PLX5200 High Frequency Mobile Digital Radiography

System Instructions for User


Nanjing Perlove Medical Equipment Co.,Ltd
• Part 1
Machine structure
Main structure

Pillar
X-ray tube

Rack
Collimator

FPD

Bumper
Brake switch
Movement
NO. Function
1 Touch screen
2 Emergency stop switch
3 Power Indicator
4 Ready indicator
5 Exposure indicator
6 Fault indicator
7 Sports battery indicator
8 Exposure battery indicator
9 Detector power off
10 Detector power on
11 Key switch off position
12 Key switch M position, can only move
the machine
13 Key switch R position for all operation
Movement
• Turn on the circuit breaker
• Turn the key switch to the M position
• Put both hands on the push rod in front of the machine, Pick up the push rod enabling
handle, and you can use the electric assist function of the machine normally.
Movement
This machine adopts the intelligent induction booster system, and there are
many ways of moving, as follows:

Status Operation
Apply force forward with both hands simultaneously to make right
Forward and left wheels advance fast at the same time.
Apply force backward with both hands simultaneously to make right
Backward and left wheels back away slowly at the same time.
Apply force forward with left hand and backward with right hand to
Right make the machine turn right.
Turn
Apply force backward with left hand and forward with right hand to
Left make the machine turn left.
Collimator
NO. Function
1 Cross arm rotating around stand column,
cross arm stretch unlock button
2 Ascending button

3 Lead door horizontally adjust

4 Radiation filed light open/close button

5 SID display

6 Lead door vertically adjust

7 Descending button

8 Tube rotating around cross arm rotate


unlock button
Up and down button Rotate button
1 and 2
Extend rack

First raise the tube, then


hold down either of the
rotary buttons, pulling
outward with both handles,
and the track will longer
Rotate
Rotate button 1

Rotate button 2
• Part 2
Preparation
Confirm installation environment

Ambient temperature: 10℃ ~ 40℃


Relative humidity: 30% ~ 75%
Power voltage: AC single-phase 220V±10%
Power frequency: 50/60±1Hz
Power supply internal resistance: no more than 1Ω
Over-current release characteristics: 10A
Have good protective grounding terminals and ensure reliable
connection of protective grounding
Installation tools
Tools you need to prepare before installing the machine: Multimeter, Crowbar, Electric
hand drill, M8 sleeve, Screwdriver, and A complete set of Allen wrenches.

Multimeter Crowbar Screwdriver

Electric hand drill M8 sleeve Allen wrenches


Check
Check whether the electric power line in the room is a 220V AC voltage not
exceeding ± 10% of the nominal value.
Check whether the room has a good protective ground wire device.
Please take photos for record.

Photos of power supply condition:

1. Single-phase voltage : 220V


±10%

2. Zero-ground voltage : less than 5V


• Part 3
Unpacking
Inspection
• Inspection of the packaging
Package inspection: whether the package is damaged or broken.
Please take photos(video) from All Positions for a record before unpacking .
Inspection

Engineer standing in
front of the goods
Shipping Mark Protection sign
Unpacking
Remove the top cover, check again whether
there is any obvious damage inside , and also
you need to take photos(video) for storage.
Then remove the surrounding side covers
Unpacking
Clear photos(video) of All Positions of the machine need to be taken.

Top

Left side

Right side

Front and Back


Unpacking

Take out all


these fixing
frame
Unpacking
Take out all accessories.
• Check whether the accessories are complete and intact one by one according to
the packing list .
• Requires the signature of responsible person for confirmation.
• If there is any discrepancy or damage, please take photos and contact us in
time.
Installation
Place the slope board, install the brake pedal and step it clockwise,
manually push the machine to slide down the slope board.
Installation
• Use the Allen wrench to install the front cover.
• Part 4
Installation
Installation
• Collimator
Use an open-end wrench to loosen the 4 screws and move the two fixing
plates to the outward edge.
Installation
• Lift the beam limiter so that it snaps into the lower tray of the tube, push the
fixing piece inward, and use an open-end wrench to fix the beam limiter.
• Use an Allen wrench to rotate clockwise to fix the beam limiter in 4
directions.
Installation

• Connect the power cord of the beam limiter and rotate clockwise to
fix it.
Installation
• Insert the power cord into the corresponding power cord holder and
rotate clockwise to fix it.

Air Switch

Power line Protective Ground Wire


Installation
• Insert the hand switch into the corresponding notch and place it on
the hand switch base
Installation
• Fix the two remote controls on the hand switch harness.
Installation
• Use a screwdriver to install the rear cover.
Installation
• Take out the charger base and adapter, Insert the output cable into the
corresponding red dot on the adapter and charger. Plug the power
cord into the adapter.
Installation
• Take out the battery, insert two batteries properly into the charger
base. Connect the power supply and charge the battery for more than
4 hours.
Installation
• Turn the Flat Panel Detector upside down, Buckle off the battery back
cover and turn it clockwise to take out.
Installation
• Remove the battery from the charger And Put the battery into the
corresponding notch.
Installation
• Close the battery cover and gently place the Flat Panel Detector in the
slot on the back of the machine.
• Part 5
Preparation
before use
Observe
Plug in the power cable, Close the air circuit-breaker The machine starts to power up,
then the fan working, and the lithium battery charging, super capacitor charging
It needs charging for more than 8 hours before the first time use, and the two lights bar
will reach the top after full charge.
• Part 6
Start up
Start up
1. Connect the power line
2. Turn the air switch to ON state
3. Hit the key to block M. Test whether movement in all directions is normal.
4. Press the power button of the flat panel detector for more than 3S to turn on the flat
panel detector
5. Press the DR "ON" start button, hit the key to block R. The computer will turn on
• Part 7
Shut down
Shut down

Click button to shut down the software.

Click button to shut down the system.

Click button “Cancel” to exit from the interface.

1. Select to turn off the whole system


2. Hit the key to block OFF.
3. Turn the air switch to OFF state
4. Disconnect the power line
• Part 8
Workstation
Software
Workstation

Click button
New patients need to register basic
information:

• Serial number
• Name
• Age
• Gender

Attention: Do not enter Capture straightly without


clicking “Save” after completing the registration
Workstation

1.Click button

2. Select corresponding body part

3. Select : position, adult/children,


oversize/normal size

4. Select corresponding photography


position

5. Adjust KV/mAs

6. Press

7. Press remote controller or


Workstation

Flip horizontally Flip vertically Move

Rotate left 90 degree Rotate 90 degree Restore

Invert WW/WL ROI

Magnifier DICOM Information Zoom


Workstation
Click button As shown in the figure below, the image is rebuilt.

Select the part of the human body and click the "rebuild" button to reconstruct the image
Workstation
1. Click button

2. When querying patients, you need to set


query criteria, which are queried by date by
default. The query conditions include
registration date, serial number, ID and
name. When the check box on the right side
of the query criteria displays "√", the criteria
is selected.

3. Click "update" to change the


basic information of the currently
selected patient.
Workstation

Double-click to select the image you want to add to the report in the
thumbnail on the right and the image will be automatically added to the
report ( 3 at most). Preview and print the report.
Workstation

1. Users can click the buttons "single


window", "double window", "three
window" and "four window" to select the
film layout mode.

2. Click the thumbnail in the lower


right corner to select the film to be
printed.

3. Click the printer drop-down box,


select the target printer, select the film
size and film direction

4.Click the button "print " to start


printing film.
Workstation
Film print settings:
1. Click the button "Print set", as shown in the figure below, to set the film printing.
2. Click “Add” to add a new parameter configuration, Click “Delete” to remove a configuration.
3. Select a configuration and the parameters will appear in the upper area.
4. Modify several parameters and click “Refresh” to update the configuration. Click “Test Print” to test
the printer
Workstation
In the list on the top of the interface,
patients in the past three days are
displayed by default. You can adjust the
patients displayed in the list by changing
the start date and end date.
Change start date: click the start date
text box first, and then select a date in the
calendar below to complete the change.
Change the end date, the same as the
start date.

Click the server drop-down box and select the transport


target server.
Click "Testing" to check whether the service is normal.
If it is normal, select the corresponding patient and click
"Transmission" to transfer the file to PACS server.
Thank you !