PROJECT: 1 X 660 MW BHUSAWAL TPS UNIT-6 TPP

PACKAGE: SEWAGE TREATMENT PLANT


SUB CONTRACTOR: EARTH WATER TECHNOLOGIES

DRAWING / DOCUMENT COMMENTS REVIEW

DOCUMENT TITLE: DATASHEET OF ANALYSERS ( EQMS) FOR SEWAGE TREATMENT PLANT

BHEL DOC NO : PE-V0-415-673-A203 RC


R02 EWT DOC NO : A4-720-12-34 RC
R02

DOC. DATE: 05.10.2021 REVIEW DATE:19.05.2022

Ref. No. CRS DATE: 03.06.2022

SR. NO. BHEL / DCPL Comments EWT Reply

1 pH Sensor Recharger to be added as per contract Ok. Noted & Added.


DATASHEET OF ANALYSERS

EARTH WATER TECHNOLOGIES

R2 03/06/2022 SNG ARG SSG 5TH SUBMISSION

R1 21/04/2022 SNG ARG SSG 4TH SUBMISSION

RB 18/01/2022 SNG ARG SSG 3RD SUBMISSION

RA 30/12/2021 SNG ARG SSG 2ND SUBMISSION

R0 09/09/2021 SNG ARG SSG 1ST SUBMISSION

REV DATE PREPARED BY CHECKED BY APPROVED BY NOTE

PACKAGE: SEWAGE TREATMENT PLANT


PROJECT: 1 X 660 MW BHUSAWAL THERMAL POWER PROJECT
OWNER:

MAHARASHTRA POWER GENERATION CO. LTD


(MAHAGENCO )

CONSULTANT:
DEVELOPMENT CONSULTANTS PVT. LTD
CONSULTANT ENGINEERS
VASHI, NAVI MUMBAI

MAIN CONTRACTOR BHEL Doc. No.

BHARAT HEAVY ELECTRICALS LIMITED PE-V0-415-673-A203


POWER SECTOR PROJECT Rev 2
ENGINEERING MANAGEMENT, NOIDA

SUB CONTRACTOR EWT Doc. No.

A4-720-12-34
EARTH WATER TECHNOLOGIES
MUMBAI Rev 2

DOCUMENT TITLE: DATASHEET OF ANALYSERS


.. current
{
SI. Project Model Installation Site Part No Installed at ref Year of Ye■r of Location-
Parameter Make
Commlsslonln1
warranty Remartcs
current
� No Reference site Installation
1 .....ect 9
NTPC Singrauli Super The part no is different due to j
Thermal Power Station configuration and range
PC1R1A+ SC200 - HACH P.O. SHAKTINAGAR, LXV404.99.00102 & 6058P4 Sep-17 Nov-17 Completed EQMS

I
LXV404.99.01102 SONEBHADRA
Uttar Pradesh-231222, Ind a
NTPC Limited
( A Government of India ::r
Enterprise ) .;
The sensor part number is
PC1RlA-33T +
Natiom1I cap;tal Power
different due to cable length [
1 PH ANALVZER
PRO-P3A1N
HACH Project PClRIA+ PRO-P3AlN Jan-18 Feb-18 Completed §
P.O. VIDYUT NAGAR �
!
DADRI GAUTAM BUDH
NAGAR(U.P.)
Uttor Pradesh- 201008,
NTPC Tamilnadu Energy
PC1RlA-33T +
HACH
Company Ltd, Vallur Therm•!
PC!RlA+ PRO-P3AlN Feb-17 Mar-17 Completed The sensor part number is �
PRO-P3AlN Power Project different due to cable length
Chennai - 600103
PC1RlA-33T + NLC Tamil Nadu Power
HACH LXV404.99.00102 & PC1R1A Sep-16 Feb-17 Completed
PRO-P3AlN Limited INTPLI.
0
NTPC Singrouli Super The part no is different due to ...,
0
0
Thermal Power Station configuration and range
LXV404.99.00102 & EQMS
3422E3A+ HACH P.O. SHAKTINAGAR, Sep-17 Nov•l7 Completed
08310=A=OOOO
SC20O- SONEBHAORA
LXV404.99.01202 Uttar Pradesh-231222, lndi1

3422B3A/3422D3A NTPC Tamilnadu Energy


/3422E3A+ Company Ltd, Vallur Thermal
2 CONDUCTIVITY HACH 343388A + PROC3AlN Feb-17 Mar-17 Completed
PRO-C3A1N Power Project
Chennal - 600103
The part no is different due to
configuration and range
NLC Tamil Nadu Power LXV404.99.00102 & EQMS
3422E3A+ HACH Sep-16 Feb-17 Completed
Limited (NTPL), 3422AlA
SC20O-
LXV404.99.01202

NTPC Limited
( A Government of India
Enterprise ) Since the requirement is 2 wire
National Capital Power system, The controller Part
PRO-P31UN • LXV404.99.00102 &
HACH Project Jan-IS Feb-18 Completed number is different, The sensor
RCIRSN RClRSN
P.O. VIDYUT NAGAR part no is different due to ORP
, OJ
,JV:, OADRI GAUTAM BUOH Rance
NAGAR(U.P.) <n
Uttar Pradesh- 201008, lJ
ca·
3 ORP
NTPC Tamilnadu Enercy The controller Part number is
PRO·P3AIN • Company Ltd, Vallur Thermal Sensor: RClRSN,
HACH Feb-17 Mar-17 Completed same, The sensor part no is
RC!RSN Power Project Transmitter: PRO-P3AlN
different due to ORP Ran1 e
Chennal - 600103
Since the requirement is 2 wire
system, The controller Part
PRO-P3AIN + NLC Tamil Nadu Power LXV404.99.00102 &
HACH Sep-16 Feb-17 Completed number is different, The sen.sor ETP - 1
RClRSN Limited (NTPL), RClRSN
part no is different due to OR?
RanD"e
--
. .
CUrrent
SI. Project Model Installation Site Part No Installedatref Yurof Yurof Location·
P amneter Make Warranty Remarb
No No Reference st
i e Insta llation Commlsslonln1 CUrrent

SSOO.SO.AlU HACH NTPC, Korba, Chhattisgarh SOOO.SO.AlU May•l3 Jan-14 Completed


4 SILICA Kudgi Super
S500.SO.AlU HACH Thermal Power S000.SO.AlU Feb-1S Apr-15 Completed It Is the upgraded Model
Plant, Biiar,ur,Karnataka
RCA ( residual ch lorine
s analvzer I
5440005 HACH NTPC. Dadri,UP 5440005 Sep-16 Feb-17 Completed
LXV404.99.01502 •
(WATER CHEMISTRY The model is different due
SC 200 Thermal Powertecl, EQMS •
HACH ANALYZ) KTD; TUS300SC 2017 03-Aug-17 Completed to configuration and range
LXV44S.99.10122 Corporation India Ltd
W/SC200 1 CH (8626500)
TUS300SC
LXV44S.99.10112
The model is different due EQMS
TU5300SC NLC Tamil Nadu Power
6 TURBIDITY HACH SC200+ 1720 Sep-16 Feb-17 Completed to configuration and range
LXV404.99.01S02 Limited (NTPL),
SC200
LXV445.99.10112 •
TUS300SC The model is different due EQMS
HACH NTPC, Dadrl,UP SC200+ 1720 Sep-16 Feb-17 Completed to configuration and range
LXV404.99.00502•
SC200
LXV441.99.21202
FP360 +
LXV404.99.00502-
The model is different due
SC200
to configuration and range EQMS •
HACH IDCL Vadodara,Gujarat SC200• FP360 Mu-17 Apr-17 Completed
LXV441.99.11202
FP 360+ 0
LXV404.99.01S02 •
SC200
LXV441.99.21202 •
FP360+
LXV404.99.00502•
SC200 The model is different due
EQMS
7 OIW HACH IOCL Paradip SC200 • FP360 Mar-17 Apr·l7 Completed to configuration and range
LXV441.99.11202 •
FP 360 +
LXV404.99.01S02 •
SC200
LXV441.99.21202 •
' FP360 +
LXV404.99.00502
SC200 The model is different due
NLC Tamil Nadu Power EQMS
HACH SC200+ FP360 Sep-16 Feb-17 Completed to configuration and range
l.XV441.99.11202 • Limited (NTPL),
FP 360+
LXV404.99.01502
SC200
,.,.. LXV423.99.00100
DVC MEJIA THERMAL
Solitu+
HACH POWER STATIDN,WEST SC200 + SOlitax Feb-15 Apr·lS Completed EQMS
LXV404.99.01502
BENGAL
SC200
LXV423.99.00100
INDIAN OIL CORPORATION
Solitax • The model is different due
8 TSS HACH LIMITED, PANIPAT , SC200 + SOlitox Sep-16 Feb-17 Completed EQMS
LXV404.99.01502 • to configuration and range
HARYANA
SC200
LXV423.99.00100
Solitax+ D ELHI JAL BDARD,NANGOU, The model is different due
HACH SC200+ SOlitax Mar-17 Apr-17 Completed to configuration and range EQMS
LXV404.99.01S02 • OKHLA,NEW DELHI
SC200
current
SI. Project Model lnstallatlon Site Part No Installed at ref Ye■rof Ye■rof location-
Parameter Ma .. Warranty Remarb
No No Reference site lnstalatlon Commlsslonlnl CUrrent
Prolect
Kindly note that the detail is
L.XV418.99.S0002 not available as BOD/COD
UVAS+ NLC Tamil Nadu Power requirements are recently
HACH 83500 & 87000 Sep-16 Feb-17 Completed EQMS
L.XV404.99.01502- Limited (NTPL), added In Power Plant
SC200 Specifications and the plants
are fewer.
LXV418.99.50002 Model number is different
UVAS+ due to configuration and
HACH !Oct. Panipat, Haryana UVAS+ SC200 Oct-16 Jan-17 Completed EQMS )1
LXV404.99.01S02 - range but same series
9 COD/BOD
SC200
L.XV418.99.50002
Model number is different
UVAS+ due to configuration and
HACH BPCl. Kochi, kerala UVAS+ SC200 Sep-17 Feb-18 Completed EQM
LXV404.99.01502 - range but same series
SC200
LXV418.99.50002
Model number is different
UVAS+ Deihl Jal Board (Pappan due to configuration and
HACH UVAS+ SC200 Feb-1S Apr-15 Completed EQMS
LXV404.99.01502 Kalon STP, Kondli STP & range but same series
SC200 Vamuna Vihar STP)

;:-

0
....
0

OJ
(I)
lJ
cci'
INDEX

S.NO DESCRIPTION PAGE NO


2
1 EFFLUENT QUALITY MONITORING STATION BILL OF MATERIAL
3
2 DATA SHEET FOR EFFLUENT QUALITY MONITORING STATION
4 TO 10
2.1 DATA SHEET & CATALOUGE FOR Ph ANALYZER
11 TO 19
2.2 DATA SHEET & CATALOUGE FOR CONDUCTIVITY ANALYZER
20 TO 25
2.3 DATA SHEET & CATALOUGE FOR TURBIDITY ANALYZER
2.4 DATA SHEET & CATALOUGE FOR BOD/COD ANALYZER 26 TO 30

2.5 DATA SHEET & CATALOUGE FOR SC200 CONTROLLER 31 TO 37


BILL OF MATERIAL EFFLUENT QUALITY MONITORING STATION
BHUSAWAL – STP PROJECT

Sl Analyzer Model .NO Item Description MAKE Total Qty

PC1R1A pH Sensor HACH 01


01 PH ANALYZER
LXV404.99.01102 sc200 CONTROLLER, AC-DC, pH,MODBUS HACH 01

3422E3A Conductivity Sensor HACH 01


CONDUCTIVITY
02
ANALYZER LXV404.99.01202 sc200 CONTROLLER, AC-DC, COND, MODBUS HACH 01

LXV445.99.10112 Turbidity Sensor HACH 01


TURBIDITY
03
ANALYZER
LXV404.99.01502 sc200 CONTROLLER,AC-DC, DIG,MODBUS HACH 01

LXV418.99.50002 COD/BOD Sensor HACH 01

04 COD/BOD ANALYZER LXV404.99.01502 sc200 CONTROLLER,AC-DC, DIG,MODBUS HACH 01

9012800 SENSOR MODULE, 4-20 mA INPUT, SC 200 HACH 01


ANALYZER DATASHEET WITH CATALOGUE
DATASHEET FOR PH ANALAYZER

Make HACH
As per Annexure-1
Tag Number
General
Service As per Annexure-1
Process Medium As per Annexure-1
Instrument Range pH 0-14 PH
Calibration Range As per Annexure-1
Model No. PC1R1A
Sensor type Flow Through
Sensor End Connection Size 3/4" NPT(M)
Measuring Electrode/ Sensor MOC General Purpose Low resistivity GPLR pH Glass
Sensor Integral Cable Length 5 Mtrs sesnor to Transmitter
pH Sensor
Sensor Installation In Flow Cell
Temperature Compensation Auto
Temperature Electrode PT1000
Accuracy pH our accuracy is loop accuracy which is 0.1 pH .(which includes inaccuracy’s due
to Controller/transmitter, pH electrode, pH buffers (pH buffers inaccuracies varies
up to +/- 0.02pH-0.05pH) and cable as well. After calibration pH inaccuracy in pH
buffer will be <1% or <0.05 Ph )

Repeatability 0.1% of span or better


Stability 0.05 % or +/- .001 PH per week
Transmitter Model No. SC200
ANNEXURE-1
Tag No. SERVICE TYPE RANGE OPER DESIGN Pro Calib
ATIN RANGE cess ratio
G Me n
From To Temp Pres Tem Pres diu Range
°C sure p °C sure m
kg/c kg/c
10GQB16CQ005 Effluent, PH measurement Flow through 0 14 pH Amb 4 60ºC 5.75 Water 0- 14
3/4-inch Combination pH and ORP Sensor Kits

pH/ORP
Use the Digital Gateway to make any Hach analog Digital combination pH and ORP sensors are available
combination pH or ORP sensor compatible with the in convertible, insertion, and sanitary mounting
Hach sc1000 Controller. styles. Choose from rugged dome electrodes
or “easy-to-clean” flat glass electrodes.

Features and Benefits


Low Price—High Performance Versatile Mounting Styles
These combination sensors are designed for specialty Sensors are available in three mounting styles—convertible,
applications for immersion or in-line mounting. The reference insertion, and sanitary. Please turn to page 3 for more
cell features a double-junction design for extended service information.
life, and a built-in solution ground. The body is molded from
chemically-resistant Ryton® or PVDF, and the reference Full-Featured “Plug and Play”
junction is coaxial porous PTFE. All sensors are rated Hach sc Digital Controllers
0 to 105ºC up to 100 psig, and have integral 4.5 m (15 ft.) There are no complicated wiring or set up procedures
cables with tinned leads. The PC-series (for pH) and RC- with any Hach sc controller. Just plug in any combination
series (for ORP) combination sensors are ideal for measuring of Hach digital sensors and it’s ready to use—
mild and aggressive media. it’s “plug and play.”

Special Electrode Configurations One or multiple sensors—The sc controller family allows


you to receive data from up to eight Hach digital sensors
Sensors with rugged dome electrodes, “easy-to-clean”
in any combination using a single controller.
flat glass electrodes, and even HF (hydrofluoric acid)
resistant glass electrodes are available for a wide variety Communications—Multiple alarm/control schemes are
of process solutions. available using the relays and PID control outputs. Available
communications include analog 4-20 mA, digital MODBUS®
Temperature Compensation Element Option (RS485 and RS232) or Profibus DP protocols. (Other digital
The PC-series combination pH sensors are available protocols are available. Contact your Hach representative
with or without a Pt 1000 ohm RTD temperature element. for details.)
The RC-series combination ORP sensors are supplied
Data logger—A built-in data logger collects measurement
without a temperature element.
data, calibration, verification points, and alarm history.
2

Specifications*
Most pH applications fall in the 2.5-12.5 pH range. General purpose pH glass electrodes perform well in this range. Some
industrial applications require accurate measurements and control at pH values below 2 or above 12. Consult Hach Technical
Support for details on these applications.

Combination pH Sensors Combination ORP Sensors


Measuring Range Measuring Range
0 to 14 pH -2000 to +2000 millivolts

Accuracy Accuracy
Less than 0.1 pH under reference conditions Limited to calibration solution accuracy (± 20 mV)

Temperature Range Temperature Range


0 to 105ºC (32 to 221ºF) 0 to 105ºC (32 to 221ºF)

Flow Rate Flow Rate


0 to 2 m/s (0 to 6.6 ft./s); non-abrasive 0 to 2 m/s (0 to 6.6 ft./s); non-abrasive

Pressure Range Pressure Range


0 to 6.9 bar at 100ºC (0 to 100 psig at 212ºF) 0 to 6.9 bar at 100ºC (0 to 100 psig at 212ºF)

Signal Transmission Distance Signal Transmission Distance


100 m (328 ft.) when used with the Hach Digital Gateway 100 m (328 ft.) when used with the Hach Digital Gateway
and a Hach sc Digital Controller. and a Hach sc Digital Controller.
1000 m (3280 ft.) when used with the Hach Digital Gateway, 1000 m (3280 ft.) when used with the Hach Digital Gateway,
Termination Box, and a Hach sc Digital Controller. Termination Box, and a Hach sc Digital Controller.

Sensor Cable Sensor Cable


Integral coaxial cable (plus two conductors for temperature Integral coaxial cable; 4.5 m (15 ft.) long; terminated with
compensator option); 4.5 m (15 ft.) long stripped and tinned wires

Wetted Materials Wetted Materials


Convertible style: Ryton® body (glass filled) Convertible style: Ryton® body (glass filled)
Insertion style: PVDF body (Kynar®) Insertion style: PVDF body (Kynar®)
Sanitary style: 316 stainless steel sleeved PVDF body Common materials for all sensor styles include PTFE double
Common materials for all sensor styles include PTFE double junction, glass with platinum process electrode, and Viton® O-
rings
junction, glass process electrode, and Viton® O-rings
Warranty
Warranty
90 days
90 days

*Specifications subject to change without notice.

Ryton® is a registered trademark of Phillips 66 Co.; Viton® is a registered trademark of E.I. DuPont de Nemours + Co.;
Kynar® is a registered trademark of Pennwalt Corp.

Engineering Specifications
1. The pH sensor shall be available in convertible, insertion 3. The convertible style pH sensor shall be available with
or sanitary styles. The ORP sensor shall be available in or without a built-in Pt 1000 ohm RTD temperature
only convertible or insertion styles. element. Insertion and sanitary style pH sensors shall
have a built-in Pt 1000 ohm RTD temperature element.
2. The convertible style sensor shall have a Ryton® body. Convertible and insertion style ORP sensors shall not
The insertion style sensor shall have a PVDF body. The have a built-in temperature element.
sanitary style sensor shall have a 316 stainless steel
sleeved PVDF body. Common materials for all sensor 4. The sensor shall communicate via MODBUS® RS-485
styles shall include a PTFE double junction, and Viton® to a Hach sc Digital Controller.
O-rings. The pH sensor shall have a glass pH electrode.
5. The sensor shall be Hach Company Model PC sc or
The ORP sensor shall have a platinum ORP electrode.
PC-series for pH measurement or Model PC sc or
RC-series for ORP measurement.
3

Dimensions
Convertible Style Sensor
The convertible style sensor has a Ryton® body that features 3/4-inch NPT threads on both ends. The sensor can be directly
mounted into a standard 3/4-inch pipe tee for flow-through mounting or fastened onto the end of a pipe for immersion mounting.
The convertible style sensor enables inventory consolidation, thereby reducing associated costs. Mounting tees and immersion
mounting hardware are offered in a variety of materials to suit application requirements.

Insertion Style Sensor


Insertion style sensors feature a longer, non-threaded PVDF body with two Viton® O-rings, providing a seal when used with the
optional Hach insertion mount hardware assembly. This ball valve hardware enables sensor insertion and retraction from a pipe
or vessel without having to stop the process flow.

Sanitary Style Sensor


The sanitary style sensor, offered for pH measurement, has a 316 stainless steel-sleeved PVDF body with a 2-inch flange. The
sensor mates to a standard 2-inch Tri-Clover fitting. The optional Hach sanitary mounting hardware includes a standard 2-inch
sanitary tee, sanitary clamp, and Viton® sanitary gasket.

Convertible Style with Dome Electrode and Built-in Protector


3/4-inch NPT
Typical
22.2 mm (0.875 inch)

22.9 mm (0.90 inch) 7.62 mm


(1.30 inches)
149.9 mm (5.9 inches)

Convertible Style with Flat Electrode

3/4-inch NPT
Typical
22.2 mm (0.875 inch)

22.9 mm (0.90 inch) 7.62 mm


(1.30 inches)
149.9 mm (5.9 inches)
4

Dimensions continued

Insertion Style with Dome Electrode and Built-In Protector

22.2 mm (0.875 inches)

149 9 mm (5.9 inches)

Sanitary Style

22.2 mm (0.875 inches) 49.1 mm (1.96 inches)

22 mm (0.87 inches)

88.9 mm (3.5 inches)

187.3 mm (7.37 inches)


5

Ordering Information
Analog PC and RC 3/4-inch Combination pH/ORP Sensors
All PC and RC 3/4-inch combination sensors come with an integral 4.5 m (15 ft.) standard length sensor cable.

Product Number Measurement Sensor Style Body Material Electrode Type Temp. Comp.
PC1R1N pH Convertible Ryton General purpose glass None
PC1R1A pH Convertible Ryton General purpose glass Pt 1000 ohm RTD
PC1R2N pH Convertible Ryton Flat glass, general purpose None
PC1R2A pH Convertible Ryton Flat glass, general purpose Pt 1000 ohm RTD
PC1R3A pH Convertible Ryton HF-resistant glass Pt 1000 ohm RTD
PC2K1A pH Insertion PVDF General purpose glass Pt 1000 ohm RTD
PC2K2A pH Insertion PVDF Flat Glass Pt 1000 ohm RTD
PC3K2A pH Sanitary 316 SS/PVDF General purpose glass Pt 1000 ohm RTD
RC1R5N ORP Convertible Ryton Platinum None
RC2K5N ORP Insertion PVDF Platinum None

Accessories for Digital and Analog 3/4-inch combination pH/ORP Sensors


Cables
Digital cables are used only with digital sensors or gateways when connecting to a Hach sc Digital Controller.
6122400 Digital Extension Cable, 1 m (3.3 ft)
5796000 Digital Extension Cable, 7.7 m (25 ft)
5796100 Digital Extension Cable, 15 m (50 ft)
5796200 Digital Extension Cable, 31 m (100 ft)
Analog cables are used only with analog sensors, junction box, and controller.
1W1100 Analog Interconnect Cable (order per foot)

Digital Termination Box


Used with digital extension cables when the desired cable length between the digital sensor/digital gateway and the
Hach sc Digital Controller is between 100 m (328 ft) and 1000 m (3280 ft).
5867000 Digital Termination Box

Analog Junction Box


Used with analog interconnect cable when the desired cable length between analog sensor and analog controller is greater
than the standard length of sensor cable. Each junction box includes terminal strip and gasket.
60A2053 Junction Box, Surface-mount, aluminum (includes mounting hardware)
60A9944 Junction Box, Pipe-mount, PVC, for 1/2-inch diameter pipe (includes mounting hardware)
60G2052 Junction Box, Pipe-mount, PVC, for 1-inch diameter pipe (includes mounting hardware)
76A4010-001 Junction Box, NEMA 4X (no mounting hardware included)
6

Ordering Information continued

1. Sanitary Mounting 2. Immersion Mounting 3. Flow-through Mounting 4. Insertion Mounting

Mounting Hardware for PC sc and RC sc Combination Sensors

Sanitary Mount Hardware Flow-through Mount Hardware


9H1310 2-inch Sanitary Tee Each tee is a standard 3/4-inch tee with 3/4-inch NPT threads
9H1132 2-inch Sanitary Clamp on all three openings.
9H1384 2-inch Sanitary Viton Gasket MH313N3NZ 316 SS Tee
MH333N3NZ CPVC Tee
MH373N3NZ PVC Tee
Immersion Mount Hardware
Each immersion hardware includes a 1/2-inch diameter
Insertion Mount Hardware
x 4 foot long pipe, 1/2 x 3/4-inch NPT coupling, and
plastic pipe-mount junction box with terminal strip. The insertion hardware includes a 1-1/2 inch ball valve,
MH432G CPVC Pipe 1-1/2 inch NPT close nipple for process connection, sensor
connection tube, stainless steel extension pipe, and stainless
steel compression fitting with washer and lock nut.
MH116M3MZ 316 SS Hardware
8

At Hach, it’s about learning


To complete your pH and ORP from our customers and
providing the right answers.
measurement system, choose from It’s more than ensuring the
these Hach controllers… quality of water—it’s about
ensuring the quality of life.
When it comes to the things
Model sc200 Controller that touch our lives...
(see Lit. #2665)
Keep it pure.
The sc200 controller platform can be configured to operate Make it simple.
either 2 Digital Sensor Inputs, or 1 or 2 Analog Sensor Be right.
Inputs, or a combination of Digital and Analog Sensor
Inputs. Customers may choose their communication
options from a variety of offerings ranging from MODBUS
RTU to Profibus DPV1. For current price information,
technical support, and ordering
sc200 for Hach Digital Sensors
assistance, contact the Hach
LXV404.99.00552 sc200 controller, 2 channel, digital office or distributor serving
LXV404.99.00502 sc200 controller, 1 channel, digital your area.
LXV404.99.00542 sc200 controller, 2 channel, digital & mA input
LXV404.99.00512 sc200 controller, 2 channel, digital & pH/DO
LXV404.99.00522 sc200 controller, 2 channel, digital & Conductivity In the United States, contact:
LXV404.99.00532 sc200 controller, 2 channel, digital & Flow
HACH COMPANY World Headquarters
sc200 for Hach Analog Sensors P.O. Box 389
LXV404.99.00102 sc200 controller, 1 channel, pH/DO Loveland, Colorado 80539-0389
LXV404.99.00112 sc200 controller, 2 channel, pH/DO U.S.A.
LXV404.99.00202 sc200 controller, 1 channel, Conductivity Telephone: 800-227-4224
LXV404.99.00222 sc200 controller, 2 channel, Conductivity Fax: 970-669-2932
LXV404.99.00212 sc200 controller, 2 channel, pH/DO & Conductivity E-mail: orders@hach.com
LXV404.99.00302 sc200 controller, 1 channel, Flow www.hach.com
LXV404.99.00332 sc200 controller, 2 channel, Flow
LXV404.99.00312 sc200 controller, 2 channel, Flow & pH/DO
LXV404.99.00322 sc200 controller, 2 channel, Flow & Conductivity U.S. exporters and customers in Canada,
Note: Other sensor combinations are available. Please contact Hach Technical Support Latin America, sub-Saharan Africa, Asia,
or your Hach representative. and Australia/New Zealand, contact:
Note: Communication options (MODBUS and Profibus DPV1) are available.
HACH COMPANY World Headquarters
Model sc1000 Controller P.O. Box 389
Loveland, Colorado 80539-0389
(see Lit. #2403)
U.S.A.
Each sc1000 Probe Module provides power to the Telephone: 970-669-3050
system and can accept up to 8 digital sensors/expansion
Fax: 970-461-3939
boards. Probe Modules can be networked together to
E-mail: intl@hach.com
accommodate up to 32 digital sensors/expansion boards
attached to the same network. www.hach.com

LXV402.99.00002 sc1000 Display Module


LXV400.99.1R572 sc1000 Probe Module, 4 sensors, In Europe, the Middle East, and
4 mA Out, 4 mA In, 4 Relays, Mediterranean Africa, contact:
110-230V HACH LANGE GmbH
LXV400.99.1B572 sc1000 Probe Module, 4 sensors, Willstätterstraße 11
4 mA Out, 4 mA In, 4 Relays, RS-485 D-40549 Düsseldorf
(MODBUS), 110-230V GERMANY
LXV400.99.1F572 sc1000 Probe Module, 4 sensors, Tel: +49 (0) 211 5288-0
4 mA Out, 4 mA In, 4 Relays, PROFIBUS DP, 110-230V Fax: +49 (0) 211 5288-143
LXV400.99.1R582 sc1000 Probe Module, 6 sensors, E- mail: info@hach-lange.de
4 mA Out, 4 mA In, 4 Relays, 110-230V
www.hach-lange.com

LIT2470 Rev 2
Printed in U.S.A.
©Hach Company, 2016. All rights reserved.
In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.
DATA SHEET FOR CONDUCTIVITY ANALYZER

Make HACH
Tag Number As per Annexure-1
Service As per Annexure-1
General
Process Medium As per Annexure-1
Instrument Range Conductivity 0–200000 µS/cm
Calibration Range As per Annexure-1
Model No. 3422E3A
Sensor Type Flow Through
Sensor End Connection Size 3/4" NPT(M)
Electrode/ Sensor MOC SS316
Sensor Insulator MOC PEEK
Sensor Integral Cable Length 5 Mtrs Sensor to transmitter
Sensor Installation In Flow Cell
Temperature Compensation Auto
Conductivity Temperature Electrode PT1000 RTD
Sensor Accuracy Loop accuracy is 2% including inaccuracies due to controller/transmitter, conductivity
electrode, conductivity standard (inaccuracy in conductivity standards ranges up to 1%
- 3%). conductivity inaccuracy in the standard will be <1%

Response Time Standard response time < 5 sec. T90 will be 1s to 30s from step change.
Repeatability ±0.5 % of reading
Sensitivity ±0.5% of reading
Flow Cell Maxi. Pressure Holding 1.7 bar at 150 °C
Capacity
Flow Cell MOC SS
Cell Constant 10
Flow Cell Maxi. Temperature Handling ‐20 ‐ 200 °C
Capacity
Model No. SC200
ANNEXURE-1
Tag No. SERVICE TYPE RANGE OPER DESIGN Pro Calib
ATIN RANGE cess ratio
G Med n
From To Temp Pres Tem Pres ium Range
°C sure p °C sure
kg/c kg/c
10GQB16CQ004 Effluent Conductivity Flow through 0 10000 Amb 2 60ºC 6 Water 10000
measurement μS/c
m
CONTACTING
Applications
• Drinking Water
• Power

CONDUCTIVITY/RESISTIVITY • Industrial Water

SENSORS

Ultimate accuracy from ultra-pure to


high conductivity applications.
universal installation with up to 4 inches (102 mm) insertion
High Performance Design depth. The 1/2-inch or 3/4-inch male NPT compression
These sensors are manufactured to exacting tolerances using fitting are o9ered in Kynar® (PVDF) or 316 stainless steel.
high quality, rugged materials for demanding applications, A longer version of this sensor is available for use with a 316
including ultra-pure water, clean-in-place (CIP), and
boiler/condensate monitoring. Each sensor is tested to
determine its unique, absolute four-digit cell constant,
Additionally, each sensor has a Pt 1000 RTD temperature
element built in for exceptionally fast response to changes in
temperature with ±0.1°C accuracy.

Resistivity and Conductivity


Measurement Capability
These enhanced performance sensors measure from
0.057 µS/cm or 18.2 MΩ (theoretically pure water) up to
200,000 µS/cm.

Versatile Mounting Styles


Compression Fitting Sensors
Feature titanium electrodes and a compression fitting for
stainless steel ball valve hardware assembly to insert/retract Designed for monitoring boiler water and condensate in
the sensor from the process without stopping the flow. return lines. They have 316 stainless steel electrodes and
The longer version can also be used for insertion through a threaded bodies (3/4-inch male NPT). They can be fastened
compression fitting. Maximum insertion depth is 7 inches into a boiler wall using a 3/4-inch weldolet or mounted into
(178 mm). a process line using a standard 3/4-inch stainless pipe tee.

Non-Metallic General Purpose Sensors Sanitary Clean-in-Place (CIP) Style Sensors


These sensors have graphite electrodes and 3/4-inch male Feature 316 stainless steel electrodes and an integral
NPT threaded Ryton® bodies. Mount into a standard 3/4-inch 1-1/2-inch or 2-inch flange. These sensors can be installed
pipe tee, 1-1/2-inch Hach union hardware (for 10 Cell using standard sanitary mounting hardware.
Constant sensor only), or fasten onto the end of the pipe.

High Pressure and High Temperature Sensors


Contacting Conductivity/Resistivity Sensors 2

Specifications*
Cell Constants and Measuring Ranges Flow Rate 0–3 m/s (0–10 ft./s), maximum, f
ully immersed
Sensor Cell Inherent Measuring Range
Constant Conductivity (µS/cm) Resistivity (Mohm) Temperature Pt 1000 RTD
0.05 0–100 0.002–20 Compensator

0.5 0–1000 0.001-20 Transmission Distance 100 m (328 ft.), maximum


1000 m (3280 ft.), maximum when
1 0–2000 not applicable used with a termination box
5 0–10000 not applicable Standard Probe Cable Digital Probe: 7 m (23 ft.)
10 0–200000 not applicable Length (integral) Analog Probe: 6 m (20 ft.)
Sensor Cable Digital: PUR (polyethylene)
5-conductor, shielded,
Temperature –20 to 200°C (–4 to 392°F) rated to 150°C (302°F)
Measurement Range
Analog: Integral (no junction box)
Accuracy ±2% of reading above 200 µS/cm 6 wire cable (4 conductors and
two isolated shield wires)
Sensitivity ±0.5% of reading
Analog with Junction Box Head:
Response Time 90% of reading within 30 seconds (optional) 6-position terminal strip
of step change supplied in integrally-mounted
Repeatability ±0.5% of reading junction box (polypropylene,
aluminum, or 316 stainless steel)
Operating Temperature –20 to 200°C (–4 to 392°F) *Subject to change without notice.

Model 3422-series Model 3433-series Model 3444-series Model 3455-series


Compression Non-metallic Boiler/ Sanitary (CIP)
Fitting General Purpose Condensate Flange
Temperature/ When used with Kynar® 150°C at 6.8 bar Sensor with integral cord When used with
Pressure Limits (PVDF) compression (302°F at 100 psi) or grip: 200°C at 20.7 bar manufacturer-supplied
(See Note 1) fitting: 150°C at 1.7 bar 20°C at 13.7 bar (392°F at sanitary mount hardware
(302°F at 25 psi) (68°F at 200 psi). 300 psi). assemblies: 125°C at

10.3 bar (257°F at 150


When used with When used with Sensor with integral psi)
manufacturer-supplied hardware, a lower rated polypropylene J-box (See Note 2)
316 stainless steel mounting hardware or Head: 92°C at 20.7 bar
compression fitting: piping material may limit (198°F at 300 psi).
150°C at 13.7 bar the temperature and
(302°F at 200 psi). pressure ratings listed Sensor with integral
above. aluminum or 316 SS
When used with J-box head: 200°C
316 stainless steel at 20.7 bar (392°F at
ball valve hardware 300 psi).
assemblies: 125°C
at 10.3 bar
(257°F at 150 psi).
Wetted Materials Titanium electrodes (316 Graphite electrodes, 316 stainless steel and 316 stainless steel
stainless steel outer Ryton® body, and titanium electrodes, electrodes, PTFE
electrode for extended Viton® O-ring seals PEEK insulator, and insulator, and
sensor body style used fluoroelastomer O-ring Pufluoroelastomer
with ball valve seals O-ring seals
assembly),
PTFE insulator, and
treated Viton®
O-ring seals

NOTES
1. For conductivity applications above 70°C (158°F), use the Digital Gateway (P/N 61207-00) with the appropriate sensor.
Please contact Hach Technical Support for further details.
2. Other brands of mounting hardware assemblies and sanitary clamps may reduce the listed rating.

Viton® is a registered trademark of E.I. DuPont de Nemours + Co.; Ryton® is a registered trademark of Phillips 66 Co.;

hach.com
Contacting Conductivity/Resistivity Sensors 3
PEEK® is a registered trademark of ICI Americas, Inc. ; Kynar® is a registered trademark of Pennwalt Corp.

hach.com
Contacting Conductivity/Resistivity Sensors 4

Dimensions
Model 3422-series Compression Fitting Sensor
om
Compression-style Sensor, 0.5-in. Diametero o , 0.5 . m

omDiameter
Compression-style Sensor, 0.75-in o o , 0.75 . m

om ®cTip
Compression-style Sensor with Teflon o o

Compression-Style Sensor with Integral Junction Box

19 mm (0.75 inches)

34 mm 41.4 mm (1.63 inches) 135.9 mm


(1.34 inches) (5.35 inches)
59.9 mm (2.32 inches)

190.5 mm (7.50 inches) 107.9 mm


(4.25 inches)
247.65 mm (9.75 inches)

hach.com
Contacting Conductivity/Resistivity Sensors 5

Dimensions
Model 3433-series Non-metallic General Purpose Sensor

Model 3444-series Boiler/Condensate Sensor


3/4-inch NPT

32.5 mm (1.28 inches)

19.05 mm (0.75 inches)


50.8 mm (2.0 inches)

72.39 mm (2.85 inches)

101.6 mm (4 inches)

149.9 mm (5.9 inches)

Model 3455-series Sanitary (CIP) Flange Sensor


49.8 mm (1.96 inches)

19.05 mm (.75 inches) 3/4-inch NPT

16.5 mm
(0.65 inches)
3.175 mm (0.125 inches)

hach.com
Contacting Conductivity/Resistivity Sensors 6
78.7 mm (3.1 inches)
106.7 mm (4.2 inches)

132.1 mm (5.2 inches)

168.9 mm (6.65 inches)

Ordering Information
Compression Fitting Sensors
Designed for ultrapure water and pure water applications, these small, enhanced performance contacting conductivity sensors
provide the required absolute cell (K) constant accuracy, and ultrafast-acting temperature compensation. Materials of construction
extend sensor operating life with no degradation in measurement reliability.

3422 sc Digital Compression Fitting Sensors


All digital compression fitting sensors have titanium electrodes and include built-in digital electronics and integral 7 m (23 ft.) cable
terminated with connector for the sc Digital Controllers. For insertion applications only. (For immersion applications, order 3422
Analog Compression Fitting Sensors.) For other cell constants and configurations, please contact your Hach Sales Representative.
Product Number Cell Constant Compression Fitting Style
D3422A1 0.05 1/2-inch NPT Kynar® (PVDF)
D3422A2 0.05 1/2-inch NPT 316 stainless steel
D3422B3 0.5 3/4-inch NPT Kynar® (PVDF)
D3422C3 1.0 3/4-inch NPT Kynar® (PVDF)
D3422D3 5.0 3/4-inch NPT Kynar® (PVDF)
D3422E3 10 3/4-inch NPT Kynar® (PVDF)

Digital Gateway
6120700 Use the Digital Gateway to connect analog Hach 3400-series conductivity sensors to the Hach sc1000 Digital Controller.

3422 Analog Compression Fitting Sensors


All analog compression fitting sensors have titanium electrodes and include an integral 6 m (20 ft.) cable terminated with stripped and
tinned wires. For other cell constants and configurations, including options for integral junction boxes,
please contact your Hach Sales Representative.
Product Number Cell Constant Compression Fitting Style
3422A1A 0.05 1/2-inch NPT Kynar® (PVDF)
3422A2A 0.05 1/2-inch NPT 316 stainless steel
3422A3A 0.05 3/4-inch NPT Kynar® (PVDF)
3422B3A 0.5 3/4-inch NPT Kynar® (PVDF)
3422C3A 1.0 3/4-inch NPT Kynar® (PVDF)
3422D3A 5.0 3/4-inch NPT Kynar® (PVDF)
3422E3A 10 3/4-inch NPT Kynar® (PVDF)

hach.com
DATA SHEET FOR TURBIDITY ANALYZER

Make HACH

Medium/ Type As per Annexure-3


General
Instrument Range As per Annexure-3

Tag NO As per Annexure-3

Location As per Annexure-3

Model Number LXV445.99.10112

Reference type Online

Range 0 - 700 NTU / FNU / TE/F / FTU

Operating principle Light reflection principle

Body Material Thermoplastic Elastomer


Turbidity
Sensor Temperature Rating 2 - 60 °C (35 - 140 °F)
Specification
Type Flow thru chamber

Resolution 0.0001 NTU / FNU / TE/F / FTU / EBC

Accuracy ±2% or 0.01 NTU from 0 - 40 NTU , ±5% from 40- 100 NTU

Repeatability Better than 1% of reading or ±0.002 NTU (TU5300) or ±0.0006


NTU (TU5400) on Formazin at 25 °C (77 °F), whichever is greater

Response Time T90 <45 seconds at 100 mL/min

Sample Flow Rate 100 - 1000 mL/min;

Pressure 6 bar (87 psi) maximum

Transmitter Model No. SC200

Annexure-3

Tag No. SERVICE TYPE INSTRUME OPERATING DESIGN PROC Calibra


NT RANGE RANGE RANGE ESS tio n
MEDI Range
Fro To Temp Press Temp Press UM
m °C ure °C ure
kg/c kg/c
10GQB16CQ001 Effluent pump Turbidity Flow 0 700 Amb.Te Atm 60ºC Atm Water 500 NTU
through NTU mp
TU5 Series Turbidimeters Applications
• Drinking Water
• Power
• Beverage
• Pharmaceutical

The next standard in the evolution of turbidity


Only the new TU5 Series Lab & Process Turbidimeters with 360° x 90° Detection™ deliver
unprecedented confidence that a change in your reading is a change in your water.

Groundbreaking 360° x 90° Detection™ Everything about turbidity – faster


Technology The TU5 Series dramatically reduces the time needed to get a
The TU5 Series employs a patented optical design that sees turbidity measurement you can rely on, with 98% less online
more of your sample than any other turbidimeter, delivering sample surface area to clean, sealed vials for calibration, and
the best low level precision and sensitivity while minimizing the elimination of the need for indexing and silicone oil in the
variability from test to test. lab. Not to mention, a smaller online sample volume means
you will detect events almost immediately.
Matching lab and online results
For the first time you will be able to remove the uncertainty No surprises
of which measurement to trust, thanks to identical 360° x 90° Prognosys™ monitors your TU5 Series online instrument,
Detection™ Technology in both instruments. proactively alerting you to maintenance needs before your
measurement becomes questionable. And a Hach Service
Agreement protects your investment and helps ensure that
you stay in compliance and on budget.

USEPA and ISO 7027 reporting: The TU5 Series Turbidimeters apply the instrument design and meet performance criteria established
by EPA Approved Hach Method 10258 and ISO 7027-1:2016, making them suitable for regulatory reporting.
Technical Data*
TU5200 TU5300sc / TU5400sc
Light Source Class 2 laser product, with Light Source Class 2 laser product, with embedded
embedded 650 nm (EPA 0.43 650 nm (EPA 0.43 mW) or Class 1
mW) or Class 1 laser product, with laser product, with embedded 850
embedded 850 nm (ISO), max. nm (ISO), max. 0.55 mW (complies
0.55 mW (complies with IEC/EN with IEC/EN 60825-1 and to 21 CFR
60825-1 and to 21 CFR 1040.10 1040.10 in accordance with Laser
in accordance with Laser Notice Notice No. 50)
No. 50) Range EPA:
Range EPA: 0 - 700 NTU / FNU / TE/F / FTU
0 - 700 NTU / FNU / TE/F / FTU 0 - 100 mg/L
0 - 100 mg/L 0 - 175 EBC
0 - 175 EBC
ISO:
ISO: 0 - 1000 NTU / FNU / TE/F / FTU
0 - 1000 NTU / FNU / TE/F / FTU 0 - 100 mg/L
0 - 100 mg/L 0 - 250 EBC
0 - 250 EBC Accuracy ±2% or 0.01 NTU from 0 - 40 NTU
Accuracy ±2 % plus 0.01 NTU from 0 - 40 NTU; ±10% of reading from 40 - 1000
NTU based on Formazin primary
±10 % of reading from 40 - 1000
standard
NTU based on Formazin primary
standard (at 25 °C) Resolution 0.0001 NTU / FNU / TE/F / FTU / EBC
Resolution 0.0001 NTU / FNU / TE/F / FTU / Repeatability Better than 1% of reading or
EBC / mg/L ±0.002 NTU (TU5300) or ±0.0006
NTU (TU5400) on Formazin at
Repeatability <40 NTU: Better than 1% of
25 °C (77 °F), whichever is greater
reading or ±0.002 NTU on
Formazin at 25 °C, whichever is Stray Light <10 mNTU
greater Units NTU, FNU, TE/F, FTU, EBC
>40 NTU: Better than 3.5% of Signal Average Time TU5300sc: 30 - 90 seconds
reading on Formazin at 25 °C TU5400sc: 1 - 90 seconds
Response Time TU5300sc:
Stray Light <10 mNTU
T90 <45 seconds at 100 mL/min
Units NTU, FNU, TE/F, FTU, EBC; TU5400sc:
mg/L if calibrated with Degrees T90 <30 seconds at 100 mL/min
calibration curve
Sample Temperature 2 - 60 °C (35 - 140 °F)

Operating Temperature 10 - 40 °C (50 - 104 °F)


Sample Pressure 6 bar (87 psi) maximum, compared
Range
to air at sample temperature range
Operating Humidity 80% at 30 °C (non condensing) from 2 - 40 °C (35.6 - 104 °F)
Sample Temperature 4 - 70 °C (39 - 158 °F) Sample Flow Rate 100 - 1000 mL/min; optimal flow
Storage Conditions -30 - 60 °C (-22 - 140 °F) rate: 200 - 500 mL/min

Power Requirements 100 - 240 VAC Operating Temperature 0 - 50 °C (32 - 122 °F)
(Voltage) Range

Power Requirements 50/60 Hz Operating Humidity Relative humidity: 5 - 95%


(Hz) at different temperatures,
non-condensing
Certifications CE compliant
US FDA accession number:
1420493-000 EPA version, www.hach.com
1420492-000 ISO version
Complies with IEC/EN 60825-1 and
to 21 CFR 1040.10 in accordance
with Laser Notice No. 50)
Australian ACMA Marking

Dimensions (H x W x D) 195 mm x 409 mm x 278 mm


Weight 2.4 kg (5.29 lbs.)
Warranty 1 year
Storage Conditions -40 - 60 °C (-40 - 140 °F)
Enclosure Rating Electronic compartment IP55;
all other functional units IP65 with
process head/ACM attached to
the TU5300sc/TU5400sc
instrument
Certifications CE compliant
US FDA accession
number: 1420493-000
EPA version,
1420492-000 ISO version
Australian ACMA Marking
Dimensions (H x W x D) 249 mm x 268 mm x 190 mm
Weight 5.95 lbs. (2.7 kg); 11 lbs. (5.0
kg) with all accessories
Warranty 1 year

*Subject to change without notice.


TU5 Series Turbidimeters 3

Principle of Operation
The TU5 Series turbidimeters measure turbidity by directing a
laser into a sample to scatter off suspended particles. The light
that is scattered at a 90° angle from the incident beam is reflected
through a conical mirror in a 360° ring around the sample before
it is captured by a detector.
The amount of light scattered is proportional to the turbidity of the
sample. If the turbidity of the sample is negligible, little light will be
scattered and detected by the photocell and the turbidity reading
will be low. High turbidity, on the other hand, will cause a high
level of light scattering and result in a high reading.
The 360° x 90° optics of the TU5 series were optimized for high
accuracy at low turbidity ranges and therefore the TU5 does not
include ratio technology. Ratio technology is only applicable for
high turbidity applications which have interference from color and
large particles.

Dimensions
TU5300sc and TU5400sc front view TU5300sc and TU5400sc rear view

TU5300sc and TU5400sc top view

www.hach.com
Order Information
TU5200 Benchtop Laser Turbidimeters
LPV442.99.03012 TU5200 Benchtop Laser Turbidimeter with RFID, EPA Version
LPV442.99.01012 TU5200 Benchtop Laser Turbidimeter without RFID, EPA Version
LPV442.99.03022 TU5200 Benchtop Laser Turbidimeter with RFID, ISO Version
LPV442.99.01022 TU5200 Benchtop Laser Turbidimeter without RFID, ISO Version

TU5300sc/TU5400sc Online Laser Turbidimeters


LXV445.99.10112 TU5300sc Low Range Laser Turbidimeter, EPA Version
LXV445.99.10212 TU5400sc Ultra-High Precision Low Range Laser Turbidimeter,
EPA Version

Please note: Other turbidimeter configurations are available and RFID may not be
available in all areas. Please contact your local Hach representative.
Please note: An SC controller is required for operation of the TU5300sc or TU5400sc.

Calibration and Verification


LZY835 Stablcal® Calibration Set with RFID
LZY898 Stablcal® Calibration Set without RFID
LZY901 Glass Rod Secondary Turbidity Standard <0.1 NTU/FNU
LZY834 Replacement Vial for TU5300sc and TU5400sc
LZV946 Sample Vials for TU5200

TU5 Series Accessories


LQV159.97.00002 Automatic Cleaning Module for TU5300sc and TU5400sc
LQV160.99.00002 Flow Sensor for TU5300sc and TU5400sc
LZY876 Desiccant Cartridge for TU5300sc and TU5400sc
LZY907.97.00002 Maintenance Kit for TU5300sc and TU5400sc
LQV157.99.50002 SIP10 Sipper Unit for TU5200
LZY903 Manual Vial Wiper for TU5200, TU5300sc, and TU5400sc

DOC050.53.10053.Sep19
With Hach Service, you have a global partner who understands your needs and cares about delivering timely, high-quality
service you can trust. Our Service Team brings unique expertise to help you maximise instrument uptime, ensure data
integrity, maintain operational stability, and reduce compliance risk.

HACH COMPANY World Headquarters: Loveland, Colorado USA


United States: 800-227-4224 tel 970-669-2932 fax orders@hach.com
Outside United States: 970-669-3050 tel 970-461-3939 fax int@hach.com
hach.com
Printed in U.S.A.
©Hach Company, 2019. All rights reserved.
In the interest of improving and updating its equipment, Hach Company reserves the right to alter specifications to equipment at any time.
DATA SHEET FOR COD BOD ANALYZER

Make HACH

General Tag.No As per Annexure-3

Service As per Annexure-3

Measurement Type As per Annexure-3

Model Number LXV418.99.50002

Measurement Technique UV absorption measurement (2-beam technique), reagent-free, with self cleaning

Measurement Method SAC 254 in accordance with DIN 38404 C3

Measurement Path Length 1, 2, 5 and 50 mm


COD BOD
Sensor
Specificati Measurement Range 0.1 to 600 m–1 at 5 mm
on 0-50 mg/L for BOD,
0-500 mg/L for COD
Accuracy < 3%

Compensation 550 nm

Measurement Interval ≥ 1 minute

Sample Temperature 2 to 40°C (35.6 to 104°F)

Probe Pressure Limit at Inlet 0.5 bar (7.25 psi) maximum

Sensor Cable Length 10 m (32.8 ft.)

Transmitter Model No. SC200

Annexure-3

INSTRUMENT OPERATING DESIGN PR Calib


ratio n
Tag No. SERVICE TYPE RANGE RANGE RANGE O
From To Temp Press Temp Press
C
°C ure °C ure
ES
kg/c kg/c
S 100
10GQB16CQ002 RO Effluent pump Dish, COD/BOD Flow 500 Amb. Atm 60ºC Atm W
/003 COD/BOD measurement through ppm Tem p ate
r
UVAS PLUS sc SENSOR Applications
• Drinking Water
• Wastewater
• Industrial

Continuously protect plant treatment processes


from high influent organic loads.
Continuous, Automatic Early Monitor E:ciency of
Warning Systems UV Disinfection Process
Use the Hach UVAS plus sc UV Absorbance/ %Transmittance Hach’s UVAS plus sc is designed to provide continuous UV
Sensor to continuously protect plant treatment processes light transmittance (UVT) measurement of pre-disinfected
from high influent organic loads. source water. Operational costs related to sampling for UVT
may be reduced with continuous on-line measurement.
Control Activated Sludge Processes
Activated sludge processes require precise balancing of Self-diagnostics and Easy Maintenance
organic load, aeration, and nutrients. Continuous trending Diagnostic routines built into the UVAS plus sc sensor reduce
of the organics using the UVAS plus sc sensor can help the need for extensive calibration and maintenance. Only
operators know how to balance other factors resulting in semi-yearly inspection and replacement of the wiper and
cost and time savings. seals is needed.

Self-cleaning Wiper System Principal of Operation


The detector windows are automatically cleaned by a The Hach UVAS plus sc UV Absorbance / %Transmittance
built-in wiper that eliminates surface films or particles that Sensor determines the Spectral Absorption Coe@cient (SAC)
can diminish accuracy. at a wavelength of 254 nm. Measurements can be expressed
in absorption units (1/m), mE, AU, %T, %T/cm, mg/L, or ppm.
This sensor requires a Hach sc200 or sc1000 Digital Controller.
Specifications*
UVAS plus sc Tank Sensors UVAS plus sc Bypass Sensors
Measurement Technique UV absorption measurement (2-beam technique), reagent-free

Measurement Method SAC 254 in accordance with DIN 38404 C3

Measurement Path Length 1, 2, 5 and 50 mm 2, 5, and 50 mm

Measurement Range Choice of: Choice of:


0.01 to 60 m–1 at 50 mm 0.01 to 60 m–1 at 50 mm
0.1 to 600 m–1 at 5 mm 0.1 to 600 m–1 at 5
0 to 1500 m–1 at 2 mm mm 0 to 1500 m–1 at 2
2 to 3000 m–1 at 1 mm mm

Compensation 550 nm

Measurement Interval ≥ 1 minute

Sample Temperature 2 to 40°C (35.6 to 104°F)

Sample pH 4.5 to 9 pH

Probe Pressure Limit at Inlet 0.5 bar (7.25 psi) maximum

Sample Flow Rate n/a 0.5 L/hour minimum

Sample Connection n/a 4 mm ID/6 mm OD hose

Sensor Cable Length 10 m (32.8 ft.)

Control Function PID, time control, 2-point controller (with sc controller)

Inspection Interval 6 months

User Maintenance 1 h / month, typical

Dimensions 70 x 333 mm (2.75 x 13.11 in.) approximate

Weight 3.6 kg (7.9 lb.) approximate


*Subject to change without notice.

NOTE
The UVAS plus sc probes cannot be used in sea water.

hach.com
UVAS plus sc Sensor 3

Dimensions
Hach UVAS plus sc UV Absorbance / %Transmittance Sensors can be installed using a fixed-point installation kit as shown in the bottom
illustration. The bypass panel below can be used for non-immersion applications. With the cable supplied, the sensor
can be used in a sample stream within 10 meters (32.8 feet) of the controller.
UVAS plus sc tank sensor

UVAS plus sc bypass sensor

20 mm
(0.79 inches) 500 mm
(19.69 inches)
400 mm
(15.75 inches) 2x Ø8 mm (0.31 inches)

210 mm
(8.27 inches)

Installation for mounting the Hach UVAS plus


sc UV Absorbance / %Transmittance Sensor
for immersion in open tanks.
Ordering Information
The following sensors include the Hach sc200 Multi-parameter Controller (see LIT2665 for complete details).
2976700 1 mm UVAS plus sc sensor
2976400 2 mm UVAS plus sc sensor
2976600 5 mm UVAS plus sc sensor
2976500 50 mm UVAS plus sc sensor

UVAS plus sc Sensor Only


LXV418.99.10002 1 mm UVAS plus sc sensor only
LXV418.99.20002 2 mm UVAS plus sc sensor only
LXV418.99.50002 5 mm UVAS plus sc sensor only
LXV418.99.90002 50 mm UVAS plus sc sensor only

Bypass Panel
LZX868 Bypass Panel for 50 mm sensor
LZX867 Bypass Panel for 5 mm sensor
LZX869 Bypass Panel for 2 mm sensor
LZX450 Sedimenter for UVAS sc sensor

Mounting Hardware
LZY714.99.53520 Mounting Hardware with 90 degree adapter

HACH COMPANY World Headquarters: Loveland, Colorado USA


United States: 800-227-4224 tel 970-669-2932 fax orders@hach.com
Outside United States: 970-669-3050 tel 970-461-3939 fax int@hach.com
hach.com

I 2485 e 3
C14 P e U. .A.
Ha C a , 2014. A e e e .
DATASHEET FOR SC200 UNIVEERSAL CONTROLLER

Model No. SC200

Type Microprocessor based SMART,

Repeatability 0.1% of range or better

Response Time 0.5 s

Power Requirement 230 VAC

Mounting Wall Mounting

Housing Material Polycarbonate

Protection Class NEMA 4X, equivalent to IP 66

Inputs Single Channel Sensor input

Outputs 4‐ 20 Ma (Glavanically isolated) + MODBUS

Load 500 Ohms Minimum

Display Type Two‐line by 16 character LCD

Calibration Manual
TRANSMITTER

Hysteresis Arrangement Provided

Cable Entry 1/2" NPT(F)

Ambient Temperature Limit 60°C

Self‐diagnostic with fault & warning type Available

Zero & Span Arrangement Provided

Accessories Mounting Bracket (CS/ MS) U bolts, nuts, screws (GI)


Washers
sc200™ UNIVERSAL CONTROLLER Applications
• Drinking Water
• Wastewater
• Industrial Water
• Power

One Controller for the Broadest Range of Sensors.


Choose from 30 digital and analog sensor families for up to 17 di:erent parameters.

Maximum Versatility
The sc200 controller allows the use of digital and analog
sensors, either alone or in combination, to provide
compatibility with Hach's broad range of sensors, eliminating
the need for dedicated, parameter-specific controllers.

Ease of Use and Confidence in Results


Large, high-resolution, transreflective display provides optimal
viewing resolution in any lighting condition. Guided
calibration procedures in 19 languages minimize complexity
and reduce operator error. Password-protected SD card
reader o:ers a simple solution for data download and
transfer. Visual warning system provides critical alerts.

Wide Variety of Communication Options


Utilize two to five analog outputs to transmit primary and
secondary values for each sensor, or integrate Hach sensors
and analyzers into MODBUS RS232/RS485, Profibus® DP, and
HART networks.
Password protected SD card reader offers a simple solution
for data download and transfer, and sc200 and digital sensor
configuration file duplication and backup.
sc200™ Universal Controller 2

Controller Comparison

Previous Models
Features sc100™ Controller GLI53 Controller sc200™ Controller Benefits

Display 64 x 128 pixels 64 x 128 pixels 160 x 240 pixels • Improved user interface—
33 x 66 mm 33 x 66 mm 48 x 68 mm 50% bigger
(1.3 x 2.6 in.) (1.3 x 2.6 in.) (1.89 x 2.67 in.) • Easier to read in daylight
Transreflective and sunlight

Data irDA Port/PDA N/A SD Card • Simplifies data transfer


Management Service Cable Service Cable • Standardized accessories/
max compatibility

Sensor Inputs 2 Max 2 Max 2 Max • Simplifies analog sensor


Direct Digital Analog Digital and/or Analog connections
Analog via Depending on with Sensor Card • Works with analog and
External Gateway Parameter digital sensors

Analog Inputs N/A N/A 1 Analog Input Signal • Enables non-sc analyzer
Analog 4-20mA Card monitoring
• Accepts mA signals from
other analyzers for local
display
• Consolidates analog mA
signals to a digital output

4-20 mA Outputs 2 Standard 2 Standard 2 Standard • Total of five (5) 4-20 mA


Optional 3 Additional outputs allows multiple mA
outputs per sensor input

Digital MODBUS RS232/RS485 HART MODBUS RS232/RS485 • Unprecedented combination


Communication Profibus DP V1.0 Profibus DP V1.0 of sensor breadth and digital
HART 7.2 communication options

hach.com
sc200™ Universal Controller 3

Choose from Hach's Broad Range of Digital and Analog Sensors


Parameter Sensor Digital or Analog
Ammonia AMTAX™ sc, NH4D sc, AISE sc, AN-ISE sc ●
Chlorine CLF10 sc, CLT10 sc, 9184 sc ●
Chlorine Dioxide 9185 sc ●
Conductivity GLI 3400 Contacting, GLI 3700 Inductive ▲
Dissolved Oxygen LDO® Model 2, 5740 sc ●
Dissolved Oxygen 5500 ▲
Flow U53, F53 Sensors ▲
Nitrate NITRATAX™ sc, NO3D sc, NISE sc, AN-ISE sc ●
Oil in Water FP360 sc ●
Organics UVAS sc ●
Ozone 9187 sc ●
pH/ORP pHD ●
pH/ORP pHD, pH Combination, LCP ▲
Phosphate PHOSPHAX™ sc ●
Sludge Level SONATAX™ sc ●
Suspended Solids SOLITAX™ sc, TSS sc ●
Turbidity 1720E, FT660 sc, SS7 sc, ULTRATURB sc, SOLITAX sc, TSS sc ●
Ultra Pure Conductivity 8310, 8311, 8312, 8315, 8316, 8317 Contacting ▲
Ultra Pure pH/ORP 8362 ▲
● = Digital ▲ = Analog

Connect up to two of any of the sensors listed above, in any 2 Channel 1 Channel
combination, to meet your application needs. The diagrams Configurations Configurations
below demonstrate the potential configurations. Operation of
analog sensors requires the controller to be equipped with the
appropriate sensor module. Contact Hach Technical Support
for help with selecting the appropriate module.

▲ ▲

hach.com
sc200™ Universal Controller 4

Specifications*
Dimensions (H x W x 5.7 in x 5.7 in x 7.1 in Relay Functions Scheduler (Timer), Alarm, Feeder
D) (144 mm x 144 mm x 181 mm) Control, Event Control, Pulse Width

Modulation, Frequency Control,


Display Graphic dot matrix LCD with LED
and Warning
backlighting, transreflective
Relays Four electromechanical SPDT
Display Size 1.9 x 2.7 in. (48 mm x 68 mm)
(Form C) contacts, 1200 W, 5 A
Display Resolution 240 x 160 pixels
Communication MODBUS RS232/RS485,
Weight 3.75 lbs. (1.70 kg) PROFIBUS DPV1, or HART 7.2

Power Requirements 100 - 240 V AC, 24 V DC optional


(Voltage) Memory Backup Flash memory

Power Requirements 50/60 Hz Electrical EMC


(Hz) Certifications
CE compliant for conducted and
Operating -20 to 60 °C , 0 to 95% RH radiated emissions:
Temperature Range non-condensing
- CISPR 11 (Class A limits)
Analog Outputs Two (Five with optional expansion
module) to isolated current - EMC Immunity EN 61326-1
outputs, max 550 , Accuracy: (Industrial limits)
± 0.1% of FS (20mA) at 25 °C, Safety
± 0.5% of FS over -20 °C to 60 °C
range cETLus safety mark for:
Operational Mode: measurement - General Locations per ANSI/UL
or calculated value 61010-1 & CAN/CSA C22.2. No.

Analog Output 61010-1


Linear, Logarithmic, Bi-linear, PID
Functional Mode - Hazardous Location Class I,

Security Levels 2 password-protected levels Division 2, Groups A,B,C & D


(Zone 2, Group IIC) per FM 3600 /
Mounting Wall, pole, and panel mounting FM 3611 & CSA C22.2 No. 213
Configurations M1987 with approved options

Enclosure Rating NEMA 4X/IP66 and appropriately rated Class I,


Division 2 or Zone 2 sensors
Conduit Openings 1/2 in NPT Conduit

cULus safety mark


Relay: Operational Primary or secondary
Mode measurement, calculated value - General Locations per UL 61010-
(dual channel only) or timer 1 & CAN/CSA C22.2. No. 61010-1
*Subject to change without notice.

hach.com
sc200™ Universal Controller 5

Dimensions

Surface Mounting Dimensions

OR

Panel Mounting Dimensions

Top and Bottom Views

hach.com
sc200™ Universal Controller 6

Ordering Information
sc200 for Hach Digital and Analog Sensors
LXV404.99.00552 sc200 controller, 2 channels, digital
LXV404.99.00502 sc200 controller, 1 channel, digital
LXV404.99.00102 sc200 controller, 1 channel, pH/DO
LXV404.99.00202 sc200 controller, 1 channel, Conductivity
LXV404.99.01552 sc200 controller, 2 channels, digital, Modbus RS232/RS485
LXV404.99.00112 sc200 controller, 2 channel, pH/DO

Note: Other Sensor combinations are available. Please contact Hach Technical Support
or your Hach representative.
Note: Communication options (MODBUS, Profibus DPV1, and HART) are available.
Please contact Hach Technical Support or your Hach representative.

sc200 for Ultrapure Sensors


9500.99.00602 sc200 controller, 1 channel, ultrapure conductivity
9500.99.00702 sc200 controller, 1 channel, ultrapure pH
9500.99.00662 sc200 controller, 2 channel, ultrapure conductivity
9500.99.00772 sc200 controller, 2 channel, ultrapure pH

Sensor and Communication Modules


9012900 Analog pH/ORP and DO module for GLI Sensors
9013000 Analog Conductivity module for GLI Sensors
9012700 Flow module
9012800 4-20 mA Input Module
9525700 Analog pH/ORP Module for Polymetron Sensors
9525800 Analog Conductivity Module for Polymetron Sensors
9013200 Modbus 232/485 Module
9173900 Profibus DP Module
9328100 HART Module
9334600 4-20 mA Output Module (Provides 3 additional mA Outputs)

Accessories
9220600 sc200 Weather and Sun Shield with UV Protection Screen
8809200 sc200 UV Protection Screen
9218200 SD card reader (USB) for connection to PC
9218100 4 GB SD card

HACH COMPANY World Headquarters: Loveland, Colorado USA


United States: 800-227-4224 tel 970-669-2932 fax orders@hach.com
Outside United States: 970-669-3050 tel 970-461-3939 fax int@hach.com
hach.com

I 2665 e 7
13 P e U. .A.
Ha C a , 2013. A e e e .

hach.com
800

1000

1800
1. DEGREE OF PRO : IP54
2. MAIN FRAME : 2.0MM CRCA SHEET
3. DOORS,PARTITIONS : 1.6MM CRCA SHEET CUSTOMER: DRAWN BY GD
PANEL GENERAL ARRANGEMENT DRAWING
4. MOUNTING TYPE : WALL MOUNTED EARTH WATER TECHNOLOGIES CHECKED BY SK
LOCATION: EQMS ROOM 5. CABLE ENTRY : BOTTOM ENTRY
APPROVED BY SK
TITLE :
1. CHEMICAL TREATMENT : 7- TANK PROCESS DATED 04-09-2021
2. PAINTING : POWDER COATING DRG NO
Note: 3. COLOUR : RAL 7035 / SIEMENS GRAY DOCUMENT NO :
4. EARTH BUSBAR : 25x03 GI Flat PAGE NO 1 OF 1
All dimensions are in mm
REV 00
MANDATORY SPARE

SR NO DESCRIPTION MODEL NO QUANTITY UNIT


1 CONDUCTIVITY
1.1 Conductivity Sensor/cell for each type of Cell Constant 3422E3A 2 Nos
1.2 Conductivity Transmitter Complete Set SC200 2 Nos
2 pH
2.1 pH Sensor PC1R1A 2 Nos
2.2 pH Transmitter Complete Set SC200 2 Nos
2.3 Sensor Recharger …. 2 Nos