Philips Medical Systems

FIELD CHANGE ORDER Service


Issued by : PMSN Best Reference No. : FCO781 00083
Date : May 2004
Publication No. : 4522 981 16991 Product Group : 781

Intera Achieva
RF coils

APPLIES TO:
First delivered Intera Achieva systems

Trace items:
Trace item New trace Name trace item
item
989603011662 989603011663 Synergy spine coil

989603011612 989603011613 Synergy Body coil

989603011632 989603011633 Synergy shoulder flex-M

989603011652 989603011653 Synergy H/N coils

989603011622 989603011623 Synergy Flex-L

989603011642 989603011643 Synergy Flex-S

989603011691 989603011692 3T QHC

989603012062 989603012063 Synergy cardiac coil

989603011281 989603011281 8 channel Sense Head coil 1.5T

989603011601 989603011601 8 channel Sense Knee coil 1.5T

The System Reference Numbers (SRN) of the affected systems are:


11009, 11010, 11012, 12017, 13005, 13006, 17000, 17002

Link to: Allocation list

TITLE:
ODU connector RF coil
FCO781 00083 (04.0) 1
(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
LIST OF PAGES & DRAWINGS:
FCO781 00083 1 (04.0)
2 (04.0)
3 (04.0)
4 (04.0)
5 (04.0)
6 (04.0)
7 (04.0)
8 (04.0)
9 (04.0)
10 (04.0)
ANR (04.0)

INTRODUCTION:
Symptom : Low reliability of the ODU connector on the RF coils
Cause : Too high connector chassis tolerances
Remedy : Replace chassis and keyings of ODU connectors on RF coils

This FCO must be executed together with FCO781 00081 and FCO781 00082

MANPOWER / TIME TO COMPLETE:


30 minutes per coil for 1 engineer (15 min. modification + 15 min. testing).

TOOLS & TEST EQUIPMENT:


Only the standard tool kit is required.

MODIFICATION KIT / PARTS REQUIRED: 4522 132 23141 RETROFIT KIT COIL CONNECTOR
PHILIPS
Kit contents
12NC Description Qty
4522 132 12131 Set labels retrofit PICU 1
4522 132 22991 Retrofit kit coil connector containing: 1
4522 090 12321 - Keying module female 3
4522 090 12191 - Chassis 1
4522 090 12341 - Dummy coax contacts 15
4522 981 16991 This FCO 1

Ordering Info: Please investigate the required quantity of modification kits and supply a purchase
order at SLI Best.

2 (04.0) FCO781 00083


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
PROCEDURE:
1. Remove the three screws (figure 7: item 6 and 7) from the
coil connector. Figure 1 – ODU Coil connector

Remove screws

2. Lift the chassis (figure 7:item 2 and 3) until you have


access to the two screws (figure 7:item 5) that hold the Figure 2 – ODU Coil connector
chassis together.
Lift chassis

Remove screws
3. Carefully remove the chassis part (figure 7: item 2) from
the connector modules. Do not change the order of the Figure 3 – ODU Coil connector
connector modules and do not rotate the modules.
Remove chassis

Connector module

FCO781 00083 (04.0) 3


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
4. The connector is now disassembled. Note that not all
connector modules have wires connected to them. The Figure 4 – ODU Coil connector
three keying strips are also removed.
Keying strips
Chassis
CAUTION

Be very carefull not to break a wire


________________

5. Some connector coax blocks do not contain coax pins.


Insert dummy coax contacts into these empty positions. Figure 5 – ODU Coil connector

All these coax


pins must be
present. If
not, fill with
dummy coax
pins

Push the RF pin into the connector module (Figure 6) untill


you hear a click. Figure 6 - Insert RF pin

Connector module RF pin

Insert RF pin

6. Mount the new supplied chassis and the new keying strips
in the reverse order.
Mount the connector modules and keying strips in the
correct order. For the correct order, select the table (from
table 1 to table 7) that corresponds with the 12NC code of
your adapter.
7. Remove the 12NC label from the adaptor and glue the
supplied label with the same 11NC and increased last digit.
Refer to table 8 for an overview of labels.

4 (04.0) FCO781 00083


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Figure 7 – Machanical layout of coil connector

Figure 8 – Old and new keying Figure 9 – Old and new frame

OLD FRAME
OLD KEYING NEW KEYING

One dot

Two dots
NEW FRAME

FCO781 00083 (04.0) 5


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
8. The tables 1 to 7 show the connector modules and the Figure 10 – Facing the coil connector
wiring attached at the inside. For the viewing direction
see the arrow in Figure 10.

Table 1 – Layout of connector for coil Synergy Spine 9896 030 1166x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2 C1 pin 1..4 Coax 01 Coax 02 coax 03 coax 04
3
4
5 C2 pin 1..4 spare spare spare spare
6
si 7 k_b1
K1
de
8 pin 1..7 RD BK ORG WH ORG BK spare
of
9 P1
UL
10 pin 8..14 RD BK spare spare spare BK spare
-
m 11 K2 k_b2
ar 12 S1 pin 1..10 spare spare BK WH/BU WH/GN spare spare WH/YE WH/VT WH/GRAY
ki 13 S2 pin 1..10 spare spare spare spare spare spare spare WH/BK spare spare
14 K3 k_b3
15
16 C3 pin 1..4 spare spare spare spare
17
18
19 C4 pin 1..4 coax 05 spare spare spare
20

Note that pin 1 is on the left side

Table 2 – Layout of connector for Synergy Body Coil 9896 030 1161x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2 C1 pin 1..4 Coax 01 Coax 02 coax 03 coax 04
3
4
5 C2 pin 1..4 spare spare spare spare
6
7 K1 k_b1
side of UL-marking

8 pin 1..7 RD BK ORG WH ORG BK spare


9 P1
10 pin 8..14 RD BK spare spare spare BK spare
11 K2 k_b2
12 S1 pin 1..10 spare spare BK WH/BU WH/GN spare spare WH/YE WH/VT WH/GREY
13 S2 pin 1..10 spare spare spare spare spare spare spare WH/BK spare spare
14 K3 k_b3
15
16 C3 pin 1..4 spare spare spare spare
17
18
19 C4 pin 1..4 spare spare spare spare
20

6 (04.0) FCO781 00083


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Table 3 – Layout of connector for Synergy Body Coil 9896 030 1163x
Synergy Flex L 9896 030 1162x
Synergy Flex S 9896 030 1164x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2 C1 pin 1..4 Coax 01 Coax 02 spare spare
3
4
5 C2 pin 1..4 spare spare spare spare
6
7 K1 k_b1
side of UL-marking

8 pin 1..7 RD BK ORG WH ORG BK spare


9 P1
10 pin 8..14 RD BK spare spare spare BK spare
11 K2 k_b2
12 S1 pin 1..10 spare spare BK WH/BU WH/GRN spare spare WH/YE WH/VT WH/GREY
13 S2 pin 1..10 spare spare spare spare spare spare spare WH/BK spare spare
14 K3 k_b3
15
16 C3 pin 1..4 spare spare spare spare
17
18
19 C4 pin 1..4 spare spare spare spare
20

Table 4 – Layout of connector for Synergy Body Coil 9896 030 1165x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2 C1 pin 1..4 Coax 01 Coax 02 Coax 03 spare
3
4
5 C2 pin 1..4 spare spare spare spare
6
7 K1 k_b1
side of UL-marking

8 pin 1..7 RD BK ORG WH ORG BK spare


9 P1
10 pin 8..14 RD BK spare spare spare BK spare
11 K2 k_b2
12 S1 pin 1..10 spare spare BK WH/BU WH/GRN spare spare WH/YE WH/VT WH/GREY
13 S2 pin 1..10 spare spare spare spare spare spare spare WH/BK spare spare
14 K3 k_b3
15
16 C3 pin 1..4 spare spare spare spare
17
18
19 C4 pin 1..4 spare spare spare spare
20

Table 5 – Layout of connector for 3T Quad Head Coil 9896 030 1129x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2
3 PWR1 pin 1..2 coax spare
4
5
6
7 C1 pin 1..4 spare spare spare spare
side of UL-marking

8
9 K1 k_b1
10 pin 1..7 BN BU-VT OG RD OG BU-VT spare
11 P1
12 pin 8..14 BN BU-VT spare spare spare BU-VT spare
13 K2 k_b2
14 S1 pin 1..10 spare spare BU-VT YE spare spare spare BU-VT BK WH/RD
15 S2 pin 1..10 spare spare spare spare spare spare spare BU-VT spare spare
16 S3 pin 1..10 spare spare spare spare spare spare spare spare spare spare
17 K3 k_b3
18
19 C2 pin 1..4 Coax spare Coax spare
20

FCO781 00083 (04.0) 7


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
Table 6 – Layout of connector for Synergy Cardiac coil 9896 030 1206x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2 C1 pin 1..4 Coax 01 Coax 02 Coax 03 Coax 04
3
4
5 C2 pin 1..4 spare spare spare spare
6
7 K1 k_b1
side of UL-marking

8 pin 1..7 spare spare spare WH spare spare spare


9 P1
10 pin 8..14 spare spare spare spare spare spare spare
11 K2 k_b2
12 S1 pin 1..10 spare spare spare WH/BU WH/GRN spare spare WH/YE WH/VT spare
13 S2 pin 1..10 spare spare spare spare spare spare spare WH/BK spare spare
14 K3 k_b3
15
16 C3 pin 1..4 spare spare spare spare
17
18
19 C4 pin 1..4 Coax 05 spare spare spare
20

Table 7 – Layout of connector for 8 ch Sense Knee Coil 1.5T 9896 030 1160x
8 ch Sense Head Coil 1.5T 9896 030 1128x
Module 1 2 3 4 5 6 7 8 9 10

Unit
1
2 C1 pin 1..4 Coax 01 Coax 02 Coax 03 Coax 04
3
4
5 C2 pin 1..4 spare spare spare spare
6
7 K1 k_b1
side of UL-marking

8 pin 1..7 -5VB GND +12V +300V +12V GND spare


9 P1
10 pin 8..14 -5VB GND spare curr spare GND spare
11 K2 k_b2
12 S1 pin 1..10 spare spare screen sel-A sel-B spare spare Ref-GND Tune/ID/M alf ID2-sel
13 S2 pin 1..10 spare spare spare spare spare spare spare SY-status spare spare
14 K3 k_b3
15
16 C3 pin 1..4 spare spare spare spare
17
18
19 C4 pin 1..4 Coax 05 Coax 06 Coax 07 Coax 08
20

Table 8 – Overview of labels


New trace Name trace item New label 1 Name label 1 New label 2 Name label 2
item
989603011663 Synergy spine coil 452213219873 Syn.assy pc 452213231223 Syn.cable+trap

989603011613 Synergy Body coil 452213219853 Syn.body coil 452213231013 driverbox

989603011633 Synergy shoulder flex-M 452213231173 Syn.flex-M coil 452213231183 Syn.FL-M driverbox

989603011653 Synergy H/N coils 452213219863 Syn.head/neck coil 452213231133 Driverbox SHNC

989603011623 Synergy Flex-L 452213231153 Sense flex-L coil 452213231183 Syn.FL-L driverbix

989603011643 Synergy Flex-S 452213231193 Syn flex-S coil 452213231203 Syn.FL-S driverbox

989603011692 3T QHC 452213219832 3T quad head coil T/R - -

989603012063 Synergy cardiac coil 452213219883 Syn.card.coil (CZ) 452213231253 Onderbak cardiac

989603011281 8 ch. Sense Head coil 1.5T

989603011601 8 ch. Sense Knee coil 1.5T

8 (04.0) FCO781 00083


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
PARTS DISPOSAL:

Old chassis part and keying.

DOCUMENTATION:

Log this action in the section "History Record" of the System Reference Manual.
File this FCO in the section "Service Information" of the System Reference Manual.
Fill out the attached Action Notification Report and send it to your SSD Customer Services Manager.

FCO781 00083 (04.0) 9


(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
10 (04.0) FCO781 00083
(c) 2004 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED
FCO ACTION NOTIFICATION REPORT
For local SSD use only; do not return to PMG.
TITLE : ODU connector RF coil

CLASSIFICATION : Proactive Action for Performance FCO REF. NO.: FCO781 00083

APPLIES TO : ODU connector RF coil

HOSPITAL / ADDRESS :

LOCATION / FW SITE NO.: SALES ORDER NO. / OA NO.:

PRODUCT NUMBER :

UNIT SERIAL NUMBER :

ACTION ON THIS UNIT WAS: (select one) JOB NO. / SERVICE INCIDENT NO.:

Completed per instruction on


DATE

Completed by the factory prior to delivery.

Not completed as this unit is not affected per instruction because: (state reason)

Not completed because customer has unit in storage.


Required parts & instructions received by the customer.

CUSTOMER ACKNOWLEDGEMENT (Required for MANDATORY ACTIONS only).


The REASON and PURPOSE of this modification have been explained to me.

CUSTOMER NAME (PLEASE PRINT) TITLE

CUSTOMER SIGNATURE DATE

BRANCH
REGION / SERVICE UNIT /
DEALER : SERVICE AREA NO.:

SIGNATURE CUSTOMER SERVICES ENGINEER DATE

SIGNATURE CUSTOMER SERVICES MANAGER MAIL TO : SSD Customer Services Manager

FCO781 00026 (03.0) ANR


(c) 2003 Koninklijke Philips Electronics N.V.
ALL RIGHTS RESERVED