Ventilogic LS
Outlet
Card reader
P mask
Flow
Leakage
VT
200 cm
22 M / 15 F
200 cm
200 cm
22 M
180 cm / Ø 22mm 22 F 15 F
200 cm
200 cm
22 M
15 F
180 cm / Ø 15 mm 22 F
200 cm
180 cm / Ø 22 mm
22 M
15 F
22 M 22 F 180 cm / Ø 22 mm
200 cm
180 cm / Ø 15 mm
15 F
180 cm / Ø 15 mm
22 M 22 F
On / Off key
Quick access
for O2
calibration
Selection of
three different Operating keys
treatments
Dial knob
Trigger
The sensitivity of the
trigger can be
adjusted on 8 levels
and trigger on
exhalation on 14
levels.
Parameters
AirTrap Control
Monitors ventilation with
regards to possible air-
trapping to prevent over
inflation of the lung.
Parameters
Volume compensation
Ensures that the patient is
supplied a minimum
volume. To this end, the
tidal volume determined
for the patient is
compared to a set
minimum volume.
Parameters
LIAM (Lung Insufflation
Assist Maneuver)
Can be used to support
coughing (similar to sigh)
by initially expanding
lung and thorax for
hyper-insufflation.
Monitoring: Start up page
Settings
IPAP / PEEP in
pressure modes
VT / PEEP in
volume mode
Bargraph
Online monitor for pressure
with alarm limits
VT: Calculated volume, out VTi: Measured inspiratory VTe: Measured expiratory
of total inspiratory delivered delivered volume to the tidal volume from the patient
gasflow patient
O2 - Measuring
Possible with all hose
circuits:
Leakage system
Single patient valve
Double patient valve
Necessary:
• SpO2 module box
• SpO2 sensor
SpO2 / Pulse - display
appears only when the
sensor is connected
Information
Low Priority
Medium Priority
(Yellow frame is flashing)
High Priority
(Red frame is flashing)
1s
2s
1s
Alarm window
After confirmation with the silence button, a small alarm window will
overwrite for 2 s the top part of the display.
Autolock off
Avoid that the display is
blocked automatically after
15 min
Available only in clinic mode