You are on page 1of 4

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/267337818

Measurement of density using oscillation-type density meters. Calibration,


traceability and Uncertainties

Conference Paper · June 2009

CITATIONS READS

6 12,102

4 authors, including:

Andreia Furtado Elsa Batista


Instituto Português da Qualidade Instituto Português da Qualidade
59 PUBLICATIONS 182 CITATIONS 72 PUBLICATIONS 337 CITATIONS

SEE PROFILE SEE PROFILE

Eduarda Filipe
Instituto Português da Qualidade
80 PUBLICATIONS 392 CITATIONS

SEE PROFILE

All content following this page was uploaded by Andreia Furtado on 27 October 2014.

The user has requested enhancement of the downloaded file.


MEASUREMENT OF DENSITY USING OSCILLATION-TYPE DENSITY
METERS
CALIBRATION, TRACEABILITY AND UNCERTAINTIES
A. Furtado, E. Batista, I. Spohr, E. Filipe
Instituto Português da Qualidade (IPQ)
Rua António Gião, 2 – Caparica – Portugal
afurtado@mail.ipq.pt

quality control of fuels and additives; in the chemical and


Résumé nuclear, to determine the concentration of acids, bases and
other solutions and to determine radioactive substances
concentration; in the food industry, cosmetics, etc.
La masse volumique d'un fluide peut être déterminé en
utilisant plusieurs instruments de mesure. Récemment, les This magnitude can be determined using several
densimètres à oscillations ont montré une grande measuring instruments, such as pycnometers, hydrometers
polyvalence, ils sont utilisés en différentes branches de and oscillation-type density meters, and also using the
l'industrie. Le Laboratoire des Propriétés des Liquides method of hydrostatic weighing. Recently the oscillation-
(LPL) de l'Institut Portugais pour la Qualité (IPQ) réalise type density meters have shown great versatility, and are
l'étalonnage de ces instruments par rapport à un étalon de being used in various branches of industry. The advantages
densimètres à oscillations, dans l’intervalle de mesure de of using these instruments are its response time, its use
la masse volumique de (500 à 2 000) kg/m3. La traçabilité simplicity and the small volume of the needed sample.
des mesures est assurée par des solutions certifiées du PTB
et de la DKD. La nécessité d'assurer la traçabilité
métrologique à un système de détermination absolue de la Measurement of liquids density
masse volumique des liquides a conduit à des évolutions
récentes dans ce domaine, qui seront décrites dans ce From Archimedes to Hans Stabinger
document. Il sera également décrit le modèle
mathématique et le bilan d’'incertitude de l'étalonnage de Formally assigned to Archimedes, but invented by Hipatia
densimètres à oscillations. of Alexandria (370-415 DC), the hydrometers are the
oldest, simple, highly effective and widely used tools for
Abstract measuring the density with different levels of accuracy.
The density of a fluid can be determined using several The method for hydrometers calibration, based on
measuring instruments. Recently the oscillation-type hydrostatic weighing, was introduced by Cuckow [1] in
density meters have shown great versatility, being used in 1949. This method allows hydrometers of any range to be
various branches of industry. The Laboratory of Properties calibrated at any point of its scale through the
of Liquids (LPL) of the Portuguese Institute for Quality measurement of the force of buoyancy in the air and when
(IPQ) performs the calibration of these instruments by immersed in the reference liquid (see Figure 1A). This is
comparison with a standard density meter, in the the most accurate method and is used in most National
measuring interval of density from (500 to 2 000) kg/m3. Metrology Institutes, including the IPQ [2]. The
The metrological traceability is given by Certified participation in the EURAMET 702 Project, gave LPL the
Reference Materials (CRMs) of PTB and DKD. The need recognition of its Calibration and Measurement Capability
to ensure traceability to a system of absolute determination (CMC) in the calibration of hydrometers by the Cuckow
of the density of liquids has led to recent developments in Method’s, using n-nonane at 20 °C, in a density measuring
this area which will be described in this paper. It will be interval [3] from (750 to 1 000) kg/m3, with an expanded
also described the mathematical model and uncertainty uncertainty of 0,1 kg/m3 for an expansion coefficient,
budget of the calibration of the oscillation-type density k = 2.
meters.

Introduction
The density is defined as the property of a body, given by
the ratio of its mass to its volume (kg/m3). The fluid
density is a major key in the control of most industrial
processes, because it allows not only, a better management
of the process but also the accurate determination of the
quantity and quality of the product. The measurement of A B
the density is used in drinks industry, to control the alcohol
content in binary mixtures or the sugar content in soft
drinks and fruit juices; in the pharmaceutical industry, to Figure 1. A - System for hydrometers calibration;
determine the specific gravity of medicinal preparations; in B - System for measuring the density of
the oil industry to determine the API gravity and for the liquids by hydrostatic weighing.
In 1967, the company Anton Paar GmbH presented the Where is the sample density,  the cell volume, m the
first digital density meter for liquids and gases. It was the mass cell and C the spring constant. Applying the square
first instrument to use the principle of the U shaped 4 2v and 4 2 m we
vibrating tube, of Stabinger Hans and Hans Leopold for of Eq. 1, and substituting G H
C C
the determination of the density. These first
obtain:
oscillation-type density meters were subject to errors
induced by the viscosity in most of the samples. These ( 2  H )
errors were usually higher than 0,7 kg/m3 in instruments  (2)
G
with a resolution of 0,1 kg/m3. In 1998, Hans Stabinger
developed a density meter (Paar DMA 5000) with a new During the oscillation of the U tube, the sample shows the
measurement cell which applied the viscosity correction of effect of damping of the oscillation. This damping is a
the sample, thereby avoiding the systematic errors of the function of the viscosity sample. The sample viscosity will
others instruments. also have the effect of apparently slightly move the
oscillation nodes, thus increasing the apparent volume of
the cell. When these two effects are combined, the error k ,
Oscillation-type density meters
in kg/m3, is of the order of k  0,05  , where  is the
Principle of operation viscosity in mPa·s.
The working principle of an oscillation-type density meter The measurement cell oscillates at several frequencies in
is based on the law of harmonic oscillation, in which a U two modes. This allows the damping due to the sample to
tube is completely filled with the sample to be analyzed, be measured and properly corrected. When the standard is
and subjected to an electromagnetic force. The calibrated, the density is a function of the oscillation
measurement of the frequency and duration of vibration of period,  and the damping.
the tube filled with the sample, allows the determination of
the density value of the sample. This measuring principle The internal software of the instrument includes the
is based on the Mass-Spring Model. assumption that:
The measuring cell consists of an oscillator formed by an
hallow U-shaped borosilicate glass tube that comprises   A 2 ( 1  D  damping  E  damping 2 )  B  C 4 (3)
about 0,7 ml of the sample. This tube has double walls and
the space between them is filled with a gas with high where A, B, C and D are coefficients determined in the
coefficient of thermal conductivity. In that space is also instrument calibration using two substances of known
placed a platinum resistance thermometer that allows the densities, typically air and water. Studies made in several
temperature measurement of the fluid during the density instruments of the same type using a wide range of CRMs,
measurement. The remaining instrumentation consists of a and water suggested that the Eq. 3 with its interaction
system of electronic excitement and electrical components between the period and damping is unnecessarily complex
that provide a signal transmission of the period for the [4]. For samples of density between (690 and 1 620) kg/m3
processor unit, free of interferences (see Figure 2). and viscosity above 600 mPa·s an equation of the form
  A  B 2  C  damping is more appropriate. For
liquids in the same density range, but with viscosities
Evaluation
Display below 30 mPa·s, the Eq. 4 describes better the cell
behaviour.
Coil C
Amplifier   A  B 2  (4)
damping
Magnet
Frequency oscillator
Calibration
Glass tube
Method and Metrological Traceability
The calibrations of oscillation-type density meters held by
LPL are performed at 20 ºC, through a comparative
Figure 2. Measurement cell of an oscillation-type density
method, using aqueous solutions in a density measuring
meters.
interval from (500 to 2 000) kg/m3. It follows a procedure
which has as a reference document the ISO 15212-1:1998
The U tube oscillates at its fundamental frequency, which
[5]. The density of the prepared solutions is measured with
is a function of the system mass. If we assume that the
the standard density meter of LPL (Anton Paar DMA5000)
sample volume inside the cell is constant, it can be seen
and the density meter of the client. The density
that the oscillation frequency is therefore a function of the
measurement of the liquid samples in both density meters
sample density. The oscillation period , is given by: are held almost simultaneously, thus ensuring that the
solution is in the same conditions of homogeneity. The
v  m
  2 (1) Certification of density of liquid samples is carried out
C using an internal procedure based on the same ISO.
The results of calibration, certification and the uncertainty rectangular distribution. Table 1 describes the uncertainty
budget, which will be further presented, are made in an sources mentioned.
Excel sheet previously validated for that purpose. The
results obtained in the calibration are presented in a Table 1- Uncertainty contributions in kg/m3 for the
Certificate of Calibration, that also describes the combined standard uncertainty of oscillation-type density
characteristics of the calibrated instrument, conditions of meters calibration and density certification of liquid
operation, errors and expanded uncertainty associated with samples.
each value of density tested. The results of the certification Uncertainty
of density of liquid samples are presented in a Certificate, Uncertainty Uncertainty
Distribution Contributions
describing the characteristics of the sample, conditions of Components Category
(kg/m3)
the operation and an associated uncertainty. Standard density meter
ucalibration B Normal 2,5×10-2
The metrological traceability [3] of density measurements uCRMs B Normal 2,5×10-2
is achieved through calibration of the oscillation-type uresolution B Rectangular 2,9×10-5
density meter with density standard samples of CRMs urepeatability A Normal variable
utemperature B Rectangular negligible
from PTB and DKD. The traceability of temperature
measurements of the sample is assured by calibration of a urepeatability A Normal variable
platinum resistance thermometer in the Temperature utemperature B Rectangular negligible
Laboratory of IPQ. Client density meter
urepeatability A Normal variable
Uncertainty Budget uresolution B Rectangular 2,9×10-4
The uncertainties were calculated according to the GUM
[6]. The uncertainty associated with the calibration of an
oscillation-type density meter is calculated combining two Concluding Remarks
sources of uncertainty of measurement: one of the client The oscillating tube density meter area of the LPL is
density meter (uclient density meter) and the other of the standard growing in terms of the number of clients that asks for
density meter (ustandard density meter). The uncertainty density certification of liquid samples and calibration of
associated with the certification of density of liquid density meters. The LPL main goal is to complete the
samples is only related with the uncertainty of chain of the metrological traceability of liquid density
measurement of standard density meter (see Figure 3). measurement. Since 2005, LPL has been working to
uclient density meter improve a system that allows the determination of the
liquids density by hydrostatic weighing, using as a density
urepeatability standard, a silicon sphere, which mass and density are
uresolution knew and certified (see Figure 1.B), allowing the decrease
ucalibration of uncertainty in density certification. Up to now, the LPL
ucertification participated in two EURAMET projects, which are still
urepeatability ucalibration on-going. We are looking forward to open new calibration
utemperature uCRMs and certification areas, in particularly, in viscosity and
urepeatability utemperature surface tension, as well as to develop new CRMs for those
ustandard density meter uresolution two properties.

Figure 3. Cause and effect diagram with the uncertainty


contributions for the combined standard
References
uncertainty of oscillation-type density meters
calibration, by comparative method and [1] Cuckow, F.W., “A new method of high accuracy for
certification of liquid samples density. the calibration of reference standard hydrometers”, J.
Chem. Ind., Vol. 68, pp. 44-9, 1949.
The component of the standard uncertainty associated with [2] Ferreira, M.C. e Dias, J.M., “The role of experimental
the repeatability of the measurements (type A) is obtained design in the hydrometer field”, XVIII Imeko World
by the standard deviation of the density measurements Congress, Brazil, 2006.
average (urepeatability). The other uncertainty components are [3] ISO/IEC GUIDE 99:2007, International vocabulary
evaluated by the type B method, including the uncertainty of metrology - Basic and general concepts and associated
related with the resolution (uresolution) of both density terms (VIM).
meters, as a rectangular distribution. Usually, the density [4] Stabinger, H., "Density Measurement using modern
meters calibrated at IPQ have a resolution of 1×10-1 kg/m3, oscillating transducers", South Yorkshire Trading
a resolution 100 times higher than the resolution of the Standards Unit, Sheffield, 1994
standard density meter. The main contribution to the [5] “Oscillation type density meters – Laboratory
measurement uncertainty of the density meter standard is instruments. ISO 15212-1, 1998.
the uncertainty of the CRMs (uCRMs) used for its [6] “Guide to the Expression of uncertainty in
calibration. Finally, the uncertainty component due to the Measurement.” BIPM et al - 1.ª ed. Genève: International
sample temperature inside the cell (utemperature), has a Organization for Standardization, 1995.

View publication stats

You might also like