Professional Documents
Culture Documents
GEHC - PET CT Discovery IQ Data Sheet
GEHC - PET CT Discovery IQ Data Sheet
Discovery IQ
Data Sheet
30 recommendations.
• Summary Table for interactive, efficient review
25 of PET functional and CT morphological
ROI/VOI data.
20 • Q.Check outputs quantitative information for
PET imaging and patient parameters for each
15
loaded study.
• Adaptable workflow for clinical reading to
10
advanced research using tools supporting
5
PERCIST, EORTC, RECIST 1.0, 1.1, and WHO
criteria (optional).
0 • Direct access to Integrated Registration option.
0 5 10 15 20 25 30 35 40
Activity concentration (kBq/mL)
Figure1: NEMA NU 2-2012 count rate accuracy graph. LightBurst has been
designed to achieve accurate quantitation at any count rate up to 22 kBq/mL.
‡Optional
1
Represents typical system performance and may vary from one system configuration to the other.
LightBurst – PET/CT Innovative Acquisition
LightBurst PET Detector Q.CORE - PET Reconstruction Architecture
LightBurst PET detector features a high stopping power material Q.CORE is a hardware reconstruction technology that enables an
with Dual Acquisition Channels for optimal performance in both advanced full 3D iterative reconstruction with no data mashing
low- and high-count rate acquisitions. and spanning and all the correction in the iterative loop.
Dual Dual Acquisition Channels technology reduces Q.CLEAR‡ Full convergence iterative reconstruction
Acquisition dead time losses and pileup at high count rates. designed to provide
Channels It provides ±3.5% count rate accuracy (mean ∆r) • up to 2 times improvement in PET quantitation
up to 22 kBq/cc providing quantitative accuracy accuracy (SUVmean)
for a large spectrum of PET tracers. • up to 2 times improvement in image
It also achieves ±1.0% count rate accuracy quality (SNR)
(mean ∆r) up to 11 kBq/cc for most of your 18F SharpIR‡ Point Spread Function modeling enhances visual
and 11C clinical and research activities. contrast and resolution in both whole-body and
Q.Temp Each block detector includes four temperature brain PET images. SharpIR provides uniform High
sensors, which control every single anode of Definition resolution over a 70 cm PET FOV.
the LightBurst PET detector and, therefore,
VUE Point HD VUE Point HD utilizes a full 3D Iterative
quantitation and detectability performance.
Reconstruction technique with all corrections
Acquisition gain control: ±0.25C.
within the loop. It offers: enhanced resolution
Scalability LightBurst PET detector design allows on site detector geometry modeling; model-based 3D
field upgrade from a 2 ring axial coverage up to scatter correction; scatter estimation outside
a 5 ring extended axial FOV. the FOV; exclusive randoms corrections based
Out of field- End and front shields are designed to reduce on singles; dead time correction with pileup
of-view scatter out of the FOV. In addition, Tungsten estimates.
scatter shield has been added to the detector WideView See the entire FOV with
limitation bloc design to reduce cross crystal scatter 70 cm DFOV in PET and CT.
quantitative artifacts.
Sensitivity LightBurst achieves new performance
milestones thanks to the combination of a
compact ring geometry, high stopping power
scintillator and a large solid angle detection up
to 38.5%.
Typical 18F acquisition Typical 11C acquisition Typical 15O, 82Rb acquisition
94% procedures* 3% procedures* 3% procedures*
120
Count Rate (kcps)
100
80
40
20
0
0 2 4 6 8 10 12 14 16 18 20
Ga
68
NaF FDG F-PIB 11
C-Choline Acetate 13
NH3 82
Rb (3D) O
15 82
Rb (2D)
Activity Concentration (kBq/mL) *IMV PET 2012 Imaging Mkt Sum Rpt
Transaxial field-of-view 70 cm
Number of detector units 36
per ring
Coincidence window 9.5 ns
Energy threshold 425 keV
Overlap 24% typical
Users defined: 1% - 50% Conventional PET Iterative Reconstruction Q.CLEAR
2 iterations 25 iterations
‡Optional
3
Represents typical system performance, measured according to NEMA Standards Publication NU2-2012. Sensitivity represents the measurements at 0 cm.
CT Innovative Technologies
High IQ with Volara Digital DAS Dose Reduction Technology
CT Data Acquisition System (DAS), with an increased sampling ASiR4*‡ ASiR is a CT iterative reconstruction
rate of up to 20% and noise reduction up to 33%, results in technique that enables a reduction in image
outstanding image quality in signal starved areas (e.g., shoulder, noise and an improvement in image quality.
hip, large patient and metal). ASiR may enable an improvement of up to
20% in low contrast detectability (measured
High Speed with IQ Enhance with statistical method) compared to non-
ASiR acquisition. ASiR may help clinicians
IQ Enhance (IQE) is an advanced algorithm designed to help achieve a more confident diagnosis with up
reduce artifact in thin-slice helical scanning. Use of IQE allows
to 40% lower dose while maintaining image
fast pitch scanning for larger anatomical coverage at the same
image quality compared to slower helical scanning without IQE. performance.
For example, IQE allows a 16-slice CT system to have coverage VISR Volumetric Image Space Reconstruction
speed that is equivalent to wide detectors (50-slice equivalent) at (VISR) is a 3D filter that reduces image
the same table speed. (Equivalent IQ means equivalent noise (standard deviation) without
helical artifact level.)
compromising spatial resolution to provide
clear visualization in neurological and cardiac
IQ Enhance
imaging, delivering diagnostic image quality
Coverage speed of a 50 slice CT with potentially lower mA.
Dose Check Prior to starting the scan, Dose Check can
Helical Pitch 0.562
notify and alert the operator that is setting
the scan parameters if the estimated dose
index is above user-defined notification
Helical Pitch 1.75 values. Dose check complies with the NEMA
XR-25-2010 standard.
Beam Tracking Beam Tracking includes techniques
that provide real-time X-ray follow-up,
Volume CT Technology Inside guaranteeing that the highest spatial
Innovative CT reconstruction algorithms, Hyperplane and resolution is reached with no post-patient
Crossbeam, providing artifact-free images and optimized slice collimation and no dose penalty.
profile at any pitch, by solving the technical challenges of cone
3D mA After selecting the desired noise/IQ, the
beam and high pitch helical scanning.
modulation CT then automatically tailors the exposure
0.35 mm isotropic microVoxel image resolution. parameters, patient-to-patient and in real-
time x-y-z during each scan, for an average of
0.31 mm Superior 40% dose reduction.
0.33 mm Organ Dose ODM provides reduction of radiation dose
Modulation (ODM) via x-ray tube mA modulation for superficial
0.36 mm tissues, such as breasts, etc. for an average
0.38 mm
of 40% dose reduction.
Overlapped Reconstruction‡**
The overlapped reconstruction optional feature enables 32 slices
per rotation in axial scanning modes and delivers improved Z-axis
visualization performance.
Ultra Kernel‡**
Adaptive Enhance Level Adjustment (AELA) can improve visual
spatial resolution while maintaining pixel noise standard
deviation and artifact. The Ultra Kernel feature Conventional reconstruction ASiR
is helpful in enhancing the visualization of small anatomical
structures with high contrast.
‡Optional
4
In clinical practice, the use of ASiR may reduce CT patient dose depending on the
Emergency Patient Mode **
clinical task, patient size, anatomical location and clinical practice. A consultation
The CT Optima CT540 series has a dedicated User Interface (UIF) with a radiologist and a physicist should be made to determine the appropriate
for emergency cases to start procedure immediately. Patient dose to obtain diagnostic image quality for the particular clinical task.
**This feature is not CE marked and cannot be placed on the market or put
Name and Patient ID are assigned automatically and once a
into service until it has been made to comply with CE marking or otherwise
protocol is selected, the system is ready to scan.
obtained regulatory authorizations.
CT Specifications
CT Performance HiLight Matrix II Detector
Display field-of-view 70 cm with WideView‡ Number of 21,888
Scan field-of-view 50 cm elements
Display matrix 1024 x 1024 CTDIw expressed Axial head: 17.0 mGy/100 mAs
in mGy/100 Axial body: 5.5 mGy/100 mAs
mAs on CTDI
X-ray Tube head and body Helical head5: 19.6 mGy/100 mAs
dose reference Helical body5: 9.9 mGy/100 mAs
Performix* Plus Designed for exams requiring a large phantoms
number of scans with minimal tube
cooling delays Noise Axial6: =< 0.35% at 17.6 mGy with ASiR4‡
(ASiR4‡ 40%) Axial6: =< 0.35% at 29.3 mGy
Anode heat capacity 6.3 MHU when ASiR4‡ is disabled
Use of ASiR4‡ allows scanning at a
lower mA and less tube heat output, Helical5: =< 0.35% at 17.1 mGy with ASiR4‡
resulting in the ability for longer (ASiR4‡ 40%)
duration helical scans similar to the Helical5: =< 0.35% at 28.5 mGy when ASiR4‡
capability of an 8 MHU tube. is disabled
‡Optional
4
In clinical practice, the use of ASiR may reduce CT patient dose depending on the clinical task, patient size, anatomical location and clinical practice. A consultation
with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task.
5
CDTI Helical calculated from Axial data and adjusted for 0.9375:1 pitch and 240 mAs. CTDIw is expressed in mGy/100 mAs for IEC pitch 1: normalized to a pitch of 1.
6
On either an AAPM water phantom or GE Quality Assurance phantom.
Discovery IQ Workspace
Quantitative SUV made easy
Innovative and ergonomic working environment designs to ease the operator's workflow in daily clinical and research studies. It features
powerful and expandable GE PET reconstruction technology that streamlines PET/CT workflow by handling massive PET/CT data sets with
ease. Everything you need at your fingertips--an intuitive quantitative workflow that fully integrates into your daily routine.
ViP replay Volume Interface Protocol (ViP) is an SnapShot Provides software and hardware to
intuitive and comprehensive PET list CardIQ‡ perform retrospective helical ECG-gated CT
mode prospective reconstruction reconstructions of the heart with SnapShot
solution. Provides integrated list mode imaging modes. CT cardiac acquisition
processing for generating a variety of enables scanning patients with heart rates
scan types (static, dynamic, gated) from a from 30 to 120 BPM with gating capabilities
single acquisition. to 200 BPM, and achieves temporal
resolution of up to 44 msec.
RadRx • Create your own PET/CT protocol from
any CT protocol.
• Automated CTAC creation from any Clinical Protocols ASiR‡ Q.AC‡
CT acquisition, including contrast
enhanced, perfusion and gated CTs. Oncology
• Integrated Average Cine CT protocol for • Whole-body PET/CT (Static) Yes Yes
improved attenuation correction. • Q.Static PET/CT Yes Yes
• Q.Freeze PET/CT (Integrated) Yes Yes
ACQC‡ Attenuation Correction Quality Control • 2 meter scan whole-body PET/CT Yes Yes
ensures proper cardiac registration • MotionMatch PET/CT Yes Yes
between PET and CT.
Cardiology
PET/CT pediatric Default protocols specifically designed • 82Rb stress/rest protocol Yes Yes
protocols for FDG oncology pediatric exams based • FDG Gated PET/CT Yes Yes
on Broselow-Luten™ Pediatric System,
which is designed to minimize CT Neurology
radiation exposure in PET/CT attenuation • Amyloid PET imaging Yes Yes
correction to meet ALARA guidelines. • Brain PET only Yes Yes
‡Optional
Discovery IQ Workspace (cont.)
PET/CT Workflow Solutions (Cont.) PET/CT Workflow Solutions (Cont.)
Xtream Injector‡ A powerful integrated injection option PET Data Storage 900 GB List Data Storage, 900 GB
that starts the injection process in Sinogram Data Storage. Storage space
synchronization with “Start Scan” is amplified by utilizing GE’s patented
to simplify the enhancement exam sinogram and list compression technology
workflow. The Enhanced Xtream Injector to shrink the data size while it resides on
supports injector parameters that are the scanner. Data is backed up in a RAID
entered on console. configuration to reduce the potential for
Direct MPR with Automatic real-time direct reconstruction lost data and system downtime in the
Auto-Batch and transfer of fully corrected multiplanar event of a storage disk failure.
images, in any plane. Up to 10 fps Data Export Ensures that the relevant images and
transfer speed of images, in real-time reports can be visualized by the referrals
during acquisition, to up to four different in a PC friendly format (MPEG, AVI, etc.)
destinations. Compatible • DVD-R: Creation of interchange media
Auto Transfer Distribute images where and when you peripherals of PET and CT image data with viewer
need them. option.
Exam Split Allows multi-anatomic exams to be read • USB 2.0 external hard drive: ViP data
in separate anatomic sections. This allows offline storage.
specialists to review only those images
needed for a given requisition.
Other
SmartPrep with Auto Enables intermittent monitoring of IV
Trigger contrast enhancement in an area of Acquisitions static, gated, dynamic
interest. The contrast flow is monitored multi-static, multi-gated, multi-dynamic
by low-dose scans until the contrast List mode capability
enhancement reaches the preferred Image matrix sizes 128 x 128, 192 x 192, 256 x 256
point and then the auto trigger function
Optional • Volume Viewer PET and CT
automatically initiates the scan
post-processing • CT Perfusion
prescription.
applications‡ • CardIQ Express
DICOM Structured Generates a CT Dose Report, which can • Autobone Express
Dose Report enable tracking of dose (CTDIvol and DLP) • Advanced Vessel Analysis
for the patient by a hospital's radiation • CardIQ Plus
tracking system/RIS/HIS. • Dynamic VUE
AWE connection The software based AW Server client • Motion VUE
on the CT console allows access to • CardIQ VUE
applications hosted on an AW Server,
right from the CT console for improved
workflow and productivity.
‡Optional
Siting requirements Misc.
PET/CT Gantry and System DICOM Conformance Standards
Description Height Width Depth Weight PET sinogram data can be saved as a DICOM file.
cm inch cm inch cm inch kg (Maximum size up to 210 MB)
Table 107 42 66 26 345 136 955 Service Class User (SCU) for image send
Gantry Depth including the back trailer is 265 cm. Service Class Provider (SCP) for image receive
DICOM query/retrieve service class
Patient Table
DICOM storage commitment class push
Maximum load 500 lbs (226 kg)
DICOM modality work list
Maximum horizontal scannable Standard 170 cm with up
range to full 200 cm‡ DICOM modality performed procedure step
Vertical scannable range 8 to 205 mm below isocenter DICOM print
Maximum horizontal speed 100 mm/sec
Filming Protocol
Patient Table Compatible‡
3M-952 standard
IV Pole integrated at the foot-end of the table prevents IV lines
Supported film and hardcopy devices:
from becoming crossed and tangled, and ensures that the lines
stay securely in place on the patient. • A DICOM print interface is standard on the system
• DICOM basic grayscale and color print via ethernet
2 m table extender: Carbon fiber table extender to enable full 2 m
• Sterling/AGFA Helios via ethernet
scan range for complete head to toe studies.
• Post script level 2 printer via TCP/IP ethernet
Radiation treatment planning flat table overlay securely lock into
the PET/CT.
HIPAA
GE integration kit for Varian respiratory patient monitor
Password protected user login and authentification
composed of a table-mounting bracket with connector to support
the RPM device; allows respiratory gating with and without Image annonimization tool
radiation treatment planning flat table overlay. Product network filters restricts access to scanner system by IP
address, services type (IE ftp, telnet) and DICOM port number
Siting Requirements
User configurable
Scan room
4267 mm x 7772 mm
Minimum scan room size Standard, Selectable Items
13 ft, 4 in x 25 ft, 6 in
Scan room temperature range 18° C (64º F) – 26° C (79º F) Language selectable keyboard ConnectPro HIS/RIS interface with
performed procedure step
Scan room cooling Air cooled, recommended 37,000
requirements BTU/hr with heavy CT utilization DICOM storage service class
Control room
Regulatory Compliance
2438 mm x 4268 mm
Minimum control room size This product is designed to comply with applicable standards under
9 ft, 0 in x 13ft, 4 in
the Radiation Control for Health and Safety Act of 1968.
Control room temperature range 18° C (64º F) – 26° C (79º F)
Laser alignment devices contained within this product are
Control room cooling Air cooled
appropriately labeled according to the requirements of the Center
requirements Recommended 3,200 BTU/HR
for Devices and Radiological Health.
Power requirements 90 kVA maximum 30 kVA average
Discovery IQ has been designed to follow PET/CT international
Equipment room standards: ACR, ACRIN, EARL, QIBA
Not required. Heat and noise
output for Power Distribution
Unit and computer hardware are
Equipment Room
low enough to allow inclusion in
the patient scanning suite with
the table and gantry.
‡Optional
©2014 General Electric Company-All rights reserved.
General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue any products
described at any time without notice or obligation. Please contact your GE representative for the most current information.
GE Healthcare GE, GE Monogram and imagination at work are trademarks of General Electric Company.
3000 North Grandview Blvd All third party trademarks are property of their respective owners.
Waukesha, WI 53188 *Trademark of General Electric Company.
USA GE Healthcare, a division of General Electric Company.
www.gehealthcare.com DOC1779283