You are on page 1of 143

Clinical Laboratory Hematology

THIRD EDITION

CHAPTER 39
Automation in
Hematology

Copyright © 2015, © 2010, © 2004 by Pearson Education, Inc.


All Rights Reserved
Learning Objectives—Level I

At the end of this unit of study, the student


should be able to:
1. Cite the electrical impedance principle of
cell counting, and identify the instruments
that use this technology.
2. Describe the use of radio frequency in cell
counting, and identify the instruments
that use this technology.

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Learning Objectives—Level I

At the end of this unit of study, the student


should be able to:
3. State the principles of light scatter used
in cell counting, and identify the
instruments that use this technology.
4. List the reported parameters for each
blood cell–counting instrument.
5. Categorize cell parameters as directly
measured and derived from a histogram,
scattergram or cytogram, or calculation.
continued on next slide
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Learning Objectives—Level I

At the end of this unit of study, the student


should be able to:
6. Describe the principle of reticulocyte
count enumeration by automated blood
cell–counting instruments.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Learning Objectives—Level II

At the end of this unit of study, the student


should be able to:
1. Compare and contrast the methods of
analysis for the described histograms,
scatterplots, scattergrams, and
cytograms, and interpret the results.
2. Describe and interpret the automated
reticulocyte parameters.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Primary testing vehicle in hematology
lab
• Automated instruments can perform
– Complete blood cell count (CBC)
including
 Platelet count
 Five-part leukocyte differential
 Absolute reticulocyte count

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Two principles of blood counting
currently used
– Impedance
 Based on increased resistance that
occurs when a blood cell with poor
conductivity passes through an electrical
field

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Two principles of blood counting
currently used
– Optical light scattering
 Based on light scattering measurements
obtained as a single blood cell passes
through a beam of light (optical or laser)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Automated reticulocyte counts
– Enumerates reticulocyte
– Reticulocyte hemoglobin concentration
(CHr or Ret He)
– Reticulocyte mean cell volume (MCVr)

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Automated reticulocyte counts
– Immature reticulocyte fraction (IRF)
– Useful in
 Measuring BM engraftment following
transplant
 Determining BM response to iron or EPO
therapy
 Diagnosing iron deficient states in infants
and adolescents

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Body fluid analysis
– Newer feature available on some
instruments
– Includes
 Total nucleated cell count (TNC)
 Leukocyte counts
 Erythrocyte count
 Partial leukocyte differential counts

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Body fluid analysis
– Determined by same technologies as
blood counts
– Improves TAT, increases precision, less
variability

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Impedance instruments
– Coulter® LH Series
– Beckman-Coulter® Unicel DxH 800
– Sysmex XE-Series™
– Sysmex XN-Series™
– Abbott CELL-DYN Sapphire®

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Automated Blood
Cell–Counting Instruments
• Light-scattering instruments
– Siemens Healthcare ADVIA 120
– Siemens Healthcare ADVIA 2120

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• LH 750 and LH 780—single blood


aspiration
– Provides
 CBC
 Five-part leukocyte differential
 Reticulocyte count
– Same basic principles as earlier Coulter
models

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• LH 750 and LH 780—single blood


aspiration
– Provide additional parameters
 NRBC%, Corrected WBC, RDW-SD

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• EDTA blood sample is aspirated


– Closed tube system
– Divided into four aliquots which are
delivered to
 RBC/PLT chamber
 WBC/Hb chamber
 Orbital mixing chamber—differential
 Reticulocyte dilution chamber

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Coulter principle
– Blood cells diluted in an electrically
conductive dilutent
– Within dilution chamber is an external
electrode and three apertures with
internal electrodes

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Coulter principle
– Individual blood cells pass through an
aperture
 Increase in resistance (pulse height)
between internal and external electrode
that is proportional to cell volume
 Steady stream of diluent behind aperture
prevents cells from reentering the
aperture

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Coulter principle
– Threshold limits established for cells
based on cell volumes

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-1 Coulter principle. Increased electrical resistance, or impedance, occurs when the poorly conduc-
tive blood cell passes through the aperture.
Pierre R. Seminar and Case Studies: The Automated Differential. Hialeah, FL: Coulter Electronics, Inc.; 1985.
Reprinted with permission

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Red blood cells


– Particles > 36fL counted as RBCs
– Three RBC counts obtained
– Reported RBC count is average of three
counts
• RBC histogram
– Size distribution curve
– Based on cell volume and relative cell
number
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-2 Histogram, or size distribution curve, allows the visualization of the blood cell population based on
the relative cell number and its volume (in fL).
Significant Advances in Hematology. Hialeah, FL: Coulter Electronics, Inc.; 1983. Reprinted with permission.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-3 Red blood cell (RBC) histogram and count. The shaded area represents those cells used in the red
cell distribution width (RDW) (i.e., RDW–coefficient of variation) calculation. The excluded cells can represent
large platelets, platelet clumps, or electrical interference on the left and RBC doublets, RBC triplets, RBC
agglutinates, or aperture artifacts on the right.
Significant Advances in Hematology. Hialeah, FL: Coulter Electronics, Inc.; 1983. Reprinted with permission.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• RBC/platelet dilution
– RBC count—direct measure
– MCV and RDW—derived from RBC
histogram
– HCT—calculated from MCV and RBC
count
– MCH—calculated from RBC count and
Hgb
– MCHC—calculated from Hgb and Hct

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• RDW-CV
– Based on SD of RBC population and MCV
 RDW-CV = 1SD/MCV × 100%

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• RDW-CV
– Can be false in certain situations
 IF both SD and MCV elevated
– RDW-CV can be normal when anisocytosis
is present
 If SD is normal and MCV is decreased
– RDW-CV can be increased when no
anisocytosis is present

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• RDW-SD
– Red cell distribution width—standard
deviation
 Determined from RBC histogram by
– Calculating width in fL of RBC population
at the 20% height level on RBC histogram

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• RDW-SD
– Red cell distribution width—standard
deviation
 Reflects degree of anisocytosis
 Direct measure
 Not affected by MCV
– Better indicator of anisocytosis

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-4 Calculation of red cell distribution width (RDW)–standard deviation (RDW-SD). The RDW-SD
represents the width of the erythrocyte population at the 20% height level on the red blood cell histogram and is
reported in fL. For example, if ChannelL is 70 fL and ChannelR is 111 fL, the RDW-SD is 41 fL (i.e., 111–70 = 41).
RDW-SD = red cell distribution width–standard deviation
Red Cell Distribution Parameters—(1) RDW-SD (2) RDW (CV), Technical Bulletin 9617. Brea, CA: Beckman-
Coulter®; 2007. Reprinted with permission

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Platelet count
– Particles between 2–20 fL are counted
– Raw platelet histogram is obtained
– Evaluated to determine whether it is a
log normal curve

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Platelet count
– Raw histogram is electronically
smoothed and extrapolated over 0–
70 fL
– Platelet count derived from
extrapolated histogram

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Additional parameters
– MPV—mean platelet volume
 Analogous to RBC MCV

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-5 Normal platelet histogram. The jagged line represents the raw data collected from 2–20 fL. The
smooth line represents the extrapolated histogram from 0–70 fL.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• WBC/hemoglobin chamber
– Hemoglobin determination
 Lytic agent added to dilution
– Lyses RBCs
– Coverts released Hgb to
cynamethemoglobin
– Shrinks WBC membrane and cytoplasm

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• WBC/hemoglobin chamber
– WBC count
 Measure of cell volume not native cell
size
 Particles > 35 fL counted as WBCs by
impedence
 Three WBC counts obtained from three
apertures
 Reported WBC count is average of three
counts
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• WBC count histogram


– Size distribution curve for WBC data
– Visualization of subpopulations of cells
based on WBC relative sizes
– WBC histogram—three-part differential
 Lymphocytes 35–90 fL
 Mononuclear cells 90–160 fL
 Granulocytes 160–450 fL

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• WBC count histogram


– Interferences within cell populations
 generate region flags

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-6 White blood cell (WBC) histogram and count. In a normal patient, the lymphocyte region
represents lymphocytes, the mononuclear region represents monocytes, and the granulocyte region represents
neutrophils, eosinophils, and basophils.
Significant Advances in Hematology. Hialeah, FL: Coulter Electronics, Inc.; 1983. Reprinted with permission.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Table 39-1 Coulter® WBC Histogram Region Flags

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Hemoglobin concentration
– Measure absorbance of
cyanmethemoglobin in the WBC/Hgb
chamber at 525 nm
– Beer's law
 Absorbance reading is proportional to the
concentration of Hgb
 Reagent blank at beginning of each
cycle

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Five-part differential
– Aliquot sent to orbital mixing chamber
– Blood is mixed by gentle agitation with
heated lysing agent
 Remove RBCs, leaves WBCs in near
native state
– Second stablizing agent is added
 Stops lytic activity
 Preserves integrity of WBCs
continued on next slide
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Five-part differential
– Sent to VCS flow cell for five-part WBC
differential
 Volume, Conductivity, Scatter
 Cells pass through flow cell singly by
hydrodynamic focusing

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• VCS flow cell


– Three separate measurements taken
simultaneously
 Cell volume
– Determined by impedance
 Cell conductivity
– High-frequency electromagnetic probe
– Evaluates internal physical and chemical
constituents of cell

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• VCS flow cell


– Three separate measurements taken
simultaneously
 Cell light scatter
– Helium-neon laser
– Determines cell surface, shape, reflectivity

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• VCS flow cell


– Different WBCs are identified and
classified by contour gating
 Individualized for each samples
 Enables differentiation of overlapping cell
populations

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• NRBCs
– First suspected on WBC scatterplot
– WBC histogram examined
 Computer system looks for presence of
cells to left of lymphocyte population
– If both plots indicate NRBCs
 Giant platelets and small lymphocytes
are ruled out

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• NRBCs
– %NRBC (number of NRBCs/100 WBCs)
is derived from VCS data
– NRBC# = NRBC% × WBC count
– WBC is corrected

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-7 White blood cell (WBC) scatterplot.WBC scatterplot graphs volume versus light scatter and reveals
the locations of four leukocyte populations in two-dimensional space. The basophil population is located behind
the lymphocytes but in three-dimensional space is clearly delineated.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Reticulocytes
– Sample mixed with new methylene blue
reagent in heated reticulocyte dilution
chamber
– Residual RNA precipitated within
reticulocytes

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Reticulocytes
– Stained sample is mixed with acidic,
hypotonic solution
 Elutes Hgb from RBCs
– Precipitated RNA remains
 Spheres the RBCs
– Eliminates error of flow cytometric
examination of irregularly shaped
erythrocytes

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Reticulocytes
– Sample sent to VCS flow cell
 Contour gating
– classifies reticulocytes vs. mature RBCs
– Identifies immature reticulocyte fraction

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Reticulocytes
– Parameters
 Absolute reticulocyte count
 Reticulocyte %
 Immature reticulocyte fraction
 MCVr

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Computer system
– Compiles data
– Corrects for coincidence
 Two or more cells passing through the
aperture at the same time
– Records "vote-out”
 Two of three counts do not agree

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Coulter® LH Series

• Computer system
– Reports average of counts when there is
agreement
– Evaluates data to detect abnormalities
– Results, histograms/scatterplots
displayed on screen and printed out or
transferred to LIS
– Suspect flags (software-generated flags)
– User-defined flags (definitive flags)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-8 Coulter® LH 750 white blood cell (WBC) scatterplot and WBC histogram depiction of the presence of
nucleated erythrocytes. On the WBC scatterplot, detection of cells between the lymphocyte population and red
blood cell ghosts suggests nucleated erythrocytes, which are suspected if there is a population of cells to the left
of the lymphocyte population on the WBC histogram. The nucleated red blood cell percentage is determined if
both observations are made and other potential interferences are ruled out (e.g., giant platelets).

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-9 Coulter® LH 750 report from a normal individual.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• Determines CBC using same methods


as Coulter LH series
• Improvements to
– WBC differential
– NRBC determination
– Reticulocyte enumeration

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• VCSn module
– Next generation of VCS technology
– Diff, NRBC and reticulocyte
measurements are made here

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• Light scatter detection by flow cell


– Multitransducer module (MTM)
 Multiple sensors around flow cell
 Collect three measures of light scatter—
lower, upper, and low-angle (LMALS),
(UMALS), (LALS)

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• Light scatter detection by flow cell


– Multitransducer module (MTM)
 Additional sensor measures axial light
loss (AL2)
 Fifth light scatter measure
– Median angle light scatter (MALS)
– Sum of LMALS and UMALS

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• Multitransducer module (MTM)


– LMALS, UMALS and UMAL—detect
granularity and membrane topography
– AL2—evaluates cellular transparency
– LALS—evaluates cellular complexity
index

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• System manager
– Evaluates raw data from flow cell
– Applies algorithm
– Calculates cell frequency (percent and
absolute count)
– Data-plots
 Two-dimensional and three-dimensional

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-10 Light scatter measurements from the Unicel® DxH 800 flow cell. The upper median angle light
scatter (UMALS) and lower median angle light scatter (LMALS) detect the cell’s granularity and membrane
topography. The axial light loss (AL2) evaluates cellular transparency and the low-angle light scatter (LALS) is
used as a cellular complexity index.
Advancements in Technology: WBC Differential Methodology, Technical Bulletin 9403. Brea, CA: Beckman-
Coulter®; 2009, Reprinted with permission

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-11 Unicel® DxH 800 white blood cell differential dataplots. Diff 1 dataplot graphs volume versus
rotated median angle light scatter (RMALS) and Diff 2 dataplot graphs volume versus opacity (OP). These
dataplots reveal five leukocyte populations in two- dimensional space.
Advancements in Technology: WBC Differential Methodology, Technical Bulletin 9403. Brea, CA: Beckman-
Coulter®; 2009, Reprinted with permission

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• NRBC%
– RBCs enumerated in VCSn module

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• NRBC%
– NRBC mix chamber
 Sample is diluted
 Lyse reagent is added
– Lyses red cells but keeps NRBCs, WBCs,
Platelets, cellular debris intact
 Flow cell analysis for VCS
– AL2 and LALS—used to determine NRBCs
– Two 2-dimensional plots created
– NRBC% determined from collected data

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-12 Unicel® DxH 800 nucleated red blood cell (NRBC) dataplots. NRBC 1 dataplot graphs rotated low-
angle light scatter (RLALS) versus axial light loss (AL2) and NRBC 2 dataplot graphs rotated upper median angle
light scatter (RUMALS) versus AL2. These dataplots clearly delineate the nucleated erythrocyte population from
the leukocyte population.
Advancements in Technology: NRBC Enumeration, Technical Bulletin 9404. Brea, CA: Beckman-Coulter®; 2009,
Reprinted with permission

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• Reticulocyte mix chamber


– Blood is mixed with NMB and RNA is
precipitated
– Clearing reagent is added
 Stabilizes dye–reticulum complex
 Spheres RBCs and reticulocytes (similar
to Coulter LH)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• MTM evaluates each cell for:


– Volume, conductivity, and five light
scatter characteristics
– System manager applies algorithm
 Discriminates reticulocytes from mature
RBCs using
– RETIC1 AND RETIC2 dataplots (vol vs LLS
and vol vs OP)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• MTM evaluates each cell for:


– Volume, conductivity, and five light
scatter characteristics
– System manager applies algorithm
 Determines RETIC%, absolute count
(RETIC% xRBC count), IRF, and mean
reticulocyte volume (MRV)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-13 Unicel® DxH 800 Reticulocyte dataplots. RETIC 1 dataplot graphs volume versus linear light
scatter (LLS) and RETIC 2 dataplot graphs volume versus opacity (OP). These dataplots provide delineation of
reticulocytes from mature erythrocytes and platelets.
Advancements in Technology: Reticulocyte Methodology, Technical Bulletin 9405. Brea, CA: Beckman- Coulter®;
2009, Reprinted with permission

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• System manager
– Evaluates all data
– Determines if an interference is present
– Corrects CBC for clearly identified
interferences
 WBC for NRBCs
 Hemoglobin for WBC > 11 × 103/mcL
 RBC for WBC > 140 × 103/mcL

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• System manager
– Abnormalities generate suspect, system,
and definitive flags to alert operator

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Beckman-Coulter Unicel® DxH 800

• Body fluid analysis


– RBC and TNC on body fluids
– Uses CBC module data
 CSF, serous, synovial fluids—all
acceptable
 Pre-treat synovial fluids with
hyluronidase to reduce viscosity
– Each lab should establish acceptable
background counts

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Uses four different technologies


– Impedance (i.e., direct current)
– Radio frequency
– Absorption spectrophotometry
– Flow cytometry—fluorescent dyes

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Parameters reported
– CBC
– Five-part leukocyte differential
– Nucleated erythrocytes
– Reticulocyte count

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Hemoglobin
– Measured using the sodium lauryl
sulfate (SLS) method (cynanide free)
– RBCs are lysed—release Hgb
– SLS converts ferrous iron to ferric iron—
methemoglobin

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Hemoglobin
– Methemoglobin combines with SLS
 SLS-hemichrome molecule
 Measured at 555 nm
 Beer's law

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RBCs and platelets


– Counting using impedance with
hydrodynamic focusing

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RBCs and platelets


– Resistance detected between the
internal and external electrodes
 RBCs counted between a lower
discriminator (25–75 fL) and a upper
discriminator (200–250 fL)

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RBCs and platelets


– Resistance detected between the
internal and external electrodes
 Platelets counted between a lower
discriminator (2–6 fL) and a upper
discriminator (12–30 fL)
– Fixed discriminator at 12 fL detects
potential interferences (e.g., RBC
fragments)

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RBCs and platelets


– RBC and platelet histograms are
generated
– HCT calculated from pulse heights of
individual RBCs
 Pulse height is proportional to RBC
volume
– Other derived or calculated parameters
 MCV, MCH, MCHC, RDW-SD, RDW-CV,
MPV
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• WBC count and differential generated


from:
– Two channels
 DIFF
 WBC/BASO

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Diff channel
– Sample mixed with lyse reagent, RBCs
lysed
– Creates "holes" in cytoplasmic
membranes of nucleated cells
– Fluorescent dye enters cells, binds to
DNA and RNA
– Sent to optical detection block
 Cells pass singly through laser
continued on next slide
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Diff channel
– Each cell generates
 Forward scatter (size), side scatter (cell
complexity), fluorescent intensity
 Scattergram—side scatter vs. fluorescent
intensity

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Diff channel
– Adaptive cluster analysis system
 Clear separation of cell populations
– RBC ghosts, lymphocytes, monocytes,
eosinophils, and neutrophils + basophils

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Diff channel
– IG master software
 Immature granulocyte population
counted
– Metamyelocytes, myelocytes,
promyelocytes

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• WBC/BASO channel
– Sample mixed with lyse reagent, RBCs
lysed
– Shears all WBCs except basophils

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• WBC/BASO channel
– Analyzed by optical detection block
 Captures forward and side scatter as
cells pass through laser beam
 Adaptive cluster analysis
– Total WBC count and basophil count
(percentage and absolute)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-14 DIFF scattergram, Sysmex XE-2100™. RBC ghosts, lymphocytes, monocytes, eosinophils, and
nutrophils + basophils are identified. Basophils are specifically identified in the WBC/BASO scattergram. Courtesy
of Sysmex Corporation, Kobe, Japan.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• IMI channel
– Reagents act selectively on lipid
membranes of WBCs
– Mature WBCs are lysed, leave bare
nuclei
– Immature myeloid cells remain intact

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• IMI channel
– Cells are analyzed by direct current and
radio frequency (cells internal structure
and density)
– Determine degree of immaturity
 Immature granulocytes and blasts
 Hematopoietic progenitor cell (HPC)
count
– Correlates with CD34+ count by flow
cytometry
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XE-5000
– Multichannel messaging (eMM) software
 Improved detection of
– Blasts
– Abnormal lymphocytes/lymphoblasts
– Atypical lymphocytes

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• NRBC channel
– Counts NRBCs, corrects WBC and
lymphocyte count
– Lyse RBCs, removes NRBC cytoplasmic
membrane, shrinks nucleus, fluorescent
dye binds to nuclei

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• NRBC channel
– WBC membrane perforated, dye enters
and binds to organelles and nucleus
 Fluorescence is stronger in WBCs than
NRBCs

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• NRBC channel
– Pass through laser beam and generate
FS, SS, and fluorescence
 Used to create NRBC scattergram
 Sensitive to NRBCs as low as
0.1/100WBCs
 #NRBC used to correct WBC count and
lymphocyte count

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-15 Nucleated red blood cell (NRBC) scattergram, Sysmex XE-2100™. Nucleated erythrocytes are
clearly delineated from leukocytes based on their lower fluorescence intensity and smaller size (i.e., less forward
scatter) as seen on this schematic pattern.WBC, white blood cell; PLT, platelet
Rowan RM, Linssen J. A picture is worth a thousand words. Sysmex Journal International. 2005;15(1):27–38.
Reprinted with permission.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RET channel
– Fluorescent dye binds to residual RNA
within reticulocytes and another dye
binds to RNA and DNA in nucleated cells

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RET channel
– Each cell evaluated by diode laser for FS
and fluorescence
 RET scattergram is generated
 Mature erythrocytes, reticulocytes and
platelets
 Reticulocytes further classified for
maturity
– Immature platelet fraction (IPF)—
reticulated platelets
continued on next slide
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RET channel
– Absolute reticulocyte count, %
reticulocyte, low, middle, high
fluorescence ratio (LFR, MFR, HFR),
immature reticulocyte fraction (IRF),
reticulocyte hemoglobin content (ReT
He), and platelet count

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• RET channel
– IRF
 Represents MFR and HFR combined
 Reported as % of total reticulocyte count

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-16 Reticulocyte (RET) scattergram, Sysmex XE-2100™. Mature erythrocytes, RETs, and platelets
(PLTs) are identified.
RBC = red blood cell
Rowan RM, Linssen J. A picture is worth a thousand words. Sysmex Journal International. 2005;15(1): 27–38.
Reprinted with permission.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XN-10 instrument—latest Sysmex


(2012)
– Modular component capable of
determining
 CBC, six-part diff (neutrophils,
lymphocytes, monocytes, eosinophils,
basophils, and IGs, NRBCs, and
reticulocyte count)
 SP-10 slide maker/stainer creates
complete workstation
continued on next slide
Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XN-10 instrument
– Uses same methods for RBC count, Hgb,
RBC parameters, platelet count,
reticulocyte count and reticulocyte
parameters

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XN-10 instrument
– New methods
 Adaptive Flagging Algorithm based on
Shape recognition (SAF-LAS)
– Allows better discrimination of WBCs,
abnormal WBCs, and NRBCs
 Low WBC mode
– Accurate diffs on low WBC counts (< 0.5 ×
103/mcL)

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XN-10 instrument
– New methods
 White cell nucleated (WNR) channel
– Clear delineation between NRBC and WBC
– Total WBC count does not require
correction

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-17 Sysmex XE-Series™ report from a normal individual.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Fluorescent platelet channel (PLT-F)


– Determines IPF and platelet count
– Platelets are clearly delineated based on
size and fluorescent intensity
 Depicted in PLT-F scattergram
 Immature platelets have high
fluorescence intensities

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• Fluorescent platelet channel (PLT-F)


– Provides accurate platlet count when
 Numbers are low
 Red cell or WBC fragments are present

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XN-Series
– Have body fluid module
– No pretreatment or off-line prep is
required
– RBC-BF
 Determined by impedance

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Sysmex XE-Series™

• XN-Series
– WBC-BF, TC-BF, and two-part diff
(PMNS and mononuclears)
 Determined by fluorescent FC in WDF
channel
– HF-BF cells (e.g., macrophages,
mesothelial cells) can also be
distinguished on the BF scattergram

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-18 WDF scattergram, Sysmex XN-Series™. Lymphocytes (LYMPH), monocytes (MONO), eosinophils
(EO), neutrophils (NEUT)+basophils (BASO), and immature granulocytes (IG) are identified. Basophils are
specifically identified in the WNR scattergram.
SSC = side scatter; SFL = fluorescence intensity
Courtesy of Sysmex Corporation, Kobe, Japan.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• Technologies used
– Flow cytometry, fluorescence staining,
impedance
– Enumerate
 WBCs, RBCs, platelets, reticulocytes,
five-part differential

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• Hemoglobin dilution cup


– RBCs lysed, free Hb complexes with
imidazole
– Hgb concentration determined
spectrophotometrically at 540 nm
 Cyanide free method

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• WBC dilution cup


– Lyses RBCs, strips cytoplasmic
membrane from NRBCs and fragile
WBCs
– Stains DNA of nuclei with PI—goes to
flow cell

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• WBC dilution cup


– MAPPS technology
 Light scatter and fluorescence
characteristics determined
– 0° (forward) = cell size
– 7° = cell complexity
– 90° (side) = nuclear lobularity
– 90° depolarized light scatter = cytoplasmic
granularity

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-20 Four light scatter measurements from the CELL-DYN Sapphire ®. The 0° scatter reflects cell size;
7° scatter reflects cell complexity; 90° scatter reflects nuclear lobularity; and 90° depolarized scatter reflects
cytoplasmic granularity.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• WBC dilution cup


– 0° vs 7° = neutrophils, monocytes,
lymphocytes
– 90° D vs 90° = eosinophils
differentiated from neutrophils
– Lymphocytes separated from basophils
based on size and complexity

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• WBC dilution cup


– Total WBC, five-part diff, NRBCs,
nonviable WBCs
 NRBCs clearly distinguished from WBCs
– Total WBC and five-part differential not
affected by their presence

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-21 0° (Size) vs 7° (complexity) scatterplot, CELL-DYN Sapphire ®. Neutrophils, monocytes, and
lymphocytes are identified. Lymphocytes are located lower left in the scatterplot; the middle cell population is
composed of monocytes; and neutrophils are located upper center in the scatterplot.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• RBC/Plt cup
– One dilution—spheres RBCs, counts
using impedance
 RBC count and histogram, platelet count
and histogram
 Part of dilution sent to optical flow cell
– RBCs and platelets enumerated by light
scatter

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• RBC/Plt cup
– Second dilution—reticulocytes
 Nucleic acid stained with fluorescent dye
 Reticulocytes determined based on
fluorescence and light scatter
– Reticulocyte count and immature
reticulocyte fraction (IRF)

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-22 Reticulocyte histogram, CELL-DYN Sapphire ®. Reticulocytes are found between the two gates.
The peak to the left represents mature erythrocytes.
x-axis = fluorescence; y-axis = relative number of cells

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-23 CELL-DYN Sapphire® report depicting WBC differential.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• New software package—2010


– Extends RBC and reticulocyte
parameters
 Calculated from data collected

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• New software package—2010


– Determines microcytic RBC% (%MIC),
macrocytic RBC% (%MAC),
hypochromic RBC% (%HPO),
hyperchromic RBC% (%HPR), MCVr,
MCHr, CHCr
 Shown to correlate well with similar
parameters from ADVIA

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Abbott CELL-DYN Sapphire®

• Two additional assays


– Fluorescent immunophenotyping
methods
 Immuno T-cell assay
– CD4 and CD8 populations
 ImmunoPlt assay
– CD61

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Light-Scattering Instruments

• Siemens Healthcare ADVIA 120


• Siemens Medical ADVIA 2120

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• Methodology
– Continous flow analysis, peroxidase
staining, flow cytometry
– CBC, five-part WBC differential,
reticulocyte count

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• RBC/Plt channel
– Isovolumetric sphering of RBCs and Plts
– Diluted cells pass singly through a laser
beam

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• RBC/Plt channel
– RBCs counted and sized by high-angle
and low-angle light scatter
 MCV, CHCM
– Generate RBC cytogram, RBC histogram,
Hgb histogram
• RDW, HDW are derived from these
histograms

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• RBC/Plt channel
– Platelets measured with high-angle and
low-angle light scatter
– Platelet cytogram and RBC cytogram
• Hgb channel
– RBCs lysed, free Hgb converted to
cyanmethemoglobin
– Measured at 546 nm

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• WBC count and five-part differential


– RBCs lysed
– Peroxidase channel
 Neutrophils, monocytes, eosinophils—
degree of peroxidase positivity
 Lymphocytes and large unstained cells
(LUCs) remain unstained
 Pass through flow cell
 Peroxidase cytogram

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• WBC count and five-part differential


– Basophil/lobularity channel
 Basophil diluent lyses RBCs, platelets,
strips all WBCs except basophils of
cytoplasm
 Flow cell

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens HealthCare ADVIA 120

• Reticulocyte channel
– Cellular RNA stained
– Flow cell
– Absolute and relative reticulocyte count
– CHr™, MCVr, reticulocyte cellular Hgb
concentration mean

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Figure 39-24 ADVIA 120 report from a normal individual.

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens Medical ADVIA 2120

• Basic principles same as ADVIA 120


• Hgb measurement
– Cyanide-free methodology
• Neither ADVIA 120 and ADVIA 2120
can enumerate NRBCs

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens Medical ADVIA 2120

• First to produce extended RBC


parameters
– %HYPO, %HYPER, %MICRO, %MACRO
 Combined use of %HYPO and CHR can be
useful in early detection of iron deficiency
in chronic renal disease dialysis patients

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens Medical ADVIA 2120

• DNI index
– Delta neutrophil index
 DNI = (neutrophils + eosinophils from
peroxidase channel)—
(polymorphonuclear cells from
basophil/lobularity channel)
– Reflects number of immature granulocytes
– May be useful as prognostic indicator for
sepsis

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams
Siemens Medical ADVIA 2120

• CSF analysis
– Use dedicated CSF mode

continued on next slide


Clinical Laboratory Hematology, Third Edition
Shirlyn B. McKenzie | J. Lynne Williams
Siemens Medical ADVIA 2120

• CSF analysis
– Sample is mixed with CSF assay reagent
 Spheres and fixes cells
 Incubated for 4 minutes
 Determines cell counts and two-part diff
by light scatter and absorbance
 CSF WBC count, CSF RBC count, and
two-part differential
– PMNs and MNs reported in absolute and
relative values

Clinical Laboratory Hematology, Third Edition


Shirlyn B. McKenzie | J. Lynne Williams

You might also like