You are on page 1of 18

Operation

3.8.15 Autoleveller (ATL) – setting instructions


Machine control system, Page 3/26
Autoleveller (ATL) – calibration of the equipment, Page 3/30
The autoleveller compensates for volume deviations of the incoming sliver. Prior to delivery, each machine
is checked and set according to instruction. The periodic checking of the levelling unit must be carried out in
accordance with the setting instructions.
Preparatory work
– When performing any checking or setting task at the levelling unit, the safety instructions must be ob-
served.

Objective
– The mechanical and electrical inaccuracies are determined and the system set precisely.

General instructions on checking the levelling unit


– In case of faults at the levelling unit, the plant electrician has to carry out a check in accordance with the
setting instructions.

– The setting of the levelling unit has to be checked and if necessary adjusted whenever changes are
made at:
displacement-transducer system (ATL)
at the target (ATL)
at the D295 processor
at the scanning rollers.

– The setting of the levelling device has to be checked at regular intervals, even if no components have
been replaced or modified. This periodic checking is the only way to ensure the precise functioning of the
levelling device.
Note: The checking interval depends on the quality requirements, but its length must not exceed one
year.

– The levelling device must be checked with no material loaded and using the prescribed gauges.

– For the calibration of the autoleveller, use only


3 mm

B3 the original gauges supplied by Rieter (4 pcs.).


The gauge thicknesses are 3 mm, 4 mm, 5
mm and 6 mm.
Gauge set (4 thicknesses): 3 mm, 4 mm,
4 mm

B4
5 mm, 6 mm, Page 2/1, 4/1
5 mm

B5
6 mm

B6
Dxx 0.00

2 / 100 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

– Use the SW32 open-jaw spanner when


installing or removing the REG sensor
measuring head of the autoleveller.
Note: The SW32 open-jaw spanner is
included among the accessories supplied.
SW32 open-jaw spanner, Page 2/2, 4/1

108

IMPORTANT

– The order in which the levelling, checking and setting tasks are carried out must be adhered to and all
tasks must be completed.

1. Check the scanning roller voltage


– The scanning-roller voltage has to be checked, in order to determine whether a scanning-roller calibra-
tion procedure is required.

– Remove any material in the machine.

– Deactivate the Q1 motor protection switch in the control cabinet.

– Select data screen 4.1.2 (“Scanning-roller calibration”).


4.1.2, Page 2/206

– Relieve the load on the scanning rollers (2) by


moving the S4 ROTARY switch to position “I”.
Note: If the width between the scanning rollers
is 4.0 mm, the top side part of the scanning
6 roller (14) and the entire dust strip (6) have
to be removed prior to checking the scanning
roller voltage, and subsequently re-installed.
(see Part 6)

– Insert the 6 mm gauge between the scanning


rollers (2).
2 2
6 mm – Push the scanning rollers together by hand
14 and subject them to load by turning the S4
ROTARY switch to position “II”.
Dxx 2.30

– The actual measurement for the scanning-roller voltage with an inserted 6-mm gauge is now displayed
as the value for the “Scanning roller voltage” parameter in data screen 4.1.2 “Scanning-roller calibration”.

– For the checking procedure, the actual measured value for the scanning roller voltage must lie within the
specified range.

– If this requirement is met, the 6-mm gauge must be removed and the Q1 motor protection switch in the
control cabinet reactivated.
– If this requirement is not met, carry out a scanning-roller calibration procedure.

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 101


Operation
Operation

2. Scanning-roller calibration procedure


– The scanning-roller calibration procedure consists of two steps.
Note: If the width between the scanning rollers is 4.0 mm, the top side part of the scanning roller and the
entire dust strip have to be removed prior to carrying out the scanning-roller calibration procedure, and
subsequently re-installed. (See Part 6)

Step 1:
– Turn the key of the S17 key switch to the right.
The hood limit switch is bypassed.

– Open the hood on the left.

– Insert the 6-mm gauge between the scanning


rollers.
– Mark the field for “Start calibration” in data
screen 4.1.2 (“Scanning-roller calibration”).
Scanning-roller calibration is started.

Dxx 6.15

– Loosen clamping screw (5).

– Turn the target (4) (after inserting an Allen key


in the hexagonal recess (7)) until the scanning-
roller voltage lies within the prescribed range
(upper/lower limit: 8125 mV ±15 mV).

– Tighten clamping screw (5).

5 – In data screen 4.1.2 (“Scanning-roller calibra-


tion”), select the gauge dimension of 6 mm by
7 marking the “6 mm” field.
4 The actual value corresponding to the
3 6-mm gauge is saved.
Dxx 2.30

– Relieve the load on the scanning rollers and then subject them to load once again with the 6-mm gauge
inserted.
Note: The scanning roller voltage must still lie within the specified range. If this is not the case, then the
scanning roller voltage must be adjusted sufficiently so that it complies with the requirements.

– In analogous fashion, insert the 3-mm, 4-mm and 5-mm gauges, one after another, and mark the field for
the respective parameter each time.
The current values are saved.

– Remove the last gauge to be inserted.

2 / 102 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

– Mark the field for “Finish calibration” in data screen 4.1.2 (“Scanning-roller calibration”).
The entries are saved and the scanning-roller calibration procedure is ended.

Step 2:
– Check the red C2 LED at the D295 processor.
D295 – Option 1: The red C2 LED is not illuminated.
The scanning-roller calibration is OK.

– Re-activate the Q1 motor protection switch in


C2 the control cabinet.
III 10 538 541

– Close the hood on the left.


– Turn the key of the S17 key switch to the left
and then remove the key.
Dxx 6-40-2
The bridging of the limit switch is now
disabled.

– For information on other procedures Start-up and automatic adaptation of the autoleveller. (See descrip-
tions “5. Start-up and automatic adaptation of the autoleveller”) and "6. Activating RQM monitoring func-
tion")

– Option 2: The red C2 LED is illuminated.


There is either a fault or a defect involved.

3. Fault recognition at the D295 processor


– The following information (in Sections 3.1 to 3.3 below) is relevant to troubleshooting.

3.1 C2 LED lit – Scanning-roller calibration not OK


– For each fault present, a warning is displayed briefly (for about 5 seconds).

– The following types of fault are possible:


Offset fault
Amplification fault
Linearity fault

– Repeat the scanning-roller calibration procedure. (See description "2. Execution of the scanning-roller
calibration")

– If the scanning rollers are still not correctly calibrated, the fitting of the ATL sensor measuring head in the
scanning unit has to be checked and if necessary the head re-installed. (See description "4. Installation
of the ATL measuring head")

3.2 C2 LED is lit – REG initialisation required


– If ATL parameter values are illogical or incorrect (e.g. the value for the levelling intensity), then ATL ini-
tialisation has to be performed using data screen 4.3.2 “Initialisations”.
Note: Here, the customer-specific values in data screens 2.2 (“Levelling”) and 5.1 (“Scanning-roller ad-
aptation”) will be overwritten with default values.
4.3.2, Page 2/210

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 103


Operation
Operation

3.3 Rapid flashing of C2 and C3 LEDs


– There is a major fault at the B195 sensor unit, e.g. the temperature of the ATL sensor measuring head or
the RQM sensor measuring head is greater than 99.9 °C or one of the sensor measuring heads is defect-
ive or not connected.

IMPORTANT
63 65 64
61
62

66 – Always replace the entire B195 sensor unit


(60), even if only a part of the unit (60) is de-
fective.
– The number (67) on the ATL sensor cable
must agree with the REG number on the sign
68 (66). Connector (64) must be fitted at the red
“Autolev.” connection point (X4) of the D295
processor.
60 67
Dxx 4.60 – The number (68) on the RQM-sensor cable
must agree with the RQM number on the sign
(66). Connector (65) must be fitted at the blue
60 B195 sensor unit “Q. Mon.” connection point (X6) of the D295
B195 sensor unit sub-assemblies processor.
61 REG sensor-measuring head
62 RQM sensor-measuring head
63 B195 sensor electronics

4. Installation of the ATL measuring head


– If the scanning rollers are still not correctly calibrated (after the second attempt), the fitting of the ATL
sensor measuring head in the scanning unit has to be checked and if necessary the head re-installed.

– Check whether the ATL sensor measuring


3 56 54 head (3) is precisely fitted in the retainer (55)
55 at the scanning unit (59).

57
– Here, the ATL sensor measuring head (3) with
the rear flat (53) and the lateral end face has to
be pressed as far as the limit stop (54) inside
the retainer (55).

– The green dot (56) on the ATL sensor measur-


ing head (3) must be visible at the top.
58
53 – The disc (57) of the ATL sensor measuring
59 head (3) must be correctly fitted in the recess
Dxx 2.30 outside of retainer (55).

– The plastic nut (58) must be correctly screwed


to the disc (57).

2 / 104 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

IMPORTANT

– After installing the sensor measuring head (3), the scanning-roller calibration procedure has to be
repeated. (See description "2. Execution of the scanning-roller calibration")

5. Start-up and automatic adaptation of the autoleveller


– Deactivate the RQM monitoring functions for:
A%: In data screen "2.3.1 Settings A%"
CV%: In data screen "2.3.2 Settings
CV%"
Spectrogram: In data screen
"2.3.3 Settings spectrum"
Slubs: In data screen "2.3.4 Settings
slubs"

– Deactivate the autoleveller by removing the “Autoleveller ON” check mark in data screen 2.2
(“Levelling”).
Note: The input field for activating or deactivating the levelling function must be pressed for at least two
seconds.
– Start the machine (with fed material).

– Carry out the automatic adaptation with the machine running; to this end, select “Yes” for “Adaptation of
autoleveller” in data screen 5.1.
At the A% display for the incoming sliver on the overview screen, the green LED will light up.

– Re-activate the autoleveller by marking the “Autoleveller ON” field in data screen 2.2 (“Levelling”).
Note: The input field for activating or deactivating the levelling function must be pressed for at least two
seconds.
– With the autoleveller active, produce a piece of sliver and determine the sliver weight in the laboratory.

– In data screen 5.1 (“Scanning-roller adaptation”), enter the determined sliver weight as the value for “Ac-
tual sliver weight”.

– Check the sliver weight once more and, if required, make corrective adjustments.

6. Activating RQM monitoring functions


– If the sliver weight is OK, automatic adapta-
tion should be carried out via data screen 5.2
(“Delivery-disc adaptation”) after approx. 2000
m of sliver has been produced.
The green LED at the A% display for
incoming sliver (on the overview screen)
will light up.

– Produce a sufficient amount of sliver (more


than 100 m) and determine the CV% value in
the laboratory.

– The CV% value determined in the laboratory has to be entered as the value for “Lab” in data screen 2.3.2
(“CV% settings”).

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 105


Operation
Operation

– After about 3000 m of sliver has been produced, activate the RQM monitoring functions in data screens
2.3.1 (“A% settings”), 2.3.2 (“CV% settings”), 2.3.3 (“Spectrogram settings”) and 2.3.4 (“Slub settings”).

3.9 Rieter Quality Monitor (RQM)


3.9.1 Adapting the Rieter Quality Monitor (RQM) to the weight of the de-
livered sliver
Rieter Quality Monitor values, Page 3/29
The Rieter Quality Monitor (RQM) monitors the quality of the delivered sliver at the delivery discs. The RQM
must be adapted to the weight of the delivered sliver.
Preparatory work
– The function for adapting the RQM to the sliver count at the delivery end is activated in data screen 5.2
(“Delivery-disc adaptation”) and involves the measurement of a certain length of delivered sliver.

Note: This adjustment is normally only required for a changeover of assortment, after the levelling has been
adapted (possibly including fine correction of the weight of delivered sliver) or following a calibration of the
RQM (according to setting instructions).
Requirements for the functioning of the RQM adaptation process
– The monitoring functions of the Rieter Quality Monitor (RQM) are deactivated (appropriate setting for
“Shutdown type” parameter in data screen 2.3.1 (“A% settings”) and for “Shutdown ON” parameter in
data screen 2.3.2 (“CV% settings”)).

– The delivery discs are being subjected to load.

– The autoleveller has been adapted.

– The autoleveller of the draw frame is active.


2.2, Page 2/182

– Any necessary fine correction work has been carried out.


Checking and correcting the weight of the delivered sliver (fine correction), Page 2/83

– Activate the RQM sliver-monitor adaptation function when the machine is running with material.
5.2, Page 2/216

– The value displayed for “Delivery-disc clearance for 0 % error” in data screen 5.2 (“Delivery-disc adapta-
tion”) lies within the valid range.

RQM adaptation procedure


– Carry out the RQM adaptation procedure; to this end, check the field for “RQM adaptation (sliver weight
OK)” in data screen 5.2 (“Delivery-disc adaptation”).
The mean thickness of the delivered sliver is determined.
This value is used to determine the delivery-disc clearance for “ideal” delivered sliver (i.e. where
there is 0 % error or, in other words, no sliver weight deviations) and the calculated value is then
saved.
When the RQM adaptation procedure is complete, the “RQM adaptation sliver weight OK” check
mark will automatically be removed.
The optimal value calculated for the delivery-disc clearance is then displayed as the setting for
“Delivery-disc clearance for 0 % error” in data screen 5.2 (“Delivery-disc adaptation”).

2 / 106 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

A% display of the delivered sliver


– The A% display for the delivered sliver is an
LED display at the overview screen.
For the ideal value of “0”, the bar in the
middle is lit green.
For all sliver weight deviations (both
positive and negative) from the ideal
value, the corresponding red bar from the
“+” range or the “-” range is lit.
The bars on the far left and right
correspond to deviations of ±5 %
respectively.

Checking the position for the adaptation within the possible setting range
– The optimal value calculated for the delivery-disc clearance is displayed as the setting for “Delivery-disc
clearance for 0 % error” in data screen 5.2 (“Delivery-disc adaptation”). This value is determined when
measuring the delivered sliver for RQM adaptation.

– With this value, the position for the adaptation within the possible setting range is indicated at the same
time.

Special case: Delivery disc clearance at 0% error is a limit value


– Indicated distance = 0.278 mm (lower limit value)
Install the sliver funnel of the next size down.

– Indicated distance = 1.266 mm (upper limit value)


Install the sliver funnel of the next size up.

– If another sliver funnel has been installed, the RQM has to be adapted anew.

– The value for “Delivery-disc clearance for 0 % error” then has to be checked to see if it matches.

– If the displayed value matches, produce a sufficient amount of sliver (more than 100 m) and determine
the CV% value in the laboratory.

– Enter the CV% value determined in the laboratory as the value for “Lab” in data screen 2.3.2 (“CV% set-
tings”). (See the following section for further information.)

– Then re-activate the RQM monitoring functions.

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 107


Operation
Operation

3.9.2 Rieter Quality Monitor (RQM) – setting instructions


Coiler, Page 3/21
Machine control system, Page 3/26
Rieter Quality Monitor (RQM) – calibration of the equipment, Page 3/30
DANGER
If the main switch is set to the “on” position, the machine can be started up.
When performing checking or setting tasks at the Rieter Quality Monitor (RQM), the safety in-
structions must be adhered to.
Set the main switch (Q10) to the “off” position and secure it with a lock.
If the standard mechanical settings for the Rieter Quality Monitor (RQM) have to be checked or
made, the main switch (Q10) must be in the “off” position and secured with a padlock.

The RQM monitors the quality of the delivered sliver. Prior to delivery, each machine is checked and set ac-
cording to instruction. Periodic checking of the RQM must be carried out in accordance with the setting in-
structions.
Objective
– The mechanical and electrical inaccuracies are determined and the system set precisely.

General instructions on checking the Rieter Quality Monitor (RQM)


– In case of faults at the RQM, the plant electrician has to carry out a check in accordance with the setting
instructions.
– The setting of the RQM has to be checked and if necessary adjusted whenever changes are made:
at sliver deflection bars
at sliver guides of sliver deflection unit
at delivery discs
to load application at delivery discs
at toothed-belt drive unit
at displacement-transducer system (REG);
at target (RQM)
at the D295 processor
to sliver count at delivery end
to material being processed (changeover of assortment)

– The setting of the RQM has to be checked at regular intervals, even if no components have been
replaced or modified. This periodic checking is the only way to ensure the precise functioning of the
RQM.
Note: The checking interval depends on the quality requirements, but its length must not exceed one
year.

– The RQM must be checked with no material loaded and using the prescribed gauge.

2 / 108 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

– To set the RQM, use only the supplied gauge


indicated in the diagram (0.6-mm gauge), so
that precise measurements can be made.
Feeler gauge for calender: 0.6 mm in
thickness, Page 2/2, 4/1

0-00-2_100a

IMPORTANT

– The order in which the RQM checking and setting tasks are carried out must be adhered to and all tasks
must be completed.

1. Checking the delivery disc voltage


– The delivery-disc voltage has to be checked, in order to determine whether a delivery-disc calibration
procedure is required.

– Remove any material in the machine.

– Deactivate the Q1 motor protection switch in the control cabinet.

– Select data screen 4.1.3 (“Delivery-disc calib-


ration”) for this purpose.
4.1.3, Page 2/207

– Insert the 0.6-mm gauge between the delivery


discs.
– The actual measurement for the delivery-disc
voltage with an inserted 0.6-mm gauge is now
displayed as the value for the “Delivery-disc
voltage” parameter in data screen 4.1.3
(“Delivery-disc calibration”).
0-00-4_101e

– For the checking procedure, the actual meas-


ured value for the delivery-disc voltage must lie
within the specified range.

– If this requirement is met, the 0.6-mm gauge must be removed and the Q1 motor protection switch in the
control cabinet reactivated.
– If this requirement is not met, then a delivery-disc calibration procedure has to be carried out.

2. Delivery-disc calibration procedure


– The delivery-disc calibration procedure consists of two steps.

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 109


Operation
Operation

Step 1:
– Insert the 0.6 mm gauge between the delivery
discs (5), (6).
48
– Mark the field for “Start calibration” in data
screen 4.1.3 (“Delivery-disc calibration”).
38 Delivery-disc calibration is started.

– Loosen screw (48).


SW4
– Turn the target (38) with an SW4 Allen key
6
5 until the delivery-disc voltage lies within the
prescribed range (upper/lower limit: 2500 mV
±15 mV).

Dxx 4.60 – Tighten the screw (48).

– In data screen 4.1.3 (“Delivery-disc calibration”), select the gauge dimension of 0.6 mm by marking the
“0.6 mm” field.
The actual value corresponding to the 0.6-mm gauge is saved.

– Relieve the load on the scanning discs and then subject them to load once again with the 0.6-mm gauge
inserted.
Note: The delivery-disc voltage must still lie within the specified range. If this is not the case, then the
delivery-disc voltage must be adjusted sufficiently so that it complies with the requirements.

– Remove the 0.6-mm gauge.

– Mark the field for “End calibration” in data screen 4.1.3 (“Delivery-disc calibration”).
The entries are saved and the delivery-disc calibration procedure is ended.

Step 2:
– Check the red C4 LED at the D295 processor.
D295 – Option 1: The red C4 LED is not illuminated.
The delivery-disc calibration is OK.

– Re-activate the Q1 motor protection switch in


the control cabinet.
III 10 538 541

C4 – For information on other procedures Start up


and automatic adaptation of the RQM. (See
description " 7. Start-up and automatic adapta-
tion of the RQM"
Dxx 6-4 0-2

– Option 2: The red C4 LED is illuminated.


There is either a fault or a defect
involved.

3. Fault detection at D295 processor


– The following information (in Sections 3.1 to 3.3 below) is relevant to troubleshooting.

3.1 C4 LED lit – Delivery-disc calibration not OK


– For the fault present, a warning is displayed.

2 / 110 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

– The following type of fault is possible:


Offset fault
Note: An offset fault occurs when the current delivery-disc voltage for an inserted 0.6-mm gauge lies out-
side the prescribed range (upper/lower limit: 2500 mV ±15 mV).

– Repeat the delivery-disc calibration procedure. (See description "2. Execution of the delivery disc calib-
ration")

– If the delivery discs are still not correctly calibrated, various standard mechanical settings have to be
checked and adjustments made where necessary. (See descriptions “4. Basic mechanical settings for
the Rieter Quality Monitor (RQM)” to "6. Adjusting the stop screw for the safety clearance of the delivery
discs")

3.2 C4 LED is lit – RQM initialisation required


– If RQM parameter values are illogical or incorrect (e.g. the shut-off limit value for spectrogram monitor-
ing), then RQM initialisation has to be performed using data screen 4.3.2 (“Initialisations”).
Note: Here, the customer-specific values for RQM parameters in data screens 2.3.1 to 2.3.4 and in data
screen 5.2 (“Delivery-disc adaptation”) will be overwritten with default values.
4.3.2, Page 2/210

3.3 Rapid flashing of C2 and C3 LEDs


– There is a major fault at the B195 sensor unit, e.g. the temperature of the REG sensor measuring head
or the RQM sensor measuring head is greater than 99.9 °C or one of the sensor measuring heads is de-
fective or not connected.

IMPORTANT
63 65 64
61
62

66 – Always replace the entire B195 sensor unit


(60), even if only a part of the unit (60) is de-
fective.
– The number (67) on the REG-sensor cable
must agree with the REG number on the sign
68 (66). Connector (64) must be fitted at the red
“Autolev.” connection point (X4) of the D295
processor.
60 67
Dxx 4.60 – The number (68) on the RQM-sensor cable
must agree with the RQM number on the sign
(66). Connector (65) must be fitted at the blue
60 B195 sensor unit “Q. Mon.” connection point (X6) of the D295
B195 sensor unit sub-assemblies processor.
61 REG sensor-measuring head
62 RQM sensor-measuring head
63 B195 sensor electronics

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 111


Operation
Operation

4. Standard mechanical settings for Rieter Quality Monitor (RQM)


X – The required delivery-disc load corresponds to
a spring length of 16 mm (320 N)
Note: See information on "Setting the delivery-
disc load" in Section 6

Dxx-4-60

– Set the preliminary tension of the double


toothed belt
Note: see “Installing the double toothed belt at
the calender-roll bearing” (Part 6)

Dxx-4-60a

5. Check of standard mechanical settings for RQM sensor measuring head


– If the delivery discs are still not correctly calibrated (after the second attempt), the standard setting of the
RQM sensor measuring head has to be checked and if necessary the head re-fitted.

– Swing down the holder (with the web nozzle


29 and insert) and remove it.
4
– Swing the swivel plate (30) towards the front.

– Remove the sliver funnel (3) from the swivel


plate (30).

– Remove the swivel plate (30): To do this, un-


screw the screws (29) on the bearing saddle
3 (4).

30
Dxx 4.55

2 / 112 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Operation

– Apply load to the delivery discs (5), (6) by lock-


ing the movable calender-roll bearing (18).

8 – Set the delivery discs (5), (6) so that there is


direct contact:
19
18 – Loosen nut (19).

– Screw out the stop screw (8), so that it is at


6 least 1 mm away from the movable calender-
5 roll bearing (18).

Dxx 4.60
0 mm

– Release the movable calender-roll bearing.


47
– Loosen screw (33).

– Move the RQM sensor measuring head (9) to-


wards the target (38) and slightly above the
33 holder (47).

48 – Lock the movable calender-roll bearing.


9
– Loosen screw (48).
38

Dxx 4.60

– Using the AF4 Allen key, turn the target (38)


47 (with this moving towards the RQM sensor
33 SW4 measuring head (9)) sufficiently, so that a
discernible gap (about 0.15 mm in width) forms
between the delivery discs (5), (6) (subjected
9 38 to load).
Z Note: The gap is easy to spot by placing a
6 white sheet of paper over the opening of the
sliver duct. This will also prevent gauges from
falling through the gap.

Z – The aim is to displace the RQM sensor meas-


5 uring head (9) until it is flush with the holder
Dxx 4.60
(47).

– Tighten the screw (33).

IMPORTANT

– The delivery-disc calibration procedure subsequently has to be repeated. (See description "2. Execution
of the delivery disc calibration")

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 113


Operation
Operation

6. Setting of stop screw for safety clearance of delivery discs


– The safety clearance for the delivery discs has
to be set so that damage to the delivery discs
8 is avoided if the machine runs without material.
– Insert the 0.15 mm gauge (7) (from the feeler
gauge set) between the delivery discs (5), (6).
19
Set of parallel feeler gauges (0.05 mm ...
18 1.0 mm), Page 2/2, 4/2, 6/36

– Screw out the stop screw (8) until it firmly


6
5 presses against the movable calender-roll
bearing (18), so that the 0.15 mm gauge (7)
can be easily removed.
7
Dxx 4.60
0,15 mm – Tighten nut (19).

– Remove the 0.15 mm gauge (7) from between


the delivery discs.

– Mount the swivel plate and insert the sliver fun-


nel.
– Attach the holder (plus the web nozzle and in-
sert) and swivel it towards the drafting system.

7. Start up and automatic adaptation of the RQM


– Deactivate the RQM monitoring functions for:
A%: In data screen "2.3.1 Settings A%"
CV%: In data screen "2.3.2 Settings
CV%"
Spectrogram: In data screen
"2.3.3 Settings spectrum"
Slubs: In data screen "2.3.4 Settings
slubs"

– Start the machine (with fed material).

– Produce a piece of sliver and determine the sliver weight in the laboratory; make corrective adjustments,
where necessary.

– If the sliver weight is OK, automatic adaptation should be carried out via data screen 5.2 (“Delivery-disc
adaptation”) after approx. 2000 m of sliver has been produced.
The green LED at the A% display for incoming sliver (on the overview screen) will light up.

– Produce a sufficient amount of sliver (more than 100 m) and determine the CV% value in the laboratory.

– The CV% value determined in the laboratory has to be entered as the value for “Lab” in data screen 2.3.2
(“CV% settings”).

– After about 3000 m of sliver has been produced, activate the RQM monitoring functions in data screens
2.3.1 (“A% settings”), 2.3.2 (“CV% settings”), 2.3.3 (“Spectrogram settings”) and 2.3.4 (“Slub settings”).

2 / 114 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Display screens Operator interface panel

4.7 Menu 4: "Service"

4 Service
Main menu 4 "Service" is displayed.
(1) Button for main menu 4 (“Service”)

(2) Second menu level


4.1 Maintenance
4.2 Tests
4.3 Diagnostics
4.4 Data export / import

4.1 Maintenance
4.1.1 Servo calibration
4.1.2 Scanning-roller calibration
4.1.3 Delivery-disc calibration
4.1.4 Lubrication

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 205


Operation
Display screens Operator interface panel

4.1.2 Scanning roller calibration


Autoleveller (ATL) – setting instructions, Page 2/100
This data screen can be brought up only with the machine in “service mode”.
(1) Start of calibration
If the box is ticked, scanning-roller calibration is started.
There must be no faults pending.

(2) End of calibration


If the box is ticked, scanning-roller calibration is ended.

(3) Selection of gauge sizes


When ticked, the current scanning-roller voltage (3) for inserted check gauges (3 mm, 4 mm, 5 mm, 6 mm) is
saved.

(4) Current scanning-roller voltage (in mV) with the respective gauge inserted

(5) Scanning-roller voltage


Display of the value of the current voltage from the displacement-transducer system (REG) for the inserted
distance gauge, with scanning rollers subjected to load
To check the scanning-roller calibration, the current scanning-roller voltage is compared with the stored
voltage for the respective gauge. The deviation must not exceed ±30 mV.
Default : 8125 mV
min. - max.: 8095 - 8155 mV

2 / 206 Draw Frame RSB-D 50 13.09.2016 - en - V01


Operation
Display screens Operator interface panel

4.1.3 Delivery disc calibration


Rieter Quality Monitor (RQM) – setting instructions, Page 2/108
This data screen can be brought up only with the machine in “service mode”.
(1) Start of calibration
If the box is ticked, delivery-disc calibration is started.
There must be no faults pending.

(2) Selection of gauge size


When ticked, the current delivery-disc voltage (3) for the inserted test gauge (0.6 mm) is saved.

(3) Current delivery-disc voltage (in mV) with the respective gauge inserted

(4) End of calibration


If the field is marked, delivery-disc calibration is ended.
Attention:
If application of the voltage is stopped prior to the end of the back-up cycle (duration of approx. 2 seconds),
the voltage values for the gauges will not be permanently saved. The old values will be re-applied after the
main switch has been moved to the “off” position.

(5) Delivery-disc voltage


Display of the value of the current voltage from the displacement-transducer system (REG) for the inserted
distance gauge, with delivery discs subjected to load
To check the delivery-disc calibration, the current delivery-disc voltage is compared with the stored voltage for
the respective gauge. The deviation must not exceed ±50 mV.
Default : 2500 mV
min. - max.: 2450 - 2550 mV

13.09.2016 - en - V01 Draw Frame RSB-D 50 2 / 207


Operation

You might also like