You are on page 1of 3

Θ

MID-RANGE

BROADBAND
SEISMOMETER

THETA Θ - TT120 - VEL

THETA Θ - TT120 - VEL is a mid-range


portable feedback broadband
seismometer housed inside a waterproof
enclosure. The low noise high dynamic
range broadband feedback seismometer
is based on truly rectilinear suspension
system which does not require mass lock.

Detailed sensor calibration information SUITABLE FOR LOCAL, REGIONAL &


is provided with every sensor, including TELE-SEISMIC RECORDING, INCLUDING:
sensor dc calibration levels, frequency
• S
 ingle or large scale seismic networks
response of the instrument and the
• M
 icro-seismic monitoring
transfer function in poles/zeros notation.
• After-shock monitoring
It is extremely simple to install the sensor. • Volcanology
Three levelling feet ensure that the sensor can
• S
 tructural health monitoring
be levelled easily with the aid of the bubble level.
• H
 ydrocarbon exploration
The North/South ordination pointes are machined
on the sensor base and the sensor component • P
 ermanent reservoir monitoring
fixing crews are fixed with dowel pins to achieve • I nduced seismicity detection
highest possible orientation accuracy. The errors • E
 xplosion monitoring
in pointer to the sensor orientation is less than
• P
 ost hole seismology
± 0.1 degrees.
• M
 icro-seismic monitoring

The product specifications and the stated data are subject to change without prior notice.

www.gaiacode.com DOC-TT120 M-Version C 1


SIGNIFICANT QUALITIES AND BENEFITS OF THE SYSTEM
• High data quality broad band data • Automatic self-installation and zeroing for
quick installation
• Truly rectilinear suspension system with high
cross axis rejection • Tilt tolerance of ±10 degrees

• Broad band data with high frequency response • P


 ortable and easy to deploy
extending to 150 Hz
• Waterproof with connectors mated
• Minimises equipment used during installation
• C
 onnector turret provides easy access to the
connectors and cabling

Figure 1:
Acceleration Ground referenced Power Spectral noise estimate in 1 Hz Triple red line is sensor
bandwidth in units of (m2/s4)/Hz. power spectral noise density
over the entire seismic
spectrum. The sensor noise
Sensor noise level level crosses the low noise
model form 8 Hz to 14
NLNM seconds.

The standard frequency pass band is flat to velocity from (120 Seconds period) 0.00833 Hz to
150 Hz, other options are available. Standard Pass band Frequency is shown in the provided
frequency response plot.

MEASURED FREQUENCY RESPONSE OF THE DIGITAL SENSOR

Note: The measured frequency response is given as low and high frequency sections in order to identify
accurately the corner frequency of the transfer function.

The product specifications and the stated data are subject to change without prior notice.

www.gaiacode.com DOC-TT120 M-Version C 2


TECHNICAL SPECIFICATIONS
Standard Frequency Response: Polarity protected, over voltage protection with
Flat velocity 0.008333Hz (120s) to 150 Hz transient suppression.

Other optional responses: Power consumption at 12 Volts:


60 seconds to 150 Hz 830 Milli Watts
30 seconds to 150 Hz Low power option: 235 Milli Watts
20 seconds to 150 Hz
PHYSICAL
10 Seconds to 200 Hz
Case diameter: 138 mm
1 second to 200 Hz
Case height: 185 mm
Analogue sensor sensitivity:
Differential: 2*1500 V/m/s. North South Pointer: Machined to the base.
Other responses are available.
Base plate and top cap: Hard Anodised Aluminium.
SENSOR CONTROL LINES Environmental: IP-68
Calibration signal: Can be applied to each axis. Weight: 3.1 Kg
Calibration enable via serial Port.

Mass Centring Range: ±10 Degrees.

Mass Lock: Not required.

Lowest spurious Resonance: 340 Hz vertical


and horizontal modules.

Cross axis sensitivity: -68 dB.

Linearity: Measured at 1 Hz: -85 dB

Bubble level: 20 mm diameter

Logic Control Lines:  


Calibration enable and centre via serial interface.

CALIBRATION SIGNAL INPUT


Wave forms:
Square, Sine, Step, Pulse with duty cycle and
frequency control and pseudo random signals
can be injected to calibrate the sensor.

Operating temperature:
-20 to 75 degrees centigrade.

Power supply: Figure 2:


Sensor production at Gaiacode, ALPHA-TA120-LIGHT,
+9 - +36 V DC, Galvanically isolated analogue
THETA-TT120-VEL, SIGMA-TSD4G-ACC.
input stage of digitiser.

Galvanically isolated sensor electronics.

The product specifications and the stated data are subject to change without prior notice.

For more information contact us:


Dickers Farm, Bramley Road, Call: 07468 439741
www.gaiacode.com Silchester RG7 2LU, United Kingdom. Email: contact@gaiacode.com