You are on page 1of 104

About this Manual

P/N:01.54.455736

MPN: 01.54.455736012

Release Date: August, 2018

© Copyright EDAN INSTRUMENTS, INC. 2016-2018. All rights reserved

Statement

This manual will help you to better understand the operation and maintenance of the product. It
is reminded that the product should be used strictly in compliance with this manual. User
operation failing to comply with this manual may result in malfunction or accident for which
EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.

EDAN owns the copyrights of this manual. Without prior written consent of EDAN, any
materials contained in this manual shall not be photocopied, reproduced or translated into other
languages.

Materials protected by the copyright law, including but not limited to confidential information
such as technical information and patent information contained in this manual, shall not be
disclosed to any irrelevant third party.

The user shall understand that nothing in this manual grants him, expressly or implicitly, any
right or license to use any of the intellectual properties of EDAN.

EDAN holds the rights to modify, update, and ultimately explain this manual.

Responsibility of the Manufacturer

EDAN only considers itself responsible for any effect on safety, reliability and performance of
the equipment if:

Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by


persons authorized by EDAN, and

The electrical installation of the relevant environment complies with national standards, and

The instrument is used in accordance with the instructions for use.

EDAN will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, or other information that will assist service personnel to repair those

I
parts of the equipment that are designated by EDAN as repairable by service personnel.

Terms Used in this Manual

This guide is designed to give key concepts on safety precautions.

WARNING

A WARNING label advises against certain actions or situations that could result in personal
injury or death.

CAUTION

A CAUTION label advises against actions or situations that could damage equipment, produce
inaccurate data, or invalidate a procedure.

NOTE

A NOTE provides useful information regarding a function or a procedure.

II
Table of Contents
1 Warranty and Service ..................................................................................................................... 1
2 Safety Guide ...................................................................................................................................... 4
2.1 Introduction.............................................................................................................................. 4

2.2 Warnings and Cautions ........................................................................................................... 4

2.3 General Information ................................................................................................................ 6

3 Installation ........................................................................................................................................ 7
3.1 Unpacking Inspection.............................................................................................................. 7

3.2 Installation Requirements ....................................................................................................... 7

3.2.1 Environmental Requirements...................................................................................... 7

3.2.2 Power Requirements.................................................................................................... 8

3.3 Installing the Analyzer ............................................................................................................ 8

3.4 Connecting to AC Power ........................................................................................................ 8

3.5 Replacing a Calibrant Fluid Pack ........................................................................................... 8

4 Functional Checks.......................................................................................................................... 12
4.1 Switching On Check.............................................................................................................. 12

4.2 LCD Screen Check ................................................................................................................ 12

4.3 Touch Screen Check.............................................................................................................. 12

4.4 Diagnostics............................................................................................................................. 12

4.5 External Simulator Test......................................................................................................... 14

4.6 Control Test ........................................................................................................................... 17

5 Maintenance.................................................................................................................................... 24
5.1 Maintenance Inspection ........................................................................................................ 24

5.2 Maintenance of the Analyzer ................................................................................................ 24

6 Principle Introduction ................................................................................................................... 25


6.1 Measurement Method............................................................................................................ 25

6.2 Determination of Test Results .............................................................................................. 26

6.2.1 Determination of the Analyte Concentration ........................................................... 26

6.2.2 Determination of Cell Concentration ....................................................................... 27


III
6.3 Equations for Calculated Parameters ................................................................................... 27

6.4 System Principle Block Diagram ......................................................................................... 31

7 Troubleshooting ............................................................................................................................. 32
7.1 Booting Failures .................................................................................................................... 32

7.2 Display Failures ..................................................................................................................... 33

7.3 Touch Screen Failures........................................................................................................... 33

7.4 Operation Failures ................................................................................................................. 34

7.5 Printer Failures....................................................................................................................... 35

7.6 Network Failures ................................................................................................................... 35

7.7 Diagnosis Database ............................................................................................................... 36

7.7.1 Searching for Diagnosis Data ................................................................................... 36

7.7.2 Viewing Details of Diagnosis Data .......................................................................... 37

7.7.3 Exporting/Printing Diagnosis Data........................................................................... 38

............................................................................................. 39
.......................................................... 39

................................................................ 40

8.4 Verifying Malfunction of the Touch Screen Drive Board .................................................. 41

8.5 Verifying Malfunction of the Scanner ................................................................................. 42

9 Disassembling the Analyzer.......................................................................................................... 43


9.1 Tools Required....................................................................................................................... 43

9.2 Disassembling the Main Unit ............................................................................................... 43

9.3 Disassembling the Lower Assembly .................................................................................... 44

9.3.1 Replacing the Main Control Board........................................................................... 44

9.3.2 Replacing the Wi-Fi Board ....................................................................................... 45

9.3.3 Replacing the Chamber Door Detection Board ....................................................... 46

9.3.4 Replacing the Ambient Temperature Detection Board ........................................... 47

9.3.5 Replacing the Analog Board ..................................................................................... 47

9.3.6 Replacing the Cam Machine ..................................................................................... 51

IV
9.3.7 Replacing the Power Board....................................................................................... 56

9.3.8 Replacing the Power Interface Board....................................................................... 57

9.3.9 Replacing the Exhaust Fan ........................................................................................ 58

9.3.10 Replacing the Speaker ............................................................................................. 59

9.3.11 Replacing the Bar Code Scanner ............................................................................ 59

9.3.12 Replacing the Spring Located in Test Cartridge Base .......................................... 61

9.3.13 Cleaning and Replacing Vacuum Pump................................................................. 63

9.3.14 Replacing FPC Cable .............................................................................................. 64

9.3.15 Replacing the Cable of Temperature Detection Board ......................................... 72

9.3.16 Replacing and Cleaning Needle.............................................................................. 75

9.4 Disassembling the Upper Assembly .................................................................................... 77

9.4.1 Replacing the Touch Screen Drive Board................................................................ 77

9.4.2 Replacing the LCD .................................................................................................... 78

9.4.3 Replacing the Touch Screen ..................................................................................... 79

9.4.4 Replacing the Print Controller Board ....................................................................... 80

9.4.5 Replacing the Printer Head ....................................................................................... 81

10 Common Failures and Solutions ............................................................................................... 82


11 Replaceable Parts......................................................................................................................... 97

V
i15 Blood Gas and Chemistry Analysis System Service Manual Warranty and Service

1 Warranty and Service


Standard Service

EDAN provides a one-year-warranty for the warranted products. The warranty period begins on
the date the products are shipped to customers. If a customer promptly notifies EDAN of

replacement
will be performed by trained individuals in a workmanlike manner.

Limitation of Warranty

Direct, indirect or final damage and delay caused by the following situations for which EDAN is
not responsible may void the warranty:

Groupware is dismounted, stretched or redebugged.

Unauthorized modification or misuse.

Damage caused by operating beyond the environmental specifications for the medical
product.

Change or remove original serial number label or Manufacturer symbol.

Improper use.

Service Procedure

(1) Fill in the Service Claim Form (SCF).

Fill in the SCF with detailed information including: Model Name, Serial Number (SN) and
Problem Phenomena.

EDAN should not have any obligation to take over the case without this information. The
form can be downloaded at: http://www.edan.com.cn
Department.

(2) Send EDAN the SCF and Select a Solution.

Once the service depart


solution in three working days. EDAN will follow out the case based on the two conditions
below:

-1-
i15 Blood Gas and Chemistry Analysis System Service Manual Warranty and Service

Within Warranty:

There are two options:

i) After receiving the Return Material Authorization (RMA) form from EDAN service
department, the customer sends EDAN the defective parts and informs about the shipment
tracking number. Then we will dispatch new part(s) to your confirmed address with
confirmed shipping invoice.

ii) The customer signs the Declaration Form and sends it back by email or fax. This form is
legally certificated to make sure the customer or end-user will return the defective parts to
EDAN on time. We will, at this option, dispatch the replacement one(s) with confirmed
shipping invoice.

NOTE:

Both Return Material Authorization Form and Declaration Form are offered by
EDAN service department once the SCF is confirmed by a service engineer.
The customer is responsible for freight & insurance charges when the equipment
is shipped to EDAN for service, including custom charges. EDAN is responsible
for the freight, insurance & custom charges from EDAN to the customer.

Out of Warranty:

After receiving the RMA form from the service department, the customer sends defective
parts to EDAN in advance. We will analyze the problems and discuss with the customer
about either repairing or replacing the part(s). Once the maintenance fee is invoiced and paid,
we will dispatch good part(s) to the confirmed address.

NOTE:

The customer is responsible for any freight & insurance charge for the returned
product.

(3) Obtain the RMA Form

Before the shipment of the materials, the customer must obtain an RMA form from our
service department, in which the RMA number, description of returning parts and shipping
instructions are included. The RMA number should be indicated on the outside of the
shipping container.

NOTE:

EDAN should not have any obligation to the end-user or customer who returns the
-2-
i15 Blood Gas and Chemistry Analysis System Service Manual Warranty and Service

sender takes full


responsibility for the accounted fee.

(4) Send the Parts to EDAN.

Follow these recommended instructions:

Please disassemble the parts with anti-static facility, and never touch the parts with naked
hand.

Please pack the parts safely before return.

Please put the RMA number on the parcel.

Please confirm the invoice with EDAN before shipment.

Contact Information

If you have any question about maintenance, technical specifications or malfunctions of devices,
do not hesitate to contact us.

EDAN Instruments, Inc.

TEL: +86-755-26898321, 26899221

FAX: +86-755-26882223, 26898330

E-mail: support@edan.com.cn

-3-
i15 Blood Gas and Chemistry Analysis System Service Manual Safety Guide

2 Safety Guide
2.1 Introduction

This service manual is a reference for periodic preventive maintenance and corrective service
procedures for the i15 blood gas and chemistry analysis system. It provides information on
troubleshooting, assembly procedures, and instructions for functional testing as well as
performance verification. The manual is intended for use only by technically qualified service
personnel.

WARNING

Strictly follow the instructions as described in this manual during service. Otherwise the
system may be damaged, or personal injury may occur.

2.2 Warnings and Cautions


To avoid possible injury, please read the following warnings and cautions during the operation of
the system.

WARNING

1. The system is not intended for treatment.

2. The system is not intended for home use.

3. The system is intended for use only by trained technologists, nurses, physicians and
therapists.

4. SHOCK HAZARD Do not attempt to connect or disconnect a power cord with wet
hands. Make sure your hands are clean and dry before touching a power cord.

5. The sample is contained in the test cartridge, so test cartridges should be


disposed of as biohazardous waste, complying with loc al regulatory guidelines.

6. The system should be serviced only by authorized and qualified personnel. EDAN
does not assume any responsibility for damage or injury if modifications or repairs
are carried out by unauthorized personnel.

7. Service personnel should be familiar with the operation of the system. Refer to i15
Blood Gas and Chemistry Analysis System User Manual for details.

-4-
i15 Blood Gas and Chemistry Analysis System Service Manual Safety Guide

WARNING

8. Use and replace the substitutive parts provided or recommended by EDAN only.

9. Do not remove the battery while AC power is on.

10. Do not directly connect the battery to an electrical outlet.

11. Do not directly solder the lead wire or the battery terminal.

12. Do not touch the stitch of a calibrant fluid pack or the stitch of a vacuum pump.

13. Do not touch the roller or gear of a printer.

14. Do not touch the vane of an exhaust fan.

15. Never use an external electronic simulator under electromagnetic


environment, and never touch it by hand during an external simulator test.

CAUTION

1. The system is designed for continuous operation. Avoid splashing water over the
system.

2. Use this system at a temperature between 10°C and 31°C. Outside this range, the
system may produce inaccurate results.

3. Do not use the system in a dusty environment with bad ventilation or in the presence
of corrosives.

4. Do not operate the system rudely, or the system may produce incorrect test results.

5. Make sure that there is no intense electromagnetic interference source around the
analyzer, such as radio transmitters, mobile phones, etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment is likely to bring electromagnetic
interference.

6. Do not use the analyzer after its life cycle, and it should be disposed of according to
local regulations after its life cycle.

-5-
i15 Blood Gas and Chemistry Analysis System Service Manual Safety Guide

2.3 General Information

EN 61010-1: 2001, EN 61010-2-101: 2002,


Comply with
EN 61326-1: 2006, EN 61326-2-6: 2006
Degree of protection against
Ordinary equipment (Sealed equipment without waterproof)
harmful ingress of water
Degree of safety in presence of
Equipment not suitable for use in presence of flammable gases
flammable gases

EMC CISPR 11 Group 1, Class A

Mode of operation Continuous

-6-
i15 Blood Gas and Chemistry Analysis System Service Manual Installation

3 Installation
WARNING

To avoid electrical shock, do not remove any cover or board. Should problems or
malfunctions occur, unplug the power cord and power adaptor. Only qualified service
personnel should carry out servicing and troubleshooting.

3.1 Unpacking Inspection

Visually examine the package prior to unpacking. If there are any signs of mishandling or
damage, contact the carrier to claim for damage. After unpacking the device, customers should
follow the PACKING LIST to check the product carefully and to make sure no damage occurred
during transportation. Then, install the device according to the installation requirements and
procedures. If there is any problem, contact the manufacturer or the authorized local distributors
immediately.

3.2 Installation Requirements

3.2.1 Environmental Requirements


Location is of great importance for the smooth running of your analyzer. Prior to installing the
analyzer, choose a site that meets the following requirements:
Convenient for the analyzer to be connected to a grounded electrical receptacle in case it is
powered by AC power.

Keep the analyzer away from direct exposure to strong sunlight.


Ambient Temperature between 10°C and 31°C.

Relative Humidity of 25% - 80% (non-condensing).

Ambient Pressure within 70 - 106.6kPa (525 - 800mmHg).


Placed onto a clean and flat surface with good ventilation.

Keep the analyzer away from equipment with strong electric field and strong magnetic field.

Keep the analyzer away from explosive gases or vapors.

NOTE:

The requirements above also apply when your analyzer is powered by a rechargeable
lithium battery.
-7-
i15 Blood Gas and Chemistry Analysis System Service Manual Installation

3.2.2 Power Requirements


The analyzer needs to be connected to a grounded electrical outlet with the voltage between
100±10% VAC - 240±10% VAC and the frequency of 50/60Hz.

3.3 Installing the Analyzer

Please place the analyzer on a secure table surface with environments that meet the requirements
as described in section 3.2.

3.4 Connecting to AC Power

1. Insert the power adaptor into the power connector on the analyzer.

2. Plug the power cord into the power adaptor.

3. Plug the power cord into a grounded electrical outlet.

NOTE:

Make sure the power requirements as described in 3.2.2 are met.

To avoid the analyzer and other electronic devices being damaged by electrical
power spikes, a surge protector is recommended.

3.5 Replacing a Calibrant Fluid Pack

WARNING

1. Never replace a calibrant fluid pack when the analyzer is off.

2. A calibrant fluid pack is intended for single use only. If a calibrant fluid pack is
removed from the system, it can not be inserted into the system again.

Follow the steps below to replace a calibrant fluid pack:

1. Examine the expiration date on the package of a calibrant fluid pack to ensure it has not
expired.

2. Remove the calibrant fluid pack from its package, and equilibrate it to room temperature.
The calibrant fluid pack needs to stand at room temperature for at least 24 hours.

3. Wipe any moisture from the foil pouch with a dry clean cloth.

-8-
i15 Blood Gas and Chemistry Analysis System Service Manual Installation

4. On the Main screen, press to go to the System Setup screen.

5. Press . The system will go to the screen below:

Figure 3-1 Open Chamber Door

6. Open the foil pouch, and remove the calibrant fluid pack from it.

NOTE:

Avoid tearing the bar code on the foil pouch.

7. Remove the calibrant fluid pack cap, and remove the key by pushing it with your finger in
the direction as shown by the arrow in the picture below.

Key Valve

NOTE:
-9-
i15 Blood Gas and Chemistry Analysis System Service Manual Installation

Avoid pushing the valve when removing the key.

8. Unlock the calibrant fluid pack chamber lock with the key, and then pull the calibrant fluid
pack ejector to open the calibrant fluid pack chamber door.
9. Remove the used calibrant fluid pack from the system. The system will go to the screen for
the next procedure, and the scanner will be turned on automatically.

Figure 3-2 Remove Old Calibrant Fluid Pack


10. Scan the bar code on the new calibrant fluid pack foil pouch with the bar code scanner.

If the bar code is scanned successfully, the system will beep and the scanner will be turned
off automatically. If the scanned data is valid, the system will display the screen for the next
procedure. If the scanned data is invalid, a message will pop up to prompt you.

If the scanner is turned off automatically, press first, and then scan the bar code.

Figure 3-3 Scan Bar Code

11. Insert the new calibrant fluid pack into its chamber, and push it gently to make sure that it
- 10 -
i15 Blood Gas and Chemistry Analysis System Service Manual Installation

clicks into place. The system will go to the screen for the next procedure.

Figure 3-4 Install Calibrant Fluid Pack


12. Close the chamber door.

Figure 3-5 Close Chamber Door

13. Lock the calibrant fluid pack chamber lock to close the chamber door securely.

14. Press OK in the pop up message. The system will then go to the Main screen.

Figure 3-6 Replacement Success

- 11 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

4 Functional Checks
This section describes the procedure of a complete functional test to support recommended
preventive-maintenance schedules.

It is unnecessary to open the analyzer case for functional checks.

WARNING

Functional checks should be performed by service personnel whenever installation,


servicing, or troubleshooting is performed.

4.1 Switching On Check

Press the On/Off button on the left side of the analyzer to turn it on.

Check if the power indicator lights up, and if the screen lights up in a few seconds.

If any failure occurs, refer to section 7.1 Booting Failures for details.

4.2 LCD Screen Check

Observe if some characters are missing, or if there are bright spots or dark shadows on the LCD
screen. Observe if fonts and symbols displayed on the LCD screen are normal.

If any failure occurs, refer to section 7.2 Display Failures for details.

4.3 Touch Screen Check

Touch any available button on the touch screen to check if it is working properly.

If you touch the center of a button but the analyzer does not execute the right operation, calibrate
the touch screen. Refer to section 4.4 Dignostics for details.

If any operation failure occurs, refer to section 7.3 Touch Screen Failures for details.

4.4 Diagnostics

This function let you view the information for some modules, check the functioning of the pump,
valve, printer, and internal electronic simulator, calibrate the touch screen, export data, reset the
system password, reset the system setup to the factory setup, online update the analog board,

- 12 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

scan the system ID bar code, view software running logs, and clear data stored in the analyzer. It
helps to troubleshoot.

NOTE:
Only service engineers and engineers from the manufacturer can perform this action.

Figure 4-1 Diagnostics Screen

Follow the steps below to perform diagnostics:

1. Press to go to the Diagnostics screen.

2. Refer to the table below to perform actions.

No. To Press
View the analog board information. Analog board software
version (it should be the latest version), front end board ID (it
should be consistent with the tracing list), ambient
1. temperature (it should be 10 C - 35 C), ambient pressure (it Analog Board
should be 730mmHg - 760mmHg), the heater 1 temperature
(it should be 10 C - 35 C), and the heater 2 temperature (it
should be 10 C - 35 C).
View the power board information. Power board software
version (it should be the latest version), 5V voltage (it should
be 4.5V - 5.5V), 12V voltage (it should be 11.5V - 12.5V),
printer voltage (it should be 12V - 20V), input current (it
2. Power Board
should be smaller than 1A), power board temperature (it
should be 10 C - 50 C), the rate of the exhaust fan (it should
be 0 8000 cycles/min, and for temperature below 29 C, the
exhaust fan does not run).

- 13 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

View the main control board information. Main control board


3. Main Board
software version (it should be the latest version).
4. Check whether the pump is working properly. Pump
5. Check whether the valve is working properly. Valve
Check whether the printer is working properly.
The contents below are printed:
6. 1234567890 Printer
abcdefghijklmnopqrstuvwxyz
AB C D E F G H I J K LMN O PQ R ST U VW XY Z
Check whether the internal electronic simulator is working
7. Inter Simulator
properly.
Calibrate the touch screen (Touch the symbol on the screen,
five symbols appear in succession). It should be verified after
8. the completion of the calibration test (Touch the symbol on Screen Calibration

the screen again, five symbols appear in succession).

9. Export the original data stored in the analyzer to *.txt format. Export Data
10. Reset the system password to the factory one. Reset Password
11. Restore the system setup to the factory setup. Factory Setup
12. Online update the analog board. Analog Update
13. Scan the system ID bar code. System ID
14. View the logs about the software running. System Log
Delete all the data stored in the analyzer. Perform this action
15. Clear Data
after scanning the system ID bar code before shipment.

3. Press Return to go back to the System Setup screen.

4.5 External Simulator Test

Follow the procedures below to perform an external simulator test:

1. Press the On/Off button on the left hand side of the analyzer to turn it on.

2. Enter the user name and password manually, and then press .

To scan the bar code of the user name, press first, and then scan the bar code.

- 14 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

Figure 4-2 Enter User Name and Password

3. On the Main screen, press to go to the Quality Control screen, and then press .

4. Gently insert the external electronic simulator into the cartridge port, and press down to
ensure it clicks into place. If the electronic simulator is inserted properly, the system will
automatically perform a simulator test.

Figure 4-3 Insert External Simulator

NOTE:

Avoid touching the contact pads.

Never touch the simulator after inserting it into the system.

5. The system automatically starts to perform an external simulator test.

- 15 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

Figure 4-4 Measurement in Progress

6. The system automatically displays the test result upon the completion of the test.

Figure 4-5 External Simulator Test Result

7. View the result.

NOTE:

Upon the completion of the test, the message Please Remove Cartridge will be
displayed on the status bar at the bottom of the screen, and the simulator will be
ejected.

If the system passes the external simulator test, it can be used to analyze samples.

If the system fails the external simulator test, re-perform the test after the simulator
has been cleaned and completely dried or use another simulator to re-perform the
- 16 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

test. If the external simulator test succeeds, the analyzer can be used to run samples.
If the external simulator test still fails, run an internal simulator test. If the internal
simulator test succeeds, perform the external simulator test after the POGO PIN has
been cleaned and completely dried. If the external simulator test fails, the cam needs
to be replaced. If the external simulator test succeeds, the analyzer can be used to
run samples. If the internal simulator test fails, check the wire between the cam and
the analog board. If the wire connects well, run the internal simulator test again. If the
internal simulator test still fails, replace the cam.

8. Press Home to go to the Main screen.

9. Remove the external simulator from the system, and put it back.

NOTE:

Never remove the external simulator until the test is complete.

4.6 Control Test

Follow the steps below to perform a control test:

1. Examine the package label of controls to ensure they have not expired.

2. Remove an ampoule from the box of controls and equilibrate it to room temperature.

If oxygen is to be measured, the ampoule needs to stand at room temperature for at least 4
hours. If not, the ampoule needs to stand at room temperature for 30 minutes.

3. Press the On/Off button on the left hand side of the analyzer to turn it on.

4. Enter the user name and password manually, and then press .

To enter the user name with the bar code scanner, press first, and then scan the user
name bar code.

- 17 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

Figure 4-6 Enter User Name and Password

5. On the Main screen, press to go to the Quality Control screen.

6. Select the desired control type.

Press to perform a control test for blood gases and blood chemistries.

Press to perform a control test for Hct.

7. Press Scan Barcode, and scan the bar code on a new cartridge foil pouch.

If the bar code is scanned successfully, the system will beep and the scanner will be
automatically turned off. If the scanned data is valid, the system will display the screen for
the next procedure. If the scanned data is invalid, a message will pop up to prompt you.

Figure 4-7 Scan Bar Code

8. Open the foil pouch and remove the cartridge from it.

- 18 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

NOTE:

Avoid tearing the bar code on the foil pouch.

For sample introduction with a capillary tube or an ampoule, insert a capillary


adaptor/ampoule adaptor into the fillport after removing the cartridge.

When inserting an ampoule adaptor into the fillport of a test cartridge, make sure the
indent of the adaptor is on the upside as shown in the picture below:

9. Press Scan Barcode, and scan the bar code on the controls user manual.

If the bar code is scanned successfully, the system will beep and the scanner will be
automatically turned off. If the scanned data is valid, the system will display the screen for
the next procedure. If the scanned data is invalid, a message will pop up to prompt you.

Figure 4-8 Scan Bar Code

- 19 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

NOTE:

Keep the controls user manual for a future control test use.

10. Mix the controls completely by shaking the ampoule gently, and then tap the tip of the
ampoule carefully with your fingernail to remove any solution.

Figure 4-9 Mix Controls and Insert Cartridge

NOTE:

Avoid heating the ampoule with your hands when shaking the ampoule.

11. Open the ampoule by snapping off the top and immediately transfer control solution by
slowly drawing an appropriate amount of solution with a syringe or capillary tube from the
bottom of the ampoule.

NOTE:

When using an ampoule for sample introduction, you need not transfer control
solution. Insert the ampoule into the adaptor after opening it, and directly go to step
13.

Take protective measures when opening the ampoule, such as using gloves, tissue,
etc.

12. Insert the syringe or capillary tube into the fillport of the cartridge.

NOTE:

When using a syringe, discard the first 2 drops of solution first, then remove the
needle from it, and finally insert it into the fillport.
- 20 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

When using a capillary tube, directly insert the capillary tube into the adaptor till the
tube reaches the interface between the adaptor and the cartridge.

To avoid inaccurate test results, make sure there are no bubbles in the sample. If
bubbles continually exist, use a new ampoule and syringe or capillary tube to collect
samples again.

13. Gently insert the cartridge into the cartridge port, and carefully press down to ensure that it
clicks into place.

For a valid cartridge, the indicator in the cartridge port will turn green, and the system will
automatically aspirate calibrant. For an invalid cartridge, the indicator will turn red, the
cartridge will be ejected and a message will pop up to prompt you.

NOTE:

The cartridge can not be removed from the analyzer until the measurement is
complete.

Never inject the sample. It will be aspirated automatically.

14. The system automatically aspirates calibrant.

Figure 4-10 Aspirating Calibrant

15. The system automatically performs calibration.

- 21 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

Figure 4-11 Calibration in Progress

16. The system automatically aspirates samples when calibration is complete.

Figure 4-12 Sampling

17. The system automatically analyzes the sample upon the completion of sampling.

Figure 4-13 Measurement in Progress

- 22 -
i15 Blood Gas and Chemistry Analysis System Service Manual Functional Checks

18. The system automatically displays the test results upon the completion of the test.

Figure 4-14 Control Test Results

19. View the results.

NOTE:
The system will indicate whether the results are within or outside the acceptable
ranges with Under Control/Out of Control.
If a parameter fails calibration, the system will not be able to determine whether it is
under control, and will display Calibration Failure to prompt you.
The system will not report the result for a parameter failing the control test in the
patient sample analysis, if QC Lockout function is enabled in Setup. To report the
result for the parameter, repeat the control test till the parameter passes it.
If the results are outside the acceptable ranges, check the following items first, and
then perform another test.
Refer to the user manual to confirm that the test procedures are correct.
Test cartridges and controls are stored properly and have not expired.
The system passes the electronic simulator test.
If all the above items are verified, but the results are still outside the acceptable
ranges, please return the analyzer to EDAN for service.
20. Remove the test cartridge from the system.
21. Press Print to print the results.
22. Press Home to return to the Main screen.

- 23 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

5 Maintenance
5.1 Maintenance Inspection

(1) Visual Inspection


Before using the system, the user should:
Inspect the analyzer for signs of damage.
Check the external cables for splits, cracks or signs of twisting.
Do not use the system if any damage occurs until the system is repaired by the professional
service personnel of EDAN or the authorized distributors.
(2) Routine Inspection
The following safety checks should be performed at least once every 24 months by a
qualified person who has adequate training, knowledge, and practical experience to perform
these tests.
Inspect the analyzer and accessories for mechanical and functional damage.
Inspect the safety related labels for legibility.
Verify the analyzer functions properly as described in the instructions for use.
If the analyzer is not functioning properly or fails any of the above tests, it has to be
repaired.

WARNING

Failure on the part of the responsible individual institution employing this equipment to
implement a satisfactory maintenance schedule may cause undue equipment failures.

(3) Mechanical Inspection


Make sure all exposed screws are tightly fixed.

5.2 Maintenance of the Analyzer


Keep the surface of the analyzer clean, free of dust and dirt.
Prevent any liquid from seeping into the analyzer.
Avoid shaking the analyzer violently when moving it to another place.
Do not scratch or damage the screen.
The gathering of dew on the screen may occur with abrupt temperature or humidity changes.
It is recommended that the system should be used in a stable environment.
- 24 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

6 Principle Introduction
The system utilizes potentiometry and amperometry to determine the concentrations of blood
gases and blood chemistries, and utilizes conductivity to determine the concentration of Hct.
Many parameters can also be calculated such as cH+, HCO3 -act, HCO3 -std, BE(ecf), BE(B),
BB(B), and so on.

6.1 Measurement Method

Methods

Measurements are performed on undiluted specimens. Undiluted methods are also called direct
methods, whereas methods that dilute the sample are called indirect methods.

For electrolytes, direct methods measure the free ion concentration of analyte (apparent or free
ion activity) per unit volume of plasma water, and indirect methods measure the concentration of
analyte per unit volume of plasma. It is known that the direct method gives the clinically
significant result for electrolytes. When there is disagreement between the methods, such as when
the patient has abnormal total protein or lipid levels, it results from interference on the indirect
method. At normal levels of protein and lipids the systematic offset between methods is often
corrected for in commercial direct measuring instruments so that the normal ranges for all
instruments are in agreement. Sensors have been calibrated in the manufacture so that normal
ranges are in agreement with indirect reference methods at normal levels of total protein and
lipids.

Direct measurement of hematocrit by the conductometric technique gives a result related to the
non-conducting excluded volume fraction of the sample. Red blood cell volu me is the
predominant component of the nonconducting volume, but proteins, lipids, and white blood cells
also contribute. Elevated hematocrit readings are expected at abnormally elevated levels of these
components. Decreased hematocrit readings are expected at abnormally low levels of protein,
such as found in hemodiluted samples taken from cardiopulmonary bypass. Osmotic imbalance
causes a discrepancy between direct (conductometric, spun) and indirect (Coulter) measurements
because of variation in the meancell volume.

Sensors

Sensor refers to the electrode embedded in test cartridges. There are three different types of
sensors:

- 25 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Potentiometric Sensor

Amperometric Sensor

Conductometric Sensor

Potentiometry: A potential is recorded using a voltmeter, which relates to the concentration of


the sample. A reference electrode is used to provide a stable, fixed potential against which other
potential differences can be measured. This measurement technique is used for pH, pCO2 and
electrolytes.

Amperometry: The magnitude of an electrical flow of current is proportional to the


concentration of the substance being oxidized or reduced at an electrode. This measurement
technique is used for pO2, glucose and lactic acid.

Conductivity: The specific impedance of a sample as measured by two conducting electrodes


held at a constant voltage is directly proportional to the conductive properties of the sample. This
technique is used for Hct.

6.2 Determination of Test Results

6.2.1 Determination of the Analyte Concentration


The concentration of the analyte is determined with the potentiometric and amperometric sensors.
For both sensors, the analyte concentration is calculated using:

The known analyte concentration of calibrant solution;

The measured voltage or current of the calibrant solution.

The measured voltage or current of the sample.

For potentiometric sensors, the analyte activity in the sample is calculated from the Nerst
equation:

Esample Ecalibrant sample calibrant)

Where E denotes the potential, denotes the activity of an ion, and S denotes the slope of the
sensor.

- 26 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

6.2.2 Determination of Cell Concentration


Hct

In whole blood, the cellular constituents, red and white blood cells and platelets do not conduct
electricity, while plasma does. For a sample whose electrolyte concentration is given, the more
cells, the less conductivity. The total cell concentration of the whole blood is determined from:

The known value of the electrolyte concentration in the calibrant.

The measured electrolyte concentration of the sample.

The measured conductivity signal generated by the calibrant.

The measured conductivity signal generated by the sample.

CPB

When samples with abnormally low protein levels are tested, the system need to use the CPB
compensation algorithm. The CPB compensation algorithm is specially intended for use when
samples are taken from patients on cardiopulmonary bypass. It also applies to the adult whose
protein levels are abnormally low.

6.3 Equations for Calculated Parameters

cH+

Hydrogen ion concentration

cH+ = 10 (9 pH) [nmol/L]

HCO3-act

Bicarbonate ion concentration

HCO3-act = 0.0307 × pCO2 × 10(pH - 6.105) [mmol/L]

HCO3-std

Bicarbonate ion concentration normalized to a pCO2 40mmHg

HCO3-std = 24.5 + 0.9 × A + [(A - 2.9)2 × (2.65 + 0.31 × tHb (est))] / 1000 [mmol/L]

Where A = BE (B) [0.2 × tHb (est) × (100 - sO2 (est))] / 100, tHb (est) can be entered by users,
and the default value for tHb (est) is 15g/dL.

- 27 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

BE (ecf)

Base excess (ecf)

BE (ecf) = HCO3-act - 24.8 + (16.2 × (pH - 7.40)) [mmol/L]

BE (B)

Base excess (B)

BE (B) = (1 - 0.014 × tHb (est)) × [HCO 3-act - 24.8 + (1.43 × tHb (est) + 7.7) × (pH - 7.40)]
[mmol/L]

Where tHb (est) can be entered by users, and the default value for tHb (est) is 15g/dL.

BB (B)

Buffer Base

BB (B) = BE (B) + 41.7 + 0.42 × tHb (est) [mmol/L]

Where tHb (est) can be entered by users, and the default value for tHb (est) is 15g/dL.

ctCO2

Total carbon dioxide

ctCO2 = HCO3 -act + 0.0307 × pCO2 [mmol/L]

Ca++ (7.4)

The ionized calcium concentration of blood normalized to pH 7.4

Ca++ (7.4) = Ca ++ × 10 [0.178 × (pH - 7.40)] [mmol/L]

AnGap

An approximation of the difference between measured cations and measured anions in the
sample

AnGap = (Na+ + K+) - (Cl- + HCO3-act) [mmol/L]

tHb (est)

An estimation of the hemoglobin contained in the sample

tHb (est) = MCHC × Hct / 100 [g/dL]

- 28 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Where MCHC is the mean corpuscular hemoglobin concentration, and can be entered by users.
The default value for MCHC is 34g/dL. The unit for Hct is %PCV.

sO2 (est)

An estimation of hemoglobin oxygen saturation: a ratio of hemoglobin bound to oxygen to the


total amount of hemoglobin able to bind oxygen

[%]

pO2 (A-a)

Alveolar-arterial oxygen tension difference

pO2 (A-a) = pO2 (A) - pO2 (a) [mmHg]

pO2 (a/A)

Arterial-alveolar oxygen tension ratio

pO2 (a/A) = pO2 (a) / pO2 (A) [mmHg]

RI

Respiratory index: the ratio of the alveolar-arterial blood oxygen-pressure difference to arterial
pO2

RI = pO2 (A-a) / pO2 (a)

pO2 / FIO2

The ratio of arterial pO2 to the fraction of inspired oxygen

pO2 / FIO2 = pO2 / FIO2 [mmHg]

pH (T)

pH value corrected for entered patient temperature

pH (T) = pH [0.0147 + 0.0065 × (pH 7.4)] (T 37)

cH+ (T)

Hydrogen ion concentration corrected for entered patient temperature

- 29 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

cH+ (T) = 10 (9 pH (T))


[nmol/L]

pCO2 (T)

pCO2 corrected for entered patient temperature

pCO2 (T) = pCO2 × 100.019(T - 37) [mmHg]

pO2 (T)

pO2 corrected for entered patient temperature

[mmHg]

pO2 (A-a) (T)

Alveolar-arterial oxygen tension difference corrected for entered patient temperature

pO2 (A-a) (T) = pO2 (A) (T) - pO2 (a) (T) [mmHg]

pO2 (a/A) (T)

Arterial-alveolar oxygen tension ratio corrected for entered patient temperature

pO2 (a/A) (T) = pO2 (a) (T) / pO2 (A) (T) [mmHg]

RI (T)

Respiratory index: the ratio of the alveolar-arterial blood oxygen-pressure difference to arterial
pO2 when both values are corrected for patient temperature

RI (T) = pO2 (A-a) (T) / pO2 (a) (T)

pO2 (T)/FIO2

The ratio of arterial pO2 to the fraction of inspired oxygen corrected for the entered patient
temperature

pO2 (T) / FIO2 = pO2 (T) / FIO2 [mmHg]

mOsm

mOsm is generally determined by Na+, K+, Glu and Bun.

- 30 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

mOsm = 1.86 × (Na + + K+) +Glu + BUN +9


In the equation:
Glu is the Glu test result by performing test with Blood Gas and Chemistry Test Cartridge on i15
Blood Gas and Chemistry Analyzer and the unit for Glu test result is mmol/L. If Glu can not be
measured directly, its default value for Glul is 5mmol/L.
BUN is the BUN test result by performing test with Blood Gas and Chemistry Test Cartridge on
i15 Blood Gas and Chemistry Analyzer and the unit for BUN test result is mmol/L. If BUN can
not be measured directly, its default value for BUN is 3mmol/L.

6.4 System Principle Block Diagram

On the left side of the analyzer, there are 6 interfaces:


RJ45 Interface (Network)
DB9 Interface (RS232)
4 USB Interfaces

6.4.1 RJ45 Interface

The analyzer is configured with a RJ45 interface. It is used to connect the analyzer to a computer
for updating or to a DMS.

6.4.2 DB9 Interface

The analyzer is configured with a DB9 interface. It allows engineers from the manufacture and
service personnel to troubleshoot

6.4.3 USB Interface

The analyzer is configured with 4 USB interfaces. It is used to connect external devices
supporting USB protocol to the analyzer, such as a USB drive, a mouth, a printer, a scanner.

NOTE:

Only the following external scanners should be connected to the analyzer through
USB ports: Honeywell 1900.

Only the following external printers should be connected to the analyzer through
USB ports: HP LaserJet P401DN, and HP LaserJet P1606DN.

- 31 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

7 Troubleshooting
EDAN supports replacement of PCB and major subassemblies for the analyzer. When
replacement is necessary, follow the procedures described in chapter 9 Disassembling the
Analyzer.

7.1 Booting Failures

Phenomenon Possible Cause Solution


Check whether the power cord is
damaged, and whether the power
adaptor is damaged. If so, replace
No AC mains are them. Check whether the power
connected, and the battery cord and power adaptor are
is empty. correctly connected to the power
socket of the analyzer and the
After switching on, LCD has hospital AC outlet. If not, connect
no display; the power them correctly.
indicator is off; the exhaust Check whether the wires
fan does not run. connecting the power button are
The connecting wires are loose, and whether the wires
loose. connecting the upper assembly and
lower assembly are loose. If so,
connect them again.
Power board is defective. Replace the power board.
Main control board is
Replace the main control board.
defective.
The analyzer is struck by
strong high voltage, such as Check the power supply.
Abrupt switching off. lightening strike.
Bad power supply
Replace power supply.
performance

- 32 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Check whether the power cord and


power adaptor connect well,
whether the power adaptor and the
Bad connection power socket connect well, and
Abrupt switching off. whether the power cord and the
electrical outlet connect well. If not,
connect them well.
Main control board is
Replace the main control board.
defective.

7.2 Display Failures

Phenomenon Possible Cause Solution


Check whether the wires
After switching the The connecting wires are connecting the LCD are loose, and
on,
power indicator is on, and loose. whether the wires connecting the
the exhaust fan works upper assembly and lower assembly
properly, but LCD has no are loose. If so, connect them again.
display, or display is LCD is defective. Replace the LCD.
abnormal or dim. Main control board is
Replace the main control board.
defective.

7.3 Touch Screen Failures

Phenomenon Possible Cause Solution

Check whether the connecting


The connecting wires are
wires are loose. If so, connect them
loose.
again.
The touch screen does not
response after pressing it. The work area of the touch Adjust the silicon loop so that the
But it responses when screen is covered by the work area of the touch screen is not
operating the system with a silicon loop. covered.

mouse. Touch screen drive board is Replace the touch screen drive
defective. board.

Touch screen is defective. Replace the touch screen.

- 33 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Main control board is


Replace the main control board.
defective.

Calibrate the touch screen


The touch position is The touch screen is not
according to the section 4.5
ineffective. calibrated.
Diagnostics.

7.4 Operation Failures

Phenomenon Possible Cause Solution


Remove the test cartridge ejector cap,
The test cartridge is
pull the middle hole with a pointed
inserted into the system
object, and remove the test cartridge
prior to a test.
from the system.
A test cartridge can not be Reboot the system. After about 1
The system is shut down
ejected automatically. minute, the test cartridge will be
during a test.
ejected automatically.
Reboot the system. After about 1
The system is dead during
minute, the test cartridge will be
a test.
ejected automatically.
The The ambient temperature Use the system in an environment
is not within 10°C - with the ambient temperature within
appears. 31°C. 10°C - 31°C.
Forget to connect a
Connect a syringe/capillary tube to
syringe/capillary tube to
the fillport.
the fillport.
Perform the test with enough
The sample is not enough.
samples.

Replace the calibrant fluid pack


Improper Removal. Please The calibrant fluid pack is
following the user manual
removed improperly.
instructions.
appears.
The date setup is
Set the date to be the current date.
incorrect.

- 34 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Use a new test cartridge that has not


The test cartridge expired.
expired.
The date setup is
Set the date to be the current date.
incorrect.
The calibrant fluid pack Use a new calibrant fluid pack that
expired. has not expired.

7.5 Printer Failures

Phenomenon Possible Cause Solution

Check whether the connecting


The connecting wires are
There is no response after wires are loose. If so, connect them
loose.
pressing Print. again.

The printer is defective. Replace the printer.

The printer casing was not Open the printer casing, and shut
Paper jam or slant
shut properly. the casing again.

The printer head is


The printout is not clear. Replace the printer head.
defective.

7.6 Network Failures

Phenomenon Possible Cause Solution


The network wire is
Replace the network wire.
damaged.
Set the network according to the
The analyzer can not be Network setup error
user manual.
connected to a network.
Main control board is
Replace the main control board.
defective.
Power board is defective. Replace the power board.
Set the network according to the
Network setup error
Wi-Fi connection failure user manual.
Wi-Fi board is defective. Replace the Wi-Fi board.

- 35 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Main control board is


Replace the main control board.
defective.

7.7 Diagnosis Database

NOTE:
Only service engineers and engineers from the manufacturer can enter the diagnosis
database.
Diagnosis data help to troubleshoot. The diagnosis database displays diagnosis data for the most
recent month by default, and it can store up to 500 data entries. The following operations can be
performed in the diagnosis database: exporting diagnosis data to a removable disk (such as a
USB drive), viewing the details of diagnosis data, searching for and printing diagnosis data, etc.

7.7.1 Searching for Diagnosis Data

1. On the Database screen, press to open the diagnosis database.

Figure 7-1 Diagnosis Database Screen

2. Press Search.

- 36 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Figure 7-2 Enter Search Conditions

3. Enter the search conditions, and press OK.

4. The system displays the results.

Figure 7-3 Search Results

5. Press Return to go back to the Database screen.

7.7.2 Viewing Details of Diagnosis Data


1. Select the desired diagnosis data.

2. Press Details.

3. The system displays the details.

- 37 -
i15 Blood Gas and Chemistry Analysis System Service Manual Troubleshooting

Figure 7-4 Details of Diagnosis Data

4. View the details.

5. Press Return to go back to the Database screen.

7.7.3 Exporting/Printing Diagnosis Data

1. On the Database screen, press to open the diagnosis database.

2. Select the desired diagnosis data.

To Do this

Export Insert a removable disk into the analyzer, and then press Export.

Print Press Print.

NOTE:

If no diagnosis data is selected before pressing Export/Print, all of the data stored in the
diagnosis database will be exported / printed.

Here is an example of diagnosis report:

- 38 -
i15 Blood Gas and Chemistry Analysis System Service Manual Module s Malfunction Verification

8 Module s Malfunction Verification


When a module is suspected of malfunction, verify it with the methods described in this chapter.

NOTE:

For all the interfaces, the first pin is the one with a square solder pad.

8.1 Verification of Main Control Board

Follow the steps below to verify the main control board:

1) Disassemble the analyzer following the procedures described in section 9.2.

2) Turn on the analyzer (Powered by AC or battery).

3) Measure the voltage to earth of the pins listed below with a multimeter:

Item Pin Reference Range

1. TP8 1.71~1.89V

2. TP9 3.13~3.47V

3. TP12 0.95~1.05V

4. TP13 0.95~1.05V

5. TP14 2.37~2.63V

6. TP49 3.13~3.47V

- 39 -
i15 Blood Gas and Chemistry Analysis System Service Manual Module s Malfunction Verification

4) Compare the measurement results with the reference ranges in the list. If there is a result
exceeding its reference range, it is confirmed that the main control board malfunctions, and
it is recommended to replace the main control board.

8.2 Verification of the Power Board it Failure

Follow the steps below to verify the power board:

1) Disassemble the analyzer following the procedures described in section 9.2.

2) Remove the upper assembly.

3) Turn on the analyzer (Powered by AC or battery).

4) Measure the voltage to earth of the pins listed below with a multimeter.

Item Pin Reference Range

1. TP12 11.40~12.60V

2. TP13 4.75~5.25V

3. TP9 4.75~5.25V

- 40 -
i15 Blood Gas and Chemistry Analysis System Service Manual Module s Malfunction Verification

5) Compare the measurement results with the reference ranges in the list. If there is a result
exceeding its reference range, it is confirmed that the power board malfunctions, and it is
recommended to replace the power board.

8.4 Verifying Malfunction of the Touch Screen Drive Board

Follow the steps below to verity the touch screen drive board:

1) Disassemble the analyzer following the procedures described in section 9.2.

2) Remove the upper assembly.

3) Turn on the analyzer (Powered by AC or battery).

4) Measure the voltage to earth of the pins listed below with a multimeter.

Item Pin Reference Result

1. TP<+3V3> 3.3V±0.165V

- 41 -
i15 Blood Gas and Chemistry Analysis System Service Manual Module s Malfunction Verification

5) Compare the measurement results with the reference ranges in the list. If there is a result
exceeding its reference range, it is confirmed that the touch screen drive board malfunctions,
and it is recommended to replace the touch screen drive board.

8.5 Verifying Malfunction of the Scanner

Follow the steps below to verify the bar code scanner:

1) Disassemble the analyzer following the procedures described in section 9.2.

2) Remove the bar code scanner.

3) Connect the bar code scanner to a computer through the USB port.

4) Operate the bar code scanner with the computer.

5) If you fail to operate the bar code scanner, it is confirmed that the bar code scanner
malfunctions, and it is recommended to replace the bar code scanner.

NOTE:

It is necessary that the SNAPI Sample Application software has been loaded on the
computer before connecting the bar code scanner to the computer. If the software has

been loaded, the symbol will appear.

- 42 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

9 Disassembling the Analyzer


WARNING

1. Only qualified service personnel should open the analyzer housing.

2. Switch off the analyzer and remove the power supply before disassembling the
analyzer.

3. After any repair of the analyzer, perform functional checks prior to use.

9.1 Tools Required

1. A cross-head screwdriver

2. A M3 nut driver

9.2 Disassembling the Main Unit


The main unit consists of two parts: upper assembly and lower assembly.

Follow the steps below to disassemble the main unit:

1. Carefully place the analyzer upside down on a flat surface covered with cloth or other
protecting pad.

2. With a cross-head screwdriver, remove the 7 screws securing the upper assembly to the lower
assembly.

3. Hold both upper and lower assembly simultaneously, and place the analyzer upward again.

4. Unlock the calibrant fluid pack chamber lock with the key, and then pull the calibrant fluid
pack ejector to open the calibrant fluid pack chamber door.

5. Slightly lift up the upper assembly, and disconnect the connecting wire on the shrouded &
ejector IDC header to separate the upper assembly and the lower assembly.

6. Carefully lift up the upper assembly, and place it upside down on the left side of the lower
assembly.

7. Disconnect the printer cable to earth to disassemble the upper and lower assembly.
- 43 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

2 4

1 Upper assembly

2 Upper assembly & lower assembly connecting wire

3 Printer cable to earth

4 Shrouded & ejector IDC header

5 Lower assembly

9.3 Disassembling the Lower Assembly

9.3.1 Replacing the Main Control Board

Follow the steps below to remove the main control board:

1. Separate the upper assembly and the lower assembly.

2. Disconnect all the wires on the main control board.

3. Remove the 5 screws securing the main control board.

4. Hold the left side of the main control board with your hand, sway it back and forth, and
remove the main control board carefully.

- 44 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

To install a main control board:

Install a main control board in the reverse order. Refer to the following illustration when
connecting wires, and ensure they are all correctly connected.

1 2 3 4 5

No. Connect to

1 Upper assembly

2 Analog communication board

3 Speaker

4 Bar code scanner

5 Wi-Fi board

9.3.2 Replacing the Wi-Fi Board

Follow the steps below to remove the Wi-Fi board:

1. Separate the upper assembly and the lower assembly.

2. Disconnect the wire on the Wi-Fi board.

3. Remove the Wi-Fi antenna socket.

4. Remove the 2 screws securing the Wi-Fi board.

5. Remove the Wi-Fi board carefully.


- 45 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

To install a Wi-Fi board:

Install a Wi-Fi board in the reverse order.

9.3.3 Replacing the Chamber Door Detection Board

Follow the steps below to remove the chamber door detection board:

1. Separate the upper assembly and the lower assembly.

2. Disconnect the wire on the chamber door detection board.

3. Remove the screw securing the chamber door detection board.

4. Remove the chamber door detection board carefully.

To install a chamber door detection board:

Install a chamber door detection board in the reverse order.

- 46 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

9.3.4 Replacing the Ambient Temperature Detection Board

Follow the steps below to remove the ambient temperature detection board:

1. Separate the upper assembly and the lower assembly.

2. Disconnect the wire on the ambient temperature detection board.

3. Remove the screw securing the ambient temperature detection board.

4. Remove the ambient temperature detection board carefully.

To install an ambient temperature detection board:

Install an ambient temperature detection board in the reverse order.

9.3.5 Replacing the Analog Board

Operation Procedures NOTE

Remove the screws located


at the left picture.

- 47 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the connection


cable between the main
control board and analog
board as the left picture
shows.

Remove the connection


cable between the main
control board and the
trumpet as the left picture
shows.

Open the cover of the FPC


connector which connect
the main control board and
barcode module.

- 48 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Open the cover of the FPC


connector which connect
the main control board and
WIFI module.

Lift the main control board


upward.

- 49 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the connection


cable of chamber detection
board.

Remove the FPC cable


between the analog board
and the cam.

- 50 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the copper pillars


from the board with the Copper pillar
help of some tools. Then
the board can be separated.

Reassemble a new analog


board as the reverse
procedures.

9.3.6 Replacing the Cam Machine

Steps NOTE

Remove the seven screws as the


picture shows with a cross-head
screwdriver.

Dismantle the
upper
assembly

- 51 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Lift the upper assembly for 10cm.

Disconnect the cable from the socket


located at the lower assembly.

Disconnect the connection cable


which connects the upper and lower
cover.

Separate the upper and lower cover


and then pull out the ground wire.

- 52 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Open the black cover of the FPC


connector.

Gently remove the FPC cable.

Dismantle the Pull out the serial cable and trumpet


main control wire.
board (PCBA)
Remove the five screws which are
used to fix the main control board
(PCBA).

Gently lift the main control board


(PCBA) upward.

Remove the screw which is used to fix


the detection board and then the
detection board can be separated from
the module.

- 53 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the four screws which are


used to fix the cam and then the cam
be removed.

Open the black cover of the FPC


connector and then remove the FPC
cable from the connector.

Hold the cam connection cable tightly


and then disconnect it from the cable.

Connect the FPC cable with a new


cam and then close the black cover of
the connector.

Press down the cam connection cable


and fix the cam assembly with the
4-M3*8 screw.

Assemble
Cam

- 54 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Fix the temperature detection board


with 1-M3*8 screw.

Mount the main control board onto the


analog board (please pay attention to
the connection board between the
main control board and the analog
board).

Assemble
main control
board.
Tightly secure the analog board and
the main control board with 5-M3*6
screws screwed by torque (3.0 4.0
kg.cm)

Connect serial cable, trumpet wire,


FPC cables.

Connect the grounding cover

Assemble
upper and
lower cover

- 55 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Connect the connection cable which


connects the upper and lower cover.

Secure the screws.

9.3.7 Replacing the Power Board

Follow the steps below to remove the power board:

1. Separate the upper assembly and the lower assembly.

2. Remove the main control board and the cam machine.

3. Disconnect all the wires on the power board.

4. Remove the 5 screws securing the power board.

5. Remove the two screws fixing the fan and then the fan can be removed.

6. Remove the power board carefully.

To install a power board:

Install a power board in the reverse order. Refer to the following illustration when connecting
wires, and ensure they are all correctly connected.

- 56 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

1 4
3

No. Connect to

1 Interface connection board

2 Exhaust fan

3 Battery

4 NTC thermal resistor

9.3.8 Replacing the Power Interface Board

Follow the steps below to remove the power interface board:

1. Separate the upper assembly and the lower assembly.

2. Remove the main control board and the analog board.

3. Disconnect all the wires on the power interface board.

4. Remove the 2 screws securing the power interface board.

5. Remove the power interface board carefully.

To install a power interface board:

Install a power interface board in the reverse order. Ensure the wire is correctly connected.

- 57 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

9.3.9 Replacing the Exhaust Fan

Follow the steps below to remove the exhaust fan:

1. Separate the upper assembly and the lower assembly.

2. Disconnect the wire on the exhaust fan.

3. Remove the 2 screws securing the exhaust fan.

4. Remove the exhaust fan carefully.

To install an exhaust fan:

Install an exhaust fan in the reverse order. Ensure the wire is on the left side of the exhaust fan as
shown in the picture above.

- 58 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

9.3.10 Replacing the Speaker

Follow the steps below to remove the speaker:

1. Separate the upper assembly and the lower assembly.

2. Disconnect the wire connecting the speaker on the main control board.

3. Remove the 2 screws securing the speaker.

4. Remove the speaker carefully.

To install a speaker:

Install a speaker in the reverse order. Ensure that the wire is correctly connected to the main
control board.

9.3.11 Replacing the Bar Code Scanner

Follow the steps below to remove the bar code scanner:

1. Separate the upper assembly and the lower assembly.

2. Disconnect the wires connecting the bar code scanner on the main control board.

3. Remove the 2 screws securing the bar code scanner.

4. Remove the bar code scanner carefully.

To install a bar code scanner:

Install a bar code scanner in the reverse order. Ensure that the wire is correctly connected to the
main control board.

- 59 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

NOTE:

After installing a new bar code scanner, the bar code scanner should be set.

Follow the steps below to set a bar code scanner:

1. Press the power button on the left side of the analyzer to turn it on.

2. Press in the User Login screen.

3. Scan the following bar codes in sequence:

- 60 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

4. Enter the user name and password in the User Login screen.

5. Press Setup Diagnostics System ID. The bar code scanner will be turned on
automatically.

6. Scan the system ID bar code.

7. Check whether the last 12 digits of the scanned data are consistent with those of the system
ID. If they are consistent, the bar code scanner is successfully set.

9.3.12 Replacing the Spring Located in Test Cartridge Base

Operation Procedures Note

- 61 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the four screws as the right picture


shows.

Remove the two screws as the right picture


shows.

Disconnect the connection wire of vacuum


pump from the PCB board and remove the
vacuum pump from the slot as the right picture
shows.

- 62 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the test cartridge base from its slot as


the right picture shows.

Lift the spring support upward.

Remove the spring from its support.

Insert a new spring into the support and


reassemble as the reverse procedures.

9.3.13 Cleaning and Replacing Vacuum Pump

Operation Procedures

If there is any need to replace the vacuum pump because of some damages, please replace the
vacuum pump as the following procedures:

Disconnect the tube from the vacuum pump

- 63 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Separate the vacuum pump from its motor


Remove the piston head by turning it anti-
Replace a new piston head and assemble it into the
Insert
Insert the tube into the connector of the vacuum pump and seal the joint with AB adhesive.
Assemble the vacuum pump as reverse procedures.

9.3.14 Replacing FPC Cable

Change FPC Cable

Operation Procedures Note

- 64 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Disassemble the Remove the five screws as


cam module the right picture shows.

Remove the cam Remove the four screws as


cover the right picture shows.

Remove the cover module.

- 65 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the FPC Smear some alcohol with


cable from the cotton swab on the
heater hot-melt adhesive which is
used to secure the FPC
connector.

Remove the holt-melt


adhesive after several
seconds with hands or
some devices, and then
open the cover of the
connector as the right
picture shows.

Remove the FPC cable


from the connector as the
right picture shows.

- 66 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the motor Remove the two screws as


the right picture shows

Disconnect the cable as the


right picture shows

- 67 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the motor from


the cam as the right picture
shows

Remove the FPC Smear alcohol with the


cable from the cotton swab on the
Cam hot-melt adhesive located
which is used to secure the
FPC connector as the right
picture shows.

- 68 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the holt-melt


adhesive after several
seconds as the right picture
shows.

Open the cover of the


connector and disconnect
the FPC cable from the
connector.

Disassemble front As the right picture shows,


board assembly. turn the turntable
anticlockwise to lower the
valve rod.

- 69 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

As the right picture shows,


support up front board with
hands to remove the valve
rod from the turn table.

As the right picture shows,


remove the front board
from the cam assembly.

The front board is as the


right picture shows.

Disassemble the As the following picture shows, remove the two screws and separate the
front board cover from the front board.

- 70 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

As the following picture shows, remove the four screws and separate the
cover from the front board.

Disassemble FPC As the right picture shows,


cable from the pull out the FPC cable
front board. from the heater.

- 71 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

As the right picture shows,


open the black cover of the
FPC cover.

As the right picture shows,


pull out the FPC cable.

As the right pictures, pull


back the FPC cable.

9.3.15 Replacing the Cable of Temperature Detection Board

Operation Procedures NOTE

- 72 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Disassemble the cable of temperature detection


board.

Remove the screw as the right picture shows


and then remove the cover.

- 73 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Smear alcohol with the cotton swab on the


hot-melt adhesive and then remove the
adhesive after several seconds.

Open the cover of the FPC terminal and then


remove the FPC cable from the terminal.

- 74 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

Remove the cable of the temperature detection


board

9.3.16 Replacing and Cleaning Needle

Clean L-shape Needle

1. Check whether L-shape needle has been blocked.

1) Observe whether there are foreign objectives around the internal surface of the needle (in
many cases, the foreign objects will be accumulated in the point of the needle).

2) If the operator can obviously feel the obvious resistance when pulling or pushing the piston
of the syringe, the point of the needle has been blocked by the foreign objects.

2. If the needle has been blocked, clean the needle as the following procedures:

1) Slowly pull the piston of vacuum pump to the maximum range.

2) Insert the 1mL syringe into the L-shaped needle and then move out the 1mL syringe from the
needle.

3) Quickly push the piston of the vacuum pump to blow out the foreign objects from the
needle.

- 75 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

If the L-shaped needle has been blocked heavily, the needle can be changed as the following
procedures:

1. Disco
2.

3.
the joint to prevent air leakage.

Clean Cali

1.
needle in accordance with Step 5 and Step 6.
2. As Picture 12 shows, make sure the valve of the calibrant fluid pack is closed.
3. Use the syringe to aspirate the air. If the piston of the syringe can be moved smoothly, it
proves the distance between the needle and the air valve has not been blocked.
4. As Picture 13 shows, close air valve by pressing it with hands and then open the valve of
calibrant fluid pack.
5. Use the syringe to aspirate calibration solutions. If the piston of the syringe can be moved
smoothly and the calibration solutions can be aspirated smoothly, it proves that the distance
between the valve of the calibrant fluid pack and the needle has not been blocked.
6. As Picture 14 shows, insert 1mL or 2mL syringe into the needle. Quickly pull the piston
outward so that the foreign object can be aspirated out.
NOTE:
After the valve of the calibrant fluid pack has been opened, the calibration solution can only be
aspirated out, or it will be contaminated.

- 76 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

9.4 Disassembling the Upper Assembly

9.4.1 Replacing the Touch Screen Drive Board

Follow the steps below to remove the touch screen drive board:

1. Separate the upper assembly and the lower assembly.

2. Disconnect all the wires on the touch screen drive board.

3. Remove the 4 screws securing the touch screen drive board.

4. Remove the touch screen drive board carefully.

To install a touch screen drive board:

Install a touch screen drive board in the reverse order. Refer to the following illustration when
- 77 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

connecting wires, and ensure they are all correctly connected.

1 2 3

No. Connect to

1 Touch screen

2 Lower assembly

3 LCD screen

4 Printer

5 Power button

9.4.2 Replacing the LCD

Follow the steps below to remove the LCD:

1. Separate the upper assembly and the lower assembly.

2. Remove the touch screen drive board.

3. Remove the 7 screws securing the LCD stander.

- 78 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

LCD Stander

4. Remove the LCD stander, and place it on a flat surface.

5. Remove the LCD carefully.

To install a LCD:

Install a LCD in the reverse order. Ensure the wires are all correctly connected.

9.4.3 Replacing the Touch Screen

Follow the steps below to replace the touch screen:

- 79 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

1. Separate the upper assembly and the lower assembly.

2. Remove the LCD.

3. Remove the touch screen.

To install a touch screen:

Install a touch screen in the reverse order.

9.4.4 Replacing the Print Controller Board

Follow the steps below to remove the print controller board:

1. Separate the upper assembly and the lower assembly.

2. Disconnect all the wires on the print controller board.

3. Remove the screw securing the print controller board.

4. Remove the print controller board carefully.

To install a print controller board:

Install a print controller board in the reverse order. Refer to the following illustration when
connecting wires, and ensure they are all correctly connected.

- 80 -
i15 Blood Gas and Chemistry Analysis System Service Manual Disassembling the Analyzer

9.4.5 Replacing the Printer Head

Follow the steps below to remove the printer head:

1. Separate the upper assembly and the lower assembly.

2. Open the printer casing, and remove the 2 screws.

3. Pull the printer head carefully, and remove wires on the printer head.

To install a printer head:

Install a printer head in the reverse order. Refer to the following illustration when connecting
wires, and ensure they are all correctly connected.

- 81 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

10 Common Failures and Solutions


Error Prompts/Failure
Causes Solutions
Code Phenomenon
1) The test cartridge
Cartridge Expired, 1) Change a new test
expired.
2 Please Change cartridge.
2) The system date is
Cartridge. 2) Correct the system date.
wrong.

3 Cartridge Type Error. support this kind of test Update the analog software.
cartridge.
1) The barcode scanner
Maybe Cartridge 1) Clean the barcode
has been
Barcode Destroyed. scanner.
4 contaminated.
Scan Again or Try 2) Change a new test
2) The barcode has been
Another Cartridge. cartridge.
damaged.
The Software Version
does Not Match with The main control software Update the main control
5 Test Cartridge. Please and the analog software software and the analog
Update the Software have not been updated. software.
Immediately.
Calibrant Pack 1) The calibrant fluid
1) Change a new calibrant
Expired.Confirm Date pack expired.
21 fluid pack.
Setup and Calibrant 2) The system date is
2) Correct system date.
Pack's Expiration Date. wrong.

Calibrant Pack Type


22 support this kind of Update analog software.
Error, Please Replace it.
calibrant fluid pack.
1) The barcode scanner
Maybe Calibrant 3) Clean the barcode
has been
Barcode Destroyed, scanner.
23 contaminated.
Scan Again or Try 4) Change a new calibrant
2) The barcode has been
Another. fluid pack.
damaged.
Quality Control
Solution Expired;
41 The controls expired. Change the controls.
Please Change Quality
Control Solution.
The controls test mode

Quality Control Train the operator how to


42 selected kind of controls.
Solution Type Error. operate correctively.

recognize its barcode.

- 82 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

The controls test mode

Quality Control Train the operator how to


43 selected kind of controls.
Solution Level Error operate correctively.

recognize its barcode.

Check whether the system


61 Current Date Error. valid period is inconsistent date is right. If not, correct
with the system date. the date.
Heating Abnormally,
2005 Please Call for Heater failure Change a new cam.
Technical Assistance.
1)
temperature is not
within the required
working temperature.
2) The syst
temperature is within 1) Ensure the system
the required working works under the
Ambient Temperature
temperature while the required working
2008 Out of Range
temperature sensor temperature.
(10-31 ).
works improperly. 2) Re-connect the wire.
3) 3) Change a new wire.
temperature is within
the required working
temperature while the
wire connection is not
secure or correct.
1) The actual ambient 1) Ensure the system
pressure is not within works within the
the range. required pressure range.
Ambient pressure Out
2) The actual ambient 2) Check whether the
2009 of Range
pressure is within the pressure sensor has
(405-800mmHg).
range while the been blocked.
pressure sensor has 3) Change a new analog
some damages. board.
1) The connection wire is
Front Board does Not not secured between 1) Re-connect the wire.
Contact Well with the front-end board 2) Change the connection
2040 Analog Board. Please and the analog board. wire between the
Call for Technical 2) The connection wire is front-end board and the
Assistance. damaged between the analog board.
front-end board and

- 83 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

the analog board.

The test cartridge has not The operator needs to


Cartridge is Improperly been clicked into the perform the operation under
2050
Ejected During Test targeted place because of the requirements specified
improper operations. in the User Manual.
Change calibrant fluid pack
Calibrant Pack
under the requirements
2051 Improper Removal, Improper operation
specified in the User
Please Replace it.
Manual.
1) Re-conduct test with a
new test cartridge
following the
procedures described in
the User Manual.
2) Train the operator not
to contact the test
1) Improper operation
car
2) The contact is poor
contacts to avoid
between the
contamination.
POGO_PIN does Not POGOPIN and the
3) If the external simulator
2052 Contact Well, Please targeted part.
test succeeds, the
Try Another Cartridge. 3) The electrical contacts
analyzer works
of the test cartridge
normally. If fails, the
have been
internal simulator test
contaminated.
should be conducted. If
the internal simulator
test succeeds, check and
clean the POGOPIN. If
the internal simulator
test fails, change the
cam.
Chamber Door
2053 Improper Open. Please Improper operation Close the chamber door.
Close Chamber Door.
1) The test cartridge has 1) Conduct test with a new
been used. test cartridge.
Cartridge is Not
2) There is some water 2) Remove the water with
2054 Qualified, Please Try
on the surface of the clean cloth from the
Another Cartridge.
electrical contacts electrical contacts.
because the 3) Train the operator not

- 84 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

refrigerated test to directly contact the


cartridge has not been electrical contacts.
returned to room
temperature as the
proper procedures or

handle it properly.
3)
electrical contacts
have been
contaminated.
1) 1) Check whether the
properly because the structural parts have
poor connection of been damaged or the
Abnormal Cam wire. wire connection is not
Location, Please Call 2) Some structural parts secured.
2056
for Technical have been damaged. 2) Clean the sensor used to
Assistance. 3)
properly to detect the position.
corresponding 3) Reboot the analyzer for
position. testing.
1) If this failure occurs

probably caused by test


1) The test cartridge has cartridge. Change a new
some problems. test cartridge.
2) The calibrant fluid 2) If this failure occurs

washer is missing or probably caused by


the pack has been
Calibrant is Not blocked. operation. Remove the
Aspirated Properly; 3) calibrant fluid pack cap
2064
Please Try Another been blocked partially and check whether the
Cartridge. or completely or plastic washer is
damaged. missing. Aspirate
4) The syringe pump has calibration solutions
been damaged which with syringe to
will cause the failure determine whether the
of calibration calibrant fluid pack is
aspiration absolutely. blocking.
3) Check the needle which
is connected to the test
cartridge and located at

- 85 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

the cam assembly. If it


is contaminated, clean
it. If it is blocking or
damaged, please replace
it.
4) Change the syringe
pump.

1) The sample is not 1) The blood sample


connected into the test should be correctly
cartridge. connected into the test
2) Improper Operations cartridge.
2) The operations should
blood sample be correct.
timely after The bubbles
contained in the
discard the first blood sample
drop of blood; the should be removed
blood sample and the blood
contains bubble. sample after
The collection should be
anti-coagulant mixed timely.
heparin is not The best blood
added into the collection device is
Sample Error, Maybe
blood sample. the arterial blood
2068 Clots are Detected in
The blood sample collection needle. If
Sample.
contains the arterial blood
anticoagulant collection needle is
heparin but its unavailable, use
concentration is heparin as
not enough. coagulant. We
The blood recommend that two
vials of 12500U/mL
is not enough. The lithium heparin or
heparin sodium
volume collected should be added
by 1mL syringe is into physiological
less than 0.5mL. saline.
The blood The volume of the
collected blood
collected by 2mL sample should be
syringe is less enough.

- 86 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

than 0.8mL. The

volume collected
by 5mL syringe is
less than 1.5mL.
Under electronic simulator Re-conduct electronic
Insertion Error,
test mode, the test simulator test with a
2082 Cartridge is Inserted
simulator.
during Simulator Test.
into the analyzer.
1) Improper operation
Insertion Error, 1) Operate properly
2) The test cartridge is
2084 Simulator is Inserted 2) Change a new test
contaminated leading
during Sample Test. cartridge
to a poor connection
1) Poor wire connection
causes that the motor
fails to work normally. 1) Check whether the
2) structural part has been
slips off. damaged, whether the
3) There is a lot of dust wire connection is not
Pump Sensor
secured and whether the
Malfunctions, Please
2086 pump. screw slips off. If yes,
Call for Technical
4) The structural part is please fix it.
Assistance.
damaged. 2) Remove the dust within
5) The sensor has some
3) Clean the pump and
detect the check the sensor
corresponding
position.
Input Current
The power adaptor or the Change power adaptor or
7001 Abnormal
battery has some problems. battery.

The battery has some Change battery or ensure the


Battery Voltage
7002 problems and the voltage analyzer is powered by
Abnormal
is abnormal. power adaptor.
If power voltage is
abnormal, the power Change the power adaptor
7003 5V Voltage Abnormal
adaptor or the battery has or the battery.
some problems.
The printer has some
Printer Voltage
7004 problems and the voltage Change or fix printer.
Abnormal
is abnormal.
The power voltage is Change the power adaptor
7005 12V Voltage Abnormal
abnormal, the power or the battery.
- 87 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

adaptor or the battery has


some problems.
Battery Charge The battery has some Change the battery or reboot
7006
Abnormal problems. the analyzer.

USB Disk Cannot be recognized. The system Use the correct type of the
Identified USB Disk.
USB Disk.
Format USB disk or change
U disk Error
recognize the USB disk. a new USB disk.

The USB port has some Change the USB port or


Can Not Get USB Disk
problems. read date via network port.
Information
The function of USB disk
Export Failure Disable the function of USB
write protection has been
disk write protection.
abled.
The internet cable is not
Connect the network cable
connected well; or the
well and check the network
network disconnects; or
Transmitting Failure setup. Select the correct
the selected protocol is not
protocol and set the correct
correct; or the server IP
server IP address.
address is not correct.
The targeted database is Disable the function of USB
empty or the USB disk is disk write protection and
Backup Error
turned off so that it can check that the database is
only be read. empty.
The model of the Only the printer with
connected printer is not designated model can be
Printer Connection
supported by the analyzer connected to the analyzer.
Failure
or the USB port works Change the USB port or
improperly. reconnect USB cable.
Enter the correct IP. The
IP Format Error The enter IP is 192.168.1.*
entered IP is 192.168.1.*
and * is not within 0-255.
and * is within 0-255.
Re-enter a new IP. The
re-entered IP should not be
The entered IP has been
IP Conflict used or the IP can be
used.
automatically obtained by
the analyzer.

- 88 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

1) The analyzer might


have some problems.
Please do external
simulator test.
2) The calibrant fluid 1) Repair cam assembly if
pack might have some the external simulator
problems, such as test fails.
being stored not under 2) Change calibrant fluid
the required pack.
All Parameters Fail
conditions, being 3) Change test cartridge.
Calibration; Please Try
contaminated or 4) Store and use the test
Another Cartridge
introduced into blood. cartridge and calibrant
3) fluid pack in
transport conditions accordance with the
requirements listed in
requirements. the user manual.
4) The failure might be
caused by the
analyzer. The external
simulator test fails.
1) Reboot i15 analyzer or
enter into Diagnostic
Screen to press Valve,
The barcode of test
the cartridge will be
Please Reboot the cartridge has not been
ejected out.
System, and Remove scanned before the
2) Require operator to
Cartridge. cartridge has been inserted
operate in accordance
into i15 analyzer.
with procedures
described in user
manual.
Check whether the network
setup is correct. Make sure
The server IP is not whether the server IP is
192.168.1.93, or the tftpd 192.168.1.93. Make sure the
Fail to Get Update File,
program has not been local IP is 192.168.1.* (*
Please Check the
opened, or the software can be any figure within
Network Setup
upgrading package is not 0-255). Open the tftpd
contained in the target file. program. Place the software
upgrading package in the
target file.
The network disconnects Make sure the connected
Update Failure when upgrading the analog network is stable when the
software. software is updated, or

- 89 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

update the analog software


via USB disk.

The barcode of calibrant 1) Reboot the analyzer or


scan the recognizable
successfully. The barcode of the calibrant
Replacement Failure.
installation of calibrant fluid pack
fluid pack fails more than 2) Change the calibrant
two times. fluid pack.
1) Conduct external
simulator test.
2) If the external simulator
test passes, disassemble
the cam assembly and
then check whether the
1)
zero-calibration board
loose.
Inter-Simulator Fails, is well connected, and
2) The POGOPIN of cam
Perform External finally re-assemble the
gets dampened.
Simulator Test cam.
3) The failure of front
3) If the external simulator
board
test fails, check whether
there are foreign objects
around the POGOPIN,
and dry up it and
re-conduct test.
4) Change Cam.
1) There are a lot of dust
on the power board,
1) Clean the analyzer and
which result in the
Power State is reboot it.
failure of the power
Abnormal, Please Call 2) Re-install lithium
board.
for Technical battery.
2)
Assistance 3) Change lithium battery.
properly.
4) Change power board.
3) The po
detect the voltage.
1) The barcode is
1) Train the operator how
contaminated or
to operate properly and
damaged.
Barcode Check Error change the barcode.
2) The window of the
2) Clean the window of
scanner is
the scanner.
contaminated.
The indicator on the 1) The connection of 1) Re-connect the power
power cord is poor. cord.

- 90 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

indicate the battery has 2) Failure of battery 2) Change the battery


been charged,

powered by the battery


while can be powered Failure of battery Change the battery
by AC power supply
normally.
The power adaptor
indicator is not on. The
Failure of power adaptor Change the power adaptor
powered and the battery
ged.
The touch screen
1) Failure of touch screen 1) change touch screen
2) The screen board has 2) Change the screen
occasionally or
some problems board
consistently.
1) The installation of
1) Correctly install printer
printer paper is wrong.
paper
Fail to print 2) The printer board
2) Re-connect the printer
connection wire is
board connection wire
loose.
The DMS software
Reinstall DMS software
lost.
i15 analyzer.
1) There are a lot of
1) Delete the unchecked
unchecked sample
sample information.
T information.
2) Ensure the analyzer is
transmitted to the 2) The analyzer is
located in a place with
located in a place with
strong and stable WiFi
unstable and weak
signal.
WiFi signal.
Re-
opened. system has some problems. operating system.
1) The database service
The system hints the disconnects.
1) Re-connect the database
data has been 2) Other opened software
service and re-transmit
transmitted successfully occupies 8000 port
the database.
while the DMS such as other
2) Change the connection
t
port.
receive the data. station. Hospital
LIS/HIS, etc.
i1
IP setting error Reset the IP address
connected to the
- 91 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

computer.

transmitted from i15


IP setting error Reset the IP address
analyzer to DMS
software.

The DMS software needs Update the DMS software


transmitted to DMS
to be updated. into the latest version
software.
DMS software hints the
The test application time
test application time is
should not be earlier than Reset the correct time
earlier than the test
the test inspection time
inspection time.

Reassemble the spring.


is not in normal condition.
normally.
1) Check whether the

needle has blood, some


contamination or
1) The test cartridge has blockade. If yes, change
not been removed the calibrant fluid pack.
The test cartridge has
immediately after 2) Check whether the
solution leakage.
completion of test. POGOPIN has blood,
2) Improper operation some contamination or
blockade. If yes, clean
it.
3) Change a new test
cartridge

The door of calibrant fluid


The calibrant fluid Change the door of calibrant

problems.
be opened.
1) Remove the test
cartridge firstly and
1) The key is not then turn off the
removed from the analyzer and finally
The calibrant fluid pack
calibrant fluid pack remove the calibrant
when installing the fluid pack from the
the analyzer.
pack. analyzer.
2) Improper operation 2) Instruct the operator to
correctly install the
calibrant fluid pack.

- 92 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

The spring which is


located in the cam and
be removed from the used to pull out the test Change a new spring.
analyzer occasionally. cartridge after completion
of test has some problems.
The test cartrid
be inserted into the
Improper operation and the Normalize the operation and
targeted position from
ensure the test cartridge
into place. clicks into place.
it will be ejected out
automatically.
The system halts when
Upgrade main control
its data exported to the
failure software
USB disk.
Upgrade main control
switched off. failure software
1) Conduct internal
1) The POGO PIN has simulator test
some foreign matter 2) If the internal simulator
which causes the test succeeds, check
failure of external whether the POGO PIN
The external simulator
simulator test has some foreign matter
test fails.
2) The channel and clean the
inspection results POGOPIN
3) If the internal simulator
requirements. test fails, change the
cam.
1) The scanned barcode 1) Scan the barcode of
controls used for the
barcode used for the test.
test. 2) Conduct tests with the
2) proper controls
QC tests with controls
with the proper 3) Re-conduct test with a
fail.
controls. new calibrant fluid
3) The calibrant solutions pack.
are contaminated. 4) Conduct simulator test.
4) The simulator test may If the test fails, maintain
be out of control. or replace the cam.

- 93 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

1) The operator needs to


operate in accordance
with the user manual.
2) Check whether the
detection slot of the test
1) The test cartridge cartridge has some
foreign objects. If yes,
place because of the remove the foreign
be recognized after its
objects.
barcode has been
operation. 3) Check whether wires
scanned.
2) The test cartridge has connected to the
some problems. detection board of the
test cartridge become
loose. Reconnect the
FPC board which
detects pump.

1) Train the operator how


1)
to operate correctly.
do the operation in
2) The distance between
accordance with
the barcode and scanner
standard procedures.
should be within
2) The barcode is not
10-15cm.
clean enough.
3) Contact manufacture
1) Check whether the
detection slot of the test
cartridge has some
foreign objects. If yes,
1) The detection slot of
remove the foreign
the test cartridge has
The system hints to objects.
some foreign objects.
remove the test 2) Check whether wires
Therefore the test
cartridge after being connected to the
logged while there is no detection board of the
work normally.
test cartridge actually. test cartridge become
2) The wiring connection
loose. Reconnect the
is poor.
FPC board which
detects pump.
3) Replace FPC board
which detects pump.

- 94 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

1) The electrical contacts


of test cartridge are
contaminated.
2) The electronic
simulator test may be
out of control and the 1) Train the customers not
POGOPIN may be to touch the electrical
contaminated. contacts of test
3) There are bubbles cartridge to avoid
when calibrant contamination.
solutions are aspirated 2) Conduct electronic
into the test cartridge. simulator test. If the test
This case may be fails, firstly clean the
caused by the POGOPIN, then
following reasons: re-conduct the
Occasional leaks electronic simulator
HCT AC impedance is
or blocking of test test. If the test fails
too high.
cartridge again, the cam needs to
The calibrant fluid be replaced.
pack is blocked 3) Re-conduct tests with a
partially. new test cartridge;
The valve Re-conduct tests with a
connecting new calibrant fluid
analyzer and pack; Re-conduct tests
calibrant fluid after the valve has been
pack or test positioned correctly.
cartridge is not 4) Re-conduct test with a
positioned in a new test cartridge
correction
position.
4) The occasional failure

sensor.
Train the personnel who is
The calibration The volume of blood
responsible for sample
succeeds while one or sample is not enough or
collection to normalize the
several items have no there are bubbles in the
procedures of sample
test results. blood sample.
collection and test.
1) Re-conduct test with a
pH has no test result 1) The occasional failure
new test cartridge.
occasionally or other of test cartridge.
2) Train the personnel who
parameters have no test 2) The sample has some
is responsible for
results occasionally. problems.
sample collection to

- 95 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

normalize the
procedures of sample
collection and test.
1) Re-conduct QC test.
The test results are 2) Train the personnel who
Controls test fails.
displayed with *** is responsible for
sample collection.
The barcode scanner Re-
connection is loose. wire,

- 96 -
i15 Blood Gas and Chemistry Analysis System Service Manual Replaceable Parts

11 Replaceable Parts
WARNING

Only the replaceable parts supplied by EDAN or its authorized distributors should be
connected to the analyzer.

Part Part Number

Main Control Board PCBA 02.03.451152-11

Power Board PCBA 02.02.451154-11

Analog Board PCBA 02.03.451128-11

Touch Screen Drive Board PCBA 02.02.451532-11

Wi-Fi Board PCBA 12.02.451363

Chamber Door Detection Board PCBA 02.02.451663-11

Power Interface Board PCBA 02.02.451357-11

Ambient Temperature Detection Board 02.02.451751-11

Cam Machine 02.01.210726

Speaker 01.14.038010

Exhaust Fan 21.58.472029

LCD 01.16.045068

Touch Screen 21.16.045069

Bar Code Scanner 21.17.047380

Print Controller Board 02.02.220047-11

Printer Head 01.17.078173

- 97 -
- 98 -

You might also like