OPERATING MANUAL

With a description of kINPen®IND - E

Picture may differ from the original

Not permitted for use on humans!

neoplas GmbH Tel.: +49 3834 515 210


Walther-Rathenau-Str. 49 a Fax: +49 3834 515 209
D-17489 Greifswald E-Mail: contact@neoplas.eu
GERMANY www.neoplas.eu
kINPen® IND – operating manual – state: 2022-02-23 2
3 kINPen® IND – operating manual – state: 2022-02-23

Content

Content ............................................................................................................................ 3
1 Description of operating mode ................................................................................... 4
2 General ...................................................................................................................... 5
2.1 Signs and symbols ........................................................................................................ 5
2.2 Scope of application .................................................................................................... 6
3 Technical Data ........................................................................................................... 7
3.1 Ambient Conditions ..................................................................................................... 8
3.2 Scope of delivery ......................................................................................................... 8
3.3 Consumables and spare parts, optional accessories ................................................... 8
3.4 Functioning and Operating Elements .......................................................................... 9
3.5 Display elements and error messages ....................................................................... 11
4 Transport and Storage.............................................................................................. 12
4.1 Transport ................................................................................................................... 12
4.2 Storage ....................................................................................................................... 13
5 Installation .............................................................................................................. 14
5.1 Setting up the device ................................................................................................. 14
5.2 Change of electrode head (optional)......................................................................... 15
5.3 Installation of the helium insert (optional) ............................................................... 16
5.4 Connection to gas supply .......................................................................................... 17
5.5 Connection to voltage supply .................................................................................... 18
5.6 Performance check of the device .............................................................................. 20
6 Commissioning and Application ............................................................................... 21
6.1 Step-by-step instructions........................................................................................... 23
6.2 How to control kINPen® IND-E ................................................................................... 26
7 Liability .................................................................................................................... 28
8 Cleaning / Inspection / Maintenance ........................................................................ 29
9 Disposal ................................................................................................................... 30
10 Manufacturer .......................................................................................................... 30
Appendix ........................................................................................................................ 31
kINPen® IND – operating manual – state: 2022-02-23 4

1 Description of operating mode

The plasma source kINPen® IND consists of an operating device and a connected handheld
device to generate a plasma jet. Both devices are connected through a combined service pipe
(gas, voltage, earth conductor). The operating device is attached to the external voltage supply
via a power cord.

Inside the plasma source kINPen® IND a high electric voltage (2 − 3 kV) is generated by a high
frequency generator (1 MHz) connected to a metal electrode inside of a ceramic capillary. The
working gas argon flows through the capillary. This capillary at the distal end of the handheld
device is surrounded by a grounded ring electrode. The high frequency discharge between the
electrodes converts the flowing working gas into physical plasma continuously. Due to the gas
flow the plasma is jet-like carried out of the handheld device.

A master power switch puts the device in the state of operational readiness. The actual plasma
operation then is triggered by a start/stop button. With start of the treatment gas flow starts
and the voltage is applied. Triggering the master power switch or the start/stop button the
voltage as well as the gas flow is stopped. The active species disappear, and the treatment
time is controlled precisely and in an easy way.
5 kINPen® IND – operating manual – state: 2022-02-23

2 General

The Manual is to be read carefully and completely prior to commissioning. The kINPen® IND
is an electric device and requires careful and responsible handling for transport, installation,
maintenance, use and disposal of equipment and the device itself.
Keep this manual for later references.

2.1 Signs and symbols

WARNING Danger to life or risk of severe injury

CAUTION Risk of injury

NOTICE Possibility of technical defect

Important information concerning the understanding of the device or the


appropriate usage thereof.

High voltage or risk of electric strike.

Electronic device, do not dispose of in regular household garbage.


kINPen® IND – operating manual – state: 2022-02-23 6

2.2 Scope of application

Surfaces are cleaned, activated and decontaminated at atmospheric pressure with the handy
kINPen® IND. The device is particularly used for surface treatment of temperature-sensitive
materials as, for instance, plastics. In addition, kINPen® IND provides an easy access to
geometric challenging surfaces such as narrow clefts, capillaries or subtlest bores.

The kINPen® IND base model is argon-powered. The admixture of reducing or oxidizing gases
can be done up to the percentage range. Furthermore, the device is convertible through the
simple change of its electrode head to operate with molecular gases like air or nitrogen.
The plasma is automatically controlled. Continuous use of kINPen® IND is possible.

The kINPen® IND can be expended modularly by means of additional devices:

1) Gas mixing unit (GMU): to admix neutral process gas (e. g. argon) with none corrosive
molecular gases (air, nitrogen, oxygen). The GMU II is controlled exclusively via USB
interface.

2) Gas supply unit (GSU): one gas channel supply unit with a flow indicator and flow
control for none corrosive molecular gases (air, nitrogen, oxygen) or argon. The GSU
is recommended for usage of kINPen® handheld device with its reactive/shielding gas
head. It can be attached to the kINPen® main unit and to the GMU (gas mixing unit)
and is also separately usable for other experiments and processes.

Figure 1: Products of kINPen® IND family


7 kINPen® IND – operating manual – state: 2022-02-23

3 Technical Data

Weight: 4,0 kg

Power supply
− Supply voltage, -frequency: 100 − 240 VAC, 50/60 Hz
− Mains Fuse: 2 x T 1A E 250V
− Max. power consumption: 50 W
− Protection Class: I
− Operation mode: continuous operation

Gas supply
Argon Argon Helium Com- Nitrogen Nitrogen
with max. pressed with max.
5 % H2 + air 5 % H2
2 % O2
Purity min. 1)

4.6
Gas flow rate (l/min) 5±2 5±2 3±1 5±2 5±2 5±2
Inlet pressure 2 – 3 bar
Noble gas head X X X
Molecular gas head X X X
Helium insert X

Base Unit (operating device)


− Dimension L x W x H: 330 mm x 180 mm x 105 mm
− Protection category: IP40

Plasma Generator (handheld device)


− Dimension Ø x L: 20 mm x 180 mm
− Supply Line Length: 1,50 m
− Protection category: IP30

1)

Quality of pressured air according DIN ISO 8573-1:2010 for kINPen:

Feststoffpartikel particles 1
Wasser humidity 2
Öl residual oil 1
kINPen® IND – operating manual – state: 2022-02-23 8

3.1 Ambient Conditions

Transport
− Temperature: -40 – 70 °C
− Relative air humidity: 10 – 100 %

Operation:
− Temperature: 15 – 40 °C
− Relative air humidity: 15 – 75 %
− Air pressure: 800 – 1060 hPa

NOTICE Ambient Condition

Excessive heat / cold or excessive humidity can impair the functionality of the
device or destroy it. Make sure to use the device with regards to the ambient
conditions.

3.2 Scope of delivery

- kINPen® IND-supply unit with permanently connected plasma generator (base unit
and handheld unit)
- built-in flow indicator with throttle and safety shutdown at subthreshold gas flows
- Electrode head for argon gas
Please note: If you would like to operate the electrode head with other gases, please
contact the manufacturer in advance.
- Protective cap for the electrode head
- Supply line from kINPen® IND main unit to handheld device (length: 1,5 m)
- Gas supply line (length: 1,5 m, Ø 6 mm)
- Plug for gas connection of the device (rear of device)
- Power cord
- Manual
- Spare Parts List

3.3 Consumables and spare parts, optional accessories

See Price and Spare Parts List

NOTICE Spare parts

The device is intended for use with the supplied accessories only.
9 kINPen® IND – operating manual – state: 2022-02-23

3.4 Functioning and Operating Elements

Figure 2: Operating unit with handheld device, stand (optional) and protection cap

Picture may differ from the original


kINPen® IND – operating manual – state: 2022-02-23 10

Figure 3: Front view of the operating unit with operating elements

Picture may differ from the original


1. Power Switch
2. Start/Stopp Button
3. Error Display (LED)
4. Gas Flow Control
5. Supply tube for handheld device

Figure 4: Display- and operating elements of the gas flow control

Picture may differ from the original


4a. Display
4b. Adjusting screw
4c. Clamp screw
11 kINPen® IND – operating manual – state: 2022-02-23

Figure 5: Rear view of operating device

Picture may differ from the original


.

6. Power supply
7. Fuses
8. Type plate
9. Connecting socket for additional potential equalization (ZPA)
10. Gas connection

3.5 Display elements and error messages

The operating mode of the device is displayed via various indicators on the front (see page 9
and page 10).

Main power switch


The main power switch lights up when the device is put to operational readiness via this
switch.

Start/Stop Button
The start/stop button includes an illuminated ring (LED). By pressing the button, the gas flow
to the handheld device is started and the plasma is ignited. During the operation, the LED
glows. The LED turns off after pressing the start/stop button a second time for ending the
treatment.
kINPen® IND – operating manual – state: 2022-02-23 12

Error indication (LED)


The error indication lights up if the gas flow rate is out of the designated range. (see chapter 3:
Technical Data). In this case, the voltage feed to the handheld device is interrupted or not
released.

Check your gas supply for proper operation (sufficient gas pressure, position of valves) and
control the gas flow rate via the adjusting screw (4b) at the gas flow control (4) of the device.
Please note, that for operating the adjusting screw (4b) the clamp screw (4c) has to be
released.

Gas flow control


The gas flow control features a display (4a) which shows the gas flow rate as litre per minute
(l/min). The display shows green numbers if the value is within the designated range and red
numbers otherwise.

4 Transport and Storage

4.1 Transport

NOTICE Transport – Transportation case

The device may be damaged due to improper transport. The device kINPen® IND
can be supplied in a transport case. In order to prevent damages, the transport
case is recommended for transportation.

Note

Inside the devices gas supply lines condensation of air humidity can occur,
caused by changing temperature during the transport. This may influence the
plasma characteristics directly after starting the device. To avoid:

• make sure to protect the electrode head during the transport with the
provided cap,

• for the duration of the transport seal the gas inlet on the rear of the
device with the provided plug,

• run the device idle for 15 min before starting the treatment to flush out
existing humidity.
13 kINPen® IND – operating manual – state: 2022-02-23

4.2 Storage

NOTICE Dropping

The gas outlet on the electrode head of the handheld device may be damaged
through shock or falling down, so that a proper function of the device may not
be guaranteed anymore. In order to prevent damage:

• when not in use, protect electrode head of the handheld device with the
protection cap provided by the manufacturer

• place the handheld device in the stand (optional)

Note

When the device is not being used for a longer time, store device in the transport
case provide for by the manufacturer.
kINPen® IND – operating manual – state: 2022-02-23 14

5 Installation

Responsibility of the user

A user qualified for electrical appliances and gas supply must set up the device.
The parameters for the electrical connection and the gas connection (see
chapter 3: Technical Data) must be observed.
If the customer renounces the on-site introduction, necessary repairs caused
by faulty operation of the device go fully at the expense of the customer.
Responsibility for the gas supply (pressure conditions, gas flow rate, gas purity)
of the device lies with the user.

Decommission and reinstallation

If the device is to be used after full decommission (device was completely


disconnected from gas and power supplies), it is mandatory to follow the details
in the proceeding of this chapter.

5.1 Setting up the device

Place the base unit close to its gas and power supply on a dry, clean, horizontal and stable
surface. Place the handheld device in the optional delivered stand.
The kINPen® IND is supplied as standard with an electrode head for noble gases (Argon).
Depending on the desired application of the kINPen® IND, you may install an electrode head
for molecular gases or a helium insert (both optional, see chapter 5.2: Change of electrode
head and chapter 5.3: Installing the helium insert).
Otherwise, follow the instructions in chapter 5.4: Connection to the gas supply and chapter
5.5: Connection to the line voltage.
15 kINPen® IND – operating manual – state: 2022-02-23

5.2 Change of electrode head (optional)

Change of electrode head

To change the electrode head separate the device from the mains!

Figure 6: Electrode heads for noble gases and molecular gases

Standard: electrode head for noble gas (e. g. Optional: electrode head for molecular
argon) gases (e. g. compressed air, nitrogen)
Feature: red gasket Feature: blue gasket

Figure 7: disassembled handheld

1. To change the head lose the locking screw (11) at the front end of the handheld
device tube (13). Use a 0,9 mm-Allen-key.

2. Remove the electrode head (14).

3. The other electrode head has to be screwed hand-firmly to the handheld device.

4. Replace the safety screw (11) into the opening (12), tighten it lightly.
kINPen® IND – operating manual – state: 2022-02-23 16

5.3 Installation of the helium insert (optional)

NOTICE Helium insert

When using the helium insert, note the following:


• The adapter may only be used in combination with the noble gas head.
• The kINPen® IND must be operated with another gas (e. g. air) to cool the
handheld device (see chapter 5.4: Connection to the gas supply).
• The safety screw (11) is not used during the use of the helium insert. The
opening (12) is kept free.

1. Remove the electrode head (see chapter 5.2: Change of electrode head), loosen the
locking screw at the front end of the handheld device tube.
2. The helium insert is placed between the handheld device and the noble gas head.

Figure 8: Handheld disassembled with adapter for additional gases

3. The helium insert and the inert gas head, have to be screwed hand-firmly to the
handheld device.

Figure 9: Handheld assembled with adapter for additional gases

4. The basic unit is to be connected due to the cooling of the handheld device to a gas
supply (e. g. compressed air).
5. The process gas (helium) is introduced via the connection through the helium insert
in the kINPen® IND for plasma generation.

The recommended gas flow using the insert is at about 5 l/min air and 3 l/min helium (see
chapter 3: Technical Data). Again, for both gases, the use of a pressure regulator or
17 kINPen® IND – operating manual – state: 2022-02-23

the connection to a pressure-reduced central gas supply is required (see chapter 5.4:
Connection to gas supply).
5.4 Connection to gas supply

From a purely technical point of view, the device may be operated with different types of
gases. The plasma source kINPen® IND is being delivered with a gas supply hose (Ø 6 mm) for
the connection of the operating device to an appropriate gas supply unit. When connecting to
a gas cylinder or a central gas supply system, an adequate pressure regulator is necessary for
a solid pressure reduction.
Remove the equipment from the packaging and create a steady connection between the
operating unit and the gas supply.

CAUTION Type of gas

The device was designed for the application range described in this manual.
Through a use of a different gas, than the one foreseen in this manual, the
device can be damaged or it produces a plasma, which does not feature all
attributes needed for the treatment. Ensure the connection of the device to a
gas supply apparatus, which supplies the gas in a specific purity according to
chapter 3: Technical Data.

ACHTUNG Gas pressure

The device was developed for the gas pressure described in chapter 3: Technical
Data. Considerably higher inlet pressure may damage the device. Compare the
information in this manual with the gas pressure on the pressure regulator of
your gas supply unit and create an appropriate gas pressure, in order to prevent
damages at the device.
kINPen® IND – operating manual – state: 2022-02-23 18

5.5 Connection to voltage supply

The plasma source kINPen® IND is supplied with a suitable power cord.
Use the enclosed power cord to connect the unit to your voltage supply.

WARNING Risk of explosion

The device produces gaseous plasma by electrical ionization of gas and thus
produces an ignition source. The device is only meant to be used in dry rooms
that do not present a risk of explosion. Use of the device in the vicinity of
anesthetic gases and pure oxygen is expressly prohibited.

Electric shock via humidity, liquids, or water

The device is not protected against humidity, splashing water, or the


permeation of liquids. The device may be damaged by ingress of liquids or water
and may cause electric shocks.

− Protect the device from liquids and moisture.


− Do not submerge the device in water or other liquids.
In case of liquid entry, disconnect the power cable and consult the manufacturer
service about inspection of the device.
19 kINPen® IND – operating manual – state: 2022-02-23

WARNING Protective ground and additional potential


equalization

For the protection of all persons, in the event of a malfunction of the device,
dangerous electrical buildups will be discharged through a ground connection.
The effectiveness of the ground connection is dependent on the connection of
the device to a wall socket with a ground wire. A further protection mechanism
is realized by connecting the device to an equipotential bus bar.

− Exclusively use the included power cord to establish an electrical connection.


− The plug may only be plugged into a socket with a protective earth
conductor.
− Use of an extension cable without protective ground is prohibited as this
may decrease the effectiveness of protective measures.
− Establish a connection between the device and your facility’s equipotential
bus bar. To do so, use the correspondingly marked socket located at the back
of the device and a (optionally available) potential equalization cable.
kINPen® IND – operating manual – state: 2022-02-23 20

WARNING High voltage

When turned on, high voltage is generated inside the handheld device.

− The device may only be used and operated by properly instructed personnel.
− Do not open the device. Any tampering with the device is strictly prohibited.
− Do not insert sharp objects into the gas emission nozzle on the electrode
head of the applicator.
− Protect the device from mechanical damage! In particular, secure the device
to prevent impacts and droppings.
− If visible damage to the gas emission nozzle is evident, immediately
disconnect the device from the power source and contact manufacturer
service.

NOTICE Supply voltage

Connection to an unspecific supply voltage may damage the device or may result
in plasma with properties that are not suitable for treatment.

Connect the device to a voltage supply as designated on the device label.

Note

The device’s positioning must enable easy access to the power switch, the plug,
and the additional potential equalization.

5.6 Performance check of the device

In order to perform a performance check of the device after initial installation, the device is
to be commissioned as described in chapter 6: Commissioning and Application.
21 kINPen® IND – operating manual – state: 2022-02-23

6 Commissioning and Application

Note

Mind the following safety information in order to protect yourself during the
application.

CAUTION Ozone

During treatment, the plasma gives rise to low concentrations of ozone


(detectable by scent). At distance of 10 cm from the handheld unit, the safety
limit is upheld. Past this point, the ozone is sufficiently dissipated in the
surrounding air. At a shorter distance to the handheld device, an exhaust is
recommended. In small rooms and with multiple treatments, sufficient
ventilation (e. g. an open window) is required.

Impact of environmental conditions

The device produces gaseous plasma. The plasma properties may depend on
environmental conditions. Consider the permissible operation conditions (see
chapter 3.1). Consider also, that after the plasma is switched off, due to its
design, a portion of the gas line will be aerated. As a result, moisture from the
ambient air may collect in the gas line and may influence the plasma properties
upon subsequent usage. This may affect the intensity of the produced UV
radiation as well as the quantity or produced reactive oxygen species,
particularly OH radicals.

Stable operation of the device may be achieved as follows:

− Place the included protective cap on the electrode head when not in use.
− To ensure sufficient gas purity before the start of treatment, turn the
plasma on before the initial treatment of the day, until a stable plasma jet
occurred. Before further treatments on the same day, a warm-up time of
ca. 1 − 2 minutes before the start of treatment is recommended.
kINPen® IND – operating manual – state: 2022-02-23 22

WARNING Improper use

Use the device exclusively for the intended purpose.

CAUTION Dropping

The device may fall if placed on an unsuitable surface and may thereby causing
injuries.

− Ensure that the main unit always rests on a level, stable surface and that the
use of the handheld unit does not result in the danger that the device may
fell down.
− The bearing surface must be dry and clean.
− Place the handheld device on the (optional) stand when not in use.
23 kINPen® IND – operating manual – state: 2022-02-23

6.1 Step-by-step instructions

1. Place the handheld unit on the stand (optional).

2. Remove the protection cap from the handheld device.

Figure 10: Handheld device with protection cap

Picture may differ from the original

3. Check and look for possible mechanical damages at the gas outlet at the electrode head
on the handheld unit.
4. Check the type of gas and its purity at your gas supply unit.
5. Guarantee the gas supply from the gas supply unit to the device, by opening the
respective valve/s.
6. Check the gas inlet pressure at the pressure regulator of your gas supply unit and
control the value stated in this manual.
7. Make sure that the power switch is switched to zero (0).

Figure 11: Power switch


kINPen® IND – operating manual – state: 2022-02-23 24

8. Connect the device with the power cord to the AC/AC converter. Connect the AC/AC
convertor to the line voltage. (Optional: Connect the device to the additional potential
equalization panel of your facility).
9. Switch the kINPen® IND device on, by switching the power switch to position one (1).
The lighting of the power switch as well as the display of the gas flow regulator are
now active and the device is ready for use.
10. By shortly pushing the start/stop button, the gas flow to the handheld unit is being
started and the plasma is generated. The green ring LED of the button is now
illuminated.
11. Make sure (through visual inspection) that a visible plasma of approx. 7-11 mm is
generated and released from the handheld device. Control the gas flow on the display
of the gas flow controller and adjust if needed to the required gas flow rate by turning
the valve screw of the gas flow controller.

Figure 12: Start/Stopp Button with activated plasma

12. In order to achieve sufficient gas purity in the pipelines and thus ensure reproducible
results, the unit is allowed to run once for at least 15 minutes on the application day.

13. After the treatment switch off the plasma by pushing the start/stop button. The
button-light will also turn off.

14. Upon termination of the treatment, set the power switch to zero (0), disconnect the
device from the power line.

15. For protection purposes, place the protection cap on the electrode head at the front
part of the handheld unit.
25 kINPen® IND – operating manual – state: 2022-02-23

Note concerning gas flow regulation

The intended use of the kINPen® IND requires the compliance with the required
gas flow rate within the stated tolerances. The gas flow controller only sets the
supply voltage for the handheld device free to generate a plasma, if the current
gas flow rate lies within the specified range. In this case, the display of the gas
flow controller is illuminated by a green light. In case the gas flow is too high or
too low, the display will be red and the error display will be illuminated as well.

Note concerning the generation of plasma

Occasionally, the plasma will not be generated immediately after pushing the
start/stop button. This may be the case if the gas supply line was not cleaned
with an adequate amount of clean gas. Should this problem remain after several
tries and thus after being adequately rinsed with gas, the problem may be a
defect of the device. In this case, please contact the manufacturer.
kINPen® IND – operating manual – state: 2022-02-23 26

6.2 How to control kINPen® IND-E

Figure 13: kINPen®IND-E

The BNC-connectors replace the manual button Start/Stopp and the LED that display a
wrong gas flow outside of the working window.
Figure 14: Front view of the operating unit with operating elements

Picture may differ from the original

1. Main Power Switch


2. BNC Input
3. Error Out
4. Gasflow switch with flow meter and needle valve
5. Cable to the plasma unit
27 kINPen® IND – operating manual – state: 2022-02-23

Switch on the main power switch (1) to bring the kINPenIND-E in standby position.
Provide Nitrogen, pressured air or synthetic air (2-3 bar) on the backside of the kINPen.
Provide a positive voltage between 5V and 24V (inner contact +, shield ground) to start the
kINPen.
If the gas flow is between 3-7 liters per minute (working window can vary) the Error Out
should be lead 0V, in the case of error the BNC output is feed 24V.
kINPen® IND – operating manual – state: 2022-02-23 28

7 Liability

The manufacturer recommends to only use the device kINPen® IND under the designated
conditions and for the designated purposes, and to take care of it as described in this manual.
The manufacturer cannot be held liable for damages resulting from improper usage of
kINPen® IND or any usage not agreed to by the manufacturer.

Furthermore, the general terms and conditions of the manufacturer apply.


29 kINPen® IND – operating manual – state: 2022-02-23

8 Cleaning / Inspection / Maintenance

WARNING Cleaning / Disinfection

The device must be protected against the permeation of liquids. The device may
be damaged due to the permeation of liquids or water and can cause an electric
shock.

• Disconnect the device from the power supply before


cleaning/disinfection.

• For the disinfection of the device kINPen® IND, use a towel which has
been moistened with a disinfectant – never spray the disinfectant
directly on the operating unit or the handheld device.

Maintenance
The electrode head of the kINPen® IND is subject to slow wear. The wear of the head varies
according to the gases used and the duration of use of the device. A spent head ignites only if
delayed or the plasma jet is irregular.
A replacement head is available from the manufacturer, a change by the customer is possible.
A maintenance of the device by the manufacturer with adjustment of the handheld on the
new head is recommended.

Error/ Customer Services/ Reparation


An error may be present, if despite of a sufficient gas flow rate and repeated start trials the
plasma is not being generated or if it extinguishes prior to the time control of 60 seconds or if
the plasma is very inconsistent (changing length, colour, sounds).
In case of an error, please contact the manufacturer.

Repairs to the device must be carried out exclusively by the manufacturer, any manipulation
of the device is prohibited.
kINPen® IND – operating manual – state: 2022-02-23 30

9 Disposal

Storage or disposal of the sales packaging


The sales packaging is made according to environment-friendly standards and is thus
recyclable. Sending back the packaging to the manufacturer safes resources and reduces
waste accumulation. For potential returns, in-house transports, or storage, we advise you to
safe the original packaging.

Disposal of the device after usage


Electric or electronic devices often contain valuable materials. However, they also contain
dangerous substances, which were crucial for its function and its security. When disposed of
as general waste or when treated in a wrong way, they can damage human health and the
environment. Therefore, do not dispose of the devices in residual waste.

This product is not meant to be disposed of as residual waste.

In case this device must be disposed of, send it to the address below for an
adequate disposal.

According to the WEEE Directive. 2002/95/EG.

10 Manufacturer

neoplas GmbH
Walther-Rathenau-Str. 49 a
17489 Greifswald
GERMANY

CEO: Ulrike Sailer


Tel.:
E-Mail: contact@neoplas.eu

www.neoplas.eu
31 kINPen® IND – operating manual – state: 2022-02-23

Appendix

Installation protocol
Serial-Number: _______________________

Nr. Procedure done

Unpacking the device

1 Check that delivery is complete according to chapter 3.2: Scope of delivery

2 Check that device and equipment are in sound condition

Set-up of device
Installation location and environmental conditions correspond to the
3 requirements of chapter 3: Technical Data

Connection to the gas supply (see Section 5.4)

4 Check gas supply unit, type of gas = argon

5 Check gas supply unit, gas purity min Ar 4.6

Connect gas supply unit with device with the help of the provided gas
6 connection line (plastic hose)

7 Check gas pressure regulator: pressure reduction to 2-3 bar

Connection to power supply voltage (see Section 5.5)

Check if voltage and frequency of power supply line correspond to the


8 specifications on type plate

9 Connection of device to the potential equalization panel of the facility

10 Connection of power supply line to the device


11 Check the functional capacity of the grounded safety power socket
12 Connect the power supply line to power socket
kINPen® IND – operating manual – state: 2022-02-23 32

Serial-Number: _______________________

Performance Test

Check of the according function, procedure according to chapter 6:


13 Commissioning and Application, Numbers 1-15 of the list

14 Regulate Gas flow controller to a flow rate of 5 ± 1 l/min


15 Plasma shows typical blue-purple colour
16 The handheld device generates a plasma of about 7-11 mm
17 Plasma can be manually switched off and relit repeatedly.

➔ The device was properly installed and tested successfully as to its function.

-----------------------------------
Place, Date