You are on page 1of 39

The

LOGIQ 200
PRO Series
LOGIQ 200 PRO Series

Professional
Image Quality

Investment Productivity
Progress
Protection

The PRO Series


represents outstanding value for the most discerning professional,
delivering increased productivity and superior diagnostic outcomes
for you and your patients.
SCAN Board
CON 1

RELAY
CON 2
-SHV

+SHV

USC BUS
SCANNER INTERFACE
TBF

PGC
TX DRV PRE AMP

\
+THV
CROSS
P GUN ASSY POINT
-THV

WATCH DOG INTERFACE

USC BUS

NOVA BUS
RBF
PGC
ACC GEN

TO MVP

AAF ADC DDS COSMO


TGC

IIC USC BUS


NOVA BUS
Digital Beamformer
MR Technology WOW!!
(Wide Aperture Technology) 40MHz Ultra-High frequency
Sampling Rate

011010011010
Flicker-free Monitor!!
011010011010

011010011010

Σ
Target 011010011010

011010011010

011010011010

Probe Digital Time Delay Digital Adder

Catch Axial, Lateral, and


Temporal
Resolution….
Digital Beamformer
Flicker-free
Monitor
ATO
Axial Resolution 3CB(2-5MHz Probe)

Auto Wave
Lateral Resolution Shaping
Best B-Mode IQ!
Contrast Resolution Wide Aperture
Technology
Temporal Resolution
40MHz
Converter
High Precision Digital Beamformer
ATO
Acoustic Data
LOGIQ200MD PRO
Image Processor Image

ATO gives you the best


contrast resolution
using One-Touch on keyboard Image
Analysis

Optimization based on Digital Adaptive Processing


ATO
Conventional Manual
Image Optimization

2-5 Minutes

ROI
In seconds

ATO

Less Operator Dependent… Faster... Better Image Quality


NOVA BUS
IMAGE
MVP Board STORAGE
RAM
IMAGE
IMAGE FRAME XY
ALM ITP DISPLAY
SOFTENER AVERAGE MEMORY
CONTROL

VIDEO Video
PEAK CRDTF ZOOMER LUT DAC
HOLD
B/M FRONT
END CONTROL
XY ADDR NON-INT INTER
LACE VID
GEN CONVERT
CONVERT DAC

VIDEO
CINE CINE DECODER
CONTROL MEMORY

SAP SYNC
GENERATOR

XY
USC BUS FIFO MEMORY

USC BUS PROGRAM


I/F DSP
MEMORY
MSTE
RESET CPU

ROM RTC

SUPER
RAM
I/O

VRAM IDE
AGDC
CTL

FROM NOVA BUS


HV

+/-100V

+/- 65 V
LV Voltages From System
Password : Antony
COMPARISION OF LOGIC 200 WITH 200 PRO

Logic 200 Alpha Logic 200 Pro


MST MST
DSC MVP
ESP RBF
RTC TBF
CONN SCAN
HV unit BPHV UNIT
MONITOR Non Interlace
Monitor
Follow these steps to enter the Service Menu.
Note
Probes should be installed in the port and Blue Shift key should be off before take these steps.
1. Pressing Shift + Ctrl keys and then press the Shift +s keys, and then Shift + 1 keys.
2. Input the PASSWORD and press Return key.
The Service Top Menu will appear on the monitor as shown in ILLUSTRATION 2--5.
Service Utility MENU : Top Menu 99/03/12 09:16:37
Current Software Version is 3.0
*System Diagnosis
Board/Program Version
System Reset
Key Operation History
Option Set
Initialize :1 (1:America 2:Europe 3:Asia 4:Japan)
System Clock Adjust
Exit
TEST Conditions
Follow these steps to exit the Service Menu.
1. Press Shift + R keys. If this step will be performed again, the Service Menu will be entered again.
OR
2. Select Exit on the Top Menu, and press Set or Return keys.
The Service Menu will be exited.

Follow these steps to operate the Service Menu.


1. Using Arrow Up/down keys, select one of seven items.
2. Press Set or Return keys.
The Service Sub Menu will be appeared or Service Menu will be operated.
Note
Resetting the system using the System Reset, to reboot System, does not affect on the system
presets.
Follow these steps to enter the System Diagnosis menu.
1. Using the Arrow up/down keys, select the “System Diagnosis”menu in the Top Menu. Refer to the ILLUSTRATION 2--5. The left side of
selected menu, here will be appeared the cursor (*).
2. Press Set or Return keys.
The System Diagnosis menu will appear on the monitor as shown in the ILLUSTRATION 2--6.
Service Utility MENU : System Diagnosis 99/03/12 09:16:37
Current Software Version is 3.0
NAME Time Status
*System test 21 M
Scan Test 0 M
TBF Test 0 M
Linear/Convex Path 0 M
Sector Probe Path 0 M
BPHV test 1 M
MVP Hardware Test 1 M
B mode Function 1 M
B mode Cine 8 M
M mode Function 2 M
MSTE test 4 M
Keyboard 0 M
PIOP 1 M
Monitor 0 M
Power Supply 1 M
Exit
Press ’ctrl’+’X’to
Failure Codes
Wide Range of Probes

CBF Convex Probe CAE Convex Probe LB Linear Probe


- Frequency:3.5Mhz -Frequency:3.5Mhz
-Frequency:5Mhz
- Bandwidth: 2.5 - 5.0 Mhz -Bandwidth: 2.5 - 5.0 Mhz
-Bandwidth: 3.7- 7.4 Mhz
-FOV:72deg -FOV:94mm
-FOV:72deg
-Convex ROC:50mmR
-Convex ROC:72mmR
-Biopsy guide: Available
-Biopsy guide: Available

LE Linear Probe LD Linear Probe MTZ Micro-Convex Probe

-Frequency:5.0Mhz -Frequency:6.5Mhz
-Frequency:3.5Mhz
-Bandwidth: 3.7 -7.4 Mhz -Bandwidth: 4.3 -8.6 Mhz
-Banswidth: 2.5 -5.0 Mhz
-FOV:83mm -FOV:124deg
-FOV:94mm
-Convex ROC:10mmR
-Biopsy guide:Available
Wide Range of Probes

LHLinear Probe
-Frequency:7.7Mhz
-FOV:63mm
-Biopsy guide:Available CZBMicro-Convex Probe
-Frequency:6.5Mhz
-Bandwidth: 4.3 -8.6 Mhz
-FOV:124deg
-Convex ROC:10mmR

LI Linear Probe
-Frequency:7.5Mhz
-Bandwidth: 5.2 -9.5Mhz
-FOV: 35mm
LTLinear Probe
-Frequency:7.5Mhz
-Bandwidth: 5.2 -9.5 Mhz
-FOV:35mm
Wide Range of Probes

ERB Micro convex/Linear 10L Linear probe S317 Phased Array Sector probe
Array probe
-Frequency : 9.0Mhz -Frequency : 3.5Mhz
-Frequency : 7.0Mh
-Bandwidth : 5.2-11.3Mhz -Bandwidth : 2.5-5.0Mhz
-Bandwidth : 5.2-11.3Mhz
-FOV : 40mm -FOV : 90 deg
-FOV : 12 deg

3Cb Convex Array Probe CS Microconvex Array probe


-Frequency : 3.5Mhz -Frequency : 3.5Mhz
-Bandwidth : 2.5-5.0Mhz -Bandwidth : 2.7-4.2Mhz
-FOV : 59 deg -FOV : 81 deg
e
“We BringGoodThings ToLife”

You might also like