Professional Documents
Culture Documents
• CHAIRPERSON: DR PAVANI
• ASST.CHAIRPERSON: DR SHIVANI
• MODERATOR: DR. INDIRA
• POSTGRADUATE: DR SUDHA
GOALS
• NORMAL LUNGS:
• VT OF10-12 ML/KG IBW
• RATE 8-12/MIN
• RESTRICTIVE LUNGS:
• VT OF 4-8ML/KG IBW
• RATE 15-25 (WATCH I:E RATIO FOR ENOUGH
EXHALATION TIME)
• AIRWAYS OBSTRUCTION AND RESISTANCE:
• VT OF 8-10 ML/KG IBW
• RATE 8-12/MIN
MINUTE VENTILATION
• MINUTE VENTILATION IS THE AMOUNT OF AIR INSPIRED OR
EXPIRED IN ONE MINUTE.
• MV = [TIDAL VOL MINUS DEADSPACE VOL] * RESPIRATORY RATE.
• 5 TO 6 LITRES/MIN IN HEALTHY INDIVIDUALS AT REST
• INCREASES AMONG PATIENTS WHO ARE MECHANICALLY
VENTILATED OR HAVE INCREASED CARBONDIOXIDE
PRODUCTION
• -FEVER
• -HYPERMETABOLIC STATES
• -HYPOXEMIA
• INCREASED DEAD SPACE
• INCREASED CENTRAL RESPIRATORY DRIVE.
RATE OF GAS FLOW
THE FLOW SETTING ESTIMATES THE
DELIVERED FLOW OF INSPIRED GASES.
INITIAL PEAK FLOW SETTING IS
60ML/MIN[40-80]
• I:E = TI/TE
• TI=VT/FLOW
INSPIRATORY TIME: EXPIRATORY
TIME RELATIONSHIP (I:E RATIO)
• DURING SPONTANEOUS BREATHING,
-- THE NORMAL I:E RATIO IS 1:2- INDICATING THAT
FOR NORMAL PATIENTS THE EXHALATION TIME IS
ABOUT TWICE AS LONG AS INHALATION TIME.
• IF EXHALATION TIME IS TOO SHORT--- “BREATH
STACKING” OCCURS RESULTING IN AN INCREASE IN
END-EXPIRATORY PRESSURE ALSO CALLED AUTO-
PEEP.
• DEPENDING ON THE DISEASE PROCESS, -SUCH AS IN
ARDS, THE I:E RATIO CAN BE CHANGED TO IMPROVE
VENTILATION
INSPIRATORY PAUSE
• A MANEUVER THAT PREVENTS THE
EXPIRATORY VALVE FROM OPENING FOR A
SHORT TIME AT THE END OF
INSPIRATION,WHEN THE INSPIRATORY VALVE
IS ALSO CLOSED.
• MOST FREQUENTLY USED TO OBTAIN AN
ESTIMATE OF THE PLATEAU PRESSURE
• IT IS USED TO IMPROVE
- DISTRIBUTION OF AIR IN THE LUNGS,
-PROVIDE OPTIMUM V/Q MATCHING AND
- REDUCE VD/VT RATIOS.
• IT INCREASES PAW AND REDUCES
PULMONARY BLOOD FLOW..
INITIAL SETTINGS DURING
PRESSURE VENTILATION
• PRESSURE VENTILATION SETTINGS
HAS THE ADVANTAGE OF
LIMITING PRESSURES TO • BASELINE PRESSURE (PEEP)
AVOID OVER-INFLATION • IP IS SET AT A LOW
AND PROVIDING FLOW
PRESSURE (10-15CMH20)
ON DEMAND
AND ADJUSTED TO ATTAIN
• THE CHANGE IN THE DESIRED VOLUME
PRESSURE BETWEEN THE
BASELINE AND PIP IS SET • RATE, TI, AND I:E ARE SET
TO ESTABLISH THE VT JUST AS IN VOLUME
DELIVERY (PEEP VENTILATION
COMPENSATION)
POSITIVE END-EXPIRATORY
PRESSURE (PEEP)
• PEEP IS ADDED TO MITIGATE END-EXPIRATORY ALVEOLAR
COLLAPSE
• . NORMALLY APPLIED PEEP IS 5 CMH2O.
• HOWEVER, UP TO 20 CMH2O MAY BE USED IN PATIENTS
UNDERGOING LOW TIDAL VOLUME VENTILATION FOR ACUTE
RESPIRATORY DISTRESS SYNDROME (ARDS)
• AUTO-PEEP:
• INCOMPLETE EXPIRATION PRIOR TO THE INITIATION OF NEXT
BREATH CAUSES PROGRESSIVE AIR
TRAPPING(HYPERINFLATION). THIS ACCUMULATION OF AIR
INCREASES ALVEOLAR PRESSURE AT END OF
EXPIRATION,WHICH IS REFERRED TO AS AUTO-PEEP..
PRESSURE SUPPORT
VENTILATION
THE VENTILATOR PROVIDES A
CONSTANT PRESSURE DURING
INSPIRATION ONCE IT SENSES
THE PATIENT HAS MADE AN
INSPIRATORY EFFORT
PSV GOALS
1.TO HELP INCREASE THE VT (4-8ML/KG)
2.TO DECREASE THE RESIRATORY RATE (<30 BREATHS/MIN)
3.TO DECREASE THE WORK OF BREATHING .
INITIAL SETTINGS DURING
PRESSURE SUPPORT
VENTILATION
• PSV IS USED BEFORE THE • FOR PATIENTS WITHOUT
PROCESS OF WEANING LUNG DISEASE, ABOUT 5CM
H2O IS USED TO
• THE PRESSURE IS SET AT A
COMPENSATE FOR
LEVEL TO PREVENT A
ADDITIONAL WORK OF
FATIGUING WORKLOAD ON BREATHING.
THE RESPIRATORY MUSCLES
• FOR PATIENTS WITH LUNG
• LEVEL OF PS CAN BE SET DISEASE ,8-14 CM OF H2O
BASED ON AIRWAY ARE USED TO COMPENSATE
RESISTANCE OR EQUAL TO FOR ADDITIONAL WORK OF
PTA[PIP-PPLAT] BREATHING.
SYNCHRONIZED INTERMITTENT
MANDATORY VENTILATION
BREATHS ARE GIVEN AT A SET MINIMAL RATE, HOWEVER IF
THE PATIENT CHOOSES TO BREATH OVER THE SET RATE NO
ADDITIONAL SUPPORT IS GIVEN
ONE ADVANTAGE OF SIMV IS THAT IT ALLOWS PATIENTS TO
ASSUME A PORTION OF THEIR VENTILATORY DRIVE
SIMV IS USUALLY ASSOCIATED WITH GREATER WORK OF
BREATHING THAN AC VENTILATION AND THEREFORE IS LESS
FREQUENTLY USED AS THE INITIAL VENTILATOR MODE
LIKE AC, SIMV CAN DELIVER SET TIDAL VOLUMES (VOLUME
CONTROL) OR A SET PRESSURE AND TIME (PRESSURE CONTROL)
NEGATIVE INSPIRATORY PRESSURE GENERATED BY
SPONTANEOUS BREATHING LEADS TO INCREASED VENOUS
RETURN, WHICH HELP CARDIAC OUTPUT AND FUNCTION .
FRACTION OF INSPIRED OXYGEN
• FIO2 IS DEFINED AS THE PERCENTAGE OR CONCENTRATION OF
OXYGEN THAT IS DELIVERED.
• THE LOWEST POSSIBLE FIO2 SHOULD BE USED TO MEET
OXYGENATION GOALS .
• THIS WILL DECREASE THE ADVERSE CONSEQUENCES OF
SUPPLEMENTAL OXYGEN TO DEVELOP -SUCH AS
-- ABSORPTION ATELECTASIS,
-- ACCENTUATION OF HYPERCAPNIA,
-- AIRWAY INJURY,
-- PARENCHYMAL INJURY
ADVANTAGES OF EACH MODE
Mode Advantages
Inspiratory
pause
1
2
A = MAP
B
1
5
5
PEEP +5
No patient Patient
effort effort
•The baseline for the pressure waveform increases when PEEP is added.
•There will be a negative deflection just before the waveform with
patient triggered breaths.
PRESSURE/TIME
SCALAR
•A-An increase in airway resistance causes the PIP to increase, but Pplat
pressure remains normal.
•B-A decrease in lung compliance causes the entire waveform to increase in
size. The difference between PIP and Pplat remain normal.
VOLUME/TIME
SCALAR
•There will also be a plateau if an inspiratory pause set or inspiratory hold maneuver
is applied to the breath.
VOLUME/TIME SCALAR
•Tidal Volume
•Active Exhalation
•Asynchrony
VOLUME/TIME
SCALAR
Air-Trapping or
Leak
Loss of
volume
•If the exhalation side of the waveform doesn’t return to baseline, it could
be from air-trapping or there could be a leak (ETT, vent circuit, chest tube,
etc.)
FLOW/TIME
SCALAR
Volume Pressur
e
FLOW/TIME SCALAR
=
Normal
Expiratory flow
doesn’t return
to baseline
Start of next
breath
•If expiratory flow doesn’t return to baseline before the next breath starts,that means
there’s AUTO- PEEP (air trapping) present , e.g. emphysema.
FLOW/TIME
SCALAR
Bronchodilator Response
Pre-Bronchodilator Post-Bronchodilator
Longer Shorte
E- r E-
time time
Peak Exp.
Flow
Improved Peak Exp. Flow
Dynamic
Complianc
B e
B = Exp. (Cdyn)
ion
500
t A = Inspiratory
Resistance/ a
pir A Resistance/
Elastic WOB ex Resistive WOB
tion
p ira
250 ins
5 15 30
•Lung Overdistention
•Airway Obstruction
•Bronchodilator Response
•Respiratory Mechanics (C/Raw)
•WOB
•Flow Starvation
•Leaks
•Triggering Effort
PRESSURE/VOLUME LOOPS
OVERDISTENTION
“beaking”
500
250
5 15 30
AIRWAY RESISTANCE
n ce
a
ist
500
e s
p .r
ex i s ”
s
re
s te
y
250 “h c e
tan
is
es
p.r
s
in
5 15 30
500 500
250 250
15 30 15 30
60
40
20
Start of Start of
Inspiration Expiratio
0 n
200 400 600
-20
-40
-60
PE
F
FLOW/VOLUME
LOOPS
•Air trapping
•Airway Obstruction
•Airway Resistance
•Bronchodilator Response
•Insp/Exp Flow
•Flow Starvation
•Leaks
•Water or Secretion accumulation
•Asynchrony
FLOW/VOLUME
LOOPS
•The shape of the inspiratory curve will match the flow setting on the
ventilator.
FLOW /VOLUME LOOPS
Airway
Obstruction
Reduce
“scooping” d
PEF
•The expiratory curve “scoops” with diseases with small airway obstruction
(high expiratory resistance). e.g. asthma, emphysema.
FLOW/ VOLUME
LOOPS
A
60
Leak
=
40 Normal
20
0
200 400 600
Expiratory -20
portion of
loop does not -40
return to
starting point,
-60
indicating a
leak.
•If there is a leak, the loop will not meet at the starting point where inhalation starts
and exhalation ends. It can also occur with air-trapping.
PRESSURE/VOLUME
LOOPS A
Leak
500
250
5 15 30
•The expiratory portion of the loop doesn’t return to baseline. This indicates a
leak.
L EAKS
• Expiratory leak: ETT cuff leak , chest tube leak, Broncho Pleural
Causes
fistula, NG tube in trachea
•Inspiratory leak: loose connections, ventilator malfunction, faulty
flow sensor
How to ID
it • Pressure wave: Decreased PIP
• Volume wave:Expiratory side of wave doesn’t return to
• baseline Flow wave: PEF decreased
• Pressure/Volume loop: exp side doesn’t return to the baseline
• Flow/Volume loop: exp side doesn’t return to baseline
How to fix it
• Check possible causes listed above
• Do a leak test and make sure all connections are
tight
AIRWAY RESISTANCE
CHANGES
• Bronchospasm
Causes:
• ETT problems (too small, kinked, obstructed, patient
• biting) High flow rate
• Secretion build-up
• Damp or blocked expiratory valve/filter Water in the HME
•
How to Identify it on the graphics
• Pressure wave: PIP increases, but the plateau stays the same
• Flow wave: it takes longer for the exp side to reach baseline/exp flow
rate is reduced
• Volume wave: it takes longer for the exp curve to reach the baseline
• Pressure/Volume loop: the loop will be wider. Increase Insp. Resistance
will cause it to bulge to the right. Exp resistance, bulges to the left.
• Flow/Volume loop: decreased exp flow with a scoop in the exp curve
How to fix
• Give treatment, suction patient, drain water, change HME, change ETT,
add a bite block, reduce PF rate, change exp filter.
ASYNCHRONY
THIS OCCURS WHEN THE TIMING OF THE VENTILATOR CYCLE IS NOT
SIMULTANEOUS WITH THE TIMING OF THE PATIENTS RESPIRATORY CYCLE
Causes (Flow, Rate, or Triggering)
Air hunger (flow
•
starvation)
• Neurological Injury
How•toImproperly
ID it set sensitivity
•
• PRESSURE WAVE: PATIENT TRIES TO INHALE/EXHALE IN THE MIDDLE OF THE
WAVEFORM, CAUSING A DIP IN THE PRESSURE
•
• FLOW WAVE: PATIENT TRIES TO INHALE/EXHALE IN THE MIDDLE OF THE
WAVEFORM, CAUSING ERRATIC FLOWS/DIPS IN THE WAVEFORM
•
• PRESSURE/VOLUME LOOP: PATIENT MAKES EFFORT TO BREATH CAUSING DIPS IN
LOOP
• EITHER INSP/EXP.
• FLOW/VOLUME LOOP: PATIENT MAKES EFFORT TO BREATH CAUSING DIPS IN
LOOP EITHER INSP/EXP.
•
•
ASYNCHRONY
Flow
Starvation
•The inspiratory portion of the pressure wave shows a scooping or “dip”, due
to inadequate flow.
ASYNCHRONY
F/V P/V
Loop Loop
RISE TIME cc
•The amount of time to reach
•the desired airway pressure
•or peak flow rate.
Inspiration
ends
pressur
e
flow
•The breath ends when the ventilator detects inspiratory flow has dropped to a
specific flow value.
INSPIRATORY CYCLE OFF
100% of Patient’s
Peak Inspiratory Flow
100
%
75%
50%
30%
•In the above example, the machine is set to cycle inspiration off at 30% of the
patient’s peak inspiratory flow.
INSPIRATORY CYCLE OFF
Exhalatio
A B
n spike
100 100
% %
60
% 10
%
•A –The cycle off percentage is too high, cycling off too soon. This makes the breath
too small. (not enough Vt.)
•B – The cycle off percentage is too low, making the breath too long. This forces the
patient to actively exhale (increase WOB), creating an exhalation “spike”.
THANK YOU…..