Professional Documents
Culture Documents
Nordtest Uncertainty Example Acrylamide Lc-Ms Solved 0
Nordtest Uncertainty Example Acrylamide Lc-Ms Solved 0
Reproducibility, u(Rw)
From repeated analysis of one of the CRMs
u(Rw) = s_Rw = 31 μg/kg
u(Rw)_rel = 2.70 %
Bias, u(bias)
From measurements with 2 different CRMs
u(Cref)_rel= 2.67 %
RMSbias_rel= 2.87 %
u(bias)_rel = 3.92 %
Result:
uc_rel = 4.8 %
uc = 48 μg/kg
U_rel (k=2) = 9.5 %
U (k=2) = 95 μg/kg