” screws in Pic. A (only 2 of the 4 screws are shown in
the picture, the other 2 are positioned symmetrically on the longitudinal sides
of the table).

- remove the protective frame of the panel movement chain (Pic. B)

- Loosen the two “c” screws in Pic. B in order to reduce as much as possible
the tautness of the mylar sheet
Page 90 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°

The picture shows one of the two screws enabling loosening of the tautness of
the mylar sheet.

Page 91 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
- Connect up the power supply to the machine again following the Setup
procedure described below:

o After turning on the device, keep pressed keys 1 (panoramic) and 6


(division into 6) placed on the main control panel of the unit
simultaneously.

6 7

o First release key 1 and then key 6

- ATTENTION: with the BRF in Set Up, control of the movements using the
software is deactivated. In addition it’s possible to control the X-ray
exposure solely by means of the high voltage generator, not from the SFD

Move the table top towards the “foot end” (right with operator in front of
unit). Until the microswitch SW105 will be activated; when happened so
press and keep pressed the recovery button and move again the table top
towards the “foot end” making it protrude by 55 cm, not more

Unscrew the two holding screws, remove the protective frame from the chain
moving the table top, placed at “the head end” of the table (Pic. A; left with
operator in front of unit).

- Unscrew the five screws holding the mylar sheet

- Raise the mylar sheet and position it on the examination table as shown in
the picture hereunder, blocking it in place.

These operations give full access to the X-ray tube assembly already fitted
under the table with the attachment flange of the collimator, so as to first verify
the orthogonality of the axis of the X-ray with the SFD aperture window, then
assemble the X-ray tube collimator.

Page 92 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°

A) B)

Accessing the X-ray tube assembly under the table and the attachment flange
of the collimator: second phase.

These operations give full access to the X-ray tube assembly already fitted
under the table along with the attachment flange of the collimator, to first check
the orthogonality of the axis of the X-ray with the SFD aperture window, then
assemble the X-ray tube collimator.

Page 93 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
IX.6 – Verification of centring of the collimator and of the format of
the imaging field

To check the centring of the collimator use the test phantom shown hereunder.

Test phantom for checking the centring of the collimator, composed of a series
of radio-opaque rounds, symmetrically positioned

The procedure for such verification is as follows:

- Attach the aforesaid phantom to the centre of the aperture window of the
SFD using adhesive tape

- Perform a fluoroscopy emission and regulate the collimator chest (as describe
below) until the fluoroscopy image is centred with the central circle of the
phantom.

- Take two radiographic images, using cassettes with film 35x35 cm, one
image with the SFD positioned at the minimum distance from the
examination table and the other with the SFD positioned at the maximum
distance

- ATTENTION: to load cassettes in the SFD you must access the collimator
setup procedure (see setup chapter XVI.7 – Collimator) and follows the setup
procedure described in section “XVI.7.3 – Collimator exposition”.

For any adjustments required:

o in a longitudinal direction: adjust the four bolts attaching the support


frame of the X-ray tube and flange of the collimator
Page 94 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°

Adjustment and centring operation of the collimator in longitudinal direction

o transversely: adjust the four holding screws of the collimator on the


relative support flange

- Repeat the above operation until the correct result is achieved

Page 95 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
- On the radiographs thus obtained check the perfect symmetry of the images
of the rounds in the test.
As an example the following image shows a negative radiographic result with
inadequate centring requiring adjustment

Radiographic image of the test phantom showing the absence of centring in a


longitudinal direction

The aforesaid verification must be followed by verification of the format of the


imaging field, with reference to the field 30x30 cm, which is the one set for the
system, proceed as follows:

- Take two X-rays (without using sights or phantoms) using cassettes with
35x35 cm film: one image with the SFD positioned at the minimum distance
from the examination table and the other with the SFD positioned at the
maximum distance

- Then check that the resulting X-rays (black area on the film) are 30x30 cm

For any corrections follow the “Set Up Procedure”:

Page 96 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
- Scroll through the menu using keys 15 (kV+) – scroll upwards – and 17 (kV
-) – scroll downwards – placed on the main control panel of the unit, and
select the relevant function.
To confirm selection press key 13 (AUTO)

• Set Up Collimator 300X300 at the minimum distance

< COL.300x300MIN >


- XXXX YYYY -

After selecting the Set Up of the collimator 300X300 at the minimum


distance, the operator must operate manually on the collimator joystick,
with transversal and longitudinal movements, until a radiographic
projection of 300mm X 300mm is achieved at the minimum distance.
ATTENTION: Once the correct positions of the leaves of the collimator
have been identified, to save the settings press key 13 (AUTO), while to
abort the procedure, press key 15 (KV+= or 17 (KV-).

• Set Up Collimator 300X300 at the maximum distance

< COL.300x300MAX >


- XXXX YYYY -

After selecting the Set Up of the collimator 300X300 at the maximum


distance, the operator must operate manually on the collimator joystick,
so as to regulate the aperture of the collimator leaves and achieve a
radiographic projection of 300mm X 300mm at the maximum distance.
ATTENTION: Once the correct positions of the leaves of the collimator
have been identified, to save the settings press key 13 (AUTO), while to
abort the procedure, press key 15 (KV+) or 17 (KV-).

Finally follow the procedure to restore BRF to normal operating conditions:

- Re assemble the mylar sheet at the “head end” of the table, attaching it with
the screws for the purpose, and the protective frame of the chain moving the
tabletop.
Slide the examination table towards the “head end” (left with operator in
front of unit) making it out of the limit of the table by not more than 60 cm.
- Disconnect the power supply
Tighten the mylar sheet from “foot end”, (right with operator in front of unit)
using the screws for the purpose
- Fit the micro-switch SW105 and restore the relative electric connection using
the faston.

Page 97 of 227
Technical Manual code: DBQ117-EN

IX - INSTALLATION AND CENTRING


XRAY COMPONENTS
BRF 90°/ 15°
Remount the chain protective frame moving the tabletop from “foot end”
(right with operator in front of unit).
Normal operating conditions are restored with the supply of electricity to the
machine.

IX.7 - Verification of compressor correct working

- At the BRF unit start up check if the compressor is in field.


- If the compressor is in field, after the initialization of the unit put the
compressor in parking position.
- REPEAT THESE CHECK AT EVERY START UP TO ENSURE THE CORRECT
WORKING OF THE TABLETOP AND SFD SAFETY WHEN THE
COMPRESSOR FORCES ARE ACTIVE (movement lock of tabletop and
SFD).

Page 98 of 227
Technical Manual code: DBQ117-EN

X - SYSTEM COMPONENTS BRF 90°/ 15°


X – System Components

On following tables are listed all electro-mechanical components of the BRF


unit.

X.1 - Spot Film Device

Electrical boards
Component Id Type / Model / Comment Connected to
SFD keyboard PCB1
SFD control board PCB2
Sensor X&Y board PCB3
Joystick board PCB4

X.RAY Commands
Component Id Type / Model / Comment Connected to
PCB1
Prep-Exe button SW1 1° click e 2° click (CN5
PIN 1° 1-2 /2° 3-4)
Fluoro button SW2 PCB1 (CN6 PIN 1-2)

Table controls
Component Id Type / Model / Comment Connected to
Joystick tilting JSTK1 PCB1 (CN7)

Joystick tabletop JSTK2 PCB1 (CN7)

Joystick collimator JSTK3 PCB1(CN7)

Joystick sfd JSTK4 PCB1 (CN7)

Grid
Component Id Type / Model / Comment Connected to
Motor MOT1 24V= / 30W PCB2 (CN5 PIN 7-3)
FC zero SW3 PCB2 (CN9 PIN 5-13)

FC XRAY SW4 PCB2 (CN9 PIN 6-14)

Compressor
Component Id Type / Model / Comment Connected to
Motor MOT2 24V= / 30W PCB2 (CN5 PIN 8-4)

FC Parking SW5 PCB2 (CN9 PIN 3-11)

FC in field SW6 PCB2 (CN9 PIN 4-12)

Cassette tray
Component Id Type / Model / Comment Connected to
Motor MOT3 24V= / 60W PCB2 (CN6 PIN 1-3)
PCB2
(CN6
PIN 2 -
Motor Position control ENC1 Encoder 25 pulses / turn
PIN 6 OUT 2
PIN 4 OUT 1
PIN 5 +
FC safety min SW7 Parking side PCB2 (CN9 PIN 1-10)
FC safety max SW8 Cassette load side PCB2 (CN9 PIN 2-9)

Page 99 of 227
Technical Manual code: DBQ117-EN

X - SYSTEM COMPONENTS BRF 90°/ 15°


Transversal shutter (parallel and cross divisions)
Component Id Type / Model / Comment Connected to
Motor MOT4 24V= / 60W PCB2 (CN7 PIN 1-3)
PCB2
(CN7
PIN 2 -
Position control ENC2 Encoder 25 pulses / turn
PIN 6 OUT 2
PIN 4 OUT 1
PIN5 +

FC zero position SW9 Shutters completely open PCB2 (CN9 PIN 7-15)

Cassette tray devation


Component Id Type / Model / Comment Connected to
Solenoid SX SLN1 24V= / 12W PCB2 (CN5 PIN 1-5)

Solenoid DX SLN2 24V / 12W PCB2 (CN5 PIN 1-5)

Cassette size sensors (magnetic reed)


Component Id Type / Model / Comment Connected to

Cassette trasv. (x) 18, 24, 30, 35, 43 cm PCB3

Cassette long. (y) 24, 30 cm PCB3

X.2 - Table

Electrical boards
Component Id Type / Model / Comment Connected to
Table control board PCB101
Base keyboard PCB102
Interface board PCB103
Column board PCB104

Emergency
Component Id Type / Model / Comment Connected to
Emergency button SW101 PCB101 (CN2 PIN 1-9)
Recovery button SWREC PCB101 (M1 PIN 1-2)

Tilting (TLITING UP/DOWN)


Component Id Type / Model / Comment Connected to
400V~ / 1,1 KW 6 poli
Motor MOT101 MD101
900g/min
Motor brake BRK101 Brake handled by inverter MD101

Position control POT101 Potentiometer 3 turn – 1K PCB101 (M2)

PCB101 (CN4
Motor drive MD101 Inverter 1.5KW
PIN 14-6-1-9)

FC safety min SW103 Tilt down (“–15°”)


PCB101(CN2 PIN 2-10)
FC safety max SW102 Tilt up (“+90°”)

Page 100 of 227


Technical Manual code: DBQ117-EN

X - SYSTEM COMPONENTS BRF 90°/ 15°


Longitudinal tabletop (PANEL HEAD/FOOT)
Component Id Type / Model / Comment Connected to
Motor MOT102 400V~ / 0,25 KW MD102
Position control POT102 Potentiometer 10 turn – 1K PCB101 (M3)
PCB101 (CN4 PIN 10-2-
Motor drive MD102 Inverter 0.4KW
15-7)

FC safety min SW104 Panel HEAD


PCB101 (CN2 PIN 3-11)
FC safety max SW105 Panel FOOT

Transversal tabletop (PANEL IN/OUT)


Component Id Type / Model / Comment Connected to
Motor MOT103 OSLV MR62, 24V= / 100W PCB101(CN3 PIN 4-8)

Motor brake BRK103 PCB101(CN5 PIN 4-8)


Position control POT103 Potentiometer 3 turn – 1K PCB101 (M4)

FC safety min SW106 Panel IN PCB101 (CN2 PIN 6-14)

FC safety max SW107 Panel OUT

SFD longitudinal (SFD HEAD/FOOT)


Component Id Type / Model / Comment Connected to

Motor MOT104 400V~ / 1,5 KW


MD103

Motor brake BRK104


Position control POT104 Potentiometer 3 turn – 1K PCB101 (M5)
PCB101 (CN4 PIN 11-3-
Motor drive MD103 Inverter 1.5KW
8-16)

FC safety min SW108 SFD HEAD


PCB101 (CN2 PIN 4-12)
FC safety max SW109 SFD FOOT

Collimator tracking (COLLIMATOR IN/OUT)


Component Id Type / Model / Comment Connected to

Motor MOT106 24V= PCB101 (M8)

Motor brake BRK106 PCB101 (CN5 PIN 3-7)


Position control POT106 Potentiometer 3 turn – 1K PCB101 (M6)
FC safety min SW115 Collimator IN PCB101 (CN6 PIN 3-7)
FC safety max SW116 Collimator OUT PCB101 (CN6 PIN 4-8)

SFD compression (SFD UP/DOWN)


Component Id Type / Model / Comment Connected to
Motor MOT105 24V= PCB104 (CN1 PIN 1-9)
Position control BRK105 PCB104 (CN1 PIN 5-13)
Position control POT105 Potentiometer 3 turn - 1K PCB104 (CN2 15-13-10)
PCB104 (CN2 PIN 7-
FC Safety min SW110 SFD DOWN
COMM)
PCB104 (CN2 PIN 8-
FC Safety max SW111 SFD UP
COMM)
PCB104 (CN2 PIN 4-
FC rope rupture SW114 SFD COLUMN ROPE
COMM)
FC compression PCB104(CN2 PIN 6-
SW112 SFD COMP1
strength 1 COMM)
FC compression SW113 SFD COMP2 PCB104 (CN2 PIN 5-

Page 101 of 227


Technical Manual code: DBQ117-EN

X - SYSTEM COMPONENTS BRF 90°/ 15°


strength 2 COMM)

SFD transversal (SFD IN/OUT)


Component Id Type / Model / Comment Connected to
Brake BRK102 PCB104 (CN1 PIN 4-12)
PCB104 (CN2 PIN 14-12-
Position control POT107 Potentiometer 10 turn – 1K
9)

Footrest
Component Id Type / Model / Comment Connected to
Footrest present SW119A
PCB101 (CN2 PIN 2-6)
Footrest present SW119B

Collimator
Component Id Type / Model / Comment Connected to
RALCO R302 MP/A PCB101 (CN8)

Potter is an optional item according to the version.

Potter
Component Id Type / Model / Comment Connected to
Potter board PTT101 PCB103 - M7

Brake unlock button SW118 PCB101 (CN5 PIN 2-6)

Brake BRK107 24V= / 12W PCB101 (CN5 PIN 1-5)

FC parcking security SW117 PCB101 (CN6 PIN 1-5)

Page 102 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI - COMPONENTS POSITION

XI.1 – Brakes position

BRK105

BRK102

BRK106

BRK107

BRK104

BRK101

BRK103

Page 103 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI.2 – Microswitches position

SW114

SW112

SW111
SW113

SW110
SW119
(A e B)
SW109

SW115
SW117 SW106

SW105

SW108 SW116

SW104

SW102

SW103

Page 104 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI.3 – Motors position

MOT105

MOT106

MOT103

MOT104

MOT101

MOT102

Page 105 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI.4 – Potentiometers position

POT107

POT105

POT103

POT106
POT104

POT102

POT101

Page 106 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI.5 – Summary of components position

Page 107 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI.6 – SFD components position

MOT1
MOT4
MOT2 +
ENC2

SW7

SW6
MOT3
SW5 +
ENC1
PCB2
SFD PCB3
Control SW3 (up)
Cassette +
Board X&Y SW4 (down)
format

PCB1
SFD Keyboard

SW8

SW9
PCB4
SW1
X Ray button Joystick board
SW2
Fluoro button

Page 108 of 227


Technical Manual code: DBQ117-EN

XI - COMPONENTS POSITION BRF 90°/ 15°


XI.7 – Column components position

MOT105
SW114

SW111 BRK107

PCB104 POT107

BRK105

SW112
SW113

SW110
POT105

Page 109 of 227


Technical Manual code: DBQ117-EN

XII - FUSES BRF 90°/ 15°


XII – Fuses

Fuse Value Protected


Electrical board Type
name (delayed) power supply

10,3x38
F1 8A
(ceramic)
LINE 230 Vac Input line
10,3x38
F2 8A
(ceramic)

F1 6.3 A 5x20 (glass) power supply: 19 Vac - 6A

F2 3.15 A 5x20 (glass) power supply: 19 Vac - 1A

PCB101 F3 3.15 A 5x20 (glass) PCB104 power supply

F4 3.15 A 5x20 (glass) collimator power supply

F5 3.15 A 5x20 (glass) MOT103 power supply

PCB103 F1 3.15 A 5x20 (glass) power supply: 19 Vac - 6A

PCB2 F1 4A 5x20 (glass) power supply: 19 Vac - 3A

Page 110 of 227


Technical Manual code: DBQ117-EN

XIII - INTERFACE SIGNALS BRF 90°/ 15°


XIII – Interface signals
All incoming and outcoming signals to other devices are galvanically
insulated and implemented through opto-isolator or relay.
XIII.1 - X-ray HT generator
Signal Input / Output Type
Preparation ( 1° click ) OUTPUT relay
Exposition ( 2° click) OUTPUT relay
KV + OUTPUT relay
KV - OUTPUT relay
KV AUTO OUTPUT relay
DIGITALE Mode (activation) OUTPUT relay
Generator READY INPUT Opto-isolator
Generator XRAY ON INPUT Opto-isolator
DIGITAL mode (request) INPUT Opto-isolator
Tilting movement Block INPUT Opto-isolator
RX button Prep INPUT Opto-isolator
XRAY Button Exposition INPUT Opto-isolator

XIII.2 - Image intensifier / TV chain / collimator


Signal Input / Output Type
ZOOM1 OUTPUT relay
ZOOM2 OUTPUT relay
ZOOM3 OUTPUT relay
PULSED FLUOROSCOPY OUTPUT relay
AUTOMATIC FLUOROSCOPY OUTPUT relay
HORIZZONTAL INVERSION OUTPUT relay
VERTICAL INVERSION OUTPUT relay
FLUOROSCOPY PEDAL INPUT Opto-isolator

XIII.3 - Various
Signal Input / Output Type
Room light OUTPUT relay
RELAY 1 OUTPUT relay
RELAY 2 OUTPUT relay
RELAY 3 OUTPUT relay
RELAY 4 OUTPUT relay
SPARE 1 INPUT Opto-isolator
SPARE 2 INPUT Opto-isolator
SPARE 3 INPUT Opto-isolator
SPARE 4 INPUT Opto-isolator

XIII.4 - Potter (support connections)


Signal Input / Output Type
Power Supply --- ---
POTTER START INPUT ---
X RAY ON OUTPUT ---

Page 111 of 227


Technical Manual code: DBQ117-EN

XIV - INVERTERS BRF 90°/ 15°


XIV - Inverters

XIV.1 - Inverters description

Name Type Power supply Max power Used for drive


Tilting motor
MD101 VZAB1P5BAA 230Vac Monophase 1.5 KW
(MOT101)
Longitudinal panel motor
MD102 VZAB0P4BAA 230Vac Monophase 0.4 KW
(MOT102)
Longitudinal SFD motor
MD103 VZAB1P5BAA 230Vac Monophase 1.5 KW
(MOT104)

XIV.2 - Inverter connection signals

Name Description Connected to


Phase from terminal block of
L1/R
the electrical panel
Power Supply: 230Vac single-phase
Neutre from terminal block of
L2/S
the electrical panel
T1/U
Tri-Phase power Supply of Motor AC
T2/V Motor phases
at variable amplitude and frequency
T3/W

S1 digital command for FWD direction PCB101 – CN4

S2 digital command for REV direction PCB101 – CN4

SC common direction command PCB101 – CN4 & Inverter AC

A1 speed analog signal PCB101 – CN4


GND relay (A2 - KA1 or KA2)
AC common speed signal
& Inverter SC
MB and MC are used only with inverters MD101 and MD103

MB 24V command of 230Vac brake PCB101 – CN4


+24V relay (A1 - KA1 or KA2)
MC 24V command of 230Vac brake
& Inverter SC

Page 112 of 227


Technical Manual code: DBQ117-EN

XIV - INVERTERS BRF 90°/ 15°


XIV.3 - Inverters setting

Starting from manufacturing of the BRF tilting table serial number A04155 on, a
new type inverter is mounted. All subsequent serial numbers feature the
inverters here described.

Following the new setting procedure:

1. Press the arrows “UP” and ”DOWN” to scroll the menu:

2. Press them until the string “PAR” appears on display:

3. Press enter button to confirm.

4. Select the default parameter (A1-03) in the below “SETTINGS TABLE”:

Page 113 of 227


Technical Manual code: DBQ117-EN

XIV - INVERTERS BRF 90°/ 15°


a. To modify the alphanumeric character use arrorw “UP” and
”DOWN”
b. To shift form left to right the alphanumeric character use the
“RESET/LEFT” arrow
5. Press “ENTER” to confirm.

6. Now you could set the default parameter “A1-03”.

7. Select the value “2220”

8. Press “ENTER” to set the value.

9. Repeat the procedure (from point 1 to point 5) for all parameters,


inserting the value like in the follow “setting table”

SETTINGS TABLE – MD101 (Tilting inverter)

Parameter Value Description


A1-02 2 Control method selection
Acceleration time 1
C1-01 0.0
(Ramp)
Deceleration time 1
C1-02 0.0
(Ramp)
Carrier frequency
C6-02 4
selection
E1-05 230.0 Max output voltage
E1-13 230.0 Base voltage
E2-01 5.36 Motor rated current
H2-01 1 DO MA/MB Function

SETTINGS TABLE – MD102 (Long. Panel inverter)

Parameter Value Description


A1-02 2 Control method selection
Acceleration time 1
C1-01 0.0
(Ramp)
Deceleration time 1
C1-02 0.0
(Ramp)
Carrier frequency
C6-02 6
selection
E1-05 230.0 Max output voltage
E2-01 2.2 Motor rated current

Page 114 of 227


Technical Manual code: DBQ117-EN

XIV - INVERTERS BRF 90°/ 15°


SETTINGS TABLE – MD103 (Column inverter)

Parameter Value Description


A1-02 2 Control method selection
Acceleration time 1
C1-01 0.5
(Ramp)
Deceleration time 1
C1-02 0.0
(Ramp)
Carrier frequency
C6-02 4
selection
E1-05 230.0 Max output voltage
E1-08 16.6 Middle output voltage
E1-10 3.5 Min. output voltage
E1-13 230.0 Base voltage
E2-01 5.90 Motor rated current
E2-02 1.96 Motor rated slip
E2-03 3.80 Motor no-load current
Motor line-to-line
E2-05 3.500
resistance
E2-06 13.8 Motor leakage inductance
E2-07 0.35 Coefficient of glut 1
E2-08 0.61 Coefficient of glut 2
E2-11 1.10 Motor rated output
H2-01 1 DO MA/MB Function
L3-24 0.142 Motor acceleration time

Page 115 of 227


Technical Manual code: DBQ117-EN

XIV - INVERTERS BRF 90°/ 15°


XIV.4 – Moving motors by inverter

1. Check that the inverter is switched on

2. Press the arrows until “PAR” appears

3. Press to confirm

4. Select the parameter B1-01 using the arrows and

press to confirm

5. Set the value to 00 and press

6. Select the parameter D1-01 using the arrows and

press to confirm

7. Set the value to 10.00 and press

8. Press until you return to the principal menu

[on display appears F 10.00]

9. Press and check that the led is lighted on

WARNING

IN THIS PHASE THE SOFTWARE AND THE SAFETY MICROSWITCHES ARE


DISABLED, SO PLEASE PAY THE UTMOST ATTENTION : THE DEVICE
MUST NOT BE MOVED IN ORDER TO AVOID MECHANICAL DAMAGE

10. Press and be ready to press immediately if the device

should go in the opposite direction you requested

11. After stopping the device going in the opposite direction you requested,

follow the instructions “CHANGE DIRECTION” hereunder indicated

Page 116 of 227


Technical Manual code: DBQ117-EN

XIV - INVERTERS BRF 90°/ 15°


12. AT THE END OF ALL OPERATIONS AND BEFORE USING THE UNIT IN
“NORMAL MODE”, SET THE INVERTER AS BEFORE BEING MANUAL MOVED AS
DESCRIBED HEREUNDER

13. Press and check that the led is lighted off

14. Set the parameter B1-01 to 01

15. Set the parameter D1-01 to 00.00

16. Press until you return to the principal menu

CHANGE DIRECTION:

Press

Press the arrow until “FOR” or “REW” appear on display

Press ; “FOR” or “REW” will start blinking

Press once the arrow until the message changes (From FOR to REW or
from REW to FOR)

Press

Press

Move the device with and as described before in the procedure

WARNING

ALWAYS PAY THE UTMOST ATTENTION BECAUSE IN THIS PHASE THE


SOFTWARE AND THE SAFETY MICROSWITCHES ARE DISABLED, SO
PLEASE PAY THE UTMOST ATTENTION : THE DEVICE MUST NOT BE
MOVED IN ORDER TO AVOID MECHANICAL DAMAGE

Page 117 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV – How to program electrical boards

To upload the firmware on the electrical boards is necessary have the special
programming box (as shown in the picture)

PIN 1

The electrical boards program procedure is the same for every BRF electrical
boards.
However is essential have access to the boards how is explained in the following
section.

XV.1 - Electrical boards position

PCB3
STS00427ast

Page 118 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°

XV.2 – Access to table control board (pcb101)

To reach PCB101, remove the tilting table front cover. Please look at the
following pictures that will show the parts to remove (refer to section VII –
Installation):

1st STEP:
remove
this cover

2st STEP:
remove
this cover

Reset

Programming
connector
PIN1

Page 119 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.3 – Access to base keyboard (pcb102)

To reach PCB102, remove the tilting table front cover. Please look at the
following pictures that will show the parts to remove (refer to section VII –
Installation):

1st STEP:
remove
this cover

PCB102

PIN 1

Programming
connector

Reset

Page 120 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.4 – Access to generator interface board (pcb103)

To reach PCB103, remove the INTERFACE BOARD front cover. Please look at the
following pictures that will show the parts to remove (refer to section VII –
Installation):

1st STEP:
remove
this cover

PCB103

Programming
connector
Reset

PIN 1

Page 121 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.5 – Access to column control board (pcb104)

To reach PCB104, remove the column front cover. Please look at the following
pictures that will show the parts to remove (refer to section VII – Installation):

1st STEP:
remove
this cover

2st STEP:
remove
this cover

PCB104

Reset

Programming
connector

PIN 1

Page 122 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.6 – Access to SFD keyboard (pcb1)

To reach PCB1, remove the SFD front cover. Please look at the following
pictures that will show the parts to remove (refer to section VII – Installation):

1st STEP:
remove
this cover

PCB1

Reset
PIN 1
Programming
connector

Page 123 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.7 – Access to sfd control board (pcb2)

To reach PCB2, remove the SFD front cover. Please look at the following
pictures that will show the parts to remove (refer to section VII – Installation):

1st STEP:
remove
this cover

PCB2

PIN 1

Programming
connector

Reset

Page 124 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.8 - Electrical boards firmware description

Last
Description Identify Board Firmware Last Firmware
change
Board code code code release address
date
SFD KEYBOARD PCB1 STS00421 STS00423 R15 16/12/16 ADDR = 01
SFD CONTROL
PCB2 STS00424 STS00426 R02 23/11/12 ADDR = 02
board
COLUMN
PCB104 STS00432 STS00434 R04 16/02/11 ADDR = 06
CONTROL board
BASE CONTROL
PCB101 STS00435 STS00437 R07 21/04/10 ADDR = 05
board
INTERFACE STS00440_
PCB103 STS00438 R04 16/02/11 ADDR = 03
board V00
BASE KEYBOARD PCB102 STS00443 STS00445 R02 16/12/16 ADDR = 04

Page 125 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
XV.9 - Firmware installation procedure

The following procedure show how to upload the firmware on electrical boards:

1. Open the CD folder and copy the installation software on your desktop
(the icon below is shown)

2. Take care that the BRF tilting table is switched OFF and connect the
programming box as described herein:

(2) BRF board


(1) PC side side

(1) to the PC serial port (the programming box has female serial
connector on this side)

(2) to the electrical board programming connector:

THIS ATTENTION: be careful to


SIDE connect the black wire to
PIN 1 pin1 (see images); a wrong
connection could damage
the electrical board
permanently!

Page 126 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°
3. SWITCH ON BRF tilting table

4. Double-click on the installation software icon: the following window will


appear.

5. Reset the electrical board by pressing the dedicated button (marked with
a Yellow circle in previously given images and described as RESET)
6. Select the correct “COM” port on PC where serial cable is connected and
press “OK”.
7. A new window will appear. Press “Refer…” and chose the path on your PC
where the firmware is saved. An error message (“Can not found the ID
file.”) will appear, as shown in the picture below: ignore it and press
“OK”:

Page 127 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°

Description Identify Board Firmware


Board code code code
SFD
PCB1 STS00421 STS00423
KEYBOARD
SFD
CONTROL PCB2 STS00424 STS00426
board
COLUMN
CONTROL PCB104 STS00432 STS00434
board
BASE
CONTROL PCB101 STS00435 STS00437
board
INTERFACE
PCB103 STS00438 STS00440
board
BASE
PCB102 STS00443 STS00445
KEYBOARD

8. MCU Type: do not change the selection M16C/20 62. In the field ID, fill
the spaces with pairs of zeroes as shown in the picture below, and then
press “OK”:

9. The following window will appear:

Page 128 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°

10. Then clear the µC memory before programming, press “Erase”, and
press “OK” to confirm.

11. Wait until “Erase OK.” message appears, as shown below. Press
“OK” to confirm: you will be redirected to the previous window.

Page 129 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°

12. Then, to set the download baud-rate, press “Setting…”. The


following window will open:

13. Change the Baud rate into 57600, than press “OK”. You will be
redirected to the main programming page.
14. Press “Program…”: the following window will appear:

Page 130 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°

15. Without changing any values, press “OK”: the window will change
into:

16. Press “OK” again to start the firmware upload:

Page 131 of 227


Technical Manual code: DBQ117-EN

XV - HOW TO PROGRAM ELECTRICAL


BOARDS
BRF 90°/ 15°

17. At the upload’s end, a message “Program OK” will appear:

18. Press “OK”, and the download will be completed. Finally, press
“Exit” and wait until the programming software closes automatically.

ATTENTION: Always remember to press “Exit” after program is OK,


otherwise the programming process won’t be terminated successfully!

19. Switch off BRF tilting table and remove the programming box from
the connector on the electrical board
Page 132 of 227
Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI – Setup and check procedures

To store in the system memory the reference positions and dead ends for the
movements of each part of the BRF, a SETUP procedure is required before
starting normal operation of the tilting table.

ATTENTION: during every start up system and after every reset of the boards,
the first 40 second will be needed by the system to enable the acoustic and
visual signal of safety (microswitches, emergency button, etc.). Nevertheless,
every safety feature is active and the system is perfectly safe even in these 30
seconds.

XVI.1 – Access to setup mode

a. Switch off the BRF.

b. keeping buttons 1 (DIV1) and button 6 (DIV6X) pressed contemporarily

6 7

c. Switch on the BRF.

d. The following message will appear on the SFD keyboard screen:

STS00423R13 IT
SETUP BRF

This message identifies the firmware code of the board (STS00423)


immediately followed by the release code (R14) and the selected language
code (IT or else).

Page 133 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


e. Releasing FIRST button 1…
f. …THEN Releasing button 6
g. Shutter and carriage will moved to the parking position
h. After that the display will show the following message:

SFD
SETUP BRF

XVI.2 – Menu and submenu list

STS00423R11 IT
SETUP BRF

< SELECT SETUP > < SELECT SETUP > < SELECT SETUP > < SELECT SETUP >
- SFD - - TABLE - - COLLIMATOR - - COLUMN -

MOTOR CASSETTE TILTING COL. OPEN/CLOSE COLUMN UP/DW

MOTOR SHUTTER LONG. PANEL COL. 300x300 MIN POSITION SFD Z

LANGUAGE TRAS. PANEL COL. 300x300 MAX SPEED DOWN

EJECT CASSETTE LONG. SFD POS. TUBE RX EXIT SETUP

IMAGE INTENS. DEFAULT PARAM. EXPOSITION

DEFAULT PARAM. EXIT SETUP EXIT SETUP

EXIT SETUP

Page 134 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.3 – Menu and submenu navigation buttons list

NUMBER NAME ACTION


13 KV AUTO Enter / Confirm
15 KV + Forward
17 KV - Backward

13 15 17

XVI.4 – SDF setup

By selecting SFD setup, the operator will enter a new scroll menu with the
following options:

< SFD >


- MOTOR CASSETTE -
< SFD >
- MOTOR SHUTTER -
< SFD >
- LANGUAGE -
< SFD >
- EJECT CASSETTE -
< SFD >
- IMAGE INTENS. -
< SFD >
- DEFAULT PARAM. -
< SFD >
- EXIT SETUP

Page 135 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.4.1 – MOTOR CASSETTE setup menu

By selecting the MOTOR CASSETTE setup, the operator will access a second
level sub-menu which will allow to set the correct reference values for the
cassette motor movements.

– Setup DISTANCE:

a. Scrolling menu until the following message appear (DISTANCE – Motor


Cassette submenu):

<MOTOR CASSETTE>
- DISTANCE -

b. DO NOT PRESS ANY KEY

c. insert a meter into the cassette inlet and measure the distance between
the cassette carriage end (arrow 1) and the edge of the cassette inlet
itself (arrow 2), as shown in the picture:

X1

Page 136 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


Reference for the measurement:
• Point 1: SFD external side
• Point 2: carriage external side

X1

d. This measure will be called X1

e. Press key 13 (KV AUTO).

f. The cassette carriage will move toward the operator and the following
message will appear on the SFD keyboard screen:

< DISTANCE >


- WAIT POSITION -

g. When the movement stops, the message on the SFD keyboard screen
will change in:

< DISTANCE >


- INSERT POS : 420 -

Page 137 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


h. Measure again the distance between the cassette carriage end and the
cassette inlet edge (It’s very important use the same reference as
before)

i. This measure will be called X2

j. Calculate the difference ∆X=X1-X2, and compare this measure with


the value shown on the display. If the two values coincide, press key 13
to confirm, otherwise increase or decrease the value shown on the
display using keys 15 and 17 until it reaches ∆X, and then press key 13
to confirm.

k. The display will come back to “Motor Cassette” menu

l. Scrolling menu until will appear the following message (SAVE


PARAMETER – Motor Cassette submenu):

<MOTOR CASSETTE>
- SAVE PARAMETER -

m. Press key 13 (KV AUTO) to confirm and save the parameter

n. Select “EXIT SETUP”


<MOTOR CASSETTE>
- EXIT SETUP -

Page 138 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


o. Press key 13 (KV AUTO) for exit to the submenu

– Setup OFFSET:

The procedure is very similar to the one followed to set the Motor Cassette
DISTANCE. Scrolling onto the following menu:

a. Scrolling menu until the following message will appear (OFFSET – Motor
Cassette submenu):

<MOTOR CASSETTE>
- OFFSET -

b. NOT PRESS ANY KEY

c. insert a meter into the cassette inlet and measure the distance between
the cassette carriage end and the edge of the cassette inlet itself, as
explain in the Cassette DISTANCE setup

d. This measure will be called X1

e. Press key 13 (KV AUTO)

f. The cassette carriage will move toward the operator and the following
message will appear on the SFD keyboard screen:

< OFFSET >


- WAIT POSITION -

g. When the movement stops, the message on the SFD keyboard screen
will change in:

< OFFSET >


- INSERT POS : 400 -
h. Measure again the distance between the cassette carriage end and the
cassette inlet edge (It’s very important use the same reference as
before)

i. This measure will be called X2

j. Calculate the difference ∆X=X1-X2, and compare this measure with the
value shown on the display. If the two values coincide, press key 13 to
confirm, otherwise increase or decrease the value shown on the display

Page 139 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


using keys 15 and 17 until it reaches ∆X, and then press key 13 to
confirm.

k. The display will come back to “Motor Cassette” menu

l. Scrolling menu until will appear the following message (SAVE


PARAMETER – Motor Cassette submenu):

<MOTOR CASSETTE>
- SAVE PARAMETER -

m. Press key 13 (KV AUTO) to confirm and save the parameter

n. Select “EXIT SETUP”

<MOTOR CASSETTE>
- EXIT SETUP -

o. Press key 13 (KV AUTO) for exit to the submenu

Page 140 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.4.2 – MOTOR SHUTTER setup menu

By selecting the MOTOR SHUTTER setup, the operator will access a second level
sub-menu which will allow him to set the correct reference values for the
shutter motor movements.

– Setup DISTANCE:

a. Scrolling menu until will appear the following message (DISTANCE –


Motor Shutter submenu):

<MOTOR SHUTTER>
- DISTANCE -

b. NOT PRESS ANY KEY.

c. Insert a meter into the SFD and measure the distance as shown in the
picture:

Reference for the measurements:


• Point 1: SFD internal side
• Point 2: Shutter external side

d. This measure will be called S1.


Page 141 of 227
Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


e. Press key 13 (KV AUTO).

f. The shutters will move toward the x-ray field and the following message
will appear on the SFD keyboard screen:

< DISTANCE >


- WAIT POSITION -
g. When the movement stops, the message on the SFD keyboard screen
will change in:

< DISTANCE >


- INSERT POS : 220 -
h. Measure again the distance between the shutter external side and the
SFD internal side (It’s very important use the same reference as
before):

1 2

S2

Reference for the measurements:


• Point 1: SFD internal side
• Point 2: Shutter external side

i. This measure will be called S2.

Page 142 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


j. For calculate the difference ∆S it’s important consider the shutter’s
width: 150mm

k. Calculate the difference ∆S=S1-(S2-150mm), and compare this


measure with the value shown on the display. If the two values
coincide, press key 13 to confirm, otherwise increase or decrease the
value shown on the display using keys 15 and 17 until it reaches ∆S,
and then press key 13 to confirm.

l. The display will come back to “Motor Shutter” menu

m. Scrolling menu until the following message will appear (SAVE


PARAMETER – Motor Shutter submenu):

<MOTOR SHUTTER>
- SAVE PARAMETER -

p. Press key 13 (KV AUTO) to confirm and save the parameter

q. Select “EXIT SETUP”

<MOTOR SHUTTER>
- EXIT SETUP -

r. Press key 13 (KV AUTO) for exit to the submenu

– Setup OFFSET:

The procedure is very similar to the one followed to set the Motor Shutter
DISTANCE. Scrolling onto the following menu:

a. Scrolling menu until will appear the following message (OFFSET – Motor
Shutter submenu):

<MOTOR SHUTTER>
- OFFSET -

b. NOT PRESS ANY KEY

c. insert a meter into the cassette inlet and measure the distance between
the shutter external side and the SFD internal side, as explain in the
Shutter DISTANCE setup

d. This measure will be called S1

Page 143 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


e. Press key 13 (KV AUTO)

f. The shutters will move toward the x-ray field and the following message
will appear on the SFD keyboard screen:

< OFFSET >


- WAIT POSITION -

g. When the movement stops, the message on the SFD keyboard screen
will change in:

< OFFSET >


- INSERT POS : 400 -

h. Measure again the distance between the shutter external side and the
SFD internal side (It’s very important use the same reference as before)

i. This measure will be called S2

n. For calculate the difference ∆S it’s important consider the shutter’s


width: 150mm

o. Calculate the difference ∆S=S1-(S2-150mm), and compare this


measure with the value shown on the display. If the two values
coincide, press key 13 to confirm, otherwise increase or decrease the
value shown on the display using keys 15 and 17 until it reaches ∆S,
and then press key 13 to confirm

j. The display will come back to “Motor Shutter” menu

k. Scrolling menu until the following message will appear (SAVE


PARAMETER – Motor Shutter submenu):

<MOTOR SHUTTER>
- SAVE PARAMETER -

l. Press key 13 (KV AUTO) to confirm and save the parameter

m. Select “EXIT SETUP”

<MOTOR SHUTTER>
- EXIT SETUP -

n. Press key 13 (KV AUTO) for exit to the submenu

Page 144 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.4.3 – LANGUAGE setup menu

By selecting the LANGUAGE setup menu, the operator will access a second level
sub-menu which will allow him to select the language of the messages which will
appear on the BRF display during the “normal” function (the setup menu it’s
always in English).

a. Select “LANGUAGE”

< LANGUAGE >


- ITALIANO -
< LANGUAGE >
- ENGLISH -
< LANGUAGE >
- NOT USED 01 -
< LANGUAGE >
- NOT USED 02 -
< LANGUAGE >
- SAVE PARAMETER -
< LANGUAGE >
- EXIT SETUP -

b. Scrolling menu until will appear the desired language (ITALIAN or


ENGLISH).

< LANGUAGE > or < LANGUAGE >


- ITALIANO - - ENGLISH -

c. Press key 13 (KV AUTO) to confirm the selection.

d. Scrolling menu until will appear the following message (SAVE


PARAMETER – Language submenu)

< LANGUAGE >


- SAVE PARAMETER-

e. Press key 13 (KV AUTO) to confirm and save the parameter

f. Select “EXIT SETUP”

< LANGUAGE >


- EXIT SETUP -

g. Press 13 (KV AUTO) for exit to the submenu

Page 145 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.4.4 – EJECT CASSETTE setup menu

By selecting the EJECT CASSETTE setup menu, the operator will access a second
level sub-menu which will allow him to choose if the cassette carriage must exit
to the operator or must rest in the parking position when the cassette’s
exposures are finished and all divisions are full.

a. Select “EJECT CASSETTE”

< EJECT CASSETTE >


- ACTIVATED -
< EJECT CASSETTE >
- DISACTIVATED -
< EJECT CASSETTE >
- NOT USED 01 -
< EJECT CASSETTE >
- NOT USED 02 -
< EJECT CASSETTE >
- SAVE PARAMETER -
< EJECT CASSETTE >
- EXIT SETUP -

b. Scrolling menu until will appear the desired function of the cassette
carriage:
• ACTIVATED: the cassette carriage exit to the operator when the
cassette’s exposures are finished and all divisions are full
• DISACTIVATED: the cassette carriage rest in the parking position when
the cassette’s exposures are finished and all divisions are full

< EJECT CASSETTE > or < EJECT CASSETTE >


- ACTIVATED - - DISACTIVATED -

c. Press key 13 (KV AUTO) to confirm the selection.

d. Scrolling menu until will appear the following message (SAVE


PARAMETER – Eject cassette submenu)

< EJECT CASSETTE >


- SAVE PARAMETER -

e. Press key 13 (KV AUTO) to confirm and save the parameter

Page 146 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


f. Select “EXIT SETUP”

< EJECT CASSETTE >


- EXIT SETUP -

g. Press 13 (KV AUTO) for exit to the submenu

XVI.4.5 – IMAGE INTENSIFIER setup menu

By selecting the IMAGE INTENSIFIER (I.I.) setup menu, the operator will access
a second level sub-menu which will allow him to specify the size of the Image
Intensifier in use.

a. Select “IMAGE INTENSIFIER”

< IMAGE INTENS. >


- I.I. 12inch -
< IMAGE INTENS. >
- I.I. 9inch -
< IMAGE INTENS. >
- I.I. 6inch -
< IMAGE INTENS. >
- SAVE PARAMETER -
< IMAGE INTENS. >
- EXIT SETUP -

b.Scrolling menu until will appear the size of image intensifier mounted:

< IMAGE INTENS. > or


- I.I. 12inch -

< IMAGE INTENS. > or


- I.I. 9inch -

< IMAGE INTENS. >


- I.I. 6inch -

c. Press key 13 (KV AUTO) to confirm the selection.

d. Scrolling menu until the following message will appear (SAVE


PARAMETER – Image intensifier submenu)

< IMAGE INTENS. >


- SAVE PARAMETER -
Page 147 of 227
Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


e. Press key 13 (KV AUTO) to confirm and save the parameter

f. Select “EXIT SETUP”

< IMAGE INTENS. >


- EXIT SETUP -

g. Press 13 (KV AUTO) for exit to the submenu

XVI.4.6 – DEFAULT PARAMETERS setup menu

By selecting the DEFAULT PARAMETER setting, the operator will access a second
level sub-menu which will allow him to choose between two options:

< DEFAULT PARAM. >


- SAVE PARAMETER -
< DEFAULT PARAM. >
- EXIT SETUP -

a. Select DEFAULT PARAMETER – SAVE PARAMETER and press key 13


(KV AUTO) to confirm the selection for load the default parameters

< DEFAULT PARAM. >


- SAVE PARAMETER -

b. Select “EXIT SETUP”

< DEFAULT PARAM. >


- EXIT SETUP -

c. Press 13 (KV AUTO) for exit to the submenu

XVI.4.7 – EXIT to SFD setup menu

a. Select “EXIT SETUP”

< SFD >


- EXIT SETUP -

b. Press key 13 (KV AUTO) for return to the main setup menu

Page 148 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.5 – Table setup

Selecting the TABLE setup, the operator will enter a new scroll menu with the
following options:
< TABLE >
- TILTING -
< TABLE >
- LONG. PANEL -
< TABLE >
- TRAS. PANEL -
< TABLE >
- LONG. SFD -
< TABLE >
- DEFAULT PARAM. -
< TABLE >
- EXIT SETUP -

ATTENTION: During table setup procedures all movements and related


safety means are not controlled by software: in this procedures the only
safety systems that remain active are mechanical.

JSTK3
JSTK2

JSTK1

JSTK4

JSTK1: Tilting up/down movement command


JSTK2: Tabletop in/out and head/foot movements command
JSTK3: Collimator’s shutters movement command
JSTK4: SFD up/down and heed/foot movements command

Page 149 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.5.1 – TILTING setup menu

By selecting the TILTING movement setup menu, the operator will access to a
second level sub-menu which will allow to set all working parameters and limits
of this movement

– Setting tilting to 0°: MEDIUM

a. Scrolling menu until will appear the following message “POSITION


MED – VALUE XXXX”:

< POSITION MED >


- VALUE : XXXX -

b. The showed value is the value (measured in ADC points) of the


potentiometer POT101 connect to the tilting movement

c. Tilt the table to 0° (±1°)

d. Access into the BRF and turn the potentiometer POT101 until will
appear on display a value of about 270.

e. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

f. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Setting tilting to -15°: MINIMUM

a. Scrolling menu until will appear the following message “POSITION


MIN – VALUE XXXX”

< POSITION MIN >


- VALUE : XXXX -

b. Tilt the table to -15° (±1°)


NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 150 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


c. If during this procedure happened that the micro switch SW103 (safety
tilting down) will be pressed, follow the following procedure to recover
the system:
1 - Press and keep pressed the recovery button (SWREC) inside
the table (showed in the following image).

2 - Tilt the table just a little bit to the opposite side until the
micro switch isn’t pressed.

d. When the BRF is in the tilting minimum position (about -15° ±1°) Press
key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-) o
17 (KV+) to exit procedure without save the inserted value)

e. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Setting tilting to +90°: MAXIMUM

a. Scrolling menu until will appear the following message “POSITION


MAX – VALUE XXXX”

< POSITION MAX >


- VALUE : XXXX -

Tilt the table to +89° (±1°)


NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 151 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


b. If during this procedure happen that the micro switch SW102 (safety
tilting up) will be pressed, follow the following procedure to recover the
system:
1 - Press and keep pressed the recovery button (SWREC) inside
the table (as showed before).
2 - Tilt the table just a little bit to the opposite side until the
micro switch isn’t pressed

c. When the BRF is in the tilting minimum position (about +89° ±1°)
press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

d. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Save TILTING movement parameters

a. Select “EXIT SETUP”.

< TILTING >


- EXIT SETUP -

b. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing
keys 15 (KV-) o 17 (KV+) to exit procedure without save the inserted
value)

Page 152 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.5.2 – LONGITUDINAL PANEL setup menu

By selecting the LONG. PANEL movement setup menu, the operator will access
to a second level sub-menu which will allow to set all working parameters and
limits of this movement

< LONG. PANEL >


- MIN -
< LONG. PANEL >
- MED -
< LONG. PANEL >
- MAX -
< LONG. PANEL >
- EXIT SETUP -

– Setting longitudinal panel movement: MEDIUM

a. Scrolling menu until will appear the following message “POSITION MED
– VALUE XXXX”:

< POSITION MED >


- VALUE : XXXX -

b. The showed value is the value (measured in ADC points) of the


potentiometer POT102 connect to the longitudinal tabletop movement

c. Move longitudinally the tabletop until it’s in the centre

d. Access into the BRF and turn the potentiometer POT102 until will appear
on display a value of about 570.

e. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-) o
17 (KV+) to exit procedure without save the inserted value)

f. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

Page 153 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


– Setting longitudinal panel movement: MINIMUM

a. Scrolling menu until will appear the following message “POSITION MIX
– VALUE XXXX”

< POSITION MIN >


- VALUE : XXXX -

b. Move the tabletop (toward head side) of 600mm respect the centre
position;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

c. If during this procedure happened that the micro switch SW104 (safety
panel head) will be pressed, follow the following procedure to recover the
system:
1 - Press and keep pressed the recovery button (SWREC) inside
the table (showed in the following image).

Page 154 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°

2 - Move the tabletop just a little bit to the opposite side until
the micro switch isn’t pressed

d. When the tabletop is in the minimum position (about 600mm respect the
centre position) press key 13 (KV AUTO) to confirm the value. (Pressing
keys 15 (KV-) o 17 (KV+) to exit procedure without save the inserted
value)

e. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Setting longitudinal panel movement: MAXIMUM

a. Scrolling menu until will appear the following message “POSITION


MAX – VALUE XXXX”

< POSITION MAX >


- VALUE : XXXX -

b. Move the tabletop (toward foot side) of 400mm respect the centre
position;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 155 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°

c. If during this procedure happened that the micro switch SW105 (safety
panel foot) will be pressed, follow the following procedure to recover
the system:
1 - Press and keep pressed the recovery button (SWREC) inside
the table (as showed before).
2 - Move the tabletop just a little bit to the opposite side until
the micro switch isn’t pressed

d. When the tabletop is in the maximum position (about 400mm respect


the centre position) press key 13 (KV AUTO) to confirm the value.
(Pressing keys 15 (KV-) o 17 (KV+) to exit procedure without save
the inserted value)

e. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Save LONGITUDINAL PANEL movement parameters

a. Select “EXIT SETUP”.

< LONG. PANEL >


- EXIT SETUP -

b. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing
keys 15 (KV-) o 17 (KV+) to exit procedure without save the inserted
value)
Page 156 of 227
Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.5.3 – TRANSVERSAL PANEL setup menu

By selecting the TRAS. PANEL movement setup menu, the operator will access
to a second level sub-menu which will allow him to set all working parameters
and limits of this movement.

< TRAS. PANEL >


- MIN -
< TRAS. PANEL >
- MED -
< TRAS. PANEL >
- MAX -
< TRAS. PANEL >
- EXIT SETUP -

– Setting transversal panel movements: MINIUM and MEDIUM

a. Scrolling menu until will appear the following message “POSITION


MIN – VALUE XXXX”:

< POSITION MIN >


- VALUE : XXXX -

b. The showed value is the value (measured in ADC points) of the


potentiometer POT103 connect to the transversal tabletop movement

c. Move transversally the tabletop until it’s in the centre position (as
showed in the following image),
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 157 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°

d. Access into the BRF and turn the potentiometer POT103 until will
appear on display a value of about 500

e. Move the tabletop (toward column side) of 100mm respect the centre
position;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 158 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


f. If during this procedure happened that the micro switch SW106 (safety
panel IN) will be pressed, follow the following procedure to recover the
system:
1 - Press and keep pressed the recovery button (SWREC) inside
the table (showed in the following image).

2 - Tilt the table just a little bit to the opposite side until the
micro switch isn’t pressed.

g. When the tabletop is in the minimum position (about 100mm respect


the centre position) press key 13 (KV AUTO) to confirm the value.
(Pressing keys 15 (KV-) o 17 (KV+) to exit procedure without save
the inserted value)

h. WITHOUT MAKE ANY MOVEMENT OF THE TABLETOP, scrolling menu


until will appear the following message “POSITION MED – VALUE
XXXX”:

< POSITION MED >


- VALUE : XXXX -

i. Press key 13 (KV AUTO) and confirm the value. (The medium value
must be as the same as the minimum value).

j. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

Page 159 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


– Setting transversal panel movement: MAXIMUM

a. Scrolling menu until will appear the following message “POSITION


MAX – VALUE XXXX”:

< POSITION MAX >


- VALUE : XXXX -

b. Move the tabletop (toward operator side) of 100mm respect the


centre position;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

c. If during this procedure happened that the micro switch SW107


(safety panel OUT) will be pressed, follow the following procedure to
recover the system:
1 - Press and keep pressed the recovery button (SWREC) inside
the table (as showed before).
2 - Move the tabletop just a little bit to the opposite side until
the micro switch isn’t pressed

d. When the tabletop is in the maximum position (about 100mm


respect the centre position) press key 13 (KV AUTO) to confirm the
value. (Pressing keys 15 (KV-) o 17 (KV+) to exit procedure
without save the inserted value)

e. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will
be automatically saved.

Page 160 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


– Save TRANSVERSAL PANEL movement parameters

a. Select “EXIT SETUP”.

< TRAS. PANEL >


- EXIT SETUP -

b. Press key 13 (KV AUTO) to exit setup and save the values.
(Pressing keys 15 (KV-) o 17 (KV+) to exit procedure without save
the inserted value)

Page 161 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.5.4 – LONGITUDINAL SFD setup menu

By selecting the LONG. SFD movement setup menu, the operator will access to
a second level sub-menu which will allow him to set all working parameters and
limits of this movement.

< LONG. SFD >


- MIN -
< LONG. SFD >
- MAX -
< LONG. SFD >
- EXIT SETUP -

– Setting longitudinal SFD movement: MINIMUM

a. Scrolling menu until will appear the following message “POSITION


MIN – VALUE XXXX”

< POSITION MIN >


- VALUE : XXXX -

b. The showed value is the value (measured in ADC points) of the


potentiometer POT104 connect to the longitudinal SFD movement

c. Move longitudinally the SFD near the centre of its stroke

d. Access into the BRF and turn the potentiometer POT104 until will
appear on display a value of about 500

e. Check that the micro switches SW108 and SW109 and their relative
cams are properly fixed and regulated (fix and regulate the micro
switches and the cams if isn’t so).

f. Move longitudinally the SFD (toward head side) until the micro switch
SW108 is pressed;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 162 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


g. Press and keep pressed the recovery button (SWREC) inside the table
(showed in the following image).

h. Move longitudinally the SFD just a little bit to the opposite side until
the micro switch isn’t pressed.

i. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

j. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Setting longitudinal SFD movement: MAXIMUM

a. Scrolling menu until will appear the following message “POSITION


MAX – VALUE XXXX”

< POSITION MAX >


- VALUE : XXXX -

b. Move longitudinally the SFD (toward foot side) until the micro switch
SW109 is pressed;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

Page 163 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


c. Press and keep pressed the recovery button (SWREC) inside the table
(showed in the following image).

d. Move longitudinally the SFD just a little bit to the opposite side until
the micro switch isn’t pressed.

e. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

f. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Save LONGITUDINAL SFD movement parameters

a. Select “EXIT SETUP

< LONG. SFD >


- EXIT SETUP -

b. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing
keys 15 (KV-) o 17 (KV+) to exit procedure without save the
inserted value)

Page 164 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.5.5 – DEFAULT PARAMETERS setup menu

By selecting the DEFAULT PARAMETER setting, the operator will access a second
level sub-menu which will allow him to choose between two options:

< DEFAULT PARAM. >


- SAVE PARAMETER -
< DEFAULT PARAM. >
- EXIT SETUP -

d. Select DEFAULT PARAMETER – SAVE PARAMETER and press key 13


(KV AUTO) to confirm the selection for load the default parameters

< DEFAULT PARAM. >


- SAVE PARAMETER -

e. Select “EXIT SETUP”

< DEFAULT PARAM. >


- EXIT SETUP -

f. Press 13 (KV AUTO) for exit to the submenu

XVI.5.6 – EXIT to table setup menu

a. Select “EXIT SETUP”

< TABLE >


- EXIT SETUP -

b. Press key 13 (KV AUTO) for return to the main setup menu

Page 165 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.6 – Column setup

Selecting the COLUMN setup, the operator will enter a new scroll menu with the
following options

< COLUMN >


- COLUMN UP/DW -
< COLUMN >
- POSITION SFD Z -
< COLUMN >
- SPEED DOWN -
< COLUMN >
- EXIT SETUP -

ATTENTION: During column setup procedures all movements and


related safety means are not controlled by software: by mean of this,
the only safety systems still active are those mechanical.

XVI.6.1 – COLUMN UP/DOWN setup menu

By selecting the COLUMN UP/DW movement setup menu, the operator will
access to a second level sub-menu which will allow him to set all working
parameters and limits of the SFD UP and DOWN movements.

< COLUMN UP/DW >


- MIN -
< COLUMN UP/DW >
- MAX -
< COLUMN UP/DW >
- EXIT SETUP -

– Setting SFD vertical movement: MINIMUM

a. Scrolling menu until will appear the following message “POSITION


MIN – VALUE XXXX”

< POSITION MIN >


- VALUE : XXXX -

Page 166 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


b. The showed value is the value (measured in ADC points) of the
potentiometer POT105 connect to the SFD up/down movement

c. Move vertically the SFD near the centre of its stroke

d. Access into (behind) the BRF and turn the potentiometer POT105 until
will appear on display a value of about 500

e. Move vertically the SFD (toward lower position) until the micro switch
SW110 is pressed;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

f. Move vertically the SFD just a little bit to the opposite side until the
micro switch isn’t pressed.

g. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

h. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will
be automatically saved.

– Setting SFD vertical movement: MAXIMUM

a. Scrolling menu until will appear the following message “POSITION


MAX – VALUE XXXX”

< POSITION MAX >


- VALUE : XXXX -

b. Move vertically the SFD (toward upper position) until the micro switch
SW111 is pressed;
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

c. Move vertically the SFD just a little bit to the opposite side until the
micro switch isn’t pressed.

d. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

Page 167 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


e. Scrolling menu until will appear EXIT SETUP and press key 13 (KV AUTO) to
confirm the selection. In this way the inserted value will be automatically saved.

Page 168 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


– Save SFD UP/DOWN movement parameters

a. Select “EXIT SETUP

< COLUMN UP/DW >


- EXIT SETUP -

b. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing
keys 15 (KV-) o 17 (KV+) to exit procedure without save the inserted
value)

XVI.6.2 – COLUMN POSITION SFD Z setup menu: SFD in/out

By selecting the POSITION SFD Z setup menu, the operator will access to a
second level sub-menu which will allow him to set all working parameters and
limits of the SFD IN and OUT movements.

< POSITION SFD Z >


- MIN -
< POSITION SFD Z >
- MED -
< POSITION SFD Z >
- MAX -
< POSITION SFD Z >
- EXIT SETUP -

– Setting position SFD Z (out): MINIMUM

a. Scrolling menu until will appear the following message “POSITION


MIN – VALUE XXXX”

< POSITION MIN >


- VALUE : XXXX -

b. The showed value is the value (measured in ADC points) of the


potentiometer POT107 connect to the SFD in/out movement

c. Check that the mechanical blocks are correctly mounted and fixed. If
isn’t so mounted those in the right way (as show in the following
images)

Page 169 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°

d. Unlock brake BR102 pressing key 21 on the SFD keyboard (as show in
the following image):

Key led light ON = brake activated


Key led light OFF = brake deactivated

e. Move SFD manually toward the operator side until its stroke will stop
by the mechanical block,
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

f. Press key 13 (KV AUTO) to confirm the value (about 50 ADC points).
(Pressing keys 15 (KV-) o 17 (KV+) to exit procedure without save
the inserted value)

g. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

Page 170 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


– Setting position SFD Z: MEDIMUM

a. Scrolling menu until will appear the following message “POSITION


MED – VALUE XXXX”

< POSITION MED >


- VALUE : XXXX -

b. Unlock brake BR102 pressing key 21 on the SFD keyboard (as show in
the precedent image)

c. Move SFD manually toward the operator side until its stroke will stop
by the mechanical block. After that press key 21 on the SFD keyboard
and break the SFD.

d. Take a meter and measure the distance between the SFD external side
and the SFD support (as show in the following image):

Z1 Z2

e. This measure will be called Z1 (about 740mm)

f. Unlock the brake BR102 and move the SFD manually of 250mm

g. This measure will be called Z2 (about 490mm)

h. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15 (KV-
) o 17 (KV+) to exit procedure without save the inserted value)

i. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

Page 171 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


– Setting position SFD Z (IN): MAXIMUM

a. Scrolling menu until will appear the following message “POSITION


MAX – VALUE XXXX”

< POSITION MAX >


- VALUE : XXXX -

b. Unlock brake BR102 pressing key 21 on the SFD keyboard

c. Move SFD manually toward the column side until its stroke will stop by
the mechanical block,
NOTE – move the table taking hi care of the potentiometer value:
this value must NEVER be about 0, after that the potentiometer will
be broken!

d. Press key 13 (KV AUTO) to confirm the value (about 1000 ADC
points). (Pressing keys 15 (KV-) o 17 (KV+) to exit procedure
without save the inserted value)

e. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

– Save POSITION SFD Z parameters

a. Select “EXIT SETUP

< POSITION SFD Z >


- EXIT SETUP -

b. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing
keys 15 (KV-) o 17 (KV+) to exit procedure without save the inserted
value)

Page 172 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.6.3 – COLUMN SPEED DOWN menu

By selecting the SPEED DOWN menu, the operator will access to a second level
sub-menu which will allow him to set the speed of SFD compression (SFD down
movement speed)

< SPEED SETUP >


- MIN -
< SPEED SETUP >
- MED -
< SPEED SETUP >
- MAX -
< SPEED SETUP >
- SAVE PARAMETER -
< SPEED SETUP >
- EXIT SETUP -

a. Scrolling menu until will appear the three different speed of


compression (MIN, MED and MAX):

< SPEED SETUP > or


- MIN -

< SPEED SETUP > or


- MED -

< SPEED SETUP >


- MAX -

b. Press key 13 (KV AUTO) to confirm the selection.

c. Scrolling menu until will appear the following message (SAVE


PARAMETER – Column Speed Down submenu)

< SPEED SETUP >


- SAVE PARAMETER -

d. Press key 13 (KV AUTO) to confirm and save the parameter

e. The system automatically come back to COLUMN MENU

f. Scrolling menu and select EXIT SETUP for exit from SPEED DOWN
menu

< SPEED SETUP >


- EXIT SETUP -
Page 173 of 227
Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.6.4 – EXIT to column setup menu

a. Select “EXIT SETUP”

< COLUMN >


- EXIT SETUP -

b. Press key 13 (KV AUTO) for return to the main setup menu

Page 174 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7 – Collimator setup

Selecting the COLLIMATOR setup, the operator will enter a new scroll menu with
the following options:

< COLLIMATOR >


- COL.OPEN/CLOSE-
< COLLIMATOR >
- COL.300x300MIN -
< COLLIMATOR >
- COL.300x300MAX -
< COLLIMATOR >
- POS. TUBE RX -
< COLLIMATOR >
- EXPOSITION -
< COLLIMATOR >
- EXIT SETUP -

Page 175 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7.1 – COLLIMATOR OPEN AND CLOSE setup

a. Scrolling menu until will appear the following message “COLLIMATOR


– COL. OPEN/CLOSE”

< COLLIMATOR >


- COL.OPEN/CLOSE-

b. Press key 13 (KV AUTO)

• The collimator will open automatically its shutters and the following
message will appear on the SFD display:

< COLL. OPEN >


- LONG LATR -

• When the collimator shutters are completely opened the collimator will
close them automatically and the following message will appear on the
SFD display:
< COLL. OPEN >
- XXXX YYYY -
c. When the collimator shutters are completely closed the x-ray led will
light on and the calibration procedure will automatically exit the
preview menu:

< COLLIMATOR >


- COL.OPEN/CLOSE-

d. Select “EXIT SETUP

< COLLIMATOR >


- EXIT SETUP -

e. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing keys 15
(KV-) o 17 (KV+) to exit procedure without save the inserted value)

Page 176 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7.2 – COLLIMATOR POSITION TUBE RX setup

Before to do this calibration it’s important that the SFD in/out movement
is already calibrated (as describe before in “COLUMN – POSITION SFD
Z” setup).

a. Scrolling menu until will appear the following message “COLLIMATOR


– POS. TUBE RX”

< COLLIMATOR >


- POS. TUBE RX -

b. Press key 13 (KW AUTO)

• The tracking tray will automatically move toward the column side
until the micro switch SW115 is pressed. During this operation on
the display will appear:

< TUBE RX IN >


- VALUE : XXXX -

The showed value is the value (measured in ADC points) of the


potentiometer POT106 connect to the tracking movement

• After that, the tracking tray will automatically move toward the
operator side until the micro switch SW116 is pressed. During this
operation on the display will appear:

< TUBE RX OUT >


- VALUE : XXXX -

The showed value is the value (measured in ADC points) of the


potentiometer POT106 connect to the tracking movement

c. When the calibration procedure is completed it will automatically exit


the preview menu:
< COLLIMATOR >
- POS. TUBE RX -
d. Select “EXIT SETUP
< COLLIMATOR >
- EXIT SETUP -
e. Press key 13 (KV AUTO) to exit setup and save the values. (Pressing keys 15
(KV-) o 17 (KV+) to exit procedure without save the inserted value)

Page 177 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7.3 – COLLIMATOR EXPOSITION setup

This option is intended to let the operator make exposure even in the SETUP
operation mode, for example during Collimator setup – imaging field
300x300mm, minimum distance SFD-tabletop and Collimator setup – imaging
field 300x300mm, maximum distance SFD-tabletop procedures.

ATTENTION: Please note that, during Collimator setup, X-Ray exposure test
procedure, X-ray exposure can be commanded only from HT generator console
(and not from the tilting table console). Verify also that brake BR102 of
transversal SFD movement (Key 21, described before in the “COLUMN –
POSITION SFD Z” setup) is active.

a. Scrolling menu until will appear the following message “COLLIMATOR


– EXPOSITION”

< COLLIMATOR >


- EXPOSITION -

b. Press key 13 (KV AUTO)

c. The cassette tray will be positioned in load position and the collimator
shutters will be positioned for 30X30 cassette. (it is recommended to
use for this check a larger cassette: e.g. 35x35)

d. Load the cassette (35X35) and press again key 13 (KV AUTO)

e. The cassette tray will be automatically placed in the imaging field with
a “find zero” movement.

f. Make the need exposure

g. After exposure, press key 13 (KV AUTO)

h. The cassette tray will move again in the load\unload position

Page 178 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7.4 – COLLIMATOR 300x300 MIN setup

By selecting the COLLIMATOR 300X300 MIN setup menu, the operator will
access to a second level sub-menu which will allow him to regulate the
collimator shutters.
Before make this calibration it is necessary setting BRF by the “COLLIMATOR -
EXPOSITION” so that is possible make exposure (e.g. before in “COLLIMATOR
– EXPOSITION”).

a. Move SFD in the lowest position respect to the tabletop

b. Select COLLIMATOR – EXPOSITION and make an exposure by HT


generator (as explain before in “COLLIMATOR – EXPOSITION”)

c. Develop the film and measure the dimensions of the square. The
measured dimensions should be 300x300mm; if isn’t so scrolling
menu until will appear the following message “COLLIMATOR – COL.
300X300 MIN”

< COLLIMATOR >


- COL.300x300MIN -

d. Press key 13 (KV AUTO)

e. On the SFD display will appear the following message:

< COL.300x300MIN >


- XXXX YYYY -
f. The showed value is the value (measured in ADC points) of the
potentiometers connect to the collimator’s shutters

g. Move collimator shutters for adjust their position so that the square of
the exposure will be 300x300mm

h. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15


(KV-) o 17 (KV+) to exit procedure without save the inserted value)

i. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

j. Repeat the exposure (as explain before).

k. Repeat this procedure until the square of the exposure will be 300x300mm

Page 179 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7.5 – COLLIMATOR 300X300 MAX setup

By selecting the COLLIMATOR 300X300 MAX setup menu, the operator will
access to a second level sub-menu which will allow him to regulate the
collimator shutters.
Before make this calibration it is necessary setting BRF by the “COLLIMATOR -
EXPOSITION” so that is possible make exposure (as explain before in
“COLLIMATOR – EXPOSITION”).

l. Move SFD in the highest position respect to the tabletop

m. Select COLLIMATOR – EXPOSITION and make an exposure by HT


generator (as explain before in “COLLIMATOR – EXPOSITION”)

n. Develop the film and measure the dimensions of the square. The
measured dimensions should be 300x300mm; if isn’t so scrolling
menu until will appear the following message “COLLIMATOR – COL.
300X300 MAX”

< COLLIMATOR >


- COL.300x300MAX -

o. Press key 13 (KV AUTO)

p. On the SFD display will appear the following message:

< COL.300x300MAX >


- XXXX YYYY -

q. The showed value is the value (measured in ADC points) of the


potentiometers connect to the collimator’s shutters

r. Move collimator shutters for adjust their position so that the square of
the exposure will be 300x300mm

s. Press key 13 (KV AUTO) to confirm the value. (Pressing keys 15


(KV-) o 17 (KV+) to exit procedure without save the inserted value)

t. Scrolling menu until will appear EXIT SETUP and press key 13 (KV
AUTO) to confirm the selection. In this way the inserted value will be
automatically saved.

u. Repeat the exposure (as explain before).

v. Repeat this procedure until the square of the exposure will be


300x300mm

Page 180 of 227


Technical Manual code: DBQ117-EN

XVI – SETUP AND CHECK PROCEDURES BRF 90°/ 15°


XVI.7.6 – EXIT to collimator setup menu

a. Select “EXIT SETUP”

< COLLIMATOR >


- EXIT SETUP -

b. Press key 13 (KV AUTO) for return to the main setup menu

Page 181 of 227


Technical Manual code: DBQ117-EN

XVII - CLEANING OPERATIONS AND


ORDINARY MAINTENANCE
BRF 90°/ 15°
XVII – Cleaning Operations and ordinary maintenance

The user is required of a few cleaning and ordinary maintenance operations, and
a few visual check of some moving part. These operations are really simple, but
highly important for a safe use and a long life cycle of the device.

All controls and maintenance operation which requires the access to the internal
parts of the devices, as well as some ordinary and extraordinary maintenance
operations are exclusive duty of authorized personnel.

In the following table we report all common cleaning operation and visual check
recommended for a correct use of the devices.

Before attempting any cleaning or disinfection operation of the structures:

o Disconnect all devices from power supply


o Take care that cleaning and disinfection products are not aggressive and
not abrading
o Avoid that all liquids used for cleaning get in touch with powered parts of
the structures

Frequency Operation

Cleaning and disinfection of all surfaces in contact with the


After every use
patient (use a clean cloth and non aggressive product)

Check the correct work of all light signals and the correct
Daily readability of all labels and indications placed on the
structures.

Check eventual unusual noises, during movement or


Weekly
during brake activation / deactivation

Page 182 of 227


Technical Manual code: DBQ117-EN

XVIII - ORDINARY AND EXTRAORDINARY


MAINTENANCE
BRF 90°/ 15°
XVIII – Ordinary and extraordinary maintenance

All the following activities of ordinary and extraordinary maintenance must be


performed only by technical, duly trained personnel.

In following table are listed all recommended operations for periodically ordinary
maintenance:

Frequency Operations

- Control of brake efficiency and regularity of movements


- Cleaning of guides and mechanisms
Each six months
- Check of all electrical connections to the line (visual
check of plugs, wires and cables)

- Check centring of collimator and X-ray tube assembly


(refer to section IX - Installation and centring xray
Yearly
components)
- Check for wear and stretching of cables (visual check)

Every three years - Replacement of ropes (regardless of condition)

Description of ordinary maintenance

- Performances of brakes and movements tuning

1) Unlock the tabletop brake and move it: check the smoothness and the
noisiness. If necessary, remove the tabletop and clean the guides where
run the bearings and the bearings themselves;

2) With the tabletop braked, pull it and with a dynamometer check that the
braking power is around 14-18kg.

Page 183 of 227


Technical Manual code: DBQ117-EN

XVIII - ORDINARY AND EXTRAORDINARY


MAINTENANCE
BRF 90°/ 15°
- Cleaning of guide, rails and inner workings;

Check clearing and


consumption

- Check the power supply connection (cabling, plug and so on)

For every extraordinary maintenance operation as repair, spare parts


substitution or every operation that needs technical service not
described in this manual, please contact the manufacturer service staff.

Page 184 of 227


Technical Manual code: DBQ117-EN

XIX - WARRANTY BRF 90°/ 15°


XIX – Warranty

Page 185 of 227


Technical Manual code: DBQ117-EN

XIX - WARRANTY BRF 90°/ 15°

Page 186 of 227


Technical Manual code: DBQ117-EN

XX - LABELLING AND IDENTIFICATION


OF PRODUCTS
BRF 90°/ 15°
XX – Labelling and identification of products

All BRF units are identifiable by a label positioned as per IMAGE BELOW

Every unit could be shipped dismantled in 2 pieces (base and tabletop) in the
same package, every component is labelled with S/N placed on:

• Package
• Single part

This care avoids exchanging the single components with ones from other table
of the same model.

Page 187 of 227


Technical Manual code: DBQ117-EN

XXI - DISPOSAL BRF 90°/ 15°


XXI – Disposal

This device is compliant with RoHS Directive Recast (Directive 2011/65/EU)


related to reduction of the use of dangerous material on electrical and electronic
devices.
Disposal of the device is subject to WEEE Directive Recast (Directive
2012/19/EU), related to waste generated by electrical and electronic equipment.
Batteries (if included) are compliant with the Battery Directive (Directive
2006/66/EC).

The symbol with the crossed garbage bin indicate that the product, at the end of
its shelf life must be collected in separated way from the other garbage.

The separate collection of this device must be arranged and managed and by
the manufacture.
The user that want to get rid off this device will contact the manufacture and
follow is protocol for the garbage disposal.

A correct separate collection to perform a compatible recycling, treatment and


disposal of the device, aids to avoid negative effects to the environment and to
the health, it also supports a reuse and/or a recycling of the materials of the
device.
The abusive disposal of the device by the side of user involves the applying of
fines estimated by the in force law.

Page 188 of 227


Technical Manual code: DBQ117-EN

XXII - SYMBOLS BRF 90°/ 15°


XXII – Symbols

BRF tilting table presents several symbols, that here follow with detailed
explanations. Users and technicians are carefully intended to be aware of the
meaning of each symbol and to respect them in every operation.
User and technician are forbidden to remove any of the signals present on the
device.

Earth protection

Applied Part Type B

DANGEROUS TENSION PRESENCE


In presence of this symbol do not get close to the area and do not
attempt any intervention.

CE conformity mark
0051

Instruction

WARNING! REFERENCE ON THE MANUALS


Carefully read the accompanying documents before attempting any
intervention of maintenance or any operations in the area marked
with this symbol.

WAEE Disposal

WARNING! IONISING RADIATION PRESENT.


Use the appropriate anti X-ray protection and adopt the necessary
precautions in conformity with current legislation.

Page 189 of 227


Technical Manual code: DBQ117-EN

XXIII - REFERENCES AND


MANUFACTURER ADDRESS
BRF 90°/ 15°
XIII - References and manufacturer address

For all additional information or to receive technical assistance in case of


damages, failure or spare needs, please contact our offices as follows.

Via Enrico Fermi 52 Q/R


24035 Curno (Bergamo) Italy
Telephone: +39 035 4376381
Fax: +39 035 4376401
Web site: www.bmibiomedical.it
E-Mail: info@bmibiomedical.it

Page 190 of 227


Technical Manual code: DBQ117-EN

XXIV - CE MARK BRF 90°/ 15°


XIV – CE Mark

BRF 90°/15° is a medical device with CE mark achieve with support of a


notified body of the European Union.

0051

Page 191 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°


ANNEX A – Block Diagram

Page 192 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 193 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 194 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 195 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 196 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 197 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 198 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 199 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 200 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 201 of 227


Technical Manual code: DBQ117-EN

ANNEX A – Block Diagram BRF 90°/ 15°

Page 202 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


ANNEX B – ELECTRICAL BOARDS

B.1 - SFD KEYBOARD (PCB1)


- LAYOUT

Page 203 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°

Page 204 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


- CONNECTIONS

device connector pin positive negative description


Buttons
1–2–3–
SW1 (N.O.) CN5 4 1-2: 1° step / 3-4: 2° step X ray button
SW2 (N.O.) CN6 1–2 # # Fluoro button
Joysticks
JSTK1 CN7 13 - 14 # # tilting movement
panel head/foot
JSTK2 CN7 1-2 # # movement
CN7 3-4 # # panel in/out movement
CN7 5-6 # # collimator X movement
JSTK3
CN7 7-8 # # collimator Y movement
SFD head/foot
JSTK4 CN7 9 10 # # movement
CN7 11 - 12 # # SFD up/down movement

Not Used

TILTING TABLE

GND 15 16 SFD UP/DOWN


13 14
11 12 SFD HEAD/FOOT
9 10
7 8 COLLIMATOR Y
5 6
3 4 COLLIMATOR X
+5V
1 2
PANEL IN/OUT

PANEL HEAD/FOOT
From top

Page 205 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.2 - SFD CONTROL BOARD (PCB2) - LAYOUT

Page 206 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


- CONNECTIONS

device connector pin positive negative description


Microswitches
SW3 (N.O.) CN9 5 - 13 # # Grid zero switch
SW4 (N.O.) CN9 6- 14 # # Grid X-Ray switch
Compressor parking
SW5 (N.O.) CN9 3 -11 # # switch
Compressor in field
SW6 (N.O.) CN9 4 - 12 # # switch
SW7 (N.O.) CN9 1-9 # # Cassette parking switch
SW8 (N.O.) CN9 2 -10 # # Cassette out switch
SW9 (N.O.) CN9 7 - 15 # # Shutter parking switch
Motors
MOT1 CN5 7-3 3 7 Grid motor
MOT2 CN5 8-4 4 8 Compressor motor
MOT3 CN6 1-3 1=red 3=black Cassette motor
MOT4 CN7 1-3 1=red 3=black Shutter motor
2=(-); 4=OUT2; 5=(+);
ENC1 CN6 2-4-5-6 6=OUT1 Cassette encoder
2=(-); 4=OUT2; 5=(+);
ENC2 CN7 2-4-5-6 6=OUT2 Shutter encoder
Solenoids
SLN1 (outside) CN5 2-6 # # Solenoid SX
SLN2 (inside) CN5 1-5 # # Solenoid DX

CARRIGE/SHUTTER

Page 207 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.3 - SFD-CASSETTE X & Y FORMAT BOARD (PCB3)
- LAYOUT

Page 208 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.4 - SFD-JOYSTICK BOARD (PCB4)
- LAYOUT

Page 209 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


- CONNECTIONS

SFD Keyboard
connector pin connect to PCB1 pin description
CN1 1 CN7 13 GND SFD UP/DOWN
CN1 2 CN7 12 Signal SFD UP/DOWN
CN1 3 CN7 5 + SFD UP/DOWN
CN1 4 CN7 7 + SFD HEAD/FOOT
CN1 5 CN7 10 Signal SFD HEAD/FOOT
CN1 6 CN7 15 GND SFD HEAD/FOOT
CN2 1 CN7 6 Signal COLLIMATOR X
CN2 2 CN7 14 Signal TILTING
CN2 3 CN7 2 Signal PANEL HEAD/FOOT
CN2 4 CN7 9 GND 3JOYSTICK LONG.
CN2 5 CN7 1 + 3JOYSTICK LONG.
CN2 6 CN7 11 GND JOYSTICK TRANSV.
CN2 7 CN7 3 + 3JOYSTICK TRANSV.
CN2 8 CN7 4 Signal PANEL IN/OUT
CN2 9 CN7 16 Signal SPARE
CN2 10 CN7 8 Signal COLLIMATOR Y

Page 210 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.5 - BASE CONTROL BOARD (PCB101)

- LAYOUT

Page 211 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


– CONNECTIONS

device connector pin positive negative description


Microswitches
SW102 (N.C.) +90° tilting safety switch
CN2 2 - 10 in series
SW103 (N.C.) -15° tilting safety switch
SW104 (N.C.) Head panel safety switch
CN2 3 - 11 in series
SW105 (N.C.) Foot panel safety switch
SW106 (N.C.) IN panel safety switch
CN2 6 - 14 in series
SW107 (N.C.) OUT panel safety switch
SW108 (N.C.) Head SFD safety switch
CN2 4 - 12 in series
SW109 (N.C.) Foot SFD safety switch
SW115 (N.C.) CN6 3-7 # # IN collimator switch
SW116 (N.C.) CN6 4-8 # # OUT collimator switch
SW117 (N.O.) CN6 1-5 shortcirciuted without optional potter Potter parking switch
SW118 (N.C.) CN5 2-6 (only with optional potter bucky) Potter unlock brake switch
SW119A/B (N.O.) CN6 2-6 # # Footrest present switch
Special switches
SW101 (N.C.)
CN2 1-9 # # Emergency button
[emergency, pin 1-3]
SWREC (N.O.)
M1 1-2 # # Recovery button
[recovery]
Motors
MOT103 CN3 4-8 4 8 IN/OUT panel motor
MOT106 M8 1-3 1 3 Tracking motor
Brakes
BRK103 CN5 4-8 # # IN/OUT panel brake
BRK106 CN5 3-7 # # Tracking brake
BRK107 CN5 1-5 (only with optional potter bucky) Potter unlock brake
Potentiometers
POT101 M2 1-2-3 1=green; 2=brown; 3=white Tilting potentiometer
Head/foot panel
POT102 M3 1-2-3 1=white; 2=brown; 3=green
potentiometer
POT103 M4 1-2-3 1=white; 2=brown; 3=green In/out panel potentiometer
POT104 M5 1-2-3 1=green; 2=brown; 3=white Head/foot SFD potentiometer
POT106 M6 1-2-3 1=green; 2=brown; 3=white Tracking potentiometer

- EMERGENCY SIGNALS

Page 212 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.6 - BASE KEYBOARD (PCB102)

- LAYOUTS

Page 213 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°

Page 214 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.7 - INTERFACE BOARD (PCB103)

- LAYOUT

Page 215 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


- LEDS DESCRIPTION

X-ray generator ready

X-ray generator on

K15

Fluoroscopy pedal K14


VERTICAL REVERSE

K13
HORIZONTAL REVERSE

FLUORO

K11

K10
ZOOM 3

K9
ZOOM 2

ZOOM 1

K5

K4 KV AUTO

K3
KV-

K2
KV+

K1
X-RAY execution (2°step)

X-RAY preparation (1°step)

Page 216 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


- CONNECTIONS

type signal connector pin signal description


Output M1 2-3 X ray preparation (1° step)
Output M1 5-6 X ray execution (2° step)
Output M1 8-9 KW +
Output M1 11 - 12 KW -
Output M2 2-3 KW AUTO
Output M3 2-3 Zoom 1 (MAX field)
Output M3 5-6 Zoom 2 (MED field)
Output M3 8-9 Zoom 3 (MIN field)
Output M4 2-3 Fluoroscopy
Output M4 5-6 Horizontal reverse
Output M4 8-9 Vertical reverse
Input M5 1-9 Generator ready
Input M5 2 - 10 Generator XRAY ON
Input M5 7 - 10 Fluoroscopy pedal

Page 217 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


– ELECTRICAL SCHEMES

Page 218 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°

Page 219 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°

Page 220 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°

Page 221 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°

Page 222 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


B.8 - COLUMN & SFD CONTROL BOARD (PCB104)
- LAYOUT

Page 223 of 227


Technical Manual code: DBQ117-EN

ANNEX B – Electrical boards BRF 90°/ 15°


- CONNECTIONS

device connector pin positive negative description


Microswitches
CN2 - terminal
SW110 (N.C.) block 7 - com SFD down switch
CN2 - terminal
SW111 (N.C.) block 8 - com SFD up switch
Com = +24A from CN2, pin
CN2 - terminal
16
SW112 (N.C.) block 6 - com SFD compression1 switch
linked to terminal block
CN2 - terminal
SW113 (N.C.) block 5 - com SFD compression2 switch
CN2 - terminal
SW114 (N.O.) block 4 - com SFD column rope switch
Emergency
CN2 - terminal
SW101 block 1 - com # # Emergency button
Motors
MOT105 CN1 1-9 9 1 Up/down motor
Brakes
BRK105 CN1 5 - 13 # # Up/down movement brake
BRK102 CN1 4 - 12 # # In/out movement brake
Potentiometers
10=green; 13=brown; Up/down movement
POT105 CN2 10 - 13 - 15 15=white potentiometer
POT107 CN2 9 - 12 - 14 9=green; 12=brown; 14=white In/out movement potentiometer

Page 224 of 227


Technical Manual code: DBQ117-EN

ANNEX C – Installation report BRF 90°/ 15°


ANNEX C – Installation report

Via Enrico Fermi 52 Q/R


24035 Curno (Bergamo) Italy
Telephone: +39 035 4376381
INSTALLATION REPORT Fax: +39 035 4376401
Web site: www.bmibiomedical.it
E-Mail: info@bmibiomedical.it

Product
Code:
Serial
Customer: Number:

Manufacturing Installation
Date: Date:

Technician: Installation requested by:

Measur. Registered
Test and check Test Reference C NC NA Note
instrument Values

1.0 Review
1.1 Mechanical structure integrity visual - - -
1.2 Painting integrity visual - - -

2.0 First start


Base keyboard display
2.1 visual - - -
powered
SFD keyboard display
2.2 visual - - -
powered
2.3 System initialization visual - - -
Page 225 of 227
Technical Manual code: DBQ117-EN

ANNEX C – Installation report BRF 90°/ 15°


completed without any errors
Potter bucky brake green led
2.4 turned on (only BLD001 and visual - - -
BLD003 model)
2.5 Spot film device braked visual - - -

Mechanical
3.0
Movements
Spot film device
3.1 Brake unlock by button visual - - -
3.2 In/Out visual - - -
Potter bucky (only BLD001
and BLD003 model)
3.4 Brake unlock visual - - -
3.5 DX/SX visual - - -
Grid (only BLD001 and
BLD003 model)
3.6 IN/OUT visual - - -

Elettromechanical
4.0
Movements
Tabletop
Brake unlock with starting
4.1 visual - - -
movement by joystick
In/Out and automatic stop at
4.2 visual - - -
the limit setted
DX/SX and automatic stop at
4.3
the limit setted
-15/+89° and automatic stop
4.4
at the limit setted
Page 226 of 227
Technical Manual code: DBQ117-EN

ANNEX C – Installation report BRF 90°/ 15°


Spot film device
Brake unlock with starting
4.5 visual - - -
movement by joystick
DX/SX and automatic stop at
4.6 visual - - -
the limit setted
UP/DOWN and automatic
4.7 visual - - -
stop at the limit setted
Compressor
4.8 IN/OUT activated by button

5.0 User training


Train authorized final user on
5.1 correct use of this medical
device.

C: Complying - NC: Not Complying - NA: Not applicable


NOTE:

TEST RESULT Complying Not Complying

DATE Tester Signature

Note: send to BMI BIOMEDICAL INTERNATIONAL Srl by fax at the following number +39 035 4376401
e-mail quality@bmibiomedical.it

Page 227 of 227

You might also like