Professional Documents
Culture Documents
● An isolated CVP measurement has very little meaning unless the information
is interpreted in the context of some estimation of cardiac function.
CVP and Venous Return
● Venous return is mostly determined by the gradient between the mean
circulatory filling pressure (MCFP) and CVP.
● MCFP results from the elastic recoil pressure from distended small veins and
venules and is the force that drives blood back to the right atrium.
● This explains why a patient may lose 10% to 12% of his circulating blood
volume without exhibiting changes in blood pressure or CVP.
● The difference between MCFP and CVP is only 6 to 8 mm Hg, and hence
small changes in CVP may have profound hemodynamic consequences.
CVP and Cardiac Function
● Changes in CVP may be the sole result of changes in inotropic state or
compliance of the ventricle, independent of the total circulating volume or
venous return to the heart.
● CVP is the result of a complex and diverse interplay among many different
physiologic variables, many of which are impossible to measure in the
operating room or ICU.
● It is therefore not surprising that studies assessing the value of CVP as a
predictor of volume status or fluid responsiveness have failed to demonstrate
a relationship.
Waveform of CVP
Components of physiological
Pressure measurement
● Invasive catheter
● Pressure transducer
● Normal saline flush
● Pressure infusion bag
● Reusable pressure cable
● Bedside physiological monitor
Levelling pressure transducer system
● Intravascular monitoring should be at the level of heart or phlebostatic axis.
● Should be done with a Carpenter’s level.
● Errors in pressure reading may occur if alignment with phlebostatic axis is not
maintained.
● For every inch (2.5 cm) the heart is offset from the reference point of the
transducer, a 2 mmHg of error will be introduced.
Levelling pressure transducer system
Levelling pressure transducer system
Zeroing
● Zero referencing eliminates the effects of atmospheric and hydrostatic
pressure
● Open the reference stopcock to air by removing the non-vented cap, keeping
sterility intact
● After removing non-vented cap, turn stopcock off to the patient
● Initiate “Zero” function on bedside monitor and confirm pressure waveform
and numeric value display 0 mmHg
● Once the “zero” is observed, turn the stopcock back to the vent port and
replace the non-vented cap
Square wave test and damping
1. Activate snap or pull tab on flush device