Professional Documents
Culture Documents
Roshan Chaudhary
Introduction
Consists of:
-X-ray tube &
-Fluorescent screen (Cu: ZnCdS)
Early Fluoroscopic Systems
Early fluoroscopy = Image was viewed directly – X-ray photons strike the
fluoroscopic screen – emits visible light.
Early Fluoroscopic Systems
Fluoroscopic Image viewing based on Human Vision.
Cones: Photopic (daylight) Vision.
Rods: Scotopic (night) Vision;
Most sensitive to blue-green light – Daylight levels reduce the
sensitivity to low illumination levels; \
Required red goggles for dark adaptation (to filter out blue green
wavelengths)
The dim fluoroscopic vision required use of rod vision, with its poor
visual acuity and poor ability to detect shades of gray (contrast).
DISADVANTAGES:
Requires complete darkroom.
High Radiation dose to the patient and the operator.
Only single person can view the image.
Requirements
2 Types:
Over couch X ray Tube.
Under Couch X ray Tube.
Offers less scatter radiation to the operator than Over-couch X
ray tube systems
Preferable to have X ray tube mounted below the Patient Couch.
Other C Arm Fluoroscopic/Interventional Suite.
Fluoroscopic X ray Tubes
Fluoroscopic Couches/Table
Mostly tilting Type
Identified by their degrees of tilt.
Allows patient to be lowered and raised.
Image Intensifier Tube
Developed in 1953.
Evacuated vacuum tube.
Converts x-ray into visible light image.
Amplifies the image brightness approx. 5000 to 20000 times.
II Tube Design
Input phosphor
Initial input phosphor used in early image intensifiers was
zinc-cadmium sulfide (ZnCdS),
Current phosphor of choice is cesium iodide (CsI:Na).
CsI was deposited on a thin aluminum substrate by a process
called vapour deposition.
CsI grew in tiny needles perpendicular to the substrate, .
The phosphor thickness have been reduced from 0.3mm to
0.1, which produced increased resolution.
Vacuum window (thin Al window); curved for accurate
electron focusing.
Components of II Tube
Function: Input phosphor absorbs x-ray photons and converts their energy into light
photons.
Components of II Tube
Accelerating Anode
A circular plate with a hole in the middle to allow the e-s through
to the output phosphor.
Located in the neck of the II tube
The potential applied at the anode is +25 to +35 KV relative to the
photocathode..
This results in gain of kinetic energy by the electrons .
Components of II Tube
Brightness gain
Flux gain
Minification gain
Conversion factor
II Tube Performance/Characteristics
Brightness Gain
Ability of the II tube to increase the illumination level of the image.
BG deteriorates as II Tube ages.
Ranges from 5000-30000.
Minification gain: Increase in image brightness that results from
reduction in image size from the input phosphor to the output
phosphor.
Minification gain does not change the contrast of the image makes the
image brighter , does not improve the quality of image.
Brightness Gain = (Minification gain) x (Flux-gain) where,
Conversion Factor:
Ratio of output phosphor illumination to the input exposure rate.
Brightness gain/ Intensification factor
10 to 30 Cd/m2 per μGy/s..
Conversion Factor ={output phosphor illumination (Cd/m2)}/{input
exposure rate(mR/s)}
Multifield II Tubes
Attempts to resolve the conflicts between image size & quality.
Operated in several modes, including the 4.5 inch, a 6 inch, or a 9 inch
mode.
9 inch mode is used to view large anatomic areas with minified image.
4.5 inch or 6inch mode produces magnified resultant dim image. ,
minification gain reduced and maintained with increased mA
determined by ABC that increases spatial resolution.
Field size is changed by a simple electronic principle: the higher the
voltage on the electrostatic focusing lens, the more the electron beam is
focused.
Greater voltage to electrostatic lenses:
– Increases acceleration of electrons
– Shifts focal point away from anode
– Used to enlarge small structure or to penetration through larger parts
Multifield II Tubes
Television Monitoring
With the television monitoring system of a fluoroscopic image,
the output phosphor of the image-intensifier tube is coupled
directly to a television camera tube.
Brightness level and contrast can be controlled electronically.
Several observers can view the fluoroscopic image at the same
time.
Allows for storage of the image in its electronic form for later
playback and image manipulation.
Viewing the Fluoroscopic Image
Viewing the Fluoroscopic Image
TV IMAGE
TV Image is similar to a screened print.
Made up of mosaic of hundreds of thousands of tiny dots of
different brightness.
Each dots contributes a minute bit to the total picture.
When viewed from a distance , the individual dots disappear ,
but at close range or with magnification, they are clearly
visible.
Recording the Fluoroscopic Image
Photospot camera
Records the the image output of an II on a film
Film – role film/cut film(10 cm)
Advantage
Reduction in pt exposure
Film does not have to be changed b/w exposures
Exposure times are shorter-motion is less likely problem
Films can be taken more rapidly
Possible to record & view image at same time
Recording the Fluoroscopic Image
Cineflourography
Process of recording fluoroscopic images on movie (cine) film
Two film sizes- 16 mm, 35 mm
Cine camera-components are lens, iris diaphragm, shutter,
aperture, pressure plate, pull down arm & film transport
mechanism
Recording the Fluoroscopic Image
Resolution
-Horizontal Resolution
-Vertical Resolution
Electronic Noise
Contrast
Lag
TV Image Quality
HORIZONTAL RESOLUTION
Limited by factors such as frequency bandwidth of the video system.
Concept of video bandwidth can be explained by drawing an analogy
with the stereo systems that many peoples have in their homes.
This analogy can be extended to the video systems by imaging a
low pass filter at the end of the amplifier chain that transmits the
signal from the camera to the monitor.
If the bandwidth is too low, a rounded –off blurry representation of
the gauge will be passed to the video monitor.
If the bandwidth is too high ,there is a great deal of electronic
noise ,generated by the video system at high frequencies.
TV Image Quality
HORIZONTAL RESOLUTION
Dictated by the number of video lines.
Since ,the electron beam spot in the pick up tube is not pill-
box shaped (with sharp edges ) but tends to be more Gaussian
,resulting in overlap betn the lines(esp. during interlaced
scanning) reduces the vertical resolution.
Also affected by the raster employed.
ELECTRONIC NOISE
Since ,the output phosphor of an II tube may exhibit quantum
noise & structure mottle ,it reduces sharpness of the final
image.
The TV system itself produces noise during the electronic
amplification of the weak video signal from the pick up tube
& degrades the quality of the image.
This produces on the TV monitor screen a “ Snowstorm”
effect similar to quantum noise.
TV Image Quality
CONTRAST
Contrast of the image on a TV camera depends on the contrast of
the original image focused onto the signal plate of the TV pick up
tube.
Contrast adjustments cause a change in average screen luminences
i.e. reducing the contrast setting also causes a reduction in image
brightness & a compensatory adjustment to the brightness control is
required to maintain a satisfactory image.
Distortion of contrast occurs in TV pick up tubes unless they
produce a dark current.
TV image contrast is also affected by the ambient lighting
conditions.
TV Image Quality
LAG
Term used to describe the inability of an imaging system to follow
rapid changes in its output image.
Refers to the condition in which the video signal output rises or
falls to the new value more slowly than the changing light input.
Undesirable property of most Vidicon tubes ,which becomes apparent
when the camera is moved rapidly results in the blurring of the
image.
Is signal dependent i.e. it is much higher in the dark parts of the
image than in bright parts.
CCD cameras has no lag.
TV Image Quality
Digital Fluoroscopy
Specific Advantages:
Pulsed Progressive Fluoroscopy:
High mA Mode- risk of tube over load avoided by pulsed mode, reduced exposure time per
pulse creating frames
Short Tube interrogation and extinction time
Temporal Frame averaging:
Averaging displayed image with previous frame
Decrease image noise. Works well for static imaging but increased image lag unacceptable for
dynamic studies
Digital Fluoroscopy
Operating Modes
i) Continuous Fluoroscopy
the most basic form of fluoroscopic imaging
X-ray beam is constantly on.
A frame rate of 25 or 30 fps yields a frame integration time of 40 or
33 msec, which can lead to blurring of moving objects
ii) Pulsed Fluoroscopy
Short Pulses of X-rays used.
Pulsed fluoroscopy can be accomplished either by
a) Operating the Generator in pulsed mode.
b) Using a Grid-Controlled or Grid-Switched XRT.
Digital Fluoroscopy
Operating Modes
Pulsed Fluoroscopy
Lower radiation dose when fluoroscopic frames are acquired in a
fraction of the time used in continuous operation.
X-ray production is needed only immediately prior to the Readout
of the video camera.
Improved Image Quality owing to reduction in motion blur because
of the reduced integration time.
Pulsed mode operation freezes the motion of objects in the image,
resulting in sharper images and improved image quality.
Reduced tube loading owing to a lower duty cycle as low as 5-8%.
Digital Fluoroscopy
Silicon(a-Si).
o Direct Conversion FPIR: Amorphous Selenium (a-se)
Digital Fluoroscopy
FPIR is much small and lighter and is manipulated more easily
than an image intensifier.
FPIR imaging suite provides easier patient manipulation and
radiologist/technologist movement.
No radiographic cassettes.
Image Intensifier is limited by non-uniform spatial resolution and
contrast resolution form the center to periphery of the circular
image and Veiling glare and pincushion distortion increase with the
age but the response of an FPIR is uniform over the entire receptor
and does not degrade with age.
The image captured by an FPIR is square or rectangular similar to
the associated flat panel monitors.
Digital Fluoroscopy
CsI/a-Si is an indirect DF process by which
X-rays are converted first to light by CsI
and then to electrical signal by a-Si.
The image receptor is fabricated into
individual pixels which has light sensitive
face of a-Si with a capacitor and a TFT
embedded in it.
Digital Fluoroscopy
Amorphous Selenium is the direct DF process
by which X-rays are converted to electric
signal as no scintillation phosphor is involved.
The imaging forming X-ray beam interacts
directly with a-Se producing a charged pair.
a-Se is both the capture element and the
coupling element.
a-Se is approx. 200 m thick and is sand-
wiched between charged electrodes.
X-rays incident on a-Se create electron hole
pairs through direct ionization of Selenium.
The created charge is collected by a storage
capacitor and remains there until the signal is
read by the switching action of the TFT.
Digital Fluoroscopy
The geometry of each individual pixel is very
important as a portion of the pixel face is occupied
by conductors, capacitors and TFT.
It is not totally sensitive to the incident image
forming X-ray beam.
The percentage of the pixel face that is sensitive to
X-rays is the fill factor, which is approximately
80%; therefore 20% of the X-ray beam does not
contribute to the image.
This represents one of the dilemmas for Digital
Imaging.
As the pixel size is reduced, spatial resolution
improves but at the expense of the patient
radiation dose.
With smaller pixels, the fill factor is reduced and
the X-ray intensity must be increased to maintain
adequate signal strength.
Digital Fluoroscopy
Digital Fluoroscopy
Digital Fluoroscopy
Flat Panel Display technology is rapidly replacing the Cathode Ray Tube
(CRT) in all applications like Television, Computer.
Radiography and fluoroscopy is the similar field in which CRTs are rapidly
being replaced by the Flat Panel Image Display.
In fluoroscopic image viewing, Flat Panel Image Display is usually Active
Matrix Liquid Crystal Display (AMLCD)
A liquid crystal has the property of a highly ordered molecular structure-
a crystal- and the property of viscosity-a fluid.
Liquid crystals materials are the linear organic molecules that are
electrically charged, forming a natural molecular dipole.
Consequently, the liquid crystal can be aligned through the action of an
external electrical field.
Active matrix liquid crystal displays are superior to cathode ray tube
displays.
Digital Fluoroscopy
o Easier to view
o Easier to manipulate
o Provide better images
o Can be readily mounted suspended in an angiographic room