Professional Documents
Culture Documents
Service Manual
Intellectual Property Statement
All information contained in this manual is believed to be correct. SK Medical shall not be
liable for errors contained herein or for incidental or consequential damages in connection
with the furnishing, performance, or use of this manual.
SK Medical is responsible for the effects on safety, reliability and performance of this
product, only if:
all installation operations, expansions, changes, modifications and repairs of this
product are conducted by SK Medical authorized personnel;
the electrical installation of the relevant room complies with the applicable
national and local requirements
2
Foreword Introduction
This Service Manual provides detailed introduction of the product and its components, as
well as the procedures for installation, removal, testing and troubleshooting, intended to
help the service technicians solve the common problems effectively. This Service Manual
does not provide depth introduction on the product structure and design principle. In case
of any problems you cannot solve, please contact our After-sale Service Division.
This Service Manual describes the product based on most complete configurations, so
that some parts might be unsuitable to your product. If you have any questions, please
contact our After-sale Service Division.
Before repairing the product, please read this Service Manual thoroughly until full
understanding of the contents herein, so as to ensure correct repair and avoid causing
damage to the product or human body.
Intended Audience
This Service Manual is intended for the biomedical engineers, authorized repairmen or
service representatives responsible for maintenance of this product.
Version Information
With the change of software or technical specifications, the version number of this Service
Manual is subject to updating without prior notice. Version information of this Service
Manual is as follows:
Version: 1.0
All rights reserved. Shenzhen Shenke Medical Instrument Technical Development Co., Ltd.
3
Contents
Intellectual Property Statement ......................................................................................... 2
Responsibility on the Manufacturer Party .......................................................................... 2
Contents of this manual are subject to changes without prior notice................................. 2
Foreword Introduction ....................................................................................................... 3
Intended Audience .......................................................................................................... 3
Version Information ........................................................................................................ 3
Contents ............................................................................................................................ 4
Chapter 1 Safety ............................................................................................................... 1
1.1 Safety information ........................................................................................... 1
1.2 Danger ............................................................................................................ 1
1.3 Warning ........................................................................................................... 2
1.4 Caution ............................................................................................................ 3
1.5 Note................................................................................................................. 4
1.6 Equipment symbols ......................................................................................... 4
Chapter 2 Principle of design ............................................................................................ 6
2.1 Overview ......................................................................................................... 6
2.2 Appearance ..................................................................................................... 8
2.2.1 Front view ............................................................................................. 8
2.3 Rear view ...................................................................................................... 10
2.4 Hardware principle ........................................................................................ 11
Chapter 3 Test and Maintenance .................................................................................... 16
3.1 Introduction ................................................................................................... 16
3.1.1 Test Report ......................................................................................... 17
3.1.2 Recommended Interval ...................................................................... 17
3.2 Appearance Inspection.................................................................................. 18
3.3 Power-On Self-Test (POST) .......................................................................... 18
3.4 Calibration ..................................................................................................... 19
3.4.1 Syringe calibration .............................................................................. 19
3.4.2 Pressure calibration ............................................................................ 22
3.4.3 Displacement sensor calibration ......................................................... 24
3.5 Operation test ................................................................................................ 28
3.6 Battery test .................................................................................................... 28
4
3.7 Electrical Safety Test ..................................................................................... 29
3.7.1 Electric Leakage Test of Enclosure ..................................................... 30
3.7.2 Electric leakage test of earth .............................................................. 30
3.7.3 Electric Leakage Test of Patient ......................................................... 31
3.8 Cleaning and Sterilizing................................................................................. 31
3.8.1 Cleaning ............................................................................................. 31
3.8.2 Sterilizing ............................................................................................ 32
Chapter 4 Fault Elimination ............................................................................................. 33
4.1 Overview ....................................................................................................... 33
4.2 Replacing Components ................................................................................. 33
4.3 Check syringe pump status ........................................................................... 34
4.4 Fault list ......................................................................................................... 36
4.4.1 Alarm information and solution ................................................................. 36
Please refer to “BeneFusion SP5 Operator’s Manual” ...................................... 36
4.4.2 Screen malfunction .................................................................................. 37
4.4.3 Operation, alarm fault .............................................................................. 37
4.4.4 Monitor fault ............................................................................................. 38
Chapter 5 Repair and Disassembly................................................................................. 39
5.1 Tools .............................................................................................................. 39
5.2 Preparations for Disassembly ....................................................................... 39
5.3 Disassembly Steps ........................................................................................ 40
5.3.1 Disassembly of UP cover assembly ................................................... 40
5.3.2 Door assembly.................................................................................... 41
5.3.3 Disassembly of pump body assembly ................................................ 42
5.3.4 Disassembly of rear cover assembly .................................................. 44
5.3.5 Removal of push box assembly, convex detection assembly, pump body
transmission assembly...................................................................................... 44
5.3.6 Disassembly of pressure sensor assembly ........................................ 45
5.3.7 Disassembly of screen, keypad board PCBA ..................................... 48
Notes: Disassembly of screen, keypad board PCBA, make it directly in whole
machine, don’t need to take out of door assembly. ........................................... 48
5.3.8 Disassembly of main board PCBA...................................................... 50
Chapter 6 Components ................................................................................................... 52
6.1 Overview ....................................................................................................... 52
6.2 Host ............................................................................................................... 52
6.2.1 Explosive view .................................................................................... 52
6.2.2 Parts list .............................................................................................. 53
5
6.3 SP5 UP cover assembly ................................................................................ 54
6.3.1 Explosive view .................................................................................... 54
6.3.2 Parts list .............................................................................................. 54
Chapter 7 Upgrade.......................................................................................................... 55
7.1 Tools .............................................................................................................. 55
7.2 Software upgrade .......................................................................................... 55
7.3 Software burning method .............................................................................. 55
6
Chapter 1 Safety
DANGER
● To indicate immediate danger. If not avoided, it might cause death, severe human
injury or property loss.
WARNING
● To indicate the potential hazard or unsafe operation. If not avoided, it might cause
death, severe human injury or property loss.
CAUTION
● To indicate the potential hazard or unsafe operation. If not avoided, it might cause
slight human injury, product problem, damage or property loss.
NOTES
● To emphasize the important considerations, provide description or explanation
for better use of this product.
1.2 Danger
In this Service Manual there is no safety information on danger level.
1
1.3 Warning
WARNING
2
1.4 Caution
CAUTION
3
1.5 Note
NOTE
NOTE!
Refer to the accompanying ON/OFF
document (This Manual)
Alternating current power
Battery
supply (AC)
Alarms Silence
Clear/Back Start
Bolus Confirm
Stop Menu
4
Move left Move right
Environmentally-friendly use
periods of electronic products
(20 years)
5
Chapter 2 Principle of design
2.1 Overview
The Syringe Pump is used in connecting with the syringe to control the dose of liquid
infused into the patient’s body.
The Syringe Pump is suitable for adults, children and newborns in clinical departments.
The Syringe Pump primarily consists of a housing, a motor drive system, an input system,
a memory system, a control system, a display system, a sensor-based monitor system
and an alarm system. Since some parts are optional parts, the Syringe Pump you
purchased may not contain these additional parts and their relevant functions.
6
Memory function, can record the last infusion parameter;
Manual bolus and auto bolus function;
Standby function, keep the parameter set of previous standby;
Purge function, air bubbles should be eliminated before the infusion;
Dynamic pressure scanning function, monitor real-time pressure changes, as per
pressure icon, to display the normal pressure status, near occlusion pressure
status and occlusion alarm pressure status.
Automatic locking and manual locking function, avoid the accident operation and
misuse.
Anti-bolus, the cannula pressure is released after occlusion;
Power-on self-test function;
Protected against defibrillation function;
KVO function, to keep vein open;
Fixing clamp can be installed horizontally and vertically, support at least 4 pumps
overlapped;
High-flux LED display;
Alarm: visible alarm, audible alarm and alarm information, alarm volume with
10 levels selectable.
Network communication: be connected with the central transfusion monitoring
management system, infusion supervision system, and BB machine, through the
approach of wireless networking. Data exported or upgraded by connecting to PC.
Optional functions of the software comprise Rate Mode, Time Mode, Bodyweight
Mode, TIVA Mode, Ramp Up/Down Mode, Sequential Mode, Loading Dose
Mode, Micro-infusion Mode.
7
2.2 Appearance
8
7. BOLUS
During infusion, press this key to enter the Bolus Settings screen.
When the pump is stopped, press this key to enter the purge settings screen.
Power on, if installing the syringes, long press or short press this key to enter
standby mode; if without syringe installed, short press this key to enter standby
mode, long press to power off.
8. POWER
Under power off status, used for turning power on;
Under power on status, if install the syringes, long press or short press this key to
enter in standby state; if no install the syringe, short press this key to enter in
standby state, long pressing to power off.
9. AC power source indicator lights
Once the pump is connected to an AC power supply, this light will be green,
otherwise, this light will be off.
10. Clip Clamps the syringe plunger
11. Handle
Controls clip opening/closing and slider action.
12. OK
Used for confirming input operation.
13. Stop
During infusion, press this key to stop infusion. Infusion stops caused by alarms, such as
occlusion and so on, press this key to cancel the alarm.
14. Silence
For any alarm, press this key to silence; without new alarm, press this key again to cancel
alarm.
15. Menu
Under non-operation status, used for switching [Main Menu] interface and other
interfaces.
Under operation status, press and hold this key to lock; in locked state, press
and hold to unlock.
16. Battery indicator
Steady green indicates that the battery is charging.
Flashing indicates that the battery is providing power.
Light off indicates that there is no battery or battery damaged.
17. Spindle clamp
Used along with flange plate to secure syringe flange and clamps the syringe.
18. Slider
Secures syringe and drives plunger assembly.
9
2.3 Rear view
1. Handle
2. Slider
3. Speaker hole
4. Alternating current
power supply
(AC) port
5. Battery box
6. Multifunction
interface
7. Product label
8. Multi-channel
pumps connect rail
10
2.4 Hardware principle
SP5 hardware schematic diagram as shown below: including main board, keypad board,
pressure board, etc.
1. Main board
SP5 main board includes 3 parts: power, master control and monitor.
Power part includes ACDC conversion module, switch startup & shutdown control circuit,
DC power conversion circuit, power management CPU, standby sound-light alarm circuit,
charge and discharge control circuit, etc.
Switch control circuit can combine three-way merge power input into VPC, including the
ACDC input, external DC input, internal battery; the ACDC input, external DC input priority
is higher than the internal battery; VPC can be converted into VPP output by switch signal.
VPC is not controlled by switch signal, VPP is controlled by switch signal. VPC is reduced
to VCCB (5V) by LDO circuit(U7 ZSR500GTA). VCCB is reduced to VBB(3.3V) by LDO
circuit (U3 SPX5205); VPP is reduced to VCC(5V) by Buck circuit (U12 TPS5410D); VCC
is reduced to VDD(3.3V) by LDO circuit (U13 SPX1117M3); VPP is increased to
VMM(+24V) by Boost circuit (U26 LT3757EMSE).
11
The power management CPU is responsible for completing the switch control,
under a single fault condition motor power-down protection, battery charging status
monitoring, and the brightest state detection module power, the battery light control,
and with the main control module CPU via the serial port. CPU Power Management
selection Cortex M0 core chip LPC1112 (U10), powered by VBB supply.
12
The main section is responsible for LCD display driver, the speaker alarm circuit,
nurse call, Dock reign detection and power CPU, the key CPU, vice CPU UART
communication, etc. The host CPU Cortex M3 core chip STM32F103VE (U14),
using Intel 8080 system bus interface as the display driver interface
Monitoring module is consist of vice CPU, stepper motor drive circuit, encoder
detection circuit, the pressure detection circuit, the detection circuit reign syringe
flange detection circuit. Vice CPU using STM32F100C8 (U1), stepper motor driver
circuit using motor control chip TMC260-PA (U2). Dual encoder detection circuit using
optocoupler, through the comparison of two sequences of pulses determines the
motion direction of the motor by measuring the frequency of a pulse sequence
determines the motor speed.
13
2. Pressure board
Pressure monitoring board includes a pressure sensing circuit and syringes reign
detection circuit, the pressure detection circuit composed by the ADS1100 and
SPX5205, SPX5205 will convert to +3.3 V +5 V output, +3.3 V power supply to the
ADS1100 and pressure sensor; ADS1100 analog pressure sensor will input signal is
converted into a digital signal, transmitted to the sub CPU through the I2C interface.
3. Keypad board
Key management using Cortex M0 core CPU chip LPC1112 (U4) is responsible for
the keys (of 12 keys, including the power switch key directly lead to the power supply
module CPU; other keys by the key management of the CPU processing) scan, alarm
lamp drivers, LCD backlight control (8 levels adjustable) and other functions.
14
Key signal testing points:
TP10 VDD 3.3V
TP11 VCC 5.0V
TP12 VPP
TP2 LCD backlight power supply
TP20 battery indicator light signal
TP21 AC indicator light signal
15
Chapter 3 Test and Maintenance
3.1 Introduction
To ensure long reliable run of the infusion pump, the service people shall carry out
periodical inspection, maintenance and testing. This section provides the basic
testing methods for infusion pump. In this section, we have also recommended the
suitable testing intervals and tools. The service people are required to repair and
test the infusion pump as needed, and select the suitable testing tools.
The testing and inspection methods provided in this chapter are mainly used for
verifying whether the infusion pump reaches the performance requirements. If the
testing result does not meet the requirements, it indicates that some function
module of the infusion pump is failed. In this case, immediate repair or replacement
is required. If you have any other questions, please contact our After-sale Service
Division.
CAUTION
16
3.1.1 Test Report
After finishing the test, the service people approved by SK Medical are required to
complete the following test report and send it back to the After-sale Service
Division.
Test Equipment
Name Model / Number Effective Date
Test Records
S/N Test Item Test Point Test Result
1
2
3
4
5
Judgment
17
Check / Maintenance Item Interval
Electric Leakage
Test of Patient
Cleaning and Sterilizing 1. Every month recommended.
2. Thorough cleaning of the pump surface is
required before or after a long storage.
2) The system will give out a sound “di” and buzzer “di di”.
3) During power on, the alarm light will changed from red to yellow orderly.
18
NOTE
3.4 Calibration
In order to improve the injection accuracy, make the calibration for the non-defined
and empty position, the data of brand definition is empty in the factory; if the defined
syringe brand is not calibrated, during using, the syringe brand list won’t show this
defined brand.
1. No syringe installed:
19
b) Press “C” key to skip the step, enter the mode operation interface
c) Press “menu” key to enter the main menu, choose “system maintenance”.
d) Input factory password “6686” to enter system maintenance menu
e) Choose syringe calibration to enter
2. With a syringe installed:
a) After starting, enter brand select interface
20
c) Pull the syringe to the full dimensions and install them on the syringe pump,
take 50ml syringe as an example, below refers to the full dimension of the
50 ml syringe:
d) Press start key and the syringe pump starts calibration as shown below:
21
e) The system will fill in the calibration data after the calibration is successful,
22
Calibration method:
shown below:
b) The syringe filled with fluid, connect one side of syringe to fluke, then install it to
c) When fluke pressure value reach 150mmHg, press stop key, the first calibration
completed;
d) Continue to press start key to start the second calibration, when fluke pressure
value reach 950mmHg, press stop key, the second calibration completed;
e) Calibration completed, the system will fill in calibration data P value automatically,
23
3.4.3 Displacement sensor calibration
The purpose of calibration:
Because there is mechanical error between short potentiometer (user Identity syringe
size) and emptying position in each machine, in order to identify the outer diameter of
the syringe and syringe emptying position more precise, each machine will calibrate
the displacement sensor at the factory.
Calibration method:
a) Under the system maintenance menu, enter the displacement sensor calibration,
b) Install standard component I correctly, press bolus key to calibrate the standard
component 1
24
c) After standard component 1 calibrated, the screen will prompt “please install
d) Install standard component 2 correctly, press bolus key to calibrate the standard
component 2
successful;
f) Calibration succeed, the current outer diameter will show that it’s closed to the
25
g) Push the handle of slider, pull out the slider, and put the slider card into the jaws
h) Move the cursor to the standard emptying position, install the standard 1 or
standard 2, then pull out the slider, so that the top surface of the fixture and clamp
26
i) Press “Start” key to start the calibration, the words “being calibrated, press the
stop key to cancel the calibration” appears on the screen, after calibration
calibration fixture 13.5mm just withstood the knob of slider, as shown below;
27
3.5 Operation test
General features:
1) For each key operation, MR pump should be able to press the “User's Guide”
normal action or reaction, abnormal operation should not cause crashes.
2) Press the “User Manual” operation of the pump, check the sound and light alarm
levels are normal, whether silence alarm is workable or not.
3) Press the “User Manual” operation of the pump and trigger the alarm, check
whether the normal recording operation history and alarm;
4) After the injection parameters set correctly, press the start button to enter
operation interface, run Fig standard display correctly.;
5) Injection process runs Fig. Standard, the speed, the cumulative amount of
pressure values displayed correctly
3) The central station receive the pump status, infusion parameters, alarm
information correctly
4) If the connecting link is disconnected, the central station can indicate
correctly, when restore the connection, the network will work properly.
2) When the battery capacity exhausts, the pump won’t start or stop operation
28
3.7 Electrical Safety Test
WARNING
The electrical safety testing is intended for finding out the abnormal
electrical hazards by using the accredited tester. Failure to discover these
hazards might cause injury to the patients or operators.
The electrical safety testing may be done by using the commercially
provided safety analyzer and other testing devices. The service people
must guarantee the suitability, functional integrity and safety of these
testing devices, while having good skills on their use.
The electrical safety testing shall follow the standards of the latest version,
that is, EN 60601-1.
Please also observe the local laws on electrical safety testing, if any.
In the patient ward, all the equipments that are powered by city grid and
connected to the medical instrument must comply with IEC 60601-1
standard. Meanwhile, electrical safety testing must be done to these
equipments according to the testing interval of the tester.
The electrical safety testing is intended for finding the potential hazards that might
cause electrical safety risk to the patients, operators and repairmen. Please carry
out the electrical safety testing under normal environment (including: temperature,
humidity and atmospheric pressure).
The electrical safety testing described in this chapter is based on 601 safety
analyzer. Different safety analyzers might be used in different regions. Please
ensure the suitability of your electrical safety testing program.
T
e
29
Testing tools:
Safety analyzer
Isolation transformer
2. Connect the applied part of the tested equipment by using the applied part
connection tooling. The SUM end of the connection tooling for applied part is
connected to the RA end of the safety analyzer.
3. Via power cord, connect the tested equipment to the auxiliary power output
port on 601 safety analyzer.
4. Connect one end of the red testing wire to the “Red input terminal” on the
safety analyzer, and clip another end tightly onto the metal foil on the surface
of tested equipment enclosure.
5. Turn on the power source of 601 safety analyzer and press “5-Enclosureleakage”
on the panel of 601 safety analyzer to enter the enclosure leakage current testing
interface.
6. The enclosure leakage current shall not exceed 100μA under normal
conditions and shall not exceed 300μA in event of a single fault.
4. Turn on the power source of 601 safety analyzer and press “4-Earth leakage”
on the panel of 601 safety analyzer to enter the earth leakage current testing
interface.
5. The earth leakage current shall not exceed 500μA under normal conditions
and shall not exceed 1000μA in event of a single fault.
30
3.7.3 Electric Leakage Test of Patient
1. Connect the 601 safety analyzer to 264 VAC/60 Hz power source.
2. Connect the applied part of the tested equipment by using the applied part
connection tooling. The SUM end of the connection tooling for the applied part
is connected to the RA end of the safety analyzer.
3. Via power cord, connect the tested equipment to the auxiliary power output
port on 601 safety analyzer.
4. Turn on the power source of 601 safety analyzer and press “6-Patient leakage”
on the panel of 601 safety analyzer.
6. The patient leakage current shall not exceed 10μA under normal conditions
and shall not exceed 50μA in case of a single fault.
If the electrical safety testing fails, please contact our technical support people.
3.8.1 Cleaning
The syringe pump shall be cleaned periodically. The cleaning interval shall be
shortened in the areas with severe pollution or great sandy wind. Before cleaning,
please consult or check the regulations for cleaning of medical equipment.
Warm water
Dilute soapy water
Dilute aqua ammonia
Sodium hypochlorite (bleaching powder for washing)
Hydrogen peroxide (3%)
Ethanol (70%)
Isopropanol (70%)
31
When cleaning the device:
1. Turn off the pump and disconnect the AC power source line.
2. Wipe the display screen after soft cotton balls absorb an appropriate amount of
detergent.
3. Use a piece of soft cloth which absorbs a modest amount of cleaning agent to
wipe the surface of the device.
4. When necessary, use a piece of cloth to wipe off any excess cleaning agents.
5. Place the equipment in a cool and ventilated environment to dry.
3.8.2 Sterilizing
The operation of disinfection may cause certain damage to the infusion pump. You are
recommended to disinfect only when it is necessary in your desired maintenance plan.
Clean the equipment before disinfection.
CAUTION
32
Chapter 4 Fault Elimination
4.1 Overview
This chapter provides categorized description of the infusion pump troubles by fault
position and symptom. When eliminating the trouble, please refer to the
corresponding fault sheet and check, locate and eliminate the fault in order.
The recommended methods in this chapter can help you solve most of the
equipment troubles, but exclusive of any problem that might occur. In case of any
fault not described in this chapter, please contact our After-sale Service Division.
To know and obtain the components you need to replace, please refer to Chapter 6
“Components”.
33
4.3 Check syringe pump status
The instruction of AD value access and checkout content
1. The instruction of AD value screen: System Maintenance- password—sensor AD
value 2AD value screen as shown below:
34
Inc: Syringe installation, wing handle, nut
Dstep: Lose count, Encoder testing and the actual value difference
ODia: Short potentiometer AD value
4. ODia means AD value of short potentiometer, pull the short potentiometer handle
smoothly, the ODia value will be larger, the scope should be between 0-4095;
6. PreTable AD showing pressure sensor output value, when the pressure to the
pressure sensor, this value should be increased;
7. CodeTable code shown optocoupler output disc detection AD value, under normal
circumstances, the motor rotates, this value should be between 0,1,2,3 cycle;
35
4.4 Fault list
SP5 in the transportation, storage, use, due to various factors (such as the grid
voltage instability, changes in ambient temperature, drop and shock, component
aging, etc.) influence may malfunction and affect the normal use. At this point there
should be some component level repair of electronic medical equipment maintenance
experienced professionals refer to the following fault classification. The so-called after
the component level repair refers to the parts of the analysis, substitution,
commissioning, fault locate the parts of the instrument, such as the power supply
section, Master control part, monitoring part, LCD and so on. Component repair itself
that component level repair, must have extensive experience in the maintenance by
the manufacturer's engineers use specialized equipment, carried out under specific
circumstances and conditions.
AC power supply is 1. Bad connect with the wire of 1. Repair or replace the FPC wire;
connected, the AC keypad board FPC; 2. Repair or replace the wire of AC
indicator light isn’t on 2. Bad connect with the wire of AC power supply;
power supply; 3. Repair or replace the ACDC
3. Malfunction of ACDC module. module
Didn’t pass the 1. Bad connect with keypad board 1. Repair or replace the FPC wire;
Self-inspection 3, alarm FPC wire; 2. Replace main board;
2. Malfunction of main board; 3. Replace keypad board
3. Malfunction of keypad board
36
Symptom Possible Cause Fault Elimination
Didn’t pass the 1. Bad connect with FPC wire of 1. Repair or replace the FPC wire
Self-inspection 5, alarm lead screw clutch board; of lead screw clutch board
2. Bad connect with 6pin wire 2. Repair or replace the 6pin wire
Cannot be recharged 1. Bad connect with wire of battery 1. Repair or replace the wire of
interface board; battery interface board;
Some operation or Main board or specific parts Check the main board and
measure function failed damaged specific parts
Power on normally, but 1. Bad connect with keypad board 1. Repair or replace the FPC wire;
with blank screen or FPC wire; 2. Replace the display
during operation with 2. Malfunction of display 3. Replace keypad board;
white screen, or blank
3. Malfunction of keypad board 4. Replace main board
screen
4. Malfunction of main board
The letter or pictures 1. Resource file burning error 1. Burning the resource file again
shown wrong 2. Malfunction of main board 2. Replace main board
Keypad invalid 1. Bad connect with keypad board 1. Repair or replace the FPC wire;
FPC wire; 2. Replace keypad board
2. Malfunction of keypad board
37
Symptom Possible Cause Fault Elimination
Red alarm indicator 1. Malfunction of alarm light board; 1. Repair the wire or replace the
invalid or yellow alarm 2. Bad connect with keypad board alarm light board
indicator invalid FPC wire; 2. Repair or replace the FPC wire;
3. Malfunction of keypad board 3. Replace keypad board
4. Malfunction of main board 4. Replace main board
Button backlight failure 1. Bad connect with keypad board 1. Repair or replace the FPC wire;
FPC wire; 2. Replace keypad board;
2. Malfunction of keypad board 3. Replace main board
3. Malfunction of main board
keep the same pressure 1. Malfunction of pressure sensor 1. Replace pressure sensor
value 2. Malfunction of pressure test 2. Replace pressure test board
board
Pressure over range 1. Malfunction of pressure sensor 1. Replace pressure sensor
2. Malfunction of pressure test 2. Replace pressure test board
board
Can’t identify the 1. Short potentiometer slip handle 1. Mechanical card sets short
syringes and mechanical card sets out potentiometer slip handle;
2. Malfunction of short 2. Replace short potentiometer
potentiometer
Malfunction of pump 1. Malfunction of motor test board; 1. Replace motor test board
motor 2. Bad connect with motor wire 2. Repair or replace the
3. Malfunction of stepper motor motor wire;
circuit 3. Replace main board;
4. Malfunction of pump motor; 4. Replace pump motor;
Detecting faulty in 1. Bad connect with screw 1. Repair or replace the FPC wire;
syringe placing. clutch FPC; 2. Repair or replace the 6Pin wire;
2. Bad connect with 6Pin wire; 3. Replace pressure test board
3. Malfunction of pressure test
board
Screw clutch fault 1. Bad connect with screw clutch 1. Repair or replace the FPC wire;
detection FPC wire; 2. Replace screw clutch board;
2. Malfunction of screw clutch
board
Flange detect failures 1. Malfunction of flange detect 1. Repair wire or replace flange
detect board
38
Chapter 5 Repair and Disassembly
NOTES
Be sure to wear the antistatic wrist strap during the whole process of repair
and disassembly.
5.1 Tools
You might need the following tools during disassembly and replacement of the
components:
Forceps
Side cutter
Wrist strap
CAUTION
39
CAUTION
1. As shown below: remove 8 double gasket screws (M3x8) fixing the UP cover
40
2. Vertically upward pull up cover assembly, as shown below:
3. Remove the battery adapter board wire from the J3 connector on the battery adapter
board PCBA, as shown below:
J3 connector
41
2. Take out of the lock of keypad board 30pin connector, separate the door assembly
and main board, as shown below:
JP5 30pin connector
1. Cup the strap, remove the convex detection wire and the wire of battery adapter board
from JP1 and JP10 in main board, as shown below:
Cut the strap
JP1
JP10
2. Take out of the screw clutch test FPC wire from the JP9 in main board, as shown below:
JP9
42
3. Remove 1 screw M3x8, as shown below:
43
5. Remove the encoder detection optocoupler cable and motor cable from the J8,
J7 connector on the main board, take out of the double gasket screw M3x8, as
shown below:
Encoder detection optocoupler cJ8 M3*8 double gasket screw
Remove up cover assembly, door assembly and pump body assembly, the rest is rear
cover assembly.
44
2. Take out of the hollow shaft fixing piece, as shown below:
3. Cut the strap in nut clutch detection board PCBA by Side cutter, then remove pressure
detection wire from J1 connector of nut clutch detection board PCBA, as shown
below:
4. Remove the slider assembly from sliding block assembly in force, then get the push
box assembly, convex detection assembly, pump body transmission assembly
Note: Remove the pressure sensor assembly, make it directly in whole machine,
don’t need to take out of the slider assembly.
1. Point the 3 slider rubber plugs by forceps, as shown below:
45
2. Loose 3 PT3×8 self-tapping screw, as shown below:
3. Remove the back cover of slider, then you can see the pressure sensor and pressure
detection board PCBA, as shown below:
4. Take out of 2 M3x8 double gasket screw in pressure sensor assembly, as shown below:
46
5. Take out of the pressure detection board PCBA from slider, as shown below:
Notes: As show above, the assembly should be as a Min maintenance unit, if there
is one material damaged, need to replace one assembly completely.
47
5.3.7 Disassembly of screen, keypad board PCBA
2. Remove the alarm indicator board PCBA from J3 connector of keypad board PCBA,
as shown below:
48
3. Remove FPC wire of LCD from J5 connector of keypad board, as shown below:
LCD FPC wire J5 connector
4. Remove 4 PT2.6×6 self-tapping screw from keypad board PCBA, as shown below:
49
5.3.8 Disassembly of main board PCBA
1. Remove the rear assembly firstly, then take out of main board PCBA
2. Take short potentiometer fork, don’t lock the handle of hort potentiometer, as shown
below:
50
4. Remove 2 M3×12 double gasket screw, as shown below:
51
Chapter 6 Components
6.1 Overview
This chapter lists the main components of syringe pump, including the Explosive view and
parts material code, it’s more convenient for maintenance personnel to identify the parts
name when removal and replacement of parts.
6.2 Host
52
6.2.2 Parts list
S/N Part Number Part Name & Spec. QTY
1 115-021570-00 SP5 rear enclosure assembly 1
2 043-004208-00 Multi-channel pumps hasp (common mode material) 1
3 043-004180-00 Multifunction interface cover 1
Pan head screw M3x8(GB/T818) Nickel-plated
4 030-000305-00 16
steel double gasket
5 115-021569-00 SP5 UP cover assembly 1
6 115-021563-00 SP5 pump body assembly 1
7 049-000662-00 SP5 slider screw rubber plug 3
8 049-000664-00 Foot pad 4
9 115-021571-00 SP5 door assembly 1
Wire FPC electron wire 10P separation distance
10 050-001401-00 1
1.0mm length120mm
11 009-002714-00 Strap,3x150mm 2
Wire electron wire 12P double head connection wire
12 009-004120-00 1
separation distance 2.0mm length100mm
53
6.3 SP5 UP cover assembly
54
Chapter 7 Upgrade
7.1 Tools
During upgrade, you may need to use the following tools:
PC computer
55
2. The other side of multifunction serial port is standard BD9 serial port, as shown below,
connect this side to computer;
3. No insert AC power line, open the battery box then remove the lithium battery;
4. Press “start” key and “power” key together;
5. After loosening the keys, install the battery, then the machine will enter boot, the
BootLoader will be shown in display, as shown in Fig.1.3;
56
6. Open the PC burning tool (G-GCRJ-0004), as shown below, at this time the “upgrade”
menu is invalidly and can’t be selected.
7. Click “administrator login”, the login password dialog box will be popped shown, as
shown below:
57
8. Enter password, right now the upgrade menu can be selected, as shown below:
58
10. In upgrade screen, choose “browse”, then choose the burning software box needed,,
as shown below;
11. Click the “software” on the top right corner of PC burning tool software, software start
burning;
12. After completed, the upgrade will display “successful”, the grade progress will show
“100%”, as shown below;
13. Power off, disconnect the multifunction serial port line. The software upgrade is
successfully finished.
59