Primo Therasonic 360 and 460

Ultrasound
The Primo Therasonic has been designed to offer
you a wide range of choices for your ultrasound
treatment programme. Widely used to improve
healing time on soft tissue injuries, it has integrated
evidence based algorithms and clinical protocols.

 Touch screen technology


 Dual frequency ultrasound (1MHz and/or
3MHz)
 Programmable memory
 Easy wipe clean reducing the risk of cross-
infection
Treatment Protocols: Pressure Ulcers, Chronic Venous Ulcers, Trigger Point
Low irritability, Tendon Injury/ Lesion, Chronic Osteoarthritis,
Rheumatoid Arthritis, Tennis Elbow, Carpal Tunnel, Fibromyalgia, Soft Tissue
Stretch, Impingement Syndrome, Calcific Tendonitis, Perineal Pain (Persistent),
Perineal Pain (Partum),

Mains powered as standard, optional battery available

Technical Specification
Size (h/w/d) 108 x 237 x 333mm
Weight 1.3 kg
Treatment Time 0-30 min
Treatment Programs 10 user-defined set-ups

Ultrasound Frequency 1MHz for 360, 1MHz and 3MHz for 460
Maximum Intensity 1.5 W/cm² in CW, 3.0 W/cm² in pulsed
Maximum Output Power 6W average (Mains operated only)
Output Modes CW pulsed 1:1, 1:2, 1:4 and 1:9
Pulse Duration 2ms
Contact Monitor Light on transducer or audio signal Transducer: Large transducer ERA 4 cm2,
Small transducer 1MHz 0.6 cm2 and 3MHz 0.4 cm2 Accessory Ordering Information
Large treatment head PMA912
Small treatment head PMA913
Ordering Information 8x250ml bottle of gel EMS502 Large Ultrasound Transducer
Primo Therasonic 360 1 Litre bottle of gel EMS502A ERA 4 cm2
Padded carrying case EMS530 BNR <5
Primo Therasonic 360 Single frequency ultrasound unit (1MHz) complete with large treatment
Beam Type Collimated
head, one bottle of coupling gel (250ml) mains only EMS3601 Treatment trolley EMS158
Primo Therasonic 460 Rechargeable battery pack (factory fitted)
Primo Therasonic 460 Dual frequency ultrasound unit mains only (1 MHz and 3MHz) complete with EMS161 Small Ultrasound Transducer
large treatment head, one bottle of coupling gel (250ml) EMS460 Ultrasound output indicator RTB01 1MHz 3MHz
Primo Therasonic 460 therapy unit as above.
ERA 0.6 cm2 0.4 cm2
BNR <5 <5
Mains and battery powered version
EMS460B Beam Type Divergent Collimated

Transducers for use with the Primo Therasonic 360 and 460 are fully
interchangeable and suitable for underwater treatment (IPx7 rated).
Further Technical Information The Primo Therasonic 360 or 460 power supply is designed to operate from any
50/60 Hz single phase supply between 100 and 240 Vac capable of supplying 50
Power Input 18V, 3.33A external PSU VA.
Battery (optional) Internal Rechargeable (NiMh) Connection is via an IEC socket on the power supply.
Internal Fuse T5A
Classification (EN60601-1) Class 1, Type BF
Size (height x width x depth) 90 x 230 x 290 mm
Weight 1.3 kg (excluding battery)
Treatment Programs 10 user-defined set-ups
Protocols (optional) 14 treatments
Dose Algorithm (optional) Calculates treatment settings from user entered
parameters.
Ultrasound
Frequency 1.1 MHz ±5% and 3.4 MHz ±5%
Maximum Intensity 1.5 W/cm2 in CW
3.0 W/cm2 in pulsed modes
Maximum Output Power 6 W average (Mains operation only)
Output Modes CW and pulsed 1:1, 1:2, 1:4 and 1:9
Pulse Duration 2 ms
Treatment Timer 0 to 30 minutes (treatment linked)
Contact Monitor Light on transducer (audio option available).
EMS PHYSIO LTD UK WWW.EMSPHYSIO.CO.UK GROVE TECHNOLOGY PARK, DOWNSVIEW ROAD, WANTAGE, OXFORDSHIRE OX12 9FE UNITED KINDGOM JANUARY 2016

You might also like