Professional Documents
Culture Documents
User Manual
User Manual
User Manual
Experiment Platform
Operating Guide
ME1758/B
Contents
Chapters Pages
Appendices Pages
Introduction
This User Manual covers the following type of D3000 Base Unit:
In addition to providing an overview of the D3000 system, the following topics are
covered:
i
Operating Guide D3000 ExP
Introduction
ii
D3000 ExP Operating Guide
Overview of the D3000 System
Chapter 1
Overview of the D3000 System
Each study module provides you with a printed circuit board (PCB)
containing electrical components that allow you to practically
investigate a particular topic area. Also included is a Laboratory
Manual, which contains theory and practical instruction, and a plastic
storage case.
1.1
Operating Guide D3000 ExP
Overview of the D3000 System
Each Study Module has a name and a module number. The module
number is especially important, as it is used in a computer managed
laboratory to identify the module that you are studying.
is 0.11.
In order to use a printed circuit board from the range of D3000 Study
Modules, you will need a D3000 ExP at your workstation. The ExP
provides:
1.2
D3000 ExP Operating Guide
Overview of the D3000 System
Operating Modes
1.3
Operating Guide D3000 ExP
Overview of the D3000 System
1.4
D3000 ExP Operating Guide
Overview of the D3000 System
The Base Unit Accessory Kit contains 2mm connecting leads and
shorting links which are used to build circuits on any circuit board that
is fitted to the D3000 ExP.
1.5
Operating Guide D3000 ExP
Overview of the D3000 System
1.6
D3000 ExP Operating Guide
Switching On the D3000 ExP
Chapter 2
Switching On the D3000 ExP
AC Power Line
The ExP uses the AC Line supply. Power enters via a three-pin IEC
inlet on the rear panel of the unit. The main power switch is adjacent
to the inlet.
Switching On
• All the front-panel LEDs and displays should light for a second to
show that the unit is working.
Switching Off
• Switch off the ExP by setting the rear-panel main power switch to
the 0 position.
2.1
Operating Guide D3000 ExP
Switching On the D3000 ExP
2.2
D3000 ExP Operating Guide
The Module Power Supplies
Chapter 3
The Module Power Supplies
Module Supplies
The D3000 ExP contains all the power supplies that are required for
powering any of the Experiment Boards in the D3000 range. These
are as follows:
• +5V 1.5A DC
• -5V 0.1A DC
• +12V 1A DC
• -12V 1A DC
• 0V to +12.7V 1A Variable DC
• 12-0-12V 0.5A 50/60 Hz AC
Note that one of the power rails provided by the ExP is a Variable DC
supply that has an output voltage adjustable between 0V and +12.7V.
Instructions will be given in the Laboratory Manual or on-screen
learning materials whenever adjustment of this control is required.
Operation
3.1
Operating Guide D3000 ExP
The Module Power Supplies
Fig 3.1
Note:
It is important not to confuse operation of the Module Supplies with
operation of the main power switch on the rear panel of the ExP.
The main power switch should not be switched off until you are told
to do so by your instructor.
3.2
D3000 ExP Operating Guide
Loading a Circuit Board
Chapter 4
Loading a Circuit Board
• Select the D3000 Study Module that is to be used, and remove the
circuit board from its plastic storage case.
• Place the bottom of the circuit board onto the two spring-loaded
guides just above the control panel.
Fig 4.1
4.1
Operating Guide D3000 ExP
Loading a Circuit Board
• Press down on the board and ease its black guide rails under the
pegs on the ExP side panels.
Fig 4.2
Fig 4.3
4.2
D3000 ExP Operating Guide
Loading a Circuit Board
• When you are sure the circuit board is properly seated, push the
lever on the latching mechanism fully away from you. The circuit
board connector should engage cleanly with the one inside the
latching mechanism as the board is pulled up into its operating
position.
Fig 4.4
4.3
Operating Guide D3000 ExP
Loading a Circuit Board
4.4
D3000 ExP Operating Guide
Unloading a Circuit Board
Chapter 5
Unloading a Circuit Board
Fig 5.1
5.1
Operating Guide D3000 ExP
Unloading a Circuit Board
• Press down on the board and ease it towards you until it is clear
of the pegs on the ExP side panels.
Fig 5.2
• Lift the circuit board off the spring-loaded guides and return it to
its plastic storage case.
Fig 5.3
5.2
D3000 ExP Operating Guide
Using a Multimeter
Chapter 6
Using a Multimeter
Multimeters
The term multimeter derives from the ability to use one instrument for
a multitude of different measurements. One instrument is capable of
taking measurements of all three of the above quantities, and switches
are provided for a wide range of values of each quantity, from the very
small (µ - micro or m - milli) to the large (k - kilo or M - Mega). Also
both direct current and voltage (DC) and alternating current and
voltage (AC) measurements can be taken with the same instrument.
6.1
Operating Guide D3000 ExP
Using a Multimeter
Types of Multimeter
There are two basic types of instrument, those that give a digital
display of the reading, and those in which a pointer is moved across a
scale by an angle that is analogous to the quantity being measured.
POWER
OFF ON
V V
200 1000 750 200
20 20
2 2
200m 200m
20M 200µ
2M 2m
20m A
200K 10A
200m
20K
2
Ω 2K
200 2
200m
200
2m
200µ 20m
10A
A
Fig. 6.1
6.2
D3000 ExP Operating Guide
Using a Multimeter
Fig 6.2
If the selected range is 100mA then the 0-10 scale is used and the
pointer is half way between 6 & 8. The scale reading gives us 7. The
scale factor is determined by dividing the full-scale marked value into
the range value, 100mA ÷ 10 = 10mA. Multiply the reading by this
factor: 7 x 10mA = 70mA.
6.3
Operating Guide D3000 ExP
Using a Multimeter
If the selected range had been 50µA then the 0-50 scale should be
used and the pointer is half way between 30 and 40. The scale reading
gives us 35. The scale factor is 35 x 1µA = 35µA.
6.4
D3000 ExP Operating Guide
Using a Multimeter
Measuring Voltage
The meter connections for voltage measurement are shown below.
POWER
OFF ON
V V
200 1000 750 200
20 20
2 2
200m 200m
20M 200µ
2M 2m
20m A
(a) Ω
200K
20K
2
10A
200m
2K
200 2
200m
200
2m
200µ 20m
10A
A
POWER
OFF ON
V V
200 1000 750 200
20 20
2 2
200m 200m
20M 200µ
2m
(b)
2M
20m A
200K 10A
200m
20K
2
Ω 2K
200 2
200m
200
2m
200µ 20m
10A
A
The voltage appears across the component. Therefore the meter must
be connected in parallel with (or across) the component to measure
the volt drop across it with the circuit still connected to the supply.
6.5
Operating Guide D3000 ExP
Using a Multimeter
6.6
D3000 ExP Operating Guide
Using a Multimeter
Measuring Current
The meter connections for current measurement are shown below.
PO W ER
O FF ON
V V
200 1000 7 50 200
20 20
2 2
200 m 20 0m
20M 200 µ
2M 2m
20m A
2 00K 10A
200 m
(a)
20K
2
Ω 2K
20 0 2
2 00m
200
2m
200µ 20m
10A
A
V-Ω CO M A 10A
PO W ER
O FF ON
V V
2 00 1000 7 50 200
20 20
2 2
200m 200m
20M 2 00µ
2M 2m
20 m A
(b)
200 K 1 0A
2 00m
20K
2
Ω 2K
200 2
200m
2 00
2m
2 00µ 20m
10A
A
V -Ω C OM A 1 0A
6.7
Operating Guide D3000 ExP
Using a Multimeter
Measuring Resistance
The meter connections for resistance measurement are shown below.
POWER
OFF ON
V V
200 1000 750 200
20 20
2 2
200m 200m
20M 200µ
2M 2m
20m A
200K 10A
200m
20K
2
Ω 2K
200 2
200m
200
2m
200µ 20m
10A
A
6.8
D3000 ExP Operating Guide
Using a Multimeter
Testing Continuity
When troubleshooting electronic circuits, it is often necessary to
detect a short circuit fault condition. This is a very low resistance
(typically less than 5Ω) that is usually caused by a conductor bridging
two connections in a circuit and causing a fault.
6.9
Operating Guide D3000 ExP
Using a Multimeter
Testing a Diode
A diode can be tested using the diode tester on a digital multimeter.
Examine the markings around your multimeter range switch. The
diode tester range is indicated by a diode symbol. It may share a
switch position with another meter range (such as a resistance range),
or it may have its own, separate switch position.
To test the diode in the forward direction connect the positive (red)
lead to the anode, and the common (black) lead to the cathode, as
shown in Fig 6.6.
With the diode tester range selected, the meter will display a value for
the forward-biased diode. How you interpret this display will depend
on the model of digital multimeter that you are using.
6.10
D3000 ExP Operating Guide
Using a Multimeter
If your meter has a separate diode tester range, this reading will
usually indicate the forward voltage across the diode in millivolts
(mV). A typical measurement for a forward-biased diode would be
600-700mV.
Whichever type of diode tester you have, you will need to test a
working diode to find out the value and units of the reading you can
expect from your meter.
The diode can also be tested in the reverse direction by connecting the
positive (red) meter lead to the anode and the common (black) lead to
the cathode.
With the diode tester range selected, the meter should give an out-of-
range indication. This will be the case whether the diode tester display
is in mV or kΩ.
The method described above for testing a diode can also be used to
test the PN junctions of a transistor. A typical forward bias
measurement for a transistor would be 700-800mV or a few kΩ. The
reverse bias will again be indicated by an out-of-range measurement.
6.11
Operating Guide D3000 ExP
Using a Multimeter
Note.
For use on a D3000 circuit board, 4mm to 2mm leads have
been provided in your D3000 Base Unit Accessory kit to allow
the connection of a Testmeter to your Base Unit.
6.12
D3000 ExP Operating Guide
Using an Oscilloscope
Chapter 7
Using an Oscilloscope
How it Works
Your understanding of the operation of this most valuable item of test
equipment will be greatly enhanced if you have at least a basic
knowledge of its fundamentals.
glass envelope
screen
cathode focus Y plates
grid X plates
electron beam
heater
Fig 7.1
The heated cathode emits electrons that are attracted by the high
potentials on succeeding electrodes to form a divergent electron
stream or beam.
The electric field of the focus assembly accelerates the electrons in the
beam and converges them so that they all meet at one spot at the
screen.
7.1
Operating Guide D3000 ExP
Using an Oscilloscope
The grid, which surrounds the cathode, allows control of the number
of electrons leaving the cathode, and therefore the strength of the
electron beam, and the intensity or brightness of the spot. The group
of electrodes which generate the beam are known collectively as the
electron gun.
voltage
stroke or flyback
scan
time
Fig 7.2
7.2
D3000 ExP Operating Guide
Using an Oscilloscope
7.3
Operating Guide D3000 ExP
Using an Oscilloscope
Practical Oscilloscope
It is now time to examine the layout of the front panel of a typical
oscilloscope and its controls. These may seem a little awe-inspiring at
first, but you will find that you can easily master them.
Fig 7.3
7.4
D3000 ExP Operating Guide
Using an Oscilloscope
7.5
Operating Guide D3000 ExP
Using an Oscilloscope
1 3
2 1
.5 .2
.1
VAR.
1
.5 .2
.1 2
2.5:1
CH.I 2 50 2 50 CH.II
DC DC
5 20 CAL. 5 20
AC AC
GD GD
10 10 10 10
1MΩ 30pF V 5 mV V 5 mV 1MΩ 30pF
20 20
400Vp-p 400Vp-p
max. INV.I CHI/II DUAL ADD max.
HOR.
TRIG.I/II CHOP INP.
4 4
Fig 7.4
Note that the lower panel in Fig 7.4 above contains the controls for
two Y amplifiers. There is provision to operate the oscilloscope with
either one or two traces (graphs) so that two waveforms of the same
frequency (or harmonically related) can be observed at the same time.
7.6
D3000 ExP Operating Guide
Using an Oscilloscope
The upper panel contains the controls for the screen and for the
timebase settings. You will also see some controls marked TRIG or
TRIGGER. These are to maintain a stable trace. More will be said
about this function later.
7.7
Operating Guide D3000 ExP
Using an Oscilloscope
Operation
• Locate the power switch ( 1 in Fig 7.5 below) and switch ON.
1 4
2 20
50
5
2
INTENS. 100 1
200 .5 TRIG. INP.
FOCUS AC 100Vp-p max.
TRIGGER DC ms µs
HF
SELECTOR LF CAL.
EXT.
LINE
Fig 7.5
After a brief warm-up period you will find that you have a line across
the screen caused by the spot moving from left to right across the
screen under the influence of the internal timebase.
If you have a watch or clock available with a second hand, time how
long it takes for - say - five passes across the screen. You should find
that it takes about ten seconds for five scans.
7.8
D3000 ExP Operating Guide
Using an Oscilloscope
Timebase
Examine the timebase control switch. This is pointing at 200ms/div.
There are ten divisions across the screen. Count them. So it takes 10 x
200ms for one scan. 2000ms is 2 seconds, so 5 x 2 = 10 seconds for
five scans.
• Return it counter-clockwise.
Look to the left of the tip of the pointer and you will see a C (for
calibrated) under a dot. There is one of these symbols to the left of
each of the variable controls, including the two on the lower panel, to
indicate the calibration position.
• Switch the timebase selector to 100ms/div. Note that the spot now
travels across the screen in about one second. Gradually increase
the speed.
When you get to 20ms/div the spot has become a short line. This is
due to two factors, one being the afterglow of the phosphor (which
takes a small time to die away) and the other is the persistence of
vision (where our eyes retains an image for a small period of time).
This latter is what makes it possible for us to see apparently moving
pictures on a television screen from a rapid sequence of still pictures.
7.9
Operating Guide D3000 ExP
Using an Oscilloscope
Frequency Measurement
Please note that if it takes 5µs (millionths of a second) for one trace
and the traces follow each other continuously then there will be
200,000 scans in one second (200,000 x 5µs = 1s), the frequency is
200kHz.
This concept is the one above all others that newcomers to electronics
find most difficult to accept, the speed at which electronic devices can
operate is far, far faster than our brains want to accept.
The reciprocal of the time taken for one cycle of events is the
frequency of that event. This is important and should be remembered.
1
frequency =
time period
This allows us to make measurements of frequency on an oscilloscope
by noting the time taken for one cycle and then calculating the
reciprocal of that time.
6.4 div
Fig 7.6
7.10
D3000 ExP Operating Guide
Using an Oscilloscope
Fig 7.7
You should have arrived at about 5.95kHz. The reading of the time
scale cannot be very accurate, certainly not to 5 parts in 600, so it
might be better to call this 6kHz.
7.11
Operating Guide D3000 ExP
Using an Oscilloscope
Y Amplifiers
Turn your attention now to some of the controls on the lower panel,
the Y amplifiers.
1 1
.5 .2
.1
VAR.
1
.5 .2
.1 3
2.5:1
CH.I 2 50 2 50 CH.II
DC DC
5 20 CAL. 5 20
AC AC
GD GD
10 10 10 10
1MΩ 30pF V 20 5 mV V 20 5 mV 1MΩ 30pF
400Vp-p 400Vp-p
max. INV.I CHI/II DUAL ADD max.
HOR.
TRIG.I/II CHOP INP.
2 4
Fig 7.8
The effect is that you are applying a signal to the Y plates, only
relatively very slowly. Electronics can do it much faster. Do not try to
rotate the knob too quickly or you may damage the track of the
control.
7.12
D3000 ExP Operating Guide
Using an Oscilloscope
A second trace will now have appeared on the screen (you may have
to adjust the Y2 shift control to move it onto the screen).
• Move the new trace down to the lower half of the screen with the
Y2 shift control 3 .
You will see that the oscilloscope draws the Y1 and Y2 traces
alternately.
This is the simplest form of dual mode operation, but is not very
satisfactory for low frequency signal inputs. You would have great
difficulty in comparing waveforms on the two traces.
What is happening is that the circuit chops between the two traces
many times during one scan, so quickly that you cannot see it doing it.
This is the best mode of operation for timebase speeds below 2ms/div.
You will see that operating the ALT/CHOP switch has little effect at
timebase speeds of 2ms/div and above, but the difference is easily
observed at 5ms/div and below.
7.13
Operating Guide D3000 ExP
Using an Oscilloscope
Voltage Measurements
2 2
Y-POS. I VOLTS/DIV. VOLTS/DIV. Y-POS. II
.5 .2 .5 .2
1 1 .1
VAR.
2.5:1
1 .1
1
CH.I 2 50 2 50 CH.II
DC DC
5 20 CAL. 5 20
AC AC
GD GD
10 10 10 10
1M Ω 30pF V 5 mV V 5 mV 1MΩ 30pF
20 20
400Vp-p 400Vp-p
HOR.
3
TRIG.I/II CHOP INP.
Fig 7.9
X10
2V
TESTER Note: The ground clip
is not needed since this
is completed internally.
Fig 7.10
7.14
D3000 ExP Operating Guide
Using an Oscilloscope
You will now have a square wave displayed on the upper trace. The
vertical edges of the waveform are so fast that they do not have time to
leave any evidence of their presence. It appears as though the change
from negative to positive is instantaneous. Increasing the brightness to
maximum may just show them very faintly.
Fig 7.11
• Press the INVERT 1 button and observe that the CH.1 display is
inverted, the CH.2 trace remaining unaffected.
7.15
Operating Guide D3000 ExP
Using an Oscilloscope
AC/DC Operation
7.16
D3000 ExP Operating Guide
Using an Oscilloscope
T = Time taken
T for one cycle.
Fig 7.12
7.17
Operating Guide D3000 ExP
Using an Oscilloscope
Trigger
You can see the waveform, but it is not stable. This is because the
trigger or synchronizing facility is automatically allocated to the CH.1
signal until you say otherwise.
Fig 7.13
7.18
D3000 ExP Operating Guide
Using an Oscilloscope
You will probably find two different settings within the range of the
control that will give you either one complete cycle or one and a half.
You find that there are very nearly two complete cycles when the
control is in the properly calibrated position. As the fine timebase
control is adjusted when the waveform is locked, all that happens is
that the waveform is stretched or contracted to display more or less
cycles. Note, however, that the trace always starts with the positive-
going edge of the waveform.
Fig 7.14
7.19
Operating Guide D3000 ExP
Using an Oscilloscope
This is the trigger point, at the zero crossing of the test waveform (in a
positive-going direction).
The timebase in the oscilloscope is held off until this point is reached
and then allowed to run. In this way the displayed waveform always
starts at the same point (crossing zero in a positive-going direction) so
each successive trace overlays the previous one and the display
appears stationary.
There are several features on the timebase panel which affect the
triggering.
7.20
D3000 ExP Operating Guide
Using an Oscilloscope
Triggering
3 1
INTENS.
50
100 1
2
2
200 .5 TRIG. INP.
FOCUS AC 100Vp-p max.
TRIGGER DC ms µs
HF
SELECTOR LF EXT.
CAL.
LINE
4 5 6
Fig 7.15
The effect of this is to change the starting point voltage so that the
display starts at any point you choose on the waveform. If you set the
level higher or lower than the extremities of the test waveform then
the timebase never triggers and there is no display, the screen remains
blank. With the level button pointing vertically upwards the trigger
point is the zero voltage crossing level.
You cannot see the effect of this control if you only have the
calibration waveform available. The square wave has only two levels,
ON or OFF. However, if you have a signal source with sine or
triangular waveform then connect this to one of the Y channel inputs,
adjust for a good display using timebase (X) and sensitivity (Y)
controls, then press the AT/NORM. button and adjust the level
7.21
Operating Guide D3000 ExP
Using an Oscilloscope
control. Observe the effect and then return the AT/NORM. button to
the out position.
The +/- button 3 inverts the display by selecting the zero crossing
trigger point when the waveform is negative going instead of positive.
• With any waveform displayed and locked, press the +/- button
and observe. Return to the out position.
For some displays the trigger circuits may need a little help in the
form of selecting the frequency. The calibration waveform is a middle
frequency and any setting of the TRIGGER SELECTOR 4 except
LINE will provide a stable display. The settings of this selector are:
7.22
D3000 ExP Operating Guide
Using an Oscilloscope
picture
information
line line
sync. sync.
pulse pulse
• Press the EXT. TRIG. button and note that the display is no
longer locked.
• Take the probe from the CH.2 input and plug it into the EXT.
TRIG. input socket 6 . Couple this to the cal. signal.
Note that the display is again locked and that all of the other triggering
functions can be selected with this input.
7.23
Operating Guide D3000 ExP
Using an Oscilloscope
Component Tester
Many oscilloscopes are provided with this most valuable facility,
which enables the instant display of the characteristics of many
electronic components. An alternating voltage is applied to the
component under test and also to the X plates of the oscilloscope. The
current drawn flows in a series resistor mounted inside the
oscilloscope, developing a volt drop across it that is proportional to
the current drawn. This is applied to the Y plates.
CAL. COMPONENT
X-MAG.
0.2V
2V
X10 TESTER
test
component
Fig 7.17
7.24
D3000 ExP Operating Guide
Using an Oscilloscope
Fig 7.18
7.25
Operating Guide D3000 ExP
Using an Oscilloscope
Patching Panel
D3000 boards requiring the use of an oscilloscope contain a patching
panel allowing for the attachment of conventional oscilloscope probes.
An example can be seen in Fig 7.19 below.
CH.1
CH.2
Fig 7.19
7.26
D3000 ExP Operating Guide
Using a Signal Generator
Chapter 8
Using a Signal Generator
Introduction
The Signal Generator or Function Generator is an instrument
designed to produce a periodic electrical signal. The instrument has
controls that allow you to configure the electrical signal to a specific
frequency, waveform shape, and amplitude. The signal generator is
often used to compliment the oscilloscope; the signal generator
produces the waveform that is then displayed on the oscilloscope.
0.9 1.0
1.1 1.2
1.3 TTL O/P
0.7
0.8 1.4 SYMMETRY AMPLITUDE D.C. OFFSET
1.5
0.6 1.6
ON 0.5 1.7
0.4 1.8
0.3 1.9
0.2 2.0
MIN MAX
FREQUENCY
MULTIPLIER
50Ω O/P
0.2Hz - 2MHz SIGNAL GENERATOR
Fig 8.1
8.1
Operating Guide D3000 ExP
Using a Signal Generator
Applications
The signal generator is a useful piece of test equipment and can be
used for a variety of tasks. Three of the more common applications
follow:
• Troubleshooting
• Simulation
• Calibration
8.2
D3000 ExP Operating Guide
Using a Signal Generator
Operation
• Output Connections
Fig 8.2
8.3
Operating Guide D3000 ExP
Using a Signal Generator
Sinusoid
Triangular
Square
• Frequency Selection
FREQUENCY
MULTIPLIER
Fig 8.5
8.4
D3000 ExP Operating Guide
Using a Signal Generator
• Symmetry
The SYMMETRY
control varies the mark/
space ratio of the
generated signal.
Fig 8.6
• Amplitude
AMPLITUDE
8.5
Operating Guide D3000 ExP
Using a Signal Generator
• DC Offset
0V
Fig 8.8
In Fig 8.8, the DC Offset control has been adjusted by +2.5V so that
the square wave is always positive. This signal could now be used for
logic type circuits.
• Attenuation
8.6
D3000 ExP Operating Guide
Fault-Switching and the Control Panel Application
Appendix 1
Fault-Switching and the Control Panel Application
Automatic Fault-Switching
Note: If the faults are not being activated automatically, you must
reconfigure the computer so that they are. If you have access to the
System Tray, click the icon to show the Switched Fault Routing
applet, then change the option to D3000 ExP: If you do not have
access to the System Tray you must ask your instructor to enable
fault-switching for you.
9.1
Operating Guide D3000 ExP
Fault-Switching and the Control Panel Application
Manual Fault-Switching
When you have completed the Worksheet, the switched fault must be
manually de-activated.
There are 15 fault-switching relays inside the D3000 ExP. These are
numbered
1 through 9, and A through F.
If your ExP is connected to a host computer through its USB link, you
can perform the manual fault-switching via an on-screen control panel
application.
If your ExP is not connected to a PC, the faults must be activated and
deactivated by means of the front-panel rotary control.
In either case, you must insert an Instructor’s Key into the front-
panel jack socket on the ExP.
9.2
D3000 ExP Operating Guide
Fault-Switching and the Control Panel Application
An Instructor’s Key must be inserted into the jack socket on the front
panel of your ExP before switched faults can be inserted manually.
Your instructor may issue you with a key, or may prefer to do the
fault-switching for you, in which case you must attract their attention
before attempting a Worksheet that involves a switched fault.
.
Instructor’s Key inserted into front-panel jack socket
9.3
Operating Guide D3000 ExP
Fault-Switching and the Control Panel Application
• Turn the rotary control knob until the desired fault number
appears on the seven-segment display (your instructor may wish
to do this covertly).
• Click the knob to insert the fault. The display will change to a
query mark (?) and the decimal point will light to show that a
fault has been activated.
• Click the rotary control knob. The fault will be removed and the
seven-segment display will show 0.
9.4
D3000 ExP Operating Guide
Fault-Switching and the Control Panel Application
• If you have access to the Start Menu, select D3000 ExP from the
list of available programs.
OR:
.
D3000 ExP control panel with Instructor’s Key in place
9.5
Operating Guide D3000 ExP
Fault-Switching and the Control Panel Application
• When the ExP can communicate with the host PC, its front-panel
PC Control LED is lit, and its 7-segment display shows P.
Active – On when the ExP can communicate with the host PC.
Board In – On when a circuit board is properly loaded.
Fault In – On when a switched fault is currently activated.
9.6
D3000 ExP Operating Guide
Fault-Switching and the Control Panel Application
• Click the In button to insert the fault. The on-screen display will
change to a query mark (?) and the Fault In LED will light.
.
• You can hide the control panel to reduce screen clutter. The panel
has a right-click pop-up menu with three options:
9.7
Operating Guide D3000 ExP
Fault-Switching and the Control Panel Application
9.8
D3000 ExP Operating Guide
Control Panel Options
Appendix 2
Control Panel Options
Introduction
10.1
Operating Guide D3000 ExP
Control Panel Options
Changing Settings
Before you can change any of the settings you must first activate the
control panel.
1234
• LJ Serial Number
This is set at the factory and you cannot change this number.
• Firmware Revision
This is provided by the ExP operating system for information
only.
10.2
D3000 ExP Operating Guide
Control Panel Options
• Use Password
If this box is checked, the Out button on the fault unit is disabled
and a four-digit password code followed by the RETURN key
must be entered to remove a fault. Default is off. The default
password code is:
5678
3. The upgrade will take several minutes, during which time you
must not interrupt the process. When finished, you will be
prompted to switch the ExP off for five seconds; when you switch
it back on it will use the new operating system.
10.3
Operating Guide D3000 ExP
Control Panel Options
Notes:
The D3000EXP.BIN file is renamed to D3000EXP.OLD after a
successful upgrade to avoid accidental reprogramming. Rename it
back to D3000EXP.BIN if you need to repeat the upgrade.
10.4
D3000 ExP Operating Guide
Technical Information
Appendix 3
Technical Information
• Weight: 5 kg approx.
AC Power Connection
The D3000 ExP requires one of the following AC power line inputs:
11.1
Operating Guide D3000 ExP
Technical Information
Ventilation requirements
The ExP provides the following supplies for use on a loaded circuit
board:
• +12V regulated DC at 1A
• -12V regulated DC at 1A
11.2
D3000 ExP Operating Guide
Technical Information
For this reason, the main power switch on the back panel of the
ExP should remain ON whenever the ExP is in use.
PC interface
11.3
Operating Guide D3000 ExP
Technical Information
Equipment maintenance
Before cleaning ensure that the unit is unplugged from any power
supply. Clean the equipment with a lightly damped cloth only.
Ensure that no water can enter the equipment. Ensure the equipment
is dry before any power is applied.
11.4
D3000 ExP Operating Guide
Technical Information
LJ Create has taken great care to design and construct all products
such as to ensure, so far as is reasonably practicable, safety during
normal use, any maintenance and servicing.
The above care includes ensuring electrical integrity where live mains
voltages are used to energize the equipment. Initial integrity is
established by suitable tests and by inspection prior to dispatch. It is
considered that, because it is foreseeable that damage could occur due
to misuse or accident, a regular inspection is carried out. If damage is
suspected or deterioration could have occurred then basic tests for
integrity should be carried out. These tests should be performed by a
suitably qualified person using a portable appliance tester (PAT).
During PAT testing, when insulation resistance or dielectric
withstanding voltage tests are involved, care must be taken to ensure
that low voltage circuit components are not destroyed or degraded.
Where damage due to the PAT testing could occur, or where the
reading obtained by any discrete components, then explanatory and
guidance notes are made in the appropriate sections of this manual.
11.5
Operating Guide D3000 ExP
Technical Information
Warnings
• Ensure liquids i.e. tea and coffee are kept well clear of the
equipment
11.6
D3000 ExP Operating Guide
Technical Information
DECLARATION OF CONFORMITY
Supplementary Information:
The products herewith comply with the requirements of the Low
Voltage Directive 73/23/EEC and amending Directive 93/68/EEC.
11.7
Operating Guide D3000 ExP
Technical Information
Notes
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
................................................................................................................
11.8