WORKING PROCEDURE

Heavy Mechanical Complex-3

Doc. No: WP/QA- 04-01 Rev. No : 00 Date: 09/06/2008

1
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

COVER PAGE

Controlled Copy No. (If Controlled):

Classification

Uncontrolled

This Procedure is the property of Heavy Mechanical Complex-3 and the contents are confidential. This procedure shall not be copied, reproduced or distributed in any form without the prior approval of the Management Representative.

Heavy Mechanical Complex-3 Hattar Road Taxila

Prepared By Signature Name Designation

Reviewed By

Approved By

WORKING PROCEDURE
Heavy Mechanical Complex-3

Doc. No: WP/QA- 04-01 Rev. No : 00 Date: 09/06/2008

2
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

TABLE OF CONTENTS

1. 2 3. 4. 5. 6. 7. 8.
9.

Purpose Scope Responsibilities. Definitions/ Abbreviations. References. Prerequisites. Limitations. Precautions. Procedure for Chemical Analysis. Acceptance criteria. Results and Records

10. 11.

WORKING PROCEDURE
Heavy Mechanical Complex-3

Doc. No: WP/QA- 04-01 Rev. No : 00 Date: 09/06/2008

3
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

1.0

Purpose: This procedure describes the methods used to determine the chemical composition of different types of materials with handheld XRF Metal Analyzer (X-MET3000TX).

2.0

Scope: This working procedure covers methods to perform Chemical analysis test at temperature 0oC to 300oC according to material specifications.

3.0 3.1 3.2 4.0

Responsibilities: Engineer (Material Testing) is responsible for preparation of this procedure. Manager (QA) and Sr. Manager (QA) are responsible to review and approve this procedure. Definitions / Abbreviations: HMC -3 QAD Sr. Manager. QA MT ASTM Heavy Mechanical Complex -3. Quality Assurance Division. Senior Manager. Quality Assurance Materials Testing American Society for Testing and Material

5.0

References ASTM E1476 ASTM E305 ASTM E322 ASTM E572

6.0 6.1 6.2 6.3

Prerequisites: The sample should have proper identification. Test specimen should be ground/polished and cleaned to remove dust, oil and corrosion from the surface of specimen. The machine should have valid Calibration.

WORKING PROCEDURE
Heavy Mechanical Complex-3

Doc. No: WP/QA- 04-01 Rev. No : 00 Date: 09/06/2008

4
Title: Procedure For Handheld XRF Metal Analyzer (X-MET3000TX) Page of 4

7.0 7.1 8.0 8.1 8.2 8.3

Limitations. The analyzer determines only those elements which have atomic number greater than 21. Precautions X-Ray fluorescence metal analyzer generates X-ray radiation, so unnecessary exposure to the radiation should be avoided. Check that infrared safety sensor should work properly before analysis. Use safety shield for small samples.

9.0

Procedure:
Turn the power key of analyzer to ON position. The yellow light indicator is switched on. Wait 1-2 minutes for the peltier cooler and X-rays tube to stabilize. Push the PDA power switch ON. From the Start menu tap, selectX-MET to start the program. Wait for warm-up time completion automatically. After appearing the measurement screen, name the sample to be measured, select the mode of operation, and define LOG file which saves all the measurement data, name and directory.

Modify the way of presenting the measurement and set the measurement time. Touch the nose of the analyzer to the sample and press the trigger. Result will be displayed after measuring time lapsed or when trigger is released( if Measuring time set to 0). Acceptance Criteria Acceptance is made according to the applicable code. Results and Records Chemical analysis test results are recorded on format no, CA01-01.

9.1 9.2