UT Lesson10 08 12
UT Lesson10 08 12
Introduction
• Ultrasonic transducers, though identical in
appearance and manufactured to the same
specification, usually have individual
characteristics.
• Acoustic anomalies may exist because of
variations in crystal cutting, areas of poor bonding
to the lens or backing, and misalignment of parts
in the transducer assembly.
Transducer Characteristics
Transducer Characterization
Equipment
• Equipment used to measure the sending and
receiving characteristics of an ultrasonic
transducer is capable of reproducing an exact
indication on the display screen of the signals sent
and received by the transducer.
• The movement of the transducer over the reflector
is accurately controlled.
Transducer Characteristics
Transducer Characterization
Equipment
• With data potentiometers coupled to sense the
motion, a distance amplitude plot of the sound
beam is produced on the display screen.
• An open shutter camera can then be used to
record the beam profile on self-developing film.
Transducer Characteristics
Test Setup
• The transducer is placed in a couplant tank made
of plastic or glass so that the immersed transducer
and reflector can be viewed through the couplant.
• The reflector is scanned by the sound beam with
accurate motion of the scanning transducer
ensured through milling table cross feeds to move
the transducer.
Transducer Characteristics
Test Setup
• Potentiometers coupled to the cross feeds convert
motion data into electrical signals that are fed into
the horizontal position controls of the sweep
display circuit.
• The horizontal display shows the distance of
transducer traverse in inches.
• Either X or Y directions of crystal movement are
produced by switching from the output of one
potentiometer to the output of the other.
Transducer Characteristics
Function
• Instrumentation equipment consists of a:
– Timer.
– Delay unit.
– Pulser.
– Wideband receiver.
• The unit repeatedly pulses the transducer with a
sharp spike and then amplifies the return signals
fed back through the transducer during operation
(as illustrated on the next slide).
Transducer Characteristics
Function
Recording Method
The response curve is recorded with an open
shutter camera. The sweep delay is used to delay
the presentation across the display screen, as
shown below:
Transducer Characteristics
Manipulative Equipment
• To obtain precise sound beam and focal length
measurements, precision elevating and
transversing mechanisms are required.
• Milling table cross feeds consisting of heavy
micrometer screw slides calibrated in thousandths
of an inch are used.
• Two of the slide screws are fitted with sprocket
and chain drives connected to data
potentiometers that develop the sweep signal.
Transducer Characteristics
Reflector Targets
• Reflector targets must be carefully chosen
because a bad target will seriously distort the
signal and will produce invalid information.
• In most cases, precision steel balls are used,
particularly when calibrating focused transducers.
The diameter of the ball must be as small as
possible.
Transducer Characteristics
Reflector Targets
The size of the effective reflecting surface of the
ball is held to less than one-quarter wavelength of
the transducer frequency to prevent frequency
distortion and undue influence of the target on the
measurement of the beam.
Transducer Characteristics
Reflector Targets
• Selection of ultrasonic reflectors varies with the
geometry of each crystal and lens.
• Reflectors must be small compared with the beam
size measured and about equal in size to actual
discontinuities the transducer is expected to
detect.
• For example, a flat, circular reflector of one-eighth
the crystal diameter is adequate for testing flat
disk transducers used to detect fairly large
discontinuities.
Transducer Characteristics
Pulser
• The transducer test requires a pulser with a short
pulse capability.
• To analyze the natural frequency and the damping
characteristics of the transducer, the transducer
must be excited with a voltage pulse that will not
drive the crystal into any abnormal oscillation.
• This requirement demands that the pulse duration
be as short as possible, much less than one
period of the natural resonant frequency of the
crystal.
Transducer Characteristics
Pulser
For analysis of high frequency (5 to 25 MHz)
transducers, the recommended pulse duration is
0.025 μs with a rise time of 10 ns (a microsecond is
one millionth second, and a nanosecond is one
billionth second).
Transducer Characteristics
Wideband Receiver
• To prevent the received signal from becoming
distorted, a receiver with a wideband radio
frequency amplifier is used.
• A recommended receiver is one with a bandwidth
of 1.5 to 60 MHz, a rise time response of 10 ns
and a gain of at least 40 dB.
Transducer Characteristics
Display System
• An effective display system has sufficient
bandwidth and rise time to present data without
distortion.
• Bandwidth is direct current to 30 MHz with a rise
time capability of 0.01 μs.
• This scope combination offers delay and time
base expansion features that are desirable for
recording transducer beam profiles.
Transducer Characteristics
Recording of Transducer
Beam Profiles
Transducer data sheets are prepared for mounting
of photographic records and recording of the
transducer analysis factors.
Transducer Characteristics
Cylindrically Focused
Transducer Measurements
• A wedge-shaped sound beam, focused in width
and unfocused along the length, is produced by a
cylindrically focused transducer that has a
concave lens.
• Beam width is measured by traversing the
immersed transducer across the length of the
beam over a steel ball reflector, as shown in the
next slide.
Transducer Characteristics
Cylindrically Focused
Transducer Measurements
Transducer Characteristics
Cylindrically Focused
Transducer Measurements
• Ball diameter is selected based on the frequency,
crystal size and lens radii of the transducer being
tested.
• A rule of thumb is to select as small a reflector as
possible that will still produce adequate signal
levels for profiling.
Transducer Characteristics
Cylindrically Focused
Transducer Measurements
• Because beam width is usually narrow, the
problem of maintaining ball alignment when
traversing the beam length may be avoided by
substituting a piece of wire for the ball.
• The wire diameter depends on the same factors
that determine ball diameter selection: frequency,
crystal size and lens radii.
Transducer Characteristics
Waveform
The return signal waveform is calibrated in millivolts
on the vertical scale with time on the horizontal
scale to allow a determination of crystal frequency,
damping factor and sensitivity.
Transducer Characteristics
Frequency
• In frequency analysis, the actual frequency of
transducer operation is measured and compared
to the design frequency.
• The actual frequency is a measurement of the
acoustic wave in the water medium.
• As this is the frequency of the energy used when
testing material, this is the frequency that is
recorded.
Transducer Characteristics
Frequency
• To record the acoustical frequency of the
transducer, the first reflected signal from the ball
target is analyzed.
• The frequency may be calculated if the period
(time base) is known.
• Frequency equals the number of complete cycles
per unit of time.
Transducer Characteristics
Damping Factor
• The damping factor is defined as the number of
positive half cycles within the pulse that are
greater in amplitude than the first half cycle.
• By counting the number of cycles generated by
the crystal when reacting to the reflected pulse, a
measure of the damping factor is reached.
• With this method, the damping factor is essentially
a measurement of the time required for the crystal
to return to a quiescent state after excitation.
Transducer Characteristics
Damping Factor
• The resolution of the transducer is directly related
to the damping factor.
• The smaller the damping factor, the better the
ability of the transducer to resolve two signals
arriving very close together in a given time.
Transducer Characteristics
Sensitivity
Sensitivity is a measure of the ability of the
transducer to detect the minute amount of sound
energy reflected from a relatively small target.
Transducer Characteristics
Focal Length
• The focal length information for focused transducers
is the water path distance at which a maximum
return signal is obtained.
• Focal length is the time base measurement on the
display screen between the excitation pulse and the
point of maximum amplitude response.
• The transducer is held over the center of the ball
target and moved toward or away from the ball until
the maximum reflected signal is received.
• The focal length is then noted and recorded.
Transducer Characteristics