You are on page 1of 12

Planned / Preventive

Maintenance

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Agilent 1100 Series HPLC
AN-CE-LC-1-001-E
SPIFPMs – service documents
Available for different LC and other systems
1100/1200 Series
1120/1220 Series
1260 Series
1290 Series
Scope of Work
- Scope/time (4 h for a typical system)
- Customer/service resposibilities (next slide)
- Part numbers, tools, consumables

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Maintenance Overview

Piston spring

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Agilent 1100 Series HPLC
AN-CE-LC-1-001-E
Recommeded tests after concluded
maintenance

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Customer responsibilities

• Customer has flushed the system with HPLC grade water, all samples
removed from autosampler and analytical columns removed from the column
compartment

• Customer has isopropanol and HPLC grade water in the appropriate


mobile phase containers available for the procedures

• Customer saves all methods and sequence files prior to PM start.

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Planned Maintenance Checklist

• Check the customer has removed the column, flushed the system and that
all methods and sequence files are saved.

• Checklist should be completed in the order given


1200_PM_Checklist.pdf

• If a test fails record the result in the Test Result box and continue with the
rest of the PM checklist.

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Agilent 1100 Series HPLC
AN-CE-LC-1-001-E
In case a test fails

• If customer has a service contract, support provider will log a service call.

• If customer does not have a service contract, the support provider will
inform the customer of the failure (leak, inoperable sensor, test etc.), and
offer to log a service call for the customer.

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
LC System

• Check firmware revision of modules to be serviced.

• If firmware is not current, ask customer if they would like the


firmware upgraded.

• If the firmware is upgraded, remember to fill in the “1100/1200 IQ


attachment form” (in the SPIFPM’s folder on your training DVD)

Agilent 1100 Series HPLC


AN-CE-LC-1-001-E
Use the PM Scope of Work and
PM Checklist as instructions.
Record standard values.
Agilent 1100 Series HPLC
AN-CE-LC-1-001-E

You might also like